@11:19:21.50PST09-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @11:19:21.53 >BREAK< ESP3 Gateway v1.11 for cartridge bus @11:19:26.24 Configuration for koa in flowerpot -- 10/21/20 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @11:19:26.64 Cartridge::Sampler::Warning in MAIN -- -35psi is out of range @11:19:31.60 Polling Can every 2:00 @11:19:31.74 Gate.power :analytic1,:ON,limit to 5000mA @11:19:31.75 Gate.power :analytic2,:ON,limit to 5000mA @11:19:31.76 Gate.power :analytic3,:ON,limit to 5000mA @11:19:32.07 Archive.configure :default @11:19:32.18 Archive.hold @11:19:32.25 Archive.configure :default @11:19:32.43 Air.configure :default @11:19:32.48 Air.hold @11:19:32.57 Air.configure :default @11:19:32.72 Lysis1.configure :default @11:19:32.77 Lysis1.hold @11:19:32.86 Lysis1.configure :default @11:19:32.99 Lysis2.configure :default @11:19:33.04 Lysis2.hold @11:19:33.13 Lysis2.configure :default @11:19:33.29 Plunge.configure :default @11:19:33.35 Plunge.hold @11:19:33.43 Plunge.configure :default @11:19:33.58 SamplePump.configure :default @11:19:33.74 ExtraPump.configure :default @11:19:33.92 Toroid.configure :default @11:19:33.99 Toroid.hold @11:19:34.06 Toroid.configure :default @11:19:34.25 Carousel.configure :default @11:19:34.30 Carousel.hold @11:19:34.40 Carousel.configure :default @11:19:34.45 All elf controllers running firmware version 4.61 @11:19:34.56 Intake.close @11:19:34.64 Exhaust.close @11:19:34.69 Bypass.close @11:19:34.97
Loading /home/esp/esp2/bin/processAll mission @11:19:35.43 Failed to increase Sampler's MaxPSIa! @11:19:35.44 Sampling default goal volumes @11:19:35.45 Cartridge group: dry Spares: {spare: 2} @11:19:36.36 Session Begun @11:19:37.10
Carousel.seek 13 @11:19:37.33 Selecting Cartridge 45 @11:19:37.51 -> MainThread.exit @11:19:37.78 Session Ended @11:20:49.81PST09-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @11:20:49.87 >BREAK< ESP3 Gateway v1.11 for cartridge bus @11:20:53.66 Configuration for koa in flowerpot -- 10/21/20 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @11:20:54.08 Cartridge::Sampler::Warning in MAIN -- -35psi is out of range @11:20:58.97 Polling Can every 2:00 @11:20:59.09 Gate.power :analytic1,:ON,limit to 5000mA @11:20:59.11 Gate.power :analytic2,:ON,limit to 5000mA @11:20:59.13 Gate.power :analytic3,:ON,limit to 5000mA @11:20:59.44 Archive.configure :default @11:20:59.52 Archive.hold @11:20:59.60 Archive.configure :default @11:20:59.78 Air.configure :default @11:20:59.83 Air.hold @11:20:59.92 Air.configure :default @11:21:00.05 Lysis1.configure :default @11:21:00.11 Lysis1.hold @11:21:00.20 Lysis1.configure :default @11:21:00.35 Lysis2.configure :default @11:21:00.43 Lysis2.hold @11:21:00.48 Lysis2.configure :default @11:21:00.63 Plunge.configure :default @11:21:00.72 Plunge.hold @11:21:00.77 Plunge.configure :default @11:21:00.92 SamplePump.configure :default @11:21:01.11 ExtraPump.configure :default @11:21:01.29 Toroid.configure :default @11:21:01.34 Toroid.hold @11:21:01.42 Toroid.configure :default @11:21:01.60 Carousel.configure :default @11:21:01.67 Carousel.hold @11:21:01.74 Carousel.configure :default @11:21:01.82 All elf controllers running firmware version 4.61 @11:21:01.93 Intake.close @11:21:01.97 Exhaust.close @11:21:02.11 Bypass.close @11:21:02.28
Loading /home/esp/esp2/bin/processAll mission @11:21:02.62 Failed to increase Sampler's MaxPSIa! @11:21:02.65 Sampling maximum of 800.0ml per cartridge @11:21:02.66 Cartridge group: dry Spares: {spare: 2} @11:21:03.49 Carousel.seek 13 @11:21:03.77 Selecting Cartridge 45 @11:21:04.19 Carousel.seek 45 @11:21:11.10 Lysis1.seek :locked @11:21:14.35 Plunge.seek :locked @11:21:14.57 Air.seek :locked @11:21:18.85
Mounted archive Cartridge @11:21:19.24 Duration of filtering limited to 1:05:00 @11:21:19.62 Priming sample loop w/160ml, bypass w/10ml @11:21:19.63 SamplePump.setPosition! 0ml @11:21:19.69 Bypass.open @11:21:21.12 Intake.open @11:21:22.93 SamplePump.reconfigure :prime @11:21:23.04 SamplePump.seek 10ml,1:21 @11:21:26.12 SamplePump.coast @11:21:26.19 Intake.close @11:21:26.29 Bypass.close @11:21:27.21 Cartridge::Sampler::Leaky in FILTERING -- Serious Leak or Stall -- 0.93ml/s made only 0.7psi vacuum @11:21:27.24 Retry #1 of 2 @11:21:27.25 Cmd::SpareCartridge in FILTERING -- 1 spare remaining @11:21:27.29 Cartridge::None in FILTERING -- No dry cartridges remain @11:21:27.30 Cartridge::None in FILTERING -- No dry cartridges remain @11:21:27.60
Cartridge Load: {45 => { Type: :aRNAl_bac, State: :leakyToroid}, 51 => { Type: :aRNAl_bac, State: :processed, filtered: 234.66}, 60 => { Type: :archive_bac, State: :missing}, [43, 41, 39, 37, 35, 33, 31, 23, 21, 19, 17, 15, 13, 11, 9, 7, 5, 3, 1] => { Type: :aRNAl_bac, State: :reserved, was: :dry}, [44, 42, 40, 38, 36, 34, 32, 30, 22, 20, 18, 16, 14, 12, 10, 8, 6, 4, 2] => { Type: :archive_bac, State: :reserved, was: :dry}, [49, 47] => { Type: :aRNAl_bac, State: :processed, filtered: 200}, [50, 48, 46] => { Type: :archive_bac, State: :processed, filtered: 200}} Spares: {spare: 1} @11:21:27.65 Thread::ChildDied in MAIN -- :FILTERING @11:23:29.68PST09-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @11:23:29.73 >BREAK< ESP3 Gateway v1.11 for cartridge bus @11:23:33.86 Configuration for koa in flowerpot -- 10/21/20 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @11:23:34.26 Cartridge::Sampler::Warning in MAIN -- -35psi is out of range @11:23:39.26 Polling Can every 2:00 @11:23:39.38 Gate.power :analytic1,:ON,limit to 5000mA @11:23:39.39 Gate.power :analytic2,:ON,limit to 5000mA @11:23:39.44 Gate.power :analytic3,:ON,limit to 5000mA @11:23:39.73 Archive.configure :default @11:23:39.79 Archive.hold @11:23:39.90 Archive.configure :default @11:23:40.07 Air.configure :default @11:23:40.14 Air.hold @11:23:40.22 Air.configure :default @11:23:40.34 Lysis1.configure :default @11:23:40.42 Lysis1.hold @11:23:40.49 Lysis1.configure :default @11:23:40.64 Lysis2.configure :default @11:23:40.69 Lysis2.hold @11:23:40.78 Lysis2.configure :default @11:23:40.95 Plunge.configure :default @11:23:41.00 Plunge.hold @11:23:41.10 Plunge.configure :default @11:23:41.26 SamplePump.configure :default @11:23:41.42 ExtraPump.configure :default @11:23:41.62 Toroid.configure :default @11:23:41.67 Toroid.hold @11:23:41.75 Toroid.configure :default @11:23:41.93 Carousel.configure :default @11:23:42.00 Carousel.hold @11:23:42.06 Carousel.configure :default @11:23:42.13 All elf controllers running firmware version 4.61 @11:23:42.23 Intake.close @11:23:42.33 Exhaust.close @11:23:42.38 Bypass.close @11:23:42.63
Loading /home/esp/esp2/bin/processAll mission @11:23:43.16 Failed to increase Sampler's MaxPSIa! @11:23:43.18 Sampling maximum of 800.0ml per cartridge @11:23:43.26 Session Begun @11:23:43.31
Cartridge group: dry Spares: {spare: 1} @11:23:44.49 Carousel.seek 45 @11:23:44.78 -> BACsampler.bypassVac=.2..20 @11:23:45.56
Mounted archive Cartridge @11:23:45.63 Selecting Cartridge 45 @11:23:46.46 Duration of filtering limited to 1:05:00 @11:23:46.86 Priming sample loop w/160ml, bypass w/10ml @11:23:46.93 SamplePump.setPosition! 0ml @11:23:47.07 Bypass.open @11:23:48.53 Intake.open @11:23:50.36 SamplePump.reconfigure :prime @11:23:50.49 SamplePump.seek 10ml,1:21 @11:24:01.61 SamplePump.setPosition! 0ml @11:24:01.71 Exhaust.open @11:24:03.17 Bypass.close @11:24:04.90 SamplePump.reconfigure :prime @11:24:05.02 SamplePump.seek 160ml,13:51 @11:24:36.88 -> exit @11:24:36.91 Session Ended @11:26:56.46 Bypass.open @11:26:57.15 Intake.close @11:26:58.08 Toroid.reconfigure :hold @11:26:58.16 Toroid.seek :crack @11:26:59.78 Toroid.seek :cartridge @11:27:00.71 Toroid.reconfigure :default @11:27:00.81 Toroid.seek :clear @11:27:02.45 Pressurizing cartridge to 26psi @11:27:02.46 SamplePump.setPosition! 0ml @11:27:02.54 Bypass.close @11:27:03.24 Intake.open @11:27:04.15 Exhaust.close @11:27:05.02 SamplePump.reconfigure :charge @11:27:05.09 SamplePump.seek 2.3ml,42 seconds @11:27:18.79 Intake.close @11:27:19.61 Cartridge pressurized with 0.850ml to 26.1psi @11:27:19.63 delay 5 seconds @11:27:24.65 Intake.open @11:27:25.57 Cartridge retained 22.7psi (87%) 5 seconds later @11:27:25.69 Sampling 800.0ml @11:27:25.70 SamplePump.setPosition! 0ml @11:27:25.74 Exhaust.open @11:27:26.65 SamplePump.reconfigure :sample @11:27:26.73 SamplePump.seek 800ml,1:07:11 @11:27:29.56 sampling 1/800ml at 0.54ml/s with 23.6psi vacuum @11:28:29.57 sampling 33/800ml at 0.52ml/s with 27.3psi vacuum @11:29:29.56 sampling 64/800ml at 0.52ml/s with 20.7psi vacuum @11:30:29.53 sampling 95/800ml at 0.52ml/s with 22.5psi vacuum @11:31:29.54 sampling 125/800ml at 0.52ml/s with 21.4psi vacuum @11:32:29.53 sampling 155/800ml at 0.49ml/s with 25.1psi vacuum @11:33:16.45 SamplePump.coast @11:33:16.53 Intake.close @11:33:16.59 Exhaust.close @11:33:16.85 Toroid.reconfigure :hold @11:33:16.93 Toroid.seek :bypass @11:33:19.16 Toroid.reconfigure :default @11:33:19.24 Toroid.seek :clear @11:33:21.17 Cartridge::Sampler::Leaky in FILTERING -- Serious Leak or Stall -- 0.63ml/s made only 8.3psi vacuum @11:33:21.19 Retry #1 of 2 @11:33:21.22 Cmd::SpareCartridge in FILTERING -- 0 spares remaining @11:33:21.27 Cartridge::None in FILTERING -- No dry cartridges remain @11:33:21.29 Cartridge::None in FILTERING -- No dry cartridges remain @11:33:21.60
Cartridge Load: {45 => { Type: :aRNAl_bac, State: :leaky, filtered: 179.512}, 51 => { Type: :aRNAl_bac, State: :processed, filtered: 234.66}, 60 => { Type: :archive_bac, State: :missing}, [43, 41, 39, 37, 35, 33, 31, 23, 21, 19, 17, 15, 13, 11, 9, 7, 5, 3, 1] => { Type: :aRNAl_bac, State: :reserved, was: :dry}, [44, 42, 40, 38, 36, 34, 32, 30, 22, 20, 18, 16, 14, 12, 10, 8, 6, 4, 2] => { Type: :archive_bac, State: :reserved, was: :dry}, [49, 47] => { Type: :aRNAl_bac, State: :processed, filtered: 200}, [50, 48, 46] => { Type: :archive_bac, State: :processed, filtered: 200}} Spares: {spare: 0} @11:33:21.62 Thread::ChildDied in MAIN -- :FILTERING @11:57:55.94PST09-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @11:57:55.98 >BREAK< ESP3 Gateway v1.11 for cartridge bus @11:57:59.72 Configuration for koa in flowerpot -- 10/21/20 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @11:58:00.12 Cartridge::Sampler::Warning in MAIN -- -35psi is out of range @11:58:05.18 Polling Can every 2:00 @11:58:05.32 Gate.power :analytic1,:ON,limit to 5000mA @11:58:05.33 Gate.power :analytic2,:ON,limit to 5000mA @11:58:05.34 Gate.power :analytic3,:ON,limit to 5000mA @11:58:05.64 Archive.configure :default @11:58:05.74 Archive.hold @11:58:05.84 Archive.configure :default @11:58:06.00 Air.configure :default @11:58:06.05 Air.hold @11:58:06.15 Air.configure :default @11:58:06.27 Lysis1.configure :default @11:58:06.33 Lysis1.hold @11:58:06.41 Lysis1.configure :default @11:58:06.57 Lysis2.configure :default @11:58:06.62 Lysis2.hold @11:58:06.70 Lysis2.configure :default @11:58:06.86 Plunge.configure :default @11:58:06.95 Plunge.hold @11:58:07.01 Plunge.configure :default @11:58:07.16 SamplePump.configure :default @11:58:07.34 ExtraPump.configure :default @11:58:07.53 Toroid.configure :default @11:58:07.58 Toroid.hold @11:58:07.66 Toroid.configure :default @11:58:07.83 Carousel.configure :default @11:58:07.88 Carousel.hold @11:58:07.96 Carousel.configure :default @11:58:08.05 All elf controllers running firmware version 4.61 @11:58:08.13 Intake.close @11:58:08.22 Exhaust.close @11:58:08.28 Bypass.close @11:58:08.56
Loading /home/esp/esp2/bin/processAll mission @11:58:09.06 Session Begun @11:58:09.22
Failed to increase Sampler's MaxPSIa! @11:58:09.28 Sampling maximum of 800.0ml per cartridge @11:58:09.30 Cartridge group: dry Spares: {spare: 0} @11:58:10.66 Carousel.seek 29 @11:58:10.90 Selecting Cartridge 43 @11:58:11.20 Carousel.seek 43 @11:58:11.28 -> BACsampler.bypassVac=0.2..20 @11:58:14.48
Lysis1.seek :locked @11:58:17.78 Plunge.seek :locked @11:58:18.02 Air.seek :locked @11:58:22.29
Mounted archive Cartridge @11:58:22.75 Duration of filtering limited to 1:05:00 @11:58:23.15 Priming sample loop w/160ml, bypass w/10ml @11:58:23.17 SamplePump.setPosition! 0ml @11:58:23.27 Bypass.open @11:58:24.75 Intake.open @11:58:26.26 -> BACsampler.bypassVac @11:58:26.63 SamplePump.reconfigure :prime @11:58:26.76 SamplePump.seek 10ml,1:21 @11:58:29.87 SamplePump.coast @11:58:29.98 Intake.close @11:58:30.08 Bypass.close @11:58:31.08 Cartridge::Sampler::Leaky in FILTERING -- Serious Leak or Stall -- 0.93ml/s made only -0.3psi vacuum @11:58:31.10 Retry #1 of 2 @11:58:31.17 Cmd::Error in FILTERING -- No spare cartridges remain @11:58:31.21 Cmd::Error in FILTERING -- No spare cartridges remain @11:58:31.53
Cartridge Load: {[49, 47] => { Type: :aRNAl_bac, State: :processed, filtered: 200}, [50, 48, 46] => { Type: :archive_bac, State: :processed, filtered: 200}, 29 => { Type: :unknown, State: :missing}, 43 => { Type: :aRNAl_bac, State: :leakyToroid}, 45 => { Type: :aRNAl_bac, State: :leaky, filtered: 179.512}, 51 => { Type: :aRNAl_bac, State: :processed, filtered: 234.66}, 60 => { Type: :archive_bac, State: :missing}, [41, 39, 37, 35, 33, 31, 23, 21, 19, 17, 15, 13, 11, 9, 7, 5, 3, 1] => { Type: :aRNAl_bac, State: :reserved, was: :dry}, [44, 42, 40, 38, 36, 34, 32, 30, 22, 20, 18, 16, 14, 12, 10, 8, 6, 4, 2] => { Type: :archive_bac, State: :reserved, was: :dry}} Spares: {spare: 0} @11:58:31.55 Thread::ChildDied in MAIN -- :FILTERING @11:58:31.84 Session Ended @17:56:30.73PST09-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @17:56:30.78 >BREAK< ESP3 Gateway v1.11 for cartridge bus @17:56:34.53 Configuration for koa in flowerpot -- 2/9/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @17:56:34.87 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @17:56:39.79 Polling Can every 2:00 @17:56:39.92 Gate.power :analytic1,:ON,limit to 5000mA @17:56:39.94 Gate.power :analytic2,:ON,limit to 5000mA @17:56:39.96 Gate.power :analytic3,:ON,limit to 5000mA @17:56:40.26 Archive.configure :default @17:56:40.32 Archive.hold @17:56:40.43 Archive.configure :default @17:56:40.58 Air.configure :default @17:56:40.66 Air.hold @17:56:40.72 Air.configure :default @17:56:40.86 Lysis1.configure :default @17:56:40.94 Lysis1.hold @17:56:41.00 Lysis1.configure :default @17:56:41.12 Lysis2.configure :default @17:56:41.21 Lysis2.hold @17:56:41.29 Lysis2.configure :default @17:56:41.46 Plunge.configure :default @17:56:41.51 Plunge.hold @17:56:41.59 Plunge.configure :default @17:56:41.76 SamplePump.configure :default @17:56:41.92 ExtraPump.configure :default @17:56:42.10 Toroid.configure :default @17:56:42.16 Toroid.hold @17:56:42.25 Toroid.configure :default @17:56:42.42 Carousel.configure :default @17:56:42.47 Carousel.hold @17:56:42.55 Carousel.configure :default @17:56:42.71 Pressure.configure :default @17:56:42.76 All elf controllers running firmware version 4.61 @17:56:42.88 Intake.close @17:56:42.96 Exhaust.close @17:56:43.00 Bypass.close @17:56:43.29
Waiting for espclients... @17:56:44.40 Session Begun @17:56:55.96 -> MainThread.exit @17:56:56.20 Session Ended @17:57:30.14PST09-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @17:57:30.17 >BREAK< ESP3 Gateway v1.11 for cartridge bus @17:57:34.04 Configuration for koa in flowerpot -- 2/9/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @17:57:34.40 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @17:57:39.33 Polling Can every 2:00 @17:57:39.47 Gate.power :analytic1,:ON,limit to 5000mA @17:57:39.49 Gate.power :analytic2,:ON,limit to 5000mA @17:57:39.51 Gate.power :analytic3,:ON,limit to 5000mA @17:57:39.79 Archive.configure :default @17:57:39.84 Archive.hold @17:57:39.96 Archive.configure :default @17:57:40.12 Air.configure :default @17:57:40.19 Air.hold @17:57:40.25 Air.configure :default @17:57:40.43 Lysis1.configure :default @17:57:40.48 Lysis1.hold @17:57:40.56 Lysis1.configure :default @17:57:40.70 Lysis2.configure :default @17:57:40.75 Lysis2.hold @17:57:40.83 Lysis2.configure :default @17:57:41.00 Plunge.configure :default @17:57:41.05 Plunge.hold @17:57:41.13 Plunge.configure :default @17:57:41.30 SamplePump.configure :default @17:57:41.45 ExtraPump.configure :default @17:57:41.62 Toroid.configure :default @17:57:41.68 Toroid.hold @17:57:41.77 Toroid.configure :default @17:57:41.95 Carousel.configure :default @17:57:42.00 Carousel.hold @17:57:42.08 Carousel.configure :default @17:57:42.25 Pressure.configure :default @17:57:42.31 All elf controllers running firmware version 4.61 @17:57:42.41 Intake.close @17:57:42.46 Exhaust.close @17:57:42.55 Bypass.close @17:57:42.80
Waiting for espclients... @17:57:43.59 Session Begun @17:57:48.81 -> Abs @17:58:06.66 -> Intake.open @17:58:06.68 Intake.open @17:58:09.50 -> PB.open @17:58:09.54 Bypass.open @17:58:12.91 -> Abs @17:58:20.69 -> ESP.closeValves @17:58:20.83 Intake.close @17:58:20.91 Bypass.close @17:58:30.06 -> slots @17:58:45.42 -> slot[45] @17:58:59.66 -> Cartridge.dry @17:59:05.41 -> Cmd.loadCartridge @17:59:06.36 Carousel.seek 45 @17:59:07.36 Mounted archive Cartridge @17:59:07.46 Selecting Cartridge 44 @17:59:07.95 Air.seek :clear @17:59:11.84 Lysis1.seek :clear @17:59:12.06 Plunge.seek :clear @17:59:15.08 Carousel.seek 44 @17:59:15.70 Lysis1.seek :locked @17:59:19.00 Plunge.seek :locked @17:59:19.22 Air.seek :locked @17:59:23.51 Mounted archive Cartridge @18:00:34.75 -> list method :runMission @18:01:13.58 -> list method :start @18:02:23.95 -> Cmd.status @18:02:55.59 -> start(:filter44) {Cmd.filter} @18:02:56.04 Duration of filtering limited to 1:05:00 @18:02:56.46 Priming sample loop w/100ml, bypass w/10ml @18:02:56.48 SamplePump.setPosition! 0ml @18:02:56.59 Bypass.open @18:02:58.14 Intake.open @18:02:59.06 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @18:03:01.64 Vacuum tare is 18% of full scale @18:03:01.70 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @18:03:01.93 Pressure.reconfigure :default @18:03:02.05 Pressure.hold @18:03:02.33 SamplePump.reconfigure :prime @18:03:02.46 SamplePump.seek 10ml,1:21 @18:03:13.58 Pressure.coast @18:03:13.70 SamplePump.setPosition! 0ml @18:03:13.81 Exhaust.open @18:03:15.26 Bypass.close @18:03:16.90 Pressure.reconfigure :default @18:03:17.01 Pressure.hold @18:03:17.41 SamplePump.reconfigure :prime @18:03:17.53 SamplePump.seek 100ml,8:51 @18:05:04.87 Pressure.coast @18:05:04.97 Bypass.open @18:05:05.70 Intake.close @18:05:06.68 Toroid.reconfigure :hold @18:05:06.76 Toroid.seek :crack @18:05:08.42 Toroid.seek :cartridge @18:05:09.40 Toroid.reconfigure :default @18:05:09.49 Toroid.seek :clear @18:05:11.45 Pressurizing cartridge to 26psi above 14.4psia @18:05:11.47 SamplePump.setPosition! 0ml @18:05:11.58 Intake.open @18:05:12.44 Bypass.close @18:05:13.14 Exhaust.close @18:05:14.06 Pressure.reconfigure :default @18:05:14.18 Pressure.hold @18:05:14.47 SamplePump.reconfigure :charge @18:05:14.59 SamplePump.seek 2.3ml,42 seconds @18:05:27.85 Pressure.coast @18:05:27.96 Intake.close @18:05:28.88 Pressurized cartridge with 0.802ml to 26.2psi @18:05:28.89 Isolated @36.3psia (21.9psi) @18:05:28.95 Checking for cartridge leaks from 36.3psia @18:05:29.00 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @18:05:33.64 Cartridge pressure fell 1% to 36.1psia @18:05:33.78 Sampling 1000.0ml @18:05:33.80 SamplePump.setPosition! 0ml @18:05:33.92 Exhaust.open @18:05:34.69 Intake.open @18:05:35.70 Pressure.reconfigure :default @18:05:35.81 Pressure.hold @18:05:36.20 SamplePump.reconfigure :sample @18:05:36.32 SamplePump.seek 1000ml,1:23:51 @18:05:38.82 Cartridge 44 sampling 1/1000ml at 0.50ml/s with 27.6psi vacuum @18:06:38.81 Cartridge 44 sampling 32/1000ml at 0.51ml/s with 27.7psi vacuum @18:07:38.80 Cartridge 44 sampling 62/1000ml at 0.51ml/s with 27.6psi vacuum @18:08:38.82 Cartridge 44 sampling 93/1000ml at 0.52ml/s with 25.7psi vacuum @18:09:38.79 Cartridge 44 sampling 124/1000ml at 0.50ml/s with 21.8psi vacuum @18:10:38.82 Cartridge 44 sampling 154/1000ml at 0.50ml/s with 18.3psi vacuum @18:11:38.79 Cartridge 44 sampling 185/1000ml at 0.50ml/s with 20.3psi vacuum @18:12:38.79 Cartridge 44 sampling 216/1000ml at 0.49ml/s with 26.2psi vacuum @18:13:38.79 Cartridge 44 sampling 246/1000ml at 0.50ml/s with 23.5psi vacuum @18:14:38.80 Cartridge 44 sampling 277/1000ml at 0.52ml/s with 24.7psi vacuum @18:15:38.82 Cartridge 44 sampling 308/1000ml at 0.53ml/s with 24.7psi vacuum @18:16:38.81 Cartridge 44 sampling 339/1000ml at 0.52ml/s with 27.2psi vacuum @18:17:38.79 Cartridge 44 sampling 370/1000ml at 0.54ml/s with 23.9psi vacuum @18:18:38.80 Cartridge 44 sampling 401/1000ml at 0.53ml/s with 23.0psi vacuum @18:19:38.80 Cartridge 44 sampling 432/1000ml at 0.51ml/s with 12.2psi vacuum 6.8m deep (24.8psia) @18:20:38.80 Cartridge 44 sampling 464/1000ml at 0.51ml/s with 19.0psi vacuum @18:21:38.80 Cartridge 44 sampling 495/1000ml at 0.50ml/s with 27.6psi vacuum @18:22:38.80 Cartridge 44 sampling 526/1000ml at 0.54ml/s with 25.9psi vacuum @18:23:38.82 Cartridge 44 sampling 557/1000ml at 0.53ml/s with 26.9psi vacuum @18:24:38.81 Cartridge 44 sampling 589/1000ml at 0.51ml/s with 17.8psi vacuum @18:25:38.79 Cartridge 44 sampling 620/1000ml at 0.53ml/s with 21.3psi vacuum 2.1m deep (17.7psia) @18:26:38.80 Cartridge 44 sampling 651/1000ml at 0.51ml/s with 29.4psi vacuum @18:27:38.80 Cartridge 44 sampling 683/1000ml at 0.52ml/s with 28.3psi vacuum @18:28:38.81 Cartridge 44 sampling 714/1000ml at 0.53ml/s with 26.8psi vacuum @18:29:38.79 Cartridge 44 sampling 745/1000ml at 0.49ml/s with 10.7psi vacuum 2.2m deep (17.9psia) @18:30:38.81 Cartridge 44 sampling 776/1000ml at 0.52ml/s with 26.5psi vacuum @18:31:38.81 Cartridge 44 sampling 808/1000ml at 0.53ml/s with 18.1psi vacuum 3.0m deep (19.2psia) @18:32:38.82 Cartridge 44 sampling 839/1000ml at 0.54ml/s with 23.6psi vacuum @18:33:38.79 Cartridge 44 sampling 870/1000ml at 0.50ml/s with 26.7psi vacuum @18:34:38.80 Cartridge 44 sampling 901/1000ml at 0.53ml/s with 24.1psi vacuum @18:35:39.47 Cartridge 44 sampling 932/1000ml at 0.52ml/s with 17.5psi vacuum 1.7m deep (17.3psia) @18:36:38.81 Cartridge 44 sampling 963/1000ml at 0.51ml/s with 9.1psi vacuum 8.4m deep (27.1psia) @18:37:38.79 Cartridge 44 sampling 995/1000ml at 0.52ml/s with 26.4psi vacuum @18:37:49.00 Pressure.coast @18:37:49.21 Intake.close @18:37:49.34 Exhaust.close @18:37:49.60 Toroid.reconfigure :hold @18:37:49.72 Toroid.seek :bypass @18:37:51.37 Toroid.reconfigure :default @18:37:51.48 Toroid.seek :clear @18:37:53.12 Sampled 1000.0ml @18:38:15.67 -> start(:process44) {Cmd.process} @18:38:16.27 Plunge.seek :reagent @18:38:17.19 Sample loop closed. Ready to Deliver Reagents @18:38:17.21 Archive.seek :sealed @18:38:20.91 Archive.seek :evac @18:38:34.93 delay 1:00 @18:39:34.96 Archive.seek :clear @18:39:52.11 RNAlater soak for 10:00 @18:49:52.15 Archive.seek :tVent @18:49:53.84 delay 1 second @18:49:54.91 Archive.seek :evac @18:50:10.88 delay 10 seconds @18:50:20.94 Sample preserved and evacuated @18:50:44.66 -> to 45 @18:50:45.09 Archive.seek :clear @18:50:45.32 Air.seek :clear @18:51:02.43 Lysis1.seek :clear @18:51:02.68 Plunge.seek :clear @18:51:05.70 Carousel.seek 45 @18:51:06.26 Lysis1.seek :locked @18:51:09.56 Plunge.seek :locked @18:51:09.78 Air.seek :locked @18:51:14.05 Mounted archive Cartridge @18:51:24.47 -> Cartridge.mounted @18:51:41.25 -> list Cartridge.method :process @18:52:14.40 -> start(:process45) {Cartridge.process} @18:52:14.45 Cartridge::Error in process45 -- Cannot process leaky cartridge @18:52:54.27 -> Cartridge.update :filtering @18:53:02.97 -> Cartridge.mounted @18:53:08.50 -> start(:process45) {Cartridge.process} @18:53:08.67 Plunge.seek :reagent @18:53:09.63 Sample loop closed. Ready to Deliver Reagents @18:53:09.66 Archive.seek :sealed @18:53:13.37 Archive.seek :evac @18:53:27.41 delay 1:00 @18:54:27.46 Archive.seek :clear @18:54:44.61 Sample preserved. Puck Full. @18:57:27.87 -> Cartridge.sampler.method :process @18:57:33.65 -> _.list @18:57:50.03 -> exit @18:57:50.06 Session Ended @21:28:10.73 Session Begun @21:28:25.62 -> Sampler::AbsSkew.unit @21:28:51.07 -> Sampler::AbsSkew.unit.update fixedPressure: 14.7 @21:29:17.79 -> reload Sampler%:tare @21:29:30.44 -> Sampler.tare @21:29:30.47 Cartridge::Sampler::Error in brent -- Cannot tare with Intake closed @21:29:38.81 -> Abs @21:29:43.13 -> Sampler.tare! @21:29:43.15 Bypass.open @21:29:44.45 Intake.open @21:29:45.41 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @21:29:48.02 Vacuum tare is 18% of full scale @21:29:48.03 Cartridge::Sampler::Warning in brent -- Vacuum sensor saturates below -18psi @21:29:48.28 Bypass.close @21:29:48.32 Intake.close @21:29:55.82 -> Abs @21:30:02.02 -> Abs @21:30:04.29 -> Abs @21:30:08.61 -> Sampler.tare! @21:30:08.64 Bypass.open @21:30:09.88 Intake.open @21:30:10.80 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @21:30:13.37 Vacuum tare is 18% of full scale @21:30:13.43 Cartridge::Sampler::Warning in brent -- Vacuum sensor saturates below -18psi @21:30:13.67 Bypass.close @21:30:13.74 Intake.close @21:30:16.75 -> Abs @21:30:18.00 -> Abs @21:30:18.97 -> Abs @21:30:19.92 -> Abs @21:30:57.34 -> Sampler.tare! @21:30:57.36 Bypass.open @21:30:58.66 Intake.open @21:30:59.60 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @21:31:02.20 Vacuum tare is 18% of full scale @21:31:02.23 Cartridge::Sampler::Warning in brent -- Vacuum sensor saturates below -18psi @21:31:02.46 Bypass.close @21:31:02.56 Intake.close @21:31:05.33 -> Abs @21:31:08.34 -> Abs @21:31:09.32 -> Abs @21:31:10.35 -> Abs @21:31:11.66 -> Abs @21:31:12.84 -> Abs @21:31:27.55 -> Bypass.open @21:31:27.57 Bypass.open @21:31:33.37 -> Abs @21:31:36.81 -> Intake.open @21:31:36.83 Intake.open @21:31:39.15 -> Abs @21:31:40.48 -> Abs @21:31:48.07 -> ESP.closeValves @21:31:48.21 Intake.close @21:31:48.34 Bypass.close @21:31:50.59 -> Abs @21:34:54.33 -> exit @21:34:54.38 Session Ended @21:35:08.14 Session Begun @21:35:13.90 -> Cartridge.unlock @21:35:14.24 Air.seek :clear @21:35:18.10 Lysis1.seek :clear @21:35:18.33 Plunge.seek :clear @21:35:28.65 -> ESP.main.exit @21:35:28.90 Session Ended @08:42:02.04PST10-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @08:42:02.07 >BREAK< ESP3 Gateway v1.11 for cartridge bus @08:42:05.91 Configuration for koa in flowerpot -- 2/9/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @08:42:06.26 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @08:42:11.52 Polling Can every 2:00 @08:42:11.65 Gate.power :analytic1,:ON,limit to 5000mA @08:42:11.66 Gate.power :analytic2,:ON,limit to 5000mA @08:42:11.68 Gate.power :analytic3,:ON,limit to 5000mA @08:42:12.00 Archive.configure :default @08:42:12.07 Archive.hold @08:42:12.16 Archive.configure :default @08:42:12.31 Air.configure :default @08:42:12.37 Air.hold @08:42:12.46 Air.configure :default @08:42:12.59 Lysis1.configure :default @08:42:12.65 Lysis1.hold @08:42:12.73 Lysis1.configure :default @08:42:12.86 Lysis2.configure :default @08:42:12.93 Lysis2.hold @08:42:13.01 Lysis2.configure :default @08:42:13.18 Plunge.configure :default @08:42:13.26 Plunge.hold @08:42:13.32 Plunge.configure :default @08:42:13.53 SamplePump.configure :default @08:42:13.70 ExtraPump.configure :default @08:42:13.86 Toroid.configure :default @08:42:13.93 Toroid.hold @08:42:14.01 Toroid.configure :default @08:42:14.18 Carousel.configure :default @08:42:14.26 Carousel.hold @08:42:14.34 Carousel.configure :default @08:42:14.51 Pressure.configure :default @08:42:14.57 All elf controllers running firmware version 4.61 @08:42:14.68 Intake.close @08:42:14.77 Exhaust.close @08:42:14.83 Bypass.close @08:42:15.06
Loading /home/esp/esp2/bin/processAll mission @08:42:15.42 Sampling maximum of 800.0ml per cartridge @08:42:15.44 Cartridge group: dry Spares: {spare: 0} @08:42:16.26 Carousel.seek 45 @08:42:16.45 Selecting Cartridge 43 @08:42:16.73 Carousel.seek 43 @08:42:17.50 Lysis1.seek :locked @08:42:20.74 Plunge.seek :locked @08:42:20.92 Air.seek :locked @08:42:25.19
Mounted archive Cartridge @08:42:25.56 Duration of filtering limited to 1:05:00 @08:42:25.91 Priming sample loop w/100ml, bypass w/10ml @08:42:25.92 SamplePump.setPosition! 0ml @08:42:25.99 Bypass.open @08:42:27.50 Intake.open @08:42:28.43 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @08:42:31.04 Vacuum tare is 19% of full scale @08:42:31.05 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @08:42:31.33 Pressure.reconfigure :default @08:42:31.41 Pressure.hold @08:42:31.69 SamplePump.reconfigure :prime @08:42:31.75 SamplePump.seek 10ml,1:21 @08:42:42.87 Pressure.coast @08:42:42.93 SamplePump.setPosition! 0ml @08:42:42.98 Exhaust.open @08:42:44.44 Bypass.close @08:42:46.00 Pressure.reconfigure :default @08:42:46.07 Pressure.hold @08:42:46.42 SamplePump.reconfigure :prime @08:42:46.50 SamplePump.seek 100ml,8:51 @08:44:33.78 Pressure.coast @08:44:33.84 Bypass.open @08:44:34.53 Intake.close @08:44:35.48 Toroid.reconfigure :hold @08:44:35.55 Toroid.seek :crack @08:44:37.16 Toroid.seek :cartridge @08:44:38.11 Toroid.reconfigure :default @08:44:38.20 Toroid.seek :clear @08:44:40.08 Pressurizing cartridge to 26psi above 13.4psia @08:44:40.11 SamplePump.setPosition! 0ml @08:44:40.16 Intake.open @08:44:41.51 Bypass.close @08:44:42.24 Exhaust.close @08:44:43.13 Pressure.reconfigure :default @08:44:43.20 Pressure.hold @08:44:43.44 SamplePump.reconfigure :charge @08:44:43.53 SamplePump.seek 2.3ml,42 seconds @08:44:57.52 Pressure.coast @08:44:57.60 Intake.close @08:44:58.49 Pressurized cartridge with 0.899ml to 26.2psi @08:44:58.50 Isolated @35.7psia (22.2psi) @08:44:58.51 Checking for cartridge leaks from 35.7psia @08:44:58.53 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @08:45:03.12 Cartridge pressure fell 1% to 35.4psia @08:45:03.22 Sampling 800.0ml @08:45:03.23 SamplePump.setPosition! 0ml @08:45:03.30 Exhaust.open @08:45:04.01 Intake.open @08:45:05.06 Pressure.reconfigure :default @08:45:05.14 Pressure.hold @08:45:05.46 SamplePump.reconfigure :sample @08:45:05.55 SamplePump.seek 800ml,1:07:11 @08:45:08.12 Cartridge 43 sampling 1/800ml at 0.54ml/s with 22.6psi vacuum @08:46:08.11 Cartridge 43 sampling 33/800ml at 0.53ml/s with 26.2psi vacuum @08:47:08.10 Cartridge 43 sampling 65/800ml at 0.52ml/s with 25.2psi vacuum @08:48:08.10 Cartridge 43 sampling 96/800ml at 0.51ml/s with 26.9psi vacuum @08:49:08.10 Cartridge 43 sampling 127/800ml at 0.50ml/s with 26.2psi vacuum @08:50:08.12 Cartridge 43 sampling 157/800ml at 0.50ml/s with 25.3psi vacuum @08:51:08.11 Cartridge 43 sampling 187/800ml at 0.48ml/s with 26.5psi vacuum @08:52:08.14 Cartridge 43 sampling 216/800ml at 0.48ml/s with 26.1psi vacuum @08:53:08.11 Cartridge 43 sampling 245/800ml at 0.48ml/s with 24.5psi vacuum @08:54:08.12 Cartridge 43 sampling 273/800ml at 0.47ml/s with 19.7psi vacuum @08:55:08.12 Cartridge 43 sampling 301/800ml at 0.45ml/s with 15.1psi vacuum 3.8m deep (20.3psia) @08:56:08.15 Cartridge 43 sampling 329/800ml at 0.47ml/s with 22.4psi vacuum @08:57:08.11 Cartridge 43 sampling 357/800ml at 0.47ml/s with 24.1psi vacuum @08:58:08.11 Cartridge 43 sampling 384/800ml at 0.43ml/s with 27.3psi vacuum @08:59:08.11 Cartridge 43 sampling 411/800ml at 0.41ml/s with 12.2psi vacuum 4.6m deep (21.5psia) @09:00:08.10 Cartridge 43 sampling 437/800ml at 0.43ml/s with 25.9psi vacuum @09:01:08.11 Cartridge 43 sampling 463/800ml at 0.43ml/s with 28.2psi vacuum @09:02:08.10 Cartridge 43 sampling 488/800ml at 0.37ml/s with 24.6psi vacuum @09:03:08.12 Cartridge 43 sampling 513/800ml at 0.43ml/s with 21.7psi vacuum @09:04:08.14 Cartridge 43 sampling 537/800ml at 0.41ml/s with 21.9psi vacuum @09:05:08.10 Cartridge 43 sampling 561/800ml at 0.40ml/s with 25.4psi vacuum @09:06:08.11 Cartridge 43 sampling 585/800ml at 0.37ml/s with 27.7psi vacuum @09:07:08.11 Cartridge 43 sampling 608/800ml at 0.35ml/s with 14.0psi vacuum @09:08:08.11 Cartridge 43 sampling 631/800ml at 0.36ml/s with 26.5psi vacuum @09:09:08.11 Cartridge 43 sampling 653/800ml at 0.39ml/s with 22.0psi vacuum @09:10:08.12 Cartridge 43 sampling 675/800ml at 0.35ml/s with 26.4psi vacuum @09:11:08.11 Cartridge 43 sampling 697/800ml at 0.32ml/s with 27.4psi vacuum @09:12:08.11 Cartridge 43 sampling 718/800ml at 0.37ml/s with 20.7psi vacuum @09:13:08.12 Cartridge 43 sampling 739/800ml at 0.35ml/s with 27.8psi vacuum @09:14:08.14 Cartridge 43 sampling 759/800ml at 0.33ml/s with 21.5psi vacuum @09:15:08.15 Cartridge 43 sampling 779/800ml at 0.32ml/s with 21.5psi vacuum @09:16:08.11 Cartridge 43 sampling 799/800ml at 0.31ml/s with 19.7psi vacuum @09:16:12.42 Pressure.coast @09:16:12.61 Intake.close @09:16:12.68 Exhaust.close @09:16:12.92 Toroid.reconfigure :hold @09:16:13.00 Toroid.seek :bypass @09:16:14.64 Toroid.reconfigure :default @09:16:14.70 Toroid.seek :clear @09:16:16.31 Sampled 800.0ml @09:16:16.93 Plunge.seek :reagent @09:16:17.86 Sample loop closed. Ready to Deliver Reagents @09:16:17.88 Archive.seek :sealed @09:16:21.54 Archive.seek :evac @09:16:35.51 delay 1:00 @09:17:35.56 Archive.seek :clear @09:17:52.67 Sample preserved. Puck Full. @09:17:52.98
Carousel.seek 43 @09:17:53.24 Plunge.seek :locked @09:17:54.35 Air.seek :clear @09:17:54.57 Lysis1.seek :clear @09:17:58.11
Archive.coast @09:17:58.15 Air.coast @09:17:58.19 Lysis1.coast @09:17:58.26 Lysis2.coast @09:17:58.30 Plunge.coast @09:17:58.37 SamplePump.coast @09:17:58.42 Toroid.coast @09:17:58.48 Carousel.coast @09:17:58.72 Gate.power :motor,:OFF @09:17:58.74 Gate.power :core,:OFF @09:17:58.78 Gate.power :analytic1,:OFF @09:17:58.79 Gate.power :analytic2,:OFF @09:17:58.80 Gate.power :analytic3,:OFF @09:17:58.81 Gate.power :main,:OFF @09:17:58.82 Safely stopped and ready to power off @09:17:59.23 Gate.power :main,:ON,limit to 5000mA @09:17:59.25 Gate.power :analytic1,:ON,limit to 5000mA @09:17:59.26 Gate.power :analytic2,:ON,limit to 5000mA @09:17:59.28 Gate.power :analytic3,:ON,limit to 5000mA @09:17:59.83 Powering core bus @09:18:02.17 Archive.configure :default @09:18:02.34 Air.configure :default @09:18:02.47 Lysis1.configure :default @09:18:02.61 Lysis2.configure :default @09:18:02.79 Plunge.configure :default @09:18:02.98 SamplePump.configure :default @09:18:03.14 ExtraPump.configure :default @09:18:03.31 Toroid.configure :default @09:18:03.48 Carousel.configure :default @09:18:03.67 Pressure.configure :default @09:18:03.73 All elf controllers running firmware version 4.61 @09:18:03.75 Gate.power :motor,:ON,limit to 5000mA @09:18:03.98 Archive.reconfigure :default @09:18:04.38 Archive.home.jog 0 @09:18:04.65 Plunge.reconfigure :default @09:18:04.72 Lysis1.reconfigure :default @09:18:04.81 Toroid.reconfigure :default @09:18:04.85 Lysis2.reconfigure :default @09:18:04.92 Air.reconfigure :default @09:18:05.13 Plunge.home.jog 0 @09:18:05.24 Lysis1.home.jog 0 @09:18:05.29 Lysis2.home.jog 0 @09:18:05.38 Toroid.home.jog 0 @09:18:05.41 Air.home.jog 0 @09:18:06.75 Archive.reconfigure :homing @09:18:07.05 Archive.jog 154 @09:18:07.14 Toroid.reconfigure :homing @09:18:07.17 Lysis2.reconfigure :homing @09:18:07.21 Air.reconfigure :homing @09:18:07.29 Lysis1.reconfigure :homing @09:18:07.34 Plunge.reconfigure :homing @09:18:07.63 Plunge.jog 853 @09:18:07.71 Lysis2.jog 1352 @09:18:07.74 Lysis1.jog 1347 @09:18:07.78 Toroid.jog 538 @09:18:07.85 Air.jog 1364 @09:18:08.15 Archive.setRawPosition! 50 @09:18:08.21 Archive.seek :home @09:18:08.53 Archive.reconfigure :default @09:18:08.91 Plunge.setRawPosition! 50 @09:18:08.96 Plunge.seek :home @09:18:09.09 Lysis2.setRawPosition! 50 @09:18:09.12 Lysis1.setRawPosition! 50 @09:18:09.24 Lysis2.seek :home @09:18:09.32 Lysis1.seek :home @09:18:09.50 Air.setRawPosition! 50 @09:18:09.53 Plunge.reconfigure :default @09:18:09.65 Air.seek :home @09:18:09.87 Lysis2.reconfigure :default @09:18:09.90 Lysis1.reconfigure :default @09:18:10.09 Toroid.setRawPosition! 0 @09:18:10.16 Air.reconfigure :default @09:18:10.27 Toroid.seek :home @09:18:10.43 Toroid.reconfigure :default @09:18:10.68
Toroid.seek :clear @09:18:12.29 Carousel.reconfigure :default @09:18:12.35 Carousel.home.jog 0 @09:18:12.92 Carousel.seek 43 @09:18:13.15 Selecting Cartridge 42 @09:18:13.41 Carousel.seek 42 @09:18:14.02 Lysis1.seek :locked @09:18:17.29 Plunge.seek :locked @09:18:17.46 Air.seek :locked @09:18:21.75
Mounted archive Cartridge @09:18:22.13 Duration of filtering limited to 1:05:00 @09:18:22.51 Priming sample loop w/100ml, bypass w/10ml @09:18:22.53 SamplePump.setPosition! 0ml @09:18:22.58 Bypass.open @09:18:24.13 Intake.open @09:18:25.03 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @09:18:27.63 Vacuum tare is 19% of full scale @09:18:27.66 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @09:18:27.92 Pressure.reconfigure :default @09:18:27.98 Pressure.hold @09:18:28.26 SamplePump.reconfigure :prime @09:18:28.35 SamplePump.seek 10ml,1:21 @09:18:39.46 Pressure.coast @09:18:39.53 SamplePump.setPosition! 0ml @09:18:39.59 Exhaust.open @09:18:41.03 Bypass.close @09:18:42.64 Pressure.reconfigure :default @09:18:42.72 Pressure.hold @09:18:43.09 SamplePump.reconfigure :prime @09:18:43.15 SamplePump.seek 100ml,8:51 @09:20:30.53 Pressure.coast @09:20:30.62 Bypass.open @09:20:31.35 Intake.close @09:20:32.27 Toroid.reconfigure :hold @09:20:32.35 Toroid.seek :crack @09:20:33.96 Toroid.seek :cartridge @09:20:34.89 Toroid.reconfigure :default @09:20:34.98 Toroid.seek :clear @09:20:36.87 Pressurizing cartridge to 26psi above 14.3psia @09:20:36.93 SamplePump.setPosition! 0ml @09:20:36.97 Intake.open @09:20:37.80 Bypass.close @09:20:38.56 Exhaust.close @09:20:39.44 Pressure.reconfigure :default @09:20:39.49 Pressure.hold @09:20:39.75 SamplePump.reconfigure :charge @09:20:39.81 SamplePump.seek 2.3ml,42 seconds @09:20:53.53 Pressure.coast @09:20:53.58 Intake.close @09:20:54.50 Pressurized cartridge with 0.859ml to 26.1psi @09:20:54.53 Isolated @36.5psia (22.2psi) @09:20:54.54 Checking for cartridge leaks from 36.5psia @09:20:54.57 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @09:20:59.25 Sampling 800.0ml @09:20:59.26 SamplePump.setPosition! 0ml @09:20:59.33 Exhaust.open @09:21:00.01 Intake.open @09:21:01.01 Pressure.reconfigure :default @09:21:01.10 Pressure.hold @09:21:01.44 SamplePump.reconfigure :sample @09:21:01.52 SamplePump.seek 800ml,1:07:11 @09:21:04.09 Cartridge 42 sampling 1/800ml at 0.50ml/s with 21.0psi vacuum @09:22:04.20 Cartridge 42 sampling 32/800ml at 0.53ml/s with 24.3psi vacuum @09:23:04.08 Cartridge 42 sampling 63/800ml at 0.51ml/s with 27.5psi vacuum @09:24:04.09 Cartridge 42 sampling 94/800ml at 0.51ml/s with 21.0psi vacuum 1.6m deep (17.1psia) @09:25:04.10 Cartridge 42 sampling 124/800ml at 0.50ml/s with 27.2psi vacuum @09:26:04.23 Cartridge 42 sampling 154/800ml at 0.50ml/s with 24.5psi vacuum @09:27:04.08 Cartridge 42 sampling 183/800ml at 0.48ml/s with 26.0psi vacuum @09:28:04.09 Cartridge 42 sampling 212/800ml at 0.48ml/s with 15.4psi vacuum 4.3m deep (21.1psia) @09:29:04.09 Cartridge 42 sampling 241/800ml at 0.49ml/s with 26.2psi vacuum @09:30:04.09 Cartridge 42 sampling 270/800ml at 0.47ml/s with 19.3psi vacuum @09:31:04.08 Cartridge 42 sampling 299/800ml at 0.49ml/s with 24.7psi vacuum @09:32:04.76 Cartridge 42 sampling 328/800ml at 0.47ml/s with 15.4psi vacuum 5.2m deep (22.4psia) @09:33:04.10 Cartridge 42 sampling 357/800ml at 0.49ml/s with 24.9psi vacuum @09:34:04.09 Cartridge 42 sampling 385/800ml at 0.47ml/s with 26.6psi vacuum @09:35:04.09 Cartridge 42 sampling 413/800ml at 0.46ml/s with 19.4psi vacuum 2.6m deep (18.6psia) @09:36:04.12 Cartridge 42 sampling 440/800ml at 0.46ml/s with 26.4psi vacuum @09:37:04.09 Cartridge 42 sampling 467/800ml at 0.43ml/s with 29.6psi vacuum @09:38:04.09 Cartridge 42 sampling 493/800ml at 0.44ml/s with 29.3psi vacuum @09:39:04.12 Cartridge 42 sampling 519/800ml at 0.44ml/s with 27.9psi vacuum @09:40:04.09 Cartridge 42 sampling 544/800ml at 0.40ml/s with 28.1psi vacuum @09:41:04.11 Cartridge 42 sampling 570/800ml at 0.39ml/s with 29.5psi vacuum @09:42:04.09 Cartridge 42 sampling 594/800ml at 0.38ml/s with 16.1psi vacuum @09:43:04.09 Cartridge 42 sampling 618/800ml at 0.39ml/s with 29.6psi vacuum @09:44:04.11 Cartridge 42 sampling 642/800ml at 0.39ml/s with 19.2psi vacuum @09:45:04.09 Cartridge 42 sampling 665/800ml at 0.38ml/s with 29.5psi vacuum @09:46:04.12 Cartridge 42 sampling 687/800ml at 0.34ml/s with 20.7psi vacuum @09:47:04.09 Cartridge 42 sampling 709/800ml at 0.37ml/s with 27.5psi vacuum @09:48:04.12 Cartridge 42 sampling 731/800ml at 0.36ml/s with 9.7psi vacuum 4.1m deep (20.8psia) @09:49:04.09 Cartridge 42 sampling 753/800ml at 0.36ml/s with 12.3psi vacuum 3.5m deep (19.9psia) @09:50:04.09 Cartridge 42 sampling 774/800ml at 0.34ml/s with 29.7psi vacuum @09:51:04.09 Cartridge 42 sampling 795/800ml at 0.36ml/s with 23.4psi vacuum @09:51:19.10 Pressure.coast @09:51:19.25 Intake.close @09:51:19.33 Exhaust.close @09:51:19.52 Toroid.reconfigure :hold @09:51:19.63 Toroid.seek :bypass @09:51:21.26 Toroid.reconfigure :default @09:51:21.33 Toroid.seek :clear @09:51:22.96 Sampled 800.0ml @09:51:23.56 Plunge.seek :reagent @09:51:24.47 Sample loop closed. Ready to Deliver Reagents @09:51:24.49 Archive.seek :sealed @09:51:28.17 Archive.seek :evac @09:51:42.15 delay 1:00 @09:52:42.20 Archive.seek :clear @09:52:59.33 RNAlater soak for 10:00 @10:02:59.37 Archive.seek :tVent @10:03:01.05 delay 1 second @10:03:02.08 Archive.seek :evac @10:03:18.03 delay 10 seconds @10:03:28.10 Sample preserved and evacuated @10:03:28.40
Carousel.seek 42 @10:03:28.68 Plunge.seek :locked @10:03:29.68 Archive.seek :clear @10:03:29.86 Air.seek :clear @10:03:30.07 Lysis1.seek :clear @10:03:46.89
Archive.coast @10:03:46.96 Air.coast @10:03:47.00 Lysis1.coast @10:03:47.06 Lysis2.coast @10:03:47.11 Plunge.coast @10:03:47.16 SamplePump.coast @10:03:47.22 Toroid.coast @10:03:47.27 Carousel.coast @10:03:47.49 Gate.power :motor,:OFF @10:03:47.51 Gate.power :core,:OFF @10:03:47.52 Gate.power :analytic1,:OFF @10:03:47.54 Gate.power :analytic2,:OFF @10:03:47.55 Gate.power :analytic3,:OFF @10:03:47.56 Gate.power :main,:OFF @10:03:47.60 Safely stopped and ready to power off @10:03:48.01 Gate.power :main,:ON,limit to 5000mA @10:03:48.02 Gate.power :analytic1,:ON,limit to 5000mA @10:03:48.04 Gate.power :analytic2,:ON,limit to 5000mA @10:03:48.05 Gate.power :analytic3,:ON,limit to 5000mA @10:03:48.56 Powering core bus @10:03:50.95 Archive.configure :default @10:03:51.12 Air.configure :default @10:03:51.26 Lysis1.configure :default @10:03:51.41 Lysis2.configure :default @10:03:51.59 Plunge.configure :default @10:03:51.76 SamplePump.configure :default @10:03:51.93 ExtraPump.configure :default @10:03:52.10 Toroid.configure :default @10:03:52.29 Carousel.configure :default @10:03:52.46 Pressure.configure :default @10:03:52.51 All elf controllers running firmware version 4.61 @10:03:52.55 Gate.power :motor,:ON,limit to 5000mA @10:03:52.97 Archive.reconfigure :default @10:03:53.31 Archive.home.jog 0 @10:03:53.58 Plunge.reconfigure :default @10:03:53.62 Lysis2.reconfigure :default @10:03:53.71 Air.reconfigure :default @10:03:53.75 Lysis1.reconfigure :default @10:03:53.83 Toroid.reconfigure :default @10:03:54.04 Plunge.home.jog 0 @10:03:54.11 Lysis2.home.jog 0 @10:03:54.15 Air.home.jog 0 @10:03:54.18 Lysis1.home.jog 0 @10:03:54.27 Toroid.home.jog 0 @10:03:54.81 Archive.reconfigure :homing @10:03:55.12 Archive.jog 1372 @10:03:55.20 Plunge.reconfigure :homing @10:03:55.24 Lysis2.reconfigure :homing @10:03:55.27 Air.reconfigure :homing @10:03:55.34 Toroid.reconfigure :homing @10:03:55.37 Lysis1.reconfigure :homing @10:03:55.66 Plunge.jog 852 @10:03:55.73 Lysis2.jog 1352 @10:03:55.79 Toroid.jog 539 @10:03:55.83 Lysis1.jog 1353 @10:03:55.92 Air.jog 1358 @10:03:56.59 Archive.setRawPosition! 50 @10:03:56.67 Archive.seek :home @10:03:56.77 Plunge.setRawPosition! 50 @10:03:56.84 Plunge.seek :home @10:03:56.96 Archive.reconfigure :default @10:03:57.14 Plunge.reconfigure :default @10:03:57.27 Lysis2.setRawPosition! 50 @10:03:57.33 Lysis1.setRawPosition! 50 @10:03:57.35 Air.setRawPosition! 50 @10:03:57.48 Lysis2.seek :home @10:03:57.53 Lysis1.seek :home @10:03:57.63 Air.seek :home @10:03:57.91 Toroid.setRawPosition! 0 @10:03:57.95 Lysis2.reconfigure :default @10:03:58.07 Toroid.seek :home @10:03:58.25 Lysis1.reconfigure :default @10:03:58.32 Air.reconfigure :default @10:03:58.35 Toroid.reconfigure :default @10:03:58.93
Toroid.seek :clear @10:04:00.55 Carousel.reconfigure :default @10:04:00.63 Carousel.home.jog 0 @10:04:01.15 Carousel.seek 42 @10:04:01.32 Selecting Cartridge 41 @10:04:02.37 Carousel.seek 41 @10:04:02.90 Lysis1.seek :locked @10:04:06.15 Plunge.seek :locked @10:04:06.37 Air.seek :locked @10:04:10.63
Mounted archive Cartridge @10:04:11.01 Duration of filtering limited to 1:05:00 @10:04:11.37 Priming sample loop w/100ml, bypass w/10ml @10:04:11.40 SamplePump.setPosition! 0ml @10:04:11.45 Bypass.open @10:04:12.97 Intake.open @10:04:13.90 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @10:04:16.52 Vacuum tare is 19% of full scale @10:04:16.56 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @10:04:16.80 Pressure.reconfigure :default @10:04:16.88 Pressure.hold @10:04:17.15 SamplePump.reconfigure :prime @10:04:17.22 SamplePump.seek 10ml,1:21 @10:04:28.35 Pressure.coast @10:04:28.43 SamplePump.setPosition! 0ml @10:04:28.47 Exhaust.open @10:04:29.91 Bypass.close @10:04:31.52 Pressure.reconfigure :default @10:04:31.61 Pressure.hold @10:04:31.96 SamplePump.reconfigure :prime @10:04:32.02 SamplePump.seek 100ml,8:51 @10:06:19.45 Pressure.coast @10:06:19.51 Bypass.open @10:06:20.21 Intake.close @10:06:21.15 Toroid.reconfigure :hold @10:06:21.20 Toroid.seek :crack @10:06:22.84 Toroid.seek :cartridge @10:06:23.79 Toroid.reconfigure :default @10:06:23.87 Toroid.seek :clear @10:06:25.75 Pressurizing cartridge to 26psi above 13.9psia @10:06:25.77 SamplePump.setPosition! 0ml @10:06:25.83 Intake.open @10:06:26.64 Bypass.close @10:06:27.38 Exhaust.close @10:06:28.23 Pressure.reconfigure :default @10:06:28.28 Pressure.hold @10:06:28.53 SamplePump.reconfigure :charge @10:06:28.60 SamplePump.seek 2.3ml,42 seconds @10:06:42.69 Pressure.coast @10:06:42.77 Intake.close @10:06:43.71 Pressurized cartridge with 0.909ml to 26.0psi @10:06:43.74 Isolated @35.5psia (21.6psi) @10:06:43.75 Checking for cartridge leaks from 35.5psia @10:06:43.77 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @10:06:48.45 Sampling 800.0ml @10:06:48.47 SamplePump.setPosition! 0ml @10:06:48.53 Exhaust.open @10:06:49.20 Intake.open @10:06:50.22 Pressure.reconfigure :default @10:06:50.31 Pressure.hold @10:06:50.68 SamplePump.reconfigure :sample @10:06:50.75 SamplePump.seek 800ml,1:07:11 @10:06:53.33 Cartridge 41 sampling 1/800ml at 0.49ml/s with 21.3psi vacuum @10:07:53.29 Cartridge 41 sampling 30/800ml at 0.48ml/s with 25.4psi vacuum @10:08:53.41 Cartridge 41 sampling 60/800ml at 0.47ml/s with 26.6psi vacuum @10:09:53.29 Cartridge 41 sampling 88/800ml at 0.46ml/s with 19.6psi vacuum @10:10:53.28 Cartridge 41 sampling 116/800ml at 0.47ml/s with 25.4psi vacuum @10:11:53.29 Cartridge 41 sampling 144/800ml at 0.45ml/s with 22.3psi vacuum @10:12:53.29 Cartridge 41 sampling 171/800ml at 0.46ml/s with 26.6psi vacuum @10:13:53.29 Cartridge 41 sampling 198/800ml at 0.42ml/s with 23.0psi vacuum @10:14:53.28 Cartridge 41 sampling 225/800ml at 0.44ml/s with 21.2psi vacuum @10:15:53.28 Cartridge 41 sampling 251/800ml at 0.43ml/s with 27.2psi vacuum @10:16:53.28 Cartridge 41 sampling 277/800ml at 0.43ml/s with 25.3psi vacuum @10:17:53.28 Cartridge 41 sampling 302/800ml at 0.41ml/s with 26.7psi vacuum @10:18:53.28 Cartridge 41 sampling 327/800ml at 0.42ml/s with 22.8psi vacuum @10:19:53.27 Cartridge 41 sampling 352/800ml at 0.40ml/s with 13.6psi vacuum 3.1m deep (19.3psia) @10:20:53.28 Cartridge 41 sampling 377/800ml at 0.43ml/s with 23.2psi vacuum @10:21:53.27 Cartridge 41 sampling 402/800ml at 0.39ml/s with 27.2psi vacuum @10:22:53.28 Cartridge 41 sampling 426/800ml at 0.39ml/s with 15.0psi vacuum 2.9m deep (19.0psia) @10:23:53.28 Cartridge 41 sampling 450/800ml at 0.39ml/s with 21.8psi vacuum @10:24:53.28 Cartridge 41 sampling 474/800ml at 0.39ml/s with 16.8psi vacuum @10:25:53.28 Cartridge 41 sampling 497/800ml at 0.40ml/s with 24.3psi vacuum @10:26:53.28 Cartridge 41 sampling 519/800ml at 0.37ml/s with 13.1psi vacuum 3.2m deep (19.5psia) @10:27:53.28 Cartridge 41 sampling 542/800ml at 0.36ml/s with 27.4psi vacuum @10:28:53.27 Cartridge 41 sampling 564/800ml at 0.34ml/s with 27.1psi vacuum @10:29:53.28 Cartridge 41 sampling 586/800ml at 0.35ml/s with 25.2psi vacuum @10:30:53.28 Cartridge 41 sampling 607/800ml at 0.34ml/s with 29.0psi vacuum @10:31:53.27 Cartridge 41 sampling 628/800ml at 0.36ml/s with 16.8psi vacuum @10:32:53.28 Cartridge 41 sampling 649/800ml at 0.35ml/s with 17.3psi vacuum 2.5m deep (18.4psia) @10:33:53.28 Cartridge 41 sampling 669/800ml at 0.33ml/s with 20.6psi vacuum @10:34:53.29 Cartridge 41 sampling 689/800ml at 0.36ml/s with 21.4psi vacuum @10:35:53.29 Cartridge 41 sampling 709/800ml at 0.36ml/s with 25.7psi vacuum @10:36:53.28 Cartridge 41 sampling 728/800ml at 0.32ml/s with 29.4psi vacuum @10:37:53.28 Cartridge 41 sampling 747/800ml at 0.31ml/s with 25.8psi vacuum @10:38:53.28 Cartridge 41 sampling 766/800ml at 0.30ml/s with 26.7psi vacuum @10:39:53.29 Cartridge 41 sampling 784/800ml at 0.29ml/s with 28.7psi vacuum @10:40:45.89 Pressure.coast @10:40:46.05 Intake.close @10:40:46.15 Exhaust.close @10:40:47.02 Toroid.reconfigure :hold @10:40:47.14 Toroid.seek :bypass @10:40:48.83 Toroid.reconfigure :default @10:40:48.89 Toroid.seek :clear @10:40:50.57 Sampled 800.0ml @10:40:51.17 Plunge.seek :reagent @10:40:52.07 Sample loop closed. Ready to Deliver Reagents @10:40:52.09 Archive.seek :sealed @10:40:55.75 Archive.seek :evac @10:41:09.74 delay 1:00 @10:42:09.78 Archive.seek :clear @10:42:26.90 Sample preserved. Puck Full. @10:42:27.21
Carousel.seek 41 @10:42:27.58 Plunge.seek :locked @10:42:28.66 Air.seek :clear @10:42:28.90 Lysis1.seek :clear @10:42:32.35
Archive.coast @10:42:32.39 Air.coast @10:42:32.46 Lysis1.coast @10:42:32.50 Lysis2.coast @10:42:32.57 Plunge.coast @10:42:32.62 SamplePump.coast @10:42:32.68 Toroid.coast @10:42:32.74 Carousel.coast @10:42:32.97 Gate.power :motor,:OFF @10:42:32.99 Gate.power :core,:OFF @10:42:33.01 Gate.power :analytic1,:OFF @10:42:33.02 Gate.power :analytic2,:OFF @10:42:33.03 Gate.power :analytic3,:OFF @10:42:33.05 Gate.power :main,:OFF @10:42:33.06 Safely stopped and ready to power off @10:42:33.43 Gate.power :main,:ON,limit to 5000mA @10:42:33.44 Gate.power :analytic1,:ON,limit to 5000mA @10:42:33.46 Gate.power :analytic2,:ON,limit to 5000mA @10:42:33.47 Gate.power :analytic3,:ON,limit to 5000mA @10:42:34.05 Powering core bus @10:42:36.40 Archive.configure :default @10:42:36.62 Air.configure :default @10:42:36.76 Lysis1.configure :default @10:42:36.90 Lysis2.configure :default @10:42:37.07 Plunge.configure :default @10:42:37.26 SamplePump.configure :default @10:42:37.45 ExtraPump.configure :default @10:42:37.62 Toroid.configure :default @10:42:37.79 Carousel.configure :default @10:42:38.00 Pressure.configure :default @10:42:38.06 All elf controllers running firmware version 4.61 @10:42:38.10 Gate.power :motor,:ON,limit to 5000mA @10:42:38.47 Archive.reconfigure :default @10:42:38.71 Archive.home.jog 0 @10:42:38.89 Plunge.reconfigure :default @10:42:39.01 Lysis2.reconfigure :default @10:42:39.07 Air.reconfigure :default @10:42:39.17 Toroid.reconfigure :default @10:42:39.22 Lysis1.reconfigure :default @10:42:39.48 Lysis2.home.jog 0 @10:42:39.54 Air.home.jog 0 @10:42:39.58 Toroid.home.jog 0 @10:42:39.60 Plunge.home.jog 0 @10:42:39.62 Lysis1.home.jog 0 @10:42:40.28 Archive.reconfigure :homing @10:42:40.53 Archive.jog 1369 @10:42:40.63 Plunge.reconfigure :homing @10:42:40.65 Lysis1.reconfigure :homing @10:42:40.68 Air.reconfigure :homing @10:42:40.74 Toroid.reconfigure :homing @10:42:40.82 Lysis2.reconfigure :homing @10:42:41.09 Plunge.jog 849 @10:42:41.16 Lysis1.jog 1352 @10:42:41.19 Toroid.jog 533 @10:42:41.22 Lysis2.jog 1355 @10:42:41.30 Air.jog 1362 @10:42:42.01 Archive.setRawPosition! 50 @10:42:42.05 Archive.seek :home @10:42:42.33 Plunge.setRawPosition! 50 @10:42:42.40 Plunge.seek :home @10:42:42.51 Archive.reconfigure :default @10:42:42.70 Plunge.reconfigure :default @10:42:42.81 Lysis1.setRawPosition! 50 @10:42:42.92 Lysis1.seek :home @10:42:43.02 Lysis2.setRawPosition! 50 @10:42:43.06 Air.setRawPosition! 50 @10:42:43.13 Lysis2.seek :home @10:42:43.21 Air.seek :home @10:42:43.39 Lysis1.reconfigure :default @10:42:43.62 Toroid.setRawPosition! 0 @10:42:43.64 Lysis2.reconfigure :default @10:42:43.69 Air.reconfigure :default @10:42:43.89 Toroid.seek :home @10:42:44.03 Toroid.reconfigure :default @10:42:44.24
Toroid.seek :clear @10:42:44.44 Slide::Error in MAIN -- Toroid Valve (TV) positionErr at bypassed - 27 counts (sensed bypassed + 58 counts) @10:42:44.45 Retry #1 of 2 @10:42:44.46 Archive.coast @10:42:44.55 Air.coast @10:42:44.59 Lysis1.coast @10:42:44.66 Lysis2.coast @10:42:44.71 Plunge.coast @10:42:44.78 Toroid.coast @10:42:45.05 Gate.power :motor,:OFF @10:42:45.06 Gate.power :core,:OFF @10:42:45.08 Gate.power :analytic1,:OFF @10:42:45.09 Gate.power :analytic2,:OFF @10:42:45.10 Gate.power :analytic3,:OFF @10:42:45.76 Gate.power :main,:OFF @10:42:45.94 Gate.power :main,:ON,limit to 5000mA @10:42:45.97 Gate.power :analytic1,:ON,limit to 5000mA @10:42:45.99 Gate.power :analytic2,:ON,limit to 5000mA @10:42:46.00 Gate.power :analytic3,:ON,limit to 5000mA @10:42:46.13 Powering core bus @10:42:48.46 Archive.configure :default @10:42:48.65 Air.configure :default @10:42:48.82 Lysis1.configure :default @10:42:48.96 Lysis2.configure :default @10:42:49.14 Plunge.configure :default @10:42:49.30 SamplePump.configure :default @10:42:49.47 ExtraPump.configure :default @10:42:49.67 Toroid.configure :default @10:42:49.83 Carousel.configure :default @10:42:50.01 Pressure.configure :default @10:42:50.08 All elf controllers running firmware version 4.61 @10:42:50.09 Gate.power :motor,:ON,limit to 5000mA @10:42:50.42 Archive.reconfigure :default @10:42:50.91 Archive.home.jog 0 @10:42:50.96 Toroid.reconfigure :default @10:42:51.00 Plunge.reconfigure :default @10:42:51.04 Lysis2.reconfigure :default @10:42:51.12 Air.reconfigure :default @10:42:51.18 Lysis1.reconfigure :default @10:42:51.50 Toroid.home.jog 0 @10:42:51.59 Lysis2.home.jog 0 @10:42:51.62 Plunge.home.jog 0 @10:42:51.66 Lysis1.home.jog 0 @10:42:51.77 Air.home.jog 0 @10:42:52.68 Plunge.reconfigure :homing @10:42:52.83 Toroid.reconfigure :homing @10:42:52.87 Lysis2.reconfigure :homing @10:42:52.94 Archive.reconfigure :homing @10:42:52.99 Air.reconfigure :homing @10:42:53.01 Lysis1.reconfigure :homing @10:42:53.28 Plunge.jog 313 @10:42:53.38 Toroid.jog 177 @10:42:53.42 Lysis2.jog 1354 @10:42:53.44 Archive.jog 1351 @10:42:53.48 Air.jog 1358 @10:42:53.56 Lysis1.jog 1351 @10:42:54.09 Plunge.setRawPosition! 50 @10:42:54.17 Plunge.seek :home @10:42:54.27 Toroid.setRawPosition! 0 @10:42:54.35 Toroid.seek :home @10:42:54.49 Toroid.reconfigure :default @10:42:54.53 Plunge.reconfigure :default @10:42:54.75 Slide::Error in MAIN -- Toroid Valve (TV) positionErr at bypassed (sensed bypassed + 94 counts) @10:42:54.78
Retry #2 of 2 @10:42:54.81 Archive.coast @10:42:54.88 Air.coast @10:42:54.95 Lysis1.coast @10:42:55.01 Lysis2.coast @10:42:55.06 Plunge.coast @10:42:55.16 Toroid.coast @10:42:55.36 Lysis2.setRawPosition! 50 @10:42:55.45 Lysis1.setRawPosition! 50 @10:42:55.47 Air.setRawPosition! 50 @10:42:55.49 Archive.setRawPosition! 50 @10:42:55.62 Lysis2.seek :home @10:42:55.87 Lysis1.seek :home @10:42:55.90 Air.seek :home @10:42:55.97 Archive.seek :home @10:42:56.22 Lysis1.reconfigure :default @10:42:56.30 Lysis2.reconfigure :default @10:42:56.46
Gate.power :motor,:OFF @10:42:56.49 Gate.power :core,:OFF @10:42:56.50 Gate.power :analytic1,:OFF @10:42:56.51 Gate.power :analytic2,:OFF @10:42:56.54 Gate.power :analytic3,:OFF @10:42:56.56 Gate.power :main,:OFF @10:42:56.74 Gate.power :main,:ON,limit to 5000mA @10:42:56.76 Gate.power :analytic1,:ON,limit to 5000mA @10:42:56.77 Gate.power :analytic2,:ON,limit to 5000mA @10:42:56.78 Gate.power :analytic3,:ON,limit to 5000mA @10:42:56.86 Archive.reconfigure :default @10:42:56.89 Air.reconfigure :default @10:42:56.99 BusOff in Archive_home -- core bus was not powered during Archive Syringe (ArS) reconfigure @10:42:57.02 BusOff in Air_home -- core bus was not powered during Air Syringe (AirS) reconfigure @10:42:57.55
Powering core bus @10:42:59.91 Archive.configure :default @10:43:00.10 Air.configure :default @10:43:00.25 Lysis1.configure :default @10:43:00.38 Lysis2.configure :default @10:43:00.54 Plunge.configure :default @10:43:00.74 SamplePump.configure :default @10:43:00.90 ExtraPump.configure :default @10:43:01.07 Toroid.configure :default @10:43:01.28 Carousel.configure :default @10:43:01.44 Pressure.configure :default @10:43:01.50 All elf controllers running firmware version 4.61 @10:43:01.54 Gate.power :motor,:ON,limit to 5000mA @10:43:01.96 Archive.reconfigure :default @10:43:02.30 Archive.home.jog 0 @10:43:02.43 Plunge.reconfigure :default @10:43:02.48 Air.reconfigure :default @10:43:02.51 Toroid.reconfigure :default @10:43:02.55 Lysis1.reconfigure :default @10:43:02.66 Lysis2.reconfigure :default @10:43:02.94 Plunge.home.jog 0 @10:43:02.97 Air.home.jog 0 @10:43:02.99 Lysis1.home.jog 0 @10:43:03.04 Toroid.home.jog 0 @10:43:03.07 Lysis2.home.jog 0 @10:43:04.28 Toroid.reconfigure :homing @10:43:04.34 Plunge.reconfigure :homing @10:43:04.88 Lysis2.reconfigure :homing @10:43:05.04 Air.reconfigure :homing @10:43:05.12 Lysis1.reconfigure :homing @10:43:05.15 Archive.reconfigure :homing @10:43:05.44 Toroid.jog 147 @10:43:05.49 Plunge.jog 313 @10:43:05.53 Lysis2.jog 1354 @10:43:05.57 Air.jog 1358 @10:43:05.61 Lysis1.jog 1352 @10:43:05.67 Archive.jog 1351 @10:43:06.30 Plunge.setRawPosition! 50 @10:43:06.37 Plunge.seek :home @10:43:06.48 Toroid.setRawPosition! 0 @10:43:06.57 Toroid.seek :home @10:43:06.71 Toroid.reconfigure :default @10:43:06.74 Plunge.reconfigure :default @10:43:07.09 Lysis2.setRawPosition! 50 @10:43:07.16 Air.setRawPosition! 50 @10:43:07.23 Lysis1.setRawPosition! 50 @10:43:07.42 Lysis2.seek :home @10:43:07.65 Air.seek :home @10:43:07.76 Lysis1.seek :home @10:43:08.17
Cartridge Load: {39 => { Type: :aRNAl_bac, State: :dry}, 42 => { Type: :archive_bac, State: :processed, filtered: 800.0}, 44 => { Type: :archive_bac, State: :processed, filtered: 1000}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 179.512}, 51 => { Type: :aRNAl_bac, State: :processed, filtered: 234.66}, [36, 34, 32, 30, 22, 20, 18, 16, 14, 12, 10, 8, 6, 4, 2] => { Type: :archive_bac, State: :reserved, was: :dry}, [37, 35, 33, 31, 23, 21, 19, 17, 15, 13, 11, 9, 7, 5, 3, 1] => { Type: :aRNAl_bac, State: :reserved, was: :dry}, [40, 38] => { Type: :archive_bac, State: :dry}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 800.0}, [49, 47] => { Type: :aRNAl_bac, State: :processed, filtered: 200}, [50, 48, 46] => { Type: :archive_bac, State: :processed, filtered: 200}, [60, 29] => { Type: :archive_bac, State: :missing}} Spares: {spare: 0} @10:43:08.24 Slide::Error in MAIN -- Toroid Valve (TV) positionErr at bypassed (sensed bypassed + 101 counts) @10:43:08.29 Archive.stop @10:43:08.34 Air.stop @10:43:08.43 Lysis1.stop @10:43:08.48 Lysis2.stop @08:17:46.32PST12-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @08:17:46.37 >BREAK< ESP3 Gateway v1.11 for cartridge bus @08:17:50.11 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @08:17:50.47 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @08:17:55.52 Polling Can every 2:00 @08:17:55.65 Gate.power :analytic1,:ON,limit to 5000mA @08:17:55.66 Gate.power :analytic2,:ON,limit to 5000mA @08:17:55.68 Gate.power :analytic3,:ON,limit to 5000mA @08:17:55.96 Archive.configure :default @08:17:56.06 Archive.hold @08:17:56.12 Archive.configure :default @08:17:56.30 Air.configure :default @08:17:56.36 Air.hold @08:17:56.45 Air.configure :default @08:17:56.59 Lysis1.configure :default @08:17:56.64 Lysis1.hold @08:17:56.72 Lysis1.configure :default @08:17:56.86 Lysis2.configure :default @08:17:56.91 Lysis2.hold @08:17:57.01 Lysis2.configure :default @08:17:57.17 Plunge.configure :default @08:17:57.24 Plunge.hold @08:17:57.30 Plunge.configure :default @08:17:57.45 SamplePump.configure :default @08:17:57.65 ExtraPump.configure :default @08:17:57.81 Toroid.configure :default @08:17:57.89 Toroid.hold @08:17:57.95 Toroid.configure :default @08:17:58.13 Carousel.configure :default @08:17:58.20 Carousel.hold @08:17:58.28 Carousel.configure :default @08:17:58.45 Pressure.configure :default @08:17:58.50 All elf controllers running firmware version 4.61 @08:17:58.61 Intake.close @08:17:58.72 Exhaust.close @08:17:58.76 Bypass.close @08:17:59.01
Loading /home/esp/esp2/bin/processAll mission @08:17:59.33 Sampling default goal volumes @08:17:59.34 Cartridge group: dry Spares: {spare: 0} @08:18:00.04 Session Begun @08:18:00.72
Carousel.seek 10 @08:18:01.96 Mounted archive Cartridge @08:18:02.04 Selecting Cartridge 1 @08:18:02.64 Air.seek :clear @08:18:06.51 Lysis1.seek :clear @08:18:06.75 Plunge.seek :clear @08:18:09.77
Carousel.seek 1 @08:18:12.01 Lysis1.seek :locked @08:18:15.29 Plunge.seek :locked @08:18:15.57 Air.seek :locked @08:18:19.85
Mounted archive Cartridge @08:18:20.26 Duration of filtering limited to 1:05:00 @08:18:20.67 Priming sample loop w/100ml, bypass w/10ml @08:18:20.69 SamplePump.setPosition! 0ml @08:18:20.78 Bypass.open @08:18:22.31 Intake.open @08:18:23.28 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @08:18:25.88 Vacuum tare is 19% of full scale @08:18:25.92 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @08:18:26.24 Pressure.reconfigure :default @08:18:26.36 Pressure.hold @08:18:26.65 SamplePump.reconfigure :prime @08:18:26.77 SamplePump.seek 10ml,1:21 @08:18:37.93 Pressure.coast @08:18:38.02 SamplePump.setPosition! 0ml @08:18:38.12 Exhaust.open @08:18:39.58 Bypass.close @08:18:41.79 Pressure.reconfigure :default @08:18:41.88 Pressure.hold @08:18:42.25 SamplePump.reconfigure :prime @08:18:42.38 SamplePump.seek 100ml,8:51 @08:20:29.63 Pressure.coast @08:20:29.74 Bypass.open @08:20:30.48 Intake.close @08:20:31.42 Toroid.reconfigure :hold @08:20:31.55 Toroid.seek :crack @08:20:33.22 Toroid.seek :cartridge @08:20:34.21 Toroid.reconfigure :default @08:20:34.34 Toroid.seek :clear @08:20:36.31 Pressurizing cartridge to 26psi above 14.7psia @08:20:36.34 SamplePump.setPosition! 0ml @08:20:36.42 Intake.open @08:20:37.31 Bypass.close @08:20:38.09 Exhaust.close @08:20:39.05 Pressure.reconfigure :default @08:20:39.14 Pressure.hold @08:20:39.47 SamplePump.reconfigure :charge @08:20:39.58 SamplePump.seek 2.3ml,42 seconds @08:20:53.60 Pressure.coast @08:20:53.68 Intake.close @08:20:54.63 Pressurized cartridge with 0.894ml to 26.2psi @08:20:54.64 Isolated @37.3psia (22.6psi) @08:20:54.66 Checking for cartridge leaks from 37.3psia @08:20:54.74 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @08:20:59.43 Sampling 1000.0ml @08:20:59.45 SamplePump.setPosition! 0ml @08:20:59.56 Exhaust.open @08:21:00.31 Intake.open @08:21:01.33 Pressure.reconfigure :default @08:21:01.44 Pressure.hold @08:21:01.80 SamplePump.reconfigure :sample @08:21:01.92 SamplePump.seek 1000ml,1:23:51 @08:21:04.39 Cartridge 1 sampling 1/1000ml at 0.63ml/s with 25.4psi vacuum @08:22:04.39 Cartridge 1 sampling 32/1000ml at 0.51ml/s with 24.3psi vacuum @08:23:04.43 Cartridge 1 sampling 62/1000ml at 0.51ml/s with 23.5psi vacuum @08:24:04.41 Cartridge 1 sampling 93/1000ml at 0.51ml/s with 23.0psi vacuum @08:25:04.39 Cartridge 1 sampling 123/1000ml at 0.50ml/s with 24.4psi vacuum @08:26:04.39 Cartridge 1 sampling 153/1000ml at 0.50ml/s with 26.4psi vacuum @08:27:04.43 Cartridge 1 sampling 184/1000ml at 0.50ml/s with 26.0psi vacuum @08:28:04.39 Cartridge 1 sampling 214/1000ml at 0.50ml/s with 22.5psi vacuum @08:29:04.39 Cartridge 1 sampling 244/1000ml at 0.49ml/s with 23.5psi vacuum @08:30:04.40 Cartridge 1 sampling 274/1000ml at 0.50ml/s with 20.6psi vacuum @08:31:04.41 Cartridge 1 sampling 304/1000ml at 0.51ml/s with 22.6psi vacuum @08:32:04.41 Cartridge 1 sampling 334/1000ml at 0.51ml/s with 24.3psi vacuum @08:33:04.39 Cartridge 1 sampling 365/1000ml at 0.50ml/s with 25.1psi vacuum @08:34:04.40 Cartridge 1 sampling 395/1000ml at 0.50ml/s with 25.5psi vacuum @08:35:04.40 Cartridge 1 sampling 425/1000ml at 0.49ml/s with 20.8psi vacuum @08:36:04.40 Cartridge 1 sampling 455/1000ml at 0.50ml/s with 25.9psi vacuum @08:37:04.39 Cartridge 1 sampling 485/1000ml at 0.48ml/s with 22.5psi vacuum @08:38:04.38 Cartridge 1 sampling 515/1000ml at 0.49ml/s with 19.6psi vacuum @08:39:04.41 Cartridge 1 sampling 546/1000ml at 0.48ml/s with 26.1psi vacuum @08:40:04.41 Cartridge 1 sampling 576/1000ml at 0.49ml/s with 21.1psi vacuum @08:41:04.40 Cartridge 1 sampling 606/1000ml at 0.50ml/s with 22.6psi vacuum @08:42:04.40 Cartridge 1 sampling 636/1000ml at 0.49ml/s with 27.0psi vacuum @08:43:04.41 Cartridge 1 sampling 666/1000ml at 0.49ml/s with 25.9psi vacuum @08:44:04.38 Cartridge 1 sampling 696/1000ml at 0.48ml/s with 20.0psi vacuum @08:45:04.39 Cartridge 1 sampling 726/1000ml at 0.49ml/s with 25.0psi vacuum @08:46:04.38 Cartridge 1 sampling 756/1000ml at 0.50ml/s with 20.3psi vacuum @08:47:04.41 Cartridge 1 sampling 786/1000ml at 0.49ml/s with 24.8psi vacuum @08:48:04.39 Cartridge 1 sampling 816/1000ml at 0.50ml/s with 21.4psi vacuum @08:49:04.40 Cartridge 1 sampling 845/1000ml at 0.49ml/s with 25.3psi vacuum @08:50:04.39 Cartridge 1 sampling 875/1000ml at 0.49ml/s with 27.1psi vacuum @08:51:04.39 Cartridge 1 sampling 905/1000ml at 0.51ml/s with 25.5psi vacuum @08:52:04.39 Cartridge 1 sampling 935/1000ml at 0.49ml/s with 25.2psi vacuum @08:53:04.41 Cartridge 1 sampling 965/1000ml at 0.48ml/s with 26.9psi vacuum @08:54:04.38 Cartridge 1 sampling 995/1000ml at 0.51ml/s with 21.5psi vacuum @08:54:14.34 Pressure.coast @08:54:14.53 Intake.close @08:54:14.62 Exhaust.close @08:54:14.89 Toroid.reconfigure :hold @08:54:15.00 Toroid.seek :bypass @08:54:16.65 Toroid.reconfigure :default @08:54:16.77 Toroid.seek :clear @08:54:18.39 Sampled 1000.0ml @08:54:19.02 Plunge.seek :reagent @08:54:19.94 Sample loop closed. Ready to Deliver Reagents @08:54:19.96 Archive.seek :sealed @08:54:23.66 Archive.seek :evac @08:54:37.64 delay 1:00 @08:55:37.68 Archive.seek :clear @08:55:54.89 Sample preserved. Puck Full. @08:55:55.22
Carousel.seek 1 @08:55:55.52 Plunge.seek :locked @08:55:56.66 Air.seek :clear @08:55:56.87 Lysis1.seek :clear @08:56:00.41
Archive.coast @08:56:00.50 Air.coast @08:56:00.59 Lysis1.coast @08:56:00.66 Lysis2.coast @08:56:00.75 Plunge.coast @08:56:00.83 SamplePump.coast @08:56:00.92 Toroid.coast @08:56:01.03 Carousel.coast @08:56:01.30 Gate.power :motor,:OFF @08:56:01.32 Gate.power :core,:OFF @08:56:01.34 Gate.power :analytic1,:OFF @08:56:01.41 Gate.power :analytic2,:OFF @08:56:01.45 Gate.power :analytic3,:OFF @08:56:01.50 Gate.power :main,:OFF @08:56:01.54 Safely stopped and ready to power off @08:56:02.90 Cartridge Load: {42 => { Type: :archive_bac, State: :processed, filtered: 800}, 44 => { Type: :archive_bac, State: :processed, filtered: 1000}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 179.512}, 51 => { Type: :aRNAl_bac, State: :processed, filtered: 234.66}, [39, 37, 35, 33, 31, 23, 21, 19, 17, 15, 13, 11, 9, 7, 5, 3] => { Type: :aRNAl_bac, State: :reserved, was: :dry}, [40, 38, 36, 34, 32, 30, 22, 20, 18, 16, 14, 12, 10, 8, 6, 4, 2] => { Type: :archive_bac, State: :reserved, was: :dry}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 800}, [49, 47] => { Type: :aRNAl_bac, State: :processed, filtered: 200}, [50, 48, 46] => { Type: :archive_bac, State: :processed, filtered: 200}, [60, 29, 28] => { Type: :archive_bac, State: :missing}, 1 => { Type: :aRNAl_bac, State: :processed, filtered: 1000.0}} Spares: {spare: 0} @08:56:02.92 Cartridge::None in MAIN -- Not enough dry cartridges remain @08:56:03.12 Session Ended @09:12:26.30PST12-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @09:12:26.34 >BREAK< ESP3 Gateway v1.11 for cartridge bus @09:12:30.17 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @09:12:30.55 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @09:12:35.73 Polling Can every 2:00 @09:12:35.79 Loading /home/esp/esp2/bin/processAll mission @09:12:36.15 Sampling default goal volumes @09:12:36.16 Cartridge group: dry Spares: {spare: 0} @09:12:36.34 Gate.power :main,:ON,limit to 5000mA @09:12:36.35 Gate.power :analytic1,:ON,limit to 5000mA @09:12:36.37 Gate.power :analytic2,:ON,limit to 5000mA @09:12:36.40 Gate.power :analytic3,:ON,limit to 5000mA @09:12:36.44 Powering core bus @09:12:38.83 Archive.configure :default @09:12:38.99 Air.configure :default @09:12:39.12 Lysis1.configure :default @09:12:39.26 Lysis2.configure :default @09:12:39.43 Plunge.configure :default @09:12:39.64 SamplePump.configure :default @09:12:39.81 ExtraPump.configure :default @09:12:39.98 Toroid.configure :default @09:12:40.14 Carousel.configure :default @09:12:40.31 Pressure.configure :default @09:12:40.38 All elf controllers running firmware version 4.61 @09:12:40.40 Gate.power :motor,:ON,limit to 5000mA @09:12:40.56 Archive.reconfigure :default @09:12:40.95 Archive.home.jog 0 @09:12:41.32 Plunge.reconfigure :default @09:12:41.39 Lysis2.reconfigure :default @09:12:41.41 Air.reconfigure :default @09:12:41.47 Toroid.reconfigure :default @09:12:41.50 Lysis1.reconfigure :default @09:12:41.78 Plunge.home.jog 0 @09:12:41.86 Toroid.home.jog 0 @09:12:41.88 Lysis2.home.jog 0 @09:12:41.90 Lysis1.home.jog 0 @09:12:41.92 Air.home.jog 0 @09:12:42.55 Archive.reconfigure :homing @09:12:42.82 Plunge.reconfigure :homing @09:12:42.94 Archive.jog 1374 @09:12:42.99 Lysis1.reconfigure :homing @09:12:43.04 Toroid.reconfigure :homing @09:12:43.09 Air.reconfigure :homing @09:12:43.14 Lysis2.reconfigure :homing @09:12:43.42 Plunge.jog 849 @09:12:43.48 Lysis1.jog 1351 @09:12:43.53 Toroid.jog 634 @09:12:43.61 Lysis2.jog 1357 @09:12:43.66 Air.jog 1365 @09:12:44.38 Archive.setRawPosition! 50 @09:12:44.43 Archive.seek :home @09:12:44.56 Plunge.setRawPosition! 50 @09:12:44.64 Plunge.seek :home @09:12:44.72 Archive.reconfigure :default @09:12:44.84 Lysis1.setRawPosition! 50 @09:12:44.94 Lysis1.seek :home @09:12:45.08 Plunge.reconfigure :default @09:12:45.10 Lysis2.setRawPosition! 50 @09:12:45.13 Air.setRawPosition! 50 @09:12:45.30 Lysis2.seek :home @09:12:45.36 Air.seek :home @09:12:45.52 Lysis1.reconfigure :default @09:12:45.66 Lysis2.reconfigure :default @09:12:45.79 Air.reconfigure :default @09:12:45.92 Toroid.setRawPosition! 0 @09:12:45.99 Toroid.seek :home @09:12:46.09 Toroid.reconfigure :default @09:12:46.30
Toroid.seek :clear @09:12:47.90 Carousel.reconfigure :default @09:12:47.99 Carousel.home.jog 0 @09:12:48.51 Carousel.seek 1 @09:12:48.70 Selecting Cartridge 5 @09:12:48.98 Carousel.seek 5 @09:12:50.18 Lysis1.seek :locked @09:12:53.47 Plunge.seek :locked @09:12:53.65 Air.seek :locked @09:12:57.93
Mounted archive Cartridge @09:12:58.31 Duration of filtering limited to 1:05:00 @09:12:58.66 Priming sample loop w/100ml, bypass w/10ml @09:12:58.67 SamplePump.setPosition! 0ml @09:12:58.75 Bypass.open @09:13:00.28 Intake.open @09:13:01.75 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @09:13:04.38 Vacuum tare is 19% of full scale @09:13:04.40 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @09:13:04.66 Pressure.reconfigure :default @09:13:04.71 Pressure.hold @09:13:04.99 SamplePump.reconfigure :prime @09:13:05.09 SamplePump.seek 10ml,1:21 @09:13:16.20 Pressure.coast @09:13:16.27 SamplePump.setPosition! 0ml @09:13:16.32 Exhaust.open @09:13:17.77 Bypass.close @09:13:19.37 Pressure.reconfigure :default @09:13:19.43 Pressure.hold @09:13:19.78 SamplePump.reconfigure :prime @09:13:19.84 SamplePump.seek 100ml,8:51 @09:15:07.21 Pressure.coast @09:15:07.31 Bypass.open @09:15:08.03 Intake.close @09:15:08.92 Toroid.reconfigure :hold @09:15:08.98 Toroid.seek :crack @09:15:10.59 Toroid.seek :cartridge @09:15:11.58 Toroid.reconfigure :default @09:15:11.64 Toroid.seek :clear @09:15:13.55 Pressurizing cartridge to 26psi above 14.6psia @09:15:13.57 SamplePump.setPosition! 0ml @09:15:13.62 Intake.open @09:15:14.42 Bypass.close @09:15:15.15 Exhaust.close @09:15:16.04 Pressure.reconfigure :default @09:15:16.13 Pressure.hold @09:15:16.40 SamplePump.reconfigure :charge @09:15:16.46 SamplePump.seek 2.3ml,42 seconds @09:15:30.67 Pressure.coast @09:15:30.75 Intake.close @09:15:31.67 Pressurized cartridge with 0.924ml to 26.0psi @09:15:31.68 Isolated @36.8psia (22.2psi) @09:15:31.69 Checking for cartridge leaks from 36.8psia @09:15:31.71 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @09:15:36.42 Sampling 1000.0ml @09:15:36.45 SamplePump.setPosition! 0ml @09:15:36.53 Exhaust.open @09:15:37.26 Intake.open @09:15:38.25 Pressure.reconfigure :default @09:15:38.32 Pressure.hold @09:15:38.69 SamplePump.reconfigure :sample @09:15:38.77 SamplePump.seek 1000ml,1:23:51 @09:15:41.32 Cartridge 5 sampling 1/1000ml at 0.61ml/s with 27.5psi vacuum @09:16:41.32 Cartridge 5 sampling 33/1000ml at 0.52ml/s with 20.5psi vacuum @09:17:41.34 Cartridge 5 sampling 65/1000ml at 0.54ml/s with 24.9psi vacuum @09:18:41.34 Cartridge 5 sampling 96/1000ml at 0.53ml/s with 24.4psi vacuum @09:19:41.32 Cartridge 5 sampling 128/1000ml at 0.52ml/s with 22.2psi vacuum @09:20:41.35 Cartridge 5 sampling 160/1000ml at 0.53ml/s with 25.7psi vacuum @09:21:41.46 Cartridge 5 sampling 191/1000ml at 0.51ml/s with 27.3psi vacuum @09:22:41.32 Cartridge 5 sampling 223/1000ml at 0.53ml/s with 23.5psi vacuum @09:23:41.34 Cartridge 5 sampling 254/1000ml at 0.53ml/s with 23.1psi vacuum @09:24:41.32 Cartridge 5 sampling 285/1000ml at 0.53ml/s with 25.6psi vacuum @09:25:41.35 Cartridge 5 sampling 317/1000ml at 0.51ml/s with 19.9psi vacuum @09:26:41.36 Cartridge 5 sampling 348/1000ml at 0.52ml/s with 22.4psi vacuum @09:27:41.33 Cartridge 5 sampling 379/1000ml at 0.50ml/s with 27.7psi vacuum @09:28:41.34 Cartridge 5 sampling 410/1000ml at 0.52ml/s with 20.1psi vacuum @09:29:41.33 Cartridge 5 sampling 441/1000ml at 0.49ml/s with 26.6psi vacuum @09:30:41.32 Cartridge 5 sampling 472/1000ml at 0.52ml/s with 21.6psi vacuum @09:31:41.33 Cartridge 5 sampling 503/1000ml at 0.53ml/s with 25.5psi vacuum @09:32:41.33 Cartridge 5 sampling 534/1000ml at 0.52ml/s with 21.1psi vacuum @09:33:41.33 Cartridge 5 sampling 565/1000ml at 0.49ml/s with 23.8psi vacuum @09:34:41.33 Cartridge 5 sampling 596/1000ml at 0.52ml/s with 26.4psi vacuum @09:35:41.32 Cartridge 5 sampling 627/1000ml at 0.50ml/s with 27.6psi vacuum @09:36:41.32 Cartridge 5 sampling 658/1000ml at 0.53ml/s with 24.7psi vacuum @09:37:41.32 Cartridge 5 sampling 688/1000ml at 0.52ml/s with 20.4psi vacuum 1.9m deep (17.5psia) @09:38:41.32 Cartridge 5 sampling 719/1000ml at 0.50ml/s with 25.9psi vacuum @09:39:41.33 Cartridge 5 sampling 750/1000ml at 0.51ml/s with 26.4psi vacuum @09:40:41.32 Cartridge 5 sampling 781/1000ml at 0.49ml/s with 27.7psi vacuum @09:41:41.36 Cartridge 5 sampling 812/1000ml at 0.51ml/s with 17.4psi vacuum 2.3m deep (18.0psia) @09:42:41.33 Cartridge 5 sampling 842/1000ml at 0.52ml/s with 21.6psi vacuum @09:43:41.35 Cartridge 5 sampling 873/1000ml at 0.51ml/s with 18.2psi vacuum @09:44:41.34 Cartridge 5 sampling 904/1000ml at 0.49ml/s with 27.6psi vacuum @09:45:41.34 Cartridge 5 sampling 935/1000ml at 0.48ml/s with 25.4psi vacuum @09:46:41.32 Cartridge 5 sampling 966/1000ml at 0.53ml/s with 26.8psi vacuum @09:47:41.32 Cartridge 5 sampling 996/1000ml at 0.49ml/s with 17.8psi vacuum @09:47:48.08 Pressure.coast @09:47:48.26 Intake.close @09:47:48.35 Exhaust.close @09:47:48.58 Toroid.reconfigure :hold @09:47:48.66 Toroid.seek :bypass @09:47:50.29 Toroid.reconfigure :default @09:47:50.35 Toroid.seek :clear @09:47:51.99 Sampled 1000.0ml @09:47:52.60 Plunge.seek :reagent @09:47:53.50 Sample loop closed. Ready to Deliver Reagents @09:47:53.52 Archive.seek :sealed @09:47:57.19 Archive.seek :evac @09:48:11.19 delay 1:00 @09:49:11.23 Archive.seek :clear @09:49:28.37 Sample preserved. Puck Full. @09:49:28.70
Carousel.seek 5 @09:49:28.93 Plunge.seek :locked @09:49:30.08 Air.seek :clear @09:49:31.19 Lysis1.seek :clear @09:49:34.41
Archive.coast @09:49:34.45 Air.coast @09:49:34.52 Lysis1.coast @09:49:34.57 Lysis2.coast @09:49:34.61 Plunge.coast @09:49:34.68 SamplePump.coast @09:49:34.72 Toroid.coast @09:49:34.79 Carousel.coast @09:49:35.06 Gate.power :motor,:OFF @09:49:35.07 Gate.power :core,:OFF @09:49:35.08 Gate.power :analytic1,:OFF @09:49:35.10 Gate.power :analytic2,:OFF @09:49:35.11 Gate.power :analytic3,:OFF @09:49:35.12 Gate.power :main,:OFF @09:49:35.13 Safely stopped and ready to power off @09:49:35.61 Gate.power :main,:ON,limit to 5000mA @09:49:35.63 Gate.power :analytic1,:ON,limit to 5000mA @09:49:35.65 Gate.power :analytic2,:ON,limit to 5000mA @09:49:35.66 Gate.power :analytic3,:ON,limit to 5000mA @09:49:36.13 Powering core bus @09:49:38.53 Archive.configure :default @09:49:38.69 Air.configure :default @09:49:38.87 Lysis1.configure :default @09:49:39.02 Lysis2.configure :default @09:49:39.18 Plunge.configure :default @09:49:39.38 SamplePump.configure :default @09:49:39.55 ExtraPump.configure :default @09:49:39.71 Toroid.configure :default @09:49:39.88 Carousel.configure :default @09:49:40.04 Pressure.configure :default @09:49:40.13 All elf controllers running firmware version 4.61 @09:49:40.15 Gate.power :motor,:ON,limit to 5000mA @09:49:40.41 Archive.reconfigure :default @09:49:40.85 Archive.home.jog 0 @09:49:41.14 Plunge.reconfigure :default @09:49:41.22 Lysis2.reconfigure :default @09:49:41.26 Lysis1.reconfigure :default @09:49:41.29 Air.reconfigure :default @09:49:41.34 Toroid.reconfigure :default @09:49:41.62 Plunge.home.jog 0 @09:49:41.70 Lysis1.home.jog 0 @09:49:41.73 Air.home.jog 0 @09:49:41.80 Lysis2.home.jog 0 @09:49:41.84 Toroid.home.jog 0 @09:49:42.35 Archive.reconfigure :homing @09:49:42.77 Archive.jog 1374 @09:49:42.84 Toroid.reconfigure :homing @09:49:42.87 Lysis2.reconfigure :homing @09:49:42.90 Lysis1.reconfigure :homing @09:49:43.00 Plunge.reconfigure :homing @09:49:43.04 Air.reconfigure :homing @09:49:43.34 Lysis2.jog 1358 @09:49:43.42 Toroid.jog 635 @09:49:43.46 Lysis1.jog 1346 @09:49:43.48 Plunge.jog 853 @09:49:43.52 Air.jog 1366 @09:49:44.25 Archive.setRawPosition! 50 @09:49:44.30 Archive.seek :home @09:49:44.63 Archive.reconfigure :default @09:49:44.75 Lysis2.setRawPosition! 50 @09:49:44.80 Plunge.setRawPosition! 50 @09:49:44.91 Lysis2.seek :home @09:49:44.95 Plunge.seek :home @09:49:45.18 Lysis1.setRawPosition! 50 @09:49:45.20 Air.setRawPosition! 50 @09:49:45.31 Lysis1.seek :home @09:49:45.37 Air.seek :home @09:49:45.59 Plunge.reconfigure :default @09:49:45.61 Lysis2.reconfigure :default @09:49:45.90 Lysis1.reconfigure :default @09:49:45.92 Air.reconfigure :default @09:49:45.96 Toroid.setRawPosition! 0 @09:49:46.16 Toroid.seek :home @09:49:46.32 Toroid.reconfigure :default @09:49:46.56
Toroid.seek :clear @09:49:48.18 Carousel.reconfigure :default @09:49:48.23 Carousel.home.jog 0 @09:49:48.79 Carousel.seek 5 @09:49:49.02 Selecting Cartridge 4 @09:49:49.30 Carousel.seek 4 @09:49:49.85 Lysis1.seek :locked @09:49:53.14 Plunge.seek :locked @09:49:53.32 Air.seek :locked @09:49:57.57
Mounted archive Cartridge @09:49:57.96 Duration of filtering limited to 1:05:00 @09:49:58.35 Priming sample loop w/100ml, bypass w/10ml @09:49:58.37 SamplePump.setPosition! 0ml @09:49:58.40 Bypass.open @09:50:00.01 Intake.open @09:50:01.03 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @09:50:03.65 Vacuum tare is 19% of full scale @09:50:03.67 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @09:50:03.94 Pressure.reconfigure :default @09:50:04.00 Pressure.hold @09:50:04.25 SamplePump.reconfigure :prime @09:50:04.36 SamplePump.seek 10ml,1:21 @09:50:15.49 Pressure.coast @09:50:15.56 SamplePump.setPosition! 0ml @09:50:15.62 Exhaust.open @09:50:17.05 Bypass.close @09:50:18.66 Pressure.reconfigure :default @09:50:18.72 Pressure.hold @09:50:19.06 SamplePump.reconfigure :prime @09:50:19.15 SamplePump.seek 100ml,8:51 @09:52:06.52 Pressure.coast @09:52:06.60 Bypass.open @09:52:07.29 Intake.close @09:52:08.24 Toroid.reconfigure :hold @09:52:08.30 Toroid.seek :crack @09:52:09.90 Toroid.seek :cartridge @09:52:10.87 Toroid.reconfigure :default @09:52:10.96 Toroid.seek :clear @09:52:12.85 Pressurizing cartridge to 26psi above 14.7psia @09:52:12.87 SamplePump.setPosition! 0ml @09:52:12.93 Intake.open @09:52:13.72 Bypass.close @09:52:14.44 Exhaust.close @09:52:16.01 Pressure.reconfigure :default @09:52:16.08 Pressure.hold @09:52:16.35 SamplePump.reconfigure :charge @09:52:16.43 SamplePump.seek 2.3ml,42 seconds @09:52:30.72 Pressure.coast @09:52:30.77 Intake.close @09:52:31.68 Pressurized cartridge with 0.935ml to 26.7psi @09:52:31.70 Isolated @36.8psia (22.1psi) @09:52:31.72 Checking for cartridge leaks from 36.8psia @09:52:31.74 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @09:52:36.32 Cartridge pressure fell 1% to 36.6psia @09:52:36.43 Sampling 1000.0ml @09:52:36.45 SamplePump.setPosition! 0ml @09:52:36.53 Exhaust.open @09:52:37.21 Intake.open @09:52:38.23 Pressure.reconfigure :default @09:52:38.31 Pressure.hold @09:52:38.65 SamplePump.reconfigure :sample @09:52:38.74 SamplePump.seek 1000ml,1:23:51 @09:52:41.29 Cartridge 4 sampling 1/1000ml at 0.57ml/s with 23.4psi vacuum @09:53:41.29 Cartridge 4 sampling 32/1000ml at 0.51ml/s with 20.8psi vacuum @09:54:41.29 Cartridge 4 sampling 62/1000ml at 0.50ml/s with 26.0psi vacuum @09:55:41.30 Cartridge 4 sampling 93/1000ml at 0.49ml/s with 19.9psi vacuum @09:56:41.30 Cartridge 4 sampling 123/1000ml at 0.50ml/s with 25.9psi vacuum @09:57:08.85 Session Begun @09:57:41.29 Cartridge 4 sampling 153/1000ml at 0.48ml/s with 20.0psi vacuum @09:58:41.32 Cartridge 4 sampling 183/1000ml at 0.51ml/s with 25.7psi vacuum @09:59:41.32 Cartridge 4 sampling 214/1000ml at 0.50ml/s with 20.1psi vacuum 1.6m deep (17.0psia) @10:00:41.29 Cartridge 4 sampling 244/1000ml at 0.50ml/s with 26.1psi vacuum @10:01:41.30 Cartridge 4 sampling 273/1000ml at 0.51ml/s with 24.6psi vacuum @10:02:41.30 Cartridge 4 sampling 303/1000ml at 0.50ml/s with 26.2psi vacuum @10:03:41.29 Cartridge 4 sampling 334/1000ml at 0.50ml/s with 20.2psi vacuum 1.8m deep (17.3psia) @10:04:41.29 Cartridge 4 sampling 364/1000ml at 0.52ml/s with 22.0psi vacuum @10:05:41.30 Cartridge 4 sampling 394/1000ml at 0.50ml/s with 26.8psi vacuum @10:06:41.30 Cartridge 4 sampling 424/1000ml at 0.50ml/s with 26.2psi vacuum @10:07:41.29 Cartridge 4 sampling 454/1000ml at 0.52ml/s with 25.3psi vacuum @10:08:41.29 Cartridge 4 sampling 484/1000ml at 0.51ml/s with 24.7psi vacuum @10:09:41.32 Cartridge 4 sampling 514/1000ml at 0.48ml/s with 27.1psi vacuum @10:10:41.28 Cartridge 4 sampling 545/1000ml at 0.48ml/s with 17.9psi vacuum @10:11:41.30 Cartridge 4 sampling 575/1000ml at 0.51ml/s with 21.1psi vacuum @10:12:41.28 Cartridge 4 sampling 605/1000ml at 0.51ml/s with 21.8psi vacuum 1.7m deep (17.2psia) @10:13:41.30 Cartridge 4 sampling 635/1000ml at 0.48ml/s with 18.5psi vacuum @10:14:41.29 Cartridge 4 sampling 665/1000ml at 0.48ml/s with 28.6psi vacuum @10:15:41.31 Cartridge 4 sampling 695/1000ml at 0.49ml/s with 19.1psi vacuum 1.9m deep (17.5psia) @10:16:41.30 Cartridge 4 sampling 725/1000ml at 0.51ml/s with 20.3psi vacuum 2.3m deep (18.0psia) @10:17:41.30 Cartridge 4 sampling 755/1000ml at 0.50ml/s with 26.9psi vacuum @10:18:41.29 Cartridge 4 sampling 785/1000ml at 0.51ml/s with 25.5psi vacuum @10:19:41.30 Cartridge 4 sampling 815/1000ml at 0.49ml/s with 17.7psi vacuum 3.8m deep (20.3psia) @10:20:41.31 Cartridge 4 sampling 844/1000ml at 0.50ml/s with 27.2psi vacuum @10:21:41.29 Cartridge 4 sampling 874/1000ml at 0.49ml/s with 16.1psi vacuum @10:22:41.31 Cartridge 4 sampling 904/1000ml at 0.49ml/s with 26.0psi vacuum @10:23:41.30 Cartridge 4 sampling 933/1000ml at 0.51ml/s with 21.6psi vacuum @10:24:41.28 Cartridge 4 sampling 963/1000ml at 0.51ml/s with 23.9psi vacuum @10:25:41.30 Cartridge 4 sampling 993/1000ml at 0.50ml/s with 23.4psi vacuum @10:25:55.19 Pressure.coast @10:25:55.43 Intake.close @10:25:55.54 Exhaust.close @10:25:55.80 Toroid.reconfigure :hold @10:25:55.90 Toroid.seek :bypass @10:25:57.59 Toroid.reconfigure :default @10:25:57.68 Toroid.seek :clear @10:25:59.36 Sampled 1000.0ml @10:25:59.97 Plunge.seek :reagent @10:26:00.94 Sample loop closed. Ready to Deliver Reagents @10:26:00.96 Archive.seek :sealed @10:26:04.66 Archive.seek :evac @10:26:18.69 delay 1:00 @10:27:18.73 Archive.seek :clear @10:27:35.94 RNAlater soak for 10:00 @10:31:08.65 -> exit @10:31:08.68 Session Ended @10:37:35.98 Archive.seek :tVent @10:37:37.63 delay 1 second @10:37:38.67 Archive.seek :evac @10:37:54.66 delay 10 seconds @10:38:04.70 Sample preserved and evacuated @10:38:05.04
Carousel.seek 4 @10:38:05.29 Plunge.seek :locked @10:38:06.31 Archive.seek :clear @10:38:06.51 Air.seek :clear @10:38:06.73 Lysis1.seek :clear @10:38:23.54
Archive.coast @10:38:23.59 Air.coast @10:38:23.64 Lysis1.coast @10:38:23.70 Lysis2.coast @10:38:23.74 Plunge.coast @10:38:23.81 SamplePump.coast @10:38:23.85 Toroid.coast @10:38:23.90 Carousel.coast @10:38:24.15 Gate.power :motor,:OFF @10:38:24.16 Gate.power :core,:OFF @10:38:24.19 Gate.power :analytic1,:OFF @10:38:24.26 Gate.power :analytic2,:OFF @10:38:24.34 Gate.power :analytic3,:OFF @10:38:24.36 Gate.power :main,:OFF @10:38:24.39 Safely stopped and ready to power off @10:38:24.74 Gate.power :main,:ON,limit to 5000mA @10:38:24.75 Gate.power :analytic1,:ON,limit to 5000mA @10:38:24.77 Gate.power :analytic2,:ON,limit to 5000mA @10:38:24.81 Gate.power :analytic3,:ON,limit to 5000mA @10:38:25.23 Powering core bus @10:38:27.62 Archive.configure :default @10:38:27.82 Air.configure :default @10:38:27.94 Lysis1.configure :default @10:38:28.09 Lysis2.configure :default @10:38:28.28 Plunge.configure :default @10:38:28.44 SamplePump.configure :default @10:38:28.67 ExtraPump.configure :default @10:38:28.84 Toroid.configure :default @10:38:29.04 Carousel.configure :default @10:38:29.23 Pressure.configure :default @10:38:29.30 All elf controllers running firmware version 4.61 @10:38:29.33 Gate.power :motor,:ON,limit to 5000mA @10:38:30.84 Lysis2.reconfigure :default @10:38:30.89 Air.reconfigure :default @10:38:30.93 Archive.reconfigure :default @10:38:31.00 Toroid.reconfigure :default @10:38:31.04 Lysis1.reconfigure :default @10:38:31.10 Plunge.reconfigure :default @10:38:31.39 Lysis2.home.jog 0 @10:38:31.48 Air.home.jog 0 @10:38:31.50 Archive.home.jog 0 @10:38:31.54 Toroid.home.jog 0 @10:38:31.55 Plunge.home.jog 0 @10:38:31.61 Lysis1.home.jog 0 @10:38:32.68 Lysis1.reconfigure :homing @10:38:32.78 Toroid.reconfigure :homing @10:38:32.81 Plunge.reconfigure :homing @10:38:32.89 Archive.reconfigure :homing @10:38:32.95 Lysis2.reconfigure :homing @10:38:33.03 Air.reconfigure :homing @10:38:33.30 Toroid.jog 627 @10:38:33.33 Lysis1.jog 1349 @10:38:33.41 Plunge.jog 851 @10:38:33.46 Archive.jog 1375 @10:38:33.52 Lysis2.jog 1358 @10:38:33.54 Air.jog 1367 @10:38:34.53 Plunge.setRawPosition! 50 @10:38:34.58 Plunge.seek :home @10:38:34.71 Lysis1.setRawPosition! 50 @10:38:34.78 Lysis1.seek :home @10:38:34.97 Plunge.reconfigure :default @10:38:35.00 Lysis2.setRawPosition! 50 @10:38:35.07 Archive.setRawPosition! 50 @10:38:35.21 Lysis2.seek :home @10:38:35.28 Archive.seek :home @10:38:35.48 Lysis1.reconfigure :default @10:38:35.50 Air.setRawPosition! 50 @10:38:35.66 Air.seek :home @10:38:35.86 Toroid.setRawPosition! 0 @10:38:35.89 Lysis2.reconfigure :default @10:38:35.90 Archive.reconfigure :default @10:38:36.10 Toroid.seek :home @10:38:36.32 Toroid.reconfigure :default @10:38:36.37 Air.reconfigure :default @10:38:36.87
Toroid.seek :clear @10:38:38.51 Carousel.reconfigure :default @10:38:38.56 Carousel.home.jog 0 @10:38:39.10 Carousel.seek 4 @10:38:39.31 Selecting Cartridge 3 @10:38:39.58 Carousel.seek 3 @10:38:40.13 Lysis1.seek :locked @10:38:43.40 Plunge.seek :locked @10:38:43.59 Air.seek :locked @10:38:47.87
Mounted archive Cartridge @10:38:48.46 Duration of filtering limited to 1:05:00 @10:38:48.83 Priming sample loop w/100ml, bypass w/10ml @10:38:48.85 SamplePump.setPosition! 0ml @10:38:48.91 Bypass.open @10:38:50.41 Intake.open @10:38:51.34 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @10:38:53.96 Vacuum tare is 19% of full scale @10:38:53.98 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @10:38:54.25 Pressure.reconfigure :default @10:38:54.32 Pressure.hold @10:38:54.67 SamplePump.reconfigure :prime @10:38:54.77 SamplePump.seek 10ml,1:21 @10:39:05.88 Pressure.coast @10:39:05.96 SamplePump.setPosition! 0ml @10:39:06.01 Exhaust.open @10:39:07.38 Bypass.close @10:39:08.99 Pressure.reconfigure :default @10:39:09.08 Pressure.hold @10:39:09.40 SamplePump.reconfigure :prime @10:39:09.49 SamplePump.seek 100ml,8:51 @10:40:56.80 Pressure.coast @10:40:56.88 Bypass.open @10:40:57.60 Intake.close @10:40:58.55 Toroid.reconfigure :hold @10:40:58.64 Toroid.seek :crack @10:41:00.25 Toroid.seek :cartridge @10:41:01.21 Toroid.reconfigure :default @10:41:01.27 Toroid.seek :clear @10:41:03.21 Pressurizing cartridge to 26psi above 14.6psia @10:41:03.23 SamplePump.setPosition! 0ml @10:41:03.26 Intake.open @10:41:04.10 Bypass.close @10:41:04.80 Exhaust.close @10:41:05.66 Pressure.reconfigure :default @10:41:05.73 Pressure.hold @10:41:05.99 SamplePump.reconfigure :charge @10:41:06.08 SamplePump.seek 2.3ml,42 seconds @10:41:20.28 Pressure.coast @10:41:20.33 Intake.close @10:41:21.29 Pressurized cartridge with 0.926ml to 26.1psi @10:41:21.30 Isolated @37.1psia (22.5psi) @10:41:21.31 Checking for cartridge leaks from 37.1psia @10:41:21.35 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @10:41:26.05 Sampling 1000.0ml @10:41:26.08 SamplePump.setPosition! 0ml @10:41:26.13 Exhaust.open @10:41:26.81 Intake.open @10:41:27.84 Pressure.reconfigure :default @10:41:27.92 Pressure.hold @10:41:28.26 SamplePump.reconfigure :sample @10:41:28.36 SamplePump.seek 1000ml,1:23:51 @10:41:30.93 Cartridge 3 sampling 1/1000ml at 0.61ml/s with 26.1psi vacuum @10:42:31.57 Cartridge 3 sampling 33/1000ml at 0.53ml/s with 21.3psi vacuum @10:43:30.93 Cartridge 3 sampling 65/1000ml at 0.53ml/s with 26.5psi vacuum @10:44:30.95 Cartridge 3 sampling 96/1000ml at 0.53ml/s with 20.6psi vacuum @10:45:30.96 Cartridge 3 sampling 128/1000ml at 0.53ml/s with 23.9psi vacuum @10:46:30.93 Cartridge 3 sampling 160/1000ml at 0.50ml/s with 23.2psi vacuum @10:47:30.94 Cartridge 3 sampling 191/1000ml at 0.53ml/s with 26.3psi vacuum @10:48:30.94 Cartridge 3 sampling 223/1000ml at 0.50ml/s with 20.7psi vacuum @10:49:30.95 Cartridge 3 sampling 254/1000ml at 0.54ml/s with 25.6psi vacuum @10:50:30.95 Cartridge 3 sampling 286/1000ml at 0.50ml/s with 26.8psi vacuum @10:51:30.94 Cartridge 3 sampling 317/1000ml at 0.53ml/s with 25.9psi vacuum @10:52:30.93 Cartridge 3 sampling 349/1000ml at 0.54ml/s with 20.8psi vacuum 1.6m deep (17.1psia) @10:53:30.95 Cartridge 3 sampling 380/1000ml at 0.54ml/s with 23.8psi vacuum @10:54:30.94 Cartridge 3 sampling 411/1000ml at 0.53ml/s with 23.3psi vacuum @10:55:30.95 Cartridge 3 sampling 443/1000ml at 0.52ml/s with 18.5psi vacuum 2.4m deep (18.3psia) @10:56:30.93 Cartridge 3 sampling 474/1000ml at 0.54ml/s with 20.5psi vacuum 1.6m deep (17.1psia) @10:57:30.97 Cartridge 3 sampling 506/1000ml at 0.51ml/s with 27.8psi vacuum @10:58:30.95 Cartridge 3 sampling 537/1000ml at 0.53ml/s with 20.0psi vacuum 2.3m deep (18.0psia) @10:59:30.96 Cartridge 3 sampling 569/1000ml at 0.49ml/s with 27.0psi vacuum @11:00:30.94 Cartridge 3 sampling 601/1000ml at 0.53ml/s with 26.9psi vacuum @11:01:30.95 Cartridge 3 sampling 632/1000ml at 0.53ml/s with 24.8psi vacuum @11:02:30.94 Cartridge 3 sampling 664/1000ml at 0.52ml/s with 19.1psi vacuum 3.1m deep (19.2psia) @11:03:30.94 Cartridge 3 sampling 696/1000ml at 0.53ml/s with 26.9psi vacuum @11:04:30.93 Cartridge 3 sampling 728/1000ml at 0.53ml/s with 18.4psi vacuum 3.6m deep (20.0psia) @11:05:30.93 Cartridge 3 sampling 760/1000ml at 0.55ml/s with 27.1psi vacuum @11:06:30.95 Cartridge 3 sampling 792/1000ml at 0.52ml/s with 24.8psi vacuum @11:07:30.94 Cartridge 3 sampling 824/1000ml at 0.51ml/s with 29.1psi vacuum @11:08:30.94 Cartridge 3 sampling 856/1000ml at 0.53ml/s with 21.8psi vacuum @11:09:30.93 Cartridge 3 sampling 888/1000ml at 0.55ml/s with 25.3psi vacuum @11:10:30.94 Cartridge 3 sampling 920/1000ml at 0.54ml/s with 27.0psi vacuum @11:11:30.93 Cartridge 3 sampling 952/1000ml at 0.53ml/s with 27.2psi vacuum @11:12:30.94 Cartridge 3 sampling 984/1000ml at 0.54ml/s with 17.6psi vacuum 4.1m deep (20.8psia) @11:13:00.20 Pressure.coast @11:13:00.37 Intake.close @11:13:00.43 Exhaust.close @11:13:00.67 Toroid.reconfigure :hold @11:13:00.75 Toroid.seek :bypass @11:13:02.38 Toroid.reconfigure :default @11:13:02.46 Toroid.seek :clear @11:13:04.07 Sampled 1000.0ml @11:13:04.66 Plunge.seek :reagent @11:13:05.61 Sample loop closed. Ready to Deliver Reagents @11:13:05.63 Archive.seek :sealed @11:13:09.29 Archive.seek :evac @11:13:23.31 delay 1:00 @11:14:23.34 Archive.seek :clear @11:14:40.47 Sample preserved. Puck Full. @11:14:40.77
Carousel.seek 3 @11:14:41.04 Plunge.seek :locked @11:14:42.16 Air.seek :clear @11:14:42.36 Lysis1.seek :clear @11:14:45.90
Archive.coast @11:14:45.96 Air.coast @11:14:46.01 Lysis1.coast @11:14:46.08 Lysis2.coast @11:14:46.13 Plunge.coast @11:14:46.18 SamplePump.coast @11:14:46.25 Toroid.coast @11:14:46.29 Carousel.coast @11:14:46.52 Gate.power :motor,:OFF @11:14:46.53 Gate.power :core,:OFF @11:14:46.54 Gate.power :analytic1,:OFF @11:14:46.56 Gate.power :analytic2,:OFF @11:14:46.57 Gate.power :analytic3,:OFF @11:14:46.58 Gate.power :main,:OFF @11:14:46.61 Safely stopped and ready to power off @11:14:46.97 Gate.power :main,:ON,limit to 5000mA @11:14:46.98 Gate.power :analytic1,:ON,limit to 5000mA @11:14:47.00 Gate.power :analytic2,:ON,limit to 5000mA @11:14:47.03 Gate.power :analytic3,:ON,limit to 5000mA @11:14:47.58 Powering core bus @11:14:49.96 Archive.configure :default @11:14:50.13 Air.configure :default @11:14:50.28 Lysis1.configure :default @11:14:50.41 Lysis2.configure :default @11:14:50.58 Plunge.configure :default @11:14:50.77 SamplePump.configure :default @11:14:50.97 ExtraPump.configure :default @11:14:51.14 Toroid.configure :default @11:14:51.32 Carousel.configure :default @11:14:51.51 Pressure.configure :default @11:14:51.56 All elf controllers running firmware version 4.61 @11:14:51.61 Gate.power :motor,:ON,limit to 5000mA @11:14:52.01 Archive.reconfigure :default @11:14:52.36 Archive.home.jog 0 @11:14:52.61 Lysis2.reconfigure :default @11:14:52.64 Lysis1.reconfigure :default @11:14:52.67 Air.reconfigure :default @11:14:52.75 Toroid.reconfigure :default @11:14:52.81 Plunge.reconfigure :default @11:14:53.08 Lysis1.home.jog 0 @11:14:53.15 Air.home.jog 0 @11:14:53.23 Lysis2.home.jog 0 @11:14:53.35 Plunge.home.jog 0 @11:14:54.08 Toroid.home.jog 0 @11:14:54.54 Archive.reconfigure :homing @11:14:54.96 Archive.jog 1373 @11:14:55.02 Toroid.reconfigure :homing @11:14:55.14 Lysis2.reconfigure :homing @11:14:55.25 Air.reconfigure :homing @11:14:55.40 Lysis1.reconfigure :homing @11:14:55.49 Plunge.reconfigure :homing @11:14:55.71 Toroid.jog 623 @11:14:55.92 Lysis2.jog 1360 @11:14:55.94 Plunge.jog 851 @11:14:55.97 Lysis1.jog 1348 @11:14:56.02 Air.jog 1368 @11:14:56.61 Archive.setRawPosition! 50 @11:14:56.68 Archive.seek :home @11:14:56.96 Archive.reconfigure :default @11:14:57.09 Plunge.setRawPosition! 50 @11:14:57.13 Plunge.seek :home @11:14:57.25 Lysis2.setRawPosition! 50 @11:14:57.36 Lysis2.seek :home @11:14:57.49 Plunge.reconfigure :default @11:14:57.53 Air.setRawPosition! 50 @11:14:57.56 Lysis1.setRawPosition! 50 @11:14:57.70 Air.seek :home @11:14:57.87 Lysis1.seek :home @11:14:58.06 Toroid.setRawPosition! 0 @11:14:58.11 Lysis2.reconfigure :default @11:14:58.23 Toroid.seek :home @11:14:58.41 Air.reconfigure :default @11:14:58.43 Toroid.reconfigure :default @11:14:58.46 Lysis1.reconfigure :default @11:14:59.06
Toroid.seek :clear @11:15:00.71 Carousel.reconfigure :default @11:15:00.76 Carousel.home.jog 0 @11:15:01.31 Carousel.seek 3 @11:15:01.53 Selecting Cartridge 2 @11:15:01.80 Carousel.seek 2 @11:15:02.39 Lysis1.seek :locked @11:15:05.65 Plunge.seek :locked @11:15:05.84 Air.seek :locked @11:15:10.10
Mounted archive Cartridge @11:15:10.46 Duration of filtering limited to 1:05:00 @11:15:10.85 Priming sample loop w/100ml, bypass w/10ml @11:15:10.88 SamplePump.setPosition! 0ml @11:15:10.93 Bypass.open @11:15:12.42 Intake.open @11:15:13.35 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @11:15:15.94 Vacuum tare is 19% of full scale @11:15:15.99 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @11:15:16.25 Pressure.reconfigure :default @11:15:16.30 Pressure.hold @11:15:16.58 SamplePump.reconfigure :prime @11:15:16.67 SamplePump.seek 10ml,1:21 @11:15:27.78 Pressure.coast @11:15:27.83 SamplePump.setPosition! 0ml @11:15:27.91 Exhaust.open @11:15:29.31 Bypass.close @11:15:30.93 Pressure.reconfigure :default @11:15:30.99 Pressure.hold @11:15:31.34 SamplePump.reconfigure :prime @11:15:31.41 SamplePump.seek 100ml,8:51 @11:17:18.78 Pressure.coast @11:17:18.85 Bypass.open @11:17:19.54 Intake.close @11:17:20.49 Toroid.reconfigure :hold @11:17:20.57 Toroid.seek :crack @11:17:22.20 Toroid.seek :cartridge @11:17:23.17 Toroid.reconfigure :default @11:17:23.22 Toroid.seek :clear @11:17:25.11 Pressurizing cartridge to 26psi above 14.7psia @11:17:25.12 SamplePump.setPosition! 0ml @11:17:25.19 Intake.open @11:17:25.99 Bypass.close @11:17:26.71 Exhaust.close @11:17:27.57 Pressure.reconfigure :default @11:17:27.63 Pressure.hold @11:17:27.90 SamplePump.reconfigure :charge @11:17:27.99 SamplePump.seek 2.3ml,42 seconds @11:17:42.15 Pressure.coast @11:17:42.20 Intake.close @11:17:43.08 Pressurized cartridge with 0.922ml to 26.2psi @11:17:43.11 Isolated @35.9psia (21.3psi) @11:17:43.13 Checking for cartridge leaks from 35.9psia @11:17:43.16 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @11:17:47.86 Sampling 1000.0ml @11:17:47.87 SamplePump.setPosition! 0ml @11:17:47.92 Exhaust.open @11:17:48.62 Intake.open @11:17:49.60 Pressure.reconfigure :default @11:17:49.65 Pressure.hold @11:17:50.01 SamplePump.reconfigure :sample @11:17:50.08 SamplePump.seek 1000ml,1:23:51 @11:17:52.67 Cartridge 2 sampling 1/1000ml at 0.63ml/s with 23.1psi vacuum @11:18:52.69 Cartridge 2 sampling 33/1000ml at 0.53ml/s with 25.9psi vacuum @11:19:52.66 Cartridge 2 sampling 65/1000ml at 0.53ml/s with 25.3psi vacuum @11:20:52.67 Cartridge 2 sampling 96/1000ml at 0.52ml/s with 27.2psi vacuum @11:21:52.69 Cartridge 2 sampling 128/1000ml at 0.52ml/s with 17.8psi vacuum @11:22:52.66 Cartridge 2 sampling 160/1000ml at 0.52ml/s with 27.2psi vacuum @11:23:52.66 Cartridge 2 sampling 191/1000ml at 0.53ml/s with 22.1psi vacuum @11:24:52.69 Cartridge 2 sampling 223/1000ml at 0.52ml/s with 27.1psi vacuum @11:25:52.67 Cartridge 2 sampling 254/1000ml at 0.53ml/s with 25.3psi vacuum @11:26:52.68 Cartridge 2 sampling 286/1000ml at 0.53ml/s with 19.9psi vacuum @11:27:52.67 Cartridge 2 sampling 317/1000ml at 0.52ml/s with 19.8psi vacuum 2.2m deep (17.9psia) @11:28:52.67 Cartridge 2 sampling 349/1000ml at 0.53ml/s with 25.8psi vacuum @11:29:52.67 Cartridge 2 sampling 381/1000ml at 0.53ml/s with 23.2psi vacuum @11:30:52.66 Cartridge 2 sampling 413/1000ml at 0.54ml/s with 18.8psi vacuum 3.1m deep (19.3psia) @11:31:52.66 Cartridge 2 sampling 444/1000ml at 0.54ml/s with 27.1psi vacuum @11:32:52.67 Cartridge 2 sampling 476/1000ml at 0.52ml/s with 27.7psi vacuum @11:33:52.66 Cartridge 2 sampling 508/1000ml at 0.55ml/s with 23.2psi vacuum @11:34:52.67 Cartridge 2 sampling 539/1000ml at 0.53ml/s with 20.1psi vacuum 2.3m deep (18.1psia) @11:35:52.68 Cartridge 2 sampling 571/1000ml at 0.52ml/s with 28.5psi vacuum @11:36:52.67 Cartridge 2 sampling 603/1000ml at 0.54ml/s with 19.0psi vacuum 3.4m deep (19.7psia) @11:37:52.66 Cartridge 2 sampling 635/1000ml at 0.54ml/s with 27.0psi vacuum @11:38:52.66 Cartridge 2 sampling 666/1000ml at 0.50ml/s with 27.8psi vacuum @11:39:52.67 Cartridge 2 sampling 698/1000ml at 0.55ml/s with 24.7psi vacuum @11:40:52.66 Cartridge 2 sampling 730/1000ml at 0.51ml/s with 19.8psi vacuum @11:41:52.67 Cartridge 2 sampling 762/1000ml at 0.53ml/s with 19.3psi vacuum 3.0m deep (19.2psia) @11:42:52.66 Cartridge 2 sampling 794/1000ml at 0.55ml/s with 24.8psi vacuum @11:43:52.68 Cartridge 2 sampling 825/1000ml at 0.53ml/s with 27.4psi vacuum @11:44:52.65 Cartridge 2 sampling 857/1000ml at 0.51ml/s with 26.9psi vacuum @11:45:52.66 Cartridge 2 sampling 889/1000ml at 0.50ml/s with 29.2psi vacuum @11:46:52.67 Cartridge 2 sampling 921/1000ml at 0.51ml/s with 16.1psi vacuum 2.5m deep (18.3psia) @11:47:52.70 Cartridge 2 sampling 952/1000ml at 0.53ml/s with 19.6psi vacuum @11:48:52.66 Cartridge 2 sampling 984/1000ml at 0.53ml/s with 14.7psi vacuum 5.8m deep (23.2psia) @11:49:22.24 Pressure.coast @11:49:22.43 Intake.close @11:49:22.47 Exhaust.close @11:49:22.72 Toroid.reconfigure :hold @11:49:22.78 Toroid.seek :bypass @11:49:24.45 Toroid.reconfigure :default @11:49:24.51 Toroid.seek :clear @11:49:26.12 Sampled 1000.0ml @11:49:26.71 Plunge.seek :reagent @11:49:27.67 Sample loop closed. Ready to Deliver Reagents @11:49:27.69 Archive.seek :sealed @11:49:31.33 Archive.seek :evac @11:49:45.30 delay 1:00 @11:50:45.33 Archive.seek :clear @11:51:02.46 RNAlater soak for 10:00 @12:01:02.49 Archive.seek :tVent @12:01:04.14 delay 1 second @12:01:05.17 Archive.seek :evac @12:01:21.13 delay 10 seconds @12:01:31.19 Sample preserved and evacuated @12:01:31.49
Carousel.seek 2 @12:01:31.75 Plunge.seek :locked @12:01:32.75 Archive.seek :clear @12:01:32.96 Air.seek :clear @12:01:33.16 Lysis1.seek :clear @12:01:49.93
Archive.coast @12:01:49.97 Air.coast @12:01:50.04 Lysis1.coast @12:01:50.08 Lysis2.coast @12:01:50.12 Plunge.coast @12:01:50.21 SamplePump.coast @12:01:50.25 Toroid.coast @12:01:50.32 Carousel.coast @12:01:50.54 Gate.power :motor,:OFF @12:01:50.57 Gate.power :core,:OFF @12:01:50.59 Gate.power :analytic1,:OFF @12:01:50.60 Gate.power :analytic2,:OFF @12:01:50.62 Gate.power :analytic3,:OFF @12:01:50.63 Gate.power :main,:OFF @12:01:50.64 Safely stopped and ready to power off @12:01:51.94 Cartridge Load: {[49, 47] => { Type: :aRNAl_bac, State: :processed, filtered: 200}, [50, 48, 46] => { Type: :archive_bac, State: :processed, filtered: 200}, [60, 29, 28] => { Type: :archive_bac, State: :missing}, 1 => { Type: :aRNAl_bac, State: :processed, filtered: 1000}, 42 => { Type: :archive_bac, State: :processed, filtered: 800}, 44 => { Type: :archive_bac, State: :processed, filtered: 1000}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 179.512}, 51 => { Type: :aRNAl_bac, State: :processed, filtered: 234.66}, [4, 2] => { Type: :archive_bac, State: :processed, filtered: 1000.0}, [5, 3] => { Type: :aRNAl_bac, State: :processed, filtered: 1000.0}, [39, 37, 35, 33, 31, 23, 21, 19, 17, 15, 13, 11, 9, 7] => { Type: :aRNAl_bac, State: :reserved, was: :dry}, [40, 38, 36, 34, 32, 30, 22, 20, 18, 16, 14, 12, 10, 8, 6] => { Type: :archive_bac, State: :reserved, was: :dry}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 800}} Spares: {spare: 0} @12:01:51.97 Cartridge::None in MAIN -- Not enough dry cartridges remain @12:31:13.17PST17-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @12:31:13.20 >BREAK< ESP3 Gateway v1.11 for cartridge bus @12:31:16.99 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @12:31:17.33 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @12:31:20.56 Polling Can every 2:00 @12:31:20.66 Gate.power :analytic1,:ON,limit to 5000mA @12:31:20.69 Gate.power :analytic2,:ON,limit to 5000mA @12:31:20.71 Gate.power :analytic3,:ON,limit to 5000mA @12:31:21.02 Archive.configure :default @12:31:21.07 Archive.hold @12:31:21.18 Archive.configure :default @12:31:21.34 Air.configure :default @12:31:21.42 Air.hold @12:31:21.47 Air.configure :default @12:31:21.62 Lysis1.configure :default @12:31:21.69 Lysis1.hold @12:31:21.75 Lysis1.configure :default @12:31:21.89 Lysis2.configure :default @12:31:21.96 Lysis2.hold @12:31:22.03 Lysis2.configure :default @12:31:22.22 Plunge.configure :default @12:31:22.27 Plunge.hold @12:31:22.36 Plunge.configure :default @12:31:22.52 SamplePump.configure :default @12:31:22.68 ExtraPump.configure :default @12:31:22.84 Toroid.configure :default @12:31:22.93 Toroid.hold @12:31:23.01 Toroid.configure :default @12:31:23.60 Carousel.configure :default @12:31:23.75 Carousel.hold @12:31:23.82 Carousel.configure :default @12:31:24.01 Pressure.configure :default @12:31:24.08 All elf controllers running firmware version 4.61 @12:31:24.20 Intake.close @12:31:24.25 Exhaust.close @12:31:24.34 Bypass.close @12:31:24.55
Loading /home/esp/esp2/bin/processAll mission @12:31:24.93 Sampling default goal volumes @12:31:24.94 Cartridge group: dry Spares: {spare: 0} @12:31:25.78 Carousel.seek 45 @12:31:25.98 Selecting Cartridge 45 @12:31:26.26 Carousel.seek 45 @12:31:26.52 Lysis1.seek :locked @12:31:29.76 Plunge.seek :locked @12:31:29.94 Air.seek :locked @12:31:34.22
Mounted archive Cartridge @12:31:34.60 Duration of filtering limited to 1:05:00 @12:31:34.99 Priming sample loop w/100ml, bypass w/10ml @12:31:35.01 SamplePump.setPosition! 0ml @12:31:35.05 Bypass.open @12:31:36.55 Intake.open @12:31:37.43 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:31:40.05 Vacuum tare is 17% of full scale @12:31:40.07 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @12:31:40.34 Pressure.reconfigure :default @12:31:40.40 Pressure.hold @12:31:40.66 SamplePump.reconfigure :prime @12:31:40.75 SamplePump.seek 10ml,1:21 @12:31:51.87 Pressure.coast @12:31:51.93 SamplePump.setPosition! 0ml @12:31:52.01 Exhaust.open @12:31:53.39 Bypass.close @12:31:55.03 Pressure.reconfigure :default @12:31:55.08 Pressure.hold @12:31:55.46 SamplePump.reconfigure :prime @12:31:55.52 SamplePump.seek 100ml,8:51 @12:33:42.93 Pressure.coast @12:33:42.99 Bypass.open @12:33:43.70 Intake.close @12:33:44.60 Toroid.reconfigure :hold @12:33:44.65 Toroid.seek :crack @12:33:46.31 Toroid.seek :cartridge @12:33:47.26 Toroid.reconfigure :default @12:33:47.32 Toroid.seek :clear @12:33:49.22 Pressurizing cartridge to 26psi above 14.2psia @12:33:49.24 SamplePump.setPosition! 0ml @12:33:49.30 Intake.open @12:33:50.06 Bypass.close @12:33:50.76 Exhaust.close @12:33:51.63 Pressure.reconfigure :default @12:33:51.71 Pressure.hold @12:33:51.99 SamplePump.reconfigure :charge @12:33:52.07 SamplePump.seek 2.3ml,42 seconds @12:34:00.19 Pressure.coast @12:34:00.27 Intake.close @12:34:01.15 Pressurized cartridge with 0.298ml to 26.0psi @12:34:01.17 Isolated @35.6psia (21.4psi) @12:34:01.18 Checking for cartridge leaks from 35.6psia @12:34:01.20 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:34:07.95 Toroid.reconfigure :hold @12:34:08.01 Toroid.seek :bypass @12:34:09.68 Toroid.reconfigure :default @12:34:09.74 Toroid.seek :clear @12:34:11.42 Cartridge::Sampler::Leaky in FILTERING -- Isolated cartridge pressure fell by 16% to 30.0psia @12:34:11.43 Retry #1 of 2 @12:34:11.45 Cmd::Error in FILTERING -- No spare cartridges remain @12:34:11.46 Cmd::Error in FILTERING -- No spare cartridges remain @12:34:11.59
Cartridge Load: {[43, 41] => { Type: :aRNAl_bac, State: :reserved, was: :dry}, 45 => { Type: :aRNAl_bac, State: :leaky}, [44, 42, 40..32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}} Spares: {spare: 0} @12:34:11.61 Thread::ChildDied in MAIN -- :FILTERING @12:34:11.73 SamplePump.coast @12:38:04.55PST17-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @12:38:04.57 >BREAK< ESP3 Gateway v1.11 for cartridge bus @12:38:08.42 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @12:38:08.82 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @12:38:12.16 Polling Can every 2:00 @12:38:12.28 Gate.power :analytic1,:ON,limit to 5000mA @12:38:12.30 Gate.power :analytic2,:ON,limit to 5000mA @12:38:12.31 Gate.power :analytic3,:ON,limit to 5000mA @12:38:12.62 Archive.configure :default @12:38:12.71 Archive.hold @12:38:12.78 Archive.configure :default @12:38:12.98 Air.configure :default @12:38:13.03 Air.hold @12:38:13.12 Air.configure :default @12:38:13.25 Lysis1.configure :default @12:38:13.30 Lysis1.hold @12:38:13.40 Lysis1.configure :default @12:38:13.54 Lysis2.configure :default @12:38:13.63 Lysis2.hold @12:38:13.69 Lysis2.configure :default @12:38:13.90 Plunge.configure :default @12:38:13.95 Plunge.hold @12:38:14.03 Plunge.configure :default @12:38:14.19 SamplePump.configure :default @12:38:14.35 ExtraPump.configure :default @12:38:14.55 Toroid.configure :default @12:38:14.63 Toroid.hold @12:38:14.71 Toroid.configure :default @12:38:15.39 Carousel.configure :default @12:38:15.45 Carousel.hold @12:38:15.54 Carousel.configure :default @12:38:15.70 Pressure.configure :default @12:38:15.78 All elf controllers running firmware version 4.61 @12:38:15.89 Intake.close @12:38:15.94 Exhaust.close @12:38:16.10 Bypass.close @12:38:16.27
Loading /home/esp/esp2/bin/processAll mission @12:38:16.62 Sampling default goal volumes @12:38:16.63 Cartridge group: dry Spares: {spare: 0} @12:38:17.45 Carousel.seek 45 @12:38:18.42 Mounted archive Cartridge @12:38:18.45 Selecting Cartridge 45 @12:38:19.42 Duration of filtering limited to 1:05:00 @12:38:19.89 Priming sample loop w/100ml, bypass w/10ml @12:38:19.90 SamplePump.setPosition! 0ml @12:38:19.95 Bypass.open @12:38:21.43 Intake.open @12:38:22.38 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:38:25.00 Vacuum tare is 17% of full scale @12:38:25.02 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:38:25.30 Pressure.reconfigure :default @12:38:25.36 Pressure.hold @12:38:25.63 SamplePump.reconfigure :prime @12:38:25.72 SamplePump.seek 10ml,1:21 @12:38:36.83 Pressure.coast @12:38:36.89 SamplePump.setPosition! 0ml @12:38:36.94 Exhaust.open @12:38:38.43 Bypass.close @12:38:40.01 Pressure.reconfigure :default @12:38:40.09 Pressure.hold @12:38:40.43 SamplePump.reconfigure :prime @12:38:40.52 SamplePump.seek 100ml,8:51 @12:40:27.90 Pressure.coast @12:40:27.96 Bypass.open @12:40:28.68 Intake.close @12:40:29.61 Toroid.reconfigure :hold @12:40:29.69 Toroid.seek :crack @12:40:31.29 Toroid.seek :cartridge @12:40:32.24 Toroid.reconfigure :default @12:40:32.31 Toroid.seek :clear @12:40:34.21 Pressurizing cartridge to 26psi above 14.0psia @12:40:34.23 SamplePump.setPosition! 0ml @12:40:34.29 Intake.open @12:40:35.09 Bypass.close @12:40:35.83 Exhaust.close @12:40:36.70 Pressure.reconfigure :default @12:40:36.79 Pressure.hold @12:40:37.06 SamplePump.reconfigure :charge @12:40:37.13 SamplePump.seek 2.3ml,42 seconds @12:40:43.98 Pressure.coast @12:40:44.03 Intake.close @12:40:44.90 Pressurized cartridge with 0.207ml to 26.2psi @12:40:44.91 Isolated @34.8psia (20.8psi) @12:40:44.92 Checking for cartridge leaks from 34.8psia @12:40:44.94 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:40:51.07 Cartridge pressure fell 5% to 32.9psia @12:40:51.19 Sampling 1000.0ml @12:40:51.21 SamplePump.setPosition! 0ml @12:40:51.25 Exhaust.open @12:40:51.97 Intake.open @12:40:52.95 Pressure.reconfigure :default @12:40:53.01 Pressure.hold @12:40:53.39 SamplePump.reconfigure :sample @12:40:53.48 SamplePump.seek 1000ml,1:23:51 @12:40:56.04 Cartridge 45 sampling 1/1000ml at 0.49ml/s with 21.5psi vacuum @12:41:56.01 Cartridge 45 sampling 30/1000ml at 0.47ml/s with 27.1psi vacuum @12:42:56.03 Cartridge 45 sampling 58/1000ml at 0.46ml/s with 25.3psi vacuum @12:43:56.02 Cartridge 45 sampling 85/1000ml at 0.43ml/s with 25.8psi vacuum @12:44:56.03 Cartridge 45 sampling 110/1000ml at 0.42ml/s with 22.5psi vacuum @12:45:56.03 Cartridge 45 sampling 133/1000ml at 0.37ml/s with 27.0psi vacuum @12:46:56.02 Cartridge 45 sampling 155/1000ml at 0.35ml/s with 23.4psi vacuum @12:47:56.02 Cartridge 45 sampling 175/1000ml at 0.33ml/s with 26.5psi vacuum @12:48:56.01 Cartridge 45 sampling 193/1000ml at 0.27ml/s with 26.7psi vacuum @12:49:56.02 Cartridge 45 sampling 210/1000ml at 0.29ml/s with 23.9psi vacuum @12:50:56.01 Cartridge 45 sampling 226/1000ml at 0.22ml/s with 18.7psi vacuum @12:51:56.02 Cartridge 45 sampling 240/1000ml at 0.22ml/s with 25.0psi vacuum @12:52:56.03 Cartridge 45 sampling 253/1000ml at 0.24ml/s with 25.6psi vacuum @12:53:56.02 Cartridge 45 sampling 266/1000ml at 0.21ml/s with 24.6psi vacuum @12:54:56.02 Cartridge 45 sampling 278/1000ml at 0.20ml/s with 28.8psi vacuum @12:55:56.01 Cartridge 45 sampling 290/1000ml at 0.20ml/s with 28.2psi vacuum @12:56:04.98 Flow averaged 0.32ml/s over 15:03 @12:56:05.04 Waiting up to 19 seconds for vacuum to reduce from 30.5 to 18.0psi @12:56:05.63 Relieved excess vacuum in 9.5 seconds @12:56:05.69 SamplePump.seek 1000ml,59:32 @12:56:19.35 Flow estimate reduced to 0.26ml/s @12:56:19.38 Waiting up to 21 seconds for vacuum to reduce from 31.1 to 18.0psi @12:56:20.46 Relieved excess vacuum in 9 seconds @12:56:20.49 SamplePump.seek 1000ml,59:18 @12:56:27.39 Flow estimate reduced to 0.21ml/s @12:56:27.43 Waiting up to 20 seconds for vacuum to reduce from 30.9 to 18.0psi @12:56:28.04 Relieved excess vacuum in 4.4 seconds @12:56:28.08 SamplePump.seek 1000ml,59:12 @12:56:34.17 Flow estimate reduced to 0.17ml/s @12:56:34.19 Intake.close @12:56:34.29 Exhaust.close @12:56:34.51 Toroid.reconfigure :hold @12:56:34.60 Toroid.seek :bypass @12:56:36.25 Toroid.reconfigure :default @12:56:36.31 Toroid.seek :clear @12:56:37.93 Cartridge::Sampler::Clogged in FILTERING -- Cartridge clogged @12:56:37.95 Pressure.coast @12:56:38.24 Sampled 297.0ml @12:56:38.82 Plunge.seek :reagent @12:56:39.72 Sample loop closed. Ready to Deliver Reagents @12:56:39.74 Archive.seek :sealed @12:56:43.41 Archive.seek :evac @12:56:57.42 delay 1:00 @12:57:57.46 Archive.seek :clear @12:58:14.62 Sample preserved. Puck Full. @12:58:14.92
Carousel.seek 45 @12:58:15.18 Plunge.seek :locked @12:58:16.29 Air.seek :clear @12:58:16.48 Lysis1.seek :clear @12:58:20.03
Archive.coast @12:58:20.14 Air.coast @12:58:20.23 Lysis1.coast @12:58:20.29 Lysis2.coast @12:58:20.35 Plunge.coast @12:58:20.39 SamplePump.coast @12:58:20.46 Toroid.coast @12:58:20.50 Carousel.coast @12:58:20.72 Gate.power :motor,:OFF @12:58:20.74 Gate.power :core,:OFF @12:58:20.75 Gate.power :analytic1,:OFF @12:58:20.76 Gate.power :analytic2,:OFF @12:58:20.77 Gate.power :analytic3,:OFF @12:58:20.79 Gate.power :main,:OFF @12:58:20.80 Safely stopped and ready to power off @12:58:21.19 Cartridge Load: {[43, 41] => { Type: :aRNAl_bac, State: :reserved, was: :dry}, [44, 42, 40..32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 297.0}} Spares: {spare: 0} @12:58:21.21 Cartridge::None in MAIN -- Not enough dry cartridges remain @13:25:43.88PST17-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @13:25:43.90 >BREAK< ESP3 Gateway v1.11 for cartridge bus @13:25:47.72 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @13:25:48.07 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @13:25:51.47 Polling Can every 2:00 @13:25:51.53 Loading /home/esp/esp2/bin/processAll mission @13:25:51.84 Sampling maximum of 250.0ml per cartridge @13:25:51.87 Cartridge group: dry Spares: {spare: 0} @13:25:52.05 Gate.power :main,:ON,limit to 5000mA @13:25:52.06 Gate.power :analytic1,:ON,limit to 5000mA @13:25:52.08 Gate.power :analytic2,:ON,limit to 5000mA @13:25:52.09 Gate.power :analytic3,:ON,limit to 5000mA @13:25:52.14 Powering core bus @13:25:54.50 Archive.configure :default @13:25:55.20 Air.configure :default @13:25:55.33 Lysis1.configure :default @13:25:55.47 Lysis2.configure :default @13:25:55.62 Plunge.configure :default @13:25:55.81 SamplePump.configure :default @13:25:56.01 ExtraPump.configure :default @13:25:56.18 Toroid.configure :default @13:25:56.34 Carousel.configure :default @13:25:56.54 Pressure.configure :default @13:25:56.60 All elf controllers running firmware version 4.61 @13:25:56.62 Gate.power :motor,:ON,limit to 5000mA @13:25:56.89 Air.reconfigure :default @13:25:56.96 Archive.reconfigure :default @13:25:57.38 Archive.home.jog 0 @13:25:57.43 Air.home.jog 0 @13:25:57.82 Plunge.reconfigure :default @13:25:57.87 Lysis2.reconfigure :default @13:25:57.95 Lysis1.reconfigure :default @13:25:57.98 Toroid.reconfigure :default @13:25:58.23 Plunge.home.jog 0 @13:25:58.27 Lysis2.home.jog 0 @13:25:58.29 Lysis1.home.jog 0 @13:25:58.37 Toroid.home.jog 0 @13:25:58.83 Air.reconfigure :homing @13:25:58.98 Archive.reconfigure :homing @13:25:59.17 Archive.jog 1374 @13:25:59.22 Air.jog 1363 @13:25:59.47 Toroid.reconfigure :homing @13:25:59.54 Plunge.reconfigure :homing @13:25:59.57 Lysis1.reconfigure :homing @13:25:59.64 Lysis2.reconfigure :homing @13:25:59.84 Plunge.jog 852 @13:25:59.98 Toroid.jog 607 @13:26:00.01 Lysis2.jog 1360 @13:26:00.04 Lysis1.jog 1348 @13:26:00.43 Archive.setRawPosition! 50 @13:26:00.47 Archive.seek :home @13:26:00.58 Air.setRawPosition! 50 @13:26:00.63 Air.seek :home @13:26:00.83 Archive.reconfigure :default @13:26:00.96 Plunge.setRawPosition! 50 @13:26:00.99 Air.reconfigure :default @13:26:01.11 Plunge.seek :home @13:26:01.37 Lysis1.setRawPosition! 50 @13:26:01.46 Lysis1.seek :home @13:26:01.59 Plunge.reconfigure :default @13:26:01.60 Lysis2.setRawPosition! 50 @13:26:01.73 Lysis2.seek :home @13:26:01.88 Lysis1.reconfigure :default @13:26:02.02 Lysis2.reconfigure :default @13:26:02.24 Toroid.setRawPosition! 0 @13:26:02.28 Toroid.seek :home @13:26:02.41 Toroid.reconfigure :default @13:26:02.58
Toroid.seek :clear @13:26:04.25 Carousel.reconfigure :default @13:26:04.30 Carousel.home.jog 0 @13:26:04.83 Carousel.seek 45 @13:26:05.05 Selecting Cartridge 44 @13:26:05.32 Carousel.seek 44 @13:26:05.88 Lysis1.seek :locked @13:26:09.13 Plunge.seek :locked @13:26:09.33 Air.seek :locked @13:26:13.61
Mounted archive Cartridge @13:26:13.99 Duration of filtering limited to 1:05:00 @13:26:14.34 Priming sample loop w/100ml, bypass w/10ml @13:26:14.37 SamplePump.setPosition! 0ml @13:26:14.41 Bypass.open @13:26:15.95 Intake.open @13:26:16.83 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @13:26:19.43 Vacuum tare is 17% of full scale @13:26:19.46 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @13:26:19.73 Pressure.reconfigure :default @13:26:19.80 Pressure.hold @13:26:20.10 SamplePump.reconfigure :prime @13:26:20.19 SamplePump.seek 10ml,1:21 @13:26:31.29 Pressure.coast @13:26:31.36 SamplePump.setPosition! 0ml @13:26:31.41 Exhaust.open @13:26:32.87 Bypass.close @13:26:34.47 Pressure.reconfigure :default @13:26:34.55 Pressure.hold @13:26:34.89 SamplePump.reconfigure :prime @13:26:34.98 SamplePump.seek 100ml,8:51 @13:28:22.36 Pressure.coast @13:28:22.43 Bypass.open @13:28:23.12 Intake.close @13:28:24.05 Toroid.reconfigure :hold @13:28:24.11 Toroid.seek :crack @13:28:25.77 Toroid.seek :cartridge @13:28:26.75 Toroid.reconfigure :default @13:28:26.81 Toroid.seek :clear @13:28:28.73 Pressurizing cartridge to 26psi above 14.2psia @13:28:28.75 SamplePump.setPosition! 0ml @13:28:28.79 Intake.open @13:28:29.63 Bypass.close @13:28:30.33 Exhaust.close @13:28:31.22 Pressure.reconfigure :default @13:28:31.31 Pressure.hold @13:28:31.56 SamplePump.reconfigure :charge @13:28:31.65 SamplePump.seek 2.3ml,42 seconds @13:28:37.94 Pressure.coast @13:28:37.99 Intake.close @13:28:38.88 Pressurized cartridge with 0.176ml to 26.1psi @13:28:38.93 Isolated @35.1psia (20.9psi) @13:28:38.94 Checking for cartridge leaks from 35.1psia @13:28:38.96 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @13:28:46.58 Cartridge pressure fell 8% to 32.2psia @13:28:46.69 Sampling 250.0ml @13:28:46.71 SamplePump.setPosition! 0ml @13:28:46.75 Exhaust.open @13:28:47.48 Intake.open @13:28:48.50 Pressure.reconfigure :default @13:28:48.58 Pressure.hold @13:28:48.94 SamplePump.reconfigure :sample @13:28:49.02 SamplePump.seek 250ml,21:21 @13:28:51.56 Cartridge 44 sampling 1/250ml at 0.50ml/s with 20.6psi vacuum @13:29:51.54 Cartridge 44 sampling 31/250ml at 0.50ml/s with 23.2psi vacuum @13:30:51.54 Cartridge 44 sampling 60/250ml at 0.48ml/s with 21.3psi vacuum @13:31:51.54 Cartridge 44 sampling 88/250ml at 0.45ml/s with 23.8psi vacuum @13:32:51.57 Cartridge 44 sampling 114/250ml at 0.42ml/s with 20.6psi vacuum @13:33:51.57 Cartridge 44 sampling 138/250ml at 0.38ml/s with 19.8psi vacuum @13:34:51.58 Cartridge 44 sampling 161/250ml at 0.34ml/s with 22.8psi vacuum @13:35:51.55 Cartridge 44 sampling 181/250ml at 0.33ml/s with 21.0psi vacuum @13:36:51.54 Cartridge 44 sampling 200/250ml at 0.31ml/s with 22.7psi vacuum @13:37:51.55 Cartridge 44 sampling 217/250ml at 0.29ml/s with 23.4psi vacuum @13:38:51.54 Cartridge 44 sampling 232/250ml at 0.26ml/s with 20.4psi vacuum @13:39:51.55 Cartridge 44 sampling 246/250ml at 0.22ml/s with 26.2psi vacuum @13:40:07.56 Pressure.coast @13:40:07.75 Intake.close @13:40:07.84 Exhaust.close @13:40:08.09 Toroid.reconfigure :hold @13:40:08.18 Toroid.seek :bypass @13:40:09.82 Toroid.reconfigure :default @13:40:09.88 Toroid.seek :clear @13:40:11.55 Sampled 250.0ml @13:40:12.15 Plunge.seek :reagent @13:40:13.10 Sample loop closed. Ready to Deliver Reagents @13:40:13.12 Archive.seek :sealed @13:40:16.75 Archive.seek :evac @13:40:30.76 delay 1:00 @13:41:30.80 Archive.seek :clear @13:41:48.00 RNAlater soak for 10:00 @13:51:48.04 Archive.seek :tVent @13:51:49.71 delay 1 second @13:51:50.74 Archive.seek :evac @13:52:06.71 delay 10 seconds @13:52:16.75 Sample preserved and evacuated @13:52:17.08
Carousel.seek 44 @13:52:17.32 Plunge.seek :locked @13:52:18.37 Archive.seek :clear @13:52:18.57 Air.seek :clear @13:52:18.77 Lysis1.seek :clear @13:52:35.61
Archive.coast @13:52:35.65 Air.coast @13:52:35.70 Lysis1.coast @13:52:35.77 Lysis2.coast @13:52:35.82 Plunge.coast @13:52:35.86 SamplePump.coast @13:52:35.93 Toroid.coast @13:52:35.97 Carousel.coast @13:52:36.22 Gate.power :motor,:OFF @13:52:36.24 Gate.power :core,:OFF @13:52:36.25 Gate.power :analytic1,:OFF @13:52:36.27 Gate.power :analytic2,:OFF @13:52:36.30 Gate.power :analytic3,:OFF @13:52:36.33 Gate.power :main,:OFF @13:52:36.34 Safely stopped and ready to power off @13:52:36.67 Gate.power :main,:ON,limit to 5000mA @13:52:36.70 Gate.power :analytic1,:ON,limit to 5000mA @13:52:36.72 Gate.power :analytic2,:ON,limit to 5000mA @13:52:36.74 Gate.power :analytic3,:ON,limit to 5000mA @13:52:37.28 Powering core bus @13:52:39.67 Archive.configure :default @13:52:39.83 Air.configure :default @13:52:39.97 Lysis1.configure :default @13:52:40.15 Lysis2.configure :default @13:52:40.32 Plunge.configure :default @13:52:40.49 SamplePump.configure :default @13:52:40.69 ExtraPump.configure :default @13:52:40.85 Toroid.configure :default @13:52:41.02 Carousel.configure :default @13:52:41.19 Pressure.configure :default @13:52:41.27 All elf controllers running firmware version 4.61 @13:52:41.29 Gate.power :motor,:ON,limit to 5000mA @13:52:41.70 Archive.reconfigure :default @13:52:41.92 Archive.home.jog 0 @13:52:42.20 Plunge.reconfigure :default @13:52:42.22 Lysis1.reconfigure :default @13:52:42.32 Toroid.reconfigure :default @13:52:42.38 Lysis2.reconfigure :default @13:52:42.44 Air.reconfigure :default @13:52:42.70 Plunge.home.jog 0 @13:52:42.77 Lysis1.home.jog 0 @13:52:42.80 Toroid.home.jog 0 @13:52:42.85 Lysis2.home.jog 0 @13:52:42.95 Air.home.jog 0 @13:52:43.48 Archive.reconfigure :homing @13:52:43.79 Archive.jog 1373 @13:52:43.88 Toroid.reconfigure :homing @13:52:43.90 Plunge.reconfigure :homing @13:52:43.93 Lysis2.reconfigure :homing @13:52:44.02 Air.reconfigure :homing @13:52:44.07 Lysis1.reconfigure :homing @13:52:44.36 Plunge.jog 848 @13:52:44.42 Toroid.jog 615 @13:52:44.45 Lysis2.jog 1358 @13:52:44.48 Air.jog 1366 @13:52:44.53 Lysis1.jog 1346 @13:52:45.27 Archive.setRawPosition! 50 @13:52:45.32 Archive.seek :home @13:52:45.47 Plunge.setRawPosition! 50 @13:52:45.54 Plunge.seek :home @13:52:45.67 Archive.reconfigure :default @13:52:45.83 Lysis2.setRawPosition! 50 @13:52:45.91 Lysis2.seek :home @13:52:46.02 Lysis1.setRawPosition! 50 @13:52:46.07 Air.setRawPosition! 50 @13:52:46.08 Plunge.reconfigure :default @13:52:46.25 Lysis1.seek :home @13:52:46.33 Air.seek :home @13:52:46.60 Lysis2.reconfigure :default @13:52:46.78 Air.reconfigure :default @13:52:46.82 Lysis1.reconfigure :default @13:52:47.02 Toroid.setRawPosition! 0 @13:52:47.06 Toroid.seek :home @13:52:47.20 Toroid.reconfigure :default @13:52:47.41
Toroid.seek :clear @13:52:49.05 Carousel.reconfigure :default @13:52:49.83 Carousel.home.jog 0 @13:52:50.39 Carousel.seek 44 @13:52:50.59 Selecting Cartridge 43 @13:52:50.86 Carousel.seek 43 @13:52:51.46 Lysis1.seek :locked @13:52:54.71 Plunge.seek :locked @13:52:54.92 Air.seek :locked @13:52:59.19
Mounted archive Cartridge @13:52:59.57 Duration of filtering limited to 1:05:00 @13:53:00.03 Priming sample loop w/100ml, bypass w/10ml @13:53:00.06 SamplePump.setPosition! 0ml @13:53:00.15 Bypass.open @13:53:01.72 Intake.open @13:53:02.66 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @13:53:05.26 Vacuum tare is 17% of full scale @13:53:05.29 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @13:53:05.57 Pressure.reconfigure :default @13:53:05.63 Pressure.hold @13:53:05.91 SamplePump.reconfigure :prime @13:53:06.00 SamplePump.seek 10ml,1:21 @13:53:17.12 Pressure.coast @13:53:17.17 SamplePump.setPosition! 0ml @13:53:17.26 Exhaust.open @13:53:18.69 Bypass.close @13:53:20.35 Pressure.reconfigure :default @13:53:20.43 Pressure.hold @13:53:20.80 SamplePump.reconfigure :prime @13:53:20.86 SamplePump.seek 100ml,8:51 @13:55:08.23 Pressure.coast @13:55:08.33 Bypass.open @13:55:09.06 Intake.close @13:55:10.01 Toroid.reconfigure :hold @13:55:10.08 Toroid.seek :crack @13:55:11.74 Toroid.seek :cartridge @13:55:12.70 Toroid.reconfigure :default @13:55:12.76 Toroid.seek :clear @13:55:14.69 Pressurizing cartridge to 26psi above 14.3psia @13:55:14.72 SamplePump.setPosition! 0ml @13:55:14.75 Intake.open @13:55:15.56 Bypass.close @13:55:16.32 Exhaust.close @13:55:17.21 Pressure.reconfigure :default @13:55:17.29 Pressure.hold @13:55:17.57 SamplePump.reconfigure :charge @13:55:17.64 SamplePump.seek 2.3ml,42 seconds @13:55:23.62 Pressure.coast @13:55:23.70 Intake.close @13:55:24.59 Pressurized cartridge with 0.161ml to 26.2psi @13:55:24.60 Isolated @34.3psia (20.0psi) @13:55:24.61 Checking for cartridge leaks from 34.3psia @13:55:24.63 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @13:55:36.22 Cartridge pressure fell 11% to 30.7psia @13:55:36.34 Sampling 250.0ml @13:55:36.36 SamplePump.setPosition! 0ml @13:55:36.44 Exhaust.open @13:55:37.11 Intake.open @13:55:38.13 Pressure.reconfigure :default @13:55:38.21 Pressure.hold @13:55:38.55 SamplePump.reconfigure :sample @13:55:38.64 SamplePump.seek 250ml,21:21 @13:55:41.20 Cartridge 43 sampling 1/250ml at 0.52ml/s with 25.9psi vacuum @13:56:41.20 Cartridge 43 sampling 31/250ml at 0.48ml/s with 23.8psi vacuum @13:57:41.21 Cartridge 43 sampling 61/250ml at 0.48ml/s with 25.9psi vacuum @13:58:41.21 Cartridge 43 sampling 89/250ml at 0.44ml/s with 20.1psi vacuum @13:59:41.22 Cartridge 43 sampling 115/250ml at 0.43ml/s with 20.5psi vacuum @14:00:41.20 Cartridge 43 sampling 140/250ml at 0.38ml/s with 19.0psi vacuum @14:01:41.24 Cartridge 43 sampling 163/250ml at 0.36ml/s with 24.8psi vacuum @14:02:41.22 Cartridge 43 sampling 183/250ml at 0.33ml/s with 25.9psi vacuum @14:03:41.22 Cartridge 43 sampling 202/250ml at 0.30ml/s with 26.5psi vacuum @14:04:41.21 Cartridge 43 sampling 220/250ml at 0.28ml/s with 19.6psi vacuum @14:05:41.22 Cartridge 43 sampling 235/250ml at 0.24ml/s with 27.4psi vacuum @14:06:41.20 Cartridge 43 sampling 249/250ml at 0.21ml/s with 19.6psi vacuum @14:06:43.59 Pressure.coast @14:06:43.75 Intake.close @14:06:43.85 Exhaust.close @14:06:44.04 Toroid.reconfigure :hold @14:06:44.12 Toroid.seek :bypass @14:06:45.78 Toroid.reconfigure :default @14:06:45.85 Toroid.seek :clear @14:06:47.48 Sampled 250.0ml @14:06:48.09 Plunge.seek :reagent @14:06:49.01 Sample loop closed. Ready to Deliver Reagents @14:06:49.03 Archive.seek :sealed @14:06:52.68 Archive.seek :evac @14:07:06.70 delay 1:00 @14:08:06.73 Archive.seek :clear @14:08:23.89 Sample preserved. Puck Full. @14:08:24.86
Carousel.seek 43 @14:08:25.12 Plunge.seek :locked @14:08:26.22 Air.seek :clear @14:08:26.43 Lysis1.seek :clear @14:08:29.93
Archive.coast @14:08:29.97 Air.coast @14:08:30.04 Lysis1.coast @14:08:30.11 Lysis2.coast @14:08:30.15 Plunge.coast @14:08:30.22 SamplePump.coast @14:08:30.26 Toroid.coast @14:08:30.34 Carousel.coast @14:08:30.56 Gate.power :motor,:OFF @14:08:30.58 Gate.power :core,:OFF @14:08:30.61 Gate.power :analytic1,:OFF @14:08:30.63 Gate.power :analytic2,:OFF @14:08:30.64 Gate.power :analytic3,:OFF @14:08:30.66 Gate.power :main,:OFF @14:08:30.67 Safely stopped and ready to power off @14:08:31.07 Gate.power :main,:ON,limit to 5000mA @14:08:31.08 Gate.power :analytic1,:ON,limit to 5000mA @14:08:31.09 Gate.power :analytic2,:ON,limit to 5000mA @14:08:31.11 Gate.power :analytic3,:ON,limit to 5000mA @14:08:31.69 Powering core bus @14:08:34.06 Archive.configure :default @14:08:34.25 Air.configure :default @14:08:34.39 Lysis1.configure :default @14:08:34.54 Lysis2.configure :default @14:08:34.71 Plunge.configure :default @14:08:34.87 SamplePump.configure :default @14:08:35.05 ExtraPump.configure :default @14:08:35.25 Toroid.configure :default @14:08:35.42 Carousel.configure :default @14:08:35.62 Pressure.configure :default @14:08:35.68 All elf controllers running firmware version 4.61 @14:08:35.70 Gate.power :motor,:ON,limit to 5000mA @14:08:36.12 Archive.reconfigure :default @14:08:36.43 Archive.home.jog 0 @14:08:36.77 Toroid.reconfigure :default @14:08:36.80 Lysis2.reconfigure :default @14:08:36.89 Air.reconfigure :default @14:08:36.96 Lysis1.reconfigure :default @14:08:37.02 Plunge.reconfigure :default @14:08:37.27 Toroid.home.jog 0 @14:08:37.39 Air.home.jog 0 @14:08:37.47 Lysis2.home.jog 0 @14:08:37.56 Lysis1.home.jog 0 @14:08:37.65 Plunge.home.jog 0 @14:08:38.38 Archive.reconfigure :homing @14:08:38.64 Archive.jog 1372 @14:08:38.71 Toroid.reconfigure :homing @14:08:38.75 Plunge.reconfigure :homing @14:08:38.78 Lysis2.reconfigure :homing @14:08:38.85 Air.reconfigure :homing @14:08:38.91 Lysis1.reconfigure :homing @14:08:39.21 Plunge.jog 848 @14:08:39.28 Lysis2.jog 1354 @14:08:39.31 Toroid.jog 615 @14:08:39.35 Air.jog 1363 @14:08:39.40 Lysis1.jog 1349 @14:08:40.14 Archive.setRawPosition! 50 @14:08:40.22 Archive.seek :home @14:08:40.35 Plunge.setRawPosition! 50 @14:08:40.39 Plunge.seek :home @14:08:40.51 Archive.reconfigure :default @14:08:40.66 Lysis2.setRawPosition! 50 @14:08:40.70 Lysis2.seek :home @14:08:40.85 Lysis1.setRawPosition! 50 @14:08:40.91 Air.setRawPosition! 50 @14:08:40.94 Plunge.reconfigure :default @14:08:41.11 Lysis1.seek :home @14:08:41.25 Air.seek :home @14:08:41.40 Lysis2.reconfigure :default @14:08:41.54 Lysis1.reconfigure :default @14:08:41.69 Toroid.setRawPosition! 0 @14:08:41.74 Air.reconfigure :default @14:08:41.85 Toroid.seek :home @14:08:42.00 Toroid.reconfigure :default @14:08:42.23
Toroid.seek :clear @14:08:43.89 Carousel.reconfigure :default @14:08:43.94 Carousel.home.jog 0 @14:08:44.48 Carousel.seek 43 @14:08:44.66 Selecting Cartridge 42 @14:08:44.93 Carousel.seek 42 @14:08:45.52 Lysis1.seek :locked @14:08:48.80 Plunge.seek :locked @14:08:48.97 Air.seek :locked @14:08:53.27
Mounted archive Cartridge @14:08:53.63 Duration of filtering limited to 1:05:00 @14:08:54.02 Priming sample loop w/100ml, bypass w/10ml @14:08:54.03 SamplePump.setPosition! 0ml @14:08:54.07 Bypass.open @14:08:55.58 Intake.open @14:08:56.48 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @14:08:59.09 Vacuum tare is 17% of full scale @14:08:59.11 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @14:08:59.39 Pressure.reconfigure :default @14:08:59.45 Pressure.hold @14:08:59.72 SamplePump.reconfigure :prime @14:08:59.82 SamplePump.seek 10ml,1:21 @14:09:10.95 Pressure.coast @14:09:11.01 SamplePump.setPosition! 0ml @14:09:11.06 Exhaust.open @14:09:12.50 Bypass.close @14:09:14.07 Pressure.reconfigure :default @14:09:14.16 Pressure.hold @14:09:14.52 SamplePump.reconfigure :prime @14:09:14.58 SamplePump.seek 100ml,8:51 @14:11:01.99 Pressure.coast @14:11:02.07 Bypass.open @14:11:02.78 Intake.close @14:11:03.69 Toroid.reconfigure :hold @14:11:03.74 Toroid.seek :crack @14:11:05.39 Toroid.seek :cartridge @14:11:06.32 Toroid.reconfigure :default @14:11:06.42 Toroid.seek :clear @14:11:08.99 Pressurizing cartridge to 26psi above 14.7psia @14:11:09.01 SamplePump.setPosition! 0ml @14:11:09.08 Intake.open @14:11:09.86 Bypass.close @14:11:10.60 Exhaust.close @14:11:11.47 Pressure.reconfigure :default @14:11:11.53 Pressure.hold @14:11:11.82 SamplePump.reconfigure :charge @14:11:11.90 SamplePump.seek 2.3ml,42 seconds @14:11:17.95 Pressure.coast @14:11:18.00 Intake.close @14:11:18.94 Pressurized cartridge with 0.164ml to 26.2psi @14:11:18.96 Isolated @35.3psia (20.6psi) @14:11:18.97 Checking for cartridge leaks from 35.3psia @14:11:18.99 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @14:11:29.08 Cartridge pressure fell 10% to 31.8psia @14:11:29.20 Sampling 250.0ml @14:11:29.21 SamplePump.setPosition! 0ml @14:11:29.29 Exhaust.open @14:11:29.99 Intake.open @14:11:31.03 Pressure.reconfigure :default @14:11:31.09 Pressure.hold @14:11:31.47 SamplePump.reconfigure :sample @14:11:31.53 SamplePump.seek 250ml,21:21 @14:11:34.09 Cartridge 42 sampling 1/250ml at 0.50ml/s with 24.9psi vacuum @14:12:34.10 Cartridge 42 sampling 30/250ml at 0.45ml/s with 18.7psi vacuum @14:13:34.10 Cartridge 42 sampling 57/250ml at 0.46ml/s with 24.9psi vacuum @14:14:34.11 Cartridge 42 sampling 83/250ml at 0.44ml/s with 26.0psi vacuum @14:15:34.10 Cartridge 42 sampling 108/250ml at 0.40ml/s with 23.6psi vacuum @14:16:34.10 Cartridge 42 sampling 131/250ml at 0.38ml/s with 22.1psi vacuum @14:17:34.10 Cartridge 42 sampling 152/250ml at 0.36ml/s with 22.9psi vacuum @14:18:34.10 Cartridge 42 sampling 171/250ml at 0.30ml/s with 25.6psi vacuum @14:19:34.11 Cartridge 42 sampling 189/250ml at 0.26ml/s with 27.5psi vacuum @14:20:34.09 Cartridge 42 sampling 205/250ml at 0.28ml/s with 22.7psi vacuum @14:21:34.12 Cartridge 42 sampling 220/250ml at 0.23ml/s with 18.2psi vacuum @14:22:34.09 Cartridge 42 sampling 234/250ml at 0.25ml/s with 25.2psi vacuum @14:23:34.10 Cartridge 42 sampling 247/250ml at 0.20ml/s with 25.3psi vacuum @14:23:48.79 Pressure.coast @14:23:48.99 Intake.close @14:23:49.08 Exhaust.close @14:23:49.29 Toroid.reconfigure :hold @14:23:49.35 Toroid.seek :bypass @14:23:51.03 Toroid.reconfigure :default @14:23:51.08 Toroid.seek :clear @14:23:52.71 Sampled 250.0ml @14:23:53.34 Plunge.seek :reagent @14:23:54.24 Sample loop closed. Ready to Deliver Reagents @14:23:54.26 Archive.seek :sealed @14:23:57.94 Archive.seek :evac @14:24:11.94 delay 1:00 @14:25:11.98 Archive.seek :clear @14:25:29.16 RNAlater soak for 10:00 @14:35:29.19 Archive.seek :tVent @14:35:30.84 delay 1 second @14:35:31.89 Archive.seek :evac @14:35:47.85 delay 10 seconds @14:35:57.89 Sample preserved and evacuated @14:35:58.20
Carousel.seek 42 @14:35:58.45 Plunge.seek :locked @14:35:59.48 Archive.seek :clear @14:35:59.66 Air.seek :clear @14:35:59.88 Lysis1.seek :clear @14:36:16.69
Archive.coast @14:36:16.74 Air.coast @14:36:16.81 Lysis1.coast @14:36:16.85 Lysis2.coast @14:36:16.91 Plunge.coast @14:36:16.97 SamplePump.coast @14:36:17.01 Toroid.coast @14:36:17.08 Carousel.coast @14:36:17.31 Gate.power :motor,:OFF @14:36:17.34 Gate.power :core,:OFF @14:36:17.35 Gate.power :analytic1,:OFF @14:36:17.36 Gate.power :analytic2,:OFF @14:36:17.38 Gate.power :analytic3,:OFF @14:36:17.40 Gate.power :main,:OFF @14:36:17.41 Safely stopped and ready to power off @14:36:17.77 Gate.power :main,:ON,limit to 5000mA @14:36:17.78 Gate.power :analytic1,:ON,limit to 5000mA @14:36:17.79 Gate.power :analytic2,:ON,limit to 5000mA @14:36:17.81 Gate.power :analytic3,:ON,limit to 5000mA @14:36:18.39 Powering core bus @14:36:20.81 Archive.configure :default @14:36:20.98 Air.configure :default @14:36:21.12 Lysis1.configure :default @14:36:21.25 Lysis2.configure :default @14:36:21.42 Plunge.configure :default @14:36:21.60 SamplePump.configure :default @14:36:21.78 ExtraPump.configure :default @14:36:21.96 Toroid.configure :default @14:36:22.11 Carousel.configure :default @14:36:22.30 Pressure.configure :default @14:36:22.35 All elf controllers running firmware version 4.61 @14:36:22.37 Gate.power :motor,:ON,limit to 5000mA @14:36:22.73 Archive.reconfigure :default @14:36:23.13 Archive.home.jog 0 @14:36:23.20 Plunge.reconfigure :default @14:36:23.22 Lysis2.reconfigure :default @14:36:23.98 Air.reconfigure :default @14:36:24.07 Toroid.reconfigure :default @14:36:24.15 Lysis1.reconfigure :default @14:36:24.43 Plunge.home.jog 0 @14:36:24.49 Lysis2.home.jog 0 @14:36:24.54 Toroid.home.jog 0 @14:36:24.56 Air.home.jog 0 @14:36:24.59 Lysis1.home.jog 0 @14:36:25.43 Plunge.reconfigure :homing @14:36:25.45 Lysis2.reconfigure :homing @14:36:25.56 Toroid.reconfigure :homing @14:36:25.65 Archive.reconfigure :homing @14:36:25.79 Lysis2.jog 1357 @14:36:25.96 Toroid.jog 615 @14:36:26.02 Plunge.jog 849 @14:36:26.05 Archive.jog 1373 @14:36:26.34 Lysis1.reconfigure :homing @14:36:26.36 Air.reconfigure :homing @14:36:26.53 Air.jog 1365 @14:36:26.60 Lysis1.jog 1348 @14:36:27.08 Lysis2.setRawPosition! 50 @14:36:27.13 Lysis2.seek :home @14:36:27.24 Plunge.setRawPosition! 50 @14:36:27.32 Plunge.seek :home @14:36:27.45 Archive.setRawPosition! 50 @14:36:27.51 Archive.seek :home @14:36:27.64 Lysis2.reconfigure :default @14:36:27.68 Plunge.reconfigure :default @14:36:27.91 Air.setRawPosition! 50 @14:36:27.98 Archive.reconfigure :default @14:36:28.00 Lysis1.setRawPosition! 50 @14:36:28.16 Air.seek :home @14:36:28.23 Lysis1.seek :home @14:36:28.42 Toroid.setRawPosition! 0 @14:36:28.51 Toroid.seek :home @14:36:28.70 Lysis1.reconfigure :default @14:36:28.73 Air.reconfigure :default @14:36:28.77 Toroid.reconfigure :default @14:36:29.36
Toroid.seek :clear @14:36:30.97 Carousel.reconfigure :default @14:36:31.03 Carousel.home.jog 0 @14:36:31.62 Carousel.seek 42 @14:36:31.79 Selecting Cartridge 41 @14:36:32.06 Carousel.seek 41 @14:36:32.67 Lysis1.seek :locked @14:36:35.91 Plunge.seek :locked @14:36:36.12 Air.seek :locked @14:36:40.39
Mounted archive Cartridge @14:36:40.76 Duration of filtering limited to 1:05:00 @14:36:41.16 Priming sample loop w/100ml, bypass w/10ml @14:36:41.17 SamplePump.setPosition! 0ml @14:36:41.21 Bypass.open @14:36:42.77 Intake.open @14:36:43.71 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @14:36:46.33 Vacuum tare is 17% of full scale @14:36:46.37 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @14:36:46.64 Pressure.reconfigure :default @14:36:46.71 Pressure.hold @14:36:46.97 SamplePump.reconfigure :prime @14:36:47.07 SamplePump.seek 10ml,1:21 @14:36:58.21 Pressure.coast @14:36:58.26 SamplePump.setPosition! 0ml @14:36:58.33 Exhaust.open @14:36:59.76 Bypass.close @14:37:01.37 Pressure.reconfigure :default @14:37:01.43 Pressure.hold @14:37:01.79 SamplePump.reconfigure :prime @14:37:01.88 SamplePump.seek 100ml,8:51 @14:38:49.25 Pressure.coast @14:38:49.33 Bypass.open @14:38:50.03 Intake.close @14:38:51.02 Toroid.reconfigure :hold @14:38:51.16 Toroid.seek :crack @14:38:52.78 Toroid.seek :cartridge @14:38:53.72 Toroid.reconfigure :default @14:38:53.77 Toroid.seek :clear @14:38:55.71 Pressurizing cartridge to 26psi above 14.3psia @14:38:55.73 SamplePump.setPosition! 0ml @14:38:55.77 Intake.open @14:38:56.54 Bypass.close @14:38:57.26 Exhaust.close @14:38:58.15 Pressure.reconfigure :default @14:38:58.21 Pressure.hold @14:38:58.46 SamplePump.reconfigure :charge @14:38:58.56 SamplePump.seek 2.3ml,42 seconds @14:39:10.00 Pressure.coast @14:39:10.06 Intake.close @14:39:10.94 Pressurized cartridge with 0.533ml to 26.4psi @14:39:10.97 Isolated @36.9psia (22.6psi) @14:39:10.99 Checking for cartridge leaks from 36.9psia @14:39:11.02 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @14:39:15.60 Cartridge pressure fell 1% to 36.4psia @14:39:15.70 Sampling 250.0ml @14:39:15.72 SamplePump.setPosition! 0ml @14:39:15.79 Exhaust.open @14:39:16.51 Intake.open @14:39:17.49 Pressure.reconfigure :default @14:39:17.57 Pressure.hold @14:39:17.94 SamplePump.reconfigure :sample @14:39:18.00 SamplePump.seek 250ml,21:21 @14:39:20.57 Cartridge 41 sampling 1/250ml at 0.50ml/s with 25.4psi vacuum @14:40:20.57 Cartridge 41 sampling 31/250ml at 0.49ml/s with 21.3psi vacuum @14:41:20.57 Cartridge 41 sampling 59/250ml at 0.46ml/s with 24.6psi vacuum @14:42:20.56 Cartridge 41 sampling 87/250ml at 0.43ml/s with 22.7psi vacuum @14:43:20.57 Cartridge 41 sampling 112/250ml at 0.41ml/s with 22.7psi vacuum @14:44:20.59 Cartridge 41 sampling 136/250ml at 0.36ml/s with 25.9psi vacuum @14:45:20.58 Cartridge 41 sampling 158/250ml at 0.35ml/s with 26.4psi vacuum @14:46:20.56 Cartridge 41 sampling 178/250ml at 0.33ml/s with 25.5psi vacuum @14:47:20.57 Cartridge 41 sampling 197/250ml at 0.28ml/s with 25.8psi vacuum @14:48:20.57 Cartridge 41 sampling 214/250ml at 0.28ml/s with 24.6psi vacuum @14:49:20.58 Cartridge 41 sampling 230/250ml at 0.26ml/s with 24.6psi vacuum @14:50:20.57 Cartridge 41 sampling 244/250ml at 0.22ml/s with 26.4psi vacuum @14:50:46.20 Pressure.coast @14:50:46.38 Intake.close @14:50:46.43 Exhaust.close @14:50:46.67 Toroid.reconfigure :hold @14:50:46.78 Toroid.seek :bypass @14:50:48.40 Toroid.reconfigure :default @14:50:48.48 Toroid.seek :clear @14:50:50.10 Sampled 250.0ml @14:50:50.68 Plunge.seek :reagent @14:50:51.58 Sample loop closed. Ready to Deliver Reagents @14:50:51.60 Archive.seek :sealed @14:50:55.28 Archive.seek :evac @14:51:09.25 delay 1:00 @14:52:09.28 Archive.seek :clear @14:52:26.44 Sample preserved. Puck Full. @14:52:26.74
Carousel.seek 41 @14:52:26.98 Plunge.seek :locked @14:52:28.08 Air.seek :clear @14:52:28.30 Lysis1.seek :clear @14:52:31.78
Archive.coast @14:52:31.82 Air.coast @14:52:31.86 Lysis1.coast @14:52:31.94 Lysis2.coast @14:52:31.98 Plunge.coast @14:52:32.05 SamplePump.coast @14:52:32.10 Toroid.coast @14:52:32.15 Carousel.coast @14:52:32.41 Gate.power :motor,:OFF @14:52:32.42 Gate.power :core,:OFF @14:52:32.44 Gate.power :analytic1,:OFF @14:52:32.47 Gate.power :analytic2,:OFF @14:52:32.49 Gate.power :analytic3,:OFF @14:52:32.50 Gate.power :main,:OFF @14:52:32.51 Safely stopped and ready to power off @14:52:32.91 Gate.power :main,:ON,limit to 5000mA @14:52:32.92 Gate.power :analytic1,:ON,limit to 5000mA @14:52:32.94 Gate.power :analytic2,:ON,limit to 5000mA @14:52:32.95 Gate.power :analytic3,:ON,limit to 5000mA @14:52:33.51 Powering core bus @14:52:35.87 Archive.configure :default @14:52:36.04 Air.configure :default @14:52:36.18 Lysis1.configure :default @14:52:36.33 Lysis2.configure :default @14:52:36.51 Plunge.configure :default @14:52:36.70 SamplePump.configure :default @14:52:36.87 ExtraPump.configure :default @14:52:37.06 Toroid.configure :default @14:52:37.24 Carousel.configure :default @14:52:37.40 Pressure.configure :default @14:52:37.48 All elf controllers running firmware version 4.61 @14:52:37.50 Gate.power :motor,:ON,limit to 5000mA @14:52:38.28 Air.reconfigure :default @14:52:38.30 Toroid.reconfigure :default @14:52:38.39 Plunge.reconfigure :default @14:52:38.43 Lysis1.reconfigure :default @14:52:38.48 Archive.reconfigure :default @14:52:38.55 Lysis2.reconfigure :default @14:52:38.83 Toroid.home.jog 0 @14:52:38.88 Air.home.jog 0 @14:52:38.94 Plunge.home.jog 0 @14:52:38.96 Lysis1.home.jog 0 @14:52:38.98 Archive.home.jog 0 @14:52:39.04 Lysis2.home.jog 0 @14:52:40.54 Toroid.reconfigure :homing @14:52:40.59 Plunge.reconfigure :homing @14:52:40.66 Air.reconfigure :homing @14:52:40.72 Lysis1.reconfigure :homing @14:52:40.96 Plunge.jog 849 @14:52:41.03 Toroid.jog 614 @14:52:41.09 Air.jog 1364 @14:52:41.19 Lysis1.jog 1348 @14:52:41.46 Lysis2.reconfigure :homing @14:52:41.49 Archive.reconfigure :homing @14:52:41.62 Lysis2.jog 1355 @14:52:41.66 Archive.jog 1371 @14:52:42.02 Plunge.setRawPosition! 50 @14:52:42.10 Plunge.seek :home @14:52:42.40 Air.setRawPosition! 50 @14:52:42.46 Plunge.reconfigure :default @14:52:42.53 Air.seek :home @14:52:42.73 Lysis1.setRawPosition! 50 @14:52:42.79 Lysis1.seek :home @14:52:42.97 Lysis2.setRawPosition! 50 @14:52:43.03 Archive.setRawPosition! 50 @14:52:43.05 Air.reconfigure :default @14:52:43.24 Lysis2.seek :home @14:52:43.32 Archive.seek :home @14:52:43.58 Toroid.setRawPosition! 0 @14:52:43.61 Lysis1.reconfigure :default @14:52:43.73 Toroid.seek :home @14:52:43.92 Lysis2.reconfigure :default @14:52:43.96 Archive.reconfigure :default @14:52:44.01 Toroid.reconfigure :default @14:52:44.57
Toroid.seek :clear @14:52:46.18 Carousel.reconfigure :default @14:52:46.25 Carousel.home.jog 0 @14:52:46.79 Carousel.seek 41 @14:52:47.00 Selecting Cartridge 40 @14:52:47.26 Carousel.seek 40 @14:52:47.84 Lysis1.seek :locked @14:52:51.11 Plunge.seek :locked @14:52:51.30 Air.seek :locked @14:52:55.57
Mounted archive Cartridge @14:52:55.94 Duration of filtering limited to 1:05:00 @14:52:56.33 Priming sample loop w/100ml, bypass w/10ml @14:52:56.35 SamplePump.setPosition! 0ml @14:52:56.38 Bypass.open @14:52:57.90 Intake.open @14:52:58.78 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @14:53:01.40 Vacuum tare is 17% of full scale @14:53:01.42 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @14:53:01.69 Pressure.reconfigure :default @14:53:01.76 Pressure.hold @14:53:02.01 SamplePump.reconfigure :prime @14:53:02.11 SamplePump.seek 10ml,1:21 @14:53:13.24 Pressure.coast @14:53:13.32 SamplePump.setPosition! 0ml @14:53:13.37 Exhaust.open @14:53:14.80 Bypass.close @14:53:16.45 Pressure.reconfigure :default @14:53:16.50 Pressure.hold @14:53:16.89 SamplePump.reconfigure :prime @14:53:16.99 SamplePump.seek 100ml,8:51 @14:55:04.40 Pressure.coast @14:55:04.48 Bypass.open @14:55:05.23 Intake.close @14:55:06.14 Toroid.reconfigure :hold @14:55:06.23 Toroid.seek :crack @14:55:07.83 Toroid.seek :cartridge @14:55:08.79 Toroid.reconfigure :default @14:55:08.84 Toroid.seek :clear @14:55:10.80 Pressurizing cartridge to 26psi above 14.3psia @14:55:10.82 SamplePump.setPosition! 0ml @14:55:10.85 Intake.open @14:55:11.66 Bypass.close @14:55:12.43 Exhaust.close @14:55:13.32 Pressure.reconfigure :default @14:55:13.41 Pressure.hold @14:55:13.72 SamplePump.reconfigure :charge @14:55:13.80 SamplePump.seek 2.3ml,42 seconds @14:55:19.93 Pressure.coast @14:55:19.98 Intake.close @14:55:20.91 Pressurized cartridge with 0.167ml to 26.1psi @14:55:20.92 Isolated @34.7psia (20.4psi) @14:55:20.93 Checking for cartridge leaks from 34.7psia @14:55:20.95 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @14:55:29.54 Cartridge pressure fell 8% to 32.1psia @14:55:29.66 Sampling 250.0ml @14:55:29.68 SamplePump.setPosition! 0ml @14:55:29.75 Exhaust.open @14:55:30.47 Intake.open @14:55:31.44 Pressure.reconfigure :default @14:55:31.53 Pressure.hold @14:55:31.90 SamplePump.reconfigure :sample @14:55:31.97 SamplePump.seek 250ml,21:21 @14:55:34.50 Cartridge 40 sampling 1/250ml at 0.50ml/s with 21.2psi vacuum @14:56:34.53 Cartridge 40 sampling 30/250ml at 0.47ml/s with 25.6psi vacuum @14:57:34.52 Cartridge 40 sampling 58/250ml at 0.46ml/s with 21.3psi vacuum @14:58:34.52 Cartridge 40 sampling 85/250ml at 0.43ml/s with 20.4psi vacuum @14:59:34.52 Cartridge 40 sampling 111/250ml at 0.41ml/s with 27.1psi vacuum @15:00:34.52 Cartridge 40 sampling 135/250ml at 0.38ml/s with 26.1psi vacuum @15:01:34.51 Cartridge 40 sampling 156/250ml at 0.34ml/s with 28.0psi vacuum @15:02:34.51 Cartridge 40 sampling 177/250ml at 0.31ml/s with 19.1psi vacuum @15:03:34.51 Cartridge 40 sampling 195/250ml at 0.29ml/s with 26.5psi vacuum @15:04:34.53 Cartridge 40 sampling 212/250ml at 0.26ml/s with 21.2psi vacuum @15:05:34.51 Cartridge 40 sampling 227/250ml at 0.24ml/s with 18.6psi vacuum @15:06:34.52 Cartridge 40 sampling 241/250ml at 0.20ml/s with 24.6psi vacuum @15:07:15.29 Pressure.coast @15:07:15.46 Intake.close @15:07:15.54 Exhaust.close @15:07:15.75 Toroid.reconfigure :hold @15:07:15.85 Toroid.seek :bypass @15:07:17.50 Toroid.reconfigure :default @15:07:17.56 Toroid.seek :clear @15:07:19.18 Sampled 250.0ml @15:07:19.78 Plunge.seek :reagent @15:07:20.68 Sample loop closed. Ready to Deliver Reagents @15:07:20.70 Archive.seek :sealed @15:07:24.35 Archive.seek :evac @15:07:38.35 delay 1:00 @15:08:38.38 Archive.seek :clear @15:08:55.55 RNAlater soak for 10:00 @15:18:55.58 Archive.seek :tVent @15:18:57.26 delay 1 second @15:18:58.30 Archive.seek :evac @15:19:14.30 delay 10 seconds @15:19:24.34 Sample preserved and evacuated @15:19:24.65
Carousel.seek 40 @15:19:24.91 Plunge.seek :locked @15:19:25.90 Archive.seek :clear @15:19:26.11 Air.seek :clear @15:19:26.33 Lysis1.seek :clear @15:19:43.13
Archive.coast @15:19:43.17 Air.coast @15:19:43.23 Lysis1.coast @15:19:43.28 Lysis2.coast @15:19:43.34 Plunge.coast @15:19:43.41 SamplePump.coast @15:19:43.48 Toroid.coast @15:19:43.53 Carousel.coast @15:19:43.75 Gate.power :motor,:OFF @15:19:43.76 Gate.power :core,:OFF @15:19:43.78 Gate.power :analytic1,:OFF @15:19:43.79 Gate.power :analytic2,:OFF @15:19:43.80 Gate.power :analytic3,:OFF @15:19:43.82 Gate.power :main,:OFF @15:19:43.83 Safely stopped and ready to power off @15:19:44.19 Cartridge Load: {45 => { Type: :aRNAl_bac, State: :processed, filtered: 297}, [39..32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 250.0}, [44, 42, 40] => { Type: :archive_bac, State: :processed, filtered: 250.0}} Spares: {spare: 0} @15:19:44.21 Cartridge::None in MAIN -- Not enough dry cartridges remain @09:09:53.50PST18-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @09:09:53.52 >BREAK< ESP3 Gateway v1.11 for cartridge bus @09:09:57.60 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @09:09:57.98 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @09:10:01.51 Polling Can every 2:00 @09:10:01.63 Gate.power :analytic1,:ON,limit to 5000mA @09:10:01.66 Gate.power :analytic2,:ON,limit to 5000mA @09:10:01.68 Gate.power :analytic3,:ON,limit to 5000mA @09:10:02.09 Archive.configure :default @09:10:02.17 Archive.hold @09:10:02.25 Archive.configure :default @09:10:02.39 Air.configure :default @09:10:02.47 Air.hold @09:10:02.53 Air.configure :default @09:10:02.67 Lysis1.configure :default @09:10:02.76 Lysis1.hold @09:10:02.81 Lysis1.configure :default @09:10:03.68 Lysis2.configure :default @09:10:03.75 Lysis2.hold @09:10:03.82 Lysis2.configure :default @09:10:04.01 Plunge.configure :default @09:10:04.07 Plunge.hold @09:10:04.16 Plunge.configure :default @09:10:04.33 SamplePump.configure :default @09:10:04.49 ExtraPump.configure :default @09:10:04.68 Toroid.configure :default @09:10:04.73 Toroid.hold @09:10:04.82 Toroid.configure :default @09:10:05.01 Carousel.configure :default @09:10:05.06 Carousel.hold @09:10:05.14 Carousel.configure :default @09:10:05.32 Pressure.configure :default @09:10:05.40 All elf controllers running firmware version 4.61 @09:10:05.51 Intake.close @09:10:05.56 Exhaust.close @09:10:05.66 Bypass.close @09:10:05.91
Loading /home/esp/esp2/bin/processAll mission @09:10:06.21 Sampling default goal volumes @09:10:06.23 Cartridge group: dry Spares: {spare: 0} @09:10:07.03 Carousel.seek 40 @09:10:07.20 Selecting Cartridge 39 @09:10:07.49 Carousel.seek 39 @09:10:08.08 Lysis1.seek :locked @09:10:11.34 Plunge.seek :locked @09:10:11.51 Air.seek :locked @09:10:15.79
Mounted archive Cartridge @09:10:16.17 Duration of filtering limited to 1:05:00 @09:10:16.56 Priming sample loop w/160ml, bypass w/10ml @09:10:16.58 SamplePump.setPosition! 0ml @09:10:16.62 Bypass.open @09:10:18.12 Intake.open @09:10:19.05 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @09:10:21.67 Vacuum tare is 19% of full scale @09:10:21.70 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -19psi @09:10:21.96 Pressure.reconfigure :default @09:10:22.01 Pressure.hold @09:10:22.28 SamplePump.reconfigure :prime @09:10:22.37 SamplePump.seek 10ml,1:21 @09:10:33.49 Pressure.coast @09:10:33.55 SamplePump.setPosition! 0ml @09:10:33.63 Exhaust.open @09:10:35.12 Bypass.close @09:10:36.74 Pressure.reconfigure :default @09:10:36.81 Pressure.hold @09:10:37.15 SamplePump.reconfigure :prime @09:10:37.23 SamplePump.seek 160ml,13:51 @09:13:28.63 Pressure.coast @09:13:28.69 Bypass.open @09:13:29.42 Intake.close @09:13:30.34 Toroid.reconfigure :hold @09:13:30.42 Toroid.seek :crack @09:13:32.05 Toroid.seek :cartridge @09:13:32.98 Toroid.reconfigure :default @09:13:33.06 Toroid.seek :clear @09:13:34.96 Pressurizing cartridge to 26psi above 14.5psia @09:13:34.98 SamplePump.setPosition! 0ml @09:13:35.01 Intake.open @09:13:35.83 Bypass.close @09:13:36.56 Exhaust.close @09:13:37.47 Pressure.reconfigure :default @09:13:37.56 Pressure.hold @09:13:37.82 SamplePump.reconfigure :charge @09:13:37.88 SamplePump.seek 2.3ml,42 seconds @09:13:45.08 Pressure.coast @09:13:45.13 Intake.close @09:13:46.02 Pressurized cartridge with 0.232ml to 26.1psi @09:13:46.03 Isolated @35.4psia (20.9psi) @09:13:46.04 Checking for cartridge leaks from 35.4psia @09:13:46.09 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @09:13:57.31 Toroid.reconfigure :hold @09:13:57.40 Toroid.seek :bypass @09:13:59.06 Toroid.reconfigure :default @09:13:59.11 Toroid.seek :clear @09:14:00.78 Cartridge::Sampler::Leaky in FILTERING -- Isolated cartridge pressure fell by 15% to 30.1psia @09:14:00.80 Retry #1 of 2 @09:14:00.82 Cmd::Error in FILTERING -- No spare cartridges remain @09:14:00.83 Cmd::Error in FILTERING -- No spare cartridges remain @09:14:01.00
Cartridge Load: {[43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 250}, [44, 42, 40] => { Type: :archive_bac, State: :processed, filtered: 250}, 38 => { Type: :archive_bac, State: :dry}, 39 => { Type: :archive_bac, State: :leaky}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 297}, [37..32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}} Spares: {spare: 0} @09:14:01.04 Thread::ChildDied in MAIN -- :FILTERING @09:14:01.13 SamplePump.coast @09:31:48.82PST18-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @09:31:48.84 >BREAK< ESP3 Gateway v1.11 for cartridge bus @09:31:52.71 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @09:31:53.08 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @09:31:56.70 Polling Can every 2:00 @09:31:56.82 Gate.power :analytic1,:ON,limit to 5000mA @09:31:56.84 Gate.power :analytic2,:ON,limit to 5000mA @09:31:56.85 Gate.power :analytic3,:ON,limit to 5000mA @09:31:57.16 Archive.configure :default @09:31:57.27 Archive.hold @09:31:57.35 Archive.configure :default @09:31:57.52 Air.configure :default @09:31:57.58 Air.hold @09:31:57.67 Air.configure :default @09:31:57.80 Lysis1.configure :default @09:31:57.88 Lysis1.hold @09:31:58.44 Lysis1.configure :default @09:31:58.58 Lysis2.configure :default @09:31:58.64 Lysis2.hold @09:31:58.73 Lysis2.configure :default @09:31:58.89 Plunge.configure :default @09:31:58.96 Plunge.hold @09:31:59.03 Plunge.configure :default @09:31:59.20 SamplePump.configure :default @09:31:59.39 ExtraPump.configure :default @09:31:59.55 Toroid.configure :default @09:31:59.61 Toroid.hold @09:31:59.70 Toroid.configure :default @09:31:59.88 Carousel.configure :default @09:31:59.95 Carousel.hold @09:32:00.01 Carousel.configure :default @09:32:00.21 Pressure.configure :default @09:32:00.26 All elf controllers running firmware version 4.61 @09:32:00.37 Intake.close @09:32:00.41 Exhaust.close @09:32:00.54 Bypass.close @09:32:00.78
Loading /home/esp/esp2/bin/processAll mission @09:32:01.08 Sampling default goal volumes @09:32:01.09 Cartridge group: dry Spares: {spare: 0} @09:32:02.17 Carousel.seek 39 @09:32:03.12 Mounted archive Cartridge @09:32:03.18 Selecting Cartridge 39 @09:32:03.96 Duration of filtering limited to 1:05:00 @09:32:04.34 Priming sample loop w/160ml, bypass w/10ml @09:32:04.35 SamplePump.setPosition! 0ml @09:32:04.39 Bypass.open @09:32:05.93 Intake.open @09:32:06.84 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @09:32:09.49 Vacuum tare is 19% of full scale @09:32:09.51 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -19psi @09:32:09.78 Pressure.reconfigure :default @09:32:09.83 Pressure.hold @09:32:10.10 SamplePump.reconfigure :prime @09:32:10.20 SamplePump.seek 10ml,1:21 @09:32:21.29 Pressure.coast @09:32:21.37 SamplePump.setPosition! 0ml @09:32:21.42 Exhaust.open @09:32:22.87 Bypass.close @09:32:24.50 Pressure.reconfigure :default @09:32:24.59 Pressure.hold @09:32:24.96 SamplePump.reconfigure :prime @09:32:25.03 SamplePump.seek 160ml,13:51 @09:35:16.40 Pressure.coast @09:35:16.46 Bypass.open @09:35:17.16 Intake.close @09:35:18.09 Toroid.reconfigure :hold @09:35:18.17 Toroid.seek :crack @09:35:19.78 Toroid.seek :cartridge @09:35:20.71 Toroid.reconfigure :default @09:35:20.79 Toroid.seek :clear @09:35:22.68 Pressurizing cartridge to 26psi above 14.7psia @09:35:22.70 SamplePump.setPosition! 0ml @09:35:22.76 Intake.open @09:35:23.57 Bypass.close @09:35:24.31 Exhaust.close @09:35:25.18 Pressure.reconfigure :default @09:35:25.26 Pressure.hold @09:35:25.53 SamplePump.reconfigure :charge @09:35:25.59 SamplePump.seek 2.3ml,42 seconds @09:35:31.55 Pressure.coast @09:35:31.61 Intake.close @09:35:32.52 Pressurized cartridge with 0.158ml to 26.0psi @09:35:32.53 Isolated @33.9psia (19.2psi) @09:35:32.57 Checking for cartridge leaks from 33.9psia @09:35:32.59 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @09:35:41.18 Cartridge pressure fell 7% to 31.6psia @09:35:41.29 Sampling 1000.0ml @09:35:41.30 SamplePump.setPosition! 0ml @09:35:41.37 Exhaust.open @09:35:42.02 Intake.open @09:35:43.03 Pressure.reconfigure :default @09:35:43.09 Pressure.hold @09:35:43.46 SamplePump.reconfigure :sample @09:35:43.55 SamplePump.seek 1000ml,1:23:51 @09:35:46.11 Cartridge 39 sampling 1/1000ml at 0.53ml/s with 27.1psi vacuum @09:36:46.10 Cartridge 39 sampling 32/1000ml at 0.53ml/s with 22.4psi vacuum @09:37:46.09 Cartridge 39 sampling 64/1000ml at 0.52ml/s with 22.5psi vacuum @09:38:46.12 Cartridge 39 sampling 96/1000ml at 0.53ml/s with 22.7psi vacuum @09:39:46.10 Cartridge 39 sampling 128/1000ml at 0.53ml/s with 18.0psi vacuum 3.1m deep (19.3psia) @09:40:46.09 Cartridge 39 sampling 160/1000ml at 0.53ml/s with 21.0psi vacuum 1.7m deep (17.1psia) @09:41:46.10 Cartridge 39 sampling 191/1000ml at 0.53ml/s with 18.7psi vacuum 2.9m deep (19.0psia) @09:42:46.09 Cartridge 39 sampling 223/1000ml at 0.54ml/s with 25.0psi vacuum @09:43:46.09 Cartridge 39 sampling 255/1000ml at 0.53ml/s with 18.8psi vacuum 3.4m deep (19.7psia) @09:44:46.11 Cartridge 39 sampling 286/1000ml at 0.53ml/s with 18.2psi vacuum 3.1m deep (19.2psia) @09:45:46.09 Cartridge 39 sampling 318/1000ml at 0.53ml/s with 25.2psi vacuum @09:46:46.09 Cartridge 39 sampling 350/1000ml at 0.51ml/s with 24.7psi vacuum @09:47:46.10 Cartridge 39 sampling 381/1000ml at 0.53ml/s with 19.7psi vacuum 2.7m deep (18.7psia) @09:48:46.10 Cartridge 39 sampling 413/1000ml at 0.53ml/s with 25.0psi vacuum @09:49:46.09 Cartridge 39 sampling 444/1000ml at 0.53ml/s with 18.5psi vacuum @09:50:46.09 Cartridge 39 sampling 476/1000ml at 0.50ml/s with 26.0psi vacuum @09:51:46.10 Cartridge 39 sampling 508/1000ml at 0.52ml/s with 27.3psi vacuum @09:52:46.09 Cartridge 39 sampling 539/1000ml at 0.54ml/s with 19.9psi vacuum 1.8m deep (17.3psia) @09:53:46.11 Cartridge 39 sampling 571/1000ml at 0.50ml/s with 21.1psi vacuum @09:54:46.11 Cartridge 39 sampling 602/1000ml at 0.52ml/s with 26.0psi vacuum @09:55:46.09 Cartridge 39 sampling 634/1000ml at 0.50ml/s with 21.9psi vacuum @09:56:46.08 Cartridge 39 sampling 665/1000ml at 0.54ml/s with 27.5psi vacuum @09:57:46.10 Cartridge 39 sampling 697/1000ml at 0.54ml/s with 27.1psi vacuum @09:58:46.13 Cartridge 39 sampling 728/1000ml at 0.53ml/s with 28.0psi vacuum @09:59:46.09 Cartridge 39 sampling 760/1000ml at 0.54ml/s with 24.0psi vacuum @10:00:46.09 Cartridge 39 sampling 792/1000ml at 0.53ml/s with 18.8psi vacuum @10:01:46.09 Cartridge 39 sampling 823/1000ml at 0.53ml/s with 26.1psi vacuum @10:02:46.15 Cartridge 39 sampling 855/1000ml at 0.54ml/s with 22.3psi vacuum @10:03:46.09 Cartridge 39 sampling 887/1000ml at 0.53ml/s with 17.0psi vacuum 3.0m deep (19.1psia) @10:04:46.11 Cartridge 39 sampling 919/1000ml at 0.52ml/s with 19.6psi vacuum @10:05:46.13 Cartridge 39 sampling 951/1000ml at 0.53ml/s with 17.3psi vacuum 3.7m deep (20.2psia) @10:06:46.11 Cartridge 39 sampling 983/1000ml at 0.55ml/s with 24.2psi vacuum @10:07:17.09 Pressure.coast @10:07:17.25 Intake.close @10:07:17.32 Exhaust.close @10:07:17.59 Toroid.reconfigure :hold @10:07:17.65 Toroid.seek :bypass @10:07:19.27 Toroid.reconfigure :default @10:07:19.34 Toroid.seek :clear @10:07:20.99 Sampled 1000.0ml @10:07:21.57 Plunge.seek :reagent @10:07:22.47 Sample loop closed. Ready to Deliver Reagents @10:07:22.49 Archive.seek :sealed @10:07:26.14 Archive.seek :evac @10:07:40.14 delay 1:00 @10:08:40.17 Archive.seek :clear @10:08:57.33 RNAlater soak for 10:00 @10:18:57.37 Archive.seek :tVent @10:18:59.07 delay 1 second @10:19:00.10 Archive.seek :evac @10:19:16.07 delay 10 seconds @10:19:26.11 Sample preserved and evacuated @10:19:26.40
Carousel.seek 39 @10:19:26.66 Plunge.seek :locked @10:19:27.69 Archive.seek :clear @10:19:27.90 Air.seek :clear @10:19:28.09 Lysis1.seek :clear @10:19:44.91
Archive.coast @10:19:44.95 Air.coast @10:19:45.02 Lysis1.coast @10:19:45.06 Lysis2.coast @10:19:45.14 Plunge.coast @10:19:45.19 SamplePump.coast @10:19:45.23 Toroid.coast @10:19:45.30 Carousel.coast @10:19:45.52 Gate.power :motor,:OFF @10:19:45.55 Gate.power :core,:OFF @10:19:45.57 Gate.power :analytic1,:OFF @10:19:45.58 Gate.power :analytic2,:OFF @10:19:45.59 Gate.power :analytic3,:OFF @10:19:45.60 Gate.power :main,:OFF @10:19:45.62 Safely stopped and ready to power off @10:19:46.03 Gate.power :main,:ON,limit to 5000mA @10:19:46.06 Gate.power :analytic1,:ON,limit to 5000mA @10:19:46.08 Gate.power :analytic2,:ON,limit to 5000mA @10:19:46.09 Gate.power :analytic3,:ON,limit to 5000mA @10:19:46.61 Powering core bus @10:19:48.98 Archive.configure :default @10:19:49.14 Air.configure :default @10:19:49.30 Lysis1.configure :default @10:19:49.44 Lysis2.configure :default @10:19:49.60 Plunge.configure :default @10:19:49.77 SamplePump.configure :default @10:19:49.98 ExtraPump.configure :default @10:19:50.15 Toroid.configure :default @10:19:50.33 Carousel.configure :default @10:19:50.51 Pressure.configure :default @10:19:50.57 All elf controllers running firmware version 4.61 @10:19:50.60 Gate.power :motor,:ON,limit to 5000mA @10:19:50.79 Archive.reconfigure :default @10:19:51.53 Archive.home.jog 0 @10:19:51.82 Toroid.reconfigure :default @10:19:51.91 Plunge.reconfigure :default @10:19:51.97 Lysis1.reconfigure :default @10:19:52.01 Air.reconfigure :default @10:19:52.09 Lysis2.reconfigure :default @10:19:52.37 Toroid.home.jog 0 @10:19:52.43 Plunge.home.jog 0 @10:19:52.45 Lysis1.home.jog 0 @10:19:52.50 Lysis2.home.jog 0 @10:19:52.52 Air.home.jog 0 @10:19:53.23 Archive.reconfigure :homing @10:19:53.43 Archive.jog 1374 @10:19:53.52 Plunge.reconfigure :homing @10:19:53.55 Lysis1.reconfigure :homing @10:19:53.62 Lysis2.reconfigure :homing @10:19:53.67 Air.reconfigure :homing @10:19:53.75 Toroid.reconfigure :homing @10:19:54.03 Plunge.jog 845 @10:19:54.10 Lysis2.jog 1356 @10:19:54.23 Air.jog 1366 @10:19:54.32 Toroid.jog 614 @10:19:54.36 Lysis1.jog 1352 @10:19:55.00 Archive.setRawPosition! 50 @10:19:55.06 Archive.seek :home @10:19:55.18 Plunge.setRawPosition! 50 @10:19:55.25 Plunge.seek :home @10:19:55.38 Archive.reconfigure :default @10:19:55.49 Lysis2.setRawPosition! 50 @10:19:55.58 Lysis2.seek :home @10:19:55.71 Lysis1.setRawPosition! 50 @10:19:55.74 Air.setRawPosition! 50 @10:19:55.78 Plunge.reconfigure :default @10:19:55.98 Lysis1.seek :home @10:19:56.07 Air.seek :home @10:19:56.28 Lysis2.reconfigure :default @10:19:56.40 Lysis1.reconfigure :default @10:19:56.58 Air.reconfigure :default @10:19:56.67 Toroid.setRawPosition! 0 @10:19:56.73 Toroid.seek :home @10:19:56.85 Toroid.reconfigure :default @10:19:57.04
Toroid.seek :clear @10:19:58.66 Carousel.reconfigure :default @10:19:58.73 Carousel.home.jog 0 @10:19:59.24 Carousel.seek 39 @10:20:00.18 Selecting Cartridge 38 @10:20:00.46 Carousel.seek 38 @10:20:01.07 Lysis1.seek :locked @10:20:04.33 Plunge.seek :locked @10:20:04.50 Air.seek :locked @10:20:08.79
Mounted archive Cartridge @10:20:09.15 Duration of filtering limited to 1:05:00 @10:20:09.54 Priming sample loop w/160ml, bypass w/10ml @10:20:09.56 SamplePump.setPosition! 0ml @10:20:09.62 Bypass.open @10:20:11.09 Intake.open @10:20:12.04 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @10:20:14.63 Vacuum tare is 19% of full scale @10:20:14.67 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @10:20:14.95 Pressure.reconfigure :default @10:20:15.01 Pressure.hold @10:20:15.30 SamplePump.reconfigure :prime @10:20:15.37 SamplePump.seek 10ml,1:21 @10:20:26.47 Pressure.coast @10:20:26.54 SamplePump.setPosition! 0ml @10:20:26.59 Exhaust.open @10:20:28.02 Bypass.close @10:20:29.57 Pressure.reconfigure :default @10:20:29.66 Pressure.hold @10:20:29.97 SamplePump.reconfigure :prime @10:20:30.07 SamplePump.seek 160ml,13:51 @10:23:21.53 Pressure.coast @10:23:21.61 Bypass.open @10:23:22.31 Intake.close @10:23:23.25 Toroid.reconfigure :hold @10:23:23.33 Toroid.seek :crack @10:23:24.96 Toroid.seek :cartridge @10:23:25.91 Toroid.reconfigure :default @10:23:25.99 Toroid.seek :clear @10:23:27.92 Pressurizing cartridge to 26psi above 14.4psia @10:23:27.93 SamplePump.setPosition! 0ml @10:23:27.96 Intake.open @10:23:28.76 Bypass.close @10:23:29.46 Exhaust.close @10:23:30.36 Pressure.reconfigure :default @10:23:30.41 Pressure.hold @10:23:30.69 SamplePump.reconfigure :charge @10:23:30.78 SamplePump.seek 2.3ml,42 seconds @10:23:36.79 Pressure.coast @10:23:36.87 Intake.close @10:23:37.81 Pressurized cartridge with 0.163ml to 26.2psi @10:23:37.82 Isolated @35.4psia (21.0psi) @10:23:37.83 Checking for cartridge leaks from 35.4psia @10:23:37.86 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @10:23:44.49 Cartridge pressure fell 7% to 32.9psia @10:23:44.58 Sampling 1000.0ml @10:23:44.61 SamplePump.setPosition! 0ml @10:23:44.66 Exhaust.open @10:23:45.33 Intake.open @10:23:46.42 Pressure.reconfigure :default @10:23:46.50 Pressure.hold @10:23:46.84 SamplePump.reconfigure :sample @10:23:46.90 SamplePump.seek 1000ml,1:23:51 @10:23:49.48 Cartridge 38 sampling 1/1000ml at 0.50ml/s with 26.5psi vacuum @10:24:49.48 Cartridge 38 sampling 32/1000ml at 0.54ml/s with 26.6psi vacuum @10:25:49.47 Cartridge 38 sampling 65/1000ml at 0.54ml/s with 26.8psi vacuum @10:26:49.47 Cartridge 38 sampling 97/1000ml at 0.52ml/s with 26.5psi vacuum @10:27:49.47 Cartridge 38 sampling 129/1000ml at 0.54ml/s with 24.1psi vacuum @10:28:49.48 Cartridge 38 sampling 160/1000ml at 0.53ml/s with 26.6psi vacuum @10:29:49.47 Cartridge 38 sampling 192/1000ml at 0.51ml/s with 18.3psi vacuum @10:30:49.47 Cartridge 38 sampling 224/1000ml at 0.53ml/s with 17.8psi vacuum 3.0m deep (19.2psia) @10:31:49.47 Cartridge 38 sampling 255/1000ml at 0.53ml/s with 20.5psi vacuum 1.7m deep (17.2psia) @10:32:49.47 Cartridge 38 sampling 287/1000ml at 0.54ml/s with 24.5psi vacuum @10:33:49.47 Cartridge 38 sampling 318/1000ml at 0.51ml/s with 25.9psi vacuum @10:34:49.47 Cartridge 38 sampling 350/1000ml at 0.51ml/s with 27.8psi vacuum @10:35:49.48 Cartridge 38 sampling 381/1000ml at 0.54ml/s with 25.3psi vacuum @10:36:49.48 Cartridge 38 sampling 413/1000ml at 0.52ml/s with 27.0psi vacuum @10:37:49.50 Cartridge 38 sampling 444/1000ml at 0.54ml/s with 22.7psi vacuum @10:38:49.49 Cartridge 38 sampling 476/1000ml at 0.52ml/s with 26.8psi vacuum @10:39:49.47 Cartridge 38 sampling 507/1000ml at 0.54ml/s with 19.2psi vacuum 3.1m deep (19.3psia) @10:40:49.49 Cartridge 38 sampling 538/1000ml at 0.51ml/s with 26.1psi vacuum @10:41:49.48 Cartridge 38 sampling 570/1000ml at 0.52ml/s with 27.9psi vacuum @10:42:49.47 Cartridge 38 sampling 601/1000ml at 0.51ml/s with 26.1psi vacuum @10:43:49.48 Cartridge 38 sampling 632/1000ml at 0.52ml/s with 28.1psi vacuum @10:44:49.47 Cartridge 38 sampling 664/1000ml at 0.53ml/s with 22.5psi vacuum @10:45:49.47 Cartridge 38 sampling 695/1000ml at 0.53ml/s with 27.3psi vacuum @10:46:49.47 Cartridge 38 sampling 727/1000ml at 0.52ml/s with 14.3psi vacuum 6.2m deep (23.8psia) @10:47:49.48 Cartridge 38 sampling 758/1000ml at 0.52ml/s with 26.2psi vacuum @10:48:49.48 Cartridge 38 sampling 789/1000ml at 0.53ml/s with 26.9psi vacuum @10:49:49.48 Cartridge 38 sampling 820/1000ml at 0.53ml/s with 27.1psi vacuum @10:50:49.47 Cartridge 38 sampling 852/1000ml at 0.51ml/s with 17.8psi vacuum @10:51:49.46 Cartridge 38 sampling 883/1000ml at 0.51ml/s with 27.7psi vacuum @10:52:49.47 Cartridge 38 sampling 914/1000ml at 0.53ml/s with 15.6psi vacuum 5.8m deep (23.2psia) @10:53:49.50 Cartridge 38 sampling 946/1000ml at 0.52ml/s with 14.5psi vacuum 6.1m deep (23.8psia) @10:54:49.51 Cartridge 38 sampling 977/1000ml at 0.52ml/s with 25.8psi vacuum @10:55:33.25 Pressure.coast @10:55:33.41 Intake.close @10:55:33.48 Exhaust.close @10:55:33.76 Toroid.reconfigure :hold @10:55:33.83 Toroid.seek :bypass @10:55:35.46 Toroid.reconfigure :default @10:55:35.55 Toroid.seek :clear @10:55:37.17 Sampled 1000.0ml @10:55:37.82 Plunge.seek :reagent @10:55:38.76 Sample loop closed. Ready to Deliver Reagents @10:55:38.78 Archive.seek :sealed @10:55:42.45 Archive.seek :evac @10:55:56.41 delay 1:00 @10:56:56.45 Archive.seek :clear @10:57:13.60 RNAlater soak for 10:00 @11:07:13.63 Archive.seek :tVent @11:07:15.29 delay 1 second @11:07:16.33 Archive.seek :evac @11:07:32.30 delay 10 seconds @11:07:42.34 Sample preserved and evacuated @11:07:42.64
Carousel.seek 38 @11:07:42.92 Plunge.seek :locked @11:07:43.93 Archive.seek :clear @11:07:44.13 Air.seek :clear @11:07:44.32 Lysis1.seek :clear @11:08:01.14
Archive.coast @11:08:01.18 Air.coast @11:08:01.26 Lysis1.coast @11:08:01.30 Lysis2.coast @11:08:01.37 Plunge.coast @11:08:01.41 SamplePump.coast @11:08:01.45 Toroid.coast @11:08:01.52 Carousel.coast @11:08:01.74 Gate.power :motor,:OFF @11:08:01.78 Gate.power :core,:OFF @11:08:01.79 Gate.power :analytic1,:OFF @11:08:01.80 Gate.power :analytic2,:OFF @11:08:01.82 Gate.power :analytic3,:OFF @11:08:01.83 Gate.power :main,:OFF @11:08:01.84 Safely stopped and ready to power off @11:08:02.24 Cartridge Load: {[43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 250}, [44, 42, 40] => { Type: :archive_bac, State: :processed, filtered: 250}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 297}, [37..32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}, [39, 38] => { Type: :archive_bac, State: :processed, filtered: 1000.0}} Spares: {spare: 0} @11:08:02.29 Cartridge::None in MAIN -- Not enough dry cartridges remain @11:10:21.19PST18-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @11:10:21.21 >BREAK< ESP3 Gateway v1.11 for cartridge bus @11:10:25.04 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @11:10:25.42 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @11:10:29.05 Polling Can every 2:00 @11:10:29.10 Loading /home/esp/esp2/bin/processAll mission @11:10:29.94 Sampling default goal volumes @11:10:29.96 Cartridge group: dry Spares: {spare: 0} @11:10:30.11 Gate.power :main,:ON,limit to 5000mA @11:10:30.12 Gate.power :analytic1,:ON,limit to 5000mA @11:10:30.13 Gate.power :analytic2,:ON,limit to 5000mA @11:10:30.17 Gate.power :analytic3,:ON,limit to 5000mA @11:10:30.23 Powering core bus @11:10:32.60 Archive.configure :default @11:10:32.80 Air.configure :default @11:10:32.95 Lysis1.configure :default @11:10:33.09 Lysis2.configure :default @11:10:33.28 Plunge.configure :default @11:10:33.48 SamplePump.configure :default @11:10:33.65 ExtraPump.configure :default @11:10:33.81 Toroid.configure :default @11:10:33.98 Carousel.configure :default @11:10:34.18 Pressure.configure :default @11:10:34.24 All elf controllers running firmware version 4.61 @11:10:34.30 Gate.power :motor,:ON,limit to 5000mA @11:10:34.55 Archive.reconfigure :default @11:10:34.92 Archive.home.jog 0 @11:10:35.23 Plunge.reconfigure :default @11:10:35.29 Lysis2.reconfigure :default @11:10:35.33 Air.reconfigure :default @11:10:35.40 Toroid.reconfigure :default @11:10:35.51 Lysis1.reconfigure :default @11:10:35.71 Lysis2.home.jog 0 @11:10:35.83 Toroid.home.jog 0 @11:10:35.91 Lysis1.home.jog 0 @11:10:35.94 Air.home.jog 0 @11:10:35.96 Plunge.home.jog 0 @11:10:36.62 Archive.reconfigure :homing @11:10:36.71 Archive.jog 1375 @11:10:36.94 Lysis2.reconfigure :homing @11:10:37.00 Lysis1.reconfigure :homing @11:10:37.06 Toroid.reconfigure :homing @11:10:37.09 Plunge.reconfigure :homing @11:10:37.18 Air.reconfigure :homing @11:10:37.60 Lysis2.jog 1357 @11:10:37.72 Toroid.jog 616 @11:10:37.77 Lysis1.jog 1348 @11:10:37.81 Plunge.jog 848 @11:10:37.90 Air.jog 1363 @11:10:38.16 Archive.setRawPosition! 50 @11:10:38.21 Archive.seek :home @11:10:38.49 Archive.reconfigure :default @11:10:38.97 Lysis2.setRawPosition! 50 @11:10:39.02 Lysis2.seek :home @11:10:39.15 Plunge.setRawPosition! 50 @11:10:39.21 Lysis1.setRawPosition! 50 @11:10:39.27 Plunge.seek :home @11:10:39.38 Lysis1.seek :home @11:10:39.57 Lysis2.reconfigure :default @11:10:39.63 Air.setRawPosition! 50 @11:10:39.79 Air.seek :home @11:10:39.91 Plunge.reconfigure :default @11:10:39.96 Lysis1.reconfigure :default @11:10:40.16 Toroid.setRawPosition! 0 @11:10:40.19 Air.reconfigure :default @11:10:40.30 Toroid.seek :home @11:10:40.52 Toroid.reconfigure :default @11:10:40.73
Toroid.seek :clear @11:10:42.40 Carousel.reconfigure :default @11:10:42.46 Carousel.home.jog 0 @11:10:43.00 Carousel.seek 38 @11:10:43.20 Selecting Cartridge 37 @11:10:43.50 Carousel.seek 37 @11:10:44.06 Lysis1.seek :locked @11:10:47.31 Plunge.seek :locked @11:10:47.50 Air.seek :locked @11:10:51.80
Mounted archive Cartridge @11:10:52.17 Duration of filtering limited to 1:05:00 @11:10:52.58 Priming sample loop w/160ml, bypass w/10ml @11:10:52.60 SamplePump.setPosition! 0ml @11:10:52.64 Bypass.open @11:10:54.17 Intake.open @11:10:55.08 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @11:10:57.70 Vacuum tare is 19% of full scale @11:10:57.72 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @11:10:57.99 Pressure.reconfigure :default @11:10:58.05 Pressure.hold @11:10:58.30 SamplePump.reconfigure :prime @11:10:58.40 SamplePump.seek 10ml,1:21 @11:11:09.51 Pressure.coast @11:11:09.56 SamplePump.setPosition! 0ml @11:11:09.65 Exhaust.open @11:11:11.02 Bypass.close @11:11:12.65 Pressure.reconfigure :default @11:11:12.70 Pressure.hold @11:11:13.07 SamplePump.reconfigure :prime @11:11:13.14 SamplePump.seek 160ml,13:51 @11:14:04.73 Pressure.coast @11:14:04.79 Bypass.open @11:14:05.50 Intake.close @11:14:06.44 Toroid.reconfigure :hold @11:14:06.53 Toroid.seek :crack @11:14:08.18 Toroid.seek :cartridge @11:14:09.12 Toroid.reconfigure :default @11:14:09.20 Toroid.seek :clear @11:14:11.10 Pressurizing cartridge to 26psi above 14.4psia @11:14:11.14 SamplePump.setPosition! 0ml @11:14:11.19 Intake.open @11:14:12.04 Bypass.close @11:14:12.77 Exhaust.close @11:14:13.63 Pressure.reconfigure :default @11:14:13.71 Pressure.hold @11:14:13.98 SamplePump.reconfigure :charge @11:14:14.04 SamplePump.seek 2.3ml,42 seconds @11:14:20.53 Pressure.coast @11:14:20.58 Intake.close @11:14:21.56 Pressurized cartridge with 0.152ml to 26.3psi @11:14:21.57 Isolated @34.7psia (20.3psi) @11:14:21.58 Checking for cartridge leaks from 34.7psia @11:14:21.63 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @11:14:29.73 Cartridge pressure fell 7% to 32.4psia @11:14:29.85 Sampling 1000.0ml @11:14:29.86 SamplePump.setPosition! 0ml @11:14:29.93 Exhaust.open @11:14:30.61 Intake.open @11:14:31.63 Pressure.reconfigure :default @11:14:31.77 Pressure.hold @11:14:32.28 SamplePump.reconfigure :sample @11:14:32.35 SamplePump.seek 1000ml,1:23:51 @11:14:34.67 Cartridge 37 sampling 1/1000ml at 0.53ml/s with 25.8psi vacuum @11:15:34.70 Cartridge 37 sampling 34/1000ml at 0.57ml/s with 20.4psi vacuum @11:16:34.67 Cartridge 37 sampling 68/1000ml at 0.57ml/s with 22.4psi vacuum @11:17:34.69 Cartridge 37 sampling 102/1000ml at 0.57ml/s with 21.0psi vacuum 1.8m deep (17.3psia) @11:18:34.69 Cartridge 37 sampling 136/1000ml at 0.57ml/s with 25.4psi vacuum @11:19:34.67 Cartridge 37 sampling 169/1000ml at 0.56ml/s with 21.4psi vacuum @11:20:34.68 Cartridge 37 sampling 203/1000ml at 0.56ml/s with 19.6psi vacuum 2.6m deep (18.5psia) @11:21:34.68 Cartridge 37 sampling 236/1000ml at 0.55ml/s with 27.2psi vacuum @11:22:34.68 Cartridge 37 sampling 269/1000ml at 0.54ml/s with 25.5psi vacuum @11:23:34.69 Cartridge 37 sampling 303/1000ml at 0.54ml/s with 21.8psi vacuum @11:24:34.68 Cartridge 37 sampling 336/1000ml at 0.53ml/s with 17.9psi vacuum 3.2m deep (19.4psia) @11:25:34.71 Cartridge 37 sampling 369/1000ml at 0.56ml/s with 27.8psi vacuum @11:26:34.70 Cartridge 37 sampling 402/1000ml at 0.55ml/s with 17.7psi vacuum 3.2m deep (19.5psia) @11:27:34.68 Cartridge 37 sampling 435/1000ml at 0.55ml/s with 27.7psi vacuum @11:28:34.68 Cartridge 37 sampling 468/1000ml at 0.54ml/s with 26.4psi vacuum @11:29:34.68 Cartridge 37 sampling 502/1000ml at 0.55ml/s with 19.1psi vacuum 3.1m deep (19.3psia) @11:30:34.69 Cartridge 37 sampling 535/1000ml at 0.56ml/s with 20.0psi vacuum 2.6m deep (18.5psia) @11:31:34.68 Cartridge 37 sampling 568/1000ml at 0.56ml/s with 22.5psi vacuum @11:32:34.68 Cartridge 37 sampling 601/1000ml at 0.56ml/s with 20.0psi vacuum 2.3m deep (18.0psia) @11:33:34.70 Cartridge 37 sampling 634/1000ml at 0.54ml/s with 16.6psi vacuum @11:34:34.68 Cartridge 37 sampling 667/1000ml at 0.54ml/s with 27.4psi vacuum @11:35:34.68 Cartridge 37 sampling 700/1000ml at 0.54ml/s with 17.8psi vacuum @11:36:34.68 Cartridge 37 sampling 733/1000ml at 0.54ml/s with 23.1psi vacuum @11:37:34.68 Cartridge 37 sampling 766/1000ml at 0.55ml/s with 26.4psi vacuum @11:38:34.69 Cartridge 37 sampling 799/1000ml at 0.55ml/s with 27.9psi vacuum @11:39:34.68 Cartridge 37 sampling 832/1000ml at 0.55ml/s with 14.7psi vacuum 6.4m deep (24.2psia) @11:40:34.70 Cartridge 37 sampling 865/1000ml at 0.54ml/s with 27.4psi vacuum @11:41:34.67 Cartridge 37 sampling 898/1000ml at 0.54ml/s with 27.8psi vacuum @11:42:34.68 Cartridge 37 sampling 931/1000ml at 0.55ml/s with 19.8psi vacuum 2.8m deep (18.9psia) @11:43:34.68 Cartridge 37 sampling 964/1000ml at 0.54ml/s with 27.0psi vacuum @11:44:34.68 Cartridge 37 sampling 997/1000ml at 0.54ml/s with 24.3psi vacuum @11:44:39.45 Pressure.coast @11:44:39.63 Intake.close @11:44:39.71 Exhaust.close @11:44:39.97 Toroid.reconfigure :hold @11:44:40.03 Toroid.seek :bypass @11:44:41.70 Toroid.reconfigure :default @11:44:41.76 Toroid.seek :clear @11:44:43.41 Sampled 1000.0ml @11:44:44.04 SamplePump.setPosition! 0ml @11:44:44.10 Plunge.seek :reagent @11:44:44.16 Exhaust.open @11:44:44.93 Intake.open @11:44:45.11 Sample loop closed. Ready to Deliver Reagents @11:44:45.13 Archive.seek :sealed @11:44:45.86 SamplePump.reconfigure :slow @11:44:45.95 SamplePump.seek -22ml @11:44:48.81 Archive.seek :evac @11:45:02.79 delay 1:00 @11:45:33.29 delay 1 second @11:45:34.32 Bypass.open @11:45:35.04 Exhaust.close @11:45:35.85 SamplePump.seek -20ml @11:45:40.41 delay 1 second @11:45:41.50 SamplePump.seek -28ml @11:45:58.94 delay 1 second @11:46:00.04 Intake.close @11:46:00.11 Bypass.close @11:46:02.83 Archive.seek :clear @11:46:20.00 RNAlater soak for 10:00 @11:56:20.03 Archive.seek :tVent @11:56:21.69 delay 1 second @11:56:22.73 Archive.seek :evac @11:56:38.71 delay 10 seconds @11:56:48.75 Sample preserved and evacuated @11:56:49.04
Carousel.seek 37 @11:56:49.32 Plunge.seek :locked @11:56:50.34 Archive.seek :clear @11:56:50.55 Air.seek :clear @11:56:50.76 Lysis1.seek :clear @11:57:07.57
Archive.coast @11:57:07.63 Air.coast @11:57:07.68 Lysis1.coast @11:57:07.74 Lysis2.coast @11:57:07.79 Plunge.coast @11:57:07.84 SamplePump.coast @11:57:07.91 Toroid.coast @11:57:07.95 Carousel.coast @11:57:08.18 Gate.power :motor,:OFF @11:57:08.19 Gate.power :core,:OFF @11:57:08.20 Gate.power :analytic1,:OFF @11:57:08.22 Gate.power :analytic2,:OFF @11:57:08.23 Gate.power :analytic3,:OFF @11:57:08.24 Gate.power :main,:OFF @11:57:08.25 Safely stopped and ready to power off @11:57:08.61 Cartridge Load: {[39, 38] => { Type: :archive_bac, State: :processed, filtered: 1000}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 250}, [44, 42, 40] => { Type: :archive_bac, State: :processed, filtered: 250}, [36..32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}, 37 => { Type: :archive_bac, State: :processed, filtered: 1000.0}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 297}} Spares: {spare: 0} @11:57:08.63 Cartridge::None in MAIN -- Not enough dry cartridges remain @13:33:30.55PST19-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @13:33:30.58 >BREAK< ESP3 Gateway v1.11 for cartridge bus @13:33:37.42 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @13:33:37.71 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @13:33:40.56 Polling Can every 2:00 @13:33:40.62 Waiting for espclients... @13:33:40.92 Session Begun @13:50:47.21 Session Begun @13:50:53.31 -> #Howdy! @13:51:05.57 -> ah here we are again, very expensive chat client @13:51:05.59 SyntaxError in peacock -- compile error peacock:1: syntax error, unexpected tIDENTIFIER, expecting kDO or '{' or '(' ah here we are again, very expensive chat client ^ @13:51:19.16 -> # sorry it's all coming back to me now @13:51:40.08 -> require 'vacuumCart' @13:51:53.45 -> refillSampleLines @13:51:53.48 Refilling sample & bypass lines with fluid. @13:51:53.56 Waiting for brent to answer "Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return." @13:52:00.51 Answered Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return."" @13:52:00.62 Bypass.open @13:52:00.69 delay 1 second @13:52:01.73 I2C::NodeOffline in brent -- carousel0 microcontroller never signed on while reading Storage Carousel (SC)'s status @13:52:06.66 -> ESP.ready! @13:52:07.11 Gate.power :main,:ON,limit to 5000mA @13:52:07.17 Gate.power :analytic1,:ON,limit to 5000mA @13:52:07.27 Gate.power :analytic2,:ON,limit to 5000mA @13:52:07.39 Gate.power :analytic3,:ON,limit to 5000mA @13:52:07.69 Powering core bus @13:52:10.29 Archive.configure :default @13:52:10.47 Air.configure :default @13:52:10.62 Lysis1.configure :default @13:52:10.77 Lysis2.configure :default @13:52:10.95 Plunge.configure :default @13:52:11.13 SamplePump.configure :default @13:52:11.30 ExtraPump.configure :default @13:52:11.48 Toroid.configure :default @13:52:11.65 Carousel.configure :default @13:52:11.84 Pressure.configure :default @13:52:11.93 All elf controllers running firmware version 4.61 @13:52:11.99 Gate.power :motor,:ON,limit to 5000mA @13:52:12.35 Air.reconfigure :default @13:52:12.42 Archive.reconfigure :default @13:52:12.70 Archive.home.jog 0 @13:52:12.77 Air.home.jog 0 @13:52:13.16 Plunge.reconfigure :default @13:52:13.22 Lysis2.reconfigure :default @13:52:13.29 Lysis1.reconfigure :default @13:52:13.35 Toroid.reconfigure :default @13:52:13.57 Plunge.home.jog 0 @13:52:13.63 Lysis2.home.jog 0 @13:52:13.68 Toroid.home.jog 0 @13:52:13.73 Lysis1.home.jog 0 @13:52:14.27 Archive.reconfigure :homing @13:52:14.34 Air.reconfigure :homing @13:52:14.47 Archive.jog 1372 @13:52:14.52 Air.jog 1363 @13:52:15.30 Plunge.reconfigure :homing @13:52:15.41 Lysis1.reconfigure :homing @13:52:15.57 Toroid.reconfigure :homing @13:52:15.68 Lysis2.reconfigure :homing @13:52:16.31 Plunge.jog 849 @13:52:16.42 Lysis1.jog 1348 @13:52:16.61 Toroid.jog 621 @13:52:16.91 Lysis2.jog 1355 @13:52:17.35 Air.setRawPosition! 50 @13:52:17.46 Archive.setRawPosition! 50 @13:52:17.76 Air.seek :home @13:52:18.04 Archive.seek :home @13:52:18.53 Plunge.setRawPosition! 50 @13:52:18.61 Lysis1.setRawPosition! 50 @13:52:18.68 Air.reconfigure :default @13:52:18.83 Lysis2.setRawPosition! 50 @13:52:19.40 Plunge.seek :home @13:52:19.43 Lysis1.seek :home @13:52:19.74 Lysis2.seek :home @13:52:20.28 Toroid.setRawPosition! 0 @13:52:20.38 Archive.reconfigure :default @13:52:20.50 Plunge.reconfigure :default @13:52:20.89 Toroid.seek :home @13:52:21.21 Lysis2.reconfigure :default @13:52:21.25 Lysis1.reconfigure :default @13:52:21.32 Toroid.reconfigure :default @13:52:21.86 Toroid.seek :clear @13:52:23.51 Carousel.reconfigure :default @13:52:23.61 Carousel.home.jog 0 @13:52:24.15 Carousel.seek 37 @13:52:49.57 -> #trying to do too many things at once. Must ready! the hardware before running any scripts... @13:52:55.06 -> refillSampleLines @13:52:55.09 Refilling sample & bypass lines with fluid. @13:52:55.25 Waiting for brent to answer "Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return." @13:52:56.58 Answered Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return."" @13:52:56.67 Bypass.open @13:52:57.62 Intake.open @13:52:58.51 SamplePump.setPosition! 0ml @13:52:58.60 SamplePump.reconfigure :fast @13:52:58.69 SamplePump.seek 30ml @13:53:25.32 Bypass.close @13:53:26.19 Exhaust.open @13:53:27.00 SamplePump.seek 160ml @13:53:28.40 -> # standing by @13:55:20.37 delay 1 second @13:55:21.41 Intake.close @13:55:21.98 Exhaust.close @13:55:23.16 Sampling Loop filled with water @13:57:07.00 -> #great! The pump works. :-) @13:57:52.82 -> # Want to verify can set up commands from our main computer to ESP @13:58:52.73 -> # I have notes from before a la 'echo cmd | ssh esp@IP espclient peacock @13:59:07.95 -> # what is good cmd to test with? @14:03:06.58 -> Cmd.loadCartridge @14:03:06.69 Cartridge::None in brent -- No dry cartridges remain @14:03:10.84 -> slot @14:03:10.91 Session Begun @14:03:11.01 -> ESP @14:03:11.87 Session Ended @14:03:58.33 -> to 45 @14:03:58.60 Carousel.seek 45 @14:04:01.18 Lysis1.seek :locked @14:04:04.72 Plunge.seek :locked @14:04:05.05 Air.seek :locked @14:04:09.40 Mounted archive Cartridge @14:04:17.67 -> Cartridge.unlock @14:04:17.96 Air.seek :clear @14:04:21.88 Lysis1.seek :clear @14:04:22.35 Plunge.seek :clear @14:04:30.59 -> SC.to 1 @14:04:30.63 Carousel.seek 1 @14:04:43.08 -> SC.to 60 @14:04:43.13 Carousel.seek 60 @14:06:07.05 Session Begun @14:06:07.10 -> ESP @14:06:08.54 Session Ended @14:11:33.20 Session Begun @14:11:33.29 -> ESP.Stop @14:11:33.35 NoMethodError in saildrone -- undefined method `Stop' for ESP:Module @14:11:33.40 Session Ended @14:12:28.09 Session Begun @14:12:28.23 -> ESP.Stop @14:12:28.38 NoMethodError in saildrone -- undefined method `Stop' for ESP:Module @14:12:28.61 Session Ended @14:12:29.87 Session Begun @14:12:29.95 -> Cmd.stop @14:12:30.30 Carousel.seek 60 @14:12:31.06 Lysis1.seek :locked @14:12:34.25 Plunge.seek :locked @14:12:35.86 Lysis1.seek :clear @14:12:39.45 Archive.coast @14:12:39.59 Air.coast @14:12:39.68 Lysis1.coast @14:12:39.77 Lysis2.coast @14:12:39.86 Plunge.coast @14:12:39.95 SamplePump.coast @14:12:40.04 Toroid.coast @14:12:40.13 Carousel.coast @14:12:40.37 Gate.power :motor,:OFF @14:12:40.43 Gate.power :core,:OFF @14:12:40.46 Gate.power :analytic1,:OFF @14:12:40.49 Gate.power :analytic2,:OFF @14:12:40.54 Gate.power :analytic3,:OFF @14:12:40.59 Gate.power :main,:OFF @14:12:40.63 Safely stopped and ready to power off @14:12:40.86 Session Ended @14:16:38.86 -> ping 8.8.8.8 @14:16:38.97 SyntaxError in peacock -- compile error peacock:7: syntax error, unexpected tFLOAT, expecting $end ping 8.8.8.8 ^ @14:16:47.88 -> quit @14:16:47.91 Session Ended @14:17:03.56 Session Ended @14:19:16.65 Session Begun @14:19:55.76 -> killSampleLoop! @14:19:55.79 SamplePump.setPosition! 0ml @14:19:55.94 BusOff in brent -- core bus was not powered while setting Sample Pump (SP)'s position @14:20:03.71 -> ESP.ready! @14:20:04.10 Gate.power :main,:ON,limit to 5000mA @14:20:04.18 Gate.power :analytic1,:ON,limit to 5000mA @14:20:04.28 Gate.power :analytic2,:ON,limit to 5000mA @14:20:04.38 Gate.power :analytic3,:ON,limit to 5000mA @14:20:04.60 Powering core bus @14:20:06.76 Archive.configure :default @14:20:06.94 Air.configure :default @14:20:07.07 Lysis1.configure :default @14:20:07.21 Lysis2.configure :default @14:20:07.38 Plunge.configure :default @14:20:07.55 SamplePump.configure :default @14:20:07.72 ExtraPump.configure :default @14:20:07.88 Toroid.configure :default @14:20:08.06 Carousel.configure :default @14:20:08.23 Pressure.configure :default @14:20:08.32 All elf controllers running firmware version 4.61 @14:20:08.35 Gate.power :motor,:ON,limit to 5000mA @14:20:08.70 Archive.reconfigure :default @14:20:08.91 Archive.home.jog 0 @14:20:09.19 Toroid.reconfigure :default @14:20:09.21 Plunge.reconfigure :default @14:20:09.25 Lysis1.reconfigure :default @14:20:09.29 Air.reconfigure :default @14:20:09.38 Lysis2.reconfigure :default @14:20:09.59 Toroid.home.jog 0 @14:20:09.71 Lysis1.home.jog 0 @14:20:09.74 Plunge.home.jog 0 @14:20:09.84 Air.home.jog 0 @14:20:09.88 Lysis2.home.jog 0 @14:20:10.35 Archive.reconfigure :homing @14:20:10.64 Archive.jog 1359 @14:20:10.70 Toroid.reconfigure :homing @14:20:10.76 Lysis2.reconfigure :homing @14:20:10.81 Air.reconfigure :homing @14:20:10.87 Lysis1.reconfigure :homing @14:20:10.92 Plunge.reconfigure :homing @14:20:11.48 Toroid.jog 607 @14:20:11.54 Lysis2.jog 1355 @14:20:11.64 Air.jog 1368 @14:20:11.84 Lysis1.jog 1352 @14:20:12.08 Plunge.jog 848 @14:20:12.99 Archive.setRawPosition! 50 @14:20:13.08 Archive.seek :home @14:20:13.58 Archive.reconfigure :default @14:20:13.99 Lysis1.setRawPosition! 50 @14:20:14.11 Plunge.setRawPosition! 50 @14:20:14.15 Lysis2.setRawPosition! 50 @14:20:14.29 Air.setRawPosition! 50 @14:20:14.70 Lysis1.seek :home @14:20:14.83 Plunge.seek :home @14:20:14.94 Lysis2.seek :home @14:20:15.02 Air.seek :home @14:20:15.84 Plunge.reconfigure :default @14:20:15.88 Lysis2.reconfigure :default @14:20:16.02 Toroid.setRawPosition! 0 @14:20:16.14 Lysis1.reconfigure :default @14:20:16.53 Toroid.seek :home @14:20:17.01 Air.reconfigure :default @14:20:17.08 Toroid.reconfigure :default @14:20:17.73 Toroid.seek :clear @14:20:19.37 Carousel.reconfigure :default @14:20:19.45 Carousel.home.jog 0 @14:20:19.99 Carousel.seek 60 @14:20:22.95 -> killSampleLoop! @14:20:23.00 SamplePump.setPosition! 0ml @14:20:23.16 Exhaust.open @14:20:23.97 Intake.open @14:20:25.06 SamplePump.reconfigure :slow @14:20:25.44 SamplePump.seek -22ml @14:21:12.89 delay 1 second @14:21:13.92 Bypass.open @14:21:14.66 Exhaust.close @14:21:15.47 SamplePump.seek -20ml @14:21:20.14 delay 1 second @14:21:21.30 SamplePump.seek -28ml @14:21:38.74 delay 1 second @14:21:39.88 Intake.close @14:21:39.98 Bypass.close @14:24:55.89 -> Cmd.stop @14:24:56.10 Carousel.seek 60 @14:24:56.34 Lysis1.seek :locked @14:24:59.51 Plunge.seek :locked @14:25:01.10 Lysis1.seek :clear @14:25:04.49 Archive.coast @14:25:04.66 Air.coast @14:25:04.76 Lysis1.coast @14:25:04.83 Lysis2.coast @14:25:04.91 Plunge.coast @14:25:04.98 SamplePump.coast @14:25:05.06 Toroid.coast @14:25:05.14 Carousel.coast @14:25:05.35 Gate.power :motor,:OFF @14:25:05.37 Gate.power :core,:OFF @14:25:05.41 Gate.power :analytic1,:OFF @14:25:05.45 Gate.power :analytic2,:OFF @14:25:05.48 Gate.power :analytic3,:OFF @14:25:05.52 Gate.power :main,:OFF @14:25:05.56 Safely stopped and ready to power off @14:26:46.94 -> exit @14:26:46.96 Session Ended @14:31:49.67PST19-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @14:31:49.70 >BREAK< ESP3 Gateway v1.11 for cartridge bus @14:31:52.90 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @14:31:53.21 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @14:32:00.01 Polling Can every 2:00 @14:32:00.39 Clients may connect on TCP port 9999 @14:32:00.41 Waiting for espclients... @14:50:47.38 Session Begun @14:50:56.71 -> quit @14:50:56.74 Session Ended @14:56:35.18PST19-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @14:56:35.21 >BREAK< ESP3 Gateway v1.11 for cartridge bus @14:56:41.19 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @14:56:41.82 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @14:56:45.47 Polling Can every 2:00 @14:56:45.85 Clients may connect on TCP port 9999 @14:56:45.87 Waiting for espclients... @14:57:21.59 Session Begun @14:57:21.64 -> ESP @14:57:21.67 Session Ended @14:57:35.66 Session Begun @14:57:42.84 Session Begun @14:57:43.29 -> ESP @14:57:43.34 Session Ended @14:57:59.48 Session Begun @14:57:59.55 -> Cmd.stop @14:57:59.77 ready to power off @14:58:00.10 Session Ended @15:23:22.04PST19-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @15:23:22.14 >BREAK< ESP3 Gateway v1.11 for cartridge bus @15:23:30.01 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @15:23:30.33 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @15:23:33.26 Polling Can every 2:00 @15:23:34.33 Clients may connect on TCP port 9999 @15:23:34.35 Waiting for espclients... @15:44:16.71 Session Begun @15:44:19.92 -> can @15:44:50.57 -> exit @15:44:50.63 Session Ended @17:02:01.79PST23-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @17:02:01.86 >BREAK< ESP3 Gateway v1.11 for cartridge bus @17:02:08.82 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @17:02:09.13 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @17:02:12.06 Polling Can every 2:00 @17:02:12.64 Clients may connect on TCP port 9999 @17:02:12.67 Waiting for espclients... @14:02:50.99PST24-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @14:02:51.06 >BREAK< ESP3 Gateway v1.11 for cartridge bus @14:02:58.42 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @14:02:59.02 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @14:03:02.23 Polling Can every 2:00 @14:03:03.55 Clients may connect on TCP port 9999 @14:03:03.57 Waiting for espclients... @14:03:42.44 Session Begun @14:03:48.85 -> quit @14:03:48.87 Session Ended @14:59:09.66PST24-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @14:59:09.69 >BREAK< ESP3 Gateway v1.11 for cartridge bus @14:59:16.77 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @14:59:17.09 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @14:59:20.09 Polling Can every 2:00 @14:59:20.46 Clients may connect on TCP port 9999 @14:59:20.48 Waiting for espclients... @15:26:50.99 Session Begun @15:28:28.65 -> ESP.ready! @15:28:28.79 Gate.power :main,:ON,limit to 5000mA @15:28:28.81 Gate.power :analytic1,:ON,limit to 5000mA @15:28:28.85 Gate.power :analytic2,:ON,limit to 5000mA @15:28:28.88 Gate.power :analytic3,:ON,limit to 5000mA @15:28:28.95 Powering core bus @15:28:31.59 Archive.configure :default @15:28:31.92 Air.configure :default @15:28:32.37 Lysis1.configure :default @15:28:32.69 Lysis2.configure :default @15:28:33.14 Plunge.configure :default @15:28:33.50 SamplePump.configure :default @15:28:33.82 ExtraPump.configure :default @15:28:34.14 Toroid.configure :default @15:28:34.46 Carousel.configure :default @15:28:34.96 Pressure.configure :default @15:28:35.24 All elf controllers running firmware version 4.61 @15:28:35.31 Gate.power :motor,:ON,limit to 5000mA @15:28:35.70 Archive.reconfigure :default @15:28:35.86 Air.reconfigure :default @15:28:35.97 Lysis1.reconfigure :default @15:28:36.10 Lysis2.reconfigure :default @15:28:36.20 Archive.home.jog 0 @15:28:36.28 Plunge.reconfigure :default @15:28:36.30 Air.home.jog 0 @15:28:36.38 Toroid.reconfigure :default @15:28:36.44 Lysis1.home.jog 0 @15:28:36.65 Lysis2.home.jog 0 @15:28:36.75 Plunge.home.jog 0 @15:28:36.80 Toroid.home.jog 0 @15:28:37.53 Toroid.reconfigure :homing @15:28:37.57 Plunge.reconfigure :homing @15:28:37.60 Lysis2.reconfigure :homing @15:28:37.68 Air.reconfigure :homing @15:28:37.74 Archive.reconfigure :homing @15:28:37.82 Lysis1.reconfigure :homing @15:28:38.07 Plunge.jog 841 @15:28:38.10 Lysis2.jog 1356 @15:28:38.19 Toroid.jog 616 @15:28:38.22 Archive.jog 1358 @15:28:38.30 Lysis1.jog 1350 @15:28:38.35 Air.jog 1364 @15:28:39.18 Plunge.setRawPosition! 50 @15:28:39.25 Plunge.seek :home @15:28:39.56 Plunge.reconfigure :default @15:28:40.00 Lysis2.setRawPosition! 50 @15:28:40.10 Lysis1.setRawPosition! 50 @15:28:40.26 Air.setRawPosition! 50 @15:28:40.64 Archive.setRawPosition! 50 @15:28:41.00 Lysis2.seek :home @15:28:41.21 Lysis1.seek :home @15:28:41.48 Air.seek :home @15:28:41.53 Archive.seek :home @15:28:42.30 Toroid.setRawPosition! 0 @15:28:42.38 Lysis2.reconfigure :default @15:28:42.42 Lysis1.reconfigure :default @15:28:42.89 Toroid.seek :home @15:28:43.46 Toroid.reconfigure :default @15:28:43.49 Air.reconfigure :default @15:28:43.65 Archive.reconfigure :default @15:28:44.55 Toroid.seek :clear @15:28:46.21 Carousel.reconfigure :default @15:28:46.29 Carousel.home.jog 0 @15:28:46.81 Carousel.seek 60 @15:28:50.76 -> SC @15:29:01.33 -> BACsampler.prime @15:29:01.54 Priming sample loop w/160ml, bypass w/10ml @15:29:01.59 SamplePump.setPosition! 0ml @15:29:01.75 Bypass.open @15:29:03.47 Intake.open @15:29:04.63 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:29:07.24 Vacuum tare is 17% of full scale @15:29:07.27 Cartridge::Sampler::Warning in preston -- Vacuum sensor saturates below -17psi @15:29:07.54 Pressure.reconfigure :default @15:29:07.62 Pressure.hold @15:29:07.85 SamplePump.reconfigure :prime @15:29:07.95 SamplePump.seek 10ml,1:21 @15:29:19.16 Pressure.coast @15:29:19.30 SamplePump.setPosition! 0ml @15:29:19.48 Exhaust.open @15:29:21.02 Bypass.close @15:29:23.06 Pressure.reconfigure :default @15:29:23.24 Pressure.hold @15:29:23.56 SamplePump.reconfigure :prime @15:29:23.66 SamplePump.seek 160ml,13:51 @15:29:26.77 SamplePump.coast @15:29:26.84 Intake.close @15:29:26.89 Exhaust.close @15:29:27.99 Pressure.coast @15:29:28.22 Cartridge::Sampler::Leaky in preston -- Serious Leak or Stall -- 0.93ml/s made only 7.2psi vacuum @15:30:06.25 -> require 'vacuumCart' @15:30:24.48 -> refillSampleLines @15:30:24.55 Refilling sample & bypass lines with fluid. @15:30:24.61 Waiting for preston to answer "Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return." @15:30:26.62 Answered Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return."" @15:30:26.72 Bypass.open @15:30:27.60 Intake.open @15:30:28.49 SamplePump.setPosition! 0ml @15:30:28.56 SamplePump.reconfigure :fast @15:30:28.64 SamplePump.seek 30ml @15:30:55.31 Bypass.close @15:30:56.11 Exhaust.open @15:30:56.82 SamplePump.seek 160ml @15:32:50.27 delay 1 second @15:32:51.32 Intake.close @15:32:51.51 Exhaust.close @15:32:52.91 Sampling Loop filled with water @15:33:25.20 -> BACsampler.prime @15:33:25.31 Priming sample loop w/160ml, bypass w/10ml @15:33:25.33 SamplePump.setPosition! 0ml @15:33:25.42 Bypass.open @15:33:26.94 Intake.open @15:33:27.82 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:33:30.40 Vacuum tare is 17% of full scale @15:33:30.54 Cartridge::Sampler::Warning in preston -- Vacuum sensor saturates below -17psi @15:33:31.17 Pressure.reconfigure :default @15:33:31.38 Pressure.hold @15:33:31.91 SamplePump.reconfigure :prime @15:33:32.10 SamplePump.seek 10ml,1:21 @15:33:35.35 SamplePump.coast @15:33:35.43 Intake.close @15:33:36.00 Bypass.close @15:33:36.88 Pressure.coast @15:33:36.96 Cartridge::Sampler::Leaky in preston -- Serious Leak or Stall -- 0.93ml/s made only 0.6psi vacuum @15:34:37.75 -> BACsampler.prime @15:34:37.93 Priming sample loop w/160ml, bypass w/10ml @15:34:37.98 SamplePump.setPosition! 0ml @15:34:38.19 Bypass.open @15:34:39.99 Intake.open @15:34:40.92 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:34:43.75 Vacuum tare is 17% of full scale @15:34:43.84 Cartridge::Sampler::Warning in preston -- Vacuum sensor saturates below -17psi @15:34:44.64 Pressure.reconfigure :default @15:34:44.87 Pressure.hold @15:34:45.45 SamplePump.reconfigure :prime @15:34:45.69 SamplePump.seek 10ml,1:21 @15:34:48.84 SamplePump.coast @15:34:48.92 Intake.close @15:34:48.97 Bypass.close @15:34:49.86 Pressure.coast @15:34:49.94 Cartridge::Sampler::Leaky in preston -- Serious Leak or Stall -- 0.93ml/s made only 0.1psi vacuum @15:41:26.69 Session Begun @15:41:29.08 -> Sampler.psi @15:41:45.09 -> loop {Sampler.psi, delay 1} @15:41:45.12 SyntaxError in preston#2 -- compile error preston#2:2: syntax error, unexpected tINTEGER, expecting kDO or '{' or '(' loop {Sampler.psi, delay 1} ^ @15:41:56.57 -> loop {Sampler.psi; delay 1} @15:41:56.70 delay 1 second @15:41:57.78 delay 1 second @15:41:58.96 delay 1 second @15:41:59.71 UserAbort in preston#2 -- @15:42:07.08 -> loop {puts Sampler.psi; delay 1} @15:42:07.21 delay 1 second @15:42:08.32 delay 1 second @15:42:09.57 delay 1 second @15:42:10.79 delay 1 second @15:42:11.92 UserAbort in preston#2 -- @15:42:29.82 -> BACsampler.prime @15:42:30.16 Priming sample loop w/160ml, bypass w/10ml @15:42:30.22 SamplePump.setPosition! 0ml @15:42:30.50 Bypass.open @15:42:32.31 Intake.open @15:42:33.27 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:42:33.38 -> loop {puts Sampler.psi; delay 1} @15:42:33.53 delay 1 second @15:42:34.63 delay 1 second @15:42:35.73 delay 1 second @15:42:35.86 Vacuum tare is 17% of full scale @15:42:35.90 Cartridge::Sampler::Warning in preston -- Vacuum sensor saturates below -17psi @15:42:36.18 Pressure.reconfigure :default @15:42:36.28 Pressure.hold @15:42:36.53 SamplePump.reconfigure :prime @15:42:36.68 SamplePump.seek 10ml,1:21 @15:42:37.26 delay 1 second @15:42:38.73 delay 1 second @15:42:40.02 delay 1 second @15:42:40.19 SamplePump.coast @15:42:40.44 Intake.close @15:42:40.71 Bypass.close @15:42:41.32 delay 1 second @15:42:41.72 Pressure.coast @15:42:41.81 Cartridge::Sampler::Leaky in preston -- Serious Leak or Stall -- 0.93ml/s made only 0.1psi vacuum @15:42:42.52 delay 1 second @15:42:43.63 delay 1 second @15:42:44.74 delay 1 second @15:42:45.19 UserAbort in preston#2 -- @16:09:23.35 Session Begun @16:09:28.11 -> showlog 200 @16:10:01.45 -> Cartridge.mounted @16:10:10.82 -> Cartridge.bypass @16:11:02.91 Session Begun @16:11:05.84 -> hidelog @16:11:15.81 -> Sampler.PSI @16:12:51.41 -> loop {Delay.sleep 2; puts Sampler.PSI} @16:13:24.33 -> BACsampler.prime nil, 60 @16:13:24.85 Priming bypass w/60ml @16:13:24.88 SamplePump.setPosition! 0ml @16:13:25.02 Bypass.open @16:13:26.57 Intake.open @16:13:27.52 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @16:13:30.26 Vacuum tare is 17% of full scale @16:13:30.39 Cartridge::Sampler::Warning in brent -- Vacuum sensor saturates below -17psi @16:13:31.13 Pressure.reconfigure :default @16:13:31.42 Pressure.hold @16:13:32.35 SamplePump.reconfigure :prime @16:13:32.69 SamplePump.seek 60ml,5:31 @16:13:35.93 SamplePump.coast @16:13:36.05 Intake.close @16:13:36.17 Bypass.close @16:13:37.06 Pressure.coast @16:13:37.18 Cartridge::Sampler::Leaky in brent -- Serious Leak or Stall -- 0.93ml/s made only 0.1psi vacuum @16:13:47.29 -> BACsampler.prime 60, nil @16:13:47.45 Priming sample loop w/60ml @16:13:47.48 SamplePump.setPosition! 0ml @16:13:47.61 Bypass.open @16:13:49.25 Intake.open @16:13:50.43 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @16:13:53.51 Vacuum tare is 17% of full scale @16:13:53.54 Cartridge::Sampler::Warning in brent -- Vacuum sensor saturates below -17psi @16:13:54.46 Exhaust.open @16:13:55.98 Bypass.close @16:13:57.57 Pressure.reconfigure :default @16:13:57.70 Pressure.hold @16:13:58.23 SamplePump.reconfigure :prime @16:13:58.36 SamplePump.seek 60ml,5:31 @16:14:01.52 SamplePump.coast @16:14:01.90 Intake.close @16:14:02.17 Exhaust.close @16:14:03.13 Pressure.coast @16:14:03.26 Cartridge::Sampler::Leaky in brent -- Serious Leak or Stall -- 0.93ml/s made only 0.8psi vacuum @16:14:15.53 -> Intake.open @16:14:15.55 Intake.open @16:14:19.33 -> Exhaust.open @16:14:19.42 Exhaust.open @16:14:30.25 -> SP.pump 100 @16:14:30.53 SamplePump.reconfigure :default @16:14:30.83 SamplePump.seek 102.48ml @16:14:39.25 SamplePump.hold @16:14:40.05 UserAbort in brent -- Sample Pump (SP) seek to 102.48ml Aborted! @16:14:44.68 UserAbort in brent#2 -- @16:15:37.77 -> #Brent stop. Run killSampleLoop! and monitor the pressure. We know the downstream bleach line is open. @16:15:52.22 -> #that should idicate if the intake is blockec, . @16:16:00.81 -> #OK. @16:16:03.60 -> loop {Delay.sleep 2; puts Sampler.PSI} @16:16:37.90 -> #So far, even at full tilt, pressure never exceeded .8psi. @16:17:13.64 -> killSampleLoop! @16:17:13.69 SamplePump.setPosition! 0ml @16:17:13.90 SamplePump.reconfigure :slow @16:17:14.02 SamplePump.seek -22ml @16:17:32.31 -> loop {puts Sampler.psi; delay 1} @16:17:32.46 delay 1 second @16:17:33.56 delay 1 second @16:17:34.67 delay 1 second @16:17:35.77 delay 1 second @16:17:37.03 delay 1 second @16:17:38.24 delay 1 second @16:17:39.49 delay 1 second @16:17:40.75 delay 1 second @16:17:41.91 delay 1 second @16:17:43.01 delay 1 second @16:17:44.13 delay 1 second @16:17:45.23 delay 1 second @16:17:46.42 delay 1 second @16:17:47.71 delay 1 second @16:17:48.24 UserAbort in brent#2 -- @16:17:48.90 delay 1 second @16:17:50.09 delay 1 second @16:17:51.20 delay 1 second @16:17:52.31 delay 1 second @16:17:53.42 delay 1 second @16:17:54.52 delay 1 second @16:17:55.81 delay 1 second @16:17:57.10 delay 1 second @16:17:58.29 delay 1 second @16:17:59.49 delay 1 second @16:18:00.61 delay 1 second @16:18:01.47 delay 1 second @16:18:01.71 delay 1 second @16:18:02.50 Bypass.open @16:18:02.81 delay 1 second @16:18:03.25 Exhaust.close @16:18:03.91 delay 1 second @16:18:04.09 SamplePump.seek -20ml @16:18:04.64 UserAbort in preston#2 -- @16:18:08.72 delay 1 second @16:18:09.81 SamplePump.seek -28ml @16:18:20.85 -> loop {Delay.sleep 2; puts Sampler.PSI} @16:18:27.36 delay 1 second @16:18:28.47 Intake.close @16:18:28.61 Bypass.close @16:18:40.55 UserAbort in brent#2 -- @16:19:02.35 -> #The peak pressure was 5.6psi. Does that sound about right? @16:19:59.48 -> #could be we run the pump really slow backwards so it doesn't cavetate. @16:20:54.44 -> #Sure. Just to recap.. Open Intake and Bypass then pump slowly backwards? @16:21:18.45 -> SP.cfgs @16:21:31.88 -> SP[:slow].maxSpeed @16:21:41.23 -> SP[:prime].maxSpeed @16:21:49.07 -> SP[:slow].maxSpeed=500 @16:21:51.16 -> #That's what killSampleLoop does, you'll need to look at the protocol I forget the SP speed we used. @16:22:12.55 -> exit @16:22:12.68 Session Ended @16:22:40.67 -> #It must have been SP[:slow]. I've reduced that speed from 2000 to 500. @16:22:53.44 -> # yes the protocol uses slow @16:22:57.06 -> #We should see pressure increasing. @16:23:07.46 -> Intake.open; Bypass.open @16:23:07.50 Intake.open @16:23:08.59 Bypass.open @16:23:25.84 -> loop {Delay.sleep 2; log.comment Sampler.PSI} @16:23:27.99 0.168028950333333 @16:23:30.08 0.154322661333332 @16:23:32.17 0.151276819333333 @16:23:34.21 UserAbort in brent#2 -- @16:23:38.91 -> loop {Delay.sleep 2; log.comment Sampler.PSI} @16:23:41.01 0.163460187333335 @16:23:43.09 0.160414345333333 @16:23:45.25 0.168028950333333 @16:23:47.38 0.161937266333332 @16:23:49.47 0.169551871333333 @16:23:51.55 0.148230977333334 @16:23:53.74 0.160414345333333 @16:23:55.66 -> SP.pump 100, :slow @16:23:55.96 SamplePump.reconfigure :slow @16:23:56.21 SamplePump.seek 72ml @16:23:56.55 0.181735239333335 @16:23:58.64 0.346210707333334 @16:24:00.73 0.242652079333332 @16:24:03.20 0.236560395333335 @16:24:05.13 SamplePump.hold @16:24:05.60 UserAbort in brent#2 -- Sample Pump (SP) seek to 72ml Aborted! @16:24:06.02 0.0751307693333345 @16:24:08.15 0.142139293333333 @16:24:09.78 -> SP.pump -100, :slow @16:24:09.91 SamplePump.seek -126.92ml @16:24:10.22 -0.150261538666665 @16:24:12.31 -0.0573633576666666 @16:24:14.51 0.624905250333335 @16:24:16.70 1.19447770433333 @16:24:19.40 1.51124527233333 @16:24:21.47 1.71074792333333 @16:24:23.68 1.69551871333333 @16:24:25.79 1.78993981533333 @16:24:27.87 1.77471060533333 @16:24:29.95 1.92243394233333 @16:24:32.18 1.93766315233333 @16:24:34.40 1.74882094833333 @16:24:36.51 1.64983108333333 @16:24:38.59 1.81735239333333 @16:24:39.92 SamplePump.hold @16:24:40.14 UserAbort in brent#2 -- Sample Pump (SP) seek to -126.92ml Aborted! @16:24:40.77 1.66506029333333 @16:24:43.00 0.958424949333335 @16:24:44.20 UserAbort in brent -- @16:25:49.97 -> #I agree. The intake is blocked. Pressure plateaued when pumping bleach out the Exhaust. @16:28:38.03 -> exit @16:28:38.06 Session Ended @16:44:41.19 Session Ended @16:46:04.89 Session Ended @16:51:40.23 Session Begun @16:51:50.24 -> ESP.closeValves @16:51:50.56 Intake.close @16:51:50.87 Bypass.close @16:58:01.26 Session Begun @16:58:33.76 -> Power @16:59:11.35 Session Begun @16:59:28.73 -> hidelog @16:59:37.94 -> loop {Delay.sleep 2; log.comment Sampler.PSI} @16:59:40.11 -0.176151195666666 @16:59:42.34 -0.189857484666668 @16:59:44.42 -0.183765800666666 @16:59:46.50 -0.208132536666668 @16:59:48.85 -0.188334563666668 @16:59:51.12 -0.185288721666666 @16:59:53.31 -0.167013669666666 @16:59:55.40 -0.191380405666667 @16:59:56.23 -> BACsampler.prime 60,60 @16:59:56.37 Priming sample loop w/60ml, bypass w/60ml @16:59:56.39 SamplePump.setPosition! 0ml @16:59:56.50 Bypass.open @16:59:57.49 -0.205086694666665 @16:59:58.19 Intake.open @16:59:59.35 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @16:59:59.87 0.215239501333333 @17:00:01.96 Vacuum tare is 18% of full scale @17:00:02.02 Cartridge::Sampler::Warning in brent#2 -- Vacuum sensor saturates below -17psi @17:00:02.66 -0.0126910083333307 @17:00:03.26 Pressure.reconfigure :default @17:00:03.57 Pressure.hold @17:00:03.84 SamplePump.reconfigure :prime @17:00:03.95 SamplePump.seek 60ml,5:31 @17:00:04.80 1.74019106266667 @17:00:06.88 2.28235093866667 @17:00:09.14 2.36763451466667 @17:00:11.37 2.46662437966667 @17:00:13.48 1.68688882766667 @17:00:15.56 1.18737073966667 @17:00:17.64 SamplePump.coast @17:00:18.07 Intake.close @17:00:18.29 Bypass.close @17:00:19.14 0.317782848666667 @17:00:19.50 Pressure.coast @17:00:19.76 Cartridge::Sampler::Leaky in brent#2 -- Serious Leak or Stall -- 0.93ml/s made only 0.7psi vacuum @17:00:21.39 0.337580821666666 @17:00:23.48 0.319305769666666 @17:00:25.56 0.290370270666667 @17:00:27.75 0.313214085666669 @17:00:29.93 0.328443295666666 @17:00:32.21 0.323874532666668 @17:00:34.31 0.319305769666666 @17:00:36.39 0.326920374666667 @17:00:38.16 -> BACsampler.prime 60,nil @17:00:38.48 Priming sample loop w/60ml @17:00:38.55 SamplePump.setPosition! 0ml @17:00:38.74 Bypass.open @17:00:39.11 0.331489137666669 @17:00:40.53 Intake.open @17:00:41.33 0.317782848666667 @17:00:41.79 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @17:00:43.40 0.291893191666666 @17:00:44.63 Vacuum tare is 18% of full scale @17:00:44.66 Cartridge::Sampler::Warning in brent#2 -- Vacuum sensor saturates below -17psi @17:00:44.89 Exhaust.open @17:00:45.48 -0.0147215696666692 @17:00:46.32 Bypass.close @17:00:47.70 -0.0299507796666667 @17:00:48.21 Pressure.reconfigure :default @17:00:48.47 Pressure.hold @17:00:49.30 SamplePump.reconfigure :prime @17:00:49.64 SamplePump.seek 60ml,5:31 @17:00:50.24 0.16650602933333 @17:00:52.38 5.69623218033333 @17:00:52.99 SamplePump.coast @17:00:53.09 Intake.close @17:00:53.14 Exhaust.close @17:00:54.06 Pressure.coast @17:00:54.15 Cartridge::Sampler::Leaky in brent#2 -- Serious Leak or Stall -- 0.93ml/s made only 6.4psi vacuum @17:00:54.46 4.68805847833333 @17:00:56.65 4.68196679433333 @17:00:58.81 4.64389376933333 @17:01:01.09 4.36063046333333 @17:01:03.18 3.99817526533333 @17:01:05.26 3.69815982833333 @17:01:06.40 -> BACsampler.prime 60,nil @17:01:06.90 Priming sample loop w/60ml @17:01:07.06 SamplePump.setPosition! 0ml @17:01:07.26 Bypass.open @17:01:07.65 3.41794236433333 @17:01:09.03 Intake.open @17:01:09.79 0.761968140333334 @17:01:10.13 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @17:01:11.88 0.0873141373333333 @17:01:12.75 Vacuum tare is 18% of full scale @17:01:12.78 Cartridge::Sampler::Warning in brent#2 -- Vacuum sensor saturates below -17psi @17:01:13.01 Exhaust.open @17:01:13.96 0.0131986486666698 @17:01:14.45 Bypass.close @17:01:16.42 0.00101528066666745 @17:01:16.51 Pressure.reconfigure :default @17:01:17.00 Pressure.hold @17:01:17.85 SamplePump.reconfigure :prime @17:01:18.11 SamplePump.seek 60ml,5:31 @17:01:18.64 0.288847349666668 @17:01:21.03 5.24443228366667 @17:01:21.26 SamplePump.coast @17:01:21.35 Intake.close @17:01:21.46 Exhaust.close @17:01:22.26 Pressure.coast @17:01:22.35 Cartridge::Sampler::Leaky in brent#2 -- Serious Leak or Stall -- 0.93ml/s made only 5.8psi vacuum @17:01:23.11 4.60023670066667 @17:01:25.18 4.57586996466667 @17:01:27.28 4.56368659666667 @17:01:29.38 4.53170525566667 @17:01:31.46 4.24387318666667 @17:01:33.54 3.95908695966667 @17:01:35.91 3.70628207366667 @17:01:38.11 3.47784392366667 @17:01:40.20 3.43672505666667 @17:01:42.30 3.43672505666667 @17:01:44.48 3.43063337266667 @17:01:45.41 UserAbort in brent -- @17:01:56.17 -> ESP.closeValves @17:04:08.15 -> exit @17:04:08.23 Session Ended @17:15:05.59 -> quit @17:15:05.63 Session Ended @17:39:12.68 -> exit @17:39:12.70 Session Ended @17:58:24.65PST24-Feb-21
>BREAK< ESP3 Gateway v1.11 for core bus @17:58:24.68 >BREAK< ESP3 Gateway v1.11 for cartridge bus @17:58:27.87 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @17:58:28.53 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @17:58:35.00 Polling Can every 2:00 @17:58:35.10 Gate.power :analytic1,:ON,limit to 5000mA @17:58:35.11 Gate.power :analytic2,:ON,limit to 5000mA @17:58:35.12 Gate.power :analytic3,:ON,limit to 5000mA @17:58:35.36 Archive.configure :default @17:58:35.41 Archive.hold @17:58:35.46 Archive.configure :default @17:58:36.02 Air.configure :default @17:58:36.07 Air.hold @17:58:36.13 Air.configure :default @17:58:36.23 Lysis1.configure :default @17:58:36.29 Lysis1.hold @17:58:36.34 Lysis1.configure :default @17:58:36.46 Lysis2.configure :default @17:58:36.51 Lysis2.hold @17:58:36.56 Lysis2.configure :default @17:58:36.70 Plunge.configure :default @17:58:36.75 Plunge.hold @17:58:36.81 Plunge.configure :default @17:58:36.96 SamplePump.configure :default @17:58:37.09 ExtraPump.configure :default @17:58:37.23 Toroid.configure :default @17:58:37.29 Toroid.hold @17:58:37.35 Toroid.configure :default @17:58:37.48 Carousel.configure :default @17:58:37.54 Carousel.hold @17:58:37.60 Carousel.configure :default @17:58:37.73 Pressure.configure :default @17:58:37.79 All elf controllers running firmware version 4.61 @17:58:37.87 Intake.close @17:58:37.98 Exhaust.close @17:58:38.00 Bypass.close @17:58:38.52
Clients may connect on TCP port 9999 @17:58:38.54 Waiting for espclients... @18:07:58.42 Session Begun @18:08:01.71 -> exit @18:08:01.73 Session Ended @18:08:10.17 Session Begun @18:08:12.24 -> upload @18:08:27.53 -> exit @18:08:27.55 Session Ended @10:16:38.32PST25-Feb-21 Session Begun @10:16:42.24 -> ESP::Mode @10:16:55.96 -> require 'vacuumCart' @10:19:44.14 Session Ended @10:20:58.71 Session Begun @10:21:59.38 -> require 'vacuumCart' @10:22:14.51 Session Begun @10:22:20.04 -> require 'vacuumCart' @10:22:24.73 -> refillSampleLines @10:22:24.76 Refilling sample & bypass lines with fluid. @10:22:24.96 Waiting for preston to answer "Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return." @10:22:26.31 Answered Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return."" @10:22:26.43 Bypass.open @10:22:27.23 Intake.open @10:22:28.06 SamplePump.setPosition! 0ml @10:22:28.15 SamplePump.reconfigure :fast @10:22:28.25 SamplePump.seek 30ml @10:22:54.89 Bypass.close @10:22:55.74 Exhaust.open @10:22:56.47 SamplePump.seek 160ml @10:23:16.33 SamplePump.hold @10:23:16.55 delay 1 second @10:23:17.58 Intake.close @10:23:17.62 Exhaust.close @10:23:18.65 UserAbort in preston -- Sample Pump (SP) seek to 160ml Aborted! @10:24:12.12 -> killSampleLoop! @10:24:12.14 SamplePump.setPosition! 0ml @10:24:12.26 Exhaust.open @10:24:12.96 Intake.open @10:24:14.48 SamplePump.reconfigure :slow @10:24:14.72 SamplePump.seek -22ml @10:25:02.21 delay 1 second @10:25:03.29 Bypass.open @10:25:04.14 Exhaust.close @10:25:05.22 SamplePump.seek -20ml @10:25:09.80 delay 1 second @10:25:10.96 SamplePump.seek -28ml @10:25:28.45 delay 1 second @10:25:29.70 Intake.close @10:25:29.98 Bypass.close @10:25:50.56 -> refillSampleLines @10:25:50.62 Refilling sample & bypass lines with fluid. @10:25:50.77 Waiting for preston to answer "Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return." @10:25:52.15 Answered Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return."" @10:25:52.26 Bypass.open @10:25:53.00 Intake.open @10:25:53.80 SamplePump.setPosition! 0ml @10:25:53.88 SamplePump.reconfigure :fast @10:25:53.99 SamplePump.seek 30ml @10:26:20.56 Bypass.close @10:26:21.47 Exhaust.open @10:26:22.19 SamplePump.seek 160ml @10:28:15.36 delay 1 second @10:28:16.40 Intake.close @10:28:16.62 Exhaust.close @10:28:17.69 Sampling Loop filled with water @10:35:25.58 -> refillSampleLines @10:35:25.64 Refilling sample & bypass lines with fluid. @10:35:25.74 Waiting for preston to answer "Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return." @10:35:26.88 Answered Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return."" @10:35:27.02 Bypass.open @10:35:27.85 Intake.open @10:35:28.66 SamplePump.setPosition! 0ml @10:35:28.75 SamplePump.seek 30ml @10:35:55.31 Bypass.close @10:35:56.04 Exhaust.open @10:35:56.75 SamplePump.seek 160ml @10:37:49.89 delay 1 second @10:37:50.95 Intake.close @10:37:51.22 Exhaust.close @10:37:52.41 Sampling Loop filled with water @10:38:00.17 -> BACsampler.prime @10:38:00.30 Priming sample loop w/160ml, bypass w/10ml @10:38:00.32 SamplePump.setPosition! 0ml @10:38:00.43 Bypass.open @10:38:02.00 Intake.open @10:38:03.08 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @10:38:05.84 Vacuum tare is 18% of full scale @10:38:05.91 Cartridge::Sampler::Warning in preston -- Vacuum sensor saturates below -17psi @10:38:06.19 Pressure.reconfigure :default @10:38:06.28 Pressure.hold @10:38:06.54 SamplePump.reconfigure :prime @10:38:06.64 SamplePump.seek 10ml,1:21 @10:38:17.78 Pressure.coast @10:38:17.86 SamplePump.setPosition! 0ml @10:38:17.97 Exhaust.open @10:38:19.41 Bypass.close @10:38:21.23 Pressure.reconfigure :default @10:38:21.45 Pressure.hold @10:38:22.17 SamplePump.reconfigure :prime @10:38:22.35 SamplePump.seek 160ml,13:51 @10:38:34.93 SamplePump.hold @10:38:35.17 SamplePump.coast @10:38:35.27 Intake.close @10:38:35.36 Exhaust.close @10:38:36.27 Pressure.coast @10:38:36.37 UserAbort in preston -- Sample Pump (SP) seek to 160ml Aborted! @10:39:20.22 -> Intke.close @10:39:20.31 NameError in preston -- uninitialized constant Intke @10:39:24.51 -> Intake.close @10:39:28.19 -> Exhaust.close @10:39:40.25 -> exit @10:39:40.27 Session Ended @10:51:36.90 -> #quit @10:51:38.17 -> exit @10:51:38.19 Session Ended @13:37:18.05 Session Begun @13:37:29.86 -> refillSampleLines @13:37:29.97 Refilling sample & bypass lines with fluid. @13:37:30.08 Waiting for preston to answer "Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return." @13:37:30.91 Answered Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return."" @13:37:31.02 Bypass.open @13:37:31.91 Intake.open @13:37:32.81 SamplePump.setPosition! 0ml @13:37:33.11 SamplePump.reconfigure :fast @13:37:33.29 SamplePump.seek 30ml @13:37:59.99 Bypass.close @13:38:00.85 Exhaust.open @13:38:01.56 SamplePump.seek 160ml @13:39:55.17 delay 1 second @13:39:56.20 Intake.close @13:39:56.28 Exhaust.close @13:39:57.27 Sampling Loop filled with water @13:40:02.25 -> BACsampler.prime @13:40:02.36 Priming sample loop w/160ml, bypass w/10ml @13:40:02.39 SamplePump.setPosition! 0ml @13:40:02.45 Bypass.open @13:40:04.00 Intake.open @13:40:04.89 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @13:40:07.55 Vacuum tare is 18% of full scale @13:40:07.67 Cartridge::Sampler::Warning in preston -- Vacuum sensor saturates below -17psi @13:40:08.32 Pressure.reconfigure :default @13:40:08.53 Pressure.hold @13:40:09.08 SamplePump.reconfigure :prime @13:40:09.32 SamplePump.seek 10ml,1:21 @13:40:20.56 Pressure.coast @13:40:20.71 SamplePump.setPosition! 0ml @13:40:20.90 Exhaust.open @13:40:22.41 Bypass.close @13:40:24.48 Pressure.reconfigure :default @13:40:24.56 Pressure.hold @13:40:24.86 SamplePump.reconfigure :prime @13:40:24.95 SamplePump.seek 160ml,13:51 @13:43:17.09 Pressure.coast @13:43:27.15 -> killSampleLoop! @13:43:27.17 SamplePump.setPosition! 0ml @13:43:27.32 SamplePump.reconfigure :slow @13:43:27.41 SamplePump.seek -22ml @13:44:14.91 delay 1 second @13:44:15.94 Bypass.open @13:44:16.69 Exhaust.close @13:44:17.73 SamplePump.seek -20ml @13:44:22.47 delay 1 second @13:44:23.56 SamplePump.seek -28ml @13:44:41.00 delay 1 second @13:44:42.09 Intake.close @13:44:42.18 Bypass.close @13:44:49.50 -> PB @13:44:59.03 -> exit @13:44:59.12 Session Ended @18:18:11.55 Session Begun @18:18:26.62 -> BACsampler.prime @18:18:27.07 Priming sample loop w/160ml, bypass w/10ml @18:18:27.32 SamplePump.setPosition! 0ml @18:18:27.40 Bypass.open @18:18:28.94 Intake.open @18:18:30.62 Pressure.reconfigure :default @18:18:30.71 Pressure.hold @18:18:30.96 SamplePump.reconfigure :prime @18:18:31.05 SamplePump.seek 10ml,1:21 @18:18:42.23 Pressure.coast @18:18:42.37 SamplePump.setPosition! 0ml @18:18:42.57 Exhaust.open @18:18:44.08 Bypass.close @18:18:46.03 Pressure.reconfigure :default @18:18:46.11 Pressure.hold @18:18:46.40 SamplePump.reconfigure :prime @18:18:46.49 SamplePump.seek 160ml,13:51 @18:21:38.12 Pressure.coast @18:24:39.48 -> killSampleLoop! @18:24:39.51 SamplePump.setPosition! 0ml @18:24:39.64 SamplePump.reconfigure :slow @18:24:39.73 SamplePump.seek -22ml @18:25:27.05 delay 1 second @18:25:28.08 Bypass.open @18:25:28.84 Exhaust.close @18:25:29.88 SamplePump.seek -20ml @18:25:34.45 delay 1 second @18:25:35.54 SamplePump.seek -28ml @18:25:52.92 delay 1 second @18:25:54.01 Intake.close @18:25:54.11 Bypass.close @18:26:42.47 -> Exhaust @18:26:44.23 -> exit @18:26:44.30 Session Ended @07:04:57.94PST26-Feb-21 Session Begun @07:05:17.05 -> BACsampler.prime @07:05:17.52 Priming sample loop w/160ml, bypass w/10ml @07:05:17.55 SamplePump.setPosition! 0ml @07:05:17.79 Bypass.open @07:05:19.58 Intake.open @07:05:21.48 Pressure.reconfigure :default @07:05:21.57 Pressure.hold @07:05:21.81 SamplePump.reconfigure :prime @07:05:21.91 SamplePump.seek 10ml,1:21 @07:05:33.03 Pressure.coast @07:05:33.10 SamplePump.setPosition! 0ml @07:05:33.19 Exhaust.open @07:05:34.60 Bypass.close @07:05:36.43 Pressure.reconfigure :default @07:05:36.65 Pressure.hold @07:05:37.29 SamplePump.reconfigure :prime @07:05:37.47 SamplePump.seek 160ml,13:51 @07:08:28.80 Pressure.coast @07:09:12.99 -> killSampleLoop! @07:09:13.05 SamplePump.setPosition! 0ml @07:09:13.23 SamplePump.reconfigure :slow @07:09:13.32 SamplePump.seek -22ml @07:10:00.59 delay 1 second @07:10:01.62 Bypass.open @07:10:02.32 Exhaust.close @07:10:03.11 SamplePump.seek -20ml @07:10:07.83 delay 1 second @07:10:08.92 SamplePump.seek -28ml @07:10:26.29 delay 1 second @07:10:27.38 Intake.close @07:10:27.45 Bypass.close @07:16:53.70 -> exit @07:16:53.75 Session Ended @08:06:14.68PST01-Mar-21
>BREAK< ESP3 Gateway v1.11 for core bus @08:06:14.71 >BREAK< ESP3 Gateway v1.11 for cartridge bus @08:06:21.81 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @08:06:22.13 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @08:06:25.15 Polling Can every 2:00 @08:06:25.60 Clients may connect on TCP port 9999 @08:06:25.62 Waiting for espclients... @09:23:12.23 Session Begun @09:23:15.15 -> exit @09:23:15.17 Session Ended @11:01:47.32 Session Begun @11:01:55.82 -> BACsampler.prime @11:01:55.86 I2C::NodeOffline in preston -- carousel0 microcontroller never signed on while reading Storage Carousel (SC)'s status @11:02:00.27 -> ESP.ready! @11:02:00.63 Gate.power :main,:ON,limit to 5000mA @11:02:00.69 Gate.power :analytic1,:ON,limit to 5000mA @11:02:00.75 Gate.power :analytic2,:ON,limit to 5000mA @11:02:00.79 Gate.power :analytic3,:ON,limit to 5000mA @11:02:00.94 Powering core bus @11:02:03.33 Archive.configure :default @11:02:03.50 Air.configure :default @11:02:03.64 Lysis1.configure :default @11:02:03.77 Lysis2.configure :default @11:02:03.94 Plunge.configure :default @11:02:04.10 SamplePump.configure :default @11:02:04.26 ExtraPump.configure :default @11:02:04.43 Toroid.configure :default @11:02:04.60 Carousel.configure :default @11:02:04.76 Pressure.configure :default @11:02:04.84 All elf controllers running firmware version 4.61 @11:02:04.88 Gate.power :motor,:ON,limit to 5000mA @11:02:05.14 Archive.reconfigure :default @11:02:05.56 Plunge.reconfigure :default @11:02:05.60 Archive.home.jog 0 @11:02:05.64 Lysis1.reconfigure :default @11:02:05.70 Air.reconfigure :default @11:02:05.74 Toroid.reconfigure :default @11:02:05.80 Lysis2.reconfigure :default @11:02:06.11 Plunge.home.jog 0 @11:02:06.16 Lysis1.home.jog 0 @11:02:06.20 Air.home.jog 0 @11:02:06.25 Toroid.home.jog 0 @11:02:06.29 Lysis2.home.jog 0 @11:02:08.70 Lysis2.reconfigure :homing @11:02:08.82 Plunge.reconfigure :homing @11:02:08.98 Toroid.reconfigure :homing @11:02:09.14 Archive.reconfigure :homing @11:02:09.28 Lysis1.reconfigure :homing @11:02:09.40 Air.reconfigure :homing @11:02:09.86 Lysis2.jog 1345 @11:02:10.00 Plunge.jog 300 @11:02:10.10 Toroid.jog 634 @11:02:10.28 Lysis1.jog 1353 @11:02:10.35 Archive.jog 1355 @11:02:10.60 Air.jog 1357 @11:02:11.41 Plunge.setRawPosition! 50 @11:02:11.48 Plunge.seek :home @11:02:11.76 Lysis2.setRawPosition! 50 @11:02:11.84 Lysis2.seek :home @11:02:12.32 Plunge.reconfigure :default @11:02:12.69 Lysis2.reconfigure :default @11:02:12.73 Lysis1.setRawPosition! 50 @11:02:12.77 Air.setRawPosition! 50 @11:02:12.80 Archive.setRawPosition! 50 @11:02:12.99 Lysis1.seek :home @11:02:13.10 Air.seek :home @11:02:13.15 Archive.seek :home @11:02:13.39 Toroid.setRawPosition! 0 @11:02:13.46 Toroid.seek :home @11:02:13.66 Toroid.reconfigure :default @11:02:13.69 Lysis1.reconfigure :default @11:02:13.73 Archive.reconfigure :default @11:02:13.79 Air.reconfigure :default @11:02:14.35 Toroid.seek :clear @11:02:16.00 Carousel.reconfigure :default @11:02:16.09 Carousel.home.jog 0 @11:02:16.62 Carousel.seek 60 @11:02:23.50 -> BACsampler.prime @11:02:23.61 Priming sample loop w/160ml, bypass w/10ml @11:02:23.63 SamplePump.setPosition! 0ml @11:02:23.71 Bypass.open @11:02:25.22 Intake.open @11:02:26.15 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @11:02:28.79 Vacuum tare is 18% of full scale @11:02:28.88 Cartridge::Sampler::Warning in preston -- Vacuum sensor saturates below -17psi @11:02:29.42 Pressure.reconfigure :default @11:02:29.62 Pressure.hold @11:02:30.10 SamplePump.reconfigure :prime @11:02:30.30 SamplePump.seek 10ml,1:21 @11:02:41.49 Pressure.coast @11:02:41.56 SamplePump.setPosition! 0ml @11:02:41.64 Exhaust.open @11:02:43.11 Bypass.close @11:02:44.68 Pressure.reconfigure :default @11:02:44.77 Pressure.hold @11:02:45.06 SamplePump.reconfigure :prime @11:02:45.15 SamplePump.seek 160ml,13:51 @11:05:36.70 Pressure.coast @11:05:59.32 -> killSampleLoop! @11:05:59.34 SamplePump.setPosition! 0ml @11:05:59.48 SamplePump.reconfigure :slow @11:05:59.57 SamplePump.seek -22ml @11:06:46.96 delay 1 second @11:06:48.07 Bypass.open @11:06:48.99 Exhaust.close @11:06:49.89 SamplePump.seek -20ml @11:06:54.47 delay 1 second @11:06:55.65 SamplePump.seek -28ml @11:07:13.21 delay 1 second @11:07:14.46 Intake.close @11:07:14.74 Bypass.close @11:08:45.25 -> exit @11:08:45.27 Session Ended @12:07:58.35 Session Begun @12:08:00.82 -> slots @12:08:15.74 -> Cartridge.release 36 @12:08:19.34 -> slots @12:08:41.57 -> proceessAll @12:09:29.17 -> processAll @12:09:39.90 -> execute 'processAll' @12:09:40.20 Processing dry cartridges: [36] With: {spare: 4} @12:09:40.26 Sampling default goal volume per cartridge type loaded @12:09:40.35 Fixing samples for 10:00, evacuating once @12:09:40.39 delay 30 seconds @12:10:11.19 Carousel.seek 60 @12:10:11.37 Selecting Cartridge 36 @12:10:11.62 Carousel.seek 36 @12:10:17.19 Lysis1.seek :locked @12:10:20.45 Plunge.seek :locked @12:10:20.63 Air.seek :locked @12:10:25.11 Mounted archive Cartridge @12:10:25.98 Duration of filtering limited to 1:05:00 @12:10:26.60 Bypass.open @12:10:27.58 Toroid.reconfigure :hold @12:10:27.73 Toroid.seek :crack @12:10:29.41 Toroid.seek :cartridge @12:10:30.37 Toroid.reconfigure :default @12:10:30.46 Toroid.seek :clear @12:10:32.13 Exhaust.open @12:10:33.06 Pressurizing cartridge to 26psi above 14.6psia @12:10:33.08 SamplePump.setPosition! 0ml @12:10:33.16 Intake.open @12:10:34.12 Bypass.close @12:10:34.96 Exhaust.close @12:10:36.13 Pressure.reconfigure :default @12:10:36.37 Pressure.hold @12:10:36.95 SamplePump.reconfigure :charge @12:10:37.19 SamplePump.seek 2.3ml,42 seconds @12:10:46.65 Pressure.coast @12:10:46.89 Intake.close @12:10:48.17 Pressurized cartridge with 0.398ml to 26.3psi @12:10:48.18 Isolated @34.7psia (20.1psi) @12:10:48.21 Checking for cartridge leaks from 34.7psia @12:10:48.28 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:10:52.95 Cartridge pressure fell 3% to 33.6psia @12:10:53.28 Sampling 1000.0ml @12:10:53.31 SamplePump.setPosition! 0ml @12:10:53.48 Exhaust.open @12:10:54.26 Intake.open @12:10:55.68 Pressure.reconfigure :default @12:10:55.84 Pressure.hold @12:10:56.56 SamplePump.reconfigure :sample @12:10:56.74 SamplePump.seek 1000ml,1:23:51 @12:10:58.57 Cartridge 36 sampling 1/1000ml at 0.52ml/s with 21.4psi vacuum @12:11:58.81 Cartridge 36 sampling 29/1000ml at 0.47ml/s with 25.6psi vacuum @12:12:58.58 Cartridge 36 sampling 56/1000ml at 0.44ml/s with 23.8psi vacuum @12:13:58.75 Cartridge 36 sampling 83/1000ml at 0.43ml/s with 20.6psi vacuum @12:14:58.58 Cartridge 36 sampling 108/1000ml at 0.40ml/s with 27.3psi vacuum @12:15:58.74 Cartridge 36 sampling 132/1000ml at 0.39ml/s with 18.1psi vacuum 2.2m deep (17.9psia) @12:16:58.57 Cartridge 36 sampling 155/1000ml at 0.36ml/s with 26.6psi vacuum @12:17:58.82 Cartridge 36 sampling 176/1000ml at 0.34ml/s with 25.9psi vacuum @12:18:58.57 Cartridge 36 sampling 196/1000ml at 0.32ml/s with 26.1psi vacuum @12:19:58.77 Cartridge 36 sampling 214/1000ml at 0.29ml/s with 27.3psi vacuum @12:20:58.57 Cartridge 36 sampling 232/1000ml at 0.29ml/s with 26.1psi vacuum @12:21:58.60 Cartridge 36 sampling 248/1000ml at 0.29ml/s with 22.5psi vacuum @12:22:58.57 Cartridge 36 sampling 264/1000ml at 0.25ml/s with 20.4psi vacuum @12:23:58.98 Cartridge 36 sampling 278/1000ml at 0.23ml/s with 22.2psi vacuum @12:23:59.04 Errno::EPIPE in preston's_reader -- Broken pipe @12:23:59.18 Session Ended @12:24:58.87 Cartridge 36 sampling 292/1000ml at 0.20ml/s with 26.7psi vacuum @12:25:58.75 Cartridge 36 sampling 306/1000ml at 0.20ml/s with 27.5psi vacuum @12:26:58.57 Cartridge 36 sampling 319/1000ml at 0.20ml/s with 25.2psi vacuum @12:27:58.80 Cartridge 36 sampling 331/1000ml at 0.21ml/s with 19.8psi vacuum 2.3m deep (18.0psia) @12:28:58.57 Cartridge 36 sampling 344/1000ml at 0.20ml/s with 22.8psi vacuum @12:29:58.84 Cartridge 36 sampling 356/1000ml at 0.20ml/s with 29.8psi vacuum @12:30:58.57 Cartridge 36 sampling 368/1000ml at 0.20ml/s with 27.7psi vacuum @12:31:53.93 Flow averaged 0.30ml/s over 20:53 @12:31:53.97 Waiting up to 23 seconds for vacuum to reduce from 32.0 to 18.0psi @12:31:54.55 Relieved excess vacuum in 5.8 seconds @12:31:54.58 SamplePump.seek 1000ml,52:14 @12:31:58.72 Cartridge 36 sampling 380/1000ml at 0.21ml/s with 22.0psi vacuum 3.0m deep (19.1psia) @12:32:02.56 Flow estimate reduced to 0.24ml/s @12:32:02.59 Waiting up to 15 seconds for vacuum to reduce from 28.6 to 18.0psi @12:32:03.17 Relieved excess vacuum in 4.4 seconds @12:32:03.20 SamplePump.seek 1000ml,52:06 @12:32:11.20 Flow estimate reduced to 0.19ml/s @12:32:11.23 Intake.close @12:32:11.29 Exhaust.close @12:32:11.50 Toroid.reconfigure :hold @12:32:11.56 Toroid.seek :bypass @12:32:13.19 Toroid.reconfigure :default @12:32:13.25 Toroid.seek :clear @12:32:14.85 Cartridge::Sampler::Clogged in FILTERING -- Cartridge clogged @12:32:14.87 Pressure.coast @12:32:15.09 Sampled 382.5ml @12:40:33.93 Session Begun @12:40:36.17 -> showlog @12:40:44.78 -> slots @12:40:54.58 -> exit @12:40:54.62 Session Ended @12:42:25.72 Session Begun @12:46:47.52 -> Cmd.process @12:46:47.95 SamplePump.setPosition! 0ml @12:46:48.53 Plunge.seek :reagent @12:46:48.56 Exhaust.open @12:46:49.39 Intake.open @12:46:49.61 Sample loop closed. Ready to Deliver Reagents @12:46:49.69 Archive.seek :sealed @12:46:50.56 SamplePump.reconfigure :slow @12:46:50.77 SamplePump.seek -22ml @12:46:53.45 Archive.seek :evac @12:47:07.54 delay 1:00 @12:47:38.23 delay 1 second @12:47:39.27 Bypass.open @12:47:40.00 Exhaust.close @12:47:40.90 SamplePump.seek -20ml @12:47:45.53 delay 1 second @12:47:46.70 SamplePump.seek -28ml @12:48:04.24 delay 1 second @12:48:05.52 Intake.close @12:48:05.68 Bypass.close @12:48:07.58 Archive.seek :clear @12:48:24.83 RNAlater soak for 10:00 @12:58:24.92 Archive.seek :tVent @12:58:26.67 delay 1 second @12:58:27.71 Archive.seek :evac @12:58:43.73 delay 10 seconds @12:58:53.81 Sample preserved and evacuated @13:00:08.59 -> exit @13:00:08.62 Session Ended @13:01:22.80 Session Begun @13:01:28.19 -> MainThread.exit @13:01:28.73 ThreadError in preston -- killed thread @13:01:28.88 Session Ended @14:07:46.88PST01-Mar-21
>BREAK< ESP3 Gateway v1.11 for core bus @14:07:47.05 >BREAK< ESP3 Gateway v1.11 for cartridge bus @14:07:52.85 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @14:07:53.15 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @14:07:56.04 Polling Can every 2:00 @14:07:56.42 Gate.power :analytic1,:ON,limit to 5000mA @14:07:56.46 Gate.power :analytic2,:ON,limit to 5000mA @14:07:56.50 Gate.power :analytic3,:ON,limit to 5000mA @14:07:58.66 Archive.configure :default @14:07:58.86 Archive.hold @14:07:58.95 Archive.configure :default @14:07:59.30 Air.configure :default @14:07:59.43 Air.hold @14:07:59.51 Air.configure :default @14:07:59.80 Lysis1.configure :default @14:07:59.95 Lysis1.hold @14:08:00.08 Lysis1.configure :default @14:08:00.37 Lysis2.configure :default @14:08:00.52 Lysis2.hold @14:08:00.62 Lysis2.configure :default @14:08:00.88 Plunge.configure :default @14:08:01.04 Plunge.hold @14:08:01.14 Plunge.configure :default @14:08:01.45 SamplePump.configure :default @14:08:01.90 ExtraPump.configure :default @14:08:02.44 Toroid.configure :default @14:08:02.49 Toroid.hold @14:08:02.55 Toroid.configure :default @14:08:02.70 Carousel.configure :default @14:08:02.75 Carousel.hold @14:08:02.81 Carousel.configure :default @14:08:02.95 Pressure.configure :default @14:08:03.01 All elf controllers running firmware version 4.61 @14:08:03.10 Intake.close @14:08:03.15 Exhaust.close @14:08:03.23 Bypass.close @14:08:03.40
Loading /home/esp/esp2/bin/processAll mission @14:08:04.28 Session Begun @14:08:05.02
Sampling default goal volumes @14:08:05.09 GoodNews.email "Cartridge group: dry\nSpares: {spare: 4}\n\n",Subject: "Begun processAll --mail --checkpoint" @14:08:05.82
Carousel.seek 36 @14:08:06.76 Mounted archive Cartridge @14:08:07.05 Selecting Cartridge 35 @14:08:08.26 Archive.seek :clear @14:08:08.87 Air.seek :clear @14:08:24.29 upsync @14:08:25.75 Lysis1.seek :clear @14:08:25.93 Plunge.seek :clear @14:08:29.03
Carousel.seek 35 @14:08:29.86 Lysis1.seek :locked @14:08:33.24 Plunge.seek :locked @14:08:33.41 Air.seek :locked @14:08:37.90
Mounted archive Cartridge @14:08:38.15 GoodNews.email "Filtering up to 1000ml",Subject: "Sampling archive_bac cartridge in slot 35" @14:08:38.97 Duration of filtering limited to 1:05:00 @14:08:39.76 Priming sample loop w/160ml, bypass w/10ml @14:08:39.79 SamplePump.setPosition! 0ml @14:08:40.05 Bypass.open @14:08:41.65 Intake.open @14:08:42.58 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @14:08:45.67 Vacuum tare is 19% of full scale @14:08:45.69 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @14:08:45.96 Pressure.reconfigure :default @14:08:46.26 Pressure.hold @14:08:46.97 SamplePump.reconfigure :prime @14:08:47.36 SamplePump.seek 10ml,1:21 @14:08:56.97 EOFError in Qmail -- end of file reached @14:08:57.05 Retry #1 of 10 after 10:00 @14:08:58.64 Pressure.coast @14:08:58.79 SamplePump.setPosition! 0ml @14:08:58.96 Exhaust.open @14:09:00.44 Bypass.close @14:09:02.05 Pressure.reconfigure :default @14:09:02.14 Pressure.hold @14:09:02.45 SamplePump.reconfigure :prime @14:09:02.54 SamplePump.seek 160ml,13:51 @14:09:40.35 -> GoodNews @14:10:24.07 -> SP @14:11:54.18 Pressure.coast @14:11:54.33 Bypass.open @14:11:55.10 Intake.close @14:11:56.19 Toroid.reconfigure :hold @14:11:56.29 Toroid.seek :crack @14:11:57.91 Toroid.seek :cartridge @14:11:58.87 Toroid.reconfigure :default @14:11:58.96 Toroid.seek :clear @14:12:01.37 Pressurizing cartridge to 26psi above 14.3psia @14:12:01.47 SamplePump.setPosition! 0ml @14:12:01.68 Intake.open @14:12:02.63 Bypass.close @14:12:03.39 Exhaust.close @14:12:04.45 Pressure.reconfigure :default @14:12:04.54 Pressure.hold @14:12:05.33 SamplePump.reconfigure :charge @14:12:05.42 SamplePump.seek 2.3ml,42 seconds @14:12:14.93 Pressure.coast @14:12:15.01 Intake.close @14:12:15.92 Pressurized cartridge with 0.409ml to 26.1psi @14:12:15.93 Isolated @35.1psia (20.8psi) @14:12:15.97 Checking for cartridge leaks from 35.1psia @14:12:16.01 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @14:12:20.74 Cartridge pressure fell 2% to 34.2psia @14:12:21.01 Sampling 1000.0ml @14:12:21.03 SamplePump.setPosition! 0ml @14:12:21.20 Exhaust.open @14:12:22.07 Intake.open @14:12:23.26 Pressure.reconfigure :default @14:12:23.50 Pressure.hold @14:12:24.50 SamplePump.reconfigure :sample @14:12:24.65 SamplePump.seek 1000ml,1:23:51 @14:12:26.32 Cartridge 35 sampling 0/1000ml at 0.55ml/s with 23.6psi vacuum @14:13:26.52 Cartridge 35 sampling 30/1000ml at 0.47ml/s with 22.9psi vacuum @14:14:26.32 Cartridge 35 sampling 59/1000ml at 0.46ml/s with 26.3psi vacuum @14:15:26.49 Cartridge 35 sampling 88/1000ml at 0.48ml/s with 22.9psi vacuum @14:16:26.32 Cartridge 35 sampling 115/1000ml at 0.45ml/s with 26.0psi vacuum @14:17:26.36 Cartridge 35 sampling 143/1000ml at 0.45ml/s with 23.7psi vacuum @14:18:26.32 Cartridge 35 sampling 169/1000ml at 0.44ml/s with 22.6psi vacuum @14:19:18.68 upsync @14:19:26.58 Cartridge 35 sampling 195/1000ml at 0.43ml/s with 26.0psi vacuum @14:20:26.32 Cartridge 35 sampling 221/1000ml at 0.43ml/s with 22.3psi vacuum @14:21:26.37 Cartridge 35 sampling 246/1000ml at 0.42ml/s with 25.8psi vacuum @14:22:26.32 Cartridge 35 sampling 270/1000ml at 0.40ml/s with 20.7psi vacuum @14:23:26.70 Cartridge 35 sampling 294/1000ml at 0.39ml/s with 21.1psi vacuum @14:24:26.50 Cartridge 35 sampling 317/1000ml at 0.37ml/s with 20.1psi vacuum @14:25:26.33 Cartridge 35 sampling 339/1000ml at 0.35ml/s with 26.6psi vacuum @14:26:26.69 Cartridge 35 sampling 360/1000ml at 0.34ml/s with 26.5psi vacuum @14:27:26.33 Cartridge 35 sampling 381/1000ml at 0.34ml/s with 22.9psi vacuum @14:28:26.50 Cartridge 35 sampling 401/1000ml at 0.33ml/s with 26.9psi vacuum @14:29:26.33 Cartridge 35 sampling 420/1000ml at 0.31ml/s with 26.6psi vacuum @14:30:26.48 Cartridge 35 sampling 439/1000ml at 0.31ml/s with 25.5psi vacuum @14:31:26.33 Cartridge 35 sampling 457/1000ml at 0.31ml/s with 24.2psi vacuum @14:32:26.48 Cartridge 35 sampling 474/1000ml at 0.29ml/s with 24.6psi vacuum @14:33:26.33 Cartridge 35 sampling 491/1000ml at 0.27ml/s with 25.9psi vacuum @14:34:26.54 Cartridge 35 sampling 507/1000ml at 0.26ml/s with 21.4psi vacuum @14:35:26.33 Cartridge 35 sampling 522/1000ml at 0.23ml/s with 22.6psi vacuum @14:36:26.48 Cartridge 35 sampling 536/1000ml at 0.24ml/s with 20.7psi vacuum @14:37:26.33 Cartridge 35 sampling 549/1000ml at 0.20ml/s with 25.7psi vacuum @14:38:26.56 Cartridge 35 sampling 562/1000ml at 0.22ml/s with 22.8psi vacuum @14:39:26.48 Cartridge 35 sampling 575/1000ml at 0.22ml/s with 22.0psi vacuum @14:40:26.32 Cartridge 35 sampling 587/1000ml at 0.21ml/s with 27.2psi vacuum @14:41:26.68 Cartridge 35 sampling 600/1000ml at 0.20ml/s with 26.6psi vacuum @14:42:26.32 Cartridge 35 sampling 613/1000ml at 0.21ml/s with 25.5psi vacuum @14:43:26.50 Cartridge 35 sampling 625/1000ml at 0.21ml/s with 20.9psi vacuum @14:44:26.32 Cartridge 35 sampling 637/1000ml at 0.20ml/s with 21.3psi vacuum @14:45:09.86 -> Intake.close @14:45:09.90 Intake.close @14:45:11.45 Flow averaged 0.33ml/s over 32:47 @14:45:11.52 Waiting up to 2:25 for vacuum to reduce from 80.8 to 18.0psi @14:45:13.37 Relieved excess vacuum in 2.4 seconds @14:45:13.43 SamplePump.seek 1000ml,29:58 @14:45:14.74 Flow estimate reduced to 0.26ml/s @14:45:14.80 Waiting up to 2:15 for vacuum to reduce from 76.9 to 18.0psi @14:45:17.21 Relieved excess vacuum in 3.5 seconds @14:45:17.28 SamplePump.seek 1000ml,29:57 @14:45:18.87 Flow estimate reduced to 0.21ml/s @14:45:18.97 Waiting up to 2:25 for vacuum to reduce from 80.8 to 18.0psi @14:45:20.63 Relieved excess vacuum in 2.9 seconds @14:45:20.68 SamplePump.seek 1000ml,29:56 @14:45:22.05 Flow estimate reduced to 0.17ml/s @14:45:22.08 Exhaust.close @14:45:22.30 Toroid.reconfigure :hold @14:45:22.39 Toroid.seek :bypass @14:45:24.05 Toroid.reconfigure :default @14:45:24.14 Toroid.seek :clear @14:45:25.83 Cartridge::Sampler::Clogged in FILTERING -- Cartridge clogged @14:45:25.87 Pressure.coast @14:45:26.59 Sampled 647.2ml @14:45:28.16 GoodNews.email "After filtering 647.2ml",Subject: "Processing archive_bac cartridge in slot 35" @14:45:29.05 SamplePump.setPosition! 0ml @14:45:29.22 Plunge.seek :reagent @14:45:29.44 Exhaust.open @14:45:30.22 Intake.open @14:45:30.47 Sample loop closed. Ready to Deliver Reagents @14:45:30.50 Archive.seek :sealed @14:45:31.40 SamplePump.reconfigure :slow @14:45:31.70 SamplePump.seek -22ml @14:45:34.19 Archive.seek :evac @14:45:48.22 delay 1:00 @14:45:48.48 upsync @14:46:19.27 delay 1 second @14:46:20.30 Bypass.open @14:46:21.02 Exhaust.close @14:46:21.83 SamplePump.seek -20ml @14:46:26.45 delay 1 second @14:46:27.69 SamplePump.seek -28ml @14:46:45.33 delay 1 second @14:46:46.42 Intake.close @14:46:46.49 Bypass.close @14:46:48.25 Archive.seek :clear @14:47:05.44 RNAlater soak for 10:00 @14:57:05.46 Archive.seek :tVent @14:57:07.25 delay 1 second @14:57:08.42 Archive.seek :evac @14:57:24.50 delay 10 seconds @14:57:34.54 Sample preserved and evacuated @14:57:35.32
Carousel.seek 35 @14:57:35.92 Plunge.seek :locked @14:57:37.15 Archive.seek :clear @14:57:37.50 Air.seek :clear @14:57:37.84 Lysis1.seek :clear @14:57:54.53
Archive.coast @14:57:54.70 Air.coast @14:57:54.95 Lysis1.coast @14:57:55.20 Lysis2.coast @14:57:55.41 Plunge.coast @14:57:55.55 SamplePump.coast @14:57:55.78 Toroid.coast @14:57:55.99 Carousel.coast @14:57:56.28 Gate.power :motor,:OFF @14:57:56.37 Gate.power :core,:OFF @14:57:56.40 Gate.power :analytic1,:OFF @14:57:56.43 Gate.power :analytic2,:OFF @14:57:56.51 Gate.power :analytic3,:OFF @14:57:56.60 Gate.power :main,:OFF @14:57:56.65 Safely stopped and ready to power off @14:57:57.37 Cartridge Load: {35 => { Type: :archive_bac, State: :processed, filtered: 647.2}, 39..37 => { Type: :archive_bac, State: :processed, filtered: 1000}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 250}, [44, 42, 40] => { Type: :archive_bac, State: :processed, filtered: 250}, 36 => { Type: :archive_bac, State: :processed, filtered: 382.5}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 297}, [34..32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}} Spares: {spare: 4} @14:57:57.49 BadNews.email "Not enough dry cartridges remain",Subject: "Cartridge::None in MAIN" @14:58:14.81 upsync @14:58:14.95 Session Ended @00:16:42.55PST02-Mar-21
>BREAK< ESP3 Gateway v1.11 for core bus @00:16:42.70 >BREAK< ESP3 Gateway v1.11 for cartridge bus @00:16:47.10 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @00:16:47.41 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @00:16:53.07 Polling Can every 2:00 @00:16:53.25 Loading /home/esp/esp2/bin/processAll mission @00:16:56.02 Sampling default goal volumes @00:16:56.06 GoodNews.email "Cartridge group: dry\nSpares: {spare: 4}\n\n",Subject: "Begun processAll --mail --checkpoint" @00:16:56.26
Gate.power :main,:ON,limit to 5000mA @00:16:56.28 Gate.power :analytic1,:ON,limit to 5000mA @00:16:56.29 Gate.power :analytic2,:ON,limit to 5000mA @00:16:56.30 Gate.power :analytic3,:ON,limit to 5000mA @00:16:56.35 Powering core bus @00:16:58.66 Archive.configure :default @00:16:58.81 Air.configure :default @00:16:58.92 Lysis1.configure :default @00:16:59.09 Lysis2.configure :default @00:16:59.23 Plunge.configure :default @00:16:59.37 SamplePump.configure :default @00:16:59.51 ExtraPump.configure :default @00:16:59.66 Toroid.configure :default @00:17:00.14 Carousel.configure :default @00:17:00.54 Pressure.configure :default @00:17:00.70 All elf controllers running firmware version 4.61 @00:17:00.76 Gate.power :motor,:ON,limit to 5000mA @00:17:02.42 Air.reconfigure :default @00:17:02.49 Archive.reconfigure :default @00:17:02.55 Plunge.reconfigure :default @00:17:02.71 Toroid.reconfigure :default @00:17:02.77 Lysis2.reconfigure :default @00:17:02.93 Lysis1.reconfigure :default @00:17:03.55 Air.home.jog 0 @00:17:03.73 Archive.home.jog 0 @00:17:03.78 Plunge.home.jog 0 @00:17:03.81 Toroid.home.jog 0 @00:17:04.02 Lysis2.home.jog 0 @00:17:04.08 Lysis1.home.jog 0 @00:17:05.87 Plunge.reconfigure :homing @00:17:05.93 Lysis2.reconfigure :homing @00:17:05.97 Air.reconfigure :homing @00:17:06.01 Toroid.reconfigure :homing @00:17:06.06 Archive.reconfigure :homing @00:17:06.11 Lysis1.reconfigure :homing @00:17:06.37 Lysis2.jog 1353 @00:17:06.42 Plunge.jog 856 @00:17:06.45 Air.jog 1370 @00:17:06.48 Toroid.jog 611 @00:17:06.51 Archive.jog 154 @00:17:06.54 Lysis1.jog 1352 @00:17:07.16 Archive.setRawPosition! 50 @00:17:07.24 Archive.seek :home @00:17:07.53 Plunge.setRawPosition! 50 @00:17:07.58 Plunge.seek :home @00:17:07.67 Archive.reconfigure :default @00:17:07.77 Lysis2.setRawPosition! 50 @00:17:07.80 Air.setRawPosition! 50 @00:17:07.89 Lysis2.seek :home @00:17:07.98 Air.seek :home @00:17:08.11 Lysis1.setRawPosition! 50 @00:17:08.13 Plunge.reconfigure :default @00:17:08.23 Lysis1.seek :home @00:17:08.40 Air.reconfigure :default @00:17:08.43 Lysis2.reconfigure :default @00:17:08.60 Lysis1.reconfigure :default @00:17:08.74 Toroid.setRawPosition! 0 @00:17:08.84 Toroid.seek :home @00:17:08.95 Toroid.reconfigure :default @00:17:09.12
Toroid.seek :clear @00:17:10.79 Carousel.reconfigure :default @00:17:10.94 Carousel.home.jog 0 @00:17:11.99 Carousel.seek 35 @00:17:12.65 Selecting Cartridge 34 @00:17:13.21 Carousel.seek 34 @00:17:14.01 Lysis1.seek :locked @00:17:17.32 Plunge.seek :locked @00:17:17.62 Air.seek :locked @00:17:22.13
Mounted archive Cartridge @00:17:22.51 GoodNews.email "Filtering up to 1000ml",Subject: "Sampling archive_bac cartridge in slot 34" @00:17:25.54 Duration of filtering limited to 1:05:00 @00:17:26.72 Priming sample loop w/160ml, bypass w/10ml @00:17:26.87 SamplePump.setPosition! 0ml @00:17:27.02 Bypass.open @00:17:28.53 Intake.open @00:17:29.39 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:17:32.22 Vacuum tare is 18% of full scale @00:17:32.34 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:17:32.94 Pressure.reconfigure :default @00:17:33.14 Pressure.hold @00:17:33.69 SamplePump.reconfigure :prime @00:17:33.83 SamplePump.seek 10ml,1:21 @00:17:45.02 Pressure.coast @00:17:45.11 SamplePump.setPosition! 0ml @00:17:45.35 Exhaust.open @00:17:46.88 Bypass.close @00:17:48.51 Pressure.hold @00:17:48.80 SamplePump.reconfigure :prime @00:17:48.86 SamplePump.seek 160ml,13:51 @00:18:03.03 EOFError in Qmail -- end of file reached @00:18:03.11 Retry #1 of 10 after 10:00 @00:20:40.34 Pressure.coast @00:20:40.41 Bypass.open @00:20:41.15 Intake.close @00:20:42.09 Toroid.reconfigure :hold @00:20:42.15 Toroid.seek :crack @00:20:43.73 Toroid.seek :cartridge @00:20:44.68 Toroid.reconfigure :default @00:20:44.73 Toroid.seek :clear @00:20:47.21 Pressurizing cartridge to 26psi above 14.6psia @00:20:47.23 SamplePump.setPosition! 0ml @00:20:47.28 Intake.open @00:20:48.17 Bypass.close @00:20:48.86 Exhaust.close @00:20:49.95 Pressure.reconfigure :default @00:20:50.03 Pressure.hold @00:20:50.48 SamplePump.reconfigure :charge @00:20:50.79 SamplePump.seek 2.3ml,42 seconds @00:21:00.52 Pressure.coast @00:21:00.57 Intake.close @00:21:01.44 Pressurized cartridge with 0.424ml to 26.1psi @00:21:01.45 Isolated @34.7psia (20.1psi) @00:21:01.46 Checking for cartridge leaks from 34.7psia @00:21:01.49 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:21:06.12 Cartridge pressure fell 1% to 34.2psia @00:21:06.42 Sampling 1000.0ml @00:21:06.45 SamplePump.setPosition! 0ml @00:21:06.53 Exhaust.open @00:21:07.30 Intake.open @00:21:08.53 Pressure.reconfigure :default @00:21:08.59 Pressure.hold @00:21:09.17 SamplePump.reconfigure :sample @00:21:09.25 SamplePump.seek 1000ml,1:23:51 @00:21:11.44 Cartridge 34 sampling 1/1000ml at 0.47ml/s with 25.5psi vacuum @00:22:11.67 Cartridge 34 sampling 28/1000ml at 0.44ml/s with 26.5psi vacuum @00:23:11.45 Cartridge 34 sampling 55/1000ml at 0.43ml/s with 26.6psi vacuum @00:24:11.45 Cartridge 34 sampling 82/1000ml at 0.45ml/s with 25.1psi vacuum @00:25:11.67 Cartridge 34 sampling 108/1000ml at 0.43ml/s with 26.3psi vacuum @00:26:11.57 Cartridge 34 sampling 135/1000ml at 0.43ml/s with 26.8psi vacuum @00:27:11.62 Cartridge 34 sampling 161/1000ml at 0.44ml/s with 25.1psi vacuum pppd: In file /etc/ppp/peers/iridium: unrecognized option 'AT/iridium' @00:28:11.82 Cartridge 34 sampling 187/1000ml at 0.43ml/s with 27.1psi vacuum @00:29:11.45 Cartridge 34 sampling 213/1000ml at 0.42ml/s with 18.0psi vacuum @00:30:11.61 Cartridge 34 sampling 238/1000ml at 0.43ml/s with 26.2psi vacuum @00:31:11.46 Cartridge 34 sampling 264/1000ml at 0.43ml/s with 21.0psi vacuum @00:32:11.66 Cartridge 34 sampling 289/1000ml at 0.42ml/s with 26.3psi vacuum @00:33:11.63 Cartridge 34 sampling 315/1000ml at 0.41ml/s with 27.3psi vacuum @00:34:11.45 Cartridge 34 sampling 340/1000ml at 0.42ml/s with 25.6psi vacuum @00:35:11.68 Cartridge 34 sampling 365/1000ml at 0.42ml/s with 26.1psi vacuum @00:36:11.70 Cartridge 34 sampling 389/1000ml at 0.38ml/s with 22.7psi vacuum @00:37:11.45 Cartridge 34 sampling 414/1000ml at 0.40ml/s with 17.9psi vacuum 2.5m deep (18.4psia) @00:38:11.60 Cartridge 34 sampling 438/1000ml at 0.41ml/s with 25.3psi vacuum @00:39:11.45 Cartridge 34 sampling 462/1000ml at 0.41ml/s with 25.7psi vacuum @00:40:11.67 Cartridge 34 sampling 486/1000ml at 0.41ml/s with 27.1psi vacuum @00:41:11.69 Cartridge 34 sampling 510/1000ml at 0.37ml/s with 26.3psi vacuum @00:42:11.45 Cartridge 34 sampling 533/1000ml at 0.40ml/s with 27.2psi vacuum @00:43:11.58 Cartridge 34 sampling 557/1000ml at 0.37ml/s with 26.8psi vacuum @00:44:11.87 Cartridge 34 sampling 580/1000ml at 0.37ml/s with 22.4psi vacuum @00:45:11.45 Cartridge 34 sampling 603/1000ml at 0.38ml/s with 26.4psi vacuum @00:46:11.65 Cartridge 34 sampling 626/1000ml at 0.36ml/s with 26.0psi vacuum @00:47:11.66 Cartridge 34 sampling 649/1000ml at 0.39ml/s with 26.1psi vacuum @00:48:11.45 Cartridge 34 sampling 672/1000ml at 0.39ml/s with 26.0psi vacuum @00:49:11.87 Cartridge 34 sampling 694/1000ml at 0.36ml/s with 26.9psi vacuum @00:50:11.45 Cartridge 34 sampling 717/1000ml at 0.38ml/s with 27.6psi vacuum @00:51:12.03 Cartridge 34 sampling 739/1000ml at 0.38ml/s with 21.2psi vacuum @00:52:12.04 Cartridge 34 sampling 761/1000ml at 0.37ml/s with 27.1psi vacuum @00:53:11.62 Cartridge 34 sampling 783/1000ml at 0.35ml/s with 15.9psi vacuum 4.6m deep (21.5psia) @00:54:11.45 Cartridge 34 sampling 805/1000ml at 0.37ml/s with 27.6psi vacuum @00:55:11.62 Cartridge 34 sampling 827/1000ml at 0.38ml/s with 21.6psi vacuum @00:56:11.74 Cartridge 34 sampling 849/1000ml at 0.35ml/s with 25.2psi vacuum @00:57:11.44 Cartridge 34 sampling 870/1000ml at 0.37ml/s with 26.9psi vacuum @00:58:11.63 Cartridge 34 sampling 892/1000ml at 0.37ml/s with 27.2psi vacuum @00:59:11.44 Cartridge 34 sampling 913/1000ml at 0.35ml/s with 17.8psi vacuum @01:00:11.51 Cartridge 34 sampling 934/1000ml at 0.33ml/s with 25.7psi vacuum @01:01:11.44 Cartridge 34 sampling 956/1000ml at 0.38ml/s with 26.7psi vacuum @01:02:11.68 Cartridge 34 sampling 977/1000ml at 0.33ml/s with 25.8psi vacuum @01:03:11.69 Cartridge 34 sampling 998/1000ml at 0.33ml/s with 26.3psi vacuum @01:03:17.53 Pressure.coast @01:03:17.84 Intake.close @01:03:17.94 Exhaust.close @01:03:18.42 Toroid.reconfigure :hold @01:03:18.50 Toroid.seek :bypass @01:03:20.23 Toroid.reconfigure :default @01:03:20.29 Toroid.seek :clear @01:03:21.94 Sampled 1000.0ml @01:03:22.41 GoodNews.email "After filtering 1000.0ml",Subject: "Processing archive_bac cartridge in slot 34" @01:03:22.82 SamplePump.setPosition! 0ml @01:03:22.87 Plunge.seek :reagent @01:03:22.91 Exhaust.open @01:03:23.61 Intake.open @01:03:23.82 Sample loop closed. Ready to Deliver Reagents @01:03:23.84 Archive.seek :sealed @01:03:24.19 EOFError in Qmail -- end of file reached @01:03:24.25 Retry #1 of 10 after 10:00 @01:03:24.48 SamplePump.reconfigure :slow @01:03:24.54 SamplePump.seek -22ml @01:03:27.59 Archive.seek :evac @01:03:41.64 delay 1:00 @01:04:11.83 delay 1 second @01:04:12.86 Bypass.open @01:04:13.57 Exhaust.close @01:04:14.31 SamplePump.seek -20ml @01:04:18.86 delay 1 second @01:04:20.03 SamplePump.seek -28ml @01:04:37.43 delay 1 second @01:04:38.67 Intake.close @01:04:38.92 Bypass.close @01:04:41.67 Archive.seek :clear @01:04:58.82 RNAlater soak for 10:00 @01:13:24.44 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:13:24.46 Retry #2 of 10 after 10:00 @01:14:58.89 Archive.seek :tVent @01:15:00.64 delay 1 second @01:15:01.72 Archive.seek :evac @01:15:17.80 delay 10 seconds @01:15:27.85 Sample preserved and evacuated @01:15:28.40
Carousel.seek 34 @01:15:28.84 Plunge.seek :locked @01:15:29.96 Archive.seek :clear @01:15:30.28 Air.seek :clear @01:15:30.60 Lysis1.seek :clear @01:15:47.28
Archive.coast @01:15:47.37 Air.coast @01:15:47.53 Lysis1.coast @01:15:47.57 Lysis2.coast @01:15:47.72 Plunge.coast @01:15:47.82 SamplePump.coast @01:15:47.94 Toroid.coast @01:15:48.08 Carousel.coast @01:15:48.44 Gate.power :motor,:OFF @01:15:48.46 Gate.power :core,:OFF @01:15:48.52 Gate.power :analytic1,:OFF @01:15:48.57 Gate.power :analytic2,:OFF @01:15:48.59 Gate.power :analytic3,:OFF @01:15:48.61 Gate.power :main,:OFF @01:15:48.66 Safely stopped and ready to power off @01:15:49.12 Cartridge Load: {34 => { Type: :archive_bac, State: :processed, filtered: 1000.0}, 35 => { Type: :archive_bac, State: :processed, filtered: 647.2}, 39..37 => { Type: :archive_bac, State: :processed, filtered: 1000}, [33, 32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 250}, [44, 42, 40] => { Type: :archive_bac, State: :processed, filtered: 250}, 36 => { Type: :archive_bac, State: :processed, filtered: 382.5}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 297}} Spares: {spare: 4} @01:16:19.23 Sending Email @01:21:19.27 Email::Error in MAIN -- flush stuck! @01:21:49.34 Uploading Logs @01:23:33.65 BadNews.email "Not enough dry cartridges remain",Subject: "Cartridge::None in MAIN" @01:31:49.38 Log::Error in MAIN -- upload stuck! @10:39:33.80PST02-Mar-21
>BREAK< ESP3 Gateway v1.11 for core bus @10:39:33.83 >BREAK< ESP3 Gateway v1.11 for cartridge bus @10:39:37.99 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @10:39:38.29 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @10:39:44.11 Polling Can every 2:00 @10:39:44.27 Waiting for espclients... @10:39:44.63 Session Begun @10:39:48.85 -> slots @10:41:18.67 -> Cartridge.release 33 @10:43:44.78 Session Ended Log::Locked in trapHandler -- Another process is already writing to /var/log/esp/real.log @10:56:40.59 Session Begun @10:56:46.21 -> MainThread.exit @10:56:46.49 Session Ended @10:57:38.37PST02-Mar-21
>BREAK< ESP3 Gateway v1.11 for core bus @10:57:38.55 >BREAK< ESP3 Gateway v1.11 for cartridge bus @10:57:41.59 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @10:57:41.90 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @10:57:49.02 Polling Can every 2:00 @10:57:49.07 Loading /home/esp/esp2/bin/processAll mission @10:57:50.19 Sampling maximum of 2000.0ml per cartridge @10:57:50.24 GoodNews.email "Cartridge group: dry\nSpares: {spare: 4}\n\n",Subject: "Begun processAll --mail --checkpoint 2000" @10:57:50.43
Gate.power :main,:ON,limit to 5000mA @10:57:50.44 Gate.power :analytic1,:ON,limit to 5000mA @10:57:50.46 Gate.power :analytic2,:ON,limit to 5000mA @10:57:50.47 Gate.power :analytic3,:ON,limit to 5000mA @10:57:50.51 Powering core bus @10:57:52.81 Archive.configure :default @10:57:52.96 Air.configure :default @10:57:53.07 Lysis1.configure :default @10:57:53.21 Lysis2.configure :default @10:57:54.04 Plunge.configure :default @10:57:54.60 SamplePump.configure :default @10:57:54.99 ExtraPump.configure :default @10:57:55.33 Toroid.configure :default @10:57:55.65 Carousel.configure :default @10:57:56.07 Pressure.configure :default @10:57:56.25 All elf controllers running firmware version 4.61 @10:57:56.28 Gate.power :motor,:ON,limit to 5000mA @10:57:57.52 Toroid.reconfigure :default @10:57:57.65 Lysis2.reconfigure :default @10:57:57.76 Air.reconfigure :default @10:57:57.81 Archive.reconfigure :default @10:57:57.95 Plunge.reconfigure :default @10:57:58.07 Lysis1.reconfigure :default @10:57:58.45 Toroid.home.jog 0 @10:57:58.61 Lysis2.home.jog 0 @10:57:58.67 Archive.home.jog 0 @10:57:58.75 Lysis1.home.jog 0 @10:57:58.80 Air.home.jog 0 @10:57:58.84 Plunge.home.jog 0 @10:58:00.18 Toroid.reconfigure :homing @10:58:00.27 Plunge.reconfigure :homing @10:58:00.30 Lysis1.reconfigure :homing @10:58:00.33 Archive.reconfigure :homing @10:58:00.37 Lysis2.reconfigure :homing @10:58:00.51 Air.reconfigure :homing @10:58:00.67 Toroid.jog 601 @10:58:00.82 Plunge.jog 860 @10:58:00.89 Lysis1.jog 1350 @10:58:00.92 Archive.jog 159 @10:58:00.95 Lysis2.jog 1362 @10:58:00.99 Air.jog 1373 @10:58:01.52 Archive.setRawPosition! 50 @10:58:01.57 Archive.seek :home @10:58:01.86 Archive.reconfigure :default @10:58:01.97 Plunge.setRawPosition! 50 @10:58:02.01 Plunge.seek :home @10:58:02.14 Lysis1.setRawPosition! 50 @10:58:02.19 Lysis1.seek :home @10:58:02.30 Plunge.reconfigure :default @10:58:02.35 Lysis2.setRawPosition! 50 @10:58:02.47 Lysis2.seek :home @10:58:02.57 Air.setRawPosition! 50 @10:58:02.60 Lysis1.reconfigure :default @10:58:02.67 Air.seek :home @10:58:03.06 Toroid.setRawPosition! 0 @10:58:03.10 Lysis2.reconfigure :default @10:58:03.29 Toroid.seek :home @10:58:03.58 Toroid.reconfigure :default @10:58:03.63 Air.reconfigure :default @10:58:04.41
Toroid.seek :clear @10:58:06.27 Carousel.reconfigure :default @10:58:06.56 Carousel.home.jog 0 @10:58:07.48 Carousel.seek 34 @10:58:08.09 Selecting Cartridge 33 @10:58:08.88 Carousel.seek 33 @10:58:09.59 Lysis1.seek :locked @10:58:12.83 Plunge.seek :locked @10:58:12.98 Air.seek :locked @10:58:17.37
Mounted archive Cartridge @10:58:18.80 GoodNews.email "Filtering up to 2000.0ml",Subject: "Sampling archive_bac cartridge in slot 33" @10:58:19.80 Duration of filtering limited to 2:05:00 @10:58:20.82 Priming sample loop w/160ml, bypass w/10ml @10:58:21.14 SamplePump.setPosition! 0ml @10:58:21.26 Bypass.open @10:58:22.81 Intake.open @10:58:23.72 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @10:58:26.48 Vacuum tare is 18% of full scale @10:58:26.58 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @10:58:27.14 Pressure.reconfigure :default @10:58:27.31 Pressure.hold @10:58:27.79 SamplePump.reconfigure :prime @10:58:27.99 SamplePump.seek 10ml,1:21 @10:58:29.53 upsync @10:58:39.24 Pressure.coast @10:58:39.29 SamplePump.setPosition! 0ml @10:58:39.34 Exhaust.open @10:58:40.78 Bypass.close @10:58:42.31 Pressure.reconfigure :default @10:58:42.37 Pressure.hold @10:58:42.65 SamplePump.reconfigure :prime @10:58:42.71 SamplePump.seek 160ml,13:51 @10:58:48.15 EOFError in Qmail -- end of file reached @10:58:48.18 Retry #1 of 10 after 10:00 @11:01:34.29 Pressure.coast @11:01:34.35 Bypass.open @11:01:35.02 Intake.close @11:01:35.90 Toroid.reconfigure :hold @11:01:35.96 Toroid.seek :crack @11:01:37.54 Toroid.seek :cartridge @11:01:38.48 Toroid.reconfigure :default @11:01:38.54 Toroid.seek :clear @11:01:40.80 Pressurizing cartridge to 26psi above 14.5psia @11:01:40.82 SamplePump.setPosition! 0ml @11:01:40.87 Intake.open @11:01:41.76 Bypass.close @11:01:42.44 Exhaust.close @11:01:43.70 Pressure.reconfigure :default @11:01:43.84 Pressure.hold @11:01:44.07 SamplePump.reconfigure :charge @11:01:44.13 SamplePump.seek 2.3ml,42 seconds @11:01:52.86 Pressure.coast @11:01:52.91 Intake.close @11:01:53.77 Pressurized cartridge with 0.337ml to 26.0psi @11:01:53.78 Isolated @35.3psia (20.8psi) @11:01:53.79 Checking for cartridge leaks from 35.3psia @11:01:53.81 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @11:01:58.49 Cartridge pressure fell 3% to 34.4psia @11:01:58.67 Sampling 2000.0ml @11:01:58.75 SamplePump.setPosition! 0ml @11:01:58.83 Exhaust.open @11:01:59.60 Intake.open @11:02:00.73 Pressure.reconfigure :default @11:02:00.88 Pressure.hold @11:02:01.49 SamplePump.reconfigure :sample @11:02:01.55 SamplePump.seek 2000ml,2:47:12 @11:02:03.73 Cartridge 33 sampling 1/2000ml at 0.51ml/s with 20.3psi vacuum @11:03:03.87 Cartridge 33 sampling 30/2000ml at 0.49ml/s with 26.3psi vacuum @11:04:03.73 Cartridge 33 sampling 59/2000ml at 0.46ml/s with 23.1psi vacuum @11:05:03.89 Cartridge 33 sampling 88/2000ml at 0.48ml/s with 26.0psi vacuum @11:06:03.73 Cartridge 33 sampling 117/2000ml at 0.48ml/s with 27.2psi vacuum @11:07:03.91 Cartridge 33 sampling 145/2000ml at 0.47ml/s with 26.6psi vacuum @11:08:03.73 Cartridge 33 sampling 173/2000ml at 0.48ml/s with 21.7psi vacuum @11:08:48.22 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @11:08:48.24 Retry #2 of 10 after 10:00 @11:09:03.83 Cartridge 33 sampling 202/2000ml at 0.45ml/s with 27.7psi vacuum @11:10:03.73 Cartridge 33 sampling 230/2000ml at 0.46ml/s with 25.9psi vacuum @11:11:03.89 Cartridge 33 sampling 257/2000ml at 0.48ml/s with 22.5psi vacuum @11:12:03.73 Cartridge 33 sampling 285/2000ml at 0.47ml/s with 25.0psi vacuum @11:13:03.86 Cartridge 33 sampling 312/2000ml at 0.46ml/s with 25.6psi vacuum @11:14:03.73 Cartridge 33 sampling 339/2000ml at 0.46ml/s with 20.6psi vacuum @11:15:04.00 Cartridge 33 sampling 366/2000ml at 0.44ml/s with 26.9psi vacuum @11:16:03.73 Cartridge 33 sampling 393/2000ml at 0.43ml/s with 26.9psi vacuum @11:17:03.89 Cartridge 33 sampling 420/2000ml at 0.44ml/s with 26.4psi vacuum @11:18:03.73 Cartridge 33 sampling 447/2000ml at 0.45ml/s with 26.9psi vacuum @11:19:03.95 Cartridge 33 sampling 473/2000ml at 0.46ml/s with 26.6psi vacuum @11:19:49.69 upsync @11:20:03.73 Cartridge 33 sampling 500/2000ml at 0.44ml/s with 26.8psi vacuum @11:21:03.89 Cartridge 33 sampling 526/2000ml at 0.46ml/s with 21.6psi vacuum @11:22:03.73 Cartridge 33 sampling 553/2000ml at 0.45ml/s with 21.7psi vacuum @11:23:03.90 Cartridge 33 sampling 579/2000ml at 0.45ml/s with 25.2psi vacuum @11:24:03.73 Cartridge 33 sampling 605/2000ml at 0.43ml/s with 26.7psi vacuum @11:25:03.94 Cartridge 33 sampling 631/2000ml at 0.42ml/s with 26.6psi vacuum @11:26:03.73 Cartridge 33 sampling 657/2000ml at 0.42ml/s with 27.1psi vacuum @11:27:03.92 Cartridge 33 sampling 682/2000ml at 0.44ml/s with 25.5psi vacuum @11:28:03.73 Cartridge 33 sampling 708/2000ml at 0.44ml/s with 20.2psi vacuum 2.3m deep (18.1psia) @11:29:03.82 Cartridge 33 sampling 733/2000ml at 0.43ml/s with 27.9psi vacuum @11:30:04.31 Cartridge 33 sampling 759/2000ml at 0.40ml/s with 21.1psi vacuum 1.6m deep (17.1psia) @11:31:03.82 Cartridge 33 sampling 784/2000ml at 0.41ml/s with 27.4psi vacuum @11:32:03.73 Cartridge 33 sampling 810/2000ml at 0.43ml/s with 19.6psi vacuum @11:33:03.81 Cartridge 33 sampling 835/2000ml at 0.41ml/s with 27.1psi vacuum @11:34:03.93 Cartridge 33 sampling 860/2000ml at 0.44ml/s with 25.5psi vacuum @11:35:03.78 Cartridge 33 sampling 885/2000ml at 0.41ml/s with 18.5psi vacuum 3.5m deep (19.8psia) @11:36:03.73 Cartridge 33 sampling 910/2000ml at 0.40ml/s with 27.8psi vacuum @11:37:03.94 Cartridge 33 sampling 934/2000ml at 0.39ml/s with 17.4psi vacuum @11:38:03.73 Cartridge 33 sampling 959/2000ml at 0.38ml/s with 26.1psi vacuum @11:39:03.79 Cartridge 33 sampling 983/2000ml at 0.38ml/s with 18.6psi vacuum @11:40:03.73 Cartridge 33 sampling 1007/2000ml at 0.40ml/s with 17.2psi vacuum 2.0m deep (17.7psia) @11:41:03.93 Cartridge 33 sampling 1032/2000ml at 0.37ml/s with 25.7psi vacuum @11:42:03.89 Cartridge 33 sampling 1056/2000ml at 0.40ml/s with 16.8psi vacuum 4.4m deep (21.2psia) @11:43:03.73 Cartridge 33 sampling 1080/2000ml at 0.43ml/s with 24.6psi vacuum @11:44:03.73 Cartridge 33 sampling 1105/2000ml at 0.40ml/s with 28.9psi vacuum @11:45:03.89 Cartridge 33 sampling 1129/2000ml at 0.39ml/s with 28.3psi vacuum @11:46:03.87 Cartridge 33 sampling 1153/2000ml at 0.39ml/s with 18.7psi vacuum 3.9m deep (20.5psia) @11:47:03.73 Cartridge 33 sampling 1177/2000ml at 0.38ml/s with 20.1psi vacuum @11:48:03.99 Cartridge 33 sampling 1201/2000ml at 0.37ml/s with 26.9psi vacuum @11:49:04.03 Cartridge 33 sampling 1225/2000ml at 0.42ml/s with 24.6psi vacuum @11:50:03.73 Cartridge 33 sampling 1249/2000ml at 0.38ml/s with 8.1psi vacuum 5.2m deep (22.4psia) @11:51:04.11 Cartridge 33 sampling 1273/2000ml at 0.41ml/s with 28.3psi vacuum @11:52:03.73 Cartridge 33 sampling 1296/2000ml at 0.39ml/s with 29.0psi vacuum @11:53:03.96 Cartridge 33 sampling 1320/2000ml at 0.38ml/s with 19.1psi vacuum @11:54:03.73 Cartridge 33 sampling 1343/2000ml at 0.41ml/s with 22.7psi vacuum @11:55:03.90 Cartridge 33 sampling 1366/2000ml at 0.38ml/s with 7.3psi vacuum 10.3m deep (29.9psia) @11:56:03.73 Cartridge 33 sampling 1390/2000ml at 0.41ml/s with 21.2psi vacuum @11:57:04.02 Cartridge 33 sampling 1413/2000ml at 0.38ml/s with 23.8psi vacuum @11:58:03.73 Cartridge 33 sampling 1436/2000ml at 0.41ml/s with 21.0psi vacuum @11:59:03.84 Cartridge 33 sampling 1460/2000ml at 0.37ml/s with 25.8psi vacuum @12:00:03.73 Cartridge 33 sampling 1483/2000ml at 0.38ml/s with 28.8psi vacuum @12:01:04.03 Cartridge 33 sampling 1506/2000ml at 0.41ml/s with 24.3psi vacuum @12:02:03.73 Cartridge 33 sampling 1529/2000ml at 0.37ml/s with 22.7psi vacuum @12:03:03.96 Cartridge 33 sampling 1552/2000ml at 0.41ml/s with 22.2psi vacuum @12:04:03.85 Cartridge 33 sampling 1575/2000ml at 0.40ml/s with 18.1psi vacuum 4.2m deep (20.9psia) @12:05:03.87 Cartridge 33 sampling 1598/2000ml at 0.39ml/s with 21.2psi vacuum 1.7m deep (17.3psia) @12:06:03.73 Cartridge 33 sampling 1620/2000ml at 0.35ml/s with 27.8psi vacuum @12:07:03.85 Cartridge 33 sampling 1643/2000ml at 0.38ml/s with 29.8psi vacuum @12:08:03.73 Cartridge 33 sampling 1665/2000ml at 0.35ml/s with 27.9psi vacuum @12:09:04.15 Cartridge 33 sampling 1688/2000ml at 0.41ml/s with 22.3psi vacuum @12:10:03.73 Cartridge 33 sampling 1710/2000ml at 0.37ml/s with 27.2psi vacuum @12:11:03.98 Cartridge 33 sampling 1733/2000ml at 0.37ml/s with 22.9psi vacuum @12:12:03.73 Cartridge 33 sampling 1755/2000ml at 0.39ml/s with 19.2psi vacuum @12:13:03.91 Cartridge 33 sampling 1777/2000ml at 0.35ml/s with 23.8psi vacuum @12:14:03.73 Cartridge 33 sampling 1799/2000ml at 0.31ml/s with 23.6psi vacuum @12:15:03.90 Cartridge 33 sampling 1820/2000ml at 0.32ml/s with 26.8psi vacuum @12:16:03.73 Cartridge 33 sampling 1842/2000ml at 0.36ml/s with 26.2psi vacuum @12:17:03.86 Cartridge 33 sampling 1863/2000ml at 0.38ml/s with 21.2psi vacuum 2.5m deep (18.5psia) @12:18:03.73 Cartridge 33 sampling 1884/2000ml at 0.35ml/s with 17.1psi vacuum @12:19:03.79 Cartridge 33 sampling 1905/2000ml at 0.34ml/s with 7.8psi vacuum 6.0m deep (23.6psia) @12:20:03.73 Cartridge 33 sampling 1926/2000ml at 0.37ml/s with 29.5psi vacuum @12:21:04.04 Cartridge 33 sampling 1947/2000ml at 0.30ml/s with 22.0psi vacuum @12:22:04.20 Cartridge 33 sampling 1968/2000ml at 0.29ml/s with 22.2psi vacuum @12:23:03.73 Cartridge 33 sampling 1989/2000ml at 0.32ml/s with 29.4psi vacuum @12:23:36.67 Pressure.coast @12:23:37.10 Intake.close @12:23:37.31 Exhaust.close @12:23:37.66 Toroid.reconfigure :hold @12:23:37.87 Toroid.seek :bypass @12:23:39.52 Toroid.reconfigure :default @12:23:39.58 Toroid.seek :clear @12:23:41.20 Sampled 2000.0ml @12:23:41.66 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 33" @12:23:42.10 SamplePump.setPosition! 0ml @12:23:42.14 Plunge.seek :reagent @12:23:42.21 Exhaust.open @12:23:42.94 Intake.open @12:23:43.10 Sample loop closed. Ready to Deliver Reagents @12:23:43.13 Archive.seek :sealed @12:23:43.82 SamplePump.reconfigure :slow @12:23:43.88 SamplePump.seek -22ml @12:23:46.86 Archive.seek :evac @12:24:00.88 delay 1:00 @12:24:31.18 delay 1 second @12:24:32.21 Bypass.open @12:24:32.96 Exhaust.close @12:24:33.91 SamplePump.seek -20ml @12:24:38.48 delay 1 second @12:24:39.58 SamplePump.seek -28ml @12:24:56.94 delay 1 second @12:24:58.03 Intake.close @12:24:58.09 Bypass.close @12:25:00.93 Archive.seek :clear @12:25:18.11 RNAlater soak for 10:00 @12:35:18.18 Archive.seek :tVent @12:35:19.88 delay 1 second @12:35:20.96 Archive.seek :evac @12:35:36.96 delay 10 seconds @12:35:47.03 Sample preserved and evacuated @12:35:47.69
Carousel.seek 33 @12:35:48.13 Plunge.seek :locked @12:35:50.74 Archive.seek :clear @12:35:50.95 Air.seek :clear @12:35:51.13 Lysis1.seek :clear @12:36:07.96
Archive.coast @12:36:08.06 Air.coast @12:36:08.10 Lysis1.coast @12:36:08.22 Lysis2.coast @12:36:08.35 Plunge.coast @12:36:08.39 SamplePump.coast @12:36:08.44 Toroid.coast @12:36:08.58 Carousel.coast @12:36:09.07 Gate.power :motor,:OFF @12:36:09.19 Gate.power :core,:OFF @12:36:09.21 Gate.power :analytic1,:OFF @12:36:09.23 Gate.power :analytic2,:OFF @12:36:09.25 Gate.power :analytic3,:OFF @12:36:09.26 Gate.power :main,:OFF @12:36:09.27 Safely stopped and ready to power off @12:36:09.64 Cartridge Load: {33 => { Type: :archive_bac, State: :processed, filtered: 2000.0}, 35 => { Type: :archive_bac, State: :processed, filtered: 647.2}, [39..37, 34] => { Type: :archive_bac, State: :processed, filtered: 1000}, [32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 250}, [44, 42, 40] => { Type: :archive_bac, State: :processed, filtered: 250}, 36 => { Type: :archive_bac, State: :processed, filtered: 382.5}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 297}} Spares: {spare: 4} @12:36:09.67 BadNews.email "Not enough dry cartridges remain",Subject: "Cartridge::None in MAIN" @12:36:51.40
Uploading Logs @12:46:51.44 Log::Error in MAIN -- upload stuck! @18:01:40.47PST06-Mar-21
>BREAK< ESP3 Gateway v1.11 for core bus @18:01:40.50 >BREAK< ESP3 Gateway v1.11 for cartridge bus @18:01:47.59 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @18:01:47.91 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @18:01:52.21 Polling Can every 2:00 @18:01:52.35 Loading /home/esp/esp2/bin/processAll mission @18:01:53.69 Cartridge Load: {33 => { Type: :archive_bac, State: :processed, filtered: 2000}, 35 => { Type: :archive_bac, State: :processed, filtered: 647.2}, [39..37, 34] => { Type: :archive_bac, State: :processed, filtered: 1000}, [32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 250}, [44, 42, 40] => { Type: :archive_bac, State: :processed, filtered: 250}, 36 => { Type: :archive_bac, State: :processed, filtered: 382.5}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 297}} Spares: {spare: 4} @18:01:53.71 ArgumentError in MAIN -- '—interval=4::' is neither a valid volume nor a cartridge type or state @18:05:30.12PST06-Mar-21
>BREAK< ESP3 Gateway v1.11 for core bus @18:05:30.30 >BREAK< ESP3 Gateway v1.11 for cartridge bus @18:05:34.51 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @18:05:34.82 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @18:05:40.69 Polling Can every 2:00 @18:05:40.80 Loading ././processAll mission @18:05:43.29 Sampling maximum of 2000.0ml per cartridge @18:05:43.35 GoodNews.email "Cartridge group: dry\nSpares: {spare: 4}\nSchedule: [-4:00:00]\n",Subject: "Begun processAll --interval=4:: --mail --checkpoint 2000" @18:05:43.57
Cartridge Load: {33 => { Type: :archive_bac, State: :processed, filtered: 2000}, 35 => { Type: :archive_bac, State: :processed, filtered: 647.2}, [39..37, 34] => { Type: :archive_bac, State: :processed, filtered: 1000}, [32, 16..2] => { Type: :archive_bac, State: :reserved, was: :dry}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 250}, [44, 42, 40] => { Type: :archive_bac, State: :processed, filtered: 250}, 36 => { Type: :archive_bac, State: :processed, filtered: 382.5}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 297}} Spares: {spare: 4} @18:05:52.42 BadNews.email "Not enough dry cartridges remain",Subject: "Cartridge::None in MAIN" @18:06:13.42 upsync @18:11:37.83PST06-Mar-21
>BREAK< ESP3 Gateway v1.11 for core bus @18:11:37.90 >BREAK< ESP3 Gateway v1.11 for cartridge bus @18:11:42.16 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @18:11:42.47 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @18:11:49.69 Polling Can every 2:00 @18:11:49.74 Loading ././processAll mission @18:11:50.99 Sampling maximum of 2000.0ml per cartridge @18:11:51.04 GoodNews.email "Cartridge group: dry\nSpares: {spare: 4}\nSchedule: [-4:00:00]\n",Subject: "Begun processAll --interval=4:: --mail --checkpoint 2000" @18:11:51.22
Gate.power :main,:ON,limit to 5000mA @18:11:51.23 Gate.power :analytic1,:ON,limit to 5000mA @18:11:51.24 Gate.power :analytic2,:ON,limit to 5000mA @18:11:51.26 Gate.power :analytic3,:ON,limit to 5000mA @18:11:51.31 Powering core bus @18:11:53.61 Archive.configure :default @18:11:53.79 Air.configure :default @18:11:53.91 Lysis1.configure :default @18:11:54.36 Lysis2.configure :default @18:11:55.09 Plunge.configure :default @18:11:55.62 SamplePump.configure :default @18:11:56.03 ExtraPump.configure :default @18:11:56.41 Toroid.configure :default @18:11:56.79 Carousel.configure :default @18:11:57.13 Pressure.configure :default @18:11:57.37 All elf controllers running firmware version 4.61 @18:11:57.40 Gate.power :motor,:ON,limit to 5000mA @18:11:58.63 Plunge.reconfigure :default @18:11:58.81 Lysis1.reconfigure :default @18:11:58.84 Air.reconfigure :default @18:11:58.96 Toroid.reconfigure :default @18:11:59.13 Lysis2.reconfigure :default @18:11:59.18 Archive.reconfigure :default @18:11:59.98 Lysis1.home.jog 0 @18:12:00.10 Plunge.home.jog 0 @18:12:00.12 Air.home.jog 0 @18:12:00.14 Lysis2.home.jog 0 @18:12:00.17 Toroid.home.jog 0 @18:12:00.20 Archive.home.jog 0 @18:12:00.96 Plunge.reconfigure :homing @18:12:01.03 Lysis2.reconfigure :homing @18:12:01.08 Air.reconfigure :homing @18:12:01.12 Archive.reconfigure :homing @18:12:01.16 Toroid.reconfigure :homing @18:12:01.22 Lysis1.reconfigure :homing @18:12:01.47 Lysis2.jog 1365 @18:12:01.51 Plunge.jog 858 @18:12:01.54 Archive.jog 152 @18:12:01.58 Toroid.jog 617 @18:12:01.61 Air.jog 1369 @18:12:01.65 Lysis1.jog 1354 @18:12:02.33 Archive.setRawPosition! 50 @18:12:02.44 Archive.seek :home @18:12:02.73 Plunge.setRawPosition! 50 @18:12:02.78 Plunge.seek :home @18:12:02.88 Lysis2.setRawPosition! 50 @18:12:02.91 Archive.reconfigure :default @18:12:02.99 Lysis2.seek :home @18:12:03.12 Plunge.reconfigure :default @18:12:03.25 Lysis1.setRawPosition! 50 @18:12:03.27 Air.setRawPosition! 50 @18:12:03.47 Lysis1.seek :home @18:12:03.52 Air.seek :home @18:12:03.89 Lysis2.reconfigure :default @18:12:04.22 Toroid.setRawPosition! 0 @18:12:04.33 Lysis1.reconfigure :default @18:12:04.39 Air.reconfigure :default @18:12:04.86 Toroid.seek :home @18:12:05.26 Toroid.reconfigure :default @18:12:05.85
Toroid.seek :clear @18:12:07.63 Carousel.reconfigure :default @18:12:07.69 Carousel.home.jog 0 @18:12:08.42 Carousel.seek 33 @18:12:08.90 Selecting Cartridge 32 @18:12:09.65 Carousel.seek 32 @18:12:10.31 Lysis1.seek :locked @18:12:13.69 Plunge.seek :locked @18:12:13.84 Air.seek :locked @18:12:19.45
Mounted archive Cartridge @18:12:19.81 GoodNews.email "Filtering up to 2000.0ml",Subject: "Sampling archive_bac cartridge in slot 32" @18:12:20.94 Duration of filtering limited to 2:05:00 @18:12:21.24 Priming sample loop w/160ml, bypass w/10ml @18:12:21.26 SamplePump.setPosition! 0ml @18:12:21.30 Bypass.open @18:12:22.78 Intake.open @18:12:23.70 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @18:12:26.18 upsync @18:12:26.56 Vacuum tare is 17% of full scale @18:12:26.63 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @18:12:27.15 Pressure.reconfigure :default @18:12:27.22 Pressure.hold @18:12:27.79 SamplePump.reconfigure :prime @18:12:27.93 SamplePump.seek 10ml,1:21 @18:12:36.32 EOFError in Qmail -- end of file reached @18:12:36.35 Retry #1 of 10 after 10:00 @18:12:39.14 Pressure.coast @18:12:39.21 SamplePump.setPosition! 0ml @18:12:39.26 Exhaust.open @18:12:40.65 Bypass.close @18:12:42.21 Pressure.reconfigure :default @18:12:42.27 Pressure.hold @18:12:42.55 SamplePump.reconfigure :prime @18:12:42.61 SamplePump.seek 160ml,13:51 @18:15:34.15 Pressure.coast @18:15:34.21 Bypass.open @18:15:34.91 Intake.close @18:15:35.82 Toroid.reconfigure :hold @18:15:35.88 Toroid.seek :crack @18:15:37.46 Toroid.seek :cartridge @18:15:38.39 Toroid.reconfigure :default @18:15:38.45 Toroid.seek :clear @18:15:40.53 Pressurizing cartridge to 26psi above 14.4psia @18:15:40.57 SamplePump.setPosition! 0ml @18:15:40.68 Intake.open @18:15:41.58 Bypass.close @18:15:42.31 Exhaust.close @18:15:43.61 Pressure.reconfigure :default @18:15:43.67 Pressure.hold @18:15:43.87 SamplePump.reconfigure :charge @18:15:43.94 SamplePump.seek 2.3ml,42 seconds @18:15:53.57 Pressure.coast @18:15:53.62 Intake.close @18:15:54.47 Pressurized cartridge with 0.423ml to 26.3psi @18:15:54.48 Isolated @35.5psia (21.1psi) @18:15:54.49 Checking for cartridge leaks from 35.5psia @18:15:54.51 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @18:15:59.13 Cartridge pressure fell 3% to 34.4psia @18:15:59.40 Sampling 2000.0ml @18:15:59.47 SamplePump.setPosition! 0ml @18:15:59.61 Exhaust.open @18:16:00.37 Intake.open @18:16:01.59 Pressure.reconfigure :default @18:16:01.65 Pressure.hold @18:16:02.48 SamplePump.reconfigure :sample @18:16:02.54 SamplePump.seek 2000ml,2:47:12 @18:16:04.49 Cartridge 32 sampling 1/2000ml at 0.49ml/s with 23.5psi vacuum @18:17:04.65 Cartridge 32 sampling 28/2000ml at 0.45ml/s with 22.9psi vacuum @18:18:04.48 Cartridge 32 sampling 55/2000ml at 0.44ml/s with 25.4psi vacuum @18:19:04.85 Cartridge 32 sampling 82/2000ml at 0.45ml/s with 24.3psi vacuum @18:20:04.49 Cartridge 32 sampling 109/2000ml at 0.45ml/s with 25.1psi vacuum @18:21:04.68 Cartridge 32 sampling 135/2000ml at 0.44ml/s with 22.5psi vacuum @18:22:04.49 Cartridge 32 sampling 160/2000ml at 0.42ml/s with 25.1psi vacuum @18:23:04.72 Cartridge 32 sampling 186/2000ml at 0.42ml/s with 25.5psi vacuum @18:23:07.12 upsync @18:24:04.49 Cartridge 32 sampling 211/2000ml at 0.43ml/s with 22.9psi vacuum @18:25:04.70 Cartridge 32 sampling 236/2000ml at 0.42ml/s with 25.1psi vacuum @18:26:04.49 Cartridge 32 sampling 260/2000ml at 0.41ml/s with 25.9psi vacuum @18:27:04.66 Cartridge 32 sampling 285/2000ml at 0.41ml/s with 23.3psi vacuum @18:28:04.49 Cartridge 32 sampling 309/2000ml at 0.39ml/s with 21.9psi vacuum @18:29:04.73 Cartridge 32 sampling 332/2000ml at 0.39ml/s with 24.9psi vacuum @18:30:04.48 Cartridge 32 sampling 356/2000ml at 0.40ml/s with 20.6psi vacuum @18:31:04.61 Cartridge 32 sampling 379/2000ml at 0.40ml/s with 25.3psi vacuum @18:32:04.49 Cartridge 32 sampling 402/2000ml at 0.38ml/s with 24.8psi vacuum @18:33:04.77 Cartridge 32 sampling 425/2000ml at 0.38ml/s with 25.8psi vacuum @18:34:04.49 Cartridge 32 sampling 448/2000ml at 0.37ml/s with 21.7psi vacuum @18:35:04.62 Cartridge 32 sampling 470/2000ml at 0.37ml/s with 24.9psi vacuum @18:36:04.50 Cartridge 32 sampling 492/2000ml at 0.37ml/s with 23.1psi vacuum @18:37:04.68 Cartridge 32 sampling 514/2000ml at 0.37ml/s with 26.3psi vacuum @18:38:04.49 Cartridge 32 sampling 536/2000ml at 0.36ml/s with 25.1psi vacuum @18:39:04.66 Cartridge 32 sampling 558/2000ml at 0.36ml/s with 24.3psi vacuum @18:40:04.50 Cartridge 32 sampling 579/2000ml at 0.36ml/s with 24.0psi vacuum @18:41:04.70 Cartridge 32 sampling 600/2000ml at 0.35ml/s with 26.6psi vacuum @18:42:04.50 Cartridge 32 sampling 621/2000ml at 0.35ml/s with 21.8psi vacuum @18:43:04.92 Cartridge 32 sampling 642/2000ml at 0.32ml/s with 23.1psi vacuum @18:44:05.07 Cartridge 32 sampling 663/2000ml at 0.35ml/s with 26.6psi vacuum @18:45:04.75 Cartridge 32 sampling 683/2000ml at 0.35ml/s with 25.2psi vacuum @18:46:04.50 Cartridge 32 sampling 704/2000ml at 0.34ml/s with 24.4psi vacuum @18:47:04.58 Cartridge 32 sampling 724/2000ml at 0.34ml/s with 19.5psi vacuum @18:48:04.50 Cartridge 32 sampling 744/2000ml at 0.34ml/s with 21.9psi vacuum @18:49:04.70 Cartridge 32 sampling 764/2000ml at 0.32ml/s with 20.6psi vacuum @18:50:04.62 Cartridge 32 sampling 784/2000ml at 0.35ml/s with 24.4psi vacuum @18:51:04.62 Cartridge 32 sampling 804/2000ml at 0.33ml/s with 22.1psi vacuum @18:52:04.50 Cartridge 32 sampling 824/2000ml at 0.32ml/s with 20.8psi vacuum @18:53:04.66 Cartridge 32 sampling 844/2000ml at 0.35ml/s with 24.8psi vacuum @18:54:04.50 Cartridge 32 sampling 863/2000ml at 0.33ml/s with 23.1psi vacuum @18:55:04.67 Cartridge 32 sampling 883/2000ml at 0.30ml/s with 25.7psi vacuum @18:56:04.50 Cartridge 32 sampling 902/2000ml at 0.31ml/s with 25.8psi vacuum @18:57:04.66 Cartridge 32 sampling 921/2000ml at 0.33ml/s with 21.9psi vacuum @18:58:04.50 Cartridge 32 sampling 940/2000ml at 0.29ml/s with 25.2psi vacuum @18:59:04.75 Cartridge 32 sampling 959/2000ml at 0.30ml/s with 19.4psi vacuum 2.2m deep (18.0psia) @19:00:04.50 Cartridge 32 sampling 977/2000ml at 0.30ml/s with 26.8psi vacuum @19:01:04.50 Cartridge 32 sampling 995/2000ml at 0.30ml/s with 26.4psi vacuum @19:02:04.72 Cartridge 32 sampling 1014/2000ml at 0.29ml/s with 26.0psi vacuum @19:03:04.50 Cartridge 32 sampling 1031/2000ml at 0.29ml/s with 24.5psi vacuum @19:04:04.50 Cartridge 32 sampling 1049/2000ml at 0.31ml/s with 21.7psi vacuum @19:05:04.87 Cartridge 32 sampling 1067/2000ml at 0.30ml/s with 20.8psi vacuum @19:06:04.65 Cartridge 32 sampling 1085/2000ml at 0.27ml/s with 26.0psi vacuum @19:07:04.72 Cartridge 32 sampling 1102/2000ml at 0.29ml/s with 21.4psi vacuum @19:08:04.50 Cartridge 32 sampling 1119/2000ml at 0.29ml/s with 20.3psi vacuum @19:09:04.86 Cartridge 32 sampling 1136/2000ml at 0.29ml/s with 24.3psi vacuum @19:10:04.49 Cartridge 32 sampling 1153/2000ml at 0.28ml/s with 22.5psi vacuum @19:11:04.66 Cartridge 32 sampling 1169/2000ml at 0.30ml/s with 24.3psi vacuum @19:12:04.49 Cartridge 32 sampling 1186/2000ml at 0.28ml/s with 21.0psi vacuum @19:13:04.65 Cartridge 32 sampling 1202/2000ml at 0.26ml/s with 19.9psi vacuum @19:14:04.50 Cartridge 32 sampling 1218/2000ml at 0.24ml/s with 24.4psi vacuum @19:15:04.66 Cartridge 32 sampling 1234/2000ml at 0.28ml/s with 24.9psi vacuum @19:16:04.50 Cartridge 32 sampling 1250/2000ml at 0.28ml/s with 23.0psi vacuum @19:17:04.93 Cartridge 32 sampling 1266/2000ml at 0.27ml/s with 20.6psi vacuum 1.7m deep (17.2psia) @19:18:04.50 Cartridge 32 sampling 1281/2000ml at 0.27ml/s with 22.7psi vacuum @19:19:04.66 Cartridge 32 sampling 1297/2000ml at 0.26ml/s with 28.2psi vacuum @19:20:04.50 Cartridge 32 sampling 1313/2000ml at 0.25ml/s with 14.3psi vacuum @19:21:04.67 Cartridge 32 sampling 1329/2000ml at 0.26ml/s with 21.7psi vacuum @19:22:04.50 Cartridge 32 sampling 1345/2000ml at 0.27ml/s with 28.5psi vacuum @19:23:04.72 Cartridge 32 sampling 1361/2000ml at 0.25ml/s with 21.3psi vacuum @19:24:04.49 Cartridge 32 sampling 1376/2000ml at 0.26ml/s with 23.3psi vacuum @19:25:04.65 Cartridge 32 sampling 1391/2000ml at 0.22ml/s with 25.8psi vacuum @19:26:04.50 Cartridge 32 sampling 1407/2000ml at 0.27ml/s with 23.8psi vacuum @19:27:04.60 Cartridge 32 sampling 1422/2000ml at 0.26ml/s with 28.2psi vacuum @19:28:04.52 Cartridge 32 sampling 1436/2000ml at 0.28ml/s with 25.9psi vacuum @19:29:04.69 Cartridge 32 sampling 1451/2000ml at 0.23ml/s with 25.4psi vacuum @19:30:04.50 Cartridge 32 sampling 1466/2000ml at 0.23ml/s with 21.5psi vacuum @19:31:04.64 Cartridge 32 sampling 1480/2000ml at 0.23ml/s with 28.6psi vacuum @19:32:04.48 Cartridge 32 sampling 1495/2000ml at 0.21ml/s with 28.4psi vacuum @19:33:04.71 Cartridge 32 sampling 1509/2000ml at 0.27ml/s with 19.9psi vacuum 2.3m deep (18.1psia) @19:34:04.49 Cartridge 32 sampling 1523/2000ml at 0.20ml/s with 22.3psi vacuum @19:35:04.65 Cartridge 32 sampling 1537/2000ml at 0.20ml/s with 26.5psi vacuum @19:36:04.48 Cartridge 32 sampling 1551/2000ml at 0.20ml/s with 24.8psi vacuum @19:37:04.57 Cartridge 32 sampling 1564/2000ml at 0.20ml/s with 28.8psi vacuum @19:38:04.49 Cartridge 32 sampling 1578/2000ml at 0.21ml/s with 28.5psi vacuum @19:39:04.60 Cartridge 32 sampling 1592/2000ml at 0.20ml/s with 25.4psi vacuum @19:40:04.68 Cartridge 32 sampling 1605/2000ml at 0.20ml/s with 27.4psi vacuum @19:41:04.57 Cartridge 32 sampling 1619/2000ml at 0.23ml/s with 22.9psi vacuum @19:42:04.49 Cartridge 32 sampling 1632/2000ml at 0.22ml/s with 25.2psi vacuum @19:43:04.78 Cartridge 32 sampling 1645/2000ml at 0.21ml/s with 21.8psi vacuum @19:44:04.67 Cartridge 32 sampling 1658/2000ml at 0.23ml/s with 25.3psi vacuum @19:45:04.50 Cartridge 32 sampling 1672/2000ml at 0.22ml/s with 15.2psi vacuum @19:46:04.79 Cartridge 32 sampling 1685/2000ml at 0.27ml/s with 23.1psi vacuum 2.0m deep (17.7psia) @19:47:04.50 Cartridge 32 sampling 1698/2000ml at 0.26ml/s with 21.2psi vacuum @19:48:04.65 Cartridge 32 sampling 1711/2000ml at 0.20ml/s with 26.0psi vacuum @19:49:04.50 Cartridge 32 sampling 1724/2000ml at 0.20ml/s with 26.3psi vacuum @19:50:04.73 Cartridge 32 sampling 1736/2000ml at 0.21ml/s with 29.6psi vacuum @19:51:04.50 Cartridge 32 sampling 1749/2000ml at 0.20ml/s with 26.7psi vacuum @19:52:04.76 Cartridge 32 sampling 1762/2000ml at 0.20ml/s with 26.8psi vacuum @19:53:04.49 Cartridge 32 sampling 1775/2000ml at 0.20ml/s with 22.5psi vacuum @19:54:04.66 Cartridge 32 sampling 1788/2000ml at 0.23ml/s with 17.5psi vacuum 4.0m deep (20.6psia) @19:55:04.50 Cartridge 32 sampling 1800/2000ml at 0.20ml/s with 26.8psi vacuum @19:56:04.65 Cartridge 32 sampling 1813/2000ml at 0.20ml/s with 23.6psi vacuum @19:57:04.50 Cartridge 32 sampling 1826/2000ml at 0.20ml/s with 28.8psi vacuum @19:58:04.51 Cartridge 32 sampling 1838/2000ml at 0.20ml/s with 28.0psi vacuum @19:59:04.57 Cartridge 32 sampling 1851/2000ml at 0.20ml/s with 29.0psi vacuum @20:00:04.59 Cartridge 32 sampling 1863/2000ml at 0.20ml/s with 27.5psi vacuum @20:01:04.98 Cartridge 32 sampling 1876/2000ml at 0.20ml/s with 28.1psi vacuum @20:02:04.62 Cartridge 32 sampling 1888/2000ml at 0.20ml/s with 29.5psi vacuum @20:03:04.50 Cartridge 32 sampling 1901/2000ml at 0.20ml/s with 25.6psi vacuum @20:04:04.58 Cartridge 32 sampling 1913/2000ml at 0.20ml/s with 28.2psi vacuum @20:05:04.50 Cartridge 32 sampling 1926/2000ml at 0.20ml/s with 28.8psi vacuum @20:06:04.51 Cartridge 32 sampling 1938/2000ml at 0.20ml/s with 28.0psi vacuum @20:07:04.50 Cartridge 32 sampling 1951/2000ml at 0.23ml/s with 22.5psi vacuum 1.6m deep (17.1psia) @20:08:04.75 Cartridge 32 sampling 1963/2000ml at 0.20ml/s with 29.3psi vacuum @20:09:05.10 Cartridge 32 sampling 1976/2000ml at 0.20ml/s with 22.6psi vacuum @20:10:04.48 Cartridge 32 sampling 1988/2000ml at 0.20ml/s with 29.4psi vacuum @20:11:02.90 Pressure.coast @20:11:03.04 Intake.close @20:11:03.09 Exhaust.close @20:11:03.29 Toroid.reconfigure :hold @20:11:03.35 Toroid.seek :bypass @20:11:05.01 Toroid.reconfigure :default @20:11:05.12 Toroid.seek :clear @20:11:06.85 Sampled 2000.0ml @20:11:07.78 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 32" @20:11:08.74 SamplePump.setPosition! 0ml @20:11:08.98 Plunge.seek :reagent @20:11:09.12 Exhaust.open @20:11:09.96 Intake.open @20:11:10.24 Sample loop closed. Ready to Deliver Reagents @20:11:10.26 Archive.seek :sealed @20:11:11.00 SamplePump.reconfigure :slow @20:11:11.06 SamplePump.seek -22ml @20:11:13.94 Archive.seek :evac @20:11:28.07 delay 1:00 @20:11:39.25 upsync @20:11:58.39 delay 1 second @20:11:59.42 Bypass.open @20:12:00.15 Exhaust.close @20:12:01.20 SamplePump.seek -20ml @20:12:05.81 delay 1 second @20:12:06.90 SamplePump.seek -28ml @20:12:24.27 delay 1 second @20:12:25.36 Intake.close @20:12:25.45 Bypass.close @20:12:28.12 Archive.seek :clear @20:12:45.27 RNAlater soak for 10:00 @20:22:45.33 Archive.seek :tVent @20:22:47.10 delay 1 second @20:22:48.20 Archive.seek :evac @20:23:04.28 delay 10 seconds @20:23:14.33 Sample preserved and evacuated @20:23:14.88
Carousel.seek 32 @20:23:15.41 Plunge.seek :locked @20:23:16.51 Archive.seek :clear @20:23:16.98 Air.seek :clear @20:23:17.38 Lysis1.seek :clear @20:23:33.92
Archive.coast @20:23:34.02 Air.coast @20:23:34.22 Lysis1.coast @20:23:34.26 Lysis2.coast @20:23:34.39 Plunge.coast @20:23:34.51 SamplePump.coast @20:23:34.55 Toroid.coast @20:23:34.59 Carousel.coast @20:23:34.95 Gate.power :motor,:OFF @20:23:34.96 Gate.power :core,:OFF @20:23:35.00 Gate.power :analytic1,:OFF @20:23:35.03 Gate.power :analytic2,:OFF @20:23:35.07 Gate.power :analytic3,:OFF @20:23:35.16 Gate.power :main,:OFF @20:23:35.18 Safely stopped and ready to power off @20:23:35.46 delayUntil "22:11:49.70" @20:23:35.55 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 22:11" @20:24:01.97 upsync @21:31:23.73 Session Begun @21:31:56.09 -> slots @21:33:21.32 -> Cartridge.release 11 @21:33:41.57 -> slots @21:33:44.72 -> exit @21:33:44.78 Session Ended @22:11:49.83
Gate.power :main,:ON,limit to 5000mA @22:11:49.84 Gate.power :analytic1,:ON,limit to 5000mA @22:11:49.85 Gate.power :analytic2,:ON,limit to 5000mA @22:11:49.87 Gate.power :analytic3,:ON,limit to 5000mA @22:11:49.90 Powering core bus @22:11:52.19 Archive.configure :default @22:11:52.33 Air.configure :default @22:11:52.44 Lysis1.configure :default @22:11:52.55 Lysis2.configure :default @22:11:52.68 Plunge.configure :default @22:11:52.83 SamplePump.configure :default @22:11:53.01 ExtraPump.configure :default @22:11:53.46 Toroid.configure :default @22:11:53.88 Carousel.configure :default @22:11:54.21 Pressure.configure :default @22:11:54.38 All elf controllers running firmware version 4.61 @22:11:54.43 Gate.power :motor,:ON,limit to 5000mA @22:11:54.84 Archive.reconfigure :default @22:11:55.43 Archive.home.jog 0 @22:11:55.94 Toroid.reconfigure :default @22:11:56.04 Lysis2.reconfigure :default @22:11:56.11 Air.reconfigure :default @22:11:56.24 Plunge.reconfigure :default @22:11:56.41 Lysis1.reconfigure :default @22:11:56.83 Lysis2.home.jog 0 @22:11:56.90 Toroid.home.jog 0 @22:11:57.10 Plunge.home.jog 0 @22:11:57.14 Air.home.jog 0 @22:11:57.27 Lysis1.home.jog 0 @22:11:58.12 Archive.reconfigure :homing @22:11:58.49 Archive.jog 150 @22:11:58.66 Plunge.reconfigure :homing @22:11:58.74 Lysis2.reconfigure :homing @22:11:58.77 Air.reconfigure :homing @22:11:58.81 Lysis1.reconfigure :homing @22:11:58.91 Toroid.reconfigure :homing @22:11:59.53 Plunge.jog 856 @22:11:59.64 Lysis2.jog 1359 @22:11:59.75 Air.jog 1371 @22:11:59.79 Lysis1.jog 1356 @22:11:59.84 Toroid.jog 608 @22:12:00.06 Archive.setRawPosition! 50 @22:12:00.10 Archive.seek :home @22:12:00.39 Archive.reconfigure :default @22:12:00.68 Plunge.setRawPosition! 50 @22:12:00.73 Plunge.seek :home @22:12:01.04 Plunge.reconfigure :default @22:12:01.77 Lysis2.setRawPosition! 50 @22:12:01.91 Lysis2.seek :home @22:12:02.02 Lysis1.setRawPosition! 50 @22:12:02.04 Air.setRawPosition! 50 @22:12:02.11 Lysis1.seek :home @22:12:02.16 Air.seek :home @22:12:02.34 Toroid.setRawPosition! 0 @22:12:02.36 Lysis2.reconfigure :default @22:12:02.50 Toroid.seek :home @22:12:02.62 Toroid.reconfigure :default @22:12:02.66 Lysis1.reconfigure :default @22:12:02.69 Air.reconfigure :default @22:12:03.18
Toroid.seek :clear @22:12:04.96 Carousel.reconfigure :default @22:12:05.08 Carousel.home.jog 0 @22:12:05.77 Carousel.seek 32 @22:12:06.08 Selecting Cartridge 16 @22:12:06.60 Carousel.seek 16 @22:12:10.21 Lysis1.seek :locked @22:12:13.67 Plunge.seek :locked @22:12:14.31 Air.seek :locked @22:12:18.89
Mounted archive Cartridge @22:12:19.02 GoodNews.email "Filtering up to 2000.0ml",Subject: "Sampling archive_bac cartridge in slot 16" @22:12:19.31 Duration of filtering limited to 2:05:00 @22:12:19.60 Priming sample loop w/160ml, bypass w/10ml @22:12:19.62 SamplePump.setPosition! 0ml @22:12:19.65 Bypass.open @22:12:21.12 Intake.open @22:12:22.03 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @22:12:24.77 Vacuum tare is 16% of full scale @22:12:24.80 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @22:12:25.49 Pressure.reconfigure :default @22:12:25.71 Pressure.hold @22:12:26.29 SamplePump.reconfigure :prime @22:12:26.39 SamplePump.seek 10ml,1:21 @22:12:37.60 Pressure.coast @22:12:37.65 SamplePump.setPosition! 0ml @22:12:37.70 Exhaust.open @22:12:39.14 Bypass.close @22:12:40.72 Pressure.reconfigure :default @22:12:40.78 Pressure.hold @22:12:41.06 SamplePump.reconfigure :prime @22:12:41.12 SamplePump.seek 160ml,13:51 @22:12:54.49 upsync @22:15:32.59 Pressure.coast @22:15:32.73 Bypass.open @22:15:33.56 Intake.close @22:15:34.46 Toroid.reconfigure :hold @22:15:34.52 Toroid.seek :crack @22:15:36.12 Toroid.seek :cartridge @22:15:37.05 Toroid.reconfigure :default @22:15:37.11 Toroid.seek :clear @22:15:39.52 Pressurizing cartridge to 26psi above 14.1psia @22:15:39.60 SamplePump.setPosition! 0ml @22:15:39.73 Intake.open @22:15:40.57 Bypass.close @22:15:41.37 Exhaust.close @22:15:42.37 Pressure.reconfigure :default @22:15:42.51 Pressure.hold @22:15:42.93 SamplePump.reconfigure :charge @22:15:43.06 SamplePump.seek 2.3ml,42 seconds @22:15:53.13 Pressure.coast @22:15:53.23 Intake.close @22:15:54.14 Pressurized cartridge with 0.443ml to 26.2psi @22:15:54.15 Isolated @35.5psia (21.5psi) @22:15:54.16 Checking for cartridge leaks from 35.5psia @22:15:54.18 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @22:16:02.78 Cartridge pressure fell 8% to 32.8psia @22:16:02.86 Sampling 2000.0ml @22:16:02.88 SamplePump.setPosition! 0ml @22:16:02.92 Exhaust.open @22:16:03.61 Intake.open @22:16:04.49 Pressure.reconfigure :default @22:16:04.55 Pressure.hold @22:16:04.83 SamplePump.reconfigure :sample @22:16:04.89 SamplePump.seek 2000ml,2:47:12 @22:16:07.89 Cartridge 16 sampling 1/2000ml at 0.46ml/s with 22.8psi vacuum @22:17:07.56 Cartridge 16 sampling 28/2000ml at 0.44ml/s with 25.7psi vacuum @22:18:07.74 Cartridge 16 sampling 54/2000ml at 0.44ml/s with 24.3psi vacuum @22:19:07.55 Cartridge 16 sampling 80/2000ml at 0.42ml/s with 26.2psi vacuum @22:20:07.72 Cartridge 16 sampling 105/2000ml at 0.43ml/s with 25.1psi vacuum @22:21:08.14 Cartridge 16 sampling 131/2000ml at 0.40ml/s with 20.9psi vacuum @22:22:07.80 Cartridge 16 sampling 156/2000ml at 0.41ml/s with 25.3psi vacuum @22:23:07.60 Cartridge 16 sampling 181/2000ml at 0.40ml/s with 24.1psi vacuum @22:24:07.56 Cartridge 16 sampling 205/2000ml at 0.42ml/s with 25.3psi vacuum @22:25:07.80 Cartridge 16 sampling 229/2000ml at 0.40ml/s with 23.3psi vacuum @22:26:07.56 Cartridge 16 sampling 253/2000ml at 0.39ml/s with 20.9psi vacuum @22:27:07.66 Cartridge 16 sampling 277/2000ml at 0.39ml/s with 21.5psi vacuum @22:28:07.56 Cartridge 16 sampling 301/2000ml at 0.41ml/s with 24.5psi vacuum @22:29:07.88 Cartridge 16 sampling 324/2000ml at 0.38ml/s with 21.8psi vacuum @22:30:07.56 Cartridge 16 sampling 347/2000ml at 0.39ml/s with 23.1psi vacuum @22:31:07.81 Cartridge 16 sampling 370/2000ml at 0.35ml/s with 24.0psi vacuum @22:32:07.56 Cartridge 16 sampling 393/2000ml at 0.39ml/s with 20.6psi vacuum @22:33:07.69 Cartridge 16 sampling 416/2000ml at 0.38ml/s with 22.2psi vacuum @22:34:07.56 Cartridge 16 sampling 438/2000ml at 0.37ml/s with 25.4psi vacuum @22:35:07.72 Cartridge 16 sampling 461/2000ml at 0.36ml/s with 26.7psi vacuum @22:36:07.55 Cartridge 16 sampling 483/2000ml at 0.36ml/s with 26.6psi vacuum @22:37:07.75 Cartridge 16 sampling 505/2000ml at 0.37ml/s with 25.5psi vacuum @22:38:07.56 Cartridge 16 sampling 527/2000ml at 0.34ml/s with 26.7psi vacuum @22:39:07.83 Cartridge 16 sampling 549/2000ml at 0.34ml/s with 26.5psi vacuum @22:40:07.55 Cartridge 16 sampling 570/2000ml at 0.33ml/s with 20.7psi vacuum @22:41:07.85 Cartridge 16 sampling 592/2000ml at 0.36ml/s with 23.1psi vacuum @22:42:07.72 Cartridge 16 sampling 613/2000ml at 0.35ml/s with 25.1psi vacuum @22:43:07.56 Cartridge 16 sampling 634/2000ml at 0.35ml/s with 24.9psi vacuum @22:44:07.72 Cartridge 16 sampling 654/2000ml at 0.33ml/s with 20.6psi vacuum @22:45:07.56 Cartridge 16 sampling 675/2000ml at 0.35ml/s with 20.3psi vacuum @22:46:07.88 Cartridge 16 sampling 695/2000ml at 0.34ml/s with 21.4psi vacuum @22:47:07.56 Cartridge 16 sampling 716/2000ml at 0.34ml/s with 20.5psi vacuum @22:48:07.79 Cartridge 16 sampling 736/2000ml at 0.33ml/s with 20.8psi vacuum @22:49:07.56 Cartridge 16 sampling 756/2000ml at 0.33ml/s with 24.2psi vacuum @22:50:07.65 Cartridge 16 sampling 776/2000ml at 0.33ml/s with 21.7psi vacuum @22:51:07.56 Cartridge 16 sampling 795/2000ml at 0.33ml/s with 20.0psi vacuum @22:52:07.78 Cartridge 16 sampling 815/2000ml at 0.33ml/s with 26.9psi vacuum @22:53:07.55 Cartridge 16 sampling 835/2000ml at 0.34ml/s with 23.7psi vacuum @22:54:07.88 Cartridge 16 sampling 854/2000ml at 0.34ml/s with 25.8psi vacuum @22:55:07.55 Cartridge 16 sampling 873/2000ml at 0.29ml/s with 25.3psi vacuum @22:56:07.72 Cartridge 16 sampling 892/2000ml at 0.30ml/s with 19.9psi vacuum @22:57:07.56 Cartridge 16 sampling 911/2000ml at 0.33ml/s with 25.7psi vacuum @22:58:07.69 Cartridge 16 sampling 929/2000ml at 0.33ml/s with 25.6psi vacuum @22:59:07.56 Cartridge 16 sampling 948/2000ml at 0.30ml/s with 20.6psi vacuum @23:00:07.73 Cartridge 16 sampling 966/2000ml at 0.29ml/s with 20.3psi vacuum @23:01:07.83 Cartridge 16 sampling 984/2000ml at 0.31ml/s with 22.5psi vacuum @23:02:07.55 Cartridge 16 sampling 1002/2000ml at 0.31ml/s with 25.1psi vacuum @23:03:07.56 Cartridge 16 sampling 1021/2000ml at 0.31ml/s with 22.7psi vacuum @23:04:07.78 Cartridge 16 sampling 1039/2000ml at 0.32ml/s with 23.4psi vacuum @23:05:07.55 Cartridge 16 sampling 1058/2000ml at 0.31ml/s with 23.1psi vacuum @23:06:07.60 Cartridge 16 sampling 1076/2000ml at 0.29ml/s with 26.6psi vacuum @23:07:07.56 Cartridge 16 sampling 1095/2000ml at 0.28ml/s with 26.2psi vacuum @23:08:07.70 Cartridge 16 sampling 1113/2000ml at 0.31ml/s with 24.0psi vacuum @23:09:07.56 Cartridge 16 sampling 1131/2000ml at 0.34ml/s with 24.6psi vacuum @23:10:07.55 Cartridge 16 sampling 1149/2000ml at 0.33ml/s with 27.3psi vacuum @23:11:07.94 Cartridge 16 sampling 1167/2000ml at 0.30ml/s with 23.6psi vacuum @23:12:07.69 Cartridge 16 sampling 1185/2000ml at 0.30ml/s with 21.3psi vacuum @23:13:07.55 Cartridge 16 sampling 1202/2000ml at 0.30ml/s with 21.4psi vacuum @23:14:07.55 Cartridge 16 sampling 1220/2000ml at 0.30ml/s with 21.1psi vacuum @23:15:07.62 Cartridge 16 sampling 1237/2000ml at 0.28ml/s with 21.3psi vacuum @23:16:07.55 Cartridge 16 sampling 1255/2000ml at 0.29ml/s with 20.6psi vacuum @23:17:08.14 Cartridge 16 sampling 1272/2000ml at 0.32ml/s with 20.4psi vacuum @23:18:07.95 Cartridge 16 sampling 1289/2000ml at 0.29ml/s with 21.6psi vacuum @23:19:07.72 Cartridge 16 sampling 1306/2000ml at 0.27ml/s with 19.4psi vacuum 2.4m deep (18.3psia) @23:20:07.75 Cartridge 16 sampling 1323/2000ml at 0.28ml/s with 12.7psi vacuum 5.0m deep (22.1psia) @23:21:07.56 Cartridge 16 sampling 1340/2000ml at 0.25ml/s with 28.6psi vacuum @23:22:07.77 Cartridge 16 sampling 1357/2000ml at 0.29ml/s with 28.6psi vacuum @23:23:07.83 Cartridge 16 sampling 1373/2000ml at 0.29ml/s with 25.2psi vacuum @23:24:07.73 Cartridge 16 sampling 1390/2000ml at 0.25ml/s with 26.2psi vacuum @23:25:07.56 Cartridge 16 sampling 1406/2000ml at 0.27ml/s with 27.5psi vacuum @23:26:07.81 Cartridge 16 sampling 1423/2000ml at 0.26ml/s with 26.7psi vacuum @23:27:07.56 Cartridge 16 sampling 1439/2000ml at 0.25ml/s with 21.6psi vacuum @23:28:07.55 Cartridge 16 sampling 1455/2000ml at 0.28ml/s with 16.1psi vacuum 1.9m deep (17.5psia) @23:29:07.64 Cartridge 16 sampling 1470/2000ml at 0.28ml/s with 27.9psi vacuum @23:30:07.94 Cartridge 16 sampling 1486/2000ml at 0.24ml/s with 26.3psi vacuum @23:31:07.56 Cartridge 16 sampling 1502/2000ml at 0.27ml/s with 28.7psi vacuum @23:32:07.55 Cartridge 16 sampling 1517/2000ml at 0.25ml/s with 17.8psi vacuum @23:33:07.72 Cartridge 16 sampling 1533/2000ml at 0.28ml/s with 21.1psi vacuum 2.2m deep (18.0psia) @23:34:07.54 Cartridge 16 sampling 1548/2000ml at 0.22ml/s with 24.8psi vacuum @23:35:07.72 Cartridge 16 sampling 1564/2000ml at 0.23ml/s with 25.1psi vacuum @23:36:07.55 Cartridge 16 sampling 1579/2000ml at 0.26ml/s with 28.7psi vacuum @23:37:07.97 Cartridge 16 sampling 1594/2000ml at 0.28ml/s with 27.1psi vacuum @23:38:07.55 Cartridge 16 sampling 1609/2000ml at 0.26ml/s with 20.1psi vacuum @23:39:07.68 Cartridge 16 sampling 1624/2000ml at 0.23ml/s with 21.4psi vacuum @23:40:07.98 Cartridge 16 sampling 1639/2000ml at 0.23ml/s with 25.0psi vacuum @23:41:07.55 Cartridge 16 sampling 1654/2000ml at 0.24ml/s with 17.2psi vacuum @23:42:07.91 Cartridge 16 sampling 1669/2000ml at 0.25ml/s with 28.4psi vacuum @23:43:07.55 Cartridge 16 sampling 1683/2000ml at 0.21ml/s with 23.8psi vacuum @23:44:07.73 Cartridge 16 sampling 1698/2000ml at 0.23ml/s with 25.6psi vacuum @23:45:07.55 Cartridge 16 sampling 1712/2000ml at 0.22ml/s with 23.5psi vacuum @23:46:07.77 Cartridge 16 sampling 1726/2000ml at 0.23ml/s with 27.0psi vacuum @23:47:07.56 Cartridge 16 sampling 1740/2000ml at 0.21ml/s with 20.4psi vacuum @23:48:07.56 Cartridge 16 sampling 1754/2000ml at 0.21ml/s with 18.4psi vacuum @23:49:07.72 Cartridge 16 sampling 1768/2000ml at 0.21ml/s with 20.0psi vacuum 2.6m deep (18.6psia) @23:50:07.55 Cartridge 16 sampling 1782/2000ml at 0.24ml/s with 17.4psi vacuum @23:51:07.70 Cartridge 16 sampling 1796/2000ml at 0.22ml/s with 24.3psi vacuum @23:52:07.56 Cartridge 16 sampling 1810/2000ml at 0.20ml/s with 27.5psi vacuum @23:53:07.66 Cartridge 16 sampling 1824/2000ml at 0.25ml/s with 11.3psi vacuum 6.0m deep (23.6psia) @23:54:07.73 Cartridge 16 sampling 1838/2000ml at 0.21ml/s with 25.6psi vacuum @23:55:07.56 Cartridge 16 sampling 1852/2000ml at 0.21ml/s with 19.4psi vacuum @23:56:07.83 Cartridge 16 sampling 1865/2000ml at 0.22ml/s with 29.6psi vacuum @23:57:07.56 Cartridge 16 sampling 1879/2000ml at 0.25ml/s with 20.6psi vacuum @23:58:07.87 Cartridge 16 sampling 1893/2000ml at 0.23ml/s with 28.3psi vacuum @23:59:07.55 Cartridge 16 sampling 1907/2000ml at 0.25ml/s with 24.5psi vacuum @00:00:07.72PST07-Mar-21 Cartridge 16 sampling 1920/2000ml at 0.20ml/s with 26.6psi vacuum @00:01:07.72 Cartridge 16 sampling 1934/2000ml at 0.21ml/s with 24.3psi vacuum @00:02:07.56 Cartridge 16 sampling 1947/2000ml at 0.24ml/s with 19.6psi vacuum @00:03:07.81 Cartridge 16 sampling 1961/2000ml at 0.23ml/s with 17.3psi vacuum 2.5m deep (18.3psia) @00:04:07.56 Cartridge 16 sampling 1974/2000ml at 0.20ml/s with 24.7psi vacuum @00:05:07.91 Cartridge 16 sampling 1987/2000ml at 0.24ml/s with 15.1psi vacuum 7.2m deep (25.3psia) @00:06:03.91 Pressure.coast @00:06:04.25 Intake.close @00:06:04.39 Exhaust.close @00:06:04.89 Toroid.reconfigure :hold @00:06:05.10 Toroid.seek :bypass @00:06:06.89 Toroid.reconfigure :default @00:06:06.95 Toroid.seek :clear @00:06:08.58 Sampled 2000.0ml @00:06:09.12 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 16" @00:06:09.22 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:06:09.26 Retry #1 of 10 after 10:00 @00:06:10.41 SamplePump.setPosition! 0ml @00:06:10.58 Plunge.seek :reagent @00:06:10.97 Exhaust.open @00:06:11.90 Sample loop closed. Ready to Deliver Reagents @00:06:11.95 Archive.seek :sealed @00:06:12.25 Intake.open @00:06:13.73 SamplePump.reconfigure :slow @00:06:14.01 SamplePump.seek -22ml @00:06:15.75 Archive.seek :evac @00:06:29.79 delay 1:00 @00:07:01.40 delay 1 second @00:07:02.56 Bypass.open @00:07:03.26 Exhaust.close @00:07:04.02 SamplePump.seek -20ml @00:07:08.60 delay 1 second @00:07:09.83 SamplePump.seek -28ml @00:07:27.38 delay 1 second @00:07:28.57 Intake.close @00:07:28.62 Bypass.close @00:07:29.85 Archive.seek :clear @00:07:46.98 RNAlater soak for 10:00 @00:17:47.02 Archive.seek :tVent @00:17:48.67 delay 1 second @00:17:49.71 Archive.seek :evac @00:18:05.65 delay 10 seconds @00:18:15.69 Sample preserved and evacuated @00:18:15.93
Carousel.seek 16 @00:18:16.14 Plunge.seek :locked @00:18:17.13 Archive.seek :clear @00:18:17.30 Air.seek :clear @00:18:17.49 Lysis1.seek :clear @00:18:34.31
Archive.coast @00:18:34.41 Air.coast @00:18:34.50 Lysis1.coast @00:18:34.59 Lysis2.coast @00:18:34.67 Plunge.coast @00:18:34.76 SamplePump.coast @00:18:34.85 Toroid.coast @00:18:34.94 Carousel.coast @00:18:35.27 Gate.power :motor,:OFF @00:18:35.30 Gate.power :core,:OFF @00:18:35.33 Gate.power :analytic1,:OFF @00:18:35.35 Gate.power :analytic2,:OFF @00:18:35.39 Gate.power :analytic3,:OFF @00:18:35.41 Gate.power :main,:OFF @00:18:35.44 Safely stopped and ready to power off @00:18:35.80 delayUntil "02:11:49.70" @00:18:35.85 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 02:11" @00:18:35.92 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:18:35.96 Retry #1 of 10 after 10:00 @02:11:49.99
Gate.power :main,:ON,limit to 5000mA @02:11:50.00 Gate.power :analytic1,:ON,limit to 5000mA @02:11:50.02 Gate.power :analytic2,:ON,limit to 5000mA @02:11:50.03 Gate.power :analytic3,:ON,limit to 5000mA @02:11:50.19 Powering core bus @02:11:52.52 Archive.configure :default @02:11:52.73 Air.configure :default @02:11:52.96 Lysis1.configure :default @02:11:53.07 Lysis2.configure :default @02:11:53.22 Plunge.configure :default @02:11:53.36 SamplePump.configure :default @02:11:53.50 ExtraPump.configure :default @02:11:53.63 Toroid.configure :default @02:11:53.77 Carousel.configure :default @02:11:53.92 Pressure.configure :default @02:11:53.97 All elf controllers running firmware version 4.61 @02:11:53.99 Gate.power :motor,:ON,limit to 5000mA @02:11:54.13 Archive.reconfigure :default @02:11:54.45 Archive.home.jog 0 @02:11:54.67 Toroid.reconfigure :default @02:11:54.70 Lysis2.reconfigure :default @02:11:54.74 Lysis1.reconfigure :default @02:11:54.80 Air.reconfigure :default @02:11:54.86 Plunge.reconfigure :default @02:11:55.09 Toroid.home.jog 0 @02:11:55.13 Lysis2.home.jog 0 @02:11:55.18 Lysis1.home.jog 0 @02:11:55.23 Plunge.home.jog 0 @02:11:55.26 Air.home.jog 0 @02:11:56.56 Archive.reconfigure :homing @02:11:57.29 Archive.jog 159 @02:11:57.41 Toroid.reconfigure :homing @02:11:57.55 Lysis2.reconfigure :homing @02:11:57.59 Plunge.reconfigure :homing @02:11:57.75 Air.reconfigure :homing @02:11:57.84 Lysis1.reconfigure :homing @02:11:58.37 Toroid.jog 621 @02:11:58.60 Plunge.jog 858 @02:11:58.62 Lysis2.jog 1358 @02:11:58.69 Lysis1.jog 1345 @02:11:58.87 Air.jog 1368 @02:11:59.44 Archive.setRawPosition! 50 @02:11:59.51 Archive.seek :home @02:11:59.94 Archive.reconfigure :default @02:12:00.01 Plunge.setRawPosition! 50 @02:12:00.14 Plunge.seek :home @02:12:00.36 Lysis2.setRawPosition! 50 @02:12:00.38 Lysis1.setRawPosition! 50 @02:12:00.62 Lysis1.seek :home @02:12:00.67 Lysis2.seek :home @02:12:00.88 Air.setRawPosition! 50 @02:12:01.12 Plunge.reconfigure :default @02:12:01.52 Air.seek :home @02:12:01.63 Lysis2.reconfigure :default @02:12:01.67 Lysis1.reconfigure :default @02:12:01.69 Toroid.setRawPosition! 0 @02:12:01.84 Toroid.seek :home @02:12:01.96 Toroid.reconfigure :default @02:12:01.98 Air.reconfigure :default @02:12:02.43
Toroid.seek :clear @02:12:04.05 Carousel.reconfigure :default @02:12:04.10 Carousel.home.jog 0 @02:12:04.61 Carousel.seek 16 @02:12:04.75 Selecting Cartridge 15 @02:12:04.97 Carousel.seek 15 @02:12:05.50 Lysis1.seek :locked @02:12:08.81 Plunge.seek :locked @02:12:09.28 Air.seek :locked @02:12:13.61
Mounted archive Cartridge @02:12:13.71 GoodNews.email "Filtering up to 2000.0ml",Subject: "Sampling archive_bac cartridge in slot 15" @02:12:14.03 Duration of filtering limited to 2:05:00 @02:12:14.33 Priming sample loop w/160ml, bypass w/10ml @02:12:14.35 SamplePump.setPosition! 0ml @02:12:14.38 Bypass.open @02:12:16.12 Intake.open @02:12:17.25 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @02:12:20.07 Vacuum tare is 16% of full scale @02:12:20.19 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @02:12:21.12 Pressure.reconfigure :default @02:12:21.18 Pressure.hold @02:12:21.40 SamplePump.reconfigure :prime @02:12:21.46 SamplePump.seek 10ml,1:21 @02:12:32.57 Pressure.coast @02:12:32.62 SamplePump.setPosition! 0ml @02:12:32.67 Exhaust.open @02:12:34.16 Bypass.close @02:12:36.12 Pressure.reconfigure :default @02:12:36.33 Pressure.hold @02:12:37.09 SamplePump.reconfigure :prime @02:12:37.29 SamplePump.seek 160ml,13:51 @02:15:28.78 Pressure.coast @02:15:28.84 Bypass.open @02:15:29.52 Intake.close @02:15:30.78 Toroid.reconfigure :hold @02:15:31.02 Toroid.seek :crack @02:15:32.84 Toroid.seek :cartridge @02:15:33.94 Toroid.reconfigure :default @02:15:34.05 Toroid.seek :clear @02:15:35.90 Pressurizing cartridge to 26psi above 14.3psia @02:15:35.91 SamplePump.setPosition! 0ml @02:15:35.95 Intake.open @02:15:36.66 Bypass.close @02:15:37.38 Exhaust.close @02:15:38.19 Pressure.reconfigure :default @02:15:38.25 Pressure.hold @02:15:38.45 SamplePump.reconfigure :charge @02:15:38.51 SamplePump.seek 2.3ml,42 seconds @02:15:48.50 Pressure.coast @02:15:48.55 Intake.close @02:15:49.71 Pressurized cartridge with 0.456ml to 26.0psi @02:15:49.75 Isolated @34.5psia (20.3psi) @02:15:49.79 Checking for cartridge leaks from 34.5psia @02:15:49.82 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @02:15:54.41 Cartridge pressure fell 2% to 33.9psia @02:15:54.49 Sampling 2000.0ml @02:15:54.51 SamplePump.setPosition! 0ml @02:15:54.55 Exhaust.open @02:15:55.22 Intake.open @02:15:56.16 Pressure.reconfigure :default @02:15:56.21 Pressure.hold @02:15:56.49 SamplePump.reconfigure :sample @02:15:56.55 SamplePump.seek 2000ml,2:47:12 @02:15:59.42 Cartridge 15 sampling 1/2000ml at 0.45ml/s with 26.3psi vacuum @02:16:59.22 Cartridge 15 sampling 27/2000ml at 0.44ml/s with 25.8psi vacuum @02:17:59.39 Cartridge 15 sampling 53/2000ml at 0.42ml/s with 24.1psi vacuum @02:18:59.22 Cartridge 15 sampling 79/2000ml at 0.42ml/s with 21.1psi vacuum @02:19:59.38 Cartridge 15 sampling 105/2000ml at 0.42ml/s with 21.7psi vacuum @02:20:59.21 Cartridge 15 sampling 130/2000ml at 0.42ml/s with 22.3psi vacuum @02:21:59.35 Cartridge 15 sampling 155/2000ml at 0.42ml/s with 24.7psi vacuum @02:22:59.22 Cartridge 15 sampling 180/2000ml at 0.41ml/s with 25.2psi vacuum @02:23:59.29 Cartridge 15 sampling 204/2000ml at 0.40ml/s with 21.4psi vacuum @02:24:59.22 Cartridge 15 sampling 228/2000ml at 0.40ml/s with 21.4psi vacuum @02:25:59.43 Cartridge 15 sampling 252/2000ml at 0.41ml/s with 22.8psi vacuum @02:26:59.22 Cartridge 15 sampling 276/2000ml at 0.39ml/s with 25.4psi vacuum @02:27:59.25 Cartridge 15 sampling 300/2000ml at 0.39ml/s with 26.5psi vacuum @02:28:59.22 Cartridge 15 sampling 323/2000ml at 0.38ml/s with 26.3psi vacuum @02:29:59.41 Cartridge 15 sampling 346/2000ml at 0.39ml/s with 21.3psi vacuum @02:30:59.22 Cartridge 15 sampling 369/2000ml at 0.36ml/s with 20.3psi vacuum @02:31:59.43 Cartridge 15 sampling 392/2000ml at 0.38ml/s with 20.8psi vacuum @02:32:59.22 Cartridge 15 sampling 414/2000ml at 0.37ml/s with 26.0psi vacuum @02:33:59.39 Cartridge 15 sampling 437/2000ml at 0.36ml/s with 26.5psi vacuum @02:34:59.22 Cartridge 15 sampling 459/2000ml at 0.38ml/s with 21.1psi vacuum @02:35:59.26 Cartridge 15 sampling 481/2000ml at 0.38ml/s with 25.6psi vacuum @02:36:59.22 Cartridge 15 sampling 503/2000ml at 0.36ml/s with 26.4psi vacuum @02:37:59.37 Cartridge 15 sampling 524/2000ml at 0.37ml/s with 25.7psi vacuum @02:38:59.22 Cartridge 15 sampling 546/2000ml at 0.37ml/s with 22.6psi vacuum @02:39:59.47 Cartridge 15 sampling 567/2000ml at 0.36ml/s with 23.9psi vacuum @02:40:59.22 Cartridge 15 sampling 588/2000ml at 0.34ml/s with 26.5psi vacuum @02:41:59.28 Cartridge 15 sampling 609/2000ml at 0.36ml/s with 24.3psi vacuum @02:42:59.22 Cartridge 15 sampling 630/2000ml at 0.36ml/s with 20.6psi vacuum @02:43:59.46 Cartridge 15 sampling 651/2000ml at 0.34ml/s with 20.3psi vacuum @02:44:59.21 Cartridge 15 sampling 671/2000ml at 0.32ml/s with 24.3psi vacuum @02:45:59.49 Cartridge 15 sampling 692/2000ml at 0.33ml/s with 26.9psi vacuum @02:46:59.21 Cartridge 15 sampling 712/2000ml at 0.34ml/s with 18.4psi vacuum @02:47:59.31 Cartridge 15 sampling 733/2000ml at 0.34ml/s with 17.9psi vacuum 1.7m deep (17.2psia) @02:48:59.33 Cartridge 15 sampling 753/2000ml at 0.34ml/s with 22.6psi vacuum @02:50:00.78 Cartridge 15 sampling 772/2000ml at 0.34ml/s with 22.9psi vacuum @02:50:59.21 Cartridge 15 sampling 792/2000ml at 0.31ml/s with 17.3psi vacuum @02:51:59.39 Cartridge 15 sampling 812/2000ml at 0.34ml/s with 18.3psi vacuum 2.7m deep (18.7psia) @02:52:59.34 Cartridge 15 sampling 831/2000ml at 0.32ml/s with 26.1psi vacuum @02:53:59.40 Cartridge 15 sampling 851/2000ml at 0.33ml/s with 26.8psi vacuum @02:54:59.22 Cartridge 15 sampling 870/2000ml at 0.32ml/s with 26.0psi vacuum @02:55:59.37 Cartridge 15 sampling 889/2000ml at 0.31ml/s with 26.1psi vacuum @02:56:59.34 Cartridge 15 sampling 908/2000ml at 0.32ml/s with 20.3psi vacuum @02:57:59.40 Cartridge 15 sampling 926/2000ml at 0.33ml/s with 26.3psi vacuum @02:58:59.22 Cartridge 15 sampling 945/2000ml at 0.33ml/s with 24.1psi vacuum @02:59:59.32 Cartridge 15 sampling 964/2000ml at 0.30ml/s with 17.2psi vacuum 3.6m deep (20.1psia) @03:00:59.22 Cartridge 15 sampling 983/2000ml at 0.30ml/s with 27.3psi vacuum @03:01:59.41 Cartridge 15 sampling 1002/2000ml at 0.29ml/s with 26.3psi vacuum @03:02:59.22 Cartridge 15 sampling 1021/2000ml at 0.29ml/s with 27.0psi vacuum @03:03:59.36 Cartridge 15 sampling 1039/2000ml at 0.31ml/s with 20.5psi vacuum 1.6m deep (17.1psia) @03:04:59.22 Cartridge 15 sampling 1058/2000ml at 0.33ml/s with 20.8psi vacuum @03:05:59.55 Cartridge 15 sampling 1076/2000ml at 0.31ml/s with 27.8psi vacuum @03:06:59.60 Cartridge 15 sampling 1094/2000ml at 0.32ml/s with 28.1psi vacuum @03:07:59.49 Cartridge 15 sampling 1112/2000ml at 0.27ml/s with 25.7psi vacuum @03:08:59.43 Cartridge 15 sampling 1130/2000ml at 0.28ml/s with 26.3psi vacuum @03:09:59.21 Cartridge 15 sampling 1148/2000ml at 0.27ml/s with 27.6psi vacuum @03:10:59.38 Cartridge 15 sampling 1166/2000ml at 0.27ml/s with 20.1psi vacuum @03:11:59.21 Cartridge 15 sampling 1184/2000ml at 0.26ml/s with 22.0psi vacuum @03:12:59.29 Cartridge 15 sampling 1201/2000ml at 0.28ml/s with 28.1psi vacuum @03:13:59.21 Cartridge 15 sampling 1218/2000ml at 0.29ml/s with 16.3psi vacuum 2.5m deep (18.4psia) @03:14:59.34 Cartridge 15 sampling 1236/2000ml at 0.31ml/s with 26.1psi vacuum @03:15:59.22 Cartridge 15 sampling 1253/2000ml at 0.28ml/s with 23.5psi vacuum @03:16:59.54 Cartridge 15 sampling 1270/2000ml at 0.30ml/s with 19.4psi vacuum 3.7m deep (20.2psia) @03:17:59.20 Cartridge 15 sampling 1287/2000ml at 0.27ml/s with 26.8psi vacuum @03:18:59.38 Cartridge 15 sampling 1303/2000ml at 0.25ml/s with 24.8psi vacuum @03:19:59.55 Cartridge 15 sampling 1320/2000ml at 0.30ml/s with 26.9psi vacuum @03:20:59.57 Cartridge 15 sampling 1337/2000ml at 0.27ml/s with 27.8psi vacuum @03:21:59.40 Cartridge 15 sampling 1353/2000ml at 0.27ml/s with 20.0psi vacuum 1.7m deep (17.2psia) @03:22:59.50 Cartridge 15 sampling 1369/2000ml at 0.30ml/s with 25.3psi vacuum @03:23:59.22 Cartridge 15 sampling 1385/2000ml at 0.29ml/s with 25.1psi vacuum @03:24:59.22 Cartridge 15 sampling 1401/2000ml at 0.26ml/s with 20.7psi vacuum @03:25:59.40 Cartridge 15 sampling 1417/2000ml at 0.25ml/s with 25.5psi vacuum @03:26:59.64 Cartridge 15 sampling 1433/2000ml at 0.23ml/s with 26.2psi vacuum @03:27:59.36 Cartridge 15 sampling 1448/2000ml at 0.27ml/s with 28.3psi vacuum @03:28:59.85 Cartridge 15 sampling 1464/2000ml at 0.24ml/s with 26.4psi vacuum @03:29:59.22 Cartridge 15 sampling 1479/2000ml at 0.24ml/s with 25.6psi vacuum @03:30:59.53 Cartridge 15 sampling 1494/2000ml at 0.25ml/s with 26.8psi vacuum @03:31:59.22 Cartridge 15 sampling 1509/2000ml at 0.22ml/s with 24.7psi vacuum @03:32:59.22 Cartridge 15 sampling 1524/2000ml at 0.28ml/s with 21.6psi vacuum @03:33:59.41 Cartridge 15 sampling 1539/2000ml at 0.28ml/s with 22.3psi vacuum @03:34:59.22 Cartridge 15 sampling 1554/2000ml at 0.25ml/s with 28.1psi vacuum @03:35:59.36 Cartridge 15 sampling 1569/2000ml at 0.23ml/s with 21.5psi vacuum @03:36:59.22 Cartridge 15 sampling 1583/2000ml at 0.22ml/s with 25.6psi vacuum @03:37:59.57 Cartridge 15 sampling 1598/2000ml at 0.23ml/s with 25.8psi vacuum @03:38:59.54 Cartridge 15 sampling 1613/2000ml at 0.27ml/s with 22.7psi vacuum @03:39:59.22 Cartridge 15 sampling 1627/2000ml at 0.23ml/s with 26.1psi vacuum @03:40:59.68 Cartridge 15 sampling 1641/2000ml at 0.22ml/s with 24.3psi vacuum @03:41:59.22 Cartridge 15 sampling 1655/2000ml at 0.22ml/s with 28.0psi vacuum @03:42:59.53 Cartridge 15 sampling 1670/2000ml at 0.22ml/s with 16.4psi vacuum 4.9m deep (22.0psia) @03:43:59.22 Cartridge 15 sampling 1684/2000ml at 0.26ml/s with 23.7psi vacuum @03:44:59.39 Cartridge 15 sampling 1697/2000ml at 0.22ml/s with 25.8psi vacuum @03:45:59.21 Cartridge 15 sampling 1711/2000ml at 0.21ml/s with 24.1psi vacuum @03:46:59.36 Cartridge 15 sampling 1725/2000ml at 0.26ml/s with 25.3psi vacuum @03:47:59.22 Cartridge 15 sampling 1739/2000ml at 0.23ml/s with 15.4psi vacuum @03:48:59.38 Cartridge 15 sampling 1752/2000ml at 0.21ml/s with 24.2psi vacuum @03:49:59.22 Cartridge 15 sampling 1766/2000ml at 0.20ml/s with 21.6psi vacuum @03:50:59.39 Cartridge 15 sampling 1779/2000ml at 0.23ml/s with 28.7psi vacuum @03:51:59.22 Cartridge 15 sampling 1792/2000ml at 0.24ml/s with 27.5psi vacuum @03:52:59.37 Cartridge 15 sampling 1806/2000ml at 0.20ml/s with 25.1psi vacuum @03:53:59.22 Cartridge 15 sampling 1819/2000ml at 0.20ml/s with 24.5psi vacuum @03:54:59.55 Cartridge 15 sampling 1832/2000ml at 0.22ml/s with 15.1psi vacuum 6.6m deep (24.4psia) @03:55:59.59 Cartridge 15 sampling 1845/2000ml at 0.23ml/s with 18.4psi vacuum 1.9m deep (17.4psia) @03:56:59.35 Cartridge 15 sampling 1858/2000ml at 0.22ml/s with 14.2psi vacuum 4.9m deep (22.0psia) @03:57:59.51 Cartridge 15 sampling 1871/2000ml at 0.24ml/s with 22.3psi vacuum @03:58:59.22 Cartridge 15 sampling 1885/2000ml at 0.20ml/s with 24.9psi vacuum @03:59:59.30 Cartridge 15 sampling 1898/2000ml at 0.20ml/s with 24.8psi vacuum @04:00:59.54 Cartridge 15 sampling 1911/2000ml at 0.20ml/s with 24.8psi vacuum @04:01:59.22 Cartridge 15 sampling 1925/2000ml at 0.23ml/s with 28.2psi vacuum @04:02:59.41 Cartridge 15 sampling 1938/2000ml at 0.21ml/s with 28.4psi vacuum @04:03:59.22 Cartridge 15 sampling 1951/2000ml at 0.20ml/s with 26.2psi vacuum @04:04:59.33 Cartridge 15 sampling 1965/2000ml at 0.21ml/s with 9.4psi vacuum 3.4m deep (19.7psia) @04:05:59.22 Cartridge 15 sampling 1978/2000ml at 0.20ml/s with 29.2psi vacuum @04:06:59.57 Cartridge 15 sampling 1991/2000ml at 0.21ml/s with 27.9psi vacuum @04:07:37.44 Pressure.coast @04:07:37.78 Intake.close @04:07:38.00 Exhaust.close @04:07:38.34 Toroid.reconfigure :hold @04:07:38.40 Toroid.seek :bypass @04:07:40.11 Toroid.reconfigure :default @04:07:40.17 Toroid.seek :clear @04:07:41.79 Sampled 2000.0ml @04:07:42.79 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 15" @04:07:43.23 SamplePump.setPosition! 0ml @04:07:43.27 Plunge.seek :reagent @04:07:43.36 Exhaust.open @04:07:44.14 Intake.open @04:07:44.43 Sample loop closed. Ready to Deliver Reagents @04:07:44.49 Archive.seek :sealed @04:07:45.48 SamplePump.reconfigure :slow @04:07:45.65 SamplePump.seek -22ml @04:07:48.25 Archive.seek :evac @04:08:01.75 Slide::Error in PROCESSING -- Archive Syringe (ArS) positionErr at 34uL (sensed 63uL) @04:08:01.77 Retry #1 of 2 @04:08:01.78 Archive.hold @04:08:02.19 Archive.reconfigure :default @04:08:02.25 Archive.home.seek :evac @04:08:03.13 Slide::Error in PROCESSING -- Archive Syringe (ArS) positionErr at 31uL (sensed 60uL) @04:08:03.16 Retry #2 of 2 @04:08:03.19 Archive.reconfigure :default @04:08:03.29 Archive.hold @04:08:03.49 Archive.reconfigure :default @04:08:03.63 Archive.home.seek :evac @04:08:04.37 Slide::Error in PROCESSING -- IGNORED Archive Syringe (ArS) positionErr at 30uL (sensed 59uL) @04:08:04.41 Archive.reconfigure :default @04:08:04.51 Archive.hold @04:08:04.71 Archive.reconfigure :default @04:08:04.87 Archive.home.jog 0 @04:08:05.17 delay 1:00 @04:08:33.09 delay 1 second @04:08:34.13 Bypass.open @04:08:34.92 Exhaust.close @04:08:35.90 SamplePump.seek -20ml @04:08:40.44 delay 1 second @04:08:41.53 SamplePump.seek -28ml @04:08:58.89 delay 1 second @04:08:59.98 Intake.close @04:09:00.05 Bypass.close @04:09:05.35 Archive.seek :clear @04:09:22.09 RNAlater soak for 10:00 @04:19:22.15 Archive.seek :tVent @04:19:23.87 delay 1 second @04:19:24.90 Archive.seek :evac @04:19:40.66 Slide::Error in PROCESSING -- Archive Syringe (ArS) positionErr at 15uL (sensed 45uL) @04:19:40.70 Retry #1 of 2 @04:19:40.77 Archive.hold @04:19:41.05 Archive.reconfigure :default @04:19:41.18 Archive.home.seek :evac @04:19:41.79 Slide::Error in PROCESSING -- Archive Syringe (ArS) positionErr at 15uL (sensed 44uL) @04:19:41.80 Retry #2 of 2 @04:19:41.82 Archive.reconfigure :default @04:19:41.87 Archive.hold @04:19:42.16 Archive.reconfigure :default @04:19:42.22 Archive.home.seek :evac @04:19:42.91 Slide::Error in PROCESSING -- IGNORED Archive Syringe (ArS) positionErr at 15uL (sensed 44uL) @04:19:42.94 Archive.reconfigure :default @04:19:43.00 Archive.hold @04:19:43.20 Archive.reconfigure :default @04:19:43.26 Archive.home.jog 0 @04:19:43.37 delay 10 seconds @04:19:53.41 Sample preserved and evacuated @04:19:53.65
Carousel.seek 15 @04:19:53.85 Plunge.seek :locked @04:19:55.15 Archive.seek :clear @04:19:55.54 Air.seek :clear @04:19:56.15 Lysis1.seek :clear @04:20:12.07
Archive.coast @04:20:12.16 Air.coast @04:20:12.27 Lysis1.coast @04:20:12.33 Lysis2.coast @04:20:12.38 Plunge.coast @04:20:12.42 SamplePump.coast @04:20:12.46 Toroid.coast @04:20:12.51 Carousel.coast @04:20:12.70 Gate.power :motor,:OFF @04:20:12.71 Gate.power :core,:OFF @04:20:12.73 Gate.power :analytic1,:OFF @04:20:12.74 Gate.power :analytic2,:OFF @04:20:12.75 Gate.power :analytic3,:OFF @04:20:12.76 Gate.power :main,:OFF @04:20:12.78 Safely stopped and ready to power off @04:20:12.98 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 06:11" @04:20:13.01
delayUntil "06:11:49.70" @06:11:49.82 Gate.power :main,:ON,limit to 5000mA @06:11:49.83 Gate.power :analytic1,:ON,limit to 5000mA @06:11:49.85 Gate.power :analytic2,:ON,limit to 5000mA @06:11:49.86 Gate.power :analytic3,:ON,limit to 5000mA @06:11:49.90 Powering core bus @06:11:52.18 Archive.configure :default @06:11:52.33 Air.configure :default @06:11:52.44 Lysis1.configure :default @06:11:52.55 Lysis2.configure :default @06:11:52.70 Plunge.configure :default @06:11:52.84 SamplePump.configure :default @06:11:52.98 ExtraPump.configure :default @06:11:53.12 Toroid.configure :default @06:11:53.25 Carousel.configure :default @06:11:53.40 Pressure.configure :default @06:11:53.45 All elf controllers running firmware version 4.61 @06:11:53.47 Gate.power :motor,:ON,limit to 5000mA @06:11:53.74 Air.reconfigure :default @06:11:53.95 Archive.reconfigure :default @06:11:54.10 Air.home.jog 0 @06:11:54.22 Archive.home.jog 0 @06:11:55.36 Plunge.reconfigure :default @06:11:55.45 Lysis2.reconfigure :default @06:11:55.72 Lysis1.reconfigure :default @06:11:55.75 Toroid.reconfigure :default @06:11:56.23 Lysis2.home.jog 0 @06:11:56.29 Lysis1.home.jog 0 @06:11:56.45 Plunge.home.jog 0 @06:11:56.49 Toroid.home.jog 0 @06:11:57.40 Archive.reconfigure :homing @06:11:57.52 Air.reconfigure :homing @06:11:57.94 Archive.jog 177 @06:11:58.01 Air.jog 1374 @06:11:58.38 Toroid.reconfigure :homing @06:11:58.58 Plunge.reconfigure :homing @06:11:58.64 Lysis1.reconfigure :homing @06:11:58.78 Lysis2.reconfigure :homing @06:11:58.99 Toroid.jog 611 @06:11:59.24 Plunge.jog 859 @06:11:59.29 Lysis2.jog 1363 @06:11:59.35 Lysis1.jog 1343 @06:11:59.98 Air.setRawPosition! 50 @06:12:00.03 Archive.setRawPosition! 50 @06:12:00.27 Air.seek :home @06:12:00.41 Archive.seek :home @06:12:00.67 Plunge.setRawPosition! 50 @06:12:00.71 Plunge.seek :home @06:12:00.83 Lysis1.setRawPosition! 50 @06:12:00.90 Lysis1.seek :home @06:12:01.04 Plunge.reconfigure :default @06:12:01.06 Lysis2.setRawPosition! 50 @06:12:01.10 Air.reconfigure :default @06:12:01.12 Archive.reconfigure :default @06:12:01.31 Lysis2.seek :home @06:12:01.48 Toroid.setRawPosition! 0 @06:12:01.50 Lysis1.reconfigure :default @06:12:01.60 Toroid.seek :home @06:12:01.75 Toroid.reconfigure :default @06:12:01.78 Lysis2.reconfigure :default @06:12:02.21
Toroid.seek :clear @06:12:03.83 Carousel.reconfigure :default @06:12:03.88 Carousel.home.jog 0 @06:12:04.39 Carousel.seek 15 @06:12:04.56 Selecting Cartridge 14 @06:12:04.79 Carousel.seek 14 @06:12:05.58 Lysis1.seek :locked @06:12:09.03 Plunge.seek :locked @06:12:09.33 Air.seek :locked @06:12:13.56
Mounted archive Cartridge @06:12:13.66 GoodNews.email "Filtering up to 2000.0ml",Subject: "Sampling archive_bac cartridge in slot 14" @06:12:13.98 Duration of filtering limited to 2:05:00 @06:12:14.28 Priming sample loop w/160ml, bypass w/10ml @06:12:14.30 SamplePump.setPosition! 0ml @06:12:14.33 Bypass.open @06:12:15.83 Intake.open @06:12:16.90 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @06:12:19.72 Vacuum tare is 17% of full scale @06:12:19.74 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @06:12:19.98 Pressure.reconfigure :default @06:12:20.04 Pressure.hold @06:12:20.25 SamplePump.reconfigure :prime @06:12:20.31 SamplePump.seek 10ml,1:21 @06:12:31.41 Pressure.coast @06:12:31.46 SamplePump.setPosition! 0ml @06:12:31.51 Exhaust.open @06:12:32.93 Bypass.close @06:12:34.71 Pressure.hold @06:12:35.47 SamplePump.reconfigure :prime @06:12:35.56 SamplePump.seek 160ml,13:51 @06:15:27.24 Pressure.coast @06:15:27.30 Bypass.open @06:15:28.01 Intake.close @06:15:28.98 Toroid.reconfigure :hold @06:15:29.09 Toroid.seek :crack @06:15:30.87 Toroid.seek :cartridge @06:15:31.85 Toroid.reconfigure :default @06:15:31.97 Toroid.seek :clear @06:15:34.21 Pressurizing cartridge to 26psi above 14.2psia @06:15:34.33 SamplePump.setPosition! 0ml @06:15:34.36 Intake.open @06:15:35.09 Bypass.close @06:15:35.97 Exhaust.close @06:15:37.04 Pressure.reconfigure :default @06:15:37.10 Pressure.hold @06:15:37.31 SamplePump.reconfigure :charge @06:15:37.37 SamplePump.seek 2.3ml,42 seconds @06:15:47.02 Pressure.coast @06:15:47.08 Intake.close @06:15:47.95 Pressurized cartridge with 0.426ml to 26.1psi @06:15:47.97 Isolated @35.2psia (21.1psi) @06:15:47.98 Checking for cartridge leaks from 35.2psia @06:15:48.00 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @06:15:52.60 Cartridge pressure fell 3% to 34.3psia @06:15:52.88 Sampling 2000.0ml @06:15:53.00 SamplePump.setPosition! 0ml @06:15:53.07 Exhaust.open @06:15:53.86 Intake.open @06:15:55.02 Pressure.reconfigure :default @06:15:55.08 Pressure.hold @06:15:55.67 SamplePump.reconfigure :sample @06:15:55.75 SamplePump.seek 2000ml,2:47:12 @06:15:57.97 Cartridge 14 sampling 1/2000ml at 0.45ml/s with 22.0psi vacuum @06:16:58.27 Cartridge 14 sampling 27/2000ml at 0.44ml/s with 22.6psi vacuum @06:17:58.15 Cartridge 14 sampling 53/2000ml at 0.40ml/s with 22.2psi vacuum @06:18:57.98 Cartridge 14 sampling 79/2000ml at 0.43ml/s with 25.1psi vacuum @06:19:58.26 Cartridge 14 sampling 104/2000ml at 0.41ml/s with 22.4psi vacuum @06:20:57.98 Cartridge 14 sampling 129/2000ml at 0.39ml/s with 26.1psi vacuum @06:21:58.15 Cartridge 14 sampling 154/2000ml at 0.41ml/s with 22.3psi vacuum @06:22:57.98 Cartridge 14 sampling 178/2000ml at 0.41ml/s with 24.9psi vacuum @06:23:58.72 Cartridge 14 sampling 202/2000ml at 0.41ml/s with 22.5psi vacuum @06:24:57.98 Cartridge 14 sampling 226/2000ml at 0.37ml/s with 19.8psi vacuum @06:25:58.15 Cartridge 14 sampling 250/2000ml at 0.40ml/s with 24.4psi vacuum @06:26:57.98 Cartridge 14 sampling 274/2000ml at 0.39ml/s with 22.0psi vacuum @06:27:58.14 Cartridge 14 sampling 297/2000ml at 0.39ml/s with 22.6psi vacuum @06:28:57.98 Cartridge 14 sampling 320/2000ml at 0.39ml/s with 21.9psi vacuum @06:29:58.13 Cartridge 14 sampling 343/2000ml at 0.39ml/s with 22.2psi vacuum @06:30:57.98 Cartridge 14 sampling 366/2000ml at 0.36ml/s with 25.7psi vacuum @06:31:58.03 Cartridge 14 sampling 389/2000ml at 0.37ml/s with 22.5psi vacuum @06:32:58.06 Cartridge 14 sampling 411/2000ml at 0.37ml/s with 26.2psi vacuum @06:33:57.99 Cartridge 14 sampling 433/2000ml at 0.36ml/s with 27.0psi vacuum @06:34:58.35 Cartridge 14 sampling 455/2000ml at 0.37ml/s with 28.0psi vacuum @06:35:57.99 Cartridge 14 sampling 477/2000ml at 0.36ml/s with 21.9psi vacuum @06:36:58.15 Cartridge 14 sampling 499/2000ml at 0.36ml/s with 26.1psi vacuum @06:37:57.99 Cartridge 14 sampling 521/2000ml at 0.34ml/s with 21.7psi vacuum @06:38:58.22 Cartridge 14 sampling 542/2000ml at 0.34ml/s with 25.8psi vacuum @06:39:57.99 Cartridge 14 sampling 563/2000ml at 0.36ml/s with 27.1psi vacuum @06:40:58.00 Cartridge 14 sampling 584/2000ml at 0.37ml/s with 21.9psi vacuum @06:41:58.30 Cartridge 14 sampling 605/2000ml at 0.36ml/s with 22.3psi vacuum @06:42:57.98 Cartridge 14 sampling 626/2000ml at 0.36ml/s with 22.7psi vacuum @06:43:58.38 Cartridge 14 sampling 647/2000ml at 0.34ml/s with 26.1psi vacuum @06:44:57.98 Cartridge 14 sampling 667/2000ml at 0.36ml/s with 20.0psi vacuum @06:45:58.32 Cartridge 14 sampling 687/2000ml at 0.32ml/s with 22.6psi vacuum @06:46:58.13 Cartridge 14 sampling 707/2000ml at 0.33ml/s with 21.3psi vacuum @06:47:58.23 Cartridge 14 sampling 728/2000ml at 0.36ml/s with 25.9psi vacuum @06:48:57.98 Cartridge 14 sampling 748/2000ml at 0.32ml/s with 20.1psi vacuum @06:49:58.41 Cartridge 14 sampling 768/2000ml at 0.35ml/s with 23.4psi vacuum @06:50:58.12 Cartridge 14 sampling 788/2000ml at 0.33ml/s with 28.3psi vacuum @06:51:57.99 Cartridge 14 sampling 808/2000ml at 0.35ml/s with 26.6psi vacuum @06:52:58.09 Cartridge 14 sampling 828/2000ml at 0.31ml/s with 18.4psi vacuum @06:53:57.99 Cartridge 14 sampling 848/2000ml at 0.33ml/s with 21.0psi vacuum @06:54:58.17 Cartridge 14 sampling 867/2000ml at 0.32ml/s with 26.3psi vacuum @06:55:58.03 Cartridge 14 sampling 887/2000ml at 0.35ml/s with 26.0psi vacuum @06:56:58.30 Cartridge 14 sampling 906/2000ml at 0.30ml/s with 28.1psi vacuum @06:57:57.99 Cartridge 14 sampling 925/2000ml at 0.29ml/s with 22.4psi vacuum @06:58:58.16 Cartridge 14 sampling 944/2000ml at 0.32ml/s with 27.2psi vacuum @06:59:57.99 Cartridge 14 sampling 963/2000ml at 0.29ml/s with 17.2psi vacuum @07:00:58.14 Cartridge 14 sampling 982/2000ml at 0.29ml/s with 22.8psi vacuum @07:01:57.99 Cartridge 14 sampling 1000/2000ml at 0.31ml/s with 15.1psi vacuum 2.2m deep (17.9psia) @07:02:58.11 Cartridge 14 sampling 1019/2000ml at 0.28ml/s with 19.4psi vacuum @07:03:57.98 Cartridge 14 sampling 1037/2000ml at 0.30ml/s with 28.5psi vacuum @07:04:58.17 Cartridge 14 sampling 1055/2000ml at 0.30ml/s with 21.1psi vacuum @07:05:57.99 Cartridge 14 sampling 1073/2000ml at 0.29ml/s with 25.9psi vacuum @07:06:58.26 Cartridge 14 sampling 1091/2000ml at 0.31ml/s with 27.0psi vacuum @07:07:57.99 Cartridge 14 sampling 1109/2000ml at 0.28ml/s with 25.2psi vacuum @07:08:58.22 Cartridge 14 sampling 1127/2000ml at 0.28ml/s with 25.6psi vacuum @07:09:57.99 Cartridge 14 sampling 1144/2000ml at 0.31ml/s with 23.7psi vacuum @07:10:58.18 Cartridge 14 sampling 1162/2000ml at 0.30ml/s with 23.0psi vacuum @07:11:58.11 Cartridge 14 sampling 1179/2000ml at 0.31ml/s with 26.5psi vacuum @07:12:58.32 Cartridge 14 sampling 1196/2000ml at 0.26ml/s with 24.9psi vacuum @07:13:57.99 Cartridge 14 sampling 1213/2000ml at 0.27ml/s with 25.6psi vacuum @07:14:58.21 Cartridge 14 sampling 1230/2000ml at 0.30ml/s with 24.8psi vacuum @07:15:57.99 Cartridge 14 sampling 1247/2000ml at 0.28ml/s with 27.2psi vacuum @07:16:58.11 Cartridge 14 sampling 1264/2000ml at 0.26ml/s with 26.5psi vacuum @07:17:57.98 Cartridge 14 sampling 1281/2000ml at 0.27ml/s with 20.4psi vacuum @07:18:58.33 Cartridge 14 sampling 1297/2000ml at 0.29ml/s with 26.6psi vacuum @07:19:57.99 Cartridge 14 sampling 1314/2000ml at 0.29ml/s with 24.7psi vacuum @07:20:58.09 Cartridge 14 sampling 1330/2000ml at 0.28ml/s with 27.3psi vacuum @07:21:57.99 Cartridge 14 sampling 1346/2000ml at 0.28ml/s with 27.3psi vacuum @07:22:58.30 Cartridge 14 sampling 1363/2000ml at 0.29ml/s with 19.7psi vacuum 1.7m deep (17.1psia) @07:23:58.07 Cartridge 14 sampling 1379/2000ml at 0.26ml/s with 27.5psi vacuum @07:24:57.99 Cartridge 14 sampling 1395/2000ml at 0.28ml/s with 28.1psi vacuum @07:25:58.46 Cartridge 14 sampling 1410/2000ml at 0.29ml/s with 22.8psi vacuum @07:26:58.02 Cartridge 14 sampling 1426/2000ml at 0.27ml/s with 17.2psi vacuum 1.6m deep (17.0psia) @07:27:58.03 Cartridge 14 sampling 1442/2000ml at 0.24ml/s with 16.4psi vacuum @07:28:58.25 Cartridge 14 sampling 1457/2000ml at 0.23ml/s with 24.9psi vacuum @07:29:57.98 Cartridge 14 sampling 1473/2000ml at 0.27ml/s with 23.9psi vacuum @07:30:58.06 Cartridge 14 sampling 1488/2000ml at 0.27ml/s with 26.8psi vacuum @07:31:58.10 Cartridge 14 sampling 1504/2000ml at 0.25ml/s with 10.7psi vacuum 7.2m deep (25.4psia) @07:32:58.18 Cartridge 14 sampling 1519/2000ml at 0.25ml/s with 29.3psi vacuum @07:33:58.07 Cartridge 14 sampling 1535/2000ml at 0.24ml/s with 29.0psi vacuum @07:34:57.99 Cartridge 14 sampling 1550/2000ml at 0.24ml/s with 13.1psi vacuum @07:35:57.98 Cartridge 14 sampling 1566/2000ml at 0.22ml/s with 18.3psi vacuum @07:36:58.22 Cartridge 14 sampling 1581/2000ml at 0.27ml/s with 22.1psi vacuum @07:37:57.99 Cartridge 14 sampling 1596/2000ml at 0.23ml/s with 28.9psi vacuum @07:38:58.23 Cartridge 14 sampling 1611/2000ml at 0.23ml/s with 26.3psi vacuum @07:39:58.07 Cartridge 14 sampling 1626/2000ml at 0.23ml/s with 21.4psi vacuum @07:40:57.99 Cartridge 14 sampling 1641/2000ml at 0.27ml/s with 25.7psi vacuum @07:41:58.15 Cartridge 14 sampling 1656/2000ml at 0.24ml/s with 10.4psi vacuum 4.6m deep (21.4psia) @07:42:57.97 Cartridge 14 sampling 1670/2000ml at 0.22ml/s with 25.8psi vacuum @07:43:58.24 Cartridge 14 sampling 1685/2000ml at 0.21ml/s with 17.8psi vacuum @07:44:57.98 Cartridge 14 sampling 1699/2000ml at 0.26ml/s with 15.3psi vacuum 2.5m deep (18.3psia) @07:45:58.33 Cartridge 14 sampling 1714/2000ml at 0.27ml/s with 22.7psi vacuum @07:46:58.21 Cartridge 14 sampling 1728/2000ml at 0.24ml/s with 28.1psi vacuum @07:47:57.99 Cartridge 14 sampling 1742/2000ml at 0.25ml/s with 27.8psi vacuum @07:48:58.14 Cartridge 14 sampling 1757/2000ml at 0.20ml/s with 25.7psi vacuum @07:49:57.98 Cartridge 14 sampling 1771/2000ml at 0.23ml/s with 28.0psi vacuum @07:50:58.12 Cartridge 14 sampling 1785/2000ml at 0.27ml/s with 26.3psi vacuum @07:51:58.35 Cartridge 14 sampling 1799/2000ml at 0.27ml/s with 28.0psi vacuum @07:52:57.98 Cartridge 14 sampling 1813/2000ml at 0.28ml/s with 20.1psi vacuum 2.1m deep (17.8psia) @07:53:58.15 Cartridge 14 sampling 1827/2000ml at 0.23ml/s with 16.9psi vacuum 4.9m deep (21.9psia) @07:54:57.98 Cartridge 14 sampling 1841/2000ml at 0.29ml/s with 27.0psi vacuum @07:55:58.19 Cartridge 14 sampling 1855/2000ml at 0.22ml/s with 27.5psi vacuum @07:56:57.97 Cartridge 14 sampling 1869/2000ml at 0.21ml/s with 24.9psi vacuum @07:57:58.17 Cartridge 14 sampling 1883/2000ml at 0.29ml/s with 21.1psi vacuum 3.5m deep (19.9psia) @07:58:57.98 Cartridge 14 sampling 1897/2000ml at 0.28ml/s with 20.9psi vacuum @07:59:58.14 Cartridge 14 sampling 1911/2000ml at 0.21ml/s with 24.0psi vacuum @08:00:57.98 Cartridge 14 sampling 1925/2000ml at 0.24ml/s with 23.6psi vacuum @08:01:58.15 Cartridge 14 sampling 1939/2000ml at 0.20ml/s with 26.4psi vacuum @08:02:57.98 Cartridge 14 sampling 1953/2000ml at 0.22ml/s with 27.9psi vacuum @08:03:58.17 Cartridge 14 sampling 1967/2000ml at 0.22ml/s with 30.0psi vacuum @08:04:58.08 Cartridge 14 sampling 1980/2000ml at 0.24ml/s with 21.2psi vacuum @08:05:58.20 Cartridge 14 sampling 1994/2000ml at 0.25ml/s with 22.8psi vacuum @08:06:24.37 Pressure.coast @08:06:24.74 Intake.close @08:06:24.90 Exhaust.close @08:06:27.07 Toroid.reconfigure :hold @08:06:27.21 Toroid.seek :bypass @08:06:28.89 Toroid.reconfigure :default @08:06:28.96 Toroid.seek :clear @08:06:30.57 Sampled 2000.0ml @08:06:31.07 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 14" @08:06:31.44 SamplePump.setPosition! 0ml @08:06:31.50 Plunge.seek :reagent @08:06:31.54 Exhaust.open @08:06:32.26 Intake.open @08:06:32.45 Sample loop closed. Ready to Deliver Reagents @08:06:32.47 Archive.seek :sealed @08:06:33.48 SamplePump.reconfigure :slow @08:06:33.66 SamplePump.seek -22ml @08:06:36.17 Archive.seek :evac @08:06:50.21 delay 1:00 @08:07:21.05 delay 1 second @08:07:22.08 Bypass.open @08:07:22.84 Exhaust.close @08:07:23.87 SamplePump.seek -20ml @08:07:28.49 delay 1 second @08:07:29.67 SamplePump.seek -28ml @08:07:47.07 delay 1 second @08:07:48.31 Intake.close @08:07:48.44 Bypass.close @08:07:50.24 Archive.seek :clear @08:08:07.49 RNAlater soak for 10:00 @08:16:31.81 upsync /usr/bin/dbclient: Connection to ESPkoa@espshore:2323 exited: Error reading: Connection reset by peer rsync: connection unexpectedly closed (0 bytes received so far) [sender] rsync error: error in rsync protocol data stream (code 12) at io.c(226) [sender=3.1.3] @08:18:07.53 Archive.seek :tVent @08:18:09.21 delay 1 second @08:18:10.23 Archive.seek :evac @08:18:26.26 delay 10 seconds @08:18:36.32 Sample preserved and evacuated @08:18:36.84
Carousel.seek 14 @08:18:37.23 Plunge.seek :locked @08:18:38.27 Archive.seek :clear @08:18:38.43 Air.seek :clear @08:18:38.59 Lysis1.seek :clear @08:18:55.54
Archive.coast @08:18:55.74 Air.coast @08:18:55.91 Lysis1.coast @08:18:56.02 Lysis2.coast @08:18:56.06 Plunge.coast @08:18:56.11 SamplePump.coast @08:18:56.15 Toroid.coast @08:18:56.20 Carousel.coast @08:18:56.39 Gate.power :motor,:OFF @08:18:56.40 Gate.power :core,:OFF @08:18:56.42 Gate.power :analytic1,:OFF @08:18:56.43 Gate.power :analytic2,:OFF @08:18:56.44 Gate.power :analytic3,:OFF @08:18:56.45 Gate.power :main,:OFF @08:18:56.47 Safely stopped and ready to power off @08:18:56.67 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 10:11" @08:18:56.70
delayUntil "10:11:49.70" @08:18:56.74 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @08:18:56.76 Retry #1 of 10 after 10:00 @10:11:49.82
Gate.power :main,:ON,limit to 5000mA @10:11:49.83 Gate.power :analytic1,:ON,limit to 5000mA @10:11:49.84 Gate.power :analytic2,:ON,limit to 5000mA @10:11:49.86 Gate.power :analytic3,:ON,limit to 5000mA @10:11:49.90 Powering core bus @10:11:52.48 Archive.configure :default @10:11:52.82 Air.configure :default @10:11:53.07 Lysis1.configure :default @10:11:54.47 Lysis2.configure :default @10:11:54.76 Plunge.configure :default @10:11:55.07 SamplePump.configure :default @10:11:55.32 ExtraPump.configure :default @10:11:55.75 Toroid.configure :default @10:11:56.03 Carousel.configure :default @10:11:56.17 Pressure.configure :default @10:11:56.22 All elf controllers running firmware version 4.61 @10:11:56.24 Gate.power :motor,:ON,limit to 5000mA @10:11:56.38 Archive.reconfigure :default @10:11:56.74 Archive.home.jog 0 @10:11:57.00 Toroid.reconfigure :default @10:11:57.03 Plunge.reconfigure :default @10:11:57.06 Lysis2.reconfigure :default @10:11:57.12 Lysis1.reconfigure :default @10:11:57.19 Air.reconfigure :default @10:11:57.42 Plunge.home.jog 0 @10:11:57.46 Toroid.home.jog 0 @10:11:57.48 Lysis2.home.jog 0 @10:11:57.50 Lysis1.home.jog 0 @10:11:57.56 Air.home.jog 0 @10:11:57.99 Archive.reconfigure :homing @10:11:58.24 Archive.jog 159 @10:11:58.35 Toroid.reconfigure :homing @10:11:58.38 Lysis2.reconfigure :homing @10:11:58.41 Air.reconfigure :homing @10:11:58.45 Plunge.reconfigure :homing @10:11:58.50 Lysis1.reconfigure :homing @10:11:58.75 Toroid.jog 617 @10:11:58.80 Lysis2.jog 1363 @10:11:58.86 Plunge.jog 859 @10:11:58.89 Lysis1.jog 1344 @10:11:58.92 Air.jog 1368 @10:11:59.19 Archive.setRawPosition! 50 @10:11:59.23 Archive.seek :home @10:11:59.50 Archive.reconfigure :default @10:11:59.97 Plunge.setRawPosition! 50 @10:12:00.02 Plunge.seek :home @10:12:00.15 Lysis2.setRawPosition! 50 @10:12:00.48 Lysis2.seek :home @10:12:00.87 Plunge.reconfigure :default @10:12:00.91 Lysis1.setRawPosition! 50 @10:12:00.99 Air.setRawPosition! 50 @10:12:01.31 Lysis1.seek :home @10:12:01.79 Air.seek :home @10:12:02.09 Toroid.setRawPosition! 0 @10:12:02.17 Lysis2.reconfigure :default @10:12:02.23 Air.reconfigure :default @10:12:02.30 Lysis1.reconfigure :default @10:12:02.61 Toroid.seek :home @10:12:03.14 Toroid.reconfigure :default @10:12:03.50
Toroid.seek :clear @10:12:05.16 Carousel.reconfigure :default @10:12:05.23 Carousel.home.jog 0 @10:12:05.88 Carousel.seek 14 @10:12:06.11 Selecting Cartridge 13 @10:12:06.32 Carousel.seek 13 @10:12:06.85 Lysis1.seek :locked @10:12:10.07 Plunge.seek :locked @10:12:10.22 Air.seek :locked @10:12:14.46
Mounted archive Cartridge @10:12:14.84 GoodNews.email "Filtering up to 2000.0ml",Subject: "Sampling archive_bac cartridge in slot 13" @10:12:15.43 Duration of filtering limited to 2:05:00 @10:12:15.72 Priming sample loop w/160ml, bypass w/10ml @10:12:15.74 SamplePump.setPosition! 0ml @10:12:15.77 Bypass.open @10:12:17.21 Intake.open @10:12:18.12 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @10:12:20.78 Vacuum tare is 16% of full scale @10:12:20.84 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @10:12:21.40 Pressure.reconfigure :default @10:12:21.64 Pressure.hold @10:12:22.35 SamplePump.reconfigure :prime @10:12:22.59 SamplePump.seek 10ml,1:21 @10:12:33.83 Pressure.coast @10:12:33.98 SamplePump.setPosition! 0ml @10:12:34.03 Exhaust.open @10:12:35.48 Bypass.close @10:12:37.03 Pressure.reconfigure :default @10:12:37.09 Pressure.hold @10:12:37.38 SamplePump.reconfigure :prime @10:12:37.44 SamplePump.seek 160ml,13:51 @10:15:28.96 Pressure.coast @10:15:29.02 Bypass.open @10:15:29.71 Intake.close @10:15:30.63 Toroid.reconfigure :hold @10:15:30.69 Toroid.seek :crack @10:15:32.27 Toroid.seek :cartridge @10:15:33.27 Toroid.reconfigure :default @10:15:33.46 Toroid.seek :clear @10:15:35.73 Pressurizing cartridge to 26psi above 14.4psia @10:15:35.78 SamplePump.setPosition! 0ml @10:15:35.85 Intake.open @10:15:36.75 Bypass.close @10:15:37.53 Exhaust.close @10:15:38.37 Pressure.reconfigure :default @10:15:38.43 Pressure.hold @10:15:38.63 SamplePump.reconfigure :charge @10:15:38.69 SamplePump.seek 2.3ml,42 seconds @10:15:48.85 Pressure.coast @10:15:48.89 Intake.close @10:15:49.74 Pressurized cartridge with 0.471ml to 26.1psi @10:15:49.75 Isolated @34.6psia (20.3psi) @10:15:49.76 Checking for cartridge leaks from 34.6psia @10:15:49.78 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @10:15:54.51 Cartridge pressure fell 3% to 33.7psia @10:15:54.65 Sampling 2000.0ml @10:15:54.68 SamplePump.setPosition! 0ml @10:15:54.80 Exhaust.open @10:15:55.55 Intake.open @10:15:57.13 Pressure.reconfigure :default @10:15:57.18 Pressure.hold @10:15:57.45 SamplePump.reconfigure :sample @10:15:57.52 SamplePump.seek 2000ml,2:47:12 @10:15:59.88 Cartridge 13 sampling 1/2000ml at 0.47ml/s with 24.3psi vacuum @10:17:00.15 Cartridge 13 sampling 28/2000ml at 0.44ml/s with 26.4psi vacuum @10:17:59.88 Cartridge 13 sampling 54/2000ml at 0.44ml/s with 24.8psi vacuum @10:18:59.92 Cartridge 13 sampling 81/2000ml at 0.44ml/s with 20.8psi vacuum @10:19:59.89 Cartridge 13 sampling 107/2000ml at 0.44ml/s with 20.4psi vacuum @10:21:00.04 Cartridge 13 sampling 133/2000ml at 0.44ml/s with 24.0psi vacuum @10:22:00.33 Cartridge 13 sampling 158/2000ml at 0.42ml/s with 22.9psi vacuum @10:22:59.88 Cartridge 13 sampling 183/2000ml at 0.42ml/s with 26.1psi vacuum @10:24:00.09 Cartridge 13 sampling 208/2000ml at 0.42ml/s with 24.4psi vacuum @10:24:59.88 Cartridge 13 sampling 233/2000ml at 0.42ml/s with 23.3psi vacuum @10:26:00.28 Cartridge 13 sampling 257/2000ml at 0.41ml/s with 25.9psi vacuum @10:26:59.87 Cartridge 13 sampling 282/2000ml at 0.41ml/s with 24.0psi vacuum @10:28:00.05 Cartridge 13 sampling 306/2000ml at 0.39ml/s with 26.3psi vacuum @10:28:59.87 Cartridge 13 sampling 330/2000ml at 0.40ml/s with 22.8psi vacuum @10:30:00.05 Cartridge 13 sampling 353/2000ml at 0.39ml/s with 26.0psi vacuum @10:30:59.88 Cartridge 13 sampling 377/2000ml at 0.38ml/s with 17.3psi vacuum @10:32:00.03 Cartridge 13 sampling 400/2000ml at 0.39ml/s with 26.6psi vacuum @10:32:59.88 Cartridge 13 sampling 423/2000ml at 0.36ml/s with 24.1psi vacuum @10:34:00.19 Cartridge 13 sampling 446/2000ml at 0.37ml/s with 25.9psi vacuum @10:34:59.88 Cartridge 13 sampling 468/2000ml at 0.38ml/s with 26.0psi vacuum @10:36:00.16 Cartridge 13 sampling 491/2000ml at 0.38ml/s with 26.7psi vacuum @10:36:59.87 Cartridge 13 sampling 514/2000ml at 0.38ml/s with 24.1psi vacuum @10:38:00.02 Cartridge 13 sampling 536/2000ml at 0.35ml/s with 25.7psi vacuum @10:38:59.88 Cartridge 13 sampling 558/2000ml at 0.36ml/s with 23.5psi vacuum @10:40:00.20 Cartridge 13 sampling 581/2000ml at 0.38ml/s with 18.8psi vacuum 3.8m deep (20.3psia) @10:40:59.88 Cartridge 13 sampling 603/2000ml at 0.36ml/s with 26.7psi vacuum @10:41:59.92 Cartridge 13 sampling 624/2000ml at 0.35ml/s with 21.6psi vacuum @10:43:00.11 Cartridge 13 sampling 646/2000ml at 0.36ml/s with 17.6psi vacuum 4.5m deep (21.4psia) @10:43:59.88 Cartridge 13 sampling 667/2000ml at 0.33ml/s with 26.3psi vacuum @10:45:00.13 Cartridge 13 sampling 689/2000ml at 0.36ml/s with 27.3psi vacuum @10:45:59.89 Cartridge 13 sampling 710/2000ml at 0.37ml/s with 22.6psi vacuum @10:47:00.09 Cartridge 13 sampling 731/2000ml at 0.35ml/s with 18.6psi vacuum @10:47:59.89 Cartridge 13 sampling 752/2000ml at 0.35ml/s with 26.8psi vacuum @10:49:00.07 Cartridge 13 sampling 773/2000ml at 0.33ml/s with 24.0psi vacuum @10:49:59.89 Cartridge 13 sampling 794/2000ml at 0.33ml/s with 17.7psi vacuum @10:51:00.22 Cartridge 13 sampling 814/2000ml at 0.35ml/s with 16.2psi vacuum 5.7m deep (23.1psia) @10:51:59.88 Cartridge 13 sampling 835/2000ml at 0.33ml/s with 28.4psi vacuum @10:53:00.03 Cartridge 13 sampling 855/2000ml at 0.34ml/s with 17.5psi vacuum 2.5m deep (18.4psia) @10:53:59.89 Cartridge 13 sampling 876/2000ml at 0.34ml/s with 26.2psi vacuum @10:55:00.04 Cartridge 13 sampling 896/2000ml at 0.34ml/s with 27.1psi vacuum @10:55:59.89 Cartridge 13 sampling 916/2000ml at 0.31ml/s with 14.7psi vacuum @10:57:00.34 Cartridge 13 sampling 936/2000ml at 0.31ml/s with 27.0psi vacuum @10:58:00.35 Cartridge 13 sampling 955/2000ml at 0.32ml/s with 28.3psi vacuum @10:59:00.04 Cartridge 13 sampling 975/2000ml at 0.30ml/s with 16.7psi vacuum 1.8m deep (17.4psia) @11:00:00.18 Cartridge 13 sampling 994/2000ml at 0.30ml/s with 26.3psi vacuum @11:00:59.88 Cartridge 13 sampling 1014/2000ml at 0.33ml/s with 15.5psi vacuum 2.8m deep (18.9psia) @11:02:00.22 Cartridge 13 sampling 1033/2000ml at 0.30ml/s with 12.5psi vacuum 8.4m deep (27.1psia) @11:02:59.88 Cartridge 13 sampling 1053/2000ml at 0.32ml/s with 28.0psi vacuum @11:04:00.10 Cartridge 13 sampling 1072/2000ml at 0.32ml/s with 27.8psi vacuum @11:04:59.88 Cartridge 13 sampling 1091/2000ml at 0.30ml/s with 26.7psi vacuum @11:06:00.14 Cartridge 13 sampling 1111/2000ml at 0.30ml/s with 26.2psi vacuum @11:06:59.88 Cartridge 13 sampling 1130/2000ml at 0.32ml/s with 26.8psi vacuum @11:08:00.05 Cartridge 13 sampling 1148/2000ml at 0.30ml/s with 11.2psi vacuum 7.5m deep (25.7psia) @11:08:59.88 Cartridge 13 sampling 1167/2000ml at 0.30ml/s with 28.0psi vacuum @11:10:00.07 Cartridge 13 sampling 1185/2000ml at 0.30ml/s with 27.2psi vacuum @11:10:59.88 Cartridge 13 sampling 1203/2000ml at 0.31ml/s with 16.6psi vacuum 3.2m deep (19.5psia) @11:12:00.05 Cartridge 13 sampling 1222/2000ml at 0.33ml/s with 25.0psi vacuum @11:12:59.88 Cartridge 13 sampling 1240/2000ml at 0.31ml/s with 16.3psi vacuum 2.6m deep (18.6psia) @11:14:00.41 Cartridge 13 sampling 1257/2000ml at 0.29ml/s with 18.3psi vacuum 4.0m deep (20.5psia) @11:15:00.07 Cartridge 13 sampling 1275/2000ml at 0.32ml/s with 23.4psi vacuum @11:16:00.25 Cartridge 13 sampling 1293/2000ml at 0.31ml/s with 24.7psi vacuum @11:17:00.08 Cartridge 13 sampling 1310/2000ml at 0.28ml/s with 10.7psi vacuum 8.9m deep (27.9psia) @11:17:59.96 Cartridge 13 sampling 1327/2000ml at 0.27ml/s with 29.2psi vacuum @11:19:00.15 Cartridge 13 sampling 1344/2000ml at 0.31ml/s with 21.0psi vacuum 2.3m deep (18.1psia) @11:19:59.89 Cartridge 13 sampling 1361/2000ml at 0.29ml/s with 21.3psi vacuum @11:21:00.04 Cartridge 13 sampling 1378/2000ml at 0.24ml/s with 24.7psi vacuum @11:21:59.88 Cartridge 13 sampling 1394/2000ml at 0.27ml/s with 16.3psi vacuum @11:23:00.10 Cartridge 13 sampling 1411/2000ml at 0.27ml/s with 28.4psi vacuum @11:23:59.88 Cartridge 13 sampling 1427/2000ml at 0.28ml/s with 19.3psi vacuum @11:25:00.04 Cartridge 13 sampling 1443/2000ml at 0.24ml/s with 26.3psi vacuum @11:25:59.89 Cartridge 13 sampling 1459/2000ml at 0.29ml/s with 16.9psi vacuum 2.4m deep (18.3psia) @11:27:00.08 Cartridge 13 sampling 1476/2000ml at 0.25ml/s with 26.7psi vacuum @11:27:59.89 Cartridge 13 sampling 1491/2000ml at 0.30ml/s with 21.6psi vacuum @11:29:00.02 Cartridge 13 sampling 1507/2000ml at 0.25ml/s with 26.6psi vacuum @11:29:59.89 Cartridge 13 sampling 1523/2000ml at 0.27ml/s with 29.5psi vacuum @11:30:59.99 Cartridge 13 sampling 1539/2000ml at 0.27ml/s with 27.4psi vacuum @11:31:59.89 Cartridge 13 sampling 1554/2000ml at 0.24ml/s with 26.7psi vacuum @11:33:00.04 Cartridge 13 sampling 1570/2000ml at 0.23ml/s with 26.4psi vacuum @11:34:00.22 Cartridge 13 sampling 1585/2000ml at 0.29ml/s with 26.2psi vacuum @11:34:59.98 Cartridge 13 sampling 1600/2000ml at 0.24ml/s with 27.8psi vacuum @11:36:00.17 Cartridge 13 sampling 1616/2000ml at 0.26ml/s with 26.6psi vacuum @11:36:59.89 Cartridge 13 sampling 1631/2000ml at 0.29ml/s with 26.9psi vacuum @11:38:00.14 Cartridge 13 sampling 1646/2000ml at 0.25ml/s with 12.5psi vacuum 7.4m deep (25.7psia) @11:38:59.89 Cartridge 13 sampling 1661/2000ml at 0.21ml/s with 24.9psi vacuum @11:39:59.97 Cartridge 13 sampling 1676/2000ml at 0.24ml/s with 30.4psi vacuum @11:40:59.89 Cartridge 13 sampling 1691/2000ml at 0.22ml/s with 29.3psi vacuum @11:42:00.03 Cartridge 13 sampling 1705/2000ml at 0.29ml/s with 24.0psi vacuum 1.7m deep (17.3psia) @11:42:59.89 Cartridge 13 sampling 1720/2000ml at 0.27ml/s with 21.8psi vacuum @11:44:00.08 Cartridge 13 sampling 1735/2000ml at 0.26ml/s with 13.5psi vacuum 7.7m deep (26.1psia) @11:44:59.89 Cartridge 13 sampling 1750/2000ml at 0.26ml/s with 15.1psi vacuum 1.9m deep (17.6psia) @11:46:00.22 Cartridge 13 sampling 1764/2000ml at 0.25ml/s with 27.9psi vacuum @11:46:59.88 Cartridge 13 sampling 1778/2000ml at 0.23ml/s with 29.8psi vacuum @11:48:00.05 Cartridge 13 sampling 1793/2000ml at 0.26ml/s with 23.0psi vacuum @11:48:59.89 Cartridge 13 sampling 1807/2000ml at 0.26ml/s with 26.0psi vacuum @11:50:00.00 Cartridge 13 sampling 1821/2000ml at 0.20ml/s with 28.2psi vacuum @11:50:59.88 Cartridge 13 sampling 1835/2000ml at 0.26ml/s with 22.6psi vacuum @11:52:00.20 Cartridge 13 sampling 1850/2000ml at 0.26ml/s with 16.4psi vacuum 5.7m deep (23.1psia) @11:53:00.26 Cartridge 13 sampling 1864/2000ml at 0.27ml/s with 18.9psi vacuum @11:53:59.88 Cartridge 13 sampling 1877/2000ml at 0.22ml/s with 8.9psi vacuum 1.8m deep (17.3psia) @11:55:00.16 Cartridge 13 sampling 1891/2000ml at 0.21ml/s with 26.4psi vacuum @11:55:59.88 Cartridge 13 sampling 1905/2000ml at 0.29ml/s with 22.5psi vacuum @11:57:00.01 Cartridge 13 sampling 1920/2000ml at 0.20ml/s with 21.6psi vacuum @11:57:59.89 Cartridge 13 sampling 1934/2000ml at 0.22ml/s with 9.5psi vacuum @11:59:00.11 Cartridge 13 sampling 1948/2000ml at 0.25ml/s with 24.8psi vacuum @11:59:59.89 Cartridge 13 sampling 1962/2000ml at 0.25ml/s with 28.3psi vacuum @12:01:00.07 Cartridge 13 sampling 1976/2000ml at 0.20ml/s with 22.4psi vacuum @12:01:59.89 Cartridge 13 sampling 1990/2000ml at 0.27ml/s with 16.1psi vacuum 2.5m deep (18.4psia) @12:02:43.04 Pressure.coast @12:02:43.45 Intake.close @12:02:43.58 Exhaust.close @12:02:43.76 Toroid.reconfigure :hold @12:02:43.82 Toroid.seek :bypass @12:02:45.44 Toroid.reconfigure :default @12:02:45.50 Toroid.seek :clear @12:02:47.12 Sampled 2000.0ml @12:02:47.60 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 13" @12:02:48.34 SamplePump.setPosition! 0ml @12:02:48.52 Plunge.seek :reagent @12:02:48.58 Exhaust.open @12:02:49.48 Intake.open @12:02:49.66 Sample loop closed. Ready to Deliver Reagents @12:02:49.69 Archive.seek :sealed @12:02:50.67 SamplePump.reconfigure :slow @12:02:50.88 SamplePump.seek -22ml @12:02:53.54 Archive.seek :evac @12:03:07.57 delay 1:00 @12:03:38.21 delay 1 second @12:03:39.23 Bypass.open @12:03:40.01 Exhaust.close @12:03:40.76 SamplePump.seek -20ml @12:03:45.35 delay 1 second @12:03:46.58 SamplePump.seek -28ml @12:04:04.02 delay 1 second @12:04:05.19 Intake.close @12:04:05.33 Bypass.close @12:04:07.60 Archive.seek :clear @12:04:24.77 RNAlater soak for 10:00 @12:14:24.82 Archive.seek :tVent @12:14:26.47 delay 1 second @12:14:27.52 Archive.seek :evac @12:14:43.56 delay 10 seconds @12:14:53.60 Sample preserved and evacuated @12:14:53.83
Carousel.seek 13 @12:14:54.04 Plunge.seek :locked @12:14:55.37 Archive.seek :clear @12:14:55.79 Air.seek :clear @12:14:56.17 Lysis1.seek :clear @12:15:12.66
Archive.coast @12:15:12.70 Air.coast @12:15:12.74 Lysis1.coast @12:15:12.79 Lysis2.coast @12:15:12.83 Plunge.coast @12:15:12.87 SamplePump.coast @12:15:12.92 Toroid.coast @12:15:12.96 Carousel.coast @12:15:13.15 Gate.power :motor,:OFF @12:15:13.17 Gate.power :core,:OFF @12:15:13.25 Gate.power :analytic1,:OFF @12:15:13.28 Gate.power :analytic2,:OFF @12:15:13.36 Gate.power :analytic3,:OFF @12:15:13.42 Gate.power :main,:OFF @12:15:13.44 Safely stopped and ready to power off @12:15:13.91 delayUntil "14:11:49.70" @12:15:14.01 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 14:11" @14:11:49.82
Gate.power :main,:ON,limit to 5000mA @14:11:49.84 Gate.power :analytic1,:ON,limit to 5000mA @14:11:49.85 Gate.power :analytic2,:ON,limit to 5000mA @14:11:49.86 Gate.power :analytic3,:ON,limit to 5000mA @14:11:49.90 Powering core bus @14:11:52.72 Archive.configure :default @14:11:53.02 Air.configure :default @14:11:53.31 Lysis1.configure :default @14:11:53.63 Lysis2.configure :default @14:11:54.03 Plunge.configure :default @14:11:54.37 SamplePump.configure :default @14:11:54.64 ExtraPump.configure :default @14:11:55.14 Toroid.configure :default @14:11:55.46 Carousel.configure :default @14:11:55.85 Pressure.configure :default @14:11:55.95 All elf controllers running firmware version 4.61 @14:11:55.97 Gate.power :motor,:ON,limit to 5000mA @14:11:56.83 Archive.reconfigure :default @14:11:57.09 Archive.home.jog 0 @14:11:57.31 Toroid.reconfigure :default @14:11:57.36 Plunge.reconfigure :default @14:11:57.40 Lysis2.reconfigure :default @14:11:57.46 Air.reconfigure :default @14:11:57.50 Lysis1.reconfigure :default @14:11:57.72 Plunge.home.jog 0 @14:11:57.77 Lysis2.home.jog 0 @14:11:57.80 Toroid.home.jog 0 @14:11:57.83 Lysis1.home.jog 0 @14:11:57.87 Air.home.jog 0 @14:11:58.38 Archive.reconfigure :homing @14:11:59.06 Toroid.reconfigure :homing @14:11:59.22 Archive.jog 158 @14:11:59.26 Plunge.reconfigure :homing @14:11:59.29 Air.reconfigure :homing @14:11:59.34 Lysis1.reconfigure :homing @14:11:59.37 Lysis2.reconfigure :homing @14:11:59.62 Toroid.jog 614 @14:11:59.65 Plunge.jog 860 @14:11:59.71 Air.jog 1373 @14:11:59.73 Lysis1.jog 1347 @14:11:59.79 Lysis2.jog 1360 @14:12:00.10 Archive.setRawPosition! 50 @14:12:00.14 Archive.seek :home @14:12:00.42 Archive.reconfigure :default @14:12:00.80 Plunge.setRawPosition! 50 @14:12:00.85 Plunge.seek :home @14:12:01.18 Air.setRawPosition! 50 @14:12:01.31 Air.seek :home @14:12:01.68 Lysis2.setRawPosition! 50 @14:12:01.76 Lysis1.setRawPosition! 50 @14:12:01.82 Plunge.reconfigure :default @14:12:02.10 Lysis2.seek :home @14:12:02.42 Lysis1.seek :home @14:12:02.87 Air.reconfigure :default @14:12:02.93 Toroid.setRawPosition! 0 @14:12:03.17 Toroid.seek :home @14:12:03.58 Toroid.reconfigure :default @14:12:03.60 Lysis2.reconfigure :default @14:12:03.65 Lysis1.reconfigure :default @14:12:04.53
Toroid.seek :clear @14:12:06.26 Carousel.reconfigure :default @14:12:06.39 Carousel.home.jog 0 @14:12:07.05 Carousel.seek 13 @14:12:07.21 Selecting Cartridge 12 @14:12:07.43 Carousel.seek 12 @14:12:07.94 Lysis1.seek :locked @14:12:11.17 Plunge.seek :locked @14:12:11.60 Air.seek :locked @14:12:16.00
Mounted archive Cartridge @14:12:16.33 GoodNews.email "Filtering up to 2000.0ml",Subject: "Sampling archive_bac cartridge in slot 12" @14:12:16.65 Duration of filtering limited to 2:05:00 @14:12:16.95 Priming sample loop w/160ml, bypass w/10ml @14:12:16.97 SamplePump.setPosition! 0ml @14:12:17.00 Bypass.open @14:12:18.55 Intake.open @14:12:19.47 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @14:12:22.07 Vacuum tare is 16% of full scale @14:12:22.13 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @14:12:22.79 Pressure.reconfigure :default @14:12:23.05 Pressure.hold @14:12:23.74 SamplePump.reconfigure :prime @14:12:23.88 SamplePump.seek 10ml,1:21 @14:12:35.09 Pressure.coast @14:12:35.22 SamplePump.setPosition! 0ml @14:12:35.27 Exhaust.open @14:12:36.66 Bypass.close @14:12:38.22 Pressure.reconfigure :default @14:12:38.28 Pressure.hold @14:12:38.55 SamplePump.reconfigure :prime @14:12:38.61 SamplePump.seek 160ml,13:51 @14:15:30.15 Pressure.coast @14:15:30.21 Bypass.open @14:15:30.90 Intake.close @14:15:31.80 Toroid.reconfigure :hold @14:15:31.86 Toroid.seek :crack @14:15:33.44 Toroid.seek :cartridge @14:15:34.40 Toroid.reconfigure :default @14:15:34.63 Toroid.seek :clear @14:15:36.85 Pressurizing cartridge to 26psi above 14.2psia @14:15:36.87 SamplePump.setPosition! 0ml @14:15:36.92 Intake.open @14:15:37.79 Bypass.close @14:15:38.53 Exhaust.close @14:15:40.47 Pressure.reconfigure :default @14:15:40.53 Pressure.hold @14:15:40.75 SamplePump.reconfigure :charge @14:15:40.81 SamplePump.seek 2.3ml,42 seconds @14:15:51.36 Pressure.coast @14:15:51.41 Intake.close @14:15:52.25 Pressurized cartridge with 0.509ml to 26.1psi @14:15:52.26 Isolated @33.9psia (19.7psi) @14:15:52.27 Checking for cartridge leaks from 33.9psia @14:15:52.29 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @14:15:56.95 Cartridge pressure fell 3% to 32.8psia @14:15:57.19 Sampling 2000.0ml @14:15:57.22 SamplePump.setPosition! 0ml @14:15:57.26 Exhaust.open @14:15:57.99 Intake.open @14:15:58.94 Pressure.reconfigure :default @14:15:59.00 Pressure.hold @14:15:59.28 SamplePump.reconfigure :sample @14:15:59.35 SamplePump.seek 2000ml,2:47:12 @14:16:02.03 Cartridge 12 sampling 1/2000ml at 0.48ml/s with 26.0psi vacuum @14:17:02.09 Cartridge 12 sampling 28/2000ml at 0.46ml/s with 20.8psi vacuum 1.7m deep (17.2psia) @14:18:02.01 Cartridge 12 sampling 56/2000ml at 0.45ml/s with 20.0psi vacuum @14:19:02.39 Cartridge 12 sampling 83/2000ml at 0.45ml/s with 26.1psi vacuum @14:20:02.25 Cartridge 12 sampling 109/2000ml at 0.51ml/s with 24.3psi vacuum @14:21:02.10 Cartridge 12 sampling 142/2000ml at 0.52ml/s with 21.0psi vacuum @14:21:20.35 SamplePump.coast @14:21:20.41 Intake.close @14:21:20.49 Exhaust.close @14:21:20.68 Pressure.coast @14:21:21.01 Slide::Error in FILTERING -- Storage Carousel (SC) Not yet Homed! @14:21:21.02 Waiting to Resume from Checkpoint " during Sample Pump (SP) move" @14:21:21.04 BadNews.email "Storage Carousel (SC) Not yet Homed!",Subject: "Slide::Error in FILTERING" @14:21:21.30 Exhaust.close @14:21:21.32 Intake.close @14:21:21.47 Exhaust.close @14:21:21.51 Intake.close @16:28:09.60 Session Begun @16:28:17.76 -> showlog 100 @16:28:50.97 -> can @16:29:17.29 upsync /usr/bin/dbclient: Connection to ESPkoa@espshore:2323 exited: Error reading: Connection reset by peer rsync: [sender] write error: Broken pipe (32) rsync error: error in socket IO (code 10) at io.c(820) [sender=3.1.3] @16:31:22.54 -> exit @16:31:22.56 Session Ended @20:29:12.23 Session Begun @20:29:19.25 -> slots @20:29:27.89 -> Cartridge.reserved @20:29:36.91 -> Cartridge.dry @20:30:27.21 -> Cartridge.unlock @20:30:27.33 Slide::Error in brent -- Storage Carousel (SC) Not yet Homed! @20:30:45.02 -> SC.home.to 12 @20:30:45.05 Carousel.reconfigure :default @20:30:45.15 Carousel.home.seek 12 @20:30:54.65 -> Cartridge.unlock @20:30:54.77 Toroid.reconfigure :hold @20:30:55.39 Toroid.seek :bypass @20:30:57.04 Toroid.reconfigure :default @20:30:57.13 Toroid.seek :clear @20:30:59.32 Air.seek :clear @20:31:03.17 Lysis1.seek :clear @20:31:03.35 Plunge.seek :clear @20:31:41.41 -> ExternalValve @20:32:19.97 -> Cmd.stop @20:32:20.17 Carousel.seek 12 @20:32:20.51 Lysis1.seek :locked @20:32:23.88 Plunge.seek :locked @20:32:25.41 Lysis1.seek :clear @20:32:28.80 Archive.coast @20:32:28.88 Air.coast @20:32:28.95 Lysis1.coast @20:32:29.03 Lysis2.coast @20:32:29.11 Plunge.coast @20:32:29.18 Toroid.coast @20:32:29.26 Carousel.coast @20:32:29.48 Gate.power :motor,:OFF @20:32:29.50 Gate.power :core,:OFF @20:32:29.53 Gate.power :analytic1,:OFF @20:32:29.56 Gate.power :analytic2,:OFF @20:32:29.59 Gate.power :analytic3,:OFF @20:32:29.63 Gate.power :main,:OFF @20:32:29.67 Safely stopped and ready to power off @20:32:31.49 -> exit @20:32:31.54 Session Ended Log::Locked in trapHandler -- Another process is already writing to /var/log/esp/real.log @20:34:26.42 Session Begun @20:34:33.67 Session Ended @20:35:06.20 Session Begun @20:35:11.98 -> MainThread.exit @20:35:12.25 Session Ended @20:35:27.34PST07-Mar-21
>BREAK< ESP3 Gateway v1.11 for core bus @20:35:27.36 >BREAK< ESP3 Gateway v1.11 for cartridge bus @20:35:30.98 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @20:35:32.04 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @20:35:40.04 Polling Can every 2:00 @20:35:40.10 Loading /home/esp/esp2/bin/processAll mission @20:35:42.00 Sampling maximum of 2000.0ml per cartridge @20:35:42.14 GoodNews.email "Cartridge group: dry\nSpares: {spare: 4}\nSchedule: [-4:00:00]\n",Subject: "Begun processAll --interval=4:: --mail --checkpoint 2000" @20:35:42.40 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @20:35:42.55 Retry #1 of 10 after 10:00 @20:35:42.85
Gate.power :main,:ON,limit to 5000mA @20:35:42.88 Gate.power :analytic1,:ON,limit to 5000mA @20:35:42.92 Gate.power :analytic2,:ON,limit to 5000mA @20:35:42.96 Gate.power :analytic3,:ON,limit to 5000mA @20:35:43.10 Powering core bus @20:35:45.94 Archive.configure :default @20:35:46.65 Air.configure :default @20:35:47.04 Lysis1.configure :default @20:35:47.27 Lysis2.configure :default @20:35:47.58 Plunge.configure :default @20:35:48.17 SamplePump.configure :default @20:35:48.60 ExtraPump.configure :default @20:35:48.74 Toroid.configure :default @20:35:48.92 Carousel.configure :default @20:35:49.19 Pressure.configure :default @20:35:49.29 All elf controllers running firmware version 4.61 @20:35:49.34 Gate.power :motor,:ON,limit to 5000mA @20:35:50.40 Plunge.reconfigure :default @20:35:50.48 Lysis2.reconfigure :default @20:35:50.58 Lysis1.reconfigure :default @20:35:50.64 Air.reconfigure :default @20:35:50.76 Toroid.reconfigure :default @20:35:50.88 Archive.reconfigure :default @20:35:51.34 Plunge.home.jog 0 @20:35:51.39 Lysis1.home.jog 0 @20:35:51.43 Lysis2.home.jog 0 @20:35:51.47 Air.home.jog 0 @20:35:51.51 Toroid.home.jog 0 @20:35:51.56 Archive.home.jog 0 @20:35:52.43 Plunge.reconfigure :homing @20:35:52.48 Lysis1.reconfigure :homing @20:35:52.52 Air.reconfigure :homing @20:35:52.65 Lysis2.reconfigure :homing @20:35:53.22 Plunge.jog 856 @20:35:53.44 Lysis1.jog 1354 @20:35:53.63 Air.jog 1369 @20:35:53.76 Lysis2.jog 1361 @20:35:54.42 Toroid.reconfigure :homing @20:35:54.45 Archive.reconfigure :homing @20:35:54.71 Toroid.jog 623 @20:35:54.96 Archive.jog 1368 @20:35:55.15 Plunge.setRawPosition! 50 @20:35:55.27 Plunge.seek :home @20:35:55.60 Lysis1.setRawPosition! 50 @20:35:55.75 Air.setRawPosition! 50 @20:35:55.77 Lysis2.setRawPosition! 50 @20:35:56.03 Lysis1.seek :home @20:35:56.17 Air.seek :home @20:35:56.24 Lysis2.seek :home @20:35:56.88 Lysis2.reconfigure :default @20:35:57.09 Lysis1.reconfigure :default @20:35:57.12 Archive.setRawPosition! 50 @20:35:57.16 Plunge.reconfigure :default @20:35:57.19 Air.reconfigure :default @20:35:57.71 Archive.seek :home @20:35:57.95 Toroid.setRawPosition! 0 @20:35:58.15 Toroid.seek :home @20:35:58.60 Archive.reconfigure :default @20:35:58.62 Toroid.reconfigure :default @20:35:59.04
Toroid.seek :clear @20:36:00.67 Carousel.reconfigure :default @20:36:00.72 Carousel.home.jog 0 @20:36:01.23 Carousel.seek 12 @20:36:01.39 Selecting Cartridge 11 @20:36:01.61 Carousel.seek 11 @20:36:02.14 Lysis1.seek :locked @20:36:05.56 Plunge.seek :locked @20:36:05.93 Air.seek :locked @20:36:10.23
Mounted archive Cartridge @20:36:10.60 Duration of filtering limited to 2:05:00 @20:36:10.91 Priming sample loop w/160ml, bypass w/10ml @20:36:10.93 SamplePump.setPosition! 0ml @20:36:10.96 Bypass.open @20:36:12.52 Intake.open @20:36:13.58 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @20:36:16.38 Vacuum tare is 17% of full scale @20:36:16.48 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @20:36:16.97 Pressure.reconfigure :default @20:36:17.12 Pressure.hold @20:36:17.57 SamplePump.reconfigure :prime @20:36:17.89 SamplePump.seek 10ml,1:21 @20:36:29.05 Pressure.coast @20:36:29.10 SamplePump.setPosition! 0ml @20:36:29.15 Exhaust.open @20:36:30.58 Bypass.close @20:36:32.58 Pressure.reconfigure :default @20:36:32.74 Pressure.hold @20:36:33.46 SamplePump.reconfigure :prime @20:36:33.59 SamplePump.seek 160ml,13:51 @20:39:25.00 Pressure.coast @20:39:25.06 Bypass.open @20:39:25.72 Intake.close @20:39:26.60 Toroid.reconfigure :hold @20:39:26.66 Toroid.seek :crack @20:39:28.24 Toroid.seek :cartridge @20:39:29.48 Toroid.reconfigure :default @20:39:29.67 Toroid.seek :clear @20:39:31.92 Pressurizing cartridge to 26psi above 13.4psia @20:39:32.00 SamplePump.setPosition! 0ml @20:39:32.12 Intake.open @20:39:32.94 Bypass.close @20:39:33.67 Exhaust.close @20:39:34.52 Pressure.reconfigure :default @20:39:34.57 Pressure.hold @20:39:34.78 SamplePump.reconfigure :charge @20:39:34.85 SamplePump.seek 2.3ml,42 seconds @20:39:46.96 Pressure.coast @20:39:47.01 Intake.close @20:39:48.65 Pressurized cartridge with 0.672ml to 26.2psi @20:39:48.69 Isolated @35.7psia (22.3psi) @20:39:48.81 Checking for cartridge leaks from 35.7psia @20:39:48.89 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @20:39:59.15 Cartridge pressure fell 9% to 32.4psia @20:39:59.43 Sampling 2000.0ml @20:39:59.46 SamplePump.setPosition! 0ml @20:39:59.61 Exhaust.open @20:40:00.38 Intake.open @20:40:01.94 Pressure.reconfigure :default @20:40:02.02 Pressure.hold @20:40:02.71 SamplePump.reconfigure :sample @20:40:02.84 SamplePump.seek 2000ml,2:47:12 @20:40:04.76 Cartridge 11 sampling 1/2000ml at 0.58ml/s with 28.9psi vacuum @20:41:04.76 Cartridge 11 sampling 29/2000ml at 0.46ml/s with 23.7psi vacuum @20:42:05.16 Cartridge 11 sampling 56/2000ml at 0.44ml/s with 22.6psi vacuum @20:43:05.15 Cartridge 11 sampling 83/2000ml at 0.45ml/s with 24.6psi vacuum @20:44:04.76 Cartridge 11 sampling 110/2000ml at 0.44ml/s with 21.6psi vacuum @20:45:05.45 Cartridge 11 sampling 136/2000ml at 0.44ml/s with 25.4psi vacuum @20:45:50.66 GoodNews.email "Filtering up to 2000.0ml",Subject: "Sampling archive_bac cartridge in slot 11" @20:46:04.95 Cartridge 11 sampling 162/2000ml at 0.44ml/s with 24.2psi vacuum @20:46:14.70 upsync @20:47:04.89 Cartridge 11 sampling 188/2000ml at 0.43ml/s with 25.0psi vacuum @20:48:04.92 Cartridge 11 sampling 214/2000ml at 0.43ml/s with 24.0psi vacuum @20:49:04.76 Cartridge 11 sampling 239/2000ml at 0.42ml/s with 24.5psi vacuum @20:50:04.76 Cartridge 11 sampling 265/2000ml at 0.42ml/s with 24.4psi vacuum @20:51:04.95 Cartridge 11 sampling 289/2000ml at 0.41ml/s with 21.3psi vacuum @20:52:04.88 Cartridge 11 sampling 314/2000ml at 0.41ml/s with 21.7psi vacuum @20:53:04.76 Cartridge 11 sampling 339/2000ml at 0.41ml/s with 22.9psi vacuum @20:53:09.87 Session Begun @20:53:10.41 Session Ended @20:53:56.86 Session Begun @20:54:04.46 -> SC.config.pwm @20:54:04.56 NoMethodError in brent -- undefined method `pwm' for Carousel[:default]:I2C::Servo::Configuration @20:54:05.00 Cartridge 11 sampling 363/2000ml at 0.41ml/s with 21.3psi vacuum @20:54:15.20 -> SC.config.maxPWM @20:54:22.33 Session Ended @20:55:04.99 Cartridge 11 sampling 388/2000ml at 0.40ml/s with 26.4psi vacuum @20:56:04.76 Cartridge 11 sampling 412/2000ml at 0.41ml/s with 24.2psi vacuum @20:57:04.76 Cartridge 11 sampling 436/2000ml at 0.41ml/s with 23.4psi vacuum @20:58:07.33 Cartridge 11 sampling 460/2000ml at 0.39ml/s with 21.0psi vacuum @20:59:04.76 Cartridge 11 sampling 484/2000ml at 0.40ml/s with 25.4psi vacuum @21:00:04.90 Cartridge 11 sampling 508/2000ml at 0.40ml/s with 23.5psi vacuum @21:01:04.76 Cartridge 11 sampling 532/2000ml at 0.38ml/s with 20.8psi vacuum @21:02:04.92 Cartridge 11 sampling 555/2000ml at 0.40ml/s with 22.3psi vacuum @21:03:04.92 Cartridge 11 sampling 579/2000ml at 0.40ml/s with 26.0psi vacuum @21:04:04.76 Cartridge 11 sampling 602/2000ml at 0.39ml/s with 18.2psi vacuum @21:05:05.13 Cartridge 11 sampling 625/2000ml at 0.38ml/s with 17.8psi vacuum @21:06:05.14 Cartridge 11 sampling 648/2000ml at 0.39ml/s with 22.9psi vacuum @21:07:04.84 Cartridge 11 sampling 671/2000ml at 0.38ml/s with 21.0psi vacuum @21:08:04.76 Cartridge 11 sampling 694/2000ml at 0.38ml/s with 21.6psi vacuum @21:09:04.92 Cartridge 11 sampling 717/2000ml at 0.37ml/s with 26.6psi vacuum @21:10:04.76 Cartridge 11 sampling 740/2000ml at 0.38ml/s with 25.2psi vacuum @21:11:05.36 Cartridge 11 sampling 762/2000ml at 0.39ml/s with 26.4psi vacuum @21:12:05.14 Cartridge 11 sampling 785/2000ml at 0.34ml/s with 23.3psi vacuum @21:13:04.80 Cartridge 11 sampling 807/2000ml at 0.39ml/s with 24.6psi vacuum @21:14:05.06 Cartridge 11 sampling 829/2000ml at 0.39ml/s with 22.3psi vacuum @21:15:04.76 Cartridge 11 sampling 851/2000ml at 0.33ml/s with 26.1psi vacuum @21:16:05.21 Cartridge 11 sampling 873/2000ml at 0.36ml/s with 26.8psi vacuum @21:17:05.00 Cartridge 11 sampling 894/2000ml at 0.36ml/s with 22.9psi vacuum @21:18:04.75 Cartridge 11 sampling 916/2000ml at 0.37ml/s with 24.8psi vacuum @21:18:38.38 Session Begun @21:18:42.21 -> slots @21:18:49.14 -> SP @21:18:53.01 -> SP @21:19:04.97 Cartridge 11 sampling 937/2000ml at 0.38ml/s with 20.6psi vacuum 2.5m deep (18.4psia) @21:19:42.99 -> Cartridge.release 10..7 @21:20:04.76 Cartridge 11 sampling 958/2000ml at 0.38ml/s with 23.6psi vacuum @21:21:04.88 Cartridge 11 sampling 979/2000ml at 0.35ml/s with 26.7psi vacuum @21:21:28.71 -> exit @21:21:28.73 Session Ended @21:22:04.76 Cartridge 11 sampling 1000/2000ml at 0.36ml/s with 22.4psi vacuum @21:23:04.92 Cartridge 11 sampling 1021/2000ml at 0.38ml/s with 25.4psi vacuum @21:24:04.76 Cartridge 11 sampling 1041/2000ml at 0.32ml/s with 26.8psi vacuum @21:25:04.80 Cartridge 11 sampling 1062/2000ml at 0.37ml/s with 20.6psi vacuum 1.8m deep (17.3psia) @21:26:04.76 Cartridge 11 sampling 1082/2000ml at 0.32ml/s with 26.5psi vacuum @21:27:05.08 Cartridge 11 sampling 1102/2000ml at 0.34ml/s with 20.9psi vacuum @21:28:05.06 Cartridge 11 sampling 1123/2000ml at 0.32ml/s with 23.2psi vacuum @21:29:04.92 Cartridge 11 sampling 1142/2000ml at 0.33ml/s with 23.8psi vacuum @21:30:05.05 Cartridge 11 sampling 1162/2000ml at 0.31ml/s with 20.8psi vacuum @21:31:04.92 Cartridge 11 sampling 1181/2000ml at 0.36ml/s with 25.1psi vacuum @21:32:04.76 Cartridge 11 sampling 1201/2000ml at 0.35ml/s with 21.9psi vacuum @21:33:04.82 Cartridge 11 sampling 1220/2000ml at 0.31ml/s with 27.0psi vacuum @21:34:04.75 Cartridge 11 sampling 1239/2000ml at 0.35ml/s with 25.7psi vacuum @21:35:05.14 Cartridge 11 sampling 1257/2000ml at 0.31ml/s with 26.3psi vacuum @21:36:04.76 Cartridge 11 sampling 1276/2000ml at 0.29ml/s with 27.0psi vacuum @21:37:05.00 Cartridge 11 sampling 1294/2000ml at 0.28ml/s with 18.9psi vacuum @21:38:04.76 Cartridge 11 sampling 1313/2000ml at 0.30ml/s with 26.8psi vacuum @21:39:04.88 Cartridge 11 sampling 1331/2000ml at 0.34ml/s with 24.6psi vacuum @21:40:04.75 Cartridge 11 sampling 1349/2000ml at 0.31ml/s with 13.9psi vacuum 3.0m deep (19.2psia) @21:41:04.98 Cartridge 11 sampling 1367/2000ml at 0.30ml/s with 24.4psi vacuum @21:42:04.76 Cartridge 11 sampling 1384/2000ml at 0.28ml/s with 19.2psi vacuum @21:43:04.92 Cartridge 11 sampling 1402/2000ml at 0.29ml/s with 26.9psi vacuum @21:44:04.75 Cartridge 11 sampling 1420/2000ml at 0.30ml/s with 26.2psi vacuum @21:45:04.96 Cartridge 11 sampling 1438/2000ml at 0.28ml/s with 26.4psi vacuum @21:46:04.75 Cartridge 11 sampling 1455/2000ml at 0.28ml/s with 26.9psi vacuum @21:47:05.02 Cartridge 11 sampling 1473/2000ml at 0.25ml/s with 22.9psi vacuum @21:48:04.74 Cartridge 11 sampling 1490/2000ml at 0.28ml/s with 8.4psi vacuum 2.6m deep (18.5psia) @21:49:04.99 Cartridge 11 sampling 1507/2000ml at 0.25ml/s with 25.8psi vacuum @21:50:04.75 Cartridge 11 sampling 1525/2000ml at 0.29ml/s with 20.9psi vacuum @21:51:04.91 Cartridge 11 sampling 1542/2000ml at 0.29ml/s with 25.9psi vacuum @21:52:04.75 Cartridge 11 sampling 1559/2000ml at 0.33ml/s with 22.6psi vacuum @21:53:04.76 Cartridge 11 sampling 1576/2000ml at 0.28ml/s with 23.9psi vacuum @21:54:05.07 Cartridge 11 sampling 1592/2000ml at 0.28ml/s with 25.5psi vacuum @21:55:04.75 Cartridge 11 sampling 1609/2000ml at 0.32ml/s with 24.8psi vacuum @21:56:05.13 Cartridge 11 sampling 1626/2000ml at 0.26ml/s with 19.5psi vacuum 2.0m deep (17.7psia) @21:57:04.88 Cartridge 11 sampling 1642/2000ml at 0.27ml/s with 26.9psi vacuum @21:58:04.76 Cartridge 11 sampling 1659/2000ml at 0.27ml/s with 29.4psi vacuum @21:59:05.31 Cartridge 11 sampling 1675/2000ml at 0.28ml/s with 20.7psi vacuum @22:00:05.02 Cartridge 11 sampling 1691/2000ml at 0.25ml/s with 26.1psi vacuum @22:01:04.78 Cartridge 11 sampling 1707/2000ml at 0.27ml/s with 26.7psi vacuum @22:02:04.96 Cartridge 11 sampling 1723/2000ml at 0.23ml/s with 25.3psi vacuum @22:03:04.76 Cartridge 11 sampling 1739/2000ml at 0.26ml/s with 9.7psi vacuum 2.8m deep (18.8psia) @22:04:05.00 Cartridge 11 sampling 1755/2000ml at 0.31ml/s with 24.0psi vacuum @22:05:04.76 Cartridge 11 sampling 1771/2000ml at 0.23ml/s with 22.3psi vacuum @22:06:04.98 Cartridge 11 sampling 1786/2000ml at 0.25ml/s with 27.0psi vacuum @22:07:04.76 Cartridge 11 sampling 1802/2000ml at 0.28ml/s with 26.8psi vacuum @22:08:04.95 Cartridge 11 sampling 1817/2000ml at 0.24ml/s with 27.0psi vacuum @22:09:04.76 Cartridge 11 sampling 1833/2000ml at 0.22ml/s with 23.1psi vacuum @22:10:04.92 Cartridge 11 sampling 1848/2000ml at 0.22ml/s with 25.2psi vacuum @22:11:04.75 Cartridge 11 sampling 1863/2000ml at 0.26ml/s with 24.4psi vacuum @22:12:04.99 Cartridge 11 sampling 1878/2000ml at 0.24ml/s with 14.7psi vacuum 6.2m deep (23.8psia) @22:13:04.76 Cartridge 11 sampling 1893/2000ml at 0.23ml/s with 18.6psi vacuum @22:14:04.94 Cartridge 11 sampling 1907/2000ml at 0.27ml/s with 21.9psi vacuum @22:15:04.76 Cartridge 11 sampling 1922/2000ml at 0.22ml/s with 19.1psi vacuum @22:16:04.90 Cartridge 11 sampling 1937/2000ml at 0.22ml/s with 10.5psi vacuum 1.6m deep (17.1psia) @22:17:04.76 Cartridge 11 sampling 1951/2000ml at 0.24ml/s with 9.6psi vacuum 1.9m deep (17.4psia) @22:18:04.99 Cartridge 11 sampling 1966/2000ml at 0.22ml/s with 26.8psi vacuum @22:19:04.84 Cartridge 11 sampling 1980/2000ml at 0.22ml/s with 28.1psi vacuum @22:20:04.76 Cartridge 11 sampling 1995/2000ml at 0.23ml/s with 22.1psi vacuum @22:20:27.64 Pressure.coast @22:20:27.84 Intake.close @22:20:28.10 Exhaust.close @22:20:28.57 Toroid.reconfigure :hold @22:20:28.63 Toroid.seek :bypass @22:20:30.25 Toroid.reconfigure :default @22:20:30.31 Toroid.seek :clear @22:20:31.93 Sampled 2000.0ml @22:20:32.38 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 11" @22:20:32.88 SamplePump.setPosition! 0ml @22:20:33.04 Plunge.seek :reagent @22:20:33.15 Exhaust.open @22:20:34.23 Intake.open @22:20:34.26 Sample loop closed. Ready to Deliver Reagents @22:20:34.50 Archive.seek :sealed @22:20:36.34 SamplePump.reconfigure :slow @22:20:36.47 SamplePump.seek -22ml @22:20:39.61 Archive.seek :evac @22:20:53.62 delay 1:00 @22:21:10.85 upsync @22:21:23.82 delay 1 second @22:21:24.84 Bypass.open @22:21:25.60 Exhaust.close @22:21:26.37 SamplePump.seek -20ml @22:21:30.95 delay 1 second @22:21:32.14 SamplePump.seek -28ml @22:21:49.60 delay 1 second @22:21:50.74 Intake.close @22:21:50.95 Bypass.close @22:21:53.66 Archive.seek :clear @22:22:10.86 RNAlater soak for 10:00 @22:32:10.92 Archive.seek :tVent @22:32:12.70 delay 1 second @22:32:13.81 Archive.seek :evac @22:32:29.88 delay 10 seconds @22:32:39.96 Sample preserved and evacuated @22:32:40.83
Carousel.seek 11 @22:32:41.45 Plunge.seek :locked @22:32:42.76 Archive.seek :clear @22:32:43.35 Air.seek :clear @22:32:44.02 Lysis1.seek :clear @22:33:00.28
Archive.coast @22:33:00.38 Air.coast @22:33:00.64 Lysis1.coast @22:33:00.74 Lysis2.coast @22:33:00.88 Plunge.coast @22:33:01.07 SamplePump.coast @22:33:01.22 Toroid.coast @22:33:01.37 Carousel.coast @22:33:01.73 Gate.power :motor,:OFF @22:33:01.74 Gate.power :core,:OFF @22:33:01.75 Gate.power :analytic1,:OFF @22:33:01.77 Gate.power :analytic2,:OFF @22:33:01.87 Gate.power :analytic3,:OFF @22:33:01.94 Gate.power :main,:OFF @22:33:01.99 Safely stopped and ready to power off @22:33:02.56 delayUntil "00:35:40.05PST08-Mar-21" @22:33:02.70 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:35PST08-Mar-21" @22:33:02.86 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @22:33:02.92 Retry #1 of 10 after 10:00 @22:43:11.28 EOFError in Qmail -- end of file reached @22:43:11.34 Retry #2 of 10 after 10:00 @22:53:11.50 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @22:53:11.54 Retry #3 of 10 after 10:00 @23:03:40.37 upsync @00:35:40.30PST08-Mar-21
Gate.power :main,:ON,limit to 5000mA @00:35:40.35 Gate.power :analytic1,:ON,limit to 5000mA @00:35:40.36 Gate.power :analytic2,:ON,limit to 5000mA @00:35:40.38 Gate.power :analytic3,:ON,limit to 5000mA @00:35:40.41 Powering core bus @00:35:42.91 Archive.configure :default @00:35:43.10 Air.configure :default @00:35:43.22 Lysis1.configure :default @00:35:43.33 Lysis2.configure :default @00:35:43.47 Plunge.configure :default @00:35:43.61 SamplePump.configure :default @00:35:43.75 ExtraPump.configure :default @00:35:43.90 Toroid.configure :default @00:35:44.04 Carousel.configure :default @00:35:44.18 Pressure.configure :default @00:35:44.23 All elf controllers running firmware version 4.61 @00:35:44.25 Gate.power :motor,:ON,limit to 5000mA @00:35:44.53 Archive.reconfigure :default @00:35:44.80 Toroid.reconfigure :default @00:35:44.89 Archive.home.jog 0 @00:35:44.92 Lysis2.reconfigure :default @00:35:44.95 Lysis1.reconfigure :default @00:35:45.00 Air.reconfigure :default @00:35:45.04 Plunge.reconfigure :default @00:35:45.28 Toroid.home.jog 0 @00:35:45.36 Lysis2.home.jog 0 @00:35:45.44 Plunge.home.jog 0 @00:35:45.48 Lysis1.home.jog 0 @00:35:45.53 Air.home.jog 0 @00:35:47.66 Lysis1.reconfigure :homing @00:35:47.84 Air.reconfigure :homing @00:35:47.91 Toroid.reconfigure :homing @00:35:49.65 Plunge.reconfigure :homing @00:35:49.69 Archive.reconfigure :homing @00:35:49.89 Lysis2.reconfigure :homing @00:35:50.37 Air.jog 1368 @00:35:50.48 Lysis1.jog 1342 @00:35:50.53 Toroid.jog 636 @00:35:50.55 Plunge.jog 858 @00:35:50.59 Archive.jog 159 @00:35:50.62 Lysis2.jog 1352 @00:35:51.51 Archive.setRawPosition! 50 @00:35:51.61 Archive.seek :home @00:35:51.86 Air.setRawPosition! 50 @00:35:51.91 Air.seek :home @00:35:52.03 Plunge.setRawPosition! 50 @00:35:52.07 Lysis1.setRawPosition! 50 @00:35:52.13 Archive.reconfigure :default @00:35:52.20 Plunge.seek :home @00:35:52.27 Lysis1.seek :home @00:35:52.47 Lysis2.setRawPosition! 50 @00:35:52.49 Air.reconfigure :default @00:35:52.56 Lysis2.seek :home @00:35:52.76 Plunge.reconfigure :default @00:35:52.78 Lysis1.reconfigure :default @00:35:52.99 Toroid.setRawPosition! 0 @00:35:53.01 Lysis2.reconfigure :default @00:35:53.10 Toroid.seek :home @00:35:53.23 Toroid.reconfigure :default @00:35:53.48
Toroid.seek :clear @00:35:55.10 Carousel.reconfigure :default @00:35:55.15 Carousel.home.jog 0 @00:35:55.67 Carousel.seek 11 @00:35:55.81 Selecting Cartridge 10 @00:35:56.10 Carousel.seek 10 @00:35:57.36 Lysis1.seek :locked @00:36:00.93 Plunge.seek :locked @00:36:01.26 Air.seek :locked @00:36:05.53
Mounted archive Cartridge @00:36:05.86 GoodNews.email "Filtering up to 2000.0ml",Subject: "Sampling archive_bac cartridge in slot 10" @00:36:07.27 Duration of filtering limited to 2:05:00 @00:36:08.11 Priming sample loop w/160ml, bypass w/10ml @00:36:08.13 SamplePump.setPosition! 0ml @00:36:08.18 Bypass.open @00:36:09.83 Intake.open @00:36:11.23 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:36:13.84 Vacuum tare is 16% of full scale @00:36:13.87 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @00:36:14.08 Pressure.reconfigure :default @00:36:14.14 Pressure.hold @00:36:14.35 SamplePump.reconfigure :prime @00:36:14.42 SamplePump.seek 10ml,1:21 @00:36:25.51 Pressure.coast @00:36:25.56 SamplePump.setPosition! 0ml @00:36:25.61 Exhaust.open @00:36:26.78 upsync @00:36:27.28 Bypass.close @00:36:29.28 Pressure.reconfigure :default @00:36:29.40 Pressure.hold @00:36:30.00 SamplePump.reconfigure :prime @00:36:30.06 SamplePump.seek 160ml,13:51 @00:36:33.15 SamplePump.coast @00:36:33.20 Intake.close @00:36:33.25 Exhaust.close @00:36:34.10 Pressure.coast @00:36:34.24 Cartridge::Sampler::Leaky in FILTERING -- Serious Leak or Stall -- 0.93ml/s made only 8.8psi vacuum @00:36:34.26 Retry #1 of 2 @00:36:34.28 Cmd::SpareCartridge in FILTERING -- 3 spares remaining @00:36:34.31 Selecting Cartridge 9 @00:36:34.68 Air.seek :clear @00:36:38.65 Lysis1.seek :clear @00:36:39.09 Plunge.seek :clear @00:36:41.96 Carousel.seek 9 @00:36:42.48 Lysis1.seek :locked @00:36:45.95 Plunge.seek :locked @00:36:46.41 Air.seek :locked @00:36:50.64 Mounted archive Cartridge @00:36:50.69 Duration of filtering limited to 2:05:00 @00:36:50.97 Priming sample loop w/160ml, bypass w/10ml @00:36:50.99 SamplePump.setPosition! 0ml @00:36:51.02 Bypass.open @00:36:52.49 Intake.open @00:36:53.37 Sea 1.7m deep (17.3psia) @00:36:53.40 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:36:57.02 Vacuum tare is 19% of full scale @00:36:57.06 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @00:36:57.63 Pressure.reconfigure :default @00:36:57.76 Pressure.hold @00:36:58.31 SamplePump.reconfigure :prime @00:36:58.54 SamplePump.seek 10ml,1:21 @00:37:01.62 SamplePump.coast @00:37:01.67 Intake.close @00:37:01.75 Bypass.close @00:37:02.57 Pressure.coast @00:37:02.72 Cartridge::Sampler::Leaky in FILTERING -- Serious Leak or Stall -- 0.93ml/s made only 1.4psi vacuum @00:37:02.74 Retry #2 of 2 @00:37:02.76 Cmd::SpareCartridge in FILTERING -- 2 spares remaining @00:37:02.78 Selecting Cartridge 8 @00:37:03.16 Air.seek :clear @00:37:07.14 Lysis1.seek :clear @00:37:07.47 Plunge.seek :clear @00:37:10.43 Carousel.seek 8 @00:37:10.96 Lysis1.seek :locked @00:37:14.58 Plunge.seek :locked @00:37:15.08 Air.seek :locked @00:37:19.32 Mounted archive Cartridge @00:37:19.37 Duration of filtering limited to 2:05:00 @00:37:19.64 Priming sample loop w/160ml, bypass w/10ml @00:37:19.66 SamplePump.setPosition! 0ml @00:37:19.69 Bypass.open @00:37:21.17 Intake.open @00:37:22.05 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:37:24.70 Vacuum tare is 19% of full scale @00:37:24.76 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -19psi @00:37:25.25 Pressure.reconfigure :default @00:37:25.38 Pressure.hold @00:37:25.85 SamplePump.reconfigure :prime @00:37:26.00 SamplePump.seek 10ml,1:21 @00:37:37.19 Pressure.coast @00:37:37.38 SamplePump.setPosition! 0ml @00:37:37.67 Exhaust.open @00:37:39.11 Bypass.close @00:37:40.68 Pressure.reconfigure :default @00:37:40.74 Pressure.hold @00:37:41.02 SamplePump.reconfigure :prime @00:37:41.08 SamplePump.seek 160ml,13:51 @00:37:44.28 SamplePump.coast @00:37:44.34 Intake.close @00:37:44.55 Exhaust.close @00:37:45.39 Pressure.coast @00:37:45.70 Cartridge::Sampler::Leaky in FILTERING -- Serious Leak or Stall -- 0.93ml/s made only 5.7psi vacuum @00:37:45.72 Waiting to Resume from Checkpoint "Sample Pump (SP) pressureOutOfBounds at 2.47ml" @00:37:45.77 BadNews.email "Serious Leak or Stall -- 0.93ml/s made only 5.7psi vacuum",Subject: "Cartridge::Sampler::Leaky in FILTERING" @00:38:16.21 upsync @01:16:48.74 Session Begun @01:16:54.70 Session Ended @01:18:48.64 Session Begun @01:19:02.86 -> Cartridge.sampler.prime @01:19:03.05 Priming sample loop w/160ml, bypass w/10ml @01:19:03.07 SamplePump.setPosition! 0ml @01:19:03.16 Bypass.open @01:19:04.63 Intake.open @01:19:05.46 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @01:19:08.37 Vacuum tare is 19% of full scale @01:19:08.46 Cartridge::Sampler::Warning in brent -- Vacuum sensor saturates below -19psi @01:19:09.07 Pressure.reconfigure :default @01:19:09.33 Pressure.hold @01:19:10.00 SamplePump.reconfigure :prime @01:19:10.16 SamplePump.seek 10ml,1:21 @01:19:13.31 SamplePump.coast @01:19:13.39 Intake.close @01:19:13.50 Bypass.close @01:19:14.24 Pressure.coast @01:19:14.32 Cartridge::Sampler::Leaky in brent -- Serious Leak or Stall -- 0.93ml/s made only 0.2psi vacuum @01:20:01.71 -> Bypass.open; Intake.open; SP.pump 100 @01:20:01.73 Bypass.open @01:20:02.47 Intake.open @01:20:03.77 SamplePump.reconfigure :default @01:20:03.99 SamplePump.seek 102.48ml @01:22:08.82 -> Cartridge.sampler.prime @01:22:08.93 Priming sample loop w/160ml, bypass w/10ml @01:22:08.95 SamplePump.setPosition! 0ml @01:22:09.88 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @01:22:13.24 Vacuum tare is 17% of full scale @01:22:13.27 Cartridge::Sampler::Warning in brent -- Vacuum sensor saturates below -17psi @01:22:13.86 Pressure.reconfigure :default @01:22:14.09 Pressure.hold @01:22:14.67 SamplePump.reconfigure :prime @01:22:14.98 SamplePump.seek 10ml,1:21 @01:22:26.25 Pressure.coast @01:22:26.48 SamplePump.setPosition! 0ml @01:22:26.58 Exhaust.open @01:22:28.06 Bypass.close @01:22:28.18 Errno::EPIPE in brent's_reader -- Broken pipe @01:22:28.26 Session Ended @01:23:06.86 Session Begun @01:23:14.66 -> showlog 20 @01:23:30.54 -> Cartridge.sampler.prime @01:23:30.79 Priming sample loop w/160ml, bypass w/10ml @01:23:30.85 SamplePump.setPosition! 0ml @01:23:31.01 Exhaust.close @01:23:31.82 Bypass.open @01:23:33.48 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @01:23:36.07 Vacuum tare is 18% of full scale @01:23:36.10 Cartridge::Sampler::Warning in brent -- Vacuum sensor saturates below -17psi @01:23:36.36 Pressure.reconfigure :default @01:23:36.44 Pressure.hold @01:23:36.67 SamplePump.reconfigure :prime @01:23:36.77 SamplePump.seek 10ml,1:21 @01:23:47.90 Pressure.coast @01:23:48.09 SamplePump.setPosition! 0ml @01:23:48.39 Exhaust.open @01:23:49.96 Bypass.close @01:23:51.79 Pressure.reconfigure :default @01:23:51.88 Pressure.hold @01:23:52.18 SamplePump.reconfigure :prime @01:23:52.27 SamplePump.seek 160ml,13:51 @01:23:55.38 SamplePump.coast @01:23:55.46 Intake.close @01:23:55.57 Exhaust.close @01:23:56.52 Pressure.coast @01:23:56.76 Cartridge::Sampler::Leaky in brent -- Serious Leak or Stall -- 0.93ml/s made only 6.9psi vacuum @01:24:21.71 -> SC @01:24:37.86 Session Ended @01:25:25.76 Session Begun @01:25:57.23 -> Bypass.open;Intake.open; SP.pump 100, :fast @01:25:57.32 Bypass.open @01:25:58.12 Intake.open @01:25:59.15 SamplePump.reconfigure :fast @01:25:59.33 SamplePump.seek 102.48ml @01:27:23.19 Session Begun @01:27:26.75 -> Sampler.psi @01:27:35.48 -> SP.rate @01:27:41.70 -> Bypass.open;Intake.open; SP.pump 100, :fast @01:27:42.04 SamplePump.seek 202.48ml @01:27:45.66 -> SP.rate @01:27:49.71 -> Sampler.psi @01:28:03.02 SamplePump.hold @01:28:03.81 UserAbort in brent -- Sample Pump (SP) seek to 202.48ml Aborted! @01:28:14.99 -> Bypass.close @01:28:15.06 Bypass.close @01:28:19.31 -> Exhaust.open @01:28:19.37 Exhaust.open @01:28:25.16 -> Intake.open @01:29:09.03 -> P.pump 50 @01:29:09.09 NameError in brent -- uninitialized constant P @01:29:13.81 Session Ended @01:29:14.14 Session Ended @01:29:14.25 Errno::EPIPE in brent's_reader -- Broken pipe @01:29:57.06 Session Begun @01:30:03.65 -> ExternalValve @01:30:10.67 -> SP.pump 50 @01:30:10.94 SamplePump.reconfigure :default @01:30:11.13 SamplePump.seek 176.37ml @01:30:23.00 Session Begun @01:30:26.50 -> Sampler.psi @01:31:06.86 -> Sampler.psi @01:31:27.60 -> Cartridge.sampler.prime @01:31:27.94 Priming sample loop w/160ml, bypass w/10ml @01:31:28.02 SamplePump.setPosition! 0ml @01:31:28.24 Exhaust.close @01:31:29.10 Bypass.open @01:31:30.77 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @01:31:33.35 Vacuum tare is 17% of full scale @01:31:33.38 Cartridge::Sampler::Warning in brent -- Vacuum sensor saturates below -17psi @01:31:33.67 Pressure.reconfigure :default @01:31:33.77 Pressure.hold @01:31:34.06 SamplePump.reconfigure :prime @01:31:34.16 SamplePump.seek 10ml,1:21 @01:31:38.04 -> Sampler.psi @01:31:43.35 -> Sampler.psi @01:31:45.37 Pressure.coast @01:31:45.71 SamplePump.setPosition! 0ml @01:31:45.86 Exhaust.open @01:31:47.46 Bypass.close @01:31:47.74 -> Sampler.psi @01:31:49.55 Pressure.reconfigure :default @01:31:49.65 Pressure.hold @01:31:49.96 SamplePump.reconfigure :prime @01:31:50.07 SamplePump.seek 160ml,13:51 @01:31:50.56 -> Sampler.psi @01:31:53.20 SamplePump.coast @01:31:53.30 Intake.close @01:31:54.20 -> Sampler.psi @01:31:54.31 Exhaust.close @01:31:55.77 Pressure.coast @01:31:55.92 Cartridge::Sampler::Leaky in brent -- Serious Leak or Stall -- 0.93ml/s made only 8.0psi vacuum @01:32:27.66 -> SC.config.minPWM @01:33:52.07 -> Cmd.loadCartridge @01:33:52.12 Selecting Cartridge 7 @01:33:52.58 Air.seek :clear @01:33:56.57 Lysis1.seek :clear @01:33:57.16 Plunge.seek :clear @01:33:57.44 -> hidelog @01:33:59.96 Carousel.seek 7 @01:34:00.50 Lysis1.seek :locked @01:34:04.14 Plunge.seek :locked @01:34:04.57 Air.seek :locked @01:34:09.01 Mounted archive Cartridge @01:34:20.33 -> Cartridge.sampler.prime @01:34:20.44 Priming sample loop w/160ml, bypass w/10ml @01:34:20.46 SamplePump.setPosition! 0ml @01:34:20.54 Bypass.open @01:34:21.10 Session Ended @01:34:21.19 Errno::EPIPE in brent's_reader -- Broken pipe @01:34:21.23 Session Ended @01:35:16.24 Session Begun @01:35:21.62 Session Begun @01:35:26.22 -> Cartridge.sampler.prime @01:35:26.42 Priming sample loop w/160ml, bypass w/10ml @01:35:26.49 SamplePump.setPosition! 0ml @01:35:26.69 Bypass.open @01:35:27.66 Intake.open @01:35:28.53 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @01:35:31.11 Vacuum tare is 19% of full scale @01:35:31.14 Cartridge::Sampler::Warning in brent -- Vacuum sensor saturates below -18psi @01:35:31.43 Pressure.reconfigure :default @01:35:31.54 Pressure.hold @01:35:32.00 SamplePump.reconfigure :prime @01:35:32.24 SamplePump.seek 10ml,1:21 @01:35:35.51 SamplePump.coast @01:35:35.77 Intake.close @01:35:35.91 Bypass.close @01:35:37.08 Pressure.coast @01:35:37.23 Cartridge::Sampler::Leaky in brent -- Serious Leak or Stall -- 0.93ml/s made only 1.3psi vacuum @01:35:57.81 -> Cmd.stop @01:35:58.06 Carousel.seek 7 @01:35:58.31 Plunge.seek :locked @01:35:58.59 Air.seek :clear @01:35:58.79 Lysis1.seek :clear @01:36:02.35 Archive.coast @01:36:02.58 Air.coast @01:36:02.83 Lysis1.coast @01:36:03.09 Lysis2.coast @01:36:03.28 Plunge.coast @01:36:03.52 Toroid.coast @01:36:03.69 Carousel.coast @01:36:04.43 Gate.power :motor,:OFF @01:36:04.49 Gate.power :core,:OFF @01:36:04.55 Gate.power :analytic1,:OFF @01:36:04.60 Gate.power :analytic2,:OFF @01:36:04.73 Gate.power :analytic3,:OFF @01:36:04.81 Gate.power :main,:OFF @01:36:04.84 Safely stopped and ready to power off @01:36:38.09 Session Ended @01:36:38.20 Session Ended @01:36:38.46 Errno::EPIPE in brent's_reader -- Broken pipe @01:43:18.23 Session Begun @01:43:20.09 -> slots @01:45:16.54 Session Ended @11:33:35.87PDT25-May-21
>BREAK< ESP3 Gateway v1.11 for core bus @11:33:35.90 >BREAK< ESP3 Gateway v1.11 for cartridge bus @11:33:42.98 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @11:33:43.46 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @11:33:47.91 Polling Can every 2:00 @11:33:48.15 Waiting for espclients... @11:33:49.02 Session Begun @11:34:44.45 -> ESP.ready! @11:34:44.86 Gate.power :main,:ON,limit to 5000mA @11:34:44.88 Gate.power :analytic1,:ON,limit to 5000mA @11:34:44.96 Gate.power :analytic2,:ON,limit to 5000mA @11:34:45.05 Gate.power :analytic3,:ON,limit to 5000mA @11:34:45.20 Powering core bus @11:34:47.84 Archive.configure :default @11:34:48.12 Air.configure :default @11:34:48.69 Lysis1.configure :default @11:34:48.82 Lysis2.configure :default @11:34:48.99 Plunge.configure :default @11:34:49.16 SamplePump.configure :default @11:34:49.33 ExtraPump.configure :default @11:34:49.51 Toroid.configure :default @11:34:49.68 Carousel.configure :default @11:34:49.85 Pressure.configure :default @11:34:49.93 All elf controllers running firmware version 4.61 @11:34:49.97 Gate.power :motor,:ON,limit to 5000mA @11:34:50.18 Archive.reconfigure :default @11:34:50.53 Archive.home.jog 0 @11:34:50.78 Toroid.reconfigure :default @11:34:50.83 Lysis2.reconfigure :default @11:34:50.87 Lysis1.reconfigure :default @11:34:50.91 Air.reconfigure :default @11:34:50.97 Plunge.reconfigure :default @11:34:51.22 Toroid.home.jog 0 @11:34:51.26 Lysis1.home.jog 0 @11:34:51.33 Lysis2.home.jog 0 @11:34:51.40 Plunge.home.jog 0 @11:34:51.43 Air.home.jog 0 @11:34:51.92 Archive.reconfigure :homing @11:34:52.18 Archive.jog 1358 @11:34:52.24 Toroid.reconfigure :homing @11:34:52.28 Lysis2.reconfigure :homing @11:34:52.32 Air.reconfigure :homing @11:34:52.37 Plunge.reconfigure :homing @11:34:52.45 Lysis1.reconfigure :homing @11:34:52.71 Toroid.jog 645 @11:34:52.77 Lysis2.jog 1346 @11:34:52.80 Air.jog 1358 @11:34:52.85 Lysis1.jog 1360 @11:34:52.89 Plunge.jog 306 @11:34:53.67 Archive.setRawPosition! 50 @11:34:54.00 Archive.seek :home @11:34:54.29 Plunge.setRawPosition! 50 @11:34:54.55 Lysis2.setRawPosition! 50 @11:34:54.76 Plunge.seek :home @11:34:54.89 Lysis2.seek :home @11:34:55.45 Toroid.setRawPosition! 0 @11:34:55.47 Lysis1.setRawPosition! 50 @11:34:55.58 Air.setRawPosition! 50 @11:34:55.67 Archive.reconfigure :default @11:34:56.13 Toroid.seek :home @11:34:56.28 Air.seek :home @11:34:56.33 Lysis1.seek :home @11:34:57.18 Toroid.reconfigure :default @11:34:57.29 Lysis2.reconfigure :default @11:34:57.43 Air.reconfigure :default @11:34:57.48 Plunge.reconfigure :default @11:34:57.68 Lysis1.reconfigure :default @11:34:58.69 Toroid.seek :clear @11:35:00.39 Carousel.reconfigure :default @11:35:00.47 Carousel.home.jog 0 @11:35:01.00 Carousel.seek 7 @11:37:52.08 -> exit @11:37:52.13 Session Ended @11:39:54.36 Session Begun @11:40:05.92 -> ESP @11:41:26.48 -> exit @11:41:26.50 Session Ended @11:44:11.70 Session Begun @11:46:04.68 -> yes=:no @11:53:54.85 -> exit @11:53:54.87 Session Ended @15:46:36.83 Session Begun @15:49:57.97 -> def shoot *args; Cmd.loadCartridge *args; Cmd.filter; Cmd.process;Cmd.stop; end @15:50:45.92 -> exit @15:50:45.99 Session Ended @17:42:17.93 Session Begun @17:44:22.62 -> slots @17:44:59.14 -> slot @17:45:48.29 -> BACsampler.goalVolume @17:45:58.72 -> BACsampler.goalVolume=2000 @17:46:31.94 -> ESP.ready! @17:46:32.67 Carousel.seek 7 @17:51:05.20 -> BACsampler.prime @17:51:05.31 Priming sample loop w/160ml, bypass w/10ml @17:51:05.33 SamplePump.setPosition! 0ml @17:51:05.41 Bypass.open @17:51:06.95 Intake.open @17:51:07.86 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @17:51:10.70 Vacuum tare is 17% of full scale @17:51:10.76 Cartridge::Sampler::Warning in preston -- Vacuum sensor saturates below -17psi @17:51:11.43 Pressure.reconfigure :default @17:51:11.65 Pressure.hold @17:51:12.27 SamplePump.reconfigure :prime @17:51:12.52 SamplePump.seek 10ml,1:21 @17:51:23.73 Pressure.coast @17:51:24.09 SamplePump.setPosition! 0ml @17:51:24.17 Exhaust.open @17:51:25.61 Bypass.close @17:51:27.18 Pressure.reconfigure :default @17:51:27.26 Pressure.hold @17:51:27.57 SamplePump.reconfigure :prime @17:51:27.66 SamplePump.seek 160ml,13:51 @17:54:18.99 Pressure.coast @17:54:32.27 -> showlog -500 @17:55:15.47 -> Intake.close @17:55:15.50 Intake.close @17:55:19.65 -> Exhaust.close @17:55:19.70 Exhaust.close @17:55:29.44 -> shoot @17:55:31.13 Carousel.seek 7 @17:55:31.68 Selecting Cartridge 7 @17:55:32.30 Carousel.seek 7 @17:55:32.70 Lysis1.seek :locked @17:55:36.12 Plunge.seek :locked @17:55:36.30 Air.seek :locked @17:55:40.77 Mounted archive Cartridge @17:55:41.48 Duration of filtering limited to 4:05:00 @17:55:42.04 Bypass.open @17:55:43.16 Toroid.reconfigure :hold @17:55:43.25 Toroid.seek :crack @17:55:44.86 Toroid.seek :cartridge @17:55:45.82 Toroid.reconfigure :default @17:55:45.90 Toroid.seek :clear @17:55:47.62 Exhaust.open @17:55:49.06 Pressurizing cartridge to 26psi above 14.6psia @17:55:49.13 SamplePump.setPosition! 0ml @17:55:49.32 Intake.open @17:55:50.32 Bypass.close @17:55:51.12 Exhaust.close @17:55:52.14 Pressure.reconfigure :default @17:55:52.32 Pressure.hold @17:55:52.90 SamplePump.reconfigure :charge @17:55:53.20 SamplePump.seek 2.3ml,42 seconds @17:56:08.57 Pressure.coast @17:56:08.67 Intake.close @17:56:09.84 Pressurized cartridge with 1.068ml to 26.4psi @17:56:09.86 Isolated @34.9psia (20.3psi) @17:56:09.88 Checking for cartridge leaks from 34.9psia @17:56:10.00 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @17:56:14.68 Cartridge pressure fell 2% to 34.1psia @17:56:14.78 Sampling 2000.0ml @17:56:14.80 SamplePump.setPosition! 0ml @17:56:14.89 Exhaust.open @17:56:15.60 Intake.open @17:56:17.22 Pressure.reconfigure :default @17:56:17.44 Pressure.hold @17:56:18.11 SamplePump.reconfigure :sample @17:56:18.33 SamplePump.seek 2000ml,2:47:12 @17:56:20.13 Cartridge 7 sampling 0/2000ml at 0.63ml/s with 16.8psi vacuum @17:57:19.95 Cartridge 7 sampling 30/2000ml at 0.49ml/s with 23.9psi vacuum @17:58:20.12 Cartridge 7 sampling 59/2000ml at 0.48ml/s with 24.5psi vacuum @17:59:19.95 Cartridge 7 sampling 89/2000ml at 0.50ml/s with 23.9psi vacuum @18:00:20.40 Cartridge 7 sampling 119/2000ml at 0.50ml/s with 25.3psi vacuum @18:01:19.95 Cartridge 7 sampling 149/2000ml at 0.50ml/s with 22.7psi vacuum @18:02:19.99 Cartridge 7 sampling 179/2000ml at 0.48ml/s with 23.1psi vacuum @18:03:20.36 Cartridge 7 sampling 208/2000ml at 0.49ml/s with 24.0psi vacuum @18:04:19.95 Cartridge 7 sampling 237/2000ml at 0.49ml/s with 23.4psi vacuum @18:05:20.27 Cartridge 7 sampling 266/2000ml at 0.48ml/s with 25.4psi vacuum @18:06:19.95 Cartridge 7 sampling 295/2000ml at 0.47ml/s with 25.8psi vacuum @18:07:20.12 Cartridge 7 sampling 324/2000ml at 0.49ml/s with 22.0psi vacuum @18:08:19.95 Cartridge 7 sampling 353/2000ml at 0.49ml/s with 26.6psi vacuum @18:09:20.27 Cartridge 7 sampling 383/2000ml at 0.48ml/s with 22.6psi vacuum @18:10:19.95 Cartridge 7 sampling 412/2000ml at 0.49ml/s with 25.9psi vacuum @18:11:20.17 Cartridge 7 sampling 441/2000ml at 0.48ml/s with 21.8psi vacuum @18:12:19.95 Cartridge 7 sampling 469/2000ml at 0.48ml/s with 23.8psi vacuum @18:13:20.12 Cartridge 7 sampling 498/2000ml at 0.47ml/s with 23.8psi vacuum @18:14:19.95 Cartridge 7 sampling 526/2000ml at 0.46ml/s with 25.4psi vacuum @18:15:20.13 Cartridge 7 sampling 554/2000ml at 0.47ml/s with 23.1psi vacuum @18:16:19.95 Cartridge 7 sampling 582/2000ml at 0.47ml/s with 23.0psi vacuum @18:17:20.19 Cartridge 7 sampling 610/2000ml at 0.46ml/s with 22.5psi vacuum @18:18:20.48 Cartridge 7 sampling 637/2000ml at 0.46ml/s with 21.7psi vacuum @18:19:20.11 Cartridge 7 sampling 664/2000ml at 0.45ml/s with 24.7psi vacuum @18:20:19.96 Cartridge 7 sampling 691/2000ml at 0.44ml/s with 20.5psi vacuum @18:21:20.03 Cartridge 7 sampling 717/2000ml at 0.43ml/s with 21.5psi vacuum @18:22:19.96 Cartridge 7 sampling 743/2000ml at 0.41ml/s with 27.6psi vacuum @18:23:20.03 Cartridge 7 sampling 769/2000ml at 0.43ml/s with 22.1psi vacuum @18:24:19.95 Cartridge 7 sampling 795/2000ml at 0.41ml/s with 27.3psi vacuum @18:25:20.30 Cartridge 7 sampling 820/2000ml at 0.43ml/s with 23.2psi vacuum @18:26:20.19 Cartridge 7 sampling 845/2000ml at 0.43ml/s with 22.2psi vacuum @18:27:19.94 Cartridge 7 sampling 871/2000ml at 0.42ml/s with 21.4psi vacuum @18:28:20.18 Cartridge 7 sampling 895/2000ml at 0.43ml/s with 23.2psi vacuum @18:29:19.94 Cartridge 7 sampling 920/2000ml at 0.40ml/s with 25.3psi vacuum @18:30:20.29 Cartridge 7 sampling 944/2000ml at 0.39ml/s with 27.9psi vacuum @18:31:19.94 Cartridge 7 sampling 968/2000ml at 0.38ml/s with 26.1psi vacuum @18:32:20.11 Cartridge 7 sampling 991/2000ml at 0.40ml/s with 22.8psi vacuum @18:33:19.94 Cartridge 7 sampling 1014/2000ml at 0.37ml/s with 27.5psi vacuum @18:34:20.14 Cartridge 7 sampling 1037/2000ml at 0.36ml/s with 21.7psi vacuum @18:35:19.95 Cartridge 7 sampling 1059/2000ml at 0.37ml/s with 24.0psi vacuum @18:36:20.16 Cartridge 7 sampling 1081/2000ml at 0.38ml/s with 25.4psi vacuum @18:37:19.94 Cartridge 7 sampling 1104/2000ml at 0.36ml/s with 23.3psi vacuum @18:38:20.06 Cartridge 7 sampling 1125/2000ml at 0.36ml/s with 27.1psi vacuum @18:39:19.94 Cartridge 7 sampling 1147/2000ml at 0.38ml/s with 25.7psi vacuum @18:40:20.14 Cartridge 7 sampling 1168/2000ml at 0.37ml/s with 21.3psi vacuum @18:41:19.95 Cartridge 7 sampling 1189/2000ml at 0.36ml/s with 24.8psi vacuum @18:42:20.12 Cartridge 7 sampling 1210/2000ml at 0.35ml/s with 21.5psi vacuum 1.7m deep (17.2psia) @18:43:19.95 Cartridge 7 sampling 1230/2000ml at 0.35ml/s with 20.5psi vacuum @18:44:20.12 Cartridge 7 sampling 1251/2000ml at 0.31ml/s with 25.5psi vacuum @18:45:19.95 Cartridge 7 sampling 1270/2000ml at 0.34ml/s with 25.1psi vacuum @18:46:20.31 Cartridge 7 sampling 1290/2000ml at 0.34ml/s with 24.2psi vacuum @18:47:19.95 Cartridge 7 sampling 1310/2000ml at 0.33ml/s with 27.8psi vacuum @18:48:20.07 Cartridge 7 sampling 1330/2000ml at 0.33ml/s with 27.3psi vacuum @18:49:19.94 Cartridge 7 sampling 1350/2000ml at 0.33ml/s with 22.5psi vacuum @18:50:20.37 Cartridge 7 sampling 1369/2000ml at 0.32ml/s with 21.5psi vacuum @18:51:20.09 Cartridge 7 sampling 1388/2000ml at 0.31ml/s with 26.0psi vacuum @18:52:19.95 Cartridge 7 sampling 1407/2000ml at 0.34ml/s with 25.7psi vacuum @18:53:20.11 Cartridge 7 sampling 1425/2000ml at 0.31ml/s with 19.2psi vacuum @18:54:19.95 Cartridge 7 sampling 1444/2000ml at 0.33ml/s with 27.1psi vacuum @18:55:20.30 Cartridge 7 sampling 1462/2000ml at 0.30ml/s with 28.2psi vacuum @18:56:19.95 Cartridge 7 sampling 1480/2000ml at 0.31ml/s with 22.7psi vacuum @18:57:20.20 Cartridge 7 sampling 1498/2000ml at 0.32ml/s with 25.1psi vacuum @18:58:19.95 Cartridge 7 sampling 1516/2000ml at 0.27ml/s with 27.2psi vacuum @18:59:20.11 Cartridge 7 sampling 1533/2000ml at 0.30ml/s with 21.2psi vacuum @19:00:19.95 Cartridge 7 sampling 1550/2000ml at 0.26ml/s with 27.4psi vacuum @19:01:20.10 Cartridge 7 sampling 1566/2000ml at 0.29ml/s with 23.9psi vacuum @19:02:19.96 Cartridge 7 sampling 1583/2000ml at 0.28ml/s with 27.9psi vacuum @19:03:20.14 Cartridge 7 sampling 1599/2000ml at 0.30ml/s with 27.5psi vacuum @19:04:19.95 Cartridge 7 sampling 1615/2000ml at 0.31ml/s with 23.8psi vacuum @19:05:20.14 Cartridge 7 sampling 1631/2000ml at 0.26ml/s with 20.9psi vacuum @19:06:19.95 Cartridge 7 sampling 1647/2000ml at 0.26ml/s with 25.1psi vacuum @19:07:20.06 Cartridge 7 sampling 1663/2000ml at 0.27ml/s with 24.8psi vacuum @19:08:19.96 Cartridge 7 sampling 1678/2000ml at 0.23ml/s with 27.9psi vacuum @19:09:20.01 Cartridge 7 sampling 1694/2000ml at 0.22ml/s with 23.1psi vacuum @19:10:19.95 Cartridge 7 sampling 1709/2000ml at 0.24ml/s with 27.0psi vacuum @19:11:20.11 Cartridge 7 sampling 1724/2000ml at 0.21ml/s with 27.3psi vacuum @19:12:19.95 Cartridge 7 sampling 1739/2000ml at 0.23ml/s with 20.2psi vacuum @19:13:20.10 Cartridge 7 sampling 1754/2000ml at 0.22ml/s with 28.0psi vacuum @19:14:20.16 Cartridge 7 sampling 1768/2000ml at 0.24ml/s with 16.2psi vacuum 2.7m deep (18.7psia) @19:15:19.95 Cartridge 7 sampling 1783/2000ml at 0.26ml/s with 29.2psi vacuum @19:16:20.22 Cartridge 7 sampling 1798/2000ml at 0.25ml/s with 24.0psi vacuum @19:17:19.95 Cartridge 7 sampling 1812/2000ml at 0.23ml/s with 24.9psi vacuum @19:18:20.32 Cartridge 7 sampling 1826/2000ml at 0.21ml/s with 19.6psi vacuum 1.9m deep (17.5psia) @19:19:19.95 Cartridge 7 sampling 1840/2000ml at 0.21ml/s with 16.6psi vacuum @19:20:20.15 Cartridge 7 sampling 1854/2000ml at 0.27ml/s with 26.9psi vacuum @19:21:19.95 Cartridge 7 sampling 1868/2000ml at 0.20ml/s with 28.0psi vacuum @19:22:20.28 Cartridge 7 sampling 1882/2000ml at 0.23ml/s with 22.8psi vacuum @19:23:19.94 Cartridge 7 sampling 1896/2000ml at 0.21ml/s with 25.5psi vacuum @19:24:20.14 Cartridge 7 sampling 1909/2000ml at 0.26ml/s with 19.4psi vacuum 1.7m deep (17.1psia) @19:25:19.94 Cartridge 7 sampling 1923/2000ml at 0.21ml/s with 22.3psi vacuum @19:26:20.19 Cartridge 7 sampling 1936/2000ml at 0.20ml/s with 26.7psi vacuum @19:27:19.95 Cartridge 7 sampling 1950/2000ml at 0.22ml/s with 24.6psi vacuum @19:28:20.16 Cartridge 7 sampling 1963/2000ml at 0.24ml/s with 17.4psi vacuum @19:29:19.94 Cartridge 7 sampling 1977/2000ml at 0.20ml/s with 30.1psi vacuum @19:30:20.07 Cartridge 7 sampling 1990/2000ml at 0.26ml/s with 28.1psi vacuum @19:31:04.63 Pressure.coast @19:31:05.11 Intake.close @19:31:05.27 Exhaust.close @19:31:05.70 Toroid.reconfigure :hold @19:31:05.91 Toroid.seek :bypass @19:31:07.58 Toroid.reconfigure :default @19:31:07.66 Toroid.seek :clear @19:31:09.29 Sampled 2000.0ml @19:31:09.81 SamplePump.setPosition! 0ml @19:31:09.86 Plunge.seek :reagent @19:31:09.97 Exhaust.open @19:31:10.70 Intake.open @19:31:10.84 Sample loop closed. Ready to Deliver Reagents @19:31:10.86 Archive.seek :sealed @19:31:11.59 SamplePump.reconfigure :slow @19:31:11.68 SamplePump.seek -35ml @19:31:14.59 Archive.seek :evac @19:31:28.67 delay 1:00 @19:32:26.79 delay 1 second @19:32:27.81 Bypass.open @19:32:28.60 Exhaust.close @19:32:28.81 Archive.seek :clear @19:32:29.45 SamplePump.seek -33ml @19:32:34.09 delay 1 second @19:32:35.42 SamplePump.seek -41ml @19:32:46.09 RNAlater soak for 10:00 @19:32:52.91 delay 1 second @19:32:54.13 Intake.close @19:32:54.47 Bypass.close @19:42:46.19 Archive.seek :tVent @19:42:47.87 delay 1 second @19:42:48.90 Archive.seek :evac @19:43:04.91 delay 10 seconds @19:43:14.98 Sample preserved and evacuated @19:43:15.95 Carousel.seek 7 @19:43:16.85 Plunge.seek :locked @19:43:18.11 Archive.seek :clear @19:43:18.32 Air.seek :clear @19:43:19.26 Lysis1.seek :clear @19:43:35.51 Archive.coast @19:43:35.70 Air.coast @19:43:35.77 Lysis1.coast @19:43:35.85 Lysis2.coast @19:43:35.91 Plunge.coast @19:43:35.98 SamplePump.coast @19:43:36.06 Toroid.coast @19:43:36.13 Carousel.coast @19:43:36.34 Gate.power :motor,:OFF @19:43:36.36 Gate.power :core,:OFF @19:43:36.39 Gate.power :analytic1,:OFF @19:43:36.42 Gate.power :analytic2,:OFF @19:43:36.45 Gate.power :analytic3,:OFF @19:43:36.48 Gate.power :main,:OFF @19:43:36.51 Safely stopped and ready to power off @19:46:42.93 -> show -200 @19:46:42.97 NoMethodError in preston -- undefined method `show' for Object:Class @19:48:18.34 -> exit @19:48:18.41 Session Ended @09:05:43.79PDT26-May-21 Session Begun @09:05:48.12 -> slots @09:06:00.53 -> Cartridge.release 6 @10:21:57.71 -> showlog -200 @10:22:22.20 -> showlog -200 @10:23:32.57 -> shoot @10:23:32.95 Gate.power :main,:ON,limit to 5000mA @10:23:33.01 Gate.power :analytic1,:ON,limit to 5000mA @10:23:33.07 Gate.power :analytic2,:ON,limit to 5000mA @10:23:33.16 Gate.power :analytic3,:ON,limit to 5000mA @10:23:33.33 Powering core bus @10:23:35.73 Archive.configure :default @10:23:35.90 Air.configure :default @10:23:36.04 Lysis1.configure :default @10:23:36.18 Lysis2.configure :default @10:23:36.34 Plunge.configure :default @10:23:36.51 SamplePump.configure :default @10:23:36.68 ExtraPump.configure :default @10:23:36.84 Toroid.configure :default @10:23:37.01 Carousel.configure :default @10:23:37.18 Pressure.configure :default @10:23:37.27 All elf controllers running firmware version 4.61 @10:23:37.30 Gate.power :motor,:ON,limit to 5000mA @10:23:37.52 Archive.reconfigure :default @10:23:37.86 Archive.home.jog 0 @10:23:38.09 Plunge.reconfigure :default @10:23:38.16 Lysis1.reconfigure :default @10:23:38.23 Air.reconfigure :default @10:23:38.29 Lysis2.reconfigure :default @10:23:38.35 Toroid.reconfigure :default @10:23:38.60 Lysis1.home.jog 0 @10:23:38.65 Plunge.home.jog 0 @10:23:38.68 Lysis2.home.jog 0 @10:23:38.72 Air.home.jog 0 @10:23:38.82 Toroid.home.jog 0 @10:23:39.31 Archive.reconfigure :homing @10:23:39.56 Archive.jog 152 @10:23:40.65 Toroid.reconfigure :homing @10:23:41.12 Lysis2.reconfigure :homing @10:23:41.24 Lysis1.reconfigure :homing @10:23:41.44 Plunge.reconfigure :homing @10:23:41.53 Air.reconfigure :homing @10:23:42.11 Toroid.jog 628 @10:23:42.31 Lysis1.jog 1348 @10:23:42.37 Lysis2.jog 1358 @10:23:42.45 Air.jog 1366 @10:23:42.53 Plunge.jog 855 @10:23:43.25 Archive.setRawPosition! 50 @10:23:43.36 Archive.seek :home @10:23:43.80 Archive.reconfigure :default @10:23:44.10 Lysis1.setRawPosition! 50 @10:23:44.22 Lysis1.seek :home @10:23:44.48 Plunge.setRawPosition! 50 @10:23:44.63 Lysis2.setRawPosition! 50 @10:23:44.71 Air.setRawPosition! 50 @10:23:45.16 Plunge.seek :home @10:23:45.31 Lysis2.seek :home @10:23:45.49 Air.seek :home @10:23:46.26 Plunge.reconfigure :default @10:23:46.35 Lysis2.reconfigure :default @10:23:46.39 Lysis1.reconfigure :default @10:23:46.46 Toroid.setRawPosition! 0 @10:23:46.50 Air.reconfigure :default @10:23:46.74 Toroid.seek :home @10:23:46.88 Toroid.reconfigure :default @10:23:47.10 Toroid.seek :clear @10:23:48.74 Carousel.reconfigure :default @10:23:48.83 Carousel.home.jog 0 @10:23:49.35 Carousel.seek 7 @10:23:49.55 Selecting Cartridge 6 @10:23:49.80 Carousel.seek 6 @10:23:50.35 Lysis1.seek :locked @10:23:53.79 Plunge.seek :locked @10:23:54.27 Air.seek :locked @10:23:58.51 Mounted archive Cartridge @10:23:58.87 Duration of filtering limited to 4:05:00 @10:23:59.20 Priming sample loop w/160ml, bypass w/10ml @10:23:59.22 SamplePump.setPosition! 0ml @10:23:59.30 Bypass.open @10:24:01.00 Intake.open @10:24:02.08 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @10:24:04.67 Vacuum tare is 17% of full scale @10:24:04.80 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @10:24:05.50 Pressure.reconfigure :default @10:24:05.58 Pressure.hold @10:24:05.82 SamplePump.reconfigure :prime @10:24:05.91 SamplePump.seek 10ml,1:21 @10:24:17.04 Pressure.coast @10:24:17.12 SamplePump.setPosition! 0ml @10:24:17.19 Exhaust.open @10:24:18.62 Bypass.close @10:24:20.53 Pressure.reconfigure :default @10:24:20.77 Pressure.hold @10:24:21.56 SamplePump.reconfigure :prime @10:24:21.81 SamplePump.seek 160ml,13:51 @10:27:13.34 Pressure.coast @10:27:13.42 Bypass.open @10:27:14.15 Intake.close @10:27:15.39 Toroid.reconfigure :hold @10:27:15.55 Toroid.seek :crack @10:27:17.51 Toroid.seek :cartridge @10:27:18.56 Toroid.reconfigure :default @10:27:18.70 Toroid.seek :clear @10:27:20.59 Pressurizing cartridge to 26psi above 14.6psia @10:27:20.61 SamplePump.setPosition! 0ml @10:27:20.69 Intake.open @10:27:21.44 Bypass.close @10:27:22.15 Exhaust.close @10:27:22.96 Pressure.reconfigure :default @10:27:23.04 Pressure.hold @10:27:23.29 SamplePump.reconfigure :charge @10:27:23.38 SamplePump.seek 2.3ml,42 seconds @10:27:38.12 Pressure.coast @10:27:38.20 Intake.close @10:27:39.07 Pressurized cartridge with 0.995ml to 26.0psi @10:27:39.09 Isolated @35.1psia (20.5psi) @10:27:39.12 Checking for cartridge leaks from 35.1psia @10:27:39.18 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @10:27:43.89 Cartridge pressure fell 3% to 34.0psia @10:27:44.16 Sampling 2000.0ml @10:27:44.20 SamplePump.setPosition! 0ml @10:27:44.38 Exhaust.open @10:27:45.09 Intake.open @10:27:46.34 Pressure.reconfigure :default @10:27:46.42 Pressure.hold @10:27:46.74 Sea 1.8m deep (17.3psia) @10:27:47.35 SamplePump.reconfigure :sample @10:27:47.57 SamplePump.seek 2000ml,2:47:12 @10:27:49.24 Cartridge 6 sampling 1/2000ml at 0.62ml/s with 25.5psi vacuum @10:28:49.41 Cartridge 6 sampling 26/2000ml at 0.43ml/s with 25.1psi vacuum @10:29:49.24 Cartridge 6 sampling 51/2000ml at 0.42ml/s with 22.4psi vacuum @10:30:49.42 Cartridge 6 sampling 77/2000ml at 0.44ml/s with 23.2psi vacuum @10:31:49.23 Cartridge 6 sampling 103/2000ml at 0.44ml/s with 25.1psi vacuum @10:32:49.45 Cartridge 6 sampling 131/2000ml at 0.46ml/s with 26.0psi vacuum @10:33:49.23 Cartridge 6 sampling 158/2000ml at 0.45ml/s with 21.2psi vacuum @10:34:49.37 Cartridge 6 sampling 185/2000ml at 0.44ml/s with 20.1psi vacuum @10:35:49.23 Cartridge 6 sampling 213/2000ml at 0.43ml/s with 25.2psi vacuum @10:36:49.38 Cartridge 6 sampling 240/2000ml at 0.46ml/s with 26.1psi vacuum @10:37:49.24 Cartridge 6 sampling 268/2000ml at 0.46ml/s with 25.3psi vacuum @10:38:49.49 Cartridge 6 sampling 295/2000ml at 0.46ml/s with 24.5psi vacuum @10:39:49.24 Cartridge 6 sampling 322/2000ml at 0.43ml/s with 16.8psi vacuum @10:40:49.26 Cartridge 6 sampling 349/2000ml at 0.44ml/s with 23.4psi vacuum @10:41:49.50 Cartridge 6 sampling 376/2000ml at 0.45ml/s with 21.9psi vacuum @10:42:49.23 Cartridge 6 sampling 403/2000ml at 0.43ml/s with 17.3psi vacuum @10:43:49.44 Cartridge 6 sampling 430/2000ml at 0.46ml/s with 24.2psi vacuum @10:44:49.23 Cartridge 6 sampling 457/2000ml at 0.45ml/s with 24.5psi vacuum @10:45:49.61 Cartridge 6 sampling 484/2000ml at 0.45ml/s with 23.0psi vacuum @10:46:49.78 Cartridge 6 sampling 511/2000ml at 0.44ml/s with 23.5psi vacuum @10:47:49.41 Cartridge 6 sampling 537/2000ml at 0.44ml/s with 22.2psi vacuum @10:48:49.24 Cartridge 6 sampling 563/2000ml at 0.41ml/s with 21.9psi vacuum @10:49:49.42 Cartridge 6 sampling 587/2000ml at 0.42ml/s with 25.7psi vacuum @10:50:49.24 Cartridge 6 sampling 612/2000ml at 0.40ml/s with 19.8psi vacuum @10:51:49.40 Cartridge 6 sampling 636/2000ml at 0.38ml/s with 23.3psi vacuum @10:52:49.24 Cartridge 6 sampling 660/2000ml at 0.40ml/s with 25.2psi vacuum @10:53:49.33 Cartridge 6 sampling 683/2000ml at 0.37ml/s with 24.3psi vacuum @10:54:49.23 Cartridge 6 sampling 706/2000ml at 0.39ml/s with 25.8psi vacuum @10:55:49.40 Cartridge 6 sampling 729/2000ml at 0.38ml/s with 24.5psi vacuum @10:56:49.23 Cartridge 6 sampling 751/2000ml at 0.39ml/s with 24.4psi vacuum @10:57:49.44 Cartridge 6 sampling 775/2000ml at 0.40ml/s with 27.2psi vacuum @10:58:49.23 Cartridge 6 sampling 798/2000ml at 0.38ml/s with 26.0psi vacuum @10:59:49.26 Cartridge 6 sampling 821/2000ml at 0.37ml/s with 22.8psi vacuum @11:00:49.23 Cartridge 6 sampling 844/2000ml at 0.37ml/s with 21.1psi vacuum @11:01:49.31 Cartridge 6 sampling 866/2000ml at 0.38ml/s with 24.1psi vacuum @11:02:49.23 Cartridge 6 sampling 889/2000ml at 0.38ml/s with 24.0psi vacuum @11:03:49.58 Cartridge 6 sampling 911/2000ml at 0.37ml/s with 27.0psi vacuum @11:04:49.41 Cartridge 6 sampling 934/2000ml at 0.37ml/s with 24.8psi vacuum @11:05:49.39 Cartridge 6 sampling 955/2000ml at 0.36ml/s with 26.4psi vacuum @11:06:49.23 Cartridge 6 sampling 977/2000ml at 0.38ml/s with 23.9psi vacuum @11:07:49.46 Cartridge 6 sampling 999/2000ml at 0.37ml/s with 22.4psi vacuum @11:08:49.49 Cartridge 6 sampling 1021/2000ml at 0.35ml/s with 27.0psi vacuum @11:09:49.24 Cartridge 6 sampling 1043/2000ml at 0.36ml/s with 22.6psi vacuum @11:10:49.77 Cartridge 6 sampling 1064/2000ml at 0.34ml/s with 27.4psi vacuum @11:11:49.56 Cartridge 6 sampling 1085/2000ml at 0.36ml/s with 23.6psi vacuum @11:12:49.52 Cartridge 6 sampling 1106/2000ml at 0.36ml/s with 22.6psi vacuum @11:13:49.41 Cartridge 6 sampling 1127/2000ml at 0.34ml/s with 25.2psi vacuum @11:14:49.54 Cartridge 6 sampling 1148/2000ml at 0.34ml/s with 22.8psi vacuum @11:15:49.30 Cartridge 6 sampling 1168/2000ml at 0.35ml/s with 25.5psi vacuum @11:16:49.68 Cartridge 6 sampling 1188/2000ml at 0.33ml/s with 22.5psi vacuum @11:17:49.24 Cartridge 6 sampling 1208/2000ml at 0.32ml/s with 20.8psi vacuum @11:18:49.44 Cartridge 6 sampling 1228/2000ml at 0.33ml/s with 21.1psi vacuum @11:19:49.24 Cartridge 6 sampling 1247/2000ml at 0.33ml/s with 27.3psi vacuum @11:20:49.42 Cartridge 6 sampling 1266/2000ml at 0.33ml/s with 22.6psi vacuum @11:21:49.23 Cartridge 6 sampling 1286/2000ml at 0.35ml/s with 21.1psi vacuum @11:22:49.49 Cartridge 6 sampling 1305/2000ml at 0.31ml/s with 21.4psi vacuum @11:23:49.24 Cartridge 6 sampling 1324/2000ml at 0.33ml/s with 25.6psi vacuum @11:24:49.42 Cartridge 6 sampling 1343/2000ml at 0.31ml/s with 21.8psi vacuum @11:25:49.24 Cartridge 6 sampling 1362/2000ml at 0.32ml/s with 24.0psi vacuum @11:26:49.41 Cartridge 6 sampling 1381/2000ml at 0.31ml/s with 21.8psi vacuum @11:27:49.24 Cartridge 6 sampling 1399/2000ml at 0.31ml/s with 24.7psi vacuum @11:28:49.51 Cartridge 6 sampling 1418/2000ml at 0.29ml/s with 20.4psi vacuum @11:29:49.24 Cartridge 6 sampling 1436/2000ml at 0.30ml/s with 21.2psi vacuum @11:30:49.61 Cartridge 6 sampling 1454/2000ml at 0.28ml/s with 25.2psi vacuum @11:31:49.23 Cartridge 6 sampling 1472/2000ml at 0.32ml/s with 23.1psi vacuum @11:32:49.39 Cartridge 6 sampling 1490/2000ml at 0.28ml/s with 27.7psi vacuum @11:33:49.24 Cartridge 6 sampling 1507/2000ml at 0.31ml/s with 22.1psi vacuum @11:34:49.32 Cartridge 6 sampling 1525/2000ml at 0.29ml/s with 25.0psi vacuum @11:35:49.23 Cartridge 6 sampling 1542/2000ml at 0.29ml/s with 25.1psi vacuum @11:36:49.59 Cartridge 6 sampling 1559/2000ml at 0.30ml/s with 20.7psi vacuum 2.2m deep (17.9psia) @11:37:49.43 Cartridge 6 sampling 1575/2000ml at 0.31ml/s with 22.1psi vacuum @11:38:49.23 Cartridge 6 sampling 1592/2000ml at 0.26ml/s with 27.6psi vacuum @11:39:49.48 Cartridge 6 sampling 1608/2000ml at 0.28ml/s with 25.4psi vacuum @11:40:49.23 Cartridge 6 sampling 1625/2000ml at 0.27ml/s with 27.6psi vacuum @11:41:49.58 Cartridge 6 sampling 1641/2000ml at 0.23ml/s with 21.0psi vacuum @11:42:49.24 Cartridge 6 sampling 1657/2000ml at 0.25ml/s with 21.0psi vacuum @11:43:49.45 Cartridge 6 sampling 1673/2000ml at 0.26ml/s with 17.5psi vacuum 2.0m deep (17.6psia) @11:44:49.24 Cartridge 6 sampling 1689/2000ml at 0.28ml/s with 22.2psi vacuum @11:45:49.37 Cartridge 6 sampling 1704/2000ml at 0.29ml/s with 24.0psi vacuum @11:46:49.23 Cartridge 6 sampling 1720/2000ml at 0.24ml/s with 29.2psi vacuum @11:47:49.40 Cartridge 6 sampling 1735/2000ml at 0.27ml/s with 23.6psi vacuum @11:48:49.23 Cartridge 6 sampling 1751/2000ml at 0.25ml/s with 18.9psi vacuum @11:49:49.44 Cartridge 6 sampling 1766/2000ml at 0.26ml/s with 22.7psi vacuum @11:50:49.24 Cartridge 6 sampling 1781/2000ml at 0.26ml/s with 23.5psi vacuum @11:51:49.47 Cartridge 6 sampling 1796/2000ml at 0.23ml/s with 21.2psi vacuum @11:52:49.23 Cartridge 6 sampling 1811/2000ml at 0.23ml/s with 21.2psi vacuum @11:53:49.39 Cartridge 6 sampling 1826/2000ml at 0.25ml/s with 22.8psi vacuum @11:54:49.24 Cartridge 6 sampling 1841/2000ml at 0.29ml/s with 27.0psi vacuum @11:55:49.26 Cartridge 6 sampling 1855/2000ml at 0.26ml/s with 22.3psi vacuum @11:56:49.23 Cartridge 6 sampling 1870/2000ml at 0.24ml/s with 26.4psi vacuum @11:57:49.61 Cartridge 6 sampling 1884/2000ml at 0.26ml/s with 23.1psi vacuum @11:58:49.43 Cartridge 6 sampling 1898/2000ml at 0.22ml/s with 26.6psi vacuum @11:59:49.30 Cartridge 6 sampling 1913/2000ml at 0.21ml/s with 16.0psi vacuum @12:00:49.56 Cartridge 6 sampling 1927/2000ml at 0.28ml/s with 22.7psi vacuum @12:01:49.23 Cartridge 6 sampling 1940/2000ml at 0.20ml/s with 23.8psi vacuum @12:02:49.63 Cartridge 6 sampling 1954/2000ml at 0.25ml/s with 20.9psi vacuum 3.2m deep (19.4psia) @12:03:49.23 Cartridge 6 sampling 1968/2000ml at 0.23ml/s with 29.1psi vacuum @12:04:49.38 Cartridge 6 sampling 1982/2000ml at 0.25ml/s with 20.4psi vacuum 1.9m deep (17.5psia) @12:05:49.23 Cartridge 6 sampling 1996/2000ml at 0.22ml/s with 23.9psi vacuum @12:06:05.47 Pressure.coast @12:06:05.80 Intake.close @12:06:05.92 Exhaust.close @12:06:06.83 Toroid.reconfigure :hold @12:06:06.92 Toroid.seek :bypass @12:06:08.56 Toroid.reconfigure :default @12:06:08.64 Toroid.seek :clear @12:06:10.28 Sampled 2000.0ml @12:06:10.81 SamplePump.setPosition! 0ml @12:06:10.84 Plunge.seek :reagent @12:06:11.26 Exhaust.open @12:06:12.24 Sample loop closed. Ready to Deliver Reagents @12:06:12.27 Intake.open @12:06:12.32 Archive.seek :sealed @12:06:14.76 SamplePump.reconfigure :slow @12:06:15.10 SamplePump.seek -35ml @12:06:16.34 Archive.seek :evac @12:06:30.47 delay 1:00 @12:07:30.22 delay 1 second @12:07:30.57 Archive.seek :clear @12:07:31.24 Bypass.open @12:07:31.94 Exhaust.close @12:07:32.74 SamplePump.seek -33ml @12:07:37.36 delay 1 second @12:07:38.49 SamplePump.seek -41ml @12:07:47.74 RNAlater soak for 10:00 @12:07:55.95 delay 1 second @12:07:57.11 Intake.close @12:07:57.23 Bypass.close @12:17:47.78 Archive.seek :tVent @12:17:49.47 delay 1 second @12:17:50.51 Archive.seek :evac @12:18:06.53 delay 10 seconds @12:18:16.57 Sample preserved and evacuated @12:18:16.83 Carousel.seek 6 @12:18:17.07 Plunge.seek :locked @12:18:18.08 Archive.seek :clear @12:18:18.61 Air.seek :clear @12:18:19.28 Lysis1.seek :clear @12:18:35.54 Archive.coast @12:18:35.62 Air.coast @12:18:35.70 Lysis1.coast @12:18:35.77 Lysis2.coast @12:18:35.85 Plunge.coast @12:18:35.92 SamplePump.coast @12:18:36.00 Toroid.coast @12:18:36.08 Carousel.coast @12:18:36.30 Gate.power :motor,:OFF @12:18:36.33 Gate.power :core,:OFF @12:18:36.36 Gate.power :analytic1,:OFF @12:18:36.39 Gate.power :analytic2,:OFF @12:18:36.43 Gate.power :analytic3,:OFF @12:18:36.46 Gate.power :main,:OFF @12:18:36.50 Safely stopped and ready to power off @12:18:56.36 -> exit @12:18:56.43 Session Ended @17:35:14.18 Session Begun @17:35:17.28 -> slots @17:35:24.87 -> Cartridge.dry @17:35:38.73 -> Cartrige.release 5 @17:35:38.77 NameError in preston -- uninitialized constant Cartrige @17:35:50.76 -> Cartridge.release 5 @17:35:54.38 -> Cartridge.dry @17:36:19.74 -> shoot @17:36:19.89 Gate.power :main,:ON,limit to 5000mA @17:36:19.91 Gate.power :analytic1,:ON,limit to 5000mA @17:36:19.94 Gate.power :analytic2,:ON,limit to 5000mA @17:36:19.99 Gate.power :analytic3,:ON,limit to 5000mA @17:36:20.05 Powering core bus @17:36:22.73 Archive.configure :default @17:36:23.37 Air.configure :default @17:36:23.74 Lysis1.configure :default @17:36:24.10 Lysis2.configure :default @17:36:24.51 Plunge.configure :default @17:36:24.92 SamplePump.configure :default @17:36:25.28 ExtraPump.configure :default @17:36:25.74 Toroid.configure :default @17:36:26.22 Carousel.configure :default @17:36:26.55 Pressure.configure :default @17:36:26.76 All elf controllers running firmware version 4.61 @17:36:26.81 Gate.power :motor,:ON,limit to 5000mA @17:36:27.90 Lysis2.reconfigure :default @17:36:28.05 Air.reconfigure :default @17:36:28.19 Archive.reconfigure :default @17:36:28.36 Plunge.reconfigure :default @17:36:28.41 Lysis1.reconfigure :default @17:36:28.48 Toroid.reconfigure :default @17:36:28.76 Lysis2.home.jog 0 @17:36:28.81 Air.home.jog 0 @17:36:28.84 Plunge.home.jog 0 @17:36:28.88 Archive.home.jog 0 @17:36:28.92 Toroid.home.jog 0 @17:36:28.97 Lysis1.home.jog 0 @17:36:29.79 Toroid.reconfigure :homing @17:36:29.84 Plunge.reconfigure :homing @17:36:29.88 Lysis2.reconfigure :homing @17:36:29.93 Air.reconfigure :homing @17:36:29.99 Archive.reconfigure :homing @17:36:30.04 Lysis1.reconfigure :homing @17:36:30.85 Plunge.jog 859 @17:36:30.90 Toroid.jog 624 @17:36:30.95 Lysis2.jog 1358 @17:36:31.00 Air.jog 1369 @17:36:31.04 Archive.jog 161 @17:36:31.10 Lysis1.jog 1355 @17:36:31.76 Archive.setRawPosition! 50 @17:36:31.83 Archive.seek :home @17:36:32.07 Plunge.setRawPosition! 50 @17:36:32.15 Plunge.seek :home @17:36:32.27 Archive.reconfigure :default @17:36:32.40 Lysis2.setRawPosition! 50 @17:36:32.49 Lysis2.seek :home @17:36:32.65 Lysis1.setRawPosition! 50 @17:36:32.89 Air.setRawPosition! 50 @17:36:32.97 Plunge.reconfigure :default @17:36:33.36 Lysis1.seek :home @17:36:33.64 Air.seek :home @17:36:34.42 Lysis2.reconfigure :default @17:36:34.61 Toroid.setRawPosition! 0 @17:36:34.69 Lysis1.reconfigure :default @17:36:35.11 Toroid.seek :home @17:36:35.49 Toroid.reconfigure :default @17:36:35.59 Air.reconfigure :default @17:36:36.38 Toroid.seek :clear @17:36:38.13 Carousel.reconfigure :default @17:36:38.34 Carousel.home.jog 0 @17:36:38.94 Carousel.seek 6 @17:36:39.11 Selecting Cartridge 5 @17:36:39.36 Carousel.seek 5 @17:36:39.91 Lysis1.seek :locked @17:36:43.38 Plunge.seek :locked @17:36:44.07 Air.seek :locked @17:36:48.48 Mounted archive Cartridge @17:36:48.80 Duration of filtering limited to 4:05:00 @17:36:49.13 Priming sample loop w/160ml, bypass w/10ml @17:36:49.15 SamplePump.setPosition! 0ml @17:36:49.22 Bypass.open @17:36:50.72 Intake.open @17:36:51.57 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @17:36:54.22 Vacuum tare is 17% of full scale @17:36:54.26 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @17:36:55.17 Pressure.reconfigure :default @17:36:55.40 Pressure.hold @17:36:56.01 SamplePump.reconfigure :prime @17:36:56.20 SamplePump.seek 10ml,1:21 @17:37:07.33 Pressure.coast @17:37:07.41 SamplePump.setPosition! 0ml @17:37:07.48 Exhaust.open @17:37:08.95 Bypass.close @17:37:10.53 Pressure.reconfigure :default @17:37:10.61 Pressure.hold @17:37:10.91 SamplePump.reconfigure :prime @17:37:11.01 SamplePump.seek 160ml,13:51 @17:40:02.58 Pressure.coast @17:40:02.66 Bypass.open @17:40:03.38 Intake.close @17:40:04.24 Toroid.reconfigure :hold @17:40:04.32 Toroid.seek :crack @17:40:05.95 Toroid.seek :cartridge @17:40:07.11 Toroid.reconfigure :default @17:40:07.27 Toroid.seek :clear @17:40:09.60 Pressurizing cartridge to 26psi above 14.5psia @17:40:09.64 SamplePump.setPosition! 0ml @17:40:09.80 Intake.open @17:40:10.66 Bypass.close @17:40:11.59 Exhaust.close @17:40:12.45 Pressure.reconfigure :default @17:40:12.53 Pressure.hold @17:40:12.77 SamplePump.reconfigure :charge @17:40:12.86 SamplePump.seek 2.3ml,42 seconds @17:40:27.71 Pressure.coast @17:40:27.94 Intake.close @17:40:28.78 Pressurized cartridge with 0.996ml to 26.1psi @17:40:28.83 Isolated @35.6psia (21.2psi) @17:40:28.91 Checking for cartridge leaks from 35.6psia @17:40:29.03 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @17:40:33.62 Cartridge pressure fell 2% to 35.0psia @17:40:33.73 Sampling 2000.0ml @17:40:33.75 SamplePump.setPosition! 0ml @17:40:33.84 Exhaust.open @17:40:34.63 Intake.open @17:40:35.94 Pressure.reconfigure :default @17:40:36.15 Pressure.hold @17:40:37.87 SamplePump.reconfigure :sample @17:40:38.17 SamplePump.seek 2000ml,2:47:12 @17:40:39.08 Cartridge 5 sampling 0/2000ml at 0.24ml/s with 4.0psi vacuum @17:41:38.92 Cartridge 5 sampling 28/2000ml at 0.46ml/s with 24.4psi vacuum @17:42:39.08 Cartridge 5 sampling 55/2000ml at 0.45ml/s with 25.1psi vacuum @17:43:38.91 Cartridge 5 sampling 82/2000ml at 0.45ml/s with 24.8psi vacuum @17:44:39.08 Cartridge 5 sampling 110/2000ml at 0.46ml/s with 24.7psi vacuum @17:45:38.91 Cartridge 5 sampling 137/2000ml at 0.45ml/s with 25.0psi vacuum @17:46:39.20 Cartridge 5 sampling 164/2000ml at 0.44ml/s with 25.9psi vacuum @17:47:39.19 Cartridge 5 sampling 191/2000ml at 0.44ml/s with 22.2psi vacuum @17:48:39.02 Cartridge 5 sampling 218/2000ml at 0.45ml/s with 24.7psi vacuum @17:49:39.07 Cartridge 5 sampling 245/2000ml at 0.44ml/s with 25.5psi vacuum @17:50:38.92 Cartridge 5 sampling 271/2000ml at 0.42ml/s with 26.1psi vacuum @17:51:39.10 Cartridge 5 sampling 297/2000ml at 0.42ml/s with 25.0psi vacuum @17:52:38.92 Cartridge 5 sampling 323/2000ml at 0.43ml/s with 24.7psi vacuum @17:53:39.13 Cartridge 5 sampling 348/2000ml at 0.41ml/s with 23.5psi vacuum @17:54:38.91 Cartridge 5 sampling 373/2000ml at 0.42ml/s with 25.8psi vacuum @17:55:39.08 Cartridge 5 sampling 398/2000ml at 0.40ml/s with 25.3psi vacuum @17:56:38.91 Cartridge 5 sampling 422/2000ml at 0.42ml/s with 21.6psi vacuum @17:57:39.12 Cartridge 5 sampling 447/2000ml at 0.42ml/s with 24.1psi vacuum @17:58:38.91 Cartridge 5 sampling 473/2000ml at 0.43ml/s with 26.2psi vacuum @17:59:39.13 Cartridge 5 sampling 498/2000ml at 0.43ml/s with 25.9psi vacuum @18:00:38.91 Cartridge 5 sampling 524/2000ml at 0.42ml/s with 25.8psi vacuum @18:01:39.17 Cartridge 5 sampling 550/2000ml at 0.43ml/s with 23.9psi vacuum @18:02:38.91 Cartridge 5 sampling 575/2000ml at 0.43ml/s with 20.0psi vacuum @18:03:39.04 Cartridge 5 sampling 600/2000ml at 0.42ml/s with 20.8psi vacuum @18:04:38.92 Cartridge 5 sampling 624/2000ml at 0.40ml/s with 22.0psi vacuum @18:05:39.07 Cartridge 5 sampling 648/2000ml at 0.38ml/s with 18.0psi vacuum @18:06:38.91 Cartridge 5 sampling 673/2000ml at 0.41ml/s with 24.6psi vacuum @18:07:39.04 Cartridge 5 sampling 696/2000ml at 0.38ml/s with 26.7psi vacuum @18:08:38.92 Cartridge 5 sampling 720/2000ml at 0.38ml/s with 21.3psi vacuum @18:09:39.20 Cartridge 5 sampling 743/2000ml at 0.39ml/s with 25.9psi vacuum @18:10:39.20 Cartridge 5 sampling 766/2000ml at 0.37ml/s with 26.4psi vacuum @18:11:38.91 Cartridge 5 sampling 788/2000ml at 0.36ml/s with 20.9psi vacuum @18:12:39.30 Cartridge 5 sampling 810/2000ml at 0.38ml/s with 25.3psi vacuum @18:13:38.91 Cartridge 5 sampling 832/2000ml at 0.32ml/s with 24.7psi vacuum @18:14:39.11 Cartridge 5 sampling 854/2000ml at 0.35ml/s with 22.8psi vacuum @18:15:38.92 Cartridge 5 sampling 875/2000ml at 0.36ml/s with 23.5psi vacuum @18:16:39.17 Cartridge 5 sampling 896/2000ml at 0.36ml/s with 23.9psi vacuum @18:17:38.91 Cartridge 5 sampling 917/2000ml at 0.35ml/s with 21.1psi vacuum @18:18:39.06 Cartridge 5 sampling 937/2000ml at 0.35ml/s with 25.7psi vacuum @18:19:38.92 Cartridge 5 sampling 957/2000ml at 0.33ml/s with 27.2psi vacuum @18:20:39.05 Cartridge 5 sampling 977/2000ml at 0.34ml/s with 22.1psi vacuum @18:21:38.91 Cartridge 5 sampling 996/2000ml at 0.33ml/s with 23.0psi vacuum @18:22:39.04 Cartridge 5 sampling 1016/2000ml at 0.34ml/s with 20.2psi vacuum 2.4m deep (18.3psia) @18:23:38.91 Cartridge 5 sampling 1035/2000ml at 0.35ml/s with 25.3psi vacuum @18:24:39.11 Cartridge 5 sampling 1054/2000ml at 0.32ml/s with 21.2psi vacuum 1.6m deep (17.1psia) @18:25:38.91 Cartridge 5 sampling 1073/2000ml at 0.30ml/s with 19.2psi vacuum @18:26:39.02 Cartridge 5 sampling 1092/2000ml at 0.30ml/s with 17.6psi vacuum 2.0m deep (17.6psia) @18:27:38.92 Cartridge 5 sampling 1110/2000ml at 0.29ml/s with 20.8psi vacuum @18:28:38.98 Cartridge 5 sampling 1127/2000ml at 0.28ml/s with 20.8psi vacuum @18:29:38.91 Cartridge 5 sampling 1145/2000ml at 0.31ml/s with 21.8psi vacuum @18:30:39.10 Cartridge 5 sampling 1162/2000ml at 0.25ml/s with 28.1psi vacuum @18:31:39.29 Cartridge 5 sampling 1178/2000ml at 0.27ml/s with 23.9psi vacuum @18:32:38.98 Cartridge 5 sampling 1195/2000ml at 0.28ml/s with 25.4psi vacuum @18:33:39.56 Cartridge 5 sampling 1211/2000ml at 0.24ml/s with 23.1psi vacuum @18:34:39.01 Cartridge 5 sampling 1226/2000ml at 0.27ml/s with 22.1psi vacuum @18:35:39.08 Cartridge 5 sampling 1242/2000ml at 0.26ml/s with 29.1psi vacuum @18:36:38.91 Cartridge 5 sampling 1257/2000ml at 0.27ml/s with 19.4psi vacuum @18:37:39.20 Cartridge 5 sampling 1272/2000ml at 0.24ml/s with 21.7psi vacuum @18:38:38.92 Cartridge 5 sampling 1286/2000ml at 0.23ml/s with 20.6psi vacuum @18:39:39.03 Cartridge 5 sampling 1300/2000ml at 0.21ml/s with 20.7psi vacuum @18:40:38.92 Cartridge 5 sampling 1315/2000ml at 0.21ml/s with 30.2psi vacuum @18:41:39.06 Cartridge 5 sampling 1329/2000ml at 0.23ml/s with 25.2psi vacuum @18:42:38.91 Cartridge 5 sampling 1343/2000ml at 0.22ml/s with 20.8psi vacuum @18:43:39.11 Cartridge 5 sampling 1356/2000ml at 0.23ml/s with 28.8psi vacuum @18:44:38.92 Cartridge 5 sampling 1370/2000ml at 0.21ml/s with 20.8psi vacuum @18:45:39.19 Cartridge 5 sampling 1384/2000ml at 0.23ml/s with 28.8psi vacuum @18:46:38.91 Cartridge 5 sampling 1397/2000ml at 0.24ml/s with 19.3psi vacuum @18:47:39.13 Cartridge 5 sampling 1411/2000ml at 0.21ml/s with 22.3psi vacuum @18:48:38.91 Cartridge 5 sampling 1425/2000ml at 0.22ml/s with 32.3psi vacuum @18:49:39.10 Cartridge 5 sampling 1438/2000ml at 0.21ml/s with 24.9psi vacuum @18:50:38.91 Cartridge 5 sampling 1451/2000ml at 0.27ml/s with 27.6psi vacuum @18:51:39.18 Cartridge 5 sampling 1465/2000ml at 0.28ml/s with 20.5psi vacuum 3.7m deep (20.2psia) @18:52:38.92 Cartridge 5 sampling 1478/2000ml at 0.23ml/s with 26.9psi vacuum @18:53:38.98 Cartridge 5 sampling 1491/2000ml at 0.20ml/s with 26.7psi vacuum @18:54:38.91 Cartridge 5 sampling 1504/2000ml at 0.20ml/s with 29.0psi vacuum @18:55:39.31 Cartridge 5 sampling 1517/2000ml at 0.20ml/s with 23.5psi vacuum @18:56:38.91 Cartridge 5 sampling 1530/2000ml at 0.20ml/s with 29.6psi vacuum @18:57:38.91 Cartridge 5 sampling 1543/2000ml at 0.20ml/s with 27.5psi vacuum @18:58:39.16 Cartridge 5 sampling 1556/2000ml at 0.20ml/s with 30.7psi vacuum @18:59:38.91 Cartridge 5 sampling 1569/2000ml at 0.22ml/s with 32.6psi vacuum @19:00:39.28 Cartridge 5 sampling 1582/2000ml at 0.20ml/s with 29.1psi vacuum @19:01:38.91 Cartridge 5 sampling 1595/2000ml at 0.20ml/s with 32.0psi vacuum @19:02:39.14 Cartridge 5 sampling 1608/2000ml at 0.20ml/s with 32.0psi vacuum @19:03:38.91 Cartridge 5 sampling 1620/2000ml at 0.22ml/s with 27.4psi vacuum @19:04:40.47 Cartridge 5 sampling 1633/2000ml at 0.20ml/s with 33.2psi vacuum @19:05:38.91 Cartridge 5 sampling 1646/2000ml at 0.20ml/s with 33.6psi vacuum @19:06:39.11 Cartridge 5 sampling 1659/2000ml at 0.20ml/s with 34.0psi vacuum @19:07:38.92 Cartridge 5 sampling 1671/2000ml at 0.20ml/s with 34.1psi vacuum @19:08:39.19 Cartridge 5 sampling 1684/2000ml at 0.20ml/s with 34.1psi vacuum @19:09:38.92 Cartridge 5 sampling 1697/2000ml at 0.22ml/s with 26.4psi vacuum @19:10:39.10 Cartridge 5 sampling 1709/2000ml at 0.20ml/s with 34.2psi vacuum @19:11:38.92 Cartridge 5 sampling 1722/2000ml at 0.20ml/s with 35.5psi vacuum @19:12:39.11 Cartridge 5 sampling 1735/2000ml at 0.20ml/s with 34.7psi vacuum @19:13:38.92 Cartridge 5 sampling 1747/2000ml at 0.20ml/s with 28.4psi vacuum @19:14:39.09 Cartridge 5 sampling 1760/2000ml at 0.20ml/s with 33.1psi vacuum @19:15:38.91 Cartridge 5 sampling 1772/2000ml at 0.20ml/s with 27.5psi vacuum @19:16:39.12 Cartridge 5 sampling 1785/2000ml at 0.20ml/s with 35.5psi vacuum @19:17:38.92 Cartridge 5 sampling 1798/2000ml at 0.20ml/s with 37.3psi vacuum @19:18:34.55 Flow averaged 0.31ml/s over 1:37:52 @19:18:34.60 Waiting up to 24 seconds for vacuum to reduce from 32.6 to 18.0psi @19:18:35.28 Relieved excess vacuum in 6.1 seconds @19:18:35.44 SamplePump.seek 2000ml,16:25 @19:18:39.07 Cartridge 5 sampling 1810/2000ml at 0.20ml/s with 21.0psi vacuum 6.5m deep (24.2psia) @19:18:47.82 Flow averaged 0.22ml/s over 12 seconds @19:18:47.89 Waiting up to 8 seconds for vacuum to reduce from 26.2 to 18.0psi @19:18:48.48 Relieved excess vacuum in 1.9 seconds @19:18:48.58 SamplePump.seek 2000ml,16:12 @19:19:38.91 Cartridge 5 sampling 1822/2000ml at 0.25ml/s with 23.5psi vacuum @19:20:30.44 Flow averaged 0.21ml/s over 1:41 @19:20:30.48 Waiting up to 36 seconds for vacuum to reduce from 37.2 to 18.0psi @19:20:31.08 Relieved excess vacuum in 2 seconds @19:20:31.12 SamplePump.seek 2000ml,14:27 @19:20:35.94 Flow estimate reduced to 0.17ml/s @19:20:35.97 Intake.close @19:20:36.11 Exhaust.close @19:20:36.80 Toroid.reconfigure :hold @19:20:37.03 Toroid.seek :bypass @19:20:38.69 Toroid.reconfigure :default @19:20:38.87 Toroid.seek :clear @19:20:39.32 Cartridge 5 sampling 1834/2000ml at 0.00ml/s with 3.9psi vacuum @19:20:40.57 Cartridge::Sampler::Clogged in FILTERING -- Cartridge clogged @19:20:40.59 Pressure.coast @19:20:40.85 Sampled 1833.7ml @19:20:41.36 SamplePump.setPosition! 0ml @19:20:41.41 Plunge.seek :reagent @19:20:41.46 Exhaust.open @19:20:42.22 Intake.open @19:20:42.41 Sample loop closed. Ready to Deliver Reagents @19:20:42.43 Archive.seek :sealed @19:20:43.10 SamplePump.reconfigure :slow @19:20:43.18 SamplePump.seek -35ml @19:20:46.19 Archive.seek :evac @19:21:00.31 delay 1:00 @19:21:58.44 delay 1 second @19:21:59.49 Bypass.open @19:22:00.28 Exhaust.close @19:22:00.46 Archive.seek :clear @19:22:01.29 SamplePump.seek -33ml @19:22:05.95 delay 1 second @19:22:07.04 SamplePump.seek -41ml @19:22:17.84 RNAlater soak for 10:00 @19:22:24.43 delay 1 second @19:22:25.52 Intake.close @19:22:25.59 Bypass.close @19:32:17.95 Archive.seek :tVent @19:32:19.70 delay 1 second @19:32:20.73 Archive.seek :evac @19:32:36.78 delay 10 seconds @19:32:46.84 Sample preserved and evacuated @19:32:47.17 Carousel.seek 5 @19:32:47.39 Plunge.seek :locked @19:32:48.41 Archive.seek :clear @19:32:48.59 Air.seek :clear @19:32:48.79 Lysis1.seek :clear @19:33:05.62 Archive.coast @19:33:05.70 Air.coast @19:33:05.77 Lysis1.coast @19:33:05.84 Lysis2.coast @19:33:05.91 Plunge.coast @19:33:05.99 SamplePump.coast @19:33:06.06 Toroid.coast @19:33:06.13 Carousel.coast @19:33:06.35 Gate.power :motor,:OFF @19:33:06.38 Gate.power :core,:OFF @19:33:06.41 Gate.power :analytic1,:OFF @19:33:06.44 Gate.power :analytic2,:OFF @19:33:06.47 Gate.power :analytic3,:OFF @19:33:06.51 Gate.power :main,:OFF @19:33:06.53 Safely stopped and ready to power off @21:43:22.02 -> Cartrige.release 4 @21:43:22.05 NameError in preston -- uninitialized constant Cartrige @21:43:33.19 -> Cartridge.release 4 @21:43:46.53 -> showlog -200 @21:44:27.65 -> showlog -300 @21:45:20.40 -> slots @21:46:49.48 -> shoot @21:46:49.81 Gate.power :main,:ON,limit to 5000mA @21:46:49.85 Gate.power :analytic1,:ON,limit to 5000mA @21:46:49.94 Gate.power :analytic2,:ON,limit to 5000mA @21:46:49.99 Gate.power :analytic3,:ON,limit to 5000mA @21:46:50.06 Powering core bus @21:46:52.37 Archive.configure :default @21:46:52.54 Air.configure :default @21:46:52.68 Lysis1.configure :default @21:46:52.82 Lysis2.configure :default @21:46:52.99 Plunge.configure :default @21:46:53.15 SamplePump.configure :default @21:46:53.32 ExtraPump.configure :default @21:46:53.49 Toroid.configure :default @21:46:53.65 Carousel.configure :default @21:46:53.82 Pressure.configure :default @21:46:53.90 All elf controllers running firmware version 4.61 @21:46:53.93 Gate.power :motor,:ON,limit to 5000mA @21:46:54.40 Archive.reconfigure :default @21:46:54.63 Plunge.reconfigure :default @21:46:54.65 Air.reconfigure :default @21:46:54.69 Archive.home.jog 0 @21:46:54.73 Toroid.reconfigure :default @21:46:54.78 Lysis2.reconfigure :default @21:46:54.85 Lysis1.reconfigure :default @21:46:55.10 Plunge.home.jog 0 @21:46:55.18 Air.home.jog 0 @21:46:55.23 Toroid.home.jog 0 @21:46:55.27 Lysis2.home.jog 0 @21:46:55.37 Lysis1.home.jog 0 @21:46:57.53 Toroid.reconfigure :homing @21:46:57.70 Lysis2.reconfigure :homing @21:46:57.80 Lysis1.reconfigure :homing @21:46:57.85 Air.reconfigure :homing @21:46:58.10 Plunge.reconfigure :homing @21:46:58.20 Archive.reconfigure :homing @21:46:58.73 Toroid.jog 628 @21:46:58.96 Lysis1.jog 1342 @21:46:59.02 Air.jog 1368 @21:46:59.18 Lysis2.jog 1366 @21:46:59.25 Plunge.jog 860 @21:46:59.50 Archive.jog 158 @21:47:00.53 Archive.setRawPosition! 50 @21:47:00.66 Archive.seek :home @21:47:01.40 Plunge.setRawPosition! 50 @21:47:01.43 Lysis2.setRawPosition! 50 @21:47:01.46 Lysis1.setRawPosition! 50 @21:47:01.49 Air.setRawPosition! 50 @21:47:01.55 Archive.reconfigure :default @21:47:01.78 Plunge.seek :home @21:47:01.83 Lysis2.seek :home @21:47:01.87 Lysis1.seek :home @21:47:01.93 Air.seek :home @21:47:02.21 Toroid.setRawPosition! 0 @21:47:02.29 Toroid.seek :home @21:47:02.49 Plunge.reconfigure :default @21:47:02.54 Lysis2.reconfigure :default @21:47:02.60 Toroid.reconfigure :default @21:47:02.65 Air.reconfigure :default @21:47:02.72 Lysis1.reconfigure :default @21:47:03.32 Toroid.seek :clear @21:47:04.96 Carousel.reconfigure :default @21:47:05.05 Carousel.home.jog 0 @21:47:05.56 Carousel.seek 5 @21:47:05.88 Selecting Cartridge 4 @21:47:06.76 Carousel.seek 4 @21:47:07.59 Lysis1.seek :locked @21:47:11.17 Plunge.seek :locked @21:47:11.34 Air.seek :locked @21:47:15.71 Mounted archive Cartridge @21:47:16.70 Duration of filtering limited to 4:05:00 @21:47:17.61 Priming sample loop w/160ml, bypass w/10ml @21:47:17.64 SamplePump.setPosition! 0ml @21:47:17.81 Bypass.open @21:47:19.55 Intake.open @21:47:20.65 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @21:47:23.25 Vacuum tare is 16% of full scale @21:47:23.28 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @21:47:23.54 Pressure.reconfigure :default @21:47:23.62 Pressure.hold @21:47:23.86 SamplePump.reconfigure :prime @21:47:23.95 SamplePump.seek 10ml,1:21 @21:47:35.10 Pressure.coast @21:47:35.27 SamplePump.setPosition! 0ml @21:47:36.72 Exhaust.open @21:47:38.26 Bypass.close @21:47:40.31 Pressure.reconfigure :default @21:47:40.39 Pressure.hold @21:47:40.70 SamplePump.reconfigure :prime @21:47:40.80 SamplePump.seek 160ml,13:51 @21:50:32.35 Pressure.coast @21:50:32.50 Bypass.open @21:50:33.30 Intake.close @21:50:34.33 Toroid.reconfigure :hold @21:50:34.58 Toroid.seek :crack @21:50:36.24 Toroid.seek :cartridge @21:50:37.20 Toroid.reconfigure :default @21:50:37.29 Toroid.seek :clear @21:50:39.17 Pressurizing cartridge to 26psi above 15.6psia @21:50:39.20 SamplePump.setPosition! 0ml @21:50:39.27 Intake.open @21:50:40.20 Bypass.close @21:50:41.13 Exhaust.close @21:50:42.28 Pressure.reconfigure :default @21:50:42.46 Pressure.hold @21:50:43.06 SamplePump.reconfigure :charge @21:50:43.29 SamplePump.seek 2.3ml,42 seconds @21:50:57.44 Pressure.coast @21:50:57.52 Intake.close @21:50:58.55 Pressurized cartridge with 0.901ml to 26.0psi @21:50:58.58 Isolated @36.8psia (21.1psi) @21:50:58.65 Checking for cartridge leaks from 36.8psia @21:50:58.77 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @21:51:03.62 Cartridge pressure fell 3% to 35.6psia @21:51:03.89 Sampling 2000.0ml @21:51:03.91 SamplePump.setPosition! 0ml @21:51:04.00 Exhaust.open @21:51:04.69 Intake.open @21:51:05.61 Pressure.reconfigure :default @21:51:05.70 Pressure.hold @21:51:05.85 Sea 3.1m deep (19.2psia) @21:51:06.04 SamplePump.reconfigure :sample @21:51:06.13 SamplePump.seek 2000ml,2:47:12 @21:51:08.84 Cartridge 4 sampling 1/2000ml at 0.44ml/s with 25.0psi vacuum @21:52:08.69 Cartridge 4 sampling 27/2000ml at 0.44ml/s with 24.7psi vacuum @21:53:08.91 Cartridge 4 sampling 53/2000ml at 0.44ml/s with 22.9psi vacuum @21:54:08.68 Cartridge 4 sampling 80/2000ml at 0.46ml/s with 24.2psi vacuum @21:55:08.89 Cartridge 4 sampling 107/2000ml at 0.46ml/s with 24.5psi vacuum @21:56:08.68 Cartridge 4 sampling 134/2000ml at 0.46ml/s with 26.8psi vacuum @21:57:09.16 Cartridge 4 sampling 162/2000ml at 0.48ml/s with 25.2psi vacuum @21:58:08.68 Cartridge 4 sampling 190/2000ml at 0.44ml/s with 27.7psi vacuum @21:59:08.88 Cartridge 4 sampling 218/2000ml at 0.48ml/s with 24.1psi vacuum @22:00:08.68 Cartridge 4 sampling 247/2000ml at 0.47ml/s with 26.9psi vacuum @22:01:08.93 Cartridge 4 sampling 275/2000ml at 0.49ml/s with 22.8psi vacuum @22:02:08.69 Cartridge 4 sampling 304/2000ml at 0.48ml/s with 23.7psi vacuum @22:03:08.95 Cartridge 4 sampling 332/2000ml at 0.46ml/s with 19.1psi vacuum @22:04:08.68 Cartridge 4 sampling 360/2000ml at 0.49ml/s with 22.1psi vacuum @22:05:08.85 Cartridge 4 sampling 389/2000ml at 0.46ml/s with 19.1psi vacuum 2.8m deep (18.8psia) @22:06:08.68 Cartridge 4 sampling 417/2000ml at 0.48ml/s with 19.1psi vacuum @22:07:08.84 Cartridge 4 sampling 445/2000ml at 0.49ml/s with 24.0psi vacuum @22:08:08.69 Cartridge 4 sampling 473/2000ml at 0.46ml/s with 27.9psi vacuum @22:09:08.84 Cartridge 4 sampling 501/2000ml at 0.44ml/s with 25.5psi vacuum @22:10:08.69 Cartridge 4 sampling 528/2000ml at 0.46ml/s with 18.9psi vacuum 1.8m deep (17.4psia) @22:11:08.96 Cartridge 4 sampling 556/2000ml at 0.45ml/s with 24.7psi vacuum @22:12:09.19 Cartridge 4 sampling 583/2000ml at 0.43ml/s with 28.1psi vacuum @22:13:08.69 Cartridge 4 sampling 610/2000ml at 0.46ml/s with 23.3psi vacuum @22:14:09.01 Cartridge 4 sampling 637/2000ml at 0.43ml/s with 27.3psi vacuum @22:15:08.68 Cartridge 4 sampling 663/2000ml at 0.42ml/s with 26.0psi vacuum @22:16:09.05 Cartridge 4 sampling 690/2000ml at 0.44ml/s with 21.9psi vacuum @22:17:08.68 Cartridge 4 sampling 716/2000ml at 0.44ml/s with 26.1psi vacuum @22:18:08.91 Cartridge 4 sampling 742/2000ml at 0.40ml/s with 22.0psi vacuum @22:19:08.68 Cartridge 4 sampling 767/2000ml at 0.43ml/s with 26.7psi vacuum @22:20:08.84 Cartridge 4 sampling 793/2000ml at 0.43ml/s with 24.1psi vacuum @22:21:08.69 Cartridge 4 sampling 817/2000ml at 0.42ml/s with 27.5psi vacuum @22:22:08.82 Cartridge 4 sampling 842/2000ml at 0.40ml/s with 23.3psi vacuum @22:23:08.68 Cartridge 4 sampling 867/2000ml at 0.39ml/s with 16.3psi vacuum @22:24:08.86 Cartridge 4 sampling 891/2000ml at 0.39ml/s with 24.4psi vacuum @22:25:08.68 Cartridge 4 sampling 914/2000ml at 0.39ml/s with 28.1psi vacuum @22:26:08.91 Cartridge 4 sampling 938/2000ml at 0.40ml/s with 22.6psi vacuum @22:27:08.69 Cartridge 4 sampling 962/2000ml at 0.37ml/s with 29.2psi vacuum @22:28:08.87 Cartridge 4 sampling 985/2000ml at 0.40ml/s with 23.8psi vacuum @22:29:08.68 Cartridge 4 sampling 1008/2000ml at 0.37ml/s with 27.9psi vacuum @22:30:08.75 Cartridge 4 sampling 1031/2000ml at 0.38ml/s with 20.0psi vacuum 2.2m deep (17.9psia) @22:31:08.68 Cartridge 4 sampling 1053/2000ml at 0.39ml/s with 26.2psi vacuum @22:32:08.72 Cartridge 4 sampling 1076/2000ml at 0.37ml/s with 21.8psi vacuum @22:33:08.68 Cartridge 4 sampling 1097/2000ml at 0.38ml/s with 23.9psi vacuum @22:34:08.92 Cartridge 4 sampling 1119/2000ml at 0.35ml/s with 26.0psi vacuum @22:35:09.12 Cartridge 4 sampling 1141/2000ml at 0.36ml/s with 27.9psi vacuum @22:36:08.68 Cartridge 4 sampling 1162/2000ml at 0.35ml/s with 28.2psi vacuum @22:37:08.90 Cartridge 4 sampling 1183/2000ml at 0.37ml/s with 24.1psi vacuum @22:38:08.69 Cartridge 4 sampling 1204/2000ml at 0.31ml/s with 25.1psi vacuum @22:39:09.21 Cartridge 4 sampling 1225/2000ml at 0.32ml/s with 14.1psi vacuum 5.4m deep (22.6psia) @22:40:08.69 Cartridge 4 sampling 1246/2000ml at 0.32ml/s with 14.2psi vacuum 1.8m deep (17.4psia) @22:41:08.90 Cartridge 4 sampling 1266/2000ml at 0.35ml/s with 23.0psi vacuum @22:42:08.69 Cartridge 4 sampling 1286/2000ml at 0.31ml/s with 17.9psi vacuum @22:43:08.90 Cartridge 4 sampling 1306/2000ml at 0.35ml/s with 25.1psi vacuum @22:44:08.69 Cartridge 4 sampling 1326/2000ml at 0.34ml/s with 20.5psi vacuum @22:45:08.89 Cartridge 4 sampling 1345/2000ml at 0.31ml/s with 28.8psi vacuum @22:46:08.69 Cartridge 4 sampling 1364/2000ml at 0.30ml/s with 19.6psi vacuum @22:47:08.88 Cartridge 4 sampling 1383/2000ml at 0.33ml/s with 23.5psi vacuum @22:48:08.69 Cartridge 4 sampling 1402/2000ml at 0.34ml/s with 22.4psi vacuum @22:49:08.91 Cartridge 4 sampling 1421/2000ml at 0.32ml/s with 19.0psi vacuum 3.5m deep (19.9psia) @22:50:08.68 Cartridge 4 sampling 1439/2000ml at 0.28ml/s with 13.2psi vacuum @22:51:08.90 Cartridge 4 sampling 1458/2000ml at 0.31ml/s with 22.8psi vacuum @22:52:08.69 Cartridge 4 sampling 1476/2000ml at 0.29ml/s with 18.2psi vacuum @22:53:08.85 Cartridge 4 sampling 1494/2000ml at 0.32ml/s with 26.1psi vacuum @22:54:08.68 Cartridge 4 sampling 1511/2000ml at 0.27ml/s with 24.5psi vacuum @22:55:08.82 Cartridge 4 sampling 1529/2000ml at 0.28ml/s with 23.3psi vacuum @22:56:08.69 Cartridge 4 sampling 1546/2000ml at 0.27ml/s with 28.6psi vacuum @22:57:08.84 Cartridge 4 sampling 1563/2000ml at 0.31ml/s with 25.6psi vacuum @22:58:08.78 Cartridge 4 sampling 1580/2000ml at 0.29ml/s with 19.5psi vacuum @22:59:08.78 Cartridge 4 sampling 1596/2000ml at 0.28ml/s with 17.1psi vacuum 4.1m deep (20.7psia) @23:00:09.02 Cartridge 4 sampling 1613/2000ml at 0.27ml/s with 27.4psi vacuum @23:01:08.67 Cartridge 4 sampling 1629/2000ml at 0.25ml/s with 28.1psi vacuum @23:02:09.14 Cartridge 4 sampling 1646/2000ml at 0.28ml/s with 28.2psi vacuum @23:03:08.69 Cartridge 4 sampling 1662/2000ml at 0.27ml/s with 22.4psi vacuum @23:04:08.90 Cartridge 4 sampling 1678/2000ml at 0.29ml/s with 23.2psi vacuum 2.1m deep (17.9psia) @23:05:08.67 Cartridge 4 sampling 1693/2000ml at 0.28ml/s with 28.3psi vacuum @23:06:08.88 Cartridge 4 sampling 1708/2000ml at 0.21ml/s with 27.7psi vacuum @23:07:08.67 Cartridge 4 sampling 1723/2000ml at 0.28ml/s with 23.6psi vacuum @23:08:08.83 Cartridge 4 sampling 1738/2000ml at 0.28ml/s with 26.2psi vacuum @23:09:08.67 Cartridge 4 sampling 1753/2000ml at 0.20ml/s with 27.2psi vacuum @23:10:08.93 Cartridge 4 sampling 1768/2000ml at 0.24ml/s with 26.6psi vacuum @23:11:08.68 Cartridge 4 sampling 1783/2000ml at 0.23ml/s with 18.9psi vacuum @23:12:08.87 Cartridge 4 sampling 1797/2000ml at 0.21ml/s with 20.0psi vacuum @23:13:08.67 Cartridge 4 sampling 1812/2000ml at 0.21ml/s with 16.5psi vacuum @23:14:08.85 Cartridge 4 sampling 1826/2000ml at 0.20ml/s with 28.3psi vacuum @23:15:08.67 Cartridge 4 sampling 1840/2000ml at 0.22ml/s with 16.6psi vacuum @23:16:08.81 Cartridge 4 sampling 1854/2000ml at 0.25ml/s with 15.4psi vacuum @23:17:08.67 Cartridge 4 sampling 1868/2000ml at 0.27ml/s with 17.9psi vacuum @23:18:08.77 Cartridge 4 sampling 1882/2000ml at 0.20ml/s with 18.6psi vacuum @23:19:08.67 Cartridge 4 sampling 1896/2000ml at 0.24ml/s with 23.8psi vacuum @23:20:08.84 Cartridge 4 sampling 1910/2000ml at 0.23ml/s with 31.9psi vacuum @23:21:08.67 Cartridge 4 sampling 1924/2000ml at 0.24ml/s with 23.4psi vacuum @23:22:09.21 Cartridge 4 sampling 1937/2000ml at 0.20ml/s with 28.5psi vacuum @23:23:08.74 Cartridge 4 sampling 1951/2000ml at 0.20ml/s with 28.8psi vacuum @23:24:08.85 Cartridge 4 sampling 1965/2000ml at 0.23ml/s with 21.3psi vacuum 1.8m deep (17.4psia) @23:25:09.00 Cartridge 4 sampling 1978/2000ml at 0.21ml/s with 28.4psi vacuum @23:26:08.70 Cartridge 4 sampling 1992/2000ml at 0.25ml/s with 30.5psi vacuum @23:26:45.77 Pressure.coast @23:26:45.93 Intake.close @23:26:45.98 Exhaust.close @23:26:46.19 Toroid.reconfigure :hold @23:26:46.28 Toroid.seek :bypass @23:26:47.92 Toroid.reconfigure :default @23:26:48.00 Toroid.seek :clear @23:26:49.75 Sampled 2000.0ml @23:26:51.32 SamplePump.setPosition! 0ml @23:26:51.53 Plunge.seek :reagent @23:26:51.69 Exhaust.open @23:26:52.60 Sample loop closed. Ready to Deliver Reagents @23:26:52.63 Archive.seek :sealed @23:26:52.90 Intake.open @23:26:53.89 SamplePump.reconfigure :slow @23:26:54.04 SamplePump.seek -35ml @23:26:56.36 Archive.seek :evac @23:27:10.39 delay 1:00 @23:28:09.25 delay 1 second @23:28:10.27 Bypass.open @23:28:10.44 Archive.seek :clear @23:28:11.02 Exhaust.close @23:28:11.79 SamplePump.seek -33ml @23:28:16.35 delay 1 second @23:28:17.70 SamplePump.seek -41ml @23:28:27.60 RNAlater soak for 10:00 @23:28:35.18 delay 1 second @23:28:36.27 Intake.close @23:28:36.31 Bypass.close @23:38:27.63 Archive.seek :tVent @23:38:29.38 delay 1 second @23:38:30.47 Archive.seek :evac @23:38:46.51 delay 10 seconds @23:38:56.55 Sample preserved and evacuated @23:38:57.13 Carousel.seek 4 @23:38:57.98 Plunge.seek :locked @23:38:59.41 Archive.seek :clear @23:39:01.03 Air.seek :clear @23:39:01.47 Lysis1.seek :clear @23:39:17.95 Archive.coast @23:39:18.09 Air.coast @23:39:18.28 Lysis1.coast @23:39:18.50 Lysis2.coast @23:39:18.66 Plunge.coast @23:39:18.91 SamplePump.coast @23:39:19.07 Toroid.coast @23:39:19.28 Carousel.coast @23:39:19.68 Gate.power :motor,:OFF @23:39:19.73 Gate.power :core,:OFF @23:39:19.79 Gate.power :analytic1,:OFF @23:39:19.85 Gate.power :analytic2,:OFF @23:39:19.90 Gate.power :analytic3,:OFF @23:39:19.95 Gate.power :main,:OFF @23:39:20.02 Safely stopped and ready to power off @01:26:30.76PDT27-May-21 -> Cartridge.dry @01:27:27.31 -> Cartridge.leakyToroid @01:27:45.94 -> slots 8..10=>nil @01:28:10.93 -> slots 8..10=>:dry @01:28:22.55 -> Cartridge.reserve 8..9 @01:28:30.74 -> Cartridge.dry @01:30:05.78 -> delay '3:00:00'; shoot @01:30:05.92 delay 3:00:00 @04:30:06.23 Gate.power :main,:ON,limit to 5000mA @04:30:06.30 Gate.power :analytic1,:ON,limit to 5000mA @04:30:06.38 Gate.power :analytic2,:ON,limit to 5000mA @04:30:06.44 Gate.power :analytic3,:ON,limit to 5000mA @04:30:06.56 Powering core bus @04:30:09.09 Archive.configure :default @04:30:09.25 Air.configure :default @04:30:09.39 Lysis1.configure :default @04:30:09.53 Lysis2.configure :default @04:30:09.69 Plunge.configure :default @04:30:09.86 SamplePump.configure :default @04:30:10.02 ExtraPump.configure :default @04:30:10.18 Toroid.configure :default @04:30:10.34 Carousel.configure :default @04:30:10.51 Pressure.configure :default @04:30:10.59 All elf controllers running firmware version 4.61 @04:30:10.62 Gate.power :motor,:ON,limit to 5000mA @04:30:10.90 Archive.reconfigure :default @04:30:11.13 Archive.home.jog 0 @04:30:11.43 Toroid.reconfigure :default @04:30:11.45 Plunge.reconfigure :default @04:30:11.49 Lysis1.reconfigure :default @04:30:11.54 Air.reconfigure :default @04:30:11.60 Lysis2.reconfigure :default @04:30:11.83 Toroid.home.jog 0 @04:30:11.89 Plunge.home.jog 0 @04:30:11.94 Lysis1.home.jog 0 @04:30:11.97 Air.home.jog 0 @04:30:12.05 Lysis2.home.jog 0 @04:30:12.47 Archive.reconfigure :homing @04:30:12.79 Archive.jog 156 @04:30:12.85 Plunge.reconfigure :homing @04:30:12.90 Lysis2.reconfigure :homing @04:30:12.94 Air.reconfigure :homing @04:30:13.01 Lysis1.reconfigure :homing @04:30:13.06 Toroid.reconfigure :homing @04:30:13.37 Plunge.jog 858 @04:30:13.42 Lysis2.jog 1365 @04:30:13.61 Toroid.jog 628 @04:30:13.76 Lysis1.jog 1347 @04:30:13.82 Air.jog 1369 @04:30:14.70 Archive.setRawPosition! 50 @04:30:15.00 Archive.seek :home @04:30:15.22 Plunge.setRawPosition! 50 @04:30:15.39 Plunge.seek :home @04:30:15.72 Lysis2.setRawPosition! 50 @04:30:15.82 Lysis1.setRawPosition! 50 @04:30:15.94 Air.setRawPosition! 50 @04:30:15.98 Archive.reconfigure :default @04:30:16.38 Lysis2.seek :home @04:30:16.56 Lysis1.seek :home @04:30:16.73 Air.seek :home @04:30:17.41 Toroid.setRawPosition! 0 @04:30:17.49 Plunge.reconfigure :default @04:30:17.59 Lysis1.reconfigure :default @04:30:17.69 Air.reconfigure :default @04:30:17.77 Lysis2.reconfigure :default @04:30:18.32 Toroid.seek :home @04:30:18.66 Toroid.reconfigure :default @04:30:19.00 Toroid.seek :clear @04:30:20.67 Carousel.reconfigure :default @04:30:20.76 Carousel.home.jog 0 @04:30:21.79 Carousel.seek 4 @04:30:22.00 Selecting Cartridge 10 @04:30:22.24 Carousel.seek 10 @04:30:23.81 Lysis1.seek :locked @04:30:27.25 Plunge.seek :locked @04:30:27.73 Air.seek :locked @04:30:32.00 Mounted archive Cartridge @04:30:32.36 Duration of filtering limited to 4:05:00 @04:30:32.84 Priming sample loop w/160ml, bypass w/10ml @04:30:32.85 SamplePump.setPosition! 0ml @04:30:32.93 Bypass.open @04:30:34.62 Intake.open @04:30:35.70 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @04:30:38.43 Vacuum tare is 17% of full scale @04:30:38.56 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @04:30:38.80 Pressure.reconfigure :default @04:30:38.88 Pressure.hold @04:30:39.12 SamplePump.reconfigure :prime @04:30:39.21 SamplePump.seek 10ml,1:21 @04:30:50.35 Pressure.coast @04:30:50.41 SamplePump.setPosition! 0ml @04:30:50.48 Exhaust.open @04:30:51.91 Bypass.close @04:30:53.69 Pressure.reconfigure :default @04:30:53.81 Pressure.hold @04:30:54.58 SamplePump.reconfigure :prime @04:30:54.71 SamplePump.seek 160ml,13:51 @04:30:57.87 SamplePump.coast @04:30:57.94 Intake.close @04:30:58.01 Exhaust.close @04:30:58.80 Pressure.coast @04:30:58.97 Cartridge::Sampler::Leaky in FILTERING -- Serious Leak or Stall -- 0.93ml/s made only 3.7psi vacuum @04:30:58.99 Retry #1 of 2 @04:30:59.02 Cmd::SpareCartridge in FILTERING -- 1 spare remaining @04:30:59.08 Cartridge::None in FILTERING -- No dry cartridges remain @04:30:59.13 Thread::ChildDied in preston -- :FILTERING @06:54:53.90 -> slots @06:55:28.82 -> BACsampler.prime @06:55:29.14 Priming sample loop w/160ml, bypass w/10ml @06:55:29.21 SamplePump.setPosition! 0ml @06:55:29.36 Bypass.open @06:55:31.02 Intake.open @06:55:32.08 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @06:55:34.66 Vacuum tare is 18% of full scale @06:55:34.69 Cartridge::Sampler::Warning in preston -- Vacuum sensor saturates below -18psi @06:55:34.94 Pressure.reconfigure :default @06:55:35.02 Pressure.hold @06:55:35.26 SamplePump.reconfigure :prime @06:55:35.35 SamplePump.seek 10ml,1:21 @06:55:38.58 SamplePump.coast @06:55:38.78 Intake.close @06:55:38.95 Bypass.close @06:55:40.06 Pressure.coast @06:55:40.21 Cartridge::Sampler::Leaky in preston -- Serious Leak or Stall -- 0.93ml/s made only -0.2psi vacuum @06:58:02.73 -> require vacuumCart @06:58:02.76 TypeError in preston -- can't convert Symbol into String @06:58:18.39 -> require 'vacuumCart' @06:58:23.79 -> refillSampleLines @06:58:23.89 Refilling sample & bypass lines with fluid. @06:58:23.96 Waiting for preston to answer "Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return." @06:58:24.13 Answered Hook Intake and Exhaust Lines to MillQ water. Open Pressure Interface Valve. Press Return."" @06:58:24.20 Bypass.open @06:58:25.10 Intake.open @06:58:26.04 SamplePump.setPosition! 0ml @06:58:26.17 SamplePump.reconfigure :fast @06:58:26.39 SamplePump.seek 30ml @06:58:53.06 Bypass.close @06:58:53.83 Exhaust.open @06:58:54.71 SamplePump.seek 160ml @07:00:47.93 delay 1 second @07:00:48.96 Intake.close @07:00:49.54 Exhaust.close @07:00:50.37 Sampling Loop filled with water @07:01:00.19 -> BACsampler.prime @07:01:00.31 Priming sample loop w/160ml, bypass w/10ml @07:01:00.36 SamplePump.setPosition! 0ml @07:01:00.60 Bypass.open @07:01:02.37 Intake.open @07:01:03.44 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @07:01:06.05 Vacuum tare is 18% of full scale @07:01:06.08 Cartridge::Sampler::Warning in preston -- Vacuum sensor saturates below -17psi @07:01:06.33 Pressure.reconfigure :default @07:01:06.42 Pressure.hold @07:01:06.64 SamplePump.reconfigure :prime @07:01:06.74 SamplePump.seek 10ml,1:21 @07:01:17.87 Pressure.coast @07:01:17.93 SamplePump.setPosition! 0ml @07:01:18.01 Exhaust.open @07:01:19.46 Bypass.close @07:01:21.48 Pressure.reconfigure :default @07:01:21.66 Pressure.hold @07:01:22.42 SamplePump.reconfigure :prime @07:01:22.58 SamplePump.seek 160ml,13:51 @07:04:14.13 Pressure.coast @07:04:38.00 -> killSampleLoop! @07:04:38.03 SamplePump.setPosition! 0ml @07:04:38.16 SamplePump.reconfigure :slow @07:04:38.25 SamplePump.seek -35ml @07:05:53.35 delay 1 second @07:05:54.37 Bypass.open @07:05:55.14 Exhaust.close @07:05:56.08 SamplePump.seek -33ml @07:06:00.76 delay 1 second @07:06:01.85 SamplePump.seek -41ml @07:06:19.23 delay 1 second @07:06:20.32 Intake.close @07:06:20.40 Bypass.close @07:06:49.96 -> BACsampler.prime @07:06:50.08 Priming sample loop w/160ml, bypass w/10ml @07:06:50.10 SamplePump.setPosition! 0ml @07:06:50.17 Bypass.open @07:06:52.02 Intake.open @07:06:53.97 Pressure.reconfigure :default @07:06:54.12 Pressure.hold @07:06:54.67 SamplePump.reconfigure :prime @07:06:54.87 SamplePump.seek 10ml,1:21 @07:07:06.03 Pressure.coast @07:07:06.11 SamplePump.setPosition! 0ml @07:07:06.18 Exhaust.open @07:07:07.59 Bypass.close @07:07:09.17 Pressure.reconfigure :default @07:07:09.25 Pressure.hold @07:07:09.54 SamplePump.reconfigure :prime @07:07:09.63 SamplePump.seek 160ml,13:51 @07:10:01.07 Pressure.coast @07:10:21.56 -> Sampler.resset @07:10:21.60 NoMethodError in preston -- undefined method `resset' for Cartridge::Sampler:Class @07:10:30.20 -> Sampler.reset @07:11:02.60 -> slot 10=>nil @07:11:13.67 -> slot 10=>:dry @07:11:50.91 -> Cartridge.release 9 @07:11:58.28 -> Cartridge.dry @07:12:07.41 -> shoot @07:12:07.55 Selecting Cartridge 10 @07:12:09.02 Duration of filtering limited to 4:05:00 @07:12:09.84 Bypass.open @07:12:10.55 Intake.close @07:12:11.45 Toroid.reconfigure :hold @07:12:11.53 Toroid.seek :crack @07:12:13.15 Toroid.seek :cartridge @07:12:14.19 Toroid.reconfigure :default @07:12:14.52 Toroid.seek :clear @07:12:16.88 Pressurizing cartridge to 26psi above 13.8psia @07:12:17.00 SamplePump.setPosition! 0ml @07:12:17.10 Intake.open @07:12:18.02 Bypass.close @07:12:18.84 Exhaust.close @07:12:19.66 Pressure.reconfigure :default @07:12:19.75 Pressure.hold @07:12:19.99 SamplePump.reconfigure :charge @07:12:20.55 SamplePump.seek 2.3ml,42 seconds @07:12:36.69 Pressure.coast @07:12:36.77 Intake.close @07:12:38.02 Pressurized cartridge with 1.186ml to 26.0psi @07:12:38.04 Isolated @36.5psia (22.6psi) @07:12:38.06 Checking for cartridge leaks from 36.5psia @07:12:38.12 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @07:12:43.15 Sampling 2000.0ml @07:12:43.19 SamplePump.setPosition! 0ml @07:12:43.41 Exhaust.open @07:12:44.19 Intake.open @07:12:45.59 Pressure.reconfigure :default @07:12:45.75 Pressure.hold @07:12:46.57 SamplePump.reconfigure :sample @07:12:46.76 SamplePump.seek 2000ml,2:47:12 @07:12:48.49 Cartridge 10 sampling 1/2000ml at 0.63ml/s with 24.7psi vacuum @07:13:48.71 Cartridge 10 sampling 28/2000ml at 0.45ml/s with 25.1psi vacuum @07:14:48.50 Cartridge 10 sampling 55/2000ml at 0.45ml/s with 25.6psi vacuum @07:15:48.71 Cartridge 10 sampling 82/2000ml at 0.44ml/s with 23.3psi vacuum @07:16:48.50 Cartridge 10 sampling 109/2000ml at 0.44ml/s with 25.5psi vacuum @07:17:48.60 Cartridge 10 sampling 135/2000ml at 0.44ml/s with 23.3psi vacuum @07:18:48.49 Cartridge 10 sampling 162/2000ml at 0.45ml/s with 19.7psi vacuum @07:19:48.79 Cartridge 10 sampling 190/2000ml at 0.47ml/s with 24.5psi vacuum @07:20:48.49 Cartridge 10 sampling 218/2000ml at 0.45ml/s with 19.8psi vacuum @07:21:48.64 Cartridge 10 sampling 245/2000ml at 0.46ml/s with 25.0psi vacuum @07:22:48.49 Cartridge 10 sampling 273/2000ml at 0.47ml/s with 24.6psi vacuum @07:23:48.67 Cartridge 10 sampling 300/2000ml at 0.46ml/s with 19.9psi vacuum @07:24:48.49 Cartridge 10 sampling 327/2000ml at 0.43ml/s with 27.0psi vacuum @07:25:48.59 Cartridge 10 sampling 353/2000ml at 0.42ml/s with 19.7psi vacuum @07:26:48.49 Cartridge 10 sampling 380/2000ml at 0.43ml/s with 26.6psi vacuum @07:27:48.54 Cartridge 10 sampling 407/2000ml at 0.45ml/s with 20.0psi vacuum @07:28:48.50 Cartridge 10 sampling 434/2000ml at 0.43ml/s with 23.3psi vacuum @07:29:48.71 Cartridge 10 sampling 460/2000ml at 0.42ml/s with 21.5psi vacuum @07:30:48.83 Cartridge 10 sampling 486/2000ml at 0.43ml/s with 23.6psi vacuum @07:31:48.49 Cartridge 10 sampling 512/2000ml at 0.42ml/s with 28.3psi vacuum @07:32:48.73 Cartridge 10 sampling 538/2000ml at 0.44ml/s with 26.5psi vacuum @07:33:48.49 Cartridge 10 sampling 565/2000ml at 0.44ml/s with 24.1psi vacuum @07:34:48.77 Cartridge 10 sampling 590/2000ml at 0.40ml/s with 27.7psi vacuum @07:35:48.49 Cartridge 10 sampling 616/2000ml at 0.41ml/s with 27.1psi vacuum @07:36:48.69 Cartridge 10 sampling 641/2000ml at 0.41ml/s with 26.4psi vacuum @07:37:48.49 Cartridge 10 sampling 666/2000ml at 0.43ml/s with 20.7psi vacuum @07:38:48.64 Cartridge 10 sampling 691/2000ml at 0.43ml/s with 20.7psi vacuum 2.2m deep (17.9psia) @07:39:48.49 Cartridge 10 sampling 716/2000ml at 0.41ml/s with 26.6psi vacuum @07:40:48.75 Cartridge 10 sampling 740/2000ml at 0.40ml/s with 27.7psi vacuum @07:41:48.50 Cartridge 10 sampling 765/2000ml at 0.40ml/s with 26.7psi vacuum @07:42:48.67 Cartridge 10 sampling 789/2000ml at 0.40ml/s with 27.0psi vacuum @07:43:48.50 Cartridge 10 sampling 812/2000ml at 0.39ml/s with 27.8psi vacuum @07:44:48.76 Cartridge 10 sampling 836/2000ml at 0.38ml/s with 16.9psi vacuum 4.2m deep (21.0psia) @07:45:48.49 Cartridge 10 sampling 859/2000ml at 0.38ml/s with 18.3psi vacuum @07:46:48.64 Cartridge 10 sampling 883/2000ml at 0.37ml/s with 21.4psi vacuum @07:47:48.50 Cartridge 10 sampling 906/2000ml at 0.40ml/s with 22.6psi vacuum @07:48:48.63 Cartridge 10 sampling 929/2000ml at 0.38ml/s with 18.0psi vacuum 2.3m deep (18.1psia) @07:49:48.49 Cartridge 10 sampling 951/2000ml at 0.38ml/s with 21.5psi vacuum @07:50:48.66 Cartridge 10 sampling 974/2000ml at 0.39ml/s with 21.5psi vacuum @07:51:48.49 Cartridge 10 sampling 996/2000ml at 0.39ml/s with 24.4psi vacuum @07:52:48.90 Cartridge 10 sampling 1018/2000ml at 0.35ml/s with 16.0psi vacuum 5.4m deep (22.7psia) @07:53:48.69 Cartridge 10 sampling 1041/2000ml at 0.36ml/s with 22.7psi vacuum @07:54:48.53 Cartridge 10 sampling 1063/2000ml at 0.37ml/s with 26.9psi vacuum @07:55:48.77 Cartridge 10 sampling 1085/2000ml at 0.37ml/s with 26.7psi vacuum @07:56:48.50 Cartridge 10 sampling 1106/2000ml at 0.34ml/s with 28.7psi vacuum @07:57:48.71 Cartridge 10 sampling 1128/2000ml at 0.34ml/s with 15.0psi vacuum 5.9m deep (23.3psia) @07:58:48.49 Cartridge 10 sampling 1149/2000ml at 0.34ml/s with 28.7psi vacuum @07:59:48.49 Cartridge 10 sampling 1170/2000ml at 0.37ml/s with 22.5psi vacuum @08:00:48.65 Cartridge 10 sampling 1191/2000ml at 0.36ml/s with 25.1psi vacuum @08:01:48.83 Cartridge 10 sampling 1212/2000ml at 0.36ml/s with 23.4psi vacuum @08:02:48.49 Cartridge 10 sampling 1232/2000ml at 0.36ml/s with 21.8psi vacuum @08:03:48.83 Cartridge 10 sampling 1253/2000ml at 0.30ml/s with 19.2psi vacuum 3.3m deep (19.5psia) @08:04:48.50 Cartridge 10 sampling 1273/2000ml at 0.35ml/s with 26.3psi vacuum @08:05:48.68 Cartridge 10 sampling 1293/2000ml at 0.35ml/s with 24.6psi vacuum @08:06:48.50 Cartridge 10 sampling 1313/2000ml at 0.34ml/s with 22.0psi vacuum @08:07:48.65 Cartridge 10 sampling 1333/2000ml at 0.35ml/s with 24.2psi vacuum @08:08:48.49 Cartridge 10 sampling 1352/2000ml at 0.34ml/s with 25.8psi vacuum @08:09:48.60 Cartridge 10 sampling 1371/2000ml at 0.31ml/s with 17.1psi vacuum 4.0m deep (20.7psia) @08:10:48.49 Cartridge 10 sampling 1390/2000ml at 0.31ml/s with 28.2psi vacuum @08:11:48.70 Cartridge 10 sampling 1409/2000ml at 0.30ml/s with 15.2psi vacuum 2.9m deep (19.0psia) @08:12:48.50 Cartridge 10 sampling 1428/2000ml at 0.35ml/s with 26.5psi vacuum @08:13:48.64 Cartridge 10 sampling 1447/2000ml at 0.29ml/s with 14.6psi vacuum 5.4m deep (22.7psia) @08:14:48.49 Cartridge 10 sampling 1466/2000ml at 0.27ml/s with 28.4psi vacuum @08:15:48.66 Cartridge 10 sampling 1484/2000ml at 0.32ml/s with 26.8psi vacuum @08:16:48.49 Cartridge 10 sampling 1502/2000ml at 0.28ml/s with 28.1psi vacuum @08:17:48.62 Cartridge 10 sampling 1520/2000ml at 0.33ml/s with 26.3psi vacuum @08:18:48.49 Cartridge 10 sampling 1538/2000ml at 0.32ml/s with 27.8psi vacuum @08:19:48.61 Cartridge 10 sampling 1556/2000ml at 0.30ml/s with 28.4psi vacuum @08:20:48.49 Cartridge 10 sampling 1574/2000ml at 0.31ml/s with 19.7psi vacuum @08:21:48.79 Cartridge 10 sampling 1591/2000ml at 0.28ml/s with 29.4psi vacuum @08:22:48.90 Cartridge 10 sampling 1608/2000ml at 0.24ml/s with 27.4psi vacuum @08:23:48.89 Cartridge 10 sampling 1625/2000ml at 0.23ml/s with 26.1psi vacuum @08:24:48.80 Cartridge 10 sampling 1642/2000ml at 0.29ml/s with 28.1psi vacuum @08:25:48.64 Cartridge 10 sampling 1659/2000ml at 0.25ml/s with 28.0psi vacuum @08:26:48.80 Cartridge 10 sampling 1676/2000ml at 0.27ml/s with 14.7psi vacuum 4.2m deep (20.9psia) @08:27:48.49 Cartridge 10 sampling 1692/2000ml at 0.22ml/s with 26.7psi vacuum @08:28:48.81 Cartridge 10 sampling 1708/2000ml at 0.27ml/s with 16.7psi vacuum 5.9m deep (23.5psia) @08:29:48.49 Cartridge 10 sampling 1724/2000ml at 0.25ml/s with 25.7psi vacuum @08:30:48.74 Cartridge 10 sampling 1740/2000ml at 0.27ml/s with 20.3psi vacuum 1.9m deep (17.5psia) @08:31:48.49 Cartridge 10 sampling 1756/2000ml at 0.24ml/s with 28.4psi vacuum @08:32:48.68 Cartridge 10 sampling 1772/2000ml at 0.26ml/s with 18.5psi vacuum @08:33:48.49 Cartridge 10 sampling 1787/2000ml at 0.26ml/s with 19.0psi vacuum @08:34:48.67 Cartridge 10 sampling 1803/2000ml at 0.21ml/s with 26.9psi vacuum @08:35:48.48 Cartridge 10 sampling 1818/2000ml at 0.23ml/s with 24.7psi vacuum @08:36:48.81 Cartridge 10 sampling 1833/2000ml at 0.25ml/s with 14.3psi vacuum 7.3m deep (25.4psia) @08:37:48.48 Cartridge 10 sampling 1848/2000ml at 0.24ml/s with 27.7psi vacuum @08:38:48.70 Cartridge 10 sampling 1863/2000ml at 0.28ml/s with 21.4psi vacuum 3.1m deep (19.2psia) @08:39:48.48 Cartridge 10 sampling 1878/2000ml at 0.29ml/s with 21.5psi vacuum @08:40:48.72 Cartridge 10 sampling 1893/2000ml at 0.23ml/s with 30.4psi vacuum @08:41:48.48 Cartridge 10 sampling 1908/2000ml at 0.28ml/s with 26.8psi vacuum @08:42:48.65 Cartridge 10 sampling 1923/2000ml at 0.26ml/s with 27.3psi vacuum @08:43:48.48 Cartridge 10 sampling 1937/2000ml at 0.27ml/s with 23.2psi vacuum @08:44:48.68 Cartridge 10 sampling 1952/2000ml at 0.28ml/s with 23.0psi vacuum @08:45:48.48 Cartridge 10 sampling 1966/2000ml at 0.22ml/s with 15.4psi vacuum @08:46:48.71 Cartridge 10 sampling 1981/2000ml at 0.27ml/s with 23.7psi vacuum 1.6m deep (17.1psia) @08:47:48.48 Cartridge 10 sampling 1995/2000ml at 0.23ml/s with 31.1psi vacuum @08:48:10.39 Pressure.coast @08:48:10.78 Intake.close @08:48:10.97 Exhaust.close @08:48:11.49 Toroid.reconfigure :hold @08:48:11.61 Toroid.seek :bypass @08:48:13.26 Toroid.reconfigure :default @08:48:13.35 Toroid.seek :clear @08:48:14.98 Sampled 2000.0ml @08:48:15.40 Cartridge::Error in PROCESSING -- Cannot process leakyToroid cartridge @08:48:15.46 Thread::ChildDied in preston -- :PROCESSING @09:20:32.11 -> slots @09:21:08.29 -> slot 10=>:archive_bac @09:21:19.00 -> Cmd.process @09:21:19.03 Cmd::Error in preston -- Cannot process while in PROCESS_FAILED state @09:21:36.10 -> archiveEvacAUV @09:21:36.19 Plunge.seek :reagent @09:21:37.30 Sample loop closed. Ready to Deliver Reagents @09:21:37.32 Archive.seek :sealed @09:21:41.01 Archive.seek :evac @09:21:54.69 Slide::Error in preston -- Archive Syringe (ArS) positionErr at 29uL (sensed 57uL) @09:21:54.83 Retry #1 of 2 @09:21:54.89 Archive.hold @09:21:55.23 Archive.reconfigure :default @09:21:55.35 Archive.home.seek :evac @09:21:55.98 Slide::Error in preston -- Archive Syringe (ArS) positionErr at 26uL (sensed 55uL) @09:21:56.00 Retry #2 of 2 @09:21:56.03 Archive.reconfigure :default @09:21:56.13 Archive.hold @09:21:56.42 Archive.reconfigure :default @09:21:56.50 Archive.home.seek :evac @09:21:57.15 Slide::Error in preston -- IGNORED Archive Syringe (ArS) positionErr at 25uL (sensed 55uL) @09:21:57.18 Archive.reconfigure :default @09:21:57.27 Archive.hold @09:21:57.56 Archive.reconfigure :default @09:21:57.64 Archive.home.jog 0 @09:21:57.78 delay 1:00 @09:22:57.83 Archive.seek :clear @09:23:14.58 RNAlater soak for 10:00 @09:33:14.66 Archive.seek :tVent @09:33:16.33 delay 1 second @09:33:17.36 Archive.seek :evac @09:33:33.00 Slide::Error in preston -- Archive Syringe (ArS) positionErr at 24uL (sensed 53uL) @09:33:33.02 Retry #1 of 2 @09:33:33.05 Archive.hold @09:33:33.40 Archive.reconfigure :default @09:33:33.58 Archive.home.seek :evac @09:33:34.33 Slide::Error in preston -- Archive Syringe (ArS) positionErr at 21uL (sensed 52uL) @09:33:34.35 Retry #2 of 2 @09:33:34.38 Archive.reconfigure :default @09:33:34.47 Archive.hold @09:33:34.76 Archive.reconfigure :default @09:33:34.85 Archive.home.seek :evac @09:33:35.50 Slide::Error in preston -- IGNORED Archive Syringe (ArS) positionErr at 22uL (sensed 52uL) @09:33:35.52 Archive.reconfigure :default @09:33:35.62 Archive.hold @09:33:35.93 Archive.reconfigure :default @09:33:36.01 Archive.home.jog 0 @09:33:36.14 delay 10 seconds @09:33:46.18 Sample preserved and evacuated @09:58:52.49 -> Cartridge.dry @09:59:04.89 -> Cartridge.release 8 @09:59:23.88 -> slot 8..9=>:archiveBac @09:59:23.93 Cartridge::InvalidTypeOrState in preston -- :archiveBac is neither type nor state @09:59:41.25 -> slot 8..9=>:archive_bac @10:00:33.09 -> slots @10:00:58.41 -> Cmd.stop @10:00:59.08 Carousel.seek 10 @10:00:59.67 Plunge.seek :locked @10:01:00.87 Archive.seek :clear @10:01:01.50 Air.seek :clear @10:01:01.69 Lysis1.seek :clear @10:01:17.80 Archive.coast @10:01:18.00 Air.coast @10:01:18.18 Lysis1.coast @10:01:18.41 Lysis2.coast @10:01:18.52 Plunge.coast @10:01:18.73 SamplePump.coast @10:01:18.79 Toroid.coast @10:01:18.99 Carousel.coast @10:01:19.36 Gate.power :motor,:OFF @10:01:19.38 Gate.power :core,:OFF @10:01:19.41 Gate.power :analytic1,:OFF @10:01:19.45 Gate.power :analytic2,:OFF @10:01:19.50 Gate.power :analytic3,:OFF @10:01:19.61 Gate.power :main,:OFF @10:01:19.66 Safely stopped and ready to power off @13:18:48.97 -> Cartridge.dry @13:19:00.86 -> Cartridge.reserve 9..8 @13:19:06.43 -> slots @13:19:22.47 -> require 'vacuumCart' @13:19:37.67 -> Cartridge.dry @13:20:41.19 -> slots 30..25=>[:dry,:archiveNull_bac] @13:23:07.71 -> 6.times {shoot; delay 120} @13:23:07.88 Gate.power :main,:ON,limit to 5000mA @13:23:07.89 Gate.power :analytic1,:ON,limit to 5000mA @13:23:07.92 Gate.power :analytic2,:ON,limit to 5000mA @13:23:07.96 Gate.power :analytic3,:ON,limit to 5000mA @13:23:08.03 Powering core bus @13:23:10.83 Archive.configure :default @13:23:11.31 Air.configure :default @13:23:11.67 Lysis1.configure :default @13:23:12.14 Lysis2.configure :default @13:23:12.46 Plunge.configure :default @13:23:12.84 SamplePump.configure :default @13:23:13.25 ExtraPump.configure :default @13:23:13.62 Toroid.configure :default @13:23:13.94 Carousel.configure :default @13:23:14.34 Pressure.configure :default @13:23:14.55 All elf controllers running firmware version 4.61 @13:23:14.60 Gate.power :motor,:ON,limit to 5000mA @13:23:15.48 Toroid.reconfigure :default @13:23:15.57 Air.reconfigure :default @13:23:15.62 Plunge.reconfigure :default @13:23:15.69 Lysis2.reconfigure :default @13:23:15.74 Lysis1.reconfigure :default @13:23:15.80 Archive.reconfigure :default @13:23:16.06 Toroid.home.jog 0 @13:23:16.13 Air.home.jog 0 @13:23:16.17 Lysis2.home.jog 0 @13:23:16.24 Plunge.home.jog 0 @13:23:16.29 Lysis1.home.jog 0 @13:23:16.33 Archive.home.jog 0 @13:23:17.11 Plunge.reconfigure :homing @13:23:17.16 Lysis2.reconfigure :homing @13:23:17.21 Lysis1.reconfigure :homing @13:23:17.27 Air.reconfigure :homing @13:23:17.31 Toroid.reconfigure :homing @13:23:17.39 Archive.reconfigure :homing @13:23:17.68 Plunge.jog 859 @13:23:17.71 Lysis2.jog 1359 @13:23:17.76 Lysis1.jog 1351 @13:23:17.81 Air.jog 1371 @13:23:17.86 Toroid.jog 626 @13:23:17.89 Archive.jog 179 @13:23:18.59 Archive.setRawPosition! 50 @13:23:18.66 Archive.seek :home @13:23:18.86 Plunge.setRawPosition! 50 @13:23:18.94 Plunge.seek :home @13:23:19.06 Archive.reconfigure :default @13:23:19.20 Lysis2.setRawPosition! 50 @13:23:19.25 Lysis1.setRawPosition! 50 @13:23:19.32 Air.setRawPosition! 50 @13:23:19.98 Lysis2.seek :home @13:23:20.10 Lysis1.seek :home @13:23:20.28 Air.seek :home @13:23:20.98 Toroid.setRawPosition! 0 @13:23:21.15 Plunge.reconfigure :default @13:23:21.22 Lysis1.reconfigure :default @13:23:21.30 Air.reconfigure :default @13:23:21.38 Lysis2.reconfigure :default @13:23:21.77 Toroid.seek :home @13:23:22.45 Toroid.reconfigure :default @13:23:24.17 Toroid.seek :clear @13:23:25.85 Carousel.reconfigure :default @13:23:25.94 Carousel.home.jog 0 @13:23:26.45 Carousel.seek 10 @13:23:26.64 Selecting Cartridge 30 @13:23:26.91 Carousel.seek 30 @13:23:31.56 Lysis1.seek :locked @13:23:34.92 Plunge.seek :locked @13:23:35.10 Air.seek :locked @13:23:39.38 Mounted archive Cartridge @13:23:40.78 Duration of filtering limited to 4:05:00 @13:23:41.60 Priming sample loop w/160ml, bypass w/10ml @13:23:41.77 SamplePump.setPosition! 0ml @13:23:41.93 Bypass.open @13:23:43.53 Intake.open @13:23:44.60 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @13:23:47.18 Vacuum tare is 18% of full scale @13:23:47.21 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @13:23:47.47 Pressure.reconfigure :default @13:23:47.56 Pressure.hold @13:23:47.80 SamplePump.reconfigure :prime @13:23:47.89 SamplePump.seek 10ml,1:21 @13:23:59.08 Pressure.coast @13:23:59.24 SamplePump.setPosition! 0ml @13:23:59.42 Exhaust.open @13:24:01.11 Bypass.close @13:24:03.11 Pressure.reconfigure :default @13:24:03.19 Pressure.hold @13:24:03.50 SamplePump.reconfigure :prime @13:24:03.59 SamplePump.seek 160ml,13:51 @13:26:55.31 Pressure.coast @13:26:55.50 Bypass.open @13:26:56.31 Intake.close @13:26:57.60 Toroid.reconfigure :hold @13:26:57.68 Toroid.seek :crack @13:26:59.30 Toroid.seek :cartridge @13:27:00.25 Toroid.reconfigure :default @13:27:00.35 Toroid.seek :clear @13:27:02.56 Pressurizing cartridge to 26psi above 14.4psia @13:27:02.63 SamplePump.setPosition! 0ml @13:27:02.84 Intake.open @13:27:03.83 Bypass.close @13:27:04.67 Exhaust.close @13:27:05.77 Pressure.reconfigure :default @13:27:05.98 Pressure.hold @13:27:06.49 SamplePump.reconfigure :charge @13:27:06.63 SamplePump.seek 2.3ml,42 seconds @13:27:16.11 Pressure.coast @13:27:16.21 Intake.close @13:27:17.18 Pressurized cartridge with 0.400ml to 26.2psi @13:27:17.20 Isolated @32.2psia (17.7psi) @13:27:17.23 Checking for cartridge leaks from 32.2psia @13:27:17.29 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @13:27:21.97 Cartridge pressure fell 1% to 31.8psia @13:27:22.38 Sampling 2000.0ml @13:27:22.42 SamplePump.setPosition! 0ml @13:27:22.68 Exhaust.open @13:27:23.66 Intake.open @13:27:24.92 Pressure.reconfigure :default @13:27:25.09 Pressure.hold @13:27:25.79 SamplePump.reconfigure :sample @13:27:26.01 SamplePump.seek 2000ml,2:47:12 @13:27:27.85 Cartridge 30 sampling 1/2000ml at 0.63ml/s with 14.7psi vacuum @13:28:28.04 Cartridge 30 sampling 39/2000ml at 0.63ml/s with 18.7psi vacuum @13:29:27.84 Cartridge 30 sampling 76/2000ml at 0.63ml/s with 17.4psi vacuum @13:30:27.99 Cartridge 30 sampling 114/2000ml at 0.63ml/s with 14.1psi vacuum @13:31:27.84 Cartridge 30 sampling 152/2000ml at 0.63ml/s with 12.8psi vacuum @13:32:28.01 Cartridge 30 sampling 190/2000ml at 0.63ml/s with 15.3psi vacuum @13:33:27.85 Cartridge 30 sampling 228/2000ml at 0.63ml/s with 16.6psi vacuum @13:34:28.09 Cartridge 30 sampling 266/2000ml at 0.63ml/s with 16.9psi vacuum @13:35:27.84 Cartridge 30 sampling 304/2000ml at 0.63ml/s with 11.1psi vacuum @13:36:28.03 Cartridge 30 sampling 342/2000ml at 0.63ml/s with 17.9psi vacuum @13:37:27.85 Cartridge 30 sampling 380/2000ml at 0.63ml/s with 18.1psi vacuum @13:38:28.05 Cartridge 30 sampling 418/2000ml at 0.63ml/s with 11.0psi vacuum @13:39:27.85 Cartridge 30 sampling 456/2000ml at 0.63ml/s with 12.0psi vacuum @13:40:27.88 Cartridge 30 sampling 494/2000ml at 0.63ml/s with 14.9psi vacuum @13:41:27.84 Cartridge 30 sampling 531/2000ml at 0.63ml/s with 11.2psi vacuum @13:42:29.08 Cartridge 30 sampling 569/2000ml at 0.63ml/s with 16.5psi vacuum @13:43:27.84 Cartridge 30 sampling 607/2000ml at 0.63ml/s with 15.1psi vacuum @13:44:28.15 Cartridge 30 sampling 645/2000ml at 0.63ml/s with 17.6psi vacuum @13:45:28.02 Cartridge 30 sampling 683/2000ml at 0.63ml/s with 12.5psi vacuum @13:46:17.57 SamplePump.coast @13:46:17.76 Intake.close @13:46:17.86 Exhaust.close @13:46:18.58 Toroid.reconfigure :hold @13:46:18.77 Toroid.seek :bypass @13:46:20.42 Toroid.reconfigure :default @13:46:20.51 Toroid.seek :clear @13:46:22.14 Pressure.coast @13:46:22.47 Cartridge::Sampler::Leaky in FILTERING -- Serious Leak or Stall -- 0.63ml/s made only 11.8psi vacuum @13:46:22.49 Retry #1 of 2 @13:46:22.52 Cmd::SpareCartridge in FILTERING -- 0 spares remaining @13:46:22.59 Selecting Cartridge 29 @13:46:23.19 Air.seek :clear @13:46:27.34 Lysis1.seek :clear @13:46:27.80 Plunge.seek :clear @13:46:30.65 Carousel.seek 29 @13:46:31.19 Lysis1.seek :locked @13:46:34.77 Plunge.seek :locked @13:46:35.18 Air.seek :locked @13:46:39.53 Mounted archive Cartridge @13:46:39.62 Duration of filtering limited to 4:05:00 @13:46:39.93 Priming sample loop w/160ml, bypass w/10ml @13:46:39.95 SamplePump.setPosition! 0ml @13:46:40.02 Bypass.open @13:46:41.51 Intake.open @13:46:42.75 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @13:46:45.51 Vacuum tare is 18% of full scale @13:46:45.54 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @13:46:46.11 Pressure.reconfigure :default @13:46:46.21 Pressure.hold @13:46:46.74 SamplePump.reconfigure :prime @13:46:46.83 SamplePump.seek 10ml,1:21 @13:46:58.01 Pressure.coast @13:46:58.08 SamplePump.setPosition! 0ml @13:46:58.15 Exhaust.open @13:46:59.58 Bypass.close @13:47:01.41 Pressure.reconfigure :default @13:47:01.67 Pressure.hold @13:47:02.76 SamplePump.reconfigure :prime @13:47:03.04 SamplePump.seek 160ml,13:51 @13:49:54.66 Pressure.coast @13:49:54.75 Bypass.open @13:49:55.48 Intake.close @13:49:56.82 Toroid.reconfigure :hold @13:49:57.01 Toroid.seek :crack @13:49:58.86 Toroid.seek :cartridge @13:50:00.00 Toroid.reconfigure :default @13:50:00.14 Toroid.seek :clear @13:50:02.15 Pressurizing cartridge to 26psi above 14.4psia @13:50:02.18 SamplePump.setPosition! 0ml @13:50:02.25 Intake.open @13:50:03.02 Bypass.close @13:50:03.76 Exhaust.close @13:50:04.63 Pressure.reconfigure :default @13:50:04.71 Pressure.hold @13:50:04.96 SamplePump.reconfigure :charge @13:50:05.05 SamplePump.seek 2.3ml,42 seconds @13:50:14.48 Pressure.coast @13:50:14.56 Intake.close @13:50:15.71 Pressurized cartridge with 0.403ml to 26.1psi @13:50:15.74 Isolated @34.1psia (19.7psi) @13:50:16.04 Checking for cartridge leaks from 34.1psia @13:50:16.16 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @13:50:20.78 Cartridge pressure fell 2% to 33.4psia @13:50:20.89 Sampling 2000.0ml @13:50:20.91 SamplePump.setPosition! 0ml @13:50:20.98 Exhaust.open @13:50:21.72 Intake.open @13:50:22.66 Pressure.reconfigure :default @13:50:22.74 Pressure.hold @13:50:23.04 SamplePump.reconfigure :sample @13:50:23.13 SamplePump.seek 2000ml,2:47:12 @13:50:26.25 Cartridge 29 sampling 1/2000ml at 0.63ml/s with 18.2psi vacuum @13:51:25.72 Cartridge 29 sampling 39/2000ml at 0.63ml/s with 17.9psi vacuum @13:52:26.23 Cartridge 29 sampling 77/2000ml at 0.63ml/s with 15.8psi vacuum @13:53:25.72 Cartridge 29 sampling 115/2000ml at 0.63ml/s with 10.9psi vacuum 1.7m deep (17.2psia) @13:54:25.97 Cartridge 29 sampling 153/2000ml at 0.63ml/s with 13.2psi vacuum 1.9m deep (17.5psia) @13:55:25.72 Cartridge 29 sampling 191/2000ml at 0.63ml/s with 17.0psi vacuum @13:56:25.89 Cartridge 29 sampling 229/2000ml at 0.63ml/s with 13.0psi vacuum 2.5m deep (18.4psia) @13:57:25.72 Cartridge 29 sampling 267/2000ml at 0.63ml/s with 16.4psi vacuum @13:58:25.95 Cartridge 29 sampling 305/2000ml at 0.63ml/s with 12.1psi vacuum @13:59:25.82 Cartridge 29 sampling 343/2000ml at 0.63ml/s with 12.8psi vacuum @14:00:25.98 Cartridge 29 sampling 381/2000ml at 0.63ml/s with 16.1psi vacuum @14:01:25.72 Cartridge 29 sampling 418/2000ml at 0.63ml/s with 13.0psi vacuum @14:02:25.93 Cartridge 29 sampling 456/2000ml at 0.63ml/s with 12.3psi vacuum 2.5m deep (18.4psia) @14:03:25.72 Cartridge 29 sampling 494/2000ml at 0.63ml/s with 14.4psi vacuum @14:04:25.86 Cartridge 29 sampling 532/2000ml at 0.63ml/s with 17.7psi vacuum @14:05:25.71 Cartridge 29 sampling 570/2000ml at 0.63ml/s with 18.0psi vacuum @14:06:25.89 Cartridge 29 sampling 608/2000ml at 0.63ml/s with 12.3psi vacuum @14:07:25.72 Cartridge 29 sampling 646/2000ml at 0.63ml/s with 13.9psi vacuum @14:07:29.92 SamplePump.coast @14:07:30.08 Intake.close @14:07:30.22 Exhaust.close @14:07:30.76 Toroid.reconfigure :hold @14:07:30.89 Toroid.seek :bypass @14:07:32.54 Toroid.reconfigure :default @14:07:32.63 Toroid.seek :clear @14:07:34.26 Pressure.coast @14:07:34.60 Cartridge::Sampler::Leaky in FILTERING -- Serious Leak or Stall -- 0.63ml/s made only 16.3psi vacuum @14:07:34.62 Retry #2 of 2 @14:07:34.66 Cmd::Error in FILTERING -- No spare cartridges remain @14:07:34.71 Thread::ChildDied in preston -- :FILTERING @15:13:25.49 -> BACsampler.prime @15:13:25.59 Priming sample loop w/160ml, bypass w/10ml @15:13:25.62 SamplePump.setPosition! 0ml @15:13:25.70 Bypass.open @15:13:27.27 Intake.open @15:13:28.97 Pressure.reconfigure :default @15:13:29.06 Pressure.hold @15:13:29.63 SamplePump.reconfigure :prime @15:13:29.93 SamplePump.seek 10ml,1:21 @15:13:41.18 Pressure.coast @15:13:41.35 SamplePump.setPosition! 0ml @15:13:41.49 Exhaust.open @15:13:43.07 Bypass.close @15:13:44.73 Pressure.reconfigure :default @15:13:44.82 Pressure.hold @15:13:45.12 SamplePump.reconfigure :prime @15:13:45.21 SamplePump.seek 160ml,13:51 @15:16:36.83 Pressure.coast @15:16:59.16 -> Cartridge.dry @15:18:07.92 -> PhytoSampler.vacuum @15:18:37.54 -> BACsampler.vacuum @15:19:38.63 -> slots 25..28=>:archiveNull_phyto @15:19:38.73 Cartridge::InvalidTypeOrState in preston -- :archiveNull_phyto is neither type nor state @15:20:29.92 -> slots 25..28=>:archiveNull_phyto @15:20:30.09 Cartridge::InvalidTypeOrState in preston -- :archiveNull_phyto is neither type nor state @15:21:25.28 -> slots 25..28=>[:archiveNull_phyto, :dry] @15:21:25.42 Cartridge::TypeUnknown in preston -- :archiveNull_phyto is not aRNAl_bac, aRNAl_phyto, aRNAl_zoo, archiveHiBiomass_bac, archiveNull_bac, archive_bac, archive_phyto, archive_zoo, daSPR_bac, daSPR_phyto, daSPR_phyto9, daSPR_zoo, lyseHold_bac, lyseHold_phyto, lyseHold_zoo, lyseNull_bac, microSPR_bac, microSPR_phyto, microSPR_zoo or unknown @15:22:01.20 -> BACsampler.vacuum @15:22:48.53 -> PhytoSampler.vacuum @15:23:02.21 -> BACsampler.vacuum=10..14 @15:23:17.15 -> Cartridge.type @15:23:26.07 -> slots @15:25:47.37 -> 4.times {shoot; delay 120} @15:25:47.43 Selecting Cartridge 28 @15:25:47.84 Air.seek :clear @15:25:51.77 Lysis1.seek :clear @15:25:52.11 Plunge.seek :clear @15:25:55.15 Carousel.seek 28 @15:25:55.70 Lysis1.seek :locked @15:25:59.34 Plunge.seek :locked @15:25:59.75 Air.seek :locked @15:26:04.06 Mounted archive Cartridge @15:26:04.39 Duration of filtering limited to 4:05:00 @15:26:04.71 Priming sample loop w/160ml, bypass w/10ml @15:26:04.74 SamplePump.setPosition! 0ml @15:26:04.81 Exhaust.close @15:26:05.52 Bypass.open @15:26:07.14 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:26:09.83 Vacuum tare is 17% of full scale @15:26:09.91 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @15:26:10.52 Pressure.reconfigure :default @15:26:10.65 Pressure.hold @15:26:11.18 SamplePump.reconfigure :prime @15:26:11.42 SamplePump.seek 10ml,1:21 @15:26:22.65 Pressure.coast @15:26:22.73 SamplePump.setPosition! 0ml @15:26:22.80 Exhaust.open @15:26:24.20 Bypass.close @15:26:25.79 Pressure.reconfigure :default @15:26:25.88 Pressure.hold @15:26:26.18 SamplePump.reconfigure :prime @15:26:26.27 SamplePump.seek 160ml,13:51 @15:29:17.77 Pressure.coast @15:29:17.85 Bypass.open @15:29:19.15 Intake.close @15:29:20.06 Toroid.reconfigure :hold @15:29:20.15 Toroid.seek :crack @15:29:21.97 Toroid.seek :cartridge @15:29:23.08 Toroid.reconfigure :default @15:29:23.30 Toroid.seek :clear @15:29:25.48 Pressurizing cartridge to 14psi above 15.0psia @15:29:25.52 SamplePump.setPosition! 0ml @15:29:25.59 Intake.open @15:29:26.52 Bypass.close @15:29:27.29 Exhaust.close @15:29:28.15 Pressure.reconfigure :default @15:29:28.22 Pressure.hold @15:29:28.46 SamplePump.reconfigure :charge @15:29:28.55 SamplePump.seek 2.3ml,42 seconds @15:29:36.72 Pressure.coast @15:29:36.80 Intake.close @15:29:37.68 Pressurized cartridge with 0.301ml to 14.0psi @15:29:37.69 Isolated @26.1psia (11.1psi) @15:29:37.71 Checking for cartridge leaks from 26.1psia @15:29:37.76 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @15:29:44.01 Cartridge pressure fell 4% to 25.0psia @15:29:44.28 Sampling 2000.0ml @15:29:44.29 SamplePump.setPosition! 0ml @15:29:44.54 Exhaust.open @15:29:45.29 Intake.open @15:29:46.32 Pressure.reconfigure :default @15:29:46.41 Pressure.hold @15:29:46.71 SamplePump.reconfigure :sample @15:29:46.81 SamplePump.seek 2000ml,2:47:12 @15:29:49.37 Cartridge 28 sampling 1/2000ml at 0.52ml/s with 14.9psi vacuum @15:30:49.61 Cartridge 28 sampling 32/2000ml at 0.51ml/s with 11.9psi vacuum @15:31:49.38 Cartridge 28 sampling 63/2000ml at 0.53ml/s with 12.5psi vacuum @15:32:29.46 Session Begun @15:32:38.94 -> slot spare:2 @15:32:45.23 -> exit @15:32:45.25 Session Ended @15:32:49.55 Cartridge 28 sampling 95/2000ml at 0.54ml/s with 10.3psi vacuum @15:33:49.39 Cartridge 28 sampling 128/2000ml at 0.53ml/s with 10.2psi vacuum @15:34:49.77 Cartridge 28 sampling 160/2000ml at 0.56ml/s with 11.6psi vacuum @15:35:49.64 Cartridge 28 sampling 193/2000ml at 0.54ml/s with 14.2psi vacuum @15:36:49.38 Cartridge 28 sampling 227/2000ml at 0.56ml/s with 14.0psi vacuum @15:37:49.63 Cartridge 28 sampling 260/2000ml at 0.54ml/s with 11.4psi vacuum @15:38:49.39 Cartridge 28 sampling 293/2000ml at 0.55ml/s with 11.4psi vacuum @15:39:49.54 Cartridge 28 sampling 327/2000ml at 0.56ml/s with 11.4psi vacuum @15:40:49.38 Cartridge 28 sampling 361/2000ml at 0.57ml/s with 14.8psi vacuum @15:41:49.63 Cartridge 28 sampling 394/2000ml at 0.56ml/s with 14.0psi vacuum @15:42:49.38 Cartridge 28 sampling 427/2000ml at 0.57ml/s with 14.9psi vacuum @15:43:49.58 Cartridge 28 sampling 461/2000ml at 0.57ml/s with 12.7psi vacuum @15:44:49.39 Cartridge 28 sampling 494/2000ml at 0.55ml/s with 13.5psi vacuum @15:45:49.53 Cartridge 28 sampling 528/2000ml at 0.55ml/s with 10.6psi vacuum @15:46:49.38 Cartridge 28 sampling 561/2000ml at 0.55ml/s with 10.8psi vacuum @15:47:49.62 Cartridge 28 sampling 595/2000ml at 0.56ml/s with 10.8psi vacuum 1.6m deep (17.1psia) @15:48:49.38 Cartridge 28 sampling 628/2000ml at 0.56ml/s with 10.9psi vacuum @15:49:49.50 Cartridge 28 sampling 662/2000ml at 0.57ml/s with 13.3psi vacuum @15:50:49.38 Cartridge 28 sampling 695/2000ml at 0.56ml/s with 12.2psi vacuum @15:51:49.59 Cartridge 28 sampling 729/2000ml at 0.57ml/s with 13.8psi vacuum @15:52:49.38 Cartridge 28 sampling 763/2000ml at 0.56ml/s with 15.1psi vacuum @15:53:49.50 Cartridge 28 sampling 796/2000ml at 0.56ml/s with 14.6psi vacuum @15:54:49.38 Cartridge 28 sampling 830/2000ml at 0.55ml/s with 13.2psi vacuum @15:55:49.42 Cartridge 28 sampling 863/2000ml at 0.55ml/s with 14.1psi vacuum @15:56:49.38 Cartridge 28 sampling 896/2000ml at 0.56ml/s with 10.9psi vacuum @15:57:49.52 Cartridge 28 sampling 929/2000ml at 0.55ml/s with 14.1psi vacuum @15:58:49.38 Cartridge 28 sampling 963/2000ml at 0.55ml/s with 13.7psi vacuum @15:59:49.47 Cartridge 28 sampling 996/2000ml at 0.54ml/s with 11.3psi vacuum @16:00:49.38 Cartridge 28 sampling 1030/2000ml at 0.56ml/s with 12.4psi vacuum @16:01:49.58 Cartridge 28 sampling 1063/2000ml at 0.55ml/s with 14.1psi vacuum @16:02:49.51 Cartridge 28 sampling 1096/2000ml at 0.55ml/s with 14.2psi vacuum @16:03:49.38 Cartridge 28 sampling 1130/2000ml at 0.55ml/s with 12.4psi vacuum @16:04:49.74 Cartridge 28 sampling 1163/2000ml at 0.56ml/s with 13.9psi vacuum @16:05:49.38 Cartridge 28 sampling 1197/2000ml at 0.56ml/s with 11.1psi vacuum @16:06:49.56 Cartridge 28 sampling 1230/2000ml at 0.56ml/s with 13.4psi vacuum @16:07:49.38 Cartridge 28 sampling 1264/2000ml at 0.55ml/s with 14.0psi vacuum @16:08:49.55 Cartridge 28 sampling 1297/2000ml at 0.55ml/s with 13.5psi vacuum @16:09:49.38 Cartridge 28 sampling 1331/2000ml at 0.55ml/s with 14.4psi vacuum @16:10:49.54 Cartridge 28 sampling 1364/2000ml at 0.56ml/s with 11.4psi vacuum @16:11:49.49 Cartridge 28 sampling 1398/2000ml at 0.55ml/s with 12.8psi vacuum @16:12:49.54 Cartridge 28 sampling 1431/2000ml at 0.55ml/s with 11.3psi vacuum @16:13:49.38 Cartridge 28 sampling 1464/2000ml at 0.56ml/s with 14.3psi vacuum @16:14:49.58 Cartridge 28 sampling 1498/2000ml at 0.56ml/s with 11.6psi vacuum @16:15:49.38 Cartridge 28 sampling 1531/2000ml at 0.56ml/s with 13.0psi vacuum @16:16:49.54 Cartridge 28 sampling 1565/2000ml at 0.55ml/s with 12.6psi vacuum @16:17:49.38 Cartridge 28 sampling 1599/2000ml at 0.56ml/s with 13.2psi vacuum @16:18:50.68 Cartridge 28 sampling 1632/2000ml at 0.57ml/s with 11.9psi vacuum @16:19:49.38 Cartridge 28 sampling 1666/2000ml at 0.56ml/s with 12.3psi vacuum @16:20:49.63 Cartridge 28 sampling 1699/2000ml at 0.55ml/s with 12.4psi vacuum @16:21:49.62 Cartridge 28 sampling 1733/2000ml at 0.56ml/s with 13.9psi vacuum @16:22:49.56 Cartridge 28 sampling 1767/2000ml at 0.56ml/s with 12.1psi vacuum @16:23:49.79 Cartridge 28 sampling 1800/2000ml at 0.55ml/s with 14.0psi vacuum @16:24:49.40 Cartridge 28 sampling 1833/2000ml at 0.55ml/s with 14.9psi vacuum @16:25:49.57 Cartridge 28 sampling 1867/2000ml at 0.55ml/s with 12.1psi vacuum @16:26:49.37 Cartridge 28 sampling 1900/2000ml at 0.56ml/s with 12.4psi vacuum @16:27:49.68 Cartridge 28 sampling 1934/2000ml at 0.55ml/s with 13.7psi vacuum @16:28:49.38 Cartridge 28 sampling 1967/2000ml at 0.56ml/s with 11.5psi vacuum @16:29:48.97 Pressure.coast @16:29:49.64 Intake.close @16:29:50.06 Exhaust.close @16:29:50.66 Toroid.reconfigure :hold @16:29:50.90 Toroid.seek :bypass @16:29:52.62 Toroid.reconfigure :default @16:29:52.80 Toroid.seek :clear @16:29:54.72 Sampled 2000.0ml @16:29:55.22 Plunge.seek :reagent @16:29:56.14 Sample loop closed. Ready to Deliver Reagents @16:29:56.41 Carousel.seek 28 @16:29:56.66 Plunge.seek :locked @16:29:57.94 Air.seek :clear @16:29:58.68 Lysis1.seek :clear @16:30:02.05 Archive.coast @16:30:02.13 Air.coast @16:30:02.34 Lysis1.coast @16:30:02.49 Lysis2.coast @16:30:02.75 Plunge.coast @16:30:02.98 SamplePump.coast @16:30:03.17 Toroid.coast @16:30:03.24 Carousel.coast @16:30:03.45 Gate.power :motor,:OFF @16:30:03.48 Gate.power :core,:OFF @16:30:03.51 Gate.power :analytic1,:OFF @16:30:03.54 Gate.power :analytic2,:OFF @16:30:03.58 Gate.power :analytic3,:OFF @16:30:03.60 Gate.power :main,:OFF @16:30:03.63 Safely stopped and ready to power off @16:30:03.82 delay 2:00 @16:32:03.97 Gate.power :main,:ON,limit to 5000mA @16:32:03.99 Gate.power :analytic1,:ON,limit to 5000mA @16:32:04.02 Gate.power :analytic2,:ON,limit to 5000mA @16:32:04.05 Gate.power :analytic3,:ON,limit to 5000mA @16:32:04.12 Powering core bus @16:32:06.43 Archive.configure :default @16:32:06.59 Air.configure :default @16:32:06.86 Lysis1.configure :default @16:32:07.43 Lysis2.configure :default @16:32:07.79 Plunge.configure :default @16:32:08.26 SamplePump.configure :default @16:32:08.64 ExtraPump.configure :default @16:32:09.02 Toroid.configure :default @16:32:09.36 Carousel.configure :default @16:32:09.73 Pressure.configure :default @16:32:09.92 All elf controllers running firmware version 4.61 @16:32:09.96 Gate.power :motor,:ON,limit to 5000mA @16:32:10.53 Archive.reconfigure :default @16:32:11.27 Archive.home.jog 0 @16:32:11.82 Toroid.reconfigure :default @16:32:11.89 Plunge.reconfigure :default @16:32:11.95 Lysis1.reconfigure :default @16:32:12.01 Air.reconfigure :default @16:32:12.16 Lysis2.reconfigure :default @16:32:12.54 Toroid.home.jog 0 @16:32:12.75 Plunge.home.jog 0 @16:32:13.18 Lysis2.home.jog 0 @16:32:13.22 Lysis1.home.jog 0 @16:32:13.25 Air.home.jog 0 @16:32:13.72 Archive.reconfigure :homing @16:32:13.87 Archive.jog 1355 @16:32:14.04 Plunge.reconfigure :homing @16:32:14.13 Toroid.reconfigure :homing @16:32:14.18 Lysis1.reconfigure :homing @16:32:14.22 Air.reconfigure :homing @16:32:14.27 Lysis2.reconfigure :homing @16:32:14.49 Plunge.jog 850 @16:32:14.56 Toroid.jog 622 @16:32:14.65 Lysis2.jog 1358 @16:32:14.68 Lysis1.jog 1352 @16:32:14.72 Air.jog 1367 @16:32:15.02 Archive.setRawPosition! 50 @16:32:15.09 Archive.seek :home @16:32:15.38 Archive.reconfigure :default @16:32:15.52 Plunge.setRawPosition! 50 @16:32:15.58 Plunge.seek :home @16:32:15.89 Plunge.reconfigure :default @16:32:16.04 Lysis1.setRawPosition! 50 @16:32:16.07 Air.setRawPosition! 50 @16:32:16.11 Lysis2.setRawPosition! 50 @16:32:16.22 Lysis1.seek :home @16:32:16.28 Air.seek :home @16:32:16.38 Lysis2.seek :home @16:32:16.76 Lysis1.reconfigure :default @16:32:16.98 Toroid.setRawPosition! 0 @16:32:17.01 Lysis2.reconfigure :default @16:32:17.05 Air.reconfigure :default @16:32:17.19 Toroid.seek :home @16:32:19.23 Toroid.reconfigure :default @16:32:19.76 Toroid.seek :clear @16:32:21.50 Carousel.reconfigure :default @16:32:21.66 Carousel.home.jog 0 @16:32:22.27 Carousel.seek 28 @16:32:22.83 Selecting Cartridge 27 @16:32:23.24 Carousel.seek 27 @16:32:23.79 Lysis1.seek :locked @16:32:27.05 Plunge.seek :locked @16:32:27.22 Air.seek :locked @16:32:31.52 Mounted archive Cartridge @16:32:32.39 Duration of filtering limited to 4:05:00 @16:32:32.71 Priming sample loop w/160ml, bypass w/10ml @16:32:32.73 SamplePump.setPosition! 0ml @16:32:32.80 Bypass.open @16:32:34.30 Intake.open @16:32:35.17 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @16:32:37.91 Vacuum tare is 18% of full scale @16:32:37.96 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @16:32:38.46 Pressure.reconfigure :default @16:32:38.72 Pressure.hold @16:32:39.20 SamplePump.reconfigure :prime @16:32:39.43 SamplePump.seek 10ml,1:21 @16:32:50.58 Pressure.coast @16:32:50.96 SamplePump.setPosition! 0ml @16:32:51.26 Exhaust.open @16:32:52.70 Bypass.close @16:32:54.24 Pressure.reconfigure :default @16:32:54.32 Pressure.hold @16:32:54.61 SamplePump.reconfigure :prime @16:32:54.70 SamplePump.seek 160ml,13:51 @16:35:46.39 Pressure.coast @16:35:46.47 Bypass.open @16:35:47.18 Intake.close @16:35:48.06 Toroid.reconfigure :hold @16:35:48.14 Toroid.seek :crack @16:35:49.76 Toroid.seek :cartridge @16:35:50.76 Toroid.reconfigure :default @16:35:51.07 Toroid.seek :clear @16:35:53.31 Pressurizing cartridge to 14psi above 14.6psia @16:35:53.37 SamplePump.setPosition! 0ml @16:35:53.57 Intake.open @16:35:54.56 Bypass.close @16:35:55.26 Exhaust.close @16:35:56.12 Pressure.reconfigure :default @16:35:56.20 Pressure.hold @16:35:56.44 SamplePump.reconfigure :charge @16:35:56.53 SamplePump.seek 2.3ml,42 seconds @16:36:04.06 Pressure.coast @16:36:04.50 Intake.close @16:36:05.39 Pressurized cartridge with 0.249ml to 14.1psi @16:36:05.40 Isolated @24.2psia (9.6psi) @16:36:05.43 Checking for cartridge leaks from 24.2psia @16:36:05.48 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @16:36:10.15 Cartridge pressure fell 6% to 22.7psia @16:36:10.48 Sampling 2000.0ml @16:36:10.55 SamplePump.setPosition! 0ml @16:36:10.70 Exhaust.open @16:36:11.52 Intake.open @16:36:12.84 Pressure.reconfigure :default @16:36:13.01 Pressure.hold @16:36:13.64 SamplePump.reconfigure :sample @16:36:13.76 SamplePump.seek 2000ml,2:47:12 @16:36:15.82 Cartridge 27 sampling 1/2000ml at 0.57ml/s with 13.0psi vacuum @16:37:16.09 Cartridge 27 sampling 33/2000ml at 0.54ml/s with 11.8psi vacuum @16:38:15.84 Cartridge 27 sampling 65/2000ml at 0.56ml/s with 12.8psi vacuum @16:39:16.01 Cartridge 27 sampling 99/2000ml at 0.56ml/s with 9.9psi vacuum @16:40:15.83 Cartridge 27 sampling 133/2000ml at 0.58ml/s with 14.3psi vacuum @16:41:16.00 Cartridge 27 sampling 167/2000ml at 0.57ml/s with 10.8psi vacuum @16:42:15.83 Cartridge 27 sampling 202/2000ml at 0.57ml/s with 11.8psi vacuum @16:43:15.96 Cartridge 27 sampling 236/2000ml at 0.60ml/s with 11.7psi vacuum @16:44:15.83 Cartridge 27 sampling 270/2000ml at 0.57ml/s with 13.8psi vacuum @16:45:15.88 Cartridge 27 sampling 305/2000ml at 0.56ml/s with 13.3psi vacuum @16:46:15.83 Cartridge 27 sampling 339/2000ml at 0.57ml/s with 13.9psi vacuum @16:47:15.96 Cartridge 27 sampling 373/2000ml at 0.57ml/s with 13.9psi vacuum @16:48:15.84 Cartridge 27 sampling 408/2000ml at 0.59ml/s with 11.2psi vacuum @16:49:15.86 Cartridge 27 sampling 442/2000ml at 0.58ml/s with 10.5psi vacuum @16:50:17.18 Cartridge 27 sampling 477/2000ml at 0.56ml/s with 15.8psi vacuum @16:51:15.94 Cartridge 27 sampling 511/2000ml at 0.57ml/s with 12.9psi vacuum @16:52:15.84 Cartridge 27 sampling 545/2000ml at 0.56ml/s with 14.6psi vacuum @16:53:16.06 Cartridge 27 sampling 580/2000ml at 0.57ml/s with 10.6psi vacuum @16:54:16.01 Cartridge 27 sampling 614/2000ml at 0.57ml/s with 12.9psi vacuum @16:55:15.83 Cartridge 27 sampling 649/2000ml at 0.56ml/s with 10.8psi vacuum @16:56:16.22 Cartridge 27 sampling 683/2000ml at 0.57ml/s with 14.1psi vacuum @16:57:15.83 Cartridge 27 sampling 717/2000ml at 0.58ml/s with 11.7psi vacuum @16:58:16.02 Cartridge 27 sampling 752/2000ml at 0.58ml/s with 14.3psi vacuum @16:59:15.83 Cartridge 27 sampling 786/2000ml at 0.58ml/s with 14.6psi vacuum @17:00:16.17 Cartridge 27 sampling 820/2000ml at 0.55ml/s with 11.3psi vacuum @17:01:15.83 Cartridge 27 sampling 854/2000ml at 0.57ml/s with 13.5psi vacuum @17:02:16.01 Cartridge 27 sampling 888/2000ml at 0.58ml/s with 10.7psi vacuum @17:03:15.83 Cartridge 27 sampling 923/2000ml at 0.58ml/s with 11.4psi vacuum @17:04:16.04 Cartridge 27 sampling 957/2000ml at 0.58ml/s with 12.6psi vacuum @17:05:15.82 Cartridge 27 sampling 991/2000ml at 0.57ml/s with 13.1psi vacuum @17:06:15.98 Cartridge 27 sampling 1026/2000ml at 0.57ml/s with 14.2psi vacuum @17:07:15.82 Cartridge 27 sampling 1060/2000ml at 0.57ml/s with 11.2psi vacuum @17:08:16.04 Cartridge 27 sampling 1094/2000ml at 0.57ml/s with 12.7psi vacuum @17:09:15.83 Cartridge 27 sampling 1129/2000ml at 0.57ml/s with 13.9psi vacuum @17:10:16.00 Cartridge 27 sampling 1163/2000ml at 0.57ml/s with 11.2psi vacuum @17:11:15.83 Cartridge 27 sampling 1197/2000ml at 0.57ml/s with 12.5psi vacuum @17:12:16.23 Cartridge 27 sampling 1231/2000ml at 0.57ml/s with 14.0psi vacuum @17:13:16.07 Cartridge 27 sampling 1265/2000ml at 0.57ml/s with 12.2psi vacuum @17:14:16.07 Cartridge 27 sampling 1300/2000ml at 0.57ml/s with 11.7psi vacuum @17:15:16.11 Cartridge 27 sampling 1334/2000ml at 0.58ml/s with 13.2psi vacuum @17:16:15.84 Cartridge 27 sampling 1368/2000ml at 0.57ml/s with 12.7psi vacuum @17:17:16.19 Cartridge 27 sampling 1402/2000ml at 0.56ml/s with 12.1psi vacuum @17:18:15.83 Cartridge 27 sampling 1436/2000ml at 0.56ml/s with 11.6psi vacuum @17:19:16.29 Cartridge 27 sampling 1471/2000ml at 0.57ml/s with 13.8psi vacuum @17:20:15.83 Cartridge 27 sampling 1505/2000ml at 0.57ml/s with 12.6psi vacuum @17:21:16.00 Cartridge 27 sampling 1539/2000ml at 0.57ml/s with 12.9psi vacuum @17:22:15.83 Cartridge 27 sampling 1573/2000ml at 0.58ml/s with 13.1psi vacuum @17:23:16.07 Cartridge 27 sampling 1608/2000ml at 0.57ml/s with 11.1psi vacuum @17:24:15.83 Cartridge 27 sampling 1642/2000ml at 0.57ml/s with 11.7psi vacuum @17:25:16.11 Cartridge 27 sampling 1677/2000ml at 0.57ml/s with 11.5psi vacuum @17:26:15.84 Cartridge 27 sampling 1711/2000ml at 0.57ml/s with 12.1psi vacuum @17:27:16.12 Cartridge 27 sampling 1745/2000ml at 0.57ml/s with 14.3psi vacuum @17:28:15.83 Cartridge 27 sampling 1780/2000ml at 0.57ml/s with 12.2psi vacuum @17:29:16.03 Cartridge 27 sampling 1814/2000ml at 0.58ml/s with 14.4psi vacuum @17:30:15.83 Cartridge 27 sampling 1849/2000ml at 0.57ml/s with 12.2psi vacuum @17:31:15.89 Cartridge 27 sampling 1883/2000ml at 0.57ml/s with 11.5psi vacuum @17:32:15.84 Cartridge 27 sampling 1918/2000ml at 0.57ml/s with 14.0psi vacuum @17:33:15.88 Cartridge 27 sampling 1952/2000ml at 0.57ml/s with 13.9psi vacuum @17:34:15.83 Cartridge 27 sampling 1986/2000ml at 0.57ml/s with 12.2psi vacuum @17:34:39.44 Pressure.coast @17:34:39.60 Intake.close @17:34:39.71 Exhaust.close @17:34:39.90 Toroid.reconfigure :hold @17:34:39.98 Toroid.seek :bypass @17:34:41.63 Toroid.reconfigure :default @17:34:41.71 Toroid.seek :clear @17:34:43.34 Sampled 2000.0ml @17:34:43.88 Plunge.seek :reagent @17:34:44.91 Sample loop closed. Ready to Deliver Reagents @17:34:45.55 Carousel.seek 27 @17:34:46.21 Plunge.seek :locked @17:34:47.67 Air.seek :clear @17:34:48.09 Lysis1.seek :clear @17:34:51.54 Archive.coast @17:34:51.62 Air.coast @17:34:51.70 Lysis1.coast @17:34:51.76 Lysis2.coast @17:34:51.83 Plunge.coast @17:34:51.90 SamplePump.coast @17:34:51.97 Toroid.coast @17:34:52.04 Carousel.coast @17:34:52.27 Gate.power :motor,:OFF @17:34:52.29 Gate.power :core,:OFF @17:34:52.32 Gate.power :analytic1,:OFF @17:34:52.35 Gate.power :analytic2,:OFF @17:34:52.39 Gate.power :analytic3,:OFF @17:34:52.42 Gate.power :main,:OFF @17:34:52.45 Safely stopped and ready to power off @17:34:52.74 delay 2:00 @17:36:52.87 Gate.power :main,:ON,limit to 5000mA @17:36:52.91 Gate.power :analytic1,:ON,limit to 5000mA @17:36:52.93 Gate.power :analytic2,:ON,limit to 5000mA @17:36:52.96 Gate.power :analytic3,:ON,limit to 5000mA @17:36:53.03 Powering core bus @17:36:55.72 Archive.configure :default @17:36:56.15 Air.configure :default @17:36:56.47 Lysis1.configure :default @17:36:56.77 Lysis2.configure :default @17:36:57.24 Plunge.configure :default @17:36:57.55 SamplePump.configure :default @17:36:57.88 ExtraPump.configure :default @17:36:58.27 Toroid.configure :default @17:36:58.93 Carousel.configure :default @17:36:59.27 Pressure.configure :default @17:36:59.96 All elf controllers running firmware version 4.61 @17:36:59.99 Gate.power :motor,:ON,limit to 5000mA @17:37:00.44 Archive.reconfigure :default @17:37:00.71 Archive.home.jog 0 @17:37:00.76 Toroid.reconfigure :default @17:37:00.79 Plunge.reconfigure :default @17:37:00.83 Lysis1.reconfigure :default @17:37:00.88 Air.reconfigure :default @17:37:00.93 Lysis2.reconfigure :default @17:37:01.19 Toroid.home.jog 0 @17:37:01.34 Plunge.home.jog 0 @17:37:01.37 Lysis2.home.jog 0 @17:37:01.41 Air.home.jog 0 @17:37:01.45 Lysis1.home.jog 0 @17:37:02.20 Toroid.reconfigure :homing @17:37:02.25 Plunge.reconfigure :homing @17:37:02.30 Lysis2.reconfigure :homing @17:37:02.36 Air.reconfigure :homing @17:37:02.40 Archive.reconfigure :homing @17:37:02.45 Lysis1.reconfigure :homing @17:37:02.72 Toroid.jog 619 @17:37:02.80 Plunge.jog 841 @17:37:02.83 Lysis2.jog 1358 @17:37:02.86 Archive.jog 1355 @17:37:02.97 Lysis1.jog 1351 @17:37:03.00 Air.jog 1363 @17:37:03.99 Plunge.setRawPosition! 50 @17:37:04.17 Plunge.seek :home @17:37:04.82 Lysis2.setRawPosition! 50 @17:37:04.94 Lysis1.setRawPosition! 50 @17:37:05.05 Air.setRawPosition! 50 @17:37:05.13 Plunge.reconfigure :default @17:37:05.18 Archive.setRawPosition! 50 @17:37:05.86 Lysis2.seek :home @17:37:06.03 Lysis1.seek :home @17:37:06.13 Air.seek :home @17:37:06.24 Archive.seek :home @17:37:07.00 Lysis2.reconfigure :default @17:37:07.02 Lysis1.reconfigure :default @17:37:07.24 Toroid.setRawPosition! 0 @17:37:07.57 Toroid.seek :home @17:37:07.88 Air.reconfigure :default @17:37:07.93 Archive.reconfigure :default @17:37:08.03 Toroid.reconfigure :default @17:37:08.98 Toroid.seek :clear @17:37:10.62 Carousel.reconfigure :default @17:37:10.70 Carousel.home.jog 0 @17:37:11.23 Carousel.seek 27 @17:37:11.43 Selecting Cartridge 26 @17:37:11.67 Carousel.seek 26 @17:37:12.22 Lysis1.seek :locked @17:37:15.80 Plunge.seek :locked @17:37:16.14 Air.seek :locked @17:37:20.45 Mounted archive Cartridge @17:37:20.78 Duration of filtering limited to 4:05:00 @17:37:21.10 Priming sample loop w/160ml, bypass w/10ml @17:37:21.12 SamplePump.setPosition! 0ml @17:37:21.19 Bypass.open @17:37:22.68 Intake.open @17:37:24.13 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @17:37:26.92 Vacuum tare is 18% of full scale @17:37:26.97 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @17:37:27.37 Pressure.reconfigure :default @17:37:27.56 Pressure.hold @17:37:28.06 SamplePump.reconfigure :prime @17:37:28.27 SamplePump.seek 10ml,1:21 @17:37:39.55 Pressure.coast @17:37:39.62 SamplePump.setPosition! 0ml @17:37:39.69 Exhaust.open @17:37:41.12 Bypass.close @17:37:43.33 Pressure.reconfigure :default @17:37:43.51 Pressure.hold @17:37:44.20 SamplePump.reconfigure :prime @17:37:44.40 SamplePump.seek 160ml,13:51 @17:40:36.12 Pressure.coast @17:40:36.20 Bypass.open @17:40:36.90 Intake.close @17:40:38.07 Toroid.reconfigure :hold @17:40:38.25 Toroid.seek :crack @17:40:40.16 Toroid.seek :cartridge @17:40:41.21 Toroid.reconfigure :default @17:40:41.39 Toroid.seek :clear @17:40:43.32 Pressurizing cartridge to 14psi above 14.4psia @17:40:43.34 SamplePump.setPosition! 0ml @17:40:43.41 Intake.open @17:40:44.23 Bypass.close @17:40:44.95 Exhaust.close @17:40:46.30 Pressure.reconfigure :default @17:40:46.37 Pressure.hold @17:40:46.65 SamplePump.reconfigure :charge @17:40:47.02 SamplePump.seek 2.3ml,42 seconds @17:40:55.29 Pressure.coast @17:40:55.37 Intake.close @17:40:56.45 Pressurized cartridge with 0.298ml to 14.0psi @17:40:56.48 Isolated @24.9psia (10.5psi) @17:40:56.53 Checking for cartridge leaks from 24.9psia @17:40:56.67 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @17:41:01.39 Cartridge pressure fell 4% to 23.8psia @17:41:01.62 Sampling 2000.0ml @17:41:01.67 SamplePump.setPosition! 0ml @17:41:01.87 Exhaust.open @17:41:02.57 Intake.open @17:41:03.50 Pressure.reconfigure :default @17:41:03.59 Pressure.hold @17:41:03.89 SamplePump.reconfigure :sample @17:41:03.98 SamplePump.seek 2000ml,2:47:12 @17:41:06.80 Cartridge 26 sampling 1/2000ml at 0.54ml/s with 13.6psi vacuum @17:42:06.56 Cartridge 26 sampling 32/2000ml at 0.54ml/s with 12.5psi vacuum @17:43:06.77 Cartridge 26 sampling 65/2000ml at 0.54ml/s with 14.1psi vacuum @17:44:06.57 Cartridge 26 sampling 99/2000ml at 0.58ml/s with 13.9psi vacuum @17:45:06.70 Cartridge 26 sampling 133/2000ml at 0.55ml/s with 10.9psi vacuum @17:46:06.56 Cartridge 26 sampling 167/2000ml at 0.56ml/s with 12.9psi vacuum @17:47:06.74 Cartridge 26 sampling 201/2000ml at 0.58ml/s with 12.6psi vacuum @17:48:06.56 Cartridge 26 sampling 235/2000ml at 0.57ml/s with 14.9psi vacuum @17:49:06.73 Cartridge 26 sampling 270/2000ml at 0.57ml/s with 15.6psi vacuum @17:50:06.56 Cartridge 26 sampling 304/2000ml at 0.57ml/s with 14.8psi vacuum @17:51:06.88 Cartridge 26 sampling 338/2000ml at 0.57ml/s with 15.4psi vacuum @17:52:06.56 Cartridge 26 sampling 372/2000ml at 0.57ml/s with 14.3psi vacuum @17:53:06.80 Cartridge 26 sampling 406/2000ml at 0.56ml/s with 12.5psi vacuum @17:54:06.56 Cartridge 26 sampling 441/2000ml at 0.56ml/s with 12.7psi vacuum @17:55:06.72 Cartridge 26 sampling 475/2000ml at 0.56ml/s with 14.0psi vacuum @17:56:06.59 Cartridge 26 sampling 509/2000ml at 0.56ml/s with 13.5psi vacuum @17:57:06.62 Cartridge 26 sampling 543/2000ml at 0.57ml/s with 11.2psi vacuum @17:58:06.56 Cartridge 26 sampling 577/2000ml at 0.57ml/s with 14.9psi vacuum @17:59:06.72 Cartridge 26 sampling 611/2000ml at 0.57ml/s with 11.2psi vacuum @18:00:06.56 Cartridge 26 sampling 645/2000ml at 0.56ml/s with 15.8psi vacuum @18:01:06.89 Cartridge 26 sampling 680/2000ml at 0.57ml/s with 11.6psi vacuum @18:02:06.77 Cartridge 26 sampling 714/2000ml at 0.56ml/s with 10.9psi vacuum @18:03:06.57 Cartridge 26 sampling 748/2000ml at 0.57ml/s with 12.3psi vacuum @18:04:06.84 Cartridge 26 sampling 782/2000ml at 0.56ml/s with 11.4psi vacuum @18:05:06.57 Cartridge 26 sampling 816/2000ml at 0.57ml/s with 11.8psi vacuum @18:06:06.99 Cartridge 26 sampling 850/2000ml at 0.56ml/s with 10.8psi vacuum @18:07:06.56 Cartridge 26 sampling 884/2000ml at 0.55ml/s with 14.8psi vacuum @18:08:06.73 Cartridge 26 sampling 918/2000ml at 0.57ml/s with 13.4psi vacuum @18:09:06.55 Cartridge 26 sampling 952/2000ml at 0.57ml/s with 14.6psi vacuum @18:10:06.73 Cartridge 26 sampling 986/2000ml at 0.57ml/s with 11.7psi vacuum @18:11:06.55 Cartridge 26 sampling 1020/2000ml at 0.56ml/s with 12.7psi vacuum @18:12:06.81 Cartridge 26 sampling 1054/2000ml at 0.57ml/s with 14.0psi vacuum @18:13:06.56 Cartridge 26 sampling 1089/2000ml at 0.56ml/s with 11.4psi vacuum @18:14:06.72 Cartridge 26 sampling 1123/2000ml at 0.57ml/s with 15.1psi vacuum @18:15:06.55 Cartridge 26 sampling 1157/2000ml at 0.57ml/s with 11.3psi vacuum @18:16:06.87 Cartridge 26 sampling 1191/2000ml at 0.57ml/s with 11.5psi vacuum @18:17:06.57 Cartridge 26 sampling 1225/2000ml at 0.57ml/s with 13.2psi vacuum @18:18:06.73 Cartridge 26 sampling 1259/2000ml at 0.57ml/s with 13.2psi vacuum @18:19:06.56 Cartridge 26 sampling 1293/2000ml at 0.57ml/s with 11.8psi vacuum @18:20:06.64 Cartridge 26 sampling 1327/2000ml at 0.57ml/s with 12.2psi vacuum @18:21:06.56 Cartridge 26 sampling 1362/2000ml at 0.57ml/s with 13.1psi vacuum @18:22:06.95 Cartridge 26 sampling 1396/2000ml at 0.57ml/s with 13.5psi vacuum @18:23:06.57 Cartridge 26 sampling 1430/2000ml at 0.58ml/s with 14.1psi vacuum @18:24:06.86 Cartridge 26 sampling 1464/2000ml at 0.57ml/s with 11.6psi vacuum @18:25:06.91 Cartridge 26 sampling 1499/2000ml at 0.57ml/s with 12.7psi vacuum @18:26:06.83 Cartridge 26 sampling 1533/2000ml at 0.57ml/s with 11.8psi vacuum @18:27:06.82 Cartridge 26 sampling 1567/2000ml at 0.59ml/s with 11.6psi vacuum @18:28:06.56 Cartridge 26 sampling 1601/2000ml at 0.57ml/s with 11.6psi vacuum @18:29:07.00 Cartridge 26 sampling 1636/2000ml at 0.57ml/s with 12.1psi vacuum @18:30:06.56 Cartridge 26 sampling 1670/2000ml at 0.57ml/s with 12.8psi vacuum @18:31:06.73 Cartridge 26 sampling 1704/2000ml at 0.56ml/s with 13.2psi vacuum @18:32:06.57 Cartridge 26 sampling 1738/2000ml at 0.57ml/s with 11.2psi vacuum @18:33:06.73 Cartridge 26 sampling 1772/2000ml at 0.56ml/s with 14.8psi vacuum @18:34:06.56 Cartridge 26 sampling 1806/2000ml at 0.55ml/s with 11.4psi vacuum @18:35:06.80 Cartridge 26 sampling 1840/2000ml at 0.57ml/s with 12.3psi vacuum @18:36:06.57 Cartridge 26 sampling 1873/2000ml at 0.56ml/s with 13.7psi vacuum @18:37:06.79 Cartridge 26 sampling 1907/2000ml at 0.57ml/s with 13.3psi vacuum @18:38:06.56 Cartridge 26 sampling 1941/2000ml at 0.57ml/s with 13.0psi vacuum @18:39:06.80 Cartridge 26 sampling 1975/2000ml at 0.57ml/s with 13.5psi vacuum @18:39:50.21 Pressure.coast @18:39:50.56 Intake.close @18:39:50.74 Exhaust.close @18:39:51.47 Toroid.reconfigure :hold @18:39:51.66 Toroid.seek :bypass @18:39:53.47 Toroid.reconfigure :default @18:39:53.70 Toroid.seek :clear @18:39:56.44 Sampled 2000.0ml @18:39:56.95 Plunge.seek :reagent @18:39:57.88 Sample loop closed. Ready to Deliver Reagents @18:39:58.15 Carousel.seek 26 @18:39:58.39 Plunge.seek :locked @18:39:59.48 Air.seek :clear @18:39:59.67 Lysis1.seek :clear @18:40:03.23 Archive.coast @18:40:03.38 Air.coast @18:40:03.54 Lysis1.coast @18:40:03.60 Lysis2.coast @18:40:03.78 Plunge.coast @18:40:03.91 SamplePump.coast @18:40:03.98 Toroid.coast @18:40:04.21 Carousel.coast @18:40:04.52 Gate.power :motor,:OFF @18:40:04.56 Gate.power :core,:OFF @18:40:04.62 Gate.power :analytic1,:OFF @18:40:04.70 Gate.power :analytic2,:OFF @18:40:04.80 Gate.power :analytic3,:OFF @18:40:04.85 Gate.power :main,:OFF @18:40:04.90 Safely stopped and ready to power off @18:40:05.41 delay 2:00 @18:42:05.55 Gate.power :main,:ON,limit to 5000mA @18:42:05.59 Gate.power :analytic1,:ON,limit to 5000mA @18:42:05.61 Gate.power :analytic2,:ON,limit to 5000mA @18:42:05.64 Gate.power :analytic3,:ON,limit to 5000mA @18:42:05.72 Powering core bus @18:42:08.02 Archive.configure :default @18:42:08.20 Air.configure :default @18:42:08.34 Lysis1.configure :default @18:42:08.49 Lysis2.configure :default @18:42:08.65 Plunge.configure :default @18:42:08.84 SamplePump.configure :default @18:42:09.33 ExtraPump.configure :default @18:42:09.87 Toroid.configure :default @18:42:10.51 Carousel.configure :default @18:42:10.99 Pressure.configure :default @18:42:11.13 All elf controllers running firmware version 4.61 @18:42:11.20 Gate.power :motor,:ON,limit to 5000mA @18:42:11.61 Archive.reconfigure :default @18:42:12.59 Archive.home.jog 0 @18:42:13.22 Toroid.reconfigure :default @18:42:13.25 Plunge.reconfigure :default @18:42:13.36 Lysis1.reconfigure :default @18:42:13.52 Lysis2.reconfigure :default @18:42:13.59 Air.reconfigure :default @18:42:13.96 Toroid.home.jog 0 @18:42:14.18 Plunge.home.jog 0 @18:42:14.22 Lysis1.home.jog 0 @18:42:14.26 Lysis2.home.jog 0 @18:42:14.35 Air.home.jog 0 @18:42:15.35 Archive.reconfigure :homing @18:42:15.61 Archive.jog 1354 @18:42:15.67 Toroid.reconfigure :homing @18:42:15.74 Lysis2.reconfigure :homing @18:42:15.78 Air.reconfigure :homing @18:42:15.84 Lysis1.reconfigure :homing @18:42:15.90 Plunge.reconfigure :homing @18:42:16.18 Toroid.jog 617 @18:42:16.21 Lysis2.jog 1355 @18:42:16.25 Lysis1.jog 1352 @18:42:16.29 Air.jog 1363 @18:42:16.34 Plunge.jog 844 @18:42:17.09 Archive.setRawPosition! 50 @18:42:17.16 Archive.seek :home @18:42:17.44 Plunge.setRawPosition! 50 @18:42:17.50 Plunge.seek :home @18:42:17.66 Lysis1.setRawPosition! 50 @18:42:17.70 Air.setRawPosition! 50 @18:42:17.73 Lysis2.setRawPosition! 50 @18:42:17.77 Archive.reconfigure :default @18:42:17.92 Lysis1.seek :home @18:42:18.01 Air.seek :home @18:42:18.07 Lysis2.seek :home @18:42:18.30 Plunge.reconfigure :default @18:42:18.45 Lysis1.reconfigure :default @18:42:18.60 Toroid.setRawPosition! 0 @18:42:18.63 Lysis2.reconfigure :default @18:42:18.66 Air.reconfigure :default @18:42:18.80 Toroid.seek :home @18:42:18.99 Toroid.reconfigure :default @18:42:19.24 Toroid.seek :clear @18:42:20.93 Carousel.reconfigure :default @18:42:21.16 Carousel.home.jog 0 @18:42:21.91 Carousel.seek 26 @18:42:22.35 Selecting Cartridge 25 @18:42:22.99 Carousel.seek 25 @18:42:23.84 Lysis1.seek :locked @18:42:27.10 Plunge.seek :locked @18:42:27.42 Air.seek :locked @18:42:31.81 Mounted archive Cartridge @18:42:32.55 Duration of filtering limited to 4:05:00 @18:42:34.66 Priming sample loop w/160ml, bypass w/10ml @18:42:34.68 SamplePump.setPosition! 0ml @18:42:34.75 Bypass.open @18:42:36.25 Intake.open @18:42:37.18 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @18:42:39.87 Vacuum tare is 18% of full scale @18:42:39.90 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @18:42:40.56 Pressure.reconfigure :default @18:42:40.73 Pressure.hold @18:42:41.32 SamplePump.reconfigure :prime @18:42:41.54 SamplePump.seek 10ml,1:21 @18:42:52.69 Pressure.coast @18:42:52.93 SamplePump.setPosition! 0ml @18:42:53.19 Exhaust.open @18:42:54.67 Bypass.close @18:42:56.22 Pressure.reconfigure :default @18:42:56.30 Pressure.hold @18:42:56.59 SamplePump.reconfigure :prime @18:42:56.68 SamplePump.seek 160ml,13:51 @18:45:48.32 Pressure.coast @18:45:48.40 Bypass.open @18:45:49.10 Intake.close @18:45:50.01 Toroid.reconfigure :hold @18:45:50.09 Toroid.seek :crack @18:45:51.70 Toroid.seek :cartridge @18:45:52.73 Toroid.reconfigure :default @18:45:53.18 Toroid.seek :clear @18:45:55.34 Pressurizing cartridge to 14psi above 13.9psia @18:45:55.35 SamplePump.setPosition! 0ml @18:45:55.56 Intake.open @18:45:56.40 Bypass.close @18:45:57.27 Exhaust.close @18:45:58.18 Pressure.reconfigure :default @18:45:58.26 Pressure.hold @18:45:58.49 SamplePump.reconfigure :charge @18:45:58.58 SamplePump.seek 2.3ml,42 seconds @18:46:06.68 Pressure.coast @18:46:06.76 Intake.close @18:46:07.65 Pressurized cartridge with 0.296ml to 14.0psi @18:46:07.66 Isolated @24.8psia (11.0psi) @18:46:07.69 Checking for cartridge leaks from 24.8psia @18:46:07.74 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @18:46:12.46 Cartridge pressure fell 4% to 23.9psia @18:46:12.73 Sampling 2000.0ml @18:46:12.76 SamplePump.setPosition! 0ml @18:46:12.98 Exhaust.open @18:46:13.76 Intake.open @18:46:14.99 Pressure.reconfigure :default @18:46:15.20 Pressure.hold @18:46:15.77 SamplePump.reconfigure :sample @18:46:16.32 SamplePump.seek 2000ml,2:47:12 @18:46:18.03 Cartridge 25 sampling 1/2000ml at 0.57ml/s with 14.1psi vacuum @18:47:18.14 Cartridge 25 sampling 32/2000ml at 0.52ml/s with 11.7psi vacuum @18:48:18.05 Cartridge 25 sampling 63/2000ml at 0.54ml/s with 12.3psi vacuum @18:49:18.21 Cartridge 25 sampling 97/2000ml at 0.56ml/s with 10.9psi vacuum @18:50:18.05 Cartridge 25 sampling 130/2000ml at 0.56ml/s with 11.8psi vacuum @18:51:18.23 Cartridge 25 sampling 165/2000ml at 0.58ml/s with 11.2psi vacuum @18:52:18.05 Cartridge 25 sampling 199/2000ml at 0.57ml/s with 10.5psi vacuum @18:53:18.27 Cartridge 25 sampling 233/2000ml at 0.56ml/s with 10.7psi vacuum @18:54:18.05 Cartridge 25 sampling 268/2000ml at 0.57ml/s with 12.5psi vacuum @18:55:18.20 Cartridge 25 sampling 302/2000ml at 0.57ml/s with 13.4psi vacuum @18:56:18.04 Cartridge 25 sampling 337/2000ml at 0.56ml/s with 12.5psi vacuum @18:57:18.22 Cartridge 25 sampling 371/2000ml at 0.56ml/s with 10.5psi vacuum @18:58:18.04 Cartridge 25 sampling 405/2000ml at 0.57ml/s with 13.9psi vacuum @18:59:18.04 Cartridge 25 sampling 440/2000ml at 0.58ml/s with 11.7psi vacuum @19:00:18.20 Cartridge 25 sampling 474/2000ml at 0.57ml/s with 12.0psi vacuum @19:01:18.04 Cartridge 25 sampling 509/2000ml at 0.56ml/s with 12.7psi vacuum @19:02:18.34 Cartridge 25 sampling 543/2000ml at 0.58ml/s with 15.7psi vacuum @19:03:18.05 Cartridge 25 sampling 578/2000ml at 0.57ml/s with 11.8psi vacuum @19:04:18.24 Cartridge 25 sampling 612/2000ml at 0.57ml/s with 11.5psi vacuum @19:05:18.04 Cartridge 25 sampling 647/2000ml at 0.57ml/s with 13.1psi vacuum @19:06:18.21 Cartridge 25 sampling 681/2000ml at 0.57ml/s with 14.3psi vacuum @19:07:18.05 Cartridge 25 sampling 716/2000ml at 0.58ml/s with 14.2psi vacuum @19:08:18.15 Cartridge 25 sampling 750/2000ml at 0.58ml/s with 12.9psi vacuum @19:09:18.04 Cartridge 25 sampling 786/2000ml at 0.59ml/s with 13.4psi vacuum @19:10:18.24 Cartridge 25 sampling 820/2000ml at 0.55ml/s with 11.4psi vacuum @19:11:18.04 Cartridge 25 sampling 853/2000ml at 0.56ml/s with 13.6psi vacuum @19:12:18.21 Cartridge 25 sampling 887/2000ml at 0.56ml/s with 12.2psi vacuum @19:13:18.04 Cartridge 25 sampling 920/2000ml at 0.56ml/s with 13.0psi vacuum @19:14:18.16 Cartridge 25 sampling 954/2000ml at 0.57ml/s with 11.3psi vacuum @19:15:18.05 Cartridge 25 sampling 988/2000ml at 0.56ml/s with 12.0psi vacuum @19:16:18.21 Cartridge 25 sampling 1021/2000ml at 0.56ml/s with 13.3psi vacuum @19:17:18.04 Cartridge 25 sampling 1055/2000ml at 0.55ml/s with 11.6psi vacuum @19:18:19.22 Cartridge 25 sampling 1089/2000ml at 0.56ml/s with 11.3psi vacuum @19:19:18.32 Cartridge 25 sampling 1122/2000ml at 0.56ml/s with 11.8psi vacuum @19:20:18.21 Cartridge 25 sampling 1156/2000ml at 0.55ml/s with 11.0psi vacuum @19:21:18.36 Cartridge 25 sampling 1189/2000ml at 0.56ml/s with 12.6psi vacuum @19:22:18.04 Cartridge 25 sampling 1223/2000ml at 0.57ml/s with 11.4psi vacuum @19:23:18.40 Cartridge 25 sampling 1256/2000ml at 0.55ml/s with 13.1psi vacuum @19:24:18.06 Cartridge 25 sampling 1290/2000ml at 0.56ml/s with 12.4psi vacuum @19:25:18.42 Cartridge 25 sampling 1324/2000ml at 0.55ml/s with 14.3psi vacuum @19:26:18.05 Cartridge 25 sampling 1357/2000ml at 0.55ml/s with 11.9psi vacuum @19:27:18.22 Cartridge 25 sampling 1391/2000ml at 0.56ml/s with 14.4psi vacuum @19:28:18.06 Cartridge 25 sampling 1424/2000ml at 0.55ml/s with 11.9psi vacuum @19:29:18.21 Cartridge 25 sampling 1458/2000ml at 0.55ml/s with 13.8psi vacuum @19:30:18.05 Cartridge 25 sampling 1492/2000ml at 0.56ml/s with 12.1psi vacuum @19:31:18.25 Cartridge 25 sampling 1525/2000ml at 0.56ml/s with 13.3psi vacuum @19:32:18.04 Cartridge 25 sampling 1559/2000ml at 0.56ml/s with 14.3psi vacuum @19:33:18.23 Cartridge 25 sampling 1593/2000ml at 0.56ml/s with 13.9psi vacuum @19:34:18.05 Cartridge 25 sampling 1627/2000ml at 0.56ml/s with 13.3psi vacuum @19:35:18.20 Cartridge 25 sampling 1661/2000ml at 0.56ml/s with 12.3psi vacuum @19:36:18.05 Cartridge 25 sampling 1695/2000ml at 0.57ml/s with 11.8psi vacuum @19:37:18.19 Cartridge 25 sampling 1728/2000ml at 0.56ml/s with 13.6psi vacuum @19:38:18.05 Cartridge 25 sampling 1762/2000ml at 0.56ml/s with 13.5psi vacuum @19:39:18.42 Cartridge 25 sampling 1796/2000ml at 0.56ml/s with 12.4psi vacuum @19:40:18.33 Cartridge 25 sampling 1830/2000ml at 0.55ml/s with 11.3psi vacuum @19:41:18.09 Cartridge 25 sampling 1864/2000ml at 0.56ml/s with 11.3psi vacuum @19:42:18.30 Cartridge 25 sampling 1898/2000ml at 0.56ml/s with 11.7psi vacuum @19:43:18.03 Cartridge 25 sampling 1932/2000ml at 0.56ml/s with 11.8psi vacuum @19:44:18.59 Cartridge 25 sampling 1965/2000ml at 0.56ml/s with 14.4psi vacuum @19:45:18.03 Cartridge 25 sampling 1999/2000ml at 0.56ml/s with 12.6psi vacuum @19:45:19.33 Pressure.coast @19:45:19.50 Intake.close @19:45:19.55 Exhaust.close @19:45:19.79 Toroid.reconfigure :hold @19:45:19.87 Toroid.seek :bypass @19:45:21.51 Toroid.reconfigure :default @19:45:21.60 Toroid.seek :clear @19:45:23.26 Sampled 2000.0ml @19:45:24.56 Plunge.seek :reagent @19:45:25.52 Sample loop closed. Ready to Deliver Reagents @19:45:26.10 Carousel.seek 25 @19:45:26.76 Plunge.seek :locked @19:45:27.91 Air.seek :clear @19:45:28.10 Lysis1.seek :clear @19:45:31.71 Archive.coast @19:45:31.95 Air.coast @19:45:32.12 Lysis1.coast @19:45:32.40 Lysis2.coast @19:45:32.61 Plunge.coast @19:45:32.88 SamplePump.coast @19:45:33.07 Toroid.coast @19:45:33.30 Carousel.coast @19:45:33.71 Gate.power :motor,:OFF @19:45:33.75 Gate.power :core,:OFF @19:45:33.81 Gate.power :analytic1,:OFF @19:45:33.89 Gate.power :analytic2,:OFF @19:45:33.96 Gate.power :analytic3,:OFF @19:45:33.97 Gate.power :main,:OFF @19:45:34.05 Safely stopped and ready to power off @19:45:34.40 delay 2:00 @20:13:40.21 -> Cartridge.dry @20:13:54.36 -> slots @20:16:02.53 -> slots @20:16:18.08 -> Cartridge.release 9 @20:16:22.57 -> Cartridge.dry @20:26:26.85 -> delay '2:00:00'; shoot @20:26:26.88 delay 2:00:00 @22:22:57.11 Session Begun @22:23:15.51 -> BACsampler.vacuum=18..26 @22:23:24.13 -> BACsampler.vacuum @22:23:36.85 -> exit @22:23:36.89 Session Ended @22:26:27.05 Gate.power :main,:ON,limit to 5000mA @22:26:27.07 Gate.power :analytic1,:ON,limit to 5000mA @22:26:27.12 Gate.power :analytic2,:ON,limit to 5000mA @22:26:27.14 Gate.power :analytic3,:ON,limit to 5000mA @22:26:27.22 Powering core bus @22:26:29.90 Archive.configure :default @22:26:30.34 Air.configure :default @22:26:30.86 Lysis1.configure :default @22:26:31.28 Lysis2.configure :default @22:26:31.67 Plunge.configure :default @22:26:33.47 SamplePump.configure :default @22:26:33.91 ExtraPump.configure :default @22:26:34.30 Toroid.configure :default @22:26:34.63 Carousel.configure :default @22:26:34.96 Pressure.configure :default @22:26:35.04 All elf controllers running firmware version 4.61 @22:26:35.09 Gate.power :motor,:ON,limit to 5000mA @22:26:36.15 Toroid.reconfigure :default @22:26:36.19 Plunge.reconfigure :default @22:26:36.25 Lysis2.reconfigure :default @22:26:36.29 Lysis1.reconfigure :default @22:26:36.35 Archive.reconfigure :default @22:26:36.40 Air.reconfigure :default @22:26:36.70 Toroid.home.jog 0 @22:26:36.74 Plunge.home.jog 0 @22:26:36.80 Lysis2.home.jog 0 @22:26:36.83 Lysis1.home.jog 0 @22:26:36.88 Archive.home.jog 0 @22:26:36.95 Air.home.jog 0 @22:26:37.70 Plunge.reconfigure :homing @22:26:37.75 Lysis2.reconfigure :homing @22:26:37.79 Toroid.reconfigure :homing @22:26:37.86 Lysis1.reconfigure :homing @22:26:38.04 Plunge.jog 848 @22:26:38.12 Lysis2.jog 1358 @22:26:38.20 Toroid.jog 620 @22:26:38.25 Lysis1.jog 1350 @22:26:38.52 Air.reconfigure :homing @22:26:38.55 Archive.reconfigure :homing @22:26:38.69 Archive.jog 1361 @22:26:38.72 Air.jog 1365 @22:26:39.09 Plunge.setRawPosition! 50 @22:26:39.16 Plunge.seek :home @22:26:39.46 Plunge.reconfigure :default @22:26:39.48 Lysis2.setRawPosition! 50 @22:26:39.62 Lysis2.seek :home @22:26:39.76 Lysis1.setRawPosition! 50 @22:26:39.83 Lysis1.seek :home @22:26:39.98 Air.setRawPosition! 50 @22:26:40.02 Archive.setRawPosition! 50 @22:26:40.06 Lysis2.reconfigure :default @22:26:40.69 Air.seek :home @22:26:40.76 Archive.seek :home @22:26:41.69 Toroid.setRawPosition! 0 @22:26:41.76 Lysis1.reconfigure :default @22:26:41.91 Archive.reconfigure :default @22:26:42.00 Air.reconfigure :default @22:26:42.47 Toroid.seek :home @22:26:42.99 Toroid.reconfigure :default @22:26:43.51 Toroid.seek :clear @22:26:45.28 Carousel.reconfigure :default @22:26:45.58 Carousel.home.jog 0 @22:26:46.19 Carousel.seek 25 @22:26:46.40 Selecting Cartridge 9 @22:26:46.66 Carousel.seek 9 @22:26:50.67 Lysis1.seek :locked @22:26:54.04 Plunge.seek :locked @22:26:54.37 Air.seek :locked @22:26:58.65 Mounted archive Cartridge @22:26:59.17 Duration of filtering limited to 4:05:00 @22:26:59.49 Priming sample loop w/160ml, bypass w/10ml @22:26:59.51 SamplePump.setPosition! 0ml @22:26:59.59 Bypass.open @22:27:01.18 Intake.open @22:27:02.35 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @22:27:04.96 Vacuum tare is 18% of full scale @22:27:04.98 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @22:27:05.24 Pressure.reconfigure :default @22:27:05.32 Pressure.hold @22:27:05.55 SamplePump.reconfigure :prime @22:27:05.64 SamplePump.seek 10ml,1:21 @22:27:16.76 Pressure.coast @22:27:16.83 SamplePump.setPosition! 0ml @22:27:16.91 Exhaust.open @22:27:18.34 Bypass.close @22:27:20.12 Pressure.reconfigure :default @22:27:20.30 Pressure.hold @22:27:20.94 SamplePump.reconfigure :prime @22:27:21.10 SamplePump.seek 160ml,13:51 @22:30:12.68 Pressure.coast @22:30:12.76 Bypass.open @22:30:13.47 Intake.close @22:30:14.80 Toroid.reconfigure :hold @22:30:14.92 Toroid.seek :crack @22:30:16.79 Toroid.seek :cartridge @22:30:17.78 Toroid.reconfigure :default @22:30:17.86 Toroid.seek :clear @22:30:20.32 Pressurizing cartridge to 26psi above 14.2psia @22:30:20.34 SamplePump.setPosition! 0ml @22:30:20.42 Intake.open @22:30:21.17 Bypass.close @22:30:21.92 Exhaust.close @22:30:22.79 Pressure.reconfigure :default @22:30:22.87 Pressure.hold @22:30:23.11 SamplePump.reconfigure :charge @22:30:23.20 SamplePump.seek 2.3ml,42 seconds @22:30:38.10 Pressure.coast @22:30:38.18 Intake.close @22:30:38.99 Pressurized cartridge with 1.008ml to 26.2psi @22:30:39.00 Isolated @37.0psia (22.9psi) @22:30:39.03 Checking for cartridge leaks from 37.0psia @22:30:39.07 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @22:30:43.73 Cartridge pressure fell 1% to 36.7psia @22:30:43.94 Sampling 2000.0ml @22:30:43.97 SamplePump.setPosition! 0ml @22:30:44.19 Exhaust.open @22:30:44.94 Intake.open @22:30:46.12 Pressure.reconfigure :default @22:30:46.27 Pressure.hold @22:30:46.92 SamplePump.reconfigure :sample @22:30:47.18 SamplePump.seek 2000ml,2:47:12 @22:30:49.08 Cartridge 9 sampling 1/2000ml at 0.63ml/s with 24.7psi vacuum @22:31:49.22 Cartridge 9 sampling 30/2000ml at 0.49ml/s with 24.9psi vacuum @22:32:49.10 Cartridge 9 sampling 60/2000ml at 0.49ml/s with 22.4psi vacuum @22:33:49.35 Cartridge 9 sampling 89/2000ml at 0.48ml/s with 20.9psi vacuum @22:34:49.10 Cartridge 9 sampling 118/2000ml at 0.47ml/s with 22.9psi vacuum @22:35:49.25 Cartridge 9 sampling 147/2000ml at 0.49ml/s with 23.1psi vacuum @22:36:49.09 Cartridge 9 sampling 176/2000ml at 0.48ml/s with 24.1psi vacuum @22:37:49.15 Cartridge 9 sampling 204/2000ml at 0.49ml/s with 24.3psi vacuum @22:38:49.10 Cartridge 9 sampling 233/2000ml at 0.49ml/s with 20.3psi vacuum @22:39:49.16 Cartridge 9 sampling 262/2000ml at 0.49ml/s with 25.7psi vacuum @22:40:49.09 Cartridge 9 sampling 291/2000ml at 0.47ml/s with 26.3psi vacuum @22:41:49.31 Cartridge 9 sampling 318/2000ml at 0.46ml/s with 20.3psi vacuum @22:42:49.30 Cartridge 9 sampling 345/2000ml at 0.46ml/s with 24.9psi vacuum @22:43:49.09 Cartridge 9 sampling 372/2000ml at 0.44ml/s with 21.6psi vacuum @22:44:49.38 Cartridge 9 sampling 399/2000ml at 0.43ml/s with 24.6psi vacuum @22:45:49.09 Cartridge 9 sampling 425/2000ml at 0.42ml/s with 19.8psi vacuum @22:46:49.47 Cartridge 9 sampling 451/2000ml at 0.41ml/s with 28.3psi vacuum @22:47:49.09 Cartridge 9 sampling 477/2000ml at 0.44ml/s with 20.1psi vacuum @22:48:49.34 Cartridge 9 sampling 503/2000ml at 0.45ml/s with 25.1psi vacuum @22:49:49.09 Cartridge 9 sampling 529/2000ml at 0.43ml/s with 25.9psi vacuum @22:50:49.19 Cartridge 9 sampling 554/2000ml at 0.41ml/s with 25.4psi vacuum @22:51:49.10 Cartridge 9 sampling 578/2000ml at 0.39ml/s with 28.1psi vacuum @22:52:49.25 Cartridge 9 sampling 603/2000ml at 0.41ml/s with 24.0psi vacuum @22:53:49.09 Cartridge 9 sampling 627/2000ml at 0.40ml/s with 20.2psi vacuum @22:54:49.33 Cartridge 9 sampling 651/2000ml at 0.41ml/s with 26.1psi vacuum @22:55:49.09 Cartridge 9 sampling 675/2000ml at 0.37ml/s with 20.7psi vacuum @22:56:49.26 Cartridge 9 sampling 698/2000ml at 0.40ml/s with 22.4psi vacuum @22:57:49.09 Cartridge 9 sampling 721/2000ml at 0.38ml/s with 27.6psi vacuum @22:58:49.32 Cartridge 9 sampling 744/2000ml at 0.39ml/s with 24.8psi vacuum @22:59:49.09 Cartridge 9 sampling 767/2000ml at 0.38ml/s with 27.2psi vacuum @23:00:49.20 Cartridge 9 sampling 789/2000ml at 0.34ml/s with 28.0psi vacuum @23:01:49.09 Cartridge 9 sampling 812/2000ml at 0.38ml/s with 24.5psi vacuum @23:02:49.45 Cartridge 9 sampling 834/2000ml at 0.36ml/s with 18.8psi vacuum @23:03:49.30 Cartridge 9 sampling 857/2000ml at 0.35ml/s with 17.7psi vacuum 2.5m deep (18.4psia) @23:04:49.12 Cartridge 9 sampling 878/2000ml at 0.33ml/s with 28.2psi vacuum @23:05:49.27 Cartridge 9 sampling 900/2000ml at 0.34ml/s with 17.6psi vacuum 3.3m deep (19.6psia) @23:06:49.08 Cartridge 9 sampling 921/2000ml at 0.37ml/s with 22.1psi vacuum @23:07:49.30 Cartridge 9 sampling 942/2000ml at 0.33ml/s with 16.9psi vacuum 4.0m deep (20.6psia) @23:08:49.13 Cartridge 9 sampling 963/2000ml at 0.36ml/s with 25.0psi vacuum @23:09:49.32 Cartridge 9 sampling 983/2000ml at 0.36ml/s with 25.9psi vacuum @23:10:49.08 Cartridge 9 sampling 1004/2000ml at 0.35ml/s with 23.5psi vacuum @23:11:49.27 Cartridge 9 sampling 1024/2000ml at 0.35ml/s with 26.5psi vacuum @23:12:49.09 Cartridge 9 sampling 1044/2000ml at 0.32ml/s with 29.3psi vacuum @23:13:49.32 Cartridge 9 sampling 1063/2000ml at 0.32ml/s with 17.1psi vacuum 5.5m deep (22.8psia) @23:14:49.09 Cartridge 9 sampling 1083/2000ml at 0.33ml/s with 24.6psi vacuum @23:15:49.32 Cartridge 9 sampling 1102/2000ml at 0.32ml/s with 18.4psi vacuum 4.8m deep (21.8psia) @23:16:49.08 Cartridge 9 sampling 1121/2000ml at 0.28ml/s with 29.1psi vacuum @23:17:49.26 Cartridge 9 sampling 1140/2000ml at 0.31ml/s with 16.7psi vacuum 5.1m deep (22.2psia) @23:18:49.08 Cartridge 9 sampling 1158/2000ml at 0.27ml/s with 27.3psi vacuum @23:19:49.26 Cartridge 9 sampling 1177/2000ml at 0.29ml/s with 29.3psi vacuum @23:20:49.09 Cartridge 9 sampling 1195/2000ml at 0.28ml/s with 13.1psi vacuum @23:21:49.22 Cartridge 9 sampling 1213/2000ml at 0.29ml/s with 16.8psi vacuum 2.7m deep (18.7psia) @23:22:49.09 Cartridge 9 sampling 1231/2000ml at 0.30ml/s with 19.2psi vacuum @23:23:49.26 Cartridge 9 sampling 1248/2000ml at 0.32ml/s with 23.5psi vacuum 1.7m deep (17.1psia) @23:24:49.08 Cartridge 9 sampling 1266/2000ml at 0.25ml/s with 22.6psi vacuum @23:25:49.15 Cartridge 9 sampling 1283/2000ml at 0.30ml/s with 22.9psi vacuum @23:26:49.08 Cartridge 9 sampling 1299/2000ml at 0.24ml/s with 28.4psi vacuum @23:27:49.12 Cartridge 9 sampling 1316/2000ml at 0.24ml/s with 29.2psi vacuum @23:28:49.30 Cartridge 9 sampling 1333/2000ml at 0.24ml/s with 15.4psi vacuum 1.7m deep (17.2psia) @23:29:49.09 Cartridge 9 sampling 1349/2000ml at 0.29ml/s with 19.2psi vacuum 2.2m deep (17.9psia) @23:30:49.49 Cartridge 9 sampling 1365/2000ml at 0.29ml/s with 20.0psi vacuum 4.0m deep (20.6psia) @23:31:49.09 Cartridge 9 sampling 1381/2000ml at 0.28ml/s with 29.7psi vacuum @23:32:49.42 Cartridge 9 sampling 1397/2000ml at 0.26ml/s with 29.5psi vacuum @23:33:49.09 Cartridge 9 sampling 1413/2000ml at 0.29ml/s with 25.6psi vacuum @23:34:49.42 Cartridge 9 sampling 1428/2000ml at 0.28ml/s with 25.2psi vacuum @23:35:49.10 Cartridge 9 sampling 1443/2000ml at 0.21ml/s with 12.8psi vacuum @23:36:49.25 Cartridge 9 sampling 1458/2000ml at 0.26ml/s with 28.7psi vacuum @23:37:49.09 Cartridge 9 sampling 1473/2000ml at 0.22ml/s with 13.4psi vacuum @23:38:49.26 Cartridge 9 sampling 1488/2000ml at 0.23ml/s with 14.6psi vacuum 2.3m deep (18.0psia) @23:39:49.10 Cartridge 9 sampling 1502/2000ml at 0.21ml/s with 18.5psi vacuum @23:40:49.21 Cartridge 9 sampling 1517/2000ml at 0.27ml/s with 28.1psi vacuum @23:41:49.10 Cartridge 9 sampling 1531/2000ml at 0.24ml/s with 13.0psi vacuum 1.8m deep (17.4psia) @23:42:49.34 Cartridge 9 sampling 1546/2000ml at 0.25ml/s with 18.3psi vacuum 3.3m deep (19.6psia) @23:43:49.09 Cartridge 9 sampling 1560/2000ml at 0.25ml/s with 17.9psi vacuum @23:44:49.29 Cartridge 9 sampling 1574/2000ml at 0.25ml/s with 29.1psi vacuum @23:45:49.10 Cartridge 9 sampling 1588/2000ml at 0.25ml/s with 25.9psi vacuum @23:46:49.20 Cartridge 9 sampling 1602/2000ml at 0.20ml/s with 30.7psi vacuum @23:47:49.09 Cartridge 9 sampling 1617/2000ml at 0.23ml/s with 10.2psi vacuum 1.6m deep (17.0psia) @23:48:49.12 Cartridge 9 sampling 1631/2000ml at 0.23ml/s with 15.9psi vacuum @23:49:49.09 Cartridge 9 sampling 1645/2000ml at 0.27ml/s with 21.9psi vacuum @23:50:49.19 Cartridge 9 sampling 1659/2000ml at 0.20ml/s with 30.8psi vacuum @23:51:49.36 Cartridge 9 sampling 1673/2000ml at 0.27ml/s with 18.8psi vacuum 6.1m deep (23.7psia) @23:52:49.09 Cartridge 9 sampling 1687/2000ml at 0.23ml/s with 28.3psi vacuum @23:53:49.45 Cartridge 9 sampling 1701/2000ml at 0.21ml/s with 15.1psi vacuum 5.4m deep (22.7psia) @23:54:49.10 Cartridge 9 sampling 1714/2000ml at 0.20ml/s with 30.7psi vacuum @23:55:49.28 Cartridge 9 sampling 1728/2000ml at 0.22ml/s with 28.6psi vacuum @23:56:49.10 Cartridge 9 sampling 1742/2000ml at 0.21ml/s with 29.0psi vacuum @23:57:49.36 Cartridge 9 sampling 1755/2000ml at 0.25ml/s with 22.6psi vacuum 2.3m deep (18.1psia) @23:58:49.09 Cartridge 9 sampling 1769/2000ml at 0.20ml/s with 29.6psi vacuum @23:59:49.34 Cartridge 9 sampling 1782/2000ml at 0.20ml/s with 29.0psi vacuum @00:00:49.09PDT28-May-21 Cartridge 9 sampling 1796/2000ml at 0.24ml/s with 27.4psi vacuum @00:01:49.23 Cartridge 9 sampling 1809/2000ml at 0.23ml/s with 27.6psi vacuum @00:02:49.10 Cartridge 9 sampling 1823/2000ml at 0.20ml/s with 29.4psi vacuum @00:03:49.22 Cartridge 9 sampling 1836/2000ml at 0.20ml/s with 30.2psi vacuum @00:04:49.10 Cartridge 9 sampling 1849/2000ml at 0.22ml/s with 30.8psi vacuum @00:05:49.20 Cartridge 9 sampling 1862/2000ml at 0.26ml/s with 25.5psi vacuum 2.4m deep (18.2psia) @00:06:49.09 Cartridge 9 sampling 1876/2000ml at 0.21ml/s with 13.9psi vacuum 1.9m deep (17.6psia) @00:07:49.38 Cartridge 9 sampling 1889/2000ml at 0.20ml/s with 32.5psi vacuum @00:08:49.10 Cartridge 9 sampling 1902/2000ml at 0.20ml/s with 30.4psi vacuum @00:09:49.20 Cartridge 9 sampling 1915/2000ml at 0.20ml/s with 32.3psi vacuum @00:10:49.25 Cartridge 9 sampling 1928/2000ml at 0.23ml/s with 28.8psi vacuum @00:11:49.18 Cartridge 9 sampling 1941/2000ml at 0.20ml/s with 31.4psi vacuum @00:12:49.09 Cartridge 9 sampling 1954/2000ml at 0.20ml/s with 20.6psi vacuum @00:13:49.15 Cartridge 9 sampling 1967/2000ml at 0.20ml/s with 19.0psi vacuum @00:14:49.39 Cartridge 9 sampling 1980/2000ml at 0.20ml/s with 30.4psi vacuum @00:15:49.09 Cartridge 9 sampling 1993/2000ml at 0.20ml/s with 32.1psi vacuum @00:16:21.58 Pressure.coast @00:16:21.98 Intake.close @00:16:22.12 Exhaust.close @00:16:22.65 Toroid.reconfigure :hold @00:16:22.90 Toroid.seek :bypass @00:16:24.78 Toroid.reconfigure :default @00:16:24.85 Toroid.seek :clear @00:16:26.52 Sampled 2000.0ml @00:16:27.03 SamplePump.setPosition! 0ml @00:16:27.11 Plunge.seek :reagent @00:16:27.14 Exhaust.open @00:16:27.92 Intake.open @00:16:28.08 Sample loop closed. Ready to Deliver Reagents @00:16:28.10 Archive.seek :sealed @00:16:28.81 SamplePump.reconfigure :slow @00:16:28.89 SamplePump.seek -35ml @00:16:31.78 Archive.seek :evac @00:16:45.90 delay 1:00 @00:17:44.06 delay 1 second @00:17:45.15 Bypass.open @00:17:45.98 Archive.seek :clear @00:17:46.16 Exhaust.close @00:17:47.17 SamplePump.seek -33ml @00:17:51.77 delay 1 second @00:17:52.85 SamplePump.seek -41ml @00:18:03.55 RNAlater soak for 10:00 @00:18:10.24 delay 1 second @00:18:11.32 Intake.close @00:18:11.37 Bypass.close @00:28:03.60 Archive.seek :tVent @00:28:05.27 delay 1 second @00:28:06.30 Archive.seek :evac @00:28:22.26 delay 10 seconds @00:28:32.34 Sample preserved and evacuated @00:28:32.66 Carousel.seek 9 @00:28:32.89 Plunge.seek :locked @00:28:33.90 Archive.seek :clear @00:28:34.09 Air.seek :clear @00:28:34.27 Lysis1.seek :clear @00:28:51.10 Archive.coast @00:28:51.16 Air.coast @00:28:51.23 Lysis1.coast @00:28:51.83 Lysis2.coast @00:28:51.92 Plunge.coast @00:28:51.99 SamplePump.coast @00:28:52.06 Toroid.coast @00:28:52.14 Carousel.coast @00:28:52.35 Gate.power :motor,:OFF @00:28:52.38 Gate.power :core,:OFF @00:28:52.41 Gate.power :analytic1,:OFF @00:28:52.44 Gate.power :analytic2,:OFF @00:28:52.47 Gate.power :analytic3,:OFF @00:28:52.50 Gate.power :main,:OFF @00:28:52.54 Safely stopped and ready to power off @05:59:32.67 -> Cartridge.dry @05:59:46.80 -> Cartridge.release 8 @05:59:54.99 -> Cartridge.dry @06:10:41.75 -> delay '1:00:00';shoot @06:10:41.79 delay 1:00:00 @07:10:42.00 Gate.power :main,:ON,limit to 5000mA @07:10:42.03 Gate.power :analytic1,:ON,limit to 5000mA @07:10:42.07 Gate.power :analytic2,:ON,limit to 5000mA @07:10:42.10 Gate.power :analytic3,:ON,limit to 5000mA @07:10:42.17 Powering core bus @07:10:44.46 Archive.configure :default @07:10:44.62 Air.configure :default @07:10:44.76 Lysis1.configure :default @07:10:45.12 Lysis2.configure :default @07:10:45.56 Plunge.configure :default @07:10:46.16 SamplePump.configure :default @07:10:46.60 ExtraPump.configure :default @07:10:47.00 Toroid.configure :default @07:10:47.50 Carousel.configure :default @07:10:47.92 Pressure.configure :default @07:10:48.07 All elf controllers running firmware version 4.61 @07:10:48.14 Gate.power :motor,:ON,limit to 5000mA @07:10:49.63 Archive.reconfigure :default @07:10:49.67 Toroid.reconfigure :default @07:10:49.81 Lysis2.reconfigure :default @07:10:49.96 Lysis1.reconfigure :default @07:10:50.02 Air.reconfigure :default @07:10:50.15 Plunge.reconfigure :default @07:10:50.84 Archive.home.jog 0 @07:10:50.95 Toroid.home.jog 0 @07:10:51.05 Air.home.jog 0 @07:10:51.14 Lysis2.home.jog 0 @07:10:51.18 Lysis1.home.jog 0 @07:10:51.22 Plunge.home.jog 0 @07:10:52.09 Toroid.reconfigure :homing @07:10:52.13 Plunge.reconfigure :homing @07:10:52.17 Lysis1.reconfigure :homing @07:10:52.22 Air.reconfigure :homing @07:10:52.28 Archive.reconfigure :homing @07:10:52.34 Lysis2.reconfigure :homing @07:10:52.61 Plunge.jog 855 @07:10:52.66 Toroid.jog 622 @07:10:52.69 Lysis1.jog 1351 @07:10:52.73 Lysis2.jog 1363 @07:10:52.78 Air.jog 1367 @07:10:52.82 Archive.jog 154 @07:10:53.49 Archive.setRawPosition! 50 @07:10:53.56 Archive.seek :home @07:10:53.79 Plunge.setRawPosition! 50 @07:10:53.87 Plunge.seek :home @07:10:53.97 Archive.reconfigure :default @07:10:54.13 Lysis1.setRawPosition! 50 @07:10:54.22 Lysis1.seek :home @07:10:54.34 Lysis2.setRawPosition! 50 @07:10:54.37 Air.setRawPosition! 50 @07:10:54.40 Plunge.reconfigure :default @07:10:54.55 Lysis2.seek :home @07:10:54.61 Air.seek :home @07:10:54.83 Lysis1.reconfigure :default @07:10:54.95 Lysis2.reconfigure :default @07:10:55.07 Toroid.setRawPosition! 0 @07:10:55.13 Air.reconfigure :default @07:10:55.19 Toroid.seek :home @07:10:55.41 Toroid.reconfigure :default @07:10:56.20 Toroid.seek :clear @07:10:57.96 Carousel.reconfigure :default @07:10:58.15 Carousel.home.jog 0 @07:10:58.77 Carousel.seek 9 @07:10:59.30 Selecting Cartridge 8 @07:10:59.81 Carousel.seek 8 @07:11:01.50 Lysis1.seek :locked @07:11:04.76 Plunge.seek :locked @07:11:04.94 Air.seek :locked @07:11:09.47 Mounted archive Cartridge @07:11:10.37 Duration of filtering limited to 4:05:00 @07:11:10.69 Priming sample loop w/160ml, bypass w/10ml @07:11:10.72 SamplePump.setPosition! 0ml @07:11:10.79 Bypass.open @07:11:12.29 Intake.open @07:11:13.22 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @07:11:16.05 Vacuum tare is 17% of full scale @07:11:16.13 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @07:11:16.76 Pressure.reconfigure :default @07:11:16.91 Pressure.hold @07:11:17.47 SamplePump.reconfigure :prime @07:11:17.66 SamplePump.seek 10ml,1:21 @07:11:28.82 Pressure.coast @07:11:28.92 SamplePump.setPosition! 0ml @07:11:29.08 Exhaust.open @07:11:30.58 Bypass.close @07:11:32.15 Pressure.reconfigure :default @07:11:32.23 Pressure.hold @07:11:32.52 SamplePump.reconfigure :prime @07:11:32.61 SamplePump.seek 160ml,13:51 @07:14:24.10 Pressure.coast @07:14:24.18 Bypass.open @07:14:24.88 Intake.close @07:14:25.81 Toroid.reconfigure :hold @07:14:25.89 Toroid.seek :crack @07:14:27.51 Toroid.seek :cartridge @07:14:28.52 Toroid.reconfigure :default @07:14:28.80 Toroid.seek :clear @07:14:31.03 Pressurizing cartridge to 26psi above 14.5psia @07:14:31.08 SamplePump.setPosition! 0ml @07:14:31.27 Intake.open @07:14:32.20 Bypass.close @07:14:33.03 Exhaust.close @07:14:33.87 Pressure.reconfigure :default @07:14:33.95 Pressure.hold @07:14:34.18 SamplePump.reconfigure :charge @07:14:34.27 SamplePump.seek 2.3ml,42 seconds @07:14:49.26 Pressure.coast @07:14:49.40 Intake.close @07:14:50.53 Pressurized cartridge with 1.016ml to 26.2psi @07:14:50.57 Isolated @35.6psia (21.1psi) @07:14:50.60 Checking for cartridge leaks from 35.6psia @07:14:50.78 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @07:14:55.39 Cartridge pressure fell 1% to 35.3psia @07:14:55.50 Sampling 2000.0ml @07:14:55.52 SamplePump.setPosition! 0ml @07:14:55.60 Exhaust.open @07:14:56.32 Intake.open @07:14:57.49 Pressure.reconfigure :default @07:14:57.71 Pressure.hold @07:14:58.49 SamplePump.reconfigure :sample @07:14:58.72 SamplePump.seek 2000ml,2:47:12 @07:15:00.59 Cartridge 8 sampling 1/2000ml at 0.61ml/s with 24.5psi vacuum @07:16:00.39 Cartridge 8 sampling 28/2000ml at 0.46ml/s with 24.2psi vacuum @07:17:00.57 Cartridge 8 sampling 55/2000ml at 0.46ml/s with 23.6psi vacuum @07:18:00.39 Cartridge 8 sampling 82/2000ml at 0.44ml/s with 23.0psi vacuum @07:19:00.42 Cartridge 8 sampling 110/2000ml at 0.46ml/s with 19.2psi vacuum @07:20:00.40 Cartridge 8 sampling 138/2000ml at 0.46ml/s with 24.1psi vacuum @07:21:00.93 Cartridge 8 sampling 166/2000ml at 0.46ml/s with 24.9psi vacuum @07:22:00.40 Cartridge 8 sampling 194/2000ml at 0.46ml/s with 26.4psi vacuum @07:23:00.62 Cartridge 8 sampling 221/2000ml at 0.46ml/s with 24.2psi vacuum @07:24:00.94 Cartridge 8 sampling 248/2000ml at 0.45ml/s with 24.5psi vacuum @07:25:00.63 Cartridge 8 sampling 275/2000ml at 0.46ml/s with 24.2psi vacuum @07:26:00.70 Cartridge 8 sampling 302/2000ml at 0.46ml/s with 23.3psi vacuum @07:27:00.39 Cartridge 8 sampling 328/2000ml at 0.43ml/s with 24.7psi vacuum @07:28:00.64 Cartridge 8 sampling 355/2000ml at 0.43ml/s with 23.4psi vacuum @07:29:00.39 Cartridge 8 sampling 381/2000ml at 0.43ml/s with 21.7psi vacuum @07:30:00.64 Cartridge 8 sampling 407/2000ml at 0.43ml/s with 19.2psi vacuum @07:31:00.40 Cartridge 8 sampling 432/2000ml at 0.41ml/s with 26.7psi vacuum @07:32:00.67 Cartridge 8 sampling 458/2000ml at 0.42ml/s with 26.1psi vacuum @07:33:00.40 Cartridge 8 sampling 483/2000ml at 0.42ml/s with 22.7psi vacuum @07:34:00.61 Cartridge 8 sampling 508/2000ml at 0.41ml/s with 22.8psi vacuum @07:35:00.40 Cartridge 8 sampling 532/2000ml at 0.41ml/s with 22.9psi vacuum @07:36:00.70 Cartridge 8 sampling 557/2000ml at 0.39ml/s with 27.4psi vacuum @07:37:00.39 Cartridge 8 sampling 581/2000ml at 0.39ml/s with 23.1psi vacuum @07:38:00.53 Cartridge 8 sampling 605/2000ml at 0.39ml/s with 22.0psi vacuum @07:39:00.41 Cartridge 8 sampling 628/2000ml at 0.40ml/s with 23.1psi vacuum @07:40:00.51 Cartridge 8 sampling 652/2000ml at 0.37ml/s with 20.1psi vacuum @07:41:00.40 Cartridge 8 sampling 675/2000ml at 0.38ml/s with 27.1psi vacuum @07:42:00.62 Cartridge 8 sampling 698/2000ml at 0.38ml/s with 22.2psi vacuum @07:43:00.39 Cartridge 8 sampling 720/2000ml at 0.37ml/s with 20.8psi vacuum @07:44:00.65 Cartridge 8 sampling 743/2000ml at 0.37ml/s with 22.5psi vacuum @07:45:00.40 Cartridge 8 sampling 764/2000ml at 0.34ml/s with 19.4psi vacuum @07:46:00.78 Cartridge 8 sampling 786/2000ml at 0.33ml/s with 26.6psi vacuum @07:47:00.63 Cartridge 8 sampling 807/2000ml at 0.34ml/s with 22.5psi vacuum @07:48:00.40 Cartridge 8 sampling 828/2000ml at 0.34ml/s with 26.4psi vacuum @07:49:00.88 Cartridge 8 sampling 849/2000ml at 0.35ml/s with 21.1psi vacuum @07:50:00.40 Cartridge 8 sampling 869/2000ml at 0.35ml/s with 25.3psi vacuum @07:51:00.65 Cartridge 8 sampling 890/2000ml at 0.33ml/s with 26.8psi vacuum @07:52:00.39 Cartridge 8 sampling 910/2000ml at 0.31ml/s with 26.0psi vacuum @07:53:00.57 Cartridge 8 sampling 930/2000ml at 0.31ml/s with 26.9psi vacuum @07:54:00.40 Cartridge 8 sampling 949/2000ml at 0.34ml/s with 26.3psi vacuum @07:55:00.58 Cartridge 8 sampling 969/2000ml at 0.31ml/s with 21.2psi vacuum @07:56:00.40 Cartridge 8 sampling 988/2000ml at 0.30ml/s with 25.4psi vacuum @07:57:00.67 Cartridge 8 sampling 1007/2000ml at 0.34ml/s with 24.3psi vacuum @07:58:00.39 Cartridge 8 sampling 1026/2000ml at 0.29ml/s with 19.3psi vacuum @07:59:00.53 Cartridge 8 sampling 1045/2000ml at 0.31ml/s with 22.7psi vacuum @08:00:00.39 Cartridge 8 sampling 1064/2000ml at 0.31ml/s with 26.9psi vacuum @08:01:00.57 Cartridge 8 sampling 1082/2000ml at 0.30ml/s with 23.2psi vacuum @08:02:00.39 Cartridge 8 sampling 1101/2000ml at 0.26ml/s with 27.7psi vacuum @08:03:00.63 Cartridge 8 sampling 1119/2000ml at 0.31ml/s with 26.0psi vacuum @08:04:00.40 Cartridge 8 sampling 1136/2000ml at 0.31ml/s with 25.8psi vacuum @08:05:00.55 Cartridge 8 sampling 1154/2000ml at 0.31ml/s with 25.9psi vacuum @08:06:00.39 Cartridge 8 sampling 1171/2000ml at 0.26ml/s with 27.2psi vacuum @08:07:00.50 Cartridge 8 sampling 1188/2000ml at 0.27ml/s with 27.5psi vacuum @08:08:00.50 Cartridge 8 sampling 1205/2000ml at 0.24ml/s with 19.4psi vacuum @08:09:00.55 Cartridge 8 sampling 1221/2000ml at 0.27ml/s with 20.5psi vacuum @08:10:00.55 Cartridge 8 sampling 1238/2000ml at 0.30ml/s with 23.7psi vacuum @08:11:00.39 Cartridge 8 sampling 1254/2000ml at 0.25ml/s with 20.0psi vacuum @08:12:00.82 Cartridge 8 sampling 1270/2000ml at 0.28ml/s with 22.2psi vacuum @08:13:00.39 Cartridge 8 sampling 1286/2000ml at 0.29ml/s with 25.6psi vacuum @08:14:00.65 Cartridge 8 sampling 1301/2000ml at 0.26ml/s with 24.3psi vacuum @08:15:00.39 Cartridge 8 sampling 1317/2000ml at 0.21ml/s with 27.4psi vacuum @08:16:00.53 Cartridge 8 sampling 1332/2000ml at 0.23ml/s with 17.6psi vacuum @08:17:00.40 Cartridge 8 sampling 1347/2000ml at 0.26ml/s with 27.7psi vacuum @08:18:00.49 Cartridge 8 sampling 1362/2000ml at 0.25ml/s with 27.9psi vacuum @08:19:00.39 Cartridge 8 sampling 1377/2000ml at 0.26ml/s with 19.8psi vacuum @08:20:00.55 Cartridge 8 sampling 1392/2000ml at 0.27ml/s with 24.7psi vacuum @08:21:00.39 Cartridge 8 sampling 1406/2000ml at 0.26ml/s with 21.1psi vacuum @08:22:00.61 Cartridge 8 sampling 1421/2000ml at 0.23ml/s with 29.4psi vacuum @08:23:00.40 Cartridge 8 sampling 1435/2000ml at 0.26ml/s with 29.1psi vacuum @08:24:00.58 Cartridge 8 sampling 1449/2000ml at 0.21ml/s with 26.6psi vacuum @08:25:00.39 Cartridge 8 sampling 1463/2000ml at 0.21ml/s with 21.0psi vacuum @08:26:00.55 Cartridge 8 sampling 1477/2000ml at 0.26ml/s with 24.9psi vacuum @08:27:00.40 Cartridge 8 sampling 1490/2000ml at 0.26ml/s with 24.2psi vacuum @08:28:00.47 Cartridge 8 sampling 1504/2000ml at 0.22ml/s with 25.2psi vacuum @08:29:00.39 Cartridge 8 sampling 1517/2000ml at 0.25ml/s with 27.8psi vacuum @08:30:00.83 Cartridge 8 sampling 1531/2000ml at 0.25ml/s with 27.1psi vacuum @08:31:00.89 Cartridge 8 sampling 1544/2000ml at 0.20ml/s with 26.8psi vacuum @08:32:00.65 Cartridge 8 sampling 1557/2000ml at 0.23ml/s with 24.5psi vacuum @08:33:00.71 Cartridge 8 sampling 1571/2000ml at 0.22ml/s with 17.3psi vacuum 2.7m deep (18.7psia) @08:34:00.40 Cartridge 8 sampling 1584/2000ml at 0.22ml/s with 21.3psi vacuum @08:35:00.71 Cartridge 8 sampling 1597/2000ml at 0.22ml/s with 22.2psi vacuum @08:36:00.39 Cartridge 8 sampling 1610/2000ml at 0.25ml/s with 27.1psi vacuum @08:37:00.80 Cartridge 8 sampling 1623/2000ml at 0.21ml/s with 25.0psi vacuum @08:38:00.39 Cartridge 8 sampling 1636/2000ml at 0.23ml/s with 26.2psi vacuum @08:39:00.92 Cartridge 8 sampling 1648/2000ml at 0.20ml/s with 29.2psi vacuum @08:40:00.40 Cartridge 8 sampling 1661/2000ml at 0.20ml/s with 30.5psi vacuum @08:41:00.68 Cartridge 8 sampling 1674/2000ml at 0.20ml/s with 29.2psi vacuum @08:42:00.39 Cartridge 8 sampling 1687/2000ml at 0.21ml/s with 27.2psi vacuum @08:43:00.57 Cartridge 8 sampling 1699/2000ml at 0.22ml/s with 27.9psi vacuum @08:44:00.39 Cartridge 8 sampling 1712/2000ml at 0.20ml/s with 28.2psi vacuum @08:45:00.56 Cartridge 8 sampling 1725/2000ml at 0.20ml/s with 30.0psi vacuum @08:46:00.39 Cartridge 8 sampling 1737/2000ml at 0.20ml/s with 30.4psi vacuum @08:47:00.57 Cartridge 8 sampling 1750/2000ml at 0.20ml/s with 26.9psi vacuum @08:48:00.39 Cartridge 8 sampling 1762/2000ml at 0.20ml/s with 31.6psi vacuum @08:49:00.73 Cartridge 8 sampling 1775/2000ml at 0.21ml/s with 21.0psi vacuum 2.9m deep (19.0psia) @08:50:00.39 Cartridge 8 sampling 1787/2000ml at 0.20ml/s with 31.9psi vacuum @08:51:00.43 Cartridge 8 sampling 1800/2000ml at 0.20ml/s with 25.2psi vacuum @08:52:00.39 Cartridge 8 sampling 1812/2000ml at 0.20ml/s with 32.4psi vacuum @08:53:00.56 Cartridge 8 sampling 1824/2000ml at 0.20ml/s with 30.9psi vacuum @08:54:00.39 Cartridge 8 sampling 1837/2000ml at 0.20ml/s with 33.2psi vacuum @08:55:00.75 Cartridge 8 sampling 1850/2000ml at 0.21ml/s with 24.6psi vacuum 2.2m deep (18.0psia) @08:56:00.63 Cartridge 8 sampling 1862/2000ml at 0.22ml/s with 19.6psi vacuum 3.1m deep (19.3psia) @08:57:00.42 Cartridge 8 sampling 1875/2000ml at 0.25ml/s with 27.1psi vacuum @08:58:00.66 Cartridge 8 sampling 1887/2000ml at 0.20ml/s with 34.3psi vacuum @08:59:00.41 Cartridge 8 sampling 1900/2000ml at 0.24ml/s with 20.8psi vacuum 2.0m deep (17.7psia) @09:00:00.83 Cartridge 8 sampling 1913/2000ml at 0.21ml/s with 27.5psi vacuum 2.0m deep (17.6psia) @09:01:00.39 Cartridge 8 sampling 1925/2000ml at 0.20ml/s with 27.8psi vacuum @09:02:00.55 Cartridge 8 sampling 1938/2000ml at 0.20ml/s with 35.6psi vacuum @09:03:00.39 Cartridge 8 sampling 1950/2000ml at 0.21ml/s with 32.5psi vacuum @09:04:00.60 Cartridge 8 sampling 1963/2000ml at 0.22ml/s with 25.4psi vacuum @09:05:00.96 Cartridge 8 sampling 1975/2000ml at 0.20ml/s with 31.9psi vacuum @09:06:00.56 Cartridge 8 sampling 1987/2000ml at 0.20ml/s with 36.1psi vacuum @09:07:00.40 Cartridge 8 sampling 2000/2000ml at 0.20ml/s with 37.3psi vacuum @09:07:00.79 Pressure.coast @09:07:00.95 Intake.close @09:07:01.01 Exhaust.close @09:07:01.21 Toroid.reconfigure :hold @09:07:01.30 Toroid.seek :bypass @09:07:02.94 Toroid.reconfigure :default @09:07:03.02 Toroid.seek :clear @09:07:04.87 Sampled 2000.0ml @09:07:06.12 SamplePump.setPosition! 0ml @09:07:06.31 Plunge.seek :reagent @09:07:06.59 Exhaust.open @09:07:07.54 Sample loop closed. Ready to Deliver Reagents @09:07:07.60 Archive.seek :sealed @09:07:07.77 Intake.open @09:07:08.70 SamplePump.reconfigure :slow @09:07:08.78 SamplePump.seek -35ml @09:07:11.27 Archive.seek :evac @09:07:25.26 delay 1:00 @09:08:23.90 delay 1 second @09:08:24.92 Bypass.open @09:08:25.35 Archive.seek :clear @09:08:25.64 Exhaust.close @09:08:26.71 SamplePump.seek -33ml @09:08:31.56 delay 1 second @09:08:32.65 SamplePump.seek -41ml @09:08:42.50 RNAlater soak for 10:00 @09:08:50.03 delay 1 second @09:08:51.12 Intake.close @09:08:51.19 Bypass.close @09:18:42.54 Archive.seek :tVent @09:18:44.27 delay 1 second @09:18:45.32 Archive.seek :evac @09:19:01.31 delay 10 seconds @09:19:11.36 Sample preserved and evacuated @09:19:12.25 Carousel.seek 8 @09:19:12.73 Plunge.seek :locked @09:19:13.95 Archive.seek :clear @09:19:14.36 Air.seek :clear @09:19:14.74 Lysis1.seek :clear @09:19:31.42 Archive.coast @09:19:31.58 Air.coast @09:19:31.72 Lysis1.coast @09:19:31.84 Lysis2.coast @09:19:32.00 Plunge.coast @09:19:32.10 SamplePump.coast @09:19:32.34 Toroid.coast @09:19:32.50 Carousel.coast @09:19:32.94 Gate.power :motor,:OFF @09:19:32.97 Gate.power :core,:OFF @09:19:33.05 Gate.power :analytic1,:OFF @09:19:33.12 Gate.power :analytic2,:OFF @09:19:33.18 Gate.power :analytic3,:OFF @09:19:33.25 Gate.power :main,:OFF @09:19:33.28 Safely stopped and ready to power off @09:30:34.25 -> BACsampler.vacuum=10..14 @09:30:38.57 -> slots @09:33:39.49 -> slots 24..19=>[:dry,:archiveNull_bac] @09:43:05.34 -> Cartridge.dry @09:43:47.28 -> 6.times {shoot; delay 120} @09:43:47.67 Gate.power :main,:ON,limit to 5000mA @09:43:47.70 Gate.power :analytic1,:ON,limit to 5000mA @09:43:47.80 Gate.power :analytic2,:ON,limit to 5000mA @09:43:47.83 Gate.power :analytic3,:ON,limit to 5000mA @09:43:48.01 Powering core bus @09:43:50.35 Archive.configure :default @09:43:50.52 Air.configure :default @09:43:50.66 Lysis1.configure :default @09:43:50.80 Lysis2.configure :default @09:43:50.97 Plunge.configure :default @09:43:51.14 SamplePump.configure :default @09:43:51.31 ExtraPump.configure :default @09:43:51.48 Toroid.configure :default @09:43:51.65 Carousel.configure :default @09:43:51.82 Pressure.configure :default @09:43:51.90 All elf controllers running firmware version 4.61 @09:43:51.92 Gate.power :motor,:ON,limit to 5000mA @09:43:52.16 Archive.reconfigure :default @09:43:52.47 Archive.home.jog 0 @09:43:52.72 Plunge.reconfigure :default @09:43:52.77 Lysis2.reconfigure :default @09:43:52.81 Lysis1.reconfigure :default @09:43:52.87 Toroid.reconfigure :default @09:43:52.94 Air.reconfigure :default @09:43:53.18 Plunge.home.jog 0 @09:43:53.23 Lysis2.home.jog 0 @09:43:53.37 Lysis1.home.jog 0 @09:43:53.65 Air.home.jog 0 @09:43:53.73 Toroid.home.jog 0 @09:43:55.20 Archive.reconfigure :homing @09:43:55.58 Archive.jog 159 @09:43:56.15 Plunge.reconfigure :homing @09:43:56.28 Lysis2.reconfigure :homing @09:43:56.33 Air.reconfigure :homing @09:43:56.55 Toroid.reconfigure :homing @09:43:56.64 Lysis1.reconfigure :homing @09:43:57.14 Plunge.jog 858 @09:43:57.38 Lysis2.jog 1362 @09:43:57.42 Air.jog 1368 @09:43:57.50 Toroid.jog 620 @09:43:57.77 Lysis1.jog 1354 @09:43:58.25 Archive.setRawPosition! 50 @09:43:58.35 Archive.seek :home @09:43:58.80 Archive.reconfigure :default @09:43:59.41 Plunge.setRawPosition! 50 @09:43:59.43 Lysis2.setRawPosition! 50 @09:43:59.46 Lysis1.setRawPosition! 50 @09:43:59.50 Air.setRawPosition! 50 @09:43:59.67 Plunge.seek :home @09:43:59.74 Lysis2.seek :home @09:43:59.78 Lysis1.seek :home @09:43:59.93 Air.seek :home @09:44:00.21 Toroid.setRawPosition! 0 @09:44:00.25 Lysis2.reconfigure :default @09:44:00.29 Plunge.reconfigure :default @09:44:00.47 Toroid.seek :home @09:44:00.66 Lysis1.reconfigure :default @09:44:00.69 Air.reconfigure :default @09:44:00.72 Toroid.reconfigure :default @09:44:01.22 Toroid.seek :clear @09:44:02.86 Carousel.reconfigure :default @09:44:02.96 Carousel.home.jog 0 @09:44:03.53 Carousel.seek 8 @09:44:03.81 Selecting Cartridge 24 @09:44:04.46 Carousel.seek 24 @09:44:08.29 Lysis1.seek :locked @09:44:11.54 Plunge.seek :locked @09:44:11.71 Air.seek :locked @09:44:16.07 Mounted archive Cartridge @09:44:16.86 Duration of filtering limited to 4:05:00 @09:44:17.52 Priming sample loop w/160ml, bypass w/10ml @09:44:17.56 SamplePump.setPosition! 0ml @09:44:17.63 Bypass.open @09:44:19.24 Intake.open @09:44:20.16 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @09:44:22.75 Vacuum tare is 17% of full scale @09:44:22.77 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @09:44:23.12 Pressure.reconfigure :default @09:44:23.36 Pressure.hold @09:44:24.02 SamplePump.reconfigure :prime @09:44:24.29 SamplePump.seek 10ml,1:21 @09:44:35.64 Pressure.coast @09:44:35.77 SamplePump.setPosition! 0ml @09:44:35.96 Exhaust.open @09:44:37.50 Bypass.close @09:44:39.21 Pressure.reconfigure :default @09:44:39.29 Pressure.hold @09:44:39.60 SamplePump.reconfigure :prime @09:44:39.69 SamplePump.seek 160ml,13:51 @09:47:31.20 Pressure.coast @09:47:31.40 Bypass.open @09:47:32.23 Intake.close @09:47:33.27 Toroid.reconfigure :hold @09:47:33.36 Toroid.seek :crack @09:47:35.00 Toroid.seek :cartridge @09:47:35.96 Toroid.reconfigure :default @09:47:36.04 Toroid.seek :clear @09:47:38.46 Pressurizing cartridge to 14psi above 15.0psia @09:47:38.48 SamplePump.setPosition! 0ml @09:47:38.69 Intake.open @09:47:39.64 Bypass.close @09:47:40.56 Exhaust.close @09:47:41.63 Pressure.reconfigure :default @09:47:41.71 Pressure.hold @09:47:42.49 SamplePump.reconfigure :charge @09:47:42.58 SamplePump.seek 2.3ml,42 seconds @09:47:49.98 Pressure.coast @09:47:50.16 Intake.close @09:47:51.34 Pressurized cartridge with 0.243ml to 14.1psi @09:47:51.48 Isolated @24.6psia (9.6psi) @09:47:51.52 Checking for cartridge leaks from 24.6psia @09:47:51.75 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @09:47:56.55 Cartridge pressure fell 8% to 22.6psia @09:47:56.87 Sampling 2000.0ml @09:47:56.91 SamplePump.setPosition! 0ml @09:47:57.17 Exhaust.open @09:47:57.99 Intake.open @09:47:59.32 Pressure.reconfigure :default @09:47:59.50 Pressure.hold @09:48:00.28 SamplePump.reconfigure :sample @09:48:00.55 SamplePump.seek 2000ml,2:47:12 @09:48:02.29 Cartridge 24 sampling 1/2000ml at 0.52ml/s with 11.9psi vacuum @09:49:02.74 Cartridge 24 sampling 30/2000ml at 0.49ml/s with 13.1psi vacuum @09:50:02.31 Cartridge 24 sampling 60/2000ml at 0.51ml/s with 13.2psi vacuum @09:51:02.45 Cartridge 24 sampling 90/2000ml at 0.53ml/s with 14.4psi vacuum @09:52:02.30 Cartridge 24 sampling 122/2000ml at 0.53ml/s with 11.0psi vacuum @09:53:02.49 Cartridge 24 sampling 154/2000ml at 0.53ml/s with 13.3psi vacuum @09:54:02.30 Cartridge 24 sampling 185/2000ml at 0.52ml/s with 11.8psi vacuum @09:55:02.54 Cartridge 24 sampling 217/2000ml at 0.52ml/s with 12.7psi vacuum @09:56:02.30 Cartridge 24 sampling 248/2000ml at 0.53ml/s with 13.7psi vacuum @09:57:02.48 Cartridge 24 sampling 280/2000ml at 0.53ml/s with 13.8psi vacuum @09:58:02.30 Cartridge 24 sampling 311/2000ml at 0.52ml/s with 12.3psi vacuum @09:59:02.43 Cartridge 24 sampling 343/2000ml at 0.53ml/s with 10.4psi vacuum @10:00:02.31 Cartridge 24 sampling 375/2000ml at 0.52ml/s with 15.0psi vacuum @10:01:02.46 Cartridge 24 sampling 407/2000ml at 0.53ml/s with 14.3psi vacuum @10:02:02.30 Cartridge 24 sampling 439/2000ml at 0.53ml/s with 11.5psi vacuum @10:03:02.46 Cartridge 24 sampling 470/2000ml at 0.52ml/s with 11.6psi vacuum @10:04:02.31 Cartridge 24 sampling 502/2000ml at 0.53ml/s with 14.0psi vacuum @10:05:02.43 Cartridge 24 sampling 534/2000ml at 0.52ml/s with 14.2psi vacuum @10:06:02.30 Cartridge 24 sampling 565/2000ml at 0.52ml/s with 13.3psi vacuum @10:07:02.64 Cartridge 24 sampling 597/2000ml at 0.53ml/s with 13.3psi vacuum @10:08:02.31 Cartridge 24 sampling 628/2000ml at 0.53ml/s with 10.6psi vacuum @10:09:02.35 Cartridge 24 sampling 660/2000ml at 0.53ml/s with 12.6psi vacuum @10:10:02.59 Cartridge 24 sampling 692/2000ml at 0.53ml/s with 11.3psi vacuum @10:11:02.30 Cartridge 24 sampling 724/2000ml at 0.51ml/s with 15.2psi vacuum @10:12:02.63 Cartridge 24 sampling 756/2000ml at 0.53ml/s with 14.0psi vacuum @10:13:02.31 Cartridge 24 sampling 788/2000ml at 0.53ml/s with 14.4psi vacuum @10:14:02.52 Cartridge 24 sampling 820/2000ml at 0.53ml/s with 11.9psi vacuum @10:15:02.30 Cartridge 24 sampling 852/2000ml at 0.53ml/s with 12.8psi vacuum @10:16:02.42 Cartridge 24 sampling 883/2000ml at 0.53ml/s with 14.0psi vacuum @10:17:02.31 Cartridge 24 sampling 915/2000ml at 0.54ml/s with 12.9psi vacuum @10:18:02.47 Cartridge 24 sampling 947/2000ml at 0.53ml/s with 13.8psi vacuum @10:19:02.31 Cartridge 24 sampling 979/2000ml at 0.52ml/s with 12.8psi vacuum @10:20:02.64 Cartridge 24 sampling 1011/2000ml at 0.53ml/s with 11.5psi vacuum @10:21:02.31 Cartridge 24 sampling 1043/2000ml at 0.53ml/s with 14.7psi vacuum @10:22:02.44 Cartridge 24 sampling 1075/2000ml at 0.54ml/s with 12.7psi vacuum @10:23:02.30 Cartridge 24 sampling 1107/2000ml at 0.54ml/s with 12.0psi vacuum @10:24:02.37 Cartridge 24 sampling 1139/2000ml at 0.53ml/s with 12.5psi vacuum @10:25:02.31 Cartridge 24 sampling 1171/2000ml at 0.53ml/s with 13.4psi vacuum @10:26:02.37 Cartridge 24 sampling 1203/2000ml at 0.54ml/s with 13.2psi vacuum @10:27:02.30 Cartridge 24 sampling 1235/2000ml at 0.53ml/s with 13.5psi vacuum @10:28:02.47 Cartridge 24 sampling 1267/2000ml at 0.53ml/s with 13.4psi vacuum @10:29:02.30 Cartridge 24 sampling 1299/2000ml at 0.54ml/s with 13.6psi vacuum @10:30:02.47 Cartridge 24 sampling 1331/2000ml at 0.52ml/s with 11.6psi vacuum @10:31:02.34 Cartridge 24 sampling 1363/2000ml at 0.54ml/s with 12.5psi vacuum @10:32:02.31 Cartridge 24 sampling 1396/2000ml at 0.54ml/s with 12.5psi vacuum @10:33:02.55 Cartridge 24 sampling 1428/2000ml at 0.54ml/s with 11.5psi vacuum @10:34:02.31 Cartridge 24 sampling 1460/2000ml at 0.52ml/s with 14.9psi vacuum @10:35:02.66 Cartridge 24 sampling 1492/2000ml at 0.53ml/s with 13.7psi vacuum @10:36:02.30 Cartridge 24 sampling 1524/2000ml at 0.53ml/s with 13.9psi vacuum @10:37:02.59 Cartridge 24 sampling 1556/2000ml at 0.53ml/s with 12.4psi vacuum @10:38:02.30 Cartridge 24 sampling 1589/2000ml at 0.54ml/s with 13.4psi vacuum @10:39:02.42 Cartridge 24 sampling 1621/2000ml at 0.54ml/s with 12.8psi vacuum @10:40:02.30 Cartridge 24 sampling 1654/2000ml at 0.53ml/s with 13.5psi vacuum @10:41:02.50 Cartridge 24 sampling 1686/2000ml at 0.54ml/s with 14.9psi vacuum @10:42:02.30 Cartridge 24 sampling 1718/2000ml at 0.54ml/s with 13.0psi vacuum @10:43:02.60 Cartridge 24 sampling 1749/2000ml at 0.52ml/s with 12.0psi vacuum @10:44:02.30 Cartridge 24 sampling 1781/2000ml at 0.53ml/s with 13.5psi vacuum @10:45:02.42 Cartridge 24 sampling 1812/2000ml at 0.52ml/s with 14.2psi vacuum @10:46:02.31 Cartridge 24 sampling 1844/2000ml at 0.50ml/s with 14.1psi vacuum @10:47:02.54 Cartridge 24 sampling 1875/2000ml at 0.53ml/s with 13.5psi vacuum @10:48:02.31 Cartridge 24 sampling 1907/2000ml at 0.52ml/s with 13.7psi vacuum @10:49:02.44 Cartridge 24 sampling 1939/2000ml at 0.53ml/s with 14.3psi vacuum @10:50:02.31 Cartridge 24 sampling 1970/2000ml at 0.54ml/s with 13.8psi vacuum @10:50:57.82 Pressure.coast @10:50:57.97 Intake.close @10:50:58.03 Exhaust.close @10:50:58.28 Toroid.reconfigure :hold @10:50:58.56 Toroid.seek :bypass @10:51:00.35 Toroid.reconfigure :default @10:51:00.59 Toroid.seek :clear @10:51:02.23 Sampled 2000.0ml @10:51:03.60 Plunge.seek :reagent @10:51:04.53 Sample loop closed. Ready to Deliver Reagents @10:51:04.80 Carousel.seek 24 @10:51:05.05 Plunge.seek :locked @10:51:06.12 Air.seek :clear @10:51:06.32 Lysis1.seek :clear @10:51:09.87 Archive.coast @10:51:09.99 Air.coast @10:51:10.20 Lysis1.coast @10:51:10.33 Lysis2.coast @10:51:10.52 Plunge.coast @10:51:10.64 SamplePump.coast @10:51:10.84 Toroid.coast @10:51:10.99 Carousel.coast @10:51:11.56 Gate.power :motor,:OFF @10:51:11.61 Gate.power :core,:OFF @10:51:11.64 Gate.power :analytic1,:OFF @10:51:11.70 Gate.power :analytic2,:OFF @10:51:11.78 Gate.power :analytic3,:OFF @10:51:11.85 Gate.power :main,:OFF @10:51:11.94 Safely stopped and ready to power off @10:51:12.20 delay 2:00 @10:53:12.48 Gate.power :main,:ON,limit to 5000mA @10:53:12.52 Gate.power :analytic1,:ON,limit to 5000mA @10:53:12.66 Gate.power :analytic2,:ON,limit to 5000mA @10:53:12.75 Gate.power :analytic3,:ON,limit to 5000mA @10:53:12.99 Powering core bus @10:53:15.30 Archive.configure :default @10:53:15.46 Air.configure :default @10:53:15.61 Lysis1.configure :default @10:53:15.75 Lysis2.configure :default @10:53:15.92 Plunge.configure :default @10:53:16.09 SamplePump.configure :default @10:53:16.26 ExtraPump.configure :default @10:53:16.43 Toroid.configure :default @10:53:16.60 Carousel.configure :default @10:53:16.76 Pressure.configure :default @10:53:16.85 All elf controllers running firmware version 4.61 @10:53:16.87 Gate.power :motor,:ON,limit to 5000mA @10:53:17.08 Archive.reconfigure :default @10:53:17.80 Archive.home.jog 0 @10:53:18.31 Plunge.reconfigure :default @10:53:18.84 Lysis1.reconfigure :default @10:53:19.01 Toroid.reconfigure :default @10:53:19.15 Lysis2.reconfigure :default @10:53:19.32 Air.reconfigure :default @10:53:19.94 Lysis1.home.jog 0 @10:53:20.02 Plunge.home.jog 0 @10:53:20.10 Lysis2.home.jog 0 @10:53:20.18 Toroid.home.jog 0 @10:53:20.26 Air.home.jog 0 @10:53:21.71 Archive.reconfigure :homing @10:53:22.21 Plunge.reconfigure :homing @10:53:22.33 Archive.jog 1369 @10:53:22.49 Toroid.reconfigure :homing @10:53:22.55 Lysis2.reconfigure :homing @10:53:22.70 Air.reconfigure :homing @10:53:22.79 Lysis1.reconfigure :homing @10:53:23.27 Plunge.jog 852 @10:53:23.54 Lysis2.jog 1364 @10:53:23.58 Air.jog 1367 @10:53:23.63 Toroid.jog 627 @10:53:23.73 Lysis1.jog 1349 @10:53:24.19 Archive.setRawPosition! 50 @10:53:24.25 Archive.seek :home @10:53:24.55 Archive.reconfigure :default @10:53:24.66 Plunge.setRawPosition! 50 @10:53:24.73 Plunge.seek :home @10:53:24.91 Lysis2.setRawPosition! 50 @10:53:25.00 Lysis2.seek :home @10:53:25.13 Lysis1.setRawPosition! 50 @10:53:25.16 Air.setRawPosition! 50 @10:53:25.19 Plunge.reconfigure :default @10:53:25.34 Lysis1.seek :home @10:53:25.39 Air.seek :home @10:53:25.65 Lysis2.reconfigure :default @10:53:25.77 Lysis1.reconfigure :default @10:53:25.87 Air.reconfigure :default @10:53:26.00 Toroid.setRawPosition! 0 @10:53:26.07 Toroid.seek :home @10:53:26.19 Toroid.reconfigure :default @10:53:26.94 Toroid.seek :clear @10:53:28.65 Carousel.reconfigure :default @10:53:28.83 Carousel.home.jog 0 @10:53:29.59 Carousel.seek 24 @10:53:30.10 Selecting Cartridge 23 @10:53:30.68 Carousel.seek 23 @10:53:31.67 Lysis1.seek :locked @10:53:34.93 Plunge.seek :locked @10:53:35.12 Air.seek :locked @10:53:39.57 Mounted archive Cartridge @10:53:40.40 Duration of filtering limited to 4:05:00 @10:53:41.14 Priming sample loop w/160ml, bypass w/10ml @10:53:41.24 SamplePump.setPosition! 0ml @10:53:41.31 Bypass.open @10:53:42.94 Intake.open @10:53:43.86 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @10:53:46.44 Vacuum tare is 18% of full scale @10:53:46.47 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @10:53:46.72 Pressure.reconfigure :default @10:53:46.80 Pressure.hold @10:53:47.44 SamplePump.reconfigure :prime @10:53:47.66 SamplePump.seek 10ml,1:21 @10:53:58.96 Pressure.coast @10:53:59.14 SamplePump.setPosition! 0ml @10:53:59.28 Exhaust.open @10:54:00.89 Bypass.close @10:54:02.58 Pressure.reconfigure :default @10:54:02.67 Pressure.hold @10:54:02.97 SamplePump.reconfigure :prime @10:54:03.06 SamplePump.seek 160ml,13:51 @10:56:54.81 Pressure.coast @10:56:55.01 Bypass.open @10:56:55.83 Intake.close @10:56:56.80 Toroid.reconfigure :hold @10:56:56.88 Toroid.seek :crack @10:56:58.50 Toroid.seek :cartridge @10:56:59.46 Toroid.reconfigure :default @10:56:59.54 Toroid.seek :clear @10:57:02.03 Pressurizing cartridge to 14psi above 14.6psia @10:57:02.06 SamplePump.setPosition! 0ml @10:57:02.19 Intake.open @10:57:03.11 Bypass.close @10:57:03.97 Exhaust.close @10:57:05.04 Pressure.reconfigure :default @10:57:05.12 Pressure.hold @10:57:05.59 SamplePump.reconfigure :charge @10:57:05.78 SamplePump.seek 2.3ml,42 seconds @10:57:13.94 Pressure.coast @10:57:14.16 Intake.close @10:57:15.30 Pressurized cartridge with 0.289ml to 14.1psi @10:57:15.32 Isolated @26.5psia (12.0psi) @10:57:15.36 Checking for cartridge leaks from 26.5psia @10:57:15.57 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @10:57:20.54 Sampling 2000.0ml @10:57:20.57 SamplePump.setPosition! 0ml @10:57:20.83 Exhaust.open @10:57:21.70 Intake.open @10:57:23.03 Pressure.reconfigure :default @10:57:23.15 Pressure.hold @10:57:23.91 SamplePump.reconfigure :sample @10:57:24.16 SamplePump.seek 2000ml,2:47:12 @10:57:25.93 Cartridge 23 sampling 1/2000ml at 0.59ml/s with 13.5psi vacuum @10:58:26.12 Cartridge 23 sampling 31/2000ml at 0.52ml/s with 14.4psi vacuum @10:59:25.93 Cartridge 23 sampling 62/2000ml at 0.54ml/s with 11.8psi vacuum @11:00:26.10 Cartridge 23 sampling 94/2000ml at 0.54ml/s with 12.1psi vacuum @11:01:25.93 Cartridge 23 sampling 127/2000ml at 0.54ml/s with 14.3psi vacuum @11:02:26.12 Cartridge 23 sampling 160/2000ml at 0.55ml/s with 10.9psi vacuum @11:03:25.94 Cartridge 23 sampling 192/2000ml at 0.54ml/s with 12.6psi vacuum @11:04:26.08 Cartridge 23 sampling 225/2000ml at 0.54ml/s with 13.2psi vacuum @11:05:25.93 Cartridge 23 sampling 257/2000ml at 0.55ml/s with 11.0psi vacuum @11:06:26.11 Cartridge 23 sampling 289/2000ml at 0.52ml/s with 13.5psi vacuum @11:07:25.93 Cartridge 23 sampling 321/2000ml at 0.54ml/s with 14.3psi vacuum @11:08:26.18 Cartridge 23 sampling 354/2000ml at 0.57ml/s with 10.8psi vacuum 1.6m deep (17.1psia) @11:09:25.93 Cartridge 23 sampling 387/2000ml at 0.54ml/s with 11.6psi vacuum @11:10:26.11 Cartridge 23 sampling 420/2000ml at 0.57ml/s with 11.7psi vacuum @11:11:25.93 Cartridge 23 sampling 454/2000ml at 0.56ml/s with 12.1psi vacuum @11:12:27.17 Cartridge 23 sampling 488/2000ml at 0.55ml/s with 14.8psi vacuum @11:13:25.93 Cartridge 23 sampling 521/2000ml at 0.55ml/s with 11.7psi vacuum @11:14:26.32 Cartridge 23 sampling 555/2000ml at 0.56ml/s with 13.8psi vacuum @11:15:26.22 Cartridge 23 sampling 589/2000ml at 0.56ml/s with 11.2psi vacuum @11:16:26.29 Cartridge 23 sampling 622/2000ml at 0.55ml/s with 11.3psi vacuum @11:17:26.16 Cartridge 23 sampling 656/2000ml at 0.57ml/s with 11.5psi vacuum @11:18:25.95 Cartridge 23 sampling 690/2000ml at 0.56ml/s with 10.8psi vacuum @11:19:26.08 Cartridge 23 sampling 723/2000ml at 0.55ml/s with 13.8psi vacuum @11:20:25.94 Cartridge 23 sampling 757/2000ml at 0.57ml/s with 11.3psi vacuum @11:21:26.19 Cartridge 23 sampling 791/2000ml at 0.57ml/s with 13.5psi vacuum @11:22:25.95 Cartridge 23 sampling 824/2000ml at 0.56ml/s with 13.3psi vacuum @11:23:26.10 Cartridge 23 sampling 858/2000ml at 0.56ml/s with 13.5psi vacuum @11:24:25.95 Cartridge 23 sampling 892/2000ml at 0.56ml/s with 13.2psi vacuum @11:25:26.13 Cartridge 23 sampling 926/2000ml at 0.56ml/s with 13.8psi vacuum @11:26:25.94 Cartridge 23 sampling 959/2000ml at 0.56ml/s with 12.0psi vacuum @11:27:26.05 Cartridge 23 sampling 993/2000ml at 0.55ml/s with 11.5psi vacuum @11:28:25.94 Cartridge 23 sampling 1026/2000ml at 0.57ml/s with 13.0psi vacuum @11:29:26.10 Cartridge 23 sampling 1060/2000ml at 0.56ml/s with 13.3psi vacuum @11:30:25.94 Cartridge 23 sampling 1094/2000ml at 0.57ml/s with 13.8psi vacuum @11:31:26.11 Cartridge 23 sampling 1127/2000ml at 0.57ml/s with 13.0psi vacuum @11:32:25.95 Cartridge 23 sampling 1161/2000ml at 0.56ml/s with 13.7psi vacuum @11:33:26.07 Cartridge 23 sampling 1195/2000ml at 0.56ml/s with 13.1psi vacuum @11:34:25.94 Cartridge 23 sampling 1229/2000ml at 0.56ml/s with 11.5psi vacuum @11:35:26.22 Cartridge 23 sampling 1263/2000ml at 0.57ml/s with 12.6psi vacuum @11:36:25.94 Cartridge 23 sampling 1297/2000ml at 0.56ml/s with 13.9psi vacuum @11:37:26.11 Cartridge 23 sampling 1332/2000ml at 0.56ml/s with 13.9psi vacuum @11:38:25.94 Cartridge 23 sampling 1366/2000ml at 0.57ml/s with 13.1psi vacuum @11:39:26.40 Cartridge 23 sampling 1400/2000ml at 0.57ml/s with 13.9psi vacuum @11:40:26.14 Cartridge 23 sampling 1434/2000ml at 0.56ml/s with 12.4psi vacuum @11:41:25.94 Cartridge 23 sampling 1468/2000ml at 0.57ml/s with 12.7psi vacuum @11:42:26.35 Cartridge 23 sampling 1502/2000ml at 0.58ml/s with 13.8psi vacuum @11:43:25.94 Cartridge 23 sampling 1537/2000ml at 0.58ml/s with 14.8psi vacuum @11:44:26.22 Cartridge 23 sampling 1571/2000ml at 0.57ml/s with 14.4psi vacuum @11:45:25.95 Cartridge 23 sampling 1605/2000ml at 0.57ml/s with 14.1psi vacuum @11:46:26.10 Cartridge 23 sampling 1639/2000ml at 0.57ml/s with 14.6psi vacuum @11:47:25.94 Cartridge 23 sampling 1674/2000ml at 0.58ml/s with 12.7psi vacuum @11:48:26.16 Cartridge 23 sampling 1709/2000ml at 0.57ml/s with 14.0psi vacuum @11:49:25.95 Cartridge 23 sampling 1743/2000ml at 0.57ml/s with 13.8psi vacuum @11:50:26.23 Cartridge 23 sampling 1777/2000ml at 0.57ml/s with 13.1psi vacuum @11:51:25.95 Cartridge 23 sampling 1812/2000ml at 0.58ml/s with 12.4psi vacuum @11:52:26.08 Cartridge 23 sampling 1846/2000ml at 0.58ml/s with 13.9psi vacuum @11:53:25.95 Cartridge 23 sampling 1881/2000ml at 0.57ml/s with 13.8psi vacuum @11:54:26.44 Cartridge 23 sampling 1916/2000ml at 0.56ml/s with 14.6psi vacuum @11:55:25.95 Cartridge 23 sampling 1950/2000ml at 0.58ml/s with 12.6psi vacuum @11:56:26.11 Cartridge 23 sampling 1984/2000ml at 0.56ml/s with 14.4psi vacuum @11:56:54.47 Pressure.coast @11:56:54.80 Intake.close @11:56:54.87 Exhaust.close @11:56:55.61 Toroid.reconfigure :hold @11:56:55.70 Toroid.seek :bypass @11:56:57.34 Toroid.reconfigure :default @11:56:57.42 Toroid.seek :clear @11:56:59.05 Sampled 2000.0ml @11:56:59.56 Plunge.seek :reagent @11:57:00.50 Sample loop closed. Ready to Deliver Reagents @11:57:01.23 Carousel.seek 23 @11:57:01.80 Plunge.seek :locked @11:57:03.07 Air.seek :clear @11:57:03.44 Lysis1.seek :clear @11:57:06.95 Archive.coast @11:57:07.02 Air.coast @11:57:07.09 Lysis1.coast @11:57:07.17 Lysis2.coast @11:57:07.23 Plunge.coast @11:57:07.30 SamplePump.coast @11:57:07.38 Toroid.coast @11:57:07.45 Carousel.coast @11:57:07.67 Gate.power :motor,:OFF @11:57:07.69 Gate.power :core,:OFF @11:57:07.73 Gate.power :analytic1,:OFF @11:57:07.76 Gate.power :analytic2,:OFF @11:57:07.79 Gate.power :analytic3,:OFF @11:57:07.82 Gate.power :main,:OFF @11:57:07.85 Safely stopped and ready to power off @11:57:08.18 delay 2:00 @11:59:08.31 Gate.power :main,:ON,limit to 5000mA @11:59:08.33 Gate.power :analytic1,:ON,limit to 5000mA @11:59:08.37 Gate.power :analytic2,:ON,limit to 5000mA @11:59:08.40 Gate.power :analytic3,:ON,limit to 5000mA @11:59:08.47 Powering core bus @11:59:11.22 Archive.configure :default @11:59:11.59 Air.configure :default @11:59:11.94 Lysis1.configure :default @11:59:12.31 Lysis2.configure :default @11:59:12.71 Plunge.configure :default @11:59:13.02 SamplePump.configure :default @11:59:13.30 ExtraPump.configure :default @11:59:13.62 Toroid.configure :default @11:59:13.96 Carousel.configure :default @11:59:14.23 Pressure.configure :default @11:59:14.75 All elf controllers running firmware version 4.61 @11:59:15.38 Gate.power :motor,:ON,limit to 5000mA @11:59:16.00 Toroid.reconfigure :default @11:59:16.04 Plunge.reconfigure :default @11:59:16.09 Lysis2.reconfigure :default @11:59:16.14 Lysis1.reconfigure :default @11:59:16.19 Archive.reconfigure :default @11:59:16.25 Air.reconfigure :default @11:59:16.54 Plunge.home.jog 0 @11:59:16.57 Toroid.home.jog 0 @11:59:16.61 Lysis2.home.jog 0 @11:59:16.65 Lysis1.home.jog 0 @11:59:16.68 Archive.home.jog 0 @11:59:16.71 Air.home.jog 0 @11:59:17.52 Plunge.reconfigure :homing @11:59:17.57 Lysis2.reconfigure :homing @11:59:17.61 Lysis1.reconfigure :homing @11:59:17.66 Archive.reconfigure :homing @11:59:17.71 Toroid.reconfigure :homing @11:59:17.92 Plunge.jog 851 @11:59:18.00 Lysis1.jog 1349 @11:59:18.05 Toroid.jog 619 @11:59:18.10 Lysis2.jog 1361 @11:59:18.15 Archive.jog 1358 @11:59:18.45 Air.reconfigure :homing @11:59:18.53 Air.jog 1367 @11:59:19.16 Plunge.setRawPosition! 50 @11:59:19.32 Plunge.seek :home @11:59:19.77 Lysis2.setRawPosition! 50 @11:59:19.89 Lysis1.setRawPosition! 50 @11:59:19.99 Air.setRawPosition! 50 @11:59:20.13 Archive.setRawPosition! 50 @11:59:20.92 Lysis2.seek :home @11:59:21.03 Lysis1.seek :home @11:59:21.11 Air.seek :home @11:59:21.27 Archive.seek :home @11:59:22.07 Toroid.setRawPosition! 0 @11:59:22.19 Plunge.reconfigure :default @11:59:22.23 Lysis1.reconfigure :default @11:59:22.41 Lysis2.reconfigure :default @11:59:22.80 Toroid.seek :home @11:59:23.33 Air.reconfigure :default @11:59:23.42 Archive.reconfigure :default @11:59:23.58 Toroid.reconfigure :default @11:59:24.46 Toroid.seek :clear @11:59:26.10 Carousel.reconfigure :default @11:59:26.18 Carousel.home.jog 0 @11:59:26.71 Carousel.seek 23 @11:59:26.91 Selecting Cartridge 22 @11:59:27.16 Carousel.seek 22 @11:59:27.70 Lysis1.seek :locked @11:59:31.08 Plunge.seek :locked @11:59:31.54 Air.seek :locked @11:59:35.91 Mounted archive Cartridge @11:59:36.24 Duration of filtering limited to 4:05:00 @11:59:36.57 Priming sample loop w/160ml, bypass w/10ml @11:59:36.59 SamplePump.setPosition! 0ml @11:59:36.65 Bypass.open @11:59:38.16 Intake.open @11:59:39.04 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @11:59:41.75 Vacuum tare is 18% of full scale @11:59:41.81 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @11:59:42.53 Pressure.reconfigure :default @11:59:42.65 Pressure.hold @11:59:43.33 SamplePump.reconfigure :prime @11:59:43.42 SamplePump.seek 10ml,1:21 @11:59:54.60 Pressure.coast @11:59:54.67 SamplePump.setPosition! 0ml @11:59:54.74 Exhaust.open @11:59:56.14 Bypass.close @11:59:57.70 Pressure.reconfigure :default @11:59:57.78 Pressure.hold @11:59:58.08 SamplePump.reconfigure :prime @11:59:58.17 SamplePump.seek 160ml,13:51 @12:02:49.81 Pressure.coast @12:02:49.88 Bypass.open @12:02:50.62 Intake.close @12:02:51.52 Toroid.reconfigure :hold @12:02:51.60 Toroid.seek :crack @12:02:53.45 Toroid.seek :cartridge @12:02:54.58 Toroid.reconfigure :default @12:02:54.72 Toroid.seek :clear @12:02:57.01 Pressurizing cartridge to 14psi above 14.5psia @12:02:57.02 SamplePump.setPosition! 0ml @12:02:57.09 Intake.open @12:02:58.11 Bypass.close @12:02:58.88 Exhaust.close @12:03:00.23 Pressure.reconfigure :default @12:03:00.32 Pressure.hold @12:03:00.56 SamplePump.reconfigure :charge @12:03:00.65 SamplePump.seek 2.3ml,42 seconds @12:03:08.84 Pressure.coast @12:03:08.91 Intake.close @12:03:09.82 Pressurized cartridge with 0.301ml to 14.1psi @12:03:09.83 Isolated @25.0psia (10.5psi) @12:03:09.85 Checking for cartridge leaks from 25.0psia @12:03:09.91 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:03:14.61 Cartridge pressure fell 6% to 23.5psia @12:03:14.85 Sampling 2000.0ml @12:03:14.88 SamplePump.setPosition! 0ml @12:03:15.03 Exhaust.open @12:03:15.79 Intake.open @12:03:17.41 Pressure.reconfigure :default @12:03:17.49 Pressure.hold @12:03:17.80 SamplePump.reconfigure :sample @12:03:17.89 SamplePump.seek 2000ml,2:47:12 @12:03:20.33 Cartridge 22 sampling 1/2000ml at 0.54ml/s with 13.8psi vacuum @12:04:20.39 Cartridge 22 sampling 32/2000ml at 0.51ml/s with 9.9psi vacuum @12:05:20.33 Cartridge 22 sampling 64/2000ml at 0.54ml/s with 11.7psi vacuum @12:06:20.48 Cartridge 22 sampling 97/2000ml at 0.55ml/s with 13.0psi vacuum @12:07:20.34 Cartridge 22 sampling 131/2000ml at 0.57ml/s with 14.3psi vacuum @12:08:20.46 Cartridge 22 sampling 165/2000ml at 0.57ml/s with 10.6psi vacuum @12:09:20.33 Cartridge 22 sampling 199/2000ml at 0.55ml/s with 14.6psi vacuum @12:10:20.36 Cartridge 22 sampling 233/2000ml at 0.57ml/s with 12.0psi vacuum @12:11:20.68 Cartridge 22 sampling 267/2000ml at 0.57ml/s with 11.0psi vacuum @12:12:20.34 Cartridge 22 sampling 301/2000ml at 0.56ml/s with 11.1psi vacuum @12:13:20.82 Cartridge 22 sampling 334/2000ml at 0.56ml/s with 11.1psi vacuum @12:14:20.34 Cartridge 22 sampling 368/2000ml at 0.56ml/s with 12.5psi vacuum @12:15:20.47 Cartridge 22 sampling 402/2000ml at 0.57ml/s with 14.3psi vacuum @12:16:20.34 Cartridge 22 sampling 436/2000ml at 0.58ml/s with 13.9psi vacuum @12:17:20.50 Cartridge 22 sampling 470/2000ml at 0.57ml/s with 13.0psi vacuum @12:18:20.34 Cartridge 22 sampling 504/2000ml at 0.56ml/s with 12.0psi vacuum @12:19:20.54 Cartridge 22 sampling 539/2000ml at 0.57ml/s with 11.3psi vacuum @12:20:20.34 Cartridge 22 sampling 572/2000ml at 0.58ml/s with 12.9psi vacuum @12:21:20.45 Cartridge 22 sampling 606/2000ml at 0.56ml/s with 12.0psi vacuum @12:22:20.34 Cartridge 22 sampling 641/2000ml at 0.55ml/s with 11.5psi vacuum @12:23:20.62 Cartridge 22 sampling 674/2000ml at 0.55ml/s with 11.1psi vacuum @12:24:20.34 Cartridge 22 sampling 708/2000ml at 0.56ml/s with 13.9psi vacuum @12:25:20.50 Cartridge 22 sampling 742/2000ml at 0.57ml/s with 13.0psi vacuum @12:26:20.34 Cartridge 22 sampling 776/2000ml at 0.56ml/s with 12.9psi vacuum @12:27:20.57 Cartridge 22 sampling 810/2000ml at 0.56ml/s with 14.3psi vacuum @12:28:20.34 Cartridge 22 sampling 844/2000ml at 0.56ml/s with 11.1psi vacuum @12:29:20.55 Cartridge 22 sampling 878/2000ml at 0.56ml/s with 12.0psi vacuum @12:30:20.35 Cartridge 22 sampling 912/2000ml at 0.56ml/s with 11.7psi vacuum @12:31:20.46 Cartridge 22 sampling 946/2000ml at 0.55ml/s with 13.5psi vacuum @12:32:20.34 Cartridge 22 sampling 980/2000ml at 0.57ml/s with 13.7psi vacuum @12:33:20.45 Cartridge 22 sampling 1014/2000ml at 0.56ml/s with 11.3psi vacuum @12:34:20.35 Cartridge 22 sampling 1048/2000ml at 0.56ml/s with 11.8psi vacuum @12:35:20.81 Cartridge 22 sampling 1082/2000ml at 0.56ml/s with 13.8psi vacuum @12:36:20.35 Cartridge 22 sampling 1116/2000ml at 0.56ml/s with 11.4psi vacuum @12:37:20.51 Cartridge 22 sampling 1150/2000ml at 0.56ml/s with 12.8psi vacuum @12:38:20.52 Cartridge 22 sampling 1184/2000ml at 0.57ml/s with 12.4psi vacuum @12:39:20.34 Cartridge 22 sampling 1219/2000ml at 0.57ml/s with 13.4psi vacuum @12:40:20.71 Cartridge 22 sampling 1253/2000ml at 0.57ml/s with 14.8psi vacuum @12:41:20.34 Cartridge 22 sampling 1287/2000ml at 0.56ml/s with 12.1psi vacuum @12:42:20.46 Cartridge 22 sampling 1322/2000ml at 0.58ml/s with 13.2psi vacuum @12:43:20.34 Cartridge 22 sampling 1356/2000ml at 0.58ml/s with 13.5psi vacuum @12:44:20.53 Cartridge 22 sampling 1390/2000ml at 0.56ml/s with 13.8psi vacuum @12:45:20.34 Cartridge 22 sampling 1425/2000ml at 0.58ml/s with 14.7psi vacuum @12:46:20.53 Cartridge 22 sampling 1459/2000ml at 0.57ml/s with 13.3psi vacuum @12:47:20.35 Cartridge 22 sampling 1494/2000ml at 0.57ml/s with 13.0psi vacuum @12:48:20.62 Cartridge 22 sampling 1529/2000ml at 0.57ml/s with 13.7psi vacuum @12:49:20.34 Cartridge 22 sampling 1564/2000ml at 0.57ml/s with 14.2psi vacuum @12:50:20.60 Cartridge 22 sampling 1598/2000ml at 0.57ml/s with 14.8psi vacuum @12:51:20.34 Cartridge 22 sampling 1633/2000ml at 0.58ml/s with 12.9psi vacuum @12:52:20.49 Cartridge 22 sampling 1668/2000ml at 0.59ml/s with 13.5psi vacuum @12:53:20.35 Cartridge 22 sampling 1703/2000ml at 0.57ml/s with 12.7psi vacuum @12:54:20.47 Cartridge 22 sampling 1737/2000ml at 0.56ml/s with 12.4psi vacuum @12:55:20.34 Cartridge 22 sampling 1771/2000ml at 0.57ml/s with 11.4psi vacuum @12:56:20.41 Cartridge 22 sampling 1805/2000ml at 0.56ml/s with 14.7psi vacuum @12:57:20.34 Cartridge 22 sampling 1839/2000ml at 0.56ml/s with 15.0psi vacuum @12:58:20.62 Cartridge 22 sampling 1873/2000ml at 0.57ml/s with 13.1psi vacuum @12:59:20.62 Cartridge 22 sampling 1907/2000ml at 0.56ml/s with 14.4psi vacuum @13:00:20.34 Cartridge 22 sampling 1941/2000ml at 0.56ml/s with 12.4psi vacuum @13:01:20.57 Cartridge 22 sampling 1975/2000ml at 0.55ml/s with 12.5psi vacuum @13:02:05.00 Pressure.coast @13:02:05.17 Intake.close @13:02:05.22 Exhaust.close @13:02:05.45 Toroid.reconfigure :hold @13:02:05.53 Toroid.seek :bypass @13:02:07.20 Toroid.reconfigure :default @13:02:07.37 Toroid.seek :clear @13:02:10.26 Sampled 2000.0ml @13:02:11.49 Plunge.seek :reagent @13:02:12.42 Sample loop closed. Ready to Deliver Reagents @13:02:12.68 Carousel.seek 22 @13:02:12.93 Plunge.seek :locked @13:02:14.03 Air.seek :clear @13:02:14.21 Lysis1.seek :clear @13:02:17.79 Archive.coast @13:02:17.93 Air.coast @13:02:18.10 Lysis1.coast @13:02:18.32 Lysis2.coast @13:02:18.47 Plunge.coast @13:02:18.64 SamplePump.coast @13:02:18.76 Toroid.coast @13:02:18.95 Carousel.coast @13:02:19.36 Gate.power :motor,:OFF @13:02:19.48 Gate.power :core,:OFF @13:02:19.51 Gate.power :analytic1,:OFF @13:02:19.58 Gate.power :analytic2,:OFF @13:02:19.74 Gate.power :analytic3,:OFF @13:02:19.77 Gate.power :main,:OFF @13:02:19.80 Safely stopped and ready to power off @13:02:20.15 delay 2:00 @13:04:20.50 Gate.power :main,:ON,limit to 5000mA @13:04:20.56 Gate.power :analytic1,:ON,limit to 5000mA @13:04:20.64 Gate.power :analytic2,:ON,limit to 5000mA @13:04:20.74 Gate.power :analytic3,:ON,limit to 5000mA @13:04:20.89 Powering core bus @13:04:23.18 Archive.configure :default @13:04:23.35 Air.configure :default @13:04:23.50 Lysis1.configure :default @13:04:23.65 Lysis2.configure :default @13:04:23.82 Plunge.configure :default @13:04:23.98 SamplePump.configure :default @13:04:24.15 ExtraPump.configure :default @13:04:24.31 Toroid.configure :default @13:04:24.48 Carousel.configure :default @13:04:24.65 Pressure.configure :default @13:04:24.73 All elf controllers running firmware version 4.61 @13:04:24.75 Gate.power :motor,:ON,limit to 5000mA @13:04:25.39 Archive.reconfigure :default @13:04:25.64 Air.reconfigure :default @13:04:26.07 Lysis1.reconfigure :default @13:04:26.66 Archive.home.jog 0 @13:04:26.82 Lysis1.home.jog 0 @13:04:26.91 Air.home.jog 0 @13:04:27.89 Plunge.reconfigure :default @13:04:27.92 Lysis2.reconfigure :default @13:04:28.10 Toroid.reconfigure :default @13:04:28.61 Plunge.home.jog 0 @13:04:28.80 Lysis2.home.jog 0 @13:04:28.89 Toroid.home.jog 0 @13:04:29.91 Air.reconfigure :homing @13:04:30.15 Lysis1.reconfigure :homing @13:04:30.36 Archive.reconfigure :homing @13:04:30.54 Air.jog 1368 @13:04:30.89 Lysis1.jog 1351 @13:04:31.08 Archive.jog 1359 @13:04:31.46 Toroid.reconfigure :homing @13:04:31.49 Plunge.reconfigure :homing @13:04:31.53 Lysis2.reconfigure :homing @13:04:31.72 Toroid.jog 619 @13:04:31.75 Lysis2.jog 1357 @13:04:31.79 Plunge.jog 853 @13:04:32.07 Air.setRawPosition! 50 @13:04:32.14 Air.seek :home @13:04:32.46 Lysis1.setRawPosition! 50 @13:04:32.52 Lysis1.seek :home @13:04:32.65 Archive.setRawPosition! 50 @13:04:32.68 Air.reconfigure :default @13:04:32.82 Archive.seek :home @13:04:32.96 Plunge.setRawPosition! 50 @13:04:32.98 Lysis1.reconfigure :default @13:04:33.10 Plunge.seek :home @13:04:33.27 Lysis2.setRawPosition! 50 @13:04:33.29 Archive.reconfigure :default @13:04:33.41 Lysis2.seek :home @13:04:33.56 Plunge.reconfigure :default @13:04:33.75 Lysis2.reconfigure :default @13:04:33.94 Toroid.setRawPosition! 0 @13:04:34.01 Toroid.seek :home @13:04:34.12 Toroid.reconfigure :default @13:04:34.29 Toroid.seek :clear @13:04:35.96 Carousel.reconfigure :default @13:04:36.14 Carousel.home.jog 0 @13:04:36.94 Carousel.seek 22 @13:04:37.38 Selecting Cartridge 21 @13:04:37.91 Carousel.seek 21 @13:04:38.67 Lysis1.seek :locked @13:04:42.02 Plunge.seek :locked @13:04:42.19 Air.seek :locked @13:04:46.60 Mounted archive Cartridge @13:04:47.43 Duration of filtering limited to 4:05:00 @13:04:49.48 Priming sample loop w/160ml, bypass w/10ml @13:04:49.53 SamplePump.setPosition! 0ml @13:04:49.73 Bypass.open @13:04:51.34 Intake.open @13:04:52.24 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @13:04:54.83 Vacuum tare is 18% of full scale @13:04:54.86 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @13:04:55.13 Pressure.reconfigure :default @13:04:55.21 Pressure.hold @13:04:55.44 SamplePump.reconfigure :prime @13:04:55.55 SamplePump.seek 10ml,1:21 @13:05:06.82 Pressure.coast @13:05:06.91 SamplePump.setPosition! 0ml @13:05:07.07 Exhaust.open @13:05:08.57 Bypass.close @13:05:10.25 Pressure.reconfigure :default @13:05:10.33 Pressure.hold @13:05:10.63 SamplePump.reconfigure :prime @13:05:10.72 SamplePump.seek 160ml,13:51 @13:08:02.39 Pressure.coast @13:08:02.60 Bypass.open @13:08:03.46 Intake.close @13:08:04.73 Toroid.reconfigure :hold @13:08:04.93 Toroid.seek :crack @13:08:06.54 Toroid.seek :cartridge @13:08:07.51 Toroid.reconfigure :default @13:08:07.59 Toroid.seek :clear @13:08:10.05 Pressurizing cartridge to 14psi above 14.5psia @13:08:10.11 SamplePump.setPosition! 0ml @13:08:10.27 Intake.open @13:08:11.23 Bypass.close @13:08:11.97 Exhaust.close @13:08:13.28 Pressure.reconfigure :default @13:08:13.59 Pressure.hold @13:08:14.16 SamplePump.reconfigure :charge @13:08:14.24 SamplePump.seek 2.3ml,42 seconds @13:08:22.84 Pressure.coast @13:08:22.94 Intake.close @13:08:24.14 Pressurized cartridge with 0.321ml to 14.1psi @13:08:24.15 Isolated @25.2psia (10.7psi) @13:08:24.17 Checking for cartridge leaks from 25.2psia @13:08:24.24 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @13:08:28.90 Cartridge pressure fell 2% to 24.8psia @13:08:29.33 Sampling 2000.0ml @13:08:29.51 SamplePump.setPosition! 0ml @13:08:29.69 Exhaust.open @13:08:30.45 Intake.open @13:08:31.86 Pressure.reconfigure :default @13:08:32.02 Pressure.hold @13:08:32.74 SamplePump.reconfigure :sample @13:08:32.90 SamplePump.seek 2000ml,2:47:12 @13:08:34.77 Cartridge 21 sampling 1/2000ml at 0.59ml/s with 13.6psi vacuum @13:09:34.94 Cartridge 21 sampling 31/2000ml at 0.52ml/s with 14.4psi vacuum @13:10:34.79 Cartridge 21 sampling 63/2000ml at 0.54ml/s with 14.8psi vacuum @13:11:34.94 Cartridge 21 sampling 96/2000ml at 0.54ml/s with 12.9psi vacuum @13:12:34.78 Cartridge 21 sampling 129/2000ml at 0.56ml/s with 10.2psi vacuum @13:13:34.94 Cartridge 21 sampling 164/2000ml at 0.57ml/s with 12.8psi vacuum @13:14:34.78 Cartridge 21 sampling 198/2000ml at 0.57ml/s with 10.9psi vacuum @13:15:34.93 Cartridge 21 sampling 232/2000ml at 0.60ml/s with 11.9psi vacuum @13:16:34.78 Cartridge 21 sampling 268/2000ml at 0.58ml/s with 11.9psi vacuum @13:17:34.98 Cartridge 21 sampling 302/2000ml at 0.56ml/s with 11.2psi vacuum @13:18:34.78 Cartridge 21 sampling 336/2000ml at 0.57ml/s with 12.0psi vacuum @13:19:35.04 Cartridge 21 sampling 370/2000ml at 0.57ml/s with 13.8psi vacuum @13:20:34.78 Cartridge 21 sampling 404/2000ml at 0.56ml/s with 14.1psi vacuum @13:21:34.91 Cartridge 21 sampling 438/2000ml at 0.54ml/s with 11.6psi vacuum @13:22:34.79 Cartridge 21 sampling 472/2000ml at 0.55ml/s with 13.1psi vacuum @13:23:34.84 Cartridge 21 sampling 505/2000ml at 0.56ml/s with 11.5psi vacuum @13:24:34.78 Cartridge 21 sampling 539/2000ml at 0.56ml/s with 12.6psi vacuum @13:25:35.07 Cartridge 21 sampling 573/2000ml at 0.56ml/s with 12.9psi vacuum @13:26:34.99 Cartridge 21 sampling 607/2000ml at 0.56ml/s with 14.0psi vacuum @13:27:34.90 Cartridge 21 sampling 642/2000ml at 0.57ml/s with 13.6psi vacuum @13:28:35.04 Cartridge 21 sampling 676/2000ml at 0.57ml/s with 13.2psi vacuum @13:29:34.79 Cartridge 21 sampling 710/2000ml at 0.56ml/s with 12.4psi vacuum @13:30:35.08 Cartridge 21 sampling 744/2000ml at 0.57ml/s with 13.2psi vacuum @13:31:34.79 Cartridge 21 sampling 778/2000ml at 0.57ml/s with 12.3psi vacuum @13:32:34.93 Cartridge 21 sampling 813/2000ml at 0.57ml/s with 12.2psi vacuum @13:33:34.78 Cartridge 21 sampling 847/2000ml at 0.58ml/s with 13.8psi vacuum @13:34:35.00 Cartridge 21 sampling 881/2000ml at 0.57ml/s with 12.5psi vacuum @13:35:34.78 Cartridge 21 sampling 916/2000ml at 0.58ml/s with 14.3psi vacuum @13:36:34.98 Cartridge 21 sampling 951/2000ml at 0.58ml/s with 14.7psi vacuum @13:37:34.78 Cartridge 21 sampling 986/2000ml at 0.58ml/s with 12.7psi vacuum @13:38:34.97 Cartridge 21 sampling 1020/2000ml at 0.58ml/s with 12.7psi vacuum @13:39:34.79 Cartridge 21 sampling 1055/2000ml at 0.57ml/s with 13.8psi vacuum @13:40:36.16 Cartridge 21 sampling 1090/2000ml at 0.58ml/s with 12.6psi vacuum @13:41:34.79 Cartridge 21 sampling 1125/2000ml at 0.59ml/s with 14.6psi vacuum @13:42:34.93 Cartridge 21 sampling 1161/2000ml at 0.59ml/s with 12.6psi vacuum @13:43:34.79 Cartridge 21 sampling 1195/2000ml at 0.55ml/s with 11.3psi vacuum @13:44:34.95 Cartridge 21 sampling 1229/2000ml at 0.55ml/s with 14.0psi vacuum @13:45:34.78 Cartridge 21 sampling 1263/2000ml at 0.57ml/s with 11.8psi vacuum @13:46:34.93 Cartridge 21 sampling 1297/2000ml at 0.59ml/s with 14.8psi vacuum @13:47:34.79 Cartridge 21 sampling 1331/2000ml at 0.56ml/s with 14.6psi vacuum @13:48:35.33 Cartridge 21 sampling 1365/2000ml at 0.57ml/s with 13.6psi vacuum @13:49:35.05 Cartridge 21 sampling 1399/2000ml at 0.57ml/s with 13.1psi vacuum @13:50:34.79 Cartridge 21 sampling 1433/2000ml at 0.56ml/s with 11.4psi vacuum @13:51:35.20 Cartridge 21 sampling 1467/2000ml at 0.56ml/s with 13.8psi vacuum @13:52:34.80 Cartridge 21 sampling 1501/2000ml at 0.57ml/s with 12.2psi vacuum @13:53:35.08 Cartridge 21 sampling 1536/2000ml at 0.57ml/s with 12.4psi vacuum @13:54:34.77 Cartridge 21 sampling 1570/2000ml at 0.57ml/s with 11.8psi vacuum @13:55:34.86 Cartridge 21 sampling 1604/2000ml at 0.57ml/s with 14.7psi vacuum @13:56:34.78 Cartridge 21 sampling 1639/2000ml at 0.57ml/s with 12.5psi vacuum @13:57:34.90 Cartridge 21 sampling 1673/2000ml at 0.57ml/s with 12.6psi vacuum @13:58:34.78 Cartridge 21 sampling 1708/2000ml at 0.57ml/s with 12.7psi vacuum @13:59:34.99 Cartridge 21 sampling 1742/2000ml at 0.58ml/s with 13.5psi vacuum @14:00:34.78 Cartridge 21 sampling 1778/2000ml at 0.59ml/s with 12.8psi vacuum @14:01:34.99 Cartridge 21 sampling 1813/2000ml at 0.58ml/s with 14.2psi vacuum @14:02:34.77 Cartridge 21 sampling 1848/2000ml at 0.59ml/s with 13.9psi vacuum @14:03:34.94 Cartridge 21 sampling 1884/2000ml at 0.58ml/s with 13.1psi vacuum @14:04:34.78 Cartridge 21 sampling 1918/2000ml at 0.55ml/s with 12.7psi vacuum @14:05:34.93 Cartridge 21 sampling 1952/2000ml at 0.57ml/s with 14.0psi vacuum @14:06:34.79 Cartridge 21 sampling 1986/2000ml at 0.57ml/s with 13.7psi vacuum @14:06:59.05 Pressure.coast @14:06:59.54 Intake.close @14:06:59.58 Exhaust.close @14:06:59.80 Toroid.reconfigure :hold @14:06:59.88 Toroid.seek :bypass @14:07:01.51 Toroid.reconfigure :default @14:07:01.60 Toroid.seek :clear @14:07:03.23 Sampled 2000.0ml @14:07:03.73 Plunge.seek :reagent @14:07:04.72 Sample loop closed. Ready to Deliver Reagents @14:07:05.48 Carousel.seek 21 @14:07:06.11 Plunge.seek :locked @14:07:07.48 Air.seek :clear @14:07:07.89 Lysis1.seek :clear @14:07:11.31 Archive.coast @14:07:11.39 Air.coast @14:07:11.46 Lysis1.coast @14:07:11.53 Lysis2.coast @14:07:11.60 Plunge.coast @14:07:11.68 SamplePump.coast @14:07:11.75 Toroid.coast @14:07:11.82 Carousel.coast @14:07:12.04 Gate.power :motor,:OFF @14:07:12.08 Gate.power :core,:OFF @14:07:12.14 Gate.power :analytic1,:OFF @14:07:12.21 Gate.power :analytic2,:OFF @14:07:12.28 Gate.power :analytic3,:OFF @14:07:12.34 Gate.power :main,:OFF @14:07:12.39 Safely stopped and ready to power off @14:07:12.59 delay 2:00 @14:09:12.74 Gate.power :main,:ON,limit to 5000mA @14:09:12.76 Gate.power :analytic1,:ON,limit to 5000mA @14:09:12.80 Gate.power :analytic2,:ON,limit to 5000mA @14:09:12.82 Gate.power :analytic3,:ON,limit to 5000mA @14:09:12.90 Powering core bus @14:09:15.71 Archive.configure :default @14:09:16.11 Air.configure :default @14:09:16.43 Lysis1.configure :default @14:09:16.77 Lysis2.configure :default @14:09:17.13 Plunge.configure :default @14:09:17.45 SamplePump.configure :default @14:09:17.79 ExtraPump.configure :default @14:09:18.43 Toroid.configure :default @14:09:18.66 Carousel.configure :default @14:09:18.82 Pressure.configure :default @14:09:18.90 All elf controllers running firmware version 4.61 @14:09:18.93 Gate.power :motor,:ON,limit to 5000mA @14:09:19.15 Archive.reconfigure :default @14:09:19.27 Air.reconfigure :default @14:09:19.58 Air.home.jog 0 @14:09:19.61 Archive.home.jog 0 @14:09:19.95 Plunge.reconfigure :default @14:09:20.00 Lysis2.reconfigure :default @14:09:20.04 Lysis1.reconfigure :default @14:09:20.11 Toroid.reconfigure :default @14:09:20.31 Plunge.home.jog 0 @14:09:20.35 Lysis2.home.jog 0 @14:09:20.38 Lysis1.home.jog 0 @14:09:20.42 Toroid.home.jog 0 @14:09:21.08 Archive.reconfigure :homing @14:09:21.15 Air.reconfigure :homing @14:09:21.28 Archive.jog 1358 @14:09:21.32 Air.jog 1363 @14:09:21.55 Toroid.reconfigure :homing @14:09:21.60 Plunge.reconfigure :homing @14:09:21.64 Lysis2.reconfigure :homing @14:09:22.30 Lysis1.reconfigure :homing @14:09:22.47 Toroid.jog 616 @14:09:22.55 Plunge.jog 851 @14:09:22.59 Lysis2.jog 1360 @14:09:22.64 Lysis1.jog 1350 @14:09:23.08 Air.setRawPosition! 50 @14:09:23.18 Archive.setRawPosition! 50 @14:09:23.47 Air.seek :home @14:09:23.81 Archive.seek :home @14:09:24.32 Plunge.setRawPosition! 50 @14:09:24.39 Lysis2.setRawPosition! 50 @14:09:24.50 Lysis1.setRawPosition! 50 @14:09:24.56 Archive.reconfigure :default @14:09:24.61 Air.reconfigure :default @14:09:25.17 Plunge.seek :home @14:09:25.21 Lysis2.seek :home @14:09:25.34 Lysis1.seek :home @14:09:26.07 Plunge.reconfigure :default @14:09:26.18 Toroid.setRawPosition! 0 @14:09:26.32 Lysis2.reconfigure :default @14:09:26.65 Toroid.seek :home @14:09:27.07 Toroid.reconfigure :default @14:09:27.13 Lysis1.reconfigure :default @14:09:27.77 Toroid.seek :clear @14:09:29.42 Carousel.reconfigure :default @14:09:29.51 Carousel.home.jog 0 @14:09:30.02 Carousel.seek 21 @14:09:30.21 Selecting Cartridge 20 @14:09:30.47 Carousel.seek 20 @14:09:31.02 Lysis1.seek :locked @14:09:34.66 Plunge.seek :locked @14:09:35.15 Air.seek :locked @14:09:39.43 Mounted archive Cartridge @14:09:39.75 Duration of filtering limited to 4:05:00 @14:09:40.07 Priming sample loop w/160ml, bypass w/10ml @14:09:40.09 SamplePump.setPosition! 0ml @14:09:40.16 Bypass.open @14:09:41.67 Intake.open @14:09:42.59 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @14:09:45.33 Vacuum tare is 18% of full scale @14:09:45.36 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @14:09:45.95 Pressure.reconfigure :default @14:09:46.18 Pressure.hold @14:09:46.75 SamplePump.reconfigure :prime @14:09:46.95 SamplePump.seek 10ml,1:21 @14:09:58.13 Pressure.coast @14:09:58.20 SamplePump.setPosition! 0ml @14:09:58.27 Exhaust.open @14:09:59.67 Bypass.close @14:10:01.26 Pressure.reconfigure :default @14:10:01.34 Pressure.hold @14:10:01.63 SamplePump.reconfigure :prime @14:10:01.72 SamplePump.seek 160ml,13:51 @14:12:53.30 Pressure.coast @14:12:53.37 Bypass.open @14:12:54.08 Intake.close @14:12:54.98 Toroid.reconfigure :hold @14:12:55.06 Toroid.seek :crack @14:12:56.80 Toroid.seek :cartridge @14:12:58.00 Toroid.reconfigure :default @14:12:58.14 Toroid.seek :clear @14:13:00.29 Pressurizing cartridge to 14psi above 14.3psia @14:13:00.33 SamplePump.setPosition! 0ml @14:13:00.51 Intake.open @14:13:01.33 Bypass.close @14:13:02.11 Exhaust.close @14:13:02.95 Pressure.reconfigure :default @14:13:03.02 Pressure.hold @14:13:03.25 SamplePump.reconfigure :charge @14:13:03.35 SamplePump.seek 2.3ml,42 seconds @14:13:12.18 Pressure.coast @14:13:12.25 Intake.close @14:13:13.15 Pressurized cartridge with 0.352ml to 14.0psi @14:13:13.16 Isolated @25.5psia (11.2psi) @14:13:13.18 Checking for cartridge leaks from 25.5psia @14:13:13.23 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @14:13:17.93 Cartridge pressure fell 2% to 24.9psia @14:13:18.10 Sampling 2000.0ml @14:13:18.13 SamplePump.setPosition! 0ml @14:13:18.34 Exhaust.open @14:13:19.08 Intake.open @14:13:20.44 Pressure.reconfigure :default @14:13:20.51 Pressure.hold @14:13:21.36 SamplePump.reconfigure :sample @14:13:21.45 SamplePump.seek 2000ml,2:47:12 @14:13:23.50 Cartridge 20 sampling 1/2000ml at 0.59ml/s with 14.2psi vacuum @14:14:23.66 Cartridge 20 sampling 32/2000ml at 0.54ml/s with 11.7psi vacuum @14:15:23.50 Cartridge 20 sampling 65/2000ml at 0.55ml/s with 14.9psi vacuum @14:16:23.58 Cartridge 20 sampling 98/2000ml at 0.55ml/s with 14.6psi vacuum @14:17:23.49 Cartridge 20 sampling 131/2000ml at 0.56ml/s with 11.0psi vacuum @14:18:23.84 Cartridge 20 sampling 165/2000ml at 0.55ml/s with 13.4psi vacuum @14:19:23.67 Cartridge 20 sampling 198/2000ml at 0.56ml/s with 10.6psi vacuum @14:20:23.50 Cartridge 20 sampling 232/2000ml at 0.57ml/s with 13.1psi vacuum @14:21:24.10 Cartridge 20 sampling 266/2000ml at 0.55ml/s with 12.1psi vacuum @14:22:23.50 Cartridge 20 sampling 299/2000ml at 0.57ml/s with 12.9psi vacuum @14:23:23.71 Cartridge 20 sampling 333/2000ml at 0.57ml/s with 14.5psi vacuum @14:24:23.51 Cartridge 20 sampling 367/2000ml at 0.57ml/s with 11.8psi vacuum @14:25:23.75 Cartridge 20 sampling 402/2000ml at 0.56ml/s with 13.0psi vacuum @14:26:23.50 Cartridge 20 sampling 436/2000ml at 0.56ml/s with 14.5psi vacuum @14:27:23.68 Cartridge 20 sampling 470/2000ml at 0.56ml/s with 14.8psi vacuum @14:28:23.50 Cartridge 20 sampling 504/2000ml at 0.57ml/s with 11.3psi vacuum @14:29:23.67 Cartridge 20 sampling 538/2000ml at 0.56ml/s with 11.5psi vacuum @14:30:23.52 Cartridge 20 sampling 572/2000ml at 0.58ml/s with 11.2psi vacuum @14:31:23.69 Cartridge 20 sampling 606/2000ml at 0.58ml/s with 14.8psi vacuum @14:32:23.50 Cartridge 20 sampling 641/2000ml at 0.57ml/s with 12.5psi vacuum @14:33:23.72 Cartridge 20 sampling 675/2000ml at 0.57ml/s with 13.0psi vacuum @14:34:23.51 Cartridge 20 sampling 709/2000ml at 0.57ml/s with 13.8psi vacuum @14:35:23.65 Cartridge 20 sampling 744/2000ml at 0.56ml/s with 12.5psi vacuum @14:36:23.51 Cartridge 20 sampling 778/2000ml at 0.58ml/s with 14.1psi vacuum @14:37:23.70 Cartridge 20 sampling 813/2000ml at 0.57ml/s with 12.4psi vacuum @14:38:23.50 Cartridge 20 sampling 847/2000ml at 0.57ml/s with 12.0psi vacuum @14:39:23.74 Cartridge 20 sampling 882/2000ml at 0.57ml/s with 12.0psi vacuum @14:40:23.51 Cartridge 20 sampling 917/2000ml at 0.58ml/s with 14.5psi vacuum @14:41:23.65 Cartridge 20 sampling 952/2000ml at 0.58ml/s with 14.1psi vacuum @14:42:23.50 Cartridge 20 sampling 987/2000ml at 0.58ml/s with 12.6psi vacuum @14:43:23.54 Cartridge 20 sampling 1022/2000ml at 0.59ml/s with 12.7psi vacuum @14:44:23.51 Cartridge 20 sampling 1057/2000ml at 0.59ml/s with 13.6psi vacuum @14:45:23.91 Cartridge 20 sampling 1092/2000ml at 0.59ml/s with 12.6psi vacuum @14:46:23.81 Cartridge 20 sampling 1127/2000ml at 0.58ml/s with 13.4psi vacuum @14:47:23.80 Cartridge 20 sampling 1163/2000ml at 0.57ml/s with 13.2psi vacuum @14:48:23.78 Cartridge 20 sampling 1196/2000ml at 0.57ml/s with 12.6psi vacuum @14:49:23.57 Cartridge 20 sampling 1230/2000ml at 0.55ml/s with 12.5psi vacuum @14:50:23.64 Cartridge 20 sampling 1264/2000ml at 0.57ml/s with 14.0psi vacuum @14:51:23.49 Cartridge 20 sampling 1298/2000ml at 0.57ml/s with 13.2psi vacuum @14:52:23.99 Cartridge 20 sampling 1332/2000ml at 0.57ml/s with 14.9psi vacuum @14:53:23.49 Cartridge 20 sampling 1366/2000ml at 0.56ml/s with 13.8psi vacuum @14:54:23.67 Cartridge 20 sampling 1400/2000ml at 0.58ml/s with 13.0psi vacuum @14:55:23.49 Cartridge 20 sampling 1434/2000ml at 0.56ml/s with 13.1psi vacuum @14:56:23.64 Cartridge 20 sampling 1468/2000ml at 0.56ml/s with 14.5psi vacuum @14:57:23.50 Cartridge 20 sampling 1502/2000ml at 0.56ml/s with 13.6psi vacuum @14:58:23.73 Cartridge 20 sampling 1536/2000ml at 0.57ml/s with 12.8psi vacuum @14:59:23.50 Cartridge 20 sampling 1571/2000ml at 0.59ml/s with 12.8psi vacuum @15:00:23.69 Cartridge 20 sampling 1606/2000ml at 0.58ml/s with 13.0psi vacuum @15:01:23.49 Cartridge 20 sampling 1640/2000ml at 0.57ml/s with 13.8psi vacuum @15:02:23.67 Cartridge 20 sampling 1675/2000ml at 0.57ml/s with 13.7psi vacuum @15:03:23.49 Cartridge 20 sampling 1709/2000ml at 0.57ml/s with 12.4psi vacuum @15:04:23.56 Cartridge 20 sampling 1744/2000ml at 0.58ml/s with 13.0psi vacuum @15:05:23.49 Cartridge 20 sampling 1779/2000ml at 0.58ml/s with 12.0psi vacuum @15:06:23.67 Cartridge 20 sampling 1814/2000ml at 0.58ml/s with 15.0psi vacuum @15:07:23.49 Cartridge 20 sampling 1849/2000ml at 0.59ml/s with 13.1psi vacuum @15:08:23.86 Cartridge 20 sampling 1884/2000ml at 0.58ml/s with 14.3psi vacuum @15:09:23.78 Cartridge 20 sampling 1919/2000ml at 0.57ml/s with 14.6psi vacuum @15:10:23.76 Cartridge 20 sampling 1954/2000ml at 0.59ml/s with 13.2psi vacuum @15:11:23.77 Cartridge 20 sampling 1989/2000ml at 0.59ml/s with 13.0psi vacuum @15:11:42.86 Pressure.coast @15:11:43.23 Intake.close @15:11:43.43 Exhaust.close @15:11:43.95 Toroid.reconfigure :hold @15:11:44.09 Toroid.seek :bypass @15:11:45.91 Toroid.reconfigure :default @15:11:45.99 Toroid.seek :clear @15:11:47.66 Sampled 2000.0ml @15:11:48.18 Plunge.seek :reagent @15:11:49.10 Sample loop closed. Ready to Deliver Reagents @15:11:49.37 Carousel.seek 20 @15:11:49.60 Plunge.seek :locked @15:11:50.68 Air.seek :clear @15:11:50.87 Lysis1.seek :clear @15:11:54.40 Archive.coast @15:11:54.65 Air.coast @15:11:54.86 Lysis1.coast @15:11:55.04 Lysis2.coast @15:11:55.16 Plunge.coast @15:11:55.27 SamplePump.coast @15:11:55.46 Toroid.coast @15:11:55.57 Carousel.coast @15:11:57.01 Gate.power :motor,:OFF @15:11:57.03 Gate.power :core,:OFF @15:11:57.06 Gate.power :analytic1,:OFF @15:11:57.10 Gate.power :analytic2,:OFF @15:11:57.13 Gate.power :analytic3,:OFF @15:11:57.16 Gate.power :main,:OFF @15:11:57.19 Safely stopped and ready to power off @15:11:57.60 delay 2:00 @15:13:57.76 Gate.power :main,:ON,limit to 5000mA @15:13:57.79 Gate.power :analytic1,:ON,limit to 5000mA @15:13:57.83 Gate.power :analytic2,:ON,limit to 5000mA @15:13:57.87 Gate.power :analytic3,:ON,limit to 5000mA @15:13:57.93 Powering core bus @15:14:00.23 Archive.configure :default @15:14:00.40 Air.configure :default @15:14:00.80 Lysis1.configure :default @15:14:01.22 Lysis2.configure :default @15:14:01.71 Plunge.configure :default @15:14:02.26 SamplePump.configure :default @15:14:02.70 ExtraPump.configure :default @15:14:03.13 Toroid.configure :default @15:14:03.49 Carousel.configure :default @15:14:03.87 Pressure.configure :default @15:14:04.06 All elf controllers running firmware version 4.61 @15:14:04.15 Gate.power :motor,:ON,limit to 5000mA @15:14:05.29 Plunge.reconfigure :default @15:14:05.43 Toroid.reconfigure :default @15:14:05.58 Lysis1.reconfigure :default @15:14:05.64 Air.reconfigure :default @15:14:05.68 Archive.reconfigure :default @15:14:05.81 Lysis2.reconfigure :default @15:14:06.53 Plunge.home.jog 0 @15:14:06.71 Toroid.home.jog 0 @15:14:06.78 Archive.home.jog 0 @15:14:06.81 Lysis1.home.jog 0 @15:14:06.85 Air.home.jog 0 @15:14:06.90 Lysis2.home.jog 0 @15:14:07.73 Toroid.reconfigure :homing @15:14:07.76 Plunge.reconfigure :homing @15:14:07.86 Lysis1.reconfigure :homing @15:14:07.91 Air.reconfigure :homing @15:14:07.95 Lysis2.reconfigure :homing @15:14:08.02 Archive.reconfigure :homing @15:14:08.27 Toroid.jog 620 @15:14:08.31 Plunge.jog 851 @15:14:08.35 Lysis1.jog 1349 @15:14:08.39 Air.jog 1364 @15:14:08.43 Lysis2.jog 1361 @15:14:08.47 Archive.jog 1358 @15:14:09.46 Plunge.setRawPosition! 50 @15:14:09.52 Plunge.seek :home @15:14:09.71 Lysis1.setRawPosition! 50 @15:14:09.81 Lysis1.seek :home @15:14:09.95 Plunge.reconfigure :default @15:14:09.99 Lysis2.setRawPosition! 50 @15:14:10.03 Air.setRawPosition! 50 @15:14:10.06 Archive.setRawPosition! 50 @15:14:10.24 Lysis2.seek :home @15:14:10.29 Air.seek :home @15:14:10.38 Archive.seek :home @15:14:10.56 Lysis1.reconfigure :default @15:14:10.71 Toroid.setRawPosition! 0 @15:14:10.77 Lysis2.reconfigure :default @15:14:10.82 Air.reconfigure :default @15:14:10.91 Toroid.seek :home @15:14:11.30 Archive.reconfigure :default @15:14:11.37 Toroid.reconfigure :default @15:14:12.10 Toroid.seek :clear @15:14:13.88 Carousel.reconfigure :default @15:14:14.12 Carousel.home.jog 0 @15:14:14.85 Carousel.seek 20 @15:14:15.22 Selecting Cartridge 19 @15:14:15.79 Carousel.seek 19 @15:14:16.75 Lysis1.seek :locked @15:14:20.11 Plunge.seek :locked @15:14:20.28 Air.seek :locked @15:14:24.64 Mounted archive Cartridge @15:14:25.37 Duration of filtering limited to 4:05:00 @15:14:26.15 Priming sample loop w/160ml, bypass w/10ml @15:14:26.27 SamplePump.setPosition! 0ml @15:14:26.40 Bypass.open @15:14:27.91 Intake.open @15:14:28.82 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:14:31.71 Vacuum tare is 18% of full scale @15:14:31.86 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @15:14:32.49 Pressure.reconfigure :default @15:14:32.71 Pressure.hold @15:14:33.19 SamplePump.reconfigure :prime @15:14:33.52 SamplePump.seek 10ml,1:21 @15:14:44.76 Pressure.coast @15:14:45.03 SamplePump.setPosition! 0ml @15:14:45.18 Exhaust.open @15:14:46.61 Bypass.close @15:14:48.23 Pressure.reconfigure :default @15:14:48.31 Pressure.hold @15:14:49.15 SamplePump.reconfigure :prime @15:14:49.24 SamplePump.seek 160ml,13:51 @15:17:40.86 Pressure.coast @15:17:40.94 Bypass.open @15:17:41.64 Intake.close @15:17:42.55 Toroid.reconfigure :hold @15:17:42.64 Toroid.seek :crack @15:17:44.28 Toroid.seek :cartridge @15:17:45.28 Toroid.reconfigure :default @15:17:45.49 Toroid.seek :clear @15:17:47.74 Pressurizing cartridge to 14psi above 14.1psia @15:17:47.79 SamplePump.setPosition! 0ml @15:17:47.95 Intake.open @15:17:48.87 Bypass.close @15:17:49.59 Exhaust.close @15:17:50.45 Pressure.reconfigure :default @15:17:50.53 Pressure.hold @15:17:50.77 SamplePump.reconfigure :charge @15:17:50.86 SamplePump.seek 2.3ml,42 seconds @15:17:59.38 Pressure.coast @15:17:59.46 Intake.close @15:18:00.33 Pressurized cartridge with 0.328ml to 14.0psi @15:18:00.35 Isolated @25.6psia (11.5psi) @15:18:00.37 Checking for cartridge leaks from 25.6psia @15:18:00.41 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @15:18:05.10 Cartridge pressure fell 3% to 24.8psia @15:18:05.28 Sampling 2000.0ml @15:18:05.36 SamplePump.setPosition! 0ml @15:18:05.54 Exhaust.open @15:18:06.37 Intake.open @15:18:07.58 Pressure.reconfigure :default @15:18:07.68 Pressure.hold @15:18:08.35 SamplePump.reconfigure :sample @15:18:08.59 SamplePump.seek 2000ml,2:47:12 @15:18:10.59 Cartridge 19 sampling 1/2000ml at 0.60ml/s with 15.0psi vacuum @15:19:10.83 Cartridge 19 sampling 31/2000ml at 0.51ml/s with 14.5psi vacuum @15:20:10.59 Cartridge 19 sampling 63/2000ml at 0.54ml/s with 11.7psi vacuum @15:21:10.77 Cartridge 19 sampling 95/2000ml at 0.55ml/s with 10.6psi vacuum @15:22:10.59 Cartridge 19 sampling 128/2000ml at 0.54ml/s with 11.0psi vacuum @15:23:10.76 Cartridge 19 sampling 161/2000ml at 0.56ml/s with 11.2psi vacuum @15:24:10.59 Cartridge 19 sampling 194/2000ml at 0.56ml/s with 14.5psi vacuum @15:25:10.70 Cartridge 19 sampling 228/2000ml at 0.57ml/s with 14.1psi vacuum @15:26:10.59 Cartridge 19 sampling 262/2000ml at 0.56ml/s with 12.9psi vacuum @15:27:10.71 Cartridge 19 sampling 296/2000ml at 0.56ml/s with 14.2psi vacuum @15:28:10.59 Cartridge 19 sampling 330/2000ml at 0.57ml/s with 12.2psi vacuum @15:29:10.81 Cartridge 19 sampling 364/2000ml at 0.57ml/s with 10.9psi vacuum @15:30:11.08 Cartridge 19 sampling 397/2000ml at 0.56ml/s with 13.3psi vacuum @15:31:10.59 Cartridge 19 sampling 431/2000ml at 0.56ml/s with 12.6psi vacuum @15:32:10.80 Cartridge 19 sampling 465/2000ml at 0.56ml/s with 14.9psi vacuum @15:33:10.59 Cartridge 19 sampling 499/2000ml at 0.57ml/s with 12.0psi vacuum @15:34:10.91 Cartridge 19 sampling 533/2000ml at 0.55ml/s with 13.6psi vacuum @15:35:10.59 Cartridge 19 sampling 566/2000ml at 0.56ml/s with 13.9psi vacuum @15:36:10.80 Cartridge 19 sampling 600/2000ml at 0.56ml/s with 11.4psi vacuum @15:37:10.59 Cartridge 19 sampling 634/2000ml at 0.57ml/s with 11.2psi vacuum @15:38:10.79 Cartridge 19 sampling 668/2000ml at 0.57ml/s with 13.0psi vacuum @15:39:10.59 Cartridge 19 sampling 702/2000ml at 0.57ml/s with 12.5psi vacuum @15:40:10.71 Cartridge 19 sampling 736/2000ml at 0.57ml/s with 13.0psi vacuum @15:41:10.59 Cartridge 19 sampling 770/2000ml at 0.56ml/s with 11.9psi vacuum @15:42:10.80 Cartridge 19 sampling 805/2000ml at 0.57ml/s with 13.7psi vacuum @15:43:10.59 Cartridge 19 sampling 839/2000ml at 0.57ml/s with 14.0psi vacuum @15:44:10.76 Cartridge 19 sampling 873/2000ml at 0.56ml/s with 14.6psi vacuum @15:45:10.59 Cartridge 19 sampling 907/2000ml at 0.58ml/s with 13.0psi vacuum @15:46:10.76 Cartridge 19 sampling 942/2000ml at 0.58ml/s with 14.1psi vacuum @15:47:10.60 Cartridge 19 sampling 977/2000ml at 0.58ml/s with 13.5psi vacuum @15:48:10.75 Cartridge 19 sampling 1011/2000ml at 0.58ml/s with 12.6psi vacuum @15:49:10.59 Cartridge 19 sampling 1046/2000ml at 0.57ml/s with 12.9psi vacuum @15:50:10.86 Cartridge 19 sampling 1080/2000ml at 0.57ml/s with 13.7psi vacuum @15:51:10.79 Cartridge 19 sampling 1115/2000ml at 0.58ml/s with 11.0psi vacuum @15:52:10.60 Cartridge 19 sampling 1149/2000ml at 0.56ml/s with 13.2psi vacuum @15:53:11.05 Cartridge 19 sampling 1183/2000ml at 0.56ml/s with 11.6psi vacuum @15:54:10.60 Cartridge 19 sampling 1217/2000ml at 0.56ml/s with 11.6psi vacuum @15:55:10.94 Cartridge 19 sampling 1251/2000ml at 0.57ml/s with 13.9psi vacuum @15:56:10.60 Cartridge 19 sampling 1285/2000ml at 0.58ml/s with 14.1psi vacuum @15:57:10.85 Cartridge 19 sampling 1319/2000ml at 0.57ml/s with 12.5psi vacuum @15:58:10.60 Cartridge 19 sampling 1353/2000ml at 0.57ml/s with 14.5psi vacuum @15:59:10.77 Cartridge 19 sampling 1387/2000ml at 0.57ml/s with 13.5psi vacuum @16:00:10.59 Cartridge 19 sampling 1421/2000ml at 0.57ml/s with 14.6psi vacuum @16:01:10.76 Cartridge 19 sampling 1456/2000ml at 0.58ml/s with 11.9psi vacuum @16:02:10.60 Cartridge 19 sampling 1490/2000ml at 0.56ml/s with 14.0psi vacuum @16:03:10.85 Cartridge 19 sampling 1525/2000ml at 0.58ml/s with 12.5psi vacuum @16:04:10.59 Cartridge 19 sampling 1559/2000ml at 0.56ml/s with 13.5psi vacuum @16:05:10.76 Cartridge 19 sampling 1594/2000ml at 0.57ml/s with 14.0psi vacuum @16:06:10.59 Cartridge 19 sampling 1628/2000ml at 0.58ml/s with 13.3psi vacuum @16:07:10.67 Cartridge 19 sampling 1663/2000ml at 0.58ml/s with 13.5psi vacuum @16:08:10.59 Cartridge 19 sampling 1698/2000ml at 0.58ml/s with 14.2psi vacuum @16:09:10.71 Cartridge 19 sampling 1732/2000ml at 0.56ml/s with 12.2psi vacuum @16:10:10.59 Cartridge 19 sampling 1767/2000ml at 0.57ml/s with 12.4psi vacuum @16:11:11.04 Cartridge 19 sampling 1801/2000ml at 0.57ml/s with 12.5psi vacuum @16:12:10.62 Cartridge 19 sampling 1836/2000ml at 0.57ml/s with 13.9psi vacuum @16:13:10.60 Cartridge 19 sampling 1870/2000ml at 0.57ml/s with 13.0psi vacuum @16:14:10.82 Cartridge 19 sampling 1905/2000ml at 0.57ml/s with 12.9psi vacuum @16:15:10.60 Cartridge 19 sampling 1940/2000ml at 0.56ml/s with 12.3psi vacuum @16:16:10.84 Cartridge 19 sampling 1974/2000ml at 0.56ml/s with 14.8psi vacuum @16:16:57.38 Pressure.coast @16:16:57.74 Intake.close @16:16:57.93 Exhaust.close @16:16:58.34 Toroid.reconfigure :hold @16:16:58.54 Toroid.seek :bypass @16:17:00.34 Toroid.reconfigure :default @16:17:00.70 Toroid.seek :clear @16:17:02.49 Sampled 2000.0ml @16:17:02.99 Plunge.seek :reagent @16:17:03.91 Sample loop closed. Ready to Deliver Reagents @16:17:04.19 Carousel.seek 19 @16:17:04.62 Plunge.seek :locked @16:17:06.02 Air.seek :clear @16:17:06.67 Lysis1.seek :clear @16:17:10.25 Archive.coast @16:17:10.43 Air.coast @16:17:10.60 Lysis1.coast @16:17:10.66 Lysis2.coast @16:17:10.73 Plunge.coast @16:17:10.80 SamplePump.coast @16:17:10.87 Toroid.coast @16:17:10.94 Carousel.coast @16:17:11.15 Gate.power :motor,:OFF @16:17:11.17 Gate.power :core,:OFF @16:17:11.20 Gate.power :analytic1,:OFF @16:17:11.25 Gate.power :analytic2,:OFF @16:17:11.27 Gate.power :analytic3,:OFF @16:17:11.30 Gate.power :main,:OFF @16:17:11.33 Safely stopped and ready to power off @16:17:11.53 delay 2:00 @16:46:00.46 -> slots 58..53=>[:dry,:archiveNull_bac] @16:46:12.97 -> Cartridge.dry @16:46:34.92 -> 6.times {shoot; delay 120} @16:46:35.31 Gate.power :main,:ON,limit to 5000mA @16:46:35.34 Gate.power :analytic1,:ON,limit to 5000mA @16:46:35.53 Gate.power :analytic2,:ON,limit to 5000mA @16:46:35.56 Gate.power :analytic3,:ON,limit to 5000mA @16:46:35.81 Powering core bus @16:46:38.14 Archive.configure :default @16:46:38.31 Air.configure :default @16:46:38.46 Lysis1.configure :default @16:46:38.60 Lysis2.configure :default @16:46:38.77 Plunge.configure :default @16:46:38.94 SamplePump.configure :default @16:46:39.10 ExtraPump.configure :default @16:46:39.27 Toroid.configure :default @16:46:39.44 Carousel.configure :default @16:46:39.60 Pressure.configure :default @16:46:39.68 All elf controllers running firmware version 4.61 @16:46:39.71 Gate.power :motor,:ON,limit to 5000mA @16:46:39.94 Archive.reconfigure :default @16:46:40.37 Toroid.reconfigure :default @16:46:40.41 Archive.home.jog 0 @16:46:40.46 Plunge.reconfigure :default @16:46:40.51 Lysis2.reconfigure :default @16:46:40.55 Lysis1.reconfigure :default @16:46:40.59 Air.reconfigure :default @16:46:40.91 Toroid.home.jog 0 @16:46:41.02 Plunge.home.jog 0 @16:46:41.14 Lysis2.home.jog 0 @16:46:41.23 Lysis1.home.jog 0 @16:46:41.37 Air.home.jog 0 @16:46:43.36 Toroid.reconfigure :homing @16:46:43.60 Lysis2.reconfigure :homing @16:46:43.64 Lysis1.reconfigure :homing @16:46:43.78 Air.reconfigure :homing @16:46:43.91 Plunge.reconfigure :homing @16:46:44.03 Archive.reconfigure :homing @16:46:44.62 Toroid.jog 617 @16:46:44.77 Lysis2.jog 1360 @16:46:44.90 Lysis1.jog 1353 @16:46:44.98 Air.jog 1362 @16:46:45.09 Archive.jog 1360 @16:46:45.14 Plunge.jog 855 @16:46:46.53 Lysis2.setRawPosition! 50 @16:46:46.76 Lysis2.seek :home @16:46:47.34 Lysis2.reconfigure :default @16:46:47.39 Lysis1.setRawPosition! 50 @16:46:47.42 Air.setRawPosition! 50 @16:46:47.48 Plunge.setRawPosition! 50 @16:46:47.51 Archive.setRawPosition! 50 @16:46:47.73 Lysis1.seek :home @16:46:47.81 Air.seek :home @16:46:47.86 Plunge.seek :home @16:46:47.91 Archive.seek :home @16:46:48.22 Toroid.setRawPosition! 0 @16:46:48.93 Toroid.seek :home @16:46:49.09 Plunge.reconfigure :default @16:46:49.14 Lysis1.reconfigure :default @16:46:49.18 Archive.reconfigure :default @16:46:49.24 Air.reconfigure :default @16:46:49.29 Toroid.reconfigure :default @16:46:49.91 Toroid.seek :clear @16:46:51.55 Carousel.reconfigure :default @16:46:51.67 Carousel.home.jog 0 @16:46:52.54 Carousel.seek 19 @16:46:53.32 Selecting Cartridge 58 @16:46:54.08 Carousel.seek 58 @16:47:02.87 Lysis1.seek :locked @16:47:06.57 Plunge.seek :locked @16:47:06.75 Air.seek :locked @16:47:11.08 Mounted archive Cartridge @16:47:12.00 Duration of filtering limited to 4:05:00 @16:47:12.89 Priming sample loop w/160ml, bypass w/10ml @16:47:12.92 SamplePump.setPosition! 0ml @16:47:13.06 Bypass.open @16:47:14.85 Intake.open @16:47:15.93 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @16:47:18.53 Vacuum tare is 18% of full scale @16:47:18.55 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @16:47:18.83 Pressure.reconfigure :default @16:47:18.91 Pressure.hold @16:47:19.16 SamplePump.reconfigure :prime @16:47:19.25 SamplePump.seek 10ml,1:21 @16:47:30.42 Pressure.coast @16:47:30.62 SamplePump.setPosition! 0ml @16:47:31.00 Exhaust.open @16:47:32.62 Bypass.close @16:47:34.49 Pressure.reconfigure :default @16:47:34.68 Pressure.hold @16:47:35.59 SamplePump.reconfigure :prime @16:47:35.68 SamplePump.seek 160ml,13:51 @16:50:27.35 Pressure.coast @16:50:27.53 Bypass.open @16:50:28.38 Intake.close @16:50:29.46 Toroid.reconfigure :hold @16:50:29.57 Toroid.seek :crack @16:50:31.40 Toroid.seek :cartridge @16:50:32.37 Toroid.reconfigure :default @16:50:32.45 Toroid.seek :clear @16:50:34.32 Pressurizing cartridge to 14psi above 13.8psia @16:50:34.35 SamplePump.setPosition! 0ml @16:50:34.42 Intake.open @16:50:35.25 Bypass.close @16:50:36.12 Exhaust.close @16:50:37.47 Pressure.reconfigure :default @16:50:37.72 Pressure.hold @16:50:38.22 SamplePump.reconfigure :charge @16:50:38.36 SamplePump.seek 2.3ml,42 seconds @16:50:47.07 Pressure.coast @16:50:47.24 Intake.close @16:50:48.41 Pressurized cartridge with 0.329ml to 14.0psi @16:50:48.42 Isolated @25.1psia (11.3psi) @16:50:48.48 Checking for cartridge leaks from 25.1psia @16:50:48.65 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @16:50:53.26 Cartridge pressure fell 6% to 23.7psia @16:50:53.37 Sampling 2000.0ml @16:50:53.38 SamplePump.setPosition! 0ml @16:50:53.46 Exhaust.open @16:50:54.20 Intake.open @16:50:55.68 Pressure.reconfigure :default @16:50:55.98 Pressure.hold @16:50:56.72 SamplePump.reconfigure :sample @16:50:56.91 SamplePump.seek 2000ml,2:47:12 @16:50:58.64 Cartridge 58 sampling 0/2000ml at 0.56ml/s with 12.0psi vacuum @16:51:58.48 Cartridge 58 sampling 32/2000ml at 0.52ml/s with 13.4psi vacuum @16:52:58.53 Cartridge 58 sampling 64/2000ml at 0.55ml/s with 11.4psi vacuum @16:53:58.48 Cartridge 58 sampling 96/2000ml at 0.54ml/s with 10.4psi vacuum @16:54:58.63 Cartridge 58 sampling 130/2000ml at 0.56ml/s with 12.9psi vacuum @16:55:58.48 Cartridge 58 sampling 163/2000ml at 0.54ml/s with 15.5psi vacuum @16:56:58.92 Cartridge 58 sampling 197/2000ml at 0.57ml/s with 11.6psi vacuum @16:57:58.48 Cartridge 58 sampling 231/2000ml at 0.56ml/s with 12.3psi vacuum @16:58:58.49 Cartridge 58 sampling 265/2000ml at 0.55ml/s with 11.1psi vacuum @16:59:58.87 Cartridge 58 sampling 299/2000ml at 0.57ml/s with 12.5psi vacuum @17:00:59.03 Cartridge 58 sampling 333/2000ml at 0.57ml/s with 14.4psi vacuum @17:01:58.81 Cartridge 58 sampling 367/2000ml at 0.59ml/s with 11.7psi vacuum @17:02:58.48 Cartridge 58 sampling 401/2000ml at 0.58ml/s with 13.3psi vacuum @17:03:58.79 Cartridge 58 sampling 435/2000ml at 0.56ml/s with 11.9psi vacuum @17:04:58.47 Cartridge 58 sampling 469/2000ml at 0.57ml/s with 12.1psi vacuum @17:05:58.77 Cartridge 58 sampling 504/2000ml at 0.57ml/s with 13.1psi vacuum @17:06:58.47 Cartridge 58 sampling 538/2000ml at 0.57ml/s with 13.6psi vacuum @17:07:58.68 Cartridge 58 sampling 572/2000ml at 0.58ml/s with 13.0psi vacuum @17:08:58.47 Cartridge 58 sampling 607/2000ml at 0.55ml/s with 12.4psi vacuum @17:09:58.70 Cartridge 58 sampling 641/2000ml at 0.57ml/s with 13.6psi vacuum @17:10:58.47 Cartridge 58 sampling 675/2000ml at 0.57ml/s with 11.7psi vacuum @17:11:58.65 Cartridge 58 sampling 709/2000ml at 0.58ml/s with 12.7psi vacuum @17:12:58.47 Cartridge 58 sampling 744/2000ml at 0.57ml/s with 12.2psi vacuum @17:13:58.64 Cartridge 58 sampling 778/2000ml at 0.58ml/s with 12.2psi vacuum @17:14:58.47 Cartridge 58 sampling 813/2000ml at 0.57ml/s with 11.8psi vacuum @17:15:58.69 Cartridge 58 sampling 847/2000ml at 0.57ml/s with 12.0psi vacuum @17:16:58.48 Cartridge 58 sampling 882/2000ml at 0.58ml/s with 13.0psi vacuum @17:17:58.61 Cartridge 58 sampling 916/2000ml at 0.57ml/s with 12.2psi vacuum @17:18:58.47 Cartridge 58 sampling 951/2000ml at 0.58ml/s with 13.0psi vacuum @17:19:58.61 Cartridge 58 sampling 985/2000ml at 0.58ml/s with 12.7psi vacuum @17:20:58.48 Cartridge 58 sampling 1020/2000ml at 0.57ml/s with 13.8psi vacuum @17:21:58.97 Cartridge 58 sampling 1054/2000ml at 0.57ml/s with 13.4psi vacuum @17:22:58.52 Cartridge 58 sampling 1089/2000ml at 0.57ml/s with 13.9psi vacuum @17:23:58.63 Cartridge 58 sampling 1124/2000ml at 0.58ml/s with 12.9psi vacuum @17:24:58.74 Cartridge 58 sampling 1159/2000ml at 0.58ml/s with 12.2psi vacuum @17:25:58.48 Cartridge 58 sampling 1193/2000ml at 0.58ml/s with 11.6psi vacuum @17:26:58.88 Cartridge 58 sampling 1228/2000ml at 0.58ml/s with 13.8psi vacuum @17:27:58.49 Cartridge 58 sampling 1263/2000ml at 0.59ml/s with 13.4psi vacuum @17:28:58.80 Cartridge 58 sampling 1299/2000ml at 0.58ml/s with 14.3psi vacuum @17:29:58.48 Cartridge 58 sampling 1334/2000ml at 0.59ml/s with 12.9psi vacuum @17:30:58.71 Cartridge 58 sampling 1369/2000ml at 0.58ml/s with 14.6psi vacuum @17:31:58.48 Cartridge 58 sampling 1404/2000ml at 0.58ml/s with 14.4psi vacuum @17:32:58.61 Cartridge 58 sampling 1440/2000ml at 0.58ml/s with 13.1psi vacuum @17:33:58.49 Cartridge 58 sampling 1475/2000ml at 0.59ml/s with 13.2psi vacuum @17:34:58.81 Cartridge 58 sampling 1510/2000ml at 0.56ml/s with 13.1psi vacuum @17:35:58.48 Cartridge 58 sampling 1544/2000ml at 0.57ml/s with 12.2psi vacuum @17:36:58.65 Cartridge 58 sampling 1578/2000ml at 0.56ml/s with 11.2psi vacuum @17:37:58.48 Cartridge 58 sampling 1612/2000ml at 0.57ml/s with 14.0psi vacuum @17:38:58.58 Cartridge 58 sampling 1647/2000ml at 0.56ml/s with 11.4psi vacuum @17:39:58.48 Cartridge 58 sampling 1681/2000ml at 0.57ml/s with 12.8psi vacuum @17:40:58.59 Cartridge 58 sampling 1715/2000ml at 0.57ml/s with 13.8psi vacuum @17:41:58.48 Cartridge 58 sampling 1749/2000ml at 0.57ml/s with 12.0psi vacuum @17:42:58.52 Cartridge 58 sampling 1784/2000ml at 0.57ml/s with 12.1psi vacuum @17:43:58.48 Cartridge 58 sampling 1818/2000ml at 0.56ml/s with 12.7psi vacuum @17:44:58.62 Cartridge 58 sampling 1852/2000ml at 0.57ml/s with 11.5psi vacuum @17:45:58.84 Cartridge 58 sampling 1886/2000ml at 0.57ml/s with 11.6psi vacuum @17:46:58.47 Cartridge 58 sampling 1921/2000ml at 0.57ml/s with 11.8psi vacuum @17:47:58.98 Cartridge 58 sampling 1955/2000ml at 0.57ml/s with 13.8psi vacuum @17:48:58.48 Cartridge 58 sampling 1990/2000ml at 0.58ml/s with 14.0psi vacuum @17:49:15.96 Pressure.coast @17:49:16.46 Intake.close @17:49:16.53 Exhaust.close @17:49:16.73 Toroid.reconfigure :hold @17:49:16.82 Toroid.seek :bypass @17:49:18.45 Toroid.reconfigure :default @17:49:18.54 Toroid.seek :clear @17:49:20.16 Sampled 2000.0ml @17:49:20.66 Plunge.seek :reagent @17:49:21.72 Sample loop closed. Ready to Deliver Reagents @17:49:22.45 Carousel.seek 58 @17:49:23.00 Plunge.seek :locked @17:49:24.42 Air.seek :clear @17:49:24.79 Lysis1.seek :clear @17:49:28.16 Archive.coast @17:49:28.24 Air.coast @17:49:28.31 Lysis1.coast @17:49:28.38 Lysis2.coast @17:49:28.45 Plunge.coast @17:49:28.53 SamplePump.coast @17:49:28.60 Toroid.coast @17:49:28.68 Carousel.coast @17:49:28.90 Gate.power :motor,:OFF @17:49:28.92 Gate.power :core,:OFF @17:49:28.95 Gate.power :analytic1,:OFF @17:49:28.98 Gate.power :analytic2,:OFF @17:49:29.01 Gate.power :analytic3,:OFF @17:49:29.04 Gate.power :main,:OFF @17:49:29.07 Safely stopped and ready to power off @17:49:29.27 delay 2:00 @17:51:29.40 Gate.power :main,:ON,limit to 5000mA @17:51:29.43 Gate.power :analytic1,:ON,limit to 5000mA @17:51:29.46 Gate.power :analytic2,:ON,limit to 5000mA @17:51:29.49 Gate.power :analytic3,:ON,limit to 5000mA @17:51:29.57 Powering core bus @17:51:32.24 Archive.configure :default @17:51:32.57 Air.configure :default @17:51:33.03 Lysis1.configure :default @17:51:33.37 Lysis2.configure :default @17:51:33.76 Plunge.configure :default @17:51:34.29 SamplePump.configure :default @17:51:35.96 ExtraPump.configure :default @17:51:36.14 Toroid.configure :default @17:51:36.30 Carousel.configure :default @17:51:36.47 Pressure.configure :default @17:51:36.55 All elf controllers running firmware version 4.61 @17:51:36.58 Gate.power :motor,:ON,limit to 5000mA @17:51:36.90 Archive.reconfigure :default @17:51:37.12 Archive.home.jog 0 @17:51:37.36 Plunge.reconfigure :default @17:51:37.41 Lysis2.reconfigure :default @17:51:37.45 Lysis1.reconfigure :default @17:51:37.51 Air.reconfigure :default @17:51:37.56 Toroid.reconfigure :default @17:51:37.80 Plunge.home.jog 0 @17:51:37.84 Lysis2.home.jog 0 @17:51:37.88 Lysis1.home.jog 0 @17:51:37.92 Air.home.jog 0 @17:51:37.96 Toroid.home.jog 0 @17:51:38.48 Archive.reconfigure :homing @17:51:38.78 Archive.jog 1358 @17:51:38.84 Toroid.reconfigure :homing @17:51:38.88 Lysis2.reconfigure :homing @17:51:38.94 Lysis1.reconfigure :homing @17:51:38.98 Plunge.reconfigure :homing @17:51:39.05 Air.reconfigure :homing @17:51:39.29 Toroid.jog 609 @17:51:39.34 Plunge.jog 856 @17:51:39.38 Lysis2.jog 1362 @17:51:39.42 Air.jog 1366 @17:51:39.48 Lysis1.jog 1349 @17:51:40.31 Archive.setRawPosition! 50 @17:51:40.64 Archive.seek :home @17:51:41.10 Plunge.setRawPosition! 50 @17:51:41.20 Lysis2.setRawPosition! 50 @17:51:41.26 Lysis1.setRawPosition! 50 @17:51:41.32 Air.setRawPosition! 50 @17:51:41.75 Plunge.seek :home @17:51:41.85 Lysis2.seek :home @17:51:41.95 Lysis1.seek :home @17:51:42.07 Air.seek :home @17:51:42.74 Toroid.setRawPosition! 0 @17:51:42.85 Lysis2.reconfigure :default @17:51:43.01 Archive.reconfigure :default @17:51:43.14 Plunge.reconfigure :default @17:51:43.20 Lysis1.reconfigure :default @17:51:43.36 Toroid.seek :home @17:51:44.17 Toroid.reconfigure :default @17:51:44.19 Air.reconfigure :default @17:51:44.98 Toroid.seek :clear @17:51:46.61 Carousel.reconfigure :default @17:51:46.70 Carousel.home.jog 0 @17:51:47.22 Carousel.seek 58 @17:51:47.41 Selecting Cartridge 57 @17:51:47.66 Carousel.seek 57 @17:51:48.21 Lysis1.seek :locked @17:51:51.84 Plunge.seek :locked @17:51:52.18 Air.seek :locked @17:51:56.50 Mounted archive Cartridge @17:51:56.83 Duration of filtering limited to 4:05:00 @17:51:57.15 Priming sample loop w/160ml, bypass w/10ml @17:51:57.18 SamplePump.setPosition! 0ml @17:51:57.25 Bypass.open @17:51:58.72 Intake.open @17:51:59.64 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @17:52:02.34 Vacuum tare is 18% of full scale @17:52:02.38 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @17:52:02.94 Pressure.reconfigure :default @17:52:03.10 Pressure.hold @17:52:04.03 SamplePump.reconfigure :prime @17:52:04.51 SamplePump.seek 10ml,1:21 @17:52:15.70 Pressure.coast @17:52:15.77 SamplePump.setPosition! 0ml @17:52:15.84 Exhaust.open @17:52:17.28 Bypass.close @17:52:18.86 Pressure.reconfigure :default @17:52:18.94 Pressure.hold @17:52:19.25 SamplePump.reconfigure :prime @17:52:19.33 SamplePump.seek 160ml,13:51 @17:55:10.95 Pressure.coast @17:55:11.03 Bypass.open @17:55:11.78 Intake.close @17:55:12.72 Toroid.reconfigure :hold @17:55:12.80 Toroid.seek :crack @17:55:14.49 Toroid.seek :cartridge @17:55:15.65 Toroid.reconfigure :default @17:55:15.90 Toroid.seek :clear @17:55:18.07 Pressurizing cartridge to 14psi above 14.4psia @17:55:19.27 SamplePump.setPosition! 0ml @17:55:19.33 Intake.open @17:55:20.14 Bypass.close @17:55:20.92 Exhaust.close @17:55:21.79 Pressure.reconfigure :default @17:55:21.87 Pressure.hold @17:55:22.12 SamplePump.reconfigure :charge @17:55:22.20 SamplePump.seek 2.3ml,42 seconds @17:55:30.66 Pressure.coast @17:55:30.74 Intake.close @17:55:31.63 Pressurized cartridge with 0.323ml to 14.0psi @17:55:31.64 Isolated @25.4psia (11.0psi) @17:55:31.67 Checking for cartridge leaks from 25.4psia @17:55:31.72 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @17:55:36.47 Cartridge pressure fell 2% to 25.0psia @17:55:36.60 Sampling 2000.0ml @17:55:36.73 SamplePump.setPosition! 0ml @17:55:36.84 Exhaust.open @17:55:37.63 Intake.open @17:55:38.69 Pressure.reconfigure :default @17:55:38.77 Pressure.hold @17:55:39.06 SamplePump.reconfigure :sample @17:55:39.15 SamplePump.seek 2000ml,2:47:12 @17:55:41.76 Cartridge 57 sampling 1/2000ml at 0.51ml/s with 14.2psi vacuum @17:56:42.00 Cartridge 57 sampling 32/2000ml at 0.53ml/s with 13.8psi vacuum @17:57:41.82 Cartridge 57 sampling 64/2000ml at 0.54ml/s with 10.7psi vacuum @17:58:41.73 Cartridge 57 sampling 97/2000ml at 0.56ml/s with 14.7psi vacuum @17:59:42.19 Cartridge 57 sampling 131/2000ml at 0.56ml/s with 11.2psi vacuum @18:00:41.74 Cartridge 57 sampling 164/2000ml at 0.57ml/s with 14.0psi vacuum @18:01:41.97 Cartridge 57 sampling 198/2000ml at 0.56ml/s with 11.9psi vacuum @18:02:41.74 Cartridge 57 sampling 231/2000ml at 0.55ml/s with 10.4psi vacuum @18:03:41.92 Cartridge 57 sampling 265/2000ml at 0.56ml/s with 13.3psi vacuum @18:04:41.75 Cartridge 57 sampling 299/2000ml at 0.57ml/s with 11.1psi vacuum @18:05:41.94 Cartridge 57 sampling 334/2000ml at 0.57ml/s with 10.8psi vacuum @18:06:41.74 Cartridge 57 sampling 368/2000ml at 0.57ml/s with 11.7psi vacuum @18:07:41.96 Cartridge 57 sampling 402/2000ml at 0.55ml/s with 11.3psi vacuum @18:08:41.74 Cartridge 57 sampling 436/2000ml at 0.55ml/s with 11.7psi vacuum @18:09:42.25 Cartridge 57 sampling 469/2000ml at 0.57ml/s with 13.6psi vacuum @18:10:41.75 Cartridge 57 sampling 503/2000ml at 0.56ml/s with 14.1psi vacuum @18:11:41.86 Cartridge 57 sampling 537/2000ml at 0.56ml/s with 13.9psi vacuum @18:12:41.74 Cartridge 57 sampling 571/2000ml at 0.57ml/s with 11.2psi vacuum @18:13:41.92 Cartridge 57 sampling 604/2000ml at 0.57ml/s with 15.3psi vacuum @18:14:41.75 Cartridge 57 sampling 638/2000ml at 0.57ml/s with 12.6psi vacuum @18:15:41.90 Cartridge 57 sampling 672/2000ml at 0.57ml/s with 13.8psi vacuum @18:16:41.74 Cartridge 57 sampling 706/2000ml at 0.56ml/s with 13.8psi vacuum @18:17:41.77 Cartridge 57 sampling 741/2000ml at 0.57ml/s with 11.5psi vacuum @18:18:41.75 Cartridge 57 sampling 775/2000ml at 0.58ml/s with 14.3psi vacuum @18:19:42.05 Cartridge 57 sampling 809/2000ml at 0.57ml/s with 13.2psi vacuum @18:20:41.94 Cartridge 57 sampling 844/2000ml at 0.58ml/s with 12.3psi vacuum @18:21:41.76 Cartridge 57 sampling 878/2000ml at 0.58ml/s with 13.4psi vacuum @18:22:41.99 Cartridge 57 sampling 913/2000ml at 0.58ml/s with 14.1psi vacuum @18:23:41.74 Cartridge 57 sampling 947/2000ml at 0.57ml/s with 12.1psi vacuum @18:24:42.00 Cartridge 57 sampling 980/2000ml at 0.53ml/s with 15.5psi vacuum @18:25:41.74 Cartridge 57 sampling 1014/2000ml at 0.56ml/s with 12.6psi vacuum @18:26:41.95 Cartridge 57 sampling 1048/2000ml at 0.56ml/s with 11.1psi vacuum @18:27:41.74 Cartridge 57 sampling 1081/2000ml at 0.56ml/s with 14.0psi vacuum @18:28:41.98 Cartridge 57 sampling 1115/2000ml at 0.56ml/s with 13.7psi vacuum @18:29:41.74 Cartridge 57 sampling 1148/2000ml at 0.56ml/s with 11.3psi vacuum @18:30:42.03 Cartridge 57 sampling 1182/2000ml at 0.58ml/s with 12.8psi vacuum @18:31:41.74 Cartridge 57 sampling 1216/2000ml at 0.56ml/s with 14.5psi vacuum @18:32:41.89 Cartridge 57 sampling 1249/2000ml at 0.56ml/s with 13.7psi vacuum @18:33:41.89 Cartridge 57 sampling 1282/2000ml at 0.52ml/s with 13.1psi vacuum @18:34:41.92 Cartridge 57 sampling 1314/2000ml at 0.53ml/s with 13.3psi vacuum @18:35:41.75 Cartridge 57 sampling 1345/2000ml at 0.53ml/s with 13.4psi vacuum @18:36:42.03 Cartridge 57 sampling 1376/2000ml at 0.51ml/s with 12.4psi vacuum @18:37:41.74 Cartridge 57 sampling 1408/2000ml at 0.54ml/s with 13.4psi vacuum @18:38:41.86 Cartridge 57 sampling 1439/2000ml at 0.52ml/s with 13.6psi vacuum @18:39:41.74 Cartridge 57 sampling 1471/2000ml at 0.54ml/s with 12.2psi vacuum @18:40:41.89 Cartridge 57 sampling 1504/2000ml at 0.53ml/s with 13.5psi vacuum @18:41:41.75 Cartridge 57 sampling 1535/2000ml at 0.52ml/s with 13.8psi vacuum @18:42:41.90 Cartridge 57 sampling 1567/2000ml at 0.51ml/s with 14.3psi vacuum @18:43:41.74 Cartridge 57 sampling 1599/2000ml at 0.54ml/s with 13.6psi vacuum @18:44:41.96 Cartridge 57 sampling 1633/2000ml at 0.56ml/s with 13.2psi vacuum @18:45:42.01 Cartridge 57 sampling 1667/2000ml at 0.55ml/s with 13.8psi vacuum @18:46:41.76 Cartridge 57 sampling 1700/2000ml at 0.53ml/s with 13.1psi vacuum @18:47:42.02 Cartridge 57 sampling 1732/2000ml at 0.52ml/s with 14.4psi vacuum @18:48:41.76 Cartridge 57 sampling 1764/2000ml at 0.54ml/s with 12.8psi vacuum @18:49:41.95 Cartridge 57 sampling 1798/2000ml at 0.56ml/s with 14.1psi vacuum @18:50:41.75 Cartridge 57 sampling 1831/2000ml at 0.56ml/s with 12.7psi vacuum @18:51:42.04 Cartridge 57 sampling 1865/2000ml at 0.55ml/s with 13.7psi vacuum @18:52:41.74 Cartridge 57 sampling 1898/2000ml at 0.53ml/s with 13.4psi vacuum @18:53:41.92 Cartridge 57 sampling 1929/2000ml at 0.53ml/s with 14.4psi vacuum @18:54:41.74 Cartridge 57 sampling 1960/2000ml at 0.52ml/s with 14.0psi vacuum @18:55:41.93 Cartridge 57 sampling 1992/2000ml at 0.52ml/s with 13.4psi vacuum @18:55:57.63 Pressure.coast @18:55:57.79 Intake.close @18:55:58.46 Exhaust.close @18:55:59.15 Toroid.reconfigure :hold @18:55:59.33 Toroid.seek :bypass @18:56:01.12 Toroid.reconfigure :default @18:56:01.36 Toroid.seek :clear @18:56:03.07 Sampled 2000.0ml @18:56:04.11 Plunge.seek :reagent @18:56:05.04 Sample loop closed. Ready to Deliver Reagents @18:56:05.33 Carousel.seek 57 @18:56:05.57 Plunge.seek :locked @18:56:06.67 Air.seek :clear @18:56:06.86 Lysis1.seek :clear @18:56:10.42 Archive.coast @18:56:10.51 Air.coast @18:56:10.64 Lysis1.coast @18:56:10.81 Lysis2.coast @18:56:11.05 Plunge.coast @18:56:11.26 SamplePump.coast @18:56:11.36 Toroid.coast @18:56:11.53 Carousel.coast @18:56:12.04 Gate.power :motor,:OFF @18:56:12.13 Gate.power :core,:OFF @18:56:12.16 Gate.power :analytic1,:OFF @18:56:12.19 Gate.power :analytic2,:OFF @18:56:12.25 Gate.power :analytic3,:OFF @18:56:12.31 Gate.power :main,:OFF @18:56:12.44 Safely stopped and ready to power off @18:56:12.65 delay 2:00 @18:58:13.01 Gate.power :main,:ON,limit to 5000mA @18:58:13.09 Gate.power :analytic1,:ON,limit to 5000mA @18:58:13.19 Gate.power :analytic2,:ON,limit to 5000mA @18:58:13.22 Gate.power :analytic3,:ON,limit to 5000mA @18:58:13.29 Powering core bus @18:58:15.59 Archive.configure :default @18:58:15.75 Air.configure :default @18:58:15.90 Lysis1.configure :default @18:58:16.04 Lysis2.configure :default @18:58:16.21 Plunge.configure :default @18:58:16.38 SamplePump.configure :default @18:58:16.55 ExtraPump.configure :default @18:58:16.72 Toroid.configure :default @18:58:16.88 Carousel.configure :default @18:58:17.04 Pressure.configure :default @18:58:17.13 All elf controllers running firmware version 4.61 @18:58:17.15 Gate.power :motor,:ON,limit to 5000mA @18:58:17.32 Archive.reconfigure :default @18:58:18.11 Archive.home.jog 0 @18:58:18.99 Plunge.reconfigure :default @18:58:19.11 Air.reconfigure :default @18:58:19.19 Toroid.reconfigure :default @18:58:19.31 Lysis2.reconfigure :default @18:58:19.43 Lysis1.reconfigure :default @18:58:20.20 Plunge.home.jog 0 @18:58:20.38 Air.home.jog 0 @18:58:20.47 Toroid.home.jog 0 @18:58:20.61 Lysis2.home.jog 0 @18:58:20.74 Lysis1.home.jog 0 @18:58:22.06 Archive.reconfigure :homing @18:58:22.57 Plunge.reconfigure :homing @18:58:22.67 Archive.jog 1362 @18:58:22.83 Lysis2.reconfigure :homing @18:58:22.90 Lysis1.reconfigure :homing @18:58:22.98 Air.reconfigure :homing @18:58:23.24 Toroid.reconfigure :homing @18:58:23.72 Plunge.jog 852 @18:58:23.78 Lysis2.jog 1363 @18:58:23.82 Lysis1.jog 1351 @18:58:23.86 Toroid.jog 632 @18:58:23.90 Air.jog 1368 @18:58:24.71 Archive.setRawPosition! 50 @18:58:24.78 Archive.seek :home @18:58:24.89 Plunge.setRawPosition! 50 @18:58:24.96 Plunge.seek :home @18:58:25.10 Archive.reconfigure :default @18:58:25.24 Lysis2.setRawPosition! 50 @18:58:25.28 Lysis1.setRawPosition! 50 @18:58:25.31 Air.setRawPosition! 50 @18:58:25.45 Lysis2.seek :home @18:58:25.49 Lysis1.seek :home @18:58:25.59 Air.seek :home @18:58:25.82 Plunge.reconfigure :default @18:58:25.96 Lysis2.reconfigure :default @18:58:25.99 Lysis1.reconfigure :default @18:58:26.16 Air.reconfigure :default @18:58:26.27 Toroid.setRawPosition! 0 @18:58:26.34 Toroid.seek :home @18:58:26.46 Toroid.reconfigure :default @18:58:26.65 Toroid.seek :clear @18:58:28.39 Carousel.reconfigure :default @18:58:28.77 Carousel.home.jog 0 @18:58:29.51 Carousel.seek 57 @18:58:30.02 Selecting Cartridge 56 @18:58:30.76 Carousel.seek 56 @18:58:31.72 Lysis1.seek :locked @18:58:35.54 Plunge.seek :locked @18:58:35.74 Air.seek :locked @18:58:40.35 Mounted archive Cartridge @18:58:41.46 Duration of filtering limited to 4:05:00 @18:58:42.10 Priming sample loop w/160ml, bypass w/10ml @18:58:42.18 SamplePump.setPosition! 0ml @18:58:42.34 Bypass.open @18:58:44.01 Intake.open @18:58:44.94 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @18:58:47.61 Vacuum tare is 18% of full scale @18:58:47.67 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @18:58:48.49 Pressure.reconfigure :default @18:58:48.66 Pressure.hold @18:58:49.25 SamplePump.reconfigure :prime @18:58:49.46 SamplePump.seek 10ml,1:21 @18:59:00.65 Pressure.coast @18:59:00.79 SamplePump.setPosition! 0ml @18:59:00.86 Exhaust.open @18:59:02.52 Bypass.close @18:59:04.09 Pressure.reconfigure :default @18:59:04.17 Pressure.hold @18:59:04.46 SamplePump.reconfigure :prime @18:59:04.55 SamplePump.seek 160ml,13:51 @19:01:56.19 Pressure.coast @19:01:56.63 Bypass.open @19:01:57.40 Intake.close @19:01:58.33 Toroid.reconfigure :hold @19:01:58.41 Toroid.seek :crack @19:02:00.03 Toroid.seek :cartridge @19:02:01.23 Toroid.reconfigure :default @19:02:01.43 Toroid.seek :clear @19:02:03.88 Pressurizing cartridge to 14psi above 14.5psia @19:02:03.90 SamplePump.setPosition! 0ml @19:02:04.02 Intake.open @19:02:04.90 Bypass.close @19:02:05.72 Exhaust.close @19:02:06.70 Pressure.hold @19:02:06.94 SamplePump.reconfigure :charge @19:02:07.03 SamplePump.seek 2.3ml,42 seconds @19:02:15.32 Pressure.coast @19:02:15.40 Intake.close @19:02:16.28 Pressurized cartridge with 0.304ml to 14.0psi @19:02:16.30 Isolated @25.5psia (11.1psi) @19:02:16.32 Checking for cartridge leaks from 25.5psia @19:02:16.36 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @19:02:21.02 Cartridge pressure fell 2% to 24.9psia @19:02:21.29 Sampling 2000.0ml @19:02:21.31 SamplePump.setPosition! 0ml @19:02:21.55 Exhaust.open @19:02:22.37 Intake.open @19:02:23.73 Pressure.reconfigure :default @19:02:23.83 Pressure.hold @19:02:24.48 SamplePump.reconfigure :sample @19:02:24.78 SamplePump.seek 2000ml,2:47:12 @19:02:26.62 Cartridge 56 sampling 0/2000ml at 0.55ml/s with 13.1psi vacuum @19:03:26.83 Cartridge 56 sampling 31/2000ml at 0.51ml/s with 13.4psi vacuum @19:04:26.62 Cartridge 56 sampling 61/2000ml at 0.50ml/s with 14.7psi vacuum @19:05:26.78 Cartridge 56 sampling 92/2000ml at 0.51ml/s with 13.6psi vacuum @19:06:26.63 Cartridge 56 sampling 122/2000ml at 0.50ml/s with 14.1psi vacuum @19:07:26.83 Cartridge 56 sampling 152/2000ml at 0.50ml/s with 14.7psi vacuum @19:08:26.62 Cartridge 56 sampling 182/2000ml at 0.52ml/s with 12.9psi vacuum @19:09:26.76 Cartridge 56 sampling 214/2000ml at 0.52ml/s with 15.1psi vacuum @19:10:26.62 Cartridge 56 sampling 245/2000ml at 0.51ml/s with 13.9psi vacuum @19:11:26.78 Cartridge 56 sampling 276/2000ml at 0.52ml/s with 12.0psi vacuum @19:12:26.63 Cartridge 56 sampling 306/2000ml at 0.51ml/s with 12.7psi vacuum @19:13:26.84 Cartridge 56 sampling 336/2000ml at 0.50ml/s with 12.7psi vacuum @19:14:26.62 Cartridge 56 sampling 367/2000ml at 0.50ml/s with 13.6psi vacuum @19:15:26.69 Cartridge 56 sampling 397/2000ml at 0.50ml/s with 14.6psi vacuum @19:16:26.78 Cartridge 56 sampling 428/2000ml at 0.52ml/s with 14.0psi vacuum @19:17:26.62 Cartridge 56 sampling 459/2000ml at 0.52ml/s with 13.4psi vacuum @19:18:26.85 Cartridge 56 sampling 490/2000ml at 0.52ml/s with 13.2psi vacuum @19:19:26.62 Cartridge 56 sampling 521/2000ml at 0.50ml/s with 12.4psi vacuum @19:20:26.86 Cartridge 56 sampling 552/2000ml at 0.52ml/s with 13.5psi vacuum @19:21:26.62 Cartridge 56 sampling 583/2000ml at 0.53ml/s with 12.9psi vacuum @19:22:26.85 Cartridge 56 sampling 615/2000ml at 0.52ml/s with 12.8psi vacuum @19:23:26.62 Cartridge 56 sampling 646/2000ml at 0.53ml/s with 12.4psi vacuum @19:24:26.75 Cartridge 56 sampling 678/2000ml at 0.53ml/s with 14.0psi vacuum @19:25:26.62 Cartridge 56 sampling 710/2000ml at 0.52ml/s with 13.6psi vacuum @19:26:26.78 Cartridge 56 sampling 741/2000ml at 0.51ml/s with 13.9psi vacuum @19:27:26.63 Cartridge 56 sampling 772/2000ml at 0.51ml/s with 12.4psi vacuum @19:28:27.93 Cartridge 56 sampling 802/2000ml at 0.49ml/s with 13.2psi vacuum @19:29:26.62 Cartridge 56 sampling 833/2000ml at 0.53ml/s with 13.3psi vacuum @19:30:26.77 Cartridge 56 sampling 865/2000ml at 0.51ml/s with 11.8psi vacuum @19:31:26.63 Cartridge 56 sampling 896/2000ml at 0.52ml/s with 14.6psi vacuum @19:32:26.82 Cartridge 56 sampling 927/2000ml at 0.51ml/s with 13.6psi vacuum @19:33:26.63 Cartridge 56 sampling 957/2000ml at 0.51ml/s with 12.8psi vacuum @19:34:26.78 Cartridge 56 sampling 988/2000ml at 0.51ml/s with 14.4psi vacuum @19:35:26.62 Cartridge 56 sampling 1019/2000ml at 0.51ml/s with 14.4psi vacuum @19:36:26.65 Cartridge 56 sampling 1050/2000ml at 0.51ml/s with 14.3psi vacuum @19:37:26.62 Cartridge 56 sampling 1081/2000ml at 0.50ml/s with 14.3psi vacuum @19:38:26.87 Cartridge 56 sampling 1112/2000ml at 0.52ml/s with 14.1psi vacuum @19:39:26.80 Cartridge 56 sampling 1143/2000ml at 0.51ml/s with 14.6psi vacuum @19:40:26.63 Cartridge 56 sampling 1174/2000ml at 0.52ml/s with 12.8psi vacuum @19:41:27.00 Cartridge 56 sampling 1205/2000ml at 0.53ml/s with 13.6psi vacuum @19:42:26.63 Cartridge 56 sampling 1236/2000ml at 0.50ml/s with 14.2psi vacuum @19:43:26.86 Cartridge 56 sampling 1267/2000ml at 0.51ml/s with 14.7psi vacuum @19:44:26.62 Cartridge 56 sampling 1298/2000ml at 0.51ml/s with 13.4psi vacuum @19:45:26.82 Cartridge 56 sampling 1328/2000ml at 0.50ml/s with 12.9psi vacuum @19:46:26.63 Cartridge 56 sampling 1359/2000ml at 0.51ml/s with 13.2psi vacuum @19:47:26.93 Cartridge 56 sampling 1389/2000ml at 0.53ml/s with 13.4psi vacuum @19:48:26.62 Cartridge 56 sampling 1420/2000ml at 0.51ml/s with 14.1psi vacuum @19:49:26.87 Cartridge 56 sampling 1450/2000ml at 0.49ml/s with 12.1psi vacuum @19:50:26.62 Cartridge 56 sampling 1481/2000ml at 0.52ml/s with 13.3psi vacuum @19:51:26.76 Cartridge 56 sampling 1512/2000ml at 0.52ml/s with 14.4psi vacuum @19:52:26.63 Cartridge 56 sampling 1543/2000ml at 0.51ml/s with 13.5psi vacuum @19:53:26.78 Cartridge 56 sampling 1574/2000ml at 0.53ml/s with 13.6psi vacuum @19:54:26.62 Cartridge 56 sampling 1605/2000ml at 0.50ml/s with 14.3psi vacuum @19:55:26.81 Cartridge 56 sampling 1636/2000ml at 0.52ml/s with 12.9psi vacuum @19:56:26.63 Cartridge 56 sampling 1667/2000ml at 0.51ml/s with 14.7psi vacuum @19:57:26.67 Cartridge 56 sampling 1698/2000ml at 0.51ml/s with 13.1psi vacuum @19:58:26.62 Cartridge 56 sampling 1729/2000ml at 0.52ml/s with 13.6psi vacuum @19:59:26.74 Cartridge 56 sampling 1760/2000ml at 0.52ml/s with 13.5psi vacuum @20:00:26.62 Cartridge 56 sampling 1791/2000ml at 0.51ml/s with 14.3psi vacuum @20:01:26.62 Cartridge 56 sampling 1822/2000ml at 0.52ml/s with 14.4psi vacuum @20:02:26.87 Cartridge 56 sampling 1852/2000ml at 0.51ml/s with 13.1psi vacuum @20:03:26.62 Cartridge 56 sampling 1884/2000ml at 0.52ml/s with 14.5psi vacuum @20:04:27.08 Cartridge 56 sampling 1915/2000ml at 0.53ml/s with 13.0psi vacuum @20:05:26.62 Cartridge 56 sampling 1946/2000ml at 0.50ml/s with 13.7psi vacuum @20:06:26.86 Cartridge 56 sampling 1976/2000ml at 0.51ml/s with 13.8psi vacuum @20:07:12.55 Pressure.coast @20:07:12.88 Intake.close @20:07:13.00 Exhaust.close @20:07:13.48 Toroid.reconfigure :hold @20:07:13.69 Toroid.seek :bypass @20:07:15.35 Toroid.reconfigure :default @20:07:15.52 Toroid.seek :clear @20:07:17.22 Sampled 2000.0ml @20:07:17.74 Plunge.seek :reagent @20:07:18.66 Sample loop closed. Ready to Deliver Reagents @20:07:18.93 Carousel.seek 56 @20:07:19.17 Plunge.seek :locked @20:07:20.26 Air.seek :clear @20:07:20.45 Lysis1.seek :clear @20:07:24.10 Archive.coast @20:07:24.20 Air.coast @20:07:24.41 Lysis1.coast @20:07:24.52 Lysis2.coast @20:07:24.71 Plunge.coast @20:07:24.78 SamplePump.coast @20:07:24.98 Toroid.coast @20:07:25.08 Carousel.coast @20:07:25.47 Gate.power :motor,:OFF @20:07:25.54 Gate.power :core,:OFF @20:07:25.59 Gate.power :analytic1,:OFF @20:07:25.66 Gate.power :analytic2,:OFF @20:07:25.78 Gate.power :analytic3,:OFF @20:07:25.86 Gate.power :main,:OFF @20:07:26.01 Safely stopped and ready to power off @20:07:26.37 delay 2:00 @20:09:26.52 Gate.power :main,:ON,limit to 5000mA @20:09:26.54 Gate.power :analytic1,:ON,limit to 5000mA @20:09:26.57 Gate.power :analytic2,:ON,limit to 5000mA @20:09:26.60 Gate.power :analytic3,:ON,limit to 5000mA @20:09:26.67 Powering core bus @20:09:28.97 Archive.configure :default @20:09:29.14 Air.configure :default @20:09:29.28 Lysis1.configure :default @20:09:29.43 Lysis2.configure :default @20:09:29.59 Plunge.configure :default @20:09:29.75 SamplePump.configure :default @20:09:29.92 ExtraPump.configure :default @20:09:30.18 Toroid.configure :default @20:09:30.89 Carousel.configure :default @20:09:31.54 Pressure.configure :default @20:09:31.82 All elf controllers running firmware version 4.61 @20:09:31.85 Gate.power :motor,:ON,limit to 5000mA @20:09:32.15 Archive.reconfigure :default @20:09:32.94 Archive.home.jog 0 @20:09:33.73 Lysis2.reconfigure :default @20:09:33.76 Lysis1.reconfigure :default @20:09:35.44 Toroid.reconfigure :default @20:09:35.78 Plunge.reconfigure :default @20:09:36.01 Air.reconfigure :default @20:09:36.11 Lysis2.home.jog 0 @20:09:36.15 Lysis1.home.jog 0 @20:09:36.27 Plunge.home.jog 0 @20:09:36.32 Toroid.home.jog 0 @20:09:36.43 Air.home.jog 0 @20:09:36.87 Archive.reconfigure :homing @20:09:37.21 Archive.jog 1360 @20:09:37.33 Toroid.reconfigure :homing @20:09:37.43 Lysis2.reconfigure :homing @20:09:37.51 Lysis1.reconfigure :homing @20:09:37.57 Plunge.reconfigure :homing @20:09:37.62 Air.reconfigure :homing @20:09:37.90 Toroid.jog 612 @20:09:37.93 Lysis2.jog 1362 @20:09:37.96 Lysis1.jog 1352 @20:09:38.01 Plunge.jog 850 @20:09:38.05 Air.jog 1364 @20:09:38.81 Archive.setRawPosition! 50 @20:09:38.88 Archive.seek :home @20:09:39.12 Plunge.setRawPosition! 50 @20:09:39.19 Plunge.seek :home @20:09:39.34 Lysis2.setRawPosition! 50 @20:09:39.38 Lysis1.setRawPosition! 50 @20:09:39.42 Archive.reconfigure :default @20:09:39.56 Lysis2.seek :home @20:09:39.63 Lysis1.seek :home @20:09:39.81 Air.setRawPosition! 50 @20:09:39.84 Plunge.reconfigure :default @20:09:39.94 Air.seek :home @20:09:40.12 Lysis2.reconfigure :default @20:09:40.16 Lysis1.reconfigure :default @20:09:40.97 Toroid.setRawPosition! 0 @20:09:41.04 Air.reconfigure :default @20:09:41.55 Toroid.seek :home @20:09:41.93 Toroid.reconfigure :default @20:09:42.34 Toroid.seek :clear @20:09:44.18 Carousel.reconfigure :default @20:09:44.33 Carousel.home.jog 0 @20:09:45.11 Carousel.seek 56 @20:09:45.49 Selecting Cartridge 55 @20:09:46.13 Carousel.seek 55 @20:09:46.68 Lysis1.seek :locked @20:09:49.94 Plunge.seek :locked @20:09:50.11 Air.seek :locked @20:09:54.42 Mounted archive Cartridge @20:09:55.28 Duration of filtering limited to 4:05:00 @20:09:55.60 Priming sample loop w/160ml, bypass w/10ml @20:09:55.62 SamplePump.setPosition! 0ml @20:09:55.70 Bypass.open @20:09:57.23 Intake.open @20:09:58.15 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @20:10:00.80 Vacuum tare is 18% of full scale @20:10:00.85 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @20:10:01.53 Pressure.reconfigure :default @20:10:01.69 Pressure.hold @20:10:02.39 SamplePump.reconfigure :prime @20:10:02.59 SamplePump.seek 10ml,1:21 @20:10:13.81 Pressure.coast @20:10:14.03 SamplePump.setPosition! 0ml @20:10:14.15 Exhaust.open @20:10:15.60 Bypass.close @20:10:17.17 Pressure.reconfigure :default @20:10:17.24 Pressure.hold @20:10:17.54 SamplePump.reconfigure :prime @20:10:17.63 SamplePump.seek 160ml,13:51 @20:13:09.20 Pressure.coast @20:13:09.28 Bypass.open @20:13:10.02 Intake.close @20:13:10.92 Toroid.reconfigure :hold @20:13:11.00 Toroid.seek :crack @20:13:12.61 Toroid.seek :cartridge @20:13:13.66 Toroid.reconfigure :default @20:13:14.04 Toroid.seek :clear @20:13:16.39 Pressurizing cartridge to 14psi above 14.5psia @20:13:16.40 SamplePump.setPosition! 0ml @20:13:16.63 Intake.open @20:13:17.52 Bypass.close @20:13:18.37 Exhaust.close @20:13:19.22 Pressure.reconfigure :default @20:13:19.30 Pressure.hold @20:13:19.54 SamplePump.reconfigure :charge @20:13:19.63 SamplePump.seek 2.3ml,42 seconds @20:13:28.18 Pressure.coast @20:13:28.26 Intake.close @20:13:29.69 Pressurized cartridge with 0.331ml to 14.0psi @20:13:29.71 Isolated @25.5psia (11.0psi) @20:13:29.73 Checking for cartridge leaks from 25.5psia @20:13:29.77 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @20:13:34.49 Cartridge pressure fell 2% to 25.0psia @20:13:34.74 Sampling 2000.0ml @20:13:34.78 SamplePump.setPosition! 0ml @20:13:34.90 Exhaust.open @20:13:35.72 Intake.open @20:13:36.98 Pressure.reconfigure :default @20:13:37.06 Pressure.hold @20:13:37.38 SamplePump.reconfigure :sample @20:13:37.47 SamplePump.seek 2000ml,2:47:12 @20:13:40.03 Cartridge 55 sampling 1/2000ml at 0.52ml/s with 13.2psi vacuum @20:14:40.12 Cartridge 55 sampling 32/2000ml at 0.51ml/s with 12.8psi vacuum @20:15:40.04 Cartridge 55 sampling 62/2000ml at 0.50ml/s with 14.4psi vacuum @20:16:40.06 Cartridge 55 sampling 93/2000ml at 0.51ml/s with 14.5psi vacuum @20:17:40.03 Cartridge 55 sampling 123/2000ml at 0.51ml/s with 11.9psi vacuum @20:18:40.28 Cartridge 55 sampling 154/2000ml at 0.51ml/s with 14.9psi vacuum @20:19:40.07 Cartridge 55 sampling 186/2000ml at 0.52ml/s with 14.4psi vacuum @20:20:40.14 Cartridge 55 sampling 217/2000ml at 0.51ml/s with 14.8psi vacuum @20:21:40.30 Cartridge 55 sampling 248/2000ml at 0.51ml/s with 11.1psi vacuum @20:22:40.03 Cartridge 55 sampling 278/2000ml at 0.50ml/s with 14.1psi vacuum @20:23:40.32 Cartridge 55 sampling 309/2000ml at 0.52ml/s with 14.5psi vacuum @20:24:40.03 Cartridge 55 sampling 340/2000ml at 0.52ml/s with 13.1psi vacuum @20:25:40.22 Cartridge 55 sampling 371/2000ml at 0.52ml/s with 14.4psi vacuum @20:26:40.03 Cartridge 55 sampling 402/2000ml at 0.51ml/s with 12.0psi vacuum @20:27:40.26 Cartridge 55 sampling 433/2000ml at 0.52ml/s with 13.8psi vacuum @20:28:40.03 Cartridge 55 sampling 464/2000ml at 0.51ml/s with 13.5psi vacuum @20:29:40.21 Cartridge 55 sampling 495/2000ml at 0.52ml/s with 14.2psi vacuum @20:30:40.03 Cartridge 55 sampling 526/2000ml at 0.52ml/s with 13.2psi vacuum @20:31:40.26 Cartridge 55 sampling 558/2000ml at 0.52ml/s with 13.4psi vacuum @20:32:40.04 Cartridge 55 sampling 589/2000ml at 0.52ml/s with 14.3psi vacuum @20:33:40.37 Cartridge 55 sampling 620/2000ml at 0.53ml/s with 13.4psi vacuum @20:34:40.03 Cartridge 55 sampling 652/2000ml at 0.52ml/s with 13.6psi vacuum @20:35:40.23 Cartridge 55 sampling 684/2000ml at 0.54ml/s with 12.9psi vacuum @20:36:40.08 Cartridge 55 sampling 716/2000ml at 0.52ml/s with 13.6psi vacuum @20:37:40.24 Cartridge 55 sampling 747/2000ml at 0.53ml/s with 13.2psi vacuum @20:38:40.04 Cartridge 55 sampling 779/2000ml at 0.51ml/s with 13.4psi vacuum @20:39:40.10 Cartridge 55 sampling 810/2000ml at 0.50ml/s with 13.7psi vacuum @20:40:40.03 Cartridge 55 sampling 842/2000ml at 0.53ml/s with 13.2psi vacuum @20:41:40.08 Cartridge 55 sampling 873/2000ml at 0.53ml/s with 13.4psi vacuum @20:42:40.04 Cartridge 55 sampling 904/2000ml at 0.51ml/s with 12.6psi vacuum @20:43:40.30 Cartridge 55 sampling 935/2000ml at 0.51ml/s with 12.6psi vacuum @20:44:40.36 Cartridge 55 sampling 966/2000ml at 0.51ml/s with 13.1psi vacuum @20:45:40.13 Cartridge 55 sampling 997/2000ml at 0.51ml/s with 12.7psi vacuum @20:46:40.33 Cartridge 55 sampling 1028/2000ml at 0.52ml/s with 12.6psi vacuum @20:47:40.03 Cartridge 55 sampling 1060/2000ml at 0.53ml/s with 13.7psi vacuum @20:48:40.36 Cartridge 55 sampling 1092/2000ml at 0.52ml/s with 13.9psi vacuum @20:49:40.03 Cartridge 55 sampling 1123/2000ml at 0.52ml/s with 12.7psi vacuum @20:50:40.17 Cartridge 55 sampling 1154/2000ml at 0.52ml/s with 13.3psi vacuum @20:51:40.03 Cartridge 55 sampling 1186/2000ml at 0.52ml/s with 12.5psi vacuum @20:52:40.23 Cartridge 55 sampling 1218/2000ml at 0.54ml/s with 13.7psi vacuum @20:53:40.03 Cartridge 55 sampling 1250/2000ml at 0.54ml/s with 13.0psi vacuum @20:54:40.26 Cartridge 55 sampling 1282/2000ml at 0.52ml/s with 13.0psi vacuum @20:55:40.04 Cartridge 55 sampling 1313/2000ml at 0.54ml/s with 12.9psi vacuum @20:56:40.25 Cartridge 55 sampling 1345/2000ml at 0.54ml/s with 13.0psi vacuum @20:57:40.04 Cartridge 55 sampling 1376/2000ml at 0.51ml/s with 12.3psi vacuum @20:58:40.19 Cartridge 55 sampling 1407/2000ml at 0.51ml/s with 12.8psi vacuum @20:59:40.03 Cartridge 55 sampling 1438/2000ml at 0.52ml/s with 13.2psi vacuum @21:00:40.15 Cartridge 55 sampling 1468/2000ml at 0.52ml/s with 14.0psi vacuum @21:01:40.03 Cartridge 55 sampling 1500/2000ml at 0.52ml/s with 13.5psi vacuum @21:02:40.17 Cartridge 55 sampling 1532/2000ml at 0.53ml/s with 14.1psi vacuum @21:03:40.03 Cartridge 55 sampling 1563/2000ml at 0.53ml/s with 12.9psi vacuum @21:04:40.12 Cartridge 55 sampling 1595/2000ml at 0.53ml/s with 12.7psi vacuum @21:05:40.03 Cartridge 55 sampling 1627/2000ml at 0.52ml/s with 14.2psi vacuum @21:06:40.28 Cartridge 55 sampling 1658/2000ml at 0.53ml/s with 13.7psi vacuum @21:07:40.23 Cartridge 55 sampling 1689/2000ml at 0.53ml/s with 13.2psi vacuum @21:08:40.03 Cartridge 55 sampling 1721/2000ml at 0.51ml/s with 13.7psi vacuum @21:09:40.49 Cartridge 55 sampling 1752/2000ml at 0.54ml/s with 13.9psi vacuum @21:10:40.03 Cartridge 55 sampling 1785/2000ml at 0.54ml/s with 13.8psi vacuum @21:11:40.28 Cartridge 55 sampling 1817/2000ml at 0.54ml/s with 14.7psi vacuum @21:12:40.03 Cartridge 55 sampling 1848/2000ml at 0.52ml/s with 12.9psi vacuum @21:13:40.19 Cartridge 55 sampling 1880/2000ml at 0.53ml/s with 13.7psi vacuum @21:14:40.04 Cartridge 55 sampling 1912/2000ml at 0.53ml/s with 12.5psi vacuum @21:15:40.15 Cartridge 55 sampling 1943/2000ml at 0.51ml/s with 14.2psi vacuum @21:16:40.04 Cartridge 55 sampling 1974/2000ml at 0.53ml/s with 14.4psi vacuum @21:17:28.44 Pressure.coast @21:17:28.89 Intake.close @21:17:29.26 Exhaust.close @21:17:29.80 Toroid.reconfigure :hold @21:17:30.09 Toroid.seek :bypass @21:17:31.78 Toroid.reconfigure :default @21:17:32.03 Toroid.seek :clear @21:17:33.82 Sampled 2000.0ml @21:17:35.02 Plunge.seek :reagent @21:17:35.94 Sample loop closed. Ready to Deliver Reagents @21:17:36.20 Carousel.seek 55 @21:17:36.45 Plunge.seek :locked @21:17:37.54 Air.seek :clear @21:17:37.72 Lysis1.seek :clear @21:17:41.30 Archive.coast @21:17:41.40 Air.coast @21:17:41.61 Lysis1.coast @21:17:41.84 Lysis2.coast @21:17:41.98 Plunge.coast @21:17:42.16 SamplePump.coast @21:17:42.23 Toroid.coast @21:17:42.42 Carousel.coast @21:17:42.76 Gate.power :motor,:OFF @21:17:42.83 Gate.power :core,:OFF @21:17:42.86 Gate.power :analytic1,:OFF @21:17:42.89 Gate.power :analytic2,:OFF @21:17:42.92 Gate.power :analytic3,:OFF @21:17:43.03 Gate.power :main,:OFF @21:17:43.11 Safely stopped and ready to power off @21:17:43.44 delay 2:00 @21:19:43.60 Gate.power :main,:ON,limit to 5000mA @21:19:43.62 Gate.power :analytic1,:ON,limit to 5000mA @21:19:43.66 Gate.power :analytic2,:ON,limit to 5000mA @21:19:43.68 Gate.power :analytic3,:ON,limit to 5000mA @21:19:43.76 Powering core bus @21:19:46.06 Archive.configure :default @21:19:46.23 Air.configure :default @21:19:46.37 Lysis1.configure :default @21:19:46.52 Lysis2.configure :default @21:19:46.68 Plunge.configure :default @21:19:46.85 SamplePump.configure :default @21:19:47.19 ExtraPump.configure :default @21:19:47.60 Toroid.configure :default @21:19:48.19 Carousel.configure :default @21:19:48.74 Pressure.configure :default @21:19:48.98 All elf controllers running firmware version 4.61 @21:19:49.02 Gate.power :motor,:ON,limit to 5000mA @21:19:49.73 Archive.reconfigure :default @21:19:50.51 Archive.home.jog 0 @21:19:51.03 Lysis1.reconfigure :default @21:19:51.16 Plunge.reconfigure :default @21:19:51.29 Lysis2.reconfigure :default @21:19:51.56 Toroid.reconfigure :default @21:19:51.70 Air.reconfigure :default @21:19:52.19 Lysis1.home.jog 0 @21:19:52.33 Lysis2.home.jog 0 @21:19:52.45 Toroid.home.jog 0 @21:19:52.49 Air.home.jog 0 @21:19:52.53 Plunge.home.jog 0 @21:19:53.58 Archive.reconfigure :homing @21:19:53.88 Archive.jog 1363 @21:19:53.94 Plunge.reconfigure :homing @21:19:53.99 Lysis2.reconfigure :homing @21:19:54.03 Air.reconfigure :homing @21:19:54.08 Lysis1.reconfigure :homing @21:19:54.14 Toroid.reconfigure :homing @21:19:54.40 Lysis2.jog 1363 @21:19:54.46 Plunge.jog 852 @21:19:54.49 Air.jog 1368 @21:19:54.54 Lysis1.jog 1350 @21:19:54.57 Toroid.jog 616 @21:19:55.34 Archive.setRawPosition! 50 @21:19:55.41 Archive.seek :home @21:19:55.60 Plunge.setRawPosition! 50 @21:19:55.67 Plunge.seek :home @21:19:55.81 Archive.reconfigure :default @21:19:55.84 Lysis2.setRawPosition! 50 @21:19:55.99 Lysis2.seek :home @21:19:56.11 Lysis1.setRawPosition! 50 @21:19:56.14 Air.setRawPosition! 50 @21:19:56.18 Plunge.reconfigure :default @21:19:56.35 Lysis1.seek :home @21:19:56.38 Air.seek :home @21:19:56.59 Lysis2.reconfigure :default @21:19:56.73 Lysis1.reconfigure :default @21:19:56.86 Air.reconfigure :default @21:19:56.97 Toroid.setRawPosition! 0 @21:19:57.04 Toroid.seek :home @21:19:57.16 Toroid.reconfigure :default @21:19:57.34 Toroid.seek :clear @21:19:59.02 Carousel.reconfigure :default @21:19:59.23 Carousel.home.jog 0 @21:19:59.97 Carousel.seek 55 @21:20:00.40 Selecting Cartridge 54 @21:20:00.88 Carousel.seek 54 @21:20:01.71 Lysis1.seek :locked @21:20:05.03 Plunge.seek :locked @21:20:05.20 Air.seek :locked @21:20:09.59 Mounted archive Cartridge @21:20:10.43 Duration of filtering limited to 4:05:00 @21:20:12.33 Priming sample loop w/160ml, bypass w/10ml @21:20:12.47 SamplePump.setPosition! 0ml @21:20:12.74 Bypass.open @21:20:14.27 Intake.open @21:20:15.17 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @21:20:17.90 Vacuum tare is 18% of full scale @21:20:18.15 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @21:20:18.94 Pressure.reconfigure :default @21:20:19.18 Pressure.hold @21:20:19.75 SamplePump.reconfigure :prime @21:20:19.96 SamplePump.seek 10ml,1:21 @21:20:31.30 Pressure.coast @21:20:31.60 SamplePump.setPosition! 0ml @21:20:31.67 Exhaust.open @21:20:33.10 Bypass.close @21:20:34.68 Pressure.reconfigure :default @21:20:34.76 Pressure.hold @21:20:35.06 SamplePump.reconfigure :prime @21:20:35.15 SamplePump.seek 160ml,13:51 @21:23:26.73 Pressure.coast @21:23:26.80 Bypass.open @21:23:27.53 Intake.close @21:23:28.44 Toroid.reconfigure :hold @21:23:28.52 Toroid.seek :crack @21:23:30.13 Toroid.seek :cartridge @21:23:31.19 Toroid.reconfigure :default @21:23:31.58 Toroid.seek :clear @21:23:33.97 Pressurizing cartridge to 14psi above 14.6psia @21:23:34.05 SamplePump.setPosition! 0ml @21:23:34.16 Intake.open @21:23:35.09 Bypass.close @21:23:35.97 Exhaust.close @21:23:37.00 Pressure.reconfigure :default @21:23:37.09 Pressure.hold @21:23:37.32 SamplePump.reconfigure :charge @21:23:37.41 SamplePump.seek 2.3ml,42 seconds @21:23:45.47 Pressure.coast @21:23:45.54 Intake.close @21:23:46.42 Pressurized cartridge with 0.283ml to 14.0psi @21:23:46.44 Isolated @24.9psia (10.2psi) @21:23:46.47 Checking for cartridge leaks from 24.9psia @21:23:46.51 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @21:23:51.20 Cartridge pressure fell 6% to 23.3psia @21:23:51.48 Sampling 2000.0ml @21:23:51.51 SamplePump.setPosition! 0ml @21:23:51.64 Exhaust.open @21:23:52.47 Intake.open @21:23:53.76 Pressure.reconfigure :default @21:23:53.92 Pressure.hold @21:23:54.59 SamplePump.reconfigure :sample @21:23:54.84 SamplePump.seek 2000ml,2:47:12 @21:23:56.75 Cartridge 54 sampling 1/2000ml at 0.51ml/s with 13.1psi vacuum @21:24:57.07 Cartridge 54 sampling 31/2000ml at 0.51ml/s with 10.9psi vacuum @21:25:56.75 Cartridge 54 sampling 61/2000ml at 0.50ml/s with 12.8psi vacuum @21:26:56.93 Cartridge 54 sampling 91/2000ml at 0.50ml/s with 12.7psi vacuum @21:27:56.76 Cartridge 54 sampling 121/2000ml at 0.52ml/s with 12.9psi vacuum @21:28:56.91 Cartridge 54 sampling 150/2000ml at 0.49ml/s with 14.4psi vacuum @21:29:56.76 Cartridge 54 sampling 180/2000ml at 0.50ml/s with 13.2psi vacuum @21:30:57.04 Cartridge 54 sampling 211/2000ml at 0.49ml/s with 12.8psi vacuum @21:31:56.76 Cartridge 54 sampling 241/2000ml at 0.51ml/s with 13.8psi vacuum @21:32:56.93 Cartridge 54 sampling 271/2000ml at 0.51ml/s with 13.0psi vacuum @21:33:56.77 Cartridge 54 sampling 301/2000ml at 0.51ml/s with 13.2psi vacuum @21:34:56.93 Cartridge 54 sampling 333/2000ml at 0.54ml/s with 12.4psi vacuum @21:35:56.76 Cartridge 54 sampling 365/2000ml at 0.55ml/s with 12.8psi vacuum @21:36:57.09 Cartridge 54 sampling 398/2000ml at 0.53ml/s with 13.0psi vacuum @21:37:56.76 Cartridge 54 sampling 430/2000ml at 0.54ml/s with 13.0psi vacuum @21:38:57.20 Cartridge 54 sampling 463/2000ml at 0.55ml/s with 12.5psi vacuum @21:39:56.95 Cartridge 54 sampling 496/2000ml at 0.54ml/s with 13.8psi vacuum @21:40:56.82 Cartridge 54 sampling 528/2000ml at 0.54ml/s with 13.5psi vacuum @21:41:56.96 Cartridge 54 sampling 561/2000ml at 0.54ml/s with 12.9psi vacuum @21:42:56.76 Cartridge 54 sampling 594/2000ml at 0.54ml/s with 11.7psi vacuum @21:43:57.09 Cartridge 54 sampling 627/2000ml at 0.54ml/s with 12.1psi vacuum @21:44:56.76 Cartridge 54 sampling 660/2000ml at 0.54ml/s with 13.6psi vacuum @21:45:57.24 Cartridge 54 sampling 693/2000ml at 0.54ml/s with 13.0psi vacuum @21:46:56.77 Cartridge 54 sampling 726/2000ml at 0.55ml/s with 14.3psi vacuum @21:47:56.97 Cartridge 54 sampling 759/2000ml at 0.55ml/s with 13.2psi vacuum @21:48:56.77 Cartridge 54 sampling 792/2000ml at 0.55ml/s with 13.7psi vacuum @21:49:56.97 Cartridge 54 sampling 826/2000ml at 0.56ml/s with 13.1psi vacuum @21:50:56.76 Cartridge 54 sampling 859/2000ml at 0.55ml/s with 12.4psi vacuum @21:51:56.99 Cartridge 54 sampling 892/2000ml at 0.56ml/s with 12.4psi vacuum @21:52:56.76 Cartridge 54 sampling 926/2000ml at 0.57ml/s with 13.7psi vacuum @21:53:57.05 Cartridge 54 sampling 960/2000ml at 0.56ml/s with 12.2psi vacuum @21:54:56.76 Cartridge 54 sampling 993/2000ml at 0.56ml/s with 11.9psi vacuum @21:55:56.96 Cartridge 54 sampling 1027/2000ml at 0.56ml/s with 12.2psi vacuum @21:56:56.78 Cartridge 54 sampling 1060/2000ml at 0.57ml/s with 12.5psi vacuum @21:57:56.96 Cartridge 54 sampling 1094/2000ml at 0.55ml/s with 12.4psi vacuum @21:58:56.77 Cartridge 54 sampling 1128/2000ml at 0.55ml/s with 14.5psi vacuum @21:59:56.83 Cartridge 54 sampling 1162/2000ml at 0.55ml/s with 14.4psi vacuum @22:00:56.77 Cartridge 54 sampling 1195/2000ml at 0.53ml/s with 14.7psi vacuum @22:01:57.01 Cartridge 54 sampling 1228/2000ml at 0.56ml/s with 11.6psi vacuum @22:02:56.77 Cartridge 54 sampling 1261/2000ml at 0.53ml/s with 12.3psi vacuum @22:03:56.86 Cartridge 54 sampling 1294/2000ml at 0.54ml/s with 14.0psi vacuum @22:04:56.77 Cartridge 54 sampling 1327/2000ml at 0.56ml/s with 13.3psi vacuum @22:05:56.97 Cartridge 54 sampling 1360/2000ml at 0.55ml/s with 13.7psi vacuum @22:06:57.01 Cartridge 54 sampling 1393/2000ml at 0.55ml/s with 14.3psi vacuum @22:07:56.77 Cartridge 54 sampling 1426/2000ml at 0.55ml/s with 13.5psi vacuum @22:08:57.15 Cartridge 54 sampling 1459/2000ml at 0.53ml/s with 12.5psi vacuum @22:09:56.76 Cartridge 54 sampling 1491/2000ml at 0.54ml/s with 11.8psi vacuum @22:10:57.02 Cartridge 54 sampling 1525/2000ml at 0.55ml/s with 12.6psi vacuum @22:11:56.77 Cartridge 54 sampling 1558/2000ml at 0.55ml/s with 13.1psi vacuum @22:12:56.93 Cartridge 54 sampling 1591/2000ml at 0.55ml/s with 14.2psi vacuum @22:13:56.76 Cartridge 54 sampling 1624/2000ml at 0.54ml/s with 14.4psi vacuum @22:14:56.89 Cartridge 54 sampling 1657/2000ml at 0.55ml/s with 11.8psi vacuum @22:15:56.76 Cartridge 54 sampling 1689/2000ml at 0.55ml/s with 14.9psi vacuum @22:16:56.96 Cartridge 54 sampling 1722/2000ml at 0.55ml/s with 11.8psi vacuum @22:17:56.76 Cartridge 54 sampling 1756/2000ml at 0.55ml/s with 13.2psi vacuum @22:18:56.95 Cartridge 54 sampling 1789/2000ml at 0.55ml/s with 12.9psi vacuum @22:19:56.77 Cartridge 54 sampling 1822/2000ml at 0.56ml/s with 13.2psi vacuum @22:20:56.96 Cartridge 54 sampling 1856/2000ml at 0.55ml/s with 12.3psi vacuum @22:21:56.76 Cartridge 54 sampling 1889/2000ml at 0.56ml/s with 13.0psi vacuum @22:22:57.02 Cartridge 54 sampling 1923/2000ml at 0.55ml/s with 12.1psi vacuum @22:23:56.76 Cartridge 54 sampling 1956/2000ml at 0.56ml/s with 14.0psi vacuum @22:24:56.88 Cartridge 54 sampling 1989/2000ml at 0.56ml/s with 12.4psi vacuum @22:25:16.14 Pressure.coast @22:25:16.29 Intake.close @22:25:16.39 Exhaust.close @22:25:16.56 Toroid.reconfigure :hold @22:25:16.64 Toroid.seek :bypass @22:25:18.29 Toroid.reconfigure :default @22:25:18.37 Toroid.seek :clear @22:25:20.00 Sampled 2000.0ml @22:25:20.51 Plunge.seek :reagent @22:25:21.58 Sample loop closed. Ready to Deliver Reagents @22:25:22.16 Carousel.seek 54 @22:25:22.66 Plunge.seek :locked @22:25:24.04 Air.seek :clear @22:25:24.51 Lysis1.seek :clear @22:25:27.94 Archive.coast @22:25:28.00 Air.coast @22:25:28.07 Lysis1.coast @22:25:28.14 Lysis2.coast @22:25:28.21 Plunge.coast @22:25:28.27 SamplePump.coast @22:25:28.35 Toroid.coast @22:25:28.42 Carousel.coast @22:25:28.63 Gate.power :motor,:OFF @22:25:28.65 Gate.power :core,:OFF @22:25:28.68 Gate.power :analytic1,:OFF @22:25:28.71 Gate.power :analytic2,:OFF @22:25:28.75 Gate.power :analytic3,:OFF @22:25:28.78 Gate.power :main,:OFF @22:25:28.82 Safely stopped and ready to power off @22:25:29.02 delay 2:00 @22:27:29.15 Gate.power :main,:ON,limit to 5000mA @22:27:29.18 Gate.power :analytic1,:ON,limit to 5000mA @22:27:29.21 Gate.power :analytic2,:ON,limit to 5000mA @22:27:29.25 Gate.power :analytic3,:ON,limit to 5000mA @22:27:29.32 Powering core bus @22:27:32.16 Archive.configure :default @22:27:32.50 Air.configure :default @22:27:32.79 Lysis1.configure :default @22:27:33.16 Lysis2.configure :default @22:27:33.54 Plunge.configure :default @22:27:33.95 SamplePump.configure :default @22:27:34.32 ExtraPump.configure :default @22:27:34.73 Toroid.configure :default @22:27:35.13 Carousel.configure :default @22:27:35.61 Pressure.configure :default @22:27:35.69 All elf controllers running firmware version 4.61 @22:27:35.71 Gate.power :motor,:ON,limit to 5000mA @22:27:36.05 Archive.reconfigure :default @22:27:36.41 Plunge.reconfigure :default @22:27:36.45 Archive.home.jog 0 @22:27:36.50 Lysis1.reconfigure :default @22:27:36.54 Air.reconfigure :default @22:27:36.60 Toroid.reconfigure :default @22:27:36.66 Lysis2.reconfigure :default @22:27:36.94 Plunge.home.jog 0 @22:27:36.98 Lysis1.home.jog 0 @22:27:37.03 Toroid.home.jog 0 @22:27:37.07 Air.home.jog 0 @22:27:37.12 Lysis2.home.jog 0 @22:27:37.83 Toroid.reconfigure :homing @22:27:37.94 Plunge.reconfigure :homing @22:27:37.99 Lysis1.reconfigure :homing @22:27:38.03 Air.reconfigure :homing @22:27:38.69 Archive.reconfigure :homing @22:27:38.75 Lysis2.reconfigure :homing @22:27:39.02 Toroid.jog 629 @22:27:39.06 Plunge.jog 851 @22:27:39.10 Air.jog 1369 @22:27:39.15 Archive.jog 1362 @22:27:39.21 Lysis1.jog 1351 @22:27:39.29 Lysis2.jog 1363 @22:27:40.33 Plunge.setRawPosition! 50 @22:27:40.47 Plunge.seek :home @22:27:41.20 Plunge.reconfigure :default @22:27:41.27 Lysis1.setRawPosition! 50 @22:27:41.31 Air.setRawPosition! 50 @22:27:41.41 Archive.setRawPosition! 50 @22:27:41.48 Lysis2.setRawPosition! 50 @22:27:41.83 Lysis1.seek :home @22:27:42.05 Air.seek :home @22:27:42.34 Lysis2.seek :home @22:27:42.44 Archive.seek :home @22:27:43.30 Toroid.setRawPosition! 0 @22:27:43.34 Lysis2.reconfigure :default @22:27:43.40 Air.reconfigure :default @22:27:43.57 Lysis1.reconfigure :default @22:27:43.92 Toroid.seek :home @22:27:44.49 Toroid.reconfigure :default @22:27:44.52 Archive.reconfigure :default @22:27:45.22 Toroid.seek :clear @22:27:46.92 Carousel.reconfigure :default @22:27:47.01 Carousel.home.jog 0 @22:27:47.53 Carousel.seek 54 @22:27:47.70 Selecting Cartridge 53 @22:27:47.95 Carousel.seek 53 @22:27:48.49 Lysis1.seek :locked @22:27:51.98 Plunge.seek :locked @22:27:52.31 Air.seek :locked @22:27:56.67 Mounted archive Cartridge @22:27:57.00 Duration of filtering limited to 4:05:00 @22:27:57.32 Priming sample loop w/160ml, bypass w/10ml @22:27:57.34 SamplePump.setPosition! 0ml @22:27:57.42 Bypass.open @22:27:58.93 Intake.open @22:28:00.02 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @22:28:02.69 Vacuum tare is 18% of full scale @22:28:02.76 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @22:28:03.23 Pressure.reconfigure :default @22:28:03.40 Pressure.hold @22:28:04.04 SamplePump.reconfigure :prime @22:28:04.12 SamplePump.seek 10ml,1:21 @22:28:15.33 Pressure.coast @22:28:15.41 SamplePump.setPosition! 0ml @22:28:15.48 Exhaust.open @22:28:16.90 Bypass.close @22:28:18.48 Pressure.reconfigure :default @22:28:18.56 Pressure.hold @22:28:18.85 SamplePump.reconfigure :prime @22:28:19.01 SamplePump.seek 160ml,13:51 @22:31:10.62 Pressure.coast @22:31:10.70 Bypass.open @22:31:11.39 Intake.close @22:31:12.29 Toroid.reconfigure :hold @22:31:12.37 Toroid.seek :crack @22:31:14.11 Toroid.seek :cartridge @22:31:15.24 Toroid.reconfigure :default @22:31:15.36 Toroid.seek :clear @22:31:17.60 Pressurizing cartridge to 14psi above 14.4psia @22:31:17.67 SamplePump.setPosition! 0ml @22:31:17.74 Intake.open @22:31:18.66 Bypass.close @22:31:19.40 Exhaust.close @22:31:20.28 Pressure.reconfigure :default @22:31:20.36 Pressure.hold @22:31:20.59 SamplePump.reconfigure :charge @22:31:20.68 SamplePump.seek 2.3ml,42 seconds @22:31:29.47 Pressure.coast @22:31:29.55 Intake.close @22:31:30.42 Pressurized cartridge with 0.350ml to 14.0psi @22:31:30.44 Isolated @25.2psia (10.8psi) @22:31:30.46 Checking for cartridge leaks from 25.2psia @22:31:30.50 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @22:31:35.19 Cartridge pressure fell 1% to 24.8psia @22:31:35.46 Sampling 2000.0ml @22:31:35.49 SamplePump.setPosition! 0ml @22:31:35.72 Exhaust.open @22:31:36.48 Intake.open @22:31:38.11 Pressure.reconfigure :default @22:31:38.20 Pressure.hold @22:31:38.51 SamplePump.reconfigure :sample @22:31:38.60 SamplePump.seek 2000ml,2:47:12 @22:31:40.65 Cartridge 53 sampling 1/2000ml at 0.52ml/s with 15.0psi vacuum @22:32:40.70 Cartridge 53 sampling 31/2000ml at 0.53ml/s with 13.4psi vacuum @22:33:40.65 Cartridge 53 sampling 63/2000ml at 0.55ml/s with 13.0psi vacuum @22:34:40.79 Cartridge 53 sampling 96/2000ml at 0.53ml/s with 10.2psi vacuum @22:35:40.66 Cartridge 53 sampling 128/2000ml at 0.54ml/s with 9.4psi vacuum @22:36:40.91 Cartridge 53 sampling 161/2000ml at 0.55ml/s with 13.5psi vacuum @22:37:40.85 Cartridge 53 sampling 195/2000ml at 0.56ml/s with 13.6psi vacuum @22:38:40.89 Cartridge 53 sampling 228/2000ml at 0.56ml/s with 14.6psi vacuum @22:39:40.94 Cartridge 53 sampling 262/2000ml at 0.54ml/s with 12.0psi vacuum @22:40:40.64 Cartridge 53 sampling 295/2000ml at 0.55ml/s with 14.7psi vacuum @22:41:41.04 Cartridge 53 sampling 329/2000ml at 0.56ml/s with 11.4psi vacuum @22:42:40.65 Cartridge 53 sampling 363/2000ml at 0.56ml/s with 12.8psi vacuum @22:43:40.94 Cartridge 53 sampling 396/2000ml at 0.56ml/s with 10.8psi vacuum @22:44:40.65 Cartridge 53 sampling 430/2000ml at 0.56ml/s with 11.2psi vacuum @22:45:40.81 Cartridge 53 sampling 464/2000ml at 0.56ml/s with 12.4psi vacuum @22:46:40.64 Cartridge 53 sampling 497/2000ml at 0.57ml/s with 13.4psi vacuum @22:47:40.94 Cartridge 53 sampling 531/2000ml at 0.58ml/s with 12.3psi vacuum @22:48:40.64 Cartridge 53 sampling 565/2000ml at 0.57ml/s with 11.9psi vacuum @22:49:40.77 Cartridge 53 sampling 599/2000ml at 0.57ml/s with 12.7psi vacuum @22:50:40.65 Cartridge 53 sampling 633/2000ml at 0.55ml/s with 12.7psi vacuum @22:51:40.83 Cartridge 53 sampling 667/2000ml at 0.56ml/s with 12.2psi vacuum @22:52:40.64 Cartridge 53 sampling 700/2000ml at 0.56ml/s with 14.6psi vacuum @22:53:40.82 Cartridge 53 sampling 734/2000ml at 0.56ml/s with 12.4psi vacuum @22:54:40.65 Cartridge 53 sampling 768/2000ml at 0.57ml/s with 13.1psi vacuum @22:55:40.73 Cartridge 53 sampling 802/2000ml at 0.56ml/s with 11.6psi vacuum @22:56:40.64 Cartridge 53 sampling 836/2000ml at 0.56ml/s with 14.7psi vacuum @22:57:40.72 Cartridge 53 sampling 870/2000ml at 0.57ml/s with 12.1psi vacuum @22:58:40.65 Cartridge 53 sampling 904/2000ml at 0.57ml/s with 13.4psi vacuum @22:59:40.74 Cartridge 53 sampling 939/2000ml at 0.56ml/s with 13.5psi vacuum @23:00:40.99 Cartridge 53 sampling 973/2000ml at 0.57ml/s with 12.9psi vacuum @23:01:40.65 Cartridge 53 sampling 1007/2000ml at 0.58ml/s with 13.3psi vacuum @23:02:41.24 Cartridge 53 sampling 1041/2000ml at 0.57ml/s with 13.5psi vacuum @23:03:40.66 Cartridge 53 sampling 1076/2000ml at 0.57ml/s with 12.1psi vacuum @23:04:40.89 Cartridge 53 sampling 1110/2000ml at 0.58ml/s with 12.7psi vacuum @23:05:40.65 Cartridge 53 sampling 1145/2000ml at 0.56ml/s with 12.9psi vacuum @23:06:40.83 Cartridge 53 sampling 1179/2000ml at 0.57ml/s with 13.8psi vacuum @23:07:40.65 Cartridge 53 sampling 1213/2000ml at 0.57ml/s with 14.8psi vacuum @23:08:40.85 Cartridge 53 sampling 1247/2000ml at 0.54ml/s with 12.3psi vacuum @23:09:40.66 Cartridge 53 sampling 1280/2000ml at 0.55ml/s with 12.8psi vacuum @23:10:40.83 Cartridge 53 sampling 1313/2000ml at 0.55ml/s with 11.1psi vacuum @23:11:40.66 Cartridge 53 sampling 1347/2000ml at 0.55ml/s with 11.1psi vacuum @23:12:40.78 Cartridge 53 sampling 1380/2000ml at 0.56ml/s with 12.8psi vacuum @23:13:40.65 Cartridge 53 sampling 1413/2000ml at 0.55ml/s with 14.1psi vacuum @23:14:40.83 Cartridge 53 sampling 1447/2000ml at 0.56ml/s with 12.2psi vacuum @23:15:40.66 Cartridge 53 sampling 1480/2000ml at 0.58ml/s with 13.9psi vacuum @23:16:40.82 Cartridge 53 sampling 1514/2000ml at 0.57ml/s with 14.1psi vacuum @23:17:40.65 Cartridge 53 sampling 1548/2000ml at 0.55ml/s with 13.2psi vacuum @23:18:40.81 Cartridge 53 sampling 1582/2000ml at 0.56ml/s with 12.8psi vacuum @23:19:40.66 Cartridge 53 sampling 1615/2000ml at 0.56ml/s with 13.1psi vacuum @23:20:40.68 Cartridge 53 sampling 1649/2000ml at 0.56ml/s with 14.3psi vacuum @23:21:40.65 Cartridge 53 sampling 1683/2000ml at 0.56ml/s with 14.2psi vacuum @23:22:40.82 Cartridge 53 sampling 1716/2000ml at 0.56ml/s with 11.5psi vacuum @23:23:40.87 Cartridge 53 sampling 1750/2000ml at 0.55ml/s with 11.8psi vacuum @23:24:40.66 Cartridge 53 sampling 1784/2000ml at 0.55ml/s with 14.7psi vacuum @23:25:41.20 Cartridge 53 sampling 1817/2000ml at 0.55ml/s with 13.8psi vacuum @23:26:40.65 Cartridge 53 sampling 1851/2000ml at 0.57ml/s with 13.0psi vacuum @23:27:41.05 Cartridge 53 sampling 1885/2000ml at 0.57ml/s with 14.4psi vacuum @23:28:40.65 Cartridge 53 sampling 1919/2000ml at 0.57ml/s with 14.4psi vacuum @23:29:40.87 Cartridge 53 sampling 1952/2000ml at 0.56ml/s with 14.1psi vacuum @23:30:40.65 Cartridge 53 sampling 1986/2000ml at 0.56ml/s with 11.9psi vacuum @23:31:05.51 Pressure.coast @23:31:05.83 Intake.close @23:31:06.00 Exhaust.close @23:31:06.42 Toroid.reconfigure :hold @23:31:06.60 Toroid.seek :bypass @23:31:08.35 Toroid.reconfigure :default @23:31:08.43 Toroid.seek :clear @23:31:10.06 Sampled 2000.0ml @23:31:10.58 Plunge.seek :reagent @23:31:11.50 Sample loop closed. Ready to Deliver Reagents @23:31:12.09 Carousel.seek 53 @23:31:12.75 Plunge.seek :locked @23:31:14.35 Air.seek :clear @23:31:14.79 Lysis1.seek :clear @23:31:18.24 Archive.coast @23:31:18.31 Air.coast @23:31:18.38 Lysis1.coast @23:31:19.01 Lysis2.coast @23:31:19.08 Plunge.coast @23:31:19.16 SamplePump.coast @23:31:19.23 Toroid.coast @23:31:19.30 Carousel.coast @23:31:19.53 Gate.power :motor,:OFF @23:31:19.55 Gate.power :core,:OFF @23:31:19.58 Gate.power :analytic1,:OFF @23:31:19.62 Gate.power :analytic2,:OFF @23:31:19.66 Gate.power :analytic3,:OFF @23:31:19.69 Gate.power :main,:OFF @23:31:19.72 Safely stopped and ready to power off @23:31:19.92 delay 2:00 @23:57:07.75 -> slots 58..53=>[:dry,:archiveNull_bac] @23:57:29.77 -> Cartridge.dry @23:58:19.10 -> slot 53..58=>:wet @23:58:24.60 -> Cartridge.dry @23:59:46.67 -> slots 52..50=>[:dry,:archiveNull_bac] @23:59:53.77 -> Cartridge.dry @00:01:35.20PDT29-May-21 -> 2.times {delay '2:00:00';shoot} @00:01:35.26 delay 2:00:00 @00:06:23.21 Session Begun @00:06:35.02 -> BACsampler.goalVolume @00:06:45.36 -> BACsampler.goalVolume=3000 @00:06:49.32 -> exit @00:06:49.36 Session Ended @02:01:35.71 Gate.power :main,:ON,limit to 5000mA @02:01:35.79 Gate.power :analytic1,:ON,limit to 5000mA @02:01:35.83 Gate.power :analytic2,:ON,limit to 5000mA @02:01:35.86 Gate.power :analytic3,:ON,limit to 5000mA @02:01:35.94 Powering core bus @02:01:38.40 Archive.configure :default @02:01:38.56 Air.configure :default @02:01:38.70 Lysis1.configure :default @02:01:38.84 Lysis2.configure :default @02:01:39.01 Plunge.configure :default @02:01:39.17 SamplePump.configure :default @02:01:39.33 ExtraPump.configure :default @02:01:39.50 Toroid.configure :default @02:01:39.67 Carousel.configure :default @02:01:39.84 Pressure.configure :default @02:01:39.92 All elf controllers running firmware version 4.61 @02:01:39.95 Gate.power :motor,:ON,limit to 5000mA @02:01:40.30 Air.reconfigure :default @02:01:40.56 Toroid.reconfigure :default @02:01:40.64 Air.home.jog 0 @02:01:40.69 Lysis2.reconfigure :default @02:01:40.74 Lysis1.reconfigure :default @02:01:40.80 Plunge.reconfigure :default @02:01:40.87 Archive.reconfigure :default @02:01:41.15 Toroid.home.jog 0 @02:01:41.21 Lysis2.home.jog 0 @02:01:41.26 Lysis1.home.jog 0 @02:01:41.30 Plunge.home.jog 0 @02:01:41.35 Archive.home.jog 0 @02:01:43.59 Toroid.reconfigure :homing @02:01:43.72 Plunge.reconfigure :homing @02:01:43.84 Lysis1.reconfigure :homing @02:01:43.90 Air.reconfigure :homing @02:01:44.02 Lysis2.reconfigure :homing @02:01:44.18 Archive.reconfigure :homing @02:01:44.74 Toroid.jog 627 @02:01:44.99 Plunge.jog 855 @02:01:45.23 Lysis2.jog 1362 @02:01:45.27 Air.jog 1364 @02:01:45.42 Lysis1.jog 1351 @02:01:45.52 Archive.jog 1365 @02:01:46.63 Plunge.setRawPosition! 50 @02:01:46.75 Plunge.seek :home @02:01:47.50 Toroid.setRawPosition! 0 @02:01:47.55 Lysis2.setRawPosition! 50 @02:01:47.58 Lysis1.setRawPosition! 50 @02:01:47.62 Archive.setRawPosition! 50 @02:01:47.65 Plunge.reconfigure :default @02:01:47.69 Air.setRawPosition! 50 @02:01:47.89 Toroid.seek :home @02:01:48.00 Lysis2.seek :home @02:01:48.14 Lysis1.seek :home @02:01:48.18 Air.seek :home @02:01:48.24 Archive.seek :home @02:01:48.55 Toroid.reconfigure :default @02:01:48.74 Lysis1.reconfigure :default @02:01:48.79 Air.reconfigure :default @02:01:48.83 Lysis2.reconfigure :default @02:01:48.87 Archive.reconfigure :default @02:01:49.42 Toroid.seek :clear @02:01:51.06 Carousel.reconfigure :default @02:01:51.15 Carousel.home.jog 0 @02:01:51.67 Carousel.seek 53 @02:01:53.55 Selecting Cartridge 52 @02:01:54.12 Carousel.seek 52 @02:01:54.93 Lysis1.seek :locked @02:01:58.22 Plunge.seek :locked @02:01:58.41 Air.seek :locked @02:02:02.98 Mounted archive Cartridge @02:02:04.05 Duration of filtering limited to 4:05:00 @02:02:04.79 Priming sample loop w/160ml, bypass w/10ml @02:02:04.84 SamplePump.setPosition! 0ml @02:02:04.97 Bypass.open @02:02:06.67 Intake.open @02:02:07.70 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @02:02:10.29 Vacuum tare is 18% of full scale @02:02:10.31 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @02:02:10.57 Pressure.reconfigure :default @02:02:10.65 Pressure.hold @02:02:10.88 SamplePump.reconfigure :prime @02:02:10.97 SamplePump.seek 10ml,1:21 @02:02:22.12 Pressure.coast @02:02:22.29 SamplePump.setPosition! 0ml @02:02:22.46 Exhaust.open @02:02:23.93 Bypass.close @02:02:25.64 Pressure.reconfigure :default @02:02:25.86 Pressure.hold @02:02:26.56 SamplePump.reconfigure :prime @02:02:26.64 SamplePump.seek 160ml,13:51 @02:05:18.18 Pressure.coast @02:05:18.41 Bypass.open @02:05:19.12 Intake.close @02:05:20.18 Toroid.reconfigure :hold @02:05:20.30 Toroid.seek :crack @02:05:22.02 Toroid.seek :cartridge @02:05:22.98 Toroid.reconfigure :default @02:05:23.06 Toroid.seek :clear @02:05:24.94 Pressurizing cartridge to 14psi above 14.5psia @02:05:24.96 SamplePump.setPosition! 0ml @02:05:25.03 Intake.open @02:05:25.88 Bypass.close @02:05:26.70 Exhaust.close @02:05:27.88 Pressure.reconfigure :default @02:05:28.18 Pressure.hold @02:05:29.18 SamplePump.reconfigure :charge @02:05:29.39 SamplePump.seek 2.3ml,42 seconds @02:05:38.27 Pressure.coast @02:05:38.39 Intake.close @02:05:39.54 Pressurized cartridge with 0.338ml to 14.0psi @02:05:39.56 Isolated @25.7psia (11.2psi) @02:05:39.58 Checking for cartridge leaks from 25.7psia @02:05:39.63 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @02:05:44.26 Cartridge pressure fell 2% to 25.2psia @02:05:44.37 Sampling 3000.0ml @02:05:44.38 SamplePump.setPosition! 0ml @02:05:44.46 Exhaust.open @02:05:45.20 Intake.open @02:05:46.58 Pressure.reconfigure :default @02:05:46.82 Pressure.hold @02:05:47.56 SamplePump.reconfigure :sample @02:05:47.72 SamplePump.seek 3000ml,4:10:32 @02:05:49.65 Cartridge 52 sampling 1/3000ml at 0.59ml/s with 12.9psi vacuum @02:06:49.48 Cartridge 52 sampling 30/3000ml at 0.51ml/s with 12.9psi vacuum @02:07:49.93 Cartridge 52 sampling 62/3000ml at 0.53ml/s with 14.1psi vacuum @02:08:49.49 Cartridge 52 sampling 95/3000ml at 0.57ml/s with 13.9psi vacuum @02:09:49.47 Cartridge 52 sampling 128/3000ml at 0.55ml/s with 15.5psi vacuum @02:10:49.75 Cartridge 52 sampling 161/3000ml at 0.56ml/s with 14.7psi vacuum @02:11:49.47 Cartridge 52 sampling 195/3000ml at 0.56ml/s with 14.4psi vacuum @02:12:49.83 Cartridge 52 sampling 229/3000ml at 0.57ml/s with 12.1psi vacuum @02:13:49.48 Cartridge 52 sampling 262/3000ml at 0.55ml/s with 14.8psi vacuum @02:14:49.69 Cartridge 52 sampling 296/3000ml at 0.57ml/s with 13.9psi vacuum @02:15:49.47 Cartridge 52 sampling 330/3000ml at 0.57ml/s with 12.8psi vacuum @02:16:49.61 Cartridge 52 sampling 364/3000ml at 0.55ml/s with 13.8psi vacuum @02:17:49.47 Cartridge 52 sampling 398/3000ml at 0.54ml/s with 11.8psi vacuum @02:18:49.71 Cartridge 52 sampling 432/3000ml at 0.56ml/s with 12.9psi vacuum @02:19:49.48 Cartridge 52 sampling 466/3000ml at 0.58ml/s with 13.5psi vacuum @02:20:49.77 Cartridge 52 sampling 500/3000ml at 0.56ml/s with 11.4psi vacuum @02:21:49.47 Cartridge 52 sampling 534/3000ml at 0.58ml/s with 14.1psi vacuum @02:22:49.67 Cartridge 52 sampling 568/3000ml at 0.56ml/s with 13.3psi vacuum @02:23:49.47 Cartridge 52 sampling 602/3000ml at 0.56ml/s with 10.6psi vacuum @02:24:49.59 Cartridge 52 sampling 636/3000ml at 0.56ml/s with 14.9psi vacuum @02:25:49.49 Cartridge 52 sampling 670/3000ml at 0.57ml/s with 12.9psi vacuum @02:26:49.64 Cartridge 52 sampling 705/3000ml at 0.57ml/s with 14.0psi vacuum @02:27:49.48 Cartridge 52 sampling 739/3000ml at 0.58ml/s with 13.2psi vacuum @02:28:49.83 Cartridge 52 sampling 773/3000ml at 0.58ml/s with 13.7psi vacuum @02:29:49.71 Cartridge 52 sampling 808/3000ml at 0.57ml/s with 12.0psi vacuum @02:30:49.62 Cartridge 52 sampling 842/3000ml at 0.57ml/s with 14.0psi vacuum @02:31:49.63 Cartridge 52 sampling 877/3000ml at 0.57ml/s with 11.9psi vacuum @02:32:49.67 Cartridge 52 sampling 912/3000ml at 0.57ml/s with 14.5psi vacuum @02:33:49.73 Cartridge 52 sampling 946/3000ml at 0.59ml/s with 12.7psi vacuum @02:34:49.48 Cartridge 52 sampling 982/3000ml at 0.58ml/s with 12.9psi vacuum @02:35:49.65 Cartridge 52 sampling 1016/3000ml at 0.57ml/s with 11.3psi vacuum @02:36:49.48 Cartridge 52 sampling 1050/3000ml at 0.57ml/s with 13.6psi vacuum @02:37:49.99 Cartridge 52 sampling 1084/3000ml at 0.56ml/s with 11.6psi vacuum @02:38:49.49 Cartridge 52 sampling 1118/3000ml at 0.56ml/s with 13.7psi vacuum @02:39:49.65 Cartridge 52 sampling 1152/3000ml at 0.57ml/s with 13.8psi vacuum @02:40:49.48 Cartridge 52 sampling 1186/3000ml at 0.56ml/s with 11.5psi vacuum @02:41:49.64 Cartridge 52 sampling 1220/3000ml at 0.57ml/s with 11.5psi vacuum @02:42:49.48 Cartridge 52 sampling 1254/3000ml at 0.57ml/s with 14.7psi vacuum @02:43:49.65 Cartridge 52 sampling 1289/3000ml at 0.58ml/s with 14.0psi vacuum @02:44:49.49 Cartridge 52 sampling 1323/3000ml at 0.57ml/s with 14.6psi vacuum @02:45:49.67 Cartridge 52 sampling 1358/3000ml at 0.57ml/s with 12.2psi vacuum @02:46:49.49 Cartridge 52 sampling 1392/3000ml at 0.57ml/s with 14.6psi vacuum @02:47:49.70 Cartridge 52 sampling 1426/3000ml at 0.57ml/s with 11.8psi vacuum @02:48:49.48 Cartridge 52 sampling 1461/3000ml at 0.57ml/s with 12.1psi vacuum @02:49:49.51 Cartridge 52 sampling 1496/3000ml at 0.58ml/s with 13.5psi vacuum @02:50:49.48 Cartridge 52 sampling 1530/3000ml at 0.58ml/s with 14.4psi vacuum @02:51:49.64 Cartridge 52 sampling 1565/3000ml at 0.58ml/s with 13.4psi vacuum @02:52:49.48 Cartridge 52 sampling 1600/3000ml at 0.58ml/s with 13.0psi vacuum @02:53:49.91 Cartridge 52 sampling 1633/3000ml at 0.55ml/s with 13.7psi vacuum @02:54:49.82 Cartridge 52 sampling 1667/3000ml at 0.57ml/s with 11.2psi vacuum @02:55:49.71 Cartridge 52 sampling 1702/3000ml at 0.57ml/s with 12.9psi vacuum @02:56:49.73 Cartridge 52 sampling 1736/3000ml at 0.57ml/s with 12.5psi vacuum @02:57:49.48 Cartridge 52 sampling 1770/3000ml at 0.57ml/s with 13.7psi vacuum @02:58:49.76 Cartridge 52 sampling 1805/3000ml at 0.57ml/s with 12.4psi vacuum @02:59:49.49 Cartridge 52 sampling 1839/3000ml at 0.57ml/s with 11.3psi vacuum @03:00:49.86 Cartridge 52 sampling 1874/3000ml at 0.57ml/s with 12.5psi vacuum @03:01:49.48 Cartridge 52 sampling 1908/3000ml at 0.57ml/s with 13.9psi vacuum @03:02:49.69 Cartridge 52 sampling 1942/3000ml at 0.57ml/s with 12.9psi vacuum @03:03:49.49 Cartridge 52 sampling 1976/3000ml at 0.57ml/s with 13.0psi vacuum @03:04:49.66 Cartridge 52 sampling 2011/3000ml at 0.58ml/s with 12.8psi vacuum @03:05:49.48 Cartridge 52 sampling 2046/3000ml at 0.56ml/s with 13.4psi vacuum @03:06:49.61 Cartridge 52 sampling 2080/3000ml at 0.57ml/s with 12.8psi vacuum @03:07:49.48 Cartridge 52 sampling 2115/3000ml at 0.58ml/s with 14.5psi vacuum @03:08:49.68 Cartridge 52 sampling 2149/3000ml at 0.56ml/s with 14.0psi vacuum @03:09:49.48 Cartridge 52 sampling 2183/3000ml at 0.57ml/s with 14.3psi vacuum @03:10:49.62 Cartridge 52 sampling 2217/3000ml at 0.56ml/s with 15.2psi vacuum @03:11:49.49 Cartridge 52 sampling 2252/3000ml at 0.57ml/s with 11.7psi vacuum @03:12:49.65 Cartridge 52 sampling 2286/3000ml at 0.57ml/s with 12.1psi vacuum @03:13:49.49 Cartridge 52 sampling 2320/3000ml at 0.57ml/s with 13.6psi vacuum @03:14:49.64 Cartridge 52 sampling 2354/3000ml at 0.57ml/s with 11.4psi vacuum @03:15:49.48 Cartridge 52 sampling 2389/3000ml at 0.56ml/s with 13.6psi vacuum @03:16:49.72 Cartridge 52 sampling 2424/3000ml at 0.58ml/s with 14.3psi vacuum @03:17:50.02 Cartridge 52 sampling 2458/3000ml at 0.57ml/s with 12.2psi vacuum @03:18:49.76 Cartridge 52 sampling 2492/3000ml at 0.57ml/s with 13.1psi vacuum @03:19:49.80 Cartridge 52 sampling 2526/3000ml at 0.58ml/s with 13.8psi vacuum @03:20:49.62 Cartridge 52 sampling 2561/3000ml at 0.57ml/s with 14.0psi vacuum @03:21:49.64 Cartridge 52 sampling 2596/3000ml at 0.58ml/s with 13.0psi vacuum @03:22:49.49 Cartridge 52 sampling 2631/3000ml at 0.59ml/s with 12.6psi vacuum @03:23:49.74 Cartridge 52 sampling 2666/3000ml at 0.58ml/s with 14.1psi vacuum @03:24:49.48 Cartridge 52 sampling 2699/3000ml at 0.56ml/s with 12.3psi vacuum @03:25:49.79 Cartridge 52 sampling 2733/3000ml at 0.56ml/s with 13.6psi vacuum @03:26:49.48 Cartridge 52 sampling 2767/3000ml at 0.58ml/s with 11.4psi vacuum @03:27:49.60 Cartridge 52 sampling 2801/3000ml at 0.56ml/s with 13.0psi vacuum @03:28:49.48 Cartridge 52 sampling 2835/3000ml at 0.56ml/s with 13.2psi vacuum @03:29:49.67 Cartridge 52 sampling 2869/3000ml at 0.56ml/s with 13.2psi vacuum @03:30:49.49 Cartridge 52 sampling 2903/3000ml at 0.58ml/s with 14.0psi vacuum @03:31:49.60 Cartridge 52 sampling 2938/3000ml at 0.57ml/s with 13.9psi vacuum @03:32:49.48 Cartridge 52 sampling 2972/3000ml at 0.57ml/s with 15.1psi vacuum @03:33:38.37 Pressure.coast @03:33:38.69 Intake.close @03:33:38.84 Exhaust.close @03:33:39.38 Toroid.reconfigure :hold @03:33:39.54 Toroid.seek :bypass @03:33:41.30 Toroid.reconfigure :default @03:33:41.40 Toroid.seek :clear @03:33:43.18 Sampled 3000.0ml @03:33:43.69 Plunge.seek :reagent @03:33:44.62 Sample loop closed. Ready to Deliver Reagents @03:33:44.88 Carousel.seek 52 @03:33:45.13 Plunge.seek :locked @03:33:46.21 Air.seek :clear @03:33:46.96 Lysis1.seek :clear @03:33:50.53 Archive.coast @03:33:50.67 Air.coast @03:33:50.86 Lysis1.coast @03:33:50.94 Lysis2.coast @03:33:51.16 Plunge.coast @03:33:51.23 SamplePump.coast @03:33:51.45 Toroid.coast @03:33:51.53 Carousel.coast @03:33:52.06 Gate.power :motor,:OFF @03:33:52.15 Gate.power :core,:OFF @03:33:52.20 Gate.power :analytic1,:OFF @03:33:52.33 Gate.power :analytic2,:OFF @03:33:52.38 Gate.power :analytic3,:OFF @03:33:52.42 Gate.power :main,:OFF @03:33:52.45 Safely stopped and ready to power off @03:33:52.67 delay 2:00:00 @05:33:52.82 Gate.power :main,:ON,limit to 5000mA @05:33:52.84 Gate.power :analytic1,:ON,limit to 5000mA @05:33:52.87 Gate.power :analytic2,:ON,limit to 5000mA @05:33:52.90 Gate.power :analytic3,:ON,limit to 5000mA @05:33:52.97 Powering core bus @05:33:55.58 Archive.configure :default @05:33:56.00 Air.configure :default @05:33:56.29 Lysis1.configure :default @05:33:56.61 Lysis2.configure :default @05:33:57.05 Plunge.configure :default @05:33:57.40 SamplePump.configure :default @05:33:57.71 ExtraPump.configure :default @05:33:58.04 Toroid.configure :default @05:33:58.35 Carousel.configure :default @05:33:59.26 Pressure.configure :default @05:33:59.34 All elf controllers running firmware version 4.61 @05:33:59.37 Gate.power :motor,:ON,limit to 5000mA @05:33:59.57 Archive.reconfigure :default @05:33:59.86 Archive.home.jog 0 @05:34:00.16 Toroid.reconfigure :default @05:34:00.19 Plunge.reconfigure :default @05:34:00.23 Lysis2.reconfigure :default @05:34:00.27 Air.reconfigure :default @05:34:00.34 Lysis1.reconfigure :default @05:34:00.56 Toroid.home.jog 0 @05:34:00.62 Plunge.home.jog 0 @05:34:00.66 Lysis2.home.jog 0 @05:34:00.69 Air.home.jog 0 @05:34:00.76 Lysis1.home.jog 0 @05:34:01.26 Archive.reconfigure :homing @05:34:01.52 Archive.jog 1365 @05:34:01.57 Toroid.reconfigure :homing @05:34:01.62 Plunge.reconfigure :homing @05:34:01.66 Lysis2.reconfigure :homing @05:34:01.71 Air.reconfigure :homing @05:34:01.76 Lysis1.reconfigure :homing @05:34:02.02 Toroid.jog 635 @05:34:02.06 Lysis2.jog 1363 @05:34:02.14 Plunge.jog 851 @05:34:02.19 Air.jog 1369 @05:34:02.26 Lysis1.jog 1353 @05:34:02.98 Archive.setRawPosition! 50 @05:34:03.04 Archive.seek :home @05:34:03.25 Plunge.setRawPosition! 50 @05:34:03.32 Plunge.seek :home @05:34:03.55 Lysis2.setRawPosition! 50 @05:34:03.77 Archive.reconfigure :default @05:34:04.23 Lysis2.seek :home @05:34:04.91 Toroid.setRawPosition! 0 @05:34:04.98 Plunge.reconfigure :default @05:34:05.11 Lysis1.setRawPosition! 50 @05:34:05.27 Air.setRawPosition! 50 @05:34:05.43 Toroid.seek :home @05:34:05.62 Lysis1.seek :home @05:34:05.87 Air.seek :home @05:34:06.51 Lysis2.reconfigure :default @05:34:06.67 Lysis1.reconfigure :default @05:34:06.72 Toroid.reconfigure :default @05:34:06.83 Air.reconfigure :default @05:34:07.66 Toroid.seek :clear @05:34:09.40 Carousel.reconfigure :default @05:34:09.48 Carousel.home.jog 0 @05:34:10.01 Carousel.seek 52 @05:34:10.19 Selecting Cartridge 51 @05:34:10.43 Carousel.seek 51 @05:34:10.97 Lysis1.seek :locked @05:34:14.51 Plunge.seek :locked @05:34:15.02 Air.seek :locked @05:34:19.41 Mounted archive Cartridge @05:34:19.73 Duration of filtering limited to 4:05:00 @05:34:20.06 Priming sample loop w/160ml, bypass w/10ml @05:34:20.07 SamplePump.setPosition! 0ml @05:34:20.15 Bypass.open @05:34:21.66 Intake.open @05:34:22.57 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @05:34:26.60 Vacuum tare is 18% of full scale @05:34:26.78 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @05:34:27.38 Pressure.reconfigure :default @05:34:27.46 Pressure.hold @05:34:28.02 SamplePump.reconfigure :prime @05:34:28.11 SamplePump.seek 10ml,1:21 @05:34:39.29 Pressure.coast @05:34:39.36 SamplePump.setPosition! 0ml @05:34:39.43 Exhaust.open @05:34:40.86 Bypass.close @05:34:42.84 Pressure.reconfigure :default @05:34:43.05 Pressure.hold @05:34:44.13 SamplePump.reconfigure :prime @05:34:44.32 SamplePump.seek 160ml,13:51 @05:37:35.86 Pressure.coast @05:37:35.94 Bypass.open @05:37:36.64 Intake.close @05:37:37.91 Toroid.reconfigure :hold @05:37:38.33 Toroid.seek :crack @05:37:40.20 Toroid.seek :cartridge @05:37:41.35 Toroid.reconfigure :default @05:37:41.51 Toroid.seek :clear @05:37:43.46 Pressurizing cartridge to 14psi above 14.6psia @05:37:43.48 SamplePump.setPosition! 0ml @05:37:43.55 Intake.open @05:37:44.35 Bypass.close @05:37:45.08 Exhaust.close @05:37:45.94 Pressure.reconfigure :default @05:37:46.01 Pressure.hold @05:37:46.26 SamplePump.reconfigure :charge @05:37:46.34 SamplePump.seek 2.3ml,42 seconds @05:37:54.92 Pressure.coast @05:37:55.00 Intake.close @05:37:55.85 Pressurized cartridge with 0.334ml to 14.1psi @05:37:55.87 Isolated @25.6psia (11.1psi) @05:37:55.89 Checking for cartridge leaks from 25.6psia @05:37:55.94 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @05:38:00.66 Cartridge pressure fell 4% to 24.7psia @05:38:00.94 Sampling 3000.0ml @05:38:00.95 SamplePump.setPosition! 0ml @05:38:01.08 Exhaust.open @05:38:01.88 Intake.open @05:38:02.82 Pressure.reconfigure :default @05:38:02.90 Pressure.hold @05:38:03.19 SamplePump.reconfigure :sample @05:38:03.28 SamplePump.seek 3000ml,4:10:32 @05:38:05.88 Cartridge 51 sampling 1/3000ml at 0.51ml/s with 13.9psi vacuum @05:39:06.04 Cartridge 51 sampling 31/3000ml at 0.51ml/s with 10.1psi vacuum @05:40:06.24 Cartridge 51 sampling 64/3000ml at 0.54ml/s with 14.1psi vacuum @05:41:05.87 Cartridge 51 sampling 97/3000ml at 0.55ml/s with 12.7psi vacuum @05:42:06.29 Cartridge 51 sampling 130/3000ml at 0.55ml/s with 11.9psi vacuum @05:43:05.88 Cartridge 51 sampling 164/3000ml at 0.57ml/s with 11.8psi vacuum @05:44:06.14 Cartridge 51 sampling 198/3000ml at 0.57ml/s with 10.4psi vacuum @05:45:05.87 Cartridge 51 sampling 232/3000ml at 0.55ml/s with 14.7psi vacuum @05:46:06.11 Cartridge 51 sampling 266/3000ml at 0.56ml/s with 12.8psi vacuum @05:47:05.87 Cartridge 51 sampling 299/3000ml at 0.55ml/s with 11.1psi vacuum @05:48:06.04 Cartridge 51 sampling 333/3000ml at 0.56ml/s with 14.7psi vacuum @05:49:05.87 Cartridge 51 sampling 366/3000ml at 0.56ml/s with 13.9psi vacuum @05:50:06.12 Cartridge 51 sampling 400/3000ml at 0.56ml/s with 11.2psi vacuum @05:51:05.87 Cartridge 51 sampling 434/3000ml at 0.58ml/s with 12.6psi vacuum @05:52:06.00 Cartridge 51 sampling 468/3000ml at 0.58ml/s with 13.4psi vacuum @05:53:05.89 Cartridge 51 sampling 503/3000ml at 0.56ml/s with 12.1psi vacuum @05:54:06.05 Cartridge 51 sampling 537/3000ml at 0.58ml/s with 13.7psi vacuum @05:55:05.88 Cartridge 51 sampling 572/3000ml at 0.57ml/s with 11.9psi vacuum @05:56:06.04 Cartridge 51 sampling 606/3000ml at 0.57ml/s with 14.0psi vacuum @05:57:05.88 Cartridge 51 sampling 641/3000ml at 0.59ml/s with 12.8psi vacuum @05:58:06.13 Cartridge 51 sampling 675/3000ml at 0.57ml/s with 12.5psi vacuum @05:59:06.03 Cartridge 51 sampling 709/3000ml at 0.55ml/s with 12.0psi vacuum @06:00:05.88 Cartridge 51 sampling 743/3000ml at 0.56ml/s with 14.6psi vacuum @06:01:06.26 Cartridge 51 sampling 777/3000ml at 0.57ml/s with 12.7psi vacuum @06:02:05.88 Cartridge 51 sampling 811/3000ml at 0.56ml/s with 12.1psi vacuum @06:03:06.29 Cartridge 51 sampling 846/3000ml at 0.57ml/s with 14.7psi vacuum @06:04:05.88 Cartridge 51 sampling 880/3000ml at 0.57ml/s with 13.8psi vacuum @06:05:06.16 Cartridge 51 sampling 915/3000ml at 0.57ml/s with 13.4psi vacuum @06:06:05.88 Cartridge 51 sampling 949/3000ml at 0.56ml/s with 14.1psi vacuum @06:07:06.04 Cartridge 51 sampling 983/3000ml at 0.57ml/s with 13.3psi vacuum @06:08:05.88 Cartridge 51 sampling 1017/3000ml at 0.58ml/s with 11.6psi vacuum @06:09:06.10 Cartridge 51 sampling 1052/3000ml at 0.58ml/s with 13.4psi vacuum @06:10:05.89 Cartridge 51 sampling 1087/3000ml at 0.58ml/s with 11.9psi vacuum @06:11:06.04 Cartridge 51 sampling 1121/3000ml at 0.59ml/s with 12.8psi vacuum @06:12:05.89 Cartridge 51 sampling 1156/3000ml at 0.57ml/s with 12.3psi vacuum @06:13:06.05 Cartridge 51 sampling 1191/3000ml at 0.57ml/s with 14.0psi vacuum @06:14:05.88 Cartridge 51 sampling 1225/3000ml at 0.57ml/s with 13.1psi vacuum @06:15:06.06 Cartridge 51 sampling 1260/3000ml at 0.59ml/s with 13.0psi vacuum @06:16:05.88 Cartridge 51 sampling 1295/3000ml at 0.57ml/s with 13.9psi vacuum @06:17:07.08 Cartridge 51 sampling 1330/3000ml at 0.57ml/s with 13.7psi vacuum @06:18:05.89 Cartridge 51 sampling 1363/3000ml at 0.57ml/s with 12.4psi vacuum @06:19:06.02 Cartridge 51 sampling 1398/3000ml at 0.57ml/s with 12.9psi vacuum @06:20:05.88 Cartridge 51 sampling 1432/3000ml at 0.56ml/s with 11.4psi vacuum @06:21:05.92 Cartridge 51 sampling 1466/3000ml at 0.57ml/s with 13.4psi vacuum @06:22:05.88 Cartridge 51 sampling 1500/3000ml at 0.57ml/s with 13.1psi vacuum @06:23:06.32 Cartridge 51 sampling 1535/3000ml at 0.57ml/s with 13.0psi vacuum @06:24:06.35 Cartridge 51 sampling 1569/3000ml at 0.58ml/s with 14.2psi vacuum @06:25:05.99 Cartridge 51 sampling 1604/3000ml at 0.58ml/s with 14.4psi vacuum @06:26:06.06 Cartridge 51 sampling 1638/3000ml at 0.58ml/s with 11.4psi vacuum @06:27:05.88 Cartridge 51 sampling 1673/3000ml at 0.58ml/s with 13.9psi vacuum @06:28:06.26 Cartridge 51 sampling 1708/3000ml at 0.57ml/s with 14.8psi vacuum @06:29:05.89 Cartridge 51 sampling 1743/3000ml at 0.58ml/s with 14.2psi vacuum @06:30:06.20 Cartridge 51 sampling 1778/3000ml at 0.58ml/s with 13.7psi vacuum @06:31:05.88 Cartridge 51 sampling 1813/3000ml at 0.55ml/s with 13.4psi vacuum @06:32:06.13 Cartridge 51 sampling 1846/3000ml at 0.58ml/s with 11.3psi vacuum @06:33:05.88 Cartridge 51 sampling 1880/3000ml at 0.56ml/s with 13.8psi vacuum @06:34:06.16 Cartridge 51 sampling 1914/3000ml at 0.57ml/s with 13.8psi vacuum @06:35:05.88 Cartridge 51 sampling 1948/3000ml at 0.58ml/s with 12.2psi vacuum @06:36:06.21 Cartridge 51 sampling 1982/3000ml at 0.56ml/s with 12.5psi vacuum @06:37:05.88 Cartridge 51 sampling 2017/3000ml at 0.58ml/s with 12.8psi vacuum @06:38:06.07 Cartridge 51 sampling 2051/3000ml at 0.57ml/s with 11.7psi vacuum @06:39:05.89 Cartridge 51 sampling 2086/3000ml at 0.57ml/s with 13.2psi vacuum @06:40:06.06 Cartridge 51 sampling 2120/3000ml at 0.58ml/s with 13.4psi vacuum @06:41:05.88 Cartridge 51 sampling 2155/3000ml at 0.58ml/s with 14.3psi vacuum @06:42:06.00 Cartridge 51 sampling 2190/3000ml at 0.57ml/s with 14.2psi vacuum @06:43:05.88 Cartridge 51 sampling 2224/3000ml at 0.58ml/s with 13.2psi vacuum @06:44:06.08 Cartridge 51 sampling 2259/3000ml at 0.55ml/s with 13.1psi vacuum @06:45:05.89 Cartridge 51 sampling 2293/3000ml at 0.56ml/s with 14.1psi vacuum @06:46:06.10 Cartridge 51 sampling 2327/3000ml at 0.57ml/s with 13.9psi vacuum @06:47:05.88 Cartridge 51 sampling 2361/3000ml at 0.57ml/s with 12.3psi vacuum @06:48:06.09 Cartridge 51 sampling 2395/3000ml at 0.56ml/s with 11.3psi vacuum @06:49:05.88 Cartridge 51 sampling 2430/3000ml at 0.58ml/s with 11.6psi vacuum @06:50:06.20 Cartridge 51 sampling 2465/3000ml at 0.58ml/s with 13.4psi vacuum @06:51:05.88 Cartridge 51 sampling 2499/3000ml at 0.58ml/s with 12.5psi vacuum @06:52:06.05 Cartridge 51 sampling 2534/3000ml at 0.57ml/s with 12.1psi vacuum @06:53:05.90 Cartridge 51 sampling 2569/3000ml at 0.58ml/s with 12.8psi vacuum @06:54:06.43 Cartridge 51 sampling 2605/3000ml at 0.58ml/s with 13.3psi vacuum @06:55:05.88 Cartridge 51 sampling 2640/3000ml at 0.59ml/s with 14.1psi vacuum @06:56:05.88 Cartridge 51 sampling 2675/3000ml at 0.55ml/s with 13.0psi vacuum @06:57:06.37 Cartridge 51 sampling 2709/3000ml at 0.57ml/s with 11.8psi vacuum @06:58:05.89 Cartridge 51 sampling 2743/3000ml at 0.58ml/s with 11.8psi vacuum @06:59:06.30 Cartridge 51 sampling 2778/3000ml at 0.57ml/s with 11.7psi vacuum @07:00:05.88 Cartridge 51 sampling 2812/3000ml at 0.57ml/s with 11.6psi vacuum @07:01:06.30 Cartridge 51 sampling 2847/3000ml at 0.58ml/s with 11.6psi vacuum @07:02:05.88 Cartridge 51 sampling 2881/3000ml at 0.57ml/s with 12.5psi vacuum @07:03:06.09 Cartridge 51 sampling 2916/3000ml at 0.58ml/s with 13.1psi vacuum @07:04:05.88 Cartridge 51 sampling 2951/3000ml at 0.57ml/s with 12.1psi vacuum @07:05:06.01 Cartridge 51 sampling 2985/3000ml at 0.58ml/s with 14.5psi vacuum @07:05:31.14 Pressure.coast @07:05:31.30 Intake.close @07:05:31.37 Exhaust.close @07:05:31.56 Toroid.reconfigure :hold @07:05:31.65 Toroid.seek :bypass @07:05:33.44 Toroid.reconfigure :default @07:05:33.63 Toroid.seek :clear @07:05:35.37 Sampled 3000.0ml @07:05:36.35 Plunge.seek :reagent @07:05:37.38 Sample loop closed. Ready to Deliver Reagents @07:05:37.65 Carousel.seek 51 @07:05:37.90 Plunge.seek :locked @07:05:38.97 Air.seek :clear @07:05:39.17 Lysis1.seek :clear @07:05:42.73 Archive.coast @07:05:42.92 Air.coast @07:05:43.06 Lysis1.coast @07:05:43.26 Lysis2.coast @07:05:43.39 Plunge.coast @07:05:43.60 SamplePump.coast @07:05:43.76 Toroid.coast @07:05:43.93 Carousel.coast @07:05:44.47 Gate.power :motor,:OFF @07:05:44.63 Gate.power :core,:OFF @07:05:44.69 Gate.power :analytic1,:OFF @07:05:44.73 Gate.power :analytic2,:OFF @07:05:44.81 Gate.power :analytic3,:OFF @07:05:44.91 Gate.power :main,:OFF @07:05:44.99 Safely stopped and ready to power off @08:19:32.82 -> Cartridge.dry @08:25:11.05 -> delay '3:00:00';shoot @08:25:11.09 delay 3:00:00 @11:25:11.24 Gate.power :main,:ON,limit to 5000mA @11:25:11.26 Gate.power :analytic1,:ON,limit to 5000mA @11:25:11.29 Gate.power :analytic2,:ON,limit to 5000mA @11:25:11.33 Gate.power :analytic3,:ON,limit to 5000mA @11:25:11.40 Powering core bus @11:25:14.15 Archive.configure :default @11:25:14.53 Air.configure :default @11:25:14.87 Lysis1.configure :default @11:25:15.26 Lysis2.configure :default @11:25:15.62 Plunge.configure :default @11:25:15.94 SamplePump.configure :default @11:25:16.28 ExtraPump.configure :default @11:25:16.60 Toroid.configure :default @11:25:17.07 Carousel.configure :default @11:25:17.47 Pressure.configure :default @11:25:17.55 All elf controllers running firmware version 4.61 @11:25:17.58 Gate.power :motor,:ON,limit to 5000mA @11:25:17.81 Air.reconfigure :default @11:25:18.13 Air.home.jog 0 @11:25:18.34 Plunge.reconfigure :default @11:25:18.40 Lysis2.reconfigure :default @11:25:18.44 Toroid.reconfigure :default @11:25:18.51 Lysis1.reconfigure :default @11:25:18.56 Archive.reconfigure :default @11:25:18.82 Plunge.home.jog 0 @11:25:18.87 Lysis2.home.jog 0 @11:25:18.92 Toroid.home.jog 0 @11:25:18.97 Lysis1.home.jog 0 @11:25:19.04 Archive.home.jog 0 @11:25:19.48 Air.reconfigure :homing @11:25:19.81 Air.jog 1370 @11:25:19.88 Toroid.reconfigure :homing @11:25:19.92 Lysis2.reconfigure :homing @11:25:19.96 Archive.reconfigure :homing @11:25:20.03 Lysis1.reconfigure :homing @11:25:20.08 Plunge.reconfigure :homing @11:25:20.37 Lysis2.jog 1363 @11:25:20.44 Toroid.jog 630 @11:25:20.48 Archive.jog 1366 @11:25:20.52 Plunge.jog 859 @11:25:20.57 Lysis1.jog 1351 @11:25:21.29 Air.setRawPosition! 50 @11:25:21.36 Air.seek :home @11:25:21.62 Plunge.setRawPosition! 50 @11:25:21.89 Plunge.seek :home @11:25:22.28 Lysis2.setRawPosition! 50 @11:25:22.39 Lysis1.setRawPosition! 50 @11:25:22.49 Archive.setRawPosition! 50 @11:25:22.59 Air.reconfigure :default @11:25:23.30 Lysis2.seek :home @11:25:23.34 Lysis1.seek :home @11:25:23.48 Archive.seek :home @11:25:24.19 Toroid.setRawPosition! 0 @11:25:24.21 Plunge.reconfigure :default @11:25:24.27 Lysis1.reconfigure :default @11:25:24.46 Lysis2.reconfigure :default @11:25:24.78 Toroid.seek :home @11:25:25.29 Toroid.reconfigure :default @11:25:25.37 Archive.reconfigure :default @11:25:26.30 Toroid.seek :clear @11:25:28.04 Carousel.reconfigure :default @11:25:28.12 Carousel.home.jog 0 @11:25:28.65 Carousel.seek 51 @11:25:28.84 Selecting Cartridge 50 @11:25:29.09 Carousel.seek 50 @11:25:29.64 Lysis1.seek :locked @11:25:33.28 Plunge.seek :locked @11:25:33.79 Air.seek :locked @11:25:38.10 Mounted archive Cartridge @11:25:38.43 Duration of filtering limited to 4:05:00 @11:25:38.75 Priming sample loop w/160ml, bypass w/10ml @11:25:38.77 SamplePump.setPosition! 0ml @11:25:38.84 Bypass.open @11:25:40.37 Intake.open @11:25:41.20 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @11:25:43.82 Vacuum tare is 18% of full scale @11:25:43.86 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @11:25:44.43 Pressure.reconfigure :default @11:25:44.63 Pressure.hold @11:25:45.43 SamplePump.reconfigure :prime @11:25:45.60 SamplePump.seek 10ml,1:21 @11:25:56.80 Pressure.coast @11:25:56.87 SamplePump.setPosition! 0ml @11:25:56.95 Exhaust.open @11:25:58.35 Bypass.close @11:25:59.93 Pressure.reconfigure :default @11:26:00.53 Pressure.hold @11:26:00.83 SamplePump.reconfigure :prime @11:26:00.99 SamplePump.seek 160ml,13:51 @11:28:52.64 Pressure.coast @11:28:52.72 Bypass.open @11:28:53.40 Intake.close @11:28:54.23 Toroid.reconfigure :hold @11:28:54.31 Toroid.seek :crack @11:28:56.09 Toroid.seek :cartridge @11:28:57.40 Toroid.reconfigure :default @11:28:57.52 Toroid.seek :clear @11:28:59.80 Pressurizing cartridge to 14psi above 14.6psia @11:28:59.85 SamplePump.setPosition! 0ml @11:29:00.00 Intake.open @11:29:00.80 Bypass.close @11:29:01.59 Exhaust.close @11:29:02.44 Pressure.reconfigure :default @11:29:02.52 Pressure.hold @11:29:02.75 SamplePump.reconfigure :charge @11:29:02.84 SamplePump.seek 2.3ml,42 seconds @11:29:10.81 Pressure.coast @11:29:10.89 Intake.close @11:29:11.67 Pressurized cartridge with 0.286ml to 14.1psi @11:29:11.69 Isolated @26.0psia (11.4psi) @11:29:11.72 Checking for cartridge leaks from 26.0psia @11:29:11.77 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @11:29:16.52 Cartridge pressure fell 1% to 25.6psia @11:29:16.79 Sampling 3000.0ml @11:29:16.80 SamplePump.setPosition! 0ml @11:29:17.05 Exhaust.open @11:29:17.84 Intake.open @11:29:19.03 Pressure.reconfigure :default @11:29:19.15 Pressure.hold @11:29:19.97 SamplePump.reconfigure :sample @11:29:20.06 SamplePump.seek 3000ml,4:10:32 @11:29:21.93 Cartridge 50 sampling 1/3000ml at 0.53ml/s with 15.0psi vacuum @11:30:22.19 Cartridge 50 sampling 31/3000ml at 0.51ml/s with 13.9psi vacuum @11:31:21.93 Cartridge 50 sampling 64/3000ml at 0.55ml/s with 15.1psi vacuum @11:32:22.14 Cartridge 50 sampling 97/3000ml at 0.57ml/s with 16.2psi vacuum @11:33:21.93 Cartridge 50 sampling 131/3000ml at 0.56ml/s with 14.1psi vacuum @11:34:22.09 Cartridge 50 sampling 165/3000ml at 0.58ml/s with 14.3psi vacuum @11:35:21.93 Cartridge 50 sampling 199/3000ml at 0.57ml/s with 13.1psi vacuum @11:36:22.03 Cartridge 50 sampling 233/3000ml at 0.56ml/s with 14.2psi vacuum @11:37:21.93 Cartridge 50 sampling 267/3000ml at 0.56ml/s with 14.9psi vacuum @11:38:22.28 Cartridge 50 sampling 301/3000ml at 0.56ml/s with 12.7psi vacuum @11:39:21.93 Cartridge 50 sampling 335/3000ml at 0.57ml/s with 12.3psi vacuum @11:40:21.93 Cartridge 50 sampling 368/3000ml at 0.55ml/s with 11.8psi vacuum @11:41:22.40 Cartridge 50 sampling 401/3000ml at 0.55ml/s with 13.6psi vacuum @11:42:21.93 Cartridge 50 sampling 435/3000ml at 0.56ml/s with 11.6psi vacuum @11:43:22.21 Cartridge 50 sampling 469/3000ml at 0.56ml/s with 12.8psi vacuum @11:44:21.93 Cartridge 50 sampling 503/3000ml at 0.56ml/s with 11.8psi vacuum @11:45:22.06 Cartridge 50 sampling 536/3000ml at 0.56ml/s with 14.7psi vacuum @11:46:21.94 Cartridge 50 sampling 570/3000ml at 0.56ml/s with 13.1psi vacuum @11:47:22.09 Cartridge 50 sampling 604/3000ml at 0.56ml/s with 11.2psi vacuum @11:48:21.93 Cartridge 50 sampling 638/3000ml at 0.57ml/s with 11.5psi vacuum @11:49:22.15 Cartridge 50 sampling 672/3000ml at 0.56ml/s with 13.9psi vacuum @11:50:21.93 Cartridge 50 sampling 706/3000ml at 0.56ml/s with 13.3psi vacuum @11:51:22.10 Cartridge 50 sampling 740/3000ml at 0.57ml/s with 13.3psi vacuum @11:52:21.93 Cartridge 50 sampling 775/3000ml at 0.57ml/s with 12.9psi vacuum @11:53:22.07 Cartridge 50 sampling 809/3000ml at 0.57ml/s with 13.6psi vacuum @11:54:21.94 Cartridge 50 sampling 843/3000ml at 0.56ml/s with 11.3psi vacuum @11:55:22.06 Cartridge 50 sampling 877/3000ml at 0.58ml/s with 14.5psi vacuum @11:56:21.94 Cartridge 50 sampling 912/3000ml at 0.57ml/s with 14.6psi vacuum @11:57:22.09 Cartridge 50 sampling 946/3000ml at 0.56ml/s with 12.2psi vacuum @11:58:21.93 Cartridge 50 sampling 980/3000ml at 0.57ml/s with 12.1psi vacuum @11:59:22.28 Cartridge 50 sampling 1014/3000ml at 0.56ml/s with 13.7psi vacuum @12:00:22.11 Cartridge 50 sampling 1048/3000ml at 0.57ml/s with 14.0psi vacuum @12:01:21.94 Cartridge 50 sampling 1082/3000ml at 0.55ml/s with 13.4psi vacuum @12:02:22.31 Cartridge 50 sampling 1116/3000ml at 0.57ml/s with 11.9psi vacuum @12:03:21.93 Cartridge 50 sampling 1150/3000ml at 0.56ml/s with 14.2psi vacuum @12:04:22.09 Cartridge 50 sampling 1184/3000ml at 0.57ml/s with 11.7psi vacuum @12:05:21.95 Cartridge 50 sampling 1218/3000ml at 0.56ml/s with 11.8psi vacuum @12:06:22.22 Cartridge 50 sampling 1252/3000ml at 0.56ml/s with 14.1psi vacuum @12:07:21.94 Cartridge 50 sampling 1287/3000ml at 0.57ml/s with 13.9psi vacuum @12:08:22.16 Cartridge 50 sampling 1321/3000ml at 0.56ml/s with 14.6psi vacuum @12:09:21.93 Cartridge 50 sampling 1356/3000ml at 0.56ml/s with 12.0psi vacuum @12:10:22.17 Cartridge 50 sampling 1390/3000ml at 0.57ml/s with 13.1psi vacuum @12:11:21.93 Cartridge 50 sampling 1425/3000ml at 0.55ml/s with 11.6psi vacuum @12:12:22.17 Cartridge 50 sampling 1458/3000ml at 0.57ml/s with 12.4psi vacuum @12:13:21.94 Cartridge 50 sampling 1493/3000ml at 0.57ml/s with 13.8psi vacuum @12:14:22.17 Cartridge 50 sampling 1527/3000ml at 0.56ml/s with 11.7psi vacuum @12:15:21.93 Cartridge 50 sampling 1561/3000ml at 0.56ml/s with 12.5psi vacuum @12:16:22.10 Cartridge 50 sampling 1595/3000ml at 0.56ml/s with 12.2psi vacuum @12:17:21.93 Cartridge 50 sampling 1629/3000ml at 0.57ml/s with 12.1psi vacuum @12:18:22.09 Cartridge 50 sampling 1663/3000ml at 0.57ml/s with 12.1psi vacuum @12:19:21.93 Cartridge 50 sampling 1697/3000ml at 0.57ml/s with 13.7psi vacuum @12:20:21.97 Cartridge 50 sampling 1732/3000ml at 0.56ml/s with 12.2psi vacuum @12:21:21.94 Cartridge 50 sampling 1766/3000ml at 0.56ml/s with 13.3psi vacuum @12:22:22.30 Cartridge 50 sampling 1800/3000ml at 0.56ml/s with 13.9psi vacuum @12:23:22.11 Cartridge 50 sampling 1834/3000ml at 0.56ml/s with 12.6psi vacuum @12:24:22.01 Cartridge 50 sampling 1868/3000ml at 0.58ml/s with 14.4psi vacuum @12:25:22.07 Cartridge 50 sampling 1902/3000ml at 0.56ml/s with 12.3psi vacuum @12:26:21.94 Cartridge 50 sampling 1936/3000ml at 0.58ml/s with 13.2psi vacuum @12:27:22.25 Cartridge 50 sampling 1970/3000ml at 0.56ml/s with 14.5psi vacuum @12:28:21.93 Cartridge 50 sampling 2004/3000ml at 0.57ml/s with 14.4psi vacuum @12:29:22.12 Cartridge 50 sampling 2038/3000ml at 0.57ml/s with 11.7psi vacuum @12:30:21.94 Cartridge 50 sampling 2072/3000ml at 0.56ml/s with 12.1psi vacuum @12:31:22.10 Cartridge 50 sampling 2107/3000ml at 0.57ml/s with 13.8psi vacuum @12:32:21.94 Cartridge 50 sampling 2141/3000ml at 0.56ml/s with 12.4psi vacuum @12:33:22.10 Cartridge 50 sampling 2176/3000ml at 0.57ml/s with 13.8psi vacuum @12:34:21.93 Cartridge 50 sampling 2210/3000ml at 0.56ml/s with 12.0psi vacuum @12:35:22.12 Cartridge 50 sampling 2244/3000ml at 0.57ml/s with 12.8psi vacuum @12:36:21.93 Cartridge 50 sampling 2278/3000ml at 0.56ml/s with 13.7psi vacuum @12:37:22.08 Cartridge 50 sampling 2312/3000ml at 0.57ml/s with 14.8psi vacuum @12:38:21.94 Cartridge 50 sampling 2346/3000ml at 0.56ml/s with 12.8psi vacuum @12:39:22.09 Cartridge 50 sampling 2380/3000ml at 0.57ml/s with 13.7psi vacuum @12:40:21.93 Cartridge 50 sampling 2414/3000ml at 0.56ml/s with 13.0psi vacuum @12:41:22.08 Cartridge 50 sampling 2448/3000ml at 0.58ml/s with 12.4psi vacuum @12:42:21.93 Cartridge 50 sampling 2482/3000ml at 0.56ml/s with 13.1psi vacuum @12:43:22.22 Cartridge 50 sampling 2516/3000ml at 0.56ml/s with 12.0psi vacuum @12:44:22.14 Cartridge 50 sampling 2550/3000ml at 0.56ml/s with 11.7psi vacuum @12:45:22.26 Cartridge 50 sampling 2584/3000ml at 0.57ml/s with 14.2psi vacuum @12:46:22.14 Cartridge 50 sampling 2619/3000ml at 0.58ml/s with 13.6psi vacuum @12:47:22.21 Cartridge 50 sampling 2653/3000ml at 0.57ml/s with 12.4psi vacuum @12:48:22.16 Cartridge 50 sampling 2688/3000ml at 0.59ml/s with 12.2psi vacuum @12:49:21.93 Cartridge 50 sampling 2722/3000ml at 0.57ml/s with 13.8psi vacuum @12:50:22.40 Cartridge 50 sampling 2757/3000ml at 0.57ml/s with 14.9psi vacuum @12:51:21.93 Cartridge 50 sampling 2791/3000ml at 0.57ml/s with 13.3psi vacuum @12:52:22.30 Cartridge 50 sampling 2825/3000ml at 0.56ml/s with 14.9psi vacuum @12:53:21.93 Cartridge 50 sampling 2860/3000ml at 0.57ml/s with 12.3psi vacuum @12:54:22.10 Cartridge 50 sampling 2894/3000ml at 0.57ml/s with 14.1psi vacuum @12:55:21.93 Cartridge 50 sampling 2928/3000ml at 0.57ml/s with 13.6psi vacuum @12:56:22.04 Cartridge 50 sampling 2962/3000ml at 0.57ml/s with 11.8psi vacuum @12:57:21.93 Cartridge 50 sampling 2996/3000ml at 0.57ml/s with 13.5psi vacuum @12:57:28.52 Pressure.coast @12:57:28.81 Intake.close @12:57:28.91 Exhaust.close @12:57:29.30 Toroid.reconfigure :hold @12:57:29.52 Toroid.seek :bypass @12:57:31.27 Toroid.reconfigure :default @12:57:31.35 Toroid.seek :clear @12:57:32.98 Sampled 3000.0ml @12:57:33.51 Plunge.seek :reagent @12:57:34.46 Sample loop closed. Ready to Deliver Reagents @12:57:35.39 Carousel.seek 50 @12:57:35.95 Plunge.seek :locked @12:57:37.41 Air.seek :clear @12:57:37.84 Lysis1.seek :clear @12:57:41.20 Archive.coast @12:57:41.27 Air.coast @12:57:41.34 Lysis1.coast @12:57:41.41 Lysis2.coast @12:57:41.47 Plunge.coast @12:57:41.54 SamplePump.coast @12:57:41.61 Toroid.coast @12:57:41.68 Carousel.coast @12:57:41.89 Gate.power :motor,:OFF @12:57:41.91 Gate.power :core,:OFF @12:57:41.94 Gate.power :analytic1,:OFF @12:57:41.97 Gate.power :analytic2,:OFF @12:57:42.00 Gate.power :analytic3,:OFF @12:57:42.03 Gate.power :main,:OFF @12:57:42.06 Safely stopped and ready to power off @09:42:18.33PDT30-May-21 -> Cartridge.reserved @09:42:27.84 -> Cartridge.release 2..3 @09:42:31.56 -> Cartridge.dry @09:43:03.36 -> BACsampler.vacuum @09:43:08.91 -> BACsampler.vacuum 18..26 @09:43:08.96 ArgumentError in preston -- wrong number of arguments (1 for 0) @09:43:14.32 -> BACsampler.vacuum=18..26 @09:43:18.65 -> BACsampler.vacuum @09:43:23.84 -> BACsampler.goalVolume @09:43:45.29 -> BACsampler.goalVolume=1000 @09:50:00.02 -> 2.times {delay '2:00:00';shoot} @09:50:00.05 delay 2:00:00 @11:50:00.21 Gate.power :main,:ON,limit to 5000mA @11:50:00.23 Gate.power :analytic1,:ON,limit to 5000mA @11:50:00.26 Gate.power :analytic2,:ON,limit to 5000mA @11:50:00.30 Gate.power :analytic3,:ON,limit to 5000mA @11:50:00.38 Powering core bus @11:50:03.19 Archive.configure :default @11:50:03.52 Air.configure :default @11:50:03.99 Lysis1.configure :default @11:50:04.29 Lysis2.configure :default @11:50:04.70 Plunge.configure :default @11:50:05.11 SamplePump.configure :default @11:50:05.59 ExtraPump.configure :default @11:50:05.95 Toroid.configure :default @11:50:06.27 Carousel.configure :default @11:50:06.59 Pressure.configure :default @11:50:06.74 All elf controllers running firmware version 4.61 @11:50:06.85 Gate.power :motor,:ON,limit to 5000mA @11:50:08.03 Archive.reconfigure :default @11:50:08.52 Archive.home.jog 0 @11:50:08.77 Plunge.reconfigure :default @11:50:08.82 Lysis2.reconfigure :default @11:50:08.86 Lysis1.reconfigure :default @11:50:08.91 Air.reconfigure :default @11:50:08.96 Toroid.reconfigure :default @11:50:09.23 Lysis2.home.jog 0 @11:50:09.30 Plunge.home.jog 0 @11:50:09.33 Air.home.jog 0 @11:50:09.38 Toroid.home.jog 0 @11:50:09.42 Lysis1.home.jog 0 @11:50:10.03 Archive.reconfigure :homing @11:50:10.24 Archive.jog 1363 @11:50:10.30 Toroid.reconfigure :homing @11:50:10.35 Plunge.reconfigure :homing @11:50:10.40 Lysis1.reconfigure :homing @11:50:10.46 Air.reconfigure :homing @11:50:10.53 Lysis2.reconfigure :homing @11:50:10.79 Toroid.jog 630 @11:50:10.84 Plunge.jog 857 @11:50:10.88 Lysis1.jog 1352 @11:50:10.94 Air.jog 1369 @11:50:11.04 Lysis2.jog 1360 @11:50:11.74 Archive.setRawPosition! 50 @11:50:11.88 Archive.seek :home @11:50:12.17 Plunge.setRawPosition! 50 @11:50:12.50 Plunge.seek :home @11:50:13.06 Lysis1.setRawPosition! 50 @11:50:13.18 Air.setRawPosition! 50 @11:50:13.26 Archive.reconfigure :default @11:50:13.36 Lysis2.setRawPosition! 50 @11:50:13.71 Lysis1.seek :home @11:50:14.00 Air.seek :home @11:50:14.38 Lysis2.seek :home @11:50:15.24 Toroid.setRawPosition! 0 @11:50:15.37 Plunge.reconfigure :default @11:50:15.46 Lysis1.reconfigure :default @11:50:15.51 Lysis2.reconfigure :default @11:50:15.64 Air.reconfigure :default @11:50:16.21 Toroid.seek :home @11:50:16.64 Toroid.reconfigure :default @11:50:17.00 Toroid.seek :clear @11:50:18.79 Carousel.reconfigure :default @11:50:18.88 Carousel.home.jog 0 @11:50:19.41 Carousel.seek 50 @11:50:19.60 Selecting Cartridge 3 @11:50:19.85 Carousel.seek 3 @11:50:29.82 Lysis1.seek :locked @11:50:33.36 Plunge.seek :locked @11:50:33.86 Air.seek :locked @11:50:38.13 Mounted archive Cartridge @11:50:38.45 Duration of filtering limited to 4:05:00 @11:50:38.77 Priming sample loop w/160ml, bypass w/10ml @11:50:38.80 SamplePump.setPosition! 0ml @11:50:38.88 Bypass.open @11:50:40.41 Intake.open @11:50:41.56 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @11:50:44.20 Vacuum tare is 18% of full scale @11:50:44.27 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @11:50:45.01 Pressure.reconfigure :default @11:50:45.22 Pressure.hold @11:50:45.68 SamplePump.reconfigure :prime @11:50:45.87 SamplePump.seek 10ml,1:21 @11:50:57.01 Pressure.coast @11:50:57.09 SamplePump.setPosition! 0ml @11:50:57.16 Exhaust.open @11:50:58.63 Bypass.close @11:51:00.64 Pressure.reconfigure :default @11:51:00.80 Pressure.hold @11:51:01.94 SamplePump.reconfigure :prime @11:51:02.10 SamplePump.seek 160ml,13:51 @11:53:53.91 Pressure.coast @11:53:53.99 Bypass.open @11:53:54.73 Intake.close @11:53:55.91 Toroid.reconfigure :hold @11:53:56.08 Toroid.seek :crack @11:53:57.94 Toroid.seek :cartridge @11:53:59.10 Toroid.reconfigure :default @11:53:59.34 Toroid.seek :clear @11:54:01.20 Pressurizing cartridge to 26psi above 14.1psia @11:54:01.21 SamplePump.setPosition! 0ml @11:54:01.29 Intake.open @11:54:02.10 Bypass.close @11:54:02.83 Exhaust.close @11:54:03.69 Pressure.reconfigure :default @11:54:04.31 Pressure.hold @11:54:04.57 SamplePump.reconfigure :charge @11:54:04.66 SamplePump.seek 2.3ml,42 seconds @11:54:19.39 Pressure.coast @11:54:19.47 Intake.close @11:54:20.35 Pressurized cartridge with 0.990ml to 26.1psi @11:54:20.36 Isolated @35.7psia (21.6psi) @11:54:20.39 Checking for cartridge leaks from 35.7psia @11:54:20.45 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @11:54:26.60 Cartridge pressure fell 13% to 31.2psia @11:54:26.86 Sampling 1000.0ml @11:54:26.87 SamplePump.setPosition! 0ml @11:54:27.11 Exhaust.open @11:54:27.86 Intake.open @11:54:28.90 Pressure.reconfigure :default @11:54:28.98 Pressure.hold @11:54:29.30 SamplePump.reconfigure :sample @11:54:29.39 SamplePump.seek 1000ml,1:23:51 @11:54:31.95 Cartridge 3 sampling 1/1000ml at 0.60ml/s with 24.9psi vacuum @11:55:32.23 Cartridge 3 sampling 31/1000ml at 0.49ml/s with 24.0psi vacuum @11:56:31.96 Cartridge 3 sampling 61/1000ml at 0.53ml/s with 23.2psi vacuum @11:57:32.06 Cartridge 3 sampling 92/1000ml at 0.52ml/s with 22.0psi vacuum @11:58:31.95 Cartridge 3 sampling 121/1000ml at 0.46ml/s with 22.0psi vacuum @11:59:32.11 Cartridge 3 sampling 148/1000ml at 0.45ml/s with 21.3psi vacuum @12:00:31.95 Cartridge 3 sampling 174/1000ml at 0.41ml/s with 20.5psi vacuum @12:01:32.09 Cartridge 3 sampling 198/1000ml at 0.38ml/s with 22.2psi vacuum @12:02:31.95 Cartridge 3 sampling 220/1000ml at 0.35ml/s with 25.5psi vacuum @12:03:32.05 Cartridge 3 sampling 240/1000ml at 0.32ml/s with 24.6psi vacuum @12:04:31.95 Cartridge 3 sampling 258/1000ml at 0.30ml/s with 26.0psi vacuum @12:05:32.16 Cartridge 3 sampling 276/1000ml at 0.26ml/s with 19.7psi vacuum @12:06:32.15 Cartridge 3 sampling 291/1000ml at 0.21ml/s with 25.0psi vacuum @12:07:04.30 Flow averaged 0.40ml/s over 12:34 @12:07:04.35 Waiting up to 19 seconds for vacuum to reduce from 30.2 to 18.0psi @12:07:05.45 Relieved excess vacuum in 3.1 seconds @12:07:05.48 SamplePump.seek 1000ml,59:03 @12:07:11.81 Flow estimate reduced to 0.32ml/s @12:07:11.89 Waiting up to 21 seconds for vacuum to reduce from 31.3 to 18.0psi @12:07:12.82 Relieved excess vacuum in 4.2 seconds @12:07:12.91 SamplePump.seek 1000ml,58:57 @12:07:15.24 Flow estimate reduced to 0.25ml/s @12:07:15.29 Waiting up to 14 seconds for vacuum to reduce from 28.5 to 18.0psi @12:07:16.63 Relieved excess vacuum in 3.5 seconds @12:07:16.65 SamplePump.seek 1000ml,58:54 @12:07:19.10 Flow estimate reduced to 0.20ml/s @12:07:19.37 Waiting up to 16 seconds for vacuum to reduce from 29.1 to 18.0psi @12:07:19.99 Relieved excess vacuum in 3.1 seconds @12:07:20.16 SamplePump.seek 1000ml,58:52 @12:07:23.38 Flow estimate reduced to 0.16ml/s @12:07:23.40 Intake.close @12:07:23.44 Exhaust.close @12:07:23.72 Toroid.reconfigure :hold @12:07:23.81 Toroid.seek :bypass @12:07:25.45 Toroid.reconfigure :default @12:07:25.53 Toroid.seek :clear @12:07:27.17 Cartridge::Sampler::Clogged in FILTERING -- Cartridge clogged @12:07:27.19 Pressure.coast @12:07:27.45 Sampled 300.4ml @12:07:28.69 SamplePump.setPosition! 0ml @12:07:28.82 Plunge.seek :reagent @12:07:29.14 Exhaust.open @12:07:30.04 Intake.open @12:07:30.27 Sample loop closed. Ready to Deliver Reagents @12:07:30.34 Archive.seek :sealed @12:07:31.20 SamplePump.reconfigure :slow @12:07:31.34 SamplePump.seek -35ml @12:07:34.09 Archive.seek :evac @12:07:47.80 Slide::Error in PROCESSING -- Archive Syringe (ArS) positionErr at 28uL (sensed 57uL) @12:07:47.90 Retry #1 of 2 @12:07:47.95 Archive.hold @12:07:48.19 Archive.reconfigure :default @12:07:48.37 Archive.home.seek :evac @12:07:49.18 Slide::Error in PROCESSING -- Archive Syringe (ArS) positionErr at 26uL (sensed 56uL) @12:07:49.20 Retry #2 of 2 @12:07:49.27 Archive.reconfigure :default @12:07:49.47 Archive.hold @12:07:49.70 Archive.reconfigure :default @12:07:49.86 Archive.home.seek :evac @12:07:50.76 Slide::Error in PROCESSING -- IGNORED Archive Syringe (ArS) positionErr at 24uL (sensed 55uL) @12:07:50.79 Archive.reconfigure :default @12:07:51.01 Archive.hold @12:07:51.14 Archive.reconfigure :default @12:07:51.35 Archive.home.jog 0 @12:07:51.56 delay 1:00 @12:08:46.56 delay 1 second @12:08:47.58 Bypass.open @12:08:48.31 Exhaust.close @12:08:49.09 SamplePump.seek -33ml @12:08:51.63 Archive.seek :clear @12:08:53.65 delay 1 second @12:08:54.80 SamplePump.seek -41ml @12:09:09.91 RNAlater soak for 10:00 @12:09:12.32 delay 1 second @12:09:13.52 Intake.close @12:09:13.71 Bypass.close @12:19:09.96 Archive.seek :tVent @12:19:11.80 delay 1 second @12:19:12.94 Archive.seek :evac @12:19:28.70 Slide::Error in PROCESSING -- Archive Syringe (ArS) positionErr at 24uL (sensed 54uL) @12:19:28.76 Retry #1 of 2 @12:19:28.81 Archive.hold @12:19:29.28 Archive.reconfigure :default @12:19:29.47 Archive.home.seek :evac @12:19:30.29 Slide::Error in PROCESSING -- Archive Syringe (ArS) positionErr at 22uL (sensed 53uL) @12:19:30.31 Retry #2 of 2 @12:19:30.36 Archive.reconfigure :default @12:19:30.55 Archive.hold @12:19:30.85 Archive.reconfigure :default @12:19:30.98 Archive.home.seek :evac @12:19:31.88 Slide::Error in PROCESSING -- IGNORED Archive Syringe (ArS) positionErr at 21uL (sensed 52uL) @12:19:31.91 Archive.reconfigure :default @12:19:32.10 Archive.hold @12:19:32.38 Archive.reconfigure :default @12:19:32.51 Archive.home.jog 0 @12:19:32.78 delay 10 seconds @12:19:42.83 Sample preserved and evacuated @12:19:43.27 Carousel.seek 3 @12:19:43.51 Plunge.seek :locked @12:19:44.52 Archive.seek :clear @12:19:44.69 Air.seek :clear @12:19:44.88 Lysis1.seek :clear @12:20:01.32 Archive.coast @12:20:01.48 Air.coast @12:20:01.62 Lysis1.coast @12:20:01.69 Lysis2.coast @12:20:01.77 Plunge.coast @12:20:01.84 SamplePump.coast @12:20:01.91 Toroid.coast @12:20:01.98 Carousel.coast @12:20:02.20 Gate.power :motor,:OFF @12:20:02.23 Gate.power :core,:OFF @12:20:02.26 Gate.power :analytic1,:OFF @12:20:02.29 Gate.power :analytic2,:OFF @12:20:02.32 Gate.power :analytic3,:OFF @12:20:02.35 Gate.power :main,:OFF @12:20:02.39 Safely stopped and ready to power off @12:20:02.59 delay 2:00:00 @13:25:53.08 UserAbort in preston -- @13:27:30.89 -> MainThread.exit @13:27:31.13 Session Ended @10:42:16.35PDT02-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @10:42:16.38 >BREAK< ESP3 Gateway v1.11 for cartridge bus @10:42:20.30 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @10:42:20.71 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @10:42:26.79 Polling Can every 2:00 @10:42:26.92 Gate.power :analytic1,:ON,limit to 5000mA @10:42:26.93 Gate.power :analytic2,:ON,limit to 5000mA @10:42:26.95 Gate.power :analytic3,:ON,limit to 5000mA @10:42:27.26 Archive.configure :default @10:42:27.35 Archive.hold @10:42:27.42 Archive.configure :default @10:42:27.58 Air.configure :default @10:42:27.66 Air.hold @10:42:27.74 Air.configure :default @10:42:27.88 Lysis1.configure :default @10:42:27.96 Lysis1.hold @10:42:28.03 Lysis1.configure :default @10:42:28.18 Lysis2.configure :default @10:42:28.27 Lysis2.hold @10:42:28.34 Lysis2.configure :default @10:42:28.53 Plunge.configure :default @10:42:28.58 Plunge.hold @10:42:28.69 Plunge.configure :default @10:42:28.85 SamplePump.configure :default @10:42:29.00 ExtraPump.configure :default @10:42:29.22 Toroid.configure :default @10:42:29.27 Toroid.hold @10:42:29.36 Toroid.configure :default @10:42:29.55 Carousel.configure :default @10:42:29.63 Carousel.hold @10:42:29.68 Carousel.configure :default @10:42:29.83 Pressure.configure :default @10:42:29.94 All elf controllers running firmware version 4.61 @10:42:30.03 Intake.close @10:42:30.08 Exhaust.close @10:42:30.20 Bypass.close @10:42:30.44
Loading /home/esp/esp2/bin/processAll mission @10:42:30.86 Sampling default goal volumes @10:42:30.88 Cartridge group: dry Spares: {spare: 2} @10:42:31.70 Carousel.seek 3 @10:42:31.93 Selecting Cartridge 2 @10:42:32.22 Carousel.seek 2 @10:42:32.80 Lysis1.seek :locked @10:42:36.05 Plunge.seek :locked @10:42:36.22 Air.seek :locked @10:42:40.48
Mounted archive Cartridge @10:42:40.87 Duration of filtering limited to 4:05:00 @10:42:41.26 Priming sample loop w/160ml, bypass w/10ml @10:42:41.28 SamplePump.setPosition! 0ml @10:42:41.32 Bypass.open @10:42:42.94 Intake.open @10:42:43.87 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @10:42:46.48 Vacuum tare is 17% of full scale @10:42:46.54 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @10:42:46.80 Pressure.reconfigure :default @10:42:46.86 Pressure.hold @10:42:47.14 SamplePump.reconfigure :prime @10:42:47.25 SamplePump.seek 10ml,1:21 @10:42:58.36 Pressure.coast @10:42:58.45 SamplePump.setPosition! 0ml @10:42:58.49 Exhaust.open @10:43:00.56 Bypass.close @10:43:02.19 Pressure.reconfigure :default @10:43:02.25 Pressure.hold @10:43:02.62 SamplePump.reconfigure :prime @10:43:02.68 SamplePump.seek 160ml,13:51 @10:44:19.55 Session Begun @10:44:21.76 -> showlog @10:45:54.31 Pressure.coast @10:45:54.43 Bypass.open @10:45:55.20 Intake.close @10:45:56.18 Toroid.reconfigure :hold @10:45:56.30 Toroid.seek :crack @10:45:57.96 Toroid.seek :cartridge @10:45:58.91 Toroid.reconfigure :default @10:45:59.03 Toroid.seek :clear @10:46:01.02 Pressurizing cartridge to 26psi above 14.7psia @10:46:01.04 SamplePump.setPosition! 0ml @10:46:01.12 Intake.open @10:46:01.92 Bypass.close @10:46:02.67 Exhaust.close @10:46:03.58 Pressure.reconfigure :default @10:46:03.70 Pressure.hold @10:46:04.01 SamplePump.reconfigure :charge @10:46:04.15 SamplePump.seek 2.3ml,42 seconds @10:46:19.13 Pressure.coast @10:46:19.23 Intake.close @10:46:20.08 Pressurized cartridge with 1.026ml to 26.0psi @10:46:20.12 Isolated @36.3psia (21.6psi) @10:46:20.15 Checking for cartridge leaks from 36.3psia @10:46:20.19 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @10:46:24.82 Cartridge pressure fell 3% to 35.2psia @10:46:24.96 Sampling 1000.0ml @10:46:24.99 SamplePump.setPosition! 0ml @10:46:25.07 Exhaust.open @10:46:25.80 Intake.open @10:46:26.84 Pressure.reconfigure :default @10:46:26.98 Pressure.hold @10:46:27.38 SamplePump.reconfigure :sample @10:46:27.48 SamplePump.seek 1000ml,1:23:51 @10:46:29.91 Cartridge 2 sampling 1/1000ml at 0.57ml/s with 25.8psi vacuum @10:47:29.90 Cartridge 2 sampling 31/1000ml at 0.51ml/s with 24.2psi vacuum @10:48:29.90 Cartridge 2 sampling 62/1000ml at 0.51ml/s with 24.9psi vacuum @10:49:29.90 Cartridge 2 sampling 93/1000ml at 0.51ml/s with 23.6psi vacuum @10:50:29.91 Cartridge 2 sampling 123/1000ml at 0.53ml/s with 25.1psi vacuum @10:51:29.92 Cartridge 2 sampling 154/1000ml at 0.52ml/s with 24.8psi vacuum @10:52:29.90 Cartridge 2 sampling 186/1000ml at 0.52ml/s with 26.3psi vacuum @10:53:29.90 Cartridge 2 sampling 217/1000ml at 0.52ml/s with 22.6psi vacuum @10:54:29.91 Cartridge 2 sampling 249/1000ml at 0.56ml/s with 24.7psi vacuum @10:55:29.91 Cartridge 2 sampling 282/1000ml at 0.54ml/s with 23.5psi vacuum @10:56:29.90 Cartridge 2 sampling 315/1000ml at 0.54ml/s with 22.6psi vacuum @10:57:29.90 Cartridge 2 sampling 348/1000ml at 0.55ml/s with 26.2psi vacuum @10:58:29.91 Cartridge 2 sampling 381/1000ml at 0.54ml/s with 20.6psi vacuum @10:59:29.92 Cartridge 2 sampling 414/1000ml at 0.55ml/s with 26.7psi vacuum @11:00:29.90 Cartridge 2 sampling 447/1000ml at 0.54ml/s with 24.8psi vacuum @11:01:29.91 Cartridge 2 sampling 480/1000ml at 0.53ml/s with 20.3psi vacuum @11:02:29.92 Cartridge 2 sampling 513/1000ml at 0.54ml/s with 19.6psi vacuum @11:03:29.90 Cartridge 2 sampling 545/1000ml at 0.54ml/s with 21.0psi vacuum @11:04:29.92 Cartridge 2 sampling 578/1000ml at 0.54ml/s with 22.7psi vacuum @11:05:29.90 Cartridge 2 sampling 611/1000ml at 0.57ml/s with 27.7psi vacuum @11:06:29.91 Cartridge 2 sampling 645/1000ml at 0.54ml/s with 28.7psi vacuum @11:07:29.91 Cartridge 2 sampling 678/1000ml at 0.54ml/s with 28.3psi vacuum @11:08:29.93 Cartridge 2 sampling 712/1000ml at 0.56ml/s with 21.6psi vacuum @11:09:29.90 Cartridge 2 sampling 745/1000ml at 0.54ml/s with 18.9psi vacuum @11:10:29.91 Cartridge 2 sampling 778/1000ml at 0.55ml/s with 19.7psi vacuum @11:11:29.90 Cartridge 2 sampling 812/1000ml at 0.56ml/s with 27.5psi vacuum @11:12:29.92 Cartridge 2 sampling 845/1000ml at 0.57ml/s with 21.5psi vacuum @11:13:29.93 Cartridge 2 sampling 879/1000ml at 0.55ml/s with 27.2psi vacuum @11:14:29.91 Cartridge 2 sampling 912/1000ml at 0.56ml/s with 22.1psi vacuum @11:15:29.91 Cartridge 2 sampling 945/1000ml at 0.57ml/s with 21.8psi vacuum @11:16:29.92 Cartridge 2 sampling 978/1000ml at 0.56ml/s with 25.9psi vacuum @11:17:08.95 Pressure.coast @11:17:09.15 Intake.close @11:17:09.24 Exhaust.close @11:17:09.54 Toroid.reconfigure :hold @11:17:09.65 Toroid.seek :bypass @11:17:11.29 Toroid.reconfigure :default @11:17:11.42 Toroid.seek :clear @11:17:13.05 Sampled 1000.0ml @11:17:13.69 SamplePump.setPosition! 0ml @11:17:13.77 Plunge.seek :reagent @11:17:13.84 Exhaust.open @11:17:14.68 Intake.open @11:17:14.86 Sample loop closed. Ready to Deliver Reagents @11:17:14.88 Archive.seek :sealed @11:17:15.57 SamplePump.reconfigure :slow @11:17:15.65 SamplePump.seek -35ml @11:17:18.61 Archive.seek :evac @11:17:32.24 Slide::Error in PROCESSING -- Archive Syringe (ArS) positionErr at 24uL (sensed 54uL) @11:17:32.29 Retry #1 of 2 @11:17:32.32 Archive.hold @11:17:32.70 Archive.reconfigure :default @11:17:32.86 Archive.home.seek :evac @11:17:33.58 Slide::Error in PROCESSING -- Archive Syringe (ArS) positionErr at 23uL (sensed 53uL) @11:17:33.63 Retry #2 of 2 @11:17:33.66 Archive.reconfigure :default @11:17:33.80 Archive.hold @11:17:34.02 Archive.reconfigure :default @11:17:34.13 Archive.home.seek :evac @11:17:34.80 Slide::Error in PROCESSING -- IGNORED Archive Syringe (ArS) positionErr at 22uL (sensed 53uL) @11:17:34.82 Archive.reconfigure :default @11:17:34.95 Archive.hold @11:17:35.24 Archive.reconfigure :default @11:17:35.34 Archive.home.jog 0 @11:17:35.51 delay 1:00 @11:18:30.88 delay 1 second @11:18:31.90 Bypass.open @11:18:32.66 Exhaust.close @11:18:33.47 SamplePump.seek -33ml @11:18:35.55 Archive.seek :clear @11:18:38.07 delay 1 second @11:18:39.18 SamplePump.seek -41ml @11:18:52.23 RNAlater soak for 10:00 @11:18:56.63 delay 1 second @11:18:58.44 Intake.close @11:18:58.57 Bypass.close @11:28:52.27 Archive.seek :tVent @11:28:53.96 delay 1 second @11:28:55.01 Archive.seek :evac @11:29:11.00 delay 10 seconds @11:29:21.03 Sample preserved and evacuated @11:29:21.38
Carousel.seek 2 @11:29:21.68 Plunge.seek :locked @11:29:22.72 Archive.seek :clear @11:29:22.97 Air.seek :clear @11:29:23.21 Lysis1.seek :clear @11:29:39.97
Archive.coast @11:29:40.06 Air.coast @11:29:40.15 Lysis1.coast @11:29:40.27 Lysis2.coast @11:29:40.34 Plunge.coast @11:29:40.45 SamplePump.coast @11:29:40.54 Toroid.coast @11:29:40.62 Carousel.coast @11:29:40.92 Gate.power :motor,:OFF @11:29:40.96 Gate.power :core,:OFF @11:29:41.01 Gate.power :analytic1,:OFF @11:29:41.05 Gate.power :analytic2,:OFF @11:29:41.12 Gate.power :analytic3,:OFF @11:29:41.16 Gate.power :main,:OFF @11:29:41.20 Safely stopped and ready to power off @11:29:41.85 Cartridge Load: {58..53 => { Type: :archiveNull_bac, State: :wet}, [33, 32, 16..13, 11, 9..6, 4] => { Type: :archive_bac, State: :processed, filtered: 2000}, [39..37, 34] => { Type: :archive_bac, State: :processed, filtered: 1000}, 2 => { Type: :archive_bac, State: :processed, filtered: 1000.0}, 3 => { Type: :archive_bac, State: :processed, filtered: 300.4}, 5 => { Type: :archive_bac, State: :processed, filtered: 1833.7}, 10 => { Type: :archive_bac, State: :filtering, filtered: 2000}, 30 => { Type: :archiveNull_bac, State: :leaky}, 35 => { Type: :archive_bac, State: :processed, filtered: 647.2}, [43, 41] => { Type: :aRNAl_bac, State: :processed, filtered: 250}, [44, 42, 40] => { Type: :archive_bac, State: :processed, filtered: 250}, 28..19 => { Type: :archiveNull_bac, State: :processed, filtered: 2000}, 52..50 => { Type: :archiveNull_bac, State: :processed, filtered: 3000}, 12 => { Type: :archive_bac, State: :filtering, filtered: 151.5}, 29 => { Type: :archiveNull_bac, State: :leaky, filtered: 648.6}, 36 => { Type: :archive_bac, State: :processed, filtered: 382.5}, 45 => { Type: :aRNAl_bac, State: :processed, filtered: 297}} Spares: {spare: 2} @11:29:41.88 Cartridge::None in MAIN -- Not enough dry cartridges remain @11:29:42.16 Session Ended @15:52:13.56PDT04-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @15:52:13.59 >BREAK< ESP3 Gateway v1.11 for cartridge bus @15:52:17.50 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @15:52:17.88 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @15:52:23.25 Polling Can every 2:00 @15:52:23.38 Gate.power :analytic1,:ON,limit to 5000mA @15:52:23.39 Gate.power :analytic2,:ON,limit to 5000mA @15:52:23.40 Gate.power :analytic3,:ON,limit to 5000mA @15:52:23.69 Archive.configure :default @15:52:23.74 Archive.hold @15:52:23.85 Archive.configure :default @15:52:24.01 Air.configure :default @15:52:24.08 Air.hold @15:52:24.15 Air.configure :default @15:52:24.29 Lysis1.configure :default @15:52:24.37 Lysis1.hold @15:52:24.43 Lysis1.configure :default @15:52:24.59 Lysis2.configure :default @15:52:24.64 Lysis2.hold @15:52:24.74 Lysis2.configure :default @15:52:24.91 Plunge.configure :default @15:52:24.97 Plunge.hold @15:52:25.04 Plunge.configure :default @15:52:25.20 SamplePump.configure :default @15:52:25.36 ExtraPump.configure :default @15:52:25.52 Toroid.configure :default @15:52:25.60 Toroid.hold @15:52:25.65 Toroid.configure :default @15:52:25.86 Carousel.configure :default @15:52:25.91 Carousel.hold @15:52:26.00 Carousel.configure :default @15:52:26.16 Pressure.configure :default @15:52:26.24 All elf controllers running firmware version 4.61 @15:52:26.33 Intake.close @15:52:26.41 Exhaust.close @15:52:26.50 Bypass.close @15:52:26.73
Loading /home/esp/esp2/bin/processAll mission @15:52:27.06 Sampling default goal volumes @15:52:27.09 Cartridge group: dry Spares: {spare: 2} @15:52:27.94 Carousel.seek 6 @15:52:28.92 Mounted archive Cartridge @15:52:28.96 Selecting Cartridge 60 @15:52:29.41 Air.seek :clear @15:52:33.29 Lysis1.seek :clear @15:52:33.49 Plunge.seek :clear @15:52:36.49
Carousel.seek 60 @15:52:47.88 Lysis1.seek :locked @15:52:51.13 Plunge.seek :locked @15:52:51.30 Air.seek :locked @15:52:55.54
Mounted archive Cartridge @15:52:55.92 Duration of filtering limited to 4:05:00 @15:52:56.31 Priming sample loop w/160ml, bypass w/10ml @15:52:56.33 SamplePump.setPosition! 0ml @15:52:56.36 Bypass.open @15:52:57.89 Intake.open @15:52:58.80 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:53:01.40 Vacuum tare is 19% of full scale @15:53:01.44 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @15:53:01.91 Pressure.reconfigure :default @15:53:01.98 Pressure.hold @15:53:02.27 SamplePump.reconfigure :prime @15:53:02.33 SamplePump.seek 10ml,1:21 @15:53:13.46 Pressure.coast @15:53:13.52 SamplePump.setPosition! 0ml @15:53:13.61 Exhaust.open @15:53:15.67 Bypass.close @15:53:17.23 Pressure.reconfigure :default @15:53:17.30 Pressure.hold @15:53:17.66 SamplePump.reconfigure :prime @15:53:17.73 SamplePump.seek 160ml,13:51 @15:55:32.65 Session Begun @15:55:37.56 -> showlog @15:56:09.30 Pressure.coast @15:56:09.43 Bypass.open @15:56:10.17 Intake.close @15:56:11.14 Toroid.reconfigure :hold @15:56:11.27 Toroid.seek :crack @15:56:12.96 Toroid.seek :cartridge @15:56:13.97 Toroid.reconfigure :default @15:56:14.09 Toroid.seek :clear @15:56:15.99 Pressurizing cartridge to 26psi above 7.1psia @15:56:16.01 SamplePump.setPosition! 0ml @15:56:16.12 Intake.open @15:56:16.92 Bypass.close @15:56:17.67 Exhaust.close @15:56:18.46 Cartridge::Sampler::Warning in FILTERING -- Delta Pressure @-7.6psi @15:56:19.79 Intake.close @15:56:20.00 Toroid.reconfigure :hold @15:56:20.12 Toroid.seek :bypass @15:56:21.79 Toroid.reconfigure :default @15:56:21.91 Toroid.seek :clear @15:56:23.60 Cartridge::Sampler::Leaky in FILTERING -- Delta Pressure @-7.5psi indicates Sensor Failure or serious Leak @15:56:23.62 Retry #1 of 2 @15:56:23.67 Cmd::SpareCartridge in FILTERING -- 1 spare remaining @15:56:23.77 Cartridge::None in FILTERING -- No dry cartridges remain @15:56:23.78 Cartridge::None in FILTERING -- No dry cartridges remain @15:56:24.03
Cartridge Load: {59..1 => { Type: :archive_bac, State: :reserved, was: :dry}, 60 => { Type: :archive_bac, State: :leaky}} Spares: {spare: 1} @15:56:24.05 Thread::ChildDied in MAIN -- :FILTERING @15:56:24.28 Session Ended @16:02:23.67PDT04-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @16:02:23.70 >BREAK< ESP3 Gateway v1.11 for cartridge bus @16:02:27.52 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @16:02:27.87 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @16:02:33.35 Polling Can every 2:00 @16:02:33.47 Gate.power :analytic1,:ON,limit to 5000mA @16:02:33.49 Gate.power :analytic2,:ON,limit to 5000mA @16:02:33.50 Gate.power :analytic3,:ON,limit to 5000mA @16:02:33.80 Archive.configure :default @16:02:33.91 Archive.hold @16:02:33.99 Archive.configure :default @16:02:34.16 Air.configure :default @16:02:34.21 Air.hold @16:02:34.30 Air.configure :default @16:02:34.43 Lysis1.configure :default @16:02:34.52 Lysis1.hold @16:02:34.57 Lysis1.configure :default @16:02:34.73 Lysis2.configure :default @16:02:34.79 Lysis2.hold @16:02:34.87 Lysis2.configure :default @16:02:35.04 Plunge.configure :default @16:02:35.11 Plunge.hold @16:02:35.17 Plunge.configure :default @16:02:35.32 SamplePump.configure :default @16:02:35.53 ExtraPump.configure :default @16:02:35.70 Toroid.configure :default @16:02:35.78 Toroid.hold @16:02:35.84 Toroid.configure :default @16:02:36.00 Carousel.configure :default @16:02:36.08 Carousel.hold @16:02:36.18 Carousel.configure :default @16:02:36.34 Pressure.configure :default @16:02:36.40 All elf controllers running firmware version 4.61 @16:02:36.52 Intake.close @16:02:36.60 Exhaust.close @16:02:36.66 Bypass.close @16:02:36.91
Loading /home/esp/esp2/bin/processAll mission @16:02:37.21 Sampling default goal volumes @16:02:37.24 Cartridge group: dry Spares: {spare: 1} @16:02:38.10 Carousel.seek 60 @16:02:39.05 Mounted archive Cartridge @16:02:39.12 Selecting Cartridge 60 @16:02:39.93 Duration of filtering limited to 4:05:00 @16:02:40.31 Priming sample loop w/160ml, bypass w/10ml @16:02:40.33 SamplePump.setPosition! 0ml @16:02:40.36 Bypass.open @16:02:41.90 Intake.open @16:02:42.83 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @16:02:45.47 Vacuum tare is 15% of full scale @16:02:45.49 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -14psi @16:02:45.76 Pressure.reconfigure :default @16:02:45.83 Pressure.hold @16:02:46.13 SamplePump.reconfigure :prime @16:02:46.19 SamplePump.seek 10ml,1:21 @16:02:57.32 Pressure.coast @16:02:57.39 SamplePump.setPosition! 0ml @16:02:57.43 Exhaust.open @16:02:58.89 Bypass.close @16:03:00.48 Pressure.reconfigure :default @16:03:00.54 Pressure.hold @16:03:00.74 Sea 2.5m deep (18.4psia) @16:03:00.92 SamplePump.reconfigure :prime @16:03:01.01 SamplePump.seek 160ml,13:51 @16:05:52.52 Pressure.coast @16:05:52.58 Bypass.open @16:05:53.29 Intake.close @16:05:54.25 Toroid.reconfigure :hold @16:05:54.32 Toroid.seek :crack @16:05:55.93 Toroid.seek :cartridge @16:05:56.87 Toroid.reconfigure :default @16:05:56.96 Toroid.seek :clear @16:05:59.45 Pressurizing cartridge to 26psi above 17.0psia @16:05:59.47 SamplePump.setPosition! 0ml @16:05:59.56 Intake.open @16:06:00.37 Bypass.close @16:06:01.13 Exhaust.close @16:06:02.03 Pressure.reconfigure :default @16:06:02.09 Pressure.hold @16:06:02.37 SamplePump.reconfigure :charge @16:06:02.46 SamplePump.seek 2.3ml,42 seconds @16:06:09.33 Pressure.coast @16:06:09.37 Intake.close @16:06:10.31 Pressurized cartridge with 0.211ml to 26.3psi @16:06:10.32 Isolated @35.7psia (18.6psi) @16:06:10.34 Checking for cartridge leaks from 35.7psia @16:06:10.38 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @16:06:20.00 Cartridge pressure fell 14% to 30.7psia @16:06:20.09 Sampling 1000.0ml @16:06:20.11 SamplePump.setPosition! 0ml @16:06:20.17 Exhaust.open @16:06:20.87 Intake.open @16:06:22.06 Pressure.reconfigure :default @16:06:22.19 Pressure.hold @16:06:22.53 Sea 2.5m deep (18.4psia) @16:06:22.89 SamplePump.reconfigure :sample @16:06:23.06 SamplePump.seek 1000ml,1:23:51 @16:06:24.97 Cartridge 60 sampling 1/1000ml at 0.46ml/s with 24.6psi vacuum @16:06:30.78 Session Begun @16:06:33.56 -> showlog @16:07:24.97 Cartridge 60 sampling 28/1000ml at 0.43ml/s with 25.3psi vacuum @16:08:24.96 Cartridge 60 sampling 53/1000ml at 0.42ml/s with 19.8psi vacuum @16:09:24.96 Cartridge 60 sampling 77/1000ml at 0.39ml/s with 19.6psi vacuum @16:10:24.96 Cartridge 60 sampling 100/1000ml at 0.35ml/s with 20.9psi vacuum @16:11:24.97 Cartridge 60 sampling 121/1000ml at 0.33ml/s with 26.2psi vacuum @16:12:24.97 Cartridge 60 sampling 141/1000ml at 0.33ml/s with 23.0psi vacuum @16:13:24.96 Cartridge 60 sampling 161/1000ml at 0.32ml/s with 25.1psi vacuum @16:14:24.96 Cartridge 60 sampling 179/1000ml at 0.29ml/s with 22.1psi vacuum @16:15:25.01 Cartridge 60 sampling 196/1000ml at 0.27ml/s with 21.6psi vacuum @16:16:24.98 Cartridge 60 sampling 212/1000ml at 0.27ml/s with 27.9psi vacuum @16:17:24.96 Cartridge 60 sampling 228/1000ml at 0.24ml/s with 25.0psi vacuum @16:18:24.96 Cartridge 60 sampling 243/1000ml at 0.24ml/s with 26.5psi vacuum @16:19:24.98 Cartridge 60 sampling 257/1000ml at 0.26ml/s with 26.0psi vacuum @16:20:24.97 Cartridge 60 sampling 270/1000ml at 0.22ml/s with 26.1psi vacuum @16:21:24.97 Cartridge 60 sampling 283/1000ml at 0.20ml/s with 25.8psi vacuum @16:22:24.99 Cartridge 60 sampling 296/1000ml at 0.20ml/s with 27.2psi vacuum @16:23:24.98 Cartridge 60 sampling 309/1000ml at 0.21ml/s with 24.7psi vacuum @16:24:24.96 Cartridge 60 sampling 321/1000ml at 0.20ml/s with 28.5psi vacuum @16:25:24.96 Cartridge 60 sampling 334/1000ml at 0.20ml/s with 28.7psi vacuum @16:26:24.96 Cartridge 60 sampling 346/1000ml at 0.20ml/s with 29.4psi vacuum @16:27:24.98 Cartridge 60 sampling 358/1000ml at 0.20ml/s with 30.4psi vacuum @16:28:24.97 Cartridge 60 sampling 371/1000ml at 0.20ml/s with 30.9psi vacuum @16:29:24.99 Cartridge 60 sampling 383/1000ml at 0.21ml/s with 20.2psi vacuum 1.6m deep (17.1psia) @16:30:00.94 Flow averaged 0.28ml/s over 23:33 @16:30:00.99 Waiting up to 22 seconds for vacuum to reduce from 31.7 to 18.0psi @16:30:01.63 Relieved excess vacuum in 6.5 seconds @16:30:01.71 SamplePump.seek 1000ml,51:19 @16:30:24.95 Cartridge 60 sampling 395/1000ml at 0.20ml/s with 29.0psi vacuum @16:30:38.27 Flow averaged 0.26ml/s over 29 seconds @16:30:38.31 Waiting up to 24 seconds for vacuum to reduce from 32.5 to 18.0psi @16:30:38.98 Relieved excess vacuum in 8.6 seconds @16:30:39.04 SamplePump.seek 1000ml,50:42 @16:30:43.96 Flow estimate reduced to 0.21ml/s @16:30:44.05 Waiting up to 24 seconds for vacuum to reduce from 32.6 to 18.0psi @16:30:44.64 Relieved excess vacuum in 2.5 seconds @16:30:44.72 SamplePump.seek 1000ml,50:37 @16:30:51.15 Flow estimate reduced to 0.17ml/s @16:30:51.20 Intake.close @16:30:51.27 Exhaust.close @16:30:51.65 Toroid.reconfigure :hold @16:30:51.77 Toroid.seek :bypass @16:30:53.43 Toroid.reconfigure :default @16:30:53.55 Toroid.seek :clear @16:30:55.18 Cartridge::Sampler::Clogged in FILTERING -- Cartridge clogged @16:30:55.20 Pressure.coast @16:30:55.56 Sampled 399.9ml @16:30:56.20 Plunge.seek :reagent @16:30:57.15 Sample loop closed. Ready to Deliver Reagents @16:30:57.17 Archive.seek :sealed @16:31:00.87 Archive.seek :evac @16:31:14.90 delay 1:00 @16:32:14.94 Archive.seek :clear @16:32:32.10 RNAlater soak for 10:00 @16:42:32.13 Archive.seek :tVent @16:42:33.79 delay 1 second @16:42:34.82 Archive.seek :evac @16:42:50.83 delay 10 seconds @16:43:00.88 Sample preserved and evacuated @16:43:01.22
Carousel.seek 60 @16:43:01.52 Plunge.seek :locked @16:43:02.54 Archive.seek :clear @16:43:02.77 Air.seek :clear @16:43:03.01 Lysis1.seek :clear @16:43:19.79
Archive.coast @16:43:19.85 Air.coast @16:43:19.96 Lysis1.coast @16:43:20.07 Lysis2.coast @16:43:20.15 Plunge.coast @16:43:20.25 SamplePump.coast @16:43:20.36 Toroid.coast @16:43:20.46 Carousel.coast @16:43:20.76 Gate.power :motor,:OFF @16:43:20.78 Gate.power :core,:OFF @16:43:20.82 Gate.power :analytic1,:OFF @16:43:20.88 Gate.power :analytic2,:OFF @16:43:20.92 Gate.power :analytic3,:OFF @16:43:20.95 Gate.power :main,:OFF @16:43:21.03 Safely stopped and ready to power off @16:43:21.46 Cartridge Load: {59..1 => { Type: :archive_bac, State: :reserved, was: :dry}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 1} @16:43:21.48 Cartridge::None in MAIN -- Not enough dry cartridges remain @16:43:22.37 Session Ended @18:02:29.39PDT04-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @18:02:29.42 >BREAK< ESP3 Gateway v1.11 for cartridge bus @18:02:33.17 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @18:02:33.52 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @18:02:38.87 Polling Can every 2:00 @18:02:38.99 Gate.power :analytic1,:ON,limit to 5000mA @18:02:39.00 Gate.power :analytic2,:ON,limit to 5000mA @18:02:39.02 Gate.power :analytic3,:ON,limit to 5000mA @18:02:39.33 Archive.configure :default @18:02:39.42 Archive.hold @18:02:39.48 Archive.configure :default @18:02:39.63 Air.configure :default @18:02:39.71 Air.hold @18:02:39.77 Air.configure :default @18:02:39.93 Lysis1.configure :default @18:02:40.00 Lysis1.hold @18:02:40.08 Lysis1.configure :default @18:02:40.24 Lysis2.configure :default @18:02:40.32 Lysis2.hold @18:02:40.40 Lysis2.configure :default @18:02:40.59 Plunge.configure :default @18:02:40.64 Plunge.hold @18:02:40.72 Plunge.configure :default @18:02:40.89 SamplePump.configure :default @18:02:41.05 ExtraPump.configure :default @18:02:41.25 Toroid.configure :default @18:02:41.30 Toroid.hold @18:02:41.37 Toroid.configure :default @18:02:41.56 Carousel.configure :default @18:02:41.61 Carousel.hold @18:02:41.69 Carousel.configure :default @18:02:41.85 Pressure.configure :default @18:02:41.92 All elf controllers running firmware version 4.61 @18:02:42.03 Intake.close @18:02:42.14 Exhaust.close @18:02:42.22 Bypass.close @18:02:42.42
Loading /home/esp/esp2/bin/processAll mission @18:02:42.74 Sampling maximum of 400.0ml per cartridge @18:02:42.76 Cartridge group: dry Spares: {spare: 1} @18:02:43.60 Carousel.seek 60 @18:02:43.77 Selecting Cartridge 59 @18:02:44.03 Carousel.seek 59 @18:02:44.57 Lysis1.seek :locked @18:02:47.82 Plunge.seek :locked @18:02:47.99 Air.seek :locked @18:02:52.26
Mounted archive Cartridge @18:02:52.63 Duration of filtering limited to 4:05:00 @18:02:53.01 Priming sample loop w/160ml, bypass w/10ml @18:02:53.03 SamplePump.setPosition! 0ml @18:02:53.07 Bypass.open @18:02:54.57 Intake.open @18:02:55.49 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @18:02:58.11 Vacuum tare is 19% of full scale @18:02:58.13 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @18:02:58.40 Pressure.reconfigure :default @18:02:58.47 Pressure.hold @18:02:58.76 SamplePump.reconfigure :prime @18:02:58.82 SamplePump.seek 10ml,1:21 @18:03:09.94 Pressure.coast @18:03:10.00 SamplePump.setPosition! 0ml @18:03:10.04 Exhaust.open @18:03:11.50 Bypass.close @18:03:13.06 Pressure.reconfigure :default @18:03:13.16 Pressure.hold @18:03:13.49 SamplePump.reconfigure :prime @18:03:13.57 SamplePump.seek 160ml,13:51 @18:06:05.14 Pressure.coast @18:06:05.21 Bypass.open @18:06:06.43 Intake.close @18:06:07.39 Toroid.reconfigure :hold @18:06:07.46 Toroid.seek :crack @18:06:09.08 Toroid.seek :cartridge @18:06:10.04 Toroid.reconfigure :default @18:06:10.10 Toroid.seek :clear @18:06:12.07 Pressurizing cartridge to 26psi above 14.0psia @18:06:12.10 SamplePump.setPosition! 0ml @18:06:12.14 Intake.open @18:06:12.92 Bypass.close @18:06:13.65 Exhaust.close @18:06:14.50 Pressure.reconfigure :default @18:06:14.57 Pressure.hold @18:06:14.84 SamplePump.reconfigure :charge @18:06:14.92 SamplePump.seek 2.3ml,42 seconds @18:06:25.35 Pressure.coast @18:06:25.40 Intake.close @18:06:26.35 Pressurized cartridge with 0.496ml to 26.2psi @18:06:26.37 Isolated @36.5psia (22.5psi) @18:06:26.38 Checking for cartridge leaks from 36.5psia @18:06:26.40 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @18:06:30.98 Cartridge pressure fell 1% to 36.0psia @18:06:31.09 Sampling 400.0ml @18:06:31.10 SamplePump.setPosition! 0ml @18:06:31.18 Exhaust.open @18:06:31.84 Intake.open @18:06:32.85 Pressure.reconfigure :default @18:06:32.94 Pressure.hold @18:06:33.31 SamplePump.reconfigure :sample @18:06:33.41 SamplePump.seek 400ml,33:51 @18:06:35.92 Cartridge 59 sampling 1/400ml at 0.51ml/s with 23.4psi vacuum @18:07:35.93 Cartridge 59 sampling 31/400ml at 0.48ml/s with 25.5psi vacuum @18:08:35.91 Cartridge 59 sampling 59/400ml at 0.45ml/s with 25.2psi vacuum @18:09:35.92 Cartridge 59 sampling 85/400ml at 0.41ml/s with 26.2psi vacuum @18:10:35.90 Cartridge 59 sampling 109/400ml at 0.41ml/s with 22.4psi vacuum @18:11:35.92 Cartridge 59 sampling 132/400ml at 0.36ml/s with 21.8psi vacuum @18:12:35.92 Cartridge 59 sampling 153/400ml at 0.33ml/s with 26.0psi vacuum @18:13:35.90 Cartridge 59 sampling 174/400ml at 0.32ml/s with 26.3psi vacuum @18:14:35.91 Cartridge 59 sampling 192/400ml at 0.31ml/s with 25.7psi vacuum @18:15:35.91 Cartridge 59 sampling 210/400ml at 0.28ml/s with 23.3psi vacuum @18:16:35.94 Cartridge 59 sampling 227/400ml at 0.26ml/s with 23.3psi vacuum @18:17:35.92 Cartridge 59 sampling 243/400ml at 0.25ml/s with 23.3psi vacuum @18:18:35.91 Cartridge 59 sampling 258/400ml at 0.26ml/s with 24.7psi vacuum @18:19:35.92 Cartridge 59 sampling 272/400ml at 0.30ml/s with 25.5psi vacuum @18:20:35.91 Cartridge 59 sampling 287/400ml at 0.22ml/s with 23.5psi vacuum @18:21:35.90 Cartridge 59 sampling 301/400ml at 0.26ml/s with 25.7psi vacuum @18:22:35.92 Cartridge 59 sampling 315/400ml at 0.22ml/s with 25.5psi vacuum @18:23:35.90 Cartridge 59 sampling 328/400ml at 0.20ml/s with 26.2psi vacuum @18:24:35.92 Cartridge 59 sampling 341/400ml at 0.20ml/s with 24.3psi vacuum @18:25:35.90 Cartridge 59 sampling 354/400ml at 0.20ml/s with 23.4psi vacuum @18:26:35.92 Cartridge 59 sampling 367/400ml at 0.21ml/s with 23.9psi vacuum @18:27:35.92 Cartridge 59 sampling 380/400ml at 0.20ml/s with 24.4psi vacuum @18:28:35.91 Cartridge 59 sampling 392/400ml at 0.20ml/s with 30.6psi vacuum @18:29:13.10 Pressure.coast @18:29:13.36 Intake.close @18:29:13.44 Exhaust.close @18:29:13.83 Toroid.reconfigure :hold @18:29:13.89 Toroid.seek :bypass @18:29:15.54 Toroid.reconfigure :default @18:29:15.62 Toroid.seek :clear @18:29:17.24 Sampled 400.0ml @18:29:17.83 Plunge.seek :reagent @18:29:18.77 Sample loop closed. Ready to Deliver Reagents @18:29:18.79 Archive.seek :sealed @18:29:22.43 Archive.seek :evac @18:29:36.44 delay 1:00 @18:30:36.48 Archive.seek :clear @18:30:53.62 RNAlater soak for 10:00 @18:40:53.69 Archive.seek :tVent @18:40:55.34 delay 1 second @18:40:56.39 Archive.seek :evac @18:41:12.38 delay 10 seconds @18:41:22.40 Sample preserved and evacuated @18:41:22.69
Carousel.seek 59 @18:41:22.96 Plunge.seek :locked @18:41:23.98 Archive.seek :clear @18:41:24.21 Air.seek :clear @18:41:24.46 Lysis1.seek :clear @18:41:41.16
Archive.coast @18:41:41.23 Air.coast @18:41:41.27 Lysis1.coast @18:41:41.32 Lysis2.coast @18:41:41.39 Plunge.coast @18:41:41.43 SamplePump.coast @18:41:41.50 Toroid.coast @18:41:41.54 Carousel.coast @18:41:41.78 Gate.power :motor,:OFF @18:41:41.80 Gate.power :core,:OFF @18:41:41.81 Gate.power :analytic1,:OFF @18:41:41.82 Gate.power :analytic2,:OFF @18:41:41.83 Gate.power :analytic3,:OFF @18:41:41.85 Gate.power :main,:OFF @18:41:41.86 Safely stopped and ready to power off @18:41:42.24 Gate.power :main,:ON,limit to 5000mA @18:41:42.25 Gate.power :analytic1,:ON,limit to 5000mA @18:41:42.29 Gate.power :analytic2,:ON,limit to 5000mA @18:41:42.31 Gate.power :analytic3,:ON,limit to 5000mA @18:41:42.84 Powering core bus @18:41:45.21 Archive.configure :default @18:41:45.38 Air.configure :default @18:41:45.56 Lysis1.configure :default @18:41:45.70 Lysis2.configure :default @18:41:45.89 Plunge.configure :default @18:41:46.07 SamplePump.configure :default @18:41:46.24 ExtraPump.configure :default @18:41:46.39 Toroid.configure :default @18:41:46.60 Carousel.configure :default @18:41:46.77 Pressure.configure :default @18:41:46.85 All elf controllers running firmware version 4.61 @18:41:46.87 Gate.power :motor,:ON,limit to 5000mA @18:41:47.11 Archive.reconfigure :default @18:41:47.52 Archive.home.jog 0 @18:41:48.54 Toroid.reconfigure :default @18:41:48.62 Plunge.reconfigure :default @18:41:48.67 Lysis2.reconfigure :default @18:41:48.75 Air.reconfigure :default @18:41:48.80 Lysis1.reconfigure :default @18:41:49.09 Toroid.home.jog 0 @18:41:49.12 Lysis2.home.jog 0 @18:41:49.18 Plunge.home.jog 0 @18:41:49.21 Air.home.jog 0 @18:41:49.25 Lysis1.home.jog 0 @18:41:49.89 Archive.reconfigure :homing @18:41:50.17 Archive.jog 1373 @18:41:50.26 Toroid.reconfigure :homing @18:41:50.28 Plunge.reconfigure :homing @18:41:50.31 Lysis2.reconfigure :homing @18:41:50.40 Air.reconfigure :homing @18:41:50.45 Lysis1.reconfigure :homing @18:41:50.75 Toroid.jog 630 @18:41:50.82 Plunge.jog 847 @18:41:50.84 Lysis2.jog 1357 @18:41:50.88 Lysis1.jog 1351 @18:41:50.93 Air.jog 1362 @18:41:51.71 Archive.setRawPosition! 50 @18:41:51.80 Archive.seek :home @18:41:51.93 Plunge.setRawPosition! 50 @18:41:51.98 Plunge.seek :home @18:41:52.10 Archive.reconfigure :default @18:41:52.25 Lysis2.setRawPosition! 50 @18:41:52.33 Lysis2.seek :home @18:41:52.46 Lysis1.setRawPosition! 50 @18:41:52.50 Air.setRawPosition! 50 @18:41:52.52 Plunge.reconfigure :default @18:41:52.68 Lysis1.seek :home @18:41:52.78 Air.seek :home @18:41:52.95 Lysis2.reconfigure :default @18:41:53.14 Lysis1.reconfigure :default @18:41:53.17 Toroid.setRawPosition! 0 @18:41:53.29 Toroid.seek :home @18:41:53.47 Air.reconfigure :default @18:41:53.50 Toroid.reconfigure :default @18:41:54.01
Toroid.seek :clear @18:41:55.65 Carousel.reconfigure :default @18:41:55.70 Carousel.home.jog 0 @18:41:56.23 Carousel.seek 59 @18:41:56.41 Selecting Cartridge 58 @18:41:56.70 Carousel.seek 58 @18:41:57.27 Lysis1.seek :locked @18:42:00.52 Plunge.seek :locked @18:42:00.70 Air.seek :locked @18:42:05.01
Mounted archive Cartridge @18:42:05.35 Duration of filtering limited to 4:05:00 @18:42:05.74 Priming sample loop w/160ml, bypass w/10ml @18:42:05.76 SamplePump.setPosition! 0ml @18:42:05.82 Bypass.open @18:42:07.34 Intake.open @18:42:08.21 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @18:42:10.81 Vacuum tare is 19% of full scale @18:42:10.85 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @18:42:11.10 Pressure.reconfigure :default @18:42:11.17 Pressure.hold @18:42:11.44 SamplePump.reconfigure :prime @18:42:11.52 SamplePump.seek 10ml,1:21 @18:42:22.62 Pressure.coast @18:42:22.70 SamplePump.setPosition! 0ml @18:42:22.74 Exhaust.open @18:42:24.12 Bypass.close @18:42:25.79 Pressure.reconfigure :default @18:42:25.85 Pressure.hold @18:42:26.22 SamplePump.reconfigure :prime @18:42:26.30 SamplePump.seek 160ml,13:51 @18:45:17.85 Pressure.coast @18:45:17.94 Bypass.open @18:45:18.64 Intake.close @18:45:19.61 Toroid.reconfigure :hold @18:45:19.68 Toroid.seek :crack @18:45:21.31 Toroid.seek :cartridge @18:45:22.29 Toroid.reconfigure :default @18:45:22.35 Toroid.seek :clear @18:45:24.26 Pressurizing cartridge to 26psi above 14.1psia @18:45:24.27 SamplePump.setPosition! 0ml @18:45:24.33 Intake.open @18:45:25.14 Bypass.close @18:45:25.84 Exhaust.close @18:45:26.71 Pressure.reconfigure :default @18:45:26.80 Pressure.hold @18:45:27.06 SamplePump.reconfigure :charge @18:45:27.12 SamplePump.seek 2.3ml,42 seconds @18:45:40.08 Pressure.coast @18:45:40.13 Intake.close @18:45:41.06 Pressurized cartridge with 0.764ml to 26.0psi @18:45:41.07 Isolated @36.2psia (22.0psi) @18:45:41.08 Checking for cartridge leaks from 36.2psia @18:45:41.10 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @18:45:47.72 Cartridge pressure fell 5% to 34.3psia @18:45:47.82 Sampling 400.0ml @18:45:47.86 SamplePump.setPosition! 0ml @18:45:47.90 Exhaust.open @18:45:48.61 Intake.open @18:45:49.79 Pressure.reconfigure :default @18:45:49.84 Pressure.hold @18:45:50.23 SamplePump.reconfigure :sample @18:45:50.32 SamplePump.seek 400ml,33:51 @18:45:52.79 Cartridge 58 sampling 1/400ml at 0.48ml/s with 25.4psi vacuum @18:46:52.81 Cartridge 58 sampling 30/400ml at 0.47ml/s with 23.7psi vacuum @18:47:52.80 Cartridge 58 sampling 58/400ml at 0.44ml/s with 25.3psi vacuum @18:48:53.49 Cartridge 58 sampling 83/400ml at 0.40ml/s with 26.0psi vacuum @18:49:52.85 Cartridge 58 sampling 108/400ml at 0.40ml/s with 25.4psi vacuum @18:50:52.82 Cartridge 58 sampling 131/400ml at 0.37ml/s with 21.3psi vacuum @18:51:52.81 Cartridge 58 sampling 154/400ml at 0.35ml/s with 25.1psi vacuum @18:52:52.82 Cartridge 58 sampling 175/400ml at 0.34ml/s with 25.3psi vacuum @18:53:52.81 Cartridge 58 sampling 195/400ml at 0.32ml/s with 24.1psi vacuum @18:54:52.81 Cartridge 58 sampling 214/400ml at 0.28ml/s with 19.9psi vacuum @18:55:52.83 Cartridge 58 sampling 232/400ml at 0.30ml/s with 22.6psi vacuum @18:56:52.81 Cartridge 58 sampling 249/400ml at 0.28ml/s with 23.0psi vacuum @18:57:52.82 Cartridge 58 sampling 265/400ml at 0.28ml/s with 21.8psi vacuum @18:58:52.81 Cartridge 58 sampling 280/400ml at 0.22ml/s with 24.6psi vacuum @18:59:52.81 Cartridge 58 sampling 296/400ml at 0.26ml/s with 24.7psi vacuum @19:00:52.80 Cartridge 58 sampling 311/400ml at 0.22ml/s with 25.1psi vacuum @19:01:52.85 Cartridge 58 sampling 325/400ml at 0.27ml/s with 26.2psi vacuum @19:02:52.81 Cartridge 58 sampling 338/400ml at 0.21ml/s with 23.9psi vacuum @19:03:52.84 Cartridge 58 sampling 351/400ml at 0.20ml/s with 23.7psi vacuum @19:04:52.80 Cartridge 58 sampling 364/400ml at 0.25ml/s with 26.4psi vacuum @19:05:52.82 Cartridge 58 sampling 377/400ml at 0.20ml/s with 28.0psi vacuum @19:06:52.80 Cartridge 58 sampling 389/400ml at 0.20ml/s with 24.1psi vacuum @19:07:42.87 Pressure.coast @19:07:43.04 Intake.close @19:07:43.10 Exhaust.close @19:07:43.35 Toroid.reconfigure :hold @19:07:43.41 Toroid.seek :bypass @19:07:45.06 Toroid.reconfigure :default @19:07:45.15 Toroid.seek :clear @19:07:46.78 Sampled 400.0ml @19:07:47.41 Plunge.seek :reagent @19:07:48.37 Sample loop closed. Ready to Deliver Reagents @19:07:48.42 Archive.seek :sealed @19:07:52.05 Archive.seek :evac @19:08:06.06 delay 1:00 @19:09:06.10 Archive.seek :clear @19:09:23.28 RNAlater soak for 10:00 @19:19:23.31 Archive.seek :tVent @19:19:24.95 delay 1 second @19:19:26.00 Archive.seek :evac @19:19:41.96 delay 10 seconds @19:19:51.99 Sample preserved and evacuated @19:19:52.27
Carousel.seek 58 @19:19:52.55 Plunge.seek :locked @19:19:53.55 Archive.seek :clear @19:19:53.73 Air.seek :clear @19:19:53.96 Lysis1.seek :clear @19:20:10.79
Archive.coast @19:20:10.83 Air.coast @19:20:10.88 Lysis1.coast @19:20:10.94 Lysis2.coast @19:20:10.98 Plunge.coast @19:20:11.05 SamplePump.coast @19:20:11.09 Toroid.coast @19:20:11.14 Carousel.coast @19:20:11.40 Gate.power :motor,:OFF @19:20:11.42 Gate.power :core,:OFF @19:20:11.44 Gate.power :analytic1,:OFF @19:20:11.47 Gate.power :analytic2,:OFF @19:20:11.48 Gate.power :analytic3,:OFF @19:20:11.49 Gate.power :main,:OFF @19:20:11.51 Safely stopped and ready to power off @19:20:11.94 Cartridge Load: {57..1 => { Type: :archive_bac, State: :reserved, was: :dry}, [59, 58] => { Type: :archive_bac, State: :processed, filtered: 400.0}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 1} @19:20:11.95 Cartridge::None in MAIN -- Not enough dry cartridges remain @13:29:24.93PDT05-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @13:29:24.98 >BREAK< ESP3 Gateway v1.11 for cartridge bus @13:29:29.06 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @13:29:29.44 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @13:29:34.94 Polling Can every 2:00 @13:29:35.06 Gate.power :analytic1,:ON,limit to 5000mA @13:29:35.08 Gate.power :analytic2,:ON,limit to 5000mA @13:29:35.12 Gate.power :analytic3,:ON,limit to 5000mA @13:29:35.42 Archive.configure :default @13:29:35.53 Archive.hold @13:29:35.60 Archive.configure :default @13:29:35.79 Air.configure :default @13:29:35.84 Air.hold @13:29:35.95 Air.configure :default @13:29:36.08 Lysis1.configure :default @13:29:36.14 Lysis1.hold @13:29:36.23 Lysis1.configure :default @13:29:36.37 Lysis2.configure :default @13:29:36.43 Lysis2.hold @13:29:36.53 Lysis2.configure :default @13:29:36.71 Plunge.configure :default @13:29:36.77 Plunge.hold @13:29:36.84 Plunge.configure :default @13:29:37.00 SamplePump.configure :default @13:29:37.22 ExtraPump.configure :default @13:29:37.38 Toroid.configure :default @13:29:37.47 Toroid.hold @13:29:37.53 Toroid.configure :default @13:29:37.70 Carousel.configure :default @13:29:37.77 Carousel.hold @13:29:37.83 Carousel.configure :default @13:29:38.05 Pressure.configure :default @13:29:38.10 All elf controllers running firmware version 4.61 @13:29:38.21 Intake.close @13:29:38.25 Exhaust.close @13:29:38.35 Bypass.close @13:29:38.58
Loading /home/esp/esp2/bin/processAll mission @13:29:38.90 Sampling default goal volumes @13:29:38.92 Cartridge group: dry Spares: {spare: 1} @13:29:39.74 Carousel.seek 58 @13:29:39.94 Selecting Cartridge 57 @13:29:40.23 Carousel.seek 57 @13:29:40.79 Lysis1.seek :locked @13:29:44.09 Plunge.seek :locked @13:29:44.29 Air.seek :locked @13:29:48.56
Mounted archive Cartridge @13:29:48.93 Duration of filtering limited to 4:05:00 @13:29:49.33 Priming sample loop w/160ml, bypass w/10ml @13:29:49.35 SamplePump.setPosition! 0ml @13:29:49.41 Bypass.open @13:29:50.95 Intake.open @13:29:51.88 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @13:29:54.50 Vacuum tare is 19% of full scale @13:29:54.52 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @13:29:54.85 Pressure.reconfigure :default @13:29:55.04 Pressure.hold @13:29:55.37 SamplePump.reconfigure :prime @13:29:55.45 SamplePump.seek 10ml,1:21 @13:30:06.58 Pressure.coast @13:30:06.62 SamplePump.setPosition! 0ml @13:30:06.71 Exhaust.open @13:30:08.17 Bypass.close @13:30:09.83 Pressure.reconfigure :default @13:30:09.91 Pressure.hold @13:30:10.26 SamplePump.reconfigure :prime @13:30:10.95 SamplePump.seek 160ml,13:51 @13:33:02.37 Pressure.coast @13:33:02.46 Bypass.open @13:33:03.18 Intake.close @13:33:04.14 Toroid.reconfigure :hold @13:33:04.19 Toroid.seek :crack @13:33:05.82 Toroid.seek :cartridge @13:33:06.80 Toroid.reconfigure :default @13:33:06.85 Toroid.seek :clear @13:33:08.80 Pressurizing cartridge to 26psi above 14.5psia @13:33:08.82 SamplePump.setPosition! 0ml @13:33:08.88 Intake.open @13:33:09.67 Bypass.close @13:33:10.38 Exhaust.close @13:33:11.28 Pressure.reconfigure :default @13:33:11.34 Pressure.hold @13:33:11.59 SamplePump.reconfigure :charge @13:33:11.70 SamplePump.seek 2.3ml,42 seconds @13:33:16.88 Pressure.coast @13:33:16.92 Intake.close @13:33:17.81 Pressurized cartridge with 0.119ml to 26.2psi @13:33:17.83 Isolated @31.9psia (17.4psi) @13:33:17.85 Checking for cartridge leaks from 31.9psia @13:33:17.89 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @13:33:22.47 Cartridge pressure fell 2% to 31.2psia @13:33:22.58 Sampling 1000.0ml @13:33:22.59 SamplePump.setPosition! 0ml @13:33:22.65 Exhaust.open @13:33:23.36 Intake.open @13:33:24.38 Pressure.reconfigure :default @13:33:24.47 Pressure.hold @13:33:24.86 SamplePump.reconfigure :sample @13:33:24.92 SamplePump.seek 1000ml,1:23:51 @13:33:27.48 Cartridge 57 sampling 1/1000ml at 0.52ml/s with 15.8psi vacuum 4.9m deep (21.9psia) @13:34:27.46 Cartridge 57 sampling 34/1000ml at 0.54ml/s with 26.7psi vacuum @13:35:27.47 Cartridge 57 sampling 67/1000ml at 0.56ml/s with 26.3psi vacuum @13:36:27.45 Cartridge 57 sampling 100/1000ml at 0.55ml/s with 27.5psi vacuum @13:37:27.47 Cartridge 57 sampling 133/1000ml at 0.56ml/s with 26.2psi vacuum @13:38:27.45 Cartridge 57 sampling 166/1000ml at 0.55ml/s with 27.6psi vacuum @13:39:27.48 Cartridge 57 sampling 199/1000ml at 0.55ml/s with 20.9psi vacuum 2.1m deep (17.8psia) @13:40:27.45 Cartridge 57 sampling 232/1000ml at 0.54ml/s with 28.3psi vacuum @13:41:27.46 Cartridge 57 sampling 265/1000ml at 0.53ml/s with 29.0psi vacuum @13:42:27.48 Cartridge 57 sampling 298/1000ml at 0.56ml/s with 26.8psi vacuum @13:43:27.48 Cartridge 57 sampling 331/1000ml at 0.55ml/s with 28.4psi vacuum @13:44:27.45 Cartridge 57 sampling 364/1000ml at 0.53ml/s with 27.9psi vacuum @13:45:27.46 Cartridge 57 sampling 397/1000ml at 0.54ml/s with 12.6psi vacuum 3.6m deep (20.0psia) @13:46:27.45 Cartridge 57 sampling 430/1000ml at 0.55ml/s with 13.8psi vacuum 3.6m deep (20.1psia) @13:47:27.48 Cartridge 57 sampling 463/1000ml at 0.56ml/s with 18.3psi vacuum 1.9m deep (17.4psia) @13:48:27.45 Cartridge 57 sampling 497/1000ml at 0.56ml/s with 24.1psi vacuum @13:49:27.46 Cartridge 57 sampling 530/1000ml at 0.56ml/s with 28.3psi vacuum @13:50:27.45 Cartridge 57 sampling 563/1000ml at 0.54ml/s with 29.0psi vacuum @13:51:27.48 Cartridge 57 sampling 596/1000ml at 0.54ml/s with 28.1psi vacuum @13:52:27.45 Cartridge 57 sampling 630/1000ml at 0.56ml/s with 27.8psi vacuum @13:53:27.47 Cartridge 57 sampling 663/1000ml at 0.54ml/s with 27.9psi vacuum @13:54:27.46 Cartridge 57 sampling 696/1000ml at 0.53ml/s with 17.1psi vacuum @13:55:27.47 Cartridge 57 sampling 729/1000ml at 0.54ml/s with 14.2psi vacuum 6.1m deep (23.7psia) @13:56:27.45 Cartridge 57 sampling 763/1000ml at 0.53ml/s with 26.8psi vacuum @13:57:27.47 Cartridge 57 sampling 796/1000ml at 0.56ml/s with 26.6psi vacuum @13:58:27.46 Cartridge 57 sampling 829/1000ml at 0.57ml/s with 26.1psi vacuum @13:59:27.46 Cartridge 57 sampling 862/1000ml at 0.55ml/s with 10.0psi vacuum 6.5m deep (24.4psia) @14:00:27.46 Cartridge 57 sampling 895/1000ml at 0.56ml/s with 17.4psi vacuum 3.2m deep (19.4psia) @14:01:27.48 Cartridge 57 sampling 928/1000ml at 0.57ml/s with 22.2psi vacuum 1.9m deep (17.5psia) @14:02:27.47 Cartridge 57 sampling 962/1000ml at 0.56ml/s with 24.9psi vacuum @14:03:27.46 Cartridge 57 sampling 995/1000ml at 0.55ml/s with 12.0psi vacuum 8.2m deep (26.9psia) @14:03:36.90 Pressure.coast @14:03:37.12 Intake.close @14:03:37.25 Exhaust.close @14:03:37.41 Toroid.reconfigure :hold @14:03:37.47 Toroid.seek :bypass @14:03:39.15 Toroid.reconfigure :default @14:03:39.21 Toroid.seek :clear @14:03:40.83 Sampled 1000.0ml @14:03:41.43 SamplePump.setPosition! 0ml @14:03:41.48 Plunge.seek :reagent @14:03:41.56 Bypass.open @14:03:42.33 Intake.open @14:03:42.50 Sample loop closed. Ready to Deliver Reagents @14:03:42.52 Archive.seek :sealed @14:03:43.17 SamplePump.reconfigure :slow @14:03:43.26 SamplePump.seek -3ml @14:03:46.19 Archive.seek :evac @14:03:49.93 delay 1 second @14:03:51.04 SamplePump.seek -2.8ml @14:03:51.77 delay 1 second @14:03:52.88 SamplePump.seek -3.3ml @14:03:54.22 delay 1 second @14:03:55.25 Exhaust.open @14:03:55.92 Bypass.close @14:03:56.76 SamplePump.seek -35.3ml @14:04:00.17 delay 1:00 @14:05:00.22 Archive.seek :clear @14:05:05.50 delay 3 seconds @14:05:08.62 Intake.close @14:05:08.71 Exhaust.close @14:05:17.37 RNAlater soak for 10:00 @14:15:17.41 Archive.seek :tVent @14:15:19.09 delay 1 second @14:15:20.14 Archive.seek :evac @14:15:36.09 delay 10 seconds @14:15:46.12 Sample preserved and evacuated @14:15:46.41
Carousel.seek 57 @14:15:46.69 Plunge.seek :locked @14:15:47.70 Archive.seek :clear @14:15:47.95 Air.seek :clear @14:15:48.31 Lysis1.seek :clear @14:16:04.92
Archive.coast @14:16:04.99 Air.coast @14:16:05.04 Lysis1.coast @14:16:05.10 Lysis2.coast @14:16:05.17 Plunge.coast @14:16:05.22 SamplePump.coast @14:16:05.29 Toroid.coast @14:16:05.33 Carousel.coast @14:16:05.55 Gate.power :motor,:OFF @14:16:05.57 Gate.power :core,:OFF @14:16:05.58 Gate.power :analytic1,:OFF @14:16:05.59 Gate.power :analytic2,:OFF @14:16:05.60 Gate.power :analytic3,:OFF @14:16:05.62 Gate.power :main,:OFF @14:16:05.65 Safely stopped and ready to power off @14:16:06.09 Cartridge Load: {[59, 58] => { Type: :archive_bac, State: :processed, filtered: 400}, 56..1 => { Type: :archive_bac, State: :reserved, was: :dry}, 57 => { Type: :archive_bac, State: :processed, filtered: 1000.0}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 1} @14:16:06.11 Cartridge::None in MAIN -- Not enough dry cartridges remain @16:04:16.09PDT10-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @16:04:16.13 >BREAK< ESP3 Gateway v1.11 for cartridge bus @16:04:19.72 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @16:04:20.96 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @16:04:27.66 Polling Can every 2:00 @16:04:27.72 Loading /home/esp/esp2/bin/processAll mission @16:04:28.04 Sampling default goal volumes @16:04:28.06 Cartridge group: dry Spares: {spare: 1} @16:04:28.22 Gate.power :main,:ON,limit to 5000mA @16:04:28.23 Gate.power :analytic1,:ON,limit to 5000mA @16:04:28.24 Gate.power :analytic2,:ON,limit to 5000mA @16:04:28.25 Gate.power :analytic3,:ON,limit to 5000mA @16:04:28.29 Powering core bus @16:04:30.89 Archive.configure :default @16:04:31.15 Air.configure :default @16:04:31.48 Lysis1.configure :default @16:04:31.74 Lysis2.configure :default @16:04:32.08 Plunge.configure :default @16:04:32.47 SamplePump.configure :default @16:04:32.78 ExtraPump.configure :default @16:04:33.16 Toroid.configure :default @16:04:33.50 Carousel.configure :default @16:04:33.66 Pressure.configure :default @16:04:33.79 All elf controllers running firmware version 4.61 @16:04:33.87 Gate.power :motor,:ON,limit to 5000mA @16:04:34.38 Archive.reconfigure :default @16:04:35.21 Archive.home.jog 0 @16:04:35.25 Toroid.reconfigure :default @16:04:35.28 Lysis2.reconfigure :default @16:04:35.30 Lysis1.reconfigure :default @16:04:35.35 Air.reconfigure :default @16:04:35.41 Plunge.reconfigure :default @16:04:35.69 Lysis1.home.jog 0 @16:04:35.73 Lysis2.home.jog 0 @16:04:35.78 Toroid.home.jog 0 @16:04:35.81 Air.home.jog 0 @16:04:35.85 Plunge.home.jog 0 @16:04:36.51 Toroid.reconfigure :homing @16:04:36.56 Lysis2.reconfigure :homing @16:04:36.61 Air.reconfigure :homing @16:04:36.64 Archive.reconfigure :homing @16:04:36.67 Lysis1.reconfigure :homing @16:04:36.90 Lysis2.jog 1346 @16:04:36.94 Toroid.jog 632 @16:04:36.98 Archive.jog 1356 @16:04:37.01 Air.jog 1359 @16:04:37.05 Lysis1.jog 1350 @16:04:37.34 Plunge.reconfigure :homing @16:04:37.39 Plunge.jog 308 @16:04:37.97 Plunge.setRawPosition! 50 @16:04:38.02 Plunge.seek :home @16:04:38.23 Lysis2.setRawPosition! 50 @16:04:38.30 Lysis2.seek :home @16:04:38.44 Lysis1.setRawPosition! 50 @16:04:38.46 Air.setRawPosition! 50 @16:04:38.49 Archive.setRawPosition! 50 @16:04:38.50 Plunge.reconfigure :default @16:04:38.64 Lysis1.seek :home @16:04:38.71 Air.seek :home @16:04:38.76 Archive.seek :home @16:04:38.95 Lysis2.reconfigure :default @16:04:39.05 Lysis1.reconfigure :default @16:04:39.18 Air.reconfigure :default @16:04:39.20 Archive.reconfigure :default @16:04:39.34 Toroid.setRawPosition! 0 @16:04:39.41 Toroid.seek :home @16:04:39.68 Toroid.reconfigure :default @16:04:40.11
Toroid.seek :clear @16:04:41.84 Carousel.reconfigure :default @16:04:41.91 Carousel.home.jog 0 @16:04:42.64 Carousel.seek 60 @16:04:43.11 Selecting Cartridge 56 @16:04:43.86 Carousel.seek 56 @16:04:45.22 Lysis1.seek :locked @16:04:48.44 Plunge.seek :locked @16:04:48.59 Air.seek :locked @16:04:53.92
Mounted archive Cartridge @16:04:54.77 Duration of filtering limited to 4:05:00 @16:04:55.08 Priming sample loop w/160ml, bypass w/40ml @16:04:55.10 SamplePump.setPosition! 0ml @16:04:55.14 Bypass.open @16:04:56.65 Intake.open @16:04:57.55 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @16:05:00.34 Vacuum tare is 18% of full scale @16:05:00.36 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @16:05:00.91 Pressure.reconfigure :default @16:05:01.05 Pressure.hold @16:05:01.55 SamplePump.reconfigure :prime @16:05:01.74 SamplePump.seek 40ml,3:51 @16:05:45.06 Pressure.coast @16:05:45.11 SamplePump.setPosition! 0ml @16:05:45.16 Exhaust.open @16:05:46.69 Bypass.close @16:05:48.56 Pressure.reconfigure :default @16:05:48.77 Pressure.hold @16:05:49.37 SamplePump.reconfigure :prime @16:05:49.45 SamplePump.seek 160ml,13:51 @16:08:41.39 Pressure.coast @16:08:41.47 Bypass.open @16:08:42.32 Intake.close @16:08:43.47 Toroid.reconfigure :hold @16:08:43.60 Toroid.seek :crack @16:08:45.32 Toroid.seek :cartridge @16:08:46.25 Toroid.reconfigure :default @16:08:46.31 Toroid.seek :clear @16:08:48.16 Pressurizing cartridge to 26psi above 14.3psia @16:08:48.18 SamplePump.setPosition! 0ml @16:08:48.22 Intake.open @16:08:49.05 Bypass.close @16:08:49.84 Exhaust.close @16:08:51.25 Pressure.reconfigure :default @16:08:51.33 Pressure.hold @16:08:51.86 SamplePump.reconfigure :charge @16:08:51.95 SamplePump.seek 2.3ml,42 seconds @16:09:01.59 Pressure.coast @16:09:01.64 Intake.close @16:09:02.82 Pressurized cartridge with 0.405ml to 26.1psi @16:09:02.85 Isolated @34.9psia (20.6psi) @16:09:02.90 Checking for cartridge leaks from 34.9psia @16:09:02.97 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @16:09:09.31 Cartridge pressure fell 16% to 29.5psia @16:09:09.65 Sampling 2000.0ml @16:09:09.68 SamplePump.setPosition! 0ml @16:09:09.77 Exhaust.open @16:09:10.56 Intake.open @16:09:11.85 Pressure.reconfigure :default @16:09:11.95 Pressure.hold @16:09:12.59 SamplePump.reconfigure :sample @16:09:12.79 SamplePump.seek 2000ml,2:47:12 @16:09:14.79 Cartridge 56 sampling 1/2000ml at 0.58ml/s with 27.4psi vacuum @16:10:14.94 Cartridge 56 sampling 33/2000ml at 0.52ml/s with 24.8psi vacuum @16:11:14.79 Cartridge 56 sampling 63/2000ml at 0.48ml/s with 26.0psi vacuum @16:12:15.10 Cartridge 56 sampling 92/2000ml at 0.45ml/s with 26.1psi vacuum @16:13:14.79 Cartridge 56 sampling 118/2000ml at 0.42ml/s with 20.4psi vacuum @16:14:14.98 Cartridge 56 sampling 143/2000ml at 0.38ml/s with 25.7psi vacuum @16:15:14.80 Cartridge 56 sampling 166/2000ml at 0.38ml/s with 26.1psi vacuum @16:16:15.01 Cartridge 56 sampling 187/2000ml at 0.36ml/s with 22.0psi vacuum @16:17:14.79 Cartridge 56 sampling 207/2000ml at 0.32ml/s with 23.0psi vacuum @16:18:14.93 Cartridge 56 sampling 224/2000ml at 0.25ml/s with 27.2psi vacuum @16:19:14.79 Cartridge 56 sampling 240/2000ml at 0.26ml/s with 16.8psi vacuum @16:20:14.99 Cartridge 56 sampling 256/2000ml at 0.25ml/s with 30.7psi vacuum @16:21:14.80 Cartridge 56 sampling 270/2000ml at 0.21ml/s with 28.1psi vacuum @16:22:14.97 Cartridge 56 sampling 284/2000ml at 0.23ml/s with 27.0psi vacuum @16:23:14.80 Cartridge 56 sampling 297/2000ml at 0.20ml/s with 29.4psi vacuum @16:24:14.98 Cartridge 56 sampling 310/2000ml at 0.22ml/s with 20.1psi vacuum @16:25:14.80 Cartridge 56 sampling 323/2000ml at 0.20ml/s with 32.6psi vacuum @16:26:14.96 Cartridge 56 sampling 336/2000ml at 0.21ml/s with 13.6psi vacuum 9.9m deep (29.3psia) @16:27:14.88 Cartridge 56 sampling 349/2000ml at 0.22ml/s with 22.0psi vacuum 4.1m deep (20.7psia) @16:28:14.89 Cartridge 56 sampling 361/2000ml at 0.22ml/s with 14.2psi vacuum 6.1m deep (23.8psia) @16:29:14.80 Cartridge 56 sampling 374/2000ml at 0.20ml/s with 27.1psi vacuum @16:30:04.90 Flow averaged 0.31ml/s over 20:43 @16:30:04.93 Waiting up to 46 seconds for vacuum to reduce from 41.3 to 18.0psi @16:30:05.79 Relieved excess vacuum in 10.5 seconds @16:30:05.94 SamplePump.seek 2000ml,2:15:10 @16:30:15.12 Cartridge 56 sampling 386/2000ml at 0.20ml/s with 23.9psi vacuum 6.5m deep (24.3psia) @16:30:16.68 Flow averaged 0.23ml/s over 10 seconds @16:30:16.71 Waiting up to 46 seconds for vacuum to reduce from 41.4 to 18.0psi @16:30:17.56 Relieved excess vacuum in 2.2 seconds @16:30:17.58 SamplePump.seek 2000ml,2:14:59 @16:30:23.96 Flow estimate reduced to 0.18ml/s @16:30:24.02 Intake.close @16:30:24.18 Exhaust.close @16:30:24.66 Toroid.reconfigure :hold @16:30:24.71 Toroid.seek :bypass @16:30:26.34 Toroid.reconfigure :default @16:30:26.40 Toroid.seek :clear @16:30:28.00 Cartridge::Sampler::Clogged in FILTERING -- Cartridge clogged @16:30:28.02 Pressure.coast @16:30:28.23 Sampled 387.7ml @16:30:28.71 SamplePump.setPosition! 0ml @16:30:28.80 Plunge.seek :reagent @16:30:28.85 Bypass.open @16:30:29.80 Intake.open @16:30:30.10 Sample loop closed. Ready to Deliver Reagents @16:30:30.15 Archive.seek :sealed @16:30:32.06 SamplePump.reconfigure :slow @16:30:32.26 SamplePump.seek -3ml @16:30:33.84 Archive.seek :evac @16:30:39.02 delay 1 second @16:30:40.23 SamplePump.seek -2.8ml @16:30:41.09 delay 1 second @16:30:42.31 SamplePump.seek -3.3ml @16:30:43.77 delay 1 second @16:30:44.83 Exhaust.open @16:30:45.50 Bypass.close @16:30:46.30 SamplePump.seek -35.3ml @16:30:47.87 delay 1:00 @16:31:47.92 Archive.seek :clear @16:31:55.07 delay 3 seconds @16:31:58.19 Intake.close @16:31:58.25 Exhaust.close @16:32:05.12 RNAlater soak for 10:00 @16:35:02.70 Session Begun @16:35:04.75 -> slots @16:37:11.26 -> Cartridge.release 56 @16:37:23.76 -> Cartridge.release 55 @16:37:27.22 -> exit @16:37:27.27 Session Ended @16:42:05.17 Archive.seek :tVent @16:42:06.82 delay 1 second @16:42:07.85 Archive.seek :evac @16:42:23.81 delay 10 seconds @16:42:33.85 Sample preserved and evacuated @16:42:34.08
Carousel.seek 56 @16:42:34.30 Plunge.seek :locked @16:42:35.29 Archive.seek :clear @16:42:35.47 Air.seek :clear @16:42:35.62 Lysis1.seek :clear @16:42:52.49
Archive.coast @16:42:52.53 Air.coast @16:42:52.58 Lysis1.coast @16:42:52.62 Lysis2.coast @16:42:52.66 Plunge.coast @16:42:52.71 SamplePump.coast @16:42:52.75 Toroid.coast @16:42:52.79 Carousel.coast @16:42:52.99 Gate.power :motor,:OFF @16:42:53.00 Gate.power :core,:OFF @16:42:53.02 Gate.power :analytic1,:OFF @16:42:53.03 Gate.power :analytic2,:OFF @16:42:53.04 Gate.power :analytic3,:OFF @16:42:53.05 Gate.power :main,:OFF @16:42:53.06 Safely stopped and ready to power off @16:42:53.42 Gate.power :main,:ON,limit to 5000mA @16:42:53.43 Gate.power :analytic1,:ON,limit to 5000mA @16:42:53.45 Gate.power :analytic2,:ON,limit to 5000mA @16:42:53.46 Gate.power :analytic3,:ON,limit to 5000mA @16:42:54.06 Powering core bus @16:42:57.13 Archive.configure :default @16:42:57.46 Air.configure :default @16:42:57.78 Lysis1.configure :default @16:42:58.12 Lysis2.configure :default @16:42:58.50 Plunge.configure :default @16:42:58.95 SamplePump.configure :default @16:42:59.41 ExtraPump.configure :default @16:42:59.72 Toroid.configure :default @16:43:00.02 Carousel.configure :default @16:43:00.32 Pressure.configure :default @16:43:00.38 All elf controllers running firmware version 4.61 @16:43:00.47 Gate.power :motor,:ON,limit to 5000mA @16:43:01.25 Archive.reconfigure :default @16:43:01.50 Archive.home.jog 0 @16:43:01.78 Plunge.reconfigure :default @16:43:01.84 Lysis2.reconfigure :default @16:43:01.87 Lysis1.reconfigure :default @16:43:01.92 Toroid.reconfigure :default @16:43:01.96 Air.reconfigure :default @16:43:02.17 Plunge.home.jog 0 @16:43:02.21 Lysis2.home.jog 0 @16:43:02.23 Lysis1.home.jog 0 @16:43:02.25 Toroid.home.jog 0 @16:43:02.30 Air.home.jog 0 @16:43:02.83 Archive.reconfigure :homing @16:43:03.03 Archive.jog 1368 @16:43:03.08 Toroid.reconfigure :homing @16:43:03.09 Plunge.reconfigure :homing @16:43:03.12 Lysis1.reconfigure :homing @16:43:03.17 Air.reconfigure :homing @16:43:03.22 Lysis2.reconfigure :homing @16:43:03.49 Plunge.jog 843 @16:43:03.53 Air.jog 1358 @16:43:03.56 Toroid.jog 632 @16:43:03.61 Lysis1.jog 1354 @16:43:03.64 Lysis2.jog 1349 @16:43:04.40 Archive.setRawPosition! 50 @16:43:04.45 Archive.seek :home @16:43:04.59 Plunge.setRawPosition! 50 @16:43:04.64 Plunge.seek :home @16:43:04.74 Archive.reconfigure :default @16:43:04.92 Air.setRawPosition! 50 @16:43:04.98 Air.seek :home @16:43:05.08 Lysis2.setRawPosition! 50 @16:43:05.11 Lysis1.setRawPosition! 50 @16:43:05.13 Plunge.reconfigure :default @16:43:05.25 Lysis2.seek :home @16:43:05.29 Lysis1.seek :home @16:43:05.59 Air.reconfigure :default @16:43:06.10 Toroid.setRawPosition! 0 @16:43:06.22 Lysis2.reconfigure :default @16:43:06.36 Lysis1.reconfigure :default @16:43:06.63 Toroid.seek :home @16:43:07.12 Toroid.reconfigure :default @16:43:07.62
Toroid.seek :clear @16:43:09.37 Carousel.reconfigure :default @16:43:09.45 Carousel.home.jog 0 @16:43:10.01 Carousel.seek 56 @16:43:10.34 Selecting Cartridge 55 @16:43:11.09 Carousel.seek 55 @16:43:11.60 Lysis1.seek :locked @16:43:14.83 Plunge.seek :locked @16:43:14.97 Air.seek :locked @16:43:19.50
Mounted archive Cartridge @16:43:20.27 Duration of filtering limited to 4:05:00 @16:43:20.67 Priming sample loop w/160ml, bypass w/40ml @16:43:20.69 SamplePump.setPosition! 0ml @16:43:20.72 Bypass.open @16:43:22.20 Intake.open @16:43:23.08 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @16:43:25.88 Vacuum tare is 18% of full scale @16:43:26.84 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @16:43:27.72 Pressure.reconfigure :default @16:43:27.88 Pressure.hold @16:43:28.42 SamplePump.reconfigure :prime @16:43:28.54 SamplePump.seek 40ml,3:51 @16:44:11.84 Pressure.coast @16:44:12.00 SamplePump.setPosition! 0ml @16:44:12.32 Exhaust.open @16:44:13.95 Bypass.close @16:44:15.68 Pressure.reconfigure :default @16:44:15.83 Pressure.hold @16:44:16.41 SamplePump.reconfigure :prime @16:44:16.58 SamplePump.seek 160ml,13:51 @16:47:08.49 Pressure.coast @16:47:08.71 Bypass.open @16:47:09.47 Intake.close @16:47:10.62 Toroid.reconfigure :hold @16:47:10.71 Toroid.seek :crack @16:47:12.50 Toroid.seek :cartridge @16:47:13.46 Toroid.reconfigure :default @16:47:13.52 Toroid.seek :clear @16:47:15.37 Pressurizing cartridge to 26psi above 14.1psia @16:47:15.39 SamplePump.setPosition! 0ml @16:47:15.42 Intake.open @16:47:16.21 Bypass.close @16:47:16.98 Exhaust.close @16:47:18.24 Pressure.reconfigure :default @16:47:18.34 Pressure.hold @16:47:18.89 SamplePump.reconfigure :charge @16:47:19.11 SamplePump.seek 2.3ml,42 seconds @16:47:30.31 Pressure.coast @16:47:30.36 Intake.close @16:47:31.23 Pressurized cartridge with 0.558ml to 26.0psi @16:47:31.24 Isolated @33.8psia (19.7psi) @16:47:31.25 Checking for cartridge leaks from 33.8psia @16:47:31.27 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @16:47:36.99 Cartridge pressure fell 5% to 32.1psia @16:47:37.14 Sampling 2000.0ml @16:47:37.17 SamplePump.setPosition! 0ml @16:47:37.33 Exhaust.open @16:47:38.02 Intake.open @16:47:39.25 Pressure.reconfigure :default @16:47:39.39 Pressure.hold @16:47:39.91 SamplePump.reconfigure :sample @16:47:40.13 SamplePump.seek 2000ml,2:47:12 @16:47:42.15 Cartridge 55 sampling 1/2000ml at 0.56ml/s with 21.1psi vacuum @16:48:42.40 Cartridge 55 sampling 34/2000ml at 0.56ml/s with 20.8psi vacuum @16:49:42.16 Cartridge 55 sampling 67/2000ml at 0.54ml/s with 19.6psi vacuum @16:50:42.40 Cartridge 55 sampling 100/2000ml at 0.54ml/s with 26.0psi vacuum @16:51:42.16 Cartridge 55 sampling 132/2000ml at 0.52ml/s with 24.8psi vacuum @16:52:42.44 Cartridge 55 sampling 164/2000ml at 0.51ml/s with 20.1psi vacuum @16:53:42.16 Cartridge 55 sampling 194/2000ml at 0.49ml/s with 24.1psi vacuum @16:54:42.46 Cartridge 55 sampling 222/2000ml at 0.46ml/s with 19.5psi vacuum @16:55:42.16 Cartridge 55 sampling 248/2000ml at 0.44ml/s with 22.6psi vacuum @16:56:42.30 Cartridge 55 sampling 273/2000ml at 0.40ml/s with 26.2psi vacuum @16:57:42.16 Cartridge 55 sampling 295/2000ml at 0.37ml/s with 28.2psi vacuum @16:58:42.36 Cartridge 55 sampling 316/2000ml at 0.35ml/s with 25.9psi vacuum @16:59:42.16 Cartridge 55 sampling 335/2000ml at 0.31ml/s with 28.2psi vacuum @17:00:42.36 Cartridge 55 sampling 353/2000ml at 0.32ml/s with 20.5psi vacuum 3.5m deep (19.9psia) @17:01:42.16 Cartridge 55 sampling 370/2000ml at 0.27ml/s with 14.4psi vacuum @17:02:42.32 Cartridge 55 sampling 387/2000ml at 0.23ml/s with 29.3psi vacuum @17:03:42.16 Cartridge 55 sampling 402/2000ml at 0.29ml/s with 23.3psi vacuum @17:04:42.31 Cartridge 55 sampling 417/2000ml at 0.24ml/s with 27.1psi vacuum @17:05:42.16 Cartridge 55 sampling 431/2000ml at 0.25ml/s with 24.3psi vacuum @17:06:42.31 Cartridge 55 sampling 445/2000ml at 0.25ml/s with 26.1psi vacuum @17:07:42.16 Cartridge 55 sampling 459/2000ml at 0.22ml/s with 10.5psi vacuum @17:08:42.31 Cartridge 55 sampling 472/2000ml at 0.24ml/s with 28.1psi vacuum @17:09:42.16 Cartridge 55 sampling 486/2000ml at 0.23ml/s with 18.6psi vacuum @17:10:42.39 Cartridge 55 sampling 499/2000ml at 0.23ml/s with 20.5psi vacuum 3.8m deep (20.3psia) @17:11:42.16 Cartridge 55 sampling 512/2000ml at 0.24ml/s with 22.9psi vacuum @17:12:42.30 Cartridge 55 sampling 525/2000ml at 0.23ml/s with 21.4psi vacuum 4.7m deep (21.7psia) @17:13:42.16 Cartridge 55 sampling 538/2000ml at 0.20ml/s with 34.4psi vacuum @17:14:42.25 Cartridge 55 sampling 550/2000ml at 0.22ml/s with 26.6psi vacuum @17:15:42.16 Cartridge 55 sampling 563/2000ml at 0.20ml/s with 27.3psi vacuum @17:16:42.21 Cartridge 55 sampling 575/2000ml at 0.20ml/s with 38.6psi vacuum @17:17:42.16 Cartridge 55 sampling 588/2000ml at 0.20ml/s with 37.1psi vacuum @17:18:42.31 Cartridge 55 sampling 600/2000ml at 0.20ml/s with 36.3psi vacuum @17:18:59.98 Flow averaged 0.32ml/s over 31:13 @17:19:00.07 Waiting up to 13 seconds for vacuum to reduce from 28.0 to 18.0psi @17:19:00.78 Relieved excess vacuum in 8.2 seconds @17:19:00.80 SamplePump.seek 2000ml,1:56:51 @17:19:41.91 Flow averaged 0.25ml/s over 34 seconds @17:19:41.97 Waiting up to 50 seconds for vacuum to reduce from 42.8 to 18.0psi @17:19:43.01 Cartridge 55 sampling 612/2000ml at 0.00ml/s with 6.1psi vacuum @17:19:43.30 Relieved excess vacuum in 9 seconds @17:19:43.34 SamplePump.seek 2000ml,1:56:09 @17:20:01.96 Flow averaged 0.29ml/s over 14 seconds @17:20:02.00 Waiting up to 51 seconds for vacuum to reduce from 43.1 to 18.0psi @17:20:02.74 Relieved excess vacuum in 6.1 seconds @17:20:02.77 SamplePump.seek 2000ml,1:55:51 @17:20:25.43 Flow averaged 0.29ml/s over 16 seconds @17:20:25.46 Waiting up to 18 seconds for vacuum to reduce from 29.9 to 18.0psi @17:20:26.31 Relieved excess vacuum in 7.5 seconds @17:20:26.33 SamplePump.seek 2000ml,1:55:28 @17:20:30.30 Flow estimate reduced to 0.23ml/s @17:20:30.33 Waiting up to 10 seconds for vacuum to reduce from 27.0 to 18.0psi @17:20:31.06 Relieved excess vacuum in 1.5 seconds @17:20:31.17 SamplePump.seek 2000ml,1:55:24 @17:20:37.08 Flow estimate reduced to 0.18ml/s @17:20:37.10 Intake.close @17:20:37.18 Exhaust.close @17:20:37.37 Toroid.reconfigure :hold @17:20:37.43 Toroid.seek :bypass @17:20:39.13 Toroid.reconfigure :default @17:20:39.25 Toroid.seek :clear @17:20:40.97 Cartridge::Sampler::Clogged in FILTERING -- Cartridge clogged @17:20:41.04 Pressure.coast @17:20:42.69 Sampled 622.6ml @17:20:43.42 SamplePump.setPosition! 0ml @17:20:43.47 Plunge.seek :reagent @17:20:43.50 Bypass.open @17:20:44.25 Intake.open @17:20:44.45 Sample loop closed. Ready to Deliver Reagents @17:20:44.47 Archive.seek :sealed @17:20:45.13 SamplePump.reconfigure :slow @17:20:45.19 SamplePump.seek -3ml @17:20:48.30 Archive.seek :evac @17:20:51.98 delay 1 second @17:20:53.08 SamplePump.seek -2.8ml @17:20:53.77 delay 1 second @17:20:54.86 SamplePump.seek -3.3ml @17:20:56.19 delay 1 second @17:20:57.22 Exhaust.open @17:20:57.95 Bypass.close @17:20:58.93 SamplePump.seek -35.3ml @17:21:02.38 delay 1:00 @17:22:02.43 Archive.seek :clear @17:22:07.64 delay 3 seconds @17:22:10.93 Intake.close @17:22:11.13 Exhaust.close @17:22:19.59 RNAlater soak for 10:00 @17:25:11.40 Session Begun @17:25:15.95 -> slots @17:25:40.35 -> Cartridge.release 54 @17:25:45.05 -> Cartridge.dry @17:25:47.49 -> exit @17:25:47.51 Session Ended @17:32:19.63 Archive.seek :tVent @17:32:21.42 delay 1 second @17:32:22.47 Archive.seek :evac @17:32:38.46 delay 10 seconds @17:32:48.50 Sample preserved and evacuated @17:32:48.93
Carousel.seek 55 @17:32:49.42 Plunge.seek :locked @17:32:50.41 Archive.seek :clear @17:32:50.57 Air.seek :clear @17:32:50.73 Lysis1.seek :clear @17:33:07.69
Archive.coast @17:33:07.81 Air.coast @17:33:07.85 Lysis1.coast @17:33:07.89 Lysis2.coast @17:33:08.00 Plunge.coast @17:33:08.14 SamplePump.coast @17:33:08.51 Toroid.coast @17:33:08.62 Carousel.coast @17:33:08.81 Gate.power :motor,:OFF @17:33:08.82 Gate.power :core,:OFF @17:33:08.83 Gate.power :analytic1,:OFF @17:33:08.85 Gate.power :analytic2,:OFF @17:33:08.86 Gate.power :analytic3,:OFF @17:33:08.87 Gate.power :main,:OFF @17:33:08.88 Safely stopped and ready to power off @17:33:09.21 Gate.power :main,:ON,limit to 5000mA @17:33:09.22 Gate.power :analytic1,:ON,limit to 5000mA @17:33:09.24 Gate.power :analytic2,:ON,limit to 5000mA @17:33:09.25 Gate.power :analytic3,:ON,limit to 5000mA @17:33:09.88 Powering core bus @17:33:12.19 Archive.configure :default @17:33:12.34 Air.configure :default @17:33:12.45 Lysis1.configure :default @17:33:12.56 Lysis2.configure :default @17:33:12.71 Plunge.configure :default @17:33:12.85 SamplePump.configure :default @17:33:13.07 ExtraPump.configure :default @17:33:13.52 Toroid.configure :default @17:33:13.84 Carousel.configure :default @17:33:14.46 Pressure.configure :default @17:33:14.63 All elf controllers running firmware version 4.61 @17:33:14.65 Gate.power :motor,:ON,limit to 5000mA @17:33:15.38 Archive.reconfigure :default @17:33:16.00 Archive.home.jog 0 @17:33:16.48 Lysis2.reconfigure :default @17:33:16.78 Plunge.reconfigure :default @17:33:16.82 Toroid.reconfigure :default @17:33:17.01 Air.reconfigure :default @17:33:17.05 Lysis1.reconfigure :default @17:33:17.60 Lysis2.home.jog 0 @17:33:17.68 Toroid.home.jog 0 @17:33:17.73 Plunge.home.jog 0 @17:33:17.83 Lysis1.home.jog 0 @17:33:17.85 Air.home.jog 0 @17:33:19.00 Archive.reconfigure :homing @17:33:19.30 Archive.jog 1367 @17:33:19.36 Plunge.reconfigure :homing @17:33:19.40 Lysis1.reconfigure :homing @17:33:19.42 Air.reconfigure :homing @17:33:19.47 Toroid.reconfigure :homing @17:33:19.53 Lysis2.reconfigure :homing @17:33:19.79 Lysis1.jog 1358 @17:33:19.82 Toroid.jog 629 @17:33:19.85 Plunge.jog 845 @17:33:19.88 Air.jog 1359 @17:33:19.93 Lysis2.jog 1351 @17:33:20.70 Archive.setRawPosition! 50 @17:33:20.75 Archive.seek :home @17:33:20.91 Plunge.setRawPosition! 50 @17:33:20.96 Plunge.seek :home @17:33:21.06 Archive.reconfigure :default @17:33:21.14 Lysis1.setRawPosition! 50 @17:33:21.22 Lysis1.seek :home @17:33:21.39 Plunge.reconfigure :default @17:33:21.45 Lysis2.setRawPosition! 50 @17:33:21.55 Air.setRawPosition! 50 @17:33:21.73 Air.seek :home @17:33:21.75 Lysis2.seek :home @17:33:21.90 Lysis1.reconfigure :default @17:33:22.07 Air.reconfigure :default @17:33:22.16 Lysis2.reconfigure :default @17:33:22.19 Toroid.setRawPosition! 0 @17:33:22.28 Toroid.seek :home @17:33:22.40 Toroid.reconfigure :default @17:33:22.65
Toroid.seek :clear @17:33:24.45 Carousel.reconfigure :default @17:33:24.59 Carousel.home.jog 0 @17:33:25.35 Carousel.seek 55 @17:33:25.74 Selecting Cartridge 54 @17:33:26.28 Carousel.seek 54 @17:33:27.19 Lysis1.seek :locked @17:33:30.99 Plunge.seek :locked @17:33:31.15 Air.seek :locked @17:33:35.54
Mounted archive Cartridge @17:33:36.30 Duration of filtering limited to 4:05:00 @17:33:36.93 Priming sample loop w/160ml, bypass w/40ml @17:33:36.98 SamplePump.setPosition! 0ml @17:33:37.06 Bypass.open @17:33:38.64 Intake.open @17:33:39.50 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @17:33:42.09 Vacuum tare is 18% of full scale @17:33:42.11 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @17:33:42.32 Pressure.reconfigure :default @17:33:42.38 Pressure.hold @17:33:42.59 SamplePump.reconfigure :prime @17:33:42.75 SamplePump.seek 40ml,3:51 @17:34:26.03 Pressure.coast @17:34:26.08 SamplePump.setPosition! 0ml @17:34:26.13 Exhaust.open @17:34:27.56 Bypass.close @17:34:29.35 Pressure.reconfigure :default @17:34:29.50 Pressure.hold @17:34:30.72 SamplePump.reconfigure :prime @17:34:30.82 SamplePump.seek 160ml,13:51 @17:37:22.50 Pressure.coast @17:37:22.56 Bypass.open @17:37:23.27 Intake.close @17:37:24.43 Toroid.reconfigure :hold @17:37:24.59 Toroid.seek :crack @17:37:26.36 Toroid.seek :cartridge @17:37:27.35 Toroid.reconfigure :default @17:37:27.56 Toroid.seek :clear @17:37:29.44 Pressurizing cartridge to 26psi above 14.7psia @17:37:29.45 SamplePump.setPosition! 0ml @17:37:29.49 Intake.open @17:37:30.27 Bypass.close @17:37:30.98 Exhaust.close @17:37:31.81 Pressure.reconfigure :default @17:37:31.86 Pressure.hold @17:37:32.07 SamplePump.reconfigure :charge @17:37:32.13 SamplePump.seek 2.3ml,42 seconds @17:37:46.70 Pressure.coast @17:37:46.79 Intake.close @17:37:47.79 Pressurized cartridge with 0.958ml to 26.2psi @17:37:47.80 Isolated @36.0psia (21.3psi) @17:37:47.81 Checking for cartridge leaks from 36.0psia @17:37:47.84 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @17:37:52.51 Cartridge pressure fell 1% to 35.5psia @17:37:52.78 Sampling 2000.0ml @17:37:52.86 SamplePump.setPosition! 0ml @17:37:52.97 Exhaust.open @17:37:53.80 Intake.open @17:37:55.19 Pressure.reconfigure :default @17:37:55.26 Pressure.hold @17:37:55.99 SamplePump.reconfigure :sample @17:37:56.17 SamplePump.seek 2000ml,2:47:12 @17:37:57.93 Cartridge 54 sampling 1/2000ml at 0.60ml/s with 26.2psi vacuum @17:38:58.20 Cartridge 54 sampling 32/2000ml at 0.53ml/s with 26.3psi vacuum @17:39:57.93 Cartridge 54 sampling 63/2000ml at 0.51ml/s with 26.6psi vacuum @17:40:58.08 Cartridge 54 sampling 93/2000ml at 0.49ml/s with 25.9psi vacuum @17:41:57.93 Cartridge 54 sampling 122/2000ml at 0.48ml/s with 19.5psi vacuum 1.6m deep (17.1psia) @17:42:58.17 Cartridge 54 sampling 151/2000ml at 0.47ml/s with 26.6psi vacuum @17:43:57.93 Cartridge 54 sampling 180/2000ml at 0.47ml/s with 21.2psi vacuum @17:44:58.16 Cartridge 54 sampling 209/2000ml at 0.46ml/s with 26.3psi vacuum @17:45:57.93 Cartridge 54 sampling 236/2000ml at 0.45ml/s with 25.7psi vacuum @17:46:58.09 Cartridge 54 sampling 262/2000ml at 0.41ml/s with 27.1psi vacuum @17:47:57.93 Cartridge 54 sampling 286/2000ml at 0.37ml/s with 26.6psi vacuum @17:48:58.13 Cartridge 54 sampling 309/2000ml at 0.39ml/s with 23.9psi vacuum @17:49:57.93 Cartridge 54 sampling 331/2000ml at 0.37ml/s with 22.8psi vacuum @17:50:57.99 Cartridge 54 sampling 352/2000ml at 0.33ml/s with 26.9psi vacuum @17:51:57.93 Cartridge 54 sampling 372/2000ml at 0.30ml/s with 27.4psi vacuum @17:52:57.99 Cartridge 54 sampling 391/2000ml at 0.31ml/s with 23.4psi vacuum @17:53:57.93 Cartridge 54 sampling 409/2000ml at 0.30ml/s with 26.1psi vacuum @17:54:58.18 Cartridge 54 sampling 427/2000ml at 0.29ml/s with 18.7psi vacuum 2.2m deep (17.9psia) @17:55:57.93 Cartridge 54 sampling 444/2000ml at 0.31ml/s with 22.8psi vacuum @17:56:58.03 Cartridge 54 sampling 461/2000ml at 0.26ml/s with 21.5psi vacuum @17:57:58.23 Cartridge 54 sampling 477/2000ml at 0.23ml/s with 26.6psi vacuum @17:58:57.93 Cartridge 54 sampling 492/2000ml at 0.24ml/s with 27.4psi vacuum @17:59:58.25 Cartridge 54 sampling 506/2000ml at 0.26ml/s with 21.4psi vacuum @18:00:57.93 Cartridge 54 sampling 520/2000ml at 0.23ml/s with 28.6psi vacuum @18:01:58.17 Cartridge 54 sampling 534/2000ml at 0.26ml/s with 25.4psi vacuum @18:02:57.93 Cartridge 54 sampling 547/2000ml at 0.21ml/s with 28.4psi vacuum @18:03:58.09 Cartridge 54 sampling 561/2000ml at 0.20ml/s with 31.4psi vacuum @18:04:57.93 Cartridge 54 sampling 574/2000ml at 0.20ml/s with 29.7psi vacuum @18:05:58.09 Cartridge 54 sampling 587/2000ml at 0.21ml/s with 22.2psi vacuum @18:06:57.93 Cartridge 54 sampling 600/2000ml at 0.20ml/s with 31.1psi vacuum @18:07:58.16 Cartridge 54 sampling 612/2000ml at 0.22ml/s with 18.0psi vacuum 4.0m deep (20.6psia) @18:08:57.93 Cartridge 54 sampling 625/2000ml at 0.20ml/s with 32.4psi vacuum @18:09:58.17 Cartridge 54 sampling 637/2000ml at 0.20ml/s with 28.9psi vacuum @18:10:57.93 Cartridge 54 sampling 650/2000ml at 0.24ml/s with 23.2psi vacuum @18:11:17.62 Session Begun @18:11:20.27 -> slots @18:11:52.97 -> Cartridge.release 53 @18:11:55.52 -> exit @18:11:55.61 Session Ended @18:11:58.21 Cartridge 54 sampling 662/2000ml at 0.20ml/s with 32.9psi vacuum @18:12:57.93 Cartridge 54 sampling 675/2000ml at 0.21ml/s with 18.1psi vacuum @18:13:58.09 Cartridge 54 sampling 687/2000ml at 0.20ml/s with 38.4psi vacuum @18:14:58.01 Cartridge 54 sampling 699/2000ml at 0.20ml/s with 37.0psi vacuum @18:15:58.09 Cartridge 54 sampling 712/2000ml at 0.23ml/s with 21.3psi vacuum 6.1m deep (23.8psia) @18:16:32.28 Flow averaged 0.31ml/s over 38:27 @18:16:32.47 Waiting up to 9 seconds for vacuum to reduce from 26.3 to 18.0psi @18:16:33.27 Relieved excess vacuum in 10.1 seconds @18:16:33.41 SamplePump.seek 2000ml,1:47:17 @18:16:37.28 Flow estimate reduced to 0.25ml/s @18:16:37.31 Waiting up to 41 seconds for vacuum to reduce from 39.3 to 18.0psi @18:16:37.89 Relieved excess vacuum in 1.4 seconds @18:16:37.92 SamplePump.seek 2000ml,1:47:14 @18:16:46.79 Flow estimate reduced to 0.20ml/s @18:16:46.81 Intake.close @18:16:46.86 Exhaust.close @18:16:47.09 Toroid.reconfigure :hold @18:16:47.15 Toroid.seek :bypass @18:16:48.78 Toroid.reconfigure :default @18:16:48.84 Toroid.seek :clear @18:16:50.45 Cartridge::Sampler::Clogged in FILTERING -- Cartridge clogged @18:16:50.54 Pressure.coast @18:16:51.10 Sampled 721.3ml @18:16:52.31 SamplePump.setPosition! 0ml @18:16:52.47 Plunge.seek :reagent @18:16:52.56 Bypass.open @18:16:53.35 Intake.open @18:16:53.58 Sample loop closed. Ready to Deliver Reagents @18:16:53.60 Archive.seek :sealed @18:16:54.58 SamplePump.reconfigure :slow @18:16:54.64 SamplePump.seek -3ml @18:16:57.25 Archive.seek :evac @18:17:01.36 delay 1 second @18:17:02.53 SamplePump.seek -2.8ml @18:17:03.27 delay 1 second @18:17:04.36 SamplePump.seek -3.3ml @18:17:05.69 delay 1 second @18:17:06.72 Exhaust.open @18:17:07.40 Bypass.close @18:17:08.31 SamplePump.seek -35.3ml @18:17:11.30 delay 1:00 @18:18:11.35 Archive.seek :clear @18:18:17.17 delay 3 seconds @18:18:20.28 Intake.close @18:18:20.32 Exhaust.close @18:18:28.49 RNAlater soak for 10:00 @18:28:28.54 Archive.seek :tVent @18:28:30.36 delay 1 second @18:28:31.42 Archive.seek :evac @18:28:47.38 delay 10 seconds @18:28:57.42 Sample preserved and evacuated @18:28:58.14
Carousel.seek 54 @18:28:58.64 Plunge.seek :locked @18:28:59.80 Archive.seek :clear @18:29:00.51 Air.seek :clear @18:29:00.67 Lysis1.seek :clear @18:29:17.21
Archive.coast @18:29:17.33 Air.coast @18:29:17.41 Lysis1.coast @18:29:17.53 Lysis2.coast @18:29:17.62 Plunge.coast @18:29:17.66 SamplePump.coast @18:29:17.72 Toroid.coast @18:29:17.87 Carousel.coast @18:29:18.13 Gate.power :motor,:OFF @18:29:18.16 Gate.power :core,:OFF @18:29:18.19 Gate.power :analytic1,:OFF @18:29:18.23 Gate.power :analytic2,:OFF @18:29:18.27 Gate.power :analytic3,:OFF @18:29:18.29 Gate.power :main,:OFF @18:29:18.30 Safely stopped and ready to power off @18:29:19.09 Gate.power :main,:ON,limit to 5000mA @18:29:19.11 Gate.power :analytic1,:ON,limit to 5000mA @18:29:19.12 Gate.power :analytic2,:ON,limit to 5000mA @18:29:19.13 Gate.power :analytic3,:ON,limit to 5000mA @18:29:19.22 Powering core bus @18:29:21.50 Archive.configure :default @18:29:21.65 Air.configure :default @18:29:21.76 Lysis1.configure :default @18:29:21.87 Lysis2.configure :default @18:29:22.02 Plunge.configure :default @18:29:22.16 SamplePump.configure :default @18:29:22.83 ExtraPump.configure :default @18:29:22.98 Toroid.configure :default @18:29:23.12 Carousel.configure :default @18:29:23.26 Pressure.configure :default @18:29:23.31 All elf controllers running firmware version 4.61 @18:29:23.33 Gate.power :motor,:ON,limit to 5000mA @18:29:23.56 Archive.reconfigure :default @18:29:24.88 Archive.home.jog 0 @18:29:25.55 Toroid.reconfigure :default @18:29:25.76 Lysis2.reconfigure :default @18:29:25.81 Lysis1.reconfigure :default @18:29:25.98 Plunge.reconfigure :default @18:29:26.06 Air.reconfigure :default @18:29:26.44 Toroid.home.jog 0 @18:29:26.53 Lysis2.home.jog 0 @18:29:26.65 Lysis1.home.jog 0 @18:29:26.69 Plunge.home.jog 0 @18:29:26.90 Air.home.jog 0 @18:29:27.85 Archive.reconfigure :homing @18:29:28.32 Archive.jog 1369 @18:29:28.48 Toroid.reconfigure :homing @18:29:28.52 Lysis1.reconfigure :homing @18:29:28.56 Air.reconfigure :homing @18:29:28.69 Plunge.reconfigure :homing @18:29:28.75 Lysis2.reconfigure :homing @18:29:29.25 Lysis1.jog 1361 @18:29:29.43 Toroid.jog 629 @18:29:29.48 Plunge.jog 845 @18:29:29.67 Air.jog 1357 @18:29:29.78 Lysis2.jog 1353 @18:29:30.00 Archive.setRawPosition! 50 @18:29:30.04 Archive.seek :home @18:29:30.32 Archive.reconfigure :default @18:29:30.65 Plunge.setRawPosition! 50 @18:29:30.70 Plunge.seek :home @18:29:30.84 Lysis1.setRawPosition! 50 @18:29:30.89 Lysis1.seek :home @18:29:31.03 Air.setRawPosition! 50 @18:29:31.05 Plunge.reconfigure :default @18:29:31.16 Air.seek :home @18:29:31.26 Lysis1.reconfigure :default @18:29:31.28 Lysis2.setRawPosition! 50 @18:29:31.37 Lysis2.seek :home @18:29:31.49 Air.reconfigure :default @18:29:31.69 Lysis2.reconfigure :default @18:29:31.92 Toroid.setRawPosition! 0 @18:29:31.97 Toroid.seek :home @18:29:32.07 Toroid.reconfigure :default @18:29:32.21
Toroid.seek :clear @18:29:33.83 Carousel.reconfigure :default @18:29:33.88 Carousel.home.jog 0 @18:29:34.64 Carousel.seek 54 @18:29:35.03 Selecting Cartridge 53 @18:29:35.71 Carousel.seek 53 @18:29:36.49 Lysis1.seek :locked @18:29:39.75 Plunge.seek :locked @18:29:39.89 Air.seek :locked @18:29:44.12
Mounted archive Cartridge @18:29:45.25 Duration of filtering limited to 4:05:00 @18:29:46.07 Priming sample loop w/160ml, bypass w/40ml @18:29:46.15 SamplePump.setPosition! 0ml @18:29:46.23 Bypass.open @18:29:47.90 Intake.open @18:29:48.90 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @18:29:51.48 Vacuum tare is 18% of full scale @18:29:51.51 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @18:29:51.72 Pressure.reconfigure :default @18:29:51.78 Pressure.hold @18:29:52.00 SamplePump.reconfigure :prime @18:29:52.06 SamplePump.seek 40ml,3:51 @18:30:35.25 Pressure.coast @18:30:35.30 SamplePump.setPosition! 0ml @18:30:35.34 Exhaust.open @18:30:36.74 Bypass.close @18:30:38.30 Pressure.reconfigure :default @18:30:38.36 Pressure.hold @18:30:38.63 SamplePump.reconfigure :prime @18:30:38.70 SamplePump.seek 160ml,13:51 @18:33:30.23 Pressure.coast @18:33:30.29 Bypass.open @18:33:30.98 Intake.close @18:33:31.85 Toroid.reconfigure :hold @18:33:31.91 Toroid.seek :crack @18:33:33.66 Toroid.seek :cartridge @18:33:36.02 Toroid.reconfigure :default @18:33:36.22 Toroid.seek :clear @18:33:38.36 Pressurizing cartridge to 26psi above 14.5psia @18:33:38.49 SamplePump.setPosition! 0ml @18:33:38.59 Intake.open @18:33:39.39 Bypass.close @18:33:40.09 Exhaust.close @18:33:40.94 Pressure.reconfigure :default @18:33:40.99 Pressure.hold @18:33:41.22 SamplePump.reconfigure :charge @18:33:41.28 SamplePump.seek 2.3ml,42 seconds @18:33:51.50 Pressure.coast @18:33:51.55 Intake.close @18:33:52.51 Pressurized cartridge with 0.477ml to 26.0psi @18:33:52.53 Isolated @33.3psia (18.8psi) @18:33:52.67 Checking for cartridge leaks from 33.3psia @18:33:52.76 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @18:33:57.84 Toroid.reconfigure :hold @18:33:57.90 Toroid.seek :bypass @18:33:59.52 Toroid.reconfigure :default @18:33:59.58 Toroid.seek :clear @18:34:01.19 Cartridge::Sampler::Leaky in FILTERING -- Isolated cartridge pressure fell by 23% to 25.6psia @18:34:01.21 Retry #1 of 2 @18:34:01.23 Cmd::SpareCartridge in FILTERING -- 0 spares remaining @18:34:01.26 Cartridge::None in FILTERING -- No dry cartridges remain @18:34:01.28 Cartridge::None in FILTERING -- No dry cartridges remain @18:34:01.48
Cartridge Load: {[59, 58] => { Type: :archive_bac, State: :processed, filtered: 400}, 52..1 => { Type: :archive_bac, State: :reserved, was: :dry}, 53 => { Type: :archive_bac, State: :leaky}, 54 => { Type: :archive_bac, State: :processed, filtered: 721.3}, 55 => { Type: :archive_bac, State: :processed, filtered: 622.6}, 56 => { Type: :archive_bac, State: :processed, filtered: 387.7}, 57 => { Type: :archive_bac, State: :processed, filtered: 1000}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 0} @18:34:01.50 Thread::ChildDied in MAIN -- :FILTERING @18:34:01.58 SamplePump.coast @19:22:19.48PDT10-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @19:22:19.50 >BREAK< ESP3 Gateway v1.11 for cartridge bus @19:22:25.43 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @19:22:26.05 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @19:22:30.97 Polling Can every 2:00 @19:22:31.07 Gate.power :analytic1,:ON,limit to 5000mA @19:22:31.08 Gate.power :analytic2,:ON,limit to 5000mA @19:22:31.09 Gate.power :analytic3,:ON,limit to 5000mA @19:22:31.76 Archive.configure :default @19:22:32.01 Archive.hold @19:22:32.24 Archive.configure :default @19:22:32.77 Air.configure :default @19:22:32.93 Air.hold @19:22:33.03 Air.configure :default @19:22:33.43 Lysis1.configure :default @19:22:33.62 Lysis1.hold @19:22:33.77 Lysis1.configure :default @19:22:34.04 Lysis2.configure :default @19:22:34.15 Lysis2.hold @19:22:34.32 Lysis2.configure :default @19:22:34.69 Plunge.configure :default @19:22:34.82 Plunge.hold @19:22:35.03 Plunge.configure :default @19:22:35.33 SamplePump.configure :default @19:22:35.70 ExtraPump.configure :default @19:22:36.08 Toroid.configure :default @19:22:36.15 Toroid.hold @19:22:36.30 Toroid.configure :default @19:22:36.51 Carousel.configure :default @19:22:36.63 Carousel.hold @19:22:36.78 Carousel.configure :default @19:22:37.29 Pressure.configure :default @19:22:37.35 All elf controllers running firmware version 4.61 @19:22:37.43 Intake.close @19:22:37.49 Exhaust.close @19:22:37.58 Bypass.close @19:22:37.74
Loading /home/esp/esp2/bin/processAll mission @19:22:38.01 Sampling default goal volumes @19:22:38.02 Cartridge group: dry Spares: {spare: 0} @19:22:38.68 Carousel.seek 53 @19:22:39.54 Mounted archive Cartridge @19:22:39.59 Selecting Cartridge 53 @19:22:40.20 Duration of filtering limited to 4:05:00 @19:22:40.50 Priming sample loop w/160ml, bypass w/40ml @19:22:40.52 SamplePump.setPosition! 0ml @19:22:40.56 Bypass.open @19:22:42.12 Intake.open @19:22:43.24 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @19:22:45.87 Vacuum tare is 18% of full scale @19:22:45.93 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @19:22:46.43 Pressure.reconfigure :default @19:22:46.74 Pressure.hold @19:22:47.05 SamplePump.reconfigure :prime @19:22:47.11 SamplePump.seek 40ml,3:51 @19:23:30.30 Pressure.coast @19:23:30.49 SamplePump.setPosition! 0ml @19:23:30.56 Exhaust.open @19:23:32.11 Bypass.close @19:23:33.84 Pressure.reconfigure :default @19:23:33.90 Pressure.hold @19:23:34.17 SamplePump.reconfigure :prime @19:23:34.69 SamplePump.seek 160ml,13:51 @19:26:26.14 Pressure.coast @19:26:26.33 Bypass.open @19:26:27.13 Intake.close @19:26:28.03 Toroid.reconfigure :hold @19:26:28.10 Toroid.seek :crack @19:26:29.68 Toroid.seek :cartridge @19:26:30.61 Toroid.reconfigure :default @19:26:30.67 Toroid.seek :clear @19:26:32.97 Pressurizing cartridge to 26psi above 14.4psia @19:26:33.01 SamplePump.setPosition! 0ml @19:26:33.15 Intake.open @19:26:33.99 Bypass.close @19:26:34.76 Exhaust.close @19:26:35.75 Pressure.reconfigure :default @19:26:35.84 Pressure.hold @19:26:36.47 SamplePump.reconfigure :charge @19:26:36.57 SamplePump.seek 2.3ml,42 seconds @19:26:44.87 Pressure.coast @19:26:44.99 Intake.close @19:26:45.89 Pressurized cartridge with 0.306ml to 26.0psi @19:26:45.91 Isolated @31.5psia (17.2psi) @19:26:45.96 Checking for cartridge leaks from 31.5psia @19:26:46.02 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @19:26:54.25 Cartridge pressure fell 11% to 27.9psia @19:26:54.49 Sampling 2000.0ml @19:26:54.51 SamplePump.setPosition! 0ml @19:26:54.62 Exhaust.open @19:26:55.44 Intake.open @19:26:56.80 Pressure.reconfigure :default @19:26:56.86 Pressure.hold @19:26:57.13 SamplePump.reconfigure :sample @19:26:57.20 SamplePump.seek 2000ml,2:47:12 @19:26:59.68 Cartridge 53 sampling 1/2000ml at 0.51ml/s with 21.6psi vacuum @19:27:59.80 Cartridge 53 sampling 31/2000ml at 0.47ml/s with 26.3psi vacuum @19:28:59.68 Cartridge 53 sampling 60/2000ml at 0.48ml/s with 20.2psi vacuum @19:29:59.76 Cartridge 53 sampling 89/2000ml at 0.49ml/s with 27.5psi vacuum @19:30:59.68 Cartridge 53 sampling 119/2000ml at 0.51ml/s with 23.1psi vacuum @19:31:59.95 Cartridge 53 sampling 149/2000ml at 0.51ml/s with 19.2psi vacuum @19:32:59.92 Cartridge 53 sampling 178/2000ml at 0.48ml/s with 25.9psi vacuum @19:33:59.68 Cartridge 53 sampling 206/2000ml at 0.48ml/s with 26.6psi vacuum @19:34:59.89 Cartridge 53 sampling 234/2000ml at 0.47ml/s with 22.9psi vacuum @19:35:59.68 Cartridge 53 sampling 262/2000ml at 0.46ml/s with 25.6psi vacuum @19:36:59.94 Cartridge 53 sampling 289/2000ml at 0.44ml/s with 28.0psi vacuum @19:37:59.68 Cartridge 53 sampling 315/2000ml at 0.45ml/s with 24.6psi vacuum @19:38:59.82 Cartridge 53 sampling 341/2000ml at 0.43ml/s with 26.7psi vacuum @19:39:59.68 Cartridge 53 sampling 366/2000ml at 0.39ml/s with 29.5psi vacuum @19:40:59.84 Cartridge 53 sampling 390/2000ml at 0.42ml/s with 25.1psi vacuum @19:41:59.68 Cartridge 53 sampling 414/2000ml at 0.38ml/s with 29.2psi vacuum @19:42:59.85 Cartridge 53 sampling 437/2000ml at 0.37ml/s with 24.8psi vacuum @19:43:59.68 Cartridge 53 sampling 459/2000ml at 0.36ml/s with 29.2psi vacuum @19:44:59.94 Cartridge 53 sampling 481/2000ml at 0.36ml/s with 20.5psi vacuum 2.7m deep (18.7psia) @19:45:59.68 Cartridge 53 sampling 503/2000ml at 0.35ml/s with 28.3psi vacuum @19:46:59.82 Cartridge 53 sampling 524/2000ml at 0.35ml/s with 28.4psi vacuum @19:47:59.68 Cartridge 53 sampling 544/2000ml at 0.35ml/s with 18.3psi vacuum 1.6m deep (17.1psia) @19:48:59.96 Cartridge 53 sampling 565/2000ml at 0.34ml/s with 27.9psi vacuum @19:49:59.96 Cartridge 53 sampling 585/2000ml at 0.32ml/s with 16.8psi vacuum 5.8m deep (23.2psia) @19:51:00.00 Cartridge 53 sampling 605/2000ml at 0.32ml/s with 18.0psi vacuum @19:51:59.94 Cartridge 53 sampling 624/2000ml at 0.32ml/s with 29.1psi vacuum @19:52:59.68 Cartridge 53 sampling 642/2000ml at 0.28ml/s with 29.7psi vacuum @19:53:59.96 Cartridge 53 sampling 661/2000ml at 0.32ml/s with 24.3psi vacuum @19:54:59.68 Cartridge 53 sampling 679/2000ml at 0.32ml/s with 22.9psi vacuum @19:55:59.83 Cartridge 53 sampling 696/2000ml at 0.28ml/s with 18.1psi vacuum 1.7m deep (17.2psia) @19:56:59.68 Cartridge 53 sampling 713/2000ml at 0.26ml/s with 29.8psi vacuum @19:57:59.88 Cartridge 53 sampling 729/2000ml at 0.28ml/s with 18.2psi vacuum 3.2m deep (19.4psia) @19:58:59.68 Cartridge 53 sampling 745/2000ml at 0.25ml/s with 29.6psi vacuum @19:59:59.82 Cartridge 53 sampling 761/2000ml at 0.24ml/s with 25.9psi vacuum @20:00:59.68 Cartridge 53 sampling 776/2000ml at 0.27ml/s with 19.1psi vacuum @20:01:59.84 Cartridge 53 sampling 792/2000ml at 0.26ml/s with 29.3psi vacuum @20:02:59.68 Cartridge 53 sampling 807/2000ml at 0.29ml/s with 24.1psi vacuum @20:03:59.87 Cartridge 53 sampling 821/2000ml at 0.20ml/s with 25.0psi vacuum @20:04:59.68 Cartridge 53 sampling 836/2000ml at 0.25ml/s with 28.0psi vacuum @20:05:59.83 Cartridge 53 sampling 850/2000ml at 0.23ml/s with 14.3psi vacuum @20:06:59.68 Cartridge 53 sampling 865/2000ml at 0.28ml/s with 25.6psi vacuum @20:08:00.06 Cartridge 53 sampling 879/2000ml at 0.28ml/s with 23.0psi vacuum @20:08:59.98 Cartridge 53 sampling 893/2000ml at 0.20ml/s with 30.4psi vacuum @20:09:59.68 Cartridge 53 sampling 907/2000ml at 0.27ml/s with 26.5psi vacuum @20:10:59.95 Cartridge 53 sampling 921/2000ml at 0.20ml/s with 30.1psi vacuum @20:11:59.68 Cartridge 53 sampling 935/2000ml at 0.20ml/s with 32.1psi vacuum @20:13:00.01 Cartridge 53 sampling 949/2000ml at 0.21ml/s with 16.8psi vacuum 5.1m deep (22.2psia) @20:13:59.77 Cartridge 53 sampling 963/2000ml at 0.20ml/s with 31.2psi vacuum @20:14:59.85 Cartridge 53 sampling 976/2000ml at 0.20ml/s with 31.9psi vacuum @20:15:59.68 Cartridge 53 sampling 990/2000ml at 0.20ml/s with 32.5psi vacuum @20:16:59.91 Cartridge 53 sampling 1003/2000ml at 0.21ml/s with 31.3psi vacuum @20:17:59.68 Cartridge 53 sampling 1016/2000ml at 0.24ml/s with 24.0psi vacuum @20:18:59.93 Cartridge 53 sampling 1029/2000ml at 0.20ml/s with 31.8psi vacuum @20:19:59.68 Cartridge 53 sampling 1042/2000ml at 0.24ml/s with 22.4psi vacuum @20:20:59.88 Cartridge 53 sampling 1055/2000ml at 0.20ml/s with 33.2psi vacuum @20:21:59.68 Cartridge 53 sampling 1068/2000ml at 0.20ml/s with 27.7psi vacuum @20:22:59.80 Cartridge 53 sampling 1081/2000ml at 0.20ml/s with 24.5psi vacuum @20:23:59.68 Cartridge 53 sampling 1094/2000ml at 0.20ml/s with 33.1psi vacuum @20:25:00.00 Cartridge 53 sampling 1107/2000ml at 0.23ml/s with 21.2psi vacuum 6.2m deep (23.9psia) @20:25:59.68 Cartridge 53 sampling 1119/2000ml at 0.23ml/s with 25.5psi vacuum @20:26:59.68 Cartridge 53 sampling 1132/2000ml at 0.20ml/s with 33.7psi vacuum @20:27:59.98 Cartridge 53 sampling 1145/2000ml at 0.20ml/s with 33.7psi vacuum @20:28:59.68 Cartridge 53 sampling 1157/2000ml at 0.20ml/s with 34.9psi vacuum @20:29:59.90 Cartridge 53 sampling 1170/2000ml at 0.20ml/s with 37.3psi vacuum @20:30:59.68 Cartridge 53 sampling 1182/2000ml at 0.21ml/s with 29.3psi vacuum @20:31:59.91 Cartridge 53 sampling 1195/2000ml at 0.22ml/s with 22.5psi vacuum 6.2m deep (23.9psia) @20:32:59.68 Cartridge 53 sampling 1207/2000ml at 0.20ml/s with 34.9psi vacuum @20:33:59.89 Cartridge 53 sampling 1220/2000ml at 0.20ml/s with 33.9psi vacuum @20:34:59.67 Cartridge 53 sampling 1232/2000ml at 0.21ml/s with 19.9psi vacuum @20:35:59.82 Cartridge 53 sampling 1244/2000ml at 0.21ml/s with 26.6psi vacuum @20:36:59.68 Cartridge 53 sampling 1257/2000ml at 0.21ml/s with 25.1psi vacuum @20:37:59.83 Cartridge 53 sampling 1269/2000ml at 0.20ml/s with 32.1psi vacuum 1.7m deep (17.3psia) @20:39:00.23 Cartridge 53 sampling 1281/2000ml at 0.21ml/s with 18.4psi vacuum 5.7m deep (23.1psia) @20:39:59.85 Cartridge 53 sampling 1294/2000ml at 0.21ml/s with 23.5psi vacuum 2.2m deep (18.0psia) @20:40:59.68 Cartridge 53 sampling 1306/2000ml at 0.20ml/s with 23.0psi vacuum @20:41:59.73 Cartridge 53 sampling 1318/2000ml at 0.20ml/s with 41.9psi vacuum @20:42:59.68 Cartridge 53 sampling 1331/2000ml at 0.20ml/s with 32.6psi vacuum @20:43:59.93 Cartridge 53 sampling 1343/2000ml at 0.20ml/s with 40.3psi vacuum @20:45:00.01 Cartridge 53 sampling 1355/2000ml at 0.20ml/s with 44.0psi vacuum @20:45:59.68 Cartridge 53 sampling 1367/2000ml at 0.20ml/s with 28.6psi vacuum @20:47:00.12 Cartridge 53 sampling 1380/2000ml at 0.20ml/s with 42.2psi vacuum @20:47:59.68 Cartridge 53 sampling 1392/2000ml at 0.20ml/s with 47.7psi vacuum @20:48:14.95 Flow averaged 0.29ml/s over 1:21:13 @20:48:15.11 Waiting up to 1:01 for vacuum to reduce from 47.3 to 18.0psi @20:48:15.82 Relieved excess vacuum in 5.9 seconds @20:48:15.95 SamplePump.seek 2000ml,50:56 @20:48:20.97 Flow estimate reduced to 0.23ml/s @20:48:21.00 Waiting up to 1:03 for vacuum to reduce from 47.9 to 18.0psi @20:48:21.83 Relieved excess vacuum in 2.7 seconds @20:48:21.86 SamplePump.seek 2000ml,50:51 @20:48:26.82 Flow estimate reduced to 0.18ml/s @20:48:26.85 Intake.close @20:48:26.90 Exhaust.close @20:48:27.11 Toroid.reconfigure :hold @20:48:27.17 Toroid.seek :bypass @20:48:28.80 Toroid.reconfigure :default @20:48:28.86 Toroid.seek :clear @20:48:30.47 Cartridge::Sampler::Clogged in FILTERING -- Cartridge clogged @20:48:30.49 Pressure.coast @20:48:30.70 Sampled 1396.8ml @20:48:31.45 SamplePump.setPosition! 0ml @20:48:31.53 Plunge.seek :reagent @20:48:31.73 Bypass.open @20:48:32.57 Sample loop closed. Ready to Deliver Reagents @20:48:32.65 Intake.open @20:48:32.78 Archive.seek :sealed @20:48:33.81 SamplePump.reconfigure :slow @20:48:33.98 SamplePump.seek -3ml @20:48:36.49 Archive.seek :evac @20:48:40.70 delay 1 second @20:48:41.89 SamplePump.seek -2.8ml @20:48:42.64 delay 1 second @20:48:43.79 SamplePump.seek -3.3ml @20:48:45.22 delay 1 second @20:48:46.25 Exhaust.open @20:48:46.94 Bypass.close @20:48:47.77 SamplePump.seek -35.3ml @20:48:50.52 delay 1:00 @20:49:50.56 Archive.seek :clear @20:49:56.51 delay 3 seconds @20:49:59.60 Intake.close @20:49:59.65 Exhaust.close @20:50:07.77 RNAlater soak for 10:00 @21:00:07.82 Archive.seek :tVent @21:00:09.48 delay 1 second @21:00:10.60 Archive.seek :evac @21:00:26.56 delay 10 seconds @21:00:36.60 Sample preserved and evacuated @21:00:36.83
Carousel.seek 53 @21:00:37.04 Plunge.seek :locked @21:00:38.31 Archive.seek :clear @21:00:38.63 Air.seek :clear @21:00:39.07 Lysis1.seek :clear @21:00:55.54
Archive.coast @21:00:55.58 Air.coast @21:00:55.62 Lysis1.coast @21:00:55.66 Lysis2.coast @21:00:55.71 Plunge.coast @21:00:55.75 SamplePump.coast @21:00:55.97 Toroid.coast @21:00:56.06 Carousel.coast @21:00:56.42 Gate.power :motor,:OFF @21:00:56.45 Gate.power :core,:OFF @21:00:56.51 Gate.power :analytic1,:OFF @21:00:56.53 Gate.power :analytic2,:OFF @21:00:56.57 Gate.power :analytic3,:OFF @21:00:56.59 Gate.power :main,:OFF @21:00:56.61 Safely stopped and ready to power off @21:00:57.62 Cartridge Load: {53 => { Type: :archive_bac, State: :processed, filtered: 1396.8}, [59, 58] => { Type: :archive_bac, State: :processed, filtered: 400}, 52..1 => { Type: :archive_bac, State: :reserved, was: :dry}, 54 => { Type: :archive_bac, State: :processed, filtered: 721.3}, 55 => { Type: :archive_bac, State: :processed, filtered: 622.6}, 56 => { Type: :archive_bac, State: :processed, filtered: 387.7}, 57 => { Type: :archive_bac, State: :processed, filtered: 1000}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 0} @21:00:57.68 Cartridge::None in MAIN -- Not enough dry cartridges remain @21:56:24.06PDT10-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @21:56:24.08 >BREAK< ESP3 Gateway v1.11 for cartridge bus @21:56:30.34 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @21:56:31.20 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @21:56:36.14 Polling Can every 2:00 @21:56:36.33 Loading /home/esp/esp2/bin/processAll mission /home/esp/esp2/bin/processAll: unrecognized option `--daily=11PM' @21:56:37.58 Cartridge Load: {53 => { Type: :archive_bac, State: :processed, filtered: 1396.8}, 56 => { Type: :archive_bac, State: :processed, filtered: 387.7}, [59, 58] => { Type: :archive_bac, State: :processed, filtered: 400}, 52 => { Type: :archive_bac, State: :dry}, 54 => { Type: :archive_bac, State: :processed, filtered: 721.3}, 55 => { Type: :archive_bac, State: :processed, filtered: 622.6}, 57 => { Type: :archive_bac, State: :processed, filtered: 1000}, 51..1 => { Type: :archive_bac, State: :reserved, was: :dry}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 0} @21:56:37.60 GetoptLong::InvalidOption in MAIN -- unrecognized option `--daily=11PM' @22:01:08.98PDT10-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @22:01:09.18 >BREAK< ESP3 Gateway v1.11 for cartridge bus @22:01:12.34 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @22:01:12.64 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @22:01:21.00 Polling Can every 2:00 @22:01:21.05 Loading /home/esp/esp2/bin/processAll mission @22:01:21.32 Sampling default goal volumes @22:01:21.34 Cartridge group: dry Spares: {spare: 0} Schedule: ["11pm"] @22:01:21.39 delayUntil "23:00:00.00" @23:00:00.12 Gate.power :main,:ON,limit to 5000mA @23:00:00.13 Gate.power :analytic1,:ON,limit to 5000mA @23:00:00.14 Gate.power :analytic2,:ON,limit to 5000mA @23:00:00.16 Gate.power :analytic3,:ON,limit to 5000mA @23:00:00.19 Powering core bus @23:00:02.73 Archive.configure :default @23:00:03.08 Air.configure :default @23:00:03.32 Lysis1.configure :default @23:00:03.57 Lysis2.configure :default @23:00:03.96 Plunge.configure :default @23:00:04.23 SamplePump.configure :default @23:00:04.54 ExtraPump.configure :default @23:00:04.86 Toroid.configure :default @23:00:05.16 Carousel.configure :default @23:00:05.47 Pressure.configure :default @23:00:05.52 All elf controllers running firmware version 4.61 @23:00:05.54 Gate.power :motor,:ON,limit to 5000mA @23:00:06.47 Archive.reconfigure :default @23:00:06.62 Air.reconfigure :default @23:00:06.79 Archive.home.jog 0 @23:00:06.89 Air.home.jog 0 @23:00:07.18 Plunge.reconfigure :default @23:00:07.20 Lysis2.reconfigure :default @23:00:07.23 Toroid.reconfigure :default @23:00:07.30 Lysis1.reconfigure :default @23:00:07.42 Lysis2.home.jog 0 @23:00:07.50 Toroid.home.jog 0 @23:00:07.53 Plunge.home.jog 0 @23:00:07.56 Lysis1.home.jog 0 @23:00:07.90 Archive.reconfigure :homing @23:00:08.01 Air.reconfigure :homing @23:00:08.14 Archive.jog 1367 @23:00:08.16 Air.jog 1359 @23:00:08.37 Toroid.reconfigure :homing @23:00:08.41 Plunge.reconfigure :homing @23:00:08.45 Lysis1.reconfigure :homing @23:00:08.50 Lysis2.reconfigure :homing @23:00:08.65 Toroid.jog 636 @23:00:08.71 Plunge.jog 848 @23:00:08.78 Lysis2.jog 1353 @23:00:08.81 Lysis1.jog 1354 @23:00:09.33 Archive.setRawPosition! 50 @23:00:09.38 Archive.seek :home @23:00:09.46 Air.setRawPosition! 50 @23:00:09.51 Air.seek :home @23:00:09.67 Archive.reconfigure :default @23:00:09.77 Plunge.setRawPosition! 50 @23:00:09.81 Plunge.seek :home @23:00:09.90 Air.reconfigure :default @23:00:10.02 Lysis2.setRawPosition! 50 @23:00:10.07 Lysis2.seek :home @23:00:10.17 Plunge.reconfigure :default @23:00:10.19 Lysis1.setRawPosition! 50 @23:00:10.28 Lysis1.seek :home @23:00:10.40 Lysis2.reconfigure :default @23:00:10.58 Lysis1.reconfigure :default @23:00:11.00 Toroid.setRawPosition! 0 @23:00:11.16 Toroid.seek :home @23:00:11.44 Toroid.reconfigure :default @23:00:11.98
Toroid.seek :clear @23:00:13.76 Carousel.reconfigure :default @23:00:13.85 Carousel.home.jog 0 @23:00:14.55 Carousel.seek 53 @23:00:14.86 Selecting Cartridge 52 @23:00:16.56 Carousel.seek 52 @23:00:17.07 Lysis1.seek :locked @23:00:20.30 Plunge.seek :locked @23:00:20.45 Air.seek :locked @23:00:24.85
Mounted archive Cartridge @23:00:25.43 Duration of filtering limited to 4:05:00 @23:00:26.16 Priming sample loop w/160ml, bypass w/40ml @23:00:26.18 SamplePump.setPosition! 0ml @23:00:26.22 Bypass.open @23:00:27.72 Intake.open @23:00:28.61 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @23:00:31.30 Vacuum tare is 18% of full scale @23:00:31.36 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @23:00:31.83 Pressure.reconfigure :default @23:00:32.01 Pressure.hold @23:00:32.59 SamplePump.reconfigure :prime @23:00:32.77 SamplePump.seek 40ml,3:51 @23:01:15.98 Pressure.coast @23:01:16.03 SamplePump.setPosition! 0ml @23:01:16.07 Exhaust.open @23:01:17.51 Bypass.close @23:01:19.35 Pressure.reconfigure :default @23:01:19.49 Pressure.hold @23:01:20.17 SamplePump.reconfigure :prime @23:01:20.31 SamplePump.seek 160ml,13:51 @23:04:11.88 Pressure.coast @23:04:12.14 Bypass.open @23:04:13.02 Intake.close @23:04:14.17 Toroid.reconfigure :hold @23:04:14.24 Toroid.seek :crack @23:04:15.91 Toroid.seek :cartridge @23:04:16.84 Toroid.reconfigure :default @23:04:16.91 Toroid.seek :clear @23:04:18.76 Pressurizing cartridge to 26psi above 14.5psia @23:04:18.78 SamplePump.setPosition! 0ml @23:04:18.81 Intake.open @23:04:19.52 Bypass.close @23:04:20.23 Exhaust.close @23:04:21.08 Pressure.reconfigure :default @23:04:21.14 Pressure.hold @23:04:21.60 SamplePump.reconfigure :charge @23:04:21.78 SamplePump.seek 2.3ml,42 seconds @23:04:30.57 Pressure.coast @23:04:30.62 Intake.close @23:04:31.65 Pressurized cartridge with 0.347ml to 26.1psi @23:04:31.67 Isolated @32.7psia (18.2psi) @23:04:31.82 Checking for cartridge leaks from 32.7psia @23:04:32.00 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @23:04:41.18 Cartridge pressure fell 16% to 27.5psia @23:04:41.43 Sampling 2000.0ml @23:04:41.56 SamplePump.setPosition! 0ml @23:04:41.73 Exhaust.open @23:04:42.49 Intake.open @23:04:43.68 Pressure.reconfigure :default @23:04:43.80 Pressure.hold @23:04:44.43 SamplePump.reconfigure :sample @23:04:44.57 SamplePump.seek 2000ml,2:47:12 @23:04:46.58 Cartridge 52 sampling 1/2000ml at 0.49ml/s with 26.0psi vacuum @23:05:46.74 Cartridge 52 sampling 30/2000ml at 0.50ml/s with 27.3psi vacuum @23:06:46.58 Cartridge 52 sampling 60/2000ml at 0.50ml/s with 21.0psi vacuum @23:07:46.76 Cartridge 52 sampling 90/2000ml at 0.51ml/s with 22.7psi vacuum @23:08:46.58 Cartridge 52 sampling 120/2000ml at 0.51ml/s with 26.9psi vacuum @23:09:46.78 Cartridge 52 sampling 150/2000ml at 0.51ml/s with 26.3psi vacuum @23:10:46.58 Cartridge 52 sampling 181/2000ml at 0.50ml/s with 21.8psi vacuum @23:11:46.69 Cartridge 52 sampling 212/2000ml at 0.52ml/s with 27.9psi vacuum @23:12:46.58 Cartridge 52 sampling 244/2000ml at 0.53ml/s with 25.9psi vacuum @23:13:46.73 Cartridge 52 sampling 274/2000ml at 0.51ml/s with 27.5psi vacuum @23:14:46.57 Cartridge 52 sampling 305/2000ml at 0.51ml/s with 24.7psi vacuum @23:15:46.66 Cartridge 52 sampling 335/2000ml at 0.49ml/s with 17.8psi vacuum 3.8m deep (20.4psia) @23:16:46.58 Cartridge 52 sampling 366/2000ml at 0.50ml/s with 22.9psi vacuum @23:17:46.73 Cartridge 52 sampling 396/2000ml at 0.49ml/s with 25.8psi vacuum @23:18:46.84 Cartridge 52 sampling 426/2000ml at 0.51ml/s with 25.7psi vacuum @23:19:46.58 Cartridge 52 sampling 455/2000ml at 0.49ml/s with 17.7psi vacuum 2.8m deep (18.8psia) @23:20:46.90 Cartridge 52 sampling 485/2000ml at 0.48ml/s with 18.6psi vacuum 2.0m deep (17.6psia) @23:21:46.59 Cartridge 52 sampling 514/2000ml at 0.49ml/s with 26.1psi vacuum @23:22:46.80 Cartridge 52 sampling 543/2000ml at 0.47ml/s with 28.9psi vacuum @23:23:46.58 Cartridge 52 sampling 571/2000ml at 0.49ml/s with 25.4psi vacuum @23:24:46.74 Cartridge 52 sampling 599/2000ml at 0.47ml/s with 17.3psi vacuum 5.4m deep (22.6psia) @23:25:46.58 Cartridge 52 sampling 627/2000ml at 0.47ml/s with 26.1psi vacuum @23:26:46.73 Cartridge 52 sampling 655/2000ml at 0.46ml/s with 17.8psi vacuum 4.9m deep (22.0psia) @23:27:46.58 Cartridge 52 sampling 683/2000ml at 0.46ml/s with 22.8psi vacuum @23:28:46.80 Cartridge 52 sampling 710/2000ml at 0.44ml/s with 29.4psi vacuum @23:29:46.58 Cartridge 52 sampling 737/2000ml at 0.42ml/s with 29.1psi vacuum @23:30:46.73 Cartridge 52 sampling 764/2000ml at 0.47ml/s with 25.7psi vacuum @23:31:46.58 Cartridge 52 sampling 791/2000ml at 0.42ml/s with 29.3psi vacuum @23:32:46.72 Cartridge 52 sampling 817/2000ml at 0.44ml/s with 23.5psi vacuum @23:33:46.58 Cartridge 52 sampling 844/2000ml at 0.41ml/s with 26.2psi vacuum @23:34:46.99 Cartridge 52 sampling 870/2000ml at 0.44ml/s with 20.4psi vacuum 1.8m deep (17.3psia) @23:35:46.91 Cartridge 52 sampling 895/2000ml at 0.44ml/s with 19.6psi vacuum @23:36:46.77 Cartridge 52 sampling 921/2000ml at 0.44ml/s with 27.6psi vacuum @23:37:46.83 Cartridge 52 sampling 947/2000ml at 0.44ml/s with 24.3psi vacuum @23:38:46.58 Cartridge 52 sampling 972/2000ml at 0.43ml/s with 24.2psi vacuum @23:39:46.94 Cartridge 52 sampling 997/2000ml at 0.41ml/s with 29.0psi vacuum @23:40:46.58 Cartridge 52 sampling 1022/2000ml at 0.42ml/s with 19.5psi vacuum @23:41:48.06 Cartridge 52 sampling 1047/2000ml at 0.42ml/s with 22.5psi vacuum @23:42:46.58 Cartridge 52 sampling 1072/2000ml at 0.40ml/s with 17.5psi vacuum @23:43:46.79 Cartridge 52 sampling 1097/2000ml at 0.40ml/s with 29.4psi vacuum @23:44:46.58 Cartridge 52 sampling 1121/2000ml at 0.42ml/s with 23.9psi vacuum @23:45:46.87 Cartridge 52 sampling 1146/2000ml at 0.42ml/s with 21.5psi vacuum 2.8m deep (18.8psia) @23:46:46.58 Cartridge 52 sampling 1170/2000ml at 0.42ml/s with 24.3psi vacuum @23:47:46.72 Cartridge 52 sampling 1194/2000ml at 0.41ml/s with 29.6psi vacuum @23:48:46.58 Cartridge 52 sampling 1218/2000ml at 0.39ml/s with 28.2psi vacuum @23:49:46.77 Cartridge 52 sampling 1242/2000ml at 0.40ml/s with 27.0psi vacuum @23:50:46.58 Cartridge 52 sampling 1266/2000ml at 0.40ml/s with 27.8psi vacuum @23:51:46.78 Cartridge 52 sampling 1290/2000ml at 0.40ml/s with 26.2psi vacuum @23:52:46.58 Cartridge 52 sampling 1313/2000ml at 0.39ml/s with 28.2psi vacuum @23:53:46.67 Cartridge 52 sampling 1336/2000ml at 0.38ml/s with 29.6psi vacuum @23:54:46.59 Cartridge 52 sampling 1359/2000ml at 0.35ml/s with 15.0psi vacuum @23:55:46.69 Cartridge 52 sampling 1383/2000ml at 0.40ml/s with 21.1psi vacuum @23:56:46.58 Cartridge 52 sampling 1405/2000ml at 0.39ml/s with 22.1psi vacuum @23:57:46.99 Cartridge 52 sampling 1428/2000ml at 0.39ml/s with 23.5psi vacuum 1.6m deep (17.1psia) @23:58:46.79 Cartridge 52 sampling 1451/2000ml at 0.38ml/s with 23.0psi vacuum 1.9m deep (17.4psia) @23:59:46.81 Cartridge 52 sampling 1473/2000ml at 0.38ml/s with 18.7psi vacuum 4.3m deep (21.0psia) @00:00:47.10PDT11-Jun-21 Cartridge 52 sampling 1496/2000ml at 0.35ml/s with 29.7psi vacuum @00:01:46.58 Cartridge 52 sampling 1518/2000ml at 0.33ml/s with 24.7psi vacuum @00:02:46.75 Cartridge 52 sampling 1539/2000ml at 0.36ml/s with 28.8psi vacuum @00:03:46.58 Cartridge 52 sampling 1561/2000ml at 0.38ml/s with 24.2psi vacuum @00:04:46.74 Cartridge 52 sampling 1583/2000ml at 0.35ml/s with 16.1psi vacuum 4.1m deep (20.8psia) @00:05:46.58 Cartridge 52 sampling 1604/2000ml at 0.38ml/s with 22.2psi vacuum @00:06:46.75 Cartridge 52 sampling 1626/2000ml at 0.35ml/s with 16.4psi vacuum 6.1m deep (23.6psia) @00:07:46.58 Cartridge 52 sampling 1647/2000ml at 0.35ml/s with 28.2psi vacuum @00:08:46.78 Cartridge 52 sampling 1668/2000ml at 0.36ml/s with 22.3psi vacuum 2.2m deep (18.0psia) @00:09:46.58 Cartridge 52 sampling 1689/2000ml at 0.34ml/s with 29.6psi vacuum @00:10:46.66 Cartridge 52 sampling 1709/2000ml at 0.35ml/s with 28.5psi vacuum @00:11:46.58 Cartridge 52 sampling 1730/2000ml at 0.34ml/s with 13.8psi vacuum 2.2m deep (18.0psia) @00:12:46.74 Cartridge 52 sampling 1750/2000ml at 0.35ml/s with 19.1psi vacuum @00:13:46.58 Cartridge 52 sampling 1771/2000ml at 0.34ml/s with 17.1psi vacuum 1.9m deep (17.5psia) @00:14:47.11 Cartridge 52 sampling 1791/2000ml at 0.31ml/s with 14.0psi vacuum 5.3m deep (22.5psia) @00:15:46.91 Cartridge 52 sampling 1811/2000ml at 0.36ml/s with 23.2psi vacuum @00:16:46.69 Cartridge 52 sampling 1831/2000ml at 0.31ml/s with 12.9psi vacuum @00:17:46.75 Cartridge 52 sampling 1851/2000ml at 0.32ml/s with 13.6psi vacuum 6.6m deep (24.5psia) @00:18:46.58 Cartridge 52 sampling 1871/2000ml at 0.34ml/s with 26.2psi vacuum @00:19:46.88 Cartridge 52 sampling 1890/2000ml at 0.34ml/s with 20.6psi vacuum 2.4m deep (18.3psia) @00:20:46.58 Cartridge 52 sampling 1910/2000ml at 0.30ml/s with 29.8psi vacuum @00:21:46.72 Cartridge 52 sampling 1929/2000ml at 0.35ml/s with 22.0psi vacuum @00:22:46.58 Cartridge 52 sampling 1948/2000ml at 0.34ml/s with 24.1psi vacuum @00:23:46.76 Cartridge 52 sampling 1967/2000ml at 0.28ml/s with 13.4psi vacuum @00:24:46.58 Cartridge 52 sampling 1986/2000ml at 0.34ml/s with 20.6psi vacuum 1.9m deep (17.4psia) @00:25:31.31 Pressure.coast @00:25:31.45 Intake.close @00:25:31.52 Exhaust.close @00:25:31.67 Toroid.reconfigure :hold @00:25:31.73 Toroid.seek :bypass @00:25:33.35 Toroid.reconfigure :default @00:25:33.41 Toroid.seek :clear @00:25:35.02 Sampled 2000.0ml @00:25:36.66 SamplePump.setPosition! 0ml @00:25:36.83 Plunge.seek :reagent @00:25:36.96 Bypass.open @00:25:37.81 Intake.open @00:25:38.05 Sample loop closed. Ready to Deliver Reagents @00:25:38.07 Archive.seek :sealed @00:25:38.83 SamplePump.reconfigure :slow @00:25:39.04 SamplePump.seek -3ml @00:25:41.72 Archive.seek :evac @00:25:45.82 delay 1 second @00:25:47.00 SamplePump.seek -2.8ml @00:25:47.84 delay 1 second @00:25:48.96 SamplePump.seek -3.3ml @00:25:50.37 delay 1 second @00:25:51.40 Exhaust.open @00:25:52.07 Bypass.close @00:25:52.85 SamplePump.seek -35.3ml @00:25:55.75 delay 1:00 @00:26:55.87 Archive.seek :clear @00:27:01.53 delay 3 seconds @00:27:04.62 Intake.close @00:27:04.66 Exhaust.close @00:27:13.00 RNAlater soak for 10:00 @00:37:13.03 Archive.seek :tVent @00:37:14.74 delay 1 second @00:37:15.79 Archive.seek :evac @00:37:31.77 delay 10 seconds @00:37:41.83 Sample preserved and evacuated @00:37:42.51
Carousel.seek 52 @00:37:43.20 Plunge.seek :locked @00:37:44.41 Archive.seek :clear @00:37:44.74 Air.seek :clear @00:37:45.09 Lysis1.seek :clear @00:38:01.67
Archive.coast @00:38:01.74 Air.coast @00:38:01.85 Lysis1.coast @00:38:01.97 Lysis2.coast @00:38:02.06 Plunge.coast @00:38:02.22 SamplePump.coast @00:38:02.30 Toroid.coast @00:38:02.42 Carousel.coast @00:38:02.83 Gate.power :motor,:OFF @00:38:02.91 Gate.power :core,:OFF @00:38:02.97 Gate.power :analytic1,:OFF @00:38:02.98 Gate.power :analytic2,:OFF @00:38:03.00 Gate.power :analytic3,:OFF @00:38:03.01 Gate.power :main,:OFF @00:38:03.05 Safely stopped and ready to power off @00:38:03.92 Cartridge Load: {53 => { Type: :archive_bac, State: :processed, filtered: 1396.8}, 56 => { Type: :archive_bac, State: :processed, filtered: 387.7}, [59, 58] => { Type: :archive_bac, State: :processed, filtered: 400}, 52 => { Type: :archive_bac, State: :processed, filtered: 2000.0}, 54 => { Type: :archive_bac, State: :processed, filtered: 721.3}, 55 => { Type: :archive_bac, State: :processed, filtered: 622.6}, 57 => { Type: :archive_bac, State: :processed, filtered: 1000}, 51..1 => { Type: :archive_bac, State: :reserved, was: :dry}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 0} @00:38:04.03 Cartridge::None in MAIN -- Not enough dry cartridges remain @07:05:15.68PDT11-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @07:05:15.72 >BREAK< ESP3 Gateway v1.11 for cartridge bus @07:05:18.74 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @07:05:19.24 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @07:05:27.53 Polling Can every 2:00 @07:05:27.59 Loading /home/esp/esp2/bin/processAll mission @07:05:27.86 Sampling default goal volumes @07:05:27.88 Cartridge group: dry Spares: {spare: 0} Schedule: ["11AM"] @07:05:27.94 delayUntil "11:00:00.00" @09:29:20.75 Session Begun @09:30:17.31 -> GoodNews.email "Yes, we have e-mail (but no banannas :-)" @09:30:17.33 NameError in brent -- uninitialized constant GoodNews @09:34:03.23 -> mailcfg = File.which "phasecfg.rb", ESP::Path @09:34:21.59 -> ESP.require('email', mailcfg, 'yelp') @09:34:33.26 -> module ESP flushMail = Thread(:flushMail) {Email.flush} unless flushMail.join 30 Log.comment "Sending Email" flushMail.join 300 or Log.recordException Email::Error.new "flush stuck!" end uploading = Log.uploading #wait for last upload to complete unless uploading.join 30 Log.comment "Uploading Logs" @09:34:33.33 SyntaxError in brent -- compile error brent:20: syntax error, unexpected $end, expecting kEND Log.comment "Uploading Logs" ^ @09:35:11.21 -> module ESP flushMail = Thread(:flushMail) {Email.flush} unless flushMail.join 30 Log.comment "Sending Email" flushMail.join 300 or Log.recordException Email::Error.new "flush stuck!" end uploading = Log.uploading #wait for last upload to complete unless uploading.join 30 Log.comment "Uploading Logs" uploading.join 600 or Log.recordException Log::Error.new "upload stuck!" end end atExit {flushMail} end @09:35:11.30 SyntaxError in brent -- compile error brent:39: syntax error, unexpected kEND, expecting $end @09:35:23.28 -> GoodNews.email "Yes, we have e-mail (but no banannas :-)" @09:35:23.30 GoodNews.email "Yes, we have e-mail (but no banannas :-)" @09:36:46.30 upsync @09:37:47.77 -> GoodNews.xray @09:41:49.72 -> exit @09:41:49.74 Session Ended @09:44:17.76 Session Begun @09:44:29.56 -> processAll '--help' @09:44:29.63 NoMethodError in brent -- undefined method `processAll' for Object:Class @09:45:16.34 -> exit @09:45:16.40 Session Ended @11:00:00.14
Gate.power :main,:ON,limit to 5000mA @11:00:00.22 Gate.power :analytic1,:ON,limit to 5000mA @11:00:00.26 Gate.power :analytic2,:ON,limit to 5000mA @11:00:00.31 Gate.power :analytic3,:ON,limit to 5000mA @11:00:00.38 Powering core bus @11:00:02.67 Archive.configure :default @11:00:02.82 Air.configure :default @11:00:02.93 Lysis1.configure :default @11:00:03.04 Lysis2.configure :default @11:00:03.18 Plunge.configure :default @11:00:03.33 SamplePump.configure :default @11:00:03.46 ExtraPump.configure :default @11:00:03.61 Toroid.configure :default @11:00:03.74 Carousel.configure :default @11:00:03.88 Pressure.configure :default @11:00:03.93 All elf controllers running firmware version 4.61 @11:00:03.96 Gate.power :motor,:ON,limit to 5000mA @11:00:04.10 Archive.reconfigure :default @11:00:04.40 Archive.home.jog 0 @11:00:04.67 Toroid.reconfigure :default @11:00:04.69 Plunge.reconfigure :default @11:00:04.73 Lysis1.reconfigure :default @11:00:04.77 Air.reconfigure :default @11:00:04.83 Lysis2.reconfigure :default @11:00:05.04 Plunge.home.jog 0 @11:00:05.07 Toroid.home.jog 0 @11:00:05.11 Lysis1.home.jog 0 @11:00:05.20 Lysis2.home.jog 0 @11:00:05.28 Air.home.jog 0 @11:00:05.91 Archive.reconfigure :homing @11:00:06.20 Archive.jog 1375 @11:00:07.05 Plunge.reconfigure :homing @11:00:07.16 Lysis2.reconfigure :homing @11:00:07.33 Toroid.reconfigure :homing @11:00:07.59 Plunge.jog 855 @11:00:07.79 Lysis2.jog 1358 @11:00:07.90 Toroid.jog 621 @11:00:08.40 Archive.setRawPosition! 50 @11:00:08.49 Lysis1.reconfigure :homing @11:00:08.51 Air.reconfigure :homing @11:00:08.85 Archive.seek :home @11:00:09.04 Air.jog 1371 @11:00:09.08 Lysis1.jog 1347 @11:00:09.67 Plunge.setRawPosition! 50 @11:00:09.78 Lysis2.setRawPosition! 50 @11:00:09.80 Archive.reconfigure :default @11:00:10.04 Plunge.seek :home @11:00:10.12 Lysis2.seek :home @11:00:10.67 Air.setRawPosition! 50 @11:00:10.70 Toroid.setRawPosition! 0 @11:00:10.79 Air.seek :home @11:00:10.92 Toroid.seek :home @11:00:11.55 Lysis1.setRawPosition! 50 @11:00:11.59 Air.reconfigure :default @11:00:11.60 Toroid.reconfigure :default @11:00:11.64 Lysis2.reconfigure :default @11:00:11.69 Plunge.reconfigure :default @11:00:11.89 Lysis1.seek :home @11:00:12.24 Lysis1.reconfigure :default @11:00:12.63
Toroid.seek :clear @11:00:14.25 Carousel.reconfigure :default @11:00:14.30 Carousel.home.jog 0 @11:00:14.79 Carousel.seek 52 @11:00:14.97 Selecting Cartridge 51 @11:00:15.19 Carousel.seek 51 @11:00:15.71 Lysis1.seek :locked @11:00:19.09 Plunge.seek :locked @11:00:19.48 Air.seek :locked @11:00:23.71
Mounted archive Cartridge @11:00:24.01 Duration of filtering limited to 4:05:00 @11:00:24.30 Priming sample loop w/160ml, bypass w/40ml @11:00:24.32 SamplePump.setPosition! 0ml @11:00:24.35 Bypass.open @11:00:25.92 Intake.open @11:00:27.17 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @11:00:29.85 Vacuum tare is 18% of full scale @11:00:29.94 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @11:00:30.44 Pressure.reconfigure :default @11:00:30.49 Pressure.hold @11:00:30.71 SamplePump.reconfigure :prime @11:00:30.77 SamplePump.seek 40ml,3:51 @11:01:14.03 Pressure.coast @11:01:14.12 SamplePump.setPosition! 0ml @11:01:14.32 Exhaust.open @11:01:15.76 Bypass.close @11:01:17.53 Pressure.reconfigure :default @11:01:17.59 Pressure.hold @11:01:17.86 SamplePump.reconfigure :prime @11:01:17.92 SamplePump.seek 160ml,13:51 @11:04:09.75 Pressure.coast @11:04:09.80 Bypass.open @11:04:10.55 Intake.close @11:04:11.50 Toroid.reconfigure :hold @11:04:11.56 Toroid.seek :crack @11:04:13.14 Toroid.seek :cartridge @11:04:14.07 Toroid.reconfigure :default @11:04:14.13 Toroid.seek :clear @11:04:16.82 Pressurizing cartridge to 26psi above 14.4psia @11:04:16.87 SamplePump.setPosition! 0ml @11:04:16.95 Intake.open @11:04:17.85 Bypass.close @11:04:18.56 Exhaust.close @11:04:19.54 Pressure.reconfigure :default @11:04:19.62 Pressure.hold @11:04:19.98 SamplePump.reconfigure :charge @11:04:20.08 SamplePump.seek 2.3ml,42 seconds @11:04:30.38 Pressure.coast @11:04:30.42 Intake.close @11:04:31.28 Pressurized cartridge with 0.481ml to 26.2psi @11:04:31.30 Isolated @36.0psia (21.6psi) @11:04:31.31 Checking for cartridge leaks from 36.0psia @11:04:31.33 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @11:04:38.42 Cartridge pressure fell 15% to 30.5psia @11:04:38.59 Sampling 2000.0ml @11:04:38.67 SamplePump.setPosition! 0ml @11:04:38.72 Exhaust.open @11:04:39.40 Intake.open @11:04:40.34 Pressure.reconfigure :default @11:04:40.41 Pressure.hold @11:04:40.69 SamplePump.reconfigure :sample @11:04:40.76 SamplePump.seek 2000ml,2:47:12 @11:04:43.40 Cartridge 51 sampling 1/2000ml at 0.51ml/s with 24.8psi vacuum @11:05:43.66 Cartridge 51 sampling 32/2000ml at 0.51ml/s with 24.3psi vacuum @11:06:43.87 Cartridge 51 sampling 63/2000ml at 0.52ml/s with 24.0psi vacuum @11:07:43.40 Cartridge 51 sampling 95/2000ml at 0.51ml/s with 28.1psi vacuum @11:08:43.81 Cartridge 51 sampling 127/2000ml at 0.55ml/s with 27.3psi vacuum @11:09:43.40 Cartridge 51 sampling 160/2000ml at 0.52ml/s with 28.7psi vacuum @11:10:43.53 Cartridge 51 sampling 193/2000ml at 0.55ml/s with 18.0psi vacuum 3.8m deep (20.4psia) @11:11:43.41 Cartridge 51 sampling 226/2000ml at 0.54ml/s with 28.5psi vacuum @11:12:43.60 Cartridge 51 sampling 258/2000ml at 0.55ml/s with 28.3psi vacuum @11:13:43.41 Cartridge 51 sampling 291/2000ml at 0.55ml/s with 21.4psi vacuum 1.6m deep (17.1psia) @11:14:43.57 Cartridge 51 sampling 323/2000ml at 0.50ml/s with 29.5psi vacuum @11:15:43.41 Cartridge 51 sampling 355/2000ml at 0.54ml/s with 22.7psi vacuum @11:16:43.52 Cartridge 51 sampling 387/2000ml at 0.52ml/s with 29.0psi vacuum @11:17:43.41 Cartridge 51 sampling 418/2000ml at 0.52ml/s with 15.9psi vacuum 3.9m deep (20.4psia) @11:18:43.57 Cartridge 51 sampling 450/2000ml at 0.50ml/s with 15.4psi vacuum @11:19:43.41 Cartridge 51 sampling 481/2000ml at 0.51ml/s with 18.5psi vacuum @11:20:43.47 Cartridge 51 sampling 512/2000ml at 0.51ml/s with 16.1psi vacuum 5.2m deep (22.4psia) @11:21:43.41 Cartridge 51 sampling 543/2000ml at 0.49ml/s with 26.0psi vacuum @11:22:43.83 Cartridge 51 sampling 573/2000ml at 0.48ml/s with 15.4psi vacuum @11:23:43.60 Cartridge 51 sampling 604/2000ml at 0.48ml/s with 30.4psi vacuum @11:24:43.41 Cartridge 51 sampling 634/2000ml at 0.47ml/s with 23.9psi vacuum @11:25:43.81 Cartridge 51 sampling 664/2000ml at 0.52ml/s with 25.0psi vacuum @11:26:43.41 Cartridge 51 sampling 694/2000ml at 0.47ml/s with 30.4psi vacuum @11:27:43.51 Cartridge 51 sampling 724/2000ml at 0.49ml/s with 27.4psi vacuum @11:28:43.41 Cartridge 51 sampling 753/2000ml at 0.49ml/s with 16.7psi vacuum 3.1m deep (19.3psia) @11:29:43.68 Cartridge 51 sampling 782/2000ml at 0.46ml/s with 29.9psi vacuum @11:30:43.41 Cartridge 51 sampling 811/2000ml at 0.48ml/s with 27.8psi vacuum @11:31:43.55 Cartridge 51 sampling 840/2000ml at 0.45ml/s with 27.3psi vacuum @11:32:43.41 Cartridge 51 sampling 868/2000ml at 0.49ml/s with 26.4psi vacuum @11:33:43.50 Cartridge 51 sampling 896/2000ml at 0.46ml/s with 16.6psi vacuum 2.8m deep (18.8psia) @11:34:43.54 Cartridge 51 sampling 924/2000ml at 0.48ml/s with 18.9psi vacuum 3.6m deep (20.0psia) @11:35:43.56 Cartridge 51 sampling 952/2000ml at 0.47ml/s with 26.3psi vacuum @11:36:43.41 Cartridge 51 sampling 979/2000ml at 0.47ml/s with 19.1psi vacuum 2.6m deep (18.5psia) @11:37:43.47 Cartridge 51 sampling 1006/2000ml at 0.44ml/s with 29.2psi vacuum @11:38:43.41 Cartridge 51 sampling 1033/2000ml at 0.46ml/s with 26.2psi vacuum @11:39:43.67 Cartridge 51 sampling 1060/2000ml at 0.46ml/s with 25.9psi vacuum @11:40:43.72 Cartridge 51 sampling 1086/2000ml at 0.44ml/s with 26.6psi vacuum @11:41:43.40 Cartridge 51 sampling 1112/2000ml at 0.44ml/s with 17.6psi vacuum @11:42:43.77 Cartridge 51 sampling 1138/2000ml at 0.41ml/s with 30.0psi vacuum @11:43:43.41 Cartridge 51 sampling 1164/2000ml at 0.43ml/s with 28.0psi vacuum @11:44:43.54 Cartridge 51 sampling 1189/2000ml at 0.41ml/s with 28.1psi vacuum @11:45:43.41 Cartridge 51 sampling 1215/2000ml at 0.43ml/s with 19.3psi vacuum 2.2m deep (17.9psia) @11:46:43.69 Cartridge 51 sampling 1240/2000ml at 0.41ml/s with 29.4psi vacuum @11:47:43.41 Cartridge 51 sampling 1264/2000ml at 0.42ml/s with 24.9psi vacuum @11:48:43.57 Cartridge 51 sampling 1289/2000ml at 0.39ml/s with 20.1psi vacuum 2.0m deep (17.6psia) @11:49:43.41 Cartridge 51 sampling 1313/2000ml at 0.38ml/s with 18.0psi vacuum @11:50:43.54 Cartridge 51 sampling 1337/2000ml at 0.38ml/s with 29.4psi vacuum @11:51:43.41 Cartridge 51 sampling 1361/2000ml at 0.41ml/s with 27.3psi vacuum @11:52:43.56 Cartridge 51 sampling 1385/2000ml at 0.37ml/s with 22.4psi vacuum @11:53:43.40 Cartridge 51 sampling 1408/2000ml at 0.40ml/s with 25.5psi vacuum @11:54:43.47 Cartridge 51 sampling 1431/2000ml at 0.40ml/s with 26.5psi vacuum @11:55:43.40 Cartridge 51 sampling 1454/2000ml at 0.40ml/s with 26.6psi vacuum @11:56:43.79 Cartridge 51 sampling 1477/2000ml at 0.40ml/s with 24.4psi vacuum @11:57:43.67 Cartridge 51 sampling 1500/2000ml at 0.39ml/s with 23.6psi vacuum @11:58:43.40 Cartridge 51 sampling 1523/2000ml at 0.33ml/s with 23.6psi vacuum @11:59:43.60 Cartridge 51 sampling 1545/2000ml at 0.36ml/s with 15.8psi vacuum 6.7m deep (24.6psia) @12:00:43.41 Cartridge 51 sampling 1568/2000ml at 0.36ml/s with 28.1psi vacuum @12:01:43.56 Cartridge 51 sampling 1590/2000ml at 0.36ml/s with 24.0psi vacuum @12:02:43.41 Cartridge 51 sampling 1612/2000ml at 0.39ml/s with 26.7psi vacuum @12:03:43.67 Cartridge 51 sampling 1633/2000ml at 0.32ml/s with 16.4psi vacuum 2.8m deep (18.9psia) @12:04:43.41 Cartridge 51 sampling 1655/2000ml at 0.37ml/s with 24.5psi vacuum @12:05:43.53 Cartridge 51 sampling 1676/2000ml at 0.36ml/s with 16.2psi vacuum 6.2m deep (23.8psia) @12:06:43.40 Cartridge 51 sampling 1697/2000ml at 0.35ml/s with 20.2psi vacuum @12:07:43.59 Cartridge 51 sampling 1718/2000ml at 0.37ml/s with 25.2psi vacuum @12:08:43.41 Cartridge 51 sampling 1740/2000ml at 0.32ml/s with 30.1psi vacuum @12:09:43.55 Cartridge 51 sampling 1760/2000ml at 0.32ml/s with 30.2psi vacuum @12:10:43.41 Cartridge 51 sampling 1781/2000ml at 0.35ml/s with 25.5psi vacuum @12:11:43.55 Cartridge 51 sampling 1801/2000ml at 0.37ml/s with 21.9psi vacuum @12:12:43.41 Cartridge 51 sampling 1822/2000ml at 0.29ml/s with 17.6psi vacuum @12:13:43.58 Cartridge 51 sampling 1842/2000ml at 0.31ml/s with 14.9psi vacuum @12:14:43.40 Cartridge 51 sampling 1862/2000ml at 0.36ml/s with 23.1psi vacuum @12:15:43.56 Cartridge 51 sampling 1882/2000ml at 0.31ml/s with 30.2psi vacuum @12:16:43.41 Cartridge 51 sampling 1901/2000ml at 0.30ml/s with 29.0psi vacuum @12:17:43.66 Cartridge 51 sampling 1921/2000ml at 0.36ml/s with 25.6psi vacuum @12:18:43.41 Cartridge 51 sampling 1940/2000ml at 0.34ml/s with 23.8psi vacuum @12:19:43.48 Cartridge 51 sampling 1960/2000ml at 0.30ml/s with 29.4psi vacuum @12:20:43.40 Cartridge 51 sampling 1978/2000ml at 0.33ml/s with 23.7psi vacuum @12:21:43.66 Cartridge 51 sampling 1997/2000ml at 0.34ml/s with 21.0psi vacuum @12:21:52.52 Pressure.coast @12:21:52.96 Intake.close @12:21:53.09 Exhaust.close @12:21:53.34 Toroid.reconfigure :hold @12:21:53.39 Toroid.seek :bypass @12:21:55.01 Toroid.reconfigure :default @12:21:55.07 Toroid.seek :clear @12:21:56.67 Sampled 2000.0ml @12:21:57.15 SamplePump.setPosition! 0ml @12:21:57.18 Plunge.seek :reagent @12:21:57.27 Bypass.open @12:21:58.05 Intake.open @12:21:58.40 Sample loop closed. Ready to Deliver Reagents @12:21:58.51 Archive.seek :sealed @12:21:59.18 SamplePump.reconfigure :slow @12:21:59.25 SamplePump.seek -3ml @12:22:03.04 Archive.seek :evac @12:22:05.99 delay 1 second @12:22:07.41 SamplePump.seek -2.8ml @12:22:08.17 delay 1 second @12:22:09.36 SamplePump.seek -3.3ml @12:22:10.75 delay 1 second @12:22:11.81 Exhaust.open @12:22:12.52 Bypass.close @12:22:13.31 SamplePump.seek -35.3ml @12:22:17.09 delay 1:00 @12:23:17.14 Archive.seek :clear @12:23:22.08 delay 3 seconds @12:23:25.32 Intake.close @12:23:25.48 Exhaust.close @12:23:34.31 RNAlater soak for 10:00 @12:33:34.37 Archive.seek :tVent @12:33:36.02 delay 1 second @12:33:37.07 Archive.seek :evac @12:33:53.09 delay 10 seconds @12:34:03.13 Sample preserved and evacuated @12:34:03.36
Carousel.seek 51 @12:34:03.58 Plunge.seek :locked @12:34:04.81 Archive.seek :clear @12:34:05.28 Air.seek :clear @12:34:05.62 Lysis1.seek :clear @12:34:22.06
Archive.coast @12:34:22.10 Air.coast @12:34:22.14 Lysis1.coast @12:34:22.18 Lysis2.coast @12:34:22.23 Plunge.coast @12:34:22.27 SamplePump.coast @12:34:22.31 Toroid.coast @12:34:22.36 Carousel.coast @12:34:22.56 Gate.power :motor,:OFF @12:34:22.57 Gate.power :core,:OFF @12:34:22.59 Gate.power :analytic1,:OFF @12:34:22.60 Gate.power :analytic2,:OFF @12:34:22.61 Gate.power :analytic3,:OFF @12:34:22.62 Gate.power :main,:OFF @12:34:22.63 Safely stopped and ready to power off @12:34:23.52 Cartridge Load: {51 => { Type: :archive_bac, State: :processed, filtered: 2000.0}, 52 => { Type: :archive_bac, State: :processed, filtered: 2000}, 53 => { Type: :archive_bac, State: :processed, filtered: 1396.8}, 56 => { Type: :archive_bac, State: :processed, filtered: 387.7}, [59, 58] => { Type: :archive_bac, State: :processed, filtered: 400}, 54 => { Type: :archive_bac, State: :processed, filtered: 721.3}, 55 => { Type: :archive_bac, State: :processed, filtered: 622.6}, 57 => { Type: :archive_bac, State: :processed, filtered: 1000}, 50..1 => { Type: :archive_bac, State: :reserved, was: :dry}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 0} @12:34:23.79 BadNews.email "Not enough dry cartridges remain",Subject: "Cartridge::None in MAIN" @13:01:30.53PDT11-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @13:01:30.56 >BREAK< ESP3 Gateway v1.11 for cartridge bus @13:01:33.59 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @13:01:33.90 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @13:01:42.21 Polling Can every 2:00 @13:01:42.26 Loading /home/esp/esp2/bin/processAll mission @13:01:43.51 Sampling default goal volumes @13:01:43.55 GoodNews.email "Cartridge group: dry\nSpares: {spare: 1}\nSchedule: [\"12am\", \"12pm\"]\n",Subject: "Begun processAll --schedule=12am,12pm --checkpoint --mail" @13:01:43.66
delayUntil "00:00:00.00PDT12-Jun-21" @13:01:53.32 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT12-Jun-21" @13:02:08.14 upsync @13:02:36.30 upsync @00:00:00.16PDT12-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.21 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.25 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.36 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.45 Powering core bus @00:00:03.11 Archive.configure :default @00:00:03.40 Air.configure :default @00:00:03.69 Lysis1.configure :default @00:00:03.95 Lysis2.configure :default @00:00:04.29 Plunge.configure :default @00:00:04.59 SamplePump.configure :default @00:00:04.82 ExtraPump.configure :default @00:00:05.06 Toroid.configure :default @00:00:05.53 Carousel.configure :default @00:00:05.89 Pressure.configure :default @00:00:05.94 All elf controllers running firmware version 4.61 @00:00:05.96 Gate.power :motor,:ON,limit to 5000mA @00:00:06.15 Archive.reconfigure :default @00:00:06.55 Lysis1.reconfigure :default @00:00:06.62 Archive.home.jog 0 @00:00:06.65 Plunge.reconfigure :default @00:00:06.68 Lysis2.reconfigure :default @00:00:06.73 Toroid.reconfigure :default @00:00:06.78 Air.reconfigure :default @00:00:07.00 Lysis1.home.jog 0 @00:00:07.07 Plunge.home.jog 0 @00:00:07.09 Lysis2.home.jog 0 @00:00:07.12 Toroid.home.jog 0 @00:00:07.14 Air.home.jog 0 @00:00:07.86 Air.reconfigure :homing @00:00:07.91 Toroid.reconfigure :homing @00:00:07.94 Plunge.reconfigure :homing @00:00:07.98 Lysis1.reconfigure :homing @00:00:08.06 Lysis2.reconfigure :homing @00:00:08.08 Archive.reconfigure :homing @00:00:08.29 Toroid.jog 636 @00:00:08.32 Plunge.jog 848 @00:00:08.38 Air.jog 1359 @00:00:08.41 Archive.jog 1371 @00:00:08.49 Lysis1.jog 1352 @00:00:08.55 Lysis2.jog 1352 @00:00:09.39 Plunge.setRawPosition! 50 @00:00:09.44 Plunge.seek :home @00:00:09.75 Archive.setRawPosition! 50 @00:00:09.78 Plunge.reconfigure :default @00:00:09.89 Archive.seek :home @00:00:10.01 Lysis1.setRawPosition! 50 @00:00:10.03 Air.setRawPosition! 50 @00:00:10.05 Lysis2.setRawPosition! 50 @00:00:10.26 Lysis1.seek :home @00:00:10.73 Lysis2.seek :home @00:00:10.81 Air.seek :home @00:00:11.60 Toroid.setRawPosition! 0 @00:00:11.76 Lysis2.reconfigure :default @00:00:11.78 Air.reconfigure :default @00:00:11.90 Archive.reconfigure :default @00:00:12.00 Lysis1.reconfigure :default @00:00:12.37 Toroid.seek :home @00:00:12.90 Toroid.reconfigure :default @00:00:13.23
Toroid.seek :clear @00:00:14.87 Carousel.reconfigure :default @00:00:15.00 Carousel.home.jog 0 @00:00:15.67 Carousel.seek 51 @00:00:16.01 Selecting Cartridge 50 @00:00:16.22 Carousel.seek 50 @00:00:16.76 Lysis1.seek :locked @00:00:19.98 Plunge.seek :locked @00:00:20.14 Air.seek :locked @00:00:24.41
Mounted archive Cartridge @00:00:24.59 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 50" @00:00:25.38 Duration of filtering limited to 4:05:00 @00:00:25.67 Priming sample loop w/160ml, bypass w/40ml @00:00:25.69 SamplePump.setPosition! 0ml @00:00:26.18 Bypass.open @00:00:27.84 Intake.open @00:00:28.70 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:31.32 Vacuum tare is 18% of full scale @00:00:31.44 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:31.93 Pressure.reconfigure :default @00:00:32.07 Pressure.hold @00:00:32.55 SamplePump.reconfigure :prime @00:00:32.66 SamplePump.seek 40ml,3:51 @00:00:57.71 upsync @00:01:15.87 Pressure.coast @00:01:15.93 SamplePump.setPosition! 0ml @00:01:15.98 Exhaust.open @00:01:17.42 Bypass.close @00:01:19.30 Pressure.reconfigure :default @00:01:19.38 Pressure.hold @00:01:20.13 SamplePump.reconfigure :prime @00:01:20.20 SamplePump.seek 160ml,13:51 @00:04:11.95 Pressure.coast @00:04:12.35 Bypass.open @00:04:13.09 Intake.close @00:04:14.21 Toroid.reconfigure :hold @00:04:14.32 Toroid.seek :crack @00:04:16.24 Toroid.seek :cartridge @00:04:17.18 Toroid.reconfigure :default @00:04:17.24 Toroid.seek :clear @00:04:19.08 Pressurizing cartridge to 26psi above 14.4psia @00:04:19.10 SamplePump.setPosition! 0ml @00:04:19.13 Intake.open @00:04:19.90 Bypass.close @00:04:20.61 Exhaust.close @00:04:21.78 Pressure.reconfigure :default @00:04:21.95 Pressure.hold @00:04:22.41 SamplePump.reconfigure :charge @00:04:22.56 SamplePump.seek 2.3ml,42 seconds @00:04:32.42 Pressure.coast @00:04:32.52 Intake.close @00:04:33.51 Pressurized cartridge with 0.429ml to 26.0psi @00:04:33.54 Isolated @35.2psia (20.8psi) @00:04:33.59 Checking for cartridge leaks from 35.2psia @00:04:33.64 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:40.32 Cartridge pressure fell 19% to 28.6psia @00:04:40.66 Sampling 2000.0ml @00:04:40.73 SamplePump.setPosition! 0ml @00:04:40.90 Exhaust.open @00:04:41.58 Intake.open @00:04:42.81 Pressure.reconfigure :default @00:04:42.93 Pressure.hold @00:04:43.63 SamplePump.reconfigure :sample @00:04:43.70 SamplePump.seek 2000ml,2:47:12 @00:04:45.70 Cartridge 50 sampling 1/2000ml at 0.50ml/s with 25.7psi vacuum @00:05:45.82 Cartridge 50 sampling 30/2000ml at 0.48ml/s with 20.9psi vacuum @00:06:45.70 Cartridge 50 sampling 59/2000ml at 0.50ml/s with 21.0psi vacuum @00:07:45.85 Cartridge 50 sampling 88/2000ml at 0.49ml/s with 23.5psi vacuum @00:08:45.69 Cartridge 50 sampling 118/2000ml at 0.48ml/s with 18.0psi vacuum @00:09:45.80 Cartridge 50 sampling 148/2000ml at 0.52ml/s with 22.1psi vacuum @00:10:45.70 Cartridge 50 sampling 178/2000ml at 0.49ml/s with 28.7psi vacuum @00:11:46.14 Cartridge 50 sampling 209/2000ml at 0.52ml/s with 25.6psi vacuum @00:12:45.70 Cartridge 50 sampling 240/2000ml at 0.50ml/s with 16.4psi vacuum 3.2m deep (19.4psia) @00:13:45.82 Cartridge 50 sampling 271/2000ml at 0.50ml/s with 17.8psi vacuum 4.7m deep (21.6psia) @00:14:45.69 Cartridge 50 sampling 302/2000ml at 0.51ml/s with 21.7psi vacuum @00:15:45.75 Cartridge 50 sampling 333/2000ml at 0.51ml/s with 28.5psi vacuum @00:16:45.69 Cartridge 50 sampling 363/2000ml at 0.51ml/s with 26.9psi vacuum @00:17:45.93 Cartridge 50 sampling 394/2000ml at 0.50ml/s with 17.5psi vacuum @00:18:45.88 Cartridge 50 sampling 424/2000ml at 0.51ml/s with 26.4psi vacuum @00:19:45.70 Cartridge 50 sampling 455/2000ml at 0.51ml/s with 25.2psi vacuum @00:20:46.05 Cartridge 50 sampling 485/2000ml at 0.49ml/s with 18.6psi vacuum 2.2m deep (18.0psia) @00:21:45.70 Cartridge 50 sampling 515/2000ml at 0.49ml/s with 18.5psi vacuum @00:22:45.95 Cartridge 50 sampling 545/2000ml at 0.48ml/s with 28.5psi vacuum @00:23:45.70 Cartridge 50 sampling 575/2000ml at 0.49ml/s with 26.8psi vacuum @00:24:45.91 Cartridge 50 sampling 605/2000ml at 0.50ml/s with 27.7psi vacuum @00:25:45.69 Cartridge 50 sampling 635/2000ml at 0.49ml/s with 26.2psi vacuum @00:26:46.00 Cartridge 50 sampling 664/2000ml at 0.49ml/s with 26.4psi vacuum @00:27:45.70 Cartridge 50 sampling 693/2000ml at 0.50ml/s with 24.0psi vacuum @00:28:46.02 Cartridge 50 sampling 723/2000ml at 0.48ml/s with 20.5psi vacuum @00:29:45.70 Cartridge 50 sampling 752/2000ml at 0.48ml/s with 24.8psi vacuum @00:30:45.72 Cartridge 50 sampling 781/2000ml at 0.48ml/s with 26.9psi vacuum @00:31:45.69 Cartridge 50 sampling 810/2000ml at 0.50ml/s with 26.6psi vacuum @00:32:45.86 Cartridge 50 sampling 839/2000ml at 0.49ml/s with 24.6psi vacuum @00:33:45.70 Cartridge 50 sampling 867/2000ml at 0.48ml/s with 23.8psi vacuum @00:34:45.82 Cartridge 50 sampling 896/2000ml at 0.47ml/s with 26.1psi vacuum @00:35:45.84 Cartridge 50 sampling 924/2000ml at 0.49ml/s with 26.6psi vacuum @00:36:45.70 Cartridge 50 sampling 953/2000ml at 0.47ml/s with 23.8psi vacuum @00:37:45.99 Cartridge 50 sampling 981/2000ml at 0.48ml/s with 19.7psi vacuum @00:38:45.69 Cartridge 50 sampling 1008/2000ml at 0.44ml/s with 16.6psi vacuum @00:39:45.86 Cartridge 50 sampling 1036/2000ml at 0.47ml/s with 27.0psi vacuum @00:40:45.70 Cartridge 50 sampling 1064/2000ml at 0.46ml/s with 17.5psi vacuum 2.4m deep (18.3psia) @00:41:45.86 Cartridge 50 sampling 1091/2000ml at 0.44ml/s with 17.1psi vacuum 4.8m deep (21.8psia) @00:42:46.15 Cartridge 50 sampling 1118/2000ml at 0.46ml/s with 17.0psi vacuum 2.1m deep (17.8psia) @00:43:45.86 Cartridge 50 sampling 1146/2000ml at 0.46ml/s with 21.9psi vacuum @00:44:45.70 Cartridge 50 sampling 1173/2000ml at 0.42ml/s with 28.9psi vacuum @00:45:45.86 Cartridge 50 sampling 1200/2000ml at 0.42ml/s with 29.1psi vacuum @00:46:45.70 Cartridge 50 sampling 1227/2000ml at 0.46ml/s with 25.0psi vacuum @00:47:45.83 Cartridge 50 sampling 1254/2000ml at 0.46ml/s with 24.6psi vacuum @00:48:45.70 Cartridge 50 sampling 1280/2000ml at 0.45ml/s with 21.3psi vacuum @00:49:45.79 Cartridge 50 sampling 1307/2000ml at 0.45ml/s with 19.1psi vacuum 3.1m deep (19.3psia) @00:50:45.69 Cartridge 50 sampling 1333/2000ml at 0.42ml/s with 28.0psi vacuum @00:51:46.12 Cartridge 50 sampling 1359/2000ml at 0.43ml/s with 28.2psi vacuum @00:52:46.02 Cartridge 50 sampling 1385/2000ml at 0.43ml/s with 28.1psi vacuum @00:53:45.91 Cartridge 50 sampling 1411/2000ml at 0.44ml/s with 26.6psi vacuum @00:54:45.87 Cartridge 50 sampling 1436/2000ml at 0.44ml/s with 26.1psi vacuum @00:55:45.69 Cartridge 50 sampling 1462/2000ml at 0.41ml/s with 29.2psi vacuum @00:56:46.06 Cartridge 50 sampling 1487/2000ml at 0.39ml/s with 19.8psi vacuum 2.1m deep (17.7psia) @00:57:45.70 Cartridge 50 sampling 1512/2000ml at 0.39ml/s with 22.2psi vacuum @00:58:45.89 Cartridge 50 sampling 1538/2000ml at 0.43ml/s with 19.9psi vacuum 1.7m deep (17.3psia) @00:59:45.69 Cartridge 50 sampling 1563/2000ml at 0.43ml/s with 22.0psi vacuum @01:00:45.99 Cartridge 50 sampling 1588/2000ml at 0.41ml/s with 20.4psi vacuum @01:01:45.69 Cartridge 50 sampling 1613/2000ml at 0.42ml/s with 23.0psi vacuum @01:02:45.86 Cartridge 50 sampling 1637/2000ml at 0.42ml/s with 23.5psi vacuum @01:03:45.69 Cartridge 50 sampling 1662/2000ml at 0.41ml/s with 19.7psi vacuum @01:04:45.84 Cartridge 50 sampling 1686/2000ml at 0.39ml/s with 30.1psi vacuum @01:05:45.69 Cartridge 50 sampling 1711/2000ml at 0.40ml/s with 26.6psi vacuum @01:06:45.81 Cartridge 50 sampling 1735/2000ml at 0.39ml/s with 19.1psi vacuum @01:07:45.70 Cartridge 50 sampling 1759/2000ml at 0.40ml/s with 28.5psi vacuum @01:08:45.77 Cartridge 50 sampling 1783/2000ml at 0.40ml/s with 19.7psi vacuum @01:09:45.70 Cartridge 50 sampling 1806/2000ml at 0.41ml/s with 25.7psi vacuum @01:10:45.83 Cartridge 50 sampling 1830/2000ml at 0.37ml/s with 29.6psi vacuum @01:11:45.89 Cartridge 50 sampling 1853/2000ml at 0.35ml/s with 28.1psi vacuum @01:12:45.69 Cartridge 50 sampling 1877/2000ml at 0.40ml/s with 21.5psi vacuum @01:13:45.93 Cartridge 50 sampling 1900/2000ml at 0.36ml/s with 19.7psi vacuum 3.8m deep (20.3psia) @01:14:45.69 Cartridge 50 sampling 1923/2000ml at 0.36ml/s with 15.4psi vacuum @01:15:45.86 Cartridge 50 sampling 1946/2000ml at 0.38ml/s with 28.4psi vacuum @01:16:45.70 Cartridge 50 sampling 1969/2000ml at 0.39ml/s with 20.2psi vacuum @01:17:45.87 Cartridge 50 sampling 1991/2000ml at 0.38ml/s with 28.7psi vacuum @01:18:08.44 Pressure.coast @01:18:08.58 Intake.close @01:18:08.64 Exhaust.close @01:18:08.83 Toroid.reconfigure :hold @01:18:08.89 Toroid.seek :bypass @01:18:10.51 Toroid.reconfigure :default @01:18:10.57 Toroid.seek :clear @01:18:12.23 Sampled 2000.0ml @01:18:13.22 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 50" @01:18:14.17 SamplePump.setPosition! 0ml @01:18:14.24 Plunge.seek :reagent @01:18:14.45 Bypass.open @01:18:15.26 Intake.open @01:18:15.49 Sample loop closed. Ready to Deliver Reagents @01:18:15.51 Archive.seek :sealed @01:18:16.62 SamplePump.reconfigure :slow @01:18:16.84 SamplePump.seek -3ml @01:18:19.16 Archive.seek :evac @01:18:23.57 delay 1 second @01:18:24.82 SamplePump.seek -2.8ml @01:18:25.60 delay 1 second @01:18:26.69 SamplePump.seek -3.3ml @01:18:28.02 delay 1 second @01:18:29.05 Exhaust.open @01:18:29.09 upsync @01:18:29.74 Bypass.close @01:18:30.54 SamplePump.seek -35.3ml @01:18:33.15 delay 1:00 @01:19:33.21 Archive.seek :clear @01:19:39.30 delay 3 seconds @01:19:42.56 Intake.close @01:19:42.61 Exhaust.close @01:19:50.35 RNAlater soak for 10:00 @01:29:50.39 Archive.seek :tVent @01:29:52.17 delay 1 second @01:29:53.21 Archive.seek :evac @01:30:09.24 delay 10 seconds @01:30:19.35 Sample preserved and evacuated @01:30:19.88
Carousel.seek 50 @01:30:20.25 Plunge.seek :locked @01:30:21.48 Archive.seek :clear @01:30:21.91 Air.seek :clear @01:30:22.22 Lysis1.seek :clear @01:30:38.79
Archive.coast @01:30:38.86 Air.coast @01:30:39.02 Lysis1.coast @01:30:39.28 Lysis2.coast @01:30:39.46 Plunge.coast @01:30:39.69 SamplePump.coast @01:30:39.74 Toroid.coast @01:30:39.88 Carousel.coast @01:30:40.25 Gate.power :motor,:OFF @01:30:40.26 Gate.power :core,:OFF @01:30:40.27 Gate.power :analytic1,:OFF @01:30:40.29 Gate.power :analytic2,:OFF @01:30:40.30 Gate.power :analytic3,:OFF @01:30:40.42 Gate.power :main,:OFF @01:30:40.44 Safely stopped and ready to power off @01:30:40.79 delayUntil "12:00:00.00" @01:30:40.81 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @01:30:53.49 upsync @12:00:00.27
Gate.power :main,:ON,limit to 5000mA @12:00:00.33 Gate.power :analytic1,:ON,limit to 5000mA @12:00:00.37 Gate.power :analytic2,:ON,limit to 5000mA @12:00:00.39 Gate.power :analytic3,:ON,limit to 5000mA @12:00:00.51 Powering core bus @12:00:02.86 Archive.configure :default @12:00:03.00 Air.configure :default @12:00:03.10 Lysis1.configure :default @12:00:03.21 Lysis2.configure :default @12:00:03.35 Plunge.configure :default @12:00:03.49 SamplePump.configure :default @12:00:03.63 ExtraPump.configure :default @12:00:03.77 Toroid.configure :default @12:00:03.91 Carousel.configure :default @12:00:04.04 Pressure.configure :default @12:00:04.10 All elf controllers running firmware version 4.61 @12:00:04.12 Gate.power :motor,:ON,limit to 5000mA @12:00:04.32 Archive.reconfigure :default @12:00:04.65 Archive.home.jog 0 @12:00:04.85 Plunge.reconfigure :default @12:00:04.94 Lysis1.reconfigure :default @12:00:04.97 Air.reconfigure :default @12:00:05.02 Lysis2.reconfigure :default @12:00:05.10 Toroid.reconfigure :default @12:00:05.25 Plunge.home.jog 0 @12:00:05.30 Lysis1.home.jog 0 @12:00:05.33 Lysis2.home.jog 0 @12:00:05.35 Air.home.jog 0 @12:00:05.39 Toroid.home.jog 0 @12:00:05.88 Archive.reconfigure :homing @12:00:06.13 Archive.jog 1372 @12:00:06.18 Toroid.reconfigure :homing @12:00:06.21 Lysis2.reconfigure :homing @12:00:06.24 Lysis1.reconfigure :homing @12:00:06.28 Plunge.reconfigure :homing @12:00:06.33 Air.reconfigure :homing @12:00:06.58 Toroid.jog 620 @12:00:06.63 Lysis1.jog 1348 @12:00:06.65 Lysis2.jog 1356 @12:00:06.69 Air.jog 1364 @12:00:06.71 Plunge.jog 849 @12:00:07.55 Archive.setRawPosition! 50 @12:00:07.70 Archive.seek :home @12:00:08.18 Lysis2.setRawPosition! 50 @12:00:08.32 Lysis1.setRawPosition! 50 @12:00:08.35 Air.setRawPosition! 50 @12:00:08.40 Plunge.setRawPosition! 50 @12:00:08.42 Archive.reconfigure :default @12:00:08.94 Lysis1.seek :home @12:00:08.99 Lysis2.seek :home @12:00:09.12 Air.seek :home @12:00:09.23 Plunge.seek :home @12:00:10.03 Plunge.reconfigure :default @12:00:10.12 Lysis2.reconfigure :default @12:00:10.23 Air.reconfigure :default @12:00:10.31 Toroid.setRawPosition! 0 @12:00:10.37 Lysis1.reconfigure :default @12:00:11.00 Toroid.seek :home @12:00:11.36 Toroid.reconfigure :default @12:00:11.69
Toroid.seek :clear @12:00:13.31 Carousel.reconfigure :default @12:00:13.36 Carousel.home.jog 0 @12:00:13.88 Carousel.seek 50 @12:00:14.04 Selecting Cartridge 49 @12:00:14.26 Carousel.seek 49 @12:00:14.77 Lysis1.seek :locked @12:00:18.30 Plunge.seek :locked @12:00:18.71 Air.seek :locked @12:00:22.98
Mounted archive Cartridge @12:00:23.64 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 49" @12:00:23.73 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:00:23.75 Retry #1 of 10 after 10:00 @12:00:24.01 Duration of filtering limited to 4:05:00 @12:00:24.31 Priming sample loop w/160ml, bypass w/40ml @12:00:24.33 SamplePump.setPosition! 0ml @12:00:24.37 Bypass.open @12:00:25.89 Intake.open @12:00:27.04 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:00:29.81 Vacuum tare is 18% of full scale @12:00:29.87 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:00:30.40 Pressure.reconfigure :default @12:00:30.45 Pressure.hold @12:00:30.87 SamplePump.reconfigure :prime @12:00:31.05 SamplePump.seek 40ml,3:51 @12:01:14.32 Pressure.coast @12:01:14.38 SamplePump.setPosition! 0ml @12:01:14.59 Exhaust.open @12:01:16.03 Bypass.close @12:01:17.81 Pressure.reconfigure :default @12:01:17.86 Pressure.hold @12:01:18.14 SamplePump.reconfigure :prime @12:01:18.20 SamplePump.seek 160ml,13:51 @12:04:09.99 Pressure.coast @12:04:10.15 Bypass.open @12:04:10.94 Intake.close @12:04:11.98 Toroid.reconfigure :hold @12:04:12.07 Toroid.seek :crack @12:04:13.71 Toroid.seek :cartridge @12:04:14.64 Toroid.reconfigure :default @12:04:14.70 Toroid.seek :clear @12:04:16.55 Pressurizing cartridge to 26psi above 14.4psia @12:04:16.56 SamplePump.setPosition! 0ml @12:04:16.59 Intake.open @12:04:17.42 Bypass.close @12:04:18.41 Exhaust.close @12:04:19.51 Pressure.reconfigure :default @12:04:19.67 Pressure.hold @12:04:20.21 SamplePump.reconfigure :charge @12:04:20.31 SamplePump.seek 2.3ml,42 seconds @12:04:33.95 Pressure.coast @12:04:34.00 Intake.close @12:04:34.90 Pressurized cartridge with 0.843ml to 26.0psi @12:04:34.91 Isolated @35.7psia (21.3psi) @12:04:34.92 Checking for cartridge leaks from 35.7psia @12:04:34.94 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:04:41.53 Cartridge pressure fell 11% to 31.9psia @12:04:41.62 Sampling 2000.0ml @12:04:41.64 SamplePump.setPosition! 0ml @12:04:41.68 Exhaust.open @12:04:42.38 Intake.open @12:04:43.36 Pressure.reconfigure :default @12:04:43.41 Pressure.hold @12:04:43.70 SamplePump.reconfigure :sample @12:04:43.77 SamplePump.seek 2000ml,2:47:12 @12:04:46.82 Cartridge 49 sampling 1/2000ml at 0.51ml/s with 24.2psi vacuum @12:05:46.42 Cartridge 49 sampling 32/2000ml at 0.51ml/s with 23.1psi vacuum @12:06:46.63 Cartridge 49 sampling 64/2000ml at 0.52ml/s with 18.9psi vacuum @12:07:46.42 Cartridge 49 sampling 97/2000ml at 0.55ml/s with 18.6psi vacuum @12:08:46.59 Cartridge 49 sampling 131/2000ml at 0.55ml/s with 21.0psi vacuum @12:09:46.42 Cartridge 49 sampling 165/2000ml at 0.57ml/s with 25.3psi vacuum @12:10:23.80 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:10:23.82 Retry #2 of 10 after 10:00 @12:10:46.62 Cartridge 49 sampling 199/2000ml at 0.58ml/s with 24.8psi vacuum @12:11:46.42 Cartridge 49 sampling 233/2000ml at 0.57ml/s with 25.9psi vacuum @12:12:46.50 Cartridge 49 sampling 267/2000ml at 0.54ml/s with 26.5psi vacuum @12:13:46.42 Cartridge 49 sampling 301/2000ml at 0.56ml/s with 27.3psi vacuum @12:14:46.67 Cartridge 49 sampling 335/2000ml at 0.57ml/s with 25.0psi vacuum @12:15:46.42 Cartridge 49 sampling 368/2000ml at 0.56ml/s with 25.2psi vacuum @12:16:46.45 Cartridge 49 sampling 402/2000ml at 0.57ml/s with 24.5psi vacuum @12:17:46.41 Cartridge 49 sampling 436/2000ml at 0.56ml/s with 26.8psi vacuum @12:18:46.67 Cartridge 49 sampling 469/2000ml at 0.56ml/s with 21.7psi vacuum @12:19:46.62 Cartridge 49 sampling 502/2000ml at 0.55ml/s with 26.1psi vacuum @12:20:23.94 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:20:24.08 Retry #3 of 10 after 10:00 @12:20:46.42 Cartridge 49 sampling 535/2000ml at 0.55ml/s with 27.0psi vacuum @12:21:46.68 Cartridge 49 sampling 569/2000ml at 0.55ml/s with 20.8psi vacuum @12:22:46.42 Cartridge 49 sampling 601/2000ml at 0.55ml/s with 26.2psi vacuum @12:23:46.73 Cartridge 49 sampling 634/2000ml at 0.55ml/s with 25.1psi vacuum @12:24:46.41 Cartridge 49 sampling 667/2000ml at 0.55ml/s with 20.3psi vacuum @12:25:46.55 Cartridge 49 sampling 699/2000ml at 0.54ml/s with 26.6psi vacuum @12:26:46.42 Cartridge 49 sampling 732/2000ml at 0.55ml/s with 25.3psi vacuum @12:27:46.56 Cartridge 49 sampling 764/2000ml at 0.53ml/s with 19.7psi vacuum @12:28:46.41 Cartridge 49 sampling 797/2000ml at 0.54ml/s with 20.1psi vacuum @12:29:46.64 Cartridge 49 sampling 829/2000ml at 0.55ml/s with 26.1psi vacuum @12:30:24.20 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:30:24.26 Retry #4 of 10 after 10:00 @12:30:46.42 Cartridge 49 sampling 862/2000ml at 0.55ml/s with 25.3psi vacuum @12:31:46.54 Cartridge 49 sampling 894/2000ml at 0.54ml/s with 25.8psi vacuum @12:32:46.42 Cartridge 49 sampling 927/2000ml at 0.52ml/s with 25.7psi vacuum @12:33:46.59 Cartridge 49 sampling 959/2000ml at 0.51ml/s with 24.7psi vacuum @12:34:46.41 Cartridge 49 sampling 991/2000ml at 0.54ml/s with 22.5psi vacuum @12:35:46.56 Cartridge 49 sampling 1023/2000ml at 0.54ml/s with 25.2psi vacuum @12:36:46.42 Cartridge 49 sampling 1055/2000ml at 0.53ml/s with 21.9psi vacuum @12:37:46.67 Cartridge 49 sampling 1086/2000ml at 0.53ml/s with 24.8psi vacuum @12:38:46.59 Cartridge 49 sampling 1118/2000ml at 0.50ml/s with 26.8psi vacuum @12:39:46.42 Cartridge 49 sampling 1149/2000ml at 0.52ml/s with 26.9psi vacuum @12:40:24.31 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:40:24.33 Retry #5 of 10 after 10:00 @12:40:46.73 Cartridge 49 sampling 1180/2000ml at 0.49ml/s with 21.6psi vacuum @12:41:46.41 Cartridge 49 sampling 1211/2000ml at 0.51ml/s with 18.6psi vacuum @12:42:46.60 Cartridge 49 sampling 1242/2000ml at 0.52ml/s with 25.3psi vacuum @12:43:46.42 Cartridge 49 sampling 1272/2000ml at 0.51ml/s with 25.0psi vacuum @12:44:46.58 Cartridge 49 sampling 1303/2000ml at 0.50ml/s with 28.0psi vacuum @12:45:46.41 Cartridge 49 sampling 1333/2000ml at 0.49ml/s with 27.9psi vacuum @12:46:47.88 Cartridge 49 sampling 1363/2000ml at 0.50ml/s with 19.5psi vacuum @12:47:46.42 Cartridge 49 sampling 1393/2000ml at 0.48ml/s with 18.1psi vacuum @12:48:46.58 Cartridge 49 sampling 1423/2000ml at 0.50ml/s with 22.9psi vacuum @12:49:46.42 Cartridge 49 sampling 1452/2000ml at 0.50ml/s with 21.0psi vacuum @12:50:24.38 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:50:24.40 Retry #6 of 10 after 10:00 @12:50:46.60 Cartridge 49 sampling 1482/2000ml at 0.46ml/s with 29.4psi vacuum @12:51:46.42 Cartridge 49 sampling 1511/2000ml at 0.49ml/s with 19.5psi vacuum @12:52:46.51 Cartridge 49 sampling 1541/2000ml at 0.49ml/s with 26.4psi vacuum @12:53:46.42 Cartridge 49 sampling 1570/2000ml at 0.50ml/s with 23.5psi vacuum @12:54:46.52 Cartridge 49 sampling 1599/2000ml at 0.51ml/s with 21.5psi vacuum 2.1m deep (17.8psia) @12:55:46.42 Cartridge 49 sampling 1629/2000ml at 0.49ml/s with 23.8psi vacuum @12:56:46.60 Cartridge 49 sampling 1658/2000ml at 0.49ml/s with 21.8psi vacuum 2.5m deep (18.4psia) @12:57:46.62 Cartridge 49 sampling 1686/2000ml at 0.48ml/s with 20.0psi vacuum @12:58:46.42 Cartridge 49 sampling 1715/2000ml at 0.45ml/s with 24.3psi vacuum @12:59:46.90 Cartridge 49 sampling 1743/2000ml at 0.48ml/s with 23.6psi vacuum @13:00:24.58 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:00:24.60 Retry #7 of 10 after 10:00 @13:00:46.42 Cartridge 49 sampling 1771/2000ml at 0.47ml/s with 21.5psi vacuum @13:01:46.60 Cartridge 49 sampling 1799/2000ml at 0.46ml/s with 28.2psi vacuum @13:02:46.42 Cartridge 49 sampling 1827/2000ml at 0.46ml/s with 28.5psi vacuum @13:03:46.62 Cartridge 49 sampling 1854/2000ml at 0.45ml/s with 24.7psi vacuum @13:04:46.41 Cartridge 49 sampling 1882/2000ml at 0.46ml/s with 21.2psi vacuum @13:05:46.55 Cartridge 49 sampling 1909/2000ml at 0.44ml/s with 17.2psi vacuum 3.4m deep (19.8psia) @13:06:46.42 Cartridge 49 sampling 1936/2000ml at 0.44ml/s with 24.9psi vacuum @13:07:46.74 Cartridge 49 sampling 1963/2000ml at 0.43ml/s with 17.0psi vacuum @13:08:46.42 Cartridge 49 sampling 1990/2000ml at 0.45ml/s with 27.3psi vacuum @13:09:08.45 Pressure.coast @13:09:08.75 Intake.close @13:09:08.83 Exhaust.close @13:09:09.33 Toroid.reconfigure :hold @13:09:09.53 Toroid.seek :bypass @13:09:11.19 Toroid.reconfigure :default @13:09:11.31 Toroid.seek :clear @13:09:12.91 Sampled 2000.0ml @13:09:13.71 SamplePump.setPosition! 0ml @13:09:13.74 Plunge.seek :reagent @13:09:13.83 Bypass.open @13:09:14.54 Intake.open @13:09:14.68 Sample loop closed. Ready to Deliver Reagents @13:09:14.70 Archive.seek :sealed @13:09:15.41 SamplePump.reconfigure :slow @13:09:15.47 SamplePump.seek -3ml @13:09:18.35 Archive.seek :evac @13:09:22.15 delay 1 second @13:09:23.24 SamplePump.seek -2.8ml @13:09:23.95 delay 1 second @13:09:25.25 SamplePump.seek -3.3ml @13:09:26.61 delay 1 second @13:09:27.65 Exhaust.open @13:09:28.47 Bypass.close @13:09:29.33 SamplePump.seek -35.3ml @13:09:32.40 delay 1:00 @13:10:24.73 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:10:24.83 Retry #8 of 10 after 10:00 @13:10:32.44 Archive.seek :clear @13:10:38.18 delay 3 seconds @13:10:41.43 Intake.close @13:10:41.49 Exhaust.close @13:10:49.63 RNAlater soak for 10:00 @13:20:24.91 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:20:24.93 Retry #9 of 10 after 10:00 @13:20:49.68 Archive.seek :tVent @13:20:51.33 delay 1 second @13:20:52.36 Archive.seek :evac @13:21:08.34 delay 10 seconds @13:21:18.38 Sample preserved and evacuated @13:21:18.61
Carousel.seek 49 @13:21:18.81 Plunge.seek :locked @13:21:19.79 Archive.seek :clear @13:21:19.96 Air.seek :clear @13:21:20.12 Lysis1.seek :clear @13:21:36.96
Archive.coast @13:21:37.00 Air.coast @13:21:37.04 Lysis1.coast @13:21:37.09 Lysis2.coast @13:21:37.13 Plunge.coast @13:21:37.17 SamplePump.coast @13:21:37.22 Toroid.coast @13:21:37.26 Carousel.coast @13:21:37.45 Gate.power :motor,:OFF @13:21:37.46 Gate.power :core,:OFF @13:21:37.48 Gate.power :analytic1,:OFF @13:21:37.49 Gate.power :analytic2,:OFF @13:21:37.50 Gate.power :analytic3,:OFF @13:21:37.51 Gate.power :main,:OFF @13:21:37.53 Safely stopped and ready to power off @13:21:37.74 delayUntil "00:00:00.00PDT13-Jun-21" @13:30:25.04 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:30:25.11 Retry #10 of 10 after 10:00 @13:40:25.19 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:40:25.21 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 49" @13:40:25.24 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:40:25.26 Retry #1 of 10 after 10:00 @13:50:25.31 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:50:25.33 Retry #2 of 10 after 10:00 @14:00:25.40 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:00:25.45 Retry #3 of 10 after 10:00 @14:10:25.52 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:10:25.54 Retry #4 of 10 after 10:00 @14:20:25.60 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:20:25.64 Retry #5 of 10 after 10:00 @14:30:25.70 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:30:25.74 Retry #6 of 10 after 10:00 @14:40:25.79 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:40:25.81 Retry #7 of 10 after 10:00 @14:50:25.86 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:50:25.88 Retry #8 of 10 after 10:00 @15:00:25.93 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:00:25.95 Retry #9 of 10 after 10:00 @15:10:26.02 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:10:26.07 Retry #10 of 10 after 10:00 @15:20:26.20 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:20:26.23 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT13-Jun-21" @15:20:26.35 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:20:26.37 Retry #1 of 10 after 10:00 @15:30:26.49 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:30:26.54 Retry #2 of 10 after 10:00 @15:40:26.63 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:40:26.65 Retry #3 of 10 after 10:00 @15:50:26.70 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:50:26.73 Retry #4 of 10 after 10:00 @16:00:26.77 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:00:26.79 Retry #5 of 10 after 10:00 @16:10:26.84 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:10:26.86 Retry #6 of 10 after 10:00 @16:20:26.91 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:20:27.06 Retry #7 of 10 after 10:00 @16:30:27.14 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:30:27.16 Retry #8 of 10 after 10:00 @16:40:27.21 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:40:27.23 Retry #9 of 10 after 10:00 @16:50:27.39 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:50:27.41 Retry #10 of 10 after 10:00 @17:00:27.47 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:00:00.20PDT13-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.22 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.24 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.29 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.40 Powering core bus @00:00:02.72 Archive.configure :default @00:00:02.86 Air.configure :default @00:00:02.97 Lysis1.configure :default @00:00:03.14 Lysis2.configure :default @00:00:03.38 Plunge.configure :default @00:00:03.52 SamplePump.configure :default @00:00:03.66 ExtraPump.configure :default @00:00:03.80 Toroid.configure :default @00:00:03.95 Carousel.configure :default @00:00:04.08 Pressure.configure :default @00:00:04.14 All elf controllers running firmware version 4.61 @00:00:04.16 Gate.power :motor,:ON,limit to 5000mA @00:00:04.37 Archive.reconfigure :default @00:00:04.70 Archive.home.jog 0 @00:00:04.91 Air.reconfigure :default @00:00:04.94 Plunge.reconfigure :default @00:00:05.02 Lysis2.reconfigure :default @00:00:05.04 Lysis1.reconfigure :default @00:00:05.09 Toroid.reconfigure :default @00:00:05.69 Air.home.jog 0 @00:00:06.05 Lysis2.home.jog 0 @00:00:06.10 Plunge.home.jog 0 @00:00:06.13 Lysis1.home.jog 0 @00:00:06.27 Toroid.home.jog 0 @00:00:07.57 Archive.reconfigure :homing @00:00:07.83 Archive.jog 1367 @00:00:08.30 Plunge.reconfigure :homing @00:00:08.37 Lysis1.reconfigure :homing @00:00:08.43 Air.reconfigure :homing @00:00:08.61 Toroid.reconfigure :homing @00:00:08.64 Lysis2.reconfigure :homing @00:00:09.07 Plunge.jog 847 @00:00:09.23 Lysis1.jog 1351 @00:00:09.26 Toroid.jog 623 @00:00:09.27 Lysis2.jog 1352 @00:00:09.42 Air.jog 1359 @00:00:09.90 Archive.setRawPosition! 50 @00:00:10.04 Archive.seek :home @00:00:10.41 Plunge.setRawPosition! 50 @00:00:10.43 Archive.reconfigure :default @00:00:10.58 Plunge.seek :home @00:00:11.11 Lysis1.setRawPosition! 50 @00:00:11.22 Air.setRawPosition! 50 @00:00:11.26 Lysis2.setRawPosition! 50 @00:00:11.39 Lysis1.seek :home @00:00:11.45 Air.seek :home @00:00:11.48 Lysis2.seek :home @00:00:11.67 Plunge.reconfigure :default @00:00:12.34 Toroid.setRawPosition! 0 @00:00:12.38 Lysis2.reconfigure :default @00:00:12.40 Air.reconfigure :default @00:00:12.44 Lysis1.reconfigure :default @00:00:12.63 Toroid.seek :home @00:00:12.74 Toroid.reconfigure :default @00:00:12.98
Toroid.seek :clear @00:00:14.59 Carousel.reconfigure :default @00:00:14.65 Carousel.home.jog 0 @00:00:15.15 Carousel.seek 49 @00:00:15.32 Selecting Cartridge 48 @00:00:15.72 Carousel.seek 48 @00:00:16.82 Lysis1.seek :locked @00:00:20.25 Plunge.seek :locked @00:00:20.79 Air.seek :locked @00:00:25.02
Mounted archive Cartridge @00:00:25.38 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 48" @00:00:25.85 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:00:26.10 Retry #1 of 10 after 10:00 @00:00:26.76 Duration of filtering limited to 4:05:00 @00:00:27.52 Priming sample loop w/160ml, bypass w/40ml @00:00:27.54 SamplePump.setPosition! 0ml @00:00:27.59 Bypass.open @00:00:29.20 Intake.open @00:00:30.35 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:32.95 Vacuum tare is 18% of full scale @00:00:32.97 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:33.18 Pressure.reconfigure :default @00:00:33.24 Pressure.hold @00:00:33.44 SamplePump.reconfigure :prime @00:00:33.51 SamplePump.seek 40ml,3:51 @00:01:16.71 Pressure.coast @00:01:16.76 SamplePump.setPosition! 0ml @00:01:16.81 Exhaust.open @00:01:18.23 Bypass.close @00:01:19.78 Pressure.reconfigure :default @00:01:19.84 Pressure.hold @00:01:20.11 SamplePump.reconfigure :prime @00:01:20.17 SamplePump.seek 160ml,13:51 @00:04:11.76 Pressure.coast @00:04:11.82 Bypass.open @00:04:12.54 Intake.close @00:04:13.43 Toroid.reconfigure :hold @00:04:13.49 Toroid.seek :crack @00:04:15.35 Toroid.seek :cartridge @00:04:16.36 Toroid.reconfigure :default @00:04:16.50 Toroid.seek :clear @00:04:18.84 Pressurizing cartridge to 26psi above 14.6psia @00:04:18.88 SamplePump.setPosition! 0ml @00:04:18.92 Intake.open @00:04:19.75 Bypass.close @00:04:20.57 Exhaust.close @00:04:21.39 Pressure.reconfigure :default @00:04:21.45 Pressure.hold @00:04:21.65 SamplePump.reconfigure :charge @00:04:21.72 SamplePump.seek 2.3ml,42 seconds @00:04:35.48 Pressure.coast @00:04:35.69 Intake.close @00:04:36.65 Pressurized cartridge with 0.862ml to 26.1psi @00:04:36.72 Isolated @34.8psia (20.2psi) @00:04:36.74 Checking for cartridge leaks from 34.8psia @00:04:36.85 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:42.96 Cartridge pressure fell 11% to 31.1psia @00:04:43.04 Sampling 2000.0ml @00:04:43.06 SamplePump.setPosition! 0ml @00:04:43.10 Exhaust.open @00:04:43.82 Intake.open @00:04:45.06 Pressure.reconfigure :default @00:04:45.20 Pressure.hold @00:04:45.89 SamplePump.reconfigure :sample @00:04:46.04 SamplePump.seek 2000ml,2:47:12 @00:04:48.31 Cartridge 48 sampling 1/2000ml at 0.62ml/s with 23.2psi vacuum @00:05:48.05 Cartridge 48 sampling 32/2000ml at 0.53ml/s with 24.3psi vacuum @00:06:48.01 Cartridge 48 sampling 64/2000ml at 0.53ml/s with 21.7psi vacuum @00:07:48.25 Cartridge 48 sampling 97/2000ml at 0.55ml/s with 17.7psi vacuum 3.5m deep (19.9psia) @00:08:47.96 Cartridge 48 sampling 130/2000ml at 0.55ml/s with 16.7psi vacuum @00:09:48.17 Cartridge 48 sampling 165/2000ml at 0.58ml/s with 22.5psi vacuum @00:10:28.08 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:10:28.19 Retry #2 of 10 after 10:00 @00:10:47.97 Cartridge 48 sampling 199/2000ml at 0.57ml/s with 20.8psi vacuum @00:11:48.44 Cartridge 48 sampling 233/2000ml at 0.58ml/s with 21.7psi vacuum @00:12:47.97 Cartridge 48 sampling 266/2000ml at 0.55ml/s with 16.0psi vacuum 2.6m deep (18.5psia) @00:13:48.14 Cartridge 48 sampling 300/2000ml at 0.57ml/s with 22.1psi vacuum @00:14:47.97 Cartridge 48 sampling 334/2000ml at 0.56ml/s with 17.9psi vacuum @00:15:48.10 Cartridge 48 sampling 367/2000ml at 0.56ml/s with 20.0psi vacuum 2.3m deep (18.1psia) @00:16:47.97 Cartridge 48 sampling 400/2000ml at 0.55ml/s with 17.8psi vacuum 1.9m deep (17.6psia) @00:17:48.14 Cartridge 48 sampling 434/2000ml at 0.55ml/s with 27.7psi vacuum @00:18:47.97 Cartridge 48 sampling 467/2000ml at 0.56ml/s with 23.8psi vacuum @00:19:48.15 Cartridge 48 sampling 500/2000ml at 0.55ml/s with 22.6psi vacuum @00:20:41.80 upsync @00:20:47.97 Cartridge 48 sampling 533/2000ml at 0.52ml/s with 13.9psi vacuum 2.6m deep (18.6psia) @00:21:48.17 Cartridge 48 sampling 565/2000ml at 0.53ml/s with 29.9psi vacuum @00:22:48.11 Cartridge 48 sampling 598/2000ml at 0.55ml/s with 20.6psi vacuum @00:23:48.05 Cartridge 48 sampling 631/2000ml at 0.54ml/s with 28.4psi vacuum @00:24:47.97 Cartridge 48 sampling 663/2000ml at 0.51ml/s with 20.7psi vacuum @00:25:48.38 Cartridge 48 sampling 695/2000ml at 0.55ml/s with 26.2psi vacuum @00:26:48.28 Cartridge 48 sampling 728/2000ml at 0.54ml/s with 16.7psi vacuum @00:27:47.97 Cartridge 48 sampling 759/2000ml at 0.52ml/s with 29.0psi vacuum @00:28:48.34 Cartridge 48 sampling 791/2000ml at 0.54ml/s with 25.3psi vacuum @00:29:47.97 Cartridge 48 sampling 823/2000ml at 0.54ml/s with 16.5psi vacuum 4.2m deep (20.9psia) @00:30:48.18 Cartridge 48 sampling 855/2000ml at 0.52ml/s with 27.8psi vacuum @00:31:47.97 Cartridge 48 sampling 886/2000ml at 0.51ml/s with 16.1psi vacuum @00:32:48.10 Cartridge 48 sampling 918/2000ml at 0.52ml/s with 28.0psi vacuum @00:33:47.97 Cartridge 48 sampling 949/2000ml at 0.51ml/s with 25.7psi vacuum @00:34:48.20 Cartridge 48 sampling 980/2000ml at 0.50ml/s with 23.2psi vacuum @00:35:47.97 Cartridge 48 sampling 1011/2000ml at 0.53ml/s with 27.2psi vacuum @00:36:48.16 Cartridge 48 sampling 1042/2000ml at 0.53ml/s with 26.3psi vacuum @00:37:47.97 Cartridge 48 sampling 1072/2000ml at 0.52ml/s with 27.7psi vacuum @00:38:48.17 Cartridge 48 sampling 1103/2000ml at 0.52ml/s with 22.8psi vacuum @00:39:47.97 Cartridge 48 sampling 1133/2000ml at 0.48ml/s with 29.8psi vacuum @00:40:48.12 Cartridge 48 sampling 1164/2000ml at 0.50ml/s with 17.8psi vacuum @00:41:47.97 Cartridge 48 sampling 1194/2000ml at 0.50ml/s with 29.9psi vacuum @00:42:48.14 Cartridge 48 sampling 1224/2000ml at 0.51ml/s with 25.2psi vacuum @00:43:48.23 Cartridge 48 sampling 1254/2000ml at 0.47ml/s with 30.4psi vacuum @00:44:47.96 Cartridge 48 sampling 1284/2000ml at 0.50ml/s with 18.4psi vacuum 2.1m deep (17.8psia) @00:45:48.28 Cartridge 48 sampling 1314/2000ml at 0.49ml/s with 27.5psi vacuum @00:46:47.96 Cartridge 48 sampling 1343/2000ml at 0.51ml/s with 25.3psi vacuum @00:47:48.13 Cartridge 48 sampling 1373/2000ml at 0.50ml/s with 28.1psi vacuum @00:48:47.96 Cartridge 48 sampling 1402/2000ml at 0.50ml/s with 16.6psi vacuum 4.4m deep (21.1psia) @00:49:48.09 Cartridge 48 sampling 1431/2000ml at 0.51ml/s with 25.4psi vacuum @00:50:47.96 Cartridge 48 sampling 1460/2000ml at 0.50ml/s with 19.5psi vacuum 2.5m deep (18.5psia) @00:51:48.16 Cartridge 48 sampling 1489/2000ml at 0.46ml/s with 30.9psi vacuum @00:52:47.96 Cartridge 48 sampling 1518/2000ml at 0.48ml/s with 18.2psi vacuum @00:53:48.17 Cartridge 48 sampling 1546/2000ml at 0.47ml/s with 28.7psi vacuum @00:54:47.96 Cartridge 48 sampling 1575/2000ml at 0.47ml/s with 29.7psi vacuum @00:55:48.13 Cartridge 48 sampling 1603/2000ml at 0.46ml/s with 29.7psi vacuum @00:56:47.96 Cartridge 48 sampling 1632/2000ml at 0.46ml/s with 29.8psi vacuum @00:57:48.04 Cartridge 48 sampling 1660/2000ml at 0.49ml/s with 26.9psi vacuum @00:58:47.96 Cartridge 48 sampling 1688/2000ml at 0.47ml/s with 21.0psi vacuum @00:59:48.04 Cartridge 48 sampling 1716/2000ml at 0.48ml/s with 17.0psi vacuum 4.7m deep (21.7psia) @01:00:48.40 Cartridge 48 sampling 1743/2000ml at 0.43ml/s with 20.4psi vacuum @01:01:47.96 Cartridge 48 sampling 1771/2000ml at 0.43ml/s with 30.0psi vacuum @01:02:48.22 Cartridge 48 sampling 1798/2000ml at 0.45ml/s with 28.7psi vacuum @01:03:47.97 Cartridge 48 sampling 1826/2000ml at 0.47ml/s with 14.5psi vacuum 5.9m deep (23.4psia) @01:04:48.12 Cartridge 48 sampling 1853/2000ml at 0.45ml/s with 30.9psi vacuum @01:05:47.97 Cartridge 48 sampling 1880/2000ml at 0.46ml/s with 24.2psi vacuum @01:06:48.14 Cartridge 48 sampling 1907/2000ml at 0.45ml/s with 24.7psi vacuum @01:07:47.97 Cartridge 48 sampling 1934/2000ml at 0.45ml/s with 16.5psi vacuum 2.4m deep (18.3psia) @01:08:48.21 Cartridge 48 sampling 1961/2000ml at 0.44ml/s with 18.1psi vacuum 4.6m deep (21.5psia) @01:09:47.97 Cartridge 48 sampling 1988/2000ml at 0.45ml/s with 18.8psi vacuum 2.2m deep (17.9psia) @01:10:15.07 Pressure.coast @01:10:15.21 Intake.close @01:10:15.26 Exhaust.close @01:10:15.44 Toroid.reconfigure :hold @01:10:15.51 Toroid.seek :bypass @01:10:17.12 Toroid.reconfigure :default @01:10:17.30 Toroid.seek :clear @01:10:19.20 Sampled 2000.0ml @01:10:20.41 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 48" @01:10:22.93 SamplePump.setPosition! 0ml @01:10:22.97 Plunge.seek :reagent @01:10:23.04 Bypass.open @01:10:23.77 Intake.open @01:10:23.91 Sample loop closed. Ready to Deliver Reagents @01:10:23.93 Archive.seek :sealed @01:10:24.64 SamplePump.reconfigure :slow @01:10:24.71 SamplePump.seek -3ml @01:10:27.72 Archive.seek :evac @01:10:31.39 delay 1 second @01:10:32.49 SamplePump.seek -2.8ml @01:10:33.18 delay 1 second @01:10:34.27 SamplePump.seek -3.3ml @01:10:34.98 upsync @01:10:35.60 delay 1 second @01:10:36.63 Exhaust.open @01:10:37.38 Bypass.close @01:10:38.55 SamplePump.seek -35.3ml @01:10:41.99 delay 1:00 @01:11:42.05 Archive.seek :clear @01:11:47.34 delay 3 seconds @01:11:50.54 Intake.close @01:11:50.76 Exhaust.close @01:11:59.22 RNAlater soak for 10:00 @01:21:59.28 Archive.seek :tVent @01:22:00.99 delay 1 second @01:22:02.04 Archive.seek :evac @01:22:18.03 delay 10 seconds @01:22:28.14 Sample preserved and evacuated @01:22:28.91
Carousel.seek 48 @01:22:29.12 Plunge.seek :locked @01:22:30.10 Archive.seek :clear @01:22:30.26 Air.seek :clear @01:22:30.43 Lysis1.seek :clear @01:22:47.29
Archive.coast @01:22:47.33 Air.coast @01:22:47.37 Lysis1.coast @01:22:47.41 Lysis2.coast @01:22:47.46 Plunge.coast @01:22:47.50 SamplePump.coast @01:22:47.55 Toroid.coast @01:22:47.59 Carousel.coast @01:22:47.78 Gate.power :motor,:OFF @01:22:47.79 Gate.power :core,:OFF @01:22:47.80 Gate.power :analytic1,:OFF @01:22:47.82 Gate.power :analytic2,:OFF @01:22:47.83 Gate.power :analytic3,:OFF @01:22:47.85 Gate.power :main,:OFF @01:22:47.86 Safely stopped and ready to power off @01:22:48.10 delayUntil "12:00:00.00" @01:22:48.13 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @01:23:02.04 upsync @16:44:21.36PDT13-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @16:44:21.41 >BREAK< ESP3 Gateway v1.11 for cartridge bus @16:44:26.26 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @16:44:26.58 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @16:44:35.93 Polling Can every 2:00 @16:44:35.99 Loading /home/esp/esp2/bin/processAll mission @16:44:36.96 Session Begun @16:44:37.91
Sampling default goal volumes @16:44:37.98 GoodNews.email "Cartridge group: dry\nSpares: {spare: 1}\nSchedule: [\"12am\", \"12pm\"]\n",Subject: "Begun processAll --schedule=12am,12pm --checkpoint --mail" @16:44:38.10
delayUntil "00:00:00.00PDT14-Jun-21" @16:44:48.33 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT14-Jun-21" @16:44:58.20 upsync @16:45:27.83 upsync @16:45:57.87 -> spare @16:46:09.11 -> slot @16:47:39.95 -> exit @16:47:39.97 Session Ended @00:00:00.32PDT14-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.33 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.35 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.36 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.51 Powering core bus @00:00:02.95 Archive.configure :default @00:00:03.08 Air.configure :default @00:00:03.20 Lysis1.configure :default @00:00:03.31 Lysis2.configure :default @00:00:03.44 Plunge.configure :default @00:00:03.58 SamplePump.configure :default @00:00:03.72 ExtraPump.configure :default @00:00:03.85 Toroid.configure :default @00:00:03.99 Carousel.configure :default @00:00:04.13 Pressure.configure :default @00:00:04.18 All elf controllers running firmware version 4.61 @00:00:04.21 Gate.power :motor,:ON,limit to 5000mA @00:00:04.45 Archive.reconfigure :default @00:00:04.74 Archive.home.jog 0 @00:00:04.99 Toroid.reconfigure :default @00:00:05.05 Plunge.reconfigure :default @00:00:05.09 Lysis2.reconfigure :default @00:00:05.12 Air.reconfigure :default @00:00:05.17 Lysis1.reconfigure :default @00:00:05.39 Toroid.home.jog 0 @00:00:05.44 Plunge.home.jog 0 @00:00:05.46 Lysis2.home.jog 0 @00:00:05.49 Air.home.jog 0 @00:00:05.54 Lysis1.home.jog 0 @00:00:06.00 Archive.reconfigure :homing @00:00:06.20 Archive.jog 1372 @00:00:06.30 Toroid.reconfigure :homing @00:00:06.33 Lysis2.reconfigure :homing @00:00:06.36 Lysis1.reconfigure :homing @00:00:06.41 Plunge.reconfigure :homing @00:00:06.46 Air.reconfigure :homing @00:00:06.70 Lysis2.jog 1358 @00:00:06.74 Toroid.jog 627 @00:00:06.77 Lysis1.jog 1352 @00:00:06.80 Plunge.jog 851 @00:00:06.83 Air.jog 1363 @00:00:07.74 Archive.setRawPosition! 50 @00:00:07.86 Archive.seek :home @00:00:08.41 Plunge.setRawPosition! 50 @00:00:08.44 Lysis2.setRawPosition! 50 @00:00:08.46 Lysis1.setRawPosition! 50 @00:00:08.73 Plunge.seek :home @00:00:08.95 Lysis2.seek :home @00:00:08.99 Lysis1.seek :home @00:00:09.62 Plunge.reconfigure :default @00:00:09.80 Lysis2.reconfigure :default @00:00:09.83 Lysis1.reconfigure :default @00:00:10.04 Archive.reconfigure :default @00:00:10.09 Air.setRawPosition! 50 @00:00:10.20 Toroid.setRawPosition! 0 @00:00:10.69 Toroid.seek :home @00:00:10.83 Air.seek :home @00:00:11.43 Air.reconfigure :default @00:00:11.46 Toroid.reconfigure :default @00:00:11.90
Toroid.seek :clear @00:00:13.58 Carousel.reconfigure :default @00:00:13.63 Carousel.home.jog 0 @00:00:14.14 Carousel.seek 48 @00:00:14.31 Selecting Cartridge 47 @00:00:14.52 Carousel.seek 47 @00:00:15.03 Lysis1.seek :locked @00:00:18.43 Plunge.seek :locked @00:00:18.86 Air.seek :locked @00:00:23.14
Mounted archive Cartridge @00:00:23.29 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 47" @00:00:23.58 Duration of filtering limited to 4:05:00 @00:00:23.88 Priming sample loop w/160ml, bypass w/40ml @00:00:23.89 SamplePump.setPosition! 0ml @00:00:23.93 Bypass.open @00:00:25.42 Intake.open @00:00:26.32 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:30.24 Vacuum tare is 18% of full scale @00:00:30.30 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:30.82 Pressure.reconfigure :default @00:00:31.01 Pressure.hold @00:00:31.54 SamplePump.reconfigure :prime @00:00:31.63 SamplePump.seek 40ml,3:51 @00:00:44.24 upsync @00:01:14.90 Pressure.coast @00:01:14.98 SamplePump.setPosition! 0ml @00:01:15.16 Exhaust.open @00:01:16.60 Bypass.close @00:01:18.72 Pressure.reconfigure :default @00:01:18.77 Pressure.hold @00:01:19.05 SamplePump.reconfigure :prime @00:01:19.12 SamplePump.seek 160ml,13:51 @00:04:10.86 Pressure.coast @00:04:10.98 Bypass.open @00:04:11.67 Intake.close @00:04:12.59 Toroid.reconfigure :hold @00:04:12.71 Toroid.seek :crack @00:04:14.39 Toroid.seek :cartridge @00:04:15.33 Toroid.reconfigure :default @00:04:15.38 Toroid.seek :clear @00:04:17.23 Pressurizing cartridge to 26psi above 14.5psia @00:04:17.25 SamplePump.setPosition! 0ml @00:04:17.28 Intake.open @00:04:18.10 Bypass.close @00:04:18.85 Exhaust.close @00:04:20.02 Pressure.reconfigure :default @00:04:20.16 Pressure.hold @00:04:20.61 SamplePump.reconfigure :charge @00:04:20.79 SamplePump.seek 2.3ml,42 seconds @00:04:32.18 Pressure.coast @00:04:32.23 Intake.close @00:04:33.08 Pressurized cartridge with 0.592ml to 26.1psi @00:04:33.09 Isolated @36.3psia (21.7psi) @00:04:33.10 Checking for cartridge leaks from 36.3psia @00:04:33.12 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:42.72 Cartridge pressure fell 15% to 30.9psia @00:04:42.80 Sampling 2000.0ml @00:04:42.82 SamplePump.setPosition! 0ml @00:04:42.86 Exhaust.open @00:04:43.54 Intake.open @00:04:44.49 Pressure.reconfigure :default @00:04:44.55 Pressure.hold @00:04:44.83 SamplePump.reconfigure :sample @00:04:44.89 SamplePump.seek 2000ml,2:47:12 @00:04:47.72 Cartridge 47 sampling 1/2000ml at 0.46ml/s with 23.8psi vacuum @00:05:47.56 Cartridge 47 sampling 30/2000ml at 0.48ml/s with 25.0psi vacuum @00:06:47.74 Cartridge 47 sampling 58/2000ml at 0.48ml/s with 23.3psi vacuum @00:07:47.56 Cartridge 47 sampling 88/2000ml at 0.48ml/s with 21.5psi vacuum @00:08:47.89 Cartridge 47 sampling 118/2000ml at 0.49ml/s with 25.8psi vacuum @00:09:47.56 Cartridge 47 sampling 149/2000ml at 0.52ml/s with 25.3psi vacuum @00:10:47.64 Cartridge 47 sampling 180/2000ml at 0.53ml/s with 26.9psi vacuum @00:11:47.56 Cartridge 47 sampling 210/2000ml at 0.51ml/s with 26.2psi vacuum @00:12:47.59 Cartridge 47 sampling 241/2000ml at 0.52ml/s with 26.5psi vacuum @00:13:47.56 Cartridge 47 sampling 272/2000ml at 0.52ml/s with 26.0psi vacuum @00:14:47.56 Cartridge 47 sampling 302/2000ml at 0.52ml/s with 26.0psi vacuum @00:15:47.91 Cartridge 47 sampling 333/2000ml at 0.51ml/s with 26.9psi vacuum @00:16:47.56 Cartridge 47 sampling 364/2000ml at 0.50ml/s with 28.0psi vacuum @00:17:47.72 Cartridge 47 sampling 394/2000ml at 0.51ml/s with 20.1psi vacuum @00:18:47.56 Cartridge 47 sampling 425/2000ml at 0.50ml/s with 21.5psi vacuum @00:19:47.75 Cartridge 47 sampling 455/2000ml at 0.50ml/s with 21.7psi vacuum @00:20:47.56 Cartridge 47 sampling 485/2000ml at 0.51ml/s with 20.1psi vacuum 1.7m deep (17.1psia) @00:21:47.72 Cartridge 47 sampling 516/2000ml at 0.50ml/s with 17.7psi vacuum 3.9m deep (20.4psia) @00:22:47.56 Cartridge 47 sampling 546/2000ml at 0.50ml/s with 21.4psi vacuum @00:23:47.70 Cartridge 47 sampling 576/2000ml at 0.50ml/s with 18.0psi vacuum 5.0m deep (22.1psia) @00:24:47.56 Cartridge 47 sampling 605/2000ml at 0.49ml/s with 28.7psi vacuum @00:25:47.58 Cartridge 47 sampling 635/2000ml at 0.51ml/s with 21.7psi vacuum @00:26:47.56 Cartridge 47 sampling 665/2000ml at 0.48ml/s with 29.5psi vacuum @00:27:47.64 Cartridge 47 sampling 694/2000ml at 0.49ml/s with 27.3psi vacuum @00:28:47.73 Cartridge 47 sampling 724/2000ml at 0.50ml/s with 22.8psi vacuum 1.8m deep (17.3psia) @00:29:47.56 Cartridge 47 sampling 753/2000ml at 0.47ml/s with 29.8psi vacuum @00:30:47.82 Cartridge 47 sampling 782/2000ml at 0.49ml/s with 21.2psi vacuum @00:31:47.56 Cartridge 47 sampling 811/2000ml at 0.48ml/s with 15.7psi vacuum 4.0m deep (20.5psia) @00:32:47.75 Cartridge 47 sampling 840/2000ml at 0.46ml/s with 17.8psi vacuum @00:33:47.56 Cartridge 47 sampling 869/2000ml at 0.46ml/s with 28.5psi vacuum @00:34:47.65 Cartridge 47 sampling 898/2000ml at 0.49ml/s with 17.3psi vacuum 5.0m deep (22.1psia) @00:35:47.56 Cartridge 47 sampling 926/2000ml at 0.48ml/s with 18.6psi vacuum @00:36:47.71 Cartridge 47 sampling 954/2000ml at 0.47ml/s with 26.0psi vacuum @00:37:47.56 Cartridge 47 sampling 982/2000ml at 0.47ml/s with 16.2psi vacuum 3.4m deep (19.7psia) @00:38:47.72 Cartridge 47 sampling 1010/2000ml at 0.46ml/s with 26.2psi vacuum @00:39:47.56 Cartridge 47 sampling 1038/2000ml at 0.47ml/s with 26.5psi vacuum @00:40:47.59 Cartridge 47 sampling 1066/2000ml at 0.48ml/s with 19.2psi vacuum 2.1m deep (17.8psia) @00:41:47.62 Cartridge 47 sampling 1093/2000ml at 0.48ml/s with 24.2psi vacuum @00:42:47.56 Cartridge 47 sampling 1121/2000ml at 0.46ml/s with 18.3psi vacuum 2.3m deep (18.1psia) @00:43:47.80 Cartridge 47 sampling 1148/2000ml at 0.47ml/s with 25.5psi vacuum @00:44:47.56 Cartridge 47 sampling 1175/2000ml at 0.46ml/s with 21.7psi vacuum @00:45:48.80 Cartridge 47 sampling 1203/2000ml at 0.44ml/s with 15.7psi vacuum 3.0m deep (19.1psia) @00:46:47.56 Cartridge 47 sampling 1229/2000ml at 0.42ml/s with 28.3psi vacuum @00:47:47.83 Cartridge 47 sampling 1256/2000ml at 0.44ml/s with 27.6psi vacuum @00:48:47.56 Cartridge 47 sampling 1283/2000ml at 0.41ml/s with 20.2psi vacuum @00:49:47.72 Cartridge 47 sampling 1309/2000ml at 0.44ml/s with 17.2psi vacuum 4.6m deep (21.5psia) @00:50:47.56 Cartridge 47 sampling 1336/2000ml at 0.46ml/s with 25.5psi vacuum @00:51:47.72 Cartridge 47 sampling 1362/2000ml at 0.42ml/s with 28.9psi vacuum @00:52:47.56 Cartridge 47 sampling 1388/2000ml at 0.45ml/s with 23.1psi vacuum @00:53:47.67 Cartridge 47 sampling 1414/2000ml at 0.44ml/s with 27.7psi vacuum @00:54:47.56 Cartridge 47 sampling 1440/2000ml at 0.41ml/s with 29.7psi vacuum @00:55:47.72 Cartridge 47 sampling 1466/2000ml at 0.43ml/s with 20.1psi vacuum 2.9m deep (18.9psia) @00:56:47.56 Cartridge 47 sampling 1492/2000ml at 0.41ml/s with 30.0psi vacuum @00:57:47.73 Cartridge 47 sampling 1518/2000ml at 0.43ml/s with 27.6psi vacuum @00:58:47.82 Cartridge 47 sampling 1543/2000ml at 0.44ml/s with 22.6psi vacuum @00:59:47.56 Cartridge 47 sampling 1569/2000ml at 0.41ml/s with 28.3psi vacuum @01:00:47.72 Cartridge 47 sampling 1594/2000ml at 0.42ml/s with 19.8psi vacuum @01:01:47.56 Cartridge 47 sampling 1619/2000ml at 0.39ml/s with 22.2psi vacuum @01:02:47.80 Cartridge 47 sampling 1644/2000ml at 0.43ml/s with 25.8psi vacuum @01:03:47.56 Cartridge 47 sampling 1669/2000ml at 0.40ml/s with 19.6psi vacuum @01:04:47.74 Cartridge 47 sampling 1694/2000ml at 0.40ml/s with 29.6psi vacuum @01:05:47.58 Cartridge 47 sampling 1718/2000ml at 0.42ml/s with 20.8psi vacuum 1.9m deep (17.4psia) @01:06:47.74 Cartridge 47 sampling 1742/2000ml at 0.39ml/s with 29.2psi vacuum @01:07:47.55 Cartridge 47 sampling 1767/2000ml at 0.40ml/s with 17.4psi vacuum @01:08:47.62 Cartridge 47 sampling 1791/2000ml at 0.38ml/s with 18.7psi vacuum @01:09:47.55 Cartridge 47 sampling 1815/2000ml at 0.38ml/s with 14.7psi vacuum 1.7m deep (17.3psia) @01:10:47.67 Cartridge 47 sampling 1839/2000ml at 0.39ml/s with 29.5psi vacuum @01:11:47.55 Cartridge 47 sampling 1863/2000ml at 0.41ml/s with 26.4psi vacuum @01:12:47.81 Cartridge 47 sampling 1887/2000ml at 0.37ml/s with 29.7psi vacuum @01:13:47.82 Cartridge 47 sampling 1911/2000ml at 0.41ml/s with 24.4psi vacuum @01:14:47.56 Cartridge 47 sampling 1934/2000ml at 0.40ml/s with 23.5psi vacuum @01:15:47.73 Cartridge 47 sampling 1958/2000ml at 0.40ml/s with 28.2psi vacuum @01:16:47.56 Cartridge 47 sampling 1981/2000ml at 0.39ml/s with 19.3psi vacuum @01:17:35.74 Pressure.coast @01:17:35.88 Intake.close @01:17:35.95 Exhaust.close @01:17:36.12 Toroid.reconfigure :hold @01:17:36.18 Toroid.seek :bypass @01:17:37.88 Toroid.reconfigure :default @01:17:37.98 Toroid.seek :clear @01:17:39.76 Sampled 2000.0ml @01:17:40.73 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 47" @01:17:41.84 SamplePump.setPosition! 0ml @01:17:41.89 Plunge.seek :reagent @01:17:41.95 Bypass.open @01:17:42.67 Intake.open @01:17:42.82 Sample loop closed. Ready to Deliver Reagents @01:17:42.84 Archive.seek :sealed @01:17:43.57 SamplePump.reconfigure :slow @01:17:43.63 SamplePump.seek -3ml @01:17:46.58 Archive.seek :evac @01:17:50.34 delay 1 second @01:17:51.45 SamplePump.seek -2.8ml @01:17:52.15 delay 1 second @01:17:53.25 SamplePump.seek -3.3ml @01:17:54.61 delay 1 second @01:17:55.65 Exhaust.open @01:17:56.43 Bypass.close @01:17:57.48 SamplePump.seek -35.3ml @01:17:58.01 upsync @01:18:00.63 delay 1:00 @01:19:00.69 Archive.seek :clear @01:19:06.18 delay 3 seconds @01:19:09.49 Intake.close @01:19:09.81 Exhaust.close @01:19:17.83 RNAlater soak for 10:00 @01:29:17.88 Archive.seek :tVent @01:29:19.69 delay 1 second @01:29:20.75 Archive.seek :evac @01:29:36.70 delay 10 seconds @01:29:46.75 Sample preserved and evacuated @01:29:47.24
Carousel.seek 47 @01:29:47.82 Plunge.seek :locked @01:29:48.80 Archive.seek :clear @01:29:48.95 Air.seek :clear @01:29:49.14 Lysis1.seek :clear @01:30:06.10
Archive.coast @01:30:06.26 Air.coast @01:30:06.30 Lysis1.coast @01:30:06.34 Lysis2.coast @01:30:06.38 Plunge.coast @01:30:06.43 SamplePump.coast @01:30:06.47 Toroid.coast @01:30:06.52 Carousel.coast @01:30:06.71 Gate.power :motor,:OFF @01:30:06.72 Gate.power :core,:OFF @01:30:06.74 Gate.power :analytic1,:OFF @01:30:06.75 Gate.power :analytic2,:OFF @01:30:06.76 Gate.power :analytic3,:OFF @01:30:06.77 Gate.power :main,:OFF @01:30:06.79 Safely stopped and ready to power off @01:30:07.01 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @01:30:07.05
delayUntil "12:00:00.00" @01:30:16.39 upsync @08:27:57.28 Session Begun @08:28:04.77 -> Cartridge.dry @08:28:19.81 -> slots @08:29:07.50 -> Cartridge.release 44..43 @08:29:26.13 -> Cartridge.release 43 @08:29:29.37 -> Cartridge.release 42 @08:29:33.51 -> Cartrixge.dry @08:29:33.55 NameError in preston -- uninitialized constant Cartrixge @08:29:40.26 -> Cartridge.dry @08:29:49.82 -> exit @08:29:49.86 Session Ended @10:56:15.28 Session Begun @10:56:26.70 -> Cartridge.dry @10:56:32.55 -> exit @10:56:32.58 Session Ended @12:00:00.11
Gate.power :main,:ON,limit to 5000mA @12:00:00.13 Gate.power :analytic1,:ON,limit to 5000mA @12:00:00.14 Gate.power :analytic2,:ON,limit to 5000mA @12:00:00.15 Gate.power :analytic3,:ON,limit to 5000mA @12:00:00.19 Powering core bus @12:00:02.72 Archive.configure :default @12:00:03.01 Air.configure :default @12:00:03.29 Lysis1.configure :default @12:00:03.53 Lysis2.configure :default @12:00:03.86 Plunge.configure :default @12:00:04.27 SamplePump.configure :default @12:00:04.62 ExtraPump.configure :default @12:00:04.92 Toroid.configure :default @12:00:05.35 Carousel.configure :default @12:00:05.68 Pressure.configure :default @12:00:05.74 All elf controllers running firmware version 4.61 @12:00:05.76 Gate.power :motor,:ON,limit to 5000mA @12:00:06.25 Archive.reconfigure :default @12:00:06.98 Archive.home.jog 0 @12:00:07.20 Plunge.reconfigure :default @12:00:07.24 Air.reconfigure :default @12:00:07.29 Toroid.reconfigure :default @12:00:07.36 Lysis1.reconfigure :default @12:00:07.40 Lysis2.reconfigure :default @12:00:07.60 Plunge.home.jog 0 @12:00:07.64 Air.home.jog 0 @12:00:07.66 Toroid.home.jog 0 @12:00:07.68 Lysis1.home.jog 0 @12:00:07.73 Lysis2.home.jog 0 @12:00:08.14 Archive.reconfigure :homing @12:00:08.44 Archive.jog 1375 @12:00:08.47 Plunge.reconfigure :homing @12:00:08.50 Air.reconfigure :homing @12:00:08.54 Toroid.reconfigure :homing @12:00:08.57 Lysis1.reconfigure :homing @12:00:08.78 Plunge.jog 844 @12:00:08.83 Toroid.jog 630 @12:00:08.88 Lysis1.jog 1347 @12:00:08.94 Air.jog 1363 @12:00:09.14 Lysis2.reconfigure :homing @12:00:09.19 Lysis2.jog 1358 @12:00:09.83 Archive.setRawPosition! 50 @12:00:09.92 Archive.seek :home @12:00:10.00 Plunge.setRawPosition! 50 @12:00:10.05 Plunge.seek :home @12:00:10.14 Lysis1.setRawPosition! 50 @12:00:10.21 Lysis1.seek :home @12:00:10.31 Air.setRawPosition! 50 @12:00:10.38 Archive.reconfigure :default @12:00:10.41 Air.seek :home @12:00:10.59 Lysis2.setRawPosition! 50 @12:00:10.62 Lysis1.reconfigure :default @12:00:10.64 Plunge.reconfigure :default @12:00:10.78 Lysis2.seek :home @12:00:10.90 Air.reconfigure :default @12:00:11.12 Toroid.setRawPosition! 0 @12:00:11.24 Toroid.seek :home @12:00:11.54 Lysis2.reconfigure :default @12:00:11.60 Toroid.reconfigure :default @12:00:12.36
Toroid.seek :clear @12:00:14.22 Carousel.reconfigure :default @12:00:14.34 Carousel.home.jog 0 @12:00:14.89 Carousel.seek 47 @12:00:15.22 Selecting Cartridge 46 @12:00:15.69 Carousel.seek 46 @12:00:16.54 Lysis1.seek :locked @12:00:19.85 Plunge.seek :locked @12:00:20.57 Air.seek :locked @12:00:24.86
Mounted archive Cartridge @12:00:25.07 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 46" @12:00:25.97 Duration of filtering limited to 4:05:00 @12:00:26.42 Priming sample loop w/160ml, bypass w/40ml @12:00:26.44 SamplePump.setPosition! 0ml @12:00:26.47 Bypass.open @12:00:27.95 Intake.open @12:00:28.88 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:00:31.54 Vacuum tare is 18% of full scale @12:00:31.58 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:00:32.08 Pressure.reconfigure :default @12:00:32.21 Pressure.hold @12:00:32.74 SamplePump.reconfigure :prime @12:00:32.89 SamplePump.seek 40ml,3:51 @12:00:42.61 upsync @12:01:16.16 Pressure.coast @12:01:16.22 SamplePump.setPosition! 0ml @12:01:16.27 Exhaust.open @12:01:17.70 Bypass.close @12:01:19.60 Pressure.reconfigure :default @12:01:19.70 Pressure.hold @12:01:20.48 SamplePump.reconfigure :prime @12:01:20.61 SamplePump.seek 160ml,13:51 @12:04:12.35 Pressure.coast @12:04:12.53 Bypass.open @12:04:13.38 Intake.close @12:04:14.45 Toroid.reconfigure :hold @12:04:14.56 Toroid.seek :crack @12:04:16.33 Toroid.seek :cartridge @12:04:17.26 Toroid.reconfigure :default @12:04:17.32 Toroid.seek :clear @12:04:19.18 Pressurizing cartridge to 26psi above 14.3psia @12:04:19.20 SamplePump.setPosition! 0ml @12:04:19.23 Intake.open @12:04:20.05 Bypass.close @12:04:20.75 Exhaust.close @12:04:22.27 Pressure.reconfigure :default @12:04:22.42 Pressure.hold @12:04:23.08 SamplePump.reconfigure :charge @12:04:23.22 SamplePump.seek 2.3ml,42 seconds @12:04:37.51 Pressure.coast @12:04:37.56 Intake.close @12:04:38.45 Pressurized cartridge with 0.925ml to 26.1psi @12:04:38.46 Isolated @35.9psia (21.6psi) @12:04:38.47 Checking for cartridge leaks from 35.9psia @12:04:38.49 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:04:41.08 Toroid.reconfigure :hold @12:04:41.21 Toroid.seek :bypass @12:04:43.04 Toroid.reconfigure :default @12:04:43.13 Toroid.seek :clear @12:04:44.80 Cartridge::Sampler::Leaky in FILTERING -- Isolated cartridge pressure fell by 22% to 28.0psia @12:04:44.82 Retry #1 of 2 @12:04:44.90 Cmd::SpareCartridge in FILTERING -- 0 spares remaining @12:04:45.00 Selecting Cartridge 45 @12:04:45.79 Air.seek :clear @12:04:49.58 Lysis1.seek :clear @12:04:49.74 Plunge.seek :clear @12:04:52.78 Carousel.seek 45 @12:04:53.55 Lysis1.seek :locked @12:04:56.79 Plunge.seek :locked @12:04:56.95 Air.seek :locked @12:05:01.31 Mounted archive Cartridge @12:05:01.47 Duration of filtering limited to 4:05:00 @12:05:02.15 Bypass.open @12:05:02.97 Toroid.reconfigure :hold @12:05:03.18 Toroid.seek :crack @12:05:04.80 Toroid.seek :cartridge @12:05:05.74 Toroid.reconfigure :default @12:05:05.80 Toroid.seek :clear @12:05:07.44 Exhaust.open @12:05:08.32 Pressurizing cartridge to 26psi above 14.6psia @12:05:08.33 SamplePump.setPosition! 0ml @12:05:08.37 Intake.open @12:05:09.20 Bypass.close @12:05:09.95 Exhaust.close @12:05:10.92 Pressure.reconfigure :default @12:05:11.01 Pressure.hold @12:05:11.54 SamplePump.seek 2.3ml,42 seconds @12:05:22.42 Pressure.coast @12:05:22.47 Intake.close @12:05:23.34 Pressurized cartridge with 0.424ml to 26.1psi @12:05:23.35 Isolated @35.1psia (20.5psi) @12:05:23.36 Checking for cartridge leaks from 35.1psia @12:05:23.38 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:05:36.58 Toroid.reconfigure :hold @12:05:36.64 Toroid.seek :bypass @12:05:38.40 Toroid.reconfigure :default @12:05:38.54 Toroid.seek :clear @12:05:40.24 Cartridge::Sampler::Leaky in FILTERING -- Isolated cartridge pressure fell by 20% to 28.1psia @12:05:40.26 Retry #2 of 2 @12:05:40.28 Cmd::Error in FILTERING -- No spare cartridges remain @12:05:40.29 Waiting to Resume from Checkpoint "Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi" @12:05:40.41 BadNews.email "No spare cartridges remain",Subject: "Cmd::Error in FILTERING" @12:05:50.84 upsync @12:22:26.91 Session Begun @12:23:25.06 -> slotss @12:23:32.33 -> slots @12:24:02.22 -> slot 45=>:dry @12:24:51.08 -> slot 46=>:dry @12:25:57.14 -> slot spare:1 @12:33:01.39 -> trouble @12:33:33.98 -> slot spare:0 @12:33:43.71 -> trouble @12:35:18.11 -> Thread[:FILTERING].resume -4 @12:35:18.14 Resuming Thread (suspended since 12:04:44) @12:35:18.19 Selecting Cartridge 45 @12:35:18.60 Duration of filtering limited to 4:05:00 @12:35:18.86 Bypass.open @12:35:19.68 Toroid.reconfigure :hold @12:35:19.76 Toroid.seek :crack @12:35:21.38 Toroid.seek :cartridge @12:35:22.56 Toroid.reconfigure :default @12:35:22.80 Toroid.seek :clear @12:35:24.52 Exhaust.open @12:35:25.76 Pressurizing cartridge to 26psi above 14.6psia @12:35:25.78 SamplePump.setPosition! 0ml @12:35:25.85 Intake.open @12:35:26.85 Bypass.close @12:35:27.62 Exhaust.close @12:35:28.44 Pressure.reconfigure :default @12:35:28.52 Pressure.hold @12:35:28.75 SamplePump.seek 2.3ml,42 seconds @12:35:38.07 Pressure.coast @12:35:38.16 Intake.close @12:35:39.08 Pressurized cartridge with 0.394ml to 26.0psi @12:35:39.09 Isolated @36.0psia (21.4psi) @12:35:39.11 Checking for cartridge leaks from 36.0psia @12:35:39.16 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:35:43.94 -> slot 46=:leaky @12:35:43.98 SyntaxError in preston -- compile error preston:10: syntax error, unexpected '=', expecting $end slot 46=:leaky ^ @12:35:46.77 Cartridge pressure fell 15% to 30.6psia @12:35:46.88 Sampling 2000.0ml @12:35:46.90 SamplePump.setPosition! 0ml @12:35:46.97 Exhaust.open @12:35:47.67 Intake.open @12:35:48.69 Pressure.reconfigure :default @12:35:48.78 Pressure.hold @12:35:49.10 SamplePump.reconfigure :sample @12:35:49.19 SamplePump.seek 2000ml,2:47:12 @12:35:51.88 Cartridge 45 sampling 1/2000ml at 0.52ml/s with 25.0psi vacuum @12:35:58.79 -> slot 46=>:leaky @12:36:22.82 -> exit @12:36:22.84 Session Ended @12:36:51.75 Cartridge 45 sampling 33/2000ml at 0.53ml/s with 23.5psi vacuum @12:37:52.06 Cartridge 45 sampling 66/2000ml at 0.56ml/s with 21.6psi vacuum @12:38:51.75 Cartridge 45 sampling 100/2000ml at 0.57ml/s with 27.4psi vacuum @12:39:51.86 Cartridge 45 sampling 134/2000ml at 0.58ml/s with 18.5psi vacuum 2.6m deep (18.6psia) @12:40:51.75 Cartridge 45 sampling 169/2000ml at 0.57ml/s with 26.0psi vacuum @12:41:53.19 Cartridge 45 sampling 203/2000ml at 0.57ml/s with 26.4psi vacuum @12:42:51.75 Cartridge 45 sampling 237/2000ml at 0.57ml/s with 26.3psi vacuum @12:43:51.84 Cartridge 45 sampling 272/2000ml at 0.56ml/s with 27.1psi vacuum @12:44:51.75 Cartridge 45 sampling 306/2000ml at 0.57ml/s with 26.3psi vacuum @12:45:52.00 Cartridge 45 sampling 340/2000ml at 0.57ml/s with 24.8psi vacuum @12:46:51.87 Cartridge 45 sampling 374/2000ml at 0.56ml/s with 21.2psi vacuum @12:47:51.75 Cartridge 45 sampling 407/2000ml at 0.55ml/s with 25.8psi vacuum @12:48:51.96 Cartridge 45 sampling 441/2000ml at 0.57ml/s with 25.8psi vacuum @12:49:51.75 Cartridge 45 sampling 474/2000ml at 0.56ml/s with 17.2psi vacuum 1.6m deep (17.1psia) @12:50:51.87 Cartridge 45 sampling 507/2000ml at 0.55ml/s with 20.3psi vacuum 1.7m deep (17.2psia) @12:51:51.75 Cartridge 45 sampling 540/2000ml at 0.54ml/s with 22.0psi vacuum @12:52:51.95 Cartridge 45 sampling 573/2000ml at 0.55ml/s with 19.3psi vacuum @12:53:51.75 Cartridge 45 sampling 606/2000ml at 0.55ml/s with 16.7psi vacuum 3.6m deep (20.0psia) @12:54:51.91 Cartridge 45 sampling 639/2000ml at 0.54ml/s with 22.5psi vacuum @12:55:51.75 Cartridge 45 sampling 672/2000ml at 0.52ml/s with 26.1psi vacuum @12:56:51.90 Cartridge 45 sampling 704/2000ml at 0.55ml/s with 21.8psi vacuum @12:57:51.75 Cartridge 45 sampling 737/2000ml at 0.54ml/s with 22.5psi vacuum @12:58:52.00 Cartridge 45 sampling 769/2000ml at 0.53ml/s with 21.8psi vacuum @12:59:52.19 Cartridge 45 sampling 801/2000ml at 0.54ml/s with 27.5psi vacuum @13:00:51.83 Cartridge 45 sampling 833/2000ml at 0.53ml/s with 25.6psi vacuum @13:01:52.11 Cartridge 45 sampling 865/2000ml at 0.52ml/s with 28.0psi vacuum @13:02:51.75 Cartridge 45 sampling 896/2000ml at 0.53ml/s with 16.6psi vacuum 3.2m deep (19.4psia) @13:03:51.86 Cartridge 45 sampling 928/2000ml at 0.52ml/s with 16.7psi vacuum 3.3m deep (19.6psia) @13:04:51.75 Cartridge 45 sampling 959/2000ml at 0.52ml/s with 27.5psi vacuum @13:05:51.93 Cartridge 45 sampling 991/2000ml at 0.50ml/s with 16.5psi vacuum @13:06:51.75 Cartridge 45 sampling 1022/2000ml at 0.53ml/s with 26.1psi vacuum @13:07:51.91 Cartridge 45 sampling 1053/2000ml at 0.53ml/s with 22.9psi vacuum @13:08:51.75 Cartridge 45 sampling 1083/2000ml at 0.51ml/s with 21.9psi vacuum @13:09:51.91 Cartridge 45 sampling 1114/2000ml at 0.49ml/s with 29.3psi vacuum @13:10:51.75 Cartridge 45 sampling 1145/2000ml at 0.52ml/s with 26.6psi vacuum @13:11:51.91 Cartridge 45 sampling 1175/2000ml at 0.51ml/s with 18.6psi vacuum @13:12:51.75 Cartridge 45 sampling 1206/2000ml at 0.51ml/s with 17.2psi vacuum 2.8m deep (18.8psia) @13:13:51.95 Cartridge 45 sampling 1236/2000ml at 0.51ml/s with 26.6psi vacuum @13:14:51.95 Cartridge 45 sampling 1266/2000ml at 0.51ml/s with 21.8psi vacuum @13:15:51.75 Cartridge 45 sampling 1296/2000ml at 0.50ml/s with 18.0psi vacuum @13:16:51.90 Cartridge 45 sampling 1326/2000ml at 0.49ml/s with 29.1psi vacuum @13:17:51.75 Cartridge 45 sampling 1355/2000ml at 0.49ml/s with 16.8psi vacuum 2.7m deep (18.7psia) @13:18:51.96 Cartridge 45 sampling 1385/2000ml at 0.47ml/s with 29.1psi vacuum @13:19:51.75 Cartridge 45 sampling 1414/2000ml at 0.49ml/s with 26.1psi vacuum @13:20:51.91 Cartridge 45 sampling 1443/2000ml at 0.48ml/s with 27.9psi vacuum @13:21:51.75 Cartridge 45 sampling 1472/2000ml at 0.48ml/s with 24.8psi vacuum @13:22:51.88 Cartridge 45 sampling 1501/2000ml at 0.47ml/s with 28.0psi vacuum @13:23:51.75 Cartridge 45 sampling 1529/2000ml at 0.47ml/s with 26.9psi vacuum @13:24:51.81 Cartridge 45 sampling 1558/2000ml at 0.48ml/s with 23.6psi vacuum @13:25:51.75 Cartridge 45 sampling 1586/2000ml at 0.48ml/s with 25.9psi vacuum @13:26:51.83 Cartridge 45 sampling 1614/2000ml at 0.46ml/s with 19.8psi vacuum @13:27:51.75 Cartridge 45 sampling 1642/2000ml at 0.47ml/s with 21.4psi vacuum @13:28:51.75 Cartridge 45 sampling 1670/2000ml at 0.45ml/s with 16.6psi vacuum @13:29:52.01 Cartridge 45 sampling 1697/2000ml at 0.48ml/s with 26.8psi vacuum @13:30:51.75 Cartridge 45 sampling 1725/2000ml at 0.43ml/s with 16.5psi vacuum @13:31:51.85 Cartridge 45 sampling 1753/2000ml at 0.45ml/s with 20.0psi vacuum @13:32:51.75 Cartridge 45 sampling 1780/2000ml at 0.43ml/s with 16.4psi vacuum @13:33:51.93 Cartridge 45 sampling 1807/2000ml at 0.45ml/s with 17.6psi vacuum @13:34:51.75 Cartridge 45 sampling 1834/2000ml at 0.44ml/s with 27.7psi vacuum @13:35:51.92 Cartridge 45 sampling 1861/2000ml at 0.44ml/s with 27.6psi vacuum @13:36:51.75 Cartridge 45 sampling 1888/2000ml at 0.45ml/s with 25.5psi vacuum @13:37:51.91 Cartridge 45 sampling 1914/2000ml at 0.42ml/s with 29.1psi vacuum @13:38:51.75 Cartridge 45 sampling 1941/2000ml at 0.43ml/s with 17.1psi vacuum @13:39:51.91 Cartridge 45 sampling 1967/2000ml at 0.45ml/s with 26.5psi vacuum @13:40:51.75 Cartridge 45 sampling 1993/2000ml at 0.45ml/s with 24.6psi vacuum @13:41:07.38 Pressure.coast @13:41:07.51 Intake.close @13:41:07.59 Exhaust.close @13:41:07.76 Toroid.reconfigure :hold @13:41:07.82 Toroid.seek :bypass @13:41:09.45 Toroid.reconfigure :default @13:41:09.50 Toroid.seek :clear @13:41:11.18 Sampled 2000.0ml @13:41:12.33 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 45" @13:41:13.17 SamplePump.setPosition! 0ml @13:41:13.29 Plunge.seek :reagent @13:41:13.38 Bypass.open @13:41:14.25 Intake.open @13:41:14.41 Sample loop closed. Ready to Deliver Reagents @13:41:14.43 Archive.seek :sealed @13:41:15.44 SamplePump.reconfigure :slow @13:41:15.57 SamplePump.seek -3ml @13:41:18.07 Archive.seek :evac @13:41:22.39 delay 1 second @13:41:23.96 SamplePump.seek -2.8ml @13:41:24.83 delay 1 second @13:41:25.97 SamplePump.seek -3.3ml @13:41:27.30 delay 1 second @13:41:28.33 Exhaust.open @13:41:29.03 Bypass.close @13:41:30.17 SamplePump.seek -35.3ml @13:41:32.11 delay 1:00 @13:42:03.05 upsync @13:42:32.20 Archive.seek :clear @13:42:39.14 delay 3 seconds @13:42:42.27 Intake.close @13:42:42.31 Exhaust.close @13:42:49.35 RNAlater soak for 10:00 @13:52:49.44 Archive.seek :tVent @13:52:51.22 delay 1 second @13:52:52.35 Archive.seek :evac @13:53:08.37 delay 10 seconds @13:53:18.43 Sample preserved and evacuated @13:53:18.83
Carousel.seek 45 @13:53:19.36 Plunge.seek :locked @13:53:20.43 Archive.seek :clear @13:53:20.70 Air.seek :clear @13:53:21.00 Lysis1.seek :clear @13:53:37.81
Archive.coast @13:53:37.86 Air.coast @13:53:37.94 Lysis1.coast @13:53:38.10 Lysis2.coast @13:53:38.15 Plunge.coast @13:53:38.34 SamplePump.coast @13:53:38.40 Toroid.coast @13:53:38.53 Carousel.coast @13:53:38.89 Gate.power :motor,:OFF @13:53:38.93 Gate.power :core,:OFF @13:53:38.94 Gate.power :analytic1,:OFF @13:53:38.95 Gate.power :analytic2,:OFF @13:53:38.97 Gate.power :analytic3,:OFF @13:53:38.98 Gate.power :main,:OFF @13:53:38.99 Safely stopped and ready to power off @13:53:39.53 delayUntil "00:00:00.00PDT15-Jun-21" @13:53:39.56 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT15-Jun-21" @13:53:57.62 upsync @00:00:00.29PDT15-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.35 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.37 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.42 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.53 Powering core bus @00:00:02.91 Archive.configure :default @00:00:03.05 Air.configure :default @00:00:03.70 Lysis1.configure :default @00:00:03.81 Lysis2.configure :default @00:00:03.94 Plunge.configure :default @00:00:04.08 SamplePump.configure :default @00:00:04.21 ExtraPump.configure :default @00:00:04.34 Toroid.configure :default @00:00:04.48 Carousel.configure :default @00:00:04.61 Pressure.configure :default @00:00:04.67 All elf controllers running firmware version 4.61 @00:00:04.69 Gate.power :motor,:ON,limit to 5000mA @00:00:04.92 Archive.reconfigure :default @00:00:05.21 Archive.home.jog 0 @00:00:05.51 Plunge.reconfigure :default @00:00:05.55 Lysis2.reconfigure :default @00:00:05.58 Lysis1.reconfigure :default @00:00:05.63 Air.reconfigure :default @00:00:05.68 Toroid.reconfigure :default @00:00:05.86 Lysis2.home.jog 0 @00:00:05.90 Plunge.home.jog 0 @00:00:05.93 Lysis1.home.jog 0 @00:00:05.97 Toroid.home.jog 0 @00:00:06.13 Air.home.jog 0 @00:00:06.53 Archive.reconfigure :homing @00:00:06.72 Lysis1.reconfigure :homing @00:00:06.79 Archive.jog 1369 @00:00:06.82 Plunge.reconfigure :homing @00:00:06.85 Lysis2.reconfigure :homing @00:00:06.91 Air.reconfigure :homing @00:00:06.94 Toroid.reconfigure :homing @00:00:07.53 Plunge.jog 844 @00:00:07.81 Lysis2.jog 1352 @00:00:07.89 Lysis1.jog 1355 @00:00:07.93 Air.jog 1361 @00:00:08.01 Toroid.jog 629 @00:00:08.77 Archive.setRawPosition! 50 @00:00:08.85 Archive.seek :home @00:00:09.07 Plunge.setRawPosition! 50 @00:00:09.20 Plunge.seek :home @00:00:09.38 Archive.reconfigure :default @00:00:09.83 Lysis1.setRawPosition! 50 @00:00:09.88 Plunge.reconfigure :default @00:00:09.91 Lysis2.setRawPosition! 50 @00:00:09.97 Air.setRawPosition! 50 @00:00:10.33 Lysis1.seek :home @00:00:10.52 Lysis2.seek :home @00:00:10.60 Air.seek :home @00:00:11.30 Lysis2.reconfigure :default @00:00:11.40 Lysis1.reconfigure :default @00:00:11.45 Toroid.setRawPosition! 0 @00:00:11.53 Air.reconfigure :default @00:00:11.93 Toroid.seek :home @00:00:12.21 Toroid.reconfigure :default @00:00:12.52
Toroid.seek :clear @00:00:14.14 Carousel.reconfigure :default @00:00:14.19 Carousel.home.jog 0 @00:00:14.68 Carousel.seek 45 @00:00:14.82 Selecting Cartridge 44 @00:00:15.04 Carousel.seek 44 @00:00:15.56 Lysis1.seek :locked @00:00:18.98 Plunge.seek :locked @00:00:19.33 Air.seek :locked @00:00:23.61
Mounted archive Cartridge @00:00:23.71 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 44" @00:00:24.04 Duration of filtering limited to 4:05:00 @00:00:24.33 Priming sample loop w/160ml, bypass w/40ml @00:00:24.35 SamplePump.setPosition! 0ml @00:00:24.40 Bypass.open @00:00:25.86 Intake.open @00:00:26.74 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:29.45 Vacuum tare is 18% of full scale @00:00:29.50 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:30.18 Pressure.reconfigure :default @00:00:30.38 Pressure.hold @00:00:31.04 SamplePump.reconfigure :prime @00:00:31.13 SamplePump.seek 40ml,3:51 @00:00:39.16 upsync @00:01:15.37 Pressure.coast @00:01:15.51 SamplePump.setPosition! 0ml @00:01:15.69 Exhaust.open @00:01:17.23 Bypass.close @00:01:18.87 Pressure.reconfigure :default @00:01:18.93 Pressure.hold @00:01:19.21 SamplePump.reconfigure :prime @00:01:19.27 SamplePump.seek 160ml,13:51 @00:04:10.93 Pressure.coast @00:04:11.09 Bypass.open @00:04:11.77 Intake.close @00:04:12.89 Toroid.reconfigure :hold @00:04:13.01 Toroid.seek :crack @00:04:14.68 Toroid.seek :cartridge @00:04:15.61 Toroid.reconfigure :default @00:04:15.67 Toroid.seek :clear @00:04:17.93 Pressurizing cartridge to 26psi above 14.5psia @00:04:18.01 SamplePump.setPosition! 0ml @00:04:18.16 Intake.open @00:04:18.97 Bypass.close @00:04:19.75 Exhaust.close @00:04:20.78 Pressure.reconfigure :default @00:04:20.95 Pressure.hold @00:04:21.40 SamplePump.reconfigure :charge @00:04:21.55 SamplePump.seek 2.3ml,42 seconds @00:04:32.69 Pressure.coast @00:04:32.74 Intake.close @00:04:33.56 Pressurized cartridge with 0.568ml to 26.2psi @00:04:33.57 Isolated @36.7psia (22.2psi) @00:04:33.58 Checking for cartridge leaks from 36.7psia @00:04:33.60 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:42.20 Cartridge pressure fell 17% to 30.6psia @00:04:42.28 Sampling 2000.0ml @00:04:42.30 SamplePump.setPosition! 0ml @00:04:42.34 Exhaust.open @00:04:43.05 Intake.open @00:04:43.94 Pressure.reconfigure :default @00:04:44.00 Pressure.hold @00:04:44.28 SamplePump.reconfigure :sample @00:04:44.35 SamplePump.seek 2000ml,2:47:12 @00:04:47.29 Cartridge 44 sampling 1/2000ml at 0.48ml/s with 25.3psi vacuum @00:05:47.01 Cartridge 44 sampling 30/2000ml at 0.48ml/s with 25.1psi vacuum @00:06:47.11 Cartridge 44 sampling 58/2000ml at 0.47ml/s with 20.8psi vacuum @00:07:47.01 Cartridge 44 sampling 87/2000ml at 0.46ml/s with 25.9psi vacuum @00:08:47.17 Cartridge 44 sampling 115/2000ml at 0.50ml/s with 22.8psi vacuum @00:09:47.01 Cartridge 44 sampling 145/2000ml at 0.46ml/s with 25.1psi vacuum @00:10:47.18 Cartridge 44 sampling 173/2000ml at 0.48ml/s with 21.7psi vacuum @00:11:47.01 Cartridge 44 sampling 203/2000ml at 0.51ml/s with 27.4psi vacuum @00:12:47.19 Cartridge 44 sampling 233/2000ml at 0.50ml/s with 27.4psi vacuum @00:13:47.01 Cartridge 44 sampling 262/2000ml at 0.51ml/s with 23.8psi vacuum @00:14:47.10 Cartridge 44 sampling 292/2000ml at 0.49ml/s with 17.9psi vacuum 1.9m deep (17.5psia) @00:15:47.01 Cartridge 44 sampling 322/2000ml at 0.47ml/s with 21.1psi vacuum @00:16:47.19 Cartridge 44 sampling 351/2000ml at 0.49ml/s with 23.4psi vacuum @00:17:47.01 Cartridge 44 sampling 380/2000ml at 0.47ml/s with 15.6psi vacuum 3.1m deep (19.2psia) @00:18:47.01 Cartridge 44 sampling 409/2000ml at 0.47ml/s with 23.5psi vacuum @00:19:47.25 Cartridge 44 sampling 438/2000ml at 0.48ml/s with 25.4psi vacuum @00:20:47.01 Cartridge 44 sampling 466/2000ml at 0.48ml/s with 27.3psi vacuum @00:21:47.20 Cartridge 44 sampling 495/2000ml at 0.48ml/s with 23.4psi vacuum @00:22:47.01 Cartridge 44 sampling 523/2000ml at 0.47ml/s with 23.0psi vacuum @00:23:47.13 Cartridge 44 sampling 552/2000ml at 0.47ml/s with 25.7psi vacuum @00:24:47.01 Cartridge 44 sampling 580/2000ml at 0.47ml/s with 23.8psi vacuum @00:25:47.22 Cartridge 44 sampling 608/2000ml at 0.46ml/s with 25.6psi vacuum @00:26:47.01 Cartridge 44 sampling 636/2000ml at 0.48ml/s with 24.7psi vacuum @00:27:47.31 Cartridge 44 sampling 664/2000ml at 0.46ml/s with 25.6psi vacuum @00:28:47.01 Cartridge 44 sampling 691/2000ml at 0.47ml/s with 25.9psi vacuum @00:29:47.09 Cartridge 44 sampling 719/2000ml at 0.46ml/s with 25.3psi vacuum @00:30:47.01 Cartridge 44 sampling 746/2000ml at 0.45ml/s with 25.4psi vacuum @00:31:47.37 Cartridge 44 sampling 773/2000ml at 0.46ml/s with 18.9psi vacuum 3.4m deep (19.7psia) @00:32:47.24 Cartridge 44 sampling 800/2000ml at 0.44ml/s with 25.2psi vacuum @00:33:47.01 Cartridge 44 sampling 826/2000ml at 0.45ml/s with 26.6psi vacuum @00:34:47.47 Cartridge 44 sampling 853/2000ml at 0.45ml/s with 25.5psi vacuum @00:35:47.01 Cartridge 44 sampling 880/2000ml at 0.43ml/s with 23.1psi vacuum @00:36:47.18 Cartridge 44 sampling 906/2000ml at 0.43ml/s with 23.9psi vacuum @00:37:47.01 Cartridge 44 sampling 932/2000ml at 0.42ml/s with 20.7psi vacuum @00:38:47.16 Cartridge 44 sampling 958/2000ml at 0.43ml/s with 27.1psi vacuum @00:39:47.01 Cartridge 44 sampling 984/2000ml at 0.42ml/s with 13.6psi vacuum 4.1m deep (20.7psia) @00:40:47.23 Cartridge 44 sampling 1010/2000ml at 0.43ml/s with 25.8psi vacuum @00:41:47.00 Cartridge 44 sampling 1035/2000ml at 0.43ml/s with 23.8psi vacuum @00:42:47.14 Cartridge 44 sampling 1061/2000ml at 0.42ml/s with 25.3psi vacuum @00:43:47.56 Cartridge 44 sampling 1086/2000ml at 0.41ml/s with 21.2psi vacuum @00:44:47.06 Cartridge 44 sampling 1111/2000ml at 0.42ml/s with 25.3psi vacuum @00:45:47.00 Cartridge 44 sampling 1135/2000ml at 0.41ml/s with 26.0psi vacuum @00:46:47.14 Cartridge 44 sampling 1160/2000ml at 0.41ml/s with 25.9psi vacuum @00:47:47.27 Cartridge 44 sampling 1184/2000ml at 0.41ml/s with 17.2psi vacuum 4.0m deep (20.7psia) @00:48:47.01 Cartridge 44 sampling 1208/2000ml at 0.40ml/s with 18.0psi vacuum 3.1m deep (19.2psia) @00:49:47.32 Cartridge 44 sampling 1232/2000ml at 0.39ml/s with 18.4psi vacuum 3.6m deep (20.0psia) @00:50:47.01 Cartridge 44 sampling 1256/2000ml at 0.39ml/s with 25.8psi vacuum @00:51:47.17 Cartridge 44 sampling 1280/2000ml at 0.41ml/s with 25.5psi vacuum @00:52:47.01 Cartridge 44 sampling 1303/2000ml at 0.39ml/s with 23.1psi vacuum @00:53:47.15 Cartridge 44 sampling 1327/2000ml at 0.38ml/s with 16.3psi vacuum 4.3m deep (21.1psia) @00:54:47.01 Cartridge 44 sampling 1350/2000ml at 0.40ml/s with 25.4psi vacuum @00:55:47.14 Cartridge 44 sampling 1373/2000ml at 0.39ml/s with 19.7psi vacuum 2.2m deep (17.9psia) @00:56:47.01 Cartridge 44 sampling 1396/2000ml at 0.36ml/s with 18.3psi vacuum @00:57:47.09 Cartridge 44 sampling 1419/2000ml at 0.39ml/s with 22.3psi vacuum @00:58:47.01 Cartridge 44 sampling 1442/2000ml at 0.38ml/s with 16.2psi vacuum 4.0m deep (20.7psia) @00:59:47.11 Cartridge 44 sampling 1464/2000ml at 0.38ml/s with 23.8psi vacuum @01:00:47.01 Cartridge 44 sampling 1486/2000ml at 0.37ml/s with 24.3psi vacuum @01:01:47.00 Cartridge 44 sampling 1508/2000ml at 0.36ml/s with 25.0psi vacuum @01:02:47.43 Cartridge 44 sampling 1530/2000ml at 0.36ml/s with 22.2psi vacuum @01:03:47.00 Cartridge 44 sampling 1552/2000ml at 0.36ml/s with 25.3psi vacuum @01:04:47.17 Cartridge 44 sampling 1574/2000ml at 0.38ml/s with 26.0psi vacuum @01:05:47.00 Cartridge 44 sampling 1595/2000ml at 0.36ml/s with 27.4psi vacuum @01:06:47.19 Cartridge 44 sampling 1617/2000ml at 0.36ml/s with 27.7psi vacuum @01:07:47.00 Cartridge 44 sampling 1639/2000ml at 0.36ml/s with 25.0psi vacuum @01:08:47.29 Cartridge 44 sampling 1660/2000ml at 0.37ml/s with 25.3psi vacuum @01:09:47.00 Cartridge 44 sampling 1681/2000ml at 0.36ml/s with 25.9psi vacuum @01:10:47.16 Cartridge 44 sampling 1702/2000ml at 0.33ml/s with 27.3psi vacuum @01:11:47.00 Cartridge 44 sampling 1723/2000ml at 0.32ml/s with 18.8psi vacuum @01:12:47.13 Cartridge 44 sampling 1743/2000ml at 0.35ml/s with 27.1psi vacuum @01:13:47.00 Cartridge 44 sampling 1764/2000ml at 0.33ml/s with 23.2psi vacuum @01:14:47.27 Cartridge 44 sampling 1784/2000ml at 0.34ml/s with 26.6psi vacuum @01:15:47.24 Cartridge 44 sampling 1805/2000ml at 0.35ml/s with 24.1psi vacuum @01:16:47.16 Cartridge 44 sampling 1825/2000ml at 0.34ml/s with 26.0psi vacuum @01:17:47.28 Cartridge 44 sampling 1845/2000ml at 0.35ml/s with 24.3psi vacuum @01:18:47.01 Cartridge 44 sampling 1865/2000ml at 0.34ml/s with 25.3psi vacuum @01:19:47.17 Cartridge 44 sampling 1885/2000ml at 0.32ml/s with 22.3psi vacuum @01:20:47.01 Cartridge 44 sampling 1904/2000ml at 0.33ml/s with 20.0psi vacuum @01:21:47.22 Cartridge 44 sampling 1924/2000ml at 0.32ml/s with 23.4psi vacuum @01:22:47.01 Cartridge 44 sampling 1943/2000ml at 0.33ml/s with 22.3psi vacuum @01:23:47.18 Cartridge 44 sampling 1962/2000ml at 0.32ml/s with 23.9psi vacuum @01:24:47.01 Cartridge 44 sampling 1981/2000ml at 0.33ml/s with 16.7psi vacuum 3.1m deep (19.2psia) @01:25:46.24 Pressure.coast @01:25:46.59 Intake.close @01:25:46.68 Exhaust.close @01:25:47.13 Toroid.reconfigure :hold @01:25:47.21 Toroid.seek :bypass @01:25:48.93 Toroid.reconfigure :default @01:25:48.99 Toroid.seek :clear @01:25:50.60 Sampled 2000.0ml @01:25:51.04 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 44" @01:25:51.47 SamplePump.setPosition! 0ml @01:25:51.51 Plunge.seek :reagent @01:25:51.60 Bypass.open @01:25:52.30 Intake.open @01:25:52.44 Sample loop closed. Ready to Deliver Reagents @01:25:52.46 Archive.seek :sealed @01:25:53.23 SamplePump.reconfigure :slow @01:25:53.39 SamplePump.seek -3ml @01:25:56.18 Archive.seek :evac @01:26:00.13 delay 1 second @01:26:01.21 SamplePump.seek -2.8ml @01:26:01.90 delay 1 second @01:26:03.07 SamplePump.seek -3.3ml @01:26:04.53 delay 1 second @01:26:05.59 Exhaust.open @01:26:06.37 Bypass.close @01:26:06.83 upsync @01:26:07.28 SamplePump.seek -35.3ml @01:26:10.23 delay 1:00 @01:27:10.32 Archive.seek :clear @01:27:16.05 delay 3 seconds @01:27:19.26 Intake.close @01:27:19.34 Exhaust.close @01:27:27.53 RNAlater soak for 10:00 @01:37:27.59 Archive.seek :tVent @01:37:29.24 delay 1 second @01:37:30.28 Archive.seek :evac @01:37:46.25 delay 10 seconds @01:37:56.29 Sample preserved and evacuated @01:37:56.52
Carousel.seek 44 @01:37:56.73 Plunge.seek :locked @01:37:57.71 Archive.seek :clear @01:37:57.87 Air.seek :clear @01:37:58.03 Lysis1.seek :clear @01:38:14.90
Archive.coast @01:38:14.94 Air.coast @01:38:14.98 Lysis1.coast @01:38:15.03 Lysis2.coast @01:38:15.07 Plunge.coast @01:38:15.11 SamplePump.coast @01:38:15.16 Toroid.coast @01:38:15.20 Carousel.coast @01:38:15.39 Gate.power :motor,:OFF @01:38:15.40 Gate.power :core,:OFF @01:38:15.41 Gate.power :analytic1,:OFF @01:38:15.43 Gate.power :analytic2,:OFF @01:38:15.44 Gate.power :analytic3,:OFF @01:38:15.45 Gate.power :main,:OFF @01:38:15.47 Safely stopped and ready to power off @01:38:15.70 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @01:38:15.74
delayUntil "12:00:00.00" @01:38:28.09 upsync @12:00:00.17
Gate.power :main,:ON,limit to 5000mA @12:00:00.19 Gate.power :analytic1,:ON,limit to 5000mA @12:00:00.20 Gate.power :analytic2,:ON,limit to 5000mA @12:00:00.26 Gate.power :analytic3,:ON,limit to 5000mA @12:00:00.41 Powering core bus @12:00:02.73 Archive.configure :default @12:00:02.86 Air.configure :default @12:00:02.97 Lysis1.configure :default @12:00:03.08 Lysis2.configure :default @12:00:03.21 Plunge.configure :default @12:00:03.35 SamplePump.configure :default @12:00:03.48 ExtraPump.configure :default @12:00:03.62 Toroid.configure :default @12:00:03.75 Carousel.configure :default @12:00:03.88 Pressure.configure :default @12:00:03.94 All elf controllers running firmware version 4.61 @12:00:03.96 Gate.power :motor,:ON,limit to 5000mA @12:00:04.27 Archive.reconfigure :default @12:00:04.51 Archive.home.jog 0 @12:00:04.70 Plunge.reconfigure :default @12:00:04.74 Lysis2.reconfigure :default @12:00:04.81 Lysis1.reconfigure :default @12:00:04.87 Air.reconfigure :default @12:00:04.93 Toroid.reconfigure :default @12:00:05.06 Plunge.home.jog 0 @12:00:05.14 Lysis2.home.jog 0 @12:00:05.95 Air.home.jog 0 @12:00:06.45 Lysis1.home.jog 0 @12:00:06.51 Toroid.home.jog 0 @12:00:08.10 Archive.reconfigure :homing @12:00:08.51 Archive.jog 1374 @12:00:08.62 Plunge.reconfigure :homing @12:00:08.68 Lysis2.reconfigure :homing @12:00:08.74 Air.reconfigure :homing @12:00:08.88 Lysis1.reconfigure :homing @12:00:08.98 Toroid.reconfigure :homing @12:00:09.60 Plunge.jog 847 @12:00:09.67 Lysis1.jog 1351 @12:00:09.75 Lysis2.jog 1359 @12:00:09.77 Air.jog 1364 @12:00:09.80 Toroid.jog 623 @12:00:10.36 Archive.setRawPosition! 50 @12:00:10.43 Archive.seek :home @12:00:10.73 Archive.reconfigure :default @12:00:10.98 Plunge.setRawPosition! 50 @12:00:11.02 Plunge.seek :home @12:00:11.33 Lysis1.setRawPosition! 50 @12:00:11.35 Lysis2.setRawPosition! 50 @12:00:11.49 Lysis1.seek :home @12:00:11.59 Lysis2.seek :home @12:00:12.02 Plunge.reconfigure :default @12:00:12.15 Lysis2.reconfigure :default @12:00:12.30 Air.setRawPosition! 50 @12:00:12.45 Air.seek :home @12:00:12.55 Toroid.setRawPosition! 0 @12:00:12.58 Lysis1.reconfigure :default @12:00:12.67 Toroid.seek :home @12:00:12.81 Air.reconfigure :default @12:00:12.84 Toroid.reconfigure :default @12:00:13.28
Toroid.seek :clear @12:00:14.89 Carousel.reconfigure :default @12:00:14.94 Carousel.home.jog 0 @12:00:15.44 Carousel.seek 44 @12:00:15.60 Selecting Cartridge 43 @12:00:15.82 Carousel.seek 43 @12:00:16.33 Lysis1.seek :locked @12:00:19.77 Plunge.seek :locked @12:00:20.08 Air.seek :locked @12:00:24.31
Mounted archive Cartridge @12:00:24.43 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 43" @12:00:24.50 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:00:24.52 Retry #1 of 10 after 10:00 @12:00:24.77 Duration of filtering limited to 4:05:00 @12:00:25.06 Priming sample loop w/160ml, bypass w/40ml @12:00:25.08 SamplePump.setPosition! 0ml @12:00:25.12 Bypass.open @12:00:26.77 Intake.open @12:00:27.87 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:00:30.61 Vacuum tare is 18% of full scale @12:00:30.72 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:00:31.11 Pressure.reconfigure :default @12:00:31.17 Pressure.hold @12:00:31.37 SamplePump.reconfigure :prime @12:00:31.44 SamplePump.seek 40ml,3:51 @12:01:14.65 Pressure.coast @12:01:14.76 SamplePump.setPosition! 0ml @12:01:14.90 Exhaust.open @12:01:16.32 Bypass.close @12:01:17.88 Pressure.reconfigure :default @12:01:17.94 Pressure.hold @12:01:18.21 SamplePump.reconfigure :prime @12:01:18.27 SamplePump.seek 160ml,13:51 @12:04:09.98 Pressure.coast @12:04:10.08 Bypass.open @12:04:10.79 Intake.close @12:04:11.75 Toroid.reconfigure :hold @12:04:11.81 Toroid.seek :crack @12:04:13.39 Toroid.seek :cartridge @12:04:14.32 Toroid.reconfigure :default @12:04:14.38 Toroid.seek :clear @12:04:16.60 Pressurizing cartridge to 26psi above 14.5psia @12:04:16.70 SamplePump.setPosition! 0ml @12:04:16.80 Intake.open @12:04:17.83 Bypass.close @12:04:18.60 Exhaust.close @12:04:19.70 Pressure.reconfigure :default @12:04:19.78 Pressure.hold @12:04:20.26 SamplePump.reconfigure :charge @12:04:20.37 SamplePump.seek 2.3ml,42 seconds @12:04:33.62 Pressure.coast @12:04:34.08 Intake.close @12:04:35.26 Pressurized cartridge with 0.799ml to 26.3psi @12:04:35.29 Isolated @35.8psia (21.3psi) @12:04:35.32 Checking for cartridge leaks from 35.8psia @12:04:35.37 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:04:43.53 Cartridge pressure fell 8% to 33.0psia @12:04:43.61 Sampling 2000.0ml @12:04:43.63 SamplePump.setPosition! 0ml @12:04:43.67 Exhaust.open @12:04:44.39 Intake.open @12:04:45.64 Pressure.reconfigure :default @12:04:45.76 Pressure.hold @12:04:46.53 SamplePump.reconfigure :sample @12:04:46.65 SamplePump.seek 2000ml,2:47:12 @12:04:48.73 Cartridge 43 sampling 1/2000ml at 0.57ml/s with 29.6psi vacuum @12:05:48.58 Cartridge 43 sampling 30/2000ml at 0.48ml/s with 20.9psi vacuum @12:06:48.84 Cartridge 43 sampling 60/2000ml at 0.52ml/s with 24.1psi vacuum @12:07:49.04 Cartridge 43 sampling 91/2000ml at 0.53ml/s with 24.5psi vacuum @12:08:48.58 Cartridge 43 sampling 123/2000ml at 0.54ml/s with 25.9psi vacuum @12:09:48.95 Cartridge 43 sampling 155/2000ml at 0.54ml/s with 26.0psi vacuum @12:10:25.09 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:10:25.13 Retry #2 of 10 after 10:00 @12:10:48.57 Cartridge 43 sampling 187/2000ml at 0.54ml/s with 24.7psi vacuum @12:11:48.84 Cartridge 43 sampling 219/2000ml at 0.54ml/s with 17.9psi vacuum 4.4m deep (21.2psia) @12:12:48.57 Cartridge 43 sampling 251/2000ml at 0.54ml/s with 26.3psi vacuum @12:13:48.73 Cartridge 43 sampling 283/2000ml at 0.51ml/s with 25.0psi vacuum @12:14:48.57 Cartridge 43 sampling 315/2000ml at 0.53ml/s with 21.4psi vacuum @12:15:48.77 Cartridge 43 sampling 347/2000ml at 0.53ml/s with 18.7psi vacuum 2.3m deep (18.0psia) @12:16:48.57 Cartridge 43 sampling 379/2000ml at 0.52ml/s with 22.1psi vacuum @12:17:48.61 Cartridge 43 sampling 410/2000ml at 0.52ml/s with 21.1psi vacuum @12:18:48.57 Cartridge 43 sampling 441/2000ml at 0.53ml/s with 25.6psi vacuum @12:19:48.65 Cartridge 43 sampling 472/2000ml at 0.52ml/s with 28.1psi vacuum @12:20:25.26 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:20:25.28 Retry #3 of 10 after 10:00 @12:20:48.57 Cartridge 43 sampling 503/2000ml at 0.51ml/s with 28.0psi vacuum @12:21:48.89 Cartridge 43 sampling 534/2000ml at 0.52ml/s with 23.3psi vacuum @12:22:48.86 Cartridge 43 sampling 564/2000ml at 0.49ml/s with 29.6psi vacuum @12:23:48.58 Cartridge 43 sampling 594/2000ml at 0.47ml/s with 27.3psi vacuum @12:24:48.80 Cartridge 43 sampling 624/2000ml at 0.50ml/s with 29.3psi vacuum @12:25:48.58 Cartridge 43 sampling 654/2000ml at 0.48ml/s with 23.9psi vacuum @12:26:48.80 Cartridge 43 sampling 684/2000ml at 0.49ml/s with 17.9psi vacuum 4.4m deep (21.1psia) @12:27:48.58 Cartridge 43 sampling 713/2000ml at 0.50ml/s with 21.5psi vacuum @12:28:48.77 Cartridge 43 sampling 742/2000ml at 0.49ml/s with 25.7psi vacuum @12:29:48.58 Cartridge 43 sampling 771/2000ml at 0.47ml/s with 29.3psi vacuum @12:30:25.33 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:30:25.35 Retry #4 of 10 after 10:00 @12:30:48.85 Cartridge 43 sampling 799/2000ml at 0.47ml/s with 27.3psi vacuum @12:31:48.58 Cartridge 43 sampling 828/2000ml at 0.47ml/s with 16.3psi vacuum 2.9m deep (19.0psia) @12:32:48.63 Cartridge 43 sampling 856/2000ml at 0.47ml/s with 16.9psi vacuum 3.3m deep (19.6psia) @12:33:48.58 Cartridge 43 sampling 883/2000ml at 0.44ml/s with 28.8psi vacuum @12:34:49.05 Cartridge 43 sampling 911/2000ml at 0.46ml/s with 27.3psi vacuum @12:35:48.78 Cartridge 43 sampling 938/2000ml at 0.46ml/s with 16.5psi vacuum 6.1m deep (23.7psia) @12:36:48.63 Cartridge 43 sampling 965/2000ml at 0.46ml/s with 26.9psi vacuum @12:37:48.91 Cartridge 43 sampling 992/2000ml at 0.44ml/s with 29.2psi vacuum @12:38:48.57 Cartridge 43 sampling 1018/2000ml at 0.47ml/s with 21.5psi vacuum @12:39:48.76 Cartridge 43 sampling 1045/2000ml at 0.42ml/s with 15.4psi vacuum 3.1m deep (19.2psia) @12:40:25.50 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:40:25.53 Retry #5 of 10 after 10:00 @12:40:48.57 Cartridge 43 sampling 1071/2000ml at 0.43ml/s with 16.0psi vacuum 3.1m deep (19.3psia) @12:41:48.79 Cartridge 43 sampling 1097/2000ml at 0.42ml/s with 28.6psi vacuum @12:42:48.58 Cartridge 43 sampling 1122/2000ml at 0.42ml/s with 27.8psi vacuum @12:43:48.75 Cartridge 43 sampling 1147/2000ml at 0.41ml/s with 16.4psi vacuum 5.0m deep (22.0psia) @12:44:48.58 Cartridge 43 sampling 1172/2000ml at 0.42ml/s with 23.8psi vacuum @12:45:48.71 Cartridge 43 sampling 1197/2000ml at 0.39ml/s with 27.8psi vacuum @12:46:48.58 Cartridge 43 sampling 1222/2000ml at 0.40ml/s with 18.3psi vacuum @12:47:48.74 Cartridge 43 sampling 1246/2000ml at 0.38ml/s with 30.7psi vacuum @12:48:48.58 Cartridge 43 sampling 1271/2000ml at 0.39ml/s with 29.5psi vacuum @12:49:48.82 Cartridge 43 sampling 1294/2000ml at 0.39ml/s with 15.4psi vacuum @12:50:25.57 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:50:25.59 Retry #6 of 10 after 10:00 @12:50:48.92 Cartridge 43 sampling 1318/2000ml at 0.42ml/s with 19.7psi vacuum @12:51:48.58 Cartridge 43 sampling 1341/2000ml at 0.40ml/s with 25.4psi vacuum @12:52:48.89 Cartridge 43 sampling 1365/2000ml at 0.40ml/s with 24.4psi vacuum @12:53:48.58 Cartridge 43 sampling 1387/2000ml at 0.37ml/s with 29.2psi vacuum @12:54:48.75 Cartridge 43 sampling 1410/2000ml at 0.38ml/s with 20.5psi vacuum 3.3m deep (19.6psia) @12:55:48.58 Cartridge 43 sampling 1432/2000ml at 0.35ml/s with 16.7psi vacuum @12:56:48.74 Cartridge 43 sampling 1454/2000ml at 0.38ml/s with 24.5psi vacuum @12:57:48.58 Cartridge 43 sampling 1475/2000ml at 0.34ml/s with 30.2psi vacuum @12:58:48.73 Cartridge 43 sampling 1497/2000ml at 0.34ml/s with 15.2psi vacuum 5.6m deep (23.0psia) @12:59:48.58 Cartridge 43 sampling 1518/2000ml at 0.37ml/s with 28.0psi vacuum @13:00:25.64 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:00:25.66 Retry #7 of 10 after 10:00 @13:00:48.72 Cartridge 43 sampling 1539/2000ml at 0.35ml/s with 17.6psi vacuum 1.9m deep (17.5psia) @13:01:48.70 Cartridge 43 sampling 1560/2000ml at 0.34ml/s with 29.7psi vacuum @13:02:48.69 Cartridge 43 sampling 1581/2000ml at 0.33ml/s with 16.7psi vacuum @13:03:48.58 Cartridge 43 sampling 1601/2000ml at 0.32ml/s with 30.4psi vacuum @13:04:48.58 Cartridge 43 sampling 1621/2000ml at 0.36ml/s with 24.1psi vacuum @13:05:49.76 Cartridge 43 sampling 1641/2000ml at 0.31ml/s with 13.4psi vacuum 8.7m deep (27.5psia) @13:06:48.58 Cartridge 43 sampling 1660/2000ml at 0.30ml/s with 26.4psi vacuum @13:07:48.74 Cartridge 43 sampling 1680/2000ml at 0.29ml/s with 29.9psi vacuum @13:08:48.57 Cartridge 43 sampling 1699/2000ml at 0.33ml/s with 19.5psi vacuum @13:09:48.98 Cartridge 43 sampling 1718/2000ml at 0.31ml/s with 16.3psi vacuum 3.9m deep (20.5psia) @13:10:25.72 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:10:25.81 Retry #8 of 10 after 10:00 @13:10:48.57 Cartridge 43 sampling 1737/2000ml at 0.33ml/s with 18.2psi vacuum @13:11:48.82 Cartridge 43 sampling 1755/2000ml at 0.30ml/s with 15.4psi vacuum 7.2m deep (25.3psia) @13:12:48.57 Cartridge 43 sampling 1774/2000ml at 0.32ml/s with 24.1psi vacuum @13:13:48.71 Cartridge 43 sampling 1792/2000ml at 0.32ml/s with 21.1psi vacuum 1.8m deep (17.4psia) @13:14:48.57 Cartridge 43 sampling 1809/2000ml at 0.25ml/s with 29.8psi vacuum @13:15:48.76 Cartridge 43 sampling 1827/2000ml at 0.27ml/s with 29.7psi vacuum @13:16:48.57 Cartridge 43 sampling 1844/2000ml at 0.28ml/s with 30.3psi vacuum @13:17:48.74 Cartridge 43 sampling 1862/2000ml at 0.31ml/s with 24.1psi vacuum @13:18:48.57 Cartridge 43 sampling 1879/2000ml at 0.27ml/s with 27.8psi vacuum @13:19:49.01 Cartridge 43 sampling 1895/2000ml at 0.31ml/s with 23.4psi vacuum @13:20:25.89 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:20:25.91 Retry #9 of 10 after 10:00 @13:20:48.64 Cartridge 43 sampling 1912/2000ml at 0.22ml/s with 23.4psi vacuum @13:21:48.59 Cartridge 43 sampling 1928/2000ml at 0.24ml/s with 30.0psi vacuum @13:22:48.92 Cartridge 43 sampling 1945/2000ml at 0.29ml/s with 27.5psi vacuum @13:23:48.58 Cartridge 43 sampling 1961/2000ml at 0.22ml/s with 28.6psi vacuum @13:24:48.72 Cartridge 43 sampling 1976/2000ml at 0.27ml/s with 16.8psi vacuum 4.5m deep (21.3psia) @13:25:48.58 Cartridge 43 sampling 1992/2000ml at 0.23ml/s with 29.9psi vacuum @13:26:19.18 Pressure.coast @13:26:19.68 Intake.close @13:26:19.89 Exhaust.close @13:26:20.47 Toroid.reconfigure :hold @13:26:20.72 Toroid.seek :bypass @13:26:22.59 Toroid.reconfigure :default @13:26:22.72 Toroid.seek :clear @13:26:24.55 Sampled 2000.0ml @13:26:25.35 SamplePump.setPosition! 0ml @13:26:25.40 Plunge.seek :reagent @13:26:25.44 Bypass.open @13:26:26.21 Intake.open @13:26:26.34 Sample loop closed. Ready to Deliver Reagents @13:26:26.36 Archive.seek :sealed @13:26:27.07 SamplePump.reconfigure :slow @13:26:27.14 SamplePump.seek -3ml @13:26:30.08 Archive.seek :evac @13:26:33.81 delay 1 second @13:26:34.90 SamplePump.seek -2.8ml @13:26:35.60 delay 1 second @13:26:36.69 SamplePump.seek -3.3ml @13:26:38.02 delay 1 second @13:26:39.05 Exhaust.open @13:26:39.75 Bypass.close @13:26:40.73 SamplePump.seek -35.3ml @13:26:44.07 delay 1:00 @13:27:44.16 Archive.seek :clear @13:27:49.48 delay 3 seconds @13:27:52.73 Intake.close @13:27:52.88 Exhaust.close @13:28:01.32 RNAlater soak for 10:00 @13:30:25.97 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:30:25.99 Retry #10 of 10 after 10:00 @13:38:01.45 Archive.seek :tVent @13:38:03.16 delay 1 second @13:38:04.20 Archive.seek :evac @13:38:20.20 delay 10 seconds @13:38:30.28 Sample preserved and evacuated @13:38:30.51
Carousel.seek 43 @13:38:30.72 Plunge.seek :locked @13:38:31.70 Archive.seek :clear @13:38:31.87 Air.seek :clear @13:38:32.03 Lysis1.seek :clear @13:38:48.89
Archive.coast @13:38:48.93 Air.coast @13:38:48.98 Lysis1.coast @13:38:49.02 Lysis2.coast @13:38:49.06 Plunge.coast @13:38:49.10 SamplePump.coast @13:38:49.15 Toroid.coast @13:38:49.19 Carousel.coast @13:38:49.38 Gate.power :motor,:OFF @13:38:49.39 Gate.power :core,:OFF @13:38:49.41 Gate.power :analytic1,:OFF @13:38:49.42 Gate.power :analytic2,:OFF @13:38:49.43 Gate.power :analytic3,:OFF @13:38:49.44 Gate.power :main,:OFF @13:38:49.46 Safely stopped and ready to power off @13:38:49.68 delayUntil "00:00:00.00PDT16-Jun-21" @13:40:26.04 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:40:26.07 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 43" @13:40:26.19 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:40:26.26 Retry #1 of 10 after 10:00 @13:50:26.31 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:50:26.33 Retry #2 of 10 after 10:00 @14:00:26.49 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:00:26.58 Retry #3 of 10 after 10:00 @14:10:26.73 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:10:26.79 Retry #4 of 10 after 10:00 @14:20:26.87 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:20:26.90 Retry #5 of 10 after 10:00 @14:30:26.94 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:30:27.04 Retry #6 of 10 after 10:00 @14:40:27.09 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:40:27.11 Retry #7 of 10 after 10:00 @14:50:27.21 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:50:27.27 Retry #8 of 10 after 10:00 @15:00:27.39 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:00:27.42 Retry #9 of 10 after 10:00 @15:10:27.46 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:10:27.48 Retry #10 of 10 after 10:00 @15:20:27.58 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:20:27.60 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT16-Jun-21" @15:20:27.63 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:20:27.65 Retry #1 of 10 after 10:00 @15:30:27.70 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:30:27.72 Retry #2 of 10 after 10:00 @15:40:27.77 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:40:27.87 Retry #3 of 10 after 10:00 @15:50:27.94 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:50:27.96 Retry #4 of 10 after 10:00 @16:00:28.21 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:00:28.27 Retry #5 of 10 after 10:00 @16:10:28.37 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:10:28.43 Retry #6 of 10 after 10:00 @16:20:28.52 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:20:28.54 Retry #7 of 10 after 10:00 @16:30:28.61 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:30:28.66 Retry #8 of 10 after 10:00 @16:40:28.71 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:40:28.73 Retry #9 of 10 after 10:00 @16:50:28.80 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:50:28.83 Retry #10 of 10 after 10:00 @17:00:28.87 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:00:00.12PDT16-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.13 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.15 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.16 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.20 Powering core bus @00:00:02.76 Archive.configure :default @00:00:03.12 Air.configure :default @00:00:03.44 Lysis1.configure :default @00:00:03.66 Lysis2.configure :default @00:00:03.94 Plunge.configure :default @00:00:04.27 SamplePump.configure :default @00:00:04.52 ExtraPump.configure :default @00:00:04.85 Toroid.configure :default @00:00:05.12 Carousel.configure :default @00:00:05.53 Pressure.configure :default @00:00:05.64 All elf controllers running firmware version 4.61 @00:00:05.66 Gate.power :motor,:ON,limit to 5000mA @00:00:06.85 Archive.reconfigure :default @00:00:06.93 Toroid.reconfigure :default @00:00:07.00 Plunge.reconfigure :default @00:00:07.08 Lysis1.reconfigure :default @00:00:07.13 Air.reconfigure :default @00:00:07.27 Lysis2.reconfigure :default @00:00:07.94 Toroid.home.jog 0 @00:00:08.00 Lysis1.home.jog 0 @00:00:08.03 Archive.home.jog 0 @00:00:08.05 Plunge.home.jog 0 @00:00:08.09 Lysis2.home.jog 0 @00:00:08.13 Air.home.jog 0 @00:00:08.85 Toroid.reconfigure :homing @00:00:08.90 Lysis2.reconfigure :homing @00:00:08.93 Lysis1.reconfigure :homing @00:00:08.97 Archive.reconfigure :homing @00:00:09.03 Plunge.reconfigure :homing @00:00:09.22 Lysis2.jog 1356 @00:00:09.28 Toroid.jog 635 @00:00:09.37 Plunge.jog 849 @00:00:09.40 Lysis1.jog 1352 @00:00:09.43 Archive.jog 1369 @00:00:09.66 Air.reconfigure :homing @00:00:09.72 Air.jog 1361 @00:00:10.49 Plunge.setRawPosition! 50 @00:00:10.51 Lysis2.setRawPosition! 50 @00:00:10.59 Plunge.seek :home @00:00:10.68 Lysis2.seek :home @00:00:11.41 Plunge.reconfigure :default @00:00:11.43 Lysis1.setRawPosition! 50 @00:00:11.45 Air.setRawPosition! 50 @00:00:11.48 Archive.setRawPosition! 50 @00:00:11.60 Lysis1.seek :home @00:00:11.72 Air.seek :home @00:00:11.75 Archive.seek :home @00:00:12.00 Toroid.setRawPosition! 0 @00:00:12.04 Lysis2.reconfigure :default @00:00:12.12 Toroid.seek :home @00:00:12.78 Toroid.reconfigure :default @00:00:12.84 Lysis1.reconfigure :default @00:00:12.90 Archive.reconfigure :default @00:00:13.02 Air.reconfigure :default @00:00:14.28
Toroid.seek :clear @00:00:15.97 Carousel.reconfigure :default @00:00:16.16 Carousel.home.jog 0 @00:00:16.85 Carousel.seek 43 @00:00:17.34 Selecting Cartridge 42 @00:00:17.93 Carousel.seek 42 @00:00:18.45 Lysis1.seek :locked @00:00:21.68 Plunge.seek :locked @00:00:21.82 Air.seek :locked @00:00:26.19
Mounted archive Cartridge @00:00:26.36 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 42" @00:00:26.52 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:00:26.58 Retry #1 of 10 after 10:00 @00:00:27.27 Duration of filtering limited to 4:05:00 @00:00:27.58 Priming sample loop w/160ml, bypass w/40ml @00:00:27.60 SamplePump.setPosition! 0ml @00:00:27.63 Bypass.open @00:00:29.08 Intake.open @00:00:29.96 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:32.64 Vacuum tare is 18% of full scale @00:00:32.73 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:33.32 Pressure.reconfigure :default @00:00:33.39 Pressure.hold @00:00:33.91 SamplePump.reconfigure :prime @00:00:34.10 SamplePump.seek 40ml,3:51 @00:01:17.31 Pressure.coast @00:01:17.36 SamplePump.setPosition! 0ml @00:01:17.41 Exhaust.open @00:01:18.89 Bypass.close @00:01:20.73 Pressure.reconfigure :default @00:01:20.83 Pressure.hold @00:01:21.57 SamplePump.reconfigure :prime @00:01:21.64 SamplePump.seek 160ml,13:51 @00:04:13.27 Pressure.coast @00:04:13.41 Bypass.open @00:04:14.24 Intake.close @00:04:15.36 Toroid.reconfigure :hold @00:04:15.53 Toroid.seek :crack @00:04:17.26 Toroid.seek :cartridge @00:04:18.25 Toroid.reconfigure :default @00:04:18.31 Toroid.seek :clear @00:04:20.15 Pressurizing cartridge to 26psi above 14.5psia @00:04:20.17 SamplePump.setPosition! 0ml @00:04:20.20 Intake.open @00:04:20.99 Bypass.close @00:04:21.67 Exhaust.close @00:04:22.83 Pressure.reconfigure :default @00:04:23.01 Pressure.hold @00:04:23.74 SamplePump.reconfigure :charge @00:04:23.93 SamplePump.seek 2.3ml,42 seconds @00:04:34.05 Pressure.coast @00:04:34.11 Intake.close @00:04:35.08 Pressurized cartridge with 0.453ml to 26.1psi @00:04:35.13 Isolated @35.2psia (20.7psi) @00:04:35.14 Checking for cartridge leaks from 35.2psia @00:04:35.20 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:44.39 Cartridge pressure fell 17% to 29.2psia @00:04:44.65 Sampling 2000.0ml @00:04:44.73 SamplePump.setPosition! 0ml @00:04:44.77 Exhaust.open @00:04:45.62 Intake.open @00:04:46.64 Pressure.reconfigure :default @00:04:46.70 Pressure.hold @00:04:47.54 SamplePump.reconfigure :sample @00:04:47.61 SamplePump.seek 2000ml,2:47:12 @00:04:49.74 Cartridge 42 sampling 1/2000ml at 0.51ml/s with 23.2psi vacuum @00:05:49.74 Cartridge 42 sampling 31/2000ml at 0.50ml/s with 26.0psi vacuum @00:06:49.70 Cartridge 42 sampling 62/2000ml at 0.51ml/s with 21.1psi vacuum @00:07:50.03 Cartridge 42 sampling 92/2000ml at 0.49ml/s with 22.8psi vacuum @00:08:49.94 Cartridge 42 sampling 122/2000ml at 0.51ml/s with 26.9psi vacuum @00:09:49.71 Cartridge 42 sampling 153/2000ml at 0.52ml/s with 23.1psi vacuum @00:10:26.64 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:10:26.66 Retry #2 of 10 after 10:00 @00:10:50.13 Cartridge 42 sampling 185/2000ml at 0.52ml/s with 22.7psi vacuum @00:11:49.71 Cartridge 42 sampling 216/2000ml at 0.54ml/s with 25.8psi vacuum @00:12:49.92 Cartridge 42 sampling 248/2000ml at 0.52ml/s with 19.3psi vacuum 2.9m deep (19.0psia) @00:13:49.71 Cartridge 42 sampling 279/2000ml at 0.53ml/s with 26.2psi vacuum @00:14:49.87 Cartridge 42 sampling 310/2000ml at 0.52ml/s with 26.1psi vacuum @00:15:49.71 Cartridge 42 sampling 341/2000ml at 0.52ml/s with 21.2psi vacuum @00:16:49.87 Cartridge 42 sampling 371/2000ml at 0.52ml/s with 25.9psi vacuum @00:17:49.72 Cartridge 42 sampling 402/2000ml at 0.51ml/s with 18.1psi vacuum 3.1m deep (19.3psia) @00:18:49.90 Cartridge 42 sampling 432/2000ml at 0.50ml/s with 17.9psi vacuum 2.7m deep (18.7psia) @00:19:49.71 Cartridge 42 sampling 462/2000ml at 0.49ml/s with 27.5psi vacuum @00:20:26.72 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:20:26.74 Retry #3 of 10 after 10:00 @00:20:49.86 Cartridge 42 sampling 492/2000ml at 0.48ml/s with 28.1psi vacuum @00:21:49.71 Cartridge 42 sampling 521/2000ml at 0.49ml/s with 22.9psi vacuum @00:22:49.98 Cartridge 42 sampling 551/2000ml at 0.48ml/s with 15.6psi vacuum 3.6m deep (20.1psia) @00:23:50.08 Cartridge 42 sampling 580/2000ml at 0.48ml/s with 26.8psi vacuum @00:24:49.70 Cartridge 42 sampling 609/2000ml at 0.46ml/s with 28.2psi vacuum @00:25:50.09 Cartridge 42 sampling 638/2000ml at 0.48ml/s with 17.0psi vacuum 3.4m deep (19.8psia) @00:26:49.70 Cartridge 42 sampling 667/2000ml at 0.48ml/s with 24.5psi vacuum @00:27:49.85 Cartridge 42 sampling 696/2000ml at 0.49ml/s with 23.2psi vacuum @00:28:49.70 Cartridge 42 sampling 724/2000ml at 0.46ml/s with 18.7psi vacuum @00:29:49.85 Cartridge 42 sampling 752/2000ml at 0.46ml/s with 17.3psi vacuum 1.6m deep (17.1psia) @00:30:26.94 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:30:26.97 Retry #4 of 10 after 10:00 @00:30:49.70 Cartridge 42 sampling 780/2000ml at 0.46ml/s with 22.8psi vacuum @00:31:49.87 Cartridge 42 sampling 808/2000ml at 0.46ml/s with 24.4psi vacuum @00:32:49.70 Cartridge 42 sampling 835/2000ml at 0.45ml/s with 26.5psi vacuum @00:33:49.96 Cartridge 42 sampling 863/2000ml at 0.43ml/s with 25.5psi vacuum @00:34:49.70 Cartridge 42 sampling 890/2000ml at 0.44ml/s with 17.1psi vacuum 2.9m deep (19.0psia) @00:35:49.80 Cartridge 42 sampling 917/2000ml at 0.44ml/s with 19.4psi vacuum 2.4m deep (18.3psia) @00:36:49.70 Cartridge 42 sampling 943/2000ml at 0.44ml/s with 20.5psi vacuum @00:37:49.94 Cartridge 42 sampling 970/2000ml at 0.45ml/s with 21.4psi vacuum @00:38:50.30 Cartridge 42 sampling 996/2000ml at 0.45ml/s with 25.3psi vacuum @00:39:49.71 Cartridge 42 sampling 1022/2000ml at 0.43ml/s with 24.5psi vacuum @00:40:27.02 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:40:27.04 Retry #5 of 10 after 10:00 @00:40:49.84 Cartridge 42 sampling 1047/2000ml at 0.44ml/s with 23.5psi vacuum @00:41:49.72 Cartridge 42 sampling 1073/2000ml at 0.41ml/s with 19.3psi vacuum @00:42:49.86 Cartridge 42 sampling 1099/2000ml at 0.42ml/s with 20.4psi vacuum @00:43:49.71 Cartridge 42 sampling 1124/2000ml at 0.42ml/s with 22.3psi vacuum @00:44:49.89 Cartridge 42 sampling 1149/2000ml at 0.39ml/s with 28.3psi vacuum @00:45:49.71 Cartridge 42 sampling 1174/2000ml at 0.41ml/s with 27.6psi vacuum @00:46:49.86 Cartridge 42 sampling 1198/2000ml at 0.41ml/s with 25.3psi vacuum @00:47:49.71 Cartridge 42 sampling 1223/2000ml at 0.40ml/s with 19.1psi vacuum @00:48:49.86 Cartridge 42 sampling 1247/2000ml at 0.40ml/s with 27.2psi vacuum @00:49:49.71 Cartridge 42 sampling 1271/2000ml at 0.41ml/s with 23.8psi vacuum @00:50:27.52 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:50:27.57 Retry #6 of 10 after 10:00 @00:50:50.04 Cartridge 42 sampling 1295/2000ml at 0.40ml/s with 20.1psi vacuum @00:51:49.97 Cartridge 42 sampling 1319/2000ml at 0.39ml/s with 20.3psi vacuum 1.9m deep (17.5psia) @00:52:49.70 Cartridge 42 sampling 1342/2000ml at 0.38ml/s with 18.7psi vacuum @00:53:50.19 Cartridge 42 sampling 1366/2000ml at 0.39ml/s with 26.6psi vacuum @00:54:49.70 Cartridge 42 sampling 1388/2000ml at 0.37ml/s with 28.8psi vacuum @00:55:49.84 Cartridge 42 sampling 1411/2000ml at 0.37ml/s with 22.9psi vacuum @00:56:49.70 Cartridge 42 sampling 1433/2000ml at 0.38ml/s with 23.3psi vacuum @00:57:49.92 Cartridge 42 sampling 1456/2000ml at 0.33ml/s with 26.1psi vacuum @00:58:49.70 Cartridge 42 sampling 1477/2000ml at 0.36ml/s with 15.4psi vacuum 3.0m deep (19.1psia) @00:59:49.87 Cartridge 42 sampling 1499/2000ml at 0.34ml/s with 28.0psi vacuum @01:00:27.64 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:00:27.67 Retry #7 of 10 after 10:00 @01:00:49.72 Cartridge 42 sampling 1521/2000ml at 0.39ml/s with 25.1psi vacuum @01:01:49.99 Cartridge 42 sampling 1543/2000ml at 0.34ml/s with 26.6psi vacuum @01:02:49.72 Cartridge 42 sampling 1564/2000ml at 0.36ml/s with 23.9psi vacuum @01:03:49.80 Cartridge 42 sampling 1585/2000ml at 0.36ml/s with 25.9psi vacuum @01:04:49.72 Cartridge 42 sampling 1607/2000ml at 0.36ml/s with 22.2psi vacuum @01:05:49.72 Cartridge 42 sampling 1628/2000ml at 0.33ml/s with 27.9psi vacuum @01:06:49.89 Cartridge 42 sampling 1648/2000ml at 0.31ml/s with 28.2psi vacuum @01:07:49.71 Cartridge 42 sampling 1669/2000ml at 0.35ml/s with 20.8psi vacuum @01:08:50.02 Cartridge 42 sampling 1689/2000ml at 0.30ml/s with 27.1psi vacuum @01:09:49.71 Cartridge 42 sampling 1710/2000ml at 0.35ml/s with 19.3psi vacuum @01:10:27.72 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:10:27.74 Retry #8 of 10 after 10:00 @01:10:49.94 Cartridge 42 sampling 1730/2000ml at 0.33ml/s with 16.6psi vacuum 3.0m deep (19.1psia) @01:11:49.71 Cartridge 42 sampling 1750/2000ml at 0.35ml/s with 23.3psi vacuum @01:12:49.90 Cartridge 42 sampling 1770/2000ml at 0.32ml/s with 20.2psi vacuum 2.0m deep (17.6psia) @01:13:49.71 Cartridge 42 sampling 1789/2000ml at 0.32ml/s with 27.4psi vacuum @01:14:49.89 Cartridge 42 sampling 1808/2000ml at 0.34ml/s with 23.7psi vacuum @01:15:49.71 Cartridge 42 sampling 1828/2000ml at 0.30ml/s with 28.1psi vacuum @01:16:49.72 Cartridge 42 sampling 1847/2000ml at 0.31ml/s with 20.4psi vacuum @01:17:49.71 Cartridge 42 sampling 1866/2000ml at 0.29ml/s with 28.2psi vacuum @01:18:49.85 Cartridge 42 sampling 1885/2000ml at 0.33ml/s with 20.9psi vacuum @01:19:51.34 Cartridge 42 sampling 1903/2000ml at 0.32ml/s with 22.9psi vacuum @01:20:27.85 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:20:27.94 Retry #9 of 10 after 10:00 @01:20:49.95 Cartridge 42 sampling 1922/2000ml at 0.29ml/s with 28.9psi vacuum @01:21:49.71 Cartridge 42 sampling 1940/2000ml at 0.32ml/s with 24.2psi vacuum @01:22:49.72 Cartridge 42 sampling 1959/2000ml at 0.33ml/s with 24.7psi vacuum @01:23:50.07 Cartridge 42 sampling 1977/2000ml at 0.32ml/s with 22.9psi vacuum @01:24:49.71 Cartridge 42 sampling 1995/2000ml at 0.31ml/s with 26.9psi vacuum @01:25:06.33 Pressure.coast @01:25:06.55 Intake.close @01:25:06.65 Exhaust.close @01:25:07.29 Toroid.reconfigure :hold @01:25:07.35 Toroid.seek :bypass @01:25:08.97 Toroid.reconfigure :default @01:25:09.03 Toroid.seek :clear @01:25:10.65 Sampled 2000.0ml @01:25:11.46 SamplePump.setPosition! 0ml @01:25:11.50 Plunge.seek :reagent @01:25:11.67 Bypass.open @01:25:12.51 Intake.open @01:25:12.79 Sample loop closed. Ready to Deliver Reagents @01:25:12.87 Archive.seek :sealed @01:25:13.63 SamplePump.reconfigure :slow @01:25:13.85 SamplePump.seek -3ml @01:25:16.63 Archive.seek :evac @01:25:20.58 delay 1 second @01:25:21.85 SamplePump.seek -2.8ml @01:25:22.69 delay 1 second @01:25:23.85 SamplePump.seek -3.3ml @01:25:25.29 delay 1 second @01:25:26.32 Exhaust.open @01:25:27.02 Bypass.close @01:25:27.80 SamplePump.seek -35.3ml @01:25:30.66 delay 1:00 @01:26:30.70 Archive.seek :clear @01:26:36.60 delay 3 seconds @01:26:39.74 Intake.close @01:26:39.81 Exhaust.close @01:26:47.84 RNAlater soak for 10:00 @01:30:28.01 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:30:28.08 Retry #10 of 10 after 10:00 @01:36:47.88 Archive.seek :tVent @01:36:49.52 delay 1 second @01:36:50.57 Archive.seek :evac @01:37:06.56 delay 10 seconds @01:37:16.60 Sample preserved and evacuated @01:37:16.83
Carousel.seek 42 @01:37:17.04 Plunge.seek :locked @01:37:18.36 Archive.seek :clear @01:37:18.77 Air.seek :clear @01:37:19.30 Lysis1.seek :clear @01:37:35.73
Archive.coast @01:37:35.77 Air.coast @01:37:35.82 Lysis1.coast @01:37:35.86 Lysis2.coast @01:37:35.90 Plunge.coast @01:37:35.95 SamplePump.coast @01:37:35.99 Toroid.coast @01:37:36.04 Carousel.coast @01:37:36.23 Gate.power :motor,:OFF @01:37:36.24 Gate.power :core,:OFF @01:37:36.26 Gate.power :analytic1,:OFF @01:37:36.27 Gate.power :analytic2,:OFF @01:37:36.28 Gate.power :analytic3,:OFF @01:37:36.30 Gate.power :main,:OFF @01:37:36.31 Safely stopped and ready to power off @01:37:36.71 Cartridge Load: {53 => { Type: :archive_bac, State: :processed, filtered: 1396.8}, 52..48 => { Type: :archive_bac, State: :processed, filtered: 2000}, [59, 58] => { Type: :archive_bac, State: :processed, filtered: 400}, 41..1 => { Type: :archive_bac, State: :reserved, was: :dry}, 46 => { Type: :archive_bac, State: :leaky}, [47, 45..42] => { Type: :archive_bac, State: :processed, filtered: 2000.0}, 54 => { Type: :archive_bac, State: :processed, filtered: 721.3}, 55 => { Type: :archive_bac, State: :processed, filtered: 622.6}, 56 => { Type: :archive_bac, State: :processed, filtered: 387.7}, 57 => { Type: :archive_bac, State: :processed, filtered: 1000}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 0} @01:38:06.76 Sending Email @01:40:28.14 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:40:28.16 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 42" @01:40:28.19 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:40:28.21 Retry #1 of 10 after 10:00 @01:43:06.79 Email::Error in MAIN -- flush stuck! @11:13:18.30PDT16-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @11:13:18.42 >BREAK< ESP3 Gateway v1.11 for cartridge bus @11:13:23.57 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @11:13:23.88 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @11:13:32.76 Polling Can every 2:00 @11:13:32.82 Loading /home/esp/esp2/bin/processAll mission @11:13:34.01 Sampling default goal volumes @11:13:34.05 GoodNews.email "Cartridge group: dry\nSpares: {spare: 0}\nSchedule: [\"12pm\", \"12am\"]\n",Subject: "Begun processAll --schedule=12pm,12am --checkpoint --mail" @11:13:34.17
delayUntil "12:00:00.00" @11:13:48.13 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @11:13:57.06 upsync @11:14:13.22 upsync @12:00:00.33
Gate.power :main,:ON,limit to 5000mA @12:00:00.36 Gate.power :analytic1,:ON,limit to 5000mA @12:00:00.44 Gate.power :analytic2,:ON,limit to 5000mA @12:00:00.48 Gate.power :analytic3,:ON,limit to 5000mA @12:00:00.63 Powering core bus @12:00:02.96 Archive.configure :default @12:00:03.11 Air.configure :default @12:00:03.21 Lysis1.configure :default @12:00:03.33 Lysis2.configure :default @12:00:03.46 Plunge.configure :default @12:00:03.60 SamplePump.configure :default @12:00:03.75 ExtraPump.configure :default @12:00:03.89 Toroid.configure :default @12:00:04.03 Carousel.configure :default @12:00:04.16 Pressure.configure :default @12:00:04.22 All elf controllers running firmware version 4.61 @12:00:04.24 Gate.power :motor,:ON,limit to 5000mA @12:00:04.53 Archive.reconfigure :default @12:00:04.81 Archive.home.jog 0 @12:00:04.90 Air.reconfigure :default @12:00:04.92 Toroid.reconfigure :default @12:00:04.96 Plunge.reconfigure :default @12:00:05.00 Lysis2.reconfigure :default @12:00:05.05 Lysis1.reconfigure :default @12:00:05.30 Plunge.home.jog 0 @12:00:05.34 Toroid.home.jog 0 @12:00:05.40 Air.home.jog 0 @12:00:05.43 Lysis2.home.jog 0 @12:00:05.47 Lysis1.home.jog 0 @12:00:06.18 Plunge.reconfigure :homing @12:00:06.24 Lysis2.reconfigure :homing @12:00:06.27 Air.reconfigure :homing @12:00:06.32 Archive.reconfigure :homing @12:00:06.36 Lysis1.reconfigure :homing @12:00:06.40 Toroid.reconfigure :homing @12:00:06.64 Lysis2.jog 1358 @12:00:06.72 Air.jog 1364 @12:00:06.75 Toroid.jog 627 @12:00:06.78 Lysis1.jog 1350 @12:00:06.80 Plunge.jog 851 @12:00:06.83 Archive.jog 151 @12:00:08.12 Archive.setRawPosition! 50 @12:00:08.35 Archive.seek :home @12:00:08.75 Plunge.setRawPosition! 50 @12:00:08.85 Lysis2.setRawPosition! 50 @12:00:08.87 Lysis1.setRawPosition! 50 @12:00:08.89 Air.setRawPosition! 50 @12:00:08.93 Archive.reconfigure :default @12:00:09.55 Lysis2.seek :home @12:00:09.67 Plunge.seek :home @12:00:09.70 Lysis1.seek :home @12:00:09.75 Air.seek :home @12:00:10.64 Toroid.setRawPosition! 0 @12:00:10.75 Plunge.reconfigure :default @12:00:10.78 Lysis2.reconfigure :default @12:00:11.02 Air.reconfigure :default @12:00:11.25 Toroid.seek :home @12:00:11.64 Lysis1.reconfigure :default @12:00:11.67 Toroid.reconfigure :default @12:00:12.37
Toroid.seek :clear @12:00:14.22 Carousel.reconfigure :default @12:00:14.27 Carousel.home.jog 0 @12:00:14.77 Carousel.seek 42 @12:00:14.94 Selecting Cartridge 41 @12:00:15.15 Carousel.seek 41 @12:00:15.68 Lysis1.seek :locked @12:00:19.04 Plunge.seek :locked @12:00:19.53 Air.seek :locked @12:00:23.80
Mounted archive Cartridge @12:00:23.92 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 41" @12:00:24.23 Duration of filtering limited to 4:05:00 @12:00:24.53 Priming sample loop w/160ml, bypass w/40ml @12:00:24.55 SamplePump.setPosition! 0ml @12:00:24.58 Bypass.open @12:00:26.06 Intake.open @12:00:27.51 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:00:30.28 Vacuum tare is 18% of full scale @12:00:30.39 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:00:30.80 Pressure.reconfigure :default @12:00:30.91 Pressure.hold @12:00:31.48 SamplePump.reconfigure :prime @12:00:31.73 SamplePump.seek 40ml,3:51 @12:00:44.08 upsync @12:01:14.95 Pressure.coast @12:01:15.17 SamplePump.setPosition! 0ml @12:01:15.25 Exhaust.open @12:01:16.75 Bypass.close @12:01:18.51 Pressure.reconfigure :default @12:01:18.57 Pressure.hold @12:01:18.86 SamplePump.reconfigure :prime @12:01:18.92 SamplePump.seek 160ml,13:51 @12:04:10.63 Pressure.coast @12:04:10.84 Bypass.open @12:04:11.58 Intake.close @12:04:12.53 Toroid.reconfigure :hold @12:04:12.58 Toroid.seek :crack @12:04:14.16 Toroid.seek :cartridge @12:04:15.09 Toroid.reconfigure :default @12:04:15.15 Toroid.seek :clear @12:04:17.55 Pressurizing cartridge to 26psi above 14.6psia @12:04:17.59 SamplePump.setPosition! 0ml @12:04:17.73 Intake.open @12:04:18.57 Bypass.close @12:04:19.32 Exhaust.close @12:04:20.33 Pressure.reconfigure :default @12:04:20.51 Pressure.hold @12:04:20.98 SamplePump.reconfigure :charge @12:04:21.13 SamplePump.seek 2.3ml,42 seconds @12:04:34.51 Pressure.coast @12:04:34.56 Intake.close @12:04:35.44 Pressurized cartridge with 0.821ml to 26.1psi @12:04:35.45 Isolated @36.0psia (21.5psi) @12:04:35.46 Checking for cartridge leaks from 36.0psia @12:04:35.48 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:04:46.18 Cartridge pressure fell 17% to 30.1psia @12:04:46.38 Sampling 2000.0ml @12:04:46.42 SamplePump.setPosition! 0ml @12:04:46.54 Exhaust.open @12:04:47.40 Intake.open @12:04:48.72 Pressure.reconfigure :default @12:04:48.90 Pressure.hold @12:04:49.54 SamplePump.reconfigure :sample @12:04:49.70 SamplePump.seek 2000ml,2:47:12 @12:04:51.63 Cartridge 41 sampling 1/2000ml at 0.59ml/s with 26.9psi vacuum @12:05:51.91 Cartridge 41 sampling 30/2000ml at 0.49ml/s with 24.8psi vacuum @12:06:51.66 Cartridge 41 sampling 60/2000ml at 0.51ml/s with 25.4psi vacuum @12:07:51.76 Cartridge 41 sampling 90/2000ml at 0.49ml/s with 27.9psi vacuum @12:08:51.63 Cartridge 41 sampling 122/2000ml at 0.54ml/s with 24.6psi vacuum @12:09:51.80 Cartridge 41 sampling 154/2000ml at 0.53ml/s with 27.9psi vacuum @12:10:51.63 Cartridge 41 sampling 186/2000ml at 0.53ml/s with 28.0psi vacuum @12:11:51.80 Cartridge 41 sampling 219/2000ml at 0.55ml/s with 26.1psi vacuum @12:12:51.63 Cartridge 41 sampling 251/2000ml at 0.54ml/s with 19.3psi vacuum 2.3m deep (18.1psia) @12:13:51.77 Cartridge 41 sampling 283/2000ml at 0.56ml/s with 25.1psi vacuum @12:14:51.63 Cartridge 41 sampling 315/2000ml at 0.53ml/s with 16.5psi vacuum 1.9m deep (17.5psia) @12:15:51.68 Cartridge 41 sampling 347/2000ml at 0.54ml/s with 21.8psi vacuum 1.9m deep (17.5psia) @12:16:51.63 Cartridge 41 sampling 379/2000ml at 0.52ml/s with 28.6psi vacuum @12:17:51.63 Cartridge 41 sampling 410/2000ml at 0.51ml/s with 27.7psi vacuum @12:18:51.91 Cartridge 41 sampling 441/2000ml at 0.50ml/s with 24.5psi vacuum @12:19:51.63 Cartridge 41 sampling 471/2000ml at 0.52ml/s with 24.8psi vacuum @12:20:51.81 Cartridge 41 sampling 502/2000ml at 0.52ml/s with 26.1psi vacuum @12:21:51.63 Cartridge 41 sampling 532/2000ml at 0.52ml/s with 26.7psi vacuum @12:22:51.77 Cartridge 41 sampling 562/2000ml at 0.47ml/s with 24.7psi vacuum @12:23:51.63 Cartridge 41 sampling 592/2000ml at 0.48ml/s with 29.1psi vacuum @12:24:51.71 Cartridge 41 sampling 621/2000ml at 0.49ml/s with 20.2psi vacuum 2.7m deep (18.7psia) @12:25:51.63 Cartridge 41 sampling 650/2000ml at 0.51ml/s with 23.9psi vacuum @12:26:51.75 Cartridge 41 sampling 680/2000ml at 0.48ml/s with 21.3psi vacuum @12:27:51.63 Cartridge 41 sampling 709/2000ml at 0.49ml/s with 18.8psi vacuum 2.7m deep (18.6psia) @12:28:51.69 Cartridge 41 sampling 738/2000ml at 0.46ml/s with 14.9psi vacuum 3.7m deep (20.1psia) @12:29:51.63 Cartridge 41 sampling 766/2000ml at 0.47ml/s with 28.7psi vacuum @12:30:51.93 Cartridge 41 sampling 795/2000ml at 0.45ml/s with 16.8psi vacuum @12:31:51.81 Cartridge 41 sampling 823/2000ml at 0.48ml/s with 26.8psi vacuum @12:32:51.63 Cartridge 41 sampling 851/2000ml at 0.48ml/s with 22.7psi vacuum @12:33:51.87 Cartridge 41 sampling 879/2000ml at 0.46ml/s with 30.3psi vacuum @12:34:51.63 Cartridge 41 sampling 906/2000ml at 0.47ml/s with 20.9psi vacuum 2.0m deep (17.6psia) @12:35:51.79 Cartridge 41 sampling 934/2000ml at 0.46ml/s with 21.8psi vacuum @12:36:51.63 Cartridge 41 sampling 961/2000ml at 0.44ml/s with 30.6psi vacuum @12:37:51.75 Cartridge 41 sampling 988/2000ml at 0.44ml/s with 30.3psi vacuum @12:38:51.63 Cartridge 41 sampling 1015/2000ml at 0.45ml/s with 27.6psi vacuum @12:39:51.77 Cartridge 41 sampling 1042/2000ml at 0.45ml/s with 27.5psi vacuum @12:40:51.63 Cartridge 41 sampling 1068/2000ml at 0.44ml/s with 12.8psi vacuum 4.3m deep (21.0psia) @12:41:51.65 Cartridge 41 sampling 1095/2000ml at 0.45ml/s with 26.3psi vacuum @12:42:51.63 Cartridge 41 sampling 1121/2000ml at 0.44ml/s with 21.2psi vacuum 2.2m deep (17.9psia) @12:43:51.80 Cartridge 41 sampling 1147/2000ml at 0.44ml/s with 14.4psi vacuum 6.5m deep (24.2psia) @12:44:51.81 Cartridge 41 sampling 1172/2000ml at 0.42ml/s with 13.1psi vacuum 7.7m deep (26.1psia) @12:45:51.75 Cartridge 41 sampling 1197/2000ml at 0.43ml/s with 20.9psi vacuum @12:46:51.88 Cartridge 41 sampling 1223/2000ml at 0.42ml/s with 13.8psi vacuum 5.1m deep (22.3psia) @12:47:51.63 Cartridge 41 sampling 1248/2000ml at 0.41ml/s with 28.9psi vacuum @12:48:51.77 Cartridge 41 sampling 1273/2000ml at 0.43ml/s with 18.1psi vacuum 4.1m deep (20.8psia) @12:49:51.63 Cartridge 41 sampling 1297/2000ml at 0.41ml/s with 16.4psi vacuum 3.2m deep (19.4psia) @12:50:51.87 Cartridge 41 sampling 1322/2000ml at 0.40ml/s with 14.9psi vacuum 6.1m deep (23.7psia) @12:51:51.63 Cartridge 41 sampling 1346/2000ml at 0.41ml/s with 28.7psi vacuum @12:52:51.98 Cartridge 41 sampling 1370/2000ml at 0.35ml/s with 25.5psi vacuum @12:53:51.63 Cartridge 41 sampling 1394/2000ml at 0.37ml/s with 18.5psi vacuum @12:54:51.74 Cartridge 41 sampling 1418/2000ml at 0.40ml/s with 28.5psi vacuum @12:55:51.63 Cartridge 41 sampling 1441/2000ml at 0.35ml/s with 24.0psi vacuum @12:56:51.79 Cartridge 41 sampling 1465/2000ml at 0.37ml/s with 16.0psi vacuum 2.0m deep (17.6psia) @12:57:51.62 Cartridge 41 sampling 1488/2000ml at 0.37ml/s with 25.9psi vacuum @12:58:51.80 Cartridge 41 sampling 1512/2000ml at 0.41ml/s with 21.5psi vacuum @12:59:52.23 Cartridge 41 sampling 1535/2000ml at 0.36ml/s with 30.3psi vacuum @13:00:51.63 Cartridge 41 sampling 1558/2000ml at 0.39ml/s with 26.6psi vacuum @13:01:51.87 Cartridge 41 sampling 1580/2000ml at 0.37ml/s with 27.2psi vacuum @13:02:51.63 Cartridge 41 sampling 1603/2000ml at 0.37ml/s with 29.4psi vacuum @13:03:51.79 Cartridge 41 sampling 1625/2000ml at 0.35ml/s with 11.8psi vacuum 8.6m deep (27.4psia) @13:04:51.63 Cartridge 41 sampling 1647/2000ml at 0.36ml/s with 26.3psi vacuum @13:05:53.20 Cartridge 41 sampling 1669/2000ml at 0.33ml/s with 29.7psi vacuum @13:06:51.63 Cartridge 41 sampling 1691/2000ml at 0.39ml/s with 26.4psi vacuum @13:07:51.80 Cartridge 41 sampling 1713/2000ml at 0.34ml/s with 27.5psi vacuum @13:08:51.64 Cartridge 41 sampling 1734/2000ml at 0.34ml/s with 26.2psi vacuum @13:09:51.78 Cartridge 41 sampling 1756/2000ml at 0.32ml/s with 20.1psi vacuum @13:10:51.63 Cartridge 41 sampling 1777/2000ml at 0.34ml/s with 29.4psi vacuum @13:11:51.71 Cartridge 41 sampling 1798/2000ml at 0.37ml/s with 25.4psi vacuum @13:12:51.63 Cartridge 41 sampling 1819/2000ml at 0.32ml/s with 30.1psi vacuum @13:13:51.70 Cartridge 41 sampling 1839/2000ml at 0.34ml/s with 14.7psi vacuum 4.4m deep (21.3psia) @13:14:51.81 Cartridge 41 sampling 1860/2000ml at 0.34ml/s with 29.7psi vacuum @13:15:51.64 Cartridge 41 sampling 1881/2000ml at 0.32ml/s with 12.3psi vacuum 1.6m deep (17.0psia) @13:16:51.91 Cartridge 41 sampling 1901/2000ml at 0.32ml/s with 13.3psi vacuum 7.7m deep (26.1psia) @13:17:51.63 Cartridge 41 sampling 1921/2000ml at 0.36ml/s with 25.2psi vacuum @13:18:51.87 Cartridge 41 sampling 1941/2000ml at 0.36ml/s with 23.5psi vacuum @13:19:51.63 Cartridge 41 sampling 1961/2000ml at 0.30ml/s with 30.3psi vacuum @13:20:51.80 Cartridge 41 sampling 1980/2000ml at 0.32ml/s with 29.5psi vacuum @13:21:51.63 Cartridge 41 sampling 2000/2000ml at 0.32ml/s with 29.7psi vacuum @13:21:51.91 Pressure.coast @13:21:52.05 Intake.close @13:21:52.10 Exhaust.close @13:21:52.29 Toroid.reconfigure :hold @13:21:52.35 Toroid.seek :bypass @13:21:54.04 Toroid.reconfigure :default @13:21:54.23 Toroid.seek :clear @13:21:55.88 Sampled 2000.0ml @13:21:56.96 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 41" @13:21:57.88 SamplePump.setPosition! 0ml @13:21:57.92 Plunge.seek :reagent @13:21:58.02 Bypass.open @13:21:58.83 Intake.open @13:21:58.98 Sample loop closed. Ready to Deliver Reagents @13:21:59.00 Archive.seek :sealed @13:21:59.77 SamplePump.reconfigure :slow @13:21:59.83 SamplePump.seek -3ml @13:22:02.64 Archive.seek :evac @13:22:06.56 delay 1 second @13:22:07.88 SamplePump.seek -2.8ml @13:22:08.61 delay 1 second @13:22:09.70 SamplePump.seek -3.3ml @13:22:11.03 delay 1 second @13:22:12.06 Exhaust.open @13:22:12.84 Bypass.close @13:22:13.95 SamplePump.seek -35.3ml @13:22:15.70 upsync @13:22:16.67 delay 1:00 @13:23:16.75 Archive.seek :clear @13:23:22.71 delay 3 seconds @13:23:25.80 Intake.close @13:23:25.85 Exhaust.close @13:23:33.88 RNAlater soak for 10:00 @13:33:33.92 Archive.seek :tVent @13:33:35.66 delay 1 second @13:33:36.81 Archive.seek :evac @13:33:52.88 delay 10 seconds @13:34:02.92 Sample preserved and evacuated @13:34:03.40
Carousel.seek 41 @13:34:03.98 Plunge.seek :locked @13:34:04.96 Archive.seek :clear @13:34:05.12 Air.seek :clear @13:34:05.28 Lysis1.seek :clear @13:34:22.15
Archive.coast @13:34:22.19 Air.coast @13:34:22.24 Lysis1.coast @13:34:22.28 Lysis2.coast @13:34:22.32 Plunge.coast @13:34:22.37 SamplePump.coast @13:34:22.41 Toroid.coast @13:34:22.46 Carousel.coast @13:34:22.65 Gate.power :motor,:OFF @13:34:22.66 Gate.power :core,:OFF @13:34:22.68 Gate.power :analytic1,:OFF @13:34:22.69 Gate.power :analytic2,:OFF @13:34:22.70 Gate.power :analytic3,:OFF @13:34:22.71 Gate.power :main,:OFF @13:34:22.73 Safely stopped and ready to power off @13:34:22.95 delayUntil "00:00:00.00PDT17-Jun-21" @13:34:22.98 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT17-Jun-21" @13:34:54.51 upsync @00:00:00.11PDT17-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.13 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.14 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.15 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.19 Powering core bus @00:00:02.72 Archive.configure :default @00:00:03.02 Air.configure :default @00:00:03.30 Lysis1.configure :default @00:00:03.57 Lysis2.configure :default @00:00:03.87 Plunge.configure :default @00:00:04.19 SamplePump.configure :default @00:00:04.51 ExtraPump.configure :default @00:00:04.82 Toroid.configure :default @00:00:05.14 Carousel.configure :default @00:00:05.51 Pressure.configure :default @00:00:05.56 All elf controllers running firmware version 4.61 @00:00:05.58 Gate.power :motor,:ON,limit to 5000mA @00:00:06.05 Archive.reconfigure :default @00:00:06.97 Archive.home.jog 0 @00:00:07.13 Lysis1.reconfigure :default @00:00:07.21 Air.reconfigure :default @00:00:07.24 Plunge.reconfigure :default @00:00:07.28 Lysis2.reconfigure :default @00:00:07.35 Toroid.reconfigure :default @00:00:07.54 Air.home.jog 0 @00:00:07.63 Lysis1.home.jog 0 @00:00:07.65 Plunge.home.jog 0 @00:00:07.67 Lysis2.home.jog 0 @00:00:07.70 Toroid.home.jog 0 @00:00:08.21 Archive.reconfigure :homing @00:00:08.42 Archive.jog 1372 @00:00:08.47 Toroid.reconfigure :homing @00:00:08.50 Plunge.reconfigure :homing @00:00:08.54 Lysis1.reconfigure :homing @00:00:08.58 Air.reconfigure :homing @00:00:08.63 Lysis2.reconfigure :homing @00:00:08.87 Plunge.jog 850 @00:00:08.91 Toroid.jog 627 @00:00:08.93 Lysis1.jog 1352 @00:00:08.96 Lysis2.jog 1356 @00:00:08.99 Air.jog 1359 @00:00:09.82 Archive.setRawPosition! 50 @00:00:09.87 Archive.seek :home @00:00:09.97 Plunge.setRawPosition! 50 @00:00:10.01 Plunge.seek :home @00:00:10.18 Archive.reconfigure :default @00:00:10.33 Plunge.reconfigure :default @00:00:10.36 Lysis2.setRawPosition! 50 @00:00:10.40 Lysis1.setRawPosition! 50 @00:00:10.41 Air.setRawPosition! 50 @00:00:10.55 Lysis2.seek :home @00:00:10.61 Lysis1.seek :home @00:00:10.67 Air.seek :home @00:00:10.93 Lysis2.reconfigure :default @00:00:11.05 Lysis1.reconfigure :default @00:00:11.13 Air.reconfigure :default @00:00:11.23 Toroid.setRawPosition! 0 @00:00:11.37 Toroid.seek :home @00:00:11.62 Toroid.reconfigure :default @00:00:12.14
Toroid.seek :clear @00:00:13.91 Carousel.reconfigure :default @00:00:13.98 Carousel.home.jog 0 @00:00:14.69 Carousel.seek 41 @00:00:14.99 Selecting Cartridge 40 @00:00:15.50 Carousel.seek 40 @00:00:16.20 Lysis1.seek :locked @00:00:19.42 Plunge.seek :locked @00:00:19.56 Air.seek :locked @00:00:23.87
Mounted archive Cartridge @00:00:24.08 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 40" @00:00:24.70 Duration of filtering limited to 4:05:00 @00:00:25.53 Priming sample loop w/160ml, bypass w/40ml @00:00:25.55 SamplePump.setPosition! 0ml @00:00:25.58 Bypass.open @00:00:27.09 Intake.open @00:00:27.98 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:30.57 Vacuum tare is 18% of full scale @00:00:30.63 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:30.91 Pressure.reconfigure :default @00:00:31.18 Pressure.hold @00:00:31.92 SamplePump.reconfigure :prime @00:00:32.11 SamplePump.seek 40ml,3:51 @00:00:38.88 upsync @00:01:15.35 Pressure.coast @00:01:15.40 SamplePump.setPosition! 0ml @00:01:15.45 Exhaust.open @00:01:16.91 Bypass.close @00:01:18.84 Pressure.reconfigure :default @00:01:18.91 Pressure.hold @00:01:19.66 SamplePump.reconfigure :prime @00:01:19.79 SamplePump.seek 160ml,13:51 @00:04:11.41 Pressure.coast @00:04:11.55 Bypass.open @00:04:12.35 Intake.close @00:04:13.39 Toroid.reconfigure :hold @00:04:13.60 Toroid.seek :crack @00:04:15.46 Toroid.seek :cartridge @00:04:16.62 Toroid.reconfigure :default @00:04:16.81 Toroid.seek :clear @00:04:18.72 Pressurizing cartridge to 26psi above 14.5psia @00:04:18.74 SamplePump.setPosition! 0ml @00:04:18.77 Intake.open @00:04:19.55 Bypass.close @00:04:20.23 Exhaust.close @00:04:21.06 Pressure.reconfigure :default @00:04:21.12 Pressure.hold @00:04:21.36 SamplePump.reconfigure :charge @00:04:21.65 SamplePump.seek 2.3ml,42 seconds @00:04:33.98 Pressure.coast @00:04:34.06 Intake.close @00:04:35.07 Pressurized cartridge with 0.680ml to 26.2psi @00:04:35.09 Isolated @35.9psia (21.3psi) @00:04:35.11 Checking for cartridge leaks from 35.9psia @00:04:35.19 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:42.95 Cartridge pressure fell 12% to 31.7psia @00:04:43.08 Sampling 2000.0ml @00:04:43.12 SamplePump.setPosition! 0ml @00:04:43.30 Exhaust.open @00:04:43.99 Intake.open @00:04:45.33 Pressure.reconfigure :default @00:04:45.39 Pressure.hold @00:04:45.66 SamplePump.reconfigure :sample @00:04:45.73 SamplePump.seek 2000ml,2:47:12 @00:04:48.31 Cartridge 40 sampling 1/2000ml at 0.48ml/s with 21.6psi vacuum @00:05:48.49 Cartridge 40 sampling 31/2000ml at 0.49ml/s with 21.2psi vacuum @00:06:48.32 Cartridge 40 sampling 60/2000ml at 0.49ml/s with 23.5psi vacuum @00:07:48.63 Cartridge 40 sampling 90/2000ml at 0.49ml/s with 24.0psi vacuum @00:08:48.47 Cartridge 40 sampling 119/2000ml at 0.49ml/s with 25.0psi vacuum @00:09:48.33 Cartridge 40 sampling 149/2000ml at 0.51ml/s with 23.3psi vacuum @00:10:48.52 Cartridge 40 sampling 179/2000ml at 0.52ml/s with 21.3psi vacuum @00:11:48.33 Cartridge 40 sampling 210/2000ml at 0.51ml/s with 25.8psi vacuum @00:12:48.44 Cartridge 40 sampling 241/2000ml at 0.51ml/s with 25.6psi vacuum @00:13:48.33 Cartridge 40 sampling 271/2000ml at 0.50ml/s with 26.4psi vacuum @00:14:48.48 Cartridge 40 sampling 301/2000ml at 0.51ml/s with 21.8psi vacuum @00:15:48.33 Cartridge 40 sampling 331/2000ml at 0.49ml/s with 27.5psi vacuum @00:16:48.46 Cartridge 40 sampling 361/2000ml at 0.50ml/s with 18.5psi vacuum 3.0m deep (19.1psia) @00:17:48.33 Cartridge 40 sampling 391/2000ml at 0.51ml/s with 23.3psi vacuum @00:18:48.49 Cartridge 40 sampling 420/2000ml at 0.48ml/s with 27.6psi vacuum @00:19:48.33 Cartridge 40 sampling 450/2000ml at 0.47ml/s with 17.5psi vacuum 2.1m deep (17.8psia) @00:20:48.43 Cartridge 40 sampling 479/2000ml at 0.50ml/s with 25.0psi vacuum @00:21:48.50 Cartridge 40 sampling 508/2000ml at 0.48ml/s with 22.5psi vacuum @00:22:48.61 Cartridge 40 sampling 538/2000ml at 0.49ml/s with 25.2psi vacuum @00:23:48.48 Cartridge 40 sampling 567/2000ml at 0.47ml/s with 27.7psi vacuum @00:24:48.32 Cartridge 40 sampling 596/2000ml at 0.48ml/s with 20.6psi vacuum @00:25:48.45 Cartridge 40 sampling 625/2000ml at 0.46ml/s with 19.0psi vacuum @00:26:48.32 Cartridge 40 sampling 653/2000ml at 0.47ml/s with 27.1psi vacuum @00:27:48.53 Cartridge 40 sampling 682/2000ml at 0.47ml/s with 26.9psi vacuum @00:28:48.32 Cartridge 40 sampling 710/2000ml at 0.46ml/s with 18.0psi vacuum 2.0m deep (17.7psia) @00:29:48.49 Cartridge 40 sampling 738/2000ml at 0.46ml/s with 26.1psi vacuum @00:30:48.32 Cartridge 40 sampling 766/2000ml at 0.46ml/s with 25.8psi vacuum @00:31:48.53 Cartridge 40 sampling 794/2000ml at 0.44ml/s with 26.5psi vacuum @00:32:48.33 Cartridge 40 sampling 822/2000ml at 0.47ml/s with 25.6psi vacuum @00:33:48.36 Cartridge 40 sampling 849/2000ml at 0.46ml/s with 25.8psi vacuum @00:34:48.33 Cartridge 40 sampling 876/2000ml at 0.45ml/s with 26.4psi vacuum @00:35:48.40 Cartridge 40 sampling 903/2000ml at 0.45ml/s with 20.2psi vacuum @00:36:48.65 Cartridge 40 sampling 930/2000ml at 0.45ml/s with 26.2psi vacuum @00:37:48.33 Cartridge 40 sampling 957/2000ml at 0.44ml/s with 27.2psi vacuum @00:38:48.73 Cartridge 40 sampling 983/2000ml at 0.43ml/s with 26.3psi vacuum @00:39:48.33 Cartridge 40 sampling 1010/2000ml at 0.44ml/s with 17.1psi vacuum 3.7m deep (20.2psia) @00:40:48.45 Cartridge 40 sampling 1036/2000ml at 0.43ml/s with 17.3psi vacuum 3.9m deep (20.5psia) @00:41:48.33 Cartridge 40 sampling 1062/2000ml at 0.44ml/s with 25.6psi vacuum @00:42:48.59 Cartridge 40 sampling 1087/2000ml at 0.43ml/s with 17.4psi vacuum 2.9m deep (19.0psia) @00:43:48.33 Cartridge 40 sampling 1113/2000ml at 0.44ml/s with 23.3psi vacuum @00:44:48.43 Cartridge 40 sampling 1139/2000ml at 0.43ml/s with 24.9psi vacuum @00:45:48.33 Cartridge 40 sampling 1164/2000ml at 0.42ml/s with 16.2psi vacuum 3.6m deep (20.1psia) @00:46:48.49 Cartridge 40 sampling 1189/2000ml at 0.42ml/s with 26.4psi vacuum @00:47:48.33 Cartridge 40 sampling 1214/2000ml at 0.43ml/s with 20.0psi vacuum @00:48:48.41 Cartridge 40 sampling 1239/2000ml at 0.42ml/s with 18.8psi vacuum 2.0m deep (17.7psia) @00:49:48.33 Cartridge 40 sampling 1264/2000ml at 0.39ml/s with 27.5psi vacuum @00:50:48.33 Cartridge 40 sampling 1289/2000ml at 0.42ml/s with 24.8psi vacuum @00:51:48.70 Cartridge 40 sampling 1313/2000ml at 0.41ml/s with 24.6psi vacuum @00:52:48.33 Cartridge 40 sampling 1337/2000ml at 0.40ml/s with 25.1psi vacuum @00:53:48.48 Cartridge 40 sampling 1361/2000ml at 0.42ml/s with 26.8psi vacuum @00:54:48.92 Cartridge 40 sampling 1385/2000ml at 0.38ml/s with 17.7psi vacuum 2.9m deep (18.9psia) @00:55:48.50 Cartridge 40 sampling 1409/2000ml at 0.39ml/s with 22.7psi vacuum @00:56:48.33 Cartridge 40 sampling 1433/2000ml at 0.37ml/s with 27.3psi vacuum @00:57:48.57 Cartridge 40 sampling 1456/2000ml at 0.39ml/s with 25.2psi vacuum @00:58:48.33 Cartridge 40 sampling 1479/2000ml at 0.36ml/s with 26.8psi vacuum @00:59:48.52 Cartridge 40 sampling 1502/2000ml at 0.41ml/s with 25.5psi vacuum @01:00:48.33 Cartridge 40 sampling 1525/2000ml at 0.38ml/s with 24.7psi vacuum @01:01:48.39 Cartridge 40 sampling 1548/2000ml at 0.36ml/s with 23.0psi vacuum @01:02:48.33 Cartridge 40 sampling 1570/2000ml at 0.37ml/s with 23.0psi vacuum @01:03:48.50 Cartridge 40 sampling 1593/2000ml at 0.38ml/s with 24.6psi vacuum @01:04:48.51 Cartridge 40 sampling 1615/2000ml at 0.38ml/s with 24.4psi vacuum @01:05:48.33 Cartridge 40 sampling 1638/2000ml at 0.37ml/s with 27.6psi vacuum @01:06:48.53 Cartridge 40 sampling 1660/2000ml at 0.34ml/s with 24.5psi vacuum @01:07:48.33 Cartridge 40 sampling 1682/2000ml at 0.36ml/s with 24.9psi vacuum @01:08:48.46 Cartridge 40 sampling 1703/2000ml at 0.34ml/s with 27.2psi vacuum @01:09:48.38 Cartridge 40 sampling 1725/2000ml at 0.36ml/s with 24.6psi vacuum @01:10:48.49 Cartridge 40 sampling 1747/2000ml at 0.38ml/s with 24.9psi vacuum @01:11:48.33 Cartridge 40 sampling 1768/2000ml at 0.37ml/s with 24.9psi vacuum @01:12:48.49 Cartridge 40 sampling 1789/2000ml at 0.36ml/s with 27.1psi vacuum @01:13:48.33 Cartridge 40 sampling 1811/2000ml at 0.35ml/s with 27.1psi vacuum @01:14:48.46 Cartridge 40 sampling 1832/2000ml at 0.36ml/s with 19.5psi vacuum 2.6m deep (18.5psia) @01:15:48.33 Cartridge 40 sampling 1852/2000ml at 0.35ml/s with 23.8psi vacuum @01:16:48.66 Cartridge 40 sampling 1873/2000ml at 0.35ml/s with 26.1psi vacuum @01:17:48.61 Cartridge 40 sampling 1893/2000ml at 0.34ml/s with 26.4psi vacuum @01:18:48.51 Cartridge 40 sampling 1914/2000ml at 0.35ml/s with 23.5psi vacuum @01:19:48.65 Cartridge 40 sampling 1934/2000ml at 0.35ml/s with 24.5psi vacuum @01:20:48.32 Cartridge 40 sampling 1955/2000ml at 0.34ml/s with 23.8psi vacuum @01:21:48.54 Cartridge 40 sampling 1975/2000ml at 0.33ml/s with 22.2psi vacuum @01:22:48.31 Cartridge 40 sampling 1995/2000ml at 0.34ml/s with 25.8psi vacuum @01:23:03.11 Pressure.coast @01:23:03.50 Intake.close @01:23:03.72 Exhaust.close @01:23:04.06 Toroid.reconfigure :hold @01:23:04.27 Toroid.seek :bypass @01:23:05.89 Toroid.reconfigure :default @01:23:05.95 Toroid.seek :clear @01:23:07.56 Sampled 2000.0ml @01:23:08.02 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 40" @01:23:08.42 SamplePump.setPosition! 0ml @01:23:08.47 Plunge.seek :reagent @01:23:08.51 Bypass.open @01:23:09.24 Intake.open @01:23:09.45 Sample loop closed. Ready to Deliver Reagents @01:23:09.47 Archive.seek :sealed @01:23:10.09 SamplePump.reconfigure :slow @01:23:10.15 SamplePump.seek -3ml @01:23:13.15 Archive.seek :evac @01:23:16.83 delay 1 second @01:23:17.92 SamplePump.seek -2.8ml @01:23:18.62 delay 1 second @01:23:19.98 SamplePump.seek -3.3ml @01:23:21.46 delay 1 second @01:23:22.48 Exhaust.open @01:23:23.29 Bypass.close @01:23:23.94 upsync @01:23:24.35 SamplePump.seek -35.3ml @01:23:27.15 delay 1:00 @01:24:27.18 Archive.seek :clear @01:24:33.17 delay 3 seconds @01:24:36.39 Intake.close @01:24:36.46 Exhaust.close @01:24:44.33 RNAlater soak for 10:00 @01:34:44.38 Archive.seek :tVent @01:34:46.02 delay 1 second @01:34:47.06 Archive.seek :evac @01:35:03.02 delay 10 seconds @01:35:13.06 Sample preserved and evacuated @01:35:13.29
Carousel.seek 40 @01:35:13.50 Plunge.seek :locked @01:35:14.48 Archive.seek :clear @01:35:14.64 Air.seek :clear @01:35:14.80 Lysis1.seek :clear @01:35:31.65
Archive.coast @01:35:31.69 Air.coast @01:35:31.74 Lysis1.coast @01:35:31.78 Lysis2.coast @01:35:31.82 Plunge.coast @01:35:31.87 SamplePump.coast @01:35:31.91 Toroid.coast @01:35:31.95 Carousel.coast @01:35:32.14 Gate.power :motor,:OFF @01:35:32.15 Gate.power :core,:OFF @01:35:32.17 Gate.power :analytic1,:OFF @01:35:32.18 Gate.power :analytic2,:OFF @01:35:32.19 Gate.power :analytic3,:OFF @01:35:32.21 Gate.power :main,:OFF @01:35:32.22 Safely stopped and ready to power off @01:35:32.46 delayUntil "12:00:00.00" @01:35:32.48 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @01:35:52.24 upsync @12:00:00.12
Gate.power :main,:ON,limit to 5000mA @12:00:00.14 Gate.power :analytic1,:ON,limit to 5000mA @12:00:00.15 Gate.power :analytic2,:ON,limit to 5000mA @12:00:00.16 Gate.power :analytic3,:ON,limit to 5000mA @12:00:00.20 Powering core bus @12:00:02.48 Archive.configure :default @12:00:02.61 Air.configure :default @12:00:02.73 Lysis1.configure :default @12:00:02.84 Lysis2.configure :default @12:00:03.29 Plunge.configure :default @12:00:03.98 SamplePump.configure :default @12:00:04.47 ExtraPump.configure :default @12:00:04.83 Toroid.configure :default @12:00:05.13 Carousel.configure :default @12:00:05.45 Pressure.configure :default @12:00:05.62 All elf controllers running firmware version 4.61 @12:00:05.66 Gate.power :motor,:ON,limit to 5000mA @12:00:06.28 Archive.reconfigure :default @12:00:07.18 Air.reconfigure :default @12:00:07.26 Archive.home.jog 0 @12:00:07.33 Toroid.reconfigure :default @12:00:07.37 Lysis2.reconfigure :default @12:00:07.41 Lysis1.reconfigure :default @12:00:07.61 Plunge.reconfigure :default @12:00:08.12 Toroid.home.jog 0 @12:00:08.28 Lysis2.home.jog 0 @12:00:08.33 Air.home.jog 0 @12:00:08.40 Lysis1.home.jog 0 @12:00:08.45 Plunge.home.jog 0 @12:00:09.47 Toroid.reconfigure :homing @12:00:09.50 Plunge.reconfigure :homing @12:00:09.53 Lysis2.reconfigure :homing @12:00:09.57 Air.reconfigure :homing @12:00:09.63 Archive.reconfigure :homing @12:00:09.66 Lysis1.reconfigure :homing @12:00:09.93 Lysis2.jog 1358 @12:00:09.97 Plunge.jog 851 @12:00:09.99 Toroid.jog 624 @12:00:10.02 Archive.jog 151 @12:00:10.05 Air.jog 1364 @12:00:10.07 Lysis1.jog 1348 @12:00:10.68 Archive.setRawPosition! 50 @12:00:10.73 Archive.seek :home @12:00:11.01 Archive.reconfigure :default @12:00:11.10 Plunge.setRawPosition! 50 @12:00:11.15 Plunge.seek :home @12:00:11.24 Lysis2.setRawPosition! 50 @12:00:11.29 Lysis2.seek :home @12:00:11.44 Air.setRawPosition! 50 @12:00:11.49 Air.seek :home @12:00:11.60 Plunge.reconfigure :default @12:00:11.63 Lysis2.reconfigure :default @12:00:11.66 Lysis1.setRawPosition! 50 @12:00:11.83 Lysis1.seek :home @12:00:11.91 Air.reconfigure :default @12:00:12.11 Lysis1.reconfigure :default @12:00:12.76 Toroid.setRawPosition! 0 @12:00:12.80 Toroid.seek :home @12:00:12.90 Toroid.reconfigure :default @12:00:13.19
Toroid.seek :clear @12:00:14.92 Carousel.reconfigure :default @12:00:15.05 Carousel.home.jog 0 @12:00:15.77 Carousel.seek 40 @12:00:16.10 Selecting Cartridge 39 @12:00:16.64 Carousel.seek 39 @12:00:17.39 Lysis1.seek :locked @12:00:20.64 Plunge.seek :locked @12:00:20.78 Air.seek :locked @12:00:25.18
Mounted archive Cartridge @12:00:25.41 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 39" @12:00:26.48 Duration of filtering limited to 4:05:00 @12:00:27.07 Priming sample loop w/160ml, bypass w/40ml @12:00:27.12 SamplePump.setPosition! 0ml @12:00:27.23 Bypass.open @12:00:28.76 Intake.open @12:00:29.62 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:00:32.21 Vacuum tare is 18% of full scale @12:00:32.23 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:00:32.47 Pressure.reconfigure :default @12:00:32.53 Pressure.hold @12:00:33.06 SamplePump.reconfigure :prime @12:00:33.22 SamplePump.seek 40ml,3:51 @12:00:49.26 upsync @12:01:16.54 Pressure.coast @12:01:16.59 SamplePump.setPosition! 0ml @12:01:16.64 Exhaust.open @12:01:18.07 Bypass.close @12:01:20.16 Pressure.reconfigure :default @12:01:20.35 Pressure.hold @12:01:21.09 SamplePump.reconfigure :prime @12:01:21.21 SamplePump.seek 160ml,13:51 @12:04:12.97 Pressure.coast @12:04:13.03 Bypass.open @12:04:13.79 Intake.close @12:04:14.87 Toroid.reconfigure :hold @12:04:15.05 Toroid.seek :crack @12:04:16.79 Toroid.seek :cartridge @12:04:17.78 Toroid.reconfigure :default @12:04:17.83 Toroid.seek :clear @12:04:19.71 Pressurizing cartridge to 26psi above 14.6psia @12:04:19.73 SamplePump.setPosition! 0ml @12:04:19.77 Intake.open @12:04:20.50 Bypass.close @12:04:21.19 Exhaust.close @12:04:22.01 Pressure.reconfigure :default @12:04:22.07 Pressure.hold @12:04:22.28 SamplePump.reconfigure :charge @12:04:22.35 SamplePump.seek 2.3ml,42 seconds @12:04:34.43 Pressure.coast @12:04:34.52 Intake.close @12:04:35.49 Pressurized cartridge with 0.658ml to 26.0psi @12:04:35.57 Isolated @34.9psia (20.3psi) @12:04:35.61 Checking for cartridge leaks from 34.9psia @12:04:35.65 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:04:42.90 Cartridge pressure fell 9% to 31.6psia @12:04:43.06 Sampling 2000.0ml @12:04:43.11 SamplePump.setPosition! 0ml @12:04:43.20 Exhaust.open @12:04:43.90 Intake.open @12:04:46.12 Pressure.reconfigure :default @12:04:46.19 Pressure.hold @12:04:46.99 SamplePump.reconfigure :sample @12:04:47.06 SamplePump.seek 2000ml,2:47:12 @12:04:49.13 Cartridge 39 sampling 1/2000ml at 0.49ml/s with 27.1psi vacuum @12:05:49.24 Cartridge 39 sampling 30/2000ml at 0.48ml/s with 24.7psi vacuum @12:06:49.14 Cartridge 39 sampling 59/2000ml at 0.49ml/s with 24.0psi vacuum @12:07:49.21 Cartridge 39 sampling 88/2000ml at 0.48ml/s with 25.8psi vacuum @12:08:49.14 Cartridge 39 sampling 118/2000ml at 0.48ml/s with 26.9psi vacuum @12:09:49.46 Cartridge 39 sampling 148/2000ml at 0.50ml/s with 18.2psi vacuum 2.5m deep (18.5psia) @12:10:49.13 Cartridge 39 sampling 179/2000ml at 0.51ml/s with 23.7psi vacuum @12:11:49.14 Cartridge 39 sampling 210/2000ml at 0.49ml/s with 22.0psi vacuum @12:12:49.55 Cartridge 39 sampling 241/2000ml at 0.50ml/s with 26.3psi vacuum @12:13:49.14 Cartridge 39 sampling 271/2000ml at 0.51ml/s with 24.9psi vacuum @12:14:49.30 Cartridge 39 sampling 302/2000ml at 0.50ml/s with 27.3psi vacuum @12:15:49.14 Cartridge 39 sampling 332/2000ml at 0.50ml/s with 17.4psi vacuum @12:16:49.27 Cartridge 39 sampling 363/2000ml at 0.50ml/s with 27.3psi vacuum @12:17:49.14 Cartridge 39 sampling 393/2000ml at 0.50ml/s with 22.5psi vacuum @12:18:49.26 Cartridge 39 sampling 423/2000ml at 0.50ml/s with 25.9psi vacuum @12:19:49.14 Cartridge 39 sampling 452/2000ml at 0.48ml/s with 27.4psi vacuum @12:20:49.29 Cartridge 39 sampling 482/2000ml at 0.48ml/s with 18.5psi vacuum 2.7m deep (18.7psia) @12:21:49.14 Cartridge 39 sampling 511/2000ml at 0.49ml/s with 23.4psi vacuum @12:22:49.43 Cartridge 39 sampling 540/2000ml at 0.48ml/s with 25.5psi vacuum @12:23:49.49 Cartridge 39 sampling 569/2000ml at 0.48ml/s with 27.5psi vacuum @12:24:49.17 Cartridge 39 sampling 598/2000ml at 0.49ml/s with 23.5psi vacuum @12:25:49.40 Cartridge 39 sampling 627/2000ml at 0.47ml/s with 21.9psi vacuum @12:26:49.13 Cartridge 39 sampling 655/2000ml at 0.48ml/s with 24.8psi vacuum @12:27:49.39 Cartridge 39 sampling 684/2000ml at 0.46ml/s with 15.9psi vacuum 3.6m deep (20.0psia) @12:28:49.13 Cartridge 39 sampling 712/2000ml at 0.44ml/s with 25.0psi vacuum @12:29:49.30 Cartridge 39 sampling 740/2000ml at 0.48ml/s with 25.1psi vacuum @12:30:49.13 Cartridge 39 sampling 768/2000ml at 0.45ml/s with 27.0psi vacuum @12:31:49.38 Cartridge 39 sampling 796/2000ml at 0.44ml/s with 24.5psi vacuum @12:32:49.13 Cartridge 39 sampling 823/2000ml at 0.46ml/s with 23.1psi vacuum @12:33:49.35 Cartridge 39 sampling 851/2000ml at 0.43ml/s with 22.2psi vacuum @12:34:49.13 Cartridge 39 sampling 878/2000ml at 0.44ml/s with 27.7psi vacuum @12:35:49.30 Cartridge 39 sampling 905/2000ml at 0.45ml/s with 27.6psi vacuum @12:36:49.13 Cartridge 39 sampling 932/2000ml at 0.45ml/s with 27.1psi vacuum @12:37:49.34 Cartridge 39 sampling 958/2000ml at 0.45ml/s with 24.4psi vacuum @12:38:49.97 Cartridge 39 sampling 985/2000ml at 0.44ml/s with 22.9psi vacuum @12:39:49.14 Cartridge 39 sampling 1011/2000ml at 0.44ml/s with 22.1psi vacuum @12:40:49.33 Cartridge 39 sampling 1037/2000ml at 0.45ml/s with 25.6psi vacuum @12:41:49.14 Cartridge 39 sampling 1063/2000ml at 0.44ml/s with 24.0psi vacuum @12:42:49.30 Cartridge 39 sampling 1088/2000ml at 0.42ml/s with 22.1psi vacuum @12:43:49.14 Cartridge 39 sampling 1114/2000ml at 0.42ml/s with 23.6psi vacuum @12:44:49.26 Cartridge 39 sampling 1139/2000ml at 0.43ml/s with 23.8psi vacuum @12:45:49.14 Cartridge 39 sampling 1164/2000ml at 0.42ml/s with 17.0psi vacuum 2.7m deep (18.7psia) @12:46:49.30 Cartridge 39 sampling 1190/2000ml at 0.40ml/s with 25.7psi vacuum @12:47:49.14 Cartridge 39 sampling 1215/2000ml at 0.40ml/s with 28.1psi vacuum @12:48:49.30 Cartridge 39 sampling 1239/2000ml at 0.42ml/s with 23.7psi vacuum @12:49:49.14 Cartridge 39 sampling 1264/2000ml at 0.39ml/s with 27.8psi vacuum @12:50:49.34 Cartridge 39 sampling 1288/2000ml at 0.40ml/s with 15.9psi vacuum 4.2m deep (21.0psia) @12:51:49.41 Cartridge 39 sampling 1312/2000ml at 0.42ml/s with 24.9psi vacuum @12:52:49.36 Cartridge 39 sampling 1336/2000ml at 0.39ml/s with 16.1psi vacuum 5.7m deep (23.2psia) @12:53:49.55 Cartridge 39 sampling 1360/2000ml at 0.40ml/s with 20.8psi vacuum @12:54:49.14 Cartridge 39 sampling 1384/2000ml at 0.38ml/s with 21.5psi vacuum @12:55:49.31 Cartridge 39 sampling 1407/2000ml at 0.40ml/s with 25.8psi vacuum @12:56:49.14 Cartridge 39 sampling 1430/2000ml at 0.37ml/s with 28.2psi vacuum @12:57:49.30 Cartridge 39 sampling 1454/2000ml at 0.36ml/s with 22.2psi vacuum @12:58:49.15 Cartridge 39 sampling 1477/2000ml at 0.37ml/s with 18.4psi vacuum @12:59:49.30 Cartridge 39 sampling 1499/2000ml at 0.38ml/s with 28.9psi vacuum @13:00:49.14 Cartridge 39 sampling 1522/2000ml at 0.40ml/s with 25.4psi vacuum @13:01:49.30 Cartridge 39 sampling 1545/2000ml at 0.38ml/s with 23.7psi vacuum @13:02:49.14 Cartridge 39 sampling 1568/2000ml at 0.39ml/s with 20.8psi vacuum @13:03:49.23 Cartridge 39 sampling 1590/2000ml at 0.38ml/s with 23.0psi vacuum @13:04:49.14 Cartridge 39 sampling 1612/2000ml at 0.36ml/s with 27.9psi vacuum @13:05:49.43 Cartridge 39 sampling 1634/2000ml at 0.34ml/s with 18.8psi vacuum 3.1m deep (19.3psia) @13:06:49.53 Cartridge 39 sampling 1657/2000ml at 0.35ml/s with 29.0psi vacuum @13:07:49.14 Cartridge 39 sampling 1679/2000ml at 0.37ml/s with 27.4psi vacuum @13:08:49.46 Cartridge 39 sampling 1701/2000ml at 0.37ml/s with 21.2psi vacuum 1.8m deep (17.4psia) @13:09:49.14 Cartridge 39 sampling 1722/2000ml at 0.36ml/s with 21.6psi vacuum @13:10:49.20 Cartridge 39 sampling 1744/2000ml at 0.36ml/s with 22.3psi vacuum @13:11:49.14 Cartridge 39 sampling 1765/2000ml at 0.34ml/s with 17.2psi vacuum @13:12:49.29 Cartridge 39 sampling 1786/2000ml at 0.32ml/s with 28.7psi vacuum @13:13:49.14 Cartridge 39 sampling 1807/2000ml at 0.34ml/s with 27.4psi vacuum @13:14:49.21 Cartridge 39 sampling 1828/2000ml at 0.34ml/s with 20.4psi vacuum @13:15:49.14 Cartridge 39 sampling 1848/2000ml at 0.35ml/s with 27.0psi vacuum @13:16:49.35 Cartridge 39 sampling 1869/2000ml at 0.34ml/s with 20.8psi vacuum 2.4m deep (18.3psia) @13:17:49.14 Cartridge 39 sampling 1889/2000ml at 0.34ml/s with 28.6psi vacuum @13:18:49.19 Cartridge 39 sampling 1910/2000ml at 0.31ml/s with 28.9psi vacuum @13:19:49.14 Cartridge 39 sampling 1930/2000ml at 0.34ml/s with 25.3psi vacuum @13:20:49.21 Cartridge 39 sampling 1950/2000ml at 0.34ml/s with 16.3psi vacuum 3.5m deep (19.9psia) @13:21:49.46 Cartridge 39 sampling 1970/2000ml at 0.30ml/s with 27.1psi vacuum @13:22:49.13 Cartridge 39 sampling 1990/2000ml at 0.34ml/s with 21.0psi vacuum @13:23:18.51 Pressure.coast @13:23:18.65 Intake.close @13:23:19.19 Exhaust.close @13:23:19.45 Toroid.reconfigure :hold @13:23:19.52 Toroid.seek :bypass @13:23:21.20 Toroid.reconfigure :default @13:23:21.37 Toroid.seek :clear @13:23:23.07 Sampled 2000.0ml @13:23:24.18 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 39" @13:23:25.08 SamplePump.setPosition! 0ml @13:23:25.12 Plunge.seek :reagent @13:23:25.21 Bypass.open @13:23:26.12 Intake.open @13:23:26.18 Sample loop closed. Ready to Deliver Reagents @13:23:26.20 Archive.seek :sealed @13:23:27.00 SamplePump.reconfigure :slow @13:23:27.05 SamplePump.seek -3ml @13:23:29.85 Archive.seek :evac @13:23:33.77 delay 1 second @13:23:34.99 SamplePump.seek -2.8ml @13:23:35.71 delay 1 second @13:23:36.80 SamplePump.seek -3.3ml @13:23:38.13 delay 1 second @13:23:39.16 Exhaust.open @13:23:39.53 upsync @13:23:39.88 Bypass.close @13:23:40.85 SamplePump.seek -35.3ml @13:23:43.83 delay 1:00 @13:24:43.91 Archive.seek :clear @13:24:49.67 delay 3 seconds @13:24:52.76 Intake.close @13:24:52.82 Exhaust.close @13:25:01.04 RNAlater soak for 10:00 @13:35:01.08 Archive.seek :tVent @13:35:02.75 delay 1 second @13:35:03.79 Archive.seek :evac @13:35:19.79 delay 10 seconds @13:35:29.84 Sample preserved and evacuated @13:35:30.33
Carousel.seek 39 @13:35:30.75 Plunge.seek :locked @13:35:31.81 Archive.seek :clear @13:35:31.96 Air.seek :clear @13:35:32.13 Lysis1.seek :clear @13:35:49.01
Archive.coast @13:35:49.17 Air.coast @13:35:49.37 Lysis1.coast @13:35:49.52 Lysis2.coast @13:35:49.56 Plunge.coast @13:35:49.61 SamplePump.coast @13:35:49.65 Toroid.coast @13:35:49.70 Carousel.coast @13:35:49.89 Gate.power :motor,:OFF @13:35:49.90 Gate.power :core,:OFF @13:35:49.92 Gate.power :analytic1,:OFF @13:35:49.93 Gate.power :analytic2,:OFF @13:35:49.94 Gate.power :analytic3,:OFF @13:35:49.96 Gate.power :main,:OFF @13:35:49.97 Safely stopped and ready to power off @13:35:50.21 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT18-Jun-21" @13:35:50.24
delayUntil "00:00:00.00PDT18-Jun-21" @13:36:01.15 upsync @13:59:23.98 Session Begun @13:59:26.72 -> slots @14:00:02.09 -> Cartridge.dry @14:02:11.06 -> Cartridge.release 37..30 @14:02:28.63 -> Cartridge.dry @14:02:42.57 -> slot spare:1 @14:02:50.85 -> exit @14:02:50.88 Session Ended @00:00:00.12PDT18-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.13 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.15 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.16 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.20 Powering core bus @00:00:02.73 Archive.configure :default @00:00:03.10 Air.configure :default @00:00:03.57 Lysis1.configure :default @00:00:03.89 Lysis2.configure :default @00:00:04.13 Plunge.configure :default @00:00:04.42 SamplePump.configure :default @00:00:04.74 ExtraPump.configure :default @00:00:05.09 Toroid.configure :default @00:00:05.48 Carousel.configure :default @00:00:05.77 Pressure.configure :default @00:00:05.86 All elf controllers running firmware version 4.61 @00:00:05.91 Gate.power :motor,:ON,limit to 5000mA @00:00:06.32 Archive.reconfigure :default @00:00:06.97 Archive.home.jog 0 @00:00:07.28 Plunge.reconfigure :default @00:00:07.54 Lysis2.reconfigure :default @00:00:07.61 Air.reconfigure :default @00:00:07.65 Toroid.reconfigure :default @00:00:07.70 Lysis1.reconfigure :default @00:00:07.97 Plunge.home.jog 0 @00:00:08.08 Lysis2.home.jog 0 @00:00:08.44 Toroid.home.jog 0 @00:00:08.71 Lysis1.home.jog 0 @00:00:08.74 Air.home.jog 0 @00:00:09.18 Archive.reconfigure :homing @00:00:09.40 Archive.jog 1373 @00:00:09.46 Plunge.reconfigure :homing @00:00:09.49 Lysis2.reconfigure :homing @00:00:09.52 Air.reconfigure :homing @00:00:09.56 Toroid.reconfigure :homing @00:00:09.61 Lysis1.reconfigure :homing @00:00:09.84 Air.jog 1363 @00:00:09.88 Lysis2.jog 1356 @00:00:09.91 Plunge.jog 850 @00:00:09.94 Toroid.jog 626 @00:00:09.96 Lysis1.jog 1349 @00:00:10.78 Archive.setRawPosition! 50 @00:00:10.83 Archive.seek :home @00:00:10.95 Plunge.setRawPosition! 50 @00:00:11.00 Plunge.seek :home @00:00:11.14 Lysis2.setRawPosition! 50 @00:00:11.17 Archive.reconfigure :default @00:00:11.28 Lysis2.seek :home @00:00:11.43 Lysis1.setRawPosition! 50 @00:00:11.46 Air.setRawPosition! 50 @00:00:11.48 Plunge.reconfigure :default @00:00:11.57 Lysis1.seek :home @00:00:11.72 Air.seek :home @00:00:11.83 Lysis2.reconfigure :default @00:00:11.96 Lysis1.reconfigure :default @00:00:12.05 Air.reconfigure :default @00:00:12.21 Toroid.setRawPosition! 0 @00:00:12.26 Toroid.seek :home @00:00:12.36 Toroid.reconfigure :default @00:00:12.51
Toroid.seek :clear @00:00:14.16 Carousel.reconfigure :default @00:00:14.27 Carousel.home.jog 0 @00:00:14.97 Carousel.seek 39 @00:00:15.28 Selecting Cartridge 38 @00:00:16.13 Carousel.seek 38 @00:00:16.92 Lysis1.seek :locked @00:00:20.72 Plunge.seek :locked @00:00:20.86 Air.seek :locked @00:00:25.23
Mounted archive Cartridge @00:00:25.47 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 38" @00:00:26.33 Duration of filtering limited to 4:05:00 @00:00:27.03 Priming sample loop w/160ml, bypass w/40ml @00:00:27.09 SamplePump.setPosition! 0ml @00:00:27.13 Bypass.open @00:00:28.68 Intake.open @00:00:29.55 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:32.14 Vacuum tare is 18% of full scale @00:00:32.16 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:32.38 Pressure.reconfigure :default @00:00:32.44 Pressure.hold @00:00:33.50 SamplePump.reconfigure :prime @00:00:33.91 SamplePump.seek 40ml,3:51 @00:00:39.10 upsync @00:01:17.19 Pressure.coast @00:01:17.24 SamplePump.setPosition! 0ml @00:01:17.29 Exhaust.open @00:01:18.69 Bypass.close @00:01:20.55 Pressure.reconfigure :default @00:01:20.65 Pressure.hold @00:01:21.31 SamplePump.reconfigure :prime @00:01:21.49 SamplePump.seek 160ml,13:51 @00:04:13.17 Pressure.coast @00:04:13.23 Bypass.open @00:04:13.97 Intake.close @00:04:15.18 Toroid.reconfigure :hold @00:04:15.34 Toroid.seek :crack @00:04:17.17 Toroid.seek :cartridge @00:04:18.15 Toroid.reconfigure :default @00:04:18.21 Toroid.seek :clear @00:04:20.09 Pressurizing cartridge to 26psi above 14.6psia @00:04:20.11 SamplePump.setPosition! 0ml @00:04:20.15 Intake.open @00:04:20.93 Bypass.close @00:04:21.67 Exhaust.close @00:04:22.48 Pressure.reconfigure :default @00:04:22.54 Pressure.hold @00:04:22.75 SamplePump.reconfigure :charge @00:04:22.81 SamplePump.seek 2.3ml,42 seconds @00:04:33.83 Pressure.coast @00:04:33.93 Intake.close @00:04:34.97 Pressurized cartridge with 0.547ml to 26.2psi @00:04:35.00 Isolated @35.7psia (21.1psi) @00:04:35.03 Checking for cartridge leaks from 35.7psia @00:04:35.06 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:41.16 Cartridge pressure fell 7% to 33.4psia @00:04:41.24 Sampling 2000.0ml @00:04:41.26 SamplePump.setPosition! 0ml @00:04:41.30 Exhaust.open @00:04:41.97 Intake.open @00:04:43.34 Pressure.reconfigure :default @00:04:43.53 Pressure.hold @00:04:44.23 SamplePump.reconfigure :sample @00:04:44.36 SamplePump.seek 2000ml,2:47:12 @00:04:46.36 Cartridge 38 sampling 1/2000ml at 0.55ml/s with 26.8psi vacuum @00:05:46.25 Cartridge 38 sampling 30/2000ml at 0.48ml/s with 23.4psi vacuum @00:06:46.79 Cartridge 38 sampling 60/2000ml at 0.48ml/s with 21.9psi vacuum @00:07:46.54 Cartridge 38 sampling 89/2000ml at 0.49ml/s with 24.7psi vacuum @00:08:46.30 Cartridge 38 sampling 119/2000ml at 0.51ml/s with 18.9psi vacuum 1.8m deep (17.3psia) @00:09:46.40 Cartridge 38 sampling 149/2000ml at 0.51ml/s with 28.3psi vacuum @00:10:46.26 Cartridge 38 sampling 180/2000ml at 0.51ml/s with 24.7psi vacuum @00:11:46.54 Cartridge 38 sampling 210/2000ml at 0.48ml/s with 21.5psi vacuum @00:12:46.70 Cartridge 38 sampling 241/2000ml at 0.52ml/s with 22.7psi vacuum @00:13:46.44 Cartridge 38 sampling 272/2000ml at 0.51ml/s with 26.0psi vacuum @00:14:46.26 Cartridge 38 sampling 302/2000ml at 0.51ml/s with 28.4psi vacuum @00:15:46.41 Cartridge 38 sampling 333/2000ml at 0.47ml/s with 20.6psi vacuum 2.4m deep (18.3psia) @00:16:46.26 Cartridge 38 sampling 363/2000ml at 0.52ml/s with 19.3psi vacuum @00:17:46.46 Cartridge 38 sampling 393/2000ml at 0.51ml/s with 24.9psi vacuum @00:18:46.26 Cartridge 38 sampling 423/2000ml at 0.48ml/s with 28.7psi vacuum @00:19:46.37 Cartridge 38 sampling 453/2000ml at 0.51ml/s with 26.4psi vacuum @00:20:46.26 Cartridge 38 sampling 483/2000ml at 0.50ml/s with 23.4psi vacuum @00:21:46.32 Cartridge 38 sampling 512/2000ml at 0.50ml/s with 18.7psi vacuum 3.7m deep (20.2psia) @00:22:46.26 Cartridge 38 sampling 542/2000ml at 0.50ml/s with 19.7psi vacuum 1.7m deep (17.2psia) @00:23:46.38 Cartridge 38 sampling 571/2000ml at 0.46ml/s with 18.6psi vacuum 3.9m deep (20.4psia) @00:24:46.68 Cartridge 38 sampling 601/2000ml at 0.51ml/s with 22.3psi vacuum @00:25:46.26 Cartridge 38 sampling 630/2000ml at 0.49ml/s with 18.3psi vacuum 2.3m deep (18.1psia) @00:26:46.51 Cartridge 38 sampling 659/2000ml at 0.48ml/s with 29.5psi vacuum @00:27:46.25 Cartridge 38 sampling 688/2000ml at 0.49ml/s with 18.9psi vacuum @00:28:46.39 Cartridge 38 sampling 717/2000ml at 0.49ml/s with 26.0psi vacuum @00:29:46.26 Cartridge 38 sampling 746/2000ml at 0.47ml/s with 29.0psi vacuum @00:30:46.44 Cartridge 38 sampling 774/2000ml at 0.45ml/s with 30.1psi vacuum @00:31:46.26 Cartridge 38 sampling 803/2000ml at 0.47ml/s with 28.4psi vacuum @00:32:46.42 Cartridge 38 sampling 831/2000ml at 0.45ml/s with 27.8psi vacuum @00:33:46.26 Cartridge 38 sampling 859/2000ml at 0.47ml/s with 19.8psi vacuum 1.9m deep (17.5psia) @00:34:46.34 Cartridge 38 sampling 887/2000ml at 0.47ml/s with 28.9psi vacuum @00:35:46.26 Cartridge 38 sampling 914/2000ml at 0.47ml/s with 22.9psi vacuum @00:36:46.29 Cartridge 38 sampling 942/2000ml at 0.45ml/s with 13.7psi vacuum 2.9m deep (19.0psia) @00:37:46.41 Cartridge 38 sampling 969/2000ml at 0.45ml/s with 14.6psi vacuum 8.0m deep (26.6psia) @00:38:46.26 Cartridge 38 sampling 996/2000ml at 0.43ml/s with 29.8psi vacuum @00:39:46.53 Cartridge 38 sampling 1023/2000ml at 0.45ml/s with 28.4psi vacuum @00:40:46.25 Cartridge 38 sampling 1049/2000ml at 0.45ml/s with 17.2psi vacuum 3.7m deep (20.1psia) @00:41:46.46 Cartridge 38 sampling 1076/2000ml at 0.45ml/s with 16.3psi vacuum 5.1m deep (22.2psia) @00:42:46.26 Cartridge 38 sampling 1102/2000ml at 0.43ml/s with 12.0psi vacuum 5.1m deep (22.2psia) @00:43:46.41 Cartridge 38 sampling 1128/2000ml at 0.46ml/s with 26.0psi vacuum @00:44:46.26 Cartridge 38 sampling 1154/2000ml at 0.45ml/s with 17.2psi vacuum 4.0m deep (20.5psia) @00:45:46.43 Cartridge 38 sampling 1180/2000ml at 0.42ml/s with 28.8psi vacuum @00:46:46.25 Cartridge 38 sampling 1206/2000ml at 0.39ml/s with 26.8psi vacuum @00:47:46.36 Cartridge 38 sampling 1232/2000ml at 0.41ml/s with 12.9psi vacuum 3.7m deep (20.2psia) @00:48:46.25 Cartridge 38 sampling 1257/2000ml at 0.44ml/s with 25.5psi vacuum @00:49:46.29 Cartridge 38 sampling 1282/2000ml at 0.41ml/s with 15.0psi vacuum 3.1m deep (19.3psia) @00:50:46.25 Cartridge 38 sampling 1307/2000ml at 0.42ml/s with 28.1psi vacuum @00:51:46.53 Cartridge 38 sampling 1331/2000ml at 0.38ml/s with 19.0psi vacuum 4.1m deep (20.7psia) @00:52:46.70 Cartridge 38 sampling 1356/2000ml at 0.43ml/s with 22.9psi vacuum @00:53:46.25 Cartridge 38 sampling 1380/2000ml at 0.40ml/s with 16.0psi vacuum @00:54:46.52 Cartridge 38 sampling 1404/2000ml at 0.40ml/s with 16.5psi vacuum 6.0m deep (23.5psia) @00:55:46.24 Cartridge 38 sampling 1428/2000ml at 0.38ml/s with 13.7psi vacuum @00:56:46.41 Cartridge 38 sampling 1452/2000ml at 0.37ml/s with 18.0psi vacuum @00:57:46.25 Cartridge 38 sampling 1476/2000ml at 0.41ml/s with 27.8psi vacuum @00:58:46.35 Cartridge 38 sampling 1499/2000ml at 0.37ml/s with 13.6psi vacuum 1.8m deep (17.3psia) @00:59:46.25 Cartridge 38 sampling 1523/2000ml at 0.40ml/s with 17.2psi vacuum 1.7m deep (17.2psia) @01:00:46.35 Cartridge 38 sampling 1546/2000ml at 0.40ml/s with 29.3psi vacuum @01:01:46.24 Cartridge 38 sampling 1569/2000ml at 0.36ml/s with 29.8psi vacuum @01:02:46.36 Cartridge 38 sampling 1592/2000ml at 0.39ml/s with 17.9psi vacuum 5.4m deep (22.7psia) @01:03:46.24 Cartridge 38 sampling 1615/2000ml at 0.37ml/s with 29.2psi vacuum @01:04:46.42 Cartridge 38 sampling 1638/2000ml at 0.37ml/s with 28.1psi vacuum @01:05:46.25 Cartridge 38 sampling 1660/2000ml at 0.37ml/s with 16.4psi vacuum @01:06:46.34 Cartridge 38 sampling 1682/2000ml at 0.39ml/s with 20.7psi vacuum 3.0m deep (19.1psia) @01:07:46.42 Cartridge 38 sampling 1704/2000ml at 0.37ml/s with 16.7psi vacuum 6.2m deep (23.9psia) @01:08:46.25 Cartridge 38 sampling 1726/2000ml at 0.34ml/s with 12.5psi vacuum 2.2m deep (17.9psia) @01:09:46.50 Cartridge 38 sampling 1748/2000ml at 0.35ml/s with 28.9psi vacuum @01:10:46.26 Cartridge 38 sampling 1769/2000ml at 0.33ml/s with 30.1psi vacuum @01:11:46.58 Cartridge 38 sampling 1791/2000ml at 0.37ml/s with 18.3psi vacuum 4.9m deep (21.9psia) @01:12:46.26 Cartridge 38 sampling 1812/2000ml at 0.32ml/s with 13.5psi vacuum @01:13:46.33 Cartridge 38 sampling 1833/2000ml at 0.34ml/s with 12.4psi vacuum 4.0m deep (20.7psia) @01:14:46.26 Cartridge 38 sampling 1854/2000ml at 0.35ml/s with 29.5psi vacuum @01:15:46.38 Cartridge 38 sampling 1875/2000ml at 0.33ml/s with 11.6psi vacuum 8.3m deep (26.9psia) @01:16:46.26 Cartridge 38 sampling 1896/2000ml at 0.34ml/s with 15.5psi vacuum @01:17:46.50 Cartridge 38 sampling 1916/2000ml at 0.32ml/s with 10.5psi vacuum 4.8m deep (21.7psia) @01:18:46.26 Cartridge 38 sampling 1937/2000ml at 0.33ml/s with 19.0psi vacuum @01:19:46.57 Cartridge 38 sampling 1957/2000ml at 0.31ml/s with 11.5psi vacuum 4.8m deep (21.8psia) @01:20:46.25 Cartridge 38 sampling 1978/2000ml at 0.35ml/s with 17.1psi vacuum 2.6m deep (18.6psia) @01:21:46.25 Cartridge 38 sampling 1998/2000ml at 0.35ml/s with 29.5psi vacuum @01:21:51.77 Pressure.coast @01:21:52.11 Intake.close @01:21:52.24 Exhaust.close @01:21:52.69 Toroid.reconfigure :hold @01:21:52.80 Toroid.seek :bypass @01:21:54.46 Toroid.reconfigure :default @01:21:54.57 Toroid.seek :clear @01:21:56.23 Sampled 2000.0ml @01:21:56.68 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 38" @01:21:57.09 SamplePump.setPosition! 0ml @01:21:57.15 Plunge.seek :reagent @01:21:57.18 Bypass.open @01:21:58.53 Intake.open @01:21:58.63 Sample loop closed. Ready to Deliver Reagents @01:21:58.65 Archive.seek :sealed @01:21:59.41 SamplePump.reconfigure :slow @01:21:59.47 SamplePump.seek -3ml @01:22:02.34 Archive.seek :evac @01:22:06.17 delay 1 second @01:22:07.28 SamplePump.seek -2.8ml @01:22:08.03 delay 1 second @01:22:09.15 SamplePump.seek -3.3ml @01:22:10.01 upsync @01:22:10.59 delay 1 second @01:22:11.62 Exhaust.open @01:22:12.36 Bypass.close @01:22:13.21 SamplePump.seek -35.3ml @01:22:16.44 delay 1:00 @01:23:16.49 Archive.seek :clear @01:23:21.90 delay 3 seconds @01:23:25.16 Intake.close @01:23:25.31 Exhaust.close @01:23:33.73 RNAlater soak for 10:00 @01:33:33.77 Archive.seek :tVent @01:33:35.41 delay 1 second @01:33:36.45 Archive.seek :evac @01:33:52.40 delay 10 seconds @01:34:02.44 Sample preserved and evacuated @01:34:02.67
Carousel.seek 38 @01:34:02.87 Plunge.seek :locked @01:34:03.86 Archive.seek :clear @01:34:04.03 Air.seek :clear @01:34:04.33 Lysis1.seek :clear @01:34:21.05
Archive.coast @01:34:21.09 Air.coast @01:34:21.14 Lysis1.coast @01:34:21.18 Lysis2.coast @01:34:21.22 Plunge.coast @01:34:21.26 SamplePump.coast @01:34:21.31 Toroid.coast @01:34:21.35 Carousel.coast @01:34:21.55 Gate.power :motor,:OFF @01:34:21.56 Gate.power :core,:OFF @01:34:21.58 Gate.power :analytic1,:OFF @01:34:21.59 Gate.power :analytic2,:OFF @01:34:21.60 Gate.power :analytic3,:OFF @01:34:21.62 Gate.power :main,:OFF @01:34:21.63 Safely stopped and ready to power off @01:34:21.88 delayUntil "12:00:00.00" @01:34:21.90 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @01:34:37.42 upsync @12:00:00.29
Gate.power :main,:ON,limit to 5000mA @12:00:00.30 Gate.power :analytic1,:ON,limit to 5000mA @12:00:00.31 Gate.power :analytic2,:ON,limit to 5000mA @12:00:00.33 Gate.power :analytic3,:ON,limit to 5000mA @12:00:00.37 Powering core bus @12:00:02.65 Archive.configure :default @12:00:02.79 Air.configure :default @12:00:02.91 Lysis1.configure :default @12:00:03.02 Lysis2.configure :default @12:00:03.16 Plunge.configure :default @12:00:03.30 SamplePump.configure :default @12:00:03.43 ExtraPump.configure :default @12:00:03.57 Toroid.configure :default @12:00:03.71 Carousel.configure :default @12:00:03.85 Pressure.configure :default @12:00:03.90 All elf controllers running firmware version 4.61 @12:00:03.92 Gate.power :motor,:ON,limit to 5000mA @12:00:04.08 Archive.reconfigure :default @12:00:04.37 Archive.home.jog 0 @12:00:04.59 Plunge.reconfigure :default @12:00:04.65 Lysis2.reconfigure :default @12:00:04.74 Lysis1.reconfigure :default @12:00:04.84 Air.reconfigure :default @12:00:04.94 Toroid.reconfigure :default @12:00:05.94 Plunge.home.jog 0 @12:00:06.16 Lysis2.home.jog 0 @12:00:06.20 Lysis1.home.jog 0 @12:00:06.23 Air.home.jog 0 @12:00:06.31 Toroid.home.jog 0 @12:00:07.51 Archive.reconfigure :homing @12:00:08.00 Archive.jog 1374 @12:00:08.16 Plunge.reconfigure :homing @12:00:08.23 Lysis2.reconfigure :homing @12:00:08.40 Air.reconfigure :homing @12:00:08.43 Toroid.reconfigure :homing @12:00:08.52 Lysis1.reconfigure :homing @12:00:09.00 Lysis2.jog 1358 @12:00:09.05 Plunge.jog 852 @12:00:09.11 Air.jog 1364 @12:00:09.27 Toroid.jog 634 @12:00:09.30 Lysis1.jog 1348 @12:00:09.86 Archive.setRawPosition! 50 @12:00:09.93 Archive.seek :home @12:00:10.41 Archive.reconfigure :default @12:00:10.73 Plunge.setRawPosition! 50 @12:00:10.77 Lysis2.setRawPosition! 50 @12:00:10.79 Lysis1.setRawPosition! 50 @12:00:10.81 Air.setRawPosition! 50 @12:00:10.97 Plunge.seek :home @12:00:11.01 Lysis2.seek :home @12:00:11.04 Lysis1.seek :home @12:00:11.08 Air.seek :home @12:00:11.39 Lysis2.reconfigure :default @12:00:11.51 Plunge.reconfigure :default @12:00:11.55 Lysis1.reconfigure :default @12:00:11.69 Air.reconfigure :default @12:00:11.79 Toroid.setRawPosition! 0 @12:00:11.83 Toroid.seek :home @12:00:11.93 Toroid.reconfigure :default @12:00:12.11
Toroid.seek :clear @12:00:13.73 Carousel.reconfigure :default @12:00:13.78 Carousel.home.jog 0 @12:00:14.29 Carousel.seek 38 @12:00:14.47 Selecting Cartridge 37 @12:00:14.68 Carousel.seek 37 @12:00:15.57 Lysis1.seek :locked @12:00:18.91 Plunge.seek :locked @12:00:19.22 Air.seek :locked @12:00:23.45
Mounted archive Cartridge @12:00:23.55 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 37" @12:00:23.64 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:00:23.67 Retry #1 of 10 after 10:00 @12:00:23.92 Duration of filtering limited to 4:05:00 @12:00:24.20 Priming sample loop w/160ml, bypass w/40ml @12:00:24.22 SamplePump.setPosition! 0ml @12:00:24.25 Bypass.open @12:00:25.75 Intake.open @12:00:26.73 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:00:29.47 Vacuum tare is 18% of full scale @12:00:29.51 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:00:29.89 Pressure.reconfigure :default @12:00:29.95 Pressure.hold @12:00:30.15 SamplePump.reconfigure :prime @12:00:30.21 SamplePump.seek 40ml,3:51 @12:01:13.99 Pressure.coast @12:01:14.11 SamplePump.setPosition! 0ml @12:01:14.18 Exhaust.open @12:01:15.65 Bypass.close @12:01:17.35 Pressure.reconfigure :default @12:01:17.41 Pressure.hold @12:01:17.68 SamplePump.reconfigure :prime @12:01:17.74 SamplePump.seek 160ml,13:51 @12:04:09.45 Pressure.coast @12:04:09.51 Bypass.open @12:04:10.20 Intake.close @12:04:11.09 Toroid.reconfigure :hold @12:04:11.15 Toroid.seek :crack @12:04:12.73 Toroid.seek :cartridge @12:04:13.67 Toroid.reconfigure :default @12:04:13.73 Toroid.seek :clear @12:04:16.04 Pressurizing cartridge to 26psi above 14.5psia @12:04:16.07 SamplePump.setPosition! 0ml @12:04:16.11 Intake.open @12:04:16.91 Bypass.close @12:04:17.71 Exhaust.close @12:04:18.64 Pressure.reconfigure :default @12:04:18.76 Pressure.hold @12:04:19.15 SamplePump.reconfigure :charge @12:04:19.43 SamplePump.seek 2.3ml,42 seconds @12:04:29.62 Pressure.coast @12:04:29.67 Intake.close @12:04:30.51 Pressurized cartridge with 0.467ml to 26.1psi @12:04:30.52 Isolated @35.1psia (20.6psi) @12:04:30.53 Checking for cartridge leaks from 35.1psia @12:04:30.55 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:04:36.80 Cartridge pressure fell 17% to 29.2psia @12:04:36.93 Sampling 2000.0ml @12:04:36.98 SamplePump.setPosition! 0ml @12:04:37.10 Exhaust.open @12:04:37.87 Intake.open @12:04:38.91 Pressure.reconfigure :default @12:04:38.96 Pressure.hold @12:04:39.24 SamplePump.reconfigure :sample @12:04:39.30 SamplePump.seek 2000ml,2:47:12 @12:04:41.97 Cartridge 37 sampling 1/2000ml at 0.45ml/s with 20.3psi vacuum @12:05:42.13 Cartridge 37 sampling 29/2000ml at 0.48ml/s with 21.2psi vacuum @12:06:41.97 Cartridge 37 sampling 57/2000ml at 0.45ml/s with 24.5psi vacuum @12:07:42.15 Cartridge 37 sampling 86/2000ml at 0.47ml/s with 26.0psi vacuum @12:08:42.13 Cartridge 37 sampling 114/2000ml at 0.48ml/s with 23.5psi vacuum @12:09:41.97 Cartridge 37 sampling 142/2000ml at 0.48ml/s with 24.9psi vacuum @12:10:23.76 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:10:23.80 Retry #2 of 10 after 10:00 @12:10:42.14 Cartridge 37 sampling 171/2000ml at 0.47ml/s with 26.4psi vacuum @12:11:41.96 Cartridge 37 sampling 199/2000ml at 0.48ml/s with 25.3psi vacuum @12:12:42.15 Cartridge 37 sampling 227/2000ml at 0.47ml/s with 25.2psi vacuum @12:13:41.97 Cartridge 37 sampling 256/2000ml at 0.49ml/s with 26.4psi vacuum @12:14:42.12 Cartridge 37 sampling 284/2000ml at 0.47ml/s with 25.5psi vacuum @12:15:41.97 Cartridge 37 sampling 311/2000ml at 0.43ml/s with 26.6psi vacuum @12:16:42.18 Cartridge 37 sampling 339/2000ml at 0.47ml/s with 26.7psi vacuum @12:17:41.97 Cartridge 37 sampling 367/2000ml at 0.46ml/s with 23.5psi vacuum @12:18:42.05 Cartridge 37 sampling 395/2000ml at 0.46ml/s with 20.4psi vacuum @12:19:41.97 Cartridge 37 sampling 422/2000ml at 0.47ml/s with 26.4psi vacuum @12:20:23.90 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:20:23.97 Retry #3 of 10 after 10:00 @12:20:42.20 Cartridge 37 sampling 450/2000ml at 0.47ml/s with 26.1psi vacuum @12:21:42.08 Cartridge 37 sampling 478/2000ml at 0.45ml/s with 27.0psi vacuum @12:22:41.97 Cartridge 37 sampling 506/2000ml at 0.46ml/s with 27.2psi vacuum @12:23:42.33 Cartridge 37 sampling 534/2000ml at 0.46ml/s with 18.5psi vacuum @12:24:41.97 Cartridge 37 sampling 561/2000ml at 0.48ml/s with 20.7psi vacuum @12:25:42.10 Cartridge 37 sampling 589/2000ml at 0.44ml/s with 18.2psi vacuum @12:26:41.97 Cartridge 37 sampling 616/2000ml at 0.45ml/s with 26.4psi vacuum @12:27:42.13 Cartridge 37 sampling 643/2000ml at 0.46ml/s with 24.7psi vacuum @12:28:41.97 Cartridge 37 sampling 671/2000ml at 0.44ml/s with 21.6psi vacuum @12:29:42.11 Cartridge 37 sampling 698/2000ml at 0.44ml/s with 24.8psi vacuum @12:30:24.02 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:30:24.04 Retry #4 of 10 after 10:00 @12:30:41.97 Cartridge 37 sampling 725/2000ml at 0.44ml/s with 26.5psi vacuum @12:31:42.22 Cartridge 37 sampling 752/2000ml at 0.45ml/s with 23.5psi vacuum @12:32:41.97 Cartridge 37 sampling 779/2000ml at 0.46ml/s with 21.3psi vacuum @12:33:42.10 Cartridge 37 sampling 805/2000ml at 0.44ml/s with 27.4psi vacuum @12:34:41.97 Cartridge 37 sampling 832/2000ml at 0.42ml/s with 21.7psi vacuum @12:35:42.07 Cartridge 37 sampling 858/2000ml at 0.44ml/s with 18.0psi vacuum 2.8m deep (18.8psia) @12:36:42.11 Cartridge 37 sampling 884/2000ml at 0.44ml/s with 24.2psi vacuum @12:37:41.97 Cartridge 37 sampling 910/2000ml at 0.40ml/s with 27.4psi vacuum @12:38:42.17 Cartridge 37 sampling 936/2000ml at 0.44ml/s with 22.6psi vacuum @12:39:41.97 Cartridge 37 sampling 962/2000ml at 0.43ml/s with 27.4psi vacuum @12:40:24.13 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:40:24.19 Retry #5 of 10 after 10:00 @12:40:42.08 Cartridge 37 sampling 988/2000ml at 0.43ml/s with 25.9psi vacuum @12:41:42.51 Cartridge 37 sampling 1013/2000ml at 0.41ml/s with 27.9psi vacuum @12:42:42.13 Cartridge 37 sampling 1039/2000ml at 0.44ml/s with 22.2psi vacuum @12:43:41.97 Cartridge 37 sampling 1064/2000ml at 0.42ml/s with 22.5psi vacuum @12:44:42.19 Cartridge 37 sampling 1089/2000ml at 0.42ml/s with 23.6psi vacuum @12:45:41.97 Cartridge 37 sampling 1114/2000ml at 0.43ml/s with 25.5psi vacuum @12:46:42.06 Cartridge 37 sampling 1139/2000ml at 0.40ml/s with 28.4psi vacuum @12:47:41.97 Cartridge 37 sampling 1164/2000ml at 0.40ml/s with 28.3psi vacuum @12:48:42.46 Cartridge 37 sampling 1189/2000ml at 0.39ml/s with 28.0psi vacuum @12:49:42.24 Cartridge 37 sampling 1214/2000ml at 0.41ml/s with 25.7psi vacuum @12:50:24.25 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:50:24.27 Retry #6 of 10 after 10:00 @12:50:42.00 Cartridge 37 sampling 1238/2000ml at 0.39ml/s with 28.2psi vacuum @12:51:42.22 Cartridge 37 sampling 1263/2000ml at 0.42ml/s with 25.6psi vacuum @12:52:41.97 Cartridge 37 sampling 1287/2000ml at 0.41ml/s with 16.6psi vacuum 2.7m deep (18.7psia) @12:53:42.17 Cartridge 37 sampling 1311/2000ml at 0.39ml/s with 26.3psi vacuum @12:54:41.97 Cartridge 37 sampling 1335/2000ml at 0.40ml/s with 20.3psi vacuum @12:55:42.21 Cartridge 37 sampling 1359/2000ml at 0.42ml/s with 25.1psi vacuum @12:56:41.96 Cartridge 37 sampling 1383/2000ml at 0.38ml/s with 22.0psi vacuum @12:57:42.15 Cartridge 37 sampling 1407/2000ml at 0.40ml/s with 25.7psi vacuum @12:58:41.96 Cartridge 37 sampling 1430/2000ml at 0.38ml/s with 28.9psi vacuum @12:59:42.38 Cartridge 37 sampling 1454/2000ml at 0.38ml/s with 26.4psi vacuum @13:00:24.32 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:00:24.34 Retry #7 of 10 after 10:00 @13:00:41.97 Cartridge 37 sampling 1477/2000ml at 0.38ml/s with 26.8psi vacuum @13:01:41.99 Cartridge 37 sampling 1500/2000ml at 0.38ml/s with 20.7psi vacuum @13:02:41.96 Cartridge 37 sampling 1523/2000ml at 0.38ml/s with 21.4psi vacuum @13:03:42.14 Cartridge 37 sampling 1546/2000ml at 0.38ml/s with 20.1psi vacuum 2.7m deep (18.7psia) @13:04:42.23 Cartridge 37 sampling 1568/2000ml at 0.37ml/s with 20.7psi vacuum 1.8m deep (17.3psia) @13:05:41.97 Cartridge 37 sampling 1591/2000ml at 0.37ml/s with 20.7psi vacuum @13:06:42.28 Cartridge 37 sampling 1613/2000ml at 0.35ml/s with 28.6psi vacuum @13:07:41.97 Cartridge 37 sampling 1635/2000ml at 0.34ml/s with 26.0psi vacuum @13:08:42.07 Cartridge 37 sampling 1657/2000ml at 0.37ml/s with 26.6psi vacuum @13:09:41.97 Cartridge 37 sampling 1679/2000ml at 0.37ml/s with 21.1psi vacuum @13:10:24.45 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:10:24.48 Retry #8 of 10 after 10:00 @13:10:42.10 Cartridge 37 sampling 1701/2000ml at 0.38ml/s with 25.6psi vacuum @13:11:41.97 Cartridge 37 sampling 1723/2000ml at 0.38ml/s with 16.8psi vacuum 2.5m deep (18.3psia) @13:12:42.19 Cartridge 37 sampling 1745/2000ml at 0.37ml/s with 25.0psi vacuum @13:13:41.97 Cartridge 37 sampling 1766/2000ml at 0.38ml/s with 24.2psi vacuum @13:14:42.08 Cartridge 37 sampling 1788/2000ml at 0.35ml/s with 29.1psi vacuum @13:15:41.97 Cartridge 37 sampling 1809/2000ml at 0.37ml/s with 25.0psi vacuum @13:16:42.27 Cartridge 37 sampling 1830/2000ml at 0.37ml/s with 20.7psi vacuum 3.5m deep (19.8psia) @13:17:42.62 Cartridge 37 sampling 1851/2000ml at 0.31ml/s with 27.4psi vacuum @13:18:41.97 Cartridge 37 sampling 1872/2000ml at 0.38ml/s with 17.6psi vacuum 2.2m deep (17.9psia) @13:19:42.19 Cartridge 37 sampling 1893/2000ml at 0.37ml/s with 23.7psi vacuum @13:20:24.53 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:20:24.55 Retry #9 of 10 after 10:00 @13:20:41.97 Cartridge 37 sampling 1914/2000ml at 0.36ml/s with 28.9psi vacuum @13:21:42.16 Cartridge 37 sampling 1935/2000ml at 0.36ml/s with 19.3psi vacuum 2.0m deep (17.7psia) @13:22:41.97 Cartridge 37 sampling 1955/2000ml at 0.37ml/s with 21.9psi vacuum @13:23:42.16 Cartridge 37 sampling 1976/2000ml at 0.34ml/s with 28.1psi vacuum @13:24:41.97 Cartridge 37 sampling 1996/2000ml at 0.31ml/s with 13.8psi vacuum @13:24:53.02 Pressure.coast @13:24:53.16 Intake.close @13:24:53.26 Exhaust.close @13:24:53.76 Toroid.reconfigure :hold @13:24:53.93 Toroid.seek :bypass @13:24:55.62 Toroid.reconfigure :default @13:24:55.74 Toroid.seek :clear @13:24:57.44 Sampled 2000.0ml @13:24:59.12 SamplePump.setPosition! 0ml @13:24:59.16 Plunge.seek :reagent @13:24:59.24 Bypass.open @13:24:59.91 Intake.open @13:25:00.07 Sample loop closed. Ready to Deliver Reagents @13:25:00.09 Archive.seek :sealed @13:25:00.76 SamplePump.reconfigure :slow @13:25:00.82 SamplePump.seek -3ml @13:25:03.78 Archive.seek :evac @13:25:07.56 delay 1 second @13:25:08.66 SamplePump.seek -2.8ml @13:25:09.36 delay 1 second @13:25:10.45 SamplePump.seek -3.3ml @13:25:11.78 delay 1 second @13:25:12.81 Exhaust.open @13:25:13.73 Bypass.close @13:25:14.68 SamplePump.seek -35.3ml @13:25:17.80 delay 1:00 @13:26:17.85 Archive.seek :clear @13:26:23.47 delay 3 seconds @13:26:26.62 Intake.close @13:26:26.88 Exhaust.close @13:26:35.01 RNAlater soak for 10:00 @13:30:24.60 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:30:24.62 Retry #10 of 10 after 10:00 @13:36:35.06 Archive.seek :tVent @13:36:36.71 delay 1 second @13:36:37.74 Archive.seek :evac @13:36:53.70 delay 10 seconds @13:37:03.74 Sample preserved and evacuated @13:37:04.49
Carousel.seek 37 @13:37:04.70 Plunge.seek :locked @13:37:05.68 Archive.seek :clear @13:37:05.85 Air.seek :clear @13:37:06.03 Lysis1.seek :clear @13:37:22.86
Archive.coast @13:37:22.91 Air.coast @13:37:22.95 Lysis1.coast @13:37:22.99 Lysis2.coast @13:37:23.03 Plunge.coast @13:37:23.08 SamplePump.coast @13:37:23.12 Toroid.coast @13:37:23.17 Carousel.coast @13:37:23.36 Gate.power :motor,:OFF @13:37:23.37 Gate.power :core,:OFF @13:37:23.39 Gate.power :analytic1,:OFF @13:37:23.40 Gate.power :analytic2,:OFF @13:37:23.41 Gate.power :analytic3,:OFF @13:37:23.43 Gate.power :main,:OFF @13:37:23.44 Safely stopped and ready to power off @13:37:23.82 delayUntil "00:00:00.00PDT19-Jun-21" @13:40:24.77 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:40:24.86 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 37" @13:40:24.89 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:40:24.91 Retry #1 of 10 after 10:00 @13:50:24.96 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:50:24.99 Retry #2 of 10 after 10:00 @14:00:25.07 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:00:25.10 Retry #3 of 10 after 10:00 @14:10:25.14 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:10:25.16 Retry #4 of 10 after 10:00 @14:20:25.25 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:20:25.33 Retry #5 of 10 after 10:00 @14:30:25.42 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:30:25.45 Retry #6 of 10 after 10:00 @14:40:35.91 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT19-Jun-21" @14:40:45.79 upsync @14:41:07.12 upsync @00:00:00.28PDT19-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.30 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.33 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.41 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.52 Powering core bus @00:00:02.89 Archive.configure :default @00:00:03.03 Air.configure :default @00:00:03.13 Lysis1.configure :default @00:00:03.79 Lysis2.configure :default @00:00:03.94 Plunge.configure :default @00:00:04.07 SamplePump.configure :default @00:00:04.22 ExtraPump.configure :default @00:00:04.36 Toroid.configure :default @00:00:04.49 Carousel.configure :default @00:00:04.64 Pressure.configure :default @00:00:04.70 All elf controllers running firmware version 4.61 @00:00:04.72 Gate.power :motor,:ON,limit to 5000mA @00:00:05.03 Archive.reconfigure :default @00:00:05.28 Archive.home.jog 0 @00:00:05.52 Toroid.reconfigure :default @00:00:05.57 Lysis2.reconfigure :default @00:00:05.61 Lysis1.reconfigure :default @00:00:05.64 Air.reconfigure :default @00:00:05.68 Plunge.reconfigure :default @00:00:05.87 Toroid.home.jog 0 @00:00:05.94 Lysis1.home.jog 0 @00:00:05.97 Air.home.jog 0 @00:00:05.99 Plunge.home.jog 0 @00:00:06.02 Lysis2.home.jog 0 @00:00:06.52 Archive.reconfigure :homing @00:00:06.76 Archive.jog 1371 @00:00:06.82 Plunge.reconfigure :homing @00:00:06.85 Lysis2.reconfigure :homing @00:00:06.89 Air.reconfigure :homing @00:00:06.93 Toroid.reconfigure :homing @00:00:06.97 Lysis1.reconfigure :homing @00:00:07.55 Plunge.jog 844 @00:00:07.68 Lysis2.jog 1356 @00:00:07.74 Air.jog 1363 @00:00:07.81 Lysis1.jog 1349 @00:00:07.87 Toroid.jog 636 @00:00:08.94 Archive.setRawPosition! 50 @00:00:09.07 Archive.seek :home @00:00:09.24 Plunge.setRawPosition! 50 @00:00:09.31 Lysis2.setRawPosition! 50 @00:00:09.52 Plunge.seek :home @00:00:09.60 Lysis2.seek :home @00:00:09.89 Lysis1.setRawPosition! 50 @00:00:10.10 Air.setRawPosition! 50 @00:00:10.12 Archive.reconfigure :default @00:00:10.29 Lysis1.seek :home @00:00:10.43 Air.seek :home @00:00:10.89 Toroid.setRawPosition! 0 @00:00:11.03 Lysis2.reconfigure :default @00:00:11.06 Plunge.reconfigure :default @00:00:11.34 Toroid.seek :home @00:00:11.81 Lysis1.reconfigure :default @00:00:11.89 Air.reconfigure :default @00:00:11.92 Toroid.reconfigure :default @00:00:12.61
Toroid.seek :clear @00:00:14.28 Carousel.reconfigure :default @00:00:14.34 Carousel.home.jog 0 @00:00:14.83 Carousel.seek 37 @00:00:15.02 Selecting Cartridge 36 @00:00:15.23 Carousel.seek 36 @00:00:15.75 Lysis1.seek :locked @00:00:19.16 Plunge.seek :locked @00:00:19.47 Air.seek :locked @00:00:23.79
Mounted archive Cartridge @00:00:23.89 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 36" @00:00:24.22 Duration of filtering limited to 4:05:00 @00:00:24.52 Priming sample loop w/160ml, bypass w/40ml @00:00:24.54 SamplePump.setPosition! 0ml @00:00:24.58 Bypass.open @00:00:26.05 Intake.open @00:00:27.00 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:29.80 Vacuum tare is 18% of full scale @00:00:29.82 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:30.31 Pressure.reconfigure :default @00:00:30.45 Pressure.hold @00:00:31.01 SamplePump.reconfigure :prime @00:00:31.18 SamplePump.seek 40ml,3:51 @00:00:44.47 upsync @00:01:14.42 Pressure.coast @00:01:14.56 SamplePump.setPosition! 0ml @00:01:14.70 Exhaust.open @00:01:16.21 Bypass.close @00:01:18.06 Pressure.reconfigure :default @00:01:18.24 Pressure.hold @00:01:18.86 SamplePump.reconfigure :prime @00:01:19.25 SamplePump.seek 160ml,13:51 @00:04:10.99 Pressure.coast @00:04:11.07 Bypass.open @00:04:11.88 Intake.close @00:04:12.98 Toroid.reconfigure :hold @00:04:13.04 Toroid.seek :crack @00:04:14.62 Toroid.seek :cartridge @00:04:15.55 Toroid.reconfigure :default @00:04:15.61 Toroid.seek :clear @00:04:17.45 Pressurizing cartridge to 26psi above 14.5psia @00:04:17.47 SamplePump.setPosition! 0ml @00:04:17.51 Intake.open @00:04:18.27 Bypass.close @00:04:19.04 Exhaust.close @00:04:20.11 Pressure.reconfigure :default @00:04:20.23 Pressure.hold @00:04:20.82 SamplePump.reconfigure :charge @00:04:20.94 SamplePump.seek 2.3ml,42 seconds @00:04:31.76 Pressure.coast @00:04:31.80 Intake.close @00:04:32.60 Pressurized cartridge with 0.516ml to 26.1psi @00:04:32.61 Isolated @35.8psia (21.2psi) @00:04:32.62 Checking for cartridge leaks from 35.8psia @00:04:32.64 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:38.84 Cartridge pressure fell 14% to 30.6psia @00:04:39.09 Sampling 2000.0ml @00:04:39.11 SamplePump.setPosition! 0ml @00:04:39.28 Exhaust.open @00:04:40.03 Intake.open @00:04:41.19 Pressure.reconfigure :default @00:04:41.33 Pressure.hold @00:04:41.99 SamplePump.reconfigure :sample @00:04:42.14 SamplePump.seek 2000ml,2:47:12 @00:04:44.17 Cartridge 36 sampling 1/2000ml at 0.57ml/s with 27.8psi vacuum @00:05:44.33 Cartridge 36 sampling 31/2000ml at 0.49ml/s with 25.0psi vacuum @00:06:44.19 Cartridge 36 sampling 60/2000ml at 0.50ml/s with 25.5psi vacuum @00:07:44.41 Cartridge 36 sampling 90/2000ml at 0.49ml/s with 24.9psi vacuum @00:08:44.19 Cartridge 36 sampling 120/2000ml at 0.49ml/s with 23.5psi vacuum @00:09:44.31 Cartridge 36 sampling 150/2000ml at 0.50ml/s with 24.3psi vacuum @00:10:44.19 Cartridge 36 sampling 180/2000ml at 0.51ml/s with 22.3psi vacuum @00:11:44.26 Cartridge 36 sampling 211/2000ml at 0.51ml/s with 26.5psi vacuum @00:12:44.66 Cartridge 36 sampling 242/2000ml at 0.52ml/s with 24.9psi vacuum @00:13:44.20 Cartridge 36 sampling 272/2000ml at 0.51ml/s with 26.6psi vacuum @00:14:44.37 Cartridge 36 sampling 303/2000ml at 0.51ml/s with 26.3psi vacuum @00:15:44.19 Cartridge 36 sampling 334/2000ml at 0.52ml/s with 24.3psi vacuum @00:16:44.42 Cartridge 36 sampling 364/2000ml at 0.48ml/s with 26.8psi vacuum @00:17:44.17 Cartridge 36 sampling 394/2000ml at 0.48ml/s with 26.4psi vacuum @00:18:44.29 Cartridge 36 sampling 424/2000ml at 0.49ml/s with 26.4psi vacuum @00:19:44.18 Cartridge 36 sampling 453/2000ml at 0.49ml/s with 22.4psi vacuum @00:20:44.34 Cartridge 36 sampling 483/2000ml at 0.50ml/s with 25.3psi vacuum @00:21:44.18 Cartridge 36 sampling 513/2000ml at 0.49ml/s with 26.1psi vacuum @00:22:44.23 Cartridge 36 sampling 542/2000ml at 0.49ml/s with 25.7psi vacuum @00:23:44.18 Cartridge 36 sampling 572/2000ml at 0.48ml/s with 25.8psi vacuum @00:24:44.25 Cartridge 36 sampling 601/2000ml at 0.49ml/s with 26.5psi vacuum @00:25:44.51 Cartridge 36 sampling 630/2000ml at 0.49ml/s with 25.9psi vacuum @00:26:44.19 Cartridge 36 sampling 659/2000ml at 0.48ml/s with 24.3psi vacuum @00:27:44.35 Cartridge 36 sampling 688/2000ml at 0.49ml/s with 18.7psi vacuum 2.5m deep (18.4psia) @00:28:44.19 Cartridge 36 sampling 717/2000ml at 0.48ml/s with 23.6psi vacuum @00:29:44.30 Cartridge 36 sampling 745/2000ml at 0.48ml/s with 18.9psi vacuum 2.5m deep (18.4psia) @00:30:44.19 Cartridge 36 sampling 774/2000ml at 0.47ml/s with 19.4psi vacuum @00:31:44.31 Cartridge 36 sampling 802/2000ml at 0.46ml/s with 16.8psi vacuum 2.9m deep (19.0psia) @00:32:44.19 Cartridge 36 sampling 831/2000ml at 0.45ml/s with 23.4psi vacuum @00:33:44.39 Cartridge 36 sampling 859/2000ml at 0.48ml/s with 26.5psi vacuum @00:34:44.19 Cartridge 36 sampling 887/2000ml at 0.46ml/s with 25.5psi vacuum @00:35:44.22 Cartridge 36 sampling 914/2000ml at 0.46ml/s with 27.1psi vacuum @00:36:44.18 Cartridge 36 sampling 942/2000ml at 0.46ml/s with 25.5psi vacuum @00:37:44.60 Cartridge 36 sampling 970/2000ml at 0.44ml/s with 27.4psi vacuum @00:38:44.70 Cartridge 36 sampling 997/2000ml at 0.46ml/s with 25.5psi vacuum @00:39:44.19 Cartridge 36 sampling 1024/2000ml at 0.43ml/s with 16.9psi vacuum 1.8m deep (17.3psia) @00:40:44.65 Cartridge 36 sampling 1051/2000ml at 0.46ml/s with 26.1psi vacuum @00:41:44.18 Cartridge 36 sampling 1078/2000ml at 0.44ml/s with 17.9psi vacuum @00:42:44.34 Cartridge 36 sampling 1105/2000ml at 0.44ml/s with 25.7psi vacuum @00:43:44.18 Cartridge 36 sampling 1132/2000ml at 0.45ml/s with 26.1psi vacuum @00:44:44.35 Cartridge 36 sampling 1158/2000ml at 0.45ml/s with 23.1psi vacuum @00:45:44.18 Cartridge 36 sampling 1185/2000ml at 0.44ml/s with 27.3psi vacuum @00:46:44.40 Cartridge 36 sampling 1211/2000ml at 0.44ml/s with 17.8psi vacuum @00:47:44.18 Cartridge 36 sampling 1237/2000ml at 0.43ml/s with 26.3psi vacuum @00:48:44.35 Cartridge 36 sampling 1263/2000ml at 0.43ml/s with 24.5psi vacuum @00:49:44.18 Cartridge 36 sampling 1289/2000ml at 0.43ml/s with 26.1psi vacuum @00:50:44.34 Cartridge 36 sampling 1314/2000ml at 0.43ml/s with 22.4psi vacuum @00:51:44.18 Cartridge 36 sampling 1340/2000ml at 0.43ml/s with 19.8psi vacuum 1.7m deep (17.2psia) @00:52:44.32 Cartridge 36 sampling 1365/2000ml at 0.43ml/s with 24.9psi vacuum @00:53:44.62 Cartridge 36 sampling 1391/2000ml at 0.40ml/s with 22.9psi vacuum @00:54:44.19 Cartridge 36 sampling 1416/2000ml at 0.40ml/s with 26.7psi vacuum @00:55:44.34 Cartridge 36 sampling 1441/2000ml at 0.40ml/s with 27.3psi vacuum @00:56:44.19 Cartridge 36 sampling 1466/2000ml at 0.40ml/s with 26.8psi vacuum @00:57:44.33 Cartridge 36 sampling 1490/2000ml at 0.41ml/s with 25.5psi vacuum @00:58:44.19 Cartridge 36 sampling 1515/2000ml at 0.40ml/s with 21.8psi vacuum @00:59:44.28 Cartridge 36 sampling 1539/2000ml at 0.42ml/s with 23.5psi vacuum @01:00:44.19 Cartridge 36 sampling 1563/2000ml at 0.40ml/s with 19.1psi vacuum @01:01:44.45 Cartridge 36 sampling 1588/2000ml at 0.42ml/s with 25.1psi vacuum @01:02:44.19 Cartridge 36 sampling 1612/2000ml at 0.41ml/s with 24.4psi vacuum @01:03:44.33 Cartridge 36 sampling 1635/2000ml at 0.38ml/s with 23.9psi vacuum @01:04:44.19 Cartridge 36 sampling 1659/2000ml at 0.41ml/s with 26.5psi vacuum @01:05:44.40 Cartridge 36 sampling 1683/2000ml at 0.40ml/s with 24.5psi vacuum @01:06:44.33 Cartridge 36 sampling 1706/2000ml at 0.39ml/s with 23.8psi vacuum @01:07:44.18 Cartridge 36 sampling 1729/2000ml at 0.39ml/s with 22.5psi vacuum @01:08:44.33 Cartridge 36 sampling 1752/2000ml at 0.37ml/s with 27.1psi vacuum @01:09:44.19 Cartridge 36 sampling 1775/2000ml at 0.35ml/s with 27.4psi vacuum @01:10:44.48 Cartridge 36 sampling 1797/2000ml at 0.38ml/s with 25.2psi vacuum @01:11:44.17 Cartridge 36 sampling 1820/2000ml at 0.38ml/s with 25.4psi vacuum @01:12:44.35 Cartridge 36 sampling 1842/2000ml at 0.35ml/s with 25.3psi vacuum @01:13:44.18 Cartridge 36 sampling 1864/2000ml at 0.38ml/s with 18.2psi vacuum 2.2m deep (17.9psia) @01:14:44.35 Cartridge 36 sampling 1886/2000ml at 0.35ml/s with 17.7psi vacuum 1.6m deep (17.0psia) @01:15:44.18 Cartridge 36 sampling 1908/2000ml at 0.35ml/s with 27.5psi vacuum @01:16:44.37 Cartridge 36 sampling 1929/2000ml at 0.36ml/s with 22.4psi vacuum @01:17:44.18 Cartridge 36 sampling 1951/2000ml at 0.35ml/s with 26.9psi vacuum @01:18:44.35 Cartridge 36 sampling 1973/2000ml at 0.36ml/s with 22.9psi vacuum @01:19:44.18 Cartridge 36 sampling 1994/2000ml at 0.37ml/s with 23.7psi vacuum @01:20:01.07 Pressure.coast @01:20:01.21 Intake.close @01:20:01.27 Exhaust.close @01:20:01.45 Toroid.reconfigure :hold @01:20:01.51 Toroid.seek :bypass @01:20:03.12 Toroid.reconfigure :default @01:20:03.18 Toroid.seek :clear @01:20:04.87 Sampled 2000.0ml @01:20:05.97 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 36" @01:20:06.07 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:20:06.12 Retry #1 of 10 after 10:00 @01:20:07.05 SamplePump.setPosition! 0ml @01:20:07.11 Plunge.seek :reagent @01:20:07.19 Bypass.open @01:20:08.09 Intake.open @01:20:08.22 Sample loop closed. Ready to Deliver Reagents @01:20:08.29 Archive.seek :sealed @01:20:08.98 SamplePump.reconfigure :slow @01:20:09.04 SamplePump.seek -3ml @01:20:11.96 Archive.seek :evac @01:20:15.81 delay 1 second @01:20:17.11 SamplePump.seek -2.8ml @01:20:17.92 delay 1 second @01:20:19.05 SamplePump.seek -3.3ml @01:20:20.38 delay 1 second @01:20:21.41 Exhaust.open @01:20:22.10 Bypass.close @01:20:23.15 SamplePump.seek -35.3ml @01:20:25.99 delay 1:00 @01:21:26.02 Archive.seek :clear @01:21:31.91 delay 3 seconds @01:21:35.00 Intake.close @01:21:35.05 Exhaust.close @01:21:43.16 RNAlater soak for 10:00 @01:30:40.61 upsync @01:31:43.21 Archive.seek :tVent @01:31:44.89 delay 1 second @01:31:45.95 Archive.seek :evac @01:32:01.92 delay 10 seconds @01:32:12.01 Sample preserved and evacuated @01:32:13.61
Carousel.seek 36 @01:32:14.46 Plunge.seek :locked @01:32:15.50 Archive.seek :clear @01:32:15.83 Air.seek :clear @01:32:16.21 Lysis1.seek :clear @01:32:32.80
Archive.coast @01:32:32.92 Air.coast @01:32:33.04 Lysis1.coast @01:32:33.09 Lysis2.coast @01:32:33.25 Plunge.coast @01:32:33.34 SamplePump.coast @01:32:33.54 Toroid.coast @01:32:33.59 Carousel.coast @01:32:33.94 Gate.power :motor,:OFF @01:32:33.98 Gate.power :core,:OFF @01:32:34.08 Gate.power :analytic1,:OFF @01:32:34.10 Gate.power :analytic2,:OFF @01:32:34.11 Gate.power :analytic3,:OFF @01:32:34.13 Gate.power :main,:OFF @01:32:34.14 Safely stopped and ready to power off @01:32:34.45 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @01:32:34.48
delayUntil "12:00:00.00" @01:32:55.81 upsync @12:51:28.64PDT19-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @12:51:28.67 >BREAK< ESP3 Gateway v1.11 for cartridge bus @12:51:31.88 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @12:51:32.20 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @12:51:41.19 Polling Can every 2:00 @12:51:41.25 Loading /home/esp/esp2/bin/processAll mission @12:51:42.67 Sampling default goal volumes @12:51:42.71 GoodNews.email "Cartridge group: dry\nSpares: {spare: 1}\nSchedule: [\"1pm\", \"1am\"]\n",Subject: "Begun processAll --schedule=1pm,1am --checkpoint --mail" @12:51:42.83
delayUntil "13:00:00.00" @12:51:57.57 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 13:00" @12:52:02.91 upsync @12:52:12.47 upsync @13:00:00.28
Gate.power :main,:ON,limit to 5000mA @13:00:00.31 Gate.power :analytic1,:ON,limit to 5000mA @13:00:00.32 Gate.power :analytic2,:ON,limit to 5000mA @13:00:00.34 Gate.power :analytic3,:ON,limit to 5000mA @13:00:00.46 Powering core bus @13:00:03.03 Archive.configure :default @13:00:03.52 Air.configure :default @13:00:03.82 Lysis1.configure :default @13:00:03.93 Lysis2.configure :default @13:00:04.06 Plunge.configure :default @13:00:04.19 SamplePump.configure :default @13:00:04.33 ExtraPump.configure :default @13:00:04.47 Toroid.configure :default @13:00:04.60 Carousel.configure :default @13:00:04.74 Pressure.configure :default @13:00:04.79 All elf controllers running firmware version 4.61 @13:00:04.82 Gate.power :motor,:ON,limit to 5000mA @13:00:05.15 Archive.reconfigure :default @13:00:05.43 Toroid.reconfigure :default @13:00:05.51 Archive.home.jog 0 @13:00:05.54 Lysis2.reconfigure :default @13:00:05.57 Lysis1.reconfigure :default @13:00:05.63 Air.reconfigure :default @13:00:05.67 Plunge.reconfigure :default @13:00:05.91 Lysis2.home.jog 0 @13:00:05.96 Lysis1.home.jog 0 @13:00:06.00 Toroid.home.jog 0 @13:00:06.03 Air.home.jog 0 @13:00:06.09 Plunge.home.jog 0 @13:00:06.79 Toroid.reconfigure :homing @13:00:06.82 Lysis2.reconfigure :homing @13:00:06.86 Air.reconfigure :homing @13:00:06.89 Archive.reconfigure :homing @13:00:06.94 Lysis1.reconfigure :homing @13:00:07.14 Toroid.jog 632 @13:00:07.18 Air.jog 1364 @13:00:07.21 Lysis2.jog 1358 @13:00:07.23 Archive.jog 1372 @13:00:07.26 Lysis1.jog 1349 @13:00:07.53 Plunge.reconfigure :homing @13:00:07.58 Plunge.jog 849 @13:00:08.70 Lysis2.setRawPosition! 50 @13:00:08.77 Air.setRawPosition! 50 @13:00:09.02 Air.seek :home @13:00:09.07 Lysis2.seek :home @13:00:09.94 Plunge.setRawPosition! 50 @13:00:10.13 Lysis2.reconfigure :default @13:00:10.17 Air.reconfigure :default @13:00:10.21 Archive.setRawPosition! 50 @13:00:10.32 Toroid.setRawPosition! 0 @13:00:10.36 Lysis1.setRawPosition! 50 @13:00:10.96 Plunge.seek :home @13:00:11.17 Archive.seek :home @13:00:11.20 Toroid.seek :home @13:00:11.26 Lysis1.seek :home @13:00:12.16 Toroid.reconfigure :default @13:00:12.29 Plunge.reconfigure :default @13:00:12.37 Archive.reconfigure :default @13:00:12.43 Lysis1.reconfigure :default @13:00:13.11
Toroid.seek :clear @13:00:14.76 Carousel.reconfigure :default @13:00:14.81 Carousel.home.jog 0 @13:00:15.30 Carousel.seek 36 @13:00:15.52 Selecting Cartridge 35 @13:00:15.73 Carousel.seek 35 @13:00:16.25 Lysis1.seek :locked @13:00:19.67 Plunge.seek :locked @13:00:20.02 Air.seek :locked @13:00:24.37
Mounted archive Cartridge @13:00:24.49 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 35" @13:00:24.82 Duration of filtering limited to 4:05:00 @13:00:25.13 Priming sample loop w/160ml, bypass w/40ml @13:00:25.15 SamplePump.setPosition! 0ml @13:00:25.19 Bypass.open @13:00:26.68 Intake.open @13:00:28.01 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @13:00:30.73 Vacuum tare is 18% of full scale @13:00:30.83 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @13:00:31.70 Pressure.reconfigure :default @13:00:31.76 Pressure.hold @13:00:32.13 SamplePump.reconfigure :prime @13:00:33.40 SamplePump.seek 40ml,3:51 @13:00:42.53 upsync @13:01:16.64 Pressure.coast @13:01:16.80 SamplePump.setPosition! 0ml @13:01:16.92 Exhaust.open @13:01:18.55 Bypass.close @13:01:20.18 Pressure.reconfigure :default @13:01:20.25 Pressure.hold @13:01:20.54 SamplePump.reconfigure :prime @13:01:20.60 SamplePump.seek 160ml,13:51 @13:04:12.29 Pressure.coast @13:04:12.39 Bypass.open @13:04:13.19 Intake.close @13:04:14.18 Toroid.reconfigure :hold @13:04:14.24 Toroid.seek :crack @13:04:15.83 Toroid.seek :cartridge @13:04:16.76 Toroid.reconfigure :default @13:04:16.82 Toroid.seek :clear @13:04:19.19 Pressurizing cartridge to 26psi above 14.5psia @13:04:19.25 SamplePump.setPosition! 0ml @13:04:19.44 Intake.open @13:04:20.28 Bypass.close @13:04:21.11 Exhaust.close @13:04:22.25 Pressure.reconfigure :default @13:04:22.37 Pressure.hold @13:04:23.23 SamplePump.reconfigure :charge @13:04:23.29 SamplePump.seek 2.3ml,42 seconds @13:04:35.23 Pressure.coast @13:04:35.28 Intake.close @13:04:36.14 Pressurized cartridge with 0.652ml to 26.2psi @13:04:36.15 Isolated @36.4psia (21.9psi) @13:04:36.16 Checking for cartridge leaks from 36.4psia @13:04:36.18 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @13:04:45.78 Cartridge pressure fell 8% to 33.3psia @13:04:45.86 Sampling 2000.0ml @13:04:45.88 SamplePump.setPosition! 0ml @13:04:45.92 Exhaust.open @13:04:46.64 Intake.open @13:04:47.80 Pressure.reconfigure :default @13:04:47.93 Pressure.hold @13:04:48.91 SamplePump.reconfigure :sample @13:04:49.08 SamplePump.seek 2000ml,2:47:12 @13:04:50.99 Cartridge 35 sampling 0/2000ml at 0.58ml/s with 24.6psi vacuum @13:05:50.77 Cartridge 35 sampling 29/2000ml at 0.47ml/s with 19.8psi vacuum @13:06:50.90 Cartridge 35 sampling 57/2000ml at 0.45ml/s with 18.9psi vacuum @13:07:50.77 Cartridge 35 sampling 85/2000ml at 0.47ml/s with 22.4psi vacuum @13:08:50.85 Cartridge 35 sampling 113/2000ml at 0.47ml/s with 21.0psi vacuum @13:09:50.91 Cartridge 35 sampling 141/2000ml at 0.49ml/s with 19.2psi vacuum 2.8m deep (18.8psia) @13:10:50.78 Cartridge 35 sampling 171/2000ml at 0.48ml/s with 26.2psi vacuum @13:11:51.23 Cartridge 35 sampling 200/2000ml at 0.47ml/s with 16.4psi vacuum @13:12:50.78 Cartridge 35 sampling 228/2000ml at 0.48ml/s with 26.9psi vacuum @13:13:51.00 Cartridge 35 sampling 257/2000ml at 0.50ml/s with 21.4psi vacuum 2.7m deep (18.6psia) @13:14:50.78 Cartridge 35 sampling 286/2000ml at 0.46ml/s with 28.7psi vacuum @13:15:50.93 Cartridge 35 sampling 315/2000ml at 0.49ml/s with 16.7psi vacuum 4.7m deep (21.6psia) @13:16:50.78 Cartridge 35 sampling 344/2000ml at 0.49ml/s with 17.4psi vacuum 3.3m deep (19.6psia) @13:17:50.94 Cartridge 35 sampling 373/2000ml at 0.48ml/s with 26.7psi vacuum @13:18:50.78 Cartridge 35 sampling 402/2000ml at 0.50ml/s with 22.0psi vacuum 1.8m deep (17.4psia) @13:19:50.87 Cartridge 35 sampling 431/2000ml at 0.48ml/s with 23.0psi vacuum @13:20:50.78 Cartridge 35 sampling 459/2000ml at 0.47ml/s with 24.2psi vacuum @13:21:51.09 Cartridge 35 sampling 487/2000ml at 0.50ml/s with 21.5psi vacuum @13:22:51.15 Cartridge 35 sampling 516/2000ml at 0.49ml/s with 24.2psi vacuum @13:23:50.77 Cartridge 35 sampling 544/2000ml at 0.47ml/s with 15.1psi vacuum 4.6m deep (21.5psia) @13:24:51.47 Cartridge 35 sampling 572/2000ml at 0.44ml/s with 30.1psi vacuum @13:25:50.77 Cartridge 35 sampling 600/2000ml at 0.48ml/s with 21.2psi vacuum 1.9m deep (17.5psia) @13:26:51.08 Cartridge 35 sampling 628/2000ml at 0.45ml/s with 23.0psi vacuum 1.8m deep (17.4psia) @13:27:50.77 Cartridge 35 sampling 655/2000ml at 0.45ml/s with 27.3psi vacuum @13:28:50.89 Cartridge 35 sampling 683/2000ml at 0.43ml/s with 15.7psi vacuum @13:29:50.77 Cartridge 35 sampling 710/2000ml at 0.48ml/s with 25.6psi vacuum @13:30:50.90 Cartridge 35 sampling 737/2000ml at 0.43ml/s with 30.4psi vacuum @13:31:50.77 Cartridge 35 sampling 764/2000ml at 0.47ml/s with 18.3psi vacuum 3.4m deep (19.7psia) @13:32:50.99 Cartridge 35 sampling 791/2000ml at 0.43ml/s with 13.7psi vacuum 3.8m deep (20.3psia) @13:33:50.77 Cartridge 35 sampling 817/2000ml at 0.42ml/s with 26.8psi vacuum @13:34:51.00 Cartridge 35 sampling 844/2000ml at 0.46ml/s with 24.4psi vacuum @13:35:50.77 Cartridge 35 sampling 870/2000ml at 0.42ml/s with 30.3psi vacuum @13:36:50.82 Cartridge 35 sampling 896/2000ml at 0.42ml/s with 30.4psi vacuum @13:37:50.77 Cartridge 35 sampling 922/2000ml at 0.43ml/s with 14.4psi vacuum 2.0m deep (17.7psia) @13:38:51.13 Cartridge 35 sampling 948/2000ml at 0.43ml/s with 28.9psi vacuum @13:39:51.31 Cartridge 35 sampling 974/2000ml at 0.42ml/s with 30.1psi vacuum @13:40:50.78 Cartridge 35 sampling 999/2000ml at 0.42ml/s with 28.5psi vacuum @13:41:51.13 Cartridge 35 sampling 1024/2000ml at 0.43ml/s with 19.4psi vacuum @13:42:50.78 Cartridge 35 sampling 1050/2000ml at 0.43ml/s with 26.9psi vacuum @13:43:51.01 Cartridge 35 sampling 1075/2000ml at 0.42ml/s with 27.3psi vacuum @13:44:50.78 Cartridge 35 sampling 1099/2000ml at 0.43ml/s with 20.6psi vacuum 2.0m deep (17.7psia) @13:45:50.91 Cartridge 35 sampling 1124/2000ml at 0.43ml/s with 20.8psi vacuum 3.7m deep (20.1psia) @13:46:50.78 Cartridge 35 sampling 1148/2000ml at 0.40ml/s with 29.9psi vacuum @13:47:50.93 Cartridge 35 sampling 1173/2000ml at 0.41ml/s with 15.4psi vacuum 4.6m deep (21.4psia) @13:48:50.78 Cartridge 35 sampling 1197/2000ml at 0.37ml/s with 26.6psi vacuum @13:49:50.96 Cartridge 35 sampling 1221/2000ml at 0.36ml/s with 26.6psi vacuum @13:50:50.78 Cartridge 35 sampling 1244/2000ml at 0.39ml/s with 30.0psi vacuum @13:51:50.90 Cartridge 35 sampling 1268/2000ml at 0.39ml/s with 29.7psi vacuum @13:52:50.78 Cartridge 35 sampling 1292/2000ml at 0.39ml/s with 11.7psi vacuum 4.7m deep (21.6psia) @13:53:51.04 Cartridge 35 sampling 1315/2000ml at 0.40ml/s with 26.3psi vacuum @13:54:51.14 Cartridge 35 sampling 1339/2000ml at 0.41ml/s with 26.0psi vacuum @13:55:50.77 Cartridge 35 sampling 1362/2000ml at 0.38ml/s with 11.1psi vacuum 5.6m deep (23.0psia) @13:56:51.09 Cartridge 35 sampling 1385/2000ml at 0.35ml/s with 29.0psi vacuum @13:57:50.77 Cartridge 35 sampling 1408/2000ml at 0.39ml/s with 23.5psi vacuum @13:58:50.92 Cartridge 35 sampling 1430/2000ml at 0.37ml/s with 29.9psi vacuum @13:59:50.77 Cartridge 35 sampling 1453/2000ml at 0.39ml/s with 25.0psi vacuum @14:00:52.10 Cartridge 35 sampling 1475/2000ml at 0.33ml/s with 23.4psi vacuum @14:01:50.77 Cartridge 35 sampling 1497/2000ml at 0.34ml/s with 21.4psi vacuum @14:02:50.94 Cartridge 35 sampling 1520/2000ml at 0.38ml/s with 27.9psi vacuum @14:03:50.77 Cartridge 35 sampling 1542/2000ml at 0.35ml/s with 26.3psi vacuum @14:04:50.98 Cartridge 35 sampling 1563/2000ml at 0.36ml/s with 29.7psi vacuum @14:05:50.77 Cartridge 35 sampling 1585/2000ml at 0.38ml/s with 26.5psi vacuum @14:06:50.82 Cartridge 35 sampling 1606/2000ml at 0.37ml/s with 27.7psi vacuum @14:07:50.77 Cartridge 35 sampling 1627/2000ml at 0.36ml/s with 12.2psi vacuum 5.4m deep (22.6psia) @14:08:51.09 Cartridge 35 sampling 1648/2000ml at 0.33ml/s with 29.9psi vacuum @14:09:51.46 Cartridge 35 sampling 1670/2000ml at 0.36ml/s with 28.2psi vacuum @14:10:51.00 Cartridge 35 sampling 1690/2000ml at 0.35ml/s with 29.0psi vacuum @14:11:51.14 Cartridge 35 sampling 1711/2000ml at 0.31ml/s with 18.9psi vacuum @14:12:50.78 Cartridge 35 sampling 1732/2000ml at 0.34ml/s with 12.8psi vacuum 2.9m deep (19.0psia) @14:13:50.93 Cartridge 35 sampling 1752/2000ml at 0.36ml/s with 22.5psi vacuum @14:14:50.78 Cartridge 35 sampling 1773/2000ml at 0.35ml/s with 27.1psi vacuum @14:15:50.95 Cartridge 35 sampling 1793/2000ml at 0.34ml/s with 29.0psi vacuum @14:16:50.78 Cartridge 35 sampling 1813/2000ml at 0.34ml/s with 11.1psi vacuum 4.5m deep (21.3psia) @14:17:50.92 Cartridge 35 sampling 1833/2000ml at 0.32ml/s with 30.3psi vacuum @14:18:50.78 Cartridge 35 sampling 1853/2000ml at 0.32ml/s with 28.2psi vacuum @14:19:50.92 Cartridge 35 sampling 1873/2000ml at 0.32ml/s with 12.2psi vacuum 2.3m deep (18.1psia) @14:20:50.78 Cartridge 35 sampling 1892/2000ml at 0.35ml/s with 22.9psi vacuum @14:21:50.93 Cartridge 35 sampling 1912/2000ml at 0.35ml/s with 25.3psi vacuum @14:22:50.78 Cartridge 35 sampling 1931/2000ml at 0.29ml/s with 19.8psi vacuum @14:23:51.14 Cartridge 35 sampling 1950/2000ml at 0.33ml/s with 28.8psi vacuum @14:24:50.78 Cartridge 35 sampling 1969/2000ml at 0.34ml/s with 19.3psi vacuum @14:25:50.78 Cartridge 35 sampling 1988/2000ml at 0.28ml/s with 26.3psi vacuum @14:26:28.29 Pressure.coast @14:26:28.43 Intake.close @14:26:28.51 Exhaust.close @14:26:28.67 Toroid.reconfigure :hold @14:26:28.73 Toroid.seek :bypass @14:26:30.35 Toroid.reconfigure :default @14:26:30.41 Toroid.seek :clear @14:26:32.06 Sampled 2000.0ml @14:26:33.47 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 35" @14:26:34.52 SamplePump.setPosition! 0ml @14:26:34.58 Plunge.seek :reagent @14:26:34.73 Bypass.open @14:26:35.78 Intake.open @14:26:35.81 Sample loop closed. Ready to Deliver Reagents @14:26:35.93 Archive.seek :sealed @14:26:36.81 SamplePump.reconfigure :slow @14:26:36.98 SamplePump.seek -3ml @14:26:39.69 Archive.seek :evac @14:26:43.94 delay 1 second @14:26:45.21 SamplePump.seek -2.8ml @14:26:46.08 delay 1 second @14:26:47.21 SamplePump.seek -3.3ml @14:26:48.54 delay 1 second @14:26:49.57 Exhaust.open @14:26:50.28 Bypass.close @14:26:51.05 SamplePump.seek -35.3ml @14:26:51.61 upsync @14:26:53.72 delay 1:00 @14:27:53.76 Archive.seek :clear @14:27:59.78 delay 3 seconds @14:28:02.87 Intake.close @14:28:02.94 Exhaust.close @14:28:10.87 RNAlater soak for 10:00 @14:38:10.95 Archive.seek :tVent @14:38:12.81 delay 1 second @14:38:13.91 Archive.seek :evac @14:38:29.89 delay 10 seconds @14:38:39.93 Sample preserved and evacuated @14:38:40.48
Carousel.seek 35 @14:38:40.88 Plunge.seek :locked @14:38:42.03 Archive.seek :clear @14:38:42.32 Air.seek :clear @14:38:42.59 Lysis1.seek :clear @14:38:59.29
Archive.coast @14:38:59.39 Air.coast @14:38:59.51 Lysis1.coast @14:38:59.61 Lysis2.coast @14:38:59.68 Plunge.coast @14:38:59.88 SamplePump.coast @14:38:59.96 Toroid.coast @14:39:00.12 Carousel.coast @14:39:00.45 Gate.power :motor,:OFF @14:39:00.46 Gate.power :core,:OFF @14:39:00.47 Gate.power :analytic1,:OFF @14:39:00.49 Gate.power :analytic2,:OFF @14:39:00.50 Gate.power :analytic3,:OFF @14:39:00.51 Gate.power :main,:OFF @14:39:00.52 Safely stopped and ready to power off @14:39:00.98 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 01:00PDT20-Jun-21" @14:39:01.04
delayUntil "01:00:00.00PDT20-Jun-21" @14:39:18.24 upsync @20:41:51.52 Session Begun @20:41:57.52 -> Cartridge.dry @20:42:16.58 -> MainThread.exit @20:42:16.68 Session Ended @20:43:53.04PDT19-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @20:43:53.14 >BREAK< ESP3 Gateway v1.11 for cartridge bus @20:43:58.55 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @20:43:58.85 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @20:44:07.18 Polling Can every 2:00 @20:44:07.47 Loading /home/esp/esp2/bin/processAll mission @20:44:08.81 Sampling default goal volumes @20:44:08.85 GoodNews.email "Cartridge group: dry\nSpares: {spare: 1}\nSchedule: [\"12am\", \"12pm\"]\n",Subject: "Begun processAll --schedule=12am,12pm --checkpoint --mail" @20:44:08.95 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @20:44:08.99 Retry #1 of 10 after 10:00 @20:44:09.01
delayUntil "00:00:00.00PDT20-Jun-21" @20:54:09.18 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @20:54:09.28 Retry #2 of 10 after 10:00 @21:04:09.51 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:04:09.56 Retry #3 of 10 after 10:00 @21:14:09.62 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:14:09.64 Retry #4 of 10 after 10:00 @21:24:09.70 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:24:09.72 Retry #5 of 10 after 10:00 @21:34:09.77 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:34:09.79 Retry #6 of 10 after 10:00 @21:44:09.84 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:44:09.91 Retry #7 of 10 after 10:00 @21:54:09.96 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:54:09.98 Retry #8 of 10 after 10:00 @22:04:11.31 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @22:04:11.47 Retry #9 of 10 after 10:00 @22:14:11.54 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @22:14:11.56 Retry #10 of 10 after 10:00 @22:24:11.62 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @22:24:11.63 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT20-Jun-21" @22:24:11.66 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @22:24:11.68 Retry #1 of 10 after 10:00 @22:34:11.87 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @22:34:11.91 Retry #2 of 10 after 10:00 @22:44:11.96 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @22:44:11.98 Retry #3 of 10 after 10:00 @22:54:12.09 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @22:54:12.19 Retry #4 of 10 after 10:00 @23:04:12.28 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @23:04:12.30 Retry #5 of 10 after 10:00 @23:14:12.35 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @23:14:12.37 Retry #6 of 10 after 10:00 @23:24:12.44 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @23:24:12.50 Retry #7 of 10 after 10:00 @23:34:12.64 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @23:34:12.66 Retry #8 of 10 after 10:00 @23:44:12.78 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @23:44:12.81 Retry #9 of 10 after 10:00 @23:54:12.86 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @23:54:12.88 Retry #10 of 10 after 10:00 @00:00:00.12PDT20-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.13 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.15 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.16 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.20 Powering core bus @00:00:02.48 Archive.configure :default @00:00:02.62 Air.configure :default @00:00:02.73 Lysis1.configure :default @00:00:02.84 Lysis2.configure :default @00:00:02.98 Plunge.configure :default @00:00:03.11 SamplePump.configure :default @00:00:03.25 ExtraPump.configure :default @00:00:03.38 Toroid.configure :default @00:00:03.53 Carousel.configure :default @00:00:03.67 Pressure.configure :default @00:00:03.72 All elf controllers running firmware version 4.61 @00:00:03.75 Gate.power :motor,:ON,limit to 5000mA @00:00:04.34 Archive.reconfigure :default @00:00:04.97 Archive.home.jog 0 @00:00:05.81 Toroid.reconfigure :default @00:00:05.91 Lysis2.reconfigure :default @00:00:06.06 Lysis1.reconfigure :default @00:00:06.29 Air.reconfigure :default @00:00:06.32 Plunge.reconfigure :default @00:00:06.84 Lysis2.home.jog 0 @00:00:06.98 Toroid.home.jog 0 @00:00:07.12 Lysis1.home.jog 0 @00:00:07.20 Air.home.jog 0 @00:00:07.46 Plunge.home.jog 0 @00:00:08.85 Archive.reconfigure :homing @00:00:09.18 Archive.jog 1373 @00:00:09.25 Toroid.reconfigure :homing @00:00:09.33 Lysis1.reconfigure :homing @00:00:09.46 Plunge.reconfigure :homing @00:00:09.53 Air.reconfigure :homing @00:00:09.70 Lysis2.reconfigure :homing @00:00:10.16 Toroid.jog 629 @00:00:10.20 Lysis1.jog 1350 @00:00:10.23 Plunge.jog 844 @00:00:10.27 Lysis2.jog 1358 @00:00:10.29 Air.jog 1362 @00:00:10.86 Archive.setRawPosition! 50 @00:00:10.91 Archive.seek :home @00:00:11.19 Archive.reconfigure :default @00:00:11.31 Plunge.setRawPosition! 50 @00:00:11.36 Plunge.seek :home @00:00:11.51 Lysis1.setRawPosition! 50 @00:00:11.58 Lysis1.seek :home @00:00:11.70 Plunge.reconfigure :default @00:00:11.73 Lysis2.setRawPosition! 50 @00:00:11.76 Air.setRawPosition! 50 @00:00:11.89 Lysis2.seek :home @00:00:11.94 Air.seek :home @00:00:12.09 Lysis1.reconfigure :default @00:00:12.24 Lysis2.reconfigure :default @00:00:12.34 Air.reconfigure :default @00:00:12.44 Toroid.setRawPosition! 0 @00:00:12.49 Toroid.seek :home @00:00:12.58 Toroid.reconfigure :default @00:00:12.73
Toroid.seek :clear @00:00:14.48 Carousel.reconfigure :default @00:00:14.63 Carousel.home.jog 0 @00:00:15.67 Carousel.seek 35 @00:00:16.03 Selecting Cartridge 34 @00:00:16.57 Carousel.seek 34 @00:00:17.46 Lysis1.seek :locked @00:00:20.70 Plunge.seek :locked @00:00:20.85 Air.seek :locked @00:00:25.29
Mounted archive Cartridge @00:00:26.27 Duration of filtering limited to 4:05:00 @00:00:27.05 Priming sample loop w/160ml, bypass w/40ml @00:00:27.07 SamplePump.setPosition! 0ml @00:00:27.26 Bypass.open @00:00:28.83 Intake.open @00:00:29.76 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:32.34 Vacuum tare is 18% of full scale @00:00:32.36 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:32.57 Pressure.reconfigure :default @00:00:32.63 Pressure.hold @00:00:32.84 SamplePump.reconfigure :prime @00:00:32.90 SamplePump.seek 40ml,3:51 @00:01:16.11 Pressure.coast @00:01:16.15 SamplePump.setPosition! 0ml @00:01:16.20 Exhaust.open @00:01:17.62 Bypass.close @00:01:19.16 Pressure.reconfigure :default @00:01:19.22 Pressure.hold @00:01:19.66 SamplePump.reconfigure :prime @00:01:19.80 SamplePump.seek 160ml,13:51 @00:04:11.52 Pressure.coast @00:04:11.99 Bypass.open @00:04:12.80 Intake.close @00:04:12.93 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:04:12.95 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 34" @00:04:12.98 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:04:13.00 Retry #1 of 10 after 10:00 @00:04:13.68 Toroid.reconfigure :hold @00:04:13.74 Toroid.seek :crack @00:04:15.44 Toroid.seek :cartridge @00:04:16.45 Toroid.reconfigure :default @00:04:16.55 Toroid.seek :clear @00:04:18.77 Pressurizing cartridge to 26psi above 14.5psia @00:04:18.82 SamplePump.setPosition! 0ml @00:04:19.04 Intake.open @00:04:19.84 Bypass.close @00:04:20.54 Exhaust.close @00:04:21.37 Pressure.reconfigure :default @00:04:21.43 Pressure.hold @00:04:21.64 SamplePump.reconfigure :charge @00:04:21.70 SamplePump.seek 2.3ml,42 seconds @00:04:32.96 Pressure.coast @00:04:33.25 Intake.close @00:04:34.23 Pressurized cartridge with 0.568ml to 26.2psi @00:04:34.28 Isolated @35.0psia (20.5psi) @00:04:34.29 Checking for cartridge leaks from 35.0psia @00:04:34.32 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:39.91 Cartridge pressure fell 6% to 32.8psia @00:04:39.99 Sampling 2000.0ml @00:04:40.01 SamplePump.setPosition! 0ml @00:04:40.05 Exhaust.open @00:04:40.77 Intake.open @00:04:41.69 Pressure.reconfigure :default @00:04:41.75 Pressure.hold @00:04:42.01 SamplePump.reconfigure :sample @00:04:42.08 SamplePump.seek 2000ml,2:47:12 @00:04:44.92 Cartridge 34 sampling 1/2000ml at 0.52ml/s with 24.5psi vacuum @00:05:44.75 Cartridge 34 sampling 32/2000ml at 0.52ml/s with 24.8psi vacuum @00:06:44.88 Cartridge 34 sampling 64/2000ml at 0.52ml/s with 20.9psi vacuum @00:07:44.75 Cartridge 34 sampling 95/2000ml at 0.52ml/s with 25.3psi vacuum @00:08:44.83 Cartridge 34 sampling 126/2000ml at 0.52ml/s with 26.7psi vacuum @00:09:44.75 Cartridge 34 sampling 158/2000ml at 0.54ml/s with 25.8psi vacuum @00:10:45.04 Cartridge 34 sampling 190/2000ml at 0.53ml/s with 22.2psi vacuum @00:11:45.30 Cartridge 34 sampling 222/2000ml at 0.53ml/s with 24.2psi vacuum @00:12:44.76 Cartridge 34 sampling 254/2000ml at 0.53ml/s with 23.8psi vacuum @00:13:45.13 Cartridge 34 sampling 285/2000ml at 0.49ml/s with 28.1psi vacuum @00:14:13.09 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:14:13.12 Retry #2 of 10 after 10:00 @00:14:44.75 Cartridge 34 sampling 317/2000ml at 0.52ml/s with 20.8psi vacuum @00:15:44.90 Cartridge 34 sampling 348/2000ml at 0.53ml/s with 18.8psi vacuum 2.9m deep (19.0psia) @00:16:44.76 Cartridge 34 sampling 379/2000ml at 0.52ml/s with 25.3psi vacuum @00:17:44.89 Cartridge 34 sampling 411/2000ml at 0.52ml/s with 27.6psi vacuum @00:18:44.75 Cartridge 34 sampling 441/2000ml at 0.51ml/s with 23.9psi vacuum @00:19:44.92 Cartridge 34 sampling 473/2000ml at 0.51ml/s with 26.2psi vacuum @00:20:44.76 Cartridge 34 sampling 504/2000ml at 0.52ml/s with 19.1psi vacuum @00:21:44.83 Cartridge 34 sampling 535/2000ml at 0.51ml/s with 18.9psi vacuum 2.4m deep (18.2psia) @00:22:44.75 Cartridge 34 sampling 565/2000ml at 0.52ml/s with 26.0psi vacuum @00:23:44.84 Cartridge 34 sampling 596/2000ml at 0.50ml/s with 28.1psi vacuum @00:24:13.17 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:24:13.19 Retry #3 of 10 after 10:00 @00:24:44.94 Cartridge 34 sampling 626/2000ml at 0.49ml/s with 29.0psi vacuum @00:25:44.81 Cartridge 34 sampling 657/2000ml at 0.51ml/s with 25.0psi vacuum @00:26:45.11 Cartridge 34 sampling 687/2000ml at 0.50ml/s with 25.5psi vacuum @00:27:44.75 Cartridge 34 sampling 717/2000ml at 0.51ml/s with 18.5psi vacuum 2.1m deep (17.9psia) @00:28:45.06 Cartridge 34 sampling 747/2000ml at 0.50ml/s with 27.1psi vacuum @00:29:44.75 Cartridge 34 sampling 777/2000ml at 0.49ml/s with 26.9psi vacuum @00:30:44.96 Cartridge 34 sampling 807/2000ml at 0.51ml/s with 25.1psi vacuum @00:31:44.75 Cartridge 34 sampling 837/2000ml at 0.50ml/s with 21.6psi vacuum @00:32:44.93 Cartridge 34 sampling 866/2000ml at 0.49ml/s with 19.8psi vacuum 2.8m deep (18.9psia) @00:33:44.75 Cartridge 34 sampling 895/2000ml at 0.49ml/s with 23.1psi vacuum @00:34:13.24 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:34:13.26 Retry #4 of 10 after 10:00 @00:34:44.92 Cartridge 34 sampling 924/2000ml at 0.49ml/s with 21.4psi vacuum 2.3m deep (18.2psia) @00:35:44.76 Cartridge 34 sampling 953/2000ml at 0.49ml/s with 23.7psi vacuum @00:36:44.88 Cartridge 34 sampling 981/2000ml at 0.48ml/s with 21.8psi vacuum @00:37:44.77 Cartridge 34 sampling 1010/2000ml at 0.47ml/s with 26.9psi vacuum @00:38:44.82 Cartridge 34 sampling 1039/2000ml at 0.45ml/s with 17.1psi vacuum @00:39:44.93 Cartridge 34 sampling 1067/2000ml at 0.47ml/s with 26.1psi vacuum @00:40:44.76 Cartridge 34 sampling 1095/2000ml at 0.48ml/s with 26.4psi vacuum @00:41:45.00 Cartridge 34 sampling 1123/2000ml at 0.48ml/s with 24.2psi vacuum @00:42:44.75 Cartridge 34 sampling 1151/2000ml at 0.45ml/s with 28.4psi vacuum @00:43:44.85 Cartridge 34 sampling 1178/2000ml at 0.46ml/s with 20.1psi vacuum @00:44:13.37 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:44:13.39 Retry #5 of 10 after 10:00 @00:44:44.75 Cartridge 34 sampling 1205/2000ml at 0.45ml/s with 17.8psi vacuum 2.3m deep (18.1psia) @00:45:44.93 Cartridge 34 sampling 1233/2000ml at 0.46ml/s with 26.4psi vacuum @00:46:44.76 Cartridge 34 sampling 1260/2000ml at 0.46ml/s with 25.2psi vacuum @00:47:44.97 Cartridge 34 sampling 1287/2000ml at 0.44ml/s with 18.3psi vacuum 2.0m deep (17.7psia) @00:48:44.76 Cartridge 34 sampling 1314/2000ml at 0.46ml/s with 21.3psi vacuum @00:49:44.88 Cartridge 34 sampling 1341/2000ml at 0.44ml/s with 22.4psi vacuum @00:50:44.76 Cartridge 34 sampling 1367/2000ml at 0.43ml/s with 22.7psi vacuum @00:51:44.93 Cartridge 34 sampling 1393/2000ml at 0.45ml/s with 24.6psi vacuum @00:52:45.01 Cartridge 34 sampling 1419/2000ml at 0.44ml/s with 26.1psi vacuum @00:53:44.75 Cartridge 34 sampling 1445/2000ml at 0.44ml/s with 23.1psi vacuum @00:54:13.45 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @00:54:13.47 Retry #6 of 10 after 10:00 @00:54:44.92 Cartridge 34 sampling 1471/2000ml at 0.44ml/s with 23.6psi vacuum @00:55:44.75 Cartridge 34 sampling 1497/2000ml at 0.42ml/s with 28.8psi vacuum @00:56:44.92 Cartridge 34 sampling 1523/2000ml at 0.43ml/s with 21.1psi vacuum 1.6m deep (17.0psia) @00:57:44.82 Cartridge 34 sampling 1548/2000ml at 0.43ml/s with 24.0psi vacuum @00:58:44.97 Cartridge 34 sampling 1573/2000ml at 0.42ml/s with 20.5psi vacuum @00:59:44.75 Cartridge 34 sampling 1599/2000ml at 0.41ml/s with 16.7psi vacuum 1.8m deep (17.4psia) @01:00:44.97 Cartridge 34 sampling 1624/2000ml at 0.41ml/s with 28.7psi vacuum @01:01:44.75 Cartridge 34 sampling 1649/2000ml at 0.42ml/s with 27.4psi vacuum @01:02:44.92 Cartridge 34 sampling 1674/2000ml at 0.43ml/s with 23.9psi vacuum @01:03:44.82 Cartridge 34 sampling 1698/2000ml at 0.42ml/s with 25.6psi vacuum @01:04:13.77 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:04:13.81 Retry #7 of 10 after 10:00 @01:04:44.87 Cartridge 34 sampling 1723/2000ml at 0.39ml/s with 23.2psi vacuum @01:05:44.76 Cartridge 34 sampling 1747/2000ml at 0.40ml/s with 26.9psi vacuum @01:06:44.95 Cartridge 34 sampling 1771/2000ml at 0.41ml/s with 19.8psi vacuum @01:07:44.97 Cartridge 34 sampling 1795/2000ml at 0.40ml/s with 20.1psi vacuum @01:08:44.75 Cartridge 34 sampling 1818/2000ml at 0.37ml/s with 16.6psi vacuum @01:09:45.22 Cartridge 34 sampling 1842/2000ml at 0.37ml/s with 24.5psi vacuum @01:10:44.76 Cartridge 34 sampling 1865/2000ml at 0.40ml/s with 27.1psi vacuum @01:11:46.32 Cartridge 34 sampling 1889/2000ml at 0.38ml/s with 27.0psi vacuum @01:12:44.75 Cartridge 34 sampling 1912/2000ml at 0.39ml/s with 25.5psi vacuum @01:13:44.93 Cartridge 34 sampling 1935/2000ml at 0.38ml/s with 27.8psi vacuum @01:14:13.92 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:14:13.95 Retry #8 of 10 after 10:00 @01:14:44.75 Cartridge 34 sampling 1958/2000ml at 0.40ml/s with 22.6psi vacuum @01:15:44.93 Cartridge 34 sampling 1981/2000ml at 0.35ml/s with 16.0psi vacuum 3.9m deep (20.5psia) @01:16:34.27 Pressure.coast @01:16:34.41 Intake.close @01:16:34.45 Exhaust.close @01:16:34.66 Toroid.reconfigure :hold @01:16:34.72 Toroid.seek :bypass @01:16:36.34 Toroid.reconfigure :default @01:16:36.40 Toroid.seek :clear @01:16:38.06 Sampled 2000.0ml @01:16:40.47 SamplePump.setPosition! 0ml @01:16:40.63 Plunge.seek :reagent @01:16:40.78 Bypass.open @01:16:41.60 Intake.open @01:16:41.84 Sample loop closed. Ready to Deliver Reagents @01:16:41.86 Archive.seek :sealed @01:16:42.69 SamplePump.reconfigure :slow @01:16:42.75 SamplePump.seek -3ml @01:16:45.51 Archive.seek :evac @01:16:49.51 delay 1 second @01:16:50.75 SamplePump.seek -2.8ml @01:16:51.45 delay 1 second @01:16:52.69 SamplePump.seek -3.3ml @01:16:54.02 delay 1 second @01:16:55.05 Exhaust.open @01:16:55.78 Bypass.close @01:16:56.58 SamplePump.seek -35.3ml @01:16:59.62 delay 1:00 @01:17:59.66 Archive.seek :clear @01:18:05.39 delay 3 seconds @01:18:08.50 Intake.close @01:18:08.55 Exhaust.close @01:18:16.81 RNAlater soak for 10:00 @01:24:13.99 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:24:14.01 Retry #9 of 10 after 10:00 @01:28:16.88 Archive.seek :tVent @01:28:18.62 delay 1 second @01:28:19.68 Archive.seek :evac @01:28:35.83 delay 10 seconds @01:28:45.87 Sample preserved and evacuated @01:28:46.43
Carousel.seek 34 @01:28:46.96 Plunge.seek :locked @01:28:48.05 Archive.seek :clear @01:28:48.54 Air.seek :clear @01:28:48.95 Lysis1.seek :clear @01:29:05.48
Archive.coast @01:29:05.59 Air.coast @01:29:05.64 Lysis1.coast @01:29:05.72 Lysis2.coast @01:29:05.89 Plunge.coast @01:29:05.93 SamplePump.coast @01:29:05.98 Toroid.coast @01:29:06.18 Carousel.coast @01:29:06.53 Gate.power :motor,:OFF @01:29:06.54 Gate.power :core,:OFF @01:29:06.55 Gate.power :analytic1,:OFF @01:29:06.57 Gate.power :analytic2,:OFF @01:29:06.58 Gate.power :analytic3,:OFF @01:29:06.64 Gate.power :main,:OFF @01:29:06.66 Safely stopped and ready to power off @01:29:07.23 delayUntil "12:00:00.00" @01:34:14.06 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:34:14.08 Retry #10 of 10 after 10:00 @01:44:14.14 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:44:14.30 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 34" @01:44:14.34 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:44:14.36 Retry #1 of 10 after 10:00 @01:54:14.41 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @01:54:14.43 Retry #2 of 10 after 10:00 @02:04:14.52 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @02:04:14.59 Retry #3 of 10 after 10:00 @02:14:14.67 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @02:14:14.69 Retry #4 of 10 after 10:00 @02:24:15.25 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @02:24:15.32 Retry #5 of 10 after 10:00 @02:34:15.38 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @02:34:15.55 Retry #6 of 10 after 10:00 @02:44:15.62 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @02:44:15.64 Retry #7 of 10 after 10:00 @02:54:15.71 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @02:54:15.78 Retry #8 of 10 after 10:00 @03:04:15.83 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:04:15.85 Retry #9 of 10 after 10:00 @03:14:15.94 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:14:15.97 Retry #10 of 10 after 10:00 @03:24:16.06 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:24:16.08 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @03:24:16.11 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:24:16.13 Retry #1 of 10 after 10:00 @03:34:16.18 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:34:16.20 Retry #2 of 10 after 10:00 @03:44:16.27 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:44:16.36 Retry #3 of 10 after 10:00 @03:54:16.41 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:54:16.43 Retry #4 of 10 after 10:00 @04:04:16.49 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:04:16.52 Retry #5 of 10 after 10:00 @04:14:16.56 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:14:16.58 Retry #6 of 10 after 10:00 @04:24:16.82 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:24:16.86 Retry #7 of 10 after 10:00 @04:34:16.97 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:34:17.00 Retry #8 of 10 after 10:00 @04:44:17.06 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:44:17.08 Retry #9 of 10 after 10:00 @04:54:17.16 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:54:17.18 Retry #10 of 10 after 10:00 @05:04:17.22 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @08:26:47.32

Session Begun @08:26:58.50 -> Cartridge.dry @08:27:20.91 -> exit @08:27:20.93 Session Ended @12:00:00.29

Gate.power :main,:ON,limit to 5000mA @12:00:00.30 Gate.power :analytic1,:ON,limit to 5000mA @12:00:00.31 Gate.power :analytic2,:ON,limit to 5000mA @12:00:00.33 Gate.power :analytic3,:ON,limit to 5000mA @12:00:00.37 Powering core bus @12:00:02.65 Archive.configure :default @12:00:02.78 Air.configure :default @12:00:02.90 Lysis1.configure :default @12:00:03.01 Lysis2.configure :default @12:00:03.14 Plunge.configure :default @12:00:03.28 SamplePump.configure :default @12:00:03.43 ExtraPump.configure :default @12:00:03.58 Toroid.configure :default @12:00:03.72 Carousel.configure :default @12:00:03.85 Pressure.configure :default @12:00:03.91 All elf controllers running firmware version 4.61 @12:00:03.93 Gate.power :motor,:ON,limit to 5000mA @12:00:04.30 Archive.reconfigure :default @12:00:04.58 Archive.home.jog 0 @12:00:04.63 Plunge.reconfigure :default @12:00:04.70 Lysis2.reconfigure :default @12:00:04.83 Lysis1.reconfigure :default @12:00:05.00 Toroid.reconfigure :default @12:00:05.27 Air.reconfigure :default @12:00:05.95 Plunge.home.jog 0 @12:00:06.03 Lysis2.home.jog 0 @12:00:06.10 Toroid.home.jog 0 @12:00:06.14 Lysis1.home.jog 0 @12:00:06.24 Air.home.jog 0 @12:00:08.23 Toroid.reconfigure :homing @12:00:08.46 Plunge.reconfigure :homing @12:00:08.49 Lysis1.reconfigure :homing @12:00:08.62 Air.reconfigure :homing @12:00:08.69 Archive.reconfigure :homing @12:00:08.77 Lysis2.reconfigure :homing @12:00:09.26 Toroid.jog 622 @12:00:09.37 Plunge.jog 851 @12:00:09.39 Air.jog 1362 @12:00:09.43 Archive.jog 1374 @12:00:09.48 Lysis1.jog 1348 @12:00:09.61 Lysis2.jog 1355 @12:00:10.68 Plunge.setRawPosition! 50 @12:00:10.77 Plunge.seek :home @12:00:10.87 Air.setRawPosition! 50 @12:00:10.91 Air.seek :home @12:00:11.00 Archive.setRawPosition! 50 @12:00:11.08 Archive.seek :home @12:00:11.17 Lysis2.setRawPosition! 50 @12:00:11.21 Lysis1.setRawPosition! 50 @12:00:11.22 Plunge.reconfigure :default @12:00:11.37 Lysis2.seek :home @12:00:11.45 Lysis1.seek :home @12:00:11.57 Air.reconfigure :default @12:00:11.60 Archive.reconfigure :default @12:00:11.80 Lysis2.reconfigure :default @12:00:11.84 Lysis1.reconfigure :default @12:00:11.87 Toroid.setRawPosition! 0 @12:00:12.02 Toroid.seek :home @12:00:12.13 Toroid.reconfigure :default @12:00:12.36
Toroid.seek :clear @12:00:13.97 Carousel.reconfigure :default @12:00:14.02 Carousel.home.jog 0 @12:00:14.52 Carousel.seek 34 @12:00:14.70 Selecting Cartridge 33 @12:00:14.91 Carousel.seek 33 @12:00:15.72 Lysis1.seek :locked @12:00:19.04 Plunge.seek :locked @12:00:19.35 Air.seek :locked @12:00:23.58
Mounted archive Cartridge @12:00:23.68 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 33" @12:00:23.77 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:00:23.81 Retry #1 of 10 after 10:00 @12:00:24.23 Duration of filtering limited to 4:05:00 @12:00:24.52 Priming sample loop w/160ml, bypass w/40ml @12:00:24.54 SamplePump.setPosition! 0ml @12:00:24.57 Bypass.open @12:00:26.15 Intake.open @12:00:27.30 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:00:29.88 Vacuum tare is 18% of full scale @12:00:29.90 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:00:30.11 Pressure.reconfigure :default @12:00:30.17 Pressure.hold @12:00:30.38 SamplePump.reconfigure :prime @12:00:30.44 SamplePump.seek 40ml,3:51 @12:01:13.69 Pressure.coast @12:01:13.84 SamplePump.setPosition! 0ml @12:01:13.93 Exhaust.open @12:01:15.47 Bypass.close @12:01:17.06 Pressure.reconfigure :default @12:01:17.12 Pressure.hold @12:01:17.38 SamplePump.reconfigure :prime @12:01:17.44 SamplePump.seek 160ml,13:51 @12:04:09.22 Pressure.coast @12:04:09.28 Bypass.open @12:04:09.98 Intake.close @12:04:10.90 Toroid.reconfigure :hold @12:04:10.96 Toroid.seek :crack @12:04:12.54 Toroid.seek :cartridge @12:04:13.47 Toroid.reconfigure :default @12:04:14.10 Toroid.seek :clear @12:04:16.33 Pressurizing cartridge to 26psi above 14.5psia @12:04:16.38 SamplePump.setPosition! 0ml @12:04:16.49 Intake.open @12:04:17.35 Bypass.close @12:04:18.10 Exhaust.close @12:04:19.08 Pressure.reconfigure :default @12:04:19.21 Pressure.hold @12:04:20.06 SamplePump.reconfigure :charge @12:04:20.12 SamplePump.seek 2.3ml,42 seconds @12:04:30.76 Pressure.coast @12:04:30.81 Intake.close @12:04:31.70 Pressurized cartridge with 0.517ml to 26.1psi @12:04:31.71 Isolated @35.2psia (20.7psi) @12:04:31.72 Checking for cartridge leaks from 35.2psia @12:04:31.74 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:04:37.34 Cartridge pressure fell 7% to 32.9psia @12:04:37.54 Sampling 2000.0ml @12:04:37.58 SamplePump.setPosition! 0ml @12:04:37.65 Exhaust.open @12:04:38.33 Intake.open @12:04:39.28 Pressure.reconfigure :default @12:04:39.34 Pressure.hold @12:04:39.62 SamplePump.reconfigure :sample @12:04:39.68 SamplePump.seek 2000ml,2:47:12 @12:04:42.34 Cartridge 33 sampling 1/2000ml at 0.50ml/s with 23.1psi vacuum @12:05:42.52 Cartridge 33 sampling 31/2000ml at 0.50ml/s with 25.4psi vacuum @12:06:42.75 Cartridge 33 sampling 61/2000ml at 0.50ml/s with 24.5psi vacuum @12:07:42.35 Cartridge 33 sampling 91/2000ml at 0.49ml/s with 23.1psi vacuum @12:08:42.67 Cartridge 33 sampling 121/2000ml at 0.50ml/s with 24.0psi vacuum @12:09:42.35 Cartridge 33 sampling 151/2000ml at 0.51ml/s with 20.8psi vacuum @12:10:23.93 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:10:24.02 Retry #2 of 10 after 10:00 @12:10:42.50 Cartridge 33 sampling 181/2000ml at 0.50ml/s with 23.4psi vacuum @12:11:42.35 Cartridge 33 sampling 212/2000ml at 0.52ml/s with 24.7psi vacuum @12:12:42.50 Cartridge 33 sampling 243/2000ml at 0.51ml/s with 26.3psi vacuum @12:13:42.35 Cartridge 33 sampling 274/2000ml at 0.51ml/s with 25.9psi vacuum @12:14:42.51 Cartridge 33 sampling 304/2000ml at 0.50ml/s with 18.6psi vacuum 2.2m deep (18.0psia) @12:15:42.35 Cartridge 33 sampling 335/2000ml at 0.51ml/s with 22.4psi vacuum @12:16:42.56 Cartridge 33 sampling 365/2000ml at 0.50ml/s with 24.6psi vacuum @12:17:42.35 Cartridge 33 sampling 395/2000ml at 0.51ml/s with 21.8psi vacuum @12:18:42.42 Cartridge 33 sampling 426/2000ml at 0.50ml/s with 25.3psi vacuum @12:19:42.35 Cartridge 33 sampling 456/2000ml at 0.51ml/s with 22.8psi vacuum @12:20:24.09 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:20:24.11 Retry #3 of 10 after 10:00 @12:20:42.40 Cartridge 33 sampling 486/2000ml at 0.50ml/s with 24.9psi vacuum @12:21:42.35 Cartridge 33 sampling 516/2000ml at 0.50ml/s with 23.7psi vacuum @12:22:42.34 Cartridge 33 sampling 546/2000ml at 0.48ml/s with 21.3psi vacuum @12:23:42.80 Cartridge 33 sampling 576/2000ml at 0.50ml/s with 22.3psi vacuum @12:24:42.35 Cartridge 33 sampling 605/2000ml at 0.51ml/s with 25.9psi vacuum @12:25:42.50 Cartridge 33 sampling 635/2000ml at 0.49ml/s with 19.8psi vacuum @12:26:42.34 Cartridge 33 sampling 664/2000ml at 0.49ml/s with 21.8psi vacuum @12:27:42.51 Cartridge 33 sampling 693/2000ml at 0.49ml/s with 22.1psi vacuum @12:28:42.35 Cartridge 33 sampling 722/2000ml at 0.49ml/s with 26.6psi vacuum @12:29:42.51 Cartridge 33 sampling 751/2000ml at 0.50ml/s with 25.5psi vacuum @12:30:42.36 Cartridge 33 sampling 780/2000ml at 0.47ml/s with 28.0psi vacuum @12:30:49.68 upsync @12:31:42.53 Cartridge 33 sampling 809/2000ml at 0.47ml/s with 24.2psi vacuum @12:32:42.35 Cartridge 33 sampling 838/2000ml at 0.48ml/s with 19.8psi vacuum @12:33:42.51 Cartridge 33 sampling 866/2000ml at 0.47ml/s with 22.9psi vacuum @12:34:42.35 Cartridge 33 sampling 895/2000ml at 0.46ml/s with 22.5psi vacuum @12:35:42.41 Cartridge 33 sampling 923/2000ml at 0.48ml/s with 25.4psi vacuum @12:36:42.55 Cartridge 33 sampling 951/2000ml at 0.47ml/s with 25.6psi vacuum @12:37:42.35 Cartridge 33 sampling 979/2000ml at 0.44ml/s with 25.6psi vacuum @12:38:42.67 Cartridge 33 sampling 1006/2000ml at 0.46ml/s with 24.3psi vacuum @12:39:42.35 Cartridge 33 sampling 1033/2000ml at 0.44ml/s with 28.2psi vacuum @12:40:42.50 Cartridge 33 sampling 1061/2000ml at 0.44ml/s with 26.9psi vacuum @12:41:42.35 Cartridge 33 sampling 1088/2000ml at 0.43ml/s with 28.4psi vacuum @12:42:42.51 Cartridge 33 sampling 1115/2000ml at 0.44ml/s with 18.0psi vacuum 4.3m deep (21.1psia) @12:43:42.35 Cartridge 33 sampling 1141/2000ml at 0.45ml/s with 24.4psi vacuum @12:44:42.62 Cartridge 33 sampling 1168/2000ml at 0.45ml/s with 25.1psi vacuum @12:45:42.35 Cartridge 33 sampling 1195/2000ml at 0.45ml/s with 25.2psi vacuum @12:46:42.51 Cartridge 33 sampling 1221/2000ml at 0.44ml/s with 25.5psi vacuum @12:47:42.35 Cartridge 33 sampling 1247/2000ml at 0.44ml/s with 21.6psi vacuum @12:48:42.51 Cartridge 33 sampling 1273/2000ml at 0.42ml/s with 17.9psi vacuum 2.3m deep (18.0psia) @12:49:42.35 Cartridge 33 sampling 1299/2000ml at 0.43ml/s with 27.3psi vacuum @12:50:42.64 Cartridge 33 sampling 1325/2000ml at 0.44ml/s with 25.2psi vacuum @12:51:42.59 Cartridge 33 sampling 1351/2000ml at 0.42ml/s with 28.3psi vacuum @12:52:42.35 Cartridge 33 sampling 1376/2000ml at 0.44ml/s with 22.7psi vacuum @12:53:42.55 Cartridge 33 sampling 1402/2000ml at 0.40ml/s with 26.5psi vacuum @12:54:42.34 Cartridge 33 sampling 1427/2000ml at 0.42ml/s with 17.0psi vacuum 2.0m deep (17.6psia) @12:55:42.50 Cartridge 33 sampling 1452/2000ml at 0.42ml/s with 21.0psi vacuum 1.8m deep (17.3psia) @12:56:42.35 Cartridge 33 sampling 1477/2000ml at 0.42ml/s with 20.1psi vacuum @12:57:42.51 Cartridge 33 sampling 1502/2000ml at 0.42ml/s with 22.9psi vacuum @12:58:42.35 Cartridge 33 sampling 1527/2000ml at 0.40ml/s with 25.5psi vacuum @12:59:42.51 Cartridge 33 sampling 1551/2000ml at 0.40ml/s with 22.8psi vacuum @13:00:42.34 Cartridge 33 sampling 1576/2000ml at 0.40ml/s with 28.3psi vacuum @13:01:42.56 Cartridge 33 sampling 1600/2000ml at 0.41ml/s with 21.4psi vacuum @13:02:42.35 Cartridge 33 sampling 1624/2000ml at 0.41ml/s with 24.1psi vacuum @13:03:42.40 Cartridge 33 sampling 1648/2000ml at 0.40ml/s with 21.5psi vacuum @13:04:42.34 Cartridge 33 sampling 1671/2000ml at 0.40ml/s with 22.2psi vacuum @13:05:42.38 Cartridge 33 sampling 1695/2000ml at 0.40ml/s with 25.0psi vacuum @13:06:42.68 Cartridge 33 sampling 1718/2000ml at 0.40ml/s with 24.2psi vacuum @13:07:42.35 Cartridge 33 sampling 1741/2000ml at 0.38ml/s with 18.6psi vacuum @13:08:42.48 Cartridge 33 sampling 1764/2000ml at 0.39ml/s with 21.1psi vacuum 2.6m deep (18.5psia) @13:09:42.35 Cartridge 33 sampling 1787/2000ml at 0.36ml/s with 28.3psi vacuum @13:10:42.51 Cartridge 33 sampling 1810/2000ml at 0.39ml/s with 25.5psi vacuum @13:11:42.34 Cartridge 33 sampling 1833/2000ml at 0.40ml/s with 21.6psi vacuum @13:12:42.54 Cartridge 33 sampling 1856/2000ml at 0.39ml/s with 26.1psi vacuum @13:13:42.35 Cartridge 33 sampling 1878/2000ml at 0.38ml/s with 17.2psi vacuum @13:14:42.53 Cartridge 33 sampling 1901/2000ml at 0.38ml/s with 24.1psi vacuum @13:15:42.35 Cartridge 33 sampling 1923/2000ml at 0.37ml/s with 18.5psi vacuum @13:16:42.41 Cartridge 33 sampling 1945/2000ml at 0.39ml/s with 22.0psi vacuum @13:17:42.35 Cartridge 33 sampling 1968/2000ml at 0.37ml/s with 27.6psi vacuum @13:18:42.55 Cartridge 33 sampling 1990/2000ml at 0.39ml/s with 22.3psi vacuum @13:19:10.78 Pressure.coast @13:19:10.91 Intake.close @13:19:10.96 Exhaust.close @13:19:11.15 Toroid.reconfigure :hold @13:19:11.21 Toroid.seek :bypass @13:19:12.83 Toroid.reconfigure :default @13:19:12.89 Toroid.seek :clear @13:19:14.64 Sampled 2000.0ml @13:19:16.20 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 33" @13:19:17.39 SamplePump.setPosition! 0ml @13:19:17.53 Plunge.seek :reagent @13:19:17.64 Bypass.open @13:19:18.64 Intake.open @13:19:18.79 Sample loop closed. Ready to Deliver Reagents @13:19:18.89 Archive.seek :sealed @13:19:19.52 SamplePump.reconfigure :slow @13:19:19.68 SamplePump.seek -3ml @13:19:22.54 Archive.seek :evac @13:19:26.40 delay 1 second @13:19:27.65 SamplePump.seek -2.8ml @13:19:28.47 delay 1 second @13:19:29.70 SamplePump.seek -3.3ml @13:19:31.03 delay 1 second @13:19:32.06 Exhaust.open @13:19:32.73 Bypass.close @13:19:33.52 SamplePump.seek -35.3ml @13:19:34.40 upsync @13:19:36.53 delay 1:00 @13:20:36.57 Archive.seek :clear @13:20:42.37 delay 3 seconds @13:20:45.50 Intake.close @13:20:45.55 Exhaust.close @13:20:53.69 RNAlater soak for 10:00 @13:30:53.79 Archive.seek :tVent @13:30:55.73 delay 1 second @13:30:56.83 Archive.seek :evac @13:31:12.80 delay 10 seconds @13:31:22.88 Sample preserved and evacuated @13:31:23.56
Carousel.seek 33 @13:31:24.05 Plunge.seek :locked @13:31:25.20 Archive.seek :clear @13:31:25.52 Air.seek :clear @13:31:25.82 Lysis1.seek :clear @13:31:42.48
Archive.coast @13:31:42.61 Air.coast @13:31:42.70 Lysis1.coast @13:31:44.04 Lysis2.coast @13:31:44.13 Plunge.coast @13:31:44.18 SamplePump.coast @13:31:44.31 Toroid.coast @13:31:44.42 Carousel.coast @13:31:44.72 Gate.power :motor,:OFF @13:31:44.75 Gate.power :core,:OFF @13:31:44.76 Gate.power :analytic1,:OFF @13:31:44.77 Gate.power :analytic2,:OFF @13:31:44.78 Gate.power :analytic3,:OFF @13:31:44.80 Gate.power :main,:OFF @13:31:44.81 Safely stopped and ready to power off @13:31:45.44 delayUntil "00:00:00.00PDT21-Jun-21" @13:31:45.56 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT21-Jun-21" @13:31:58.71 upsync @00:00:00.34PDT21-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.35 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.37 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.38 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.42 Powering core bus @00:00:02.72 Archive.configure :default @00:00:02.85 Air.configure :default @00:00:02.97 Lysis1.configure :default @00:00:03.09 Lysis2.configure :default @00:00:03.24 Plunge.configure :default @00:00:03.38 SamplePump.configure :default @00:00:03.51 ExtraPump.configure :default @00:00:03.65 Toroid.configure :default @00:00:03.78 Carousel.configure :default @00:00:03.93 Pressure.configure :default @00:00:03.99 All elf controllers running firmware version 4.61 @00:00:04.01 Gate.power :motor,:ON,limit to 5000mA @00:00:04.19 Archive.reconfigure :default @00:00:04.56 Lysis2.reconfigure :default @00:00:04.63 Archive.home.jog 0 @00:00:04.67 Toroid.reconfigure :default @00:00:04.70 Plunge.reconfigure :default @00:00:04.74 Lysis1.reconfigure :default @00:00:04.78 Air.reconfigure :default @00:00:05.06 Lysis2.home.jog 0 @00:00:05.10 Toroid.home.jog 0 @00:00:05.13 Plunge.home.jog 0 @00:00:05.15 Air.home.jog 0 @00:00:05.18 Lysis1.home.jog 0 @00:00:05.92 Archive.reconfigure :homing @00:00:05.97 Toroid.reconfigure :homing @00:00:06.05 Plunge.reconfigure :homing @00:00:06.14 Lysis1.reconfigure :homing @00:00:06.21 Air.reconfigure :homing @00:00:06.45 Lysis2.reconfigure :homing @00:00:06.95 Archive.jog 1371 @00:00:07.08 Toroid.jog 620 @00:00:07.22 Plunge.jog 850 @00:00:07.29 Air.jog 1359 @00:00:07.42 Lysis1.jog 1350 @00:00:07.69 Lysis2.jog 1354 @00:00:08.69 Archive.setRawPosition! 50 @00:00:08.82 Archive.seek :home @00:00:09.00 Plunge.setRawPosition! 50 @00:00:09.23 Plunge.seek :home @00:00:09.57 Lysis2.setRawPosition! 50 @00:00:10.68 Lysis1.setRawPosition! 50 @00:00:10.92 Air.setRawPosition! 50 @00:00:10.93 Archive.reconfigure :default @00:00:11.28 Lysis2.seek :home @00:00:11.36 Lysis1.seek :home @00:00:11.43 Air.seek :home @00:00:12.15 Toroid.setRawPosition! 0 @00:00:12.18 Plunge.reconfigure :default @00:00:12.20 Lysis1.reconfigure :default @00:00:12.23 Lysis2.reconfigure :default @00:00:12.28 Air.reconfigure :default @00:00:12.51 Toroid.seek :home @00:00:12.87 Toroid.reconfigure :default @00:00:13.10
Toroid.seek :clear @00:00:14.72 Carousel.reconfigure :default @00:00:14.77 Carousel.home.jog 0 @00:00:15.28 Carousel.seek 33 @00:00:15.45 Selecting Cartridge 32 @00:00:15.67 Carousel.seek 32 @00:00:16.18 Lysis1.seek :locked @00:00:19.55 Plunge.seek :locked @00:00:19.89 Air.seek :locked @00:00:24.16
Mounted archive Cartridge @00:00:24.26 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 32" @00:00:24.58 Duration of filtering limited to 4:05:00 @00:00:24.87 Priming sample loop w/160ml, bypass w/40ml @00:00:24.89 SamplePump.setPosition! 0ml @00:00:24.92 Bypass.open @00:00:26.48 Intake.open @00:00:27.49 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:30.12 Vacuum tare is 18% of full scale @00:00:30.14 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:30.73 Pressure.reconfigure :default @00:00:30.95 Pressure.hold @00:00:31.16 SamplePump.reconfigure :prime @00:00:31.22 SamplePump.seek 40ml,3:51 @00:01:05.19 upsync @00:01:14.45 Pressure.coast @00:01:14.58 SamplePump.setPosition! 0ml @00:01:14.66 Exhaust.open @00:01:16.22 Bypass.close @00:01:17.89 Pressure.reconfigure :default @00:01:17.95 Pressure.hold @00:01:18.22 SamplePump.reconfigure :prime @00:01:18.29 SamplePump.seek 160ml,13:51 @00:04:09.97 Pressure.coast @00:04:10.03 Bypass.open @00:04:10.84 Intake.close @00:04:11.84 Toroid.reconfigure :hold @00:04:11.90 Toroid.seek :crack @00:04:13.49 Toroid.seek :cartridge @00:04:14.42 Toroid.reconfigure :default @00:04:14.48 Toroid.seek :clear @00:04:16.67 Pressurizing cartridge to 26psi above 14.6psia @00:04:16.71 SamplePump.setPosition! 0ml @00:04:16.84 Intake.open @00:04:17.75 Bypass.close @00:04:18.45 Exhaust.close @00:04:19.46 Pressure.reconfigure :default @00:04:19.59 Pressure.hold @00:04:20.04 SamplePump.reconfigure :charge @00:04:20.10 SamplePump.seek 2.3ml,42 seconds @00:04:31.26 Pressure.coast @00:04:31.31 Intake.close @00:04:32.17 Pressurized cartridge with 0.569ml to 26.3psi @00:04:32.18 Isolated @35.5psia (20.9psi) @00:04:32.19 Checking for cartridge leaks from 35.5psia @00:04:32.21 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:37.48 Cartridge pressure fell 2% to 34.8psia @00:04:37.72 Sampling 2000.0ml @00:04:37.78 SamplePump.setPosition! 0ml @00:04:37.86 Exhaust.open @00:04:38.58 Intake.open @00:04:39.71 Pressure.reconfigure :default @00:04:39.87 Pressure.hold @00:04:40.59 SamplePump.reconfigure :sample @00:04:40.66 SamplePump.seek 2000ml,2:47:12 @00:04:42.77 Cartridge 32 sampling 1/2000ml at 0.58ml/s with 26.2psi vacuum @00:05:42.98 Cartridge 32 sampling 33/2000ml at 0.54ml/s with 22.6psi vacuum @00:06:42.78 Cartridge 32 sampling 65/2000ml at 0.54ml/s with 25.4psi vacuum @00:07:42.81 Cartridge 32 sampling 98/2000ml at 0.53ml/s with 22.8psi vacuum @00:08:42.78 Cartridge 32 sampling 131/2000ml at 0.54ml/s with 21.7psi vacuum @00:09:43.13 Cartridge 32 sampling 164/2000ml at 0.58ml/s with 25.8psi vacuum @00:10:43.13 Cartridge 32 sampling 197/2000ml at 0.57ml/s with 23.1psi vacuum @00:11:42.77 Cartridge 32 sampling 230/2000ml at 0.55ml/s with 24.1psi vacuum @00:12:43.09 Cartridge 32 sampling 262/2000ml at 0.54ml/s with 22.9psi vacuum @00:13:42.76 Cartridge 32 sampling 295/2000ml at 0.53ml/s with 28.3psi vacuum @00:14:42.96 Cartridge 32 sampling 328/2000ml at 0.57ml/s with 27.0psi vacuum @00:15:42.76 Cartridge 32 sampling 361/2000ml at 0.56ml/s with 21.9psi vacuum 1.7m deep (17.2psia) @00:16:42.92 Cartridge 32 sampling 393/2000ml at 0.54ml/s with 27.2psi vacuum @00:17:42.77 Cartridge 32 sampling 426/2000ml at 0.52ml/s with 29.7psi vacuum @00:18:42.97 Cartridge 32 sampling 459/2000ml at 0.57ml/s with 19.1psi vacuum 4.1m deep (20.8psia) @00:19:42.77 Cartridge 32 sampling 491/2000ml at 0.52ml/s with 28.9psi vacuum @00:20:42.94 Cartridge 32 sampling 523/2000ml at 0.50ml/s with 13.9psi vacuum 3.8m deep (20.3psia) @00:21:42.78 Cartridge 32 sampling 555/2000ml at 0.52ml/s with 21.9psi vacuum @00:22:42.82 Cartridge 32 sampling 586/2000ml at 0.54ml/s with 28.7psi vacuum @00:23:42.81 Cartridge 32 sampling 618/2000ml at 0.53ml/s with 20.4psi vacuum @00:24:42.80 Cartridge 32 sampling 650/2000ml at 0.51ml/s with 13.9psi vacuum 2.0m deep (17.7psia) @00:25:43.10 Cartridge 32 sampling 682/2000ml at 0.53ml/s with 20.4psi vacuum @00:26:42.78 Cartridge 32 sampling 713/2000ml at 0.53ml/s with 17.8psi vacuum 3.7m deep (20.2psia) @00:27:43.08 Cartridge 32 sampling 745/2000ml at 0.53ml/s with 21.2psi vacuum @00:28:42.78 Cartridge 32 sampling 776/2000ml at 0.52ml/s with 16.9psi vacuum 2.3m deep (18.0psia) @00:29:42.94 Cartridge 32 sampling 808/2000ml at 0.53ml/s with 15.7psi vacuum 7.1m deep (25.1psia) @00:30:42.77 Cartridge 32 sampling 839/2000ml at 0.54ml/s with 20.5psi vacuum 2.3m deep (18.1psia) @00:31:42.94 Cartridge 32 sampling 870/2000ml at 0.52ml/s with 22.8psi vacuum @00:32:42.78 Cartridge 32 sampling 901/2000ml at 0.50ml/s with 14.7psi vacuum @00:33:42.94 Cartridge 32 sampling 932/2000ml at 0.52ml/s with 26.6psi vacuum @00:34:42.78 Cartridge 32 sampling 962/2000ml at 0.52ml/s with 21.8psi vacuum 1.8m deep (17.3psia) @00:35:42.93 Cartridge 32 sampling 993/2000ml at 0.51ml/s with 23.4psi vacuum 1.9m deep (17.5psia) @00:36:42.77 Cartridge 32 sampling 1023/2000ml at 0.50ml/s with 14.2psi vacuum 4.6m deep (21.4psia) @00:37:42.93 Cartridge 32 sampling 1052/2000ml at 0.50ml/s with 27.4psi vacuum @00:38:42.78 Cartridge 32 sampling 1082/2000ml at 0.49ml/s with 15.5psi vacuum 3.2m deep (19.4psia) @00:39:42.87 Cartridge 32 sampling 1112/2000ml at 0.51ml/s with 16.1psi vacuum 6.2m deep (23.8psia) @00:40:43.18 Cartridge 32 sampling 1142/2000ml at 0.51ml/s with 19.2psi vacuum 3.9m deep (20.5psia) @00:41:42.77 Cartridge 32 sampling 1172/2000ml at 0.49ml/s with 27.3psi vacuum @00:42:43.06 Cartridge 32 sampling 1201/2000ml at 0.49ml/s with 13.4psi vacuum @00:43:42.78 Cartridge 32 sampling 1230/2000ml at 0.50ml/s with 21.5psi vacuum 1.6m deep (17.0psia) @00:44:43.02 Cartridge 32 sampling 1259/2000ml at 0.47ml/s with 16.1psi vacuum 5.1m deep (22.2psia) @00:45:42.78 Cartridge 32 sampling 1287/2000ml at 0.47ml/s with 16.8psi vacuum @00:46:42.94 Cartridge 32 sampling 1316/2000ml at 0.46ml/s with 11.5psi vacuum 9.0m deep (28.0psia) @00:47:42.78 Cartridge 32 sampling 1344/2000ml at 0.47ml/s with 12.1psi vacuum 5.1m deep (22.3psia) @00:48:43.01 Cartridge 32 sampling 1372/2000ml at 0.44ml/s with 20.3psi vacuum 3.7m deep (20.1psia) @00:49:42.78 Cartridge 32 sampling 1401/2000ml at 0.49ml/s with 25.6psi vacuum @00:50:42.99 Cartridge 32 sampling 1429/2000ml at 0.45ml/s with 27.7psi vacuum @00:51:42.77 Cartridge 32 sampling 1456/2000ml at 0.45ml/s with 15.7psi vacuum @00:52:42.83 Cartridge 32 sampling 1484/2000ml at 0.46ml/s with 12.5psi vacuum 6.9m deep (24.9psia) @00:53:42.78 Cartridge 32 sampling 1511/2000ml at 0.47ml/s with 23.9psi vacuum @00:54:42.97 Cartridge 32 sampling 1539/2000ml at 0.44ml/s with 13.9psi vacuum 8.4m deep (27.1psia) @00:55:42.96 Cartridge 32 sampling 1566/2000ml at 0.44ml/s with 19.8psi vacuum 3.0m deep (19.1psia) @00:56:42.78 Cartridge 32 sampling 1593/2000ml at 0.46ml/s with 23.6psi vacuum @00:57:42.94 Cartridge 32 sampling 1620/2000ml at 0.45ml/s with 17.0psi vacuum 4.5m deep (21.3psia) @00:58:42.78 Cartridge 32 sampling 1647/2000ml at 0.45ml/s with 18.3psi vacuum 2.6m deep (18.5psia) @00:59:43.03 Cartridge 32 sampling 1673/2000ml at 0.46ml/s with 21.6psi vacuum @01:00:42.77 Cartridge 32 sampling 1700/2000ml at 0.46ml/s with 25.3psi vacuum @01:01:42.86 Cartridge 32 sampling 1726/2000ml at 0.43ml/s with 12.3psi vacuum 7.2m deep (25.3psia) @01:02:42.77 Cartridge 32 sampling 1752/2000ml at 0.45ml/s with 20.8psi vacuum 2.2m deep (17.9psia) @01:03:42.92 Cartridge 32 sampling 1778/2000ml at 0.43ml/s with 29.7psi vacuum @01:04:42.77 Cartridge 32 sampling 1804/2000ml at 0.44ml/s with 28.7psi vacuum @01:05:42.93 Cartridge 32 sampling 1830/2000ml at 0.45ml/s with 21.6psi vacuum 2.4m deep (18.3psia) @01:06:42.77 Cartridge 32 sampling 1856/2000ml at 0.45ml/s with 25.5psi vacuum @01:07:43.11 Cartridge 32 sampling 1881/2000ml at 0.43ml/s with 27.5psi vacuum @01:08:42.94 Cartridge 32 sampling 1907/2000ml at 0.43ml/s with 28.7psi vacuum @01:09:42.81 Cartridge 32 sampling 1932/2000ml at 0.42ml/s with 13.2psi vacuum 6.6m deep (24.5psia) @01:10:43.23 Cartridge 32 sampling 1957/2000ml at 0.43ml/s with 26.8psi vacuum @01:11:42.78 Cartridge 32 sampling 1981/2000ml at 0.42ml/s with 22.2psi vacuum @01:12:28.39 Pressure.coast @01:12:28.54 Intake.close @01:12:28.59 Exhaust.close @01:12:28.77 Toroid.reconfigure :hold @01:12:28.83 Toroid.seek :bypass @01:12:30.45 Toroid.reconfigure :default @01:12:30.51 Toroid.seek :clear @01:12:32.19 Sampled 2000.0ml @01:12:35.26 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 32" @01:12:35.97 SamplePump.setPosition! 0ml @01:12:36.09 Plunge.seek :reagent @01:12:36.21 Bypass.open @01:12:37.01 Intake.open @01:12:37.24 Sample loop closed. Ready to Deliver Reagents @01:12:37.28 Archive.seek :sealed @01:12:38.10 SamplePump.reconfigure :slow @01:12:38.17 SamplePump.seek -3ml @01:12:40.95 Archive.seek :evac @01:12:45.00 delay 1 second @01:12:46.27 SamplePump.seek -2.8ml @01:12:47.03 delay 1 second @01:12:48.15 SamplePump.seek -3.3ml @01:12:49.49 delay 1 second @01:12:50.52 Exhaust.open @01:12:51.25 Bypass.close @01:12:52.19 SamplePump.seek -35.3ml @01:12:53.06 upsync @01:12:55.01 delay 1:00 @01:13:55.07 Archive.seek :clear @01:14:01.08 delay 3 seconds @01:14:04.20 Intake.close @01:14:04.28 Exhaust.close @01:14:12.22 RNAlater soak for 10:00 @01:24:12.27 Archive.seek :tVent @01:24:14.00 delay 1 second @01:24:15.09 Archive.seek :evac @01:24:31.09 delay 10 seconds @01:24:41.12 Sample preserved and evacuated @01:24:41.68
Carousel.seek 32 @01:24:42.23 Plunge.seek :locked @01:24:43.50 Archive.seek :clear @01:24:43.67 Air.seek :clear @01:24:43.83 Lysis1.seek :clear @01:25:00.82
Archive.coast @01:25:00.89 Air.coast @01:25:00.93 Lysis1.coast @01:25:00.98 Lysis2.coast @01:25:01.17 Plunge.coast @01:25:01.22 SamplePump.coast @01:25:01.26 Toroid.coast @01:25:01.42 Carousel.coast @01:25:01.99 Gate.power :motor,:OFF @01:25:02.01 Gate.power :core,:OFF @01:25:02.03 Gate.power :analytic1,:OFF @01:25:02.04 Gate.power :analytic2,:OFF @01:25:02.05 Gate.power :analytic3,:OFF @01:25:02.06 Gate.power :main,:OFF @01:25:02.08 Safely stopped and ready to power off @01:25:02.32 delayUntil "12:00:00.00" @01:25:02.34 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @01:25:13.14 upsync @11:59:32.33 Session Begun @11:59:34.84 -> exit @11:59:34.87 Session Ended @12:00:00.11
Gate.power :main,:ON,limit to 5000mA @12:00:00.13 Gate.power :analytic1,:ON,limit to 5000mA @12:00:00.14 Gate.power :analytic2,:ON,limit to 5000mA @12:00:00.15 Gate.power :analytic3,:ON,limit to 5000mA @12:00:00.19 Powering core bus @12:00:02.56 Archive.configure :default @12:00:02.93 Air.configure :default @12:00:03.18 Lysis1.configure :default @12:00:03.42 Lysis2.configure :default @12:00:03.74 Plunge.configure :default @12:00:04.33 SamplePump.configure :default @12:00:04.72 ExtraPump.configure :default @12:00:05.02 Toroid.configure :default @12:00:05.32 Carousel.configure :default @12:00:05.68 Pressure.configure :default @12:00:05.83 All elf controllers running firmware version 4.61 @12:00:05.85 Gate.power :motor,:ON,limit to 5000mA @12:00:06.90 Plunge.reconfigure :default @12:00:07.03 Archive.reconfigure :default @12:00:07.22 Air.reconfigure :default @12:00:07.26 Lysis1.reconfigure :default @12:00:07.58 Plunge.home.jog 0 @12:00:07.65 Archive.home.jog 0 @12:00:07.85 Lysis1.home.jog 0 @12:00:07.89 Air.home.jog 0 @12:00:08.84 Lysis2.reconfigure :default @12:00:08.92 Toroid.reconfigure :default @12:00:09.06 Lysis2.home.jog 0 @12:00:09.09 Toroid.home.jog 0 @12:00:09.41 Plunge.reconfigure :homing @12:00:09.47 Lysis1.reconfigure :homing @12:00:09.50 Archive.reconfigure :homing @12:00:09.54 Air.reconfigure :homing @12:00:09.71 Lysis1.jog 1351 @12:00:09.77 Plunge.jog 844 @12:00:09.82 Air.jog 1358 @12:00:09.87 Archive.jog 1374 @12:00:10.10 Toroid.reconfigure :homing @12:00:10.13 Lysis2.reconfigure :homing @12:00:10.23 Lysis2.jog 1355 @12:00:10.27 Toroid.jog 613 @12:00:10.85 Plunge.setRawPosition! 50 @12:00:10.90 Plunge.seek :home @12:00:10.99 Lysis1.setRawPosition! 50 @12:00:11.05 Lysis1.seek :home @12:00:11.15 Air.setRawPosition! 50 @12:00:11.18 Archive.setRawPosition! 50 @12:00:11.25 Air.seek :home @12:00:11.32 Archive.seek :home @12:00:11.51 Lysis2.setRawPosition! 50 @12:00:11.54 Lysis1.reconfigure :default @12:00:11.57 Plunge.reconfigure :default @12:00:11.72 Lysis2.seek :home @12:00:11.89 Air.reconfigure :default @12:00:11.91 Archive.reconfigure :default @12:00:12.08 Lysis2.reconfigure :default @12:00:12.39 Toroid.setRawPosition! 0 @12:00:12.44 Toroid.seek :home @12:00:12.53 Toroid.reconfigure :default @12:00:12.68
Toroid.seek :clear @12:00:14.29 Carousel.reconfigure :default @12:00:14.45 Carousel.home.jog 0 @12:00:15.24 Carousel.seek 32 @12:00:15.64 Selecting Cartridge 31 @12:00:16.10 Carousel.seek 31 @12:00:16.78 Lysis1.seek :locked @12:00:20.03 Plunge.seek :locked @12:00:20.17 Air.seek :locked @12:00:24.64
Mounted archive Cartridge @12:00:24.97 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 31" @12:00:25.89 Duration of filtering limited to 4:05:00 @12:00:27.74 Priming sample loop w/160ml, bypass w/40ml @12:00:27.76 SamplePump.setPosition! 0ml @12:00:27.79 Bypass.open @12:00:29.36 Intake.open @12:00:30.27 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:00:32.93 Vacuum tare is 17% of full scale @12:00:32.97 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:00:33.88 Pressure.reconfigure :default @12:00:34.12 Pressure.hold @12:00:34.86 SamplePump.reconfigure :prime @12:00:35.05 SamplePump.seek 40ml,3:51 @12:00:37.32 upsync @12:01:18.25 Pressure.coast @12:01:18.30 SamplePump.setPosition! 0ml @12:01:18.35 Exhaust.open @12:01:19.76 Bypass.close @12:01:21.72 Pressure.reconfigure :default @12:01:21.84 Pressure.hold @12:01:22.44 SamplePump.reconfigure :prime @12:01:22.60 SamplePump.seek 160ml,13:51 @12:04:14.43 Pressure.coast @12:04:14.99 Bypass.open @12:04:15.75 Intake.close @12:04:16.93 Toroid.reconfigure :hold @12:04:17.06 Toroid.seek :crack @12:04:18.86 Toroid.seek :cartridge @12:04:19.85 Toroid.reconfigure :default @12:04:19.90 Toroid.seek :clear @12:04:21.74 Pressurizing cartridge to 26psi above 14.5psia @12:04:21.76 SamplePump.setPosition! 0ml @12:04:21.79 Intake.open @12:04:22.59 Bypass.close @12:04:23.31 Exhaust.close @12:04:24.43 Pressure.reconfigure :default @12:04:24.55 Pressure.hold @12:04:25.08 SamplePump.reconfigure :charge @12:04:25.27 SamplePump.seek 2.3ml,42 seconds @12:04:36.18 Pressure.coast @12:04:36.29 Intake.close @12:04:37.39 Pressurized cartridge with 0.531ml to 26.4psi @12:04:37.40 Isolated @34.9psia (20.4psi) @12:04:37.41 Checking for cartridge leaks from 34.9psia @12:04:37.43 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:04:42.52 Cartridge pressure fell 6% to 32.6psia @12:04:42.61 Sampling 2000.0ml @12:04:42.62 SamplePump.setPosition! 0ml @12:04:42.67 Exhaust.open @12:04:43.45 Intake.open @12:04:44.62 Pressure.reconfigure :default @12:04:44.76 Pressure.hold @12:04:45.45 SamplePump.reconfigure :sample @12:04:45.64 SamplePump.seek 2000ml,2:47:12 @12:04:47.75 Cartridge 31 sampling 1/2000ml at 0.57ml/s with 24.9psi vacuum @12:05:47.84 Cartridge 31 sampling 33/2000ml at 0.54ml/s with 25.8psi vacuum @12:06:47.64 Cartridge 31 sampling 65/2000ml at 0.53ml/s with 26.0psi vacuum @12:07:48.08 Cartridge 31 sampling 98/2000ml at 0.54ml/s with 25.6psi vacuum @12:08:47.63 Cartridge 31 sampling 130/2000ml at 0.55ml/s with 25.1psi vacuum @12:09:47.81 Cartridge 31 sampling 163/2000ml at 0.55ml/s with 20.6psi vacuum @12:10:47.64 Cartridge 31 sampling 196/2000ml at 0.54ml/s with 17.0psi vacuum @12:11:47.75 Cartridge 31 sampling 229/2000ml at 0.53ml/s with 25.0psi vacuum @12:12:47.63 Cartridge 31 sampling 263/2000ml at 0.54ml/s with 27.5psi vacuum @12:13:47.80 Cartridge 31 sampling 297/2000ml at 0.54ml/s with 17.8psi vacuum @12:14:47.64 Cartridge 31 sampling 330/2000ml at 0.54ml/s with 24.6psi vacuum @12:15:47.77 Cartridge 31 sampling 364/2000ml at 0.56ml/s with 23.2psi vacuum @12:16:47.64 Cartridge 31 sampling 397/2000ml at 0.57ml/s with 25.1psi vacuum @12:17:47.80 Cartridge 31 sampling 430/2000ml at 0.55ml/s with 19.2psi vacuum @12:18:47.64 Cartridge 31 sampling 464/2000ml at 0.54ml/s with 28.6psi vacuum @12:19:47.67 Cartridge 31 sampling 497/2000ml at 0.57ml/s with 23.5psi vacuum @12:20:47.87 Cartridge 31 sampling 531/2000ml at 0.57ml/s with 20.5psi vacuum @12:21:47.64 Cartridge 31 sampling 564/2000ml at 0.56ml/s with 27.6psi vacuum @12:22:47.85 Cartridge 31 sampling 598/2000ml at 0.56ml/s with 22.7psi vacuum @12:23:47.63 Cartridge 31 sampling 631/2000ml at 0.55ml/s with 16.3psi vacuum @12:24:47.90 Cartridge 31 sampling 664/2000ml at 0.55ml/s with 18.6psi vacuum 1.9m deep (17.5psia) @12:25:47.64 Cartridge 31 sampling 697/2000ml at 0.56ml/s with 25.4psi vacuum @12:26:47.82 Cartridge 31 sampling 731/2000ml at 0.58ml/s with 21.9psi vacuum @12:27:47.64 Cartridge 31 sampling 765/2000ml at 0.55ml/s with 16.6psi vacuum @12:28:47.85 Cartridge 31 sampling 798/2000ml at 0.57ml/s with 21.7psi vacuum @12:29:47.64 Cartridge 31 sampling 831/2000ml at 0.56ml/s with 25.7psi vacuum @12:30:47.78 Cartridge 31 sampling 865/2000ml at 0.56ml/s with 25.8psi vacuum @12:31:47.64 Cartridge 31 sampling 898/2000ml at 0.55ml/s with 15.8psi vacuum 3.7m deep (20.2psia) @12:32:47.87 Cartridge 31 sampling 931/2000ml at 0.56ml/s with 27.0psi vacuum @12:33:48.04 Cartridge 31 sampling 964/2000ml at 0.53ml/s with 19.1psi vacuum @12:34:47.63 Cartridge 31 sampling 997/2000ml at 0.54ml/s with 26.1psi vacuum @12:35:48.10 Cartridge 31 sampling 1030/2000ml at 0.52ml/s with 28.4psi vacuum @12:36:47.64 Cartridge 31 sampling 1063/2000ml at 0.54ml/s with 22.4psi vacuum @12:37:47.76 Cartridge 31 sampling 1095/2000ml at 0.53ml/s with 16.5psi vacuum 5.0m deep (22.1psia) @12:38:47.64 Cartridge 31 sampling 1127/2000ml at 0.56ml/s with 23.8psi vacuum @12:39:47.77 Cartridge 31 sampling 1160/2000ml at 0.54ml/s with 27.6psi vacuum @12:40:47.63 Cartridge 31 sampling 1192/2000ml at 0.54ml/s with 20.0psi vacuum 2.3m deep (18.0psia) @12:41:47.76 Cartridge 31 sampling 1224/2000ml at 0.53ml/s with 15.7psi vacuum 5.8m deep (23.3psia) @12:42:47.63 Cartridge 31 sampling 1256/2000ml at 0.54ml/s with 29.1psi vacuum @12:43:47.77 Cartridge 31 sampling 1288/2000ml at 0.54ml/s with 27.1psi vacuum @12:44:47.64 Cartridge 31 sampling 1320/2000ml at 0.51ml/s with 30.8psi vacuum @12:45:47.66 Cartridge 31 sampling 1351/2000ml at 0.53ml/s with 27.8psi vacuum @12:46:47.64 Cartridge 31 sampling 1382/2000ml at 0.51ml/s with 16.0psi vacuum 2.6m deep (18.5psia) @12:47:47.72 Cartridge 31 sampling 1414/2000ml at 0.53ml/s with 28.1psi vacuum @12:48:47.92 Cartridge 31 sampling 1445/2000ml at 0.51ml/s with 22.4psi vacuum 2.0m deep (17.6psia) @12:49:47.62 Cartridge 31 sampling 1476/2000ml at 0.49ml/s with 14.7psi vacuum @12:50:48.00 Cartridge 31 sampling 1507/2000ml at 0.52ml/s with 25.9psi vacuum @12:51:47.63 Cartridge 31 sampling 1537/2000ml at 0.52ml/s with 16.5psi vacuum 4.0m deep (20.7psia) @12:52:47.82 Cartridge 31 sampling 1568/2000ml at 0.52ml/s with 14.5psi vacuum 6.3m deep (24.0psia) @12:53:47.62 Cartridge 31 sampling 1599/2000ml at 0.51ml/s with 16.1psi vacuum 4.2m deep (20.9psia) @12:54:47.82 Cartridge 31 sampling 1629/2000ml at 0.52ml/s with 25.7psi vacuum @12:55:47.62 Cartridge 31 sampling 1659/2000ml at 0.49ml/s with 28.5psi vacuum @12:56:47.78 Cartridge 31 sampling 1689/2000ml at 0.50ml/s with 28.6psi vacuum @12:57:47.63 Cartridge 31 sampling 1719/2000ml at 0.50ml/s with 29.0psi vacuum @12:58:47.79 Cartridge 31 sampling 1749/2000ml at 0.51ml/s with 26.0psi vacuum @12:59:47.63 Cartridge 31 sampling 1779/2000ml at 0.48ml/s with 30.8psi vacuum @13:00:47.68 Cartridge 31 sampling 1808/2000ml at 0.46ml/s with 30.6psi vacuum @13:01:47.71 Cartridge 31 sampling 1838/2000ml at 0.50ml/s with 28.5psi vacuum @13:02:47.63 Cartridge 31 sampling 1867/2000ml at 0.51ml/s with 24.5psi vacuum @13:03:48.02 Cartridge 31 sampling 1896/2000ml at 0.49ml/s with 13.4psi vacuum 8.4m deep (27.1psia) @13:04:47.63 Cartridge 31 sampling 1925/2000ml at 0.48ml/s with 18.8psi vacuum @13:05:47.78 Cartridge 31 sampling 1954/2000ml at 0.50ml/s with 25.7psi vacuum @13:06:47.64 Cartridge 31 sampling 1983/2000ml at 0.45ml/s with 27.9psi vacuum @13:07:22.19 Pressure.coast @13:07:22.57 Intake.close @13:07:22.83 Exhaust.close @13:07:23.07 Toroid.reconfigure :hold @13:07:23.13 Toroid.seek :bypass @13:07:24.74 Toroid.reconfigure :default @13:07:24.80 Toroid.seek :clear @13:07:26.41 Sampled 2000.0ml @13:07:26.86 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 31" @13:07:27.53 SamplePump.setPosition! 0ml @13:07:27.62 Plunge.seek :reagent @13:07:28.16 Bypass.open @13:07:29.12 Sample loop closed. Ready to Deliver Reagents @13:07:29.14 Archive.seek :sealed @13:07:29.33 Intake.open @13:07:30.39 SamplePump.reconfigure :slow @13:07:30.61 SamplePump.seek -3ml @13:07:32.84 Archive.seek :evac @13:07:37.37 delay 1 second @13:07:38.68 SamplePump.seek -2.8ml @13:07:39.56 delay 1 second @13:07:40.70 SamplePump.seek -3.3ml @13:07:42.13 delay 1 second @13:07:43.18 Exhaust.open @13:07:43.88 Bypass.close @13:07:44.66 SamplePump.seek -35.3ml @13:07:46.86 delay 1:00 @13:07:52.63 upsync @13:08:46.91 Archive.seek :clear @13:08:53.48 delay 3 seconds @13:08:57.34 Intake.close @13:08:57.38 Exhaust.close @13:09:04.03 RNAlater soak for 10:00 @13:19:04.07 Archive.seek :tVent @13:19:05.71 delay 1 second @13:19:06.77 Archive.seek :evac @13:19:22.80 delay 10 seconds @13:19:32.84 Sample preserved and evacuated @13:19:33.07
Carousel.seek 31 @13:19:33.28 Plunge.seek :locked @13:19:34.54 Archive.seek :clear @13:19:35.01 Air.seek :clear @13:19:35.42 Lysis1.seek :clear @13:19:51.87
Archive.coast @13:19:51.91 Air.coast @13:19:51.95 Lysis1.coast @13:19:52.00 Lysis2.coast @13:19:52.04 Plunge.coast @13:19:52.08 SamplePump.coast @13:19:52.13 Toroid.coast @13:19:52.17 Carousel.coast @13:19:52.36 Gate.power :motor,:OFF @13:19:52.39 Gate.power :core,:OFF @13:19:52.42 Gate.power :analytic1,:OFF @13:19:52.44 Gate.power :analytic2,:OFF @13:19:52.46 Gate.power :analytic3,:OFF @13:19:52.50 Gate.power :main,:OFF @13:19:52.52 Safely stopped and ready to power off @13:19:53.41 delayUntil "00:00:00.00PDT22-Jun-21" @13:19:53.57 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT22-Jun-21" @13:20:10.92 upsync @22:00:48.76 Session Begun @22:00:58.55 -> Cartridge.dry @22:01:18.42 -> Cartridge.reserved @22:02:06.93 -> Cartridge.release 29..21 @22:02:14.67 -> Cartridge.dry @22:02:24.84 -> slots @22:04:39.85 -> exit @22:04:39.87 Session Ended @00:00:00.36PDT22-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.39 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.43 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.47 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.62 Powering core bus @00:00:03.31 Archive.configure :default @00:00:03.59 Air.configure :default @00:00:03.75 Lysis1.configure :default @00:00:03.98 Lysis2.configure :default @00:00:04.28 Plunge.configure :default @00:00:04.62 SamplePump.configure :default @00:00:05.19 ExtraPump.configure :default @00:00:05.46 Toroid.configure :default @00:00:05.61 Carousel.configure :default @00:00:05.74 Pressure.configure :default @00:00:05.80 All elf controllers running firmware version 4.61 @00:00:05.82 Gate.power :motor,:ON,limit to 5000mA @00:00:06.63 Archive.reconfigure :default @00:00:06.95 Archive.home.jog 0 @00:00:07.18 Plunge.reconfigure :default @00:00:07.24 Lysis2.reconfigure :default @00:00:07.28 Lysis1.reconfigure :default @00:00:07.32 Toroid.reconfigure :default @00:00:07.36 Air.reconfigure :default @00:00:07.58 Plunge.home.jog 0 @00:00:07.62 Lysis1.home.jog 0 @00:00:07.64 Lysis2.home.jog 0 @00:00:07.67 Toroid.home.jog 0 @00:00:07.72 Air.home.jog 0 @00:00:08.19 Archive.reconfigure :homing @00:00:08.42 Archive.jog 1370 @00:00:08.48 Toroid.reconfigure :homing @00:00:08.51 Plunge.reconfigure :homing @00:00:08.55 Lysis2.reconfigure :homing @00:00:08.60 Air.reconfigure :homing @00:00:08.66 Lysis1.reconfigure :homing @00:00:08.84 Toroid.jog 617 @00:00:08.89 Plunge.jog 848 @00:00:08.95 Air.jog 1360 @00:00:09.00 Lysis2.jog 1358 @00:00:09.11 Lysis1.jog 1353 @00:00:09.84 Archive.setRawPosition! 50 @00:00:10.06 Archive.seek :home @00:00:10.74 Plunge.setRawPosition! 50 @00:00:10.79 Lysis2.setRawPosition! 50 @00:00:10.84 Lysis1.setRawPosition! 50 @00:00:10.96 Archive.reconfigure :default @00:00:11.01 Air.setRawPosition! 50 @00:00:11.51 Lysis2.seek :home @00:00:11.60 Plunge.seek :home @00:00:11.65 Lysis1.seek :home @00:00:11.72 Air.seek :home @00:00:12.50 Toroid.setRawPosition! 0 @00:00:12.56 Plunge.reconfigure :default @00:00:12.60 Lysis1.reconfigure :default @00:00:12.70 Lysis2.reconfigure :default @00:00:12.83 Air.reconfigure :default @00:00:13.09 Toroid.seek :home @00:00:13.67 Toroid.reconfigure :default @00:00:14.06
Toroid.seek :clear @00:00:15.68 Carousel.reconfigure :default @00:00:15.73 Carousel.home.jog 0 @00:00:16.22 Carousel.seek 31 @00:00:16.38 Selecting Cartridge 30 @00:00:16.59 Carousel.seek 30 @00:00:17.10 Lysis1.seek :locked @00:00:20.61 Plunge.seek :locked @00:00:21.15 Air.seek :locked @00:00:25.42
Mounted archive Cartridge @00:00:25.52 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 30" @00:00:25.85 Duration of filtering limited to 4:05:00 @00:00:26.14 Priming sample loop w/160ml, bypass w/40ml @00:00:26.16 SamplePump.setPosition! 0ml @00:00:26.20 Bypass.open @00:00:27.71 Intake.open @00:00:28.57 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:31.23 Vacuum tare is 18% of full scale @00:00:31.30 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:31.91 Pressure.reconfigure :default @00:00:32.02 Pressure.hold @00:00:32.66 SamplePump.reconfigure :prime @00:00:32.74 SamplePump.seek 40ml,3:51 @00:00:39.41 upsync @00:01:16.13 Pressure.coast @00:01:16.25 SamplePump.setPosition! 0ml @00:01:16.35 Exhaust.open @00:01:17.85 Bypass.close @00:01:19.70 Pressure.reconfigure :default @00:01:19.76 Pressure.hold @00:01:21.44 SamplePump.reconfigure :prime @00:01:21.50 SamplePump.seek 160ml,13:51 @00:04:13.13 Pressure.coast @00:04:13.24 Bypass.open @00:04:14.09 Intake.close @00:04:14.95 Toroid.reconfigure :hold @00:04:15.09 Toroid.seek :crack @00:04:16.89 Toroid.seek :cartridge @00:04:17.83 Toroid.reconfigure :default @00:04:17.89 Toroid.seek :clear @00:04:19.73 Pressurizing cartridge to 26psi above 14.6psia @00:04:19.75 SamplePump.setPosition! 0ml @00:04:19.79 Intake.open @00:04:20.60 Bypass.close @00:04:21.39 Exhaust.close @00:04:22.41 Pressure.reconfigure :default @00:04:22.56 Pressure.hold @00:04:23.05 SamplePump.reconfigure :charge @00:04:23.22 SamplePump.seek 2.3ml,42 seconds @00:04:34.82 Pressure.coast @00:04:34.87 Intake.close @00:04:35.65 Pressurized cartridge with 0.600ml to 26.0psi @00:04:35.66 Isolated @35.9psia (21.4psi) @00:04:35.67 Checking for cartridge leaks from 35.9psia @00:04:35.69 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:40.43 Cartridge pressure fell 3% to 34.8psia @00:04:40.56 Sampling 2000.0ml @00:04:40.62 SamplePump.setPosition! 0ml @00:04:40.74 Exhaust.open @00:04:41.44 Intake.open @00:04:42.68 Pressure.reconfigure :default @00:04:42.82 Pressure.hold @00:04:43.38 SamplePump.reconfigure :sample @00:04:43.53 SamplePump.seek 2000ml,2:47:12 @00:04:45.56 Cartridge 30 sampling 1/2000ml at 0.57ml/s with 27.4psi vacuum @00:05:45.72 Cartridge 30 sampling 32/2000ml at 0.53ml/s with 21.2psi vacuum @00:06:45.55 Cartridge 30 sampling 64/2000ml at 0.52ml/s with 21.3psi vacuum @00:07:45.69 Cartridge 30 sampling 95/2000ml at 0.53ml/s with 24.7psi vacuum @00:08:45.56 Cartridge 30 sampling 127/2000ml at 0.52ml/s with 26.4psi vacuum @00:09:45.66 Cartridge 30 sampling 158/2000ml at 0.53ml/s with 23.5psi vacuum @00:10:45.56 Cartridge 30 sampling 190/2000ml at 0.51ml/s with 26.0psi vacuum @00:11:45.63 Cartridge 30 sampling 222/2000ml at 0.52ml/s with 22.8psi vacuum @00:12:45.56 Cartridge 30 sampling 254/2000ml at 0.54ml/s with 26.8psi vacuum @00:13:45.85 Cartridge 30 sampling 287/2000ml at 0.53ml/s with 18.3psi vacuum 2.7m deep (18.6psia) @00:14:46.05 Cartridge 30 sampling 319/2000ml at 0.54ml/s with 25.0psi vacuum @00:15:45.79 Cartridge 30 sampling 351/2000ml at 0.52ml/s with 17.2psi vacuum @00:16:45.85 Cartridge 30 sampling 383/2000ml at 0.51ml/s with 28.7psi vacuum @00:17:45.55 Cartridge 30 sampling 414/2000ml at 0.53ml/s with 19.9psi vacuum @00:18:45.83 Cartridge 30 sampling 446/2000ml at 0.52ml/s with 16.5psi vacuum 2.8m deep (18.8psia) @00:19:45.56 Cartridge 30 sampling 478/2000ml at 0.52ml/s with 27.4psi vacuum @00:20:45.75 Cartridge 30 sampling 509/2000ml at 0.52ml/s with 23.8psi vacuum @00:21:45.56 Cartridge 30 sampling 541/2000ml at 0.51ml/s with 19.5psi vacuum @00:22:45.84 Cartridge 30 sampling 572/2000ml at 0.50ml/s with 29.3psi vacuum @00:23:45.55 Cartridge 30 sampling 603/2000ml at 0.51ml/s with 28.4psi vacuum @00:24:45.72 Cartridge 30 sampling 634/2000ml at 0.52ml/s with 23.8psi vacuum @00:25:45.56 Cartridge 30 sampling 664/2000ml at 0.51ml/s with 28.5psi vacuum @00:26:45.64 Cartridge 30 sampling 695/2000ml at 0.52ml/s with 21.1psi vacuum @00:27:45.56 Cartridge 30 sampling 726/2000ml at 0.52ml/s with 19.9psi vacuum @00:28:45.56 Cartridge 30 sampling 756/2000ml at 0.52ml/s with 22.9psi vacuum @00:29:45.97 Cartridge 30 sampling 786/2000ml at 0.49ml/s with 16.8psi vacuum 5.0m deep (22.2psia) @00:30:45.55 Cartridge 30 sampling 817/2000ml at 0.51ml/s with 20.3psi vacuum 2.0m deep (17.6psia) @00:31:45.81 Cartridge 30 sampling 847/2000ml at 0.50ml/s with 28.0psi vacuum @00:32:45.56 Cartridge 30 sampling 877/2000ml at 0.50ml/s with 18.6psi vacuum 2.2m deep (18.0psia) @00:33:45.72 Cartridge 30 sampling 907/2000ml at 0.51ml/s with 19.9psi vacuum 3.2m deep (19.4psia) @00:34:45.55 Cartridge 30 sampling 936/2000ml at 0.52ml/s with 26.4psi vacuum @00:35:45.75 Cartridge 30 sampling 966/2000ml at 0.48ml/s with 15.2psi vacuum 3.6m deep (20.0psia) @00:36:45.55 Cartridge 30 sampling 996/2000ml at 0.51ml/s with 20.3psi vacuum 2.3m deep (18.0psia) @00:37:45.77 Cartridge 30 sampling 1025/2000ml at 0.49ml/s with 30.0psi vacuum @00:38:45.56 Cartridge 30 sampling 1055/2000ml at 0.48ml/s with 29.9psi vacuum @00:39:45.75 Cartridge 30 sampling 1084/2000ml at 0.49ml/s with 16.1psi vacuum 6.4m deep (24.2psia) @00:40:45.56 Cartridge 30 sampling 1113/2000ml at 0.49ml/s with 28.7psi vacuum @00:41:45.72 Cartridge 30 sampling 1142/2000ml at 0.49ml/s with 27.1psi vacuum @00:42:45.55 Cartridge 30 sampling 1171/2000ml at 0.50ml/s with 25.5psi vacuum @00:43:45.56 Cartridge 30 sampling 1200/2000ml at 0.47ml/s with 14.3psi vacuum 3.6m deep (20.0psia) @00:44:45.76 Cartridge 30 sampling 1228/2000ml at 0.45ml/s with 14.0psi vacuum 7.5m deep (25.8psia) @00:45:45.56 Cartridge 30 sampling 1257/2000ml at 0.48ml/s with 29.6psi vacuum @00:46:46.03 Cartridge 30 sampling 1285/2000ml at 0.49ml/s with 23.9psi vacuum @00:47:45.56 Cartridge 30 sampling 1314/2000ml at 0.45ml/s with 30.8psi vacuum @00:48:45.75 Cartridge 30 sampling 1342/2000ml at 0.48ml/s with 28.0psi vacuum @00:49:45.55 Cartridge 30 sampling 1370/2000ml at 0.49ml/s with 22.9psi vacuum @00:50:45.74 Cartridge 30 sampling 1398/2000ml at 0.47ml/s with 25.7psi vacuum @00:51:45.56 Cartridge 30 sampling 1426/2000ml at 0.48ml/s with 22.4psi vacuum @00:52:45.81 Cartridge 30 sampling 1454/2000ml at 0.45ml/s with 14.0psi vacuum @00:53:45.56 Cartridge 30 sampling 1481/2000ml at 0.42ml/s with 26.8psi vacuum @00:54:45.63 Cartridge 30 sampling 1509/2000ml at 0.46ml/s with 27.6psi vacuum @00:55:46.51 Cartridge 30 sampling 1536/2000ml at 0.42ml/s with 18.3psi vacuum @00:56:45.70 Cartridge 30 sampling 1564/2000ml at 0.45ml/s with 28.8psi vacuum @00:57:45.75 Cartridge 30 sampling 1591/2000ml at 0.46ml/s with 25.0psi vacuum @00:58:45.56 Cartridge 30 sampling 1618/2000ml at 0.43ml/s with 30.5psi vacuum @00:59:45.90 Cartridge 30 sampling 1645/2000ml at 0.46ml/s with 26.9psi vacuum @01:00:45.56 Cartridge 30 sampling 1671/2000ml at 0.46ml/s with 18.9psi vacuum 1.8m deep (17.3psia) @01:01:45.71 Cartridge 30 sampling 1698/2000ml at 0.42ml/s with 15.9psi vacuum 3.0m deep (19.1psia) @01:02:45.56 Cartridge 30 sampling 1724/2000ml at 0.40ml/s with 26.3psi vacuum @01:03:45.70 Cartridge 30 sampling 1751/2000ml at 0.45ml/s with 18.4psi vacuum 5.3m deep (22.5psia) @01:04:45.56 Cartridge 30 sampling 1777/2000ml at 0.44ml/s with 27.6psi vacuum @01:05:45.81 Cartridge 30 sampling 1803/2000ml at 0.44ml/s with 20.7psi vacuum 2.7m deep (18.8psia) @01:06:45.56 Cartridge 30 sampling 1829/2000ml at 0.44ml/s with 13.8psi vacuum 5.0m deep (22.1psia) @01:07:45.78 Cartridge 30 sampling 1855/2000ml at 0.43ml/s with 16.9psi vacuum 5.4m deep (22.6psia) @01:08:45.56 Cartridge 30 sampling 1880/2000ml at 0.40ml/s with 30.4psi vacuum @01:09:45.57 Cartridge 30 sampling 1906/2000ml at 0.40ml/s with 30.4psi vacuum @01:10:45.81 Cartridge 30 sampling 1931/2000ml at 0.44ml/s with 23.4psi vacuum @01:11:45.84 Cartridge 30 sampling 1957/2000ml at 0.39ml/s with 16.9psi vacuum 4.6m deep (21.5psia) @01:12:45.69 Cartridge 30 sampling 1982/2000ml at 0.44ml/s with 23.2psi vacuum @01:13:28.74 Pressure.coast @01:13:28.88 Intake.close @01:13:28.96 Exhaust.close @01:13:29.13 Toroid.reconfigure :hold @01:13:29.18 Toroid.seek :bypass @01:13:30.80 Toroid.reconfigure :default @01:13:30.86 Toroid.seek :clear @01:13:32.47 Sampled 2000.0ml @01:13:33.78 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 30" @01:13:34.90 SamplePump.setPosition! 0ml @01:13:35.23 Plunge.seek :reagent @01:13:35.27 Bypass.open @01:13:36.07 Intake.open @01:13:36.35 Sample loop closed. Ready to Deliver Reagents @01:13:36.61 Archive.seek :sealed @01:13:37.00 SamplePump.reconfigure :slow @01:13:37.07 SamplePump.seek -3ml @01:13:40.27 Archive.seek :evac @01:13:43.83 delay 1 second @01:13:44.99 SamplePump.seek -2.8ml @01:13:45.78 delay 1 second @01:13:46.01 upsync @01:13:46.96 SamplePump.seek -3.3ml @01:13:48.29 delay 1 second @01:13:49.32 Exhaust.open @01:13:50.03 Bypass.close @01:13:51.01 SamplePump.seek -35.3ml @01:13:54.26 delay 1:00 @01:14:54.30 Archive.seek :clear @01:14:59.85 delay 3 seconds @01:15:02.95 Intake.close @01:15:02.99 Exhaust.close @01:15:11.43 RNAlater soak for 10:00 @01:25:11.49 Archive.seek :tVent @01:25:13.16 delay 1 second @01:25:14.20 Archive.seek :evac @01:25:30.23 delay 10 seconds @01:25:40.32 Sample preserved and evacuated @01:25:40.74
Carousel.seek 30 @01:25:41.17 Plunge.seek :locked @01:25:42.19 Archive.seek :clear @01:25:42.36 Air.seek :clear @01:25:42.52 Lysis1.seek :clear @01:25:59.38
Archive.coast @01:25:59.44 Air.coast @01:25:59.53 Lysis1.coast @01:25:59.71 Lysis2.coast @01:25:59.80 Plunge.coast @01:26:00.01 SamplePump.coast @01:26:00.18 Toroid.coast @01:26:00.23 Carousel.coast @01:26:00.42 Gate.power :motor,:OFF @01:26:00.43 Gate.power :core,:OFF @01:26:00.45 Gate.power :analytic1,:OFF @01:26:00.46 Gate.power :analytic2,:OFF @01:26:00.47 Gate.power :analytic3,:OFF @01:26:00.49 Gate.power :main,:OFF @01:26:00.50 Safely stopped and ready to power off @01:26:00.72 delayUntil "12:00:00.00" @01:26:00.74 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @01:26:13.16 upsync @07:39:49.62 Session Begun @07:39:57.61 -> can @07:40:12.11 -> exit @07:40:12.13 Session Ended @12:00:00.12
Gate.power :main,:ON,limit to 5000mA @12:00:00.14 Gate.power :analytic1,:ON,limit to 5000mA @12:00:00.15 Gate.power :analytic2,:ON,limit to 5000mA @12:00:00.16 Gate.power :analytic3,:ON,limit to 5000mA @12:00:00.20 Powering core bus @12:00:02.77 Archive.configure :default @12:00:03.06 Air.configure :default @12:00:03.32 Lysis1.configure :default @12:00:03.51 Lysis2.configure :default @12:00:03.90 Plunge.configure :default @12:00:04.20 SamplePump.configure :default @12:00:04.50 ExtraPump.configure :default @12:00:04.91 Toroid.configure :default @12:00:05.16 Carousel.configure :default @12:00:05.41 Pressure.configure :default @12:00:05.60 All elf controllers running firmware version 4.61 @12:00:05.62 Gate.power :motor,:ON,limit to 5000mA @12:00:06.23 Archive.reconfigure :default @12:00:06.78 Plunge.reconfigure :default @12:00:06.82 Archive.home.jog 0 @12:00:06.85 Toroid.reconfigure :default @12:00:06.88 Lysis1.reconfigure :default @12:00:06.91 Air.reconfigure :default @12:00:06.97 Lysis2.reconfigure :default @12:00:07.20 Plunge.home.jog 0 @12:00:07.24 Lysis1.home.jog 0 @12:00:07.29 Toroid.home.jog 0 @12:00:07.34 Lysis2.home.jog 0 @12:00:07.38 Air.home.jog 0 @12:00:08.05 Plunge.reconfigure :homing @12:00:08.09 Lysis2.reconfigure :homing @12:00:08.12 Toroid.reconfigure :homing @12:00:08.16 Lysis1.reconfigure :homing @12:00:08.20 Archive.reconfigure :homing @12:00:08.38 Plunge.jog 847 @12:00:08.43 Lysis2.jog 1357 @12:00:08.47 Toroid.jog 617 @12:00:08.50 Archive.jog 1371 @12:00:08.54 Lysis1.jog 1350 @12:00:08.78 Air.reconfigure :homing @12:00:08.84 Air.jog 1358 @12:00:09.43 Plunge.setRawPosition! 50 @12:00:09.48 Plunge.seek :home @12:00:09.72 Lysis2.setRawPosition! 50 @12:00:09.80 Lysis2.seek :home @12:00:09.90 Lysis1.setRawPosition! 50 @12:00:09.92 Archive.setRawPosition! 50 @12:00:09.96 Plunge.reconfigure :default @12:00:10.10 Lysis1.seek :home @12:00:10.14 Archive.seek :home @12:00:10.35 Lysis2.reconfigure :default @12:00:10.37 Air.setRawPosition! 50 @12:00:10.50 Air.seek :home @12:00:12.31 Lysis1.reconfigure :default @12:00:12.40 Toroid.setRawPosition! 0 @12:00:12.45 Archive.reconfigure :default @12:00:12.74 Toroid.seek :home @12:00:13.22 Air.reconfigure :default @12:00:13.35 Toroid.reconfigure :default @12:00:13.95
Toroid.seek :clear @12:00:15.70 Carousel.reconfigure :default @12:00:15.75 Carousel.home.jog 0 @12:00:16.39 Carousel.seek 30 @12:00:16.65 Selecting Cartridge 29 @12:00:16.87 Carousel.seek 29 @12:00:17.38 Lysis1.seek :locked @12:00:20.61 Plunge.seek :locked @12:00:20.75 Air.seek :locked @12:00:25.09
Mounted archive Cartridge @12:00:25.39 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 29" @12:00:25.47 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:00:25.63 Retry #1 of 10 after 10:00 @12:00:26.10 Duration of filtering limited to 4:05:00 @12:00:26.39 Priming sample loop w/160ml, bypass w/40ml @12:00:26.41 SamplePump.setPosition! 0ml @12:00:26.44 Bypass.open @12:00:27.87 Intake.open @12:00:28.75 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:00:31.42 Vacuum tare is 18% of full scale @12:00:31.50 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:00:32.10 Pressure.reconfigure :default @12:00:32.19 Pressure.hold @12:00:32.77 SamplePump.reconfigure :prime @12:00:32.92 SamplePump.seek 40ml,3:51 @12:01:16.19 Pressure.coast @12:01:16.24 SamplePump.setPosition! 0ml @12:01:16.29 Exhaust.open @12:01:17.73 Bypass.close @12:01:19.57 Pressure.reconfigure :default @12:01:19.79 Pressure.hold @12:01:20.54 SamplePump.reconfigure :prime @12:01:20.60 SamplePump.seek 160ml,13:51 @12:04:12.38 Pressure.coast @12:04:12.52 Bypass.open @12:04:13.26 Intake.close @12:04:14.41 Toroid.reconfigure :hold @12:04:14.55 Toroid.seek :crack @12:04:16.50 Toroid.seek :cartridge @12:04:17.56 Toroid.reconfigure :default @12:04:17.61 Toroid.seek :clear @12:04:19.46 Pressurizing cartridge to 26psi above 14.7psia @12:04:19.47 SamplePump.setPosition! 0ml @12:04:19.51 Intake.open @12:04:20.28 Bypass.close @12:04:20.95 Exhaust.close @12:04:22.21 Pressure.reconfigure :default @12:04:22.30 Pressure.hold @12:04:22.85 SamplePump.reconfigure :charge @12:04:23.01 SamplePump.seek 2.3ml,42 seconds @12:04:34.03 Pressure.coast @12:04:34.08 Intake.close @12:04:35.10 Pressurized cartridge with 0.535ml to 26.5psi @12:04:35.13 Isolated @33.0psia (18.3psi) @12:04:35.15 Checking for cartridge leaks from 33.0psia @12:04:35.21 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:04:39.89 Cartridge pressure fell 5% to 31.4psia @12:04:40.29 Sampling 2000.0ml @12:04:40.33 SamplePump.setPosition! 0ml @12:04:40.43 Exhaust.open @12:04:41.27 Intake.open @12:04:42.60 Pressure.reconfigure :default @12:04:42.72 Pressure.hold @12:04:43.58 SamplePump.reconfigure :sample @12:04:43.79 SamplePump.seek 2000ml,2:47:12 @12:04:45.56 Cartridge 29 sampling 1/2000ml at 0.60ml/s with 21.9psi vacuum @12:05:45.83 Cartridge 29 sampling 32/2000ml at 0.52ml/s with 23.4psi vacuum @12:06:45.58 Cartridge 29 sampling 63/2000ml at 0.53ml/s with 24.0psi vacuum @12:07:45.74 Cartridge 29 sampling 94/2000ml at 0.53ml/s with 25.5psi vacuum @12:08:45.58 Cartridge 29 sampling 126/2000ml at 0.55ml/s with 22.4psi vacuum @12:09:45.74 Cartridge 29 sampling 158/2000ml at 0.54ml/s with 24.5psi vacuum @12:10:25.70 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:10:26.29 Retry #2 of 10 after 10:00 @12:10:45.57 Cartridge 29 sampling 190/2000ml at 0.54ml/s with 18.9psi vacuum @12:11:45.80 Cartridge 29 sampling 222/2000ml at 0.56ml/s with 25.7psi vacuum @12:12:45.58 Cartridge 29 sampling 254/2000ml at 0.54ml/s with 18.6psi vacuum @12:13:45.60 Cartridge 29 sampling 286/2000ml at 0.55ml/s with 26.7psi vacuum @12:14:45.59 Cartridge 29 sampling 319/2000ml at 0.52ml/s with 27.6psi vacuum @12:15:45.77 Cartridge 29 sampling 351/2000ml at 0.51ml/s with 27.0psi vacuum @12:16:45.98 Cartridge 29 sampling 383/2000ml at 0.55ml/s with 23.1psi vacuum @12:17:45.58 Cartridge 29 sampling 414/2000ml at 0.53ml/s with 18.5psi vacuum 2.2m deep (17.9psia) @12:18:45.83 Cartridge 29 sampling 446/2000ml at 0.52ml/s with 29.2psi vacuum @12:19:45.58 Cartridge 29 sampling 478/2000ml at 0.54ml/s with 21.6psi vacuum @12:20:26.45 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:20:26.53 Retry #3 of 10 after 10:00 @12:20:45.75 Cartridge 29 sampling 509/2000ml at 0.51ml/s with 29.1psi vacuum @12:21:45.58 Cartridge 29 sampling 541/2000ml at 0.53ml/s with 19.5psi vacuum @12:22:45.73 Cartridge 29 sampling 572/2000ml at 0.49ml/s with 27.2psi vacuum @12:23:45.58 Cartridge 29 sampling 603/2000ml at 0.49ml/s with 18.6psi vacuum @12:24:45.75 Cartridge 29 sampling 634/2000ml at 0.50ml/s with 29.1psi vacuum @12:25:45.58 Cartridge 29 sampling 665/2000ml at 0.51ml/s with 27.0psi vacuum @12:26:45.73 Cartridge 29 sampling 695/2000ml at 0.51ml/s with 16.0psi vacuum 5.6m deep (23.0psia) @12:27:45.57 Cartridge 29 sampling 726/2000ml at 0.53ml/s with 20.8psi vacuum 2.1m deep (17.9psia) @12:28:45.61 Cartridge 29 sampling 757/2000ml at 0.51ml/s with 27.3psi vacuum @12:29:45.58 Cartridge 29 sampling 787/2000ml at 0.48ml/s with 22.9psi vacuum @12:30:26.61 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:30:26.63 Retry #4 of 10 after 10:00 @12:30:45.71 Cartridge 29 sampling 818/2000ml at 0.50ml/s with 27.3psi vacuum @12:31:45.77 Cartridge 29 sampling 848/2000ml at 0.49ml/s with 13.8psi vacuum 5.6m deep (22.9psia) @12:32:45.57 Cartridge 29 sampling 878/2000ml at 0.50ml/s with 15.9psi vacuum 1.7m deep (17.2psia) @12:33:45.92 Cartridge 29 sampling 908/2000ml at 0.51ml/s with 28.8psi vacuum @12:34:45.58 Cartridge 29 sampling 938/2000ml at 0.49ml/s with 18.3psi vacuum @12:35:45.66 Cartridge 29 sampling 967/2000ml at 0.49ml/s with 28.1psi vacuum @12:36:45.57 Cartridge 29 sampling 997/2000ml at 0.48ml/s with 24.6psi vacuum @12:37:45.76 Cartridge 29 sampling 1026/2000ml at 0.49ml/s with 27.2psi vacuum @12:38:45.58 Cartridge 29 sampling 1056/2000ml at 0.48ml/s with 28.7psi vacuum @12:39:45.71 Cartridge 29 sampling 1085/2000ml at 0.50ml/s with 18.6psi vacuum 3.1m deep (19.3psia) @12:40:26.68 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:40:26.70 Retry #5 of 10 after 10:00 @12:40:45.58 Cartridge 29 sampling 1114/2000ml at 0.51ml/s with 23.0psi vacuum @12:41:45.74 Cartridge 29 sampling 1143/2000ml at 0.49ml/s with 21.2psi vacuum @12:42:45.58 Cartridge 29 sampling 1172/2000ml at 0.47ml/s with 29.1psi vacuum @12:43:45.74 Cartridge 29 sampling 1201/2000ml at 0.48ml/s with 28.3psi vacuum @12:44:45.58 Cartridge 29 sampling 1230/2000ml at 0.49ml/s with 21.7psi vacuum @12:45:45.76 Cartridge 29 sampling 1259/2000ml at 0.47ml/s with 24.8psi vacuum @12:46:46.05 Cartridge 29 sampling 1287/2000ml at 0.46ml/s with 28.2psi vacuum @12:47:45.58 Cartridge 29 sampling 1316/2000ml at 0.49ml/s with 23.7psi vacuum @12:48:45.75 Cartridge 29 sampling 1344/2000ml at 0.49ml/s with 24.9psi vacuum @12:49:45.58 Cartridge 29 sampling 1372/2000ml at 0.45ml/s with 12.4psi vacuum 2.9m deep (19.0psia) @12:50:26.78 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @12:50:26.82 Retry #6 of 10 after 10:00 @12:50:45.78 Cartridge 29 sampling 1400/2000ml at 0.46ml/s with 13.7psi vacuum 8.9m deep (27.8psia) @12:51:45.58 Cartridge 29 sampling 1428/2000ml at 0.47ml/s with 14.6psi vacuum 5.1m deep (22.2psia) @12:52:45.75 Cartridge 29 sampling 1456/2000ml at 0.45ml/s with 15.8psi vacuum 7.4m deep (25.6psia) @12:53:45.58 Cartridge 29 sampling 1484/2000ml at 0.45ml/s with 13.1psi vacuum 3.8m deep (20.3psia) @12:54:45.89 Cartridge 29 sampling 1511/2000ml at 0.44ml/s with 12.0psi vacuum 4.8m deep (21.8psia) @12:55:45.57 Cartridge 29 sampling 1539/2000ml at 0.46ml/s with 27.9psi vacuum @12:56:45.74 Cartridge 29 sampling 1566/2000ml at 0.46ml/s with 19.9psi vacuum @12:57:45.58 Cartridge 29 sampling 1593/2000ml at 0.44ml/s with 29.8psi vacuum @12:58:45.70 Cartridge 29 sampling 1620/2000ml at 0.45ml/s with 16.3psi vacuum 4.1m deep (20.7psia) @12:59:45.58 Cartridge 29 sampling 1647/2000ml at 0.45ml/s with 17.9psi vacuum 1.8m deep (17.3psia) @13:00:26.91 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:00:26.95 Retry #7 of 10 after 10:00 @13:00:45.74 Cartridge 29 sampling 1674/2000ml at 0.45ml/s with 29.6psi vacuum @13:01:45.58 Cartridge 29 sampling 1701/2000ml at 0.45ml/s with 29.8psi vacuum @13:02:45.58 Cartridge 29 sampling 1728/2000ml at 0.43ml/s with 12.4psi vacuum 3.2m deep (19.4psia) @13:03:45.81 Cartridge 29 sampling 1755/2000ml at 0.45ml/s with 26.7psi vacuum @13:04:45.57 Cartridge 29 sampling 1781/2000ml at 0.41ml/s with 30.3psi vacuum @13:05:45.73 Cartridge 29 sampling 1807/2000ml at 0.44ml/s with 20.0psi vacuum 1.7m deep (17.2psia) @13:06:45.57 Cartridge 29 sampling 1833/2000ml at 0.40ml/s with 15.1psi vacuum @13:07:45.72 Cartridge 29 sampling 1859/2000ml at 0.41ml/s with 30.2psi vacuum @13:08:45.58 Cartridge 29 sampling 1885/2000ml at 0.42ml/s with 14.1psi vacuum 3.1m deep (19.2psia) @13:09:45.75 Cartridge 29 sampling 1911/2000ml at 0.41ml/s with 30.3psi vacuum @13:10:27.00 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:10:27.02 Retry #8 of 10 after 10:00 @13:10:45.58 Cartridge 29 sampling 1936/2000ml at 0.43ml/s with 24.7psi vacuum @13:11:45.75 Cartridge 29 sampling 1962/2000ml at 0.44ml/s with 20.8psi vacuum @13:12:45.58 Cartridge 29 sampling 1987/2000ml at 0.38ml/s with 27.2psi vacuum @13:13:16.00 Pressure.coast @13:13:16.28 Intake.close @13:13:16.41 Exhaust.close @13:13:17.03 Toroid.reconfigure :hold @13:13:17.22 Toroid.seek :bypass @13:13:19.06 Toroid.reconfigure :default @13:13:19.12 Toroid.seek :clear @13:13:20.73 Sampled 2000.0ml @13:13:21.53 SamplePump.setPosition! 0ml @13:13:21.57 Plunge.seek :reagent @13:13:22.18 Bypass.open @13:13:22.91 Intake.open @13:13:23.06 Sample loop closed. Ready to Deliver Reagents @13:13:23.08 Archive.seek :sealed @13:13:24.21 SamplePump.reconfigure :slow @13:13:24.51 SamplePump.seek -3ml @13:13:26.80 Archive.seek :evac @13:13:31.29 delay 1 second @13:13:32.40 SamplePump.seek -2.8ml @13:13:33.13 delay 1 second @13:13:34.35 SamplePump.seek -3.3ml @13:13:35.70 delay 1 second @13:13:36.72 Exhaust.open @13:13:37.49 Bypass.close @13:13:38.32 SamplePump.seek -35.3ml @13:13:40.81 delay 1:00 @13:14:40.93 Archive.seek :clear @13:14:47.18 delay 3 seconds @13:14:50.49 Intake.close @13:14:50.56 Exhaust.close @13:14:58.11 RNAlater soak for 10:00 @13:20:27.09 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:20:27.15 Retry #9 of 10 after 10:00 @13:24:58.19 Archive.seek :tVent @13:24:59.83 delay 1 second @13:25:00.88 Archive.seek :evac @13:25:16.85 delay 10 seconds @13:25:26.89 Sample preserved and evacuated @13:25:27.12
Carousel.seek 29 @13:25:27.33 Plunge.seek :locked @13:25:28.31 Archive.seek :clear @13:25:28.48 Air.seek :clear @13:25:28.64 Lysis1.seek :clear @13:25:45.49
Archive.coast @13:25:45.53 Air.coast @13:25:45.57 Lysis1.coast @13:25:45.62 Lysis2.coast @13:25:45.66 Plunge.coast @13:25:45.70 SamplePump.coast @13:25:45.74 Toroid.coast @13:25:45.79 Carousel.coast @13:25:45.98 Gate.power :motor,:OFF @13:25:45.99 Gate.power :core,:OFF @13:25:46.01 Gate.power :analytic1,:OFF @13:25:46.02 Gate.power :analytic2,:OFF @13:25:46.03 Gate.power :analytic3,:OFF @13:25:46.04 Gate.power :main,:OFF @13:25:46.06 Safely stopped and ready to power off @13:25:46.43 delayUntil "00:00:00.00PDT23-Jun-21" @13:30:27.25 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:30:27.33 Retry #10 of 10 after 10:00 @13:40:27.37 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:40:27.40 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 29" @13:40:27.43 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:40:27.45 Retry #1 of 10 after 10:00 @13:50:27.51 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @13:50:27.54 Retry #2 of 10 after 10:00 @14:00:27.58 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:00:27.61 Retry #3 of 10 after 10:00 @14:10:28.03 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:10:28.10 Retry #4 of 10 after 10:00 @14:20:28.22 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:20:28.28 Retry #5 of 10 after 10:00 @14:30:28.34 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:30:28.36 Retry #6 of 10 after 10:00 @14:40:28.47 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:40:28.53 Retry #7 of 10 after 10:00 @14:50:40.78 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT23-Jun-21" @14:50:50.82 upsync @14:51:01.99 upsync @14:53:50.97 Session Begun @14:53:55.94 -> Cartridge.dry @14:54:54.45 -> exit @14:54:54.49 Session Ended @00:00:00.18PDT23-Jun-21
Gate.power :main,:ON,limit to 5000mA @00:00:00.29 Gate.power :analytic1,:ON,limit to 5000mA @00:00:00.31 Gate.power :analytic2,:ON,limit to 5000mA @00:00:00.32 Gate.power :analytic3,:ON,limit to 5000mA @00:00:00.36 Powering core bus @00:00:02.68 Archive.configure :default @00:00:02.81 Air.configure :default @00:00:02.92 Lysis1.configure :default @00:00:03.03 Lysis2.configure :default @00:00:03.18 Plunge.configure :default @00:00:03.32 SamplePump.configure :default @00:00:03.47 ExtraPump.configure :default @00:00:03.61 Toroid.configure :default @00:00:03.74 Carousel.configure :default @00:00:03.88 Pressure.configure :default @00:00:03.94 All elf controllers running firmware version 4.61 @00:00:03.96 Gate.power :motor,:ON,limit to 5000mA @00:00:04.09 Archive.reconfigure :default @00:00:04.48 Archive.home.jog 0 @00:00:04.75 Toroid.reconfigure :default @00:00:04.78 Plunge.reconfigure :default @00:00:04.81 Lysis1.reconfigure :default @00:00:04.86 Air.reconfigure :default @00:00:04.91 Lysis2.reconfigure :default @00:00:05.09 Plunge.home.jog 0 @00:00:05.14 Toroid.home.jog 0 @00:00:05.18 Lysis1.home.jog 0 @00:00:05.21 Lysis2.home.jog 0 @00:00:05.25 Air.home.jog 0 @00:00:06.29 Archive.reconfigure :homing @00:00:06.84 Archive.jog 1369 @00:00:07.53 Plunge.reconfigure :homing @00:00:07.63 Lysis2.reconfigure :homing @00:00:07.76 Lysis1.reconfigure :homing @00:00:07.83 Toroid.reconfigure :homing @00:00:07.93 Air.reconfigure :homing @00:00:08.21 Plunge.jog 849 @00:00:08.42 Lysis2.jog 1356 @00:00:08.56 Toroid.jog 627 @00:00:08.63 Lysis1.jog 1352 @00:00:08.68 Air.jog 1360 @00:00:09.30 Archive.setRawPosition! 50 @00:00:09.42 Archive.seek :home @00:00:09.74 Archive.reconfigure :default @00:00:10.01 Lysis2.setRawPosition! 50 @00:00:10.12 Plunge.setRawPosition! 50 @00:00:10.28 Lysis2.seek :home @00:00:10.34 Plunge.seek :home @00:00:10.68 Air.setRawPosition! 50 @00:00:10.74 Lysis1.setRawPosition! 50 @00:00:10.91 Air.seek :home @00:00:10.98 Lysis1.seek :home @00:00:11.73 Toroid.setRawPosition! 0 @00:00:11.78 Plunge.reconfigure :default @00:00:11.79 Lysis1.reconfigure :default @00:00:11.82 Air.reconfigure :default @00:00:11.89 Lysis2.reconfigure :default @00:00:12.08 Toroid.seek :home @00:00:12.26 Toroid.reconfigure :default @00:00:12.50
Toroid.seek :clear @00:00:14.11 Carousel.reconfigure :default @00:00:14.16 Carousel.home.jog 0 @00:00:14.67 Carousel.seek 29 @00:00:14.83 Selecting Cartridge 28 @00:00:15.06 Carousel.seek 28 @00:00:15.58 Lysis1.seek :locked @00:00:19.04 Plunge.seek :locked @00:00:19.40 Air.seek :locked @00:00:23.70
Mounted archive Cartridge @00:00:23.80 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 28" @00:00:24.68 Duration of filtering limited to 4:05:00 @00:00:25.01 Priming sample loop w/160ml, bypass w/40ml @00:00:25.03 SamplePump.setPosition! 0ml @00:00:25.07 Bypass.open @00:00:26.75 Intake.open @00:00:27.91 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @00:00:30.62 Vacuum tare is 17% of full scale @00:00:30.64 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @00:00:30.85 Pressure.reconfigure :default @00:00:30.91 Pressure.hold @00:00:31.12 SamplePump.reconfigure :prime @00:00:31.18 SamplePump.seek 40ml,3:51 @00:00:36.36 upsync @00:01:14.44 Pressure.coast @00:01:14.65 SamplePump.setPosition! 0ml @00:01:14.87 Exhaust.open @00:01:16.32 Bypass.close @00:01:17.95 Pressure.reconfigure :default @00:01:18.01 Pressure.hold @00:01:18.29 SamplePump.reconfigure :prime @00:01:18.36 SamplePump.seek 160ml,13:51 @00:04:10.02 Pressure.coast @00:04:10.08 Bypass.open @00:04:10.96 Intake.close @00:04:11.87 Toroid.reconfigure :hold @00:04:11.92 Toroid.seek :crack @00:04:13.55 Toroid.seek :cartridge @00:04:14.49 Toroid.reconfigure :default @00:04:14.54 Toroid.seek :clear @00:04:16.88 Pressurizing cartridge to 26psi above 14.7psia @00:04:16.96 SamplePump.setPosition! 0ml @00:04:17.02 Intake.open @00:04:17.94 Bypass.close @00:04:18.75 Exhaust.close @00:04:19.72 Pressure.reconfigure :default @00:04:19.83 Pressure.hold @00:04:20.34 SamplePump.reconfigure :charge @00:04:20.64 SamplePump.seek 2.3ml,42 seconds @00:04:31.75 Pressure.coast @00:04:31.80 Intake.close @00:04:32.66 Pressurized cartridge with 0.564ml to 26.2psi @00:04:32.67 Isolated @35.7psia (21.0psi) @00:04:32.68 Checking for cartridge leaks from 35.7psia @00:04:32.70 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @00:04:37.44 Cartridge pressure fell 3% to 34.8psia @00:04:37.69 Sampling 2000.0ml @00:04:37.71 SamplePump.setPosition! 0ml @00:04:37.85 Exhaust.open @00:04:38.56 Intake.open @00:04:39.48 Pressure.reconfigure :default @00:04:39.54 Pressure.hold @00:04:39.81 SamplePump.reconfigure :sample @00:04:39.88 SamplePump.seek 2000ml,2:47:12 @00:04:42.54 Cartridge 28 sampling 1/2000ml at 0.53ml/s with 23.6psi vacuum @00:05:42.67 Cartridge 28 sampling 34/2000ml at 0.54ml/s with 26.3psi vacuum @00:06:42.55 Cartridge 28 sampling 66/2000ml at 0.54ml/s with 23.0psi vacuum @00:07:42.82 Cartridge 28 sampling 98/2000ml at 0.55ml/s with 26.0psi vacuum @00:08:42.55 Cartridge 28 sampling 131/2000ml at 0.55ml/s with 21.9psi vacuum @00:09:42.59 Cartridge 28 sampling 164/2000ml at 0.52ml/s with 18.4psi vacuum @00:10:42.67 Cartridge 28 sampling 197/2000ml at 0.56ml/s with 25.3psi vacuum @00:11:42.54 Cartridge 28 sampling 230/2000ml at 0.57ml/s with 22.2psi vacuum @00:12:42.70 Cartridge 28 sampling 263/2000ml at 0.54ml/s with 17.8psi vacuum 2.8m deep (18.8psia) @00:13:42.54 Cartridge 28 sampling 297/2000ml at 0.54ml/s with 28.7psi vacuum @00:14:42.67 Cartridge 28 sampling 330/2000ml at 0.57ml/s with 25.6psi vacuum @00:15:42.54 Cartridge 28 sampling 363/2000ml at 0.54ml/s with 16.4psi vacuum 1.7m deep (17.3psia) @00:16:42.71 Cartridge 28 sampling 396/2000ml at 0.55ml/s with 26.6psi vacuum @00:17:42.53 Cartridge 28 sampling 429/2000ml at 0.55ml/s with 27.8psi vacuum @00:18:42.70 Cartridge 28 sampling 462/2000ml at 0.52ml/s with 20.3psi vacuum @00:19:42.54 Cartridge 28 sampling 494/2000ml at 0.53ml/s with 29.1psi vacuum @00:20:42.69 Cartridge 28 sampling 526/2000ml at 0.53ml/s with 17.0psi vacuum 5.4m deep (22.7psia) @00:21:42.54 Cartridge 28 sampling 559/2000ml at 0.52ml/s with 15.3psi vacuum @00:22:42.75 Cartridge 28 sampling 591/2000ml at 0.52ml/s with 15.6psi vacuum 4.8m deep (21.8psia) @00:23:42.74 Cartridge 28 sampling 623/2000ml at 0.54ml/s with 21.0psi vacuum @00:24:42.55 Cartridge 28 sampling 655/2000ml at 0.50ml/s with 25.7psi vacuum @00:25:42.72 Cartridge 28 sampling 687/2000ml at 0.50ml/s with 30.3psi vacuum @00:26:42.55 Cartridge 28 sampling 718/2000ml at 0.54ml/s with 26.2psi vacuum @00:27:42.65 Cartridge 28 sampling 750/2000ml at 0.52ml/s with 28.9psi vacuum @00:28:42.55 Cartridge 28 sampling 781/2000ml at 0.53ml/s with 17.6psi vacuum 3.5m deep (19.9psia) @00:29:42.63 Cartridge 28 sampling 812/2000ml at 0.52ml/s with 28.9psi vacuum @00:30:42.54 Cartridge 28 sampling 843/2000ml at 0.52ml/s with 26.5psi vacuum @00:31:42.73 Cartridge 28 sampling 874/2000ml at 0.53ml/s with 23.5psi vacuum @00:32:42.55 Cartridge 28 sampling 904/2000ml at 0.50ml/s with 28.9psi vacuum @00:33:42.79 Cartridge 28 sampling 935/2000ml at 0.49ml/s with 13.7psi vacuum @00:34:42.55 Cartridge 28 sampling 965/2000ml at 0.52ml/s with 25.7psi vacuum @00:35:42.67 Cartridge 28 sampling 995/2000ml at 0.49ml/s with 28.1psi vacuum @00:36:42.55 Cartridge 28 sampling 1025/2000ml at 0.48ml/s with 24.5psi vacuum @00:37:42.56 Cartridge 28 sampling 1054/2000ml at 0.51ml/s with 26.4psi vacuum @00:38:42.87 Cartridge 28 sampling 1084/2000ml at 0.50ml/s with 28.6psi vacuum @00:39:42.54 Cartridge 28 sampling 1113/2000ml at 0.48ml/s with 21.4psi vacuum @00:40:42.77 Cartridge 28 sampling 1143/2000ml at 0.48ml/s with 29.8psi vacuum @00:41:42.54 Cartridge 28 sampling 1172/2000ml at 0.48ml/s with 13.1psi vacuum 5.7m deep (23.1psia) @00:42:42.69 Cartridge 28 sampling 1201/2000ml at 0.48ml/s with 12.3psi vacuum 5.2m deep (22.4psia) @00:43:42.53 Cartridge 28 sampling 1230/2000ml at 0.50ml/s with 26.3psi vacuum @00:44:42.67 Cartridge 28 sampling 1259/2000ml at 0.50ml/s with 23.1psi vacuum @00:45:42.61 Cartridge 28 sampling 1288/2000ml at 0.47ml/s with 11.6psi vacuum 9.8m deep (29.2psia) @00:46:42.64 Cartridge 28 sampling 1316/2000ml at 0.47ml/s with 28.9psi vacuum @00:47:42.55 Cartridge 28 sampling 1345/2000ml at 0.48ml/s with 21.2psi vacuum 2.1m deep (17.9psia) @00:48:42.64 Cartridge 28 sampling 1373/2000ml at 0.45ml/s with 27.9psi vacuum @00:49:42.54 Cartridge 28 sampling 1401/2000ml at 0.47ml/s with 27.5psi vacuum @00:50:42.68 Cartridge 28 sampling 1429/2000ml at 0.46ml/s with 29.7psi vacuum @00:51:42.55 Cartridge 28 sampling 1457/2000ml at 0.46ml/s with 12.8psi vacuum 2.5m deep (18.3psia) @00:52:42.55 Cartridge 28 sampling 1485/2000ml at 0.44ml/s with 29.1psi vacuum @00:53:42.71 Cartridge 28 sampling 1512/2000ml at 0.45ml/s with 15.1psi vacuum 6.8m deep (24.7psia) @00:54:42.55 Cartridge 28 sampling 1539/2000ml at 0.44ml/s with 29.8psi vacuum @00:55:42.77 Cartridge 28 sampling 1567/2000ml at 0.44ml/s with 12.7psi vacuum 8.1m deep (26.7psia) @00:56:42.55 Cartridge 28 sampling 1594/2000ml at 0.46ml/s with 26.9psi vacuum @00:57:42.75 Cartridge 28 sampling 1621/2000ml at 0.46ml/s with 26.0psi vacuum @00:58:42.55 Cartridge 28 sampling 1648/2000ml at 0.46ml/s with 22.9psi vacuum @00:59:42.71 Cartridge 28 sampling 1675/2000ml at 0.44ml/s with 11.4psi vacuum 10.2m deep (29.8psia) @01:00:42.55 Cartridge 28 sampling 1701/2000ml at 0.46ml/s with 24.1psi vacuum @01:01:42.72 Cartridge 28 sampling 1728/2000ml at 0.45ml/s with 28.0psi vacuum @01:02:42.54 Cartridge 28 sampling 1754/2000ml at 0.44ml/s with 29.7psi vacuum @01:03:42.65 Cartridge 28 sampling 1781/2000ml at 0.43ml/s with 10.2psi vacuum 5.9m deep (23.4psia) @01:04:42.54 Cartridge 28 sampling 1807/2000ml at 0.45ml/s with 25.6psi vacuum @01:05:42.75 Cartridge 28 sampling 1832/2000ml at 0.40ml/s with 24.6psi vacuum @01:06:42.74 Cartridge 28 sampling 1858/2000ml at 0.43ml/s with 13.1psi vacuum 5.8m deep (23.2psia) @01:07:42.59 Cartridge 28 sampling 1884/2000ml at 0.39ml/s with 25.6psi vacuum @01:08:42.79 Cartridge 28 sampling 1910/2000ml at 0.41ml/s with 13.5psi vacuum 7.2m deep (25.4psia) @01:09:42.55 Cartridge 28 sampling 1935/2000ml at 0.41ml/s with 25.5psi vacuum @01:10:42.74 Cartridge 28 sampling 1960/2000ml at 0.39ml/s with 29.3psi vacuum @01:11:42.54 Cartridge 28 sampling 1985/2000ml at 0.41ml/s with 16.8psi vacuum 2.2m deep (18.0psia) @01:12:18.81 Pressure.coast @01:12:18.95 Intake.close @01:12:19.03 Exhaust.close @01:12:19.19 Toroid.reconfigure :hold @01:12:19.25 Toroid.seek :bypass @01:12:20.87 Toroid.reconfigure :default @01:12:20.93 Toroid.seek :clear @01:12:22.61 Sampled 2000.0ml @01:12:23.85 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 28" @01:12:25.05 SamplePump.setPosition! 0ml @01:12:25.12 Plunge.seek :reagent @01:12:25.20 Bypass.open @01:12:26.12 Intake.open @01:12:26.38 Sample loop closed. Ready to Deliver Reagents @01:12:26.40 Archive.seek :sealed @01:12:27.48 SamplePump.reconfigure :slow @01:12:27.71 SamplePump.seek -3ml @01:12:30.81 Archive.seek :evac @01:12:34.52 delay 1 second @01:12:34.56 upsync @01:12:35.74 SamplePump.seek -2.8ml @01:12:36.57 delay 1 second @01:12:37.67 SamplePump.seek -3.3ml @01:12:39.00 delay 1 second @01:12:40.03 Exhaust.open @01:12:40.70 Bypass.close @01:12:41.50 SamplePump.seek -35.3ml @01:12:44.90 delay 1:00 @01:13:44.99 Archive.seek :clear @01:13:50.33 delay 3 seconds @01:13:53.43 Intake.close @01:13:53.48 Exhaust.close @01:14:02.12 RNAlater soak for 10:00 @01:24:02.18 Archive.seek :tVent @01:24:03.85 delay 1 second @01:24:04.92 Archive.seek :evac @01:24:20.94 delay 10 seconds @01:24:31.00 Sample preserved and evacuated @01:24:31.57
Carousel.seek 28 @01:24:31.93 Plunge.seek :locked @01:24:32.91 Archive.seek :clear @01:24:33.07 Air.seek :clear @01:24:33.23 Lysis1.seek :clear @01:24:50.12
Archive.coast @01:24:50.29 Air.coast @01:24:50.33 Lysis1.coast @01:24:50.37 Lysis2.coast @01:24:50.50 Plunge.coast @01:24:50.63 SamplePump.coast @01:24:50.82 Toroid.coast @01:24:50.91 Carousel.coast @01:24:51.28 Gate.power :motor,:OFF @01:24:51.30 Gate.power :core,:OFF @01:24:51.31 Gate.power :analytic1,:OFF @01:24:51.32 Gate.power :analytic2,:OFF @01:24:51.34 Gate.power :analytic3,:OFF @01:24:51.35 Gate.power :main,:OFF @01:24:51.36 Safely stopped and ready to power off @01:24:51.59 delayUntil "12:00:00.00" @01:24:51.62 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 12:00" @01:25:01.05 upsync @12:00:00.12
Gate.power :main,:ON,limit to 5000mA @12:00:00.14 Gate.power :analytic1,:ON,limit to 5000mA @12:00:00.15 Gate.power :analytic2,:ON,limit to 5000mA @12:00:00.16 Gate.power :analytic3,:ON,limit to 5000mA @12:00:00.20 Powering core bus @12:00:02.83 Archive.configure :default @12:00:03.24 Air.configure :default @12:00:03.53 Lysis1.configure :default @12:00:03.80 Lysis2.configure :default @12:00:04.13 Plunge.configure :default @12:00:04.60 SamplePump.configure :default @12:00:04.90 ExtraPump.configure :default @12:00:05.21 Toroid.configure :default @12:00:06.75 Carousel.configure :default @12:00:06.90 Pressure.configure :default @12:00:06.95 All elf controllers running firmware version 4.61 @12:00:06.97 Gate.power :motor,:ON,limit to 5000mA @12:00:07.23 Archive.reconfigure :default @12:00:07.50 Archive.home.jog 0 @12:00:07.73 Toroid.reconfigure :default @12:00:07.78 Lysis2.reconfigure :default @12:00:07.81 Lysis1.reconfigure :default @12:00:07.87 Air.reconfigure :default @12:00:07.91 Plunge.reconfigure :default @12:00:08.10 Toroid.home.jog 0 @12:00:08.21 Plunge.home.jog 0 @12:00:08.26 Lysis2.home.jog 0 @12:00:08.29 Lysis1.home.jog 0 @12:00:08.31 Air.home.jog 0 @12:00:08.66 Archive.reconfigure :homing @12:00:08.98 Plunge.reconfigure :homing @12:00:09.03 Lysis2.reconfigure :homing @12:00:09.06 Archive.jog 1375 @12:00:09.10 Toroid.reconfigure :homing @12:00:09.29 Lysis2.jog 1353 @12:00:09.33 Toroid.jog 630 @12:00:09.37 Plunge.jog 847 @12:00:09.56 Air.reconfigure :homing @12:00:09.59 Lysis1.reconfigure :homing @12:00:09.69 Lysis1.jog 1349 @12:00:09.73 Air.jog 1358 @12:00:10.35 Plunge.setRawPosition! 50 @12:00:10.37 Archive.setRawPosition! 50 @12:00:10.45 Plunge.seek :home @12:00:10.52 Archive.seek :home @12:00:10.68 Lysis2.setRawPosition! 50 @12:00:10.73 Lysis2.seek :home @12:00:10.82 Plunge.reconfigure :default @12:00:11.00 Lysis1.setRawPosition! 50 @12:00:11.12 Air.setRawPosition! 50 @12:00:11.30 Archive.reconfigure :default @12:00:11.42 Lysis1.seek :home @12:00:11.76 Air.seek :home @12:00:12.50 Lysis2.reconfigure :default @12:00:12.62 Lysis1.reconfigure :default @12:00:12.71 Toroid.setRawPosition! 0 @12:00:12.74 Air.reconfigure :default @12:00:13.28 Toroid.seek :home @12:00:13.65 Toroid.reconfigure :default @12:00:14.05
Toroid.seek :clear @12:00:15.71 Carousel.reconfigure :default @12:00:15.83 Carousel.home.jog 0 @12:00:16.52 Carousel.seek 28 @12:00:16.88 Selecting Cartridge 27 @12:00:17.26 Carousel.seek 27 @12:00:17.77 Lysis1.seek :locked @12:00:21.01 Plunge.seek :locked @12:00:21.23 Air.seek :locked @12:00:25.65
Mounted archive Cartridge @12:00:26.11 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 27" @12:00:26.68 Duration of filtering limited to 4:05:00 @12:00:26.97 Priming sample loop w/160ml, bypass w/40ml @12:00:26.99 SamplePump.setPosition! 0ml @12:00:27.02 Bypass.open @12:00:28.51 Intake.open @12:00:29.39 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @12:00:32.07 Vacuum tare is 17% of full scale @12:00:32.15 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @12:00:32.69 Pressure.reconfigure :default @12:00:32.79 Pressure.hold @12:00:33.32 SamplePump.reconfigure :prime @12:00:33.54 SamplePump.seek 40ml,3:51 @12:00:45.91 upsync @12:01:16.79 Pressure.coast @12:01:16.84 SamplePump.setPosition! 0ml @12:01:16.89 Exhaust.open @12:01:18.42 Bypass.close @12:01:21.51 Pressure.reconfigure :default @12:01:21.66 Pressure.hold @12:01:22.39 SamplePump.reconfigure :prime @12:01:22.45 SamplePump.seek 160ml,13:51 @12:04:14.20 Pressure.coast @12:04:14.36 Bypass.open @12:04:15.04 Intake.close @12:04:15.95 Toroid.reconfigure :hold @12:04:16.11 Toroid.seek :crack @12:04:17.81 Toroid.seek :cartridge @12:04:18.75 Toroid.reconfigure :default @12:04:18.81 Toroid.seek :clear @12:04:20.65 Pressurizing cartridge to 26psi above 14.6psia @12:04:20.67 SamplePump.setPosition! 0ml @12:04:20.70 Intake.open @12:04:21.50 Bypass.close @12:04:22.34 Exhaust.close @12:04:23.36 Pressure.reconfigure :default @12:04:23.51 Pressure.hold @12:04:23.93 SamplePump.reconfigure :charge @12:04:24.15 SamplePump.seek 2.3ml,42 seconds @12:04:38.03 Pressure.coast @12:04:38.09 Intake.close @12:04:38.95 Pressurized cartridge with 0.884ml to 26.3psi @12:04:38.96 Isolated @35.1psia (20.5psi) @12:04:38.97 Checking for cartridge leaks from 35.1psia @12:04:38.99 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @12:04:43.57 Cartridge pressure fell 1% to 34.6psia @12:04:43.78 Sampling 2000.0ml @12:04:43.82 SamplePump.setPosition! 0ml @12:04:43.88 Exhaust.open @12:04:44.60 Intake.open @12:04:45.74 Pressure.reconfigure :default @12:04:45.79 Pressure.hold @12:04:46.06 SamplePump.reconfigure :sample @12:04:46.12 SamplePump.seek 2000ml,2:47:12 @12:04:48.80 Cartridge 27 sampling 1/2000ml at 0.53ml/s with 24.4psi vacuum @12:05:49.06 Cartridge 27 sampling 32/2000ml at 0.51ml/s with 26.2psi vacuum @12:06:48.93 Cartridge 27 sampling 63/2000ml at 0.52ml/s with 25.0psi vacuum @12:07:48.81 Cartridge 27 sampling 94/2000ml at 0.52ml/s with 20.5psi vacuum @12:08:49.19 Cartridge 27 sampling 125/2000ml at 0.52ml/s with 25.1psi vacuum @12:09:48.80 Cartridge 27 sampling 157/2000ml at 0.53ml/s with 26.1psi vacuum @12:10:48.94 Cartridge 27 sampling 188/2000ml at 0.53ml/s with 21.8psi vacuum @12:11:48.81 Cartridge 27 sampling 220/2000ml at 0.53ml/s with 20.6psi vacuum @12:12:48.96 Cartridge 27 sampling 252/2000ml at 0.54ml/s with 22.1psi vacuum @12:13:48.80 Cartridge 27 sampling 284/2000ml at 0.54ml/s with 26.9psi vacuum @12:14:48.90 Cartridge 27 sampling 316/2000ml at 0.55ml/s with 24.8psi vacuum @12:15:48.81 Cartridge 27 sampling 348/2000ml at 0.55ml/s with 20.3psi vacuum @12:16:49.10 Cartridge 27 sampling 381/2000ml at 0.52ml/s with 17.4psi vacuum 2.4m deep (18.2psia) @12:17:48.81 Cartridge 27 sampling 413/2000ml at 0.54ml/s with 20.5psi vacuum @12:18:48.86 Cartridge 27 sampling 445/2000ml at 0.54ml/s with 26.0psi vacuum @12:19:48.80 Cartridge 27 sampling 477/2000ml at 0.54ml/s with 19.9psi vacuum @12:20:49.01 Cartridge 27 sampling 509/2000ml at 0.53ml/s with 19.6psi vacuum @12:21:49.07 Cartridge 27 sampling 542/2000ml at 0.55ml/s with 27.0psi vacuum @12:22:48.81 Cartridge 27 sampling 574/2000ml at 0.51ml/s with 29.2psi vacuum @12:23:49.05 Cartridge 27 sampling 606/2000ml at 0.55ml/s with 21.0psi vacuum @12:24:48.81 Cartridge 27 sampling 638/2000ml at 0.53ml/s with 20.0psi vacuum @12:25:48.97 Cartridge 27 sampling 669/2000ml at 0.51ml/s with 29.1psi vacuum @12:26:48.81 Cartridge 27 sampling 701/2000ml at 0.53ml/s with 20.1psi vacuum 1.8m deep (17.3psia) @12:27:48.92 Cartridge 27 sampling 732/2000ml at 0.52ml/s with 27.3psi vacuum @12:28:48.81 Cartridge 27 sampling 764/2000ml at 0.54ml/s with 23.2psi vacuum @12:29:48.97 Cartridge 27 sampling 795/2000ml at 0.51ml/s with 29.7psi vacuum @12:30:48.81 Cartridge 27 sampling 827/2000ml at 0.53ml/s with 27.6psi vacuum @12:31:48.90 Cartridge 27 sampling 858/2000ml at 0.52ml/s with 27.2psi vacuum @12:32:48.80 Cartridge 27 sampling 889/2000ml at 0.52ml/s with 26.7psi vacuum @12:33:49.00 Cartridge 27 sampling 920/2000ml at 0.51ml/s with 28.3psi vacuum @12:34:48.80 Cartridge 27 sampling 952/2000ml at 0.53ml/s with 28.1psi vacuum @12:35:48.80 Cartridge 27 sampling 983/2000ml at 0.51ml/s with 21.1psi vacuum @12:36:49.26 Cartridge 27 sampling 1013/2000ml at 0.50ml/s with 29.1psi vacuum @12:37:48.80 Cartridge 27 sampling 1044/2000ml at 0.53ml/s with 23.1psi vacuum @12:38:49.05 Cartridge 27 sampling 1075/2000ml at 0.47ml/s with 23.1psi vacuum @12:39:48.80 Cartridge 27 sampling 1106/2000ml at 0.50ml/s with 29.7psi vacuum @12:40:49.23 Cartridge 27 sampling 1136/2000ml at 0.49ml/s with 18.6psi vacuum @12:41:48.80 Cartridge 27 sampling 1167/2000ml at 0.48ml/s with 17.9psi vacuum @12:42:48.95 Cartridge 27 sampling 1197/2000ml at 0.50ml/s with 25.5psi vacuum @12:43:48.80 Cartridge 27 sampling 1228/2000ml at 0.53ml/s with 23.7psi vacuum @12:44:48.96 Cartridge 27 sampling 1258/2000ml at 0.51ml/s with 20.4psi vacuum @12:45:48.80 Cartridge 27 sampling 1288/2000ml at 0.49ml/s with 23.7psi vacuum @12:46:48.88 Cartridge 27 sampling 1318/2000ml at 0.51ml/s with 29.2psi vacuum @12:47:48.80 Cartridge 27 sampling 1347/2000ml at 0.51ml/s with 24.8psi vacuum @12:48:49.00 Cartridge 27 sampling 1377/2000ml at 0.49ml/s with 21.2psi vacuum @12:49:49.02 Cartridge 27 sampling 1407/2000ml at 0.49ml/s with 27.9psi vacuum @12:50:48.80 Cartridge 27 sampling 1436/2000ml at 0.50ml/s with 25.7psi vacuum @12:51:50.03 Cartridge 27 sampling 1465/2000ml at 0.50ml/s with 24.5psi vacuum @12:52:48.79 Cartridge 27 sampling 1494/2000ml at 0.49ml/s with 28.3psi vacuum @12:53:49.00 Cartridge 27 sampling 1524/2000ml at 0.50ml/s with 19.9psi vacuum 3.9m deep (20.4psia) @12:54:48.80 Cartridge 27 sampling 1553/2000ml at 0.47ml/s with 15.5psi vacuum 2.6m deep (18.6psia) @12:55:48.97 Cartridge 27 sampling 1581/2000ml at 0.50ml/s with 25.6psi vacuum @12:56:48.80 Cartridge 27 sampling 1610/2000ml at 0.48ml/s with 29.2psi vacuum @12:57:49.01 Cartridge 27 sampling 1639/2000ml at 0.48ml/s with 15.1psi vacuum 5.2m deep (22.3psia) @12:58:48.80 Cartridge 27 sampling 1667/2000ml at 0.49ml/s with 25.6psi vacuum @12:59:48.93 Cartridge 27 sampling 1696/2000ml at 0.49ml/s with 23.0psi vacuum @13:00:48.79 Cartridge 27 sampling 1724/2000ml at 0.48ml/s with 26.0psi vacuum @13:01:48.89 Cartridge 27 sampling 1753/2000ml at 0.49ml/s with 15.7psi vacuum 6.6m deep (24.5psia) @13:02:48.80 Cartridge 27 sampling 1781/2000ml at 0.47ml/s with 25.9psi vacuum @13:03:48.95 Cartridge 27 sampling 1809/2000ml at 0.47ml/s with 23.8psi vacuum @13:04:49.00 Cartridge 27 sampling 1837/2000ml at 0.48ml/s with 24.3psi vacuum @13:05:48.81 Cartridge 27 sampling 1864/2000ml at 0.45ml/s with 29.5psi vacuum @13:06:48.99 Cartridge 27 sampling 1892/2000ml at 0.47ml/s with 27.9psi vacuum @13:07:48.81 Cartridge 27 sampling 1919/2000ml at 0.48ml/s with 25.6psi vacuum @13:08:48.92 Cartridge 27 sampling 1947/2000ml at 0.45ml/s with 12.0psi vacuum 4.1m deep (20.7psia) @13:09:48.85 Cartridge 27 sampling 1974/2000ml at 0.45ml/s with 28.6psi vacuum @13:10:45.50 Pressure.coast @13:10:45.64 Intake.close @13:10:45.70 Exhaust.close @13:10:45.87 Toroid.reconfigure :hold @13:10:45.94 Toroid.seek :bypass @13:10:47.67 Toroid.reconfigure :default @13:10:47.84 Toroid.seek :clear @13:10:49.57 Sampled 2000.0ml @13:10:50.68 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 27" @13:10:51.82 SamplePump.setPosition! 0ml @13:10:51.88 Plunge.seek :reagent @13:10:51.91 Bypass.open @13:10:52.63 Intake.open @13:10:52.82 Sample loop closed. Ready to Deliver Reagents @13:10:52.84 Archive.seek :sealed @13:10:53.50 SamplePump.reconfigure :slow @13:10:53.56 SamplePump.seek -3ml @13:10:56.52 Archive.seek :evac @13:11:00.28 delay 1 second @13:11:01.40 SamplePump.seek -2.8ml @13:11:02.09 delay 1 second @13:11:03.18 SamplePump.seek -3.3ml @13:11:04.54 delay 1 second @13:11:05.57 Exhaust.open @13:11:06.33 Bypass.close @13:11:07.30 SamplePump.seek -35.3ml @13:11:10.53 delay 1:00 @13:11:12.94 upsync @13:12:10.59 Archive.seek :clear @13:12:16.18 delay 3 seconds @13:12:19.30 Intake.close @13:12:19.42 Exhaust.close @13:12:27.79 RNAlater soak for 10:00 @13:22:27.84 Archive.seek :tVent @13:22:29.61 delay 1 second @13:22:30.70 Archive.seek :evac @13:22:46.68 delay 10 seconds @13:22:56.79 Sample preserved and evacuated @13:22:57.42
Carousel.seek 27 @13:22:57.63 Plunge.seek :locked @13:22:58.62 Archive.seek :clear @13:22:58.78 Air.seek :clear @13:22:58.96 Lysis1.seek :clear @13:23:15.82
Archive.coast @13:23:15.86 Air.coast @13:23:15.91 Lysis1.coast @13:23:15.95 Lysis2.coast @13:23:15.99 Plunge.coast @13:23:16.04 SamplePump.coast @13:23:16.08 Toroid.coast @13:23:16.12 Carousel.coast @13:23:16.31 Gate.power :motor,:OFF @13:23:16.32 Gate.power :core,:OFF @13:23:16.33 Gate.power :analytic1,:OFF @13:23:16.35 Gate.power :analytic2,:OFF @13:23:16.36 Gate.power :analytic3,:OFF @13:23:16.37 Gate.power :main,:OFF @13:23:16.39 Safely stopped and ready to power off @13:23:16.61 delayUntil "00:00:00.00PDT24-Jun-21" @13:23:16.64 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 00:00PDT24-Jun-21" @13:23:31.96 upsync @23:09:00.35 Session Begun @23:11:16.12 -> Cartridge.dry @23:11:25.48 -> MainThread.exit @23:11:25.58 ThreadError in preston -- killed thread @23:11:25.64 Session Ended @23:11:53.16PDT23-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @23:11:53.25 >BREAK< ESP3 Gateway v1.11 for cartridge bus @23:11:57.05 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @23:11:57.36 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @23:12:06.62 Polling Can every 2:00 @23:12:06.67 Loading /home/esp/esp2/bin/processAll mission @23:12:07.92 Sampling default goal volumes @23:12:07.97 GoodNews.email "Cartridge group: dry\nSpares: {spare: 1}\nSchedule: [\"3am\", \"3pm\"]\n",Subject: "Begun processAll --schedule=3am,3pm --checkpoint --mail" @23:12:08.09
delayUntil "03:00:00.00PDT24-Jun-21" @23:12:17.75 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 03:00PDT24-Jun-21" @23:12:29.57 upsync @23:12:45.47 upsync @03:00:00.35PDT24-Jun-21
Gate.power :main,:ON,limit to 5000mA @03:00:00.40 Gate.power :analytic1,:ON,limit to 5000mA @03:00:00.45 Gate.power :analytic2,:ON,limit to 5000mA @03:00:00.49 Gate.power :analytic3,:ON,limit to 5000mA @03:00:00.55 Powering core bus @03:00:02.89 Archive.configure :default @03:00:03.04 Air.configure :default @03:00:03.15 Lysis1.configure :default @03:00:03.26 Lysis2.configure :default @03:00:03.39 Plunge.configure :default @03:00:03.53 SamplePump.configure :default @03:00:03.67 ExtraPump.configure :default @03:00:03.81 Toroid.configure :default @03:00:03.94 Carousel.configure :default @03:00:04.08 Pressure.configure :default @03:00:04.14 All elf controllers running firmware version 4.61 @03:00:04.17 Gate.power :motor,:ON,limit to 5000mA @03:00:04.43 Archive.reconfigure :default @03:00:04.70 Archive.home.jog 0 @03:00:04.92 Toroid.reconfigure :default @03:00:04.98 Plunge.reconfigure :default @03:00:05.02 Lysis2.reconfigure :default @03:00:05.06 Air.reconfigure :default @03:00:05.11 Lysis1.reconfigure :default @03:00:05.31 Plunge.home.jog 0 @03:00:05.36 Lysis2.home.jog 0 @03:00:05.39 Toroid.home.jog 0 @03:00:05.46 Lysis1.home.jog 0 @03:00:05.50 Air.home.jog 0 @03:00:06.01 Archive.reconfigure :homing @03:00:06.21 Archive.jog 1366 @03:00:06.27 Plunge.reconfigure :homing @03:00:06.29 Lysis2.reconfigure :homing @03:00:06.32 Air.reconfigure :homing @03:00:06.36 Toroid.reconfigure :homing @03:00:06.50 Lysis1.reconfigure :homing @03:00:07.40 Lysis2.jog 1352 @03:00:07.48 Plunge.jog 847 @03:00:07.62 Lysis1.jog 1351 @03:00:07.76 Air.jog 1358 @03:00:07.93 Toroid.jog 609 @03:00:08.44 Archive.setRawPosition! 50 @03:00:08.54 Archive.seek :home @03:00:08.88 Archive.reconfigure :default @03:00:09.10 Plunge.setRawPosition! 50 @03:00:09.16 Lysis2.setRawPosition! 50 @03:00:09.38 Plunge.seek :home @03:00:09.60 Lysis2.seek :home @03:00:09.94 Plunge.reconfigure :default @03:00:10.01 Lysis2.reconfigure :default @03:00:10.11 Lysis1.setRawPosition! 50 @03:00:10.17 Air.setRawPosition! 50 @03:00:10.65 Lysis1.seek :home @03:00:10.73 Air.seek :home @03:00:11.05 Toroid.setRawPosition! 0 @03:00:11.11 Toroid.seek :home @03:00:11.41 Lysis1.reconfigure :default @03:00:11.44 Air.reconfigure :default @03:00:11.47 Toroid.reconfigure :default @03:00:12.37
Toroid.seek :clear @03:00:14.07 Carousel.reconfigure :default @03:00:14.12 Carousel.home.jog 0 @03:00:14.63 Carousel.seek 27 @03:00:14.81 Selecting Cartridge 26 @03:00:15.02 Carousel.seek 26 @03:00:15.56 Lysis1.seek :locked @03:00:18.92 Plunge.seek :locked @03:00:19.28 Air.seek :locked @03:00:23.56
Mounted archive Cartridge @03:00:23.68 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 26" @03:00:23.98 Duration of filtering limited to 4:05:00 @03:00:24.81 Priming sample loop w/160ml, bypass w/40ml @03:00:24.83 SamplePump.setPosition! 0ml @03:00:24.87 Bypass.open @03:00:26.60 Intake.open @03:00:27.75 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @03:00:30.45 Vacuum tare is 18% of full scale @03:00:30.47 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @03:00:30.84 Pressure.reconfigure :default @03:00:30.93 Pressure.hold @03:00:31.38 SamplePump.reconfigure :prime @03:00:31.68 SamplePump.seek 40ml,3:51 @03:00:36.62 upsync @03:01:14.96 Pressure.coast @03:01:15.12 SamplePump.setPosition! 0ml @03:01:15.18 Exhaust.open @03:01:16.67 Bypass.close @03:01:18.38 Pressure.reconfigure :default @03:01:18.45 Pressure.hold @03:01:18.73 SamplePump.reconfigure :prime @03:01:18.79 SamplePump.seek 160ml,13:51 @03:04:10.42 Pressure.coast @03:04:10.54 Bypass.open @03:04:11.27 Intake.close @03:04:12.53 Toroid.reconfigure :hold @03:04:12.59 Toroid.seek :crack @03:04:14.17 Toroid.seek :cartridge @03:04:15.10 Toroid.reconfigure :default @03:04:15.17 Toroid.seek :clear @03:04:17.44 Pressurizing cartridge to 26psi above 14.6psia @03:04:17.47 SamplePump.setPosition! 0ml @03:04:17.57 Intake.open @03:04:18.44 Bypass.close @03:04:19.25 Exhaust.close @03:04:20.36 Pressure.reconfigure :default @03:04:20.57 Pressure.hold @03:04:21.01 SamplePump.reconfigure :charge @03:04:21.16 SamplePump.seek 2.3ml,42 seconds @03:04:33.13 Pressure.coast @03:04:33.18 Intake.close @03:04:34.05 Pressurized cartridge with 0.656ml to 26.2psi @03:04:34.06 Isolated @35.8psia (21.2psi) @03:04:34.07 Checking for cartridge leaks from 35.8psia @03:04:34.09 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @03:04:39.30 Cartridge pressure fell 5% to 34.0psia @03:04:39.50 Sampling 2000.0ml @03:04:39.55 SamplePump.setPosition! 0ml @03:04:39.61 Exhaust.open @03:04:40.38 Intake.open @03:04:41.34 Pressure.reconfigure :default @03:04:41.40 Pressure.hold @03:04:41.68 SamplePump.reconfigure :sample @03:04:41.74 SamplePump.seek 2000ml,2:47:12 @03:04:44.39 Cartridge 26 sampling 1/2000ml at 0.53ml/s with 25.9psi vacuum @03:05:44.50 Cartridge 26 sampling 32/2000ml at 0.52ml/s with 25.4psi vacuum @03:06:44.40 Cartridge 26 sampling 63/2000ml at 0.52ml/s with 24.7psi vacuum @03:07:44.61 Cartridge 26 sampling 94/2000ml at 0.51ml/s with 24.2psi vacuum @03:08:44.57 Cartridge 26 sampling 125/2000ml at 0.51ml/s with 25.0psi vacuum @03:09:44.41 Cartridge 26 sampling 155/2000ml at 0.53ml/s with 26.5psi vacuum @03:10:44.63 Cartridge 26 sampling 187/2000ml at 0.54ml/s with 21.7psi vacuum @03:11:44.41 Cartridge 26 sampling 219/2000ml at 0.55ml/s with 23.9psi vacuum @03:12:44.51 Cartridge 26 sampling 251/2000ml at 0.53ml/s with 26.8psi vacuum @03:13:44.41 Cartridge 26 sampling 283/2000ml at 0.53ml/s with 22.3psi vacuum @03:14:44.62 Cartridge 26 sampling 314/2000ml at 0.51ml/s with 28.0psi vacuum @03:15:44.40 Cartridge 26 sampling 346/2000ml at 0.53ml/s with 25.2psi vacuum @03:16:44.60 Cartridge 26 sampling 378/2000ml at 0.52ml/s with 23.8psi vacuum @03:17:44.41 Cartridge 26 sampling 409/2000ml at 0.54ml/s with 27.7psi vacuum @03:18:44.57 Cartridge 26 sampling 441/2000ml at 0.54ml/s with 20.9psi vacuum @03:19:44.41 Cartridge 26 sampling 473/2000ml at 0.50ml/s with 22.0psi vacuum @03:20:44.57 Cartridge 26 sampling 505/2000ml at 0.52ml/s with 17.8psi vacuum @03:21:44.41 Cartridge 26 sampling 536/2000ml at 0.54ml/s with 20.7psi vacuum @03:22:44.59 Cartridge 26 sampling 568/2000ml at 0.51ml/s with 28.4psi vacuum @03:23:44.61 Cartridge 26 sampling 599/2000ml at 0.51ml/s with 17.7psi vacuum @03:24:44.41 Cartridge 26 sampling 631/2000ml at 0.50ml/s with 27.4psi vacuum @03:25:44.60 Cartridge 26 sampling 662/2000ml at 0.50ml/s with 28.1psi vacuum @03:26:44.40 Cartridge 26 sampling 693/2000ml at 0.52ml/s with 23.2psi vacuum @03:27:44.60 Cartridge 26 sampling 724/2000ml at 0.52ml/s with 24.3psi vacuum @03:28:44.40 Cartridge 26 sampling 756/2000ml at 0.54ml/s with 21.8psi vacuum @03:29:44.59 Cartridge 26 sampling 787/2000ml at 0.50ml/s with 20.5psi vacuum @03:30:44.41 Cartridge 26 sampling 818/2000ml at 0.53ml/s with 24.1psi vacuum @03:31:44.53 Cartridge 26 sampling 848/2000ml at 0.52ml/s with 19.5psi vacuum @03:32:44.41 Cartridge 26 sampling 879/2000ml at 0.51ml/s with 27.1psi vacuum @03:33:44.51 Cartridge 26 sampling 910/2000ml at 0.48ml/s with 25.2psi vacuum @03:34:44.41 Cartridge 26 sampling 940/2000ml at 0.50ml/s with 17.9psi vacuum 2.5m deep (18.5psia) @03:35:44.56 Cartridge 26 sampling 971/2000ml at 0.50ml/s with 29.1psi vacuum @03:36:44.49 Cartridge 26 sampling 1001/2000ml at 0.51ml/s with 25.6psi vacuum @03:37:44.55 Cartridge 26 sampling 1031/2000ml at 0.52ml/s with 24.3psi vacuum @03:38:44.71 Cartridge 26 sampling 1062/2000ml at 0.52ml/s with 23.7psi vacuum @03:39:44.39 Cartridge 26 sampling 1092/2000ml at 0.50ml/s with 27.4psi vacuum @03:40:44.55 Cartridge 26 sampling 1122/2000ml at 0.51ml/s with 20.6psi vacuum @03:41:44.40 Cartridge 26 sampling 1152/2000ml at 0.50ml/s with 24.5psi vacuum @03:42:45.90 Cartridge 26 sampling 1181/2000ml at 0.51ml/s with 26.8psi vacuum @03:43:44.39 Cartridge 26 sampling 1211/2000ml at 0.51ml/s with 21.7psi vacuum @03:44:44.59 Cartridge 26 sampling 1241/2000ml at 0.50ml/s with 27.3psi vacuum @03:45:44.40 Cartridge 26 sampling 1271/2000ml at 0.48ml/s with 29.1psi vacuum @03:46:44.58 Cartridge 26 sampling 1300/2000ml at 0.50ml/s with 18.6psi vacuum @03:47:44.40 Cartridge 26 sampling 1330/2000ml at 0.48ml/s with 16.7psi vacuum @03:48:44.66 Cartridge 26 sampling 1359/2000ml at 0.49ml/s with 24.5psi vacuum @03:49:44.40 Cartridge 26 sampling 1388/2000ml at 0.48ml/s with 25.4psi vacuum @03:50:44.51 Cartridge 26 sampling 1417/2000ml at 0.45ml/s with 29.5psi vacuum @03:51:44.40 Cartridge 26 sampling 1446/2000ml at 0.48ml/s with 28.1psi vacuum @03:52:44.53 Cartridge 26 sampling 1475/2000ml at 0.47ml/s with 28.9psi vacuum @03:53:44.40 Cartridge 26 sampling 1504/2000ml at 0.47ml/s with 28.3psi vacuum @03:54:44.63 Cartridge 26 sampling 1532/2000ml at 0.46ml/s with 24.8psi vacuum @03:55:44.62 Cartridge 26 sampling 1561/2000ml at 0.48ml/s with 28.0psi vacuum @03:56:44.41 Cartridge 26 sampling 1589/2000ml at 0.48ml/s with 18.7psi vacuum 2.3m deep (18.1psia) @03:57:44.61 Cartridge 26 sampling 1617/2000ml at 0.47ml/s with 26.2psi vacuum @03:58:44.41 Cartridge 26 sampling 1645/2000ml at 0.45ml/s with 16.1psi vacuum 1.6m deep (17.1psia) @03:59:44.84 Cartridge 26 sampling 1673/2000ml at 0.47ml/s with 26.2psi vacuum @04:00:44.40 Cartridge 26 sampling 1701/2000ml at 0.46ml/s with 29.5psi vacuum @04:01:44.55 Cartridge 26 sampling 1729/2000ml at 0.46ml/s with 29.1psi vacuum @04:02:44.41 Cartridge 26 sampling 1756/2000ml at 0.47ml/s with 20.6psi vacuum @04:03:44.51 Cartridge 26 sampling 1784/2000ml at 0.45ml/s with 15.8psi vacuum 6.4m deep (24.2psia) @04:04:44.41 Cartridge 26 sampling 1812/2000ml at 0.46ml/s with 20.4psi vacuum @04:05:44.47 Cartridge 26 sampling 1839/2000ml at 0.47ml/s with 25.6psi vacuum @04:06:44.41 Cartridge 26 sampling 1866/2000ml at 0.46ml/s with 25.6psi vacuum @04:07:44.65 Cartridge 26 sampling 1893/2000ml at 0.44ml/s with 21.4psi vacuum @04:08:44.66 Cartridge 26 sampling 1920/2000ml at 0.43ml/s with 15.2psi vacuum @04:09:44.41 Cartridge 26 sampling 1947/2000ml at 0.43ml/s with 29.7psi vacuum @04:10:44.79 Cartridge 26 sampling 1974/2000ml at 0.46ml/s with 24.6psi vacuum @04:11:44.05 Pressure.coast @04:11:44.18 Intake.close @04:11:44.26 Exhaust.close @04:11:44.43 Toroid.reconfigure :hold @04:11:44.49 Toroid.seek :bypass @04:11:46.11 Toroid.reconfigure :default @04:11:46.17 Toroid.seek :clear @04:11:47.81 Sampled 2000.0ml @04:11:49.41 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 26" @04:11:49.56 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:11:49.66 Retry #1 of 10 after 10:00 @04:11:50.41 SamplePump.setPosition! 0ml @04:11:50.54 Plunge.seek :reagent @04:11:50.68 Bypass.open @04:11:51.46 Intake.open @04:11:51.71 Sample loop closed. Ready to Deliver Reagents @04:11:51.73 Archive.seek :sealed @04:11:52.38 SamplePump.reconfigure :slow @04:11:52.56 SamplePump.seek -3ml @04:11:55.42 Archive.seek :evac @04:11:59.25 delay 1 second @04:12:00.40 SamplePump.seek -2.8ml @04:12:01.25 delay 1 second @04:12:02.57 SamplePump.seek -3.3ml @04:12:03.97 delay 1 second @04:12:05.00 Exhaust.open @04:12:05.68 Bypass.close @04:12:06.45 SamplePump.seek -35.3ml @04:12:09.42 delay 1:00 @04:13:09.47 Archive.seek :clear @04:13:15.18 delay 3 seconds @04:13:18.27 Intake.close @04:13:18.31 Exhaust.close @04:13:26.60 RNAlater soak for 10:00 @04:21:49.70 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:21:49.72 Retry #2 of 10 after 10:00 @04:23:26.65 Archive.seek :tVent @04:23:28.34 delay 1 second @04:23:29.43 Archive.seek :evac @04:23:45.42 delay 10 seconds @04:23:55.50 Sample preserved and evacuated @04:23:56.00
Carousel.seek 26 @04:23:56.45 Plunge.seek :locked @04:23:57.89 Archive.seek :clear @04:23:58.13 Air.seek :clear @04:23:58.29 Lysis1.seek :clear @04:24:15.24
Archive.coast @04:24:15.28 Air.coast @04:24:15.32 Lysis1.coast @04:24:15.40 Lysis2.coast @04:24:15.57 Plunge.coast @04:24:15.61 SamplePump.coast @04:24:15.67 Toroid.coast @04:24:15.75 Carousel.coast @04:24:16.23 Gate.power :motor,:OFF @04:24:16.31 Gate.power :core,:OFF @04:24:16.33 Gate.power :analytic1,:OFF @04:24:16.41 Gate.power :analytic2,:OFF @04:24:16.43 Gate.power :analytic3,:OFF @04:24:16.45 Gate.power :main,:OFF @04:24:16.46 Safely stopped and ready to power off @04:24:16.69 delayUntil "15:00:00.00" @04:31:49.77 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:31:49.79 Retry #3 of 10 after 10:00 @04:41:49.86 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:41:49.99 Retry #4 of 10 after 10:00 @04:51:50.05 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:51:50.07 Retry #5 of 10 after 10:00 @05:01:50.20 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:01:50.30 Retry #6 of 10 after 10:00 @05:11:50.35 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:11:50.37 Retry #7 of 10 after 10:00 @05:21:50.80 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:21:50.84 Retry #8 of 10 after 10:00 @05:31:50.92 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:31:50.98 Retry #9 of 10 after 10:00 @05:41:51.03 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:41:51.05 Retry #10 of 10 after 10:00 @05:51:51.13 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:51:51.19 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 15:00" @05:51:51.27 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:51:51.33 Retry #1 of 10 after 10:00 @06:01:51.38 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:01:51.40 Retry #2 of 10 after 10:00 @06:11:51.47 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:11:51.52 Retry #3 of 10 after 10:00 @06:21:51.57 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:21:51.60 Retry #4 of 10 after 10:00 @06:31:51.65 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:31:51.67 Retry #5 of 10 after 10:00 @06:41:51.71 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:41:51.73 Retry #6 of 10 after 10:00 @06:51:51.78 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:51:51.80 Retry #7 of 10 after 10:00 @07:01:53.25 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @07:01:53.27 Retry #8 of 10 after 10:00 @07:11:53.32 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @07:11:53.34 Retry #9 of 10 after 10:00 @07:21:53.43 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @07:21:53.47 Retry #10 of 10 after 10:00 @07:31:53.56 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:00:00.11
Gate.power :main,:ON,limit to 5000mA @15:00:00.12 Gate.power :analytic1,:ON,limit to 5000mA @15:00:00.14 Gate.power :analytic2,:ON,limit to 5000mA @15:00:00.15 Gate.power :analytic3,:ON,limit to 5000mA @15:00:00.19 Powering core bus @15:00:02.47 Archive.configure :default @15:00:02.61 Air.configure :default @15:00:02.72 Lysis1.configure :default @15:00:02.83 Lysis2.configure :default @15:00:02.96 Plunge.configure :default @15:00:03.11 SamplePump.configure :default @15:00:03.24 ExtraPump.configure :default @15:00:03.38 Toroid.configure :default @15:00:03.52 Carousel.configure :default @15:00:03.66 Pressure.configure :default @15:00:03.72 All elf controllers running firmware version 4.61 @15:00:03.74 Gate.power :motor,:ON,limit to 5000mA @15:00:03.87 Archive.reconfigure :default @15:00:04.22 Archive.home.jog 0 @15:00:04.54 Plunge.reconfigure :default @15:00:04.81 Lysis2.reconfigure :default @15:00:04.94 Lysis1.reconfigure :default @15:00:05.30 Toroid.reconfigure :default @15:00:05.58 Air.reconfigure :default @15:00:06.04 Plunge.home.jog 0 @15:00:06.22 Lysis1.home.jog 0 @15:00:06.28 Toroid.home.jog 0 @15:00:06.36 Lysis2.home.jog 0 @15:00:06.45 Air.home.jog 0 @15:00:07.49 Archive.reconfigure :homing @15:00:07.69 Archive.jog 1364 @15:00:08.08 Plunge.reconfigure :homing @15:00:08.19 Lysis2.reconfigure :homing @15:00:08.32 Air.reconfigure :homing @15:00:08.36 Toroid.reconfigure :homing @15:00:08.49 Lysis1.reconfigure :homing @15:00:08.75 Plunge.jog 843 @15:00:09.09 Lysis2.jog 1344 @15:00:09.14 Toroid.jog 605 @15:00:09.20 Air.jog 1363 @15:00:09.45 Lysis1.jog 1350 @15:00:09.78 Archive.setRawPosition! 50 @15:00:09.82 Archive.seek :home @15:00:10.42 Archive.reconfigure :default @15:00:10.59 Lysis2.setRawPosition! 50 @15:00:10.62 Plunge.setRawPosition! 50 @15:00:10.71 Lysis2.seek :home @15:00:10.74 Plunge.seek :home @15:00:10.89 Air.setRawPosition! 50 @15:00:10.93 Lysis1.setRawPosition! 50 @15:00:11.19 Air.seek :home @15:00:11.23 Lysis1.seek :home @15:00:11.44 Lysis2.reconfigure :default @15:00:11.48 Plunge.reconfigure :default @15:00:11.66 Toroid.setRawPosition! 0 @15:00:11.69 Lysis1.reconfigure :default @15:00:11.72 Air.reconfigure :default @15:00:11.84 Toroid.seek :home @15:00:11.99 Toroid.reconfigure :default @15:00:12.22
Toroid.seek :clear @15:00:13.84 Carousel.reconfigure :default @15:00:13.89 Carousel.home.jog 0 @15:00:14.38 Carousel.seek 26 @15:00:14.56 Selecting Cartridge 25 @15:00:14.77 Carousel.seek 25 @15:00:15.52 Lysis1.seek :locked @15:00:19.00 Plunge.seek :locked @15:00:19.30 Air.seek :locked @15:00:23.53
Mounted archive Cartridge @15:00:23.63 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 25" @15:00:23.71 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:00:23.73 Retry #1 of 10 after 10:00 @15:00:23.98 Duration of filtering limited to 4:05:00 @15:00:24.28 Priming sample loop w/160ml, bypass w/40ml @15:00:24.30 SamplePump.setPosition! 0ml @15:00:24.33 Bypass.open @15:00:25.92 Intake.open @15:00:27.11 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:00:29.76 Vacuum tare is 16% of full scale @15:00:29.78 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @15:00:29.99 Pressure.reconfigure :default @15:00:30.05 Pressure.hold @15:00:30.83 SamplePump.reconfigure :prime @15:00:30.89 SamplePump.seek 40ml,3:51 @15:01:14.19 Pressure.coast @15:01:14.31 SamplePump.setPosition! 0ml @15:01:14.42 Exhaust.open @15:01:15.89 Bypass.close @15:01:17.48 Pressure.reconfigure :default @15:01:17.54 Pressure.hold @15:01:17.81 SamplePump.reconfigure :prime @15:01:17.87 SamplePump.seek 160ml,13:51 @15:04:09.69 Pressure.coast @15:04:09.75 Bypass.open @15:04:10.48 Intake.close @15:04:11.36 Toroid.reconfigure :hold @15:04:11.42 Toroid.seek :crack @15:04:13.01 Toroid.seek :cartridge @15:04:13.95 Toroid.reconfigure :default @15:04:14.00 Toroid.seek :clear @15:04:16.12 Pressurizing cartridge to 26psi above 14.1psia @15:04:16.17 SamplePump.setPosition! 0ml @15:04:16.24 Intake.open @15:04:17.13 Bypass.close @15:04:17.87 Exhaust.close @15:04:19.18 Pressure.reconfigure :default @15:04:19.24 Pressure.hold @15:04:19.44 SamplePump.reconfigure :charge @15:04:19.50 SamplePump.seek 2.3ml,42 seconds @15:04:31.78 Pressure.coast @15:04:31.83 Intake.close @15:04:33.02 Pressurized cartridge with 0.689ml to 26.2psi @15:04:33.11 Isolated @33.4psia (19.2psi) @15:04:33.15 Checking for cartridge leaks from 33.4psia @15:04:33.20 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @15:04:37.96 Cartridge pressure fell 1% to 33.2psia @15:04:38.18 Sampling 2000.0ml @15:04:38.21 SamplePump.setPosition! 0ml @15:04:38.32 Exhaust.open @15:04:39.09 Intake.open @15:04:40.06 Pressure.reconfigure :default @15:04:40.12 Pressure.hold @15:04:40.40 SamplePump.reconfigure :sample @15:04:40.46 SamplePump.seek 2000ml,2:47:12 @15:04:43.32 Cartridge 25 sampling 1/2000ml at 0.60ml/s with 21.9psi vacuum @15:05:43.13 Cartridge 25 sampling 37/2000ml at 0.60ml/s with 20.6psi vacuum @15:06:43.29 Cartridge 25 sampling 73/2000ml at 0.61ml/s with 22.0psi vacuum @15:07:43.13 Cartridge 25 sampling 109/2000ml at 0.61ml/s with 25.4psi vacuum @15:08:43.29 Cartridge 25 sampling 145/2000ml at 0.60ml/s with 22.7psi vacuum @15:09:43.13 Cartridge 25 sampling 181/2000ml at 0.60ml/s with 20.3psi vacuum @15:10:23.84 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:10:23.89 Retry #2 of 10 after 10:00 @15:10:43.29 Cartridge 25 sampling 217/2000ml at 0.61ml/s with 25.4psi vacuum @15:11:43.12 Cartridge 25 sampling 253/2000ml at 0.60ml/s with 24.1psi vacuum @15:12:43.29 Cartridge 25 sampling 290/2000ml at 0.61ml/s with 22.0psi vacuum @15:13:43.13 Cartridge 25 sampling 326/2000ml at 0.62ml/s with 21.5psi vacuum @15:14:43.25 Cartridge 25 sampling 363/2000ml at 0.61ml/s with 26.0psi vacuum @15:15:43.13 Cartridge 25 sampling 399/2000ml at 0.62ml/s with 24.2psi vacuum @15:16:43.13 Cartridge 25 sampling 436/2000ml at 0.61ml/s with 19.4psi vacuum @15:17:43.31 Cartridge 25 sampling 472/2000ml at 0.61ml/s with 24.3psi vacuum @15:18:43.13 Cartridge 25 sampling 508/2000ml at 0.59ml/s with 22.5psi vacuum @15:19:43.30 Cartridge 25 sampling 544/2000ml at 0.60ml/s with 26.9psi vacuum @15:20:23.94 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:20:23.96 Retry #3 of 10 after 10:00 @15:20:43.13 Cartridge 25 sampling 580/2000ml at 0.60ml/s with 27.0psi vacuum @15:21:43.34 Cartridge 25 sampling 617/2000ml at 0.60ml/s with 23.4psi vacuum @15:22:43.11 Cartridge 25 sampling 653/2000ml at 0.60ml/s with 25.1psi vacuum @15:23:43.35 Cartridge 25 sampling 689/2000ml at 0.59ml/s with 23.1psi vacuum @15:24:43.12 Cartridge 25 sampling 724/2000ml at 0.60ml/s with 25.4psi vacuum @15:25:43.31 Cartridge 25 sampling 760/2000ml at 0.59ml/s with 26.8psi vacuum @15:26:43.12 Cartridge 25 sampling 795/2000ml at 0.58ml/s with 23.6psi vacuum @15:27:43.29 Cartridge 25 sampling 831/2000ml at 0.58ml/s with 21.7psi vacuum @15:28:43.12 Cartridge 25 sampling 866/2000ml at 0.58ml/s with 23.0psi vacuum @15:29:43.23 Cartridge 25 sampling 901/2000ml at 0.59ml/s with 26.6psi vacuum @15:30:24.01 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:30:24.03 Retry #4 of 10 after 10:00 @15:30:43.33 Cartridge 25 sampling 935/2000ml at 0.58ml/s with 27.5psi vacuum @15:31:43.13 Cartridge 25 sampling 970/2000ml at 0.57ml/s with 27.3psi vacuum @15:32:43.24 Cartridge 25 sampling 1005/2000ml at 0.56ml/s with 28.0psi vacuum @15:33:43.13 Cartridge 25 sampling 1039/2000ml at 0.58ml/s with 25.0psi vacuum @15:34:43.39 Cartridge 25 sampling 1073/2000ml at 0.56ml/s with 27.9psi vacuum @15:35:43.13 Cartridge 25 sampling 1107/2000ml at 0.58ml/s with 23.6psi vacuum @15:36:43.23 Cartridge 25 sampling 1141/2000ml at 0.57ml/s with 18.8psi vacuum @15:37:43.13 Cartridge 25 sampling 1175/2000ml at 0.54ml/s with 28.3psi vacuum @15:38:43.22 Cartridge 25 sampling 1209/2000ml at 0.55ml/s with 25.3psi vacuum @15:39:43.12 Cartridge 25 sampling 1243/2000ml at 0.57ml/s with 25.4psi vacuum @15:40:24.10 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:40:24.14 Retry #5 of 10 after 10:00 @15:40:43.19 Cartridge 25 sampling 1276/2000ml at 0.56ml/s with 24.4psi vacuum @15:41:43.13 Cartridge 25 sampling 1309/2000ml at 0.55ml/s with 26.8psi vacuum @15:42:43.41 Cartridge 25 sampling 1342/2000ml at 0.53ml/s with 28.6psi vacuum @15:43:43.64 Cartridge 25 sampling 1376/2000ml at 0.55ml/s with 26.0psi vacuum @15:44:43.23 Cartridge 25 sampling 1409/2000ml at 0.54ml/s with 20.9psi vacuum @15:45:43.46 Cartridge 25 sampling 1441/2000ml at 0.53ml/s with 28.8psi vacuum @15:46:43.12 Cartridge 25 sampling 1474/2000ml at 0.54ml/s with 22.8psi vacuum @15:47:43.28 Cartridge 25 sampling 1507/2000ml at 0.55ml/s with 20.8psi vacuum @15:48:43.12 Cartridge 25 sampling 1539/2000ml at 0.53ml/s with 18.1psi vacuum @15:49:43.33 Cartridge 25 sampling 1571/2000ml at 0.54ml/s with 19.8psi vacuum 2.1m deep (17.7psia) @15:50:43.17 Cartridge 25 sampling 1603/2000ml at 0.54ml/s with 22.7psi vacuum @15:50:51.92 upsync @15:51:43.32 Cartridge 25 sampling 1635/2000ml at 0.54ml/s with 17.7psi vacuum 1.9m deep (17.4psia) @15:52:43.13 Cartridge 25 sampling 1667/2000ml at 0.52ml/s with 28.8psi vacuum @15:53:43.27 Cartridge 25 sampling 1699/2000ml at 0.53ml/s with 17.8psi vacuum 2.7m deep (18.6psia) @15:54:43.12 Cartridge 25 sampling 1730/2000ml at 0.53ml/s with 25.0psi vacuum @15:55:43.41 Cartridge 25 sampling 1762/2000ml at 0.52ml/s with 26.9psi vacuum @15:56:43.13 Cartridge 25 sampling 1793/2000ml at 0.54ml/s with 24.9psi vacuum @15:57:43.29 Cartridge 25 sampling 1825/2000ml at 0.52ml/s with 20.6psi vacuum @15:58:43.12 Cartridge 25 sampling 1856/2000ml at 0.51ml/s with 27.8psi vacuum @15:59:43.20 Cartridge 25 sampling 1887/2000ml at 0.52ml/s with 21.5psi vacuum @16:00:43.13 Cartridge 25 sampling 1918/2000ml at 0.53ml/s with 21.9psi vacuum @16:01:43.42 Cartridge 25 sampling 1949/2000ml at 0.52ml/s with 25.4psi vacuum @16:02:43.38 Cartridge 25 sampling 1980/2000ml at 0.49ml/s with 26.6psi vacuum @16:03:22.17 Pressure.coast @16:03:22.49 Intake.close @16:03:22.64 Exhaust.close @16:03:23.05 Toroid.reconfigure :hold @16:03:23.27 Toroid.seek :bypass @16:03:24.89 Toroid.reconfigure :default @16:03:24.95 Toroid.seek :clear @16:03:26.57 Sampled 2000.0ml @16:03:27.06 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 25" @16:03:27.44 SamplePump.setPosition! 0ml @16:03:27.48 Plunge.seek :reagent @16:03:27.55 Bypass.open @16:03:28.27 Intake.open @16:03:28.42 Sample loop closed. Ready to Deliver Reagents @16:03:28.44 Archive.seek :sealed @16:03:29.17 SamplePump.reconfigure :slow @16:03:29.34 SamplePump.seek -3ml @16:03:32.11 Archive.seek :evac @16:03:36.07 delay 1 second @16:03:37.17 SamplePump.seek -2.8ml @16:03:37.87 delay 1 second @16:03:39.25 SamplePump.seek -3.3ml @16:03:40.71 delay 1 second @16:03:41.75 Exhaust.open @16:03:42.51 Bypass.close @16:03:43.53 SamplePump.seek -35.3ml @16:03:46.18 delay 1:00 @16:03:50.62 upsync @16:04:46.24 Archive.seek :clear @16:04:52.41 delay 3 seconds @16:04:55.61 Intake.close @16:04:55.81 Exhaust.close @16:05:03.47 RNAlater soak for 10:00 @16:15:03.52 Archive.seek :tVent @16:15:05.17 delay 1 second @16:15:06.20 Archive.seek :evac @16:15:22.17 delay 10 seconds @16:15:32.21 Sample preserved and evacuated @16:15:32.44
Carousel.seek 25 @16:15:32.65 Plunge.seek :locked @16:15:33.63 Archive.seek :clear @16:15:33.78 Air.seek :clear @16:15:34.02 Lysis1.seek :clear @16:15:50.82
Archive.coast @16:15:50.86 Air.coast @16:15:50.90 Lysis1.coast @16:15:50.95 Lysis2.coast @16:15:50.99 Plunge.coast @16:15:51.03 SamplePump.coast @16:15:51.08 Toroid.coast @16:15:51.12 Carousel.coast @16:15:51.32 Gate.power :motor,:OFF @16:15:51.33 Gate.power :core,:OFF @16:15:51.35 Gate.power :analytic1,:OFF @16:15:51.36 Gate.power :analytic2,:OFF @16:15:51.37 Gate.power :analytic3,:OFF @16:15:51.39 Gate.power :main,:OFF @16:15:51.40 Safely stopped and ready to power off @16:15:51.62 delayUntil "03:00:00.00PDT25-Jun-21" @16:15:51.65 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 03:00PDT25-Jun-21" @16:16:07.81 upsync @03:00:00.11PDT25-Jun-21
Gate.power :main,:ON,limit to 5000mA @03:00:00.12 Gate.power :analytic1,:ON,limit to 5000mA @03:00:00.14 Gate.power :analytic2,:ON,limit to 5000mA @03:00:00.15 Gate.power :analytic3,:ON,limit to 5000mA @03:00:00.19 Powering core bus @03:00:02.69 Archive.configure :default @03:00:03.09 Air.configure :default @03:00:03.30 Lysis1.configure :default @03:00:03.53 Lysis2.configure :default @03:00:03.79 Plunge.configure :default @03:00:04.11 SamplePump.configure :default @03:00:04.41 ExtraPump.configure :default @03:00:04.66 Toroid.configure :default @03:00:05.02 Carousel.configure :default @03:00:05.32 Pressure.configure :default @03:00:05.51 All elf controllers running firmware version 4.61 @03:00:05.55 Gate.power :motor,:ON,limit to 5000mA @03:00:05.76 Archive.reconfigure :default @03:00:06.93 Archive.home.jog 0 @03:00:07.11 Plunge.reconfigure :default @03:00:07.19 Lysis1.reconfigure :default @03:00:07.25 Air.reconfigure :default @03:00:07.28 Lysis2.reconfigure :default @03:00:07.34 Toroid.reconfigure :default @03:00:07.50 Plunge.home.jog 0 @03:00:07.61 Lysis1.home.jog 0 @03:00:07.65 Toroid.home.jog 0 @03:00:07.68 Lysis2.home.jog 0 @03:00:07.70 Air.home.jog 0 @03:00:08.14 Archive.reconfigure :homing @03:00:08.38 Archive.jog 1356 @03:00:08.44 Plunge.reconfigure :homing @03:00:08.47 Lysis2.reconfigure :homing @03:00:08.51 Air.reconfigure :homing @03:00:08.56 Lysis1.reconfigure :homing @03:00:08.60 Toroid.reconfigure :homing @03:00:08.83 Plunge.jog 836 @03:00:08.88 Lysis2.jog 1341 @03:00:08.91 Air.jog 1351 @03:00:08.94 Lysis1.jog 1352 @03:00:08.97 Toroid.jog 630 @03:00:09.84 Archive.setRawPosition! 50 @03:00:09.89 Archive.seek :home @03:00:09.99 Plunge.setRawPosition! 50 @03:00:10.03 Plunge.seek :home @03:00:10.13 Lysis2.setRawPosition! 50 @03:00:10.19 Lysis2.seek :home @03:00:10.34 Lysis1.setRawPosition! 50 @03:00:10.37 Air.setRawPosition! 50 @03:00:10.39 Archive.reconfigure :default @03:00:10.41 Plunge.reconfigure :default @03:00:10.58 Lysis1.seek :home @03:00:10.62 Air.seek :home @03:00:10.80 Lysis2.reconfigure :default @03:00:10.96 Lysis1.reconfigure :default @03:00:11.08 Air.reconfigure :default @03:00:11.35 Toroid.setRawPosition! 0 @03:00:11.45 Toroid.seek :home @03:00:11.81 Toroid.reconfigure :default @03:00:12.30
Toroid.seek :clear @03:00:14.01 Carousel.reconfigure :default @03:00:14.09 Carousel.home.jog 0 @03:00:15.82 Carousel.seek 25 @03:00:16.12 Selecting Cartridge 24 @03:00:16.88 Carousel.seek 24 @03:00:17.39 Lysis1.seek :locked @03:00:20.63 Plunge.seek :locked @03:00:20.77 Air.seek :locked @03:00:25.17
Mounted archive Cartridge @03:00:25.27 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 24" @03:00:26.25 Duration of filtering limited to 4:05:00 @03:00:26.61 Priming sample loop w/160ml, bypass w/40ml @03:00:26.63 SamplePump.setPosition! 0ml @03:00:26.67 Bypass.open @03:00:28.14 Intake.open @03:00:28.98 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @03:00:31.92 Vacuum tare is 18% of full scale @03:00:31.94 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @03:00:32.47 Pressure.reconfigure :default @03:00:32.61 Pressure.hold @03:00:33.15 SamplePump.reconfigure :prime @03:00:33.30 SamplePump.seek 40ml,3:51 @03:00:46.08 upsync @03:01:16.59 Pressure.coast @03:01:16.65 SamplePump.setPosition! 0ml @03:01:16.70 Exhaust.open @03:01:18.20 Bypass.close @03:01:20.10 Pressure.reconfigure :default @03:01:20.25 Pressure.hold @03:01:20.94 SamplePump.reconfigure :prime @03:01:21.18 SamplePump.seek 160ml,13:51 @03:04:12.96 Pressure.coast @03:04:13.14 Bypass.open @03:04:13.83 Intake.close @03:04:14.91 Toroid.reconfigure :hold @03:04:15.04 Toroid.seek :crack @03:04:16.76 Toroid.seek :cartridge @03:04:17.69 Toroid.reconfigure :default @03:04:17.75 Toroid.seek :clear @03:04:19.59 Pressurizing cartridge to 26psi above 14.5psia @03:04:19.61 SamplePump.setPosition! 0ml @03:04:19.64 Intake.open @03:04:20.46 Bypass.close @03:04:21.17 Exhaust.close @03:04:22.32 Pressure.reconfigure :default @03:04:22.50 Pressure.hold @03:04:22.98 SamplePump.reconfigure :charge @03:04:23.10 SamplePump.seek 2.3ml,42 seconds @03:04:35.44 Pressure.coast @03:04:35.54 Intake.close @03:04:36.43 Pressurized cartridge with 0.667ml to 26.1psi @03:04:36.44 Isolated @33.8psia (19.3psi) @03:04:36.45 Checking for cartridge leaks from 33.8psia @03:04:36.47 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @03:04:41.16 Cartridge pressure fell 3% to 32.8psia @03:04:41.33 Sampling 2000.0ml @03:04:41.45 SamplePump.setPosition! 0ml @03:04:41.50 Exhaust.open @03:04:42.36 Intake.open @03:04:43.54 Pressure.reconfigure :default @03:04:43.64 Pressure.hold @03:04:44.23 SamplePump.reconfigure :sample @03:04:44.30 SamplePump.seek 2000ml,2:47:12 @03:04:46.55 Cartridge 24 sampling 1/2000ml at 0.57ml/s with 26.3psi vacuum @03:05:46.70 Cartridge 24 sampling 33/2000ml at 0.53ml/s with 21.9psi vacuum @03:06:46.48 Cartridge 24 sampling 64/2000ml at 0.52ml/s with 25.3psi vacuum @03:07:46.84 Cartridge 24 sampling 96/2000ml at 0.53ml/s with 25.7psi vacuum @03:08:46.49 Cartridge 24 sampling 127/2000ml at 0.52ml/s with 25.4psi vacuum @03:09:46.80 Cartridge 24 sampling 159/2000ml at 0.55ml/s with 26.5psi vacuum @03:10:46.49 Cartridge 24 sampling 191/2000ml at 0.53ml/s with 23.4psi vacuum @03:11:46.69 Cartridge 24 sampling 224/2000ml at 0.54ml/s with 25.5psi vacuum @03:12:46.49 Cartridge 24 sampling 256/2000ml at 0.52ml/s with 27.4psi vacuum @03:13:46.56 Cartridge 24 sampling 288/2000ml at 0.54ml/s with 21.2psi vacuum @03:14:46.49 Cartridge 24 sampling 321/2000ml at 0.55ml/s with 22.4psi vacuum @03:15:46.51 Cartridge 24 sampling 353/2000ml at 0.52ml/s with 17.4psi vacuum @03:16:46.85 Cartridge 24 sampling 385/2000ml at 0.53ml/s with 18.2psi vacuum 3.9m deep (20.5psia) @03:17:46.49 Cartridge 24 sampling 417/2000ml at 0.54ml/s with 18.7psi vacuum @03:18:46.79 Cartridge 24 sampling 450/2000ml at 0.54ml/s with 26.4psi vacuum @03:19:46.49 Cartridge 24 sampling 482/2000ml at 0.52ml/s with 26.6psi vacuum @03:20:46.65 Cartridge 24 sampling 513/2000ml at 0.54ml/s with 24.6psi vacuum @03:21:46.49 Cartridge 24 sampling 545/2000ml at 0.54ml/s with 26.2psi vacuum @03:22:46.63 Cartridge 24 sampling 577/2000ml at 0.52ml/s with 22.1psi vacuum @03:23:46.49 Cartridge 24 sampling 608/2000ml at 0.51ml/s with 28.2psi vacuum @03:24:46.65 Cartridge 24 sampling 639/2000ml at 0.52ml/s with 23.2psi vacuum @03:25:46.49 Cartridge 24 sampling 671/2000ml at 0.54ml/s with 21.2psi vacuum @03:26:46.53 Cartridge 24 sampling 702/2000ml at 0.50ml/s with 27.3psi vacuum @03:27:46.49 Cartridge 24 sampling 733/2000ml at 0.51ml/s with 27.9psi vacuum @03:28:46.61 Cartridge 24 sampling 764/2000ml at 0.53ml/s with 25.7psi vacuum @03:29:46.59 Cartridge 24 sampling 794/2000ml at 0.48ml/s with 28.8psi vacuum @03:30:46.53 Cartridge 24 sampling 825/2000ml at 0.52ml/s with 26.8psi vacuum @03:31:46.69 Cartridge 24 sampling 855/2000ml at 0.52ml/s with 19.9psi vacuum @03:32:46.49 Cartridge 24 sampling 885/2000ml at 0.49ml/s with 27.9psi vacuum @03:33:46.66 Cartridge 24 sampling 915/2000ml at 0.51ml/s with 24.3psi vacuum @03:34:46.48 Cartridge 24 sampling 944/2000ml at 0.47ml/s with 15.5psi vacuum @03:35:46.62 Cartridge 24 sampling 974/2000ml at 0.50ml/s with 20.0psi vacuum 1.6m deep (17.1psia) @03:36:46.49 Cartridge 24 sampling 1003/2000ml at 0.48ml/s with 29.2psi vacuum @03:37:46.60 Cartridge 24 sampling 1033/2000ml at 0.50ml/s with 27.2psi vacuum @03:38:46.48 Cartridge 24 sampling 1062/2000ml at 0.51ml/s with 25.1psi vacuum @03:39:46.63 Cartridge 24 sampling 1091/2000ml at 0.49ml/s with 18.7psi vacuum 3.9m deep (20.4psia) @03:40:46.49 Cartridge 24 sampling 1120/2000ml at 0.47ml/s with 13.4psi vacuum 3.2m deep (19.4psia) @03:41:46.56 Cartridge 24 sampling 1149/2000ml at 0.48ml/s with 13.0psi vacuum 8.3m deep (27.0psia) @03:42:46.48 Cartridge 24 sampling 1178/2000ml at 0.49ml/s with 26.5psi vacuum @03:43:46.57 Cartridge 24 sampling 1207/2000ml at 0.48ml/s with 28.7psi vacuum @03:44:46.50 Cartridge 24 sampling 1235/2000ml at 0.45ml/s with 16.1psi vacuum @03:45:46.52 Cartridge 24 sampling 1264/2000ml at 0.47ml/s with 15.1psi vacuum 4.2m deep (20.9psia) @03:46:46.75 Cartridge 24 sampling 1292/2000ml at 0.49ml/s with 22.8psi vacuum @03:47:46.49 Cartridge 24 sampling 1320/2000ml at 0.48ml/s with 23.7psi vacuum @03:48:46.57 Cartridge 24 sampling 1348/2000ml at 0.46ml/s with 12.9psi vacuum 5.9m deep (23.4psia) @03:49:46.49 Cartridge 24 sampling 1376/2000ml at 0.48ml/s with 21.0psi vacuum 2.0m deep (17.7psia) @03:50:46.70 Cartridge 24 sampling 1403/2000ml at 0.45ml/s with 19.8psi vacuum 3.6m deep (20.1psia) @03:51:46.48 Cartridge 24 sampling 1431/2000ml at 0.47ml/s with 26.6psi vacuum @03:52:46.66 Cartridge 24 sampling 1458/2000ml at 0.42ml/s with 20.0psi vacuum 1.9m deep (17.5psia) @03:53:46.49 Cartridge 24 sampling 1485/2000ml at 0.47ml/s with 22.4psi vacuum @03:54:46.69 Cartridge 24 sampling 1513/2000ml at 0.47ml/s with 25.4psi vacuum @03:55:46.48 Cartridge 24 sampling 1540/2000ml at 0.44ml/s with 9.9psi vacuum 5.5m deep (22.8psia) @03:56:46.89 Cartridge 24 sampling 1567/2000ml at 0.47ml/s with 25.2psi vacuum @03:57:46.66 Cartridge 24 sampling 1593/2000ml at 0.44ml/s with 10.5psi vacuum 9.9m deep (29.3psia) @03:58:46.51 Cartridge 24 sampling 1620/2000ml at 0.46ml/s with 25.7psi vacuum @03:59:46.86 Cartridge 24 sampling 1647/2000ml at 0.44ml/s with 27.9psi vacuum @04:00:46.49 Cartridge 24 sampling 1673/2000ml at 0.41ml/s with 29.1psi vacuum @04:01:46.63 Cartridge 24 sampling 1699/2000ml at 0.41ml/s with 16.8psi vacuum 4.6m deep (21.5psia) @04:02:46.49 Cartridge 24 sampling 1725/2000ml at 0.40ml/s with 26.7psi vacuum @04:03:46.60 Cartridge 24 sampling 1751/2000ml at 0.41ml/s with 10.9psi vacuum 5.3m deep (22.5psia) @04:04:46.49 Cartridge 24 sampling 1776/2000ml at 0.41ml/s with 30.0psi vacuum @04:05:46.73 Cartridge 24 sampling 1802/2000ml at 0.43ml/s with 16.0psi vacuum 2.4m deep (18.3psia) @04:06:46.49 Cartridge 24 sampling 1827/2000ml at 0.42ml/s with 28.9psi vacuum @04:07:46.65 Cartridge 24 sampling 1852/2000ml at 0.44ml/s with 24.4psi vacuum @04:08:46.49 Cartridge 24 sampling 1877/2000ml at 0.43ml/s with 24.0psi vacuum @04:09:46.61 Cartridge 24 sampling 1902/2000ml at 0.42ml/s with 22.7psi vacuum 2.0m deep (17.7psia) @04:10:46.48 Cartridge 24 sampling 1927/2000ml at 0.40ml/s with 14.0psi vacuum @04:11:46.73 Cartridge 24 sampling 1952/2000ml at 0.42ml/s with 14.8psi vacuum @04:12:46.81 Cartridge 24 sampling 1976/2000ml at 0.41ml/s with 14.4psi vacuum 7.2m deep (25.4psia) @04:13:45.34 Pressure.coast @04:13:45.57 Intake.close @04:13:45.72 Exhaust.close @04:13:46.20 Toroid.reconfigure :hold @04:13:46.36 Toroid.seek :bypass @04:13:47.97 Toroid.reconfigure :default @04:13:48.03 Toroid.seek :clear @04:13:49.68 Sampled 2000.0ml @04:13:50.25 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 24" @04:13:50.91 SamplePump.setPosition! 0ml @04:13:51.31 Plunge.seek :reagent @04:13:51.68 Bypass.open @04:13:52.45 Sample loop closed. Ready to Deliver Reagents @04:13:53.78 Archive.seek :sealed @04:13:54.01 Intake.open @04:13:54.96 SamplePump.reconfigure :slow @04:13:55.08 SamplePump.seek -3ml @04:13:57.57 Archive.seek :evac @04:14:02.15 delay 1 second @04:14:03.42 SamplePump.seek -2.8ml @04:14:04.12 delay 1 second @04:14:05.41 SamplePump.seek -3.3ml @04:14:06.88 delay 1 second @04:14:07.90 Exhaust.open @04:14:08.62 Bypass.close @04:14:09.38 SamplePump.seek -35.3ml @04:14:11.58 delay 1:00 @04:14:11.78 upsync @04:15:11.63 Archive.seek :clear @04:15:18.14 delay 3 seconds @04:15:21.27 Intake.close @04:15:21.31 Exhaust.close @04:15:28.76 RNAlater soak for 10:00 @04:25:28.80 Archive.seek :tVent @04:25:30.45 delay 1 second @04:25:31.49 Archive.seek :evac @04:25:47.49 delay 10 seconds @04:25:57.53 Sample preserved and evacuated @04:25:58.30
Carousel.seek 24 @04:25:58.76 Plunge.seek :locked @04:25:59.92 Archive.seek :clear @04:26:00.32 Air.seek :clear @04:26:00.66 Lysis1.seek :clear @04:26:17.26
Archive.coast @04:26:17.34 Air.coast @04:26:17.39 Lysis1.coast @04:26:17.51 Lysis2.coast @04:26:17.63 Plunge.coast @04:26:17.68 SamplePump.coast @04:26:17.84 Toroid.coast @04:26:17.93 Carousel.coast @04:26:18.29 Gate.power :motor,:OFF @04:26:18.36 Gate.power :core,:OFF @04:26:18.44 Gate.power :analytic1,:OFF @04:26:18.45 Gate.power :analytic2,:OFF @04:26:18.46 Gate.power :analytic3,:OFF @04:26:18.48 Gate.power :main,:OFF @04:26:18.50 Safely stopped and ready to power off @04:26:19.01 delayUntil "15:00:00.00" @04:26:19.04 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 15:00" @04:26:37.44 upsync @15:00:00.12
Gate.power :main,:ON,limit to 5000mA @15:00:00.13 Gate.power :analytic1,:ON,limit to 5000mA @15:00:00.15 Gate.power :analytic2,:ON,limit to 5000mA @15:00:00.16 Gate.power :analytic3,:ON,limit to 5000mA @15:00:00.20 Powering core bus @15:00:03.05 Archive.configure :default @15:00:03.30 Air.configure :default @15:00:03.63 Lysis1.configure :default @15:00:03.92 Lysis2.configure :default @15:00:04.26 Plunge.configure :default @15:00:04.50 SamplePump.configure :default @15:00:04.83 ExtraPump.configure :default @15:00:05.26 Toroid.configure :default @15:00:05.57 Carousel.configure :default @15:00:05.81 Pressure.configure :default @15:00:05.92 All elf controllers running firmware version 4.61 @15:00:05.94 Gate.power :motor,:ON,limit to 5000mA @15:00:06.65 Archive.reconfigure :default @15:00:07.03 Archive.home.jog 0 @15:00:07.19 Lysis1.reconfigure :default @15:00:07.28 Plunge.reconfigure :default @15:00:07.34 Lysis2.reconfigure :default @15:00:07.37 Air.reconfigure :default @15:00:07.42 Toroid.reconfigure :default @15:00:07.60 Lysis1.home.jog 0 @15:00:07.75 Lysis2.home.jog 0 @15:00:07.82 Toroid.home.jog 0 @15:00:07.86 Plunge.home.jog 0 @15:00:07.91 Air.home.jog 0 @15:00:08.45 Archive.reconfigure :homing @15:00:08.67 Archive.jog 1367 @15:00:08.72 Toroid.reconfigure :homing @15:00:08.75 Plunge.reconfigure :homing @15:00:08.79 Lysis1.reconfigure :homing @15:00:08.83 Air.reconfigure :homing @15:00:08.87 Lysis2.reconfigure :homing @15:00:09.11 Plunge.jog 845 @15:00:09.14 Lysis1.jog 1355 @15:00:09.17 Toroid.jog 626 @15:00:09.20 Lysis2.jog 1342 @15:00:09.23 Air.jog 1359 @15:00:10.04 Archive.setRawPosition! 50 @15:00:10.09 Archive.seek :home @15:00:10.18 Plunge.setRawPosition! 50 @15:00:10.24 Plunge.seek :home @15:00:10.36 Archive.reconfigure :default @15:00:10.48 Lysis1.setRawPosition! 50 @15:00:10.55 Lysis1.seek :home @15:00:10.65 Lysis2.setRawPosition! 50 @15:00:10.67 Air.setRawPosition! 50 @15:00:10.69 Plunge.reconfigure :default @15:00:10.80 Lysis2.seek :home @15:00:10.86 Air.seek :home @15:00:11.06 Lysis1.reconfigure :default @15:00:11.65 Lysis2.reconfigure :default @15:00:11.85 Air.reconfigure :default @15:00:11.95 Toroid.setRawPosition! 0 @15:00:12.54 Toroid.seek :home @15:00:12.81 Toroid.reconfigure :default @15:00:13.18
Toroid.seek :clear @15:00:14.87 Carousel.reconfigure :default @15:00:14.99 Carousel.home.jog 0 @15:00:15.55 Carousel.seek 24 @15:00:15.85 Selecting Cartridge 23 @15:00:16.45 Carousel.seek 23 @15:00:17.12 Lysis1.seek :locked @15:00:20.35 Plunge.seek :locked @15:00:20.49 Air.seek :locked @15:00:24.84
Mounted archive Cartridge @15:00:25.11 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 23" @15:00:25.19 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:00:25.25 Retry #1 of 10 after 10:00 @15:00:26.00 Duration of filtering limited to 4:05:00 @15:00:26.28 Priming sample loop w/160ml, bypass w/40ml @15:00:26.30 SamplePump.setPosition! 0ml @15:00:26.34 Bypass.open @15:00:27.80 Intake.open @15:00:28.69 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:00:31.42 Vacuum tare is 17% of full scale @15:00:31.47 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -16psi @15:00:31.91 Pressure.reconfigure :default @15:00:32.10 Pressure.hold @15:00:32.66 SamplePump.reconfigure :prime @15:00:32.74 SamplePump.seek 40ml,3:51 @15:01:16.10 Pressure.coast @15:01:16.15 SamplePump.setPosition! 0ml @15:01:16.19 Exhaust.open @15:01:17.65 Bypass.close @15:01:20.72 Pressure.reconfigure :default @15:01:20.78 Pressure.hold @15:01:21.39 SamplePump.reconfigure :prime @15:01:21.50 SamplePump.seek 160ml,13:51 @15:04:13.36 Pressure.coast @15:04:13.49 Bypass.open @15:04:14.27 Intake.close @15:04:15.39 Toroid.reconfigure :hold @15:04:15.45 Toroid.seek :crack @15:04:17.16 Toroid.seek :cartridge @15:04:18.09 Toroid.reconfigure :default @15:04:18.15 Toroid.seek :clear @15:04:19.99 Pressurizing cartridge to 26psi above 14.2psia @15:04:20.01 SamplePump.setPosition! 0ml @15:04:20.04 Intake.open @15:04:20.88 Bypass.close @15:04:21.69 Exhaust.close @15:04:22.88 Pressure.reconfigure :default @15:04:23.02 Pressure.hold @15:04:23.42 SamplePump.reconfigure :charge @15:04:23.61 SamplePump.seek 2.3ml,42 seconds @15:04:36.08 Pressure.coast @15:04:36.13 Intake.close @15:04:37.04 Pressurized cartridge with 0.707ml to 26.0psi @15:04:37.05 Isolated @33.5psia (19.4psi) @15:04:37.06 Checking for cartridge leaks from 33.5psia @15:04:37.08 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @15:04:41.80 Cartridge pressure fell 1% to 33.2psia @15:04:42.07 Sampling 2000.0ml @15:04:42.09 SamplePump.setPosition! 0ml @15:04:42.15 Exhaust.open @15:04:42.96 Intake.open @15:04:44.20 Pressure.reconfigure :default @15:04:44.37 Pressure.hold @15:04:45.19 SamplePump.reconfigure :sample @15:04:45.25 SamplePump.seek 2000ml,2:47:12 @15:04:47.21 Cartridge 23 sampling 1/2000ml at 0.63ml/s with 22.9psi vacuum @15:05:47.29 Cartridge 23 sampling 34/2000ml at 0.56ml/s with 21.2psi vacuum @15:06:47.22 Cartridge 23 sampling 68/2000ml at 0.55ml/s with 24.9psi vacuum @15:07:47.35 Cartridge 23 sampling 102/2000ml at 0.58ml/s with 25.7psi vacuum @15:08:47.22 Cartridge 23 sampling 136/2000ml at 0.57ml/s with 25.1psi vacuum @15:09:47.24 Cartridge 23 sampling 171/2000ml at 0.57ml/s with 26.7psi vacuum @15:10:25.30 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:10:25.32 Retry #2 of 10 after 10:00 @15:10:47.20 Cartridge 23 sampling 205/2000ml at 0.58ml/s with 24.5psi vacuum @15:11:47.22 Cartridge 23 sampling 240/2000ml at 0.58ml/s with 21.8psi vacuum @15:12:47.35 Cartridge 23 sampling 275/2000ml at 0.57ml/s with 22.7psi vacuum @15:13:47.21 Cartridge 23 sampling 310/2000ml at 0.57ml/s with 25.1psi vacuum @15:14:47.35 Cartridge 23 sampling 344/2000ml at 0.57ml/s with 27.0psi vacuum @15:15:47.21 Cartridge 23 sampling 379/2000ml at 0.59ml/s with 25.8psi vacuum @15:16:47.37 Cartridge 23 sampling 414/2000ml at 0.57ml/s with 20.1psi vacuum @15:17:47.21 Cartridge 23 sampling 448/2000ml at 0.57ml/s with 24.6psi vacuum @15:18:47.40 Cartridge 23 sampling 483/2000ml at 0.58ml/s with 26.0psi vacuum @15:19:47.21 Cartridge 23 sampling 517/2000ml at 0.58ml/s with 25.3psi vacuum @15:20:25.37 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:20:25.39 Retry #3 of 10 after 10:00 @15:20:47.40 Cartridge 23 sampling 552/2000ml at 0.57ml/s with 24.9psi vacuum @15:21:47.21 Cartridge 23 sampling 586/2000ml at 0.57ml/s with 26.1psi vacuum @15:22:47.30 Cartridge 23 sampling 620/2000ml at 0.58ml/s with 24.5psi vacuum @15:23:47.21 Cartridge 23 sampling 654/2000ml at 0.56ml/s with 27.4psi vacuum @15:24:47.23 Cartridge 23 sampling 688/2000ml at 0.56ml/s with 23.6psi vacuum @15:25:47.52 Cartridge 23 sampling 722/2000ml at 0.57ml/s with 25.8psi vacuum @15:26:47.22 Cartridge 23 sampling 756/2000ml at 0.56ml/s with 18.4psi vacuum @15:27:47.47 Cartridge 23 sampling 790/2000ml at 0.57ml/s with 18.3psi vacuum 2.3m deep (18.0psia) @15:28:47.22 Cartridge 23 sampling 824/2000ml at 0.58ml/s with 22.9psi vacuum @15:29:47.41 Cartridge 23 sampling 858/2000ml at 0.54ml/s with 28.9psi vacuum @15:30:25.48 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:30:25.50 Retry #4 of 10 after 10:00 @15:30:47.20 Cartridge 23 sampling 891/2000ml at 0.56ml/s with 26.6psi vacuum @15:31:47.38 Cartridge 23 sampling 925/2000ml at 0.55ml/s with 17.9psi vacuum @15:32:47.22 Cartridge 23 sampling 959/2000ml at 0.56ml/s with 27.4psi vacuum @15:33:47.38 Cartridge 23 sampling 992/2000ml at 0.55ml/s with 25.1psi vacuum @15:34:47.22 Cartridge 23 sampling 1025/2000ml at 0.55ml/s with 21.9psi vacuum @15:35:47.30 Cartridge 23 sampling 1058/2000ml at 0.56ml/s with 23.2psi vacuum @15:36:47.22 Cartridge 23 sampling 1091/2000ml at 0.54ml/s with 29.3psi vacuum @15:37:47.50 Cartridge 23 sampling 1124/2000ml at 0.55ml/s with 26.1psi vacuum @15:38:47.44 Cartridge 23 sampling 1157/2000ml at 0.55ml/s with 24.7psi vacuum @15:39:47.39 Cartridge 23 sampling 1190/2000ml at 0.55ml/s with 27.8psi vacuum @15:40:25.56 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:40:25.58 Retry #5 of 10 after 10:00 @15:40:47.57 Cartridge 23 sampling 1222/2000ml at 0.54ml/s with 22.7psi vacuum @15:41:47.20 Cartridge 23 sampling 1254/2000ml at 0.54ml/s with 24.0psi vacuum @15:42:47.38 Cartridge 23 sampling 1286/2000ml at 0.53ml/s with 21.9psi vacuum @15:43:47.22 Cartridge 23 sampling 1318/2000ml at 0.53ml/s with 26.5psi vacuum @15:44:47.37 Cartridge 23 sampling 1350/2000ml at 0.51ml/s with 25.3psi vacuum @15:45:47.22 Cartridge 23 sampling 1381/2000ml at 0.51ml/s with 19.3psi vacuum @15:46:48.61 Cartridge 23 sampling 1414/2000ml at 0.54ml/s with 25.3psi vacuum @15:47:47.22 Cartridge 23 sampling 1445/2000ml at 0.53ml/s with 17.7psi vacuum @15:48:47.43 Cartridge 23 sampling 1477/2000ml at 0.53ml/s with 21.5psi vacuum @15:49:47.22 Cartridge 23 sampling 1508/2000ml at 0.52ml/s with 21.0psi vacuum @15:50:25.71 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:50:25.79 Retry #6 of 10 after 10:00 @15:50:47.37 Cartridge 23 sampling 1539/2000ml at 0.53ml/s with 23.2psi vacuum @15:51:47.22 Cartridge 23 sampling 1570/2000ml at 0.52ml/s with 28.7psi vacuum @15:52:47.46 Cartridge 23 sampling 1602/2000ml at 0.51ml/s with 20.5psi vacuum @15:53:47.50 Cartridge 23 sampling 1633/2000ml at 0.53ml/s with 23.4psi vacuum @15:54:47.22 Cartridge 23 sampling 1663/2000ml at 0.50ml/s with 27.4psi vacuum @15:55:47.57 Cartridge 23 sampling 1693/2000ml at 0.50ml/s with 18.3psi vacuum 2.6m deep (18.5psia) @15:56:47.22 Cartridge 23 sampling 1724/2000ml at 0.51ml/s with 24.9psi vacuum @15:57:47.39 Cartridge 23 sampling 1754/2000ml at 0.50ml/s with 19.6psi vacuum 1.9m deep (17.6psia) @15:58:47.22 Cartridge 23 sampling 1784/2000ml at 0.49ml/s with 18.0psi vacuum @15:59:47.36 Cartridge 23 sampling 1814/2000ml at 0.51ml/s with 22.9psi vacuum @16:00:25.86 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:00:25.94 Retry #7 of 10 after 10:00 @16:00:47.21 Cartridge 23 sampling 1844/2000ml at 0.51ml/s with 25.2psi vacuum @16:01:47.40 Cartridge 23 sampling 1874/2000ml at 0.49ml/s with 27.6psi vacuum @16:02:47.22 Cartridge 23 sampling 1903/2000ml at 0.50ml/s with 22.9psi vacuum @16:03:47.47 Cartridge 23 sampling 1933/2000ml at 0.49ml/s with 18.0psi vacuum @16:04:47.22 Cartridge 23 sampling 1963/2000ml at 0.49ml/s with 27.9psi vacuum @16:05:47.30 Cartridge 23 sampling 1993/2000ml at 0.49ml/s with 18.3psi vacuum 3.0m deep (19.1psia) @16:06:02.33 Pressure.coast @16:06:02.85 Intake.close @16:06:03.06 Exhaust.close @16:06:03.50 Toroid.reconfigure :hold @16:06:03.60 Toroid.seek :bypass @16:06:05.39 Toroid.reconfigure :default @16:06:05.47 Toroid.seek :clear @16:06:07.13 Sampled 2000.0ml @16:06:07.94 SamplePump.setPosition! 0ml @16:06:07.97 Plunge.seek :reagent @16:06:08.05 Bypass.open @16:06:08.80 Intake.open @16:06:08.92 Sample loop closed. Ready to Deliver Reagents @16:06:08.94 Archive.seek :sealed @16:06:09.67 SamplePump.reconfigure :slow @16:06:09.73 SamplePump.seek -3ml @16:06:12.74 Archive.seek :evac @16:06:16.42 delay 1 second @16:06:17.51 SamplePump.seek -2.8ml @16:06:18.20 delay 1 second @16:06:19.29 SamplePump.seek -3.3ml @16:06:20.62 delay 1 second @16:06:21.65 Exhaust.open @16:06:22.42 Bypass.close @16:06:23.50 SamplePump.seek -35.3ml @16:06:26.76 delay 1:00 @16:07:26.81 Archive.seek :clear @16:07:32.35 delay 3 seconds @16:07:35.51 Intake.close @16:07:35.65 Exhaust.close @16:07:44.10 RNAlater soak for 10:00 @16:10:26.01 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:10:26.03 Retry #8 of 10 after 10:00 @16:17:44.18 Archive.seek :tVent @16:17:45.98 delay 1 second @16:17:47.01 Archive.seek :evac @16:18:02.98 delay 10 seconds @16:18:13.02 Sample preserved and evacuated @16:18:13.39
Carousel.seek 23 @16:18:13.64 Plunge.seek :locked @16:18:14.62 Archive.seek :clear @16:18:14.79 Air.seek :clear @16:18:14.94 Lysis1.seek :clear @16:18:31.81
Archive.coast @16:18:31.85 Air.coast @16:18:31.89 Lysis1.coast @16:18:31.94 Lysis2.coast @16:18:31.98 Plunge.coast @16:18:32.02 SamplePump.coast @16:18:32.07 Toroid.coast @16:18:32.11 Carousel.coast @16:18:32.30 Gate.power :motor,:OFF @16:18:32.31 Gate.power :core,:OFF @16:18:32.32 Gate.power :analytic1,:OFF @16:18:32.34 Gate.power :analytic2,:OFF @16:18:32.35 Gate.power :analytic3,:OFF @16:18:32.36 Gate.power :main,:OFF @16:18:32.38 Safely stopped and ready to power off @16:18:32.60 delayUntil "03:00:00.00PDT26-Jun-21" @16:20:26.11 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:20:26.14 Retry #9 of 10 after 10:00 @16:30:26.18 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:30:26.20 Retry #10 of 10 after 10:00 @16:40:26.25 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:40:26.27 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 23" @16:40:26.30 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:40:26.32 Retry #1 of 10 after 10:00 @16:50:26.39 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:50:26.45 Retry #2 of 10 after 10:00 @17:00:26.96 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:00:27.09 Retry #3 of 10 after 10:00 @17:10:27.14 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:10:27.17 Retry #4 of 10 after 10:00 @17:20:27.22 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:20:27.24 Retry #5 of 10 after 10:00 @17:30:27.34 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:30:27.44 Retry #6 of 10 after 10:00 @17:40:27.53 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:40:27.58 Retry #7 of 10 after 10:00 @17:50:27.64 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:50:27.66 Retry #8 of 10 after 10:00 @18:00:27.72 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:00:27.90 Retry #9 of 10 after 10:00 @18:10:27.95 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:10:27.97 Retry #10 of 10 after 10:00 @18:20:28.03 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:20:28.11 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 03:00PDT26-Jun-21" @18:20:28.19 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:20:28.23 Retry #1 of 10 after 10:00 @18:30:28.30 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:30:28.32 Retry #2 of 10 after 10:00 @18:40:28.39 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:40:28.47 Retry #3 of 10 after 10:00 @18:50:28.54 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:50:28.58 Retry #4 of 10 after 10:00 @19:00:28.80 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:00:28.82 Retry #5 of 10 after 10:00 @19:10:28.98 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:10:29.05 Retry #6 of 10 after 10:00 @19:20:29.10 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:20:29.12 Retry #7 of 10 after 10:00 @19:30:29.24 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:30:29.30 Retry #8 of 10 after 10:00 @19:40:29.35 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:40:29.37 Retry #9 of 10 after 10:00 @19:50:29.47 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:50:29.59 Retry #10 of 10 after 10:00 @20:00:29.68 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:00:00.27PDT26-Jun-21
Gate.power :main,:ON,limit to 5000mA @03:00:00.30 Gate.power :analytic1,:ON,limit to 5000mA @03:00:00.33 Gate.power :analytic2,:ON,limit to 5000mA @03:00:00.37 Gate.power :analytic3,:ON,limit to 5000mA @03:00:00.44 Powering core bus @03:00:02.87 Archive.configure :default @03:00:03.01 Air.configure :default @03:00:03.13 Lysis1.configure :default @03:00:03.24 Lysis2.configure :default @03:00:03.37 Plunge.configure :default @03:00:03.52 SamplePump.configure :default @03:00:03.66 ExtraPump.configure :default @03:00:03.80 Toroid.configure :default @03:00:03.93 Carousel.configure :default @03:00:04.07 Pressure.configure :default @03:00:04.13 All elf controllers running firmware version 4.61 @03:00:04.15 Gate.power :motor,:ON,limit to 5000mA @03:00:04.33 Archive.reconfigure :default @03:00:04.62 Archive.home.jog 0 @03:00:04.82 Plunge.reconfigure :default @03:00:04.97 Lysis2.reconfigure :default @03:00:05.00 Lysis1.reconfigure :default @03:00:05.04 Toroid.reconfigure :default @03:00:05.06 Air.reconfigure :default @03:00:05.24 Plunge.home.jog 0 @03:00:05.29 Lysis2.home.jog 0 @03:00:05.31 Lysis1.home.jog 0 @03:00:05.35 Toroid.home.jog 0 @03:00:05.38 Air.home.jog 0 @03:00:05.81 Archive.reconfigure :homing @03:00:06.07 Archive.jog 1358 @03:00:06.17 Toroid.reconfigure :homing @03:00:06.20 Lysis2.reconfigure :homing @03:00:06.23 Lysis1.reconfigure :homing @03:00:06.28 Plunge.reconfigure :homing @03:00:06.35 Air.reconfigure :homing @03:00:06.57 Lysis2.jog 1338 @03:00:06.61 Toroid.jog 644 @03:00:06.64 Plunge.jog 830 @03:00:06.68 Lysis1.jog 1352 @03:00:06.72 Air.jog 1349 @03:00:09.34 Archive.setRawPosition! 50 @03:00:09.42 Archive.seek :home @03:00:09.82 Archive.reconfigure :default @03:00:10.07 Lysis2.setRawPosition! 50 @03:00:10.20 Lysis2.seek :home @03:00:10.43 Lysis1.setRawPosition! 50 @03:00:10.49 Air.setRawPosition! 50 @03:00:10.51 Plunge.setRawPosition! 50 @03:00:10.80 Lysis1.seek :home @03:00:11.00 Air.seek :home @03:00:11.03 Plunge.seek :home @03:00:11.58 Toroid.setRawPosition! 0 @03:00:11.63 Plunge.reconfigure :default @03:00:11.65 Lysis1.reconfigure :default @03:00:11.78 Lysis2.reconfigure :default @03:00:11.88 Air.reconfigure :default @03:00:12.16 Toroid.seek :home @03:00:12.41 Toroid.reconfigure :default @03:00:13.08
Toroid.seek :clear @03:00:14.70 Carousel.reconfigure :default @03:00:14.75 Carousel.home.jog 0 @03:00:15.26 Carousel.seek 23 @03:00:15.43 Selecting Cartridge 22 @03:00:15.65 Carousel.seek 22 @03:00:16.16 Lysis1.seek :locked @03:00:19.57 Plunge.seek :locked @03:00:19.96 Air.seek :locked @03:00:24.19
Mounted archive Cartridge @03:00:24.30 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 22" @03:00:24.36 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:00:24.38 Retry #1 of 10 after 10:00 @03:00:24.63 Duration of filtering limited to 4:05:00 @03:00:24.92 Priming sample loop w/160ml, bypass w/40ml @03:00:24.94 SamplePump.setPosition! 0ml @03:00:24.97 Bypass.open @03:00:26.43 Intake.open @03:00:27.55 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @03:00:30.30 Vacuum tare is 18% of full scale @03:00:30.35 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @03:00:30.88 Pressure.reconfigure :default @03:00:30.94 Pressure.hold @03:00:31.34 SamplePump.reconfigure :prime @03:00:31.52 SamplePump.seek 40ml,3:51 @03:01:14.77 Pressure.coast @03:01:14.82 SamplePump.setPosition! 0ml @03:01:14.98 Exhaust.open @03:01:16.48 Bypass.close @03:01:18.46 Pressure.reconfigure :default @03:01:18.52 Pressure.hold @03:01:18.80 SamplePump.reconfigure :prime @03:01:18.86 SamplePump.seek 160ml,13:51 @03:04:10.62 Pressure.coast @03:04:10.75 Bypass.open @03:04:11.54 Intake.close @03:04:12.76 Toroid.reconfigure :hold @03:04:12.94 Toroid.seek :crack @03:04:14.56 Toroid.seek :cartridge @03:04:15.50 Toroid.reconfigure :default @03:04:15.56 Toroid.seek :clear @03:04:17.61 Pressurizing cartridge to 26psi above 14.4psia @03:04:17.66 SamplePump.setPosition! 0ml @03:04:17.83 Intake.open @03:04:18.70 Bypass.close @03:04:19.49 Exhaust.close @03:04:20.54 Pressure.reconfigure :default @03:04:20.68 Pressure.hold @03:04:21.20 SamplePump.reconfigure :charge @03:04:21.42 SamplePump.seek 2.3ml,42 seconds @03:04:33.93 Pressure.coast @03:04:33.98 Intake.close @03:04:34.86 Pressurized cartridge with 0.716ml to 26.2psi @03:04:34.87 Isolated @34.5psia (20.1psi) @03:04:34.88 Checking for cartridge leaks from 34.5psia @03:04:34.90 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @03:04:39.63 Cartridge pressure fell 1% to 34.2psia @03:04:39.87 Sampling 2000.0ml @03:04:39.89 SamplePump.setPosition! 0ml @03:04:39.99 Exhaust.open @03:04:40.72 Intake.open @03:04:41.65 Pressure.reconfigure :default @03:04:41.71 Pressure.hold @03:04:42.53 SamplePump.reconfigure :sample @03:04:42.60 SamplePump.seek 2000ml,2:47:12 @03:04:44.72 Cartridge 22 sampling 1/2000ml at 0.56ml/s with 28.0psi vacuum @03:05:44.80 Cartridge 22 sampling 32/2000ml at 0.52ml/s with 25.0psi vacuum @03:06:44.72 Cartridge 22 sampling 63/2000ml at 0.50ml/s with 25.3psi vacuum @03:07:44.87 Cartridge 22 sampling 94/2000ml at 0.51ml/s with 26.1psi vacuum @03:08:44.91 Cartridge 22 sampling 125/2000ml at 0.51ml/s with 21.9psi vacuum @03:09:44.72 Cartridge 22 sampling 156/2000ml at 0.50ml/s with 25.5psi vacuum @03:10:24.43 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:10:24.45 Retry #2 of 10 after 10:00 @03:10:44.93 Cartridge 22 sampling 187/2000ml at 0.51ml/s with 25.8psi vacuum @03:11:44.72 Cartridge 22 sampling 218/2000ml at 0.51ml/s with 23.7psi vacuum @03:12:44.88 Cartridge 22 sampling 249/2000ml at 0.52ml/s with 21.2psi vacuum 2.0m deep (17.6psia) @03:13:44.72 Cartridge 22 sampling 281/2000ml at 0.51ml/s with 26.6psi vacuum @03:14:44.91 Cartridge 22 sampling 312/2000ml at 0.54ml/s with 23.4psi vacuum @03:15:44.72 Cartridge 22 sampling 344/2000ml at 0.52ml/s with 27.2psi vacuum @03:16:44.94 Cartridge 22 sampling 375/2000ml at 0.54ml/s with 18.9psi vacuum 3.2m deep (19.5psia) @03:17:44.72 Cartridge 22 sampling 407/2000ml at 0.53ml/s with 24.2psi vacuum @03:18:44.83 Cartridge 22 sampling 438/2000ml at 0.53ml/s with 26.4psi vacuum @03:19:44.72 Cartridge 22 sampling 469/2000ml at 0.52ml/s with 26.2psi vacuum @03:20:24.55 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:20:24.57 Retry #3 of 10 after 10:00 @03:20:44.88 Cartridge 22 sampling 500/2000ml at 0.53ml/s with 20.6psi vacuum 2.3m deep (18.1psia) @03:21:44.72 Cartridge 22 sampling 531/2000ml at 0.49ml/s with 28.7psi vacuum @03:22:44.72 Cartridge 22 sampling 562/2000ml at 0.50ml/s with 19.3psi vacuum @03:23:45.07 Cartridge 22 sampling 593/2000ml at 0.53ml/s with 23.7psi vacuum @03:24:44.72 Cartridge 22 sampling 623/2000ml at 0.51ml/s with 18.2psi vacuum @03:25:44.88 Cartridge 22 sampling 654/2000ml at 0.52ml/s with 23.7psi vacuum @03:26:44.72 Cartridge 22 sampling 684/2000ml at 0.49ml/s with 28.4psi vacuum @03:27:44.88 Cartridge 22 sampling 715/2000ml at 0.52ml/s with 26.4psi vacuum @03:28:44.72 Cartridge 22 sampling 745/2000ml at 0.50ml/s with 27.2psi vacuum @03:29:44.88 Cartridge 22 sampling 775/2000ml at 0.48ml/s with 28.6psi vacuum @03:30:24.63 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:30:24.65 Retry #4 of 10 after 10:00 @03:30:44.72 Cartridge 22 sampling 805/2000ml at 0.49ml/s with 27.2psi vacuum @03:31:44.88 Cartridge 22 sampling 834/2000ml at 0.51ml/s with 22.7psi vacuum @03:32:44.72 Cartridge 22 sampling 864/2000ml at 0.47ml/s with 28.9psi vacuum @03:33:44.73 Cartridge 22 sampling 893/2000ml at 0.50ml/s with 24.5psi vacuum @03:34:44.72 Cartridge 22 sampling 922/2000ml at 0.49ml/s with 26.8psi vacuum @03:35:44.90 Cartridge 22 sampling 951/2000ml at 0.47ml/s with 21.3psi vacuum @03:36:44.90 Cartridge 22 sampling 980/2000ml at 0.48ml/s with 28.5psi vacuum @03:37:44.72 Cartridge 22 sampling 1009/2000ml at 0.46ml/s with 16.5psi vacuum @03:38:44.94 Cartridge 22 sampling 1038/2000ml at 0.49ml/s with 26.9psi vacuum @03:39:44.72 Cartridge 22 sampling 1067/2000ml at 0.48ml/s with 18.9psi vacuum 1.8m deep (17.4psia) @03:40:25.29 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:40:25.34 Retry #5 of 10 after 10:00 @03:40:44.91 Cartridge 22 sampling 1095/2000ml at 0.49ml/s with 22.5psi vacuum @03:41:44.72 Cartridge 22 sampling 1124/2000ml at 0.46ml/s with 16.6psi vacuum 2.1m deep (17.9psia) @03:42:44.90 Cartridge 22 sampling 1152/2000ml at 0.48ml/s with 27.3psi vacuum @03:43:44.72 Cartridge 22 sampling 1180/2000ml at 0.48ml/s with 26.3psi vacuum @03:44:44.92 Cartridge 22 sampling 1208/2000ml at 0.47ml/s with 16.8psi vacuum 4.6m deep (21.5psia) @03:45:44.72 Cartridge 22 sampling 1236/2000ml at 0.48ml/s with 21.7psi vacuum @03:46:44.73 Cartridge 22 sampling 1263/2000ml at 0.46ml/s with 17.2psi vacuum @03:47:44.72 Cartridge 22 sampling 1291/2000ml at 0.46ml/s with 28.0psi vacuum @03:48:44.89 Cartridge 22 sampling 1318/2000ml at 0.45ml/s with 29.1psi vacuum @03:49:45.02 Cartridge 22 sampling 1345/2000ml at 0.43ml/s with 19.8psi vacuum @03:50:25.41 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:50:25.45 Retry #6 of 10 after 10:00 @03:50:44.72 Cartridge 22 sampling 1373/2000ml at 0.46ml/s with 20.3psi vacuum @03:51:45.14 Cartridge 22 sampling 1400/2000ml at 0.46ml/s with 25.7psi vacuum @03:52:44.71 Cartridge 22 sampling 1427/2000ml at 0.46ml/s with 25.6psi vacuum @03:53:44.89 Cartridge 22 sampling 1453/2000ml at 0.45ml/s with 19.7psi vacuum 3.4m deep (19.7psia) @03:54:44.71 Cartridge 22 sampling 1480/2000ml at 0.41ml/s with 29.7psi vacuum @03:55:44.89 Cartridge 22 sampling 1506/2000ml at 0.45ml/s with 27.0psi vacuum @03:56:44.71 Cartridge 22 sampling 1532/2000ml at 0.42ml/s with 16.3psi vacuum @03:57:45.02 Cartridge 22 sampling 1558/2000ml at 0.43ml/s with 27.8psi vacuum @03:58:44.71 Cartridge 22 sampling 1584/2000ml at 0.45ml/s with 26.1psi vacuum @03:59:44.88 Cartridge 22 sampling 1610/2000ml at 0.41ml/s with 30.1psi vacuum @04:00:25.51 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:00:25.53 Retry #7 of 10 after 10:00 @04:00:44.72 Cartridge 22 sampling 1635/2000ml at 0.41ml/s with 29.3psi vacuum @04:01:44.88 Cartridge 22 sampling 1661/2000ml at 0.42ml/s with 29.0psi vacuum @04:02:44.71 Cartridge 22 sampling 1686/2000ml at 0.43ml/s with 21.5psi vacuum @04:03:44.72 Cartridge 22 sampling 1711/2000ml at 0.42ml/s with 19.6psi vacuum @04:04:44.88 Cartridge 22 sampling 1736/2000ml at 0.40ml/s with 28.9psi vacuum @04:05:44.71 Cartridge 22 sampling 1761/2000ml at 0.38ml/s with 21.7psi vacuum @04:06:44.93 Cartridge 22 sampling 1786/2000ml at 0.38ml/s with 15.5psi vacuum 5.4m deep (22.7psia) @04:07:44.72 Cartridge 22 sampling 1810/2000ml at 0.42ml/s with 22.3psi vacuum @04:08:44.91 Cartridge 22 sampling 1835/2000ml at 0.40ml/s with 29.1psi vacuum @04:09:44.72 Cartridge 22 sampling 1860/2000ml at 0.39ml/s with 30.5psi vacuum @04:10:25.65 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:10:25.69 Retry #8 of 10 after 10:00 @04:10:44.94 Cartridge 22 sampling 1884/2000ml at 0.43ml/s with 24.9psi vacuum @04:11:44.72 Cartridge 22 sampling 1908/2000ml at 0.37ml/s with 27.0psi vacuum @04:12:44.87 Cartridge 22 sampling 1932/2000ml at 0.42ml/s with 18.3psi vacuum 4.6m deep (21.6psia) @04:13:44.72 Cartridge 22 sampling 1957/2000ml at 0.38ml/s with 18.0psi vacuum @04:14:44.84 Cartridge 22 sampling 1980/2000ml at 0.39ml/s with 19.4psi vacuum 4.4m deep (21.2psia) @04:15:33.96 Pressure.coast @04:15:34.10 Intake.close @04:15:34.20 Exhaust.close @04:15:34.79 Toroid.reconfigure :hold @04:15:35.02 Toroid.seek :bypass @04:15:36.68 Toroid.reconfigure :default @04:15:37.07 Toroid.seek :clear @04:15:38.78 Sampled 2000.0ml @04:15:40.77 SamplePump.setPosition! 0ml @04:15:41.01 Plunge.seek :reagent @04:15:41.11 Bypass.open @04:15:41.96 Intake.open @04:15:42.10 Sample loop closed. Ready to Deliver Reagents @04:15:42.12 Archive.seek :sealed @04:15:42.85 SamplePump.reconfigure :slow @04:15:42.91 SamplePump.seek -3ml @04:15:45.91 Archive.seek :evac @04:15:49.63 delay 1 second @04:15:50.78 SamplePump.seek -2.8ml @04:15:51.47 delay 1 second @04:15:52.56 SamplePump.seek -3.3ml @04:15:53.89 delay 1 second @04:15:54.92 Exhaust.open @04:15:55.65 Bypass.close @04:15:56.79 SamplePump.seek -35.3ml @04:15:59.97 delay 1:00 @04:17:00.02 Archive.seek :clear @04:17:05.61 delay 3 seconds @04:17:08.70 Intake.close @04:17:08.74 Exhaust.close @04:17:17.16 RNAlater soak for 10:00 @04:20:25.75 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:20:25.84 Retry #9 of 10 after 10:00 @04:27:17.22 Archive.seek :tVent @04:27:19.18 delay 1 second @04:27:20.24 Archive.seek :evac @04:27:36.30 delay 10 seconds @04:27:46.34 Sample preserved and evacuated @04:27:47.09
Carousel.seek 22 @04:27:47.30 Plunge.seek :locked @04:27:48.28 Archive.seek :clear @04:27:48.45 Air.seek :clear @04:27:48.60 Lysis1.seek :clear @04:28:05.48
Archive.coast @04:28:05.52 Air.coast @04:28:05.57 Lysis1.coast @04:28:05.61 Lysis2.coast @04:28:05.65 Plunge.coast @04:28:05.70 SamplePump.coast @04:28:05.74 Toroid.coast @04:28:05.78 Carousel.coast @04:28:05.97 Gate.power :motor,:OFF @04:28:05.98 Gate.power :core,:OFF @04:28:06.00 Gate.power :analytic1,:OFF @04:28:06.01 Gate.power :analytic2,:OFF @04:28:06.02 Gate.power :analytic3,:OFF @04:28:06.04 Gate.power :main,:OFF @04:28:06.05 Safely stopped and ready to power off @04:28:06.41 delayUntil "15:00:00.00" @04:30:25.90 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:30:25.93 Retry #10 of 10 after 10:00 @04:40:25.99 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:40:26.03 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 22" @04:40:26.13 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:40:26.19 Retry #1 of 10 after 10:00 @04:50:26.25 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:50:26.27 Retry #2 of 10 after 10:00 @05:00:26.35 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:00:26.38 Retry #3 of 10 after 10:00 @05:10:26.42 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:10:26.44 Retry #4 of 10 after 10:00 @05:20:26.49 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:20:26.51 Retry #5 of 10 after 10:00 @05:30:26.58 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:30:26.63 Retry #6 of 10 after 10:00 @05:40:26.71 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:40:26.73 Retry #7 of 10 after 10:00 @05:50:26.78 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:50:26.88 Retry #8 of 10 after 10:00 @06:00:26.93 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:00:26.95 Retry #9 of 10 after 10:00 @06:10:43.82 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 15:00" @06:11:01.33 upsync @06:11:11.46 upsync @15:00:00.24
Gate.power :main,:ON,limit to 5000mA @15:00:00.26 Gate.power :analytic1,:ON,limit to 5000mA @15:00:00.27 Gate.power :analytic2,:ON,limit to 5000mA @15:00:00.29 Gate.power :analytic3,:ON,limit to 5000mA @15:00:00.40 Powering core bus @15:00:02.72 Archive.configure :default @15:00:02.88 Air.configure :default @15:00:03.12 Lysis1.configure :default @15:00:03.28 Lysis2.configure :default @15:00:03.42 Plunge.configure :default @15:00:03.57 SamplePump.configure :default @15:00:03.71 ExtraPump.configure :default @15:00:03.85 Toroid.configure :default @15:00:03.98 Carousel.configure :default @15:00:04.12 Pressure.configure :default @15:00:04.18 All elf controllers running firmware version 4.61 @15:00:04.20 Gate.power :motor,:ON,limit to 5000mA @15:00:04.42 Archive.reconfigure :default @15:00:04.70 Archive.home.jog 0 @15:00:04.91 Lysis1.reconfigure :default @15:00:05.00 Plunge.reconfigure :default @15:00:05.05 Lysis2.reconfigure :default @15:00:05.09 Air.reconfigure :default @15:00:05.16 Toroid.reconfigure :default @15:00:05.30 Lysis1.home.jog 0 @15:00:05.39 Plunge.home.jog 0 @15:00:05.43 Lysis2.home.jog 0 @15:00:05.46 Air.home.jog 0 @15:00:05.51 Toroid.home.jog 0 @15:00:06.61 Archive.reconfigure :homing @15:00:07.18 Archive.jog 1369 @15:00:07.58 Toroid.reconfigure :homing @15:00:07.69 Lysis2.reconfigure :homing @15:00:07.79 Lysis1.reconfigure :homing @15:00:07.92 Plunge.reconfigure :homing @15:00:08.04 Air.reconfigure :homing @15:00:08.53 Toroid.jog 642 @15:00:08.58 Plunge.jog 840 @15:00:08.66 Lysis2.jog 1346 @15:00:08.73 Lysis1.jog 1359 @15:00:08.84 Air.jog 1360 @15:00:09.28 Archive.setRawPosition! 50 @15:00:09.41 Archive.seek :home @15:00:09.69 Archive.reconfigure :default @15:00:09.92 Plunge.setRawPosition! 50 @15:00:09.98 Plunge.seek :home @15:00:10.25 Lysis2.setRawPosition! 50 @15:00:10.41 Lysis2.seek :home @15:00:10.58 Plunge.reconfigure :default @15:00:10.60 Lysis1.setRawPosition! 50 @15:00:10.66 Air.setRawPosition! 50 @15:00:10.91 Lysis1.seek :home @15:00:11.21 Air.seek :home @15:00:11.41 Lysis2.reconfigure :default @15:00:11.53 Lysis1.reconfigure :default @15:00:11.60 Toroid.setRawPosition! 0 @15:00:11.94 Toroid.seek :home @15:00:12.19 Toroid.reconfigure :default @15:00:12.22 Air.reconfigure :default @15:00:12.67
Toroid.seek :clear @15:00:14.27 Carousel.reconfigure :default @15:00:14.33 Carousel.home.jog 0 @15:00:15.39 Carousel.seek 22 @15:00:15.59 Selecting Cartridge 21 @15:00:15.81 Carousel.seek 21 @15:00:16.32 Lysis1.seek :locked @15:00:19.75 Plunge.seek :locked @15:00:20.06 Air.seek :locked @15:00:24.29
Mounted archive Cartridge @15:00:24.43 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 21" @15:00:24.73 Duration of filtering limited to 4:05:00 @15:00:25.02 Priming sample loop w/160ml, bypass w/40ml @15:00:25.04 SamplePump.setPosition! 0ml @15:00:25.08 Bypass.open @15:00:26.72 Intake.open @15:00:27.80 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:00:30.44 Vacuum tare is 18% of full scale @15:00:30.46 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @15:00:31.11 Pressure.reconfigure :default @15:00:31.21 Pressure.hold @15:00:31.52 SamplePump.reconfigure :prime @15:00:31.58 SamplePump.seek 40ml,3:51 @15:00:44.91 upsync @15:01:14.83 Pressure.coast @15:01:14.96 SamplePump.setPosition! 0ml @15:01:15.10 Exhaust.open @15:01:16.60 Bypass.close @15:01:18.15 Pressure.reconfigure :default @15:01:18.20 Pressure.hold @15:01:18.49 SamplePump.reconfigure :prime @15:01:18.55 SamplePump.seek 160ml,13:51 @15:04:10.32 Pressure.coast @15:04:10.38 Bypass.open @15:04:11.13 Intake.close @15:04:12.03 Toroid.reconfigure :hold @15:04:12.09 Toroid.seek :crack @15:04:13.67 Toroid.seek :cartridge @15:04:14.61 Toroid.reconfigure :default @15:04:14.67 Toroid.seek :clear @15:04:17.00 Pressurizing cartridge to 26psi above 14.4psia @15:04:17.06 SamplePump.setPosition! 0ml @15:04:17.24 Intake.open @15:04:18.05 Bypass.close @15:04:18.87 Exhaust.close @15:04:20.03 Pressure.reconfigure :default @15:04:20.09 Pressure.hold @15:04:20.45 SamplePump.reconfigure :charge @15:04:20.59 SamplePump.seek 2.3ml,42 seconds @15:04:33.10 Pressure.coast @15:04:33.15 Intake.close @15:04:34.01 Pressurized cartridge with 0.707ml to 26.5psi @15:04:34.02 Isolated @35.8psia (21.3psi) @15:04:34.03 Checking for cartridge leaks from 35.8psia @15:04:34.05 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @15:04:39.28 Cartridge pressure fell 4% to 34.2psia @15:04:39.39 Sampling 2000.0ml @15:04:39.41 SamplePump.setPosition! 0ml @15:04:39.56 Exhaust.open @15:04:40.29 Intake.open @15:04:41.29 Pressure.reconfigure :default @15:04:41.35 Pressure.hold @15:04:41.62 SamplePump.reconfigure :sample @15:04:41.69 SamplePump.seek 2000ml,2:47:12 @15:04:44.34 Cartridge 21 sampling 1/2000ml at 0.55ml/s with 26.5psi vacuum @15:05:44.56 Cartridge 21 sampling 34/2000ml at 0.55ml/s with 23.1psi vacuum @15:06:44.54 Cartridge 21 sampling 67/2000ml at 0.55ml/s with 24.1psi vacuum @15:07:44.73 Cartridge 21 sampling 100/2000ml at 0.56ml/s with 24.8psi vacuum @15:08:44.74 Cartridge 21 sampling 133/2000ml at 0.55ml/s with 24.9psi vacuum @15:09:44.35 Cartridge 21 sampling 167/2000ml at 0.56ml/s with 20.7psi vacuum @15:10:44.67 Cartridge 21 sampling 201/2000ml at 0.55ml/s with 20.7psi vacuum @15:11:44.35 Cartridge 21 sampling 234/2000ml at 0.56ml/s with 24.7psi vacuum @15:12:44.55 Cartridge 21 sampling 268/2000ml at 0.56ml/s with 22.3psi vacuum @15:13:44.35 Cartridge 21 sampling 301/2000ml at 0.56ml/s with 19.8psi vacuum @15:14:44.52 Cartridge 21 sampling 335/2000ml at 0.57ml/s with 21.1psi vacuum @15:15:44.35 Cartridge 21 sampling 369/2000ml at 0.57ml/s with 25.9psi vacuum @15:16:44.50 Cartridge 21 sampling 404/2000ml at 0.56ml/s with 25.8psi vacuum @15:17:44.34 Cartridge 21 sampling 438/2000ml at 0.58ml/s with 23.6psi vacuum @15:18:44.48 Cartridge 21 sampling 472/2000ml at 0.58ml/s with 23.4psi vacuum @15:19:44.35 Cartridge 21 sampling 506/2000ml at 0.57ml/s with 26.3psi vacuum @15:20:44.89 Cartridge 21 sampling 540/2000ml at 0.56ml/s with 26.9psi vacuum @15:21:44.41 Cartridge 21 sampling 574/2000ml at 0.57ml/s with 20.7psi vacuum @15:22:44.47 Cartridge 21 sampling 608/2000ml at 0.57ml/s with 22.6psi vacuum @15:23:44.51 Cartridge 21 sampling 641/2000ml at 0.57ml/s with 25.9psi vacuum @15:24:44.35 Cartridge 21 sampling 675/2000ml at 0.55ml/s with 26.9psi vacuum @15:25:44.50 Cartridge 21 sampling 709/2000ml at 0.55ml/s with 24.3psi vacuum @15:26:44.35 Cartridge 21 sampling 743/2000ml at 0.57ml/s with 24.4psi vacuum @15:27:44.48 Cartridge 21 sampling 776/2000ml at 0.55ml/s with 21.9psi vacuum @15:28:44.35 Cartridge 21 sampling 810/2000ml at 0.56ml/s with 24.0psi vacuum @15:29:44.48 Cartridge 21 sampling 843/2000ml at 0.57ml/s with 26.1psi vacuum @15:30:44.34 Cartridge 21 sampling 876/2000ml at 0.55ml/s with 24.5psi vacuum @15:31:44.51 Cartridge 21 sampling 910/2000ml at 0.55ml/s with 19.7psi vacuum 2.7m deep (18.8psia) @15:32:44.35 Cartridge 21 sampling 943/2000ml at 0.54ml/s with 26.8psi vacuum @15:33:44.50 Cartridge 21 sampling 975/2000ml at 0.55ml/s with 21.4psi vacuum @15:34:44.34 Cartridge 21 sampling 1008/2000ml at 0.55ml/s with 24.9psi vacuum @15:35:44.58 Cartridge 21 sampling 1041/2000ml at 0.55ml/s with 24.3psi vacuum @15:36:44.46 Cartridge 21 sampling 1073/2000ml at 0.54ml/s with 21.0psi vacuum @15:37:44.34 Cartridge 21 sampling 1105/2000ml at 0.55ml/s with 24.6psi vacuum @15:38:44.65 Cartridge 21 sampling 1138/2000ml at 0.55ml/s with 25.1psi vacuum @15:39:44.35 Cartridge 21 sampling 1170/2000ml at 0.53ml/s with 25.9psi vacuum @15:40:44.55 Cartridge 21 sampling 1202/2000ml at 0.54ml/s with 23.0psi vacuum @15:41:44.34 Cartridge 21 sampling 1234/2000ml at 0.53ml/s with 24.6psi vacuum @15:42:44.50 Cartridge 21 sampling 1266/2000ml at 0.51ml/s with 27.8psi vacuum @15:43:44.35 Cartridge 21 sampling 1297/2000ml at 0.52ml/s with 24.5psi vacuum @15:44:44.43 Cartridge 21 sampling 1329/2000ml at 0.53ml/s with 25.9psi vacuum @15:45:44.35 Cartridge 21 sampling 1360/2000ml at 0.53ml/s with 24.3psi vacuum @15:46:44.50 Cartridge 21 sampling 1392/2000ml at 0.50ml/s with 22.1psi vacuum @15:47:44.35 Cartridge 21 sampling 1423/2000ml at 0.52ml/s with 26.2psi vacuum @15:48:44.49 Cartridge 21 sampling 1454/2000ml at 0.51ml/s with 22.4psi vacuum @15:49:44.34 Cartridge 21 sampling 1485/2000ml at 0.52ml/s with 25.3psi vacuum @15:50:44.48 Cartridge 21 sampling 1515/2000ml at 0.52ml/s with 22.9psi vacuum @15:51:44.56 Cartridge 21 sampling 1545/2000ml at 0.52ml/s with 23.6psi vacuum @15:52:44.34 Cartridge 21 sampling 1576/2000ml at 0.51ml/s with 22.6psi vacuum @15:53:44.59 Cartridge 21 sampling 1606/2000ml at 0.49ml/s with 27.8psi vacuum @15:54:44.34 Cartridge 21 sampling 1636/2000ml at 0.50ml/s with 26.4psi vacuum @15:55:44.44 Cartridge 21 sampling 1666/2000ml at 0.50ml/s with 21.4psi vacuum @15:56:44.34 Cartridge 21 sampling 1696/2000ml at 0.49ml/s with 19.6psi vacuum @15:57:44.46 Cartridge 21 sampling 1726/2000ml at 0.50ml/s with 26.1psi vacuum @15:58:44.34 Cartridge 21 sampling 1756/2000ml at 0.47ml/s with 25.4psi vacuum @15:59:44.53 Cartridge 21 sampling 1785/2000ml at 0.48ml/s with 19.4psi vacuum @16:00:44.33 Cartridge 21 sampling 1814/2000ml at 0.49ml/s with 26.0psi vacuum @16:01:44.46 Cartridge 21 sampling 1844/2000ml at 0.49ml/s with 23.1psi vacuum @16:02:44.34 Cartridge 21 sampling 1873/2000ml at 0.48ml/s with 27.3psi vacuum @16:03:44.76 Cartridge 21 sampling 1902/2000ml at 0.49ml/s with 20.1psi vacuum @16:04:44.55 Cartridge 21 sampling 1930/2000ml at 0.47ml/s with 28.5psi vacuum @16:05:44.35 Cartridge 21 sampling 1959/2000ml at 0.48ml/s with 27.5psi vacuum @16:06:44.57 Cartridge 21 sampling 1988/2000ml at 0.47ml/s with 27.3psi vacuum @16:07:09.80 Pressure.coast @16:07:09.93 Intake.close @16:07:10.00 Exhaust.close @16:07:10.18 Toroid.reconfigure :hold @16:07:10.24 Toroid.seek :bypass @16:07:12.02 Toroid.reconfigure :default @16:07:12.25 Toroid.seek :clear @16:07:13.92 Sampled 2000.0ml @16:07:15.00 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 21" @16:07:15.72 SamplePump.setPosition! 0ml @16:07:15.84 Plunge.seek :reagent @16:07:16.13 Bypass.open @16:07:17.07 Sample loop closed. Ready to Deliver Reagents @16:07:17.09 Archive.seek :sealed @16:07:17.16 Intake.open @16:07:18.02 SamplePump.reconfigure :slow @16:07:18.08 SamplePump.seek -3ml @16:07:20.73 Archive.seek :evac @16:07:24.79 delay 1 second @16:07:25.91 SamplePump.seek -2.8ml @16:07:26.60 delay 1 second @16:07:27.69 SamplePump.seek -3.3ml @16:07:29.02 delay 1 second @16:07:30.13 Exhaust.open @16:07:30.91 Bypass.close @16:07:32.01 SamplePump.seek -35.3ml @16:07:34.96 delay 1:00 @16:07:48.41 upsync @16:08:35.15 Archive.seek :clear @16:08:40.83 delay 3 seconds @16:08:44.13 Intake.close @16:08:44.34 Exhaust.close @16:08:52.32 RNAlater soak for 10:00 @16:18:52.37 Archive.seek :tVent @16:18:54.34 delay 1 second @16:18:55.43 Archive.seek :evac @16:19:11.40 delay 10 seconds @16:19:21.46 Sample preserved and evacuated @16:19:22.10
Carousel.seek 21 @16:19:22.54 Plunge.seek :locked @16:19:23.51 Archive.seek :clear @16:19:23.67 Air.seek :clear @16:19:23.86 Lysis1.seek :clear @16:19:40.70
Archive.coast @16:19:40.74 Air.coast @16:19:40.78 Lysis1.coast @16:19:40.83 Lysis2.coast @16:19:40.87 Plunge.coast @16:19:40.91 SamplePump.coast @16:19:40.95 Toroid.coast @16:19:41.00 Carousel.coast @16:19:41.19 Gate.power :motor,:OFF @16:19:41.20 Gate.power :core,:OFF @16:19:41.22 Gate.power :analytic1,:OFF @16:19:41.23 Gate.power :analytic2,:OFF @16:19:41.24 Gate.power :analytic3,:OFF @16:19:41.26 Gate.power :main,:OFF @16:19:41.27 Safely stopped and ready to power off @16:19:41.66 Cartridge Load: {[52..47, 45..27] => { Type: :archive_bac, State: :processed, filtered: 2000}, [59, 58] => { Type: :archive_bac, State: :processed, filtered: 400}, 20..1 => { Type: :archive_bac, State: :reserved, was: :dry}, 46 => { Type: :archive_bac, State: :leaky}, 53 => { Type: :archive_bac, State: :processed, filtered: 1396.8}, 54 => { Type: :archive_bac, State: :processed, filtered: 721.3}, 55 => { Type: :archive_bac, State: :processed, filtered: 622.6}, 56 => { Type: :archive_bac, State: :processed, filtered: 387.7}, 57 => { Type: :archive_bac, State: :processed, filtered: 1000}, 26..21 => { Type: :archive_bac, State: :processed, filtered: 2000.0}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 1} @16:19:41.70 BadNews.email "Not enough dry cartridges remain",Subject: "Cartridge::None in MAIN" @16:20:05.44 upsync @17:09:32.49PDT26-Jun-21
>BREAK< ESP3 Gateway v1.11 for core bus @17:09:32.51 >BREAK< ESP3 Gateway v1.11 for cartridge bus @17:09:38.15 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @17:09:38.91 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @17:09:48.15 Polling Can every 2:00 @17:09:48.26 Loading /home/esp/esp2/bin/processAll mission @17:09:51.46 Sampling default goal volumes @17:09:51.51 GoodNews.email "Cartridge group: dry\nSpares: {spare: 1}\nSchedule: [\"3am\", \"3pm\"]\n",Subject: "Begun processAll --schedule=3am,3pm --checkpoint --mail" @17:09:51.63
delayUntil "03:00:00.00PDT27-Jun-21" @17:09:55.52 Session Begun @17:10:06.68 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 03:00PDT27-Jun-21" @17:10:08.18 -> showlog 30 @17:10:21.24 upsync @17:10:38.86 upsync @17:33:46.17 -> exit @17:33:46.19 Session Ended @03:00:00.23PDT27-Jun-21
Gate.power :main,:ON,limit to 5000mA @03:00:00.27 Gate.power :analytic1,:ON,limit to 5000mA @03:00:00.30 Gate.power :analytic2,:ON,limit to 5000mA @03:00:00.33 Gate.power :analytic3,:ON,limit to 5000mA @03:00:00.37 Powering core bus @03:00:02.76 Archive.configure :default @03:00:02.89 Air.configure :default @03:00:03.00 Lysis1.configure :default @03:00:03.11 Lysis2.configure :default @03:00:03.25 Plunge.configure :default @03:00:03.39 SamplePump.configure :default @03:00:03.54 ExtraPump.configure :default @03:00:03.68 Toroid.configure :default @03:00:03.81 Carousel.configure :default @03:00:03.96 Pressure.configure :default @03:00:04.02 All elf controllers running firmware version 4.61 @03:00:04.04 Gate.power :motor,:ON,limit to 5000mA @03:00:04.30 Archive.reconfigure :default @03:00:04.63 Archive.home.jog 0 @03:00:04.72 Air.reconfigure :default @03:00:04.75 Toroid.reconfigure :default @03:00:04.79 Lysis2.reconfigure :default @03:00:04.89 Lysis1.reconfigure :default @03:00:04.97 Plunge.reconfigure :default @03:00:05.30 Toroid.home.jog 0 @03:00:05.34 Lysis2.home.jog 0 @03:00:05.39 Air.home.jog 0 @03:00:05.44 Lysis1.home.jog 0 @03:00:05.48 Plunge.home.jog 0 @03:00:06.14 Toroid.reconfigure :homing @03:00:06.20 Plunge.reconfigure :homing @03:00:06.24 Lysis2.reconfigure :homing @03:00:06.28 Air.reconfigure :homing @03:00:06.32 Archive.reconfigure :homing @03:00:06.40 Lysis1.reconfigure :homing @03:00:06.50 Toroid.jog 629 @03:00:06.63 Air.jog 1352 @03:00:06.67 Lysis2.jog 1351 @03:00:06.73 Plunge.jog 841 @03:00:06.76 Archive.jog 1362 @03:00:06.80 Lysis1.jog 1355 @03:00:07.85 Plunge.setRawPosition! 50 @03:00:08.15 Plunge.seek :home @03:00:08.49 Lysis2.setRawPosition! 50 @03:00:08.52 Lysis1.setRawPosition! 50 @03:00:08.55 Air.setRawPosition! 50 @03:00:08.58 Archive.setRawPosition! 50 @03:00:09.04 Lysis2.seek :home @03:00:09.16 Lysis1.seek :home @03:00:09.30 Air.seek :home @03:00:09.44 Archive.seek :home @03:00:11.32 Plunge.reconfigure :default @03:00:11.36 Lysis2.reconfigure :default @03:00:11.38 Toroid.setRawPosition! 0 @03:00:11.46 Lysis1.reconfigure :default @03:00:11.81 Toroid.seek :home @03:00:12.09 Toroid.reconfigure :default @03:00:12.18 Air.reconfigure :default @03:00:12.31 Archive.reconfigure :default @03:00:13.19
Toroid.seek :clear @03:00:14.80 Carousel.reconfigure :default @03:00:14.85 Carousel.home.jog 0 @03:00:15.35 Carousel.seek 21 @03:00:15.52 Selecting Cartridge 20 @03:00:15.73 Carousel.seek 20 @03:00:16.26 Lysis1.seek :locked @03:00:19.66 Plunge.seek :locked @03:00:19.97 Air.seek :locked @03:00:24.20
Mounted archive Cartridge @03:00:24.31 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 20" @03:00:24.62 Duration of filtering limited to 4:05:00 @03:00:24.90 Priming sample loop w/160ml, bypass w/40ml @03:00:24.92 SamplePump.setPosition! 0ml @03:00:24.96 Bypass.open @03:00:26.44 Intake.open @03:00:27.75 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @03:00:30.40 Vacuum tare is 18% of full scale @03:00:30.50 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @03:00:30.89 Pressure.reconfigure :default @03:00:30.95 Pressure.hold @03:00:31.31 SamplePump.reconfigure :prime @03:00:31.47 SamplePump.seek 40ml,3:51 @03:00:43.29 upsync @03:01:14.79 Pressure.coast @03:01:14.87 SamplePump.setPosition! 0ml @03:01:14.99 Exhaust.open @03:01:16.60 Bypass.close @03:01:18.35 Pressure.reconfigure :default @03:01:18.41 Pressure.hold @03:01:18.69 SamplePump.reconfigure :prime @03:01:18.76 SamplePump.seek 160ml,13:51 @03:04:10.50 Pressure.coast @03:04:10.62 Bypass.open @03:04:11.34 Intake.close @03:04:12.44 Toroid.reconfigure :hold @03:04:12.74 Toroid.seek :crack @03:04:14.39 Toroid.seek :cartridge @03:04:15.33 Toroid.reconfigure :default @03:04:15.39 Toroid.seek :clear @03:04:17.72 Pressurizing cartridge to 26psi above 14.6psia @03:04:17.78 SamplePump.setPosition! 0ml @03:04:17.88 Intake.open @03:04:18.71 Bypass.close @03:04:19.40 Exhaust.close @03:04:20.55 Pressure.reconfigure :default @03:04:20.63 Pressure.hold @03:04:21.16 SamplePump.reconfigure :charge @03:04:21.27 SamplePump.seek 2.3ml,42 seconds @03:04:33.94 Pressure.coast @03:04:33.99 Intake.close @03:04:34.85 Pressurized cartridge with 0.722ml to 26.0psi @03:04:34.86 Isolated @35.7psia (21.1psi) @03:04:34.87 Checking for cartridge leaks from 35.7psia @03:04:34.89 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @03:04:42.48 Cartridge pressure fell 7% to 33.2psia @03:04:42.57 Sampling 2000.0ml @03:04:42.58 SamplePump.setPosition! 0ml @03:04:42.62 Exhaust.open @03:04:43.33 Intake.open @03:04:44.31 Pressure.reconfigure :default @03:04:44.37 Pressure.hold @03:04:44.64 SamplePump.reconfigure :sample @03:04:44.71 SamplePump.seek 2000ml,2:47:12 @03:04:47.51 Cartridge 20 sampling 1/2000ml at 0.55ml/s with 26.0psi vacuum @03:05:47.38 Cartridge 20 sampling 33/2000ml at 0.53ml/s with 24.7psi vacuum @03:06:47.55 Cartridge 20 sampling 65/2000ml at 0.53ml/s with 21.0psi vacuum @03:07:47.38 Cartridge 20 sampling 96/2000ml at 0.54ml/s with 21.8psi vacuum @03:08:47.62 Cartridge 20 sampling 128/2000ml at 0.54ml/s with 21.8psi vacuum @03:09:47.38 Cartridge 20 sampling 160/2000ml at 0.53ml/s with 26.1psi vacuum @03:10:47.53 Cartridge 20 sampling 192/2000ml at 0.53ml/s with 26.3psi vacuum @03:11:47.37 Cartridge 20 sampling 225/2000ml at 0.53ml/s with 24.3psi vacuum @03:12:47.66 Cartridge 20 sampling 257/2000ml at 0.53ml/s with 19.2psi vacuum 2.0m deep (17.7psia) @03:13:47.42 Cartridge 20 sampling 289/2000ml at 0.55ml/s with 22.2psi vacuum @03:14:47.37 Cartridge 20 sampling 322/2000ml at 0.54ml/s with 27.5psi vacuum @03:15:47.73 Cartridge 20 sampling 354/2000ml at 0.54ml/s with 24.1psi vacuum @03:16:47.37 Cartridge 20 sampling 386/2000ml at 0.54ml/s with 25.7psi vacuum @03:17:47.61 Cartridge 20 sampling 419/2000ml at 0.54ml/s with 22.8psi vacuum @03:18:47.37 Cartridge 20 sampling 451/2000ml at 0.52ml/s with 18.7psi vacuum @03:19:47.55 Cartridge 20 sampling 483/2000ml at 0.55ml/s with 27.1psi vacuum @03:20:47.37 Cartridge 20 sampling 515/2000ml at 0.55ml/s with 21.6psi vacuum @03:21:47.67 Cartridge 20 sampling 547/2000ml at 0.54ml/s with 21.4psi vacuum @03:22:47.37 Cartridge 20 sampling 579/2000ml at 0.53ml/s with 22.6psi vacuum @03:23:47.67 Cartridge 20 sampling 611/2000ml at 0.52ml/s with 27.9psi vacuum @03:24:47.37 Cartridge 20 sampling 643/2000ml at 0.53ml/s with 20.5psi vacuum @03:25:47.46 Cartridge 20 sampling 675/2000ml at 0.50ml/s with 27.4psi vacuum @03:26:47.37 Cartridge 20 sampling 706/2000ml at 0.51ml/s with 21.2psi vacuum @03:27:47.38 Cartridge 20 sampling 737/2000ml at 0.53ml/s with 26.9psi vacuum @03:28:47.64 Cartridge 20 sampling 769/2000ml at 0.51ml/s with 28.5psi vacuum @03:29:47.38 Cartridge 20 sampling 800/2000ml at 0.53ml/s with 20.2psi vacuum @03:30:47.58 Cartridge 20 sampling 831/2000ml at 0.50ml/s with 28.1psi vacuum @03:31:47.37 Cartridge 20 sampling 861/2000ml at 0.53ml/s with 26.2psi vacuum @03:32:47.46 Cartridge 20 sampling 892/2000ml at 0.49ml/s with 28.0psi vacuum @03:33:47.38 Cartridge 20 sampling 922/2000ml at 0.51ml/s with 20.8psi vacuum @03:34:47.54 Cartridge 20 sampling 953/2000ml at 0.50ml/s with 17.8psi vacuum @03:35:47.38 Cartridge 20 sampling 983/2000ml at 0.51ml/s with 27.1psi vacuum @03:36:47.48 Cartridge 20 sampling 1013/2000ml at 0.48ml/s with 18.9psi vacuum 2.7m deep (18.7psia) @03:37:47.38 Cartridge 20 sampling 1043/2000ml at 0.50ml/s with 21.9psi vacuum @03:38:47.43 Cartridge 20 sampling 1073/2000ml at 0.51ml/s with 20.5psi vacuum 2.2m deep (18.0psia) @03:39:47.37 Cartridge 20 sampling 1103/2000ml at 0.47ml/s with 24.5psi vacuum @03:40:47.61 Cartridge 20 sampling 1132/2000ml at 0.50ml/s with 21.1psi vacuum @03:41:47.57 Cartridge 20 sampling 1162/2000ml at 0.48ml/s with 16.5psi vacuum 5.8m deep (23.3psia) @03:42:47.38 Cartridge 20 sampling 1191/2000ml at 0.50ml/s with 22.6psi vacuum @03:43:47.80 Cartridge 20 sampling 1220/2000ml at 0.49ml/s with 20.9psi vacuum @03:44:47.37 Cartridge 20 sampling 1249/2000ml at 0.47ml/s with 28.6psi vacuum @03:45:47.58 Cartridge 20 sampling 1277/2000ml at 0.48ml/s with 20.0psi vacuum 1.8m deep (17.4psia) @03:46:47.44 Cartridge 20 sampling 1306/2000ml at 0.49ml/s with 28.7psi vacuum @03:47:47.55 Cartridge 20 sampling 1335/2000ml at 0.49ml/s with 22.9psi vacuum @03:48:47.38 Cartridge 20 sampling 1364/2000ml at 0.49ml/s with 21.5psi vacuum @03:49:47.56 Cartridge 20 sampling 1392/2000ml at 0.49ml/s with 20.9psi vacuum @03:50:47.38 Cartridge 20 sampling 1421/2000ml at 0.48ml/s with 23.1psi vacuum @03:51:47.50 Cartridge 20 sampling 1449/2000ml at 0.48ml/s with 28.1psi vacuum @03:52:47.37 Cartridge 20 sampling 1477/2000ml at 0.47ml/s with 22.6psi vacuum @03:53:47.66 Cartridge 20 sampling 1505/2000ml at 0.48ml/s with 26.0psi vacuum @03:54:47.72 Cartridge 20 sampling 1533/2000ml at 0.45ml/s with 29.7psi vacuum @03:55:47.57 Cartridge 20 sampling 1560/2000ml at 0.46ml/s with 28.7psi vacuum @03:56:47.58 Cartridge 20 sampling 1588/2000ml at 0.47ml/s with 26.9psi vacuum @03:57:47.38 Cartridge 20 sampling 1615/2000ml at 0.47ml/s with 23.5psi vacuum @03:58:47.57 Cartridge 20 sampling 1643/2000ml at 0.45ml/s with 28.1psi vacuum @03:59:47.37 Cartridge 20 sampling 1669/2000ml at 0.45ml/s with 17.5psi vacuum @04:00:47.56 Cartridge 20 sampling 1696/2000ml at 0.44ml/s with 18.5psi vacuum 3.6m deep (19.9psia) @04:01:47.37 Cartridge 20 sampling 1723/2000ml at 0.46ml/s with 25.6psi vacuum @04:02:47.56 Cartridge 20 sampling 1749/2000ml at 0.45ml/s with 24.3psi vacuum @04:03:47.38 Cartridge 20 sampling 1776/2000ml at 0.44ml/s with 17.3psi vacuum 2.1m deep (17.8psia) @04:04:47.54 Cartridge 20 sampling 1802/2000ml at 0.41ml/s with 29.1psi vacuum @04:05:47.38 Cartridge 20 sampling 1829/2000ml at 0.42ml/s with 29.4psi vacuum @04:06:47.53 Cartridge 20 sampling 1855/2000ml at 0.43ml/s with 26.9psi vacuum @04:07:47.38 Cartridge 20 sampling 1881/2000ml at 0.41ml/s with 28.5psi vacuum @04:08:47.64 Cartridge 20 sampling 1907/2000ml at 0.45ml/s with 25.8psi vacuum @04:09:47.66 Cartridge 20 sampling 1932/2000ml at 0.43ml/s with 22.7psi vacuum @04:10:47.56 Cartridge 20 sampling 1958/2000ml at 0.40ml/s with 14.4psi vacuum 7.3m deep (25.5psia) @04:11:47.77 Cartridge 20 sampling 1983/2000ml at 0.42ml/s with 17.5psi vacuum @04:12:27.73 Pressure.coast @04:12:27.94 Intake.close @04:12:28.08 Exhaust.close @04:12:28.65 Toroid.reconfigure :hold @04:12:28.70 Toroid.seek :bypass @04:12:30.32 Toroid.reconfigure :default @04:12:30.38 Toroid.seek :clear @04:12:31.98 Sampled 2000.0ml @04:12:32.44 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 20" @04:12:32.85 SamplePump.setPosition! 0ml @04:12:32.96 Plunge.seek :reagent @04:12:33.43 Bypass.open @04:12:34.22 Sample loop closed. Ready to Deliver Reagents @04:12:34.30 Archive.seek :sealed @04:12:34.41 Intake.open @04:12:35.54 SamplePump.reconfigure :slow @04:12:35.65 SamplePump.seek -3ml @04:12:37.99 Archive.seek :evac @04:12:42.43 delay 1 second @04:12:43.68 SamplePump.seek -2.8ml @04:12:44.38 delay 1 second @04:12:45.77 SamplePump.seek -3.3ml @04:12:47.24 delay 1 second @04:12:48.29 Exhaust.open @04:12:49.01 Bypass.close @04:12:49.80 SamplePump.seek -35.3ml @04:12:52.13 delay 1:00 @04:12:53.65 upsync @04:13:52.17 Archive.seek :clear @04:13:58.60 delay 3 seconds @04:14:01.71 Intake.close @04:14:01.77 Exhaust.close @04:14:09.32 RNAlater soak for 10:00 @04:24:09.97 Archive.seek :tVent @04:24:11.69 delay 1 second @04:24:12.81 Archive.seek :evac @04:24:28.76 delay 10 seconds @04:24:38.82 Sample preserved and evacuated @04:24:39.47
Carousel.seek 20 @04:24:40.07 Plunge.seek :locked @04:24:41.24 Archive.seek :clear @04:24:41.56 Air.seek :clear @04:24:41.89 Lysis1.seek :clear @04:24:58.56
Archive.coast @04:24:58.70 Air.coast @04:24:58.86 Lysis1.coast @04:24:59.01 Lysis2.coast @04:24:59.10 Plunge.coast @04:24:59.22 SamplePump.coast @04:24:59.30 Toroid.coast @04:24:59.36 Carousel.coast @04:24:59.75 Gate.power :motor,:OFF @04:24:59.81 Gate.power :core,:OFF @04:24:59.84 Gate.power :analytic1,:OFF @04:24:59.85 Gate.power :analytic2,:OFF @04:24:59.87 Gate.power :analytic3,:OFF @04:24:59.88 Gate.power :main,:OFF @04:24:59.90 Safely stopped and ready to power off @04:25:00.38 delayUntil "15:00:00.00" @04:25:00.43 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 15:00" @04:25:16.27 upsync @15:00:00.36
Gate.power :main,:ON,limit to 5000mA @15:00:00.40 Gate.power :analytic1,:ON,limit to 5000mA @15:00:00.42 Gate.power :analytic2,:ON,limit to 5000mA @15:00:00.46 Gate.power :analytic3,:ON,limit to 5000mA @15:00:00.53 Powering core bus @15:00:02.82 Archive.configure :default @15:00:02.95 Air.configure :default @15:00:03.06 Lysis1.configure :default @15:00:03.18 Lysis2.configure :default @15:00:03.31 Plunge.configure :default @15:00:03.45 SamplePump.configure :default @15:00:03.59 ExtraPump.configure :default @15:00:03.73 Toroid.configure :default @15:00:03.87 Carousel.configure :default @15:00:04.01 Pressure.configure :default @15:00:04.07 All elf controllers running firmware version 4.61 @15:00:04.09 Gate.power :motor,:ON,limit to 5000mA @15:00:04.44 Archive.reconfigure :default @15:00:04.71 Archive.home.jog 0 @15:00:05.44 Toroid.reconfigure :default @15:00:05.57 Plunge.reconfigure :default @15:00:05.79 Lysis2.reconfigure :default @15:00:05.94 Lysis1.reconfigure :default @15:00:06.10 Air.reconfigure :default @15:00:06.51 Toroid.home.jog 0 @15:00:06.63 Plunge.home.jog 0 @15:00:06.75 Lysis2.home.jog 0 @15:00:06.81 Lysis1.home.jog 0 @15:00:06.86 Air.home.jog 0 @15:00:07.97 Archive.reconfigure :homing @15:00:08.83 Archive.jog 1369 @15:00:08.94 Plunge.reconfigure :homing @15:00:08.99 Lysis2.reconfigure :homing @15:00:09.06 Lysis1.reconfigure :homing @15:00:09.19 Air.reconfigure :homing @15:00:09.25 Toroid.reconfigure :homing @15:00:09.76 Plunge.jog 841 @15:00:09.84 Lysis1.jog 1358 @15:00:09.89 Lysis2.jog 1347 @15:00:09.94 Air.jog 1359 @15:00:10.14 Toroid.jog 624 @15:00:10.83 Archive.setRawPosition! 50 @15:00:10.87 Archive.seek :home @15:00:10.96 Plunge.setRawPosition! 50 @15:00:11.01 Plunge.seek :home @15:00:11.16 Archive.reconfigure :default @15:00:11.29 Plunge.reconfigure :default @15:00:12.00 Lysis2.setRawPosition! 50 @15:00:12.03 Lysis1.setRawPosition! 50 @15:00:12.04 Air.setRawPosition! 50 @15:00:12.15 Lysis2.seek :home @15:00:12.20 Lysis1.seek :home @15:00:12.28 Air.seek :home @15:00:12.51 Lysis2.reconfigure :default @15:00:12.60 Lysis1.reconfigure :default @15:00:12.70 Air.reconfigure :default @15:00:12.82 Toroid.setRawPosition! 0 @15:00:12.87 Toroid.seek :home @15:00:12.97 Toroid.reconfigure :default @15:00:13.14
Toroid.seek :clear @15:00:14.75 Carousel.reconfigure :default @15:00:14.80 Carousel.home.jog 0 @15:00:15.47 Carousel.seek 20 @15:00:16.03 Selecting Cartridge 19 @15:00:16.72 Carousel.seek 19 @15:00:17.47 Lysis1.seek :locked @15:00:20.71 Plunge.seek :locked @15:00:20.86 Air.seek :locked @15:00:25.11
Mounted archive Cartridge @15:00:25.53 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 19" @15:00:25.69 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:00:25.76 Retry #1 of 10 after 10:00 @15:00:26.43 Duration of filtering limited to 4:05:00 @15:00:27.09 Priming sample loop w/160ml, bypass w/40ml @15:00:27.14 SamplePump.setPosition! 0ml @15:00:27.23 Bypass.open @15:00:28.88 Intake.open @15:00:30.12 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:00:32.72 Vacuum tare is 18% of full scale @15:00:32.74 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @15:00:32.95 Pressure.reconfigure :default @15:00:33.00 Pressure.hold @15:00:33.21 SamplePump.reconfigure :prime @15:00:33.27 SamplePump.seek 40ml,3:51 @15:01:16.50 Pressure.coast @15:01:16.55 SamplePump.setPosition! 0ml @15:01:16.60 Exhaust.open @15:01:18.01 Bypass.close @15:01:19.57 Pressure.reconfigure :default @15:01:19.63 Pressure.hold @15:01:19.89 SamplePump.reconfigure :prime @15:01:19.96 SamplePump.seek 160ml,13:51 @15:04:11.70 Pressure.coast @15:04:11.75 Bypass.open @15:04:12.49 Intake.close @15:04:13.38 Toroid.reconfigure :hold @15:04:13.43 Toroid.seek :crack @15:04:15.26 Toroid.seek :cartridge @15:04:16.35 Toroid.reconfigure :default @15:04:16.50 Toroid.seek :clear @15:04:18.63 Pressurizing cartridge to 26psi above 14.5psia @15:04:18.65 SamplePump.setPosition! 0ml @15:04:18.72 Intake.open @15:04:19.61 Bypass.close @15:04:20.39 Exhaust.close @15:04:21.20 Pressure.reconfigure :default @15:04:21.26 Pressure.hold @15:04:21.46 SamplePump.reconfigure :charge @15:04:21.53 SamplePump.seek 2.3ml,42 seconds @15:04:34.67 Pressure.coast @15:04:34.79 Intake.close @15:04:35.85 Pressurized cartridge with 0.774ml to 26.1psi @15:04:35.91 Isolated @33.5psia (19.0psi) @15:04:35.95 Checking for cartridge leaks from 33.5psia @15:04:35.97 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @15:04:41.06 Cartridge pressure fell 8% to 30.7psia @15:04:41.15 Sampling 2000.0ml @15:04:41.16 SamplePump.setPosition! 0ml @15:04:41.20 Exhaust.open @15:04:41.92 Intake.open @15:04:42.86 Pressure.reconfigure :default @15:04:42.92 Pressure.hold @15:04:43.30 SamplePump.reconfigure :sample @15:04:43.49 SamplePump.seek 2000ml,2:47:12 @15:04:46.10 Cartridge 19 sampling 1/2000ml at 0.59ml/s with 26.9psi vacuum @15:05:45.93 Cartridge 19 sampling 33/2000ml at 0.54ml/s with 24.1psi vacuum @15:06:46.05 Cartridge 19 sampling 66/2000ml at 0.53ml/s with 24.8psi vacuum @15:07:45.93 Cartridge 19 sampling 98/2000ml at 0.53ml/s with 24.9psi vacuum @15:08:46.26 Cartridge 19 sampling 131/2000ml at 0.53ml/s with 27.2psi vacuum @15:09:45.93 Cartridge 19 sampling 164/2000ml at 0.55ml/s with 20.9psi vacuum @15:10:25.84 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:10:25.89 Retry #2 of 10 after 10:00 @15:10:46.12 Cartridge 19 sampling 197/2000ml at 0.55ml/s with 23.8psi vacuum @15:11:46.29 Cartridge 19 sampling 231/2000ml at 0.55ml/s with 23.0psi vacuum @15:12:45.92 Cartridge 19 sampling 264/2000ml at 0.55ml/s with 22.0psi vacuum @15:13:46.16 Cartridge 19 sampling 297/2000ml at 0.56ml/s with 21.9psi vacuum @15:14:45.91 Cartridge 19 sampling 330/2000ml at 0.53ml/s with 28.1psi vacuum @15:15:46.12 Cartridge 19 sampling 363/2000ml at 0.56ml/s with 21.9psi vacuum @15:16:45.92 Cartridge 19 sampling 396/2000ml at 0.56ml/s with 23.0psi vacuum @15:17:46.17 Cartridge 19 sampling 429/2000ml at 0.53ml/s with 27.7psi vacuum @15:18:45.92 Cartridge 19 sampling 462/2000ml at 0.56ml/s with 25.6psi vacuum @15:19:46.09 Cartridge 19 sampling 495/2000ml at 0.52ml/s with 28.3psi vacuum @15:20:25.95 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:20:25.97 Retry #3 of 10 after 10:00 @15:20:45.92 Cartridge 19 sampling 528/2000ml at 0.55ml/s with 23.8psi vacuum @15:21:46.05 Cartridge 19 sampling 560/2000ml at 0.52ml/s with 23.1psi vacuum @15:22:45.91 Cartridge 19 sampling 593/2000ml at 0.55ml/s with 26.0psi vacuum @15:23:46.21 Cartridge 19 sampling 626/2000ml at 0.56ml/s with 26.1psi vacuum @15:24:46.21 Cartridge 19 sampling 658/2000ml at 0.53ml/s with 20.3psi vacuum @15:25:46.50 Cartridge 19 sampling 690/2000ml at 0.53ml/s with 20.3psi vacuum @15:26:46.28 Cartridge 19 sampling 723/2000ml at 0.54ml/s with 24.3psi vacuum @15:27:45.93 Cartridge 19 sampling 755/2000ml at 0.55ml/s with 19.4psi vacuum 1.8m deep (17.4psia) @15:28:46.05 Cartridge 19 sampling 787/2000ml at 0.52ml/s with 18.8psi vacuum @15:29:45.93 Cartridge 19 sampling 819/2000ml at 0.52ml/s with 21.7psi vacuum @15:30:26.55 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:30:26.63 Retry #4 of 10 after 10:00 @15:30:46.13 Cartridge 19 sampling 850/2000ml at 0.53ml/s with 25.7psi vacuum @15:31:45.93 Cartridge 19 sampling 882/2000ml at 0.51ml/s with 28.9psi vacuum @15:32:46.05 Cartridge 19 sampling 913/2000ml at 0.53ml/s with 21.4psi vacuum @15:33:45.93 Cartridge 19 sampling 945/2000ml at 0.53ml/s with 22.1psi vacuum @15:34:46.06 Cartridge 19 sampling 976/2000ml at 0.52ml/s with 28.0psi vacuum @15:35:45.93 Cartridge 19 sampling 1007/2000ml at 0.50ml/s with 27.2psi vacuum @15:36:46.06 Cartridge 19 sampling 1038/2000ml at 0.53ml/s with 19.0psi vacuum @15:37:45.93 Cartridge 19 sampling 1069/2000ml at 0.51ml/s with 25.2psi vacuum @15:38:46.00 Cartridge 19 sampling 1100/2000ml at 0.51ml/s with 27.8psi vacuum @15:39:45.93 Cartridge 19 sampling 1131/2000ml at 0.50ml/s with 17.8psi vacuum @15:40:26.71 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:40:26.76 Retry #5 of 10 after 10:00 @15:40:46.03 Cartridge 19 sampling 1161/2000ml at 0.51ml/s with 21.8psi vacuum @15:41:45.93 Cartridge 19 sampling 1191/2000ml at 0.50ml/s with 22.1psi vacuum @15:42:45.93 Cartridge 19 sampling 1221/2000ml at 0.52ml/s with 24.6psi vacuum @15:43:46.19 Cartridge 19 sampling 1252/2000ml at 0.49ml/s with 22.5psi vacuum 1.7m deep (17.3psia) @15:44:45.93 Cartridge 19 sampling 1281/2000ml at 0.51ml/s with 18.7psi vacuum 2.3m deep (18.1psia) @15:45:46.36 Cartridge 19 sampling 1311/2000ml at 0.48ml/s with 17.5psi vacuum @15:46:45.93 Cartridge 19 sampling 1341/2000ml at 0.49ml/s with 20.2psi vacuum @15:47:46.09 Cartridge 19 sampling 1370/2000ml at 0.49ml/s with 28.0psi vacuum @15:48:45.93 Cartridge 19 sampling 1399/2000ml at 0.48ml/s with 29.1psi vacuum @15:49:46.03 Cartridge 19 sampling 1428/2000ml at 0.47ml/s with 29.4psi vacuum @15:50:26.81 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:50:26.83 Retry #6 of 10 after 10:00 @15:50:45.92 Cartridge 19 sampling 1457/2000ml at 0.50ml/s with 26.0psi vacuum @15:51:46.17 Cartridge 19 sampling 1486/2000ml at 0.49ml/s with 23.3psi vacuum @15:52:45.93 Cartridge 19 sampling 1515/2000ml at 0.48ml/s with 27.6psi vacuum @15:53:46.08 Cartridge 19 sampling 1543/2000ml at 0.49ml/s with 20.6psi vacuum @15:54:45.93 Cartridge 19 sampling 1572/2000ml at 0.48ml/s with 28.5psi vacuum @15:55:46.37 Cartridge 19 sampling 1600/2000ml at 0.47ml/s with 28.3psi vacuum @15:56:46.09 Cartridge 19 sampling 1628/2000ml at 0.45ml/s with 18.0psi vacuum 3.6m deep (20.1psia) @15:57:45.93 Cartridge 19 sampling 1656/2000ml at 0.48ml/s with 22.8psi vacuum @15:58:46.35 Cartridge 19 sampling 1684/2000ml at 0.46ml/s with 16.9psi vacuum 4.8m deep (21.8psia) @15:59:45.93 Cartridge 19 sampling 1712/2000ml at 0.48ml/s with 26.4psi vacuum @16:00:26.97 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:00:27.03 Retry #7 of 10 after 10:00 @16:00:46.16 Cartridge 19 sampling 1739/2000ml at 0.46ml/s with 27.6psi vacuum @16:01:45.93 Cartridge 19 sampling 1766/2000ml at 0.45ml/s with 28.2psi vacuum @16:02:46.17 Cartridge 19 sampling 1794/2000ml at 0.44ml/s with 28.5psi vacuum @16:03:45.93 Cartridge 19 sampling 1820/2000ml at 0.44ml/s with 17.5psi vacuum @16:04:46.10 Cartridge 19 sampling 1847/2000ml at 0.46ml/s with 23.1psi vacuum @16:05:45.93 Cartridge 19 sampling 1874/2000ml at 0.46ml/s with 26.9psi vacuum @16:06:46.11 Cartridge 19 sampling 1901/2000ml at 0.44ml/s with 18.9psi vacuum @16:07:45.93 Cartridge 19 sampling 1927/2000ml at 0.46ml/s with 24.2psi vacuum @16:08:45.99 Cartridge 19 sampling 1953/2000ml at 0.45ml/s with 20.0psi vacuum 2.4m deep (18.2psia) @16:09:45.93 Cartridge 19 sampling 1979/2000ml at 0.45ml/s with 24.0psi vacuum @16:10:27.12 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:10:27.18 Retry #8 of 10 after 10:00 @16:10:33.20 Pressure.coast @16:10:33.57 Intake.close @16:10:33.84 Exhaust.close @16:10:34.27 Toroid.reconfigure :hold @16:10:34.42 Toroid.seek :bypass @16:10:36.05 Toroid.reconfigure :default @16:10:36.26 Toroid.seek :clear @16:10:37.88 Sampled 2000.0ml @16:10:38.68 SamplePump.setPosition! 0ml @16:10:38.73 Plunge.seek :reagent @16:10:38.76 Bypass.open @16:10:39.51 Intake.open @16:10:39.71 Sample loop closed. Ready to Deliver Reagents @16:10:39.73 Archive.seek :sealed @16:10:40.35 SamplePump.reconfigure :slow @16:10:40.41 SamplePump.seek -3ml @16:10:43.43 Archive.seek :evac @16:10:47.09 delay 1 second @16:10:48.18 SamplePump.seek -2.8ml @16:10:48.88 delay 1 second @16:10:49.97 SamplePump.seek -3.3ml @16:10:51.30 delay 1 second @16:10:52.33 Exhaust.open @16:10:53.08 Bypass.close @16:10:53.99 SamplePump.seek -35.3ml @16:10:57.45 delay 1:00 @16:11:57.50 Archive.seek :clear @16:12:02.85 delay 3 seconds @16:12:06.08 Intake.close @16:12:06.38 Exhaust.close @16:12:14.70 RNAlater soak for 10:00 @16:20:27.23 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:20:27.25 Retry #9 of 10 after 10:00 @16:22:14.76 Archive.seek :tVent @16:22:16.47 delay 1 second @16:22:17.50 Archive.seek :evac @16:22:33.47 delay 10 seconds @16:22:43.53 Sample preserved and evacuated @16:22:43.76
Carousel.seek 19 @16:22:43.96 Plunge.seek :locked @16:22:44.94 Archive.seek :clear @16:22:45.08 Air.seek :clear @16:22:45.24 Lysis1.seek :clear @16:23:02.11
Archive.coast @16:23:02.15 Air.coast @16:23:02.19 Lysis1.coast @16:23:02.23 Lysis2.coast @16:23:02.28 Plunge.coast @16:23:02.32 SamplePump.coast @16:23:02.36 Toroid.coast @16:23:02.41 Carousel.coast @16:23:02.60 Gate.power :motor,:OFF @16:23:02.61 Gate.power :core,:OFF @16:23:02.62 Gate.power :analytic1,:OFF @16:23:03.18 Gate.power :analytic2,:OFF @16:23:03.19 Gate.power :analytic3,:OFF @16:23:03.21 Gate.power :main,:OFF @16:23:03.22 Safely stopped and ready to power off @16:23:03.45 delayUntil "03:00:00.00PDT28-Jun-21" @16:30:27.36 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:30:27.48 Retry #10 of 10 after 10:00 @16:40:27.54 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:40:27.56 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 19" @16:40:27.59 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:40:27.61 Retry #1 of 10 after 10:00 @16:50:27.67 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:50:27.70 Retry #2 of 10 after 10:00 @17:00:27.83 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:00:27.85 Retry #3 of 10 after 10:00 @17:10:27.90 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:10:27.93 Retry #4 of 10 after 10:00 @17:20:27.98 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:20:28.01 Retry #5 of 10 after 10:00 @17:30:28.07 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:30:28.09 Retry #6 of 10 after 10:00 @17:40:28.50 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:40:28.54 Retry #7 of 10 after 10:00 @17:50:28.67 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:50:28.75 Retry #8 of 10 after 10:00 @18:00:28.80 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:00:28.82 Retry #9 of 10 after 10:00 @18:10:28.89 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:10:28.99 Retry #10 of 10 after 10:00 @18:20:29.05 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:20:29.07 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 03:00PDT28-Jun-21" @18:20:29.10 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:20:29.12 Retry #1 of 10 after 10:00 @18:30:29.19 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:30:29.25 Retry #2 of 10 after 10:00 @18:40:29.38 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:40:29.46 Retry #3 of 10 after 10:00 @18:50:29.52 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:50:29.54 Retry #4 of 10 after 10:00 @19:00:29.58 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:00:29.60 Retry #5 of 10 after 10:00 @19:10:29.65 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:10:29.68 Retry #6 of 10 after 10:00 @19:20:29.73 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:20:29.78 Retry #7 of 10 after 10:00 @19:30:29.84 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:30:29.86 Retry #8 of 10 after 10:00 @19:40:29.96 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:40:30.09 Retry #9 of 10 after 10:00 @19:50:30.16 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:50:30.18 Retry #10 of 10 after 10:00 @20:00:30.23 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @03:00:00.41PDT28-Jun-21
Gate.power :main,:ON,limit to 5000mA @03:00:00.43 Gate.power :analytic1,:ON,limit to 5000mA @03:00:00.45 Gate.power :analytic2,:ON,limit to 5000mA @03:00:00.46 Gate.power :analytic3,:ON,limit to 5000mA @03:00:00.50 Powering core bus @03:00:02.78 Archive.configure :default @03:00:02.92 Air.configure :default @03:00:03.04 Lysis1.configure :default @03:00:03.15 Lysis2.configure :default @03:00:03.29 Plunge.configure :default @03:00:03.42 SamplePump.configure :default @03:00:03.56 ExtraPump.configure :default @03:00:03.71 Toroid.configure :default @03:00:03.84 Carousel.configure :default @03:00:04.06 Pressure.configure :default @03:00:04.17 All elf controllers running firmware version 4.61 @03:00:04.20 Gate.power :motor,:ON,limit to 5000mA @03:00:04.76 Toroid.reconfigure :default @03:00:04.88 Plunge.reconfigure :default @03:00:04.95 Lysis2.reconfigure :default @03:00:05.25 Lysis1.reconfigure :default @03:00:05.62 Archive.reconfigure :default @03:00:05.71 Air.reconfigure :default @03:00:06.52 Plunge.home.jog 0 @03:00:06.63 Toroid.home.jog 0 @03:00:06.69 Lysis2.home.jog 0 @03:00:06.74 Lysis1.home.jog 0 @03:00:06.78 Archive.home.jog 0 @03:00:06.81 Air.home.jog 0 @03:00:09.65 Lysis2.reconfigure :homing @03:00:09.90 Plunge.reconfigure :homing @03:00:09.94 Air.reconfigure :homing @03:00:10.01 Toroid.reconfigure :homing @03:00:10.16 Lysis1.reconfigure :homing @03:00:10.28 Archive.reconfigure :homing @03:00:10.83 Lysis2.jog 1349 @03:00:10.89 Plunge.jog 843 @03:00:10.93 Toroid.jog 630 @03:00:10.95 Air.jog 1352 @03:00:10.99 Lysis1.jog 1358 @03:00:11.13 Archive.jog 1362 @03:00:11.95 Plunge.setRawPosition! 50 @03:00:11.99 Plunge.seek :home @03:00:12.13 Lysis2.setRawPosition! 50 @03:00:12.18 Lysis2.seek :home @03:00:12.31 Air.setRawPosition! 50 @03:00:12.34 Plunge.reconfigure :default @03:00:12.42 Air.seek :home @03:00:12.57 Lysis1.setRawPosition! 50 @03:00:12.60 Archive.setRawPosition! 50 @03:00:12.62 Lysis2.reconfigure :default @03:00:12.73 Lysis1.seek :home @03:00:12.79 Archive.seek :home @03:00:12.96 Air.reconfigure :default @03:00:13.10 Lysis1.reconfigure :default @03:00:13.22 Archive.reconfigure :default @03:00:13.24 Toroid.setRawPosition! 0 @03:00:13.33 Toroid.seek :home @03:00:13.47 Toroid.reconfigure :default @03:00:13.73
Toroid.seek :clear @03:00:15.42 Carousel.reconfigure :default @03:00:15.63 Carousel.home.jog 0 @03:00:16.53 Carousel.seek 19 @03:00:16.85 Selecting Cartridge 18 @03:00:17.36 Carousel.seek 18 @03:00:18.17 Lysis1.seek :locked @03:00:21.49 Plunge.seek :locked @03:00:21.63 Air.seek :locked @03:00:26.11
Mounted archive Cartridge @03:00:26.29 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 18" @03:00:27.08 Duration of filtering limited to 4:05:00 @03:00:27.80 Priming sample loop w/160ml, bypass w/40ml @03:00:27.87 SamplePump.setPosition! 0ml @03:00:27.91 Bypass.open @03:00:29.68 Intake.open @03:00:30.56 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @03:00:33.18 Vacuum tare is 18% of full scale @03:00:33.20 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @03:00:33.43 Pressure.reconfigure :default @03:00:33.48 Pressure.hold @03:00:33.68 SamplePump.reconfigure :prime @03:00:33.75 SamplePump.seek 40ml,3:51 @03:00:47.46 upsync @03:01:16.94 Pressure.coast @03:01:16.99 SamplePump.setPosition! 0ml @03:01:17.04 Exhaust.open @03:01:18.46 Bypass.close @03:01:20.05 Pressure.reconfigure :default @03:01:20.11 Pressure.hold @03:01:20.38 SamplePump.reconfigure :prime @03:01:20.45 SamplePump.seek 160ml,13:51 @03:04:12.10 Pressure.coast @03:04:12.15 Bypass.open @03:04:12.87 Intake.close @03:04:13.77 Toroid.reconfigure :hold @03:04:13.83 Toroid.seek :crack @03:04:15.57 Toroid.seek :cartridge @03:04:16.58 Toroid.reconfigure :default @03:04:16.69 Toroid.seek :clear @03:04:18.98 Pressurizing cartridge to 26psi above 14.8psia @03:04:19.07 SamplePump.setPosition! 0ml @03:04:20.02 Intake.open @03:04:21.22 Bypass.close @03:04:21.93 Exhaust.close @03:04:22.74 Pressure.reconfigure :default @03:04:22.80 Pressure.hold @03:04:23.01 SamplePump.reconfigure :charge @03:04:23.07 SamplePump.seek 2.3ml,42 seconds @03:04:35.49 Pressure.coast @03:04:35.55 Intake.close @03:04:36.70 Pressurized cartridge with 0.696ml to 26.1psi @03:04:36.71 Isolated @34.4psia (19.6psi) @03:04:36.72 Checking for cartridge leaks from 34.4psia @03:04:36.77 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @03:04:41.88 Cartridge pressure fell 9% to 31.2psia @03:04:41.97 Sampling 2000.0ml @03:04:41.98 SamplePump.setPosition! 0ml @03:04:42.02 Exhaust.open @03:04:42.71 Intake.open @03:04:43.75 Pressure.reconfigure :default @03:04:43.95 Pressure.hold @03:04:44.69 SamplePump.reconfigure :sample @03:04:44.90 SamplePump.seek 2000ml,2:47:12 @03:04:46.90 Cartridge 18 sampling 1/2000ml at 0.63ml/s with 20.0psi vacuum @03:05:46.74 Cartridge 18 sampling 32/2000ml at 0.52ml/s with 23.1psi vacuum @03:06:46.90 Cartridge 18 sampling 63/2000ml at 0.52ml/s with 25.2psi vacuum @03:07:46.75 Cartridge 18 sampling 94/2000ml at 0.53ml/s with 23.5psi vacuum @03:08:46.81 Cartridge 18 sampling 126/2000ml at 0.53ml/s with 23.6psi vacuum @03:09:46.74 Cartridge 18 sampling 158/2000ml at 0.53ml/s with 25.1psi vacuum @03:10:47.06 Cartridge 18 sampling 190/2000ml at 0.54ml/s with 26.8psi vacuum @03:11:46.86 Cartridge 18 sampling 222/2000ml at 0.54ml/s with 23.0psi vacuum @03:12:46.74 Cartridge 18 sampling 255/2000ml at 0.55ml/s with 24.2psi vacuum @03:13:47.14 Cartridge 18 sampling 287/2000ml at 0.55ml/s with 21.0psi vacuum @03:14:46.74 Cartridge 18 sampling 319/2000ml at 0.52ml/s with 19.5psi vacuum @03:15:47.00 Cartridge 18 sampling 351/2000ml at 0.54ml/s with 24.7psi vacuum @03:16:46.74 Cartridge 18 sampling 383/2000ml at 0.53ml/s with 27.1psi vacuum @03:17:46.92 Cartridge 18 sampling 416/2000ml at 0.54ml/s with 21.1psi vacuum @03:18:46.75 Cartridge 18 sampling 448/2000ml at 0.54ml/s with 23.6psi vacuum @03:19:46.91 Cartridge 18 sampling 480/2000ml at 0.51ml/s with 21.8psi vacuum @03:20:46.75 Cartridge 18 sampling 512/2000ml at 0.54ml/s with 25.4psi vacuum @03:21:46.91 Cartridge 18 sampling 544/2000ml at 0.54ml/s with 25.7psi vacuum @03:22:46.79 Cartridge 18 sampling 576/2000ml at 0.51ml/s with 29.0psi vacuum @03:23:46.80 Cartridge 18 sampling 608/2000ml at 0.52ml/s with 18.7psi vacuum 2.9m deep (19.0psia) @03:24:46.74 Cartridge 18 sampling 640/2000ml at 0.52ml/s with 28.9psi vacuum @03:25:46.91 Cartridge 18 sampling 671/2000ml at 0.54ml/s with 22.4psi vacuum @03:26:46.97 Cartridge 18 sampling 703/2000ml at 0.50ml/s with 28.5psi vacuum @03:27:46.74 Cartridge 18 sampling 734/2000ml at 0.52ml/s with 18.7psi vacuum @03:28:47.07 Cartridge 18 sampling 765/2000ml at 0.53ml/s with 25.4psi vacuum @03:29:46.75 Cartridge 18 sampling 797/2000ml at 0.52ml/s with 22.7psi vacuum @03:30:46.96 Cartridge 18 sampling 828/2000ml at 0.50ml/s with 26.8psi vacuum @03:31:46.75 Cartridge 18 sampling 859/2000ml at 0.51ml/s with 21.9psi vacuum @03:32:46.94 Cartridge 18 sampling 890/2000ml at 0.51ml/s with 28.0psi vacuum @03:33:46.75 Cartridge 18 sampling 920/2000ml at 0.49ml/s with 23.2psi vacuum @03:34:46.80 Cartridge 18 sampling 951/2000ml at 0.53ml/s with 24.0psi vacuum @03:35:46.74 Cartridge 18 sampling 982/2000ml at 0.48ml/s with 21.9psi vacuum @03:36:46.80 Cartridge 18 sampling 1012/2000ml at 0.51ml/s with 19.4psi vacuum 3.2m deep (19.4psia) @03:37:46.75 Cartridge 18 sampling 1042/2000ml at 0.51ml/s with 26.3psi vacuum @03:38:46.90 Cartridge 18 sampling 1072/2000ml at 0.51ml/s with 24.2psi vacuum @03:39:46.75 Cartridge 18 sampling 1102/2000ml at 0.50ml/s with 25.6psi vacuum @03:40:46.87 Cartridge 18 sampling 1132/2000ml at 0.47ml/s with 27.5psi vacuum @03:41:46.97 Cartridge 18 sampling 1161/2000ml at 0.50ml/s with 25.2psi vacuum @03:42:46.74 Cartridge 18 sampling 1191/2000ml at 0.49ml/s with 27.6psi vacuum @03:43:46.95 Cartridge 18 sampling 1220/2000ml at 0.48ml/s with 29.6psi vacuum @03:44:46.74 Cartridge 18 sampling 1249/2000ml at 0.49ml/s with 23.3psi vacuum @03:45:46.93 Cartridge 18 sampling 1278/2000ml at 0.48ml/s with 27.6psi vacuum @03:46:46.79 Cartridge 18 sampling 1307/2000ml at 0.46ml/s with 19.2psi vacuum @03:47:46.87 Cartridge 18 sampling 1336/2000ml at 0.47ml/s with 17.4psi vacuum 4.0m deep (20.6psia) @03:48:46.74 Cartridge 18 sampling 1365/2000ml at 0.49ml/s with 23.6psi vacuum @03:49:46.93 Cartridge 18 sampling 1394/2000ml at 0.49ml/s with 26.3psi vacuum @03:50:46.73 Cartridge 18 sampling 1422/2000ml at 0.48ml/s with 25.4psi vacuum @03:51:46.92 Cartridge 18 sampling 1450/2000ml at 0.44ml/s with 28.8psi vacuum @03:52:46.73 Cartridge 18 sampling 1478/2000ml at 0.46ml/s with 27.2psi vacuum @03:53:46.81 Cartridge 18 sampling 1506/2000ml at 0.47ml/s with 20.3psi vacuum @03:54:46.74 Cartridge 18 sampling 1534/2000ml at 0.46ml/s with 28.0psi vacuum @03:55:46.97 Cartridge 18 sampling 1562/2000ml at 0.47ml/s with 19.1psi vacuum 3.5m deep (19.9psia) @03:56:46.98 Cartridge 18 sampling 1590/2000ml at 0.46ml/s with 26.0psi vacuum @03:57:46.75 Cartridge 18 sampling 1617/2000ml at 0.45ml/s with 28.1psi vacuum @03:58:47.01 Cartridge 18 sampling 1644/2000ml at 0.46ml/s with 25.9psi vacuum @03:59:46.75 Cartridge 18 sampling 1671/2000ml at 0.43ml/s with 20.1psi vacuum @04:00:46.81 Cartridge 18 sampling 1698/2000ml at 0.43ml/s with 13.8psi vacuum 4.4m deep (21.2psia) @04:01:46.75 Cartridge 18 sampling 1725/2000ml at 0.42ml/s with 16.9psi vacuum @04:02:46.90 Cartridge 18 sampling 1751/2000ml at 0.45ml/s with 25.5psi vacuum @04:03:46.75 Cartridge 18 sampling 1778/2000ml at 0.45ml/s with 16.7psi vacuum 3.6m deep (20.0psia) @04:04:46.96 Cartridge 18 sampling 1805/2000ml at 0.45ml/s with 19.3psi vacuum 3.9m deep (20.5psia) @04:05:46.75 Cartridge 18 sampling 1831/2000ml at 0.41ml/s with 28.1psi vacuum @04:06:46.83 Cartridge 18 sampling 1858/2000ml at 0.45ml/s with 25.1psi vacuum @04:07:47.30 Cartridge 18 sampling 1884/2000ml at 0.43ml/s with 17.8psi vacuum 5.3m deep (22.6psia) @04:08:47.05 Cartridge 18 sampling 1910/2000ml at 0.41ml/s with 15.6psi vacuum @04:09:47.07 Cartridge 18 sampling 1936/2000ml at 0.42ml/s with 25.0psi vacuum @04:10:46.95 Cartridge 18 sampling 1962/2000ml at 0.42ml/s with 14.0psi vacuum 7.1m deep (25.2psia) @04:11:46.98 Cartridge 18 sampling 1987/2000ml at 0.42ml/s with 29.1psi vacuum @04:12:16.54 Pressure.coast @04:12:16.84 Intake.close @04:12:17.01 Exhaust.close @04:12:17.40 Toroid.reconfigure :hold @04:12:17.52 Toroid.seek :bypass @04:12:19.21 Toroid.reconfigure :default @04:12:19.27 Toroid.seek :clear @04:12:20.87 Sampled 2000.0ml @04:12:21.33 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 18" @04:12:21.40 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:12:21.42 Retry #1 of 10 after 10:00 @04:12:21.78 SamplePump.setPosition! 0ml @04:12:21.81 Plunge.seek :reagent @04:12:21.88 Bypass.open @04:12:22.63 Intake.open @04:12:22.75 Sample loop closed. Ready to Deliver Reagents @04:12:22.77 Archive.seek :sealed @04:12:23.64 SamplePump.reconfigure :slow @04:12:23.79 SamplePump.seek -3ml @04:12:26.50 Archive.seek :evac @04:12:30.50 delay 1 second @04:12:31.60 SamplePump.seek -2.8ml @04:12:32.29 delay 1 second @04:12:33.63 SamplePump.seek -3.3ml @04:12:35.01 delay 1 second @04:12:36.07 Exhaust.open @04:12:36.81 Bypass.close @04:12:37.81 SamplePump.seek -35.3ml @04:12:40.48 delay 1:00 @04:13:40.54 Archive.seek :clear @04:13:46.63 delay 3 seconds @04:13:49.93 Intake.close @04:13:50.14 Exhaust.close @04:13:57.73 RNAlater soak for 10:00 @04:22:21.52 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:22:21.60 Retry #2 of 10 after 10:00 @04:23:57.82 Archive.seek :tVent @04:23:59.47 delay 1 second @04:24:00.50 Archive.seek :evac @04:24:16.44 delay 10 seconds @04:24:26.48 Sample preserved and evacuated @04:24:26.71
Carousel.seek 18 @04:24:26.92 Plunge.seek :locked @04:24:27.90 Archive.seek :clear @04:24:28.13 Air.seek :clear @04:24:28.48 Lysis1.seek :clear @04:24:45.09
Archive.coast @04:24:45.13 Air.coast @04:24:45.17 Lysis1.coast @04:24:45.21 Lysis2.coast @04:24:45.26 Plunge.coast @04:24:45.30 SamplePump.coast @04:24:45.34 Toroid.coast @04:24:45.39 Carousel.coast @04:24:45.58 Gate.power :motor,:OFF @04:24:45.59 Gate.power :core,:OFF @04:24:45.60 Gate.power :analytic1,:OFF @04:24:45.62 Gate.power :analytic2,:OFF @04:24:45.63 Gate.power :analytic3,:OFF @04:24:45.64 Gate.power :main,:OFF @04:24:45.66 Safely stopped and ready to power off @04:24:45.88 delayUntil "15:00:00.00" @04:32:21.64 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:32:21.66 Retry #3 of 10 after 10:00 @04:42:21.71 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:42:21.73 Retry #4 of 10 after 10:00 @04:52:21.78 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @04:52:21.89 Retry #5 of 10 after 10:00 @05:02:21.95 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:02:21.97 Retry #6 of 10 after 10:00 @05:12:22.06 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:12:22.10 Retry #7 of 10 after 10:00 @05:22:22.15 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:22:22.17 Retry #8 of 10 after 10:00 @05:32:22.30 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:32:22.38 Retry #9 of 10 after 10:00 @05:42:22.44 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:42:22.46 Retry #10 of 10 after 10:00 @05:52:22.52 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:52:22.54 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 15:00" @05:52:22.57 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @05:52:22.58 Retry #1 of 10 after 10:00 @06:02:22.67 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:02:22.69 Retry #2 of 10 after 10:00 @06:12:22.75 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:12:22.77 Retry #3 of 10 after 10:00 @06:22:22.90 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:22:22.94 Retry #4 of 10 after 10:00 @06:32:23.05 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:32:23.07 Retry #5 of 10 after 10:00 @06:42:23.64 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:42:23.69 Retry #6 of 10 after 10:00 @06:52:23.85 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:52:23.87 Retry #7 of 10 after 10:00 @07:02:23.92 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @07:02:23.94 Retry #8 of 10 after 10:00 @07:12:23.99 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @07:12:24.05 Retry #9 of 10 after 10:00 @07:22:24.10 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @07:22:24.12 Retry #10 of 10 after 10:00 @07:32:24.19 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @14:08:46.65 Session Begun @14:08:55.76 -> Cartridge.dry @14:11:07.16 -> Cartridge.release 14..9 @14:11:18.68 -> Cartridge.dry @14:11:26.24 -> exit @14:11:26.26 Session Ended @15:00:00.32
Gate.power :main,:ON,limit to 5000mA @15:00:00.34 Gate.power :analytic1,:ON,limit to 5000mA @15:00:00.43 Gate.power :analytic2,:ON,limit to 5000mA @15:00:00.47 Gate.power :analytic3,:ON,limit to 5000mA @15:00:00.52 Powering core bus @15:00:02.94 Archive.configure :default @15:00:03.09 Air.configure :default @15:00:03.20 Lysis1.configure :default @15:00:03.30 Lysis2.configure :default @15:00:03.45 Plunge.configure :default @15:00:03.59 SamplePump.configure :default @15:00:03.72 ExtraPump.configure :default @15:00:03.86 Toroid.configure :default @15:00:04.00 Carousel.configure :default @15:00:04.13 Pressure.configure :default @15:00:04.19 All elf controllers running firmware version 4.61 @15:00:04.21 Gate.power :motor,:ON,limit to 5000mA @15:00:04.35 Archive.reconfigure :default @15:00:04.71 Archive.home.jog 0 @15:00:04.97 Toroid.reconfigure :default @15:00:05.00 Plunge.reconfigure :default @15:00:05.04 Lysis1.reconfigure :default @15:00:05.10 Air.reconfigure :default @15:00:05.16 Lysis2.reconfigure :default @15:00:05.34 Toroid.home.jog 0 @15:00:05.42 Lysis1.home.jog 0 @15:00:05.44 Plunge.home.jog 0 @15:00:05.46 Air.home.jog 0 @15:00:05.48 Lysis2.home.jog 0 @15:00:05.87 Archive.reconfigure :homing @15:00:06.15 Toroid.reconfigure :homing @15:00:06.21 Archive.jog 1370 @15:00:06.25 Plunge.reconfigure :homing @15:00:06.27 Lysis1.reconfigure :homing @15:00:06.44 Toroid.jog 626 @15:00:06.47 Lysis1.jog 1362 @15:00:06.52 Plunge.jog 843 @15:00:06.72 Lysis2.reconfigure :homing @15:00:06.75 Air.reconfigure :homing @15:00:06.84 Lysis2.jog 1348 @15:00:06.94 Air.jog 1359 @15:00:07.93 Plunge.setRawPosition! 50 @15:00:08.04 Lysis1.setRawPosition! 50 @15:00:08.23 Archive.setRawPosition! 50 @15:00:08.48 Plunge.seek :home @15:00:08.59 Lysis1.seek :home @15:00:08.68 Archive.seek :home @15:00:09.38 Plunge.reconfigure :default @15:00:09.53 Lysis2.setRawPosition! 50 @15:00:09.61 Lysis1.reconfigure :default @15:00:09.63 Air.setRawPosition! 50 @15:00:09.68 Archive.reconfigure :default @15:00:09.77 Toroid.setRawPosition! 0 @15:00:10.37 Lysis2.seek :home @15:00:10.43 Air.seek :home @15:00:10.73 Toroid.seek :home @15:00:11.29 Toroid.reconfigure :default @15:00:11.32 Lysis2.reconfigure :default @15:00:11.35 Air.reconfigure :default @15:00:12.02
Toroid.seek :clear @15:00:13.69 Carousel.reconfigure :default @15:00:13.75 Carousel.home.jog 0 @15:00:14.24 Carousel.seek 18 @15:00:14.42 Selecting Cartridge 17 @15:00:15.16 Carousel.seek 17 @15:00:15.68 Lysis1.seek :locked @15:00:19.08 Plunge.seek :locked @15:00:19.39 Air.seek :locked @15:00:23.63
Mounted archive Cartridge @15:00:23.77 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 17" @15:00:23.81 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:00:23.83 Retry #1 of 10 after 10:00 @15:00:24.08 Duration of filtering limited to 4:05:00 @15:00:24.37 Priming sample loop w/160ml, bypass w/40ml @15:00:24.39 SamplePump.setPosition! 0ml @15:00:24.43 Bypass.open @15:00:25.92 Intake.open @15:00:26.98 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:00:29.71 Vacuum tare is 18% of full scale @15:00:29.75 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @15:00:30.22 Pressure.reconfigure :default @15:00:30.34 Pressure.hold @15:00:30.80 SamplePump.reconfigure :prime @15:00:30.94 SamplePump.seek 40ml,3:51 @15:01:14.19 Pressure.coast @15:01:14.31 SamplePump.setPosition! 0ml @15:01:14.40 Exhaust.open @15:01:15.90 Bypass.close @15:01:17.83 Pressure.reconfigure :default @15:01:17.99 Pressure.hold @15:01:18.32 SamplePump.reconfigure :prime @15:01:18.39 SamplePump.seek 160ml,13:51 @15:04:10.11 Pressure.coast @15:04:10.28 Bypass.open @15:04:10.99 Intake.close @15:04:12.25 Toroid.reconfigure :hold @15:04:12.30 Toroid.seek :crack @15:04:13.89 Toroid.seek :cartridge @15:04:14.82 Toroid.reconfigure :default @15:04:14.88 Toroid.seek :clear @15:04:17.09 Pressurizing cartridge to 26psi above 14.4psia @15:04:17.14 SamplePump.setPosition! 0ml @15:04:17.26 Intake.open @15:04:18.06 Bypass.close @15:04:18.96 Exhaust.close @15:04:19.95 Pressure.reconfigure :default @15:04:20.02 Pressure.hold @15:04:20.54 SamplePump.reconfigure :charge @15:04:20.75 SamplePump.seek 2.3ml,42 seconds @15:04:34.43 Pressure.coast @15:04:34.48 Intake.close @15:04:35.35 Pressurized cartridge with 0.857ml to 26.1psi @15:04:35.36 Isolated @36.3psia (21.8psi) @15:04:35.37 Checking for cartridge leaks from 36.3psia @15:04:35.39 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @15:04:43.99 Cartridge pressure fell 12% to 31.8psia @15:04:44.07 Sampling 2000.0ml @15:04:44.08 SamplePump.setPosition! 0ml @15:04:44.12 Exhaust.open @15:04:44.84 Intake.open @15:04:46.14 Pressure.reconfigure :default @15:04:46.25 Pressure.hold @15:04:47.23 SamplePump.reconfigure :sample @15:04:47.35 SamplePump.seek 2000ml,2:47:12 @15:04:49.20 Cartridge 17 sampling 0/2000ml at 0.63ml/s with 17.8psi vacuum @15:05:49.05 Cartridge 17 sampling 31/2000ml at 0.52ml/s with 23.8psi vacuum @15:06:49.16 Cartridge 17 sampling 62/2000ml at 0.50ml/s with 24.8psi vacuum @15:07:49.05 Cartridge 17 sampling 92/2000ml at 0.51ml/s with 24.2psi vacuum @15:08:49.41 Cartridge 17 sampling 122/2000ml at 0.49ml/s with 24.8psi vacuum @15:09:49.50 Cartridge 17 sampling 153/2000ml at 0.52ml/s with 25.3psi vacuum @15:10:23.87 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:10:23.89 Retry #2 of 10 after 10:00 @15:10:49.21 Cartridge 17 sampling 184/2000ml at 0.51ml/s with 25.1psi vacuum @15:11:49.58 Cartridge 17 sampling 216/2000ml at 0.52ml/s with 25.4psi vacuum @15:12:49.04 Cartridge 17 sampling 247/2000ml at 0.51ml/s with 24.1psi vacuum @15:13:49.22 Cartridge 17 sampling 278/2000ml at 0.52ml/s with 24.6psi vacuum @15:14:49.14 Cartridge 17 sampling 309/2000ml at 0.52ml/s with 23.8psi vacuum @15:15:49.20 Cartridge 17 sampling 340/2000ml at 0.51ml/s with 20.1psi vacuum @15:16:49.04 Cartridge 17 sampling 371/2000ml at 0.52ml/s with 24.8psi vacuum @15:17:50.37 Cartridge 17 sampling 402/2000ml at 0.51ml/s with 21.4psi vacuum @15:18:49.04 Cartridge 17 sampling 433/2000ml at 0.53ml/s with 24.2psi vacuum @15:19:49.25 Cartridge 17 sampling 465/2000ml at 0.52ml/s with 20.4psi vacuum @15:20:23.95 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:20:24.04 Retry #3 of 10 after 10:00 @15:20:49.04 Cartridge 17 sampling 496/2000ml at 0.52ml/s with 20.0psi vacuum @15:21:49.09 Cartridge 17 sampling 527/2000ml at 0.51ml/s with 25.7psi vacuum @15:22:49.05 Cartridge 17 sampling 557/2000ml at 0.51ml/s with 26.1psi vacuum @15:23:49.21 Cartridge 17 sampling 588/2000ml at 0.52ml/s with 24.1psi vacuum @15:24:49.29 Cartridge 17 sampling 619/2000ml at 0.51ml/s with 24.7psi vacuum @15:25:49.05 Cartridge 17 sampling 649/2000ml at 0.51ml/s with 24.3psi vacuum @15:26:49.30 Cartridge 17 sampling 680/2000ml at 0.52ml/s with 23.9psi vacuum @15:27:49.05 Cartridge 17 sampling 710/2000ml at 0.51ml/s with 20.4psi vacuum @15:28:49.22 Cartridge 17 sampling 741/2000ml at 0.51ml/s with 25.0psi vacuum @15:29:49.05 Cartridge 17 sampling 771/2000ml at 0.50ml/s with 26.4psi vacuum @15:30:24.16 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:30:24.28 Retry #4 of 10 after 10:00 @15:30:49.14 Cartridge 17 sampling 801/2000ml at 0.50ml/s with 21.9psi vacuum @15:31:49.05 Cartridge 17 sampling 831/2000ml at 0.49ml/s with 22.9psi vacuum @15:32:49.23 Cartridge 17 sampling 861/2000ml at 0.51ml/s with 25.0psi vacuum @15:33:49.05 Cartridge 17 sampling 890/2000ml at 0.48ml/s with 26.4psi vacuum @15:34:49.42 Cartridge 17 sampling 920/2000ml at 0.50ml/s with 21.7psi vacuum @15:35:49.05 Cartridge 17 sampling 949/2000ml at 0.48ml/s with 26.5psi vacuum @15:36:49.16 Cartridge 17 sampling 979/2000ml at 0.48ml/s with 23.7psi vacuum @15:37:49.05 Cartridge 17 sampling 1008/2000ml at 0.48ml/s with 26.0psi vacuum @15:38:49.21 Cartridge 17 sampling 1037/2000ml at 0.48ml/s with 24.5psi vacuum @15:39:49.05 Cartridge 17 sampling 1067/2000ml at 0.49ml/s with 25.2psi vacuum @15:40:24.33 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:40:24.35 Retry #5 of 10 after 10:00 @15:40:49.06 Cartridge 17 sampling 1096/2000ml at 0.49ml/s with 24.5psi vacuum @15:41:49.34 Cartridge 17 sampling 1125/2000ml at 0.48ml/s with 19.1psi vacuum 1.8m deep (17.3psia) @15:42:49.07 Cartridge 17 sampling 1154/2000ml at 0.48ml/s with 24.0psi vacuum @15:43:49.21 Cartridge 17 sampling 1182/2000ml at 0.48ml/s with 24.5psi vacuum @15:44:49.05 Cartridge 17 sampling 1211/2000ml at 0.48ml/s with 25.1psi vacuum @15:45:49.22 Cartridge 17 sampling 1239/2000ml at 0.46ml/s with 20.1psi vacuum @15:46:49.05 Cartridge 17 sampling 1267/2000ml at 0.47ml/s with 20.9psi vacuum @15:47:49.30 Cartridge 17 sampling 1295/2000ml at 0.46ml/s with 24.8psi vacuum @15:48:49.05 Cartridge 17 sampling 1323/2000ml at 0.46ml/s with 24.0psi vacuum @15:49:49.14 Cartridge 17 sampling 1351/2000ml at 0.46ml/s with 25.4psi vacuum @15:50:24.42 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:50:24.48 Retry #6 of 10 after 10:00 @15:50:49.05 Cartridge 17 sampling 1378/2000ml at 0.46ml/s with 24.0psi vacuum @15:51:49.08 Cartridge 17 sampling 1406/2000ml at 0.45ml/s with 26.1psi vacuum @15:52:49.05 Cartridge 17 sampling 1433/2000ml at 0.45ml/s with 25.3psi vacuum @15:53:49.26 Cartridge 17 sampling 1461/2000ml at 0.45ml/s with 26.6psi vacuum @15:54:49.29 Cartridge 17 sampling 1488/2000ml at 0.45ml/s with 26.1psi vacuum @15:55:49.05 Cartridge 17 sampling 1515/2000ml at 0.45ml/s with 24.5psi vacuum @15:56:49.38 Cartridge 17 sampling 1542/2000ml at 0.43ml/s with 26.3psi vacuum @15:57:49.05 Cartridge 17 sampling 1569/2000ml at 0.45ml/s with 24.5psi vacuum @15:58:49.21 Cartridge 17 sampling 1596/2000ml at 0.44ml/s with 23.6psi vacuum @15:59:49.04 Cartridge 17 sampling 1623/2000ml at 0.44ml/s with 26.1psi vacuum @16:00:24.54 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:00:24.56 Retry #7 of 10 after 10:00 @16:00:49.23 Cartridge 17 sampling 1649/2000ml at 0.44ml/s with 25.5psi vacuum @16:01:49.05 Cartridge 17 sampling 1676/2000ml at 0.44ml/s with 24.6psi vacuum @16:02:49.19 Cartridge 17 sampling 1702/2000ml at 0.44ml/s with 26.2psi vacuum @16:03:49.05 Cartridge 17 sampling 1728/2000ml at 0.44ml/s with 25.0psi vacuum @16:04:49.27 Cartridge 17 sampling 1754/2000ml at 0.43ml/s with 25.6psi vacuum @16:05:49.05 Cartridge 17 sampling 1780/2000ml at 0.44ml/s with 23.2psi vacuum @16:06:49.21 Cartridge 17 sampling 1806/2000ml at 0.43ml/s with 22.5psi vacuum @16:07:49.05 Cartridge 17 sampling 1832/2000ml at 0.41ml/s with 24.1psi vacuum @16:08:49.42 Cartridge 17 sampling 1857/2000ml at 0.41ml/s with 20.1psi vacuum @16:09:49.26 Cartridge 17 sampling 1882/2000ml at 0.42ml/s with 23.9psi vacuum @16:10:24.61 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:10:24.63 Retry #8 of 10 after 10:00 @16:10:49.03 Cartridge 17 sampling 1907/2000ml at 0.42ml/s with 26.4psi vacuum @16:11:49.33 Cartridge 17 sampling 1933/2000ml at 0.42ml/s with 23.1psi vacuum @16:12:49.04 Cartridge 17 sampling 1957/2000ml at 0.41ml/s with 20.3psi vacuum @16:13:49.24 Cartridge 17 sampling 1982/2000ml at 0.41ml/s with 26.8psi vacuum @16:14:32.35 Pressure.coast @16:14:32.49 Intake.close @16:14:32.55 Exhaust.close @16:14:32.77 Toroid.reconfigure :hold @16:14:32.83 Toroid.seek :bypass @16:14:34.50 Toroid.reconfigure :default @16:14:34.67 Toroid.seek :clear @16:14:36.36 Sampled 2000.0ml @16:14:38.16 SamplePump.setPosition! 0ml @16:14:38.27 Plunge.seek :reagent @16:14:38.38 Bypass.open @16:14:39.18 Intake.open @16:14:39.39 Sample loop closed. Ready to Deliver Reagents @16:14:39.45 Archive.seek :sealed @16:14:40.10 SamplePump.reconfigure :slow @16:14:40.16 SamplePump.seek -3ml @16:14:43.12 Archive.seek :evac @16:14:46.88 delay 1 second @16:14:47.98 SamplePump.seek -2.8ml @16:14:49.74 delay 1 second @16:14:50.83 SamplePump.seek -3.3ml @16:14:52.16 delay 1 second @16:14:53.19 Exhaust.open @16:14:54.07 Bypass.close @16:14:54.99 SamplePump.seek -35.3ml @16:14:57.09 delay 1:00 @16:15:57.14 Archive.seek :clear @16:16:03.76 delay 3 seconds @16:16:07.24 Intake.close @16:16:07.54 Exhaust.close @16:16:14.29 RNAlater soak for 10:00 @16:20:24.84 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:20:24.89 Retry #9 of 10 after 10:00 @16:26:14.35 Archive.seek :tVent @16:26:16.10 delay 1 second @16:26:17.13 Archive.seek :evac @16:26:33.12 delay 10 seconds @16:26:43.17 Sample preserved and evacuated @16:26:43.77
Carousel.seek 17 @16:26:44.45 Plunge.seek :locked @16:26:45.87 Archive.seek :clear @16:26:46.15 Air.seek :clear @16:26:46.30 Lysis1.seek :clear @16:27:03.15
Archive.coast @16:27:03.19 Air.coast @16:27:03.32 Lysis1.coast @16:27:03.43 Lysis2.coast @16:27:03.47 Plunge.coast @16:27:03.52 SamplePump.coast @16:27:03.64 Toroid.coast @16:27:03.83 Carousel.coast @16:27:04.34 Gate.power :motor,:OFF @16:27:04.35 Gate.power :core,:OFF @16:27:04.37 Gate.power :analytic1,:OFF @16:27:04.38 Gate.power :analytic2,:OFF @16:27:04.39 Gate.power :analytic3,:OFF @16:27:04.40 Gate.power :main,:OFF @16:27:04.42 Safely stopped and ready to power off @16:27:04.64 delayUntil "03:00:00.00PDT29-Jun-21" @16:30:24.99 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:30:25.02 Retry #10 of 10 after 10:00 @16:40:25.06 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:40:25.08 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 17" @16:40:25.12 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:40:25.14 Retry #1 of 10 after 10:00 @16:50:37.87 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 03:00PDT29-Jun-21" @16:50:49.23 upsync @16:51:01.66 upsync @03:00:00.11PDT29-Jun-21
Gate.power :main,:ON,limit to 5000mA @03:00:00.13 Gate.power :analytic1,:ON,limit to 5000mA @03:00:00.14 Gate.power :analytic2,:ON,limit to 5000mA @03:00:00.15 Gate.power :analytic3,:ON,limit to 5000mA @03:00:00.19 Powering core bus @03:00:02.47 Archive.configure :default @03:00:02.60 Air.configure :default @03:00:02.72 Lysis1.configure :default @03:00:02.83 Lysis2.configure :default @03:00:02.96 Plunge.configure :default @03:00:03.10 SamplePump.configure :default @03:00:03.24 ExtraPump.configure :default @03:00:03.38 Toroid.configure :default @03:00:03.51 Carousel.configure :default @03:00:03.66 Pressure.configure :default @03:00:03.71 All elf controllers running firmware version 4.61 @03:00:03.73 Gate.power :motor,:ON,limit to 5000mA @03:00:03.91 Archive.reconfigure :default @03:00:05.17 Archive.home.jog 0 @03:00:05.29 Toroid.reconfigure :default @03:00:05.37 Lysis2.reconfigure :default @03:00:05.54 Lysis1.reconfigure :default @03:00:05.62 Air.reconfigure :default @03:00:05.71 Plunge.reconfigure :default @03:00:06.23 Toroid.home.jog 0 @03:00:06.31 Lysis2.home.jog 0 @03:00:06.43 Lysis1.home.jog 0 @03:00:06.52 Plunge.home.jog 0 @03:00:06.65 Air.home.jog 0 @03:00:08.11 Toroid.reconfigure :homing @03:00:08.26 Plunge.reconfigure :homing @03:00:08.35 Lysis1.reconfigure :homing @03:00:08.50 Archive.reconfigure :homing @03:00:08.57 Lysis2.reconfigure :homing @03:00:08.63 Air.reconfigure :homing @03:00:09.00 Plunge.jog 837 @03:00:09.09 Toroid.jog 621 @03:00:09.26 Archive.jog 1367 @03:00:09.30 Lysis1.jog 1356 @03:00:09.35 Lysis2.jog 1346 @03:00:09.39 Air.jog 1354 @03:00:10.23 Plunge.setRawPosition! 50 @03:00:10.36 Plunge.seek :home @03:00:10.79 Plunge.reconfigure :default @03:00:10.89 Lysis2.setRawPosition! 50 @03:00:10.93 Lysis1.setRawPosition! 50 @03:00:10.95 Air.setRawPosition! 50 @03:00:10.97 Archive.setRawPosition! 50 @03:00:11.07 Lysis2.seek :home @03:00:11.13 Lysis1.seek :home @03:00:11.22 Archive.seek :home @03:00:11.27 Air.seek :home @03:00:11.56 Lysis2.reconfigure :default @03:00:11.71 Lysis1.reconfigure :default @03:00:11.75 Air.reconfigure :default @03:00:11.79 Toroid.setRawPosition! 0 @03:00:11.83 Archive.reconfigure :default @03:00:12.02 Toroid.seek :home @03:00:12.14 Toroid.reconfigure :default @03:00:12.37
Toroid.seek :clear @03:00:13.99 Carousel.reconfigure :default @03:00:14.04 Carousel.home.jog 0 @03:00:14.53 Carousel.seek 17 @03:00:14.72 Selecting Cartridge 16 @03:00:14.94 Carousel.seek 16 @03:00:15.76 Lysis1.seek :locked @03:00:19.18 Plunge.seek :locked @03:00:19.47 Air.seek :locked @03:00:23.70
Mounted archive Cartridge @03:00:23.81 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 16" @03:00:24.12 Duration of filtering limited to 4:05:00 @03:00:24.41 Priming sample loop w/160ml, bypass w/40ml @03:00:24.43 SamplePump.setPosition! 0ml @03:00:24.46 Bypass.open @03:00:26.20 Intake.open @03:00:27.35 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @03:00:30.36 Vacuum tare is 18% of full scale @03:00:30.38 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @03:00:30.62 Pressure.reconfigure :default @03:00:30.68 Pressure.hold @03:00:30.89 SamplePump.reconfigure :prime @03:00:30.96 SamplePump.seek 40ml,3:51 @03:00:46.07 upsync @03:01:14.24 Pressure.coast @03:01:14.35 SamplePump.setPosition! 0ml @03:01:14.44 Exhaust.open @03:01:16.08 Bypass.close @03:01:17.75 Pressure.reconfigure :default @03:01:17.81 Pressure.hold @03:01:18.08 SamplePump.reconfigure :prime @03:01:18.14 SamplePump.seek 160ml,13:51 @03:04:09.74 Pressure.coast @03:04:09.80 Bypass.open @03:04:10.52 Intake.close @03:04:11.50 Toroid.reconfigure :hold @03:04:11.55 Toroid.seek :crack @03:04:13.13 Toroid.seek :cartridge @03:04:14.06 Toroid.reconfigure :default @03:04:14.12 Toroid.seek :clear @03:04:16.42 Pressurizing cartridge to 26psi above 14.7psia @03:04:16.44 SamplePump.setPosition! 0ml @03:04:16.48 Intake.open @03:04:17.37 Bypass.close @03:04:18.07 Exhaust.close @03:04:19.09 Pressure.reconfigure :default @03:04:19.16 Pressure.hold @03:04:19.67 SamplePump.reconfigure :charge @03:04:19.73 SamplePump.seek 2.3ml,42 seconds @03:04:32.14 Pressure.coast @03:04:32.19 Intake.close @03:04:33.08 Pressurized cartridge with 0.698ml to 26.0psi @03:04:33.09 Isolated @36.0psia (21.3psi) @03:04:33.10 Checking for cartridge leaks from 36.0psia @03:04:33.12 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @03:04:40.21 Cartridge pressure fell 13% to 31.5psia @03:04:40.29 Sampling 2000.0ml @03:04:40.31 SamplePump.setPosition! 0ml @03:04:40.35 Exhaust.open @03:04:41.06 Intake.open @03:04:41.99 Pressure.reconfigure :default @03:04:42.05 Pressure.hold @03:04:42.35 SamplePump.reconfigure :sample @03:04:42.42 SamplePump.seek 2000ml,2:47:12 @03:04:45.14 Cartridge 16 sampling 1/2000ml at 0.54ml/s with 25.3psi vacuum @03:05:45.06 Cartridge 16 sampling 32/2000ml at 0.52ml/s with 23.2psi vacuum @03:06:45.24 Cartridge 16 sampling 63/2000ml at 0.52ml/s with 24.4psi vacuum @03:07:45.06 Cartridge 16 sampling 93/2000ml at 0.50ml/s with 23.2psi vacuum @03:08:45.32 Cartridge 16 sampling 124/2000ml at 0.51ml/s with 22.5psi vacuum @03:09:45.06 Cartridge 16 sampling 154/2000ml at 0.50ml/s with 23.6psi vacuum @03:10:45.22 Cartridge 16 sampling 185/2000ml at 0.50ml/s with 25.0psi vacuum @03:11:45.05 Cartridge 16 sampling 216/2000ml at 0.52ml/s with 25.0psi vacuum @03:12:45.13 Cartridge 16 sampling 247/2000ml at 0.53ml/s with 24.6psi vacuum @03:13:45.05 Cartridge 16 sampling 278/2000ml at 0.52ml/s with 22.8psi vacuum @03:14:45.32 Cartridge 16 sampling 310/2000ml at 0.52ml/s with 22.9psi vacuum @03:15:45.32 Cartridge 16 sampling 341/2000ml at 0.52ml/s with 21.0psi vacuum @03:16:45.05 Cartridge 16 sampling 373/2000ml at 0.52ml/s with 24.2psi vacuum @03:17:45.37 Cartridge 16 sampling 404/2000ml at 0.52ml/s with 25.7psi vacuum @03:18:45.06 Cartridge 16 sampling 435/2000ml at 0.53ml/s with 25.0psi vacuum @03:19:45.21 Cartridge 16 sampling 467/2000ml at 0.51ml/s with 25.7psi vacuum @03:20:45.07 Cartridge 16 sampling 498/2000ml at 0.51ml/s with 24.9psi vacuum @03:21:45.19 Cartridge 16 sampling 529/2000ml at 0.52ml/s with 23.4psi vacuum @03:22:45.06 Cartridge 16 sampling 560/2000ml at 0.52ml/s with 21.8psi vacuum @03:23:45.22 Cartridge 16 sampling 591/2000ml at 0.52ml/s with 24.8psi vacuum @03:24:45.06 Cartridge 16 sampling 621/2000ml at 0.51ml/s with 23.3psi vacuum @03:25:45.21 Cartridge 16 sampling 652/2000ml at 0.50ml/s with 23.4psi vacuum @03:26:45.06 Cartridge 16 sampling 682/2000ml at 0.50ml/s with 24.7psi vacuum @03:27:45.22 Cartridge 16 sampling 712/2000ml at 0.51ml/s with 24.1psi vacuum @03:28:45.06 Cartridge 16 sampling 743/2000ml at 0.50ml/s with 25.0psi vacuum @03:29:45.05 Cartridge 16 sampling 773/2000ml at 0.50ml/s with 25.7psi vacuum @03:30:45.39 Cartridge 16 sampling 803/2000ml at 0.50ml/s with 24.5psi vacuum @03:31:45.05 Cartridge 16 sampling 833/2000ml at 0.50ml/s with 24.2psi vacuum @03:32:45.25 Cartridge 16 sampling 863/2000ml at 0.49ml/s with 23.6psi vacuum @03:33:45.05 Cartridge 16 sampling 893/2000ml at 0.49ml/s with 21.8psi vacuum @03:34:46.57 Cartridge 16 sampling 923/2000ml at 0.49ml/s with 26.2psi vacuum @03:35:45.05 Cartridge 16 sampling 952/2000ml at 0.49ml/s with 21.2psi vacuum @03:36:45.22 Cartridge 16 sampling 981/2000ml at 0.48ml/s with 24.6psi vacuum @03:37:45.05 Cartridge 16 sampling 1011/2000ml at 0.48ml/s with 23.3psi vacuum @03:38:45.20 Cartridge 16 sampling 1040/2000ml at 0.48ml/s with 22.4psi vacuum @03:39:45.05 Cartridge 16 sampling 1069/2000ml at 0.47ml/s with 20.5psi vacuum @03:40:45.22 Cartridge 16 sampling 1098/2000ml at 0.49ml/s with 21.5psi vacuum @03:41:45.05 Cartridge 16 sampling 1126/2000ml at 0.48ml/s with 21.0psi vacuum @03:42:45.26 Cartridge 16 sampling 1155/2000ml at 0.48ml/s with 23.7psi vacuum @03:43:45.23 Cartridge 16 sampling 1183/2000ml at 0.48ml/s with 24.4psi vacuum @03:44:45.06 Cartridge 16 sampling 1212/2000ml at 0.47ml/s with 23.6psi vacuum @03:45:45.35 Cartridge 16 sampling 1240/2000ml at 0.47ml/s with 24.4psi vacuum @03:46:45.06 Cartridge 16 sampling 1269/2000ml at 0.47ml/s with 21.8psi vacuum @03:47:45.21 Cartridge 16 sampling 1297/2000ml at 0.47ml/s with 26.6psi vacuum @03:48:45.06 Cartridge 16 sampling 1325/2000ml at 0.47ml/s with 21.8psi vacuum @03:49:45.19 Cartridge 16 sampling 1353/2000ml at 0.46ml/s with 24.5psi vacuum @03:50:45.06 Cartridge 16 sampling 1381/2000ml at 0.45ml/s with 26.0psi vacuum @03:51:45.16 Cartridge 16 sampling 1408/2000ml at 0.46ml/s with 26.8psi vacuum @03:52:45.06 Cartridge 16 sampling 1436/2000ml at 0.46ml/s with 24.2psi vacuum @03:53:45.22 Cartridge 16 sampling 1464/2000ml at 0.46ml/s with 26.3psi vacuum @03:54:45.06 Cartridge 16 sampling 1491/2000ml at 0.46ml/s with 24.2psi vacuum @03:55:45.21 Cartridge 16 sampling 1518/2000ml at 0.45ml/s with 22.2psi vacuum @03:56:45.06 Cartridge 16 sampling 1545/2000ml at 0.45ml/s with 23.9psi vacuum @03:57:45.30 Cartridge 16 sampling 1572/2000ml at 0.44ml/s with 26.2psi vacuum @03:58:45.35 Cartridge 16 sampling 1599/2000ml at 0.44ml/s with 22.3psi vacuum @03:59:45.06 Cartridge 16 sampling 1625/2000ml at 0.44ml/s with 22.2psi vacuum @04:00:45.35 Cartridge 16 sampling 1652/2000ml at 0.45ml/s with 20.8psi vacuum @04:01:45.06 Cartridge 16 sampling 1678/2000ml at 0.44ml/s with 27.0psi vacuum @04:02:45.22 Cartridge 16 sampling 1705/2000ml at 0.44ml/s with 22.1psi vacuum @04:03:45.05 Cartridge 16 sampling 1731/2000ml at 0.44ml/s with 25.6psi vacuum @04:04:45.18 Cartridge 16 sampling 1757/2000ml at 0.43ml/s with 26.7psi vacuum @04:05:45.06 Cartridge 16 sampling 1784/2000ml at 0.42ml/s with 20.7psi vacuum @04:06:45.30 Cartridge 16 sampling 1810/2000ml at 0.42ml/s with 24.8psi vacuum @04:07:45.06 Cartridge 16 sampling 1836/2000ml at 0.42ml/s with 22.2psi vacuum @04:08:45.16 Cartridge 16 sampling 1861/2000ml at 0.43ml/s with 25.6psi vacuum @04:09:45.06 Cartridge 16 sampling 1887/2000ml at 0.43ml/s with 20.8psi vacuum @04:10:45.36 Cartridge 16 sampling 1912/2000ml at 0.42ml/s with 22.8psi vacuum @04:11:45.26 Cartridge 16 sampling 1938/2000ml at 0.41ml/s with 25.5psi vacuum @04:12:45.10 Cartridge 16 sampling 1963/2000ml at 0.42ml/s with 20.1psi vacuum @04:13:45.32 Cartridge 16 sampling 1988/2000ml at 0.42ml/s with 25.0psi vacuum @04:14:14.05 Pressure.coast @04:14:14.35 Intake.close @04:14:14.49 Exhaust.close @04:14:14.92 Toroid.reconfigure :hold @04:14:15.07 Toroid.seek :bypass @04:14:16.75 Toroid.reconfigure :default @04:14:16.94 Toroid.seek :clear @04:14:18.59 Sampled 2000.0ml @04:14:19.04 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 16" @04:14:19.47 SamplePump.setPosition! 0ml @04:14:19.50 Plunge.seek :reagent @04:14:19.59 Bypass.open @04:14:20.32 Intake.open @04:14:20.46 Sample loop closed. Ready to Deliver Reagents @04:14:20.48 Archive.seek :sealed @04:14:21.19 SamplePump.reconfigure :slow @04:14:21.26 SamplePump.seek -3ml @04:14:24.15 Archive.seek :evac @04:14:27.94 delay 1 second @04:14:29.03 SamplePump.seek -2.8ml @04:14:29.73 delay 1 second @04:14:30.81 SamplePump.seek -3.3ml @04:14:32.16 delay 1 second @04:14:33.19 Exhaust.open @04:14:33.94 Bypass.close @04:14:34.83 SamplePump.seek -35.3ml @04:14:35.18 upsync @04:14:38.12 delay 1:00 @04:15:38.20 Archive.seek :clear @04:15:43.66 delay 3 seconds @04:15:46.89 Intake.close @04:15:47.07 Exhaust.close @04:15:55.46 RNAlater soak for 10:00 @04:25:55.50 Archive.seek :tVent @04:25:57.73 delay 1 second @04:25:58.76 Archive.seek :evac @04:26:14.72 delay 10 seconds @04:26:24.76 Sample preserved and evacuated @04:26:24.99
Carousel.seek 16 @04:26:25.20 Plunge.seek :locked @04:26:26.18 Archive.seek :clear @04:26:26.34 Air.seek :clear @04:26:26.51 Lysis1.seek :clear @04:26:43.35
Archive.coast @04:26:43.39 Air.coast @04:26:43.44 Lysis1.coast @04:26:43.48 Lysis2.coast @04:26:43.52 Plunge.coast @04:26:43.56 SamplePump.coast @04:26:43.60 Toroid.coast @04:26:43.65 Carousel.coast @04:26:43.84 Gate.power :motor,:OFF @04:26:43.85 Gate.power :core,:OFF @04:26:43.87 Gate.power :analytic1,:OFF @04:26:43.88 Gate.power :analytic2,:OFF @04:26:43.89 Gate.power :analytic3,:OFF @04:26:43.90 Gate.power :main,:OFF @04:26:43.92 Safely stopped and ready to power off @04:26:44.29 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 15:00" @04:26:44.32
delayUntil "15:00:00.00" @04:26:57.78 upsync @11:37:30.09 Session Begun @11:37:32.40 -> can @11:37:44.28 -> quit @11:37:44.30 Session Ended @15:00:00.32
Gate.power :main,:ON,limit to 5000mA @15:00:00.34 Gate.power :analytic1,:ON,limit to 5000mA @15:00:00.35 Gate.power :analytic2,:ON,limit to 5000mA @15:00:00.40 Gate.power :analytic3,:ON,limit to 5000mA @15:00:00.47 Powering core bus @15:00:03.04 Archive.configure :default @15:00:03.30 Air.configure :default @15:00:03.59 Lysis1.configure :default @15:00:03.93 Lysis2.configure :default @15:00:04.10 Plunge.configure :default @15:00:04.25 SamplePump.configure :default @15:00:04.38 ExtraPump.configure :default @15:00:04.53 Toroid.configure :default @15:00:04.66 Carousel.configure :default @15:00:04.80 Pressure.configure :default @15:00:04.86 All elf controllers running firmware version 4.61 @15:00:04.88 Gate.power :motor,:ON,limit to 5000mA @15:00:05.10 Archive.reconfigure :default @15:00:05.40 Archive.home.jog 0 @15:00:05.60 Plunge.reconfigure :default @15:00:05.68 Lysis1.reconfigure :default @15:00:05.73 Air.reconfigure :default @15:00:05.76 Lysis2.reconfigure :default @15:00:05.84 Toroid.reconfigure :default @15:00:05.99 Plunge.home.jog 0 @15:00:06.62 Air.home.jog 0 @15:00:06.66 Lysis1.home.jog 0 @15:00:06.70 Lysis2.home.jog 0 @15:00:06.76 Toroid.home.jog 0 @15:00:07.26 Archive.reconfigure :homing @15:00:07.47 Archive.jog 1370 @15:00:07.53 Toroid.reconfigure :homing @15:00:07.56 Lysis2.reconfigure :homing @15:00:07.60 Air.reconfigure :homing @15:00:07.64 Lysis1.reconfigure :homing @15:00:07.69 Plunge.reconfigure :homing @15:00:07.92 Toroid.jog 626 @15:00:07.96 Lysis2.jog 1350 @15:00:08.02 Air.jog 1358 @15:00:08.06 Plunge.jog 842 @15:00:08.10 Lysis1.jog 1358 @15:00:08.99 Archive.setRawPosition! 50 @15:00:09.11 Archive.seek :home @15:00:09.48 Plunge.setRawPosition! 50 @15:00:09.59 Lysis2.setRawPosition! 50 @15:00:09.73 Plunge.seek :home @15:00:09.85 Lysis2.seek :home @15:00:10.31 Toroid.setRawPosition! 0 @15:00:10.48 Plunge.reconfigure :default @15:00:10.54 Lysis2.reconfigure :default @15:00:10.58 Lysis1.setRawPosition! 50 @15:00:10.62 Archive.reconfigure :default @15:00:10.69 Air.setRawPosition! 50 @15:00:11.21 Toroid.seek :home @15:00:11.26 Lysis1.seek :home @15:00:11.33 Air.seek :home @15:00:12.00 Toroid.reconfigure :default @15:00:12.04 Lysis1.reconfigure :default @15:00:12.41 Air.reconfigure :default @15:00:12.98
Toroid.seek :clear @15:00:14.63 Carousel.reconfigure :default @15:00:14.69 Carousel.home.jog 0 @15:00:15.18 Carousel.seek 16 @15:00:15.36 Selecting Cartridge 15 @15:00:15.58 Carousel.seek 15 @15:00:16.09 Lysis1.seek :locked @15:00:19.67 Plunge.seek :locked @15:00:20.02 Air.seek :locked @15:00:24.26
Mounted archive Cartridge @15:00:24.38 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 15" @15:00:24.45 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:00:24.46 Retry #1 of 10 after 10:00 @15:00:24.72 Duration of filtering limited to 4:05:00 @15:00:25.02 Priming sample loop w/160ml, bypass w/40ml @15:00:25.04 SamplePump.setPosition! 0ml @15:00:25.08 Bypass.open @15:00:26.58 Intake.open @15:00:27.47 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:00:30.08 Vacuum tare is 18% of full scale @15:00:30.13 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @15:00:30.66 Pressure.reconfigure :default @15:00:30.84 Pressure.hold @15:00:31.35 SamplePump.reconfigure :prime @15:00:31.47 SamplePump.seek 40ml,3:51 @15:01:14.86 Pressure.coast @15:01:15.02 SamplePump.setPosition! 0ml @15:01:15.29 Exhaust.open @15:01:16.78 Bypass.close @15:01:18.65 Pressure.reconfigure :default @15:01:18.71 Pressure.hold @15:01:19.24 SamplePump.reconfigure :prime @15:01:19.40 SamplePump.seek 160ml,13:51 @15:04:11.20 Pressure.coast @15:04:11.32 Bypass.open @15:04:12.07 Intake.close @15:04:13.03 Toroid.reconfigure :hold @15:04:13.18 Toroid.seek :crack @15:04:14.82 Toroid.seek :cartridge @15:04:15.75 Toroid.reconfigure :default @15:04:15.81 Toroid.seek :clear @15:04:17.65 Pressurizing cartridge to 26psi above 14.7psia @15:04:17.67 SamplePump.setPosition! 0ml @15:04:17.70 Intake.open @15:04:18.55 Bypass.close @15:04:19.32 Exhaust.close @15:04:20.41 Pressure.reconfigure :default @15:04:20.53 Pressure.hold @15:04:21.16 SamplePump.reconfigure :charge @15:04:21.25 SamplePump.seek 2.3ml,42 seconds @15:04:33.23 Pressure.coast @15:04:33.81 Intake.close @15:04:34.71 Pressurized cartridge with 0.651ml to 26.1psi @15:04:34.72 Isolated @34.3psia (19.6psi) @15:04:34.73 Checking for cartridge leaks from 34.3psia @15:04:34.75 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @15:04:42.86 Cartridge pressure fell 14% to 29.5psia @15:04:42.94 Sampling 2000.0ml @15:04:42.96 SamplePump.setPosition! 0ml @15:04:43.01 Exhaust.open @15:04:43.71 Intake.open @15:04:44.68 Pressure.reconfigure :default @15:04:44.74 Pressure.hold @15:04:44.87 Sea 2.4m deep (18.3psia) @15:04:45.02 SamplePump.reconfigure :sample @15:04:45.09 SamplePump.seek 2000ml,2:47:12 @15:04:48.00 Cartridge 15 sampling 1/2000ml at 0.46ml/s with 26.4psi vacuum @15:05:47.75 Cartridge 15 sampling 30/2000ml at 0.49ml/s with 25.2psi vacuum @15:06:47.87 Cartridge 15 sampling 60/2000ml at 0.49ml/s with 24.5psi vacuum @15:07:47.75 Cartridge 15 sampling 89/2000ml at 0.49ml/s with 24.3psi vacuum @15:08:47.91 Cartridge 15 sampling 119/2000ml at 0.49ml/s with 21.5psi vacuum @15:09:47.75 Cartridge 15 sampling 148/2000ml at 0.49ml/s with 22.9psi vacuum @15:10:24.51 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:10:24.53 Retry #2 of 10 after 10:00 @15:10:48.06 Cartridge 15 sampling 178/2000ml at 0.50ml/s with 23.5psi vacuum @15:11:47.74 Cartridge 15 sampling 208/2000ml at 0.49ml/s with 24.3psi vacuum @15:12:47.87 Cartridge 15 sampling 238/2000ml at 0.50ml/s with 23.5psi vacuum @15:13:47.75 Cartridge 15 sampling 268/2000ml at 0.51ml/s with 20.7psi vacuum @15:14:47.87 Cartridge 15 sampling 299/2000ml at 0.49ml/s with 25.1psi vacuum @15:15:47.75 Cartridge 15 sampling 329/2000ml at 0.50ml/s with 19.5psi vacuum 1.9m deep (17.5psia) @15:16:47.76 Cartridge 15 sampling 359/2000ml at 0.50ml/s with 26.4psi vacuum @15:17:48.20 Cartridge 15 sampling 389/2000ml at 0.50ml/s with 24.6psi vacuum @15:18:47.75 Cartridge 15 sampling 419/2000ml at 0.50ml/s with 24.6psi vacuum @15:19:47.92 Cartridge 15 sampling 449/2000ml at 0.51ml/s with 21.0psi vacuum @15:20:24.64 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:20:24.76 Retry #3 of 10 after 10:00 @15:20:47.74 Cartridge 15 sampling 479/2000ml at 0.50ml/s with 23.5psi vacuum @15:21:48.04 Cartridge 15 sampling 509/2000ml at 0.50ml/s with 23.8psi vacuum @15:22:47.75 Cartridge 15 sampling 539/2000ml at 0.50ml/s with 24.6psi vacuum @15:23:47.99 Cartridge 15 sampling 569/2000ml at 0.50ml/s with 26.2psi vacuum @15:24:47.75 Cartridge 15 sampling 598/2000ml at 0.49ml/s with 25.7psi vacuum @15:25:47.82 Cartridge 15 sampling 628/2000ml at 0.49ml/s with 25.9psi vacuum @15:26:47.75 Cartridge 15 sampling 657/2000ml at 0.49ml/s with 24.7psi vacuum @15:27:47.96 Cartridge 15 sampling 687/2000ml at 0.49ml/s with 19.1psi vacuum @15:28:47.82 Cartridge 15 sampling 716/2000ml at 0.48ml/s with 25.9psi vacuum @15:29:47.74 Cartridge 15 sampling 745/2000ml at 0.49ml/s with 24.5psi vacuum @15:30:24.81 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:30:24.83 Retry #4 of 10 after 10:00 @15:30:48.06 Cartridge 15 sampling 775/2000ml at 0.49ml/s with 24.2psi vacuum @15:31:47.73 Cartridge 15 sampling 804/2000ml at 0.48ml/s with 24.5psi vacuum @15:32:48.00 Cartridge 15 sampling 832/2000ml at 0.47ml/s with 17.6psi vacuum 1.7m deep (17.2psia) @15:33:47.74 Cartridge 15 sampling 861/2000ml at 0.47ml/s with 24.9psi vacuum @15:34:47.92 Cartridge 15 sampling 890/2000ml at 0.47ml/s with 25.9psi vacuum @15:35:47.74 Cartridge 15 sampling 919/2000ml at 0.46ml/s with 25.3psi vacuum @15:36:47.96 Cartridge 15 sampling 947/2000ml at 0.48ml/s with 25.7psi vacuum @15:37:47.73 Cartridge 15 sampling 975/2000ml at 0.47ml/s with 25.6psi vacuum @15:38:47.89 Cartridge 15 sampling 1003/2000ml at 0.48ml/s with 24.3psi vacuum @15:39:47.74 Cartridge 15 sampling 1032/2000ml at 0.47ml/s with 24.7psi vacuum @15:40:24.88 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:40:24.90 Retry #5 of 10 after 10:00 @15:40:47.81 Cartridge 15 sampling 1060/2000ml at 0.46ml/s with 25.1psi vacuum @15:41:47.74 Cartridge 15 sampling 1087/2000ml at 0.47ml/s with 24.8psi vacuum @15:42:48.23 Cartridge 15 sampling 1115/2000ml at 0.45ml/s with 19.8psi vacuum @15:43:48.12 Cartridge 15 sampling 1143/2000ml at 0.45ml/s with 26.0psi vacuum @15:44:47.75 Cartridge 15 sampling 1170/2000ml at 0.46ml/s with 25.3psi vacuum @15:45:48.16 Cartridge 15 sampling 1198/2000ml at 0.46ml/s with 25.3psi vacuum @15:46:47.75 Cartridge 15 sampling 1225/2000ml at 0.45ml/s with 19.0psi vacuum 1.9m deep (17.5psia) @15:47:47.90 Cartridge 15 sampling 1252/2000ml at 0.45ml/s with 24.9psi vacuum @15:48:47.75 Cartridge 15 sampling 1278/2000ml at 0.44ml/s with 25.2psi vacuum @15:49:47.86 Cartridge 15 sampling 1305/2000ml at 0.45ml/s with 23.3psi vacuum @15:50:25.01 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @15:50:25.05 Retry #6 of 10 after 10:00 @15:50:47.74 Cartridge 15 sampling 1332/2000ml at 0.45ml/s with 25.3psi vacuum @15:51:47.90 Cartridge 15 sampling 1358/2000ml at 0.43ml/s with 19.8psi vacuum @15:52:47.75 Cartridge 15 sampling 1385/2000ml at 0.44ml/s with 24.1psi vacuum @15:53:47.86 Cartridge 15 sampling 1411/2000ml at 0.44ml/s with 26.5psi vacuum @15:54:47.75 Cartridge 15 sampling 1438/2000ml at 0.44ml/s with 25.3psi vacuum @15:55:47.79 Cartridge 15 sampling 1464/2000ml at 0.42ml/s with 23.9psi vacuum @15:56:47.75 Cartridge 15 sampling 1490/2000ml at 0.43ml/s with 25.3psi vacuum @15:57:47.89 Cartridge 15 sampling 1516/2000ml at 0.43ml/s with 22.6psi vacuum @15:58:47.90 Cartridge 15 sampling 1542/2000ml at 0.42ml/s with 22.5psi vacuum @15:59:47.74 Cartridge 15 sampling 1567/2000ml at 0.42ml/s with 25.9psi vacuum @16:00:25.11 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:00:25.13 Retry #7 of 10 after 10:00 @16:00:47.92 Cartridge 15 sampling 1593/2000ml at 0.42ml/s with 25.9psi vacuum @16:01:47.75 Cartridge 15 sampling 1619/2000ml at 0.41ml/s with 26.6psi vacuum @16:02:47.89 Cartridge 15 sampling 1644/2000ml at 0.41ml/s with 21.7psi vacuum @16:03:47.74 Cartridge 15 sampling 1670/2000ml at 0.42ml/s with 24.3psi vacuum @16:04:47.91 Cartridge 15 sampling 1695/2000ml at 0.42ml/s with 26.3psi vacuum @16:05:47.75 Cartridge 15 sampling 1720/2000ml at 0.40ml/s with 25.9psi vacuum @16:06:47.91 Cartridge 15 sampling 1745/2000ml at 0.42ml/s with 23.2psi vacuum @16:07:47.75 Cartridge 15 sampling 1770/2000ml at 0.40ml/s with 26.1psi vacuum @16:08:47.79 Cartridge 15 sampling 1794/2000ml at 0.41ml/s with 23.9psi vacuum @16:09:47.75 Cartridge 15 sampling 1819/2000ml at 0.39ml/s with 25.9psi vacuum @16:10:25.72 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:10:25.83 Retry #8 of 10 after 10:00 @16:10:48.19 Cartridge 15 sampling 1843/2000ml at 0.40ml/s with 23.8psi vacuum @16:11:47.87 Cartridge 15 sampling 1867/2000ml at 0.41ml/s with 23.9psi vacuum @16:12:47.88 Cartridge 15 sampling 1891/2000ml at 0.41ml/s with 21.2psi vacuum @16:13:48.05 Cartridge 15 sampling 1916/2000ml at 0.39ml/s with 20.0psi vacuum 1.7m deep (17.2psia) @16:14:47.75 Cartridge 15 sampling 1940/2000ml at 0.40ml/s with 22.2psi vacuum @16:15:47.94 Cartridge 15 sampling 1964/2000ml at 0.41ml/s with 24.2psi vacuum @16:16:47.75 Cartridge 15 sampling 1988/2000ml at 0.39ml/s with 23.8psi vacuum @16:17:17.90 Pressure.coast @16:17:18.04 Intake.close @16:17:18.09 Exhaust.close @16:17:18.30 Toroid.reconfigure :hold @16:17:18.36 Toroid.seek :bypass @16:17:20.05 Toroid.reconfigure :default @16:17:20.15 Toroid.seek :clear @16:17:21.83 Sampled 2000.0ml @16:17:23.39 SamplePump.setPosition! 0ml @16:17:23.51 Bypass.open @16:17:23.65 Plunge.seek :reagent @16:17:24.42 Intake.open @16:17:24.69 Sample loop closed. Ready to Deliver Reagents @16:17:24.71 Archive.seek :sealed @16:17:25.27 SamplePump.reconfigure :slow @16:17:25.33 SamplePump.seek -3ml @16:17:28.47 Archive.seek :evac @16:17:32.08 delay 1 second @16:17:33.18 SamplePump.seek -2.8ml @16:17:33.88 delay 1 second @16:17:34.98 SamplePump.seek -3.3ml @16:17:36.31 delay 1 second @16:17:37.34 Exhaust.open @16:17:38.11 Bypass.close @16:17:39.09 SamplePump.seek -35.3ml @16:17:42.49 delay 1:00 @16:18:42.55 Archive.seek :clear @16:18:47.92 delay 3 seconds @16:18:51.01 Intake.close @16:18:51.07 Exhaust.close @16:18:59.73 RNAlater soak for 10:00 @16:20:25.91 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:20:25.98 Retry #9 of 10 after 10:00 @16:28:59.77 Archive.seek :tVent @16:29:01.44 delay 1 second @16:29:02.47 Archive.seek :evac @16:29:18.44 delay 10 seconds @16:29:28.48 Sample preserved and evacuated @16:29:29.18
Carousel.seek 15 @16:29:29.38 Plunge.seek :locked @16:29:30.36 Archive.seek :clear @16:29:30.54 Air.seek :clear @16:29:30.71 Lysis1.seek :clear @16:29:47.56
Archive.coast @16:29:47.62 Air.coast @16:29:47.66 Lysis1.coast @16:29:47.70 Lysis2.coast @16:29:47.74 Plunge.coast @16:29:47.79 SamplePump.coast @16:29:47.83 Toroid.coast @16:29:47.88 Carousel.coast @16:29:48.07 Gate.power :motor,:OFF @16:29:48.08 Gate.power :core,:OFF @16:29:48.10 Gate.power :analytic1,:OFF @16:29:48.11 Gate.power :analytic2,:OFF @16:29:48.12 Gate.power :analytic3,:OFF @16:29:48.13 Gate.power :main,:OFF @16:29:48.15 Safely stopped and ready to power off @16:29:48.38 delayUntil "03:00:00.00PDT30-Jun-21" @16:30:26.06 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:30:26.09 Retry #10 of 10 after 10:00 @16:40:26.18 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:40:26.20 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 15" @16:40:26.33 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:40:26.37 Retry #1 of 10 after 10:00 @16:50:26.45 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @16:50:26.47 Retry #2 of 10 after 10:00 @17:00:26.52 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:00:26.54 Retry #3 of 10 after 10:00 @17:10:26.59 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:10:26.61 Retry #4 of 10 after 10:00 @17:20:26.67 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:20:26.69 Retry #5 of 10 after 10:00 @17:30:26.81 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:30:26.87 Retry #6 of 10 after 10:00 @17:40:26.93 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:40:26.95 Retry #7 of 10 after 10:00 @17:50:27.42 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @17:50:27.48 Retry #8 of 10 after 10:00 @18:00:27.59 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:00:28.27 Retry #9 of 10 after 10:00 @18:10:28.31 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:10:28.33 Retry #10 of 10 after 10:00 @18:20:28.45 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:20:28.50 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 03:00PDT30-Jun-21" @18:20:28.68 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:20:28.74 Retry #1 of 10 after 10:00 @18:30:28.81 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:30:28.83 Retry #2 of 10 after 10:00 @18:40:28.95 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:40:29.00 Retry #3 of 10 after 10:00 @18:50:29.05 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:50:29.07 Retry #4 of 10 after 10:00 @19:00:29.14 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:00:29.17 Retry #5 of 10 after 10:00 @19:10:29.21 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:10:29.23 Retry #6 of 10 after 10:00 @19:20:29.28 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:20:29.30 Retry #7 of 10 after 10:00 @19:30:29.37 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:30:29.40 Retry #8 of 10 after 10:00 @19:40:29.45 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:40:29.47 Retry #9 of 10 after 10:00 @19:50:29.55 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:50:29.60 Retry #10 of 10 after 10:00 @20:00:29.65 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @22:50:39.89 Session Begun @22:50:47.42 -> GoodNews.xray @22:52:35.51 -> exit @22:52:35.55 Session Ended @23:32:49.68 Session Begun @23:33:24.57 -> GoodNews.mail "Restored networking (I think...)" @23:33:24.59 GoodNews.email "Restored networking (I think...)" @23:33:37.52 upsync @00:48:06.58PDT30-Jun-21 -> exit @00:48:06.63 Session Ended @03:00:00.38
Gate.power :main,:ON,limit to 5000mA @03:00:00.40 Gate.power :analytic1,:ON,limit to 5000mA @03:00:00.52 Gate.power :analytic2,:ON,limit to 5000mA @03:00:00.57 Gate.power :analytic3,:ON,limit to 5000mA @03:00:00.69 Powering core bus @03:00:03.25 Archive.configure :default @03:00:03.48 Air.configure :default @03:00:03.75 Lysis1.configure :default @03:00:03.95 Lysis2.configure :default @03:00:04.15 Plunge.configure :default @03:00:04.48 SamplePump.configure :default @03:00:04.95 ExtraPump.configure :default @03:00:05.10 Toroid.configure :default @03:00:05.24 Carousel.configure :default @03:00:05.38 Pressure.configure :default @03:00:05.45 All elf controllers running firmware version 4.61 @03:00:05.47 Gate.power :motor,:ON,limit to 5000mA @03:00:05.69 Archive.reconfigure :default @03:00:05.99 Archive.home.jog 0 @03:00:06.24 Toroid.reconfigure :default @03:00:06.27 Plunge.reconfigure :default @03:00:06.30 Lysis1.reconfigure :default @03:00:06.34 Air.reconfigure :default @03:00:06.40 Lysis2.reconfigure :default @03:00:06.64 Toroid.home.jog 0 @03:00:06.72 Lysis1.home.jog 0 @03:00:06.83 Plunge.home.jog 0 @03:00:06.93 Lysis2.home.jog 0 @03:00:06.97 Air.home.jog 0 @03:00:07.44 Archive.reconfigure :homing @03:00:07.61 Lysis2.reconfigure :homing @03:00:07.69 Toroid.reconfigure :homing @03:00:08.32 Archive.jog 1364 @03:00:08.36 Plunge.reconfigure :homing @03:00:08.39 Lysis1.reconfigure :homing @03:00:08.60 Lysis2.jog 1349 @03:00:08.64 Toroid.jog 636 @03:00:08.67 Plunge.jog 842 @03:00:08.70 Lysis1.jog 1352 @03:00:08.94 Air.reconfigure :homing @03:00:09.00 Air.jog 1355 @03:00:09.72 Plunge.setRawPosition! 50 @03:00:09.91 Archive.setRawPosition! 50 @03:00:10.33 Archive.seek :home @03:00:10.39 Plunge.seek :home @03:00:10.90 Lysis2.setRawPosition! 50 @03:00:11.09 Lysis1.setRawPosition! 50 @03:00:11.14 Air.setRawPosition! 50 @03:00:11.16 Archive.reconfigure :default @03:00:11.41 Lysis2.seek :home @03:00:11.58 Lysis1.seek :home @03:00:11.85 Air.seek :home @03:00:12.24 Toroid.setRawPosition! 0 @03:00:12.35 Lysis2.reconfigure :default @03:00:12.43 Plunge.reconfigure :default @03:00:12.62 Toroid.seek :home @03:00:13.18 Lysis1.reconfigure :default @03:00:13.23 Air.reconfigure :default @03:00:13.28 Toroid.reconfigure :default @03:00:14.28
Toroid.seek :clear @03:00:15.90 Carousel.reconfigure :default @03:00:15.95 Carousel.home.jog 0 @03:00:16.45 Carousel.seek 15 @03:00:16.65 Selecting Cartridge 14 @03:00:16.86 Carousel.seek 14 @03:00:17.37 Lysis1.seek :locked @03:00:20.80 Plunge.seek :locked @03:00:21.15 Air.seek :locked @03:00:25.38
Mounted archive Cartridge @03:00:25.49 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 14" @03:00:25.79 Duration of filtering limited to 4:05:00 @03:00:26.09 Priming sample loop w/160ml, bypass w/40ml @03:00:26.11 SamplePump.setPosition! 0ml @03:00:26.14 Bypass.open @03:00:27.58 Intake.open @03:00:28.46 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @03:00:31.20 Vacuum tare is 18% of full scale @03:00:31.22 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @03:00:31.75 Pressure.reconfigure :default @03:00:31.94 Pressure.hold @03:00:32.36 SamplePump.reconfigure :prime @03:00:32.56 SamplePump.seek 40ml,3:51 @03:00:49.87 upsync @03:01:15.84 Pressure.coast @03:01:15.98 SamplePump.setPosition! 0ml @03:01:16.12 Exhaust.open @03:01:17.63 Bypass.close @03:01:19.44 Pressure.reconfigure :default @03:01:19.50 Pressure.hold @03:01:19.98 SamplePump.reconfigure :prime @03:01:20.13 SamplePump.seek 160ml,13:51 @03:04:11.89 Pressure.coast @03:04:12.04 Bypass.open @03:04:12.75 Intake.close @03:04:13.73 Toroid.reconfigure :hold @03:04:13.81 Toroid.seek :crack @03:04:15.51 Toroid.seek :cartridge @03:04:16.44 Toroid.reconfigure :default @03:04:16.50 Toroid.seek :clear @03:04:18.35 Pressurizing cartridge to 26psi above 14.6psia @03:04:18.37 SamplePump.setPosition! 0ml @03:04:18.40 Intake.open @03:04:19.18 Bypass.close @03:04:19.90 Exhaust.close @03:04:22.32 Pressure.reconfigure :default @03:04:22.46 Pressure.hold @03:04:22.99 SamplePump.reconfigure :charge @03:04:23.12 SamplePump.seek 2.3ml,42 seconds @03:04:35.59 Pressure.coast @03:04:35.64 Intake.close @03:04:36.52 Pressurized cartridge with 0.704ml to 26.1psi @03:04:36.53 Isolated @35.8psia (21.2psi) @03:04:36.54 Checking for cartridge leaks from 35.8psia @03:04:36.56 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @03:04:43.89 Toroid.reconfigure :hold @03:04:43.95 Toroid.seek :bypass @03:04:45.58 Toroid.reconfigure :default @03:04:45.63 Toroid.seek :clear @03:04:47.25 Cartridge::Sampler::Leaky in FILTERING -- Isolated cartridge pressure fell by 20% to 28.5psia @03:04:47.27 Retry #1 of 2 @03:04:47.29 Cmd::SpareCartridge in FILTERING -- 0 spares remaining @03:04:47.32 Selecting Cartridge 13 @03:04:47.72 Air.seek :clear @03:04:51.67 Lysis1.seek :clear @03:04:52.02 Plunge.seek :clear @03:04:54.87 Carousel.seek 13 @03:04:55.39 Lysis1.seek :locked @03:04:58.80 Plunge.seek :locked @03:04:59.13 Air.seek :locked @03:05:03.44 Mounted archive Cartridge @03:05:03.48 Duration of filtering limited to 4:05:00 @03:05:03.71 Bypass.open @03:05:04.47 Toroid.reconfigure :hold @03:05:04.53 Toroid.seek :crack @03:05:06.12 Toroid.seek :cartridge @03:05:07.23 Toroid.reconfigure :default @03:05:07.43 Toroid.seek :clear @03:05:09.14 Exhaust.open @03:05:10.38 Pressurizing cartridge to 26psi above 14.8psia @03:05:10.39 SamplePump.setPosition! 0ml @03:05:10.53 Intake.open @03:05:11.41 Bypass.close @03:05:12.11 Exhaust.close @03:05:12.95 Pressure.reconfigure :default @03:05:13.01 Pressure.hold @03:05:13.23 SamplePump.seek 2.3ml,42 seconds @03:05:27.21 Pressure.coast @03:05:27.30 Intake.close @03:05:28.41 Pressurized cartridge with 0.880ml to 26.3psi @03:05:28.42 Isolated @35.6psia (20.7psi) @03:05:28.43 Checking for cartridge leaks from 35.6psia @03:05:28.52 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @03:05:35.26 Cartridge pressure fell 8% to 32.6psia @03:05:35.53 Sampling 2000.0ml @03:05:35.59 SamplePump.setPosition! 0ml @03:05:35.77 Exhaust.open @03:05:36.65 Intake.open @03:05:37.84 Pressure.reconfigure :default @03:05:37.98 Pressure.hold @03:05:38.66 SamplePump.reconfigure :sample @03:05:38.82 SamplePump.seek 2000ml,2:47:12 @03:05:40.82 Cartridge 13 sampling 1/2000ml at 0.63ml/s with 24.3psi vacuum @03:06:41.01 Cartridge 13 sampling 31/2000ml at 0.50ml/s with 24.6psi vacuum @03:07:40.82 Cartridge 13 sampling 62/2000ml at 0.51ml/s with 24.5psi vacuum @03:08:40.96 Cartridge 13 sampling 93/2000ml at 0.51ml/s with 24.1psi vacuum @03:09:40.82 Cartridge 13 sampling 123/2000ml at 0.50ml/s with 23.0psi vacuum @03:10:40.96 Cartridge 13 sampling 153/2000ml at 0.50ml/s with 24.4psi vacuum @03:11:40.81 Cartridge 13 sampling 183/2000ml at 0.49ml/s with 25.8psi vacuum @03:12:41.03 Cartridge 13 sampling 214/2000ml at 0.51ml/s with 25.5psi vacuum @03:13:40.82 Cartridge 13 sampling 244/2000ml at 0.51ml/s with 25.6psi vacuum @03:14:40.98 Cartridge 13 sampling 275/2000ml at 0.51ml/s with 22.2psi vacuum @03:15:40.82 Cartridge 13 sampling 306/2000ml at 0.52ml/s with 25.5psi vacuum @03:16:40.87 Cartridge 13 sampling 337/2000ml at 0.51ml/s with 23.3psi vacuum @03:17:40.82 Cartridge 13 sampling 368/2000ml at 0.51ml/s with 22.8psi vacuum @03:18:40.87 Cartridge 13 sampling 399/2000ml at 0.50ml/s with 24.5psi vacuum @03:19:41.03 Cartridge 13 sampling 429/2000ml at 0.50ml/s with 25.9psi vacuum @03:20:40.82 Cartridge 13 sampling 460/2000ml at 0.52ml/s with 25.3psi vacuum @03:21:40.94 Cartridge 13 sampling 491/2000ml at 0.51ml/s with 21.9psi vacuum @03:22:40.82 Cartridge 13 sampling 521/2000ml at 0.51ml/s with 19.9psi vacuum @03:23:40.98 Cartridge 13 sampling 552/2000ml at 0.51ml/s with 23.7psi vacuum @03:24:40.82 Cartridge 13 sampling 582/2000ml at 0.49ml/s with 26.1psi vacuum @03:25:41.18 Cartridge 13 sampling 613/2000ml at 0.51ml/s with 24.7psi vacuum @03:26:40.82 Cartridge 13 sampling 643/2000ml at 0.51ml/s with 23.5psi vacuum @03:27:41.01 Cartridge 13 sampling 673/2000ml at 0.49ml/s with 20.4psi vacuum @03:28:40.81 Cartridge 13 sampling 704/2000ml at 0.51ml/s with 23.2psi vacuum @03:29:40.86 Cartridge 13 sampling 734/2000ml at 0.49ml/s with 18.2psi vacuum @03:30:40.82 Cartridge 13 sampling 764/2000ml at 0.48ml/s with 25.6psi vacuum @03:31:40.82 Cartridge 13 sampling 794/2000ml at 0.49ml/s with 21.7psi vacuum @03:32:41.07 Cartridge 13 sampling 824/2000ml at 0.49ml/s with 26.4psi vacuum @03:33:40.82 Cartridge 13 sampling 854/2000ml at 0.49ml/s with 20.8psi vacuum @03:34:40.95 Cartridge 13 sampling 884/2000ml at 0.50ml/s with 25.7psi vacuum @03:35:40.82 Cartridge 13 sampling 913/2000ml at 0.50ml/s with 25.6psi vacuum @03:36:41.10 Cartridge 13 sampling 943/2000ml at 0.49ml/s with 23.3psi vacuum @03:37:40.81 Cartridge 13 sampling 972/2000ml at 0.49ml/s with 20.8psi vacuum @03:38:40.98 Cartridge 13 sampling 1001/2000ml at 0.49ml/s with 23.4psi vacuum @03:39:40.81 Cartridge 13 sampling 1030/2000ml at 0.49ml/s with 25.7psi vacuum @03:40:40.94 Cartridge 13 sampling 1059/2000ml at 0.48ml/s with 24.0psi vacuum @03:41:40.82 Cartridge 13 sampling 1088/2000ml at 0.47ml/s with 25.3psi vacuum @03:42:40.87 Cartridge 13 sampling 1116/2000ml at 0.48ml/s with 22.6psi vacuum @03:43:40.81 Cartridge 13 sampling 1145/2000ml at 0.48ml/s with 25.0psi vacuum @03:44:41.14 Cartridge 13 sampling 1173/2000ml at 0.45ml/s with 18.3psi vacuum @03:45:41.22 Cartridge 13 sampling 1201/2000ml at 0.48ml/s with 25.9psi vacuum @03:46:40.82 Cartridge 13 sampling 1229/2000ml at 0.46ml/s with 20.0psi vacuum @03:47:41.17 Cartridge 13 sampling 1257/2000ml at 0.46ml/s with 25.3psi vacuum @03:48:40.81 Cartridge 13 sampling 1284/2000ml at 0.46ml/s with 23.5psi vacuum @03:49:40.92 Cartridge 13 sampling 1312/2000ml at 0.46ml/s with 26.8psi vacuum @03:50:40.82 Cartridge 13 sampling 1340/2000ml at 0.47ml/s with 20.5psi vacuum @03:51:41.03 Cartridge 13 sampling 1367/2000ml at 0.45ml/s with 22.1psi vacuum @03:52:40.82 Cartridge 13 sampling 1394/2000ml at 0.45ml/s with 25.7psi vacuum @03:53:40.97 Cartridge 13 sampling 1422/2000ml at 0.46ml/s with 24.6psi vacuum @03:54:40.82 Cartridge 13 sampling 1449/2000ml at 0.44ml/s with 27.3psi vacuum @03:55:40.93 Cartridge 13 sampling 1476/2000ml at 0.43ml/s with 20.3psi vacuum @03:56:40.82 Cartridge 13 sampling 1503/2000ml at 0.44ml/s with 19.3psi vacuum @03:57:40.95 Cartridge 13 sampling 1530/2000ml at 0.45ml/s with 26.0psi vacuum @03:58:40.81 Cartridge 13 sampling 1556/2000ml at 0.45ml/s with 24.4psi vacuum @03:59:41.08 Cartridge 13 sampling 1583/2000ml at 0.44ml/s with 26.9psi vacuum @04:00:41.01 Cartridge 13 sampling 1610/2000ml at 0.44ml/s with 26.0psi vacuum @04:01:40.81 Cartridge 13 sampling 1637/2000ml at 0.45ml/s with 23.2psi vacuum @04:02:41.00 Cartridge 13 sampling 1663/2000ml at 0.44ml/s with 26.9psi vacuum @04:03:40.82 Cartridge 13 sampling 1689/2000ml at 0.43ml/s with 21.3psi vacuum @04:04:40.98 Cartridge 13 sampling 1715/2000ml at 0.41ml/s with 19.5psi vacuum @04:05:40.82 Cartridge 13 sampling 1741/2000ml at 0.42ml/s with 26.5psi vacuum @04:06:41.24 Cartridge 13 sampling 1767/2000ml at 0.44ml/s with 24.9psi vacuum @04:07:40.82 Cartridge 13 sampling 1793/2000ml at 0.42ml/s with 26.2psi vacuum @04:08:41.04 Cartridge 13 sampling 1819/2000ml at 0.42ml/s with 18.6psi vacuum 2.8m deep (18.8psia) @04:09:40.81 Cartridge 13 sampling 1844/2000ml at 0.43ml/s with 23.6psi vacuum @04:10:40.91 Cartridge 13 sampling 1869/2000ml at 0.42ml/s with 26.4psi vacuum @04:11:40.82 Cartridge 13 sampling 1894/2000ml at 0.41ml/s with 27.4psi vacuum @04:12:41.01 Cartridge 13 sampling 1919/2000ml at 0.41ml/s with 21.5psi vacuum @04:13:41.01 Cartridge 13 sampling 1944/2000ml at 0.42ml/s with 22.2psi vacuum @04:14:40.82 Cartridge 13 sampling 1968/2000ml at 0.40ml/s with 27.2psi vacuum @04:15:41.09 Cartridge 13 sampling 1993/2000ml at 0.40ml/s with 19.1psi vacuum @04:15:57.70 Pressure.coast @04:15:57.83 Intake.close @04:15:57.88 Exhaust.close @04:15:58.08 Toroid.reconfigure :hold @04:15:58.14 Toroid.seek :bypass @04:15:59.84 Toroid.reconfigure :default @04:15:59.99 Toroid.seek :clear @04:16:01.76 Sampled 2000.0ml @04:16:02.68 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 13" @04:16:03.67 SamplePump.setPosition! 0ml @04:16:03.72 Plunge.seek :reagent @04:16:03.76 Bypass.open @04:16:04.51 Intake.open @04:16:04.67 Sample loop closed. Ready to Deliver Reagents @04:16:04.69 Archive.seek :sealed @04:16:05.38 SamplePump.reconfigure :slow @04:16:05.44 SamplePump.seek -3ml @04:16:08.42 Archive.seek :evac @04:16:12.15 delay 1 second @04:16:13.30 SamplePump.seek -2.8ml @04:16:13.99 delay 1 second @04:16:15.12 SamplePump.seek -3.3ml @04:16:16.46 delay 1 second @04:16:17.48 Exhaust.open @04:16:18.39 Bypass.close @04:16:19.40 SamplePump.seek -35.3ml @04:16:21.08 upsync @04:16:22.50 delay 1:00 @04:17:22.57 Archive.seek :clear @04:17:28.25 delay 3 seconds @04:17:31.48 Intake.close @04:17:31.62 Exhaust.close @04:17:39.74 RNAlater soak for 10:00 @04:27:39.82 Archive.seek :tVent @04:27:41.52 delay 1 second @04:27:42.56 Archive.seek :evac @04:27:58.53 delay 10 seconds @04:28:08.57 Sample preserved and evacuated @04:28:09.03
Carousel.seek 13 @04:28:09.24 Plunge.seek :locked @04:28:10.22 Archive.seek :clear @04:28:10.37 Air.seek :clear @04:28:10.54 Lysis1.seek :clear @04:28:27.43
Archive.coast @04:28:27.47 Air.coast @04:28:27.51 Lysis1.coast @04:28:27.55 Lysis2.coast @04:28:27.60 Plunge.coast @04:28:27.64 SamplePump.coast @04:28:27.68 Toroid.coast @04:28:27.73 Carousel.coast @04:28:27.92 Gate.power :motor,:OFF @04:28:27.93 Gate.power :core,:OFF @04:28:27.95 Gate.power :analytic1,:OFF @04:28:27.96 Gate.power :analytic2,:OFF @04:28:27.97 Gate.power :analytic3,:OFF @04:28:27.98 Gate.power :main,:OFF @04:28:28.00 Safely stopped and ready to power off @04:28:28.22 delayUntil "15:00:00.00" @04:28:28.25 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 15:00" @04:28:44.59 upsync @15:00:00.26
Gate.power :main,:ON,limit to 5000mA @15:00:00.34 Gate.power :analytic1,:ON,limit to 5000mA @15:00:00.38 Gate.power :analytic2,:ON,limit to 5000mA @15:00:00.43 Gate.power :analytic3,:ON,limit to 5000mA @15:00:00.47 Powering core bus @15:00:03.02 Archive.configure :default @15:00:03.32 Air.configure :default @15:00:03.49 Lysis1.configure :default @15:00:03.93 Lysis2.configure :default @15:00:04.22 Plunge.configure :default @15:00:04.36 SamplePump.configure :default @15:00:05.06 ExtraPump.configure :default @15:00:05.20 Toroid.configure :default @15:00:05.35 Carousel.configure :default @15:00:05.48 Pressure.configure :default @15:00:05.54 All elf controllers running firmware version 4.61 @15:00:05.56 Gate.power :motor,:ON,limit to 5000mA @15:00:05.87 Archive.reconfigure :default @15:00:06.20 Archive.home.jog 0 @15:00:06.25 Plunge.reconfigure :default @15:00:06.28 Lysis2.reconfigure :default @15:00:06.32 Lysis1.reconfigure :default @15:00:06.35 Toroid.reconfigure :default @15:00:06.40 Air.reconfigure :default @15:00:06.64 Plunge.home.jog 0 @15:00:06.69 Lysis1.home.jog 0 @15:00:06.73 Lysis2.home.jog 0 @15:00:06.75 Toroid.home.jog 0 @15:00:06.80 Air.home.jog 0 @15:00:07.50 Toroid.reconfigure :homing @15:00:07.54 Plunge.reconfigure :homing @15:00:07.57 Lysis1.reconfigure :homing @15:00:07.61 Air.reconfigure :homing @15:00:07.67 Archive.reconfigure :homing @15:00:07.71 Lysis2.reconfigure :homing @15:00:07.93 Plunge.jog 845 @15:00:07.98 Lysis1.jog 1359 @15:00:08.02 Toroid.jog 630 @15:00:08.05 Air.jog 1357 @15:00:08.08 Lysis2.jog 1349 @15:00:08.11 Archive.jog 1367 @15:00:09.11 Plunge.setRawPosition! 50 @15:00:09.26 Plunge.seek :home @15:00:09.72 Lysis1.setRawPosition! 50 @15:00:09.79 Lysis2.setRawPosition! 50 @15:00:09.84 Archive.setRawPosition! 50 @15:00:09.90 Air.setRawPosition! 50 @15:00:10.28 Lysis1.seek :home @15:00:10.36 Lysis2.seek :home @15:00:10.45 Archive.seek :home @15:00:10.56 Air.seek :home @15:00:11.31 Toroid.setRawPosition! 0 @15:00:11.44 Plunge.reconfigure :default @15:00:11.46 Lysis1.reconfigure :default @15:00:11.65 Archive.reconfigure :default @15:00:11.69 Lysis2.reconfigure :default @15:00:11.83 Air.reconfigure :default @15:00:12.28 Toroid.seek :home @15:00:12.82 Toroid.reconfigure :default @15:00:13.25
Toroid.seek :clear @15:00:14.92 Carousel.reconfigure :default @15:00:14.97 Carousel.home.jog 0 @15:00:15.47 Carousel.seek 13 @15:00:15.65 Selecting Cartridge 12 @15:00:15.86 Carousel.seek 12 @15:00:16.37 Lysis1.seek :locked @15:00:19.83 Plunge.seek :locked @15:00:20.16 Air.seek :locked @15:00:24.39
Mounted archive Cartridge @15:00:24.49 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 12" @15:00:24.81 Duration of filtering limited to 4:05:00 @15:00:25.10 Priming sample loop w/160ml, bypass w/40ml @15:00:25.11 SamplePump.setPosition! 0ml @15:00:25.15 Bypass.open @15:00:26.66 Intake.open @15:00:27.63 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:00:30.40 Vacuum tare is 18% of full scale @15:00:30.42 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -17psi @15:00:31.00 Pressure.reconfigure :default @15:00:31.14 Pressure.hold @15:00:31.73 SamplePump.reconfigure :prime @15:00:31.82 SamplePump.seek 40ml,3:51 @15:00:42.07 upsync @15:01:15.21 Pressure.coast @15:01:15.36 SamplePump.setPosition! 0ml @15:01:16.83 Exhaust.open @15:01:18.37 Bypass.close @15:01:19.94 Pressure.reconfigure :default @15:01:19.99 Pressure.hold @15:01:20.27 SamplePump.reconfigure :prime @15:01:20.33 SamplePump.seek 160ml,13:51 @15:04:12.02 Pressure.coast @15:04:12.09 Bypass.open @15:04:12.84 Intake.close @15:04:13.89 Toroid.reconfigure :hold @15:04:13.95 Toroid.seek :crack @15:04:15.60 Toroid.seek :cartridge @15:04:16.53 Toroid.reconfigure :default @15:04:16.59 Toroid.seek :clear @15:04:18.42 Pressurizing cartridge to 26psi above 14.5psia @15:04:18.44 SamplePump.setPosition! 0ml @15:04:18.47 Intake.open @15:04:19.27 Bypass.close @15:04:20.14 Exhaust.close @15:04:21.17 Pressure.reconfigure :default @15:04:21.29 Pressure.hold @15:04:21.84 SamplePump.reconfigure :charge @15:04:21.99 SamplePump.seek 2.3ml,42 seconds @15:04:35.10 Pressure.coast @15:04:35.14 Intake.close @15:04:36.04 Pressurized cartridge with 0.784ml to 26.1psi @15:04:36.05 Isolated @35.9psia (21.4psi) @15:04:36.06 Checking for cartridge leaks from 35.9psia @15:04:36.08 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @15:04:43.18 Cartridge pressure fell 6% to 33.8psia @15:04:43.26 Sampling 2000.0ml @15:04:43.28 SamplePump.setPosition! 0ml @15:04:43.32 Exhaust.open @15:04:43.99 Intake.open @15:04:44.97 Pressure.reconfigure :default @15:04:45.03 Pressure.hold @15:04:45.15 Sea 1.7m deep (17.2psia) @15:04:45.32 SamplePump.reconfigure :sample @15:04:45.39 SamplePump.seek 2000ml,2:47:12 @15:04:48.39 Cartridge 12 sampling 1/2000ml at 0.55ml/s with 24.5psi vacuum @15:05:48.04 Cartridge 12 sampling 33/2000ml at 0.53ml/s with 25.0psi vacuum @15:06:48.27 Cartridge 12 sampling 65/2000ml at 0.53ml/s with 23.3psi vacuum @15:07:48.04 Cartridge 12 sampling 96/2000ml at 0.53ml/s with 24.7psi vacuum @15:08:48.30 Cartridge 12 sampling 128/2000ml at 0.55ml/s with 23.9psi vacuum @15:09:48.04 Cartridge 12 sampling 161/2000ml at 0.55ml/s with 21.6psi vacuum @15:10:48.17 Cartridge 12 sampling 193/2000ml at 0.55ml/s with 24.1psi vacuum @15:11:48.03 Cartridge 12 sampling 226/2000ml at 0.53ml/s with 23.8psi vacuum @15:12:48.13 Cartridge 12 sampling 258/2000ml at 0.53ml/s with 25.2psi vacuum @15:13:48.04 Cartridge 12 sampling 291/2000ml at 0.54ml/s with 24.4psi vacuum @15:14:48.19 Cartridge 12 sampling 323/2000ml at 0.54ml/s with 21.4psi vacuum @15:15:48.04 Cartridge 12 sampling 356/2000ml at 0.55ml/s with 24.3psi vacuum @15:16:48.19 Cartridge 12 sampling 389/2000ml at 0.54ml/s with 25.2psi vacuum @15:17:48.42 Cartridge 12 sampling 421/2000ml at 0.55ml/s with 22.6psi vacuum @15:18:48.03 Cartridge 12 sampling 454/2000ml at 0.55ml/s with 22.5psi vacuum @15:19:48.36 Cartridge 12 sampling 487/2000ml at 0.54ml/s with 25.6psi vacuum @15:20:48.02 Cartridge 12 sampling 519/2000ml at 0.54ml/s with 24.1psi vacuum @15:21:48.20 Cartridge 12 sampling 552/2000ml at 0.55ml/s with 21.1psi vacuum @15:22:48.03 Cartridge 12 sampling 584/2000ml at 0.54ml/s with 23.8psi vacuum @15:23:48.24 Cartridge 12 sampling 616/2000ml at 0.53ml/s with 20.1psi vacuum @15:24:48.03 Cartridge 12 sampling 649/2000ml at 0.54ml/s with 25.4psi vacuum @15:25:48.29 Cartridge 12 sampling 681/2000ml at 0.54ml/s with 24.8psi vacuum @15:26:48.03 Cartridge 12 sampling 713/2000ml at 0.54ml/s with 24.7psi vacuum @15:27:48.09 Cartridge 12 sampling 745/2000ml at 0.52ml/s with 20.3psi vacuum @15:28:48.02 Cartridge 12 sampling 777/2000ml at 0.53ml/s with 20.7psi vacuum @15:29:48.10 Cartridge 12 sampling 809/2000ml at 0.54ml/s with 24.6psi vacuum @15:30:48.03 Cartridge 12 sampling 842/2000ml at 0.54ml/s with 22.3psi vacuum @15:31:48.17 Cartridge 12 sampling 874/2000ml at 0.53ml/s with 24.5psi vacuum @15:32:48.43 Cartridge 12 sampling 906/2000ml at 0.53ml/s with 24.7psi vacuum @15:33:48.04 Cartridge 12 sampling 937/2000ml at 0.53ml/s with 26.7psi vacuum @15:34:48.21 Cartridge 12 sampling 969/2000ml at 0.52ml/s with 26.3psi vacuum @15:35:48.03 Cartridge 12 sampling 1000/2000ml at 0.52ml/s with 23.5psi vacuum @15:36:48.13 Cartridge 12 sampling 1032/2000ml at 0.52ml/s with 24.2psi vacuum @15:37:48.03 Cartridge 12 sampling 1063/2000ml at 0.53ml/s with 24.2psi vacuum @15:38:48.22 Cartridge 12 sampling 1094/2000ml at 0.52ml/s with 22.3psi vacuum @15:39:48.04 Cartridge 12 sampling 1125/2000ml at 0.51ml/s with 20.9psi vacuum @15:40:48.27 Cartridge 12 sampling 1156/2000ml at 0.51ml/s with 25.2psi vacuum @15:41:48.03 Cartridge 12 sampling 1187/2000ml at 0.52ml/s with 25.7psi vacuum @15:42:48.08 Cartridge 12 sampling 1218/2000ml at 0.51ml/s with 25.7psi vacuum @15:43:48.04 Cartridge 12 sampling 1249/2000ml at 0.50ml/s with 23.1psi vacuum @15:44:48.12 Cartridge 12 sampling 1280/2000ml at 0.50ml/s with 26.6psi vacuum @15:45:48.27 Cartridge 12 sampling 1310/2000ml at 0.50ml/s with 21.3psi vacuum @15:46:48.04 Cartridge 12 sampling 1340/2000ml at 0.50ml/s with 24.3psi vacuum @15:47:49.52 Cartridge 12 sampling 1370/2000ml at 0.50ml/s with 26.7psi vacuum @15:48:48.04 Cartridge 12 sampling 1400/2000ml at 0.51ml/s with 24.0psi vacuum @15:49:48.19 Cartridge 12 sampling 1431/2000ml at 0.49ml/s with 21.1psi vacuum @15:50:48.04 Cartridge 12 sampling 1461/2000ml at 0.50ml/s with 25.1psi vacuum @15:51:48.18 Cartridge 12 sampling 1490/2000ml at 0.50ml/s with 22.5psi vacuum @15:52:48.04 Cartridge 12 sampling 1520/2000ml at 0.49ml/s with 22.1psi vacuum @15:53:48.20 Cartridge 12 sampling 1550/2000ml at 0.48ml/s with 27.1psi vacuum @15:54:48.04 Cartridge 12 sampling 1579/2000ml at 0.50ml/s with 24.9psi vacuum @15:55:48.21 Cartridge 12 sampling 1609/2000ml at 0.47ml/s with 24.5psi vacuum @15:56:48.04 Cartridge 12 sampling 1638/2000ml at 0.49ml/s with 22.3psi vacuum @15:57:48.20 Cartridge 12 sampling 1667/2000ml at 0.49ml/s with 23.2psi vacuum @15:58:48.04 Cartridge 12 sampling 1696/2000ml at 0.48ml/s with 23.6psi vacuum @15:59:48.28 Cartridge 12 sampling 1725/2000ml at 0.48ml/s with 21.7psi vacuum @16:00:48.04 Cartridge 12 sampling 1754/2000ml at 0.47ml/s with 25.6psi vacuum @16:01:48.03 Cartridge 12 sampling 1783/2000ml at 0.48ml/s with 22.7psi vacuum @16:02:48.40 Cartridge 12 sampling 1811/2000ml at 0.46ml/s with 27.7psi vacuum @16:03:48.04 Cartridge 12 sampling 1839/2000ml at 0.47ml/s with 25.0psi vacuum @16:04:48.22 Cartridge 12 sampling 1868/2000ml at 0.45ml/s with 22.9psi vacuum @16:05:48.04 Cartridge 12 sampling 1896/2000ml at 0.46ml/s with 21.6psi vacuum @16:06:48.28 Cartridge 12 sampling 1923/2000ml at 0.46ml/s with 19.9psi vacuum @16:07:48.04 Cartridge 12 sampling 1951/2000ml at 0.46ml/s with 23.5psi vacuum @16:08:48.15 Cartridge 12 sampling 1978/2000ml at 0.46ml/s with 23.1psi vacuum @16:09:35.95 Pressure.coast @16:09:36.09 Intake.close @16:09:36.14 Exhaust.close @16:09:36.34 Toroid.reconfigure :hold @16:09:36.40 Toroid.seek :bypass @16:09:38.02 Toroid.reconfigure :default @16:09:38.08 Toroid.seek :clear @16:09:39.69 Sampled 2000.0ml @16:09:40.14 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 12" @16:09:40.91 SamplePump.setPosition! 0ml @16:09:41.09 Plunge.seek :reagent @16:09:41.24 Bypass.open @16:09:42.20 Intake.open @16:09:42.46 Sample loop closed. Ready to Deliver Reagents @16:09:42.48 Archive.seek :sealed @16:09:43.25 SamplePump.reconfigure :slow @16:09:43.41 SamplePump.seek -3ml @16:09:46.25 Archive.seek :evac @16:09:50.18 delay 1 second @16:09:51.53 SamplePump.seek -2.8ml @16:09:52.27 delay 1 second @16:09:52.68 upsync @16:09:53.43 SamplePump.seek -3.3ml @16:09:54.84 delay 1 second @16:09:55.88 Exhaust.open @16:09:56.58 Bypass.close @16:09:57.34 SamplePump.seek -35.3ml @16:10:00.26 delay 1:00 @16:11:00.34 Archive.seek :clear @16:11:06.12 delay 3 seconds @16:11:09.24 Intake.close @16:11:09.29 Exhaust.close @16:11:17.47 RNAlater soak for 10:00 @16:21:17.50 Archive.seek :tVent @16:21:19.19 delay 1 second @16:21:20.22 Archive.seek :evac @16:21:36.19 delay 10 seconds @16:21:46.23 Sample preserved and evacuated @16:21:46.76
Carousel.seek 12 @16:21:47.53 Plunge.seek :locked @16:21:48.76 Archive.seek :clear @16:21:49.25 Air.seek :clear @16:21:49.70 Lysis1.seek :clear @16:22:06.18
Archive.coast @16:22:06.40 Air.coast @16:22:06.46 Lysis1.coast @16:22:06.66 Lysis2.coast @16:22:06.71 Plunge.coast @16:22:06.86 SamplePump.coast @16:22:06.94 Toroid.coast @16:22:07.00 Carousel.coast @16:22:08.70 Gate.power :motor,:OFF @16:22:08.83 Gate.power :core,:OFF @16:22:08.85 Gate.power :analytic1,:OFF @16:22:08.89 Gate.power :analytic2,:OFF @16:22:08.92 Gate.power :analytic3,:OFF @16:22:08.93 Gate.power :main,:OFF @16:22:08.94 Safely stopped and ready to power off @16:22:09.30 delayUntil "03:00:00.00PDT01-Jul-21" @16:22:09.41 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 03:00PDT01-Jul-21" @16:22:25.79 upsync @03:00:00.26PDT01-Jul-21
Gate.power :main,:ON,limit to 5000mA @03:00:00.28 Gate.power :analytic1,:ON,limit to 5000mA @03:00:00.31 Gate.power :analytic2,:ON,limit to 5000mA @03:00:00.33 Gate.power :analytic3,:ON,limit to 5000mA @03:00:00.40 Powering core bus @03:00:03.10 Archive.configure :default @03:00:03.24 Air.configure :default @03:00:03.35 Lysis1.configure :default @03:00:03.46 Lysis2.configure :default @03:00:03.60 Plunge.configure :default @03:00:03.73 SamplePump.configure :default @03:00:03.87 ExtraPump.configure :default @03:00:04.01 Toroid.configure :default @03:00:04.15 Carousel.configure :default @03:00:04.29 Pressure.configure :default @03:00:04.34 All elf controllers running firmware version 4.61 @03:00:04.36 Gate.power :motor,:ON,limit to 5000mA @03:00:04.59 Archive.reconfigure :default @03:00:04.90 Archive.home.jog 0 @03:00:05.06 Lysis2.reconfigure :default @03:00:05.16 Plunge.reconfigure :default @03:00:05.22 Air.reconfigure :default @03:00:05.26 Lysis1.reconfigure :default @03:00:05.32 Toroid.reconfigure :default @03:00:05.47 Lysis2.home.jog 0 @03:00:05.57 Plunge.home.jog 0 @03:00:05.58 Toroid.home.jog 0 @03:00:05.63 Lysis1.home.jog 0 @03:00:05.68 Air.home.jog 0 @03:00:06.13 Archive.reconfigure :homing @03:00:06.32 Toroid.reconfigure :homing @03:00:06.42 Archive.jog 1363 @03:00:06.45 Plunge.reconfigure :homing @03:00:06.47 Lysis2.reconfigure :homing @03:00:06.50 Air.reconfigure :homing @03:00:06.55 Lysis1.reconfigure :homing @03:00:06.79 Toroid.jog 631 @03:00:06.85 Air.jog 1357 @03:00:06.88 Lysis2.jog 1349 @03:00:06.92 Plunge.jog 840 @03:00:06.96 Lysis1.jog 1361 @03:00:07.83 Archive.setRawPosition! 50 @03:00:07.96 Archive.seek :home @03:00:08.56 Lysis2.setRawPosition! 50 @03:00:08.73 Lysis1.setRawPosition! 50 @03:00:08.81 Air.setRawPosition! 50 @03:00:08.90 Archive.reconfigure :default @03:00:08.96 Plunge.setRawPosition! 50 @03:00:09.42 Lysis2.seek :home @03:00:09.60 Lysis1.seek :home @03:00:10.62 Plunge.seek :home @03:00:11.00 Air.seek :home @03:00:11.67 Toroid.setRawPosition! 0 @03:00:11.80 Plunge.reconfigure :default @03:00:11.83 Air.reconfigure :default @03:00:11.94 Lysis2.reconfigure :default @03:00:11.97 Lysis1.reconfigure :default @03:00:12.37 Toroid.seek :home @03:00:13.07 Toroid.reconfigure :default @03:00:13.66
Toroid.seek :clear @03:00:15.32 Carousel.reconfigure :default @03:00:15.37 Carousel.home.jog 0 @03:00:15.87 Carousel.seek 12 @03:00:16.04 Selecting Cartridge 11 @03:00:16.25 Carousel.seek 11 @03:00:16.76 Lysis1.seek :locked @03:00:20.15 Plunge.seek :locked @03:00:20.48 Air.seek :locked @03:00:24.84
Mounted archive Cartridge @03:00:24.94 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 11" @03:00:25.26 Duration of filtering limited to 4:05:00 @03:00:25.55 Priming sample loop w/160ml, bypass w/40ml @03:00:25.57 SamplePump.setPosition! 0ml @03:00:25.61 Bypass.open @03:00:27.06 Intake.open @03:00:28.33 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @03:00:31.07 Vacuum tare is 18% of full scale @03:00:31.18 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @03:00:31.55 Pressure.reconfigure :default @03:00:31.74 Pressure.hold @03:00:32.16 SamplePump.reconfigure :prime @03:00:32.23 SamplePump.seek 40ml,3:51 @03:00:41.31 upsync @03:01:15.51 Pressure.coast @03:01:15.57 SamplePump.setPosition! 0ml @03:01:15.78 Exhaust.open @03:01:17.30 Bypass.close @03:01:19.00 Pressure.reconfigure :default @03:01:19.06 Pressure.hold @03:01:19.33 SamplePump.reconfigure :prime @03:01:19.39 SamplePump.seek 160ml,13:51 @03:04:11.04 Pressure.coast @03:04:11.12 Bypass.open @03:04:11.98 Intake.close @03:04:13.12 Toroid.reconfigure :hold @03:04:13.27 Toroid.seek :crack @03:04:14.94 Toroid.seek :cartridge @03:04:15.87 Toroid.reconfigure :default @03:04:15.93 Toroid.seek :clear @03:04:17.78 Pressurizing cartridge to 26psi above 14.5psia @03:04:17.80 SamplePump.setPosition! 0ml @03:04:17.83 Intake.open @03:04:18.67 Bypass.close @03:04:19.66 Exhaust.close @03:04:20.75 Pressure.reconfigure :default @03:04:20.87 Pressure.hold @03:04:21.35 SamplePump.reconfigure :charge @03:04:21.57 SamplePump.seek 2.3ml,42 seconds @03:04:33.77 Pressure.coast @03:04:33.82 Intake.close @03:04:34.70 Pressurized cartridge with 0.681ml to 26.0psi @03:04:34.71 Isolated @35.8psia (21.4psi) @03:04:34.72 Checking for cartridge leaks from 35.8psia @03:04:34.74 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @03:04:42.83 Cartridge pressure fell 9% to 32.6psia @03:04:42.92 Sampling 2000.0ml @03:04:42.93 SamplePump.setPosition! 0ml @03:04:42.98 Exhaust.open @03:04:43.68 Intake.open @03:04:44.64 Pressure.reconfigure :default @03:04:44.70 Pressure.hold @03:04:44.83 Sea 2.7m deep (18.7psia) @03:04:44.99 SamplePump.reconfigure :sample @03:04:45.06 SamplePump.seek 2000ml,2:47:12 @03:04:47.84 Cartridge 11 sampling 1/2000ml at 0.46ml/s with 26.3psi vacuum @03:05:47.70 Cartridge 11 sampling 29/2000ml at 0.47ml/s with 24.4psi vacuum @03:06:47.92 Cartridge 11 sampling 58/2000ml at 0.47ml/s with 22.2psi vacuum @03:07:47.71 Cartridge 11 sampling 85/2000ml at 0.46ml/s with 24.5psi vacuum @03:08:47.86 Cartridge 11 sampling 113/2000ml at 0.47ml/s with 24.5psi vacuum @03:09:47.71 Cartridge 11 sampling 141/2000ml at 0.45ml/s with 24.5psi vacuum @03:10:47.88 Cartridge 11 sampling 168/2000ml at 0.45ml/s with 24.7psi vacuum @03:11:47.71 Cartridge 11 sampling 196/2000ml at 0.45ml/s with 20.9psi vacuum @03:12:47.87 Cartridge 11 sampling 223/2000ml at 0.45ml/s with 24.7psi vacuum @03:13:47.71 Cartridge 11 sampling 250/2000ml at 0.46ml/s with 24.4psi vacuum @03:14:47.93 Cartridge 11 sampling 278/2000ml at 0.46ml/s with 24.7psi vacuum @03:15:47.95 Cartridge 11 sampling 306/2000ml at 0.46ml/s with 26.5psi vacuum @03:16:47.71 Cartridge 11 sampling 334/2000ml at 0.47ml/s with 22.3psi vacuum @03:17:47.98 Cartridge 11 sampling 362/2000ml at 0.47ml/s with 23.9psi vacuum @03:18:47.72 Cartridge 11 sampling 390/2000ml at 0.47ml/s with 23.7psi vacuum @03:19:48.03 Cartridge 11 sampling 418/2000ml at 0.46ml/s with 24.2psi vacuum @03:20:47.70 Cartridge 11 sampling 446/2000ml at 0.45ml/s with 20.1psi vacuum @03:21:47.88 Cartridge 11 sampling 473/2000ml at 0.46ml/s with 21.5psi vacuum @03:22:47.70 Cartridge 11 sampling 501/2000ml at 0.47ml/s with 24.8psi vacuum @03:23:47.87 Cartridge 11 sampling 528/2000ml at 0.44ml/s with 25.3psi vacuum @03:24:47.70 Cartridge 11 sampling 555/2000ml at 0.45ml/s with 24.3psi vacuum @03:25:47.79 Cartridge 11 sampling 582/2000ml at 0.45ml/s with 24.4psi vacuum @03:26:47.70 Cartridge 11 sampling 609/2000ml at 0.45ml/s with 25.0psi vacuum @03:27:47.74 Cartridge 11 sampling 636/2000ml at 0.45ml/s with 24.2psi vacuum @03:28:47.70 Cartridge 11 sampling 663/2000ml at 0.44ml/s with 24.8psi vacuum @03:29:47.91 Cartridge 11 sampling 690/2000ml at 0.44ml/s with 25.7psi vacuum @03:30:48.12 Cartridge 11 sampling 716/2000ml at 0.46ml/s with 21.0psi vacuum @03:31:47.70 Cartridge 11 sampling 743/2000ml at 0.44ml/s with 24.9psi vacuum @03:32:47.87 Cartridge 11 sampling 769/2000ml at 0.44ml/s with 25.5psi vacuum @03:33:47.71 Cartridge 11 sampling 795/2000ml at 0.43ml/s with 23.7psi vacuum @03:34:47.86 Cartridge 11 sampling 821/2000ml at 0.43ml/s with 26.0psi vacuum @03:35:47.71 Cartridge 11 sampling 846/2000ml at 0.43ml/s with 24.5psi vacuum @03:36:47.89 Cartridge 11 sampling 872/2000ml at 0.42ml/s with 25.2psi vacuum @03:37:47.71 Cartridge 11 sampling 898/2000ml at 0.42ml/s with 24.7psi vacuum @03:38:47.85 Cartridge 11 sampling 923/2000ml at 0.41ml/s with 25.1psi vacuum @03:39:47.70 Cartridge 11 sampling 949/2000ml at 0.41ml/s with 25.3psi vacuum @03:40:47.77 Cartridge 11 sampling 974/2000ml at 0.42ml/s with 21.1psi vacuum @03:41:47.70 Cartridge 11 sampling 999/2000ml at 0.42ml/s with 21.3psi vacuum @03:42:47.95 Cartridge 11 sampling 1024/2000ml at 0.41ml/s with 23.7psi vacuum @03:43:48.01 Cartridge 11 sampling 1049/2000ml at 0.42ml/s with 25.8psi vacuum @03:44:47.69 Cartridge 11 sampling 1074/2000ml at 0.40ml/s with 25.5psi vacuum @03:45:48.09 Cartridge 11 sampling 1098/2000ml at 0.40ml/s with 21.3psi vacuum @03:46:47.69 Cartridge 11 sampling 1123/2000ml at 0.41ml/s with 24.1psi vacuum @03:47:47.88 Cartridge 11 sampling 1147/2000ml at 0.40ml/s with 24.3psi vacuum @03:48:47.70 Cartridge 11 sampling 1171/2000ml at 0.40ml/s with 24.7psi vacuum @03:49:47.93 Cartridge 11 sampling 1195/2000ml at 0.39ml/s with 24.7psi vacuum @03:50:47.70 Cartridge 11 sampling 1219/2000ml at 0.40ml/s with 24.2psi vacuum @03:51:47.80 Cartridge 11 sampling 1242/2000ml at 0.40ml/s with 23.7psi vacuum @03:52:47.70 Cartridge 11 sampling 1266/2000ml at 0.39ml/s with 25.0psi vacuum @03:53:47.85 Cartridge 11 sampling 1289/2000ml at 0.39ml/s with 25.3psi vacuum @03:54:47.70 Cartridge 11 sampling 1313/2000ml at 0.39ml/s with 24.9psi vacuum @03:55:47.87 Cartridge 11 sampling 1336/2000ml at 0.39ml/s with 26.0psi vacuum @03:56:47.71 Cartridge 11 sampling 1359/2000ml at 0.38ml/s with 24.8psi vacuum @03:57:47.94 Cartridge 11 sampling 1382/2000ml at 0.38ml/s with 24.5psi vacuum @03:58:47.94 Cartridge 11 sampling 1405/2000ml at 0.37ml/s with 25.0psi vacuum @03:59:47.71 Cartridge 11 sampling 1428/2000ml at 0.37ml/s with 24.1psi vacuum @04:00:48.02 Cartridge 11 sampling 1450/2000ml at 0.38ml/s with 21.2psi vacuum @04:01:47.70 Cartridge 11 sampling 1472/2000ml at 0.36ml/s with 25.1psi vacuum @04:02:47.95 Cartridge 11 sampling 1495/2000ml at 0.36ml/s with 24.6psi vacuum @04:03:47.70 Cartridge 11 sampling 1517/2000ml at 0.36ml/s with 25.4psi vacuum @04:04:47.92 Cartridge 11 sampling 1539/2000ml at 0.36ml/s with 24.3psi vacuum @04:05:47.71 Cartridge 11 sampling 1560/2000ml at 0.36ml/s with 24.4psi vacuum @04:06:47.91 Cartridge 11 sampling 1582/2000ml at 0.36ml/s with 24.5psi vacuum @04:07:47.70 Cartridge 11 sampling 1603/2000ml at 0.35ml/s with 25.3psi vacuum @04:08:48.04 Cartridge 11 sampling 1625/2000ml at 0.37ml/s with 24.2psi vacuum @04:09:47.70 Cartridge 11 sampling 1647/2000ml at 0.36ml/s with 21.9psi vacuum @04:10:47.88 Cartridge 11 sampling 1669/2000ml at 0.35ml/s with 23.3psi vacuum @04:11:47.71 Cartridge 11 sampling 1690/2000ml at 0.36ml/s with 25.4psi vacuum @04:12:47.88 Cartridge 11 sampling 1711/2000ml at 0.35ml/s with 22.9psi vacuum @04:13:48.05 Cartridge 11 sampling 1732/2000ml at 0.36ml/s with 24.2psi vacuum @04:14:47.71 Cartridge 11 sampling 1753/2000ml at 0.35ml/s with 24.5psi vacuum @04:15:47.92 Cartridge 11 sampling 1774/2000ml at 0.35ml/s with 23.9psi vacuum @04:16:47.71 Cartridge 11 sampling 1796/2000ml at 0.36ml/s with 24.5psi vacuum @04:17:47.83 Cartridge 11 sampling 1817/2000ml at 0.36ml/s with 21.3psi vacuum @04:18:47.71 Cartridge 11 sampling 1838/2000ml at 0.36ml/s with 24.7psi vacuum @04:19:47.79 Cartridge 11 sampling 1859/2000ml at 0.35ml/s with 24.5psi vacuum @04:20:48.28 Cartridge 11 sampling 1880/2000ml at 0.34ml/s with 25.2psi vacuum @04:21:47.79 Cartridge 11 sampling 1901/2000ml at 0.34ml/s with 20.3psi vacuum @04:22:47.70 Cartridge 11 sampling 1921/2000ml at 0.35ml/s with 25.4psi vacuum @04:23:47.87 Cartridge 11 sampling 1941/2000ml at 0.34ml/s with 25.6psi vacuum @04:24:47.71 Cartridge 11 sampling 1962/2000ml at 0.33ml/s with 25.8psi vacuum @04:25:47.92 Cartridge 11 sampling 1982/2000ml at 0.33ml/s with 24.9psi vacuum @04:26:41.91 Pressure.coast @04:26:42.21 Intake.close @04:26:42.33 Exhaust.close @04:26:42.70 Toroid.reconfigure :hold @04:26:42.84 Toroid.seek :bypass @04:26:44.46 Toroid.reconfigure :default @04:26:44.52 Toroid.seek :clear @04:26:46.14 Sampled 2000.0ml @04:26:46.59 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 11" @04:26:47.00 SamplePump.setPosition! 0ml @04:26:47.05 Plunge.seek :reagent @04:26:47.09 Bypass.open @04:26:47.84 Intake.open @04:26:48.05 Sample loop closed. Ready to Deliver Reagents @04:26:48.09 Archive.seek :sealed @04:26:49.11 SamplePump.reconfigure :slow @04:26:49.34 SamplePump.seek -3ml @04:26:51.84 Archive.seek :evac @04:26:56.03 delay 1 second @04:26:57.15 SamplePump.seek -2.8ml @04:26:57.95 delay 1 second @04:26:59.17 SamplePump.seek -3.3ml @04:27:00.68 delay 1 second @04:27:01.72 Exhaust.open @04:27:02.20 upsync @04:27:02.43 Bypass.close @04:27:03.31 SamplePump.seek -35.3ml @04:27:05.84 delay 1:00 @04:28:05.89 Archive.seek :clear @04:28:12.15 delay 3 seconds @04:28:15.41 Intake.close @04:28:15.47 Exhaust.close @04:28:23.10 RNAlater soak for 10:00 @04:38:23.15 Archive.seek :tVent @04:38:24.80 delay 1 second @04:38:25.84 Archive.seek :evac @04:38:41.87 delay 10 seconds @04:38:51.91 Sample preserved and evacuated @04:38:52.14
Carousel.seek 11 @04:38:52.35 Plunge.seek :locked @04:38:53.33 Archive.seek :clear @04:38:54.22 Air.seek :clear @04:38:54.57 Lysis1.seek :clear @04:39:10.48
Archive.coast @04:39:10.52 Air.coast @04:39:10.56 Lysis1.coast @04:39:10.61 Lysis2.coast @04:39:10.65 Plunge.coast @04:39:10.69 SamplePump.coast @04:39:10.74 Toroid.coast @04:39:10.78 Carousel.coast @04:39:10.97 Gate.power :motor,:OFF @04:39:10.98 Gate.power :core,:OFF @04:39:11.00 Gate.power :analytic1,:OFF @04:39:11.01 Gate.power :analytic2,:OFF @04:39:11.02 Gate.power :analytic3,:OFF @04:39:11.04 Gate.power :main,:OFF @04:39:11.05 Safely stopped and ready to power off @04:39:11.27 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 15:00" @04:39:11.30
delayUntil "15:00:00.00" @04:39:28.72 upsync @15:00:00.21
Gate.power :main,:ON,limit to 5000mA @15:00:00.24 Gate.power :analytic1,:ON,limit to 5000mA @15:00:00.30 Gate.power :analytic2,:ON,limit to 5000mA @15:00:00.31 Gate.power :analytic3,:ON,limit to 5000mA @15:00:00.35 Powering core bus @15:00:02.72 Archive.configure :default @15:00:02.86 Air.configure :default @15:00:02.98 Lysis1.configure :default @15:00:03.09 Lysis2.configure :default @15:00:03.23 Plunge.configure :default @15:00:03.37 SamplePump.configure :default @15:00:03.50 ExtraPump.configure :default @15:00:03.64 Toroid.configure :default @15:00:03.78 Carousel.configure :default @15:00:03.91 Pressure.configure :default @15:00:03.97 All elf controllers running firmware version 4.61 @15:00:03.99 Gate.power :motor,:ON,limit to 5000mA @15:00:04.28 Archive.reconfigure :default @15:00:04.50 Archive.home.jog 0 @15:00:04.68 Air.reconfigure :default @15:00:04.74 Toroid.reconfigure :default @15:00:04.77 Plunge.reconfigure :default @15:00:04.81 Lysis1.reconfigure :default @15:00:04.89 Lysis2.reconfigure :default @15:00:05.07 Toroid.home.jog 0 @15:00:05.19 Plunge.home.jog 0 @15:00:05.26 Lysis2.home.jog 0 @15:00:05.28 Lysis1.home.jog 0 @15:00:05.30 Air.home.jog 0 @15:00:05.85 Archive.reconfigure :homing @15:00:06.15 Archive.jog 1370 @15:00:06.73 Toroid.reconfigure :homing @15:00:06.83 Lysis2.reconfigure :homing @15:00:07.14 Lysis1.reconfigure :homing @15:00:07.26 Plunge.reconfigure :homing @15:00:07.64 Toroid.jog 636 @15:00:07.69 Lysis1.jog 1358 @15:00:07.79 Lysis2.jog 1344 @15:00:07.83 Plunge.jog 843 @15:00:08.27 Archive.setRawPosition! 50 @15:00:08.44 Air.reconfigure :homing @15:00:08.61 Archive.seek :home @15:00:08.71 Air.jog 1359 @15:00:09.22 Lysis2.setRawPosition! 50 @15:00:09.32 Lysis1.setRawPosition! 50 @15:00:09.34 Plunge.setRawPosition! 50 @15:00:09.51 Archive.reconfigure :default @15:00:09.69 Lysis2.seek :home @15:00:09.86 Lysis1.seek :home @15:00:09.92 Plunge.seek :home @15:00:10.53 Toroid.setRawPosition! 0 @15:00:10.71 Lysis2.reconfigure :default @15:00:10.76 Air.setRawPosition! 50 @15:00:10.78 Plunge.reconfigure :default @15:00:10.82 Lysis1.reconfigure :default @15:00:11.13 Toroid.seek :home @15:00:11.20 Air.seek :home @15:00:12.02 Air.reconfigure :default @15:00:12.04 Toroid.reconfigure :default @15:00:12.46
Toroid.seek :clear @15:00:14.06 Carousel.reconfigure :default @15:00:14.12 Carousel.home.jog 0 @15:00:14.61 Carousel.seek 11 @15:00:14.79 Selecting Cartridge 10 @15:00:15.00 Carousel.seek 10 @15:00:15.52 Lysis1.seek :locked @15:00:19.54 Plunge.seek :locked @15:00:19.84 Air.seek :locked @15:00:24.07
Mounted archive Cartridge @15:00:24.21 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 10" @15:00:24.49 Duration of filtering limited to 4:05:00 @15:00:24.78 Priming sample loop w/160ml, bypass w/40ml @15:00:24.80 SamplePump.setPosition! 0ml @15:00:24.84 Bypass.open @15:00:26.54 Intake.open @15:00:27.77 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @15:00:30.66 Vacuum tare is 18% of full scale @15:00:30.72 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @15:00:30.94 Pressure.reconfigure :default @15:00:31.00 Pressure.hold @15:00:31.31 SamplePump.reconfigure :prime @15:00:31.37 SamplePump.seek 40ml,3:51 @15:00:46.43 upsync @15:01:14.64 Pressure.coast @15:01:14.76 SamplePump.setPosition! 0ml @15:01:14.84 Exhaust.open @15:01:16.50 Bypass.close @15:01:18.25 Pressure.reconfigure :default @15:01:18.31 Pressure.hold @15:01:18.59 SamplePump.reconfigure :prime @15:01:18.65 SamplePump.seek 160ml,13:51 @15:04:10.44 Pressure.coast @15:04:10.51 Bypass.open @15:04:11.24 Intake.close @15:04:12.12 Toroid.reconfigure :hold @15:04:12.18 Toroid.seek :crack @15:04:13.76 Toroid.seek :cartridge @15:04:14.70 Toroid.reconfigure :default @15:04:14.76 Toroid.seek :clear @15:04:17.03 Pressurizing cartridge to 26psi above 14.6psia @15:04:17.08 SamplePump.setPosition! 0ml @15:04:17.12 Intake.open @15:04:17.97 Bypass.close @15:04:18.74 Exhaust.close @15:04:19.78 Pressure.reconfigure :default @15:04:19.85 Pressure.hold @15:04:20.60 SamplePump.reconfigure :charge @15:04:20.92 SamplePump.seek 2.3ml,42 seconds @15:04:33.68 Pressure.coast @15:04:33.73 Intake.close @15:04:34.93 Pressurized cartridge with 0.739ml to 26.2psi @15:04:34.95 Isolated @36.7psia (22.1psi) @15:04:34.97 Checking for cartridge leaks from 36.7psia @15:04:35.10 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @15:04:41.71 Cartridge pressure fell 8% to 33.8psia @15:04:41.79 Sampling 2000.0ml @15:04:41.80 SamplePump.setPosition! 0ml @15:04:41.85 Exhaust.open @15:04:42.54 Intake.open @15:04:43.47 Pressure.reconfigure :default @15:04:43.53 Pressure.hold @15:04:43.67 Sea 1.6m deep (17.0psia) @15:04:43.84 SamplePump.reconfigure :sample @15:04:43.91 SamplePump.seek 2000ml,2:47:12 @15:04:46.72 Cartridge 10 sampling 1/2000ml at 0.56ml/s with 25.9psi vacuum @15:05:46.54 Cartridge 10 sampling 33/2000ml at 0.53ml/s with 24.9psi vacuum @15:06:46.68 Cartridge 10 sampling 64/2000ml at 0.52ml/s with 25.0psi vacuum @15:07:46.54 Cartridge 10 sampling 95/2000ml at 0.52ml/s with 24.1psi vacuum @15:08:46.63 Cartridge 10 sampling 127/2000ml at 0.52ml/s with 23.8psi vacuum @15:09:46.53 Cartridge 10 sampling 159/2000ml at 0.54ml/s with 23.0psi vacuum @15:10:46.84 Cartridge 10 sampling 191/2000ml at 0.54ml/s with 24.3psi vacuum @15:11:47.13 Cartridge 10 sampling 224/2000ml at 0.54ml/s with 21.9psi vacuum @15:12:46.55 Cartridge 10 sampling 256/2000ml at 0.53ml/s with 23.3psi vacuum @15:13:46.86 Cartridge 10 sampling 288/2000ml at 0.54ml/s with 22.9psi vacuum @15:14:46.54 Cartridge 10 sampling 321/2000ml at 0.54ml/s with 24.7psi vacuum @15:15:46.74 Cartridge 10 sampling 353/2000ml at 0.54ml/s with 21.2psi vacuum @15:16:46.54 Cartridge 10 sampling 386/2000ml at 0.54ml/s with 24.0psi vacuum @15:17:46.72 Cartridge 10 sampling 418/2000ml at 0.53ml/s with 25.5psi vacuum @15:18:46.54 Cartridge 10 sampling 450/2000ml at 0.54ml/s with 24.0psi vacuum @15:19:46.69 Cartridge 10 sampling 483/2000ml at 0.54ml/s with 24.3psi vacuum @15:20:46.54 Cartridge 10 sampling 515/2000ml at 0.54ml/s with 24.6psi vacuum @15:21:46.73 Cartridge 10 sampling 547/2000ml at 0.54ml/s with 24.6psi vacuum @15:22:46.54 Cartridge 10 sampling 580/2000ml at 0.52ml/s with 21.8psi vacuum @15:23:46.65 Cartridge 10 sampling 612/2000ml at 0.53ml/s with 26.0psi vacuum @15:24:46.54 Cartridge 10 sampling 644/2000ml at 0.54ml/s with 24.0psi vacuum @15:25:46.77 Cartridge 10 sampling 676/2000ml at 0.53ml/s with 24.8psi vacuum @15:26:46.70 Cartridge 10 sampling 708/2000ml at 0.53ml/s with 24.0psi vacuum @15:27:46.53 Cartridge 10 sampling 740/2000ml at 0.52ml/s with 24.8psi vacuum @15:28:46.73 Cartridge 10 sampling 772/2000ml at 0.53ml/s with 25.2psi vacuum @15:29:46.53 Cartridge 10 sampling 803/2000ml at 0.53ml/s with 23.8psi vacuum @15:30:46.71 Cartridge 10 sampling 835/2000ml at 0.52ml/s with 22.8psi vacuum @15:31:46.52 Cartridge 10 sampling 866/2000ml at 0.53ml/s with 25.5psi vacuum @15:32:46.74 Cartridge 10 sampling 898/2000ml at 0.52ml/s with 26.5psi vacuum @15:33:46.53 Cartridge 10 sampling 929/2000ml at 0.51ml/s with 22.7psi vacuum @15:34:46.79 Cartridge 10 sampling 961/2000ml at 0.52ml/s with 23.8psi vacuum @15:35:46.53 Cartridge 10 sampling 992/2000ml at 0.52ml/s with 20.7psi vacuum @15:36:46.59 Cartridge 10 sampling 1023/2000ml at 0.50ml/s with 22.5psi vacuum @15:37:46.52 Cartridge 10 sampling 1054/2000ml at 0.50ml/s with 24.9psi vacuum @15:38:46.58 Cartridge 10 sampling 1085/2000ml at 0.50ml/s with 24.0psi vacuum @15:39:46.52 Cartridge 10 sampling 1115/2000ml at 0.51ml/s with 24.6psi vacuum @15:40:46.54 Cartridge 10 sampling 1146/2000ml at 0.51ml/s with 24.8psi vacuum @15:41:46.80 Cartridge 10 sampling 1176/2000ml at 0.51ml/s with 24.0psi vacuum @15:42:46.54 Cartridge 10 sampling 1206/2000ml at 0.50ml/s with 23.5psi vacuum @15:43:46.77 Cartridge 10 sampling 1236/2000ml at 0.49ml/s with 23.8psi vacuum @15:44:46.53 Cartridge 10 sampling 1266/2000ml at 0.50ml/s with 24.5psi vacuum @15:45:46.65 Cartridge 10 sampling 1296/2000ml at 0.50ml/s with 24.9psi vacuum @15:46:46.54 Cartridge 10 sampling 1326/2000ml at 0.49ml/s with 21.4psi vacuum @15:47:46.79 Cartridge 10 sampling 1356/2000ml at 0.50ml/s with 23.1psi vacuum @15:48:46.53 Cartridge 10 sampling 1385/2000ml at 0.49ml/s with 23.8psi vacuum @15:49:46.70 Cartridge 10 sampling 1415/2000ml at 0.49ml/s with 26.2psi vacuum @15:50:46.54 Cartridge 10 sampling 1444/2000ml at 0.49ml/s with 25.0psi vacuum @15:51:46.68 Cartridge 10 sampling 1473/2000ml at 0.49ml/s with 20.8psi vacuum @15:52:46.54 Cartridge 10 sampling 1502/2000ml at 0.48ml/s with 26.6psi vacuum @15:53:46.89 Cartridge 10 sampling 1531/2000ml at 0.48ml/s with 24.5psi vacuum @15:54:47.18 Cartridge 10 sampling 1560/2000ml at 0.48ml/s with 24.4psi vacuum @15:55:46.64 Cartridge 10 sampling 1589/2000ml at 0.47ml/s with 20.9psi vacuum @15:56:46.79 Cartridge 10 sampling 1617/2000ml at 0.47ml/s with 21.2psi vacuum @15:57:46.54 Cartridge 10 sampling 1646/2000ml at 0.48ml/s with 25.3psi vacuum @15:58:46.63 Cartridge 10 sampling 1674/2000ml at 0.48ml/s with 24.0psi vacuum @15:59:46.53 Cartridge 10 sampling 1702/2000ml at 0.46ml/s with 21.8psi vacuum @16:00:46.60 Cartridge 10 sampling 1730/2000ml at 0.47ml/s with 23.8psi vacuum @16:01:46.53 Cartridge 10 sampling 1758/2000ml at 0.46ml/s with 24.3psi vacuum @16:02:46.64 Cartridge 10 sampling 1786/2000ml at 0.47ml/s with 24.0psi vacuum @16:03:46.53 Cartridge 10 sampling 1813/2000ml at 0.44ml/s with 26.6psi vacuum @16:04:46.73 Cartridge 10 sampling 1841/2000ml at 0.45ml/s with 22.5psi vacuum @16:05:46.54 Cartridge 10 sampling 1868/2000ml at 0.45ml/s with 26.3psi vacuum @16:06:46.57 Cartridge 10 sampling 1895/2000ml at 0.45ml/s with 23.4psi vacuum @16:07:46.54 Cartridge 10 sampling 1922/2000ml at 0.45ml/s with 26.0psi vacuum @16:08:46.74 Cartridge 10 sampling 1949/2000ml at 0.44ml/s with 23.3psi vacuum @16:09:46.77 Cartridge 10 sampling 1976/2000ml at 0.44ml/s with 23.0psi vacuum @16:10:42.00 Pressure.coast @16:10:42.40 Intake.close @16:10:42.56 Exhaust.close @16:10:42.99 Toroid.reconfigure :hold @16:10:43.11 Toroid.seek :bypass @16:10:44.82 Toroid.reconfigure :default @16:10:44.92 Toroid.seek :clear @16:10:46.61 Sampled 2000.0ml @16:10:47.30 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 10" @16:10:47.66 SamplePump.setPosition! 0ml @16:10:47.71 Plunge.seek :reagent @16:10:47.74 Bypass.open @16:10:48.48 Intake.open @16:10:48.69 Sample loop closed. Ready to Deliver Reagents @16:10:48.71 Archive.seek :sealed @16:10:49.37 SamplePump.reconfigure :slow @16:10:49.43 SamplePump.seek -3ml @16:10:52.37 Archive.seek :evac @16:10:56.11 delay 1 second @16:10:57.19 SamplePump.seek -2.8ml @16:10:57.89 delay 1 second @16:10:58.99 SamplePump.seek -3.3ml @16:11:00.32 delay 1 second @16:11:01.35 Exhaust.open @16:11:02.17 Bypass.close @16:11:03.18 SamplePump.seek -35.3ml @16:11:06.43 delay 1:00 @16:11:17.42 upsync @16:12:06.48 Archive.seek :clear @16:12:12.01 delay 3 seconds @16:12:15.27 Intake.close @16:12:15.32 Exhaust.close @16:12:23.70 RNAlater soak for 10:00 @16:22:23.75 Archive.seek :tVent @16:22:25.40 delay 1 second @16:22:26.44 Archive.seek :evac @16:22:42.46 delay 10 seconds @16:22:52.56 Sample preserved and evacuated @16:22:52.80
Carousel.seek 10 @16:22:53.00 Plunge.seek :locked @16:22:53.98 Archive.seek :clear @16:22:54.15 Air.seek :clear @16:22:54.30 Lysis1.seek :clear @16:23:11.18
Archive.coast @16:23:11.22 Air.coast @16:23:11.27 Lysis1.coast @16:23:11.31 Lysis2.coast @16:23:11.35 Plunge.coast @16:23:11.39 SamplePump.coast @16:23:11.44 Toroid.coast @16:23:11.48 Carousel.coast @16:23:11.69 Gate.power :motor,:OFF @16:23:11.70 Gate.power :core,:OFF @16:23:11.71 Gate.power :analytic1,:OFF @16:23:11.73 Gate.power :analytic2,:OFF @16:23:11.74 Gate.power :analytic3,:OFF @16:23:11.76 Gate.power :main,:OFF @16:23:11.77 Safely stopped and ready to power off @16:23:12.00 delayUntil "03:00:00.00PDT02-Jul-21" @16:23:12.03 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 03:00PDT02-Jul-21" @16:23:29.97 upsync @03:00:00.11PDT02-Jul-21
Gate.power :main,:ON,limit to 5000mA @03:00:00.12 Gate.power :analytic1,:ON,limit to 5000mA @03:00:00.14 Gate.power :analytic2,:ON,limit to 5000mA @03:00:00.15 Gate.power :analytic3,:ON,limit to 5000mA @03:00:00.19 Powering core bus @03:00:02.47 Archive.configure :default @03:00:02.60 Air.configure :default @03:00:03.29 Lysis1.configure :default @03:00:03.40 Lysis2.configure :default @03:00:03.54 Plunge.configure :default @03:00:03.67 SamplePump.configure :default @03:00:03.81 ExtraPump.configure :default @03:00:03.96 Toroid.configure :default @03:00:04.09 Carousel.configure :default @03:00:04.43 Pressure.configure :default @03:00:04.58 All elf controllers running firmware version 4.61 @03:00:04.62 Gate.power :motor,:ON,limit to 5000mA @03:00:06.12 Lysis2.reconfigure :default @03:00:06.18 Lysis1.reconfigure :default @03:00:06.24 Air.reconfigure :default @03:00:06.36 Plunge.reconfigure :default @03:00:06.45 Archive.reconfigure :default @03:00:06.51 Toroid.reconfigure :default @03:00:06.88 Lysis2.home.jog 0 @03:00:07.08 Lysis1.home.jog 0 @03:00:07.18 Plunge.home.jog 0 @03:00:07.29 Air.home.jog 0 @03:00:07.35 Toroid.home.jog 0 @03:00:07.47 Archive.home.jog 0 @03:00:08.89 Lysis2.reconfigure :homing @03:00:09.09 Toroid.reconfigure :homing @03:00:09.21 Air.reconfigure :homing @03:00:09.31 Archive.reconfigure :homing @03:00:09.45 Plunge.reconfigure :homing @03:00:09.65 Lysis1.reconfigure :homing @03:00:10.09 Lysis2.jog 1346 @03:00:10.18 Toroid.jog 644 @03:00:10.26 Archive.jog 1364 @03:00:10.37 Air.jog 1354 @03:00:10.40 Lysis1.jog 1359 @03:00:10.51 Plunge.jog 840 @03:00:11.68 Lysis2.setRawPosition! 50 @03:00:11.73 Lysis2.seek :home @03:00:11.82 Archive.setRawPosition! 50 @03:00:11.87 Archive.seek :home @03:00:11.95 Plunge.setRawPosition! 50 @03:00:12.02 Plunge.seek :home @03:00:12.15 Lysis2.reconfigure :default @03:00:12.18 Lysis1.setRawPosition! 50 @03:00:12.22 Air.setRawPosition! 50 @03:00:12.25 Archive.reconfigure :default @03:00:12.41 Lysis1.seek :home @03:00:12.45 Air.seek :home @03:00:12.61 Plunge.reconfigure :default @03:00:12.77 Lysis1.reconfigure :default @03:00:12.87 Air.reconfigure :default @03:00:12.96 Toroid.setRawPosition! 0 @03:00:13.01 Toroid.seek :home @03:00:13.11 Toroid.reconfigure :default @03:00:13.31
Toroid.seek :clear @03:00:14.92 Carousel.reconfigure :default @03:00:14.97 Carousel.home.jog 0 @03:00:15.78 Carousel.seek 10 @03:00:16.12 Selecting Cartridge 9 @03:00:16.69 Carousel.seek 9 @03:00:17.47 Lysis1.seek :locked @03:00:20.72 Plunge.seek :locked @03:00:20.86 Air.seek :locked @03:00:25.10
Mounted archive Cartridge @03:00:25.54 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 9" @03:00:26.40 Duration of filtering limited to 4:05:00 @03:00:27.04 Priming sample loop w/160ml, bypass w/40ml @03:00:27.06 SamplePump.setPosition! 0ml @03:00:27.12 Bypass.open @03:00:28.82 Intake.open @03:00:29.81 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @03:00:32.42 Vacuum tare is 18% of full scale @03:00:32.44 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @03:00:32.65 Pressure.reconfigure :default @03:00:32.70 Pressure.hold @03:00:32.91 SamplePump.reconfigure :prime @03:00:33.50 SamplePump.seek 40ml,3:51 @03:00:40.08 upsync @03:01:16.74 Pressure.coast @03:01:16.80 SamplePump.setPosition! 0ml @03:01:16.86 Exhaust.open @03:01:18.26 Bypass.close @03:01:19.85 Pressure.reconfigure :default @03:01:19.91 Pressure.hold @03:01:20.18 SamplePump.reconfigure :prime @03:01:20.25 SamplePump.seek 160ml,13:51 @03:04:11.85 Pressure.coast @03:04:11.91 Bypass.open @03:04:12.62 Intake.close @03:04:13.48 Toroid.reconfigure :hold @03:04:13.54 Toroid.seek :crack @03:04:15.30 Toroid.seek :cartridge @03:04:16.41 Toroid.reconfigure :default @03:04:16.52 Toroid.seek :clear @03:04:18.72 Pressurizing cartridge to 26psi above 14.8psia @03:04:18.83 SamplePump.setPosition! 0ml @03:04:18.87 Intake.open @03:04:19.66 Bypass.close @03:04:20.46 Exhaust.close @03:04:21.24 Pressure.reconfigure :default @03:04:21.30 Pressure.hold @03:04:21.51 SamplePump.reconfigure :charge @03:04:21.57 SamplePump.seek 2.3ml,42 seconds @03:04:34.92 Pressure.coast @03:04:35.09 Intake.close @03:04:36.22 Pressurized cartridge with 0.804ml to 26.9psi @03:04:36.26 Isolated @36.5psia (21.7psi) @03:04:36.30 Checking for cartridge leaks from 36.5psia @03:04:36.34 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @03:04:44.02 Cartridge pressure fell 7% to 33.9psia @03:04:44.35 Sampling 2000.0ml @03:04:44.40 SamplePump.setPosition! 0ml @03:04:44.51 Exhaust.open @03:04:45.33 Intake.open @03:04:46.55 Pressure.reconfigure :default @03:04:46.70 Pressure.hold @03:04:47.06 Sea 2.8m deep (18.9psia) @03:04:47.37 SamplePump.reconfigure :sample @03:04:47.57 SamplePump.seek 2000ml,2:47:12 @03:04:49.43 Cartridge 9 sampling 1/2000ml at 0.63ml/s with 21.2psi vacuum @03:05:49.60 Cartridge 9 sampling 27/2000ml at 0.43ml/s with 24.6psi vacuum @03:06:49.44 Cartridge 9 sampling 53/2000ml at 0.42ml/s with 24.6psi vacuum @03:07:49.60 Cartridge 9 sampling 79/2000ml at 0.43ml/s with 25.4psi vacuum @03:08:49.44 Cartridge 9 sampling 105/2000ml at 0.42ml/s with 24.2psi vacuum @03:09:49.61 Cartridge 9 sampling 130/2000ml at 0.41ml/s with 23.9psi vacuum @03:10:49.43 Cartridge 9 sampling 155/2000ml at 0.43ml/s with 22.9psi vacuum @03:11:49.61 Cartridge 9 sampling 181/2000ml at 0.43ml/s with 24.3psi vacuum @03:12:49.43 Cartridge 9 sampling 207/2000ml at 0.42ml/s with 24.8psi vacuum @03:13:49.52 Cartridge 9 sampling 233/2000ml at 0.44ml/s with 20.9psi vacuum @03:14:49.44 Cartridge 9 sampling 258/2000ml at 0.43ml/s with 24.9psi vacuum @03:15:49.71 Cartridge 9 sampling 284/2000ml at 0.42ml/s with 24.8psi vacuum @03:16:49.69 Cartridge 9 sampling 309/2000ml at 0.42ml/s with 21.0psi vacuum @03:17:49.44 Cartridge 9 sampling 335/2000ml at 0.42ml/s with 25.2psi vacuum @03:18:49.90 Cartridge 9 sampling 360/2000ml at 0.43ml/s with 25.4psi vacuum @03:19:49.45 Cartridge 9 sampling 386/2000ml at 0.42ml/s with 26.0psi vacuum @03:20:49.67 Cartridge 9 sampling 411/2000ml at 0.44ml/s with 24.7psi vacuum @03:21:49.45 Cartridge 9 sampling 437/2000ml at 0.42ml/s with 23.8psi vacuum @03:22:49.68 Cartridge 9 sampling 462/2000ml at 0.41ml/s with 22.4psi vacuum @03:23:49.45 Cartridge 9 sampling 487/2000ml at 0.42ml/s with 21.3psi vacuum @03:24:49.59 Cartridge 9 sampling 512/2000ml at 0.41ml/s with 22.7psi vacuum @03:25:49.45 Cartridge 9 sampling 536/2000ml at 0.41ml/s with 23.6psi vacuum @03:26:49.60 Cartridge 9 sampling 561/2000ml at 0.40ml/s with 24.9psi vacuum @03:27:49.45 Cartridge 9 sampling 585/2000ml at 0.39ml/s with 25.4psi vacuum @03:28:49.49 Cartridge 9 sampling 609/2000ml at 0.40ml/s with 23.3psi vacuum @03:29:49.45 Cartridge 9 sampling 633/2000ml at 0.40ml/s with 24.3psi vacuum @03:30:49.64 Cartridge 9 sampling 657/2000ml at 0.40ml/s with 24.5psi vacuum @03:31:49.67 Cartridge 9 sampling 681/2000ml at 0.40ml/s with 24.7psi vacuum @03:32:49.45 Cartridge 9 sampling 704/2000ml at 0.38ml/s with 23.6psi vacuum @03:33:49.63 Cartridge 9 sampling 728/2000ml at 0.39ml/s with 23.9psi vacuum @03:34:49.45 Cartridge 9 sampling 751/2000ml at 0.39ml/s with 25.2psi vacuum @03:35:49.75 Cartridge 9 sampling 775/2000ml at 0.39ml/s with 25.3psi vacuum @03:36:49.44 Cartridge 9 sampling 798/2000ml at 0.38ml/s with 25.4psi vacuum @03:37:49.67 Cartridge 9 sampling 822/2000ml at 0.40ml/s with 25.5psi vacuum @03:38:49.45 Cartridge 9 sampling 846/2000ml at 0.40ml/s with 23.3psi vacuum @03:39:49.89 Cartridge 9 sampling 869/2000ml at 0.39ml/s with 22.9psi vacuum @03:40:49.45 Cartridge 9 sampling 892/2000ml at 0.39ml/s with 24.1psi vacuum @03:41:49.55 Cartridge 9 sampling 916/2000ml at 0.38ml/s with 25.5psi vacuum @03:42:49.45 Cartridge 9 sampling 939/2000ml at 0.38ml/s with 24.0psi vacuum @03:43:50.52 Cartridge 9 sampling 961/2000ml at 0.38ml/s with 25.2psi vacuum @03:44:49.66 Cartridge 9 sampling 984/2000ml at 0.37ml/s with 25.5psi vacuum @03:45:49.45 Cartridge 9 sampling 1007/2000ml at 0.37ml/s with 23.8psi vacuum @03:46:49.67 Cartridge 9 sampling 1029/2000ml at 0.37ml/s with 24.9psi vacuum @03:47:49.45 Cartridge 9 sampling 1052/2000ml at 0.38ml/s with 25.5psi vacuum @03:48:49.57 Cartridge 9 sampling 1074/2000ml at 0.37ml/s with 19.0psi vacuum @03:49:49.45 Cartridge 9 sampling 1097/2000ml at 0.37ml/s with 19.5psi vacuum @03:50:49.61 Cartridge 9 sampling 1119/2000ml at 0.37ml/s with 24.6psi vacuum @03:51:49.45 Cartridge 9 sampling 1141/2000ml at 0.37ml/s with 23.0psi vacuum @03:52:49.71 Cartridge 9 sampling 1163/2000ml at 0.36ml/s with 24.9psi vacuum @03:53:49.45 Cartridge 9 sampling 1185/2000ml at 0.37ml/s with 24.4psi vacuum @03:54:49.63 Cartridge 9 sampling 1207/2000ml at 0.37ml/s with 25.1psi vacuum @03:55:49.45 Cartridge 9 sampling 1229/2000ml at 0.36ml/s with 23.2psi vacuum @03:56:49.52 Cartridge 9 sampling 1250/2000ml at 0.36ml/s with 24.8psi vacuum @03:57:49.45 Cartridge 9 sampling 1272/2000ml at 0.36ml/s with 23.7psi vacuum @03:58:49.70 Cartridge 9 sampling 1293/2000ml at 0.35ml/s with 25.7psi vacuum @03:59:49.81 Cartridge 9 sampling 1314/2000ml at 0.35ml/s with 23.2psi vacuum @04:00:49.45 Cartridge 9 sampling 1335/2000ml at 0.34ml/s with 25.4psi vacuum @04:01:49.81 Cartridge 9 sampling 1357/2000ml at 0.34ml/s with 25.9psi vacuum @04:02:49.45 Cartridge 9 sampling 1378/2000ml at 0.35ml/s with 25.1psi vacuum @04:03:49.61 Cartridge 9 sampling 1399/2000ml at 0.34ml/s with 24.8psi vacuum @04:04:49.45 Cartridge 9 sampling 1420/2000ml at 0.35ml/s with 24.9psi vacuum @04:05:49.60 Cartridge 9 sampling 1441/2000ml at 0.34ml/s with 23.5psi vacuum @04:06:49.44 Cartridge 9 sampling 1461/2000ml at 0.34ml/s with 25.6psi vacuum @04:07:49.77 Cartridge 9 sampling 1482/2000ml at 0.33ml/s with 25.8psi vacuum @04:08:49.45 Cartridge 9 sampling 1502/2000ml at 0.36ml/s with 24.1psi vacuum @04:09:49.55 Cartridge 9 sampling 1523/2000ml at 0.34ml/s with 23.5psi vacuum @04:10:49.45 Cartridge 9 sampling 1543/2000ml at 0.32ml/s with 24.9psi vacuum @04:11:49.49 Cartridge 9 sampling 1563/2000ml at 0.32ml/s with 25.8psi vacuum @04:12:49.44 Cartridge 9 sampling 1583/2000ml at 0.34ml/s with 24.6psi vacuum @04:13:49.58 Cartridge 9 sampling 1603/2000ml at 0.33ml/s with 25.0psi vacuum @04:14:49.70 Cartridge 9 sampling 1622/2000ml at 0.33ml/s with 21.3psi vacuum @04:15:49.44 Cartridge 9 sampling 1642/2000ml at 0.33ml/s with 21.0psi vacuum @04:16:49.75 Cartridge 9 sampling 1661/2000ml at 0.32ml/s with 25.7psi vacuum @04:17:49.45 Cartridge 9 sampling 1680/2000ml at 0.32ml/s with 24.4psi vacuum @04:18:49.51 Cartridge 9 sampling 1700/2000ml at 0.32ml/s with 23.0psi vacuum @04:19:49.44 Cartridge 9 sampling 1719/2000ml at 0.33ml/s with 24.1psi vacuum @04:20:49.54 Cartridge 9 sampling 1738/2000ml at 0.32ml/s with 25.8psi vacuum @04:21:49.44 Cartridge 9 sampling 1757/2000ml at 0.32ml/s with 24.8psi vacuum @04:22:49.61 Cartridge 9 sampling 1776/2000ml at 0.32ml/s with 25.1psi vacuum @04:23:49.44 Cartridge 9 sampling 1795/2000ml at 0.32ml/s with 24.4psi vacuum @04:24:49.60 Cartridge 9 sampling 1814/2000ml at 0.31ml/s with 19.8psi vacuum @04:25:49.44 Cartridge 9 sampling 1832/2000ml at 0.31ml/s with 24.3psi vacuum @04:26:49.57 Cartridge 9 sampling 1851/2000ml at 0.30ml/s with 25.2psi vacuum @04:27:49.44 Cartridge 9 sampling 1870/2000ml at 0.31ml/s with 25.7psi vacuum @04:28:49.60 Cartridge 9 sampling 1888/2000ml at 0.29ml/s with 25.9psi vacuum @04:29:49.66 Cartridge 9 sampling 1906/2000ml at 0.30ml/s with 23.8psi vacuum @04:30:49.45 Cartridge 9 sampling 1924/2000ml at 0.31ml/s with 24.1psi vacuum @04:31:49.62 Cartridge 9 sampling 1942/2000ml at 0.30ml/s with 24.5psi vacuum @04:32:49.44 Cartridge 9 sampling 1960/2000ml at 0.30ml/s with 23.5psi vacuum @04:33:49.60 Cartridge 9 sampling 1978/2000ml at 0.28ml/s with 26.0psi vacuum @04:34:49.45 Cartridge 9 sampling 1996/2000ml at 0.30ml/s with 25.4psi vacuum @04:35:03.89 Pressure.coast @04:35:04.19 Intake.close @04:35:04.31 Exhaust.close @04:35:04.76 Toroid.reconfigure :hold @04:35:04.85 Toroid.seek :bypass @04:35:06.54 Toroid.reconfigure :default @04:35:06.60 Toroid.seek :clear @04:35:08.21 Sampled 2000.0ml @04:35:08.64 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 9" @04:35:09.07 SamplePump.setPosition! 0ml @04:35:09.13 Plunge.seek :reagent @04:35:09.15 Bypass.open @04:35:09.93 Intake.open @04:35:10.07 Sample loop closed. Ready to Deliver Reagents @04:35:10.09 Archive.seek :sealed @04:35:11.19 SamplePump.reconfigure :slow @04:35:11.37 SamplePump.seek -3ml @04:35:13.88 Archive.seek :evac @04:35:18.09 delay 1 second @04:35:19.90 SamplePump.seek -2.8ml @04:35:20.67 delay 1 second @04:35:21.97 SamplePump.seek -3.3ml @04:35:23.31 delay 1 second @04:35:24.34 Exhaust.open @04:35:24.52 upsync @04:35:25.06 Bypass.close @04:35:25.83 SamplePump.seek -35.3ml @04:35:27.90 delay 1:00 @04:36:27.94 Archive.seek :clear @04:36:34.55 delay 3 seconds @04:36:37.71 Intake.close @04:36:37.75 Exhaust.close @04:36:45.07 RNAlater soak for 10:00 @04:46:45.11 Archive.seek :tVent @04:46:46.76 delay 1 second @04:46:47.80 Archive.seek :evac @04:47:03.76 delay 10 seconds @04:47:13.80 Sample preserved and evacuated @04:47:14.03
Carousel.seek 9 @04:47:14.23 Plunge.seek :locked @04:47:15.21 Archive.seek :clear @04:47:15.36 Air.seek :clear @04:47:15.52 Lysis1.seek :clear @04:47:32.41
Archive.coast @04:47:32.45 Air.coast @04:47:32.49 Lysis1.coast @04:47:32.54 Lysis2.coast @04:47:32.58 Plunge.coast @04:47:32.62 SamplePump.coast @04:47:32.66 Toroid.coast @04:47:32.71 Carousel.coast @04:47:32.90 Gate.power :motor,:OFF @04:47:32.91 Gate.power :core,:OFF @04:47:32.93 Gate.power :analytic1,:OFF @04:47:32.94 Gate.power :analytic2,:OFF @04:47:32.95 Gate.power :analytic3,:OFF @04:47:32.97 Gate.power :main,:OFF @04:47:32.98 Safely stopped and ready to power off @04:47:33.40 Cartridge Load: {53 => { Type: :archive_bac, State: :processed, filtered: 1396.8}, 56 => { Type: :archive_bac, State: :processed, filtered: 387.7}, [52..47, 45..21] => { Type: :archive_bac, State: :processed, filtered: 2000}, [46, 14] => { Type: :archive_bac, State: :leaky}, [59, 58] => { Type: :archive_bac, State: :processed, filtered: 400}, 8..1 => { Type: :archive_bac, State: :reserved, was: :dry}, [20..15, 13..9] => { Type: :archive_bac, State: :processed, filtered: 2000.0}, 54 => { Type: :archive_bac, State: :processed, filtered: 721.3}, 55 => { Type: :archive_bac, State: :processed, filtered: 622.6}, 57 => { Type: :archive_bac, State: :processed, filtered: 1000}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 0} @04:47:33.43 BadNews.email "Not enough dry cartridges remain",Subject: "Cartridge::None in MAIN" @04:47:44.33 upsync @11:52:28.25PDT03-Jul-21
>BREAK< ESP3 Gateway v1.11 for core bus @11:52:28.27 >BREAK< ESP3 Gateway v1.11 for cartridge bus @11:52:34.76 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @11:52:35.63 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @11:52:44.20 Polling Can every 2:00 @11:52:44.36 Loading /home/esp/esp2/bin/processAll mission @11:52:47.02 Sampling default goal volumes @11:52:47.07 GoodNews.email "Cartridge group: dry\nSpares: {spare: 0}\nSchedule: [\"3am\", \"3pm\"]\n",Subject: "Begun processAll --schedule=3am,3pm --checkpoint --mail" @11:52:47.17 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @11:52:47.21
delayUntil "03:00:00.00PDT04-Jul-21" @11:52:47.24 Retry #1 of 10 after 10:00 @11:53:15.30 Session Begun @11:53:33.28 -> Cartridge.dry @11:53:51.64 -> Cartridge.reserved @11:53:58.19 -> exit @11:53:58.22 Session Ended @11:58:41.37 Session Begun @11:58:47.82 -> showlog @11:59:20.87 -> GoodNews.email "Mission resumed after power failure" @11:59:20.98 GoodNews.email "Mission resumed after power failure" @11:59:29.22 -> exit @11:59:29.28 Session Ended @11:59:31.29 upsync @12:02:57.16 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 03:00PDT04-Jul-21" @12:03:05.55 upsync @12:03:15.78 upsync @19:18:05.44 Session Begun @19:18:10.56 -> showlog -200 @19:18:20.91 -> MainThread.exit @19:18:21.19 Session Ended @14:39:36.02PDT05-Jul-21
>BREAK< ESP3 Gateway v1.11 for core bus @14:39:36.10 >BREAK< ESP3 Gateway v1.11 for cartridge bus @14:39:41.57 Configuration for koa in flowerpot -- 2/11/21 brent@mbari.org /home/esp/esp2/type/auv/koa/configure.rb @14:39:41.88 Cartridge::Sampler::Warning in MAIN -- Vacuum sensor saturates below -18psi @14:39:51.36 Polling Can every 2:00 @14:39:51.42 Loading /home/esp/esp2/bin/processAll mission @14:39:53.12 Sampling default goal volumes @14:39:53.17 GoodNews.email "Cartridge group: dry\nSpares: {spare: 1}\nSchedule: [-4:00:00]\n",Subject: "Begun processAll --interval=4:: --mail --checkpoint" @14:39:53.36
Gate.power :main,:ON,limit to 5000mA @14:39:53.37 Gate.power :analytic1,:ON,limit to 5000mA @14:39:53.39 Gate.power :analytic2,:ON,limit to 5000mA @14:39:53.40 Gate.power :analytic3,:ON,limit to 5000mA @14:39:53.44 Powering core bus @14:39:56.06 Archive.configure :default @14:39:56.46 Air.configure :default @14:39:56.75 Lysis1.configure :default @14:39:57.20 Lysis2.configure :default @14:39:57.57 Plunge.configure :default @14:39:57.87 SamplePump.configure :default @14:39:58.28 ExtraPump.configure :default @14:39:58.75 Toroid.configure :default @14:39:59.26 Carousel.configure :default @14:39:59.55 Pressure.configure :default @14:39:59.66 All elf controllers running firmware version 4.61 @14:39:59.72 Gate.power :motor,:ON,limit to 5000mA @14:40:00.75 Plunge.reconfigure :default @14:40:00.79 Lysis2.reconfigure :default @14:40:00.84 Lysis1.reconfigure :default @14:40:00.87 Air.reconfigure :default @14:40:00.91 Toroid.reconfigure :default @14:40:00.97 Archive.reconfigure :default @14:40:01.22 Plunge.home.jog 0 @14:40:01.26 Lysis1.home.jog 0 @14:40:01.28 Toroid.home.jog 0 @14:40:01.31 Air.home.jog 0 @14:40:01.34 Lysis2.home.jog 0 @14:40:01.39 Archive.home.jog 0 @14:40:02.06 Plunge.reconfigure :homing @14:40:02.11 Toroid.reconfigure :homing @14:40:02.16 Air.reconfigure :homing @14:40:02.31 Plunge.jog 840 @14:40:02.37 Air.jog 1358 @14:40:02.40 Toroid.jog 636 @14:40:02.67 Lysis1.reconfigure :homing @14:40:02.70 Archive.reconfigure :homing @14:40:02.73 Lysis2.reconfigure :homing @14:40:02.87 Archive.jog 1367 @14:40:02.93 Lysis2.jog 1343 @14:40:02.96 Lysis1.jog 1358 @14:40:03.36 Plunge.setRawPosition! 50 @14:40:03.46 Plunge.seek :home @14:40:03.62 Air.setRawPosition! 50 @14:40:03.67 Air.seek :home @14:40:03.78 Plunge.reconfigure :default @14:40:03.95 Air.reconfigure :default @14:40:04.12 Archive.setRawPosition! 50 @14:40:04.18 Archive.seek :home @14:40:04.28 Lysis1.setRawPosition! 50 @14:40:04.31 Lysis2.setRawPosition! 50 @14:40:04.39 Lysis1.seek :home @14:40:04.42 Lysis2.seek :home @14:40:04.70 Archive.reconfigure :default @14:40:05.20 Toroid.setRawPosition! 0 @14:40:05.38 Lysis2.reconfigure :default @14:40:05.50 Lysis1.reconfigure :default @14:40:06.04 Toroid.seek :home @14:40:06.35 Toroid.reconfigure :default @14:40:06.76
Toroid.seek :clear @14:40:07.75 upsync @14:40:08.42
Carousel.reconfigure :default @14:40:08.63 Carousel.home.jog 0 @14:40:09.23 Carousel.seek 9 @14:40:09.55 Selecting Cartridge 8 @14:40:10.51 Carousel.seek 8 @14:40:11.02 Lysis1.seek :locked @14:40:14.25 Plunge.seek :locked @14:40:14.39 Air.seek :locked @14:40:18.76
Mounted archive Cartridge @14:40:18.87 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 8" @14:40:19.76 Duration of filtering limited to 4:05:00 @14:40:20.05 Priming sample loop w/160ml, bypass w/40ml @14:40:20.07 SamplePump.setPosition! 0ml @14:40:20.11 Bypass.open @14:40:21.56 Intake.open @14:40:22.46 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @14:40:25.24 Vacuum tare is 18% of full scale @14:40:25.30 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @14:40:25.87 Pressure.reconfigure :default @14:40:26.01 Pressure.hold @14:40:26.70 SamplePump.reconfigure :prime @14:40:26.90 SamplePump.seek 40ml,3:51 @14:40:32.87 upsync @14:41:10.16 Pressure.coast @14:41:10.22 SamplePump.setPosition! 0ml @14:41:10.27 Exhaust.open @14:41:11.69 Bypass.close @14:41:13.48 Pressure.reconfigure :default @14:41:13.62 Pressure.hold @14:41:14.24 SamplePump.reconfigure :prime @14:41:14.37 SamplePump.seek 160ml,13:51 @14:44:06.19 Pressure.coast @14:44:06.27 Bypass.open @14:44:07.04 Intake.close @14:44:08.03 Toroid.reconfigure :hold @14:44:08.18 Toroid.seek :crack @14:44:10.00 Toroid.seek :cartridge @14:44:10.98 Toroid.reconfigure :default @14:44:11.04 Toroid.seek :clear @14:44:12.88 Pressurizing cartridge to 26psi above 14.4psia @14:44:12.90 SamplePump.setPosition! 0ml @14:44:12.93 Intake.open @14:44:13.73 Bypass.close @14:44:14.44 Exhaust.close @14:44:15.58 Pressure.reconfigure :default @14:44:15.68 Pressure.hold @14:44:16.21 SamplePump.reconfigure :charge @14:44:16.30 SamplePump.seek 2.3ml,42 seconds @14:44:29.61 Pressure.coast @14:44:29.66 Intake.close @14:44:30.48 Pressurized cartridge with 0.807ml to 26.0psi @14:44:30.50 Isolated @36.5psia (22.1psi) @14:44:30.51 Checking for cartridge leaks from 36.5psia @14:44:30.53 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @14:44:38.12 Cartridge pressure fell 8% to 33.5psia @14:44:38.51 Sampling 2000.0ml @14:44:38.56 SamplePump.setPosition! 0ml @14:44:38.66 Exhaust.open @14:44:39.47 Intake.open @14:44:40.41 Pressure.reconfigure :default @14:44:40.47 Pressure.hold @14:44:40.60 Sea 1.7m deep (17.2psia) @14:44:40.76 SamplePump.reconfigure :sample @14:44:40.83 SamplePump.seek 2000ml,2:47:12 @14:44:43.87 Cartridge 8 sampling 1/2000ml at 0.60ml/s with 25.5psi vacuum @14:45:43.48 Cartridge 8 sampling 34/2000ml at 0.55ml/s with 24.3psi vacuum @14:46:43.71 Cartridge 8 sampling 67/2000ml at 0.56ml/s with 22.3psi vacuum @14:47:43.48 Cartridge 8 sampling 101/2000ml at 0.55ml/s with 25.4psi vacuum @14:48:43.63 Cartridge 8 sampling 134/2000ml at 0.56ml/s with 23.9psi vacuum @14:49:43.48 Cartridge 8 sampling 168/2000ml at 0.55ml/s with 24.0psi vacuum @14:50:43.58 Cartridge 8 sampling 202/2000ml at 0.55ml/s with 22.0psi vacuum @14:51:43.48 Cartridge 8 sampling 236/2000ml at 0.57ml/s with 24.6psi vacuum @14:52:43.63 Cartridge 8 sampling 270/2000ml at 0.57ml/s with 24.7psi vacuum @14:53:43.48 Cartridge 8 sampling 304/2000ml at 0.58ml/s with 24.3psi vacuum @14:54:43.64 Cartridge 8 sampling 338/2000ml at 0.56ml/s with 23.5psi vacuum @14:55:43.48 Cartridge 8 sampling 372/2000ml at 0.57ml/s with 25.1psi vacuum @14:56:43.62 Cartridge 8 sampling 406/2000ml at 0.57ml/s with 25.1psi vacuum @14:57:43.48 Cartridge 8 sampling 440/2000ml at 0.57ml/s with 25.1psi vacuum @14:58:43.79 Cartridge 8 sampling 474/2000ml at 0.57ml/s with 23.8psi vacuum @14:59:43.89 Cartridge 8 sampling 508/2000ml at 0.57ml/s with 23.9psi vacuum @15:00:43.48 Cartridge 8 sampling 542/2000ml at 0.56ml/s with 26.1psi vacuum @15:01:44.01 Cartridge 8 sampling 576/2000ml at 0.57ml/s with 21.1psi vacuum @15:02:43.48 Cartridge 8 sampling 610/2000ml at 0.57ml/s with 22.7psi vacuum @15:03:43.67 Cartridge 8 sampling 644/2000ml at 0.56ml/s with 24.2psi vacuum @15:04:43.48 Cartridge 8 sampling 677/2000ml at 0.56ml/s with 25.2psi vacuum @15:05:43.69 Cartridge 8 sampling 711/2000ml at 0.56ml/s with 23.0psi vacuum @15:06:43.48 Cartridge 8 sampling 744/2000ml at 0.57ml/s with 24.9psi vacuum @15:07:43.67 Cartridge 8 sampling 778/2000ml at 0.57ml/s with 22.5psi vacuum @15:08:43.48 Cartridge 8 sampling 812/2000ml at 0.56ml/s with 22.7psi vacuum @15:09:43.63 Cartridge 8 sampling 846/2000ml at 0.56ml/s with 24.9psi vacuum @15:10:43.48 Cartridge 8 sampling 879/2000ml at 0.54ml/s with 20.1psi vacuum @15:11:43.76 Cartridge 8 sampling 912/2000ml at 0.55ml/s with 24.8psi vacuum @15:12:43.87 Cartridge 8 sampling 944/2000ml at 0.56ml/s with 23.2psi vacuum @15:13:43.48 Cartridge 8 sampling 977/2000ml at 0.55ml/s with 24.1psi vacuum @15:14:43.80 Cartridge 8 sampling 1010/2000ml at 0.54ml/s with 25.1psi vacuum @15:15:43.48 Cartridge 8 sampling 1043/2000ml at 0.54ml/s with 24.1psi vacuum @15:16:43.68 Cartridge 8 sampling 1076/2000ml at 0.55ml/s with 22.8psi vacuum @15:17:43.48 Cartridge 8 sampling 1109/2000ml at 0.54ml/s with 22.6psi vacuum @15:18:43.68 Cartridge 8 sampling 1141/2000ml at 0.55ml/s with 25.3psi vacuum @15:19:43.48 Cartridge 8 sampling 1174/2000ml at 0.54ml/s with 22.7psi vacuum @15:20:43.59 Cartridge 8 sampling 1207/2000ml at 0.55ml/s with 25.3psi vacuum @15:21:43.48 Cartridge 8 sampling 1239/2000ml at 0.55ml/s with 24.8psi vacuum @15:22:43.68 Cartridge 8 sampling 1271/2000ml at 0.53ml/s with 20.2psi vacuum @15:23:43.48 Cartridge 8 sampling 1303/2000ml at 0.54ml/s with 22.4psi vacuum @15:24:43.59 Cartridge 8 sampling 1335/2000ml at 0.53ml/s with 24.7psi vacuum @15:25:43.48 Cartridge 8 sampling 1367/2000ml at 0.54ml/s with 22.3psi vacuum @15:26:43.48 Cartridge 8 sampling 1398/2000ml at 0.52ml/s with 22.8psi vacuum @15:27:43.74 Cartridge 8 sampling 1430/2000ml at 0.50ml/s with 20.6psi vacuum @15:28:43.48 Cartridge 8 sampling 1462/2000ml at 0.53ml/s with 25.2psi vacuum @15:29:43.65 Cartridge 8 sampling 1493/2000ml at 0.53ml/s with 21.9psi vacuum @15:30:44.05 Cartridge 8 sampling 1524/2000ml at 0.50ml/s with 21.3psi vacuum @15:31:43.56 Cartridge 8 sampling 1555/2000ml at 0.52ml/s with 23.6psi vacuum @15:32:43.48 Cartridge 8 sampling 1587/2000ml at 0.53ml/s with 22.0psi vacuum @15:33:43.66 Cartridge 8 sampling 1618/2000ml at 0.53ml/s with 24.3psi vacuum @15:34:43.48 Cartridge 8 sampling 1649/2000ml at 0.52ml/s with 22.6psi vacuum @15:35:43.61 Cartridge 8 sampling 1680/2000ml at 0.51ml/s with 21.3psi vacuum @15:36:43.48 Cartridge 8 sampling 1711/2000ml at 0.51ml/s with 23.6psi vacuum @15:37:43.58 Cartridge 8 sampling 1742/2000ml at 0.50ml/s with 20.9psi vacuum @15:38:43.48 Cartridge 8 sampling 1773/2000ml at 0.52ml/s with 24.0psi vacuum @15:39:43.48 Cartridge 8 sampling 1804/2000ml at 0.51ml/s with 26.5psi vacuum @15:40:43.80 Cartridge 8 sampling 1834/2000ml at 0.51ml/s with 26.1psi vacuum @15:41:43.48 Cartridge 8 sampling 1865/2000ml at 0.52ml/s with 22.9psi vacuum @15:42:43.64 Cartridge 8 sampling 1895/2000ml at 0.51ml/s with 24.5psi vacuum @15:43:43.48 Cartridge 8 sampling 1926/2000ml at 0.50ml/s with 25.6psi vacuum @15:44:43.65 Cartridge 8 sampling 1956/2000ml at 0.51ml/s with 23.4psi vacuum @15:45:43.49 Cartridge 8 sampling 1987/2000ml at 0.51ml/s with 26.4psi vacuum @15:46:10.01 Pressure.coast @15:46:10.14 Intake.close @15:46:10.21 Exhaust.close @15:46:10.39 Toroid.reconfigure :hold @15:46:10.44 Toroid.seek :bypass @15:46:12.06 Toroid.reconfigure :default @15:46:12.12 Toroid.seek :clear @15:46:13.94 Sampled 2000.0ml @15:46:14.98 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 8" @15:46:15.88 SamplePump.setPosition! 0ml @15:46:16.07 Plunge.seek :reagent @15:46:16.54 Bypass.open @15:46:17.34 Sample loop closed. Ready to Deliver Reagents @15:46:17.36 Archive.seek :sealed @15:46:17.56 Intake.open @15:46:18.65 SamplePump.reconfigure :slow @15:46:18.71 SamplePump.seek -3ml @15:46:21.03 Archive.seek :evac @15:46:25.43 delay 1 second @15:46:26.70 SamplePump.seek -2.8ml @15:46:27.45 delay 1 second @15:46:28.56 SamplePump.seek -3.3ml @15:46:29.89 delay 1 second @15:46:30.92 Exhaust.open @15:46:31.61 Bypass.close @15:46:32.20 upsync @15:46:32.38 SamplePump.seek -35.3ml @15:46:35.07 delay 1:00 @15:47:35.12 Archive.seek :clear @15:47:41.22 delay 3 seconds @15:47:44.41 Intake.close @15:47:44.52 Exhaust.close @15:47:52.26 RNAlater soak for 10:00 @15:57:52.30 Archive.seek :tVent @15:57:54.04 delay 1 second @15:57:55.07 Archive.seek :evac @15:58:11.13 delay 10 seconds @15:58:21.19 Sample preserved and evacuated @15:58:21.65
Carousel.seek 8 @15:58:22.13 Plunge.seek :locked @15:58:23.33 Archive.seek :clear @15:58:23.82 Air.seek :clear @15:58:23.98 Lysis1.seek :clear @15:58:40.92
Archive.coast @15:58:41.07 Air.coast @15:58:41.11 Lysis1.coast @15:58:41.15 Lysis2.coast @15:58:41.35 Plunge.coast @15:58:41.39 SamplePump.coast @15:58:41.44 Toroid.coast @15:58:41.55 Carousel.coast @15:58:42.21 Gate.power :motor,:OFF @15:58:42.22 Gate.power :core,:OFF @15:58:42.23 Gate.power :analytic1,:OFF @15:58:42.25 Gate.power :analytic2,:OFF @15:58:42.26 Gate.power :analytic3,:OFF @15:58:42.27 Gate.power :main,:OFF @15:58:42.28 Safely stopped and ready to power off @15:58:42.50 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 18:39" @15:58:42.52
delayUntil "18:39:51.37" @15:59:14.47 upsync @18:39:51.80
Gate.power :main,:ON,limit to 5000mA @18:39:51.83 Gate.power :analytic1,:ON,limit to 5000mA @18:39:51.86 Gate.power :analytic2,:ON,limit to 5000mA @18:39:51.90 Gate.power :analytic3,:ON,limit to 5000mA @18:39:52.05 Powering core bus @18:39:54.41 Archive.configure :default @18:39:54.55 Air.configure :default @18:39:54.65 Lysis1.configure :default @18:39:54.77 Lysis2.configure :default @18:39:54.92 Plunge.configure :default @18:39:55.06 SamplePump.configure :default @18:39:55.19 ExtraPump.configure :default @18:39:55.33 Toroid.configure :default @18:39:55.47 Carousel.configure :default @18:39:55.62 Pressure.configure :default @18:39:55.67 All elf controllers running firmware version 4.61 @18:39:55.69 Gate.power :motor,:ON,limit to 5000mA @18:39:55.99 Archive.reconfigure :default @18:39:56.23 Archive.home.jog 0 @18:39:56.43 Plunge.reconfigure :default @18:39:56.50 Lysis1.reconfigure :default @18:39:56.55 Air.reconfigure :default @18:39:56.58 Toroid.reconfigure :default @18:39:56.63 Lysis2.reconfigure :default @18:39:56.84 Plunge.home.jog 0 @18:39:56.89 Air.home.jog 0 @18:39:56.91 Toroid.home.jog 0 @18:39:56.93 Lysis1.home.jog 0 @18:39:56.95 Lysis2.home.jog 0 @18:39:57.34 Archive.reconfigure :homing @18:39:57.62 Toroid.reconfigure :homing @18:39:57.68 Archive.jog 1367 @18:39:57.73 Plunge.reconfigure :homing @18:39:57.75 Lysis1.reconfigure :homing @18:39:57.79 Air.reconfigure :homing @18:39:57.99 Toroid.jog 644 @18:39:58.05 Plunge.jog 837 @18:39:58.12 Lysis1.jog 1356 @18:39:58.27 Air.jog 1355 @18:39:58.82 Lysis2.reconfigure :homing @18:39:58.96 Lysis2.jog 1346 @18:39:59.19 Archive.setRawPosition! 50 @18:39:59.30 Archive.seek :home @18:39:59.74 Plunge.setRawPosition! 50 @18:39:59.89 Plunge.seek :home @18:40:00.07 Lysis1.setRawPosition! 50 @18:40:00.17 Archive.reconfigure :default @18:40:00.27 Air.setRawPosition! 50 @18:40:00.68 Lysis1.seek :home @18:40:00.75 Air.seek :home @18:40:01.31 Toroid.setRawPosition! 0 @18:40:01.43 Plunge.reconfigure :default @18:40:01.49 Lysis1.reconfigure :default @18:40:01.58 Lysis2.setRawPosition! 50 @18:40:01.62 Air.reconfigure :default @18:40:02.08 Toroid.seek :home @18:40:02.46 Lysis2.seek :home @18:40:02.69 Toroid.reconfigure :default @18:40:02.95 Lysis2.reconfigure :default @18:40:03.44
Toroid.seek :clear @18:40:05.15 Carousel.reconfigure :default @18:40:05.20 Carousel.home.jog 0 @18:40:05.69 Carousel.seek 8 @18:40:05.87 Selecting Cartridge 7 @18:40:06.08 Carousel.seek 7 @18:40:06.59 Lysis1.seek :locked @18:40:10.05 Plunge.seek :locked @18:40:10.42 Air.seek :locked @18:40:14.65
Mounted archive Cartridge @18:40:14.77 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 7" @18:40:14.88 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:40:14.91 Retry #1 of 10 after 10:00 @18:40:15.13 Duration of filtering limited to 4:05:00 @18:40:15.42 Priming sample loop w/160ml, bypass w/40ml @18:40:15.43 SamplePump.setPosition! 0ml @18:40:15.47 Bypass.open @18:40:16.96 Intake.open @18:40:18.17 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @18:40:20.94 Vacuum tare is 18% of full scale @18:40:20.96 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @18:40:21.49 Pressure.reconfigure :default @18:40:21.61 Pressure.hold @18:40:22.26 SamplePump.reconfigure :prime @18:40:22.37 SamplePump.seek 40ml,3:51 @18:41:05.73 Pressure.coast @18:41:05.84 SamplePump.setPosition! 0ml @18:41:05.89 Exhaust.open @18:41:07.45 Bypass.close @18:41:09.24 Pressure.reconfigure :default @18:41:09.30 Pressure.hold @18:41:09.57 SamplePump.reconfigure :prime @18:41:09.64 SamplePump.seek 160ml,13:51 @18:44:01.37 Pressure.coast @18:44:01.49 Bypass.open @18:44:02.35 Intake.close @18:44:03.43 Toroid.reconfigure :hold @18:44:03.48 Toroid.seek :crack @18:44:05.07 Toroid.seek :cartridge @18:44:06.00 Toroid.reconfigure :default @18:44:06.06 Toroid.seek :clear @18:44:08.41 Pressurizing cartridge to 26psi above 14.4psia @18:44:08.45 SamplePump.setPosition! 0ml @18:44:08.53 Intake.open @18:44:09.35 Bypass.close @18:44:10.20 Exhaust.close @18:44:11.28 Pressure.reconfigure :default @18:44:11.41 Pressure.hold @18:44:11.96 SamplePump.reconfigure :charge @18:44:12.10 SamplePump.seek 2.3ml,42 seconds @18:44:25.62 Pressure.coast @18:44:26.09 Intake.close @18:44:27.24 Pressurized cartridge with 0.832ml to 26.0psi @18:44:27.28 Isolated @32.6psia (18.2psi) @18:44:27.34 Checking for cartridge leaks from 32.6psia @18:44:27.38 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @18:44:32.11 Cartridge pressure fell 4% to 31.4psia @18:44:32.19 Sampling 2000.0ml @18:44:32.21 SamplePump.setPosition! 0ml @18:44:32.25 Exhaust.open @18:44:32.95 Intake.open @18:44:33.91 Pressure.reconfigure :default @18:44:33.97 Pressure.hold @18:44:34.24 SamplePump.reconfigure :sample @18:44:34.31 SamplePump.seek 2000ml,2:47:12 @18:44:37.19 Cartridge 7 sampling 1/2000ml at 0.63ml/s with 24.7psi vacuum @18:45:36.98 Cartridge 7 sampling 34/2000ml at 0.55ml/s with 24.9psi vacuum @18:46:37.24 Cartridge 7 sampling 67/2000ml at 0.56ml/s with 23.9psi vacuum @18:47:36.98 Cartridge 7 sampling 101/2000ml at 0.56ml/s with 25.6psi vacuum @18:48:37.05 Cartridge 7 sampling 135/2000ml at 0.56ml/s with 25.9psi vacuum @18:49:36.98 Cartridge 7 sampling 169/2000ml at 0.56ml/s with 24.1psi vacuum @18:50:15.01 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @18:50:15.04 Retry #2 of 10 after 10:00 @18:50:37.16 Cartridge 7 sampling 203/2000ml at 0.57ml/s with 26.8psi vacuum @18:51:36.98 Cartridge 7 sampling 237/2000ml at 0.57ml/s with 25.1psi vacuum @18:52:37.15 Cartridge 7 sampling 271/2000ml at 0.57ml/s with 25.1psi vacuum @18:53:36.98 Cartridge 7 sampling 306/2000ml at 0.58ml/s with 23.2psi vacuum @18:54:37.32 Cartridge 7 sampling 340/2000ml at 0.57ml/s with 22.1psi vacuum @18:55:37.35 Cartridge 7 sampling 374/2000ml at 0.56ml/s with 26.1psi vacuum @18:56:36.98 Cartridge 7 sampling 408/2000ml at 0.56ml/s with 24.4psi vacuum @18:57:37.19 Cartridge 7 sampling 442/2000ml at 0.56ml/s with 18.6psi vacuum @18:58:36.98 Cartridge 7 sampling 476/2000ml at 0.55ml/s with 27.2psi vacuum @18:59:37.13 Cartridge 7 sampling 510/2000ml at 0.58ml/s with 22.9psi vacuum @19:00:15.24 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:00:15.27 Retry #3 of 10 after 10:00 @19:00:36.97 Cartridge 7 sampling 544/2000ml at 0.56ml/s with 25.8psi vacuum @19:01:37.09 Cartridge 7 sampling 578/2000ml at 0.56ml/s with 27.9psi vacuum @19:02:36.98 Cartridge 7 sampling 612/2000ml at 0.56ml/s with 21.0psi vacuum @19:03:37.07 Cartridge 7 sampling 646/2000ml at 0.55ml/s with 20.7psi vacuum 1.7m deep (17.2psia) @19:04:36.98 Cartridge 7 sampling 680/2000ml at 0.56ml/s with 27.8psi vacuum @19:05:37.12 Cartridge 7 sampling 714/2000ml at 0.58ml/s with 24.4psi vacuum @19:06:36.98 Cartridge 7 sampling 747/2000ml at 0.54ml/s with 27.4psi vacuum @19:07:37.22 Cartridge 7 sampling 781/2000ml at 0.57ml/s with 23.0psi vacuum @19:08:37.06 Cartridge 7 sampling 815/2000ml at 0.57ml/s with 25.4psi vacuum @19:09:36.97 Cartridge 7 sampling 848/2000ml at 0.55ml/s with 27.4psi vacuum @19:10:15.33 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:10:15.35 Retry #4 of 10 after 10:00 @19:10:37.23 Cartridge 7 sampling 882/2000ml at 0.54ml/s with 21.8psi vacuum @19:11:36.97 Cartridge 7 sampling 915/2000ml at 0.54ml/s with 24.6psi vacuum @19:12:37.23 Cartridge 7 sampling 949/2000ml at 0.55ml/s with 19.7psi vacuum @19:13:36.97 Cartridge 7 sampling 982/2000ml at 0.55ml/s with 26.8psi vacuum @19:14:37.14 Cartridge 7 sampling 1015/2000ml at 0.57ml/s with 26.2psi vacuum @19:15:36.98 Cartridge 7 sampling 1048/2000ml at 0.56ml/s with 26.8psi vacuum @19:16:37.10 Cartridge 7 sampling 1082/2000ml at 0.57ml/s with 25.0psi vacuum @19:17:36.98 Cartridge 7 sampling 1115/2000ml at 0.56ml/s with 24.6psi vacuum @19:18:37.01 Cartridge 7 sampling 1148/2000ml at 0.54ml/s with 19.9psi vacuum @19:19:36.98 Cartridge 7 sampling 1181/2000ml at 0.56ml/s with 28.0psi vacuum @19:20:15.45 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:20:15.47 Retry #5 of 10 after 10:00 @19:20:37.16 Cartridge 7 sampling 1213/2000ml at 0.56ml/s with 23.5psi vacuum @19:21:37.15 Cartridge 7 sampling 1246/2000ml at 0.55ml/s with 23.0psi vacuum @19:22:36.98 Cartridge 7 sampling 1278/2000ml at 0.53ml/s with 26.5psi vacuum @19:23:37.22 Cartridge 7 sampling 1311/2000ml at 0.54ml/s with 27.0psi vacuum @19:24:36.98 Cartridge 7 sampling 1343/2000ml at 0.54ml/s with 22.8psi vacuum @19:25:37.14 Cartridge 7 sampling 1375/2000ml at 0.51ml/s with 28.1psi vacuum @19:26:36.97 Cartridge 7 sampling 1406/2000ml at 0.52ml/s with 26.7psi vacuum @19:27:37.18 Cartridge 7 sampling 1438/2000ml at 0.53ml/s with 25.1psi vacuum @19:28:36.98 Cartridge 7 sampling 1470/2000ml at 0.51ml/s with 19.2psi vacuum @19:29:37.14 Cartridge 7 sampling 1502/2000ml at 0.53ml/s with 21.1psi vacuum @19:30:36.98 Cartridge 7 sampling 1533/2000ml at 0.53ml/s with 23.1psi vacuum @19:30:37.15 upsync @19:31:37.15 Cartridge 7 sampling 1565/2000ml at 0.53ml/s with 24.2psi vacuum @19:32:36.98 Cartridge 7 sampling 1596/2000ml at 0.51ml/s with 19.5psi vacuum @19:33:37.06 Cartridge 7 sampling 1628/2000ml at 0.53ml/s with 20.8psi vacuum @19:34:36.98 Cartridge 7 sampling 1659/2000ml at 0.50ml/s with 27.3psi vacuum @19:35:37.38 Cartridge 7 sampling 1690/2000ml at 0.49ml/s with 21.5psi vacuum @19:36:37.29 Cartridge 7 sampling 1721/2000ml at 0.51ml/s with 29.5psi vacuum @19:37:37.17 Cartridge 7 sampling 1752/2000ml at 0.53ml/s with 23.6psi vacuum @19:38:37.37 Cartridge 7 sampling 1783/2000ml at 0.52ml/s with 27.3psi vacuum @19:39:36.98 Cartridge 7 sampling 1814/2000ml at 0.51ml/s with 21.1psi vacuum @19:40:37.14 Cartridge 7 sampling 1844/2000ml at 0.50ml/s with 19.3psi vacuum 2.2m deep (18.0psia) @19:41:36.98 Cartridge 7 sampling 1875/2000ml at 0.50ml/s with 27.6psi vacuum @19:42:37.14 Cartridge 7 sampling 1905/2000ml at 0.50ml/s with 18.5psi vacuum 3.2m deep (19.4psia) @19:43:36.98 Cartridge 7 sampling 1935/2000ml at 0.52ml/s with 19.9psi vacuum @19:44:37.08 Cartridge 7 sampling 1965/2000ml at 0.51ml/s with 26.5psi vacuum @19:45:36.98 Cartridge 7 sampling 1995/2000ml at 0.51ml/s with 23.0psi vacuum @19:45:46.37 Pressure.coast @19:45:46.51 Intake.close @19:45:46.57 Exhaust.close @19:45:46.75 Toroid.reconfigure :hold @19:45:46.81 Toroid.seek :bypass @19:45:48.48 Toroid.reconfigure :default @19:45:48.70 Toroid.seek :clear @19:45:50.43 Sampled 2000.0ml @19:45:52.68 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 7" @19:45:52.78 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:45:52.81 Retry #1 of 10 after 10:00 @19:45:53.15 SamplePump.setPosition! 0ml @19:45:53.20 Plunge.seek :reagent @19:45:53.23 Bypass.open @19:45:54.00 Intake.open @19:45:54.18 Sample loop closed. Ready to Deliver Reagents @19:45:54.20 Archive.seek :sealed @19:45:54.84 SamplePump.reconfigure :slow @19:45:54.91 SamplePump.seek -3ml @19:45:57.90 Archive.seek :evac @19:46:01.62 delay 1 second @19:46:02.76 SamplePump.seek -2.8ml @19:46:03.46 delay 1 second @19:46:04.55 SamplePump.seek -3.3ml @19:46:05.88 delay 1 second @19:46:06.91 Exhaust.open @19:46:07.71 Bypass.close @19:46:08.56 SamplePump.seek -35.3ml @19:46:11.91 delay 1:00 @19:47:11.96 Archive.seek :clear @19:47:17.38 delay 3 seconds @19:47:20.56 Intake.close @19:47:20.78 Exhaust.close @19:47:29.11 RNAlater soak for 10:00 @19:55:52.86 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @19:55:52.88 Retry #2 of 10 after 10:00 @19:57:29.14 Archive.seek :tVent @19:57:30.86 delay 1 second @19:57:31.91 Archive.seek :evac @19:57:47.90 delay 10 seconds @19:57:57.94 Sample preserved and evacuated @19:57:58.67
Carousel.seek 7 @19:57:58.90 Plunge.seek :locked @19:57:59.89 Archive.seek :clear @19:58:00.05 Air.seek :clear @19:58:00.23 Lysis1.seek :clear @19:58:17.09
Archive.coast @19:58:17.13 Air.coast @19:58:17.18 Lysis1.coast @19:58:17.22 Lysis2.coast @19:58:17.26 Plunge.coast @19:58:17.31 SamplePump.coast @19:58:17.35 Toroid.coast @19:58:17.40 Carousel.coast @19:58:17.59 Gate.power :motor,:OFF @19:58:17.60 Gate.power :core,:OFF @19:58:17.62 Gate.power :analytic1,:OFF @19:58:17.63 Gate.power :analytic2,:OFF @19:58:17.64 Gate.power :analytic3,:OFF @19:58:17.65 Gate.power :main,:OFF @19:58:17.67 Safely stopped and ready to power off @19:58:17.87 delayUntil "22:39:51.37" @20:05:52.98 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @20:05:53.04 Retry #3 of 10 after 10:00 @20:16:33.11 EOFError in Qmail -- end of file reached @20:16:33.13 Retry #4 of 10 after 10:00 @20:26:53.24 EOFError in Qmail -- end of file reached @20:26:53.26 Retry #5 of 10 after 10:00 @20:39:52.33 EOFError in Qmail -- end of file reached @20:39:52.39 Retry #6 of 10 after 10:00 @20:49:52.47 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @20:49:52.49 Retry #7 of 10 after 10:00 @20:59:52.54 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @20:59:52.57 Retry #8 of 10 after 10:00 @21:11:23.26 EOFError in Qmail -- end of file reached @21:11:23.31 Retry #9 of 10 after 10:00 @21:21:23.37 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:21:23.40 Retry #10 of 10 after 10:00 @21:31:23.48 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:31:23.60 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 22:39" @21:31:23.65 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:31:23.67 Retry #1 of 10 after 10:00 @21:41:23.72 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:41:23.74 Retry #2 of 10 after 10:00 @21:51:23.79 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @21:51:23.81 Retry #3 of 10 after 10:00 @22:01:35.55 upsync @22:39:51.51
Gate.power :main,:ON,limit to 5000mA @22:39:51.53 Gate.power :analytic1,:ON,limit to 5000mA @22:39:51.54 Gate.power :analytic2,:ON,limit to 5000mA @22:39:51.56 Gate.power :analytic3,:ON,limit to 5000mA @22:39:51.60 Powering core bus @22:39:54.26 Archive.configure :default @22:39:54.68 Air.configure :default @22:39:54.89 Lysis1.configure :default @22:39:55.19 Lysis2.configure :default @22:39:55.52 Plunge.configure :default @22:39:55.79 SamplePump.configure :default @22:39:56.04 ExtraPump.configure :default @22:39:56.40 Toroid.configure :default @22:39:56.70 Carousel.configure :default @22:39:56.89 Pressure.configure :default @22:39:57.08 All elf controllers running firmware version 4.61 @22:39:57.12 Gate.power :motor,:ON,limit to 5000mA @22:39:58.15 Air.reconfigure :default @22:39:58.23 Toroid.reconfigure :default @22:39:58.26 Lysis2.reconfigure :default @22:39:58.30 Lysis1.reconfigure :default @22:39:58.35 Plunge.reconfigure :default @22:39:58.39 Archive.reconfigure :default @22:39:58.60 Air.home.jog 0 @22:39:58.63 Lysis2.home.jog 0 @22:39:58.67 Toroid.home.jog 0 @22:39:58.70 Lysis1.home.jog 0 @22:39:58.74 Plunge.home.jog 0 @22:39:58.94 Archive.home.jog 0 @22:39:59.57 Toroid.reconfigure :homing @22:39:59.63 Plunge.reconfigure :homing @22:39:59.66 Lysis2.reconfigure :homing @22:39:59.70 Air.reconfigure :homing @22:39:59.76 Archive.reconfigure :homing @22:39:59.79 Lysis1.reconfigure :homing @22:40:00.01 Plunge.jog 836 @22:40:00.07 Lysis2.jog 1346 @22:40:00.10 Toroid.jog 653 @22:40:00.13 Archive.jog 1359 @22:40:00.16 Lysis1.jog 1352 @22:40:00.19 Air.jog 1353 @22:40:01.11 Plunge.setRawPosition! 50 @22:40:01.16 Plunge.seek :home @22:40:01.38 Lysis2.setRawPosition! 50 @22:40:01.45 Lysis2.seek :home @22:40:01.59 Plunge.reconfigure :default @22:40:01.63 Lysis1.setRawPosition! 50 @22:40:01.65 Air.setRawPosition! 50 @22:40:01.68 Archive.setRawPosition! 50 @22:40:01.80 Lysis1.seek :home @22:40:01.92 Air.seek :home @22:40:01.97 Archive.seek :home @22:40:02.19 Lysis2.reconfigure :default @22:40:02.68 Lysis1.reconfigure :default @22:40:02.86 Air.reconfigure :default @22:40:02.98 Archive.reconfigure :default @22:40:03.46 Toroid.setRawPosition! 0 @22:40:03.61 Toroid.seek :home @22:40:03.91 Toroid.reconfigure :default @22:40:04.24
Toroid.seek :clear @22:40:05.95 Carousel.reconfigure :default @22:40:06.04 Carousel.home.jog 0 @22:40:06.62 Carousel.seek 7 @22:40:06.91 Selecting Cartridge 6 @22:40:07.27 Carousel.seek 6 @22:40:08.04 Lysis1.seek :locked @22:40:11.26 Plunge.seek :locked @22:40:11.40 Air.seek :locked @22:40:15.70
Mounted archive Cartridge @22:40:16.01 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 6" @22:40:16.50 Duration of filtering limited to 4:05:00 @22:40:17.56 Priming sample loop w/160ml, bypass w/40ml @22:40:17.58 SamplePump.setPosition! 0ml @22:40:17.61 Bypass.open @22:40:19.06 Intake.open @22:40:19.93 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @22:40:22.64 Vacuum tare is 19% of full scale @22:40:22.68 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @22:40:23.48 Pressure.reconfigure :default @22:40:23.55 Pressure.hold @22:40:24.11 SamplePump.reconfigure :prime @22:40:24.30 SamplePump.seek 40ml,3:51 @22:40:30.94 upsync @22:41:07.54 Pressure.coast @22:41:07.60 SamplePump.setPosition! 0ml @22:41:07.65 Exhaust.open @22:41:09.12 Bypass.close @22:41:10.96 Pressure.hold @22:41:11.54 SamplePump.reconfigure :prime @22:41:11.74 SamplePump.seek 160ml,13:51 @22:44:03.38 Pressure.coast @22:44:03.70 Bypass.open @22:44:04.54 Intake.close @22:44:05.70 Toroid.reconfigure :hold @22:44:05.76 Toroid.seek :crack @22:44:07.45 Toroid.seek :cartridge @22:44:08.41 Toroid.reconfigure :default @22:44:08.47 Toroid.seek :clear @22:44:10.32 Pressurizing cartridge to 26psi above 14.5psia @22:44:10.34 SamplePump.setPosition! 0ml @22:44:10.38 Intake.open @22:44:11.20 Bypass.close @22:44:11.90 Exhaust.close @22:44:12.75 Pressure.reconfigure :default @22:44:12.90 Pressure.hold @22:44:13.45 SamplePump.reconfigure :charge @22:44:13.63 SamplePump.seek 2.3ml,42 seconds @22:44:27.13 Pressure.coast @22:44:27.18 Intake.close @22:44:28.07 Pressurized cartridge with 0.825ml to 26.0psi @22:44:28.08 Isolated @35.7psia (21.2psi) @22:44:28.09 Checking for cartridge leaks from 35.7psia @22:44:28.11 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @22:44:37.81 Toroid.reconfigure :hold @22:44:37.87 Toroid.seek :bypass @22:44:39.49 Toroid.reconfigure :default @22:44:39.55 Toroid.seek :clear @22:44:41.16 Cartridge::Sampler::Leaky in FILTERING -- Isolated cartridge pressure fell by 20% to 28.5psia @22:44:41.18 Retry #1 of 2 @22:44:41.20 Cmd::SpareCartridge in FILTERING -- 0 spares remaining @22:44:41.23 Selecting Cartridge 5 @22:44:42.15 Air.seek :clear @22:44:46.53 Lysis1.seek :clear @22:44:46.89 Plunge.seek :clear @22:44:49.91 Carousel.seek 5 @22:44:50.43 Lysis1.seek :locked @22:44:53.88 Plunge.seek :locked @22:44:54.31 Air.seek :locked @22:44:58.54 Mounted archive Cartridge @22:44:58.58 Duration of filtering limited to 4:05:00 @22:44:58.80 Bypass.open @22:44:59.60 Toroid.reconfigure :hold @22:44:59.65 Toroid.seek :crack @22:45:01.44 Toroid.seek :cartridge @22:45:02.57 Toroid.reconfigure :default @22:45:02.66 Toroid.seek :clear @22:45:04.44 Exhaust.open @22:45:05.69 Pressurizing cartridge to 26psi above 14.6psia @22:45:05.71 SamplePump.setPosition! 0ml @22:45:05.75 Intake.open @22:45:06.58 Bypass.close @22:45:07.32 Exhaust.close @22:45:08.14 Pressure.reconfigure :default @22:45:08.74 Pressure.hold @22:45:08.96 SamplePump.seek 2.3ml,42 seconds @22:45:23.47 Pressure.coast @22:45:23.59 Intake.close @22:45:24.52 Pressurized cartridge with 0.950ml to 26.7psi @22:45:24.53 Isolated @36.3psia (21.7psi) @22:45:24.54 Checking for cartridge leaks from 36.3psia @22:45:24.56 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @22:45:29.75 Cartridge pressure fell 4% to 34.8psia @22:45:29.99 Sampling 2000.0ml @22:45:30.01 SamplePump.setPosition! 0ml @22:45:30.17 Exhaust.open @22:45:30.92 Intake.open @22:45:32.14 Pressure.reconfigure :default @22:45:32.23 Pressure.hold @22:45:32.84 SamplePump.reconfigure :sample @22:45:33.16 SamplePump.seek 2000ml,2:47:12 @22:45:35.17 Cartridge 5 sampling 1/2000ml at 0.63ml/s with 22.0psi vacuum @22:46:35.29 Cartridge 5 sampling 34/2000ml at 0.55ml/s with 24.4psi vacuum @22:47:35.17 Cartridge 5 sampling 67/2000ml at 0.55ml/s with 24.4psi vacuum @22:48:35.34 Cartridge 5 sampling 100/2000ml at 0.55ml/s with 23.8psi vacuum @22:49:35.16 Cartridge 5 sampling 133/2000ml at 0.55ml/s with 24.7psi vacuum @22:50:35.24 Cartridge 5 sampling 166/2000ml at 0.57ml/s with 24.1psi vacuum @22:51:35.17 Cartridge 5 sampling 200/2000ml at 0.56ml/s with 23.1psi vacuum @22:52:35.49 Cartridge 5 sampling 234/2000ml at 0.56ml/s with 24.8psi vacuum @22:53:35.31 Cartridge 5 sampling 268/2000ml at 0.55ml/s with 24.6psi vacuum @22:54:35.17 Cartridge 5 sampling 302/2000ml at 0.56ml/s with 24.9psi vacuum @22:55:35.25 Cartridge 5 sampling 336/2000ml at 0.55ml/s with 21.1psi vacuum @22:56:35.17 Cartridge 5 sampling 370/2000ml at 0.56ml/s with 23.8psi vacuum @22:57:35.45 Cartridge 5 sampling 404/2000ml at 0.56ml/s with 22.3psi vacuum @22:58:35.18 Cartridge 5 sampling 437/2000ml at 0.56ml/s with 26.5psi vacuum @22:59:35.39 Cartridge 5 sampling 471/2000ml at 0.56ml/s with 22.4psi vacuum @23:00:35.17 Cartridge 5 sampling 505/2000ml at 0.56ml/s with 21.0psi vacuum @23:01:35.36 Cartridge 5 sampling 538/2000ml at 0.56ml/s with 24.1psi vacuum @23:02:35.16 Cartridge 5 sampling 572/2000ml at 0.56ml/s with 24.2psi vacuum @23:03:35.31 Cartridge 5 sampling 606/2000ml at 0.54ml/s with 26.2psi vacuum @23:04:35.17 Cartridge 5 sampling 639/2000ml at 0.56ml/s with 24.3psi vacuum @23:05:35.31 Cartridge 5 sampling 672/2000ml at 0.56ml/s with 25.7psi vacuum @23:06:35.33 Cartridge 5 sampling 706/2000ml at 0.54ml/s with 23.6psi vacuum @23:07:35.17 Cartridge 5 sampling 739/2000ml at 0.55ml/s with 23.7psi vacuum @23:08:35.45 Cartridge 5 sampling 772/2000ml at 0.55ml/s with 21.6psi vacuum @23:09:35.17 Cartridge 5 sampling 805/2000ml at 0.56ml/s with 23.9psi vacuum @23:10:35.34 Cartridge 5 sampling 838/2000ml at 0.55ml/s with 21.7psi vacuum @23:11:35.17 Cartridge 5 sampling 871/2000ml at 0.55ml/s with 24.8psi vacuum @23:12:35.36 Cartridge 5 sampling 904/2000ml at 0.54ml/s with 24.6psi vacuum @23:13:35.16 Cartridge 5 sampling 936/2000ml at 0.54ml/s with 26.4psi vacuum @23:14:35.31 Cartridge 5 sampling 969/2000ml at 0.54ml/s with 25.3psi vacuum @23:15:35.17 Cartridge 5 sampling 1001/2000ml at 0.54ml/s with 25.0psi vacuum @23:16:35.27 Cartridge 5 sampling 1034/2000ml at 0.54ml/s with 24.4psi vacuum @23:17:35.17 Cartridge 5 sampling 1066/2000ml at 0.53ml/s with 24.0psi vacuum @23:18:35.17 Cartridge 5 sampling 1098/2000ml at 0.53ml/s with 26.0psi vacuum @23:19:35.44 Cartridge 5 sampling 1129/2000ml at 0.53ml/s with 24.4psi vacuum @23:20:35.17 Cartridge 5 sampling 1161/2000ml at 0.54ml/s with 23.1psi vacuum @23:21:35.33 Cartridge 5 sampling 1194/2000ml at 0.54ml/s with 20.6psi vacuum @23:22:35.17 Cartridge 5 sampling 1226/2000ml at 0.54ml/s with 25.6psi vacuum @23:23:35.40 Cartridge 5 sampling 1258/2000ml at 0.53ml/s with 24.1psi vacuum @23:24:35.17 Cartridge 5 sampling 1290/2000ml at 0.54ml/s with 22.9psi vacuum @23:25:35.27 Cartridge 5 sampling 1322/2000ml at 0.53ml/s with 25.1psi vacuum @23:26:35.17 Cartridge 5 sampling 1353/2000ml at 0.53ml/s with 25.2psi vacuum @23:27:35.20 Cartridge 5 sampling 1385/2000ml at 0.52ml/s with 23.9psi vacuum @23:28:35.17 Cartridge 5 sampling 1417/2000ml at 0.52ml/s with 25.6psi vacuum @23:29:35.52 Cartridge 5 sampling 1448/2000ml at 0.52ml/s with 24.0psi vacuum @23:30:35.49 Cartridge 5 sampling 1480/2000ml at 0.52ml/s with 27.1psi vacuum @23:31:35.17 Cartridge 5 sampling 1511/2000ml at 0.51ml/s with 22.5psi vacuum @23:32:35.47 Cartridge 5 sampling 1542/2000ml at 0.51ml/s with 25.0psi vacuum @23:33:35.17 Cartridge 5 sampling 1572/2000ml at 0.51ml/s with 22.5psi vacuum @23:34:35.31 Cartridge 5 sampling 1603/2000ml at 0.51ml/s with 24.7psi vacuum @23:35:35.17 Cartridge 5 sampling 1633/2000ml at 0.50ml/s with 21.5psi vacuum @23:36:35.36 Cartridge 5 sampling 1664/2000ml at 0.50ml/s with 22.4psi vacuum @23:37:35.17 Cartridge 5 sampling 1694/2000ml at 0.50ml/s with 21.1psi vacuum @23:38:35.33 Cartridge 5 sampling 1724/2000ml at 0.50ml/s with 24.9psi vacuum @23:39:35.17 Cartridge 5 sampling 1754/2000ml at 0.50ml/s with 24.9psi vacuum @23:40:35.21 Cartridge 5 sampling 1784/2000ml at 0.50ml/s with 23.0psi vacuum @23:41:35.16 Cartridge 5 sampling 1814/2000ml at 0.50ml/s with 23.2psi vacuum @23:42:35.36 Cartridge 5 sampling 1844/2000ml at 0.49ml/s with 21.8psi vacuum @23:43:35.30 Cartridge 5 sampling 1873/2000ml at 0.49ml/s with 25.8psi vacuum @23:44:35.17 Cartridge 5 sampling 1903/2000ml at 0.50ml/s with 25.5psi vacuum @23:45:35.34 Cartridge 5 sampling 1932/2000ml at 0.49ml/s with 21.9psi vacuum @23:46:35.16 Cartridge 5 sampling 1961/2000ml at 0.48ml/s with 21.2psi vacuum @23:47:35.36 Cartridge 5 sampling 1990/2000ml at 0.49ml/s with 23.1psi vacuum @23:47:54.89 Pressure.coast @23:47:55.25 Intake.close @23:47:55.35 Exhaust.close @23:47:55.67 Toroid.reconfigure :hold @23:47:57.37 Toroid.seek :bypass @23:47:59.00 Toroid.reconfigure :default @23:47:59.06 Toroid.seek :clear @23:48:00.68 Sampled 2000.0ml @23:48:01.14 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 5" @23:48:01.72 SamplePump.setPosition! 0ml @23:48:01.89 Plunge.seek :reagent @23:48:01.95 Bypass.open @23:48:02.89 Intake.open @23:48:03.06 Sample loop closed. Ready to Deliver Reagents @23:48:03.10 Archive.seek :sealed @23:48:03.97 SamplePump.reconfigure :slow @23:48:04.19 SamplePump.seek -3ml @23:48:06.86 Archive.seek :evac @23:48:10.87 delay 1 second @23:48:12.51 SamplePump.seek -2.8ml @23:48:13.39 delay 1 second @23:48:14.66 SamplePump.seek -3.3ml @23:48:16.08 delay 1 second @23:48:17.18 Exhaust.open @23:48:17.87 Bypass.close @23:48:18.33 upsync @23:48:18.67 SamplePump.seek -35.3ml @23:48:20.88 delay 1:00 @23:49:20.94 Archive.seek :clear @23:49:27.46 delay 3 seconds @23:49:30.55 Intake.close @23:49:30.61 Exhaust.close @23:49:38.15 RNAlater soak for 10:00 @23:59:38.23 Archive.seek :tVent @23:59:39.93 delay 1 second @23:59:41.01 Archive.seek :evac @23:59:57.06 delay 10 seconds @00:00:07.11PDT06-Jul-21 Sample preserved and evacuated @00:00:08.17
Carousel.seek 5 @00:00:08.63 Plunge.seek :locked @00:00:09.94 Archive.seek :clear @00:00:10.26 Air.seek :clear @00:00:10.72 Lysis1.seek :clear @00:00:27.20
Archive.coast @00:00:27.29 Air.coast @00:00:27.44 Lysis1.coast @00:00:27.50 Lysis2.coast @00:00:27.57 Plunge.coast @00:00:27.71 SamplePump.coast @00:00:27.79 Toroid.coast @00:00:27.99 Carousel.coast @00:00:28.33 Gate.power :motor,:OFF @00:00:28.34 Gate.power :core,:OFF @00:00:28.36 Gate.power :analytic1,:OFF @00:00:28.44 Gate.power :analytic2,:OFF @00:00:28.49 Gate.power :analytic3,:OFF @00:00:28.52 Gate.power :main,:OFF @00:00:28.57 Safely stopped and ready to power off @00:00:28.88 delayUntil "02:39:51.37" @00:00:28.91 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 02:39" @00:00:43.97 upsync @02:39:51.68
Gate.power :main,:ON,limit to 5000mA @02:39:51.69 Gate.power :analytic1,:ON,limit to 5000mA @02:39:51.70 Gate.power :analytic2,:ON,limit to 5000mA @02:39:51.72 Gate.power :analytic3,:ON,limit to 5000mA @02:39:51.78 Powering core bus @02:39:54.63 Archive.configure :default @02:39:54.76 Air.configure :default @02:39:54.87 Lysis1.configure :default @02:39:54.98 Lysis2.configure :default @02:39:55.12 Plunge.configure :default @02:39:55.26 SamplePump.configure :default @02:39:55.41 ExtraPump.configure :default @02:39:55.54 Toroid.configure :default @02:39:55.68 Carousel.configure :default @02:39:55.82 Pressure.configure :default @02:39:55.88 All elf controllers running firmware version 4.61 @02:39:55.90 Gate.power :motor,:ON,limit to 5000mA @02:39:56.08 Archive.reconfigure :default @02:39:56.36 Archive.home.jog 0 @02:39:56.59 Toroid.reconfigure :default @02:39:56.64 Plunge.reconfigure :default @02:39:56.67 Lysis2.reconfigure :default @02:39:56.73 Lysis1.reconfigure :default @02:39:56.77 Air.reconfigure :default @02:39:57.30 Toroid.home.jog 0 @02:39:57.38 Plunge.home.jog 0 @02:39:57.42 Lysis2.home.jog 0 @02:39:57.52 Lysis1.home.jog 0 @02:39:57.56 Air.home.jog 0 @02:39:58.93 Archive.reconfigure :homing @02:39:59.12 Archive.jog 1363 @02:39:59.52 Plunge.reconfigure :homing @02:39:59.60 Toroid.reconfigure :homing @02:39:59.73 Lysis2.reconfigure :homing @02:39:59.80 Air.reconfigure :homing @02:39:59.84 Lysis1.reconfigure :homing @02:40:00.38 Plunge.jog 843 @02:40:00.43 Toroid.jog 640 @02:40:00.49 Lysis2.jog 1346 @02:40:00.55 Air.jog 1350 @02:40:00.63 Lysis1.jog 1349 @02:40:01.22 Archive.setRawPosition! 50 @02:40:01.29 Archive.seek :home @02:40:01.73 Archive.reconfigure :default @02:40:01.84 Plunge.setRawPosition! 50 @02:40:01.89 Plunge.seek :home @02:40:02.18 Lysis2.setRawPosition! 50 @02:40:02.23 Lysis1.setRawPosition! 50 @02:40:02.32 Air.setRawPosition! 50 @02:40:02.48 Lysis2.seek :home @02:40:02.56 Lysis1.seek :home @02:40:02.81 Air.seek :home @02:40:03.40 Toroid.setRawPosition! 0 @02:40:03.44 Plunge.reconfigure :default @02:40:03.46 Lysis1.reconfigure :default @02:40:03.51 Lysis2.reconfigure :default @02:40:03.65 Toroid.seek :home @02:40:03.84 Toroid.reconfigure :default @02:40:03.87 Air.reconfigure :default @02:40:04.30
Toroid.seek :clear @02:40:05.92 Carousel.reconfigure :default @02:40:05.97 Carousel.home.jog 0 @02:40:06.46 Carousel.seek 5 @02:40:06.65 Selecting Cartridge 4 @02:40:06.88 Carousel.seek 4 @02:40:07.40 Lysis1.seek :locked @02:40:10.72 Plunge.seek :locked @02:40:11.11 Air.seek :locked @02:40:15.41
Mounted archive Cartridge @02:40:15.51 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 4" @02:40:15.82 Duration of filtering limited to 4:05:00 @02:40:16.11 Priming sample loop w/160ml, bypass w/40ml @02:40:16.14 SamplePump.setPosition! 0ml @02:40:16.21 Bypass.open @02:40:18.04 Intake.open @02:40:19.07 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @02:40:21.76 Vacuum tare is 19% of full scale @02:40:21.82 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @02:40:22.34 Pressure.reconfigure :default @02:40:22.42 Pressure.hold @02:40:22.82 SamplePump.reconfigure :prime @02:40:23.38 SamplePump.seek 40ml,3:51 @02:40:32.78 upsync @02:41:06.67 Pressure.coast @02:41:06.80 SamplePump.setPosition! 0ml @02:41:06.96 Exhaust.open @02:41:08.48 Bypass.close @02:41:10.12 Pressure.reconfigure :default @02:41:10.19 Pressure.hold @02:41:10.46 SamplePump.reconfigure :prime @02:41:10.53 SamplePump.seek 160ml,13:51 @02:44:02.30 Pressure.coast @02:44:02.35 Bypass.open @02:44:03.08 Intake.close @02:44:03.96 Toroid.reconfigure :hold @02:44:04.02 Toroid.seek :crack @02:44:05.61 Toroid.seek :cartridge @02:44:06.54 Toroid.reconfigure :default @02:44:06.60 Toroid.seek :clear @02:44:08.72 Pressurizing cartridge to 26psi above 14.5psia @02:44:08.75 SamplePump.setPosition! 0ml @02:44:08.85 Intake.open @02:44:09.69 Bypass.close @02:44:10.43 Exhaust.close @02:44:11.69 Pressure.reconfigure :default @02:44:11.93 Pressure.hold @02:44:12.29 SamplePump.reconfigure :charge @02:44:12.35 SamplePump.seek 2.3ml,42 seconds @02:44:26.16 Pressure.coast @02:44:26.30 Intake.close @02:44:27.31 Pressurized cartridge with 0.849ml to 26.2psi @02:44:27.36 Isolated @35.2psia (20.7psi) @02:44:27.44 Checking for cartridge leaks from 35.2psia @02:44:27.52 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @02:44:32.10 Cartridge pressure fell 2% to 34.6psia @02:44:32.18 Sampling 2000.0ml @02:44:32.19 SamplePump.setPosition! 0ml @02:44:32.23 Exhaust.open @02:44:32.93 Intake.open @02:44:33.88 Pressure.reconfigure :default @02:44:33.93 Pressure.hold @02:44:34.06 Sea 2.7m deep (18.7psia) @02:44:34.21 SamplePump.reconfigure :sample @02:44:34.28 SamplePump.seek 2000ml,2:47:12 @02:44:37.08 Cartridge 4 sampling 1/2000ml at 0.55ml/s with 25.7psi vacuum @02:45:36.93 Cartridge 4 sampling 32/2000ml at 0.52ml/s with 24.4psi vacuum @02:46:37.14 Cartridge 4 sampling 64/2000ml at 0.52ml/s with 24.3psi vacuum @02:47:36.94 Cartridge 4 sampling 96/2000ml at 0.53ml/s with 24.4psi vacuum @02:48:37.07 Cartridge 4 sampling 127/2000ml at 0.53ml/s with 21.9psi vacuum @02:49:36.94 Cartridge 4 sampling 160/2000ml at 0.54ml/s with 22.5psi vacuum @02:50:37.00 Cartridge 4 sampling 192/2000ml at 0.55ml/s with 25.0psi vacuum @02:51:37.01 Cartridge 4 sampling 225/2000ml at 0.55ml/s with 24.0psi vacuum @02:52:36.94 Cartridge 4 sampling 258/2000ml at 0.54ml/s with 22.2psi vacuum @02:53:37.35 Cartridge 4 sampling 290/2000ml at 0.53ml/s with 22.8psi vacuum @02:54:36.94 Cartridge 4 sampling 322/2000ml at 0.54ml/s with 23.3psi vacuum @02:55:37.10 Cartridge 4 sampling 355/2000ml at 0.54ml/s with 25.7psi vacuum @02:56:36.94 Cartridge 4 sampling 388/2000ml at 0.55ml/s with 22.0psi vacuum @02:57:37.13 Cartridge 4 sampling 421/2000ml at 0.55ml/s with 22.5psi vacuum @02:58:36.94 Cartridge 4 sampling 454/2000ml at 0.56ml/s with 24.3psi vacuum @02:59:37.03 Cartridge 4 sampling 487/2000ml at 0.54ml/s with 24.0psi vacuum @03:00:36.94 Cartridge 4 sampling 519/2000ml at 0.53ml/s with 22.4psi vacuum @03:01:37.01 Cartridge 4 sampling 551/2000ml at 0.53ml/s with 23.7psi vacuum @03:02:36.93 Cartridge 4 sampling 584/2000ml at 0.53ml/s with 22.6psi vacuum @03:03:36.97 Cartridge 4 sampling 616/2000ml at 0.53ml/s with 26.4psi vacuum @03:04:37.11 Cartridge 4 sampling 648/2000ml at 0.54ml/s with 24.1psi vacuum @03:05:36.94 Cartridge 4 sampling 681/2000ml at 0.54ml/s with 24.8psi vacuum @03:06:37.29 Cartridge 4 sampling 713/2000ml at 0.54ml/s with 24.9psi vacuum @03:07:36.94 Cartridge 4 sampling 745/2000ml at 0.54ml/s with 25.0psi vacuum @03:08:37.09 Cartridge 4 sampling 777/2000ml at 0.53ml/s with 22.9psi vacuum @03:09:36.94 Cartridge 4 sampling 810/2000ml at 0.54ml/s with 24.0psi vacuum @03:10:37.07 Cartridge 4 sampling 842/2000ml at 0.53ml/s with 25.0psi vacuum @03:11:36.94 Cartridge 4 sampling 874/2000ml at 0.54ml/s with 24.8psi vacuum @03:12:37.13 Cartridge 4 sampling 906/2000ml at 0.52ml/s with 21.3psi vacuum @03:13:36.94 Cartridge 4 sampling 937/2000ml at 0.53ml/s with 24.2psi vacuum @03:14:37.10 Cartridge 4 sampling 969/2000ml at 0.52ml/s with 25.6psi vacuum @03:15:36.94 Cartridge 4 sampling 1001/2000ml at 0.53ml/s with 22.9psi vacuum @03:16:37.01 Cartridge 4 sampling 1032/2000ml at 0.52ml/s with 25.3psi vacuum @03:17:36.94 Cartridge 4 sampling 1064/2000ml at 0.53ml/s with 25.4psi vacuum @03:18:37.05 Cartridge 4 sampling 1095/2000ml at 0.52ml/s with 23.8psi vacuum @03:19:37.33 Cartridge 4 sampling 1126/2000ml at 0.53ml/s with 24.3psi vacuum @03:20:36.94 Cartridge 4 sampling 1158/2000ml at 0.51ml/s with 22.1psi vacuum @03:21:37.09 Cartridge 4 sampling 1189/2000ml at 0.52ml/s with 22.7psi vacuum @03:22:36.94 Cartridge 4 sampling 1220/2000ml at 0.51ml/s with 25.4psi vacuum @03:23:37.10 Cartridge 4 sampling 1251/2000ml at 0.52ml/s with 25.1psi vacuum @03:24:37.49 Cartridge 4 sampling 1282/2000ml at 0.51ml/s with 21.1psi vacuum @03:25:37.05 Cartridge 4 sampling 1313/2000ml at 0.51ml/s with 21.8psi vacuum @03:26:36.93 Cartridge 4 sampling 1343/2000ml at 0.51ml/s with 22.6psi vacuum @03:27:37.08 Cartridge 4 sampling 1374/2000ml at 0.51ml/s with 22.8psi vacuum @03:28:36.93 Cartridge 4 sampling 1405/2000ml at 0.51ml/s with 21.2psi vacuum @03:29:37.07 Cartridge 4 sampling 1435/2000ml at 0.51ml/s with 24.3psi vacuum @03:30:36.93 Cartridge 4 sampling 1466/2000ml at 0.51ml/s with 22.8psi vacuum @03:31:36.98 Cartridge 4 sampling 1496/2000ml at 0.49ml/s with 26.0psi vacuum @03:32:37.43 Cartridge 4 sampling 1526/2000ml at 0.49ml/s with 26.9psi vacuum @03:33:36.94 Cartridge 4 sampling 1556/2000ml at 0.51ml/s with 24.8psi vacuum @03:34:37.14 Cartridge 4 sampling 1586/2000ml at 0.50ml/s with 24.0psi vacuum @03:35:36.94 Cartridge 4 sampling 1616/2000ml at 0.51ml/s with 23.1psi vacuum @03:36:37.20 Cartridge 4 sampling 1646/2000ml at 0.49ml/s with 21.2psi vacuum @03:37:36.94 Cartridge 4 sampling 1676/2000ml at 0.50ml/s with 22.2psi vacuum @03:38:37.04 Cartridge 4 sampling 1705/2000ml at 0.49ml/s with 20.6psi vacuum @03:39:36.94 Cartridge 4 sampling 1735/2000ml at 0.49ml/s with 21.3psi vacuum @03:40:37.13 Cartridge 4 sampling 1765/2000ml at 0.49ml/s with 26.2psi vacuum @03:41:36.93 Cartridge 4 sampling 1794/2000ml at 0.48ml/s with 27.2psi vacuum @03:42:37.11 Cartridge 4 sampling 1823/2000ml at 0.48ml/s with 21.4psi vacuum 1.6m deep (17.1psia) @03:43:36.94 Cartridge 4 sampling 1852/2000ml at 0.48ml/s with 26.5psi vacuum @03:44:37.00 Cartridge 4 sampling 1882/2000ml at 0.50ml/s with 21.6psi vacuum @03:45:36.94 Cartridge 4 sampling 1911/2000ml at 0.49ml/s with 24.9psi vacuum @03:46:37.01 Cartridge 4 sampling 1941/2000ml at 0.49ml/s with 22.9psi vacuum @03:47:37.33 Cartridge 4 sampling 1970/2000ml at 0.49ml/s with 24.1psi vacuum @03:48:36.94 Cartridge 4 sampling 1998/2000ml at 0.48ml/s with 21.5psi vacuum @03:48:40.09 Pressure.coast @03:48:40.22 Intake.close @03:48:40.27 Exhaust.close @03:48:40.46 Toroid.reconfigure :hold @03:48:40.51 Toroid.seek :bypass @03:48:42.22 Toroid.reconfigure :default @03:48:42.32 Toroid.seek :clear @03:48:44.03 Sampled 2000.0ml @03:48:44.86 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 4" @03:48:45.97 SamplePump.setPosition! 0ml @03:48:46.02 Plunge.seek :reagent @03:48:46.06 Bypass.open @03:48:46.79 Intake.open @03:48:47.03 Sample loop closed. Ready to Deliver Reagents @03:48:47.07 Archive.seek :sealed @03:48:47.69 SamplePump.reconfigure :slow @03:48:47.81 SamplePump.seek -3ml @03:48:50.12 upsync @03:48:50.79 Archive.seek :evac @03:48:54.67 delay 1 second @03:48:55.81 SamplePump.seek -2.8ml @03:48:56.50 delay 1 second @03:48:57.59 SamplePump.seek -3.3ml @03:48:58.92 delay 1 second @03:48:59.95 Exhaust.open @03:49:00.74 Bypass.close @03:49:01.66 SamplePump.seek -35.3ml @03:49:04.82 delay 1:00 @03:50:04.87 Archive.seek :clear @03:50:10.43 delay 3 seconds @03:50:13.81 Intake.close @03:50:13.93 Exhaust.close @03:50:22.03 RNAlater soak for 10:00 @04:00:22.07 Archive.seek :tVent @04:00:23.73 delay 1 second @04:00:24.77 Archive.seek :evac @04:00:40.78 delay 10 seconds @04:00:50.85 Sample preserved and evacuated @04:00:51.07
Carousel.seek 4 @04:00:51.28 Plunge.seek :locked @04:00:52.25 Archive.seek :clear @04:00:52.42 Air.seek :clear @04:00:52.61 Lysis1.seek :clear @04:01:09.42
Archive.coast @04:01:09.46 Air.coast @04:01:09.51 Lysis1.coast @04:01:09.55 Lysis2.coast @04:01:10.13 Plunge.coast @04:01:10.18 SamplePump.coast @04:01:10.22 Toroid.coast @04:01:10.27 Carousel.coast @04:01:10.46 Gate.power :motor,:OFF @04:01:10.47 Gate.power :core,:OFF @04:01:10.48 Gate.power :analytic1,:OFF @04:01:10.50 Gate.power :analytic2,:OFF @04:01:10.51 Gate.power :analytic3,:OFF @04:01:10.52 Gate.power :main,:OFF @04:01:10.54 Safely stopped and ready to power off @04:01:10.75 delayUntil "06:39:51.37" @04:01:10.77 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 06:39" @04:01:15.89 upsync @06:39:51.51
Gate.power :main,:ON,limit to 5000mA @06:39:51.53 Gate.power :analytic1,:ON,limit to 5000mA @06:39:51.54 Gate.power :analytic2,:ON,limit to 5000mA @06:39:51.55 Gate.power :analytic3,:ON,limit to 5000mA @06:39:51.59 Powering core bus @06:39:54.25 Archive.configure :default @06:39:54.55 Air.configure :default @06:39:54.80 Lysis1.configure :default @06:39:55.10 Lysis2.configure :default @06:39:55.39 Plunge.configure :default @06:39:55.92 SamplePump.configure :default @06:39:56.17 ExtraPump.configure :default @06:39:56.47 Toroid.configure :default @06:39:56.77 Carousel.configure :default @06:39:57.08 Pressure.configure :default @06:39:57.19 All elf controllers running firmware version 4.61 @06:39:57.28 Gate.power :motor,:ON,limit to 5000mA @06:39:57.95 Toroid.reconfigure :default @06:39:58.05 Archive.reconfigure :default @06:39:58.10 Plunge.reconfigure :default @06:39:58.15 Lysis1.reconfigure :default @06:39:58.19 Air.reconfigure :default @06:39:58.23 Lysis2.reconfigure :default @06:39:58.46 Toroid.home.jog 0 @06:39:58.50 Archive.home.jog 0 @06:39:58.52 Plunge.home.jog 0 @06:39:58.54 Lysis1.home.jog 0 @06:39:58.57 Lysis2.home.jog 0 @06:39:58.62 Air.home.jog 0 @06:39:59.21 Toroid.reconfigure :homing @06:39:59.36 Plunge.reconfigure :homing @06:39:59.43 Lysis1.reconfigure :homing @06:39:59.46 Archive.reconfigure :homing @06:39:59.60 Toroid.jog 637 @06:39:59.64 Plunge.jog 840 @06:39:59.73 Archive.jog 1364 @06:39:59.79 Lysis1.jog 1357 @06:39:59.97 Lysis2.reconfigure :homing @06:39:59.99 Air.reconfigure :homing @06:40:00.10 Air.jog 1355 @06:40:00.12 Lysis2.jog 1348 @06:40:00.65 Plunge.setRawPosition! 50 @06:40:00.70 Plunge.seek :home @06:40:00.99 Archive.setRawPosition! 50 @06:40:01.01 Lysis1.setRawPosition! 50 @06:40:01.10 Archive.seek :home @06:40:01.15 Lysis1.seek :home @06:40:01.37 Plunge.reconfigure :default @06:40:01.40 Lysis2.setRawPosition! 50 @06:40:01.43 Air.setRawPosition! 50 @06:40:01.57 Lysis2.seek :home @06:40:01.61 Air.seek :home @06:40:01.78 Lysis1.reconfigure :default @06:40:01.81 Archive.reconfigure :default @06:40:02.41 Toroid.setRawPosition! 0 @06:40:02.56 Lysis2.reconfigure :default @06:40:02.63 Air.reconfigure :default @06:40:02.93 Toroid.seek :home @06:40:03.49 Toroid.reconfigure :default @06:40:03.91
Toroid.seek :clear @06:40:05.61 Carousel.reconfigure :default @06:40:05.71 Carousel.home.jog 0 @06:40:06.45 Carousel.seek 4 @06:40:06.82 Selecting Cartridge 3 @06:40:07.32 Carousel.seek 3 @06:40:07.92 Lysis1.seek :locked @06:40:11.14 Plunge.seek :locked @06:40:11.82 Air.seek :locked @06:40:16.07
Mounted archive Cartridge @06:40:16.53 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 3" @06:40:16.67 Errno::ECONNREFUSED in Qmail -- Connection refused - connect(2) @06:40:16.78 Retry #1 of 10 after 10:00 @06:40:17.09 Duration of filtering limited to 4:05:00 @06:40:17.38 Priming sample loop w/160ml, bypass w/40ml @06:40:17.40 SamplePump.setPosition! 0ml @06:40:17.44 Bypass.open @06:40:18.91 Intake.open @06:40:19.79 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @06:40:22.48 Vacuum tare is 19% of full scale @06:40:22.52 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @06:40:23.01 Pressure.reconfigure :default @06:40:23.16 Pressure.hold @06:40:23.70 SamplePump.reconfigure :prime @06:40:23.89 SamplePump.seek 40ml,3:51 @06:41:07.16 Pressure.coast @06:41:07.21 SamplePump.setPosition! 0ml @06:41:07.26 Exhaust.open @06:41:08.77 Bypass.close @06:41:10.53 Pressure.reconfigure :default @06:41:10.69 Pressure.hold @06:41:11.28 SamplePump.reconfigure :prime @06:41:11.35 SamplePump.seek 160ml,13:51 @06:44:03.20 Pressure.coast @06:44:03.36 Bypass.open @06:44:04.10 Intake.close @06:44:05.14 Toroid.reconfigure :hold @06:44:05.29 Toroid.seek :crack @06:44:06.96 Toroid.seek :cartridge @06:44:07.89 Toroid.reconfigure :default @06:44:07.95 Toroid.seek :clear @06:44:09.79 Pressurizing cartridge to 26psi above 14.7psia @06:44:09.81 SamplePump.setPosition! 0ml @06:44:09.85 Intake.open @06:44:10.65 Bypass.close @06:44:11.38 Exhaust.close @06:44:12.74 Pressure.reconfigure :default @06:44:12.88 Pressure.hold @06:44:13.39 SamplePump.reconfigure :charge @06:44:13.48 SamplePump.seek 2.3ml,42 seconds @06:44:26.81 Pressure.coast @06:44:26.86 Intake.close @06:44:27.70 Pressurized cartridge with 0.805ml to 26.3psi @06:44:27.71 Isolated @36.2psia (21.5psi) @06:44:27.72 Checking for cartridge leaks from 36.2psia @06:44:27.74 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @06:44:34.83 Cartridge pressure fell 12% to 31.7psia @06:44:35.09 Sampling 2000.0ml @06:44:35.12 SamplePump.setPosition! 0ml @06:44:35.36 Exhaust.open @06:44:36.15 Intake.open @06:44:37.09 Pressure.reconfigure :default @06:44:37.15 Pressure.hold @06:44:37.43 SamplePump.reconfigure :sample @06:44:37.51 SamplePump.seek 2000ml,2:47:12 @06:44:40.31 Cartridge 3 sampling 1/2000ml at 0.63ml/s with 24.8psi vacuum @06:45:40.16 Cartridge 3 sampling 34/2000ml at 0.55ml/s with 24.2psi vacuum @06:46:40.43 Cartridge 3 sampling 67/2000ml at 0.54ml/s with 24.6psi vacuum @06:47:40.16 Cartridge 3 sampling 99/2000ml at 0.54ml/s with 22.4psi vacuum @06:48:40.26 Cartridge 3 sampling 132/2000ml at 0.54ml/s with 21.2psi vacuum @06:49:40.16 Cartridge 3 sampling 165/2000ml at 0.55ml/s with 24.7psi vacuum @06:50:32.41 upsync @06:50:40.32 Cartridge 3 sampling 198/2000ml at 0.55ml/s with 22.0psi vacuum @06:51:40.16 Cartridge 3 sampling 231/2000ml at 0.55ml/s with 24.5psi vacuum @06:52:41.69 Cartridge 3 sampling 264/2000ml at 0.54ml/s with 23.3psi vacuum @06:53:40.16 Cartridge 3 sampling 297/2000ml at 0.56ml/s with 25.8psi vacuum @06:54:40.28 Cartridge 3 sampling 331/2000ml at 0.56ml/s with 25.9psi vacuum @06:55:19.31 Session Begun @06:55:31.31 -> BACsampler.chargeBleed @06:55:40.17 Cartridge 3 sampling 364/2000ml at 0.56ml/s with 26.1psi vacuum @06:56:06.33 -> BACsampler.chargeBleed=0.75 @06:56:18.12 -> BACsampler.chargeBleed @06:56:40.30 Cartridge 3 sampling 398/2000ml at 0.57ml/s with 23.3psi vacuum @06:56:44.40 -> Cartridge.release 2 @06:56:54.38 -> Cartridge.dry @06:56:59.33 -> exit @06:56:59.35 Session Ended @06:57:40.15 Cartridge 3 sampling 432/2000ml at 0.56ml/s with 22.9psi vacuum @06:58:40.30 Cartridge 3 sampling 465/2000ml at 0.56ml/s with 25.2psi vacuum @06:59:40.16 Cartridge 3 sampling 499/2000ml at 0.55ml/s with 26.4psi vacuum @07:00:40.27 Cartridge 3 sampling 532/2000ml at 0.54ml/s with 23.0psi vacuum @07:01:40.35 Cartridge 3 sampling 566/2000ml at 0.56ml/s with 22.1psi vacuum @07:02:40.16 Cartridge 3 sampling 599/2000ml at 0.55ml/s with 26.6psi vacuum @07:03:40.48 Cartridge 3 sampling 632/2000ml at 0.56ml/s with 24.8psi vacuum @07:04:40.16 Cartridge 3 sampling 666/2000ml at 0.53ml/s with 23.4psi vacuum @07:05:40.30 Cartridge 3 sampling 699/2000ml at 0.56ml/s with 26.1psi vacuum @07:06:40.29 Cartridge 3 sampling 732/2000ml at 0.55ml/s with 22.0psi vacuum @07:07:40.33 Cartridge 3 sampling 764/2000ml at 0.55ml/s with 22.7psi vacuum @07:08:40.16 Cartridge 3 sampling 797/2000ml at 0.54ml/s with 26.5psi vacuum @07:09:40.37 Cartridge 3 sampling 830/2000ml at 0.54ml/s with 20.1psi vacuum 1.8m deep (17.3psia) @07:10:40.16 Cartridge 3 sampling 863/2000ml at 0.53ml/s with 27.5psi vacuum @07:11:40.28 Cartridge 3 sampling 895/2000ml at 0.54ml/s with 26.5psi vacuum @07:12:40.15 Cartridge 3 sampling 928/2000ml at 0.54ml/s with 24.4psi vacuum @07:13:40.32 Cartridge 3 sampling 960/2000ml at 0.54ml/s with 25.5psi vacuum @07:14:40.15 Cartridge 3 sampling 993/2000ml at 0.54ml/s with 25.7psi vacuum @07:15:40.22 Cartridge 3 sampling 1025/2000ml at 0.54ml/s with 26.2psi vacuum @07:16:40.49 Cartridge 3 sampling 1057/2000ml at 0.53ml/s with 26.4psi vacuum @07:17:40.15 Cartridge 3 sampling 1089/2000ml at 0.52ml/s with 19.7psi vacuum @07:18:40.29 Cartridge 3 sampling 1121/2000ml at 0.53ml/s with 22.9psi vacuum @07:19:40.16 Cartridge 3 sampling 1153/2000ml at 0.52ml/s with 27.5psi vacuum @07:20:40.29 Cartridge 3 sampling 1184/2000ml at 0.52ml/s with 26.2psi vacuum @07:21:40.16 Cartridge 3 sampling 1216/2000ml at 0.52ml/s with 22.4psi vacuum @07:22:40.27 Cartridge 3 sampling 1248/2000ml at 0.53ml/s with 23.3psi vacuum @07:23:40.16 Cartridge 3 sampling 1279/2000ml at 0.51ml/s with 24.2psi vacuum @07:24:40.31 Cartridge 3 sampling 1311/2000ml at 0.54ml/s with 25.4psi vacuum @07:25:40.15 Cartridge 3 sampling 1342/2000ml at 0.52ml/s with 25.9psi vacuum @07:26:40.30 Cartridge 3 sampling 1373/2000ml at 0.51ml/s with 27.7psi vacuum @07:27:40.16 Cartridge 3 sampling 1404/2000ml at 0.53ml/s with 23.9psi vacuum @07:28:40.23 Cartridge 3 sampling 1435/2000ml at 0.51ml/s with 22.0psi vacuum @07:29:40.47 Cartridge 3 sampling 1466/2000ml at 0.52ml/s with 25.3psi vacuum @07:30:40.15 Cartridge 3 sampling 1497/2000ml at 0.52ml/s with 25.1psi vacuum @07:31:40.24 Cartridge 3 sampling 1528/2000ml at 0.50ml/s with 28.1psi vacuum @07:32:40.15 Cartridge 3 sampling 1559/2000ml at 0.50ml/s with 28.0psi vacuum @07:33:40.40 Cartridge 3 sampling 1589/2000ml at 0.50ml/s with 20.8psi vacuum 2.5m deep (18.5psia) @07:34:40.15 Cartridge 3 sampling 1620/2000ml at 0.49ml/s with 21.7psi vacuum @07:35:40.48 Cartridge 3 sampling 1651/2000ml at 0.52ml/s with 21.2psi vacuum @07:36:40.15 Cartridge 3 sampling 1681/2000ml at 0.52ml/s with 22.2psi vacuum @07:37:40.32 Cartridge 3 sampling 1711/2000ml at 0.52ml/s with 20.1psi vacuum @07:38:40.15 Cartridge 3 sampling 1742/2000ml at 0.49ml/s with 19.3psi vacuum @07:39:40.34 Cartridge 3 sampling 1772/2000ml at 0.51ml/s with 24.4psi vacuum @07:40:40.16 Cartridge 3 sampling 1802/2000ml at 0.50ml/s with 26.8psi vacuum @07:41:40.28 Cartridge 3 sampling 1831/2000ml at 0.50ml/s with 26.5psi vacuum @07:42:40.15 Cartridge 3 sampling 1861/2000ml at 0.50ml/s with 19.8psi vacuum @07:43:40.25 Cartridge 3 sampling 1890/2000ml at 0.50ml/s with 19.7psi vacuum @07:44:40.15 Cartridge 3 sampling 1920/2000ml at 0.48ml/s with 28.7psi vacuum @07:45:40.16 Cartridge 3 sampling 1949/2000ml at 0.48ml/s with 19.3psi vacuum @07:46:40.45 Cartridge 3 sampling 1979/2000ml at 0.47ml/s with 28.6psi vacuum @07:47:24.21 Pressure.coast @07:47:24.35 Intake.close @07:47:24.43 Exhaust.close @07:47:24.60 Toroid.reconfigure :hold @07:47:24.66 Toroid.seek :bypass @07:47:26.37 Toroid.reconfigure :default @07:47:26.44 Toroid.seek :clear @07:47:28.21 Sampled 2000.0ml @07:47:29.28 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 3" @07:47:30.10 SamplePump.setPosition! 0ml @07:47:30.25 Plunge.seek :reagent @07:47:30.39 Bypass.open @07:47:32.03 Intake.open @07:47:32.18 Sample loop closed. Ready to Deliver Reagents @07:47:32.20 Archive.seek :sealed @07:47:32.88 SamplePump.reconfigure :slow @07:47:32.94 SamplePump.seek -3ml @07:47:35.78 upsync @07:47:36.03 Archive.seek :evac @07:47:39.68 delay 1 second @07:47:40.80 SamplePump.seek -2.8ml @07:47:41.49 delay 1 second @07:47:42.59 SamplePump.seek -3.3ml @07:47:43.92 delay 1 second @07:47:44.95 Exhaust.open @07:47:45.70 Bypass.close @07:47:46.73 SamplePump.seek -35.3ml @07:47:50.11 delay 1:00 @07:48:50.18 Archive.seek :clear @07:48:55.54 delay 3 seconds @07:48:58.71 Intake.close @07:48:58.87 Exhaust.close @07:49:07.36 RNAlater soak for 10:00 @07:59:07.41 Archive.seek :tVent @07:59:09.11 delay 1 second @07:59:10.14 Archive.seek :evac @07:59:26.16 delay 10 seconds @07:59:36.30 Sample preserved and evacuated @07:59:36.53
Carousel.seek 3 @07:59:36.74 Plunge.seek :locked @07:59:37.72 Archive.seek :clear @07:59:37.88 Air.seek :clear @07:59:38.06 Lysis1.seek :clear @07:59:54.92
Archive.coast @07:59:54.97 Air.coast @07:59:55.01 Lysis1.coast @07:59:55.05 Lysis2.coast @07:59:55.09 Plunge.coast @07:59:55.14 SamplePump.coast @07:59:55.18 Toroid.coast @07:59:55.22 Carousel.coast @07:59:55.42 Gate.power :motor,:OFF @07:59:55.43 Gate.power :core,:OFF @07:59:55.45 Gate.power :analytic1,:OFF @07:59:55.46 Gate.power :analytic2,:OFF @07:59:55.47 Gate.power :analytic3,:OFF @07:59:55.49 Gate.power :main,:OFF @07:59:55.50 Safely stopped and ready to power off @07:59:55.70 delayUntil "10:39:51.37" @07:59:55.73 GoodNews.email "Sampling next dry cartridges",Subject: "Shutdown until 10:39" @08:00:00.61 upsync @10:39:51.56
Gate.power :main,:ON,limit to 5000mA @10:39:51.57 Gate.power :analytic1,:ON,limit to 5000mA @10:39:51.60 Gate.power :analytic2,:ON,limit to 5000mA @10:39:51.64 Gate.power :analytic3,:ON,limit to 5000mA @10:39:51.74 Powering core bus @10:39:54.12 Archive.configure :default @10:39:54.26 Air.configure :default @10:39:54.37 Lysis1.configure :default @10:39:54.48 Lysis2.configure :default @10:39:54.62 Plunge.configure :default @10:39:54.76 SamplePump.configure :default @10:39:54.89 ExtraPump.configure :default @10:39:55.04 Toroid.configure :default @10:39:55.18 Carousel.configure :default @10:39:55.31 Pressure.configure :default @10:39:55.37 All elf controllers running firmware version 4.61 @10:39:55.39 Gate.power :motor,:ON,limit to 5000mA @10:39:55.65 Archive.reconfigure :default @10:39:55.99 Toroid.reconfigure :default @10:39:56.03 Archive.home.jog 0 @10:39:56.50 Plunge.reconfigure :default @10:39:56.64 Lysis2.reconfigure :default @10:39:56.68 Lysis1.reconfigure :default @10:39:56.73 Air.reconfigure :default @10:39:57.12 Toroid.home.jog 0 @10:39:57.26 Lysis2.home.jog 0 @10:39:57.38 Lysis1.home.jog 0 @10:39:57.45 Plunge.home.jog 0 @10:39:57.56 Air.home.jog 0 @10:39:59.43 Plunge.reconfigure :homing @10:39:59.57 Lysis1.reconfigure :homing @10:39:59.68 Archive.reconfigure :homing @10:39:59.75 Toroid.reconfigure :homing @10:39:59.84 Air.reconfigure :homing @10:39:59.98 Lysis2.reconfigure :homing @10:40:00.43 Plunge.jog 836 @10:40:00.60 Lysis1.jog 1358 @10:40:00.67 Archive.jog 1366 @10:40:00.69 Toroid.jog 643 @10:40:00.81 Air.jog 1354 @10:40:00.90 Lysis2.jog 1350 @10:40:01.87 Plunge.setRawPosition! 50 @10:40:02.05 Plunge.seek :home @10:40:02.14 Lysis1.setRawPosition! 50 @10:40:02.38 Lysis1.seek :home @10:40:02.57 Air.setRawPosition! 50 @10:40:02.78 Archive.setRawPosition! 50 @10:40:02.91 Plunge.reconfigure :default @10:40:03.04 Lysis2.setRawPosition! 50 @10:40:03.26 Air.seek :home @10:40:03.32 Archive.seek :home @10:40:03.36 Lysis2.seek :home @10:40:03.58 Lysis1.reconfigure :default @10:40:03.61 Toroid.setRawPosition! 0 @10:40:03.77 Toroid.seek :home @10:40:04.07 Toroid.reconfigure :default @10:40:04.11 Lysis2.reconfigure :default @10:40:04.15 Archive.reconfigure :default @10:40:04.19 Air.reconfigure :default @10:40:04.71
Toroid.seek :clear @10:40:06.33 Carousel.reconfigure :default @10:40:06.38 Carousel.home.jog 0 @10:40:06.89 Carousel.seek 3 @10:40:07.08 Selecting Cartridge 2 @10:40:07.29 Carousel.seek 2 @10:40:08.38 Lysis1.seek :locked @10:40:11.89 Plunge.seek :locked @10:40:12.10 Air.seek :locked @10:40:16.37
Mounted archive Cartridge @10:40:16.50 GoodNews.email "Filtering up to 2000ml",Subject: "Sampling archive_bac cartridge in slot 2" @10:40:16.80 Duration of filtering limited to 4:05:00 @10:40:17.48 Priming sample loop w/160ml, bypass w/40ml @10:40:17.56 SamplePump.setPosition! 0ml @10:40:17.65 Bypass.open @10:40:19.29 Intake.open @10:40:20.65 Waiting up to 20 seconds for sea pressure to stabilize +/-0.50psi @10:40:22.12 upsync @10:40:23.30 Vacuum tare is 19% of full scale @10:40:23.32 Cartridge::Sampler::Warning in FILTERING -- Vacuum sensor saturates below -18psi @10:40:23.53 Pressure.reconfigure :default @10:40:23.59 Pressure.hold @10:40:23.80 SamplePump.reconfigure :prime @10:40:23.87 SamplePump.seek 40ml,3:51 @10:41:07.14 Pressure.coast @10:41:07.19 SamplePump.setPosition! 0ml @10:41:07.40 Exhaust.open @10:41:08.88 Bypass.close @10:41:10.42 Pressure.reconfigure :default @10:41:10.47 Pressure.hold @10:41:10.74 SamplePump.reconfigure :prime @10:41:10.80 SamplePump.seek 160ml,13:51 @10:44:02.49 Pressure.coast @10:44:02.58 Bypass.open @10:44:03.32 Intake.close @10:44:04.80 Toroid.reconfigure :hold @10:44:04.86 Toroid.seek :crack @10:44:06.44 Toroid.seek :cartridge @10:44:07.38 Toroid.reconfigure :default @10:44:07.43 Toroid.seek :clear @10:44:09.64 Pressurizing cartridge to 26psi above 14.5psia @10:44:09.66 SamplePump.setPosition! 0ml @10:44:09.71 Intake.open @10:44:10.59 Bypass.close @10:44:11.31 Exhaust.close @10:44:12.53 Pressure.reconfigure :default @10:44:12.66 Pressure.hold @10:44:12.87 SamplePump.reconfigure :charge @10:44:12.94 SamplePump.seek 2.3ml,42 seconds @10:44:26.41 Pressure.coast @10:44:26.56 Intake.close @10:44:27.61 Pressurized cartridge with 0.814ml to 26.0psi @10:44:27.62 Isolated @33.4psia (18.9psi) @10:44:27.63 Checking for cartridge leaks from 33.4psia @10:44:27.66 Waiting up to 20 seconds for cartridge pressure to stabilize +/-0.75psi @10:44:32.25 Cartridge pressure fell 4% to 32.2psia @10:44:32.33 Sampling 2000.0ml @10:44:32.35 SamplePump.setPosition! 0ml @10:44:32.39 Exhaust.open @10:44:33.07 Intake.open @10:44:34.01 Pressure.reconfigure :default @10:44:34.07 Pressure.hold @10:44:34.20 Sea 2.0m deep (17.6psia) @10:44:34.36 SamplePump.reconfigure :sample @10:44:34.42 SamplePump.seek 2000ml,2:47:12 @10:44:37.27 Cartridge 2 sampling 1/2000ml at 0.58ml/s with 24.0psi vacuum @10:45:37.07 Cartridge 2 sampling 32/2000ml at 0.53ml/s with 24.9psi vacuum @10:46:37.21 Cartridge 2 sampling 64/2000ml at 0.52ml/s with 24.3psi vacuum @10:47:37.07 Cartridge 2 sampling 95/2000ml at 0.52ml/s with 24.5psi vacuum @10:48:37.20 Cartridge 2 sampling 126/2000ml at 0.52ml/s with 23.0psi vacuum @10:49:37.07 Cartridge 2 sampling 157/2000ml at 0.51ml/s with 24.9psi vacuum @10:50:37.22 Cartridge 2 sampling 188/2000ml at 0.52ml/s with 25.0psi vacuum @10:51:37.07 Cartridge 2 sampling 220/2000ml at 0.54ml/s with 26.3psi vacuum @10:52:37.24 Cartridge 2 sampling 253/2000ml at 0.55ml/s with 25.0psi vacuum @10:53:37.07 Cartridge 2 sampling 285/2000ml at 0.53ml/s with 22.4psi vacuum @10:54:37.45 Cartridge 2 sampling 318/2000ml at 0.54ml/s with 22.1psi vacuum @10:55:37.35 Cartridge 2 sampling 350/2000ml at 0.54ml/s with 25.5psi vacuum @10:56:37.08 Cartridge 2 sampling 383/2000ml at 0.54ml/s with 22.7psi vacuum @10:57:37.39 Cartridge 2 sampling 415/2000ml at 0.54ml/s with 23.6psi vacuum @10:58:37.08 Cartridge 2 sampling 447/2000ml at 0.54ml/s with 23.4psi vacuum @10:59:37.24 Cartridge 2 sampling 480/2000ml at 0.54ml/s with 24.0psi vacuum @11:00:37.07 Cartridge 2 sampling 512/2000ml at 0.54ml/s with 23.9psi vacuum @11:01:37.39 Cartridge 2 sampling 544/2000ml at 0.53ml/s with 25.6psi vacuum @11:02:37.08 Cartridge 2 sampling 576/2000ml at 0.54ml/s with 24.1psi vacuum @11:03:37.24 Cartridge 2 sampling 609/2000ml at 0.54ml/s with 22.2psi vacuum @11:04:37.08 Cartridge 2 sampling 641/2000ml at 0.54ml/s with 23.9psi vacuum @11:05:37.15 Cartridge 2 sampling 673/2000ml at 0.54ml/s with 21.6psi vacuum @11:06:37.08 Cartridge 2 sampling 705/2000ml at 0.53ml/s with 21.8psi vacuum @11:07:37.30 Cartridge 2 sampling 737/2000ml at 0.53ml/s with 22.8psi vacuum @11:08:37.56 Cartridge 2 sampling 769/2000ml at 0.53ml/s with 24.5psi vacuum @11:09:37.08 Cartridge 2 sampling 800/2000ml at 0.53ml/s with 23.1psi vacuum @11:10:37.33 Cartridge 2 sampling 832/2000ml at 0.52ml/s with 26.5psi vacuum @11:11:37.08 Cartridge 2 sampling 864/2000ml at 0.53ml/s with 22.8psi vacuum @11:12:37.24 Cartridge 2 sampling 895/2000ml at 0.54ml/s with 24.8psi vacuum @11:13:37.08 Cartridge 2 sampling 927/2000ml at 0.52ml/s with 23.3psi vacuum @11:14:37.20 Cartridge 2 sampling 958/2000ml at 0.52ml/s with 21.0psi vacuum @11:15:37.08 Cartridge 2 sampling 990/2000ml at 0.53ml/s with 24.8psi vacuum @11:16:37.35 Cartridge 2 sampling 1021/2000ml at 0.52ml/s with 24.1psi vacuum @11:17:37.08 Cartridge 2 sampling 1052/2000ml at 0.52ml/s with 23.9psi vacuum @11:18:37.21 Cartridge 2 sampling 1083/2000ml at 0.51ml/s with 26.0psi vacuum @11:19:37.08 Cartridge 2 sampling 1114/2000ml at 0.51ml/s with 22.1psi vacuum @11:20:37.45 Cartridge 2 sampling 1145/2000ml at 0.51ml/s with 25.1psi vacuum @11:21:37.43 Cartridge 2 sampling 1176/2000ml at 0.51ml/s with 20.8psi vacuum @11:22:37.08 Cartridge 2 sampling 1207/2000ml at 0.51ml/s with 22.4psi vacuum @11:23:37.34 Cartridge 2 sampling 1237/2000ml at 0.51ml/s with 24.4psi vacuum @11:24:37.08 Cartridge 2 sampling 1268/2000ml at 0.51ml/s with 23.3psi vacuum @11:25:37.39 Cartridge 2 sampling 1299/2000ml at 0.51ml/s with 24.2psi vacuum @11:26:37.07 Cartridge 2 sampling 1329/2000ml at 0.51ml/s with 21.4psi vacuum @11:27:37.21 Cartridge 2 sampling 1359/2000ml at 0.50ml/s with 24.9psi vacuum @11:28:37.07 Cartridge 2 sampling 1389/2000ml at 0.50ml/s with 22.8psi vacuum @11:29:37.25 Cartridge 2 sampling 1419/2000ml at 0.51ml/s with 24.2psi vacuum @11:30:37.08 Cartridge 2 sampling 1449/2000ml at 0.50ml/s with 23.8psi vacuum @11:31:37.24 Cartridge 2 sampling 1479/2000ml at 0.50ml/s with 24.0psi vacuum @11:32:37.08 Cartridge 2 sampling 1509/2000ml at 0.50ml/s with 22.0psi vacuum @11:33:37.13 Cartridge 2 sampling 1538/2000ml at 0.49ml/s with 23.1psi vacuum @11:34:37.08 Cartridge 2 sampling 1568/2000ml at 0.49ml/s with 24.2psi vacuum @11:35:37.10 Cartridge 2 sampling 1597/2000ml at 0.48ml/s with 25.0psi vacuum @11:36:37.28 Cartridge 2 sampling 1626/2000ml at 0.49ml/s with 22.6psi vacuum @11:37:37.07 Cartridge 2 sampling 1655/2000ml at 0.48ml/s with 26.2psi vacuum @11:38:37.41 Cartridge 2 sampling 1685/2000ml at 0.49ml/s with 23.8psi vacuum @11:39:37.07 Cartridge 2 sampling 1714/2000ml at 0.49ml/s with 23.4psi vacuum @11:40:37.32 Cartridge 2 sampling 1743/2000ml at 0.48ml/s with 26.6psi vacuum @11:41:37.08 Cartridge 2 sampling 1771/2000ml at 0.48ml/s with 20.7psi vacuum @11:42:37.26 Cartridge 2 sampling 1800/2000ml at 0.48ml/s with 22.2psi vacuum @11:43:37.08 Cartridge 2 sampling 1829/2000ml at 0.49ml/s with 24.1psi vacuum @11:44:37.25 Cartridge 2 sampling 1858/2000ml at 0.48ml/s with 24.6psi vacuum @11:45:37.08 Cartridge 2 sampling 1887/2000ml at 0.47ml/s with 20.7psi vacuum @11:46:37.22 Cartridge 2 sampling 1916/2000ml at 0.47ml/s with 23.9psi vacuum @11:47:37.08 Cartridge 2 sampling 1944/2000ml at 0.48ml/s with 24.1psi vacuum @11:48:37.34 Cartridge 2 sampling 1973/2000ml at 0.47ml/s with 25.2psi vacuum @11:49:35.53 Pressure.coast @11:49:35.67 Intake.close @11:49:35.72 Exhaust.close @11:49:36.09 Toroid.reconfigure :hold @11:49:36.15 Toroid.seek :bypass @11:49:37.82 Toroid.reconfigure :default @11:49:37.94 Toroid.seek :clear @11:49:39.65 Sampled 2000.0ml @11:49:40.70 GoodNews.email "After filtering 2000.0ml",Subject: "Processing archive_bac cartridge in slot 2" @11:49:41.79 SamplePump.setPosition! 0ml @11:49:41.84 Plunge.seek :reagent @11:49:41.90 Bypass.open @11:49:42.62 Intake.open @11:49:42.78 Sample loop closed. Ready to Deliver Reagents @11:49:42.80 Archive.seek :sealed @11:49:43.55 SamplePump.reconfigure :slow @11:49:43.61 SamplePump.seek -3ml @11:49:45.96 upsync @11:49:46.56 Archive.seek :evac @11:49:50.31 delay 1 second @11:49:51.43 SamplePump.seek -2.8ml @11:49:52.12 delay 1 second @11:49:53.21 SamplePump.seek -3.3ml @11:49:54.54 delay 1 second @11:49:55.57 Exhaust.open @11:49:56.40 Bypass.close @11:49:57.43 SamplePump.seek -35.3ml @11:50:00.61 delay 1:00 @11:51:00.67 Archive.seek :clear @11:51:06.24 delay 3 seconds @11:51:09.47 Intake.close @11:51:09.72 Exhaust.close @11:51:17.83 RNAlater soak for 10:00 @12:01:17.96 Archive.seek :tVent @12:01:19.65 delay 1 second @12:01:20.68 Archive.seek :evac @12:01:36.65 delay 10 seconds @12:01:46.69 Sample preserved and evacuated @12:01:46.92
Carousel.seek 2 @12:01:47.13 Plunge.seek :locked @12:01:48.11 Archive.seek :clear @12:01:48.29 Air.seek :clear @12:01:48.46 Lysis1.seek :clear @12:02:05.32
Archive.coast @12:02:05.36 Air.coast @12:02:05.40 Lysis1.coast @12:02:05.45 Lysis2.coast @12:02:05.49 Plunge.coast @12:02:05.53 SamplePump.coast @12:02:05.58 Toroid.coast @12:02:05.62 Carousel.coast @12:02:05.81 Gate.power :motor,:OFF @12:02:05.82 Gate.power :core,:OFF @12:02:05.83 Gate.power :analytic1,:OFF @12:02:05.85 Gate.power :analytic2,:OFF @12:02:05.86 Gate.power :analytic3,:OFF @12:02:05.87 Gate.power :main,:OFF @12:02:05.89 Safely stopped and ready to power off @12:02:06.32 Cartridge Load: {[52..47, 45..15, 13..9] => { Type: :archive_bac, State: :processed, filtered: 2000}, [59, 58] => { Type: :archive_bac, State: :processed, filtered: 400}, 1 => { Type: :archive_bac, State: :reserved, was: :dry}, 53 => { Type: :archive_bac, State: :processed, filtered: 1396.8}, 54 => { Type: :archive_bac, State: :processed, filtered: 721.3}, 55 => { Type: :archive_bac, State: :processed, filtered: 622.6}, 56 => { Type: :archive_bac, State: :processed, filtered: 387.7}, 57 => { Type: :archive_bac, State: :processed, filtered: 1000}, [8, 7, 5..2] => { Type: :archive_bac, State: :processed, filtered: 2000.0}, [46, 14, 6] => { Type: :archive_bac, State: :leaky}, 60 => { Type: :archive_bac, State: :processed, filtered: 399.9}} Spares: {spare: 0} @12:02:06.35 BadNews.email "Not enough dry cartridges remain",Subject: "Cartridge::None in MAIN"